M R I Physics Course. Jerry Allison Ph.D. Chris Wright B.S. Tom Lavin M.S.M.P. Department of Radiology Medical College of Georgia

Similar documents
Abstract. Learning Objectives 8/1/2017

RAD 465 (MRI) Lecture one (Pulse Sequences) Ruba Khushaim MSc

4/14/2009. The Big Picture of Quality. MRI Quality Assurance and ACR MRI Accreditation Program. Basic Elements for Image Quality.

Artifacts in Abdominopelvic MR A Pictorial Review

EPI. Thanks to Samantha Holdsworth!

MSK Imaging Fundamentals

Phantom Test Guidance for Use of the Small MRI Phantom for the MRI Accreditation Program

Breast MR Imaging and Quality Control

Troubleshooting Guide. Prep, Scan Errors, and Artifacts

Magnetic resonance imaging phase encoding:

Nuclear Associates and

Comparison of Robarts s 3T and 7T MRI Machines for obtaining fmri Sequences Medical Biophysics 3970: General Laboratory

Chapter 7. Scanner Controls

Release Notes. Multi-Band EPI C2P. Release February 2013

Bioengineering 508: Physical Aspects of Medical Imaging Nature of Medical Imaging. Nature of Medical Imaging

Image quality in non-gated versus gated reconstruction of tongue motion using Magnetic Resonance Imaging:

Guide to artifacts in MRI: classification, explanation, countermeasures.

Supplementary Course Notes: Continuous vs. Discrete (Analog vs. Digital) Representation of Information

PATIENT POSITION IMAGING PARAMETERS

HITACHI S FAST SPIN ECHO TECHNOLOGY

MR Accreditation Programs - E. Jackson

True comfort and flexibility with the power of 3T.

Experiences in ACR MRI Accreditation Vendor Nuances That Every Clinical MRI Physicist Should Know

Calibrate, Characterize and Emulate Systems Using RFXpress in AWG Series

NanoGiant Oscilloscope/Function-Generator Program. Getting Started

SPATIAL LIGHT MODULATORS

EMI/EMC diagnostic and debugging

2

Peacefully quiet. Remarkably fast.

ni.com Digital Signal Processing for Every Application

Procedure Manual for MRI of the Brain

Introduction to the oscilloscope and digital data acquisition

BitWise (V2.1 and later) includes features for determining AP240 settings and measuring the Single Ion Area.

Multi echo Multi slice (MEMS) High Performance fmri at CFMRI... 1

Area-Efficient Decimation Filter with 50/60 Hz Power-Line Noise Suppression for ΔΣ A/D Converters

Technology White Paper Plasma Displays. NEC Technologies Visual Systems Division

Introduction to Data Conversion and Processing

ANTENNAS, WAVE PROPAGATION &TV ENGG. Lecture : TV working

Spectrum Analyser Basics

Spatial Light Modulators XY Series

Peacefully quiet. Remarkably fast.

Premium 1.5T MRI System

Introduction: Overview. EECE 2510 Circuits and Signals: Biomedical Applications. ECG Circuit 2 Analog Filtering and A/D Conversion

DPD80 Infrared Datasheet

Audiovisual Archiving Terminology

Synthesized Clock Generator

Supplemental Material for Gamma-band Synchronization in the Macaque Hippocampus and Memory Formation

Multirate Digital Signal Processing

Chapter 6: Real-Time Image Formation

Techniques for Extending Real-Time Oscilloscope Bandwidth

GG450 4/12/2010. Today s material comes from p in the text book. Please read and understand all of this material!

PACS. Dark Current of Ge:Ga detectors from FM-ILT. J. Schreiber 1, U. Klaas 1, H. Dannerbauer 1, M. Nielbock 1, J. Bouwman 1.

An Efficient SOC approach to Design CRT controller on CPLD s

ON THE INTERPOLATION OF ULTRASONIC GUIDED WAVE SIGNALS

Preparation of the participant. EOG, ECG, HPI coils : what, why and how

SC24 Magnetic Field Cancelling System

Ultrasound instrumentation and image formation. Lecturer: Chelsea Munding September 28 th, 2017

More Info at Open Access Database Process Control for Computed Tomography using Digital Detector Arrays

DPD80 Visible Datasheet

SC24 Magnetic Field Cancelling System

Detailed Design Report

Realizing Waveform Characteristics up to a Digitizer s Full Bandwidth Increasing the effective sampling rate when measuring repetitive signals

CATHODE RAY OSCILLOSCOPE. Basic block diagrams Principle of operation Measurement of voltage, current and frequency

Requirements for Imaging

Quest Chapter 26. Flying bees buzz. What could they be doing that generates sound? What type of wave is sound?

DIGITAL COMMUNICATION

Experiment 9A: Magnetism/The Oscilloscope

Amplification. Most common signal conditioning

Operation Manual for. SCU1 Signal Conditioning Unit

Signal to noise the key to increased marine seismic bandwidth

Module 8 : Numerical Relaying I : Fundamentals

Removing the Pattern Noise from all STIS Side-2 CCD data

Digital Lock-In Amplifiers SR850 DSP lock-in amplifier with graphical display

Colour Reproduction Performance of JPEG and JPEG2000 Codecs

(12) Patent Application Publication (10) Pub. No.: US 2008/ A1. (51) Int. Cl. be maintained. GRADENT SYSTEM PHYSOLOGICAL ACQUESTION CONTROLLER

decodes it along with the normal intensity signal, to determine how to modulate the three colour beams.

10:15-11 am Digital signal processing

DVG-5000 Motion Pattern Option

Please feel free to download the Demo application software from analogarts.com to help you follow this seminar.

Fourier Transforms 1D

RSNA 2006 November 26 to December 1 Chicago. Guest author for ImPACT Dr. Koos Geleijns, Medical Physicist, Leiden University Medical Center.

Installation Evaluation Form

The Development of a Synthetic Colour Test Image for Subjective and Objective Quality Assessment of Digital Codecs

CHAPTER 4 OSCILLOSCOPES

REPORT DOCUMENTATION PAGE

S I N E V I B E S FRACTION AUDIO SLICING WORKSTATION

T ips in measuring and reducing monitor jitter

ONE SENSOR MICROPHONE ARRAY APPLICATION IN SOURCE LOCALIZATION. Hsin-Chu, Taiwan

ECE 5765 Modern Communication Fall 2005, UMD Experiment 10: PRBS Messages, Eye Patterns & Noise Simulation using PRBS

Chapter 2 Signals. 2.1 Signals in the Wild One-Dimensional Continuous Time Signals

TERMINOLOGY INDEX. DME Down Stream Keyer (DSK) Drop Shadow. A/B Roll Edit Animation Effects Anti-Alias Auto Transition

Procedures for conducting User QA on the scanner

Experiment 13 Sampling and reconstruction

Ch. 1: Audio/Image/Video Fundamentals Multimedia Systems. School of Electrical Engineering and Computer Science Oregon State University

Minimize your cost for Phased Array & TOFD

Application of HDMI Cables as an MRI Compatible Single Cable Solution for Readout and Power Supply of SiPM Based PET Detectors

Pole Zero Correction using OBSPY and PSN Data

How to Obtain a Good Stereo Sound Stage in Cars

DESIGNING OPTIMIZED MICROPHONE BEAMFORMERS

Elasticity Imaging with Ultrasound JEE 4980 Final Report. George Michaels and Mary Watts

Transcription:

M R I Physics Course Jerry Allison Ph.D. Chris Wright B.S. Tom Lavin M.S.M.P. Department of Radiology Medical College of Georgia

M R I Physics Course chapter 12 Artifacts and Suppression Techniques Artifacts / Undesirable Contrast Chemical Shift artifact Truncation (Gibbs) Artifact Errors in data Aliasing (wrap around) Zippers / Stars (Zero line artifact) Flow artifact Metallic artifact (magnetic field perturbations)

M R I Physics Course chapter 12 Artifacts and Suppression Techniques Motion Magnetic Susceptibility artifact Surface Coil artifact N/2 Ghosts in EPI

Chemical shift artifacts -At 1.5T the Larmor frequency of fat is approximately 220 Hz lower than water. MRI images can be considered to be images of free water with a superimposed spatially misregistered fat image. -Chemical shift artifacts produce high intensity and low intensity bands near fat/water interfaces, primarily in the frequency encode direction. -Dark boundaries between regions are sometimes called contour artifacts. 4

Kidney example: from the http://www.mritutor.org website ( 1994-2003 by Ray Ballinger) 5

Copyright Radiological Society of North America, 1999 Williams, R. L. et al. Radiology 1999;212:151-154

Chemical shift at 7T: oil and water

Truncation (Gibbs) artifact -Truncation artifact or Gibbs phenomena appear as a periodic ringing at high contrast interfaces. -Truncation artifact is caused by underestimation or overestimation of the signal at high contrast boundaries. 8

Truncation (Gibbs) artifact -The step function resulting from a high contrast interface cannot be accurately depicted by the finite number of sine and cosine waves prescribed by a typical MR image acquisition matrix. -Truncation artifacts can be diminished by a decrease in pixel size or raw data filtering. It is important to remember that raw data filtering affects all image data not just the artifactual portions. -Pixel size can be decreased by increasing the imaging matrix or decreasing the field of view. 9

Partial Volume Artifacts -Caused by an imaging voxel containing two different tissues and therefore possessing a signal average of both tissues. 10

11

Errors in data -Errors in data can cause striped images, washed out images and zipper artifacts. -Failure of the array processor can produce a variety of stripes and patterns (i.e. corduroy artifacts) on reconstructed images. 12

-Leaking RF shield: The exam room is RF shielded to attenuate RF energy from radio, television, etc. Failure of the shield can produce zipper artifacts. To demonstrate a leaking RF shield, scan with the door open. A simple test of the RF shield is to see if a transistor radio receives signal in the exam room with the door closed. 13

RF receiver gain -If the RF receiver gain is set such that the voltage level of the amplified signal exceeds the voltage range of the ADC, washed out images result. -The RF receiver gain settings are generally adjusted by auto prescan to avoid this artifact. 14

Aliasing (wrap-around) -Aliasing or wrap-around artifacts can occur when the anatomy exceeds the field of view. Aliasing can occur in the frequency encode direction or the phase encode direction. 15

16

Aliasing (wrap-around) -Solution 1: Sampling faster may reduce or eliminate aliasing but results in an increased receiver bandwidth with decreased SNR. 17

Aliasing (wrap-around) Solution 2:-Use of a larger FOV than actually desired and display of the central non-aliased portion of the image. This is the essence of No Phase Wrap, No Frequency Wrap and Oversampling. Aliasing in the frequency encode direction is generally prevented on modern MRI systems by oversampling. 18

19

Aliasing (wrap-around) -Siemens: Oversampling in the frequency encode direction doubles the number of points sampled. -GE: Oversampling in the frequency encode direction is always utilized. -Oversampling in the frequency encode direction does not change image acquisition time (or number of slices or echo time). 20

Aliasing (wrap-around) -Aliasing also occurs in the phase encode direction. This can be cured by use of No Phase Wrap:GE or Oversampling:Siemens. -Siemens: Oversampling in the phase encode direction increases the number of phase encodes by a user selectable percentage. -GE: No Phase Wrap doubles the number of phase encode samples (and halves the number of excitations). 21

22

Aliasing (wrap-around) -Removal of phase encode aliasing can require additional scan time. -In gradient recalled echo imaging, aliasing may produce a pronounced zebra stripe artifact. This artifact can be eliminated by oversampling ( No Phase Wrap ) or increasing the FOV. 23

Aliasing (wrap-around) -Aliasing can occur in the slice encoding direction in a 3D data set. GE routinely discards the four outermost slices in a 3D data set (acquire 128 slices; displays 124 slices: acquire 64 slices; display 60 slices: acquire 32 slices; display 28 slices). Aliasing or wrap-around in the slice encoding direction is often still visible in the outermost slices. -The phase encode axis is usually chosen as the narrower aspect of anatomy in order to minimize phase encoding aliasing. 24

Aliasing (wrap-around) -Solution 3: SAT uration bands can be applied outside the FOV to reduce aliasing artifacts. 25

Zippers/Stars (Zero line artifact) -Bands having a dashed appearance through the center of image. -Stars consist of a bright dot in the center of the image with dashed line tails along the frequency and phase encode directions. -The signal for each point in an MRI image is actually a sinc function ([sin x] / x) in both the frequency encode and the phase encode direction. 26

Zippers/Stars (Zero line artifact) -The sinc function signal actually contributes to several pixels in an image but ordinarily the signal side lobes is 27 small and is not recognized in an image.

Zippers/Stars (Zero line artifact) -A large noise signal can cause zipper or star artifacts. If the noise signal has large amplitude the signal from side lobes is large enough to show up as a dashed pattern in the reconstructed image. 28

Flow artifacts -Flow artifacts occur due to in-flow/out-flow time of flight phenomena and due to phase accumulation in a gradient magnetic field. 29

Flow artifacts -Flow compensation (gradient motion rephasing) can recover some phase dispersion attributable to flow. 1st order gradient moment nulling: compensation for constant velocity 30

Flow artifacts -Depends on -Flow velocity -Flow direction -Slice thickness -TR / TE -Direction of artifacts can be reversed by swapping the phase and frequency encode directions. 31

Metallic artifacts (magnetic field perturbations) -Metals can cause distortions in the static magnetic field. -Ferromagnetic metals can dramatically distort the static field due to the alignment of many magnetic domains. -Non-ferromagnetic metals can distort the static magnetic field due to smaller magnetic susceptibility effects. -Metal artifacts are generally worse with gradient recalled echo sequences than with spin echo 32 sequences.

Metallic artifacts -During imaging, eddy currents may be induced in some metal objects resulting in induced magnetism. -The effect of metallic objects is signal loss and image distortion: -Ferromagnetic metal -Cosmetics (Fe, Co) -Dental amalgams -Implants -Non-ferromagnetic metal -Shrapnel 33

Artifacts and Suppression Techniques Signal voids near metal implants (e.g., fillings, as in this gradient-echo picture) are due to magnetic susceptibility differences. MRI artifact website, Vancouver, BC Copyright 1996, Wayne Patola and Bruce Coulter http://wwwrad.pulmonary.ubc.ca/stpaulsstuff/mrart/gre.html 34

Motion -Patient motion produces ghost images; primarily in the phase encode direction. -In the frequency encode direction, motion produces little displacement in the 4-8 msec interval when data are collected. 35

Motion Artifacts and Suppression Techniques -In the phase encode direction, motion can produce large displacements (and associated large phase errors) since it can be as much as a few seconds between collection of adjacent lines of k-space. -Solutions for motion artifacts include: -Restraints -Physiological triggering to capture anatomy in the same aspect of the respiratory or cardiac cycle for each line of k-space. -Cardiac triggering is used frequently. -Peripheral gating is available. -Respiratory triggering is available but impractical. 36

Solutions for motion artifacts include: -Respiratory-Ordered Phase Encoding (ROPE) also known as respiratory compensation or retrospective cardiac gating. Each acquisition is assigned to k-space based upon the respiratory phase or cardiac phase when it was acquired. Retrospective cardiac gating is similar to a first pass MUGA scan in Nuclear Medicine. -Fat suppression to eliminate bright ghosting from moving fat. -ChemSAT -STIR -SAT bands 37

Magnetic susceptibility artifacts -The magnetic susceptibility of material present in an MRI image varies. Various soft tissues have various magnetic susceptibilities. Air has a very low magnetic susceptibility. Iron deposits from hemorrhage, liver iron overload, etc. have very high magnetic susceptibility. 38

Magnetic susceptibility artifacts -Perturbations in the static magnetic field due to magnetic susceptibility can cause geometric distortion and changes in signal intensity. Geometric distortions occur primarily in the frequency encode direction. Signal intensity can be increased or decreased by magnetic susceptibility. The artifacts caused by magnetic susceptibility can sometimes be observed in slices adjacent to the inhomogeneity. These artifacts may be more prominent on gradient recalled echo images or long TE spin echo images. Magnetic susceptibility may be useful contrast for imaging of calcification and hemorrhage. 39

Phantom Susceptibility in a T2* Weighted Sequence

Susceptibility Artifact in Orthogonal T2* (EPI) images

Surface coil artifact -The sensitivity of surface coils falls off dramatically with distance from the coil. Tissues close to the surface coil will have higher intensity than tissues farther from the coil. This artifact is very apparent in T 1 weighted imaging of the spine where subcutaneous fat produces very intense signal due to it s short T 1 value and proximity to the coil. -GE Signa has an option Image Intensity Correction that is designed to reduce surface coil artifacts (not used). 42

Surface Coil Artifact -Notice how the SNR varies across the FOV for an 8 channel (parallel) brain coil. 43

N / 2 Ghosts in EPI -EPI is very sensitive to ghosting artifacts. Because the MRI signal is sampled under alternating gradients, it is necessary to reverse (invert) every other line. Any imperfection in the entire MRI imaging chain can modulate the MRI signal at half the Nyquist frequency. The result can be an N / 2 ghost in the phase encode direction. - N / 2 ghosts can be minimized by: -calibration scans (with no phase encodes) -image filtering 44

GHOSTS! 45

Challenge: Dielectric Resonance Larmor frequency increases for stronger magnets RF wavelength approaches body dimensions and FOV dimensions λ air = 4.7m at 1.5T λ air = 2.35m at 3T λ tissue = 0.3m at 3T!!!!!!!! (high dielectric constant) Can produce dielectric resonances which may reduce RF penetration Standing waves Image shading Worse in body imaging than head Worse in large patients (obese) Dielectric resonance effects occur at all field strengths but become apparent at 3T 46

Challenge: Dielectric Resonance UGA (Torso Array) 47

Challenge: Dielectric Resonance MCG (Torso Array) T2 T2 STIR 48

2006 GE Healthcare

2006 GE Healthcare

2006 GE Healthcare

2006 GE Healthcare