INSTRUCTION MANUAL V Barometric Pressure Sensor. May Copyright 2007 Campbell Scientific (Canada)Corp.

Similar documents
CS210 Enclosure Relative Humidity Sensor Revision: 10/07

10164-L Sampler Control Cable for use with Isco and Sigma Autosamplers Revision: 2/11

Operating Instructions

Instruction Manual MODEL RSP SANITARY ELECTRONIC PRESSURE TRANSMITTER

Global Water Instrumentation, Inc.

Instruction Manual Model # Block Upconverter

E8Y. Micropressure Sensor with Easy-to-Read Digital Display. Differential Pressure Sensor. Ordering Information. Sensors

User s Manual. Model Digital Sound Level Meter

SCALE & WEIGHT DISPLAYS

CR-R880-BL: Indoor/Outdoor Proximity Reader with 10cm (4in) read range

Model /29S RF Splitter

Instruction Manual Model BlockUpconverter

Model Number Structure

DICKSON ES120/ES120A DICKSON. Electronic Signal Data Logger. Specifications. Applications, Features, & Getting Started. Instructions / Operating

Garmin GC 10 Marine Camera Instructions

FOTS100 User Manual. BIOPAC Systems, Inc. Opsens Inc. 42 Aero Camino, Goleta, CA Tel (805) , Fax (805)

DEX Fruit & Stem Growth Dendrometer User Manual

6170 Shiloh Road Alpharetta, Georgia (770) FAX (770) Toll Free

3M Sensored Termination (15 kv) QX-T15I-vi1-E

32 Channel CPCI Board User Manual

AE-341 SERIES PROCESS INDICATORS 2 TO 6 DIGITS 0-20 MA, 4-20 MA, 0-5V, OR 0-10V SCALED INPUT

OPERATION MANUAL JENDPM008 THERMOCOUPLE PANEL THERMOMETER. TIP TEMPerature Products 340 W Broad Street Burlington, NJ 08016

QD1 Incremental Encoder Interface Revision: 12/91

VGA Extender LR EXT-VGA-141LR. User s Manual

Digital Lighting Systems. DIGTAN4-PRO Converts 4 PROTOCOL channels to 4x 0.5 V to 10 VDC Analog

Extender w/ RS-232 and 2-way IR

User Manual. Model 979T and 979R Digital Audio Converters

BENESTON USER MANUAL AHD-ER03H. AHD to HDMI & VGA & AV Converter AHD-ER03H

Model P/03P Upconverters

DUAL/QUAD DISPLAY CONTROLLER Operation Manual

Kramer Electronics, Ltd. USER MANUAL. Model: VM Video Component Distributor

Audio. 4K Ultra HD Extender. w/ RS-232 and 2-way IR GTB-UHD2IRS-ELRPOL-BLK. User Manual. Release A2

1x3 Component / Audio Distribution Amplifier

LEVEL ADJUST POWER Shiloh Road Alpharetta, Georgia (770) FAX (770) Toll Free

Kramer Electronics, Ltd. USER MANUAL. Model: WP-210A. XGA/Audio Line Driver

Indoor/Outdoor Security System with Quad Monitor User s Manual

Model R177M and R177S Baseband Switch

ACTIVE IF SPLITTER/COMBINER UHP-IFS

VGA Extender LR EXT-VGA-141LR. User s Manual

Active VGA Managed Dispatcher (8 Ports) (16 Ports) Installation Guide

Kramer Electronics, Ltd. USER MANUAL. Models: WPN-11, XGA/Component/CV/YC Transmitter WPN-12, XGA/Component/CV/YC Receiver

WXL-2F, WXL-1F, WXL-2M, WXL-1M

FD70CV-M-C-9. Installation and Operation Manual. 7 NVG Compatible LCD. Revision Date: 06/16/2017 Page 1 of 12. Rev: A

HDMI Extender over UTP Cable

2000 Series Weather Stations Analog Temperature / RH Sensor Upgrade Kit PRODUCT MANUAL KIT # 3613WDU

Kramer Electronics, Ltd. USER MANUAL. Model: TP-43. Component S/PDIF Transceiver

Kramer Electronics, Ltd. USER MANUAL. Model: VP-100. VGA/XGA to RGBHV Converter

Kramer Electronics, Ltd. USER MANUAL. Model: WP-220. XGA/Audio/Video Line Driver

Metal Electrode Meter

Kramer Electronics, Ltd. USER MANUAL. Models: TR-1YC, s-video Isolation Transformer TR-2YC, s-video Dual Isolation Transformers

USER MANUAL. Kramer Electronics, Ltd. Models:

User Guide. HDMI Fiber Optic Extender. DVI-7350a

Kramer Electronics, Ltd. USER MANUAL. Model: VM-10xl. Video Audio Distribution Amplifier

Kramer Electronics, Ltd. USER MANUAL. Model: 482xl. Bi-directional Audio Transcoder

USER MANUAL FOR THE ANALOGIC GAUGE FIRMWARE VERSION 1.0

ZIG-POEPRO-70A. HDBaseT HDMI Extender

EZ-LIGHT K30L Series With Independent Push Button Contact

VGA & Audio over CAT5 Distribution Series

PSM-003. Micro Polarization Controller/Scrambler. User Guide

PoE Over Coax Extender Kit

DisplayPort Extender over 2 LC Fibers

INSTRUCTION MANUAL DAFM2

WINEGARD INSTALLATION MANUAL. Model GM Carryout Ladder Mount for mounting pipes with outer diameters between 1 to 1-1/8

HD VIDEO IP STREAMER CT-HDVD-HDSTR-KIT

Thank you for purchasing our product. If there is any question or request, please do not hesitate to contact us. This manual may contain technical or

LEVEL ADJUST POWER Shiloh Road Alpharetta, Georgia (770) FAX (770) Toll Free

VGA CAT-5 1:8 Distribution S VGA CAT-5 Distribution R. EXT-VGA-CAT5-148S EXT-VGA-CAT5-148R User Manual

CS x1 RS-232 Computer Controlled Video Switcher. Instruction Manual

General Specifications

CS525 ISFET ph Probe Revision: 10/11

INTRODUCTION INSTALLATION LIGHTNING RETARDING LOOPS. Figure 1. Relay Box Antenna Connections. Relay Box Cable Connections

TC Mbps - 622Mbps FIBER OPTIC MODE CONVERTER/REPEATER (Rev A0.1) User's Manual

KRAMER ELECTRONICS LTD. USER MANUAL MODEL: VP-2L UXGA Line Amplifier. P/N: Rev 3

DATA SHEET Panorama Rudder Indicator, TRI-2

SD/HD/3G-SDI Video + Power + Data RS-485 Transmission over Coax Kit

VHF + UHF Amplified HDTV Antenna Model OA8000 & OA8001 Installation Instructions Reception Frequencies

Kramer Electronics, Ltd. USER MANUAL. Model: PT-102VN. 1:2 Video DA

Warner Photoscanner MCS-500 Series LED Photoelectric Control

AEXX-200F SERIES 4-DIGIT LINE FREQUENCY MONITORS

On-screen display signal strength meter Version 1.02

SC-C1M SiriusConnect TM Vehicle Tuner

Kramer Electronics, Ltd. USER MANUAL. Models: 6808, SDI-AES De-Embedder 6809, SDI-AES Embedder

MultiView T4 / T5 Transmitter

Quick Installation Guide. Indoor / Outdoor Antenna, Antenna Cable & Surge Arrestor

SCIR. Infrared Illuminator Powerful Indoor/Outdoor IR Spotlight SCIR

Kramer Electronics, Ltd. USER MANUAL. Model: VP-12NHD. 1: UXGA / CAT5 Distributor

Kramer Electronics, Ltd. USER MANUAL. Model: VP-200Dxl. XGA Differential Amplifier / DA

UTP ACTIVE TRANSCEIVER HUB

SAFETY AND NOTICE TABLE OF CONTENTS

Kramer Electronics, Ltd. USER MANUAL. Model: 810B. Black Burst / Audio Generator

USER INSTRUCTIONS MODEL CSI-200 COAXIAL SYSTEM INTERFACE

FD171CV-C-4. Installation and Operation Manual. 17 HDSDI Special Mission Quad Monitor. Revision Date: 01/11/2017 Page 1 of 14.

Master Time Clock MTC-200 MTC-400 MTC-600. Users Manual

BASIC PH METER BASIC PH METER KIT

USER MANUAL FOR THE ANALOGIC GAUGE FIRMWARE VERSION 1.1

SD/HD-SDI 1 to 2 Distribution Amplifier VCF-1002DA-P

Kramer Electronics, Ltd. USER MANUAL. Model: Power Amplifier

PRO-CoaxExt HDMI extender over Coaxial cable with bi-directional IR User s Guide

MENU EXECUTE Shiloh Road Alpharetta, Georgia (770) FAX (770) Toll Free

INSTRUCTION MANUAL MODEL 2710 SUBCARRIER DEMODULATOR

Transcription:

INSTRUCTION MANUAL 61205V Barometric Pressure Sensor May 2007 Copyright 2007 Campbell Scientific (Canada)Corp.

WARRANTY AND ASSISTANCE This equipment is warranted by CAMPBELL SCIENTIFIC (CANADA) CORP. ( CSC ) to be free from defects in materials and workmanship under normal use and service for twelve (12) months from date of shipment unless specified otherwise. ***** Batteries are not warranted. ***** CSC's obligation under this warranty is limited to repairing or replacing (at CSC's option) defective products. The customer shall assume all costs of removing, reinstalling, and shipping defective products to CSC. CSC will return such products by surface carrier prepaid. This warranty shall not apply to any CSC products which have been subjected to modification, misuse, neglect, accidents of nature, or shipping damage. This warranty is in lieu of all other warranties, expressed or implied, including warranties of merchantability or fitness for a particular purpose. CSC is not liable for special, indirect, incidental, or consequential damages. Products may not be returned without prior authorization. To obtain a Return Merchandise Authorization (RMA), contact CAMPBELL SCIENTIFIC (CANADA) CORP., at (780) 454-2505. An RMA number will be issued in order to facilitate Repair Personnel in identifying an instrument upon arrival. Please write this number clearly on the outside of the shipping container. Include description of symptoms and all pertinent details. CAMPBELL SCIENTIFIC (CANADA) CORP. does not accept collect calls. Non-warranty products returned for repair should be accompanied by a purchase order to cover repair costs.

TABLE OF CONTENTS 1. General Description...2 1.1 Physical Dimensions...2 1.2 Specifications...3 2. Installation...3 2.1 Mounting...3 2.2 Wiring...3 Differential Wiring:...4 Single Ended Wiring:...4 3. Programming...6 3.1 Program Example:...7 CR10X Example...7 CR1000 Example...8 1

1. General Description The RM Young 61205V Barometric Pressure Sensor is a versatile electronic barometer with low power consumption and wide operating temperature range. It is designed to be installed in a weather-proof enclosure or indoors. Accuracy of better than 0.5 hpa (mb) is maintained over the entire pressure and temperature range. Signal output is linear 0-2500 mv. It can be operated in a power-up or continuous mode. In the power-up mode, the datalogger must switch 12 VDC to the sensor at least 3 seconds before the measurement, take a reading, then switch the sensor off to conserve power. This can be done using the switched 12V terminal on the CR10X, CR23X, CR800, CR850, CR1000, and CR3000 datalogger. See datalogger manual Section OV1.1.9 for a description of the power up/down programming. FIGURE 1. 61205V Barometric Pressure Sensor 1.1 Physical Dimensions Length: Width: Height: Weight: Dimensions include base plate. 14 cm (5 1/2 inches) 7.2 cm (2 13/16 inches) 2.2 cm (14/16 inch) 454 g (1 lb) 2

1.2 Specifications Operating Range Pressure Range: 600 to 1100 hpa (600 1100 mb) Operating Temperature: -50 to 60 0 C Humidity: Non-condensing Accuracy Accuracy: 0.5 hpa Resolution: 0.1 hpa Update Rate: 2 seconds Signal Output Analog 0-2500mV General Power Requirement: Power Consumption: Warm-up Time: Case Material: 7-30 VDC 9 ma 3 seconds Fibre-reinforced Thermoplastic NOTE: The black outer jacket of the cable is Santoprene rubber. This compound was chosen for its resistance to temperature extremes, moisture, and UV degradation. However, this jacket will support combustion in air. It is rated as slow burning when tested according to U.L. 94 H.B. and will pass FMVSS302. Local fire codes may preclude its use inside buildings. 2. Installation 2.1 Mounting 2.2 Wiring The sensor must be protected from the elements either by being placed in a weather proof enclosure or indoors. The 61205V has four mounting holes that are spaced 1 on centre. The screws and grommets shipped with the sensor will aid in mounting the sensor inside one of Campbell Scientific s white fibreglass enclosures. The sensor was shipped with a black rubber boot over the pressure port (upper right corner) for protection. REMOVE this boot before deploying sensor. If the 61205V is being installed in an enclosure, use the poly tube that was shipped with the sensor to vent the pressure port to the outdoors. Vent this tube about 1 out the cable entry port and insert one of the white hydrophobic filters (provided) into the end of the tube. The recommended measurement method is using a datalogger differential channel. It can be measured using a single-ended channel, but a small measurement error will be introduced due to cable lead length. 3

Differential Wiring: Colour Sensor Terminal CR10(X), CR510, CR23X 21X, CR7 White Tx-A-V Diff H Diff H Diff H CR800, CR850 CR1000, CR3000 Green Ref Diff L Diff L Diff L Red Power +12V or SW 12V +12V +12V or SW-12 Jumper from Ctrl Port to SW 12 CTRL Black Ref G G Clear Drain G for switching 12V power only Single Ended Wiring: Using single-ended measurements may be necessary in certain applications to reduce the number of analog inputs required on the datalogger. It is important to be aware of three issues concerning the 61205V Barometric Pressure Sensor and a single-ended measurement. The single-ended measurement introduces a small measurement error due to cable lead lengths. If this error is unacceptable, or if you will be powering the sensor continuously, you should use a differential measurement. Error Caused By Lead Lengths When Using Single- Ended Measurement 1. The signal reference (Green) and the power ground (Black) are in common inside the 61205V. When the barometric pressure is measured using a single-ended analog measurement, both the signal reference and power ground are connected to ground at the datalogger. The signal reference and power ground both serve as the return path for 12V. There will be a voltage drop along those leads because the wire itself has resistance. The 61205V draws approximately 9 ma when it is powered. The wire used in the 61205V (P/N L9720) has a nominal resistance of 17.5 Ω/1000 feet. Since the signal reference and the power ground are both connected to ground at the datalogger, the effective resistance of those wires together is half of 17.5 Ω/1000 feet, or 8.75 Ω/1000 feet. Using Ohm s law, the voltage drop (V d ), along the signal reference/power ground is given by the equation below: V d = I*R = 9mA * 8.75 Ohm /1000 ft. = 79 mv/1000 ft. This voltage drop will raise the apparent barometric pressure because the difference between the signal and signal reference lead at the datalogger has increased by V d. With an analog output 4

of 0.2 hpa/ mv, the approximate error in barometric pressure is 1.58 hpa per 100 feet of cable length. It is not recommended to compensate for this offset in your datalogger program without system testing for two reasons. First, the stated 9mA current draw is a maximum value subject to variation. Second, the stated cable resistance is a nominal value. In order to properly compensate for lead length on a single ended measurement, the actual current draw and cable resistance must be determined. Important Note: 2. When measuring the 61205V single-endedly, the green and black leads must both be connected to AG on the CR10(X), CR510 or to on the CR23X, CR800, CR850, CR1000, and CR3000. Failure to do so will connect the datalogger analog and power ground planes together, which in some cases can cause offsets on low level analog measurements. 3. When using single-ended measurements on a CR10(X), CR510, CR23X, CR800, CR850, CR1000, or CR3000 it is important to ensure that the sensor is powered down prior to making other analog measurements. Failure to turn the sensor off when the 61205V is connected as a single-ended measurement causes the power return to be connected to the analog ground on the datalogger. The return currents from the sensor could cause small offsets on other low-level analog measurements (i.e. radiation sensors). Colour Sensor Terminal CR10(X), CR510 CR23X, 21X, CR7 White Tx-A-V SE SE SE Green Ref AG CR800, CR850 CR1000, CR3000 Red Power SW 12V +12V +12V or SW-12 Black Ref AG Clear Drain G Jumper from Ctrl Port to SW 12 CTRL 5

3. Programming The 61205V draws 9mA of power when activated. Depending on your power supply it may be beneficial to conserve power. This can be achieved by powering the sensors on and off. This is done through the switched 12V port on the CR10X, CR23X, CR800, CR850, CR1000, and CR3000. Other loggers may require a relay (model C1701) when power consumption is a concern. The sensor is measured using Instruction 1 SINGLE-ENDED VOLTS or Instruction 2 DIFFERENTIAL VOLTS. The reading will be in millivolts and has to be converted to barometric pressure units. 0-2500mV = 600 to 1100 hpa (mb). MULTIPLIER AND OFFSET CALCULATION Output from the 61205V is 0-2500mV which corresponds to 600 to 1100 hpa (mb) of pressure. The multiplier and offset used to convert the voltage reading to hpa is: hpa = (mv x 0.2) + 600 hpa Multiplier = 0.2 Offset = 600 This results in true barometric pressure in hpa. 1 hpa = 1mb. The weather service and most airports correct the pressure reading at a particular station to what it would be if that station were located at sea level. This is done so that weather forecasters can obtain a clearer picture of what is happening in the atmosphere. Use the following equation to determine the pressure correction factor. Then add this to the offset (i.e. CorrFactor + 600 = New Adjusted Offset). Because barometric pressure is greatly effected by elevation, the elevation of the site used in the equation should be as accurate as possible. P = 1013.25 [ 1 ( 1 Elevation/44307.69231 ) ^5.253283 ] Where Elevation is in meters above sea level. To convert the reading to units other than hpa, use program instruction Z=X*F(P37) and one of the values from Table 2. TABLE 2. Multipliers to convert the hpa reading to different units For Multiply by kpa 0.1 mm of Hg 0.75006 in of Hg 0.02953 mbar 1.0 PSI 0.0145 The following program examples are not to be used verbatim. Input channels, control ports, and location labels must be adjusted for actual circumstances. 6

3.1 Program Example: CR10X Example Turn sensor on one minute before the hour then take a reading. This allows enough time for the sensor to warm up and settle (minimum 3 seconds). Conditions of Example: - Measure sensor with a differential channel - Signal wires are connected to Diff Chan #1 - Control Port #1 is jumpered to SW 12V CTRL for CR10X example only - Red wire is connected to the SW 12V terminal - Desired output is kpa (Environment Canada standard) - TO LEAVE THE SENSOR ON CONTINUALLY, CONNECT THE RED WIRE TO +12V AND REMOVE THE PROGRAM INSTRUCTIONS THAT HAVE THE <><><> SYMBOLS BESIDE THEM. ; Measure 61205V Barometric Pressure Sensor one minute before the hour and ; output the reading. ;{CR10X} ; *Table 1 Program 01: 5 Execution Interval (seconds) 1: If time is (P92) ; One minute before the hour, turn on sensor. <><><><><><> 1: 59 Minutes (Seconds --) into a 2: 60 Interval (same units as above) 3: 41 Set Port 1 High 2: If time is (P92) ; On the hour take sensor reading. <><><><><><> 1: 0 Minutes (Seconds --) into a 2: 60 Interval (same units as above) 3: 30 Then Do: ; Measure 61205V Barometric Pressure Sensor... 3: Volt (Diff) (P2) <<<<<<<<<<< For Single-ended Measure- 1: 1 Reps ment use Volt (SE) P1. 2: 25 2500 mv 60 Hz Rejection Range 3: 1 DIFF Channel 4: 1 Loc [ Air_PRESS ] 5:.2 Mult 6: 600 Offset ; Add necessary sea level correction to offset. ; Convert reading to kpa. 4: Z=X*F (P37) 1: 1 X Loc [ Air_PRESS ] 2:.1 F 3: 1 Z Loc [ Air_PRESS ] 5: Do (P86) <><><><><><> 1: 51 Set Port 1 Low 7

6: End (P95) <><><><><><> ; Hourly data output 7: If time is (P92) 1: 0 Minutes (Seconds --) into a 2: 60 Interval (same units as above) 3: 10 Set Output Flag High (Flag 0) 8: Set Active Storage Area (P80) 1: 1 Final Storage Area 1 2: 60 Array ID 9: Real Time (P77) 1: 1220 Year,Day,Hour/Minute (midnight = 2400) 10: Resolution (P78) 1: 1 High Resolution 11: Sample (P70) 1: 1 Reps 2: 1 Loc [ Air_PRESS ] 12: Resolution (P78) 1: 0 Low Resolution Input Location Labels: 1 AIR_PRESS SAMPLE OF PROCESSED DATA: Array ID YEAR JULIAN DAY TIME AIR PRESSURE 60 2003 172 1300 100.34 60 2003 172 1400 100.39 60 2003 172 1500 100.42 CR1000 Example 'Measure 61205V Barometric Pressure Sensor one minute before the hour and output 'the reading 'CR1000 'Declare Public Variables Public Batt_Volt Public Air_PRESShPa Public Air_PRESSkPa 'Define Data Tables DataTable (Table1,True,-1) DataInterval (0,60,Min,10) Sample (1,Air_PRESSkPa,FP2) EndTable 'Main Program BeginProg 8

Scan (5,Sec,0,0) 'Default Datalogger Battery Voltage measurement Batt_Volt: Battery (Batt_Volt) 'Turn 61205V on one minute before the hour, using SW12V: If TimeIntoInterval (59,60,Min) Then SW12 (1 ) '<><><><><><> '61205V air pressure measurement in hpa. Change offset from 600 hpa to the 'appropriate sea level correction for you site: If TimeIntoInterval (0,60,Min) Then VoltDiff (Air_PRESShPa,1,mV2500,1,True,0,_60H z,0.2,600) conversion factors are found in Table 2. NextScan EndProg EndIf 'Call Data Tables and Store Data CallTable (Table1) 'Convert air pressure value to kpa. Other Air_PRESSkPa = Air_PRESShPa*0.1 'Turn 61205V off after a reading is taken SW12 (0) '<><><><><><> 9