AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode)

Similar documents
SPM Training Manual Veeco Bioscope II NIFTI-NUANCE Center Northwestern University

Quick Start Bruker Dimension Icon AFM

Standard Operating Procedure of nanoir2-s

AFM Standard Operating Procedure

Figure 1. MFP-3D software tray

Standard AFM Modes User s Manual

University of MN, Minnesota Nano Center Standard Operating Procedure

Magnetic Force Microscope (MFM)

STANDARD OPERATING PROCEDURE: ASYLUM MFP-3D AFM

Quick Start ATOMIC FORCE MICROSCOPE West Campus Imaging Core

TT AFM LongBeach Procedures and Protocols V2.1

Asylum MFP-3D Standard Operating Procedures

Dektak Step by Step Instructions:

Asylum MFP-3D Standard Operating Procedures

Operation Procedure for Phillips XL30 ESEM

B-AFM. v East 33rd St., Signal Hill, CA (888)

Topic: Instructional David G. Thomas December 23, 2015

Scanning Probe Microscope Training. Wenhui Pang

INSTALATION PROCEDURE

INSTRUCTIONAL MANUAL FOR LCD ZOOM MICROSCOPE

Scanning Electron Microscopy (FEI Versa 3D Dual Beam)

SC24 Magnetic Field Cancelling System

[AMBIENT ATOMIC/MAGNETIC FORCE MICROSCOPY MANUAL]

SC24 Magnetic Field Cancelling System

Kratos AXIS Ultra DLD X-ray Photoelectron Spectrometer Instructions

SEM- EDS Instruction Manual

SIMET AVIKO D INSTRUCTION MANUAL SORTING Solutions, Ltd.

12.1 Inch CGA EGA VGA SVGA LCD Panel - ID #492

Do not disassemble any part of the device! Disassembling of the product is permitted only for experts certified with the NT-MDT company.

ME EN 363 ELEMENTARY INSTRUMENTATION Lab: Basic Lab Instruments and Data Acquisition

SX7. Saga 7" Super Bright HDMI/3G-SDI Field Monitor with 3D-LUTs. Quick Start Guide. What s Included CHECKED BY

Application Note AN-708 Vibration Measurements with the Vibration Synchronization Module

imso-104 Manual Revised August 5, 2011

9070 Smart Vibration Meter Instruction Manual

ISOMET. Compensation look-up-table (LUT) and How to Generate. Isomet: Contents:

AX20. Atlas 19.5" 3G-SDI/HDMI Field and Studio Monitor with 3D LUTs & Scopes. Quick Start Guide. What s Included CHECKED BY

Preparing for remote data collection at NE-CAT

Dektak II SOP Revision 1 05/30/12 Page 1 of 5. NRF Dektak II SOP

F7000NV ROBOT VISION OPERATING MANUAL

USER MANUEL. SNIPE 2 Ref R13

PSC300 Operation Manual

BitWise (V2.1 and later) includes features for determining AP240 settings and measuring the Single Ion Area.

Fluke 190-Series II Firmware Upgrade V11.44

Personal Protective Equipment Wear nitrile gloves, lab coat, and safety glasses as a minimum protection, unless otherwise indicated.

First Question: Camera head. Lighting unit. Shooting stage

Data Collection Using APEX3. March 30, Chemical Crystallography Laboratory

Chapter 4. Dish Antenna Installation. Installing a DISH 500 Antenna. Finding the Satellites

Three Axis Digital Readout System

NanoGiant Oscilloscope/Function-Generator Program. Getting Started

Nanonis STM Simulator Tutorial

Spectrum Analyser Basics

JS007WQK HEAVY DUTY WIRELESS REVERSING KIT 7 LCD DIGITAL QUAD RECORDING MONITOR with WATERPROOF CCD CAMERA

Technical Procedure for Scanning Electron Microscope/ Energy Dispersive X-Ray System (SEM/EDX) for non-gsr Casework

DektakXT Profilometer. Standard Operating Procedure

imso-104 Manual Revised July 19, 2012

NOTICE: This document is for use only at UNSW. No copies can be made of this document without the permission of the authors.

Spectroscopy Module. Vescent Photonics, Inc E. 41 st Ave Denver, CO Phone: (303) Fax: (303)

Video wall Precautions when moving and handling. Precaution. Please adhere to the warning signs printed on the box.

CNC Router Parts Plasma Software Setup and Usage Guide

GEOSATpro GS120. DiSEqC 1.2 Motorized H-H H Motor

Classroom Setup... 2 PC... 2 Document Camera... 3 DVD... 4 Auxiliary... 5

XRD-DSC Sample Alignment (BB) Part

Instruction manual. KUZMA 4POINT 14 inch TONEARM Serial Number:

MTI-2100 FOTONIC SENSOR. High resolution, non-contact. measurement of vibration. and displacement

J.M. Stewart Corporation 2201 Cantu Ct., Suite 218 Sarasota, FL Stewartsigns.com

Using AFM Phase Lag Data to Identify Microconstituents with Varying Values of Elastic Modulus

HRXRD Analysis of Epitaxial Thin Films

KVANT. Spectrum 1,6W RGB laser system. Product datasheet

FOSC-600 C and D I N S T A L L A T I O N I N S T R U C T I O N

nanoir2-s TM System User s Manual Part # Issued August 2016 Page 1 of 52

Software manual. ChipScan-Scanner 3.0

B2 Spice A/D Tutorial Author: B. Mealy revised: July 27, 2006

Fully ly Automaticti. Motorised Satellite t TV System. User s manual. ver 3.0.

J R Sky, Inc. Cross-Modulation Distortion Analyzer

Instructions for Use: Video Inspection Scope with Display

Operating Manual. Automated Gear. Apollo Design Technology, Inc Fourier Drive Fort Wayne, IN USA

User Manual VM700T Video Measurement Set Option 30 Component Measurements

LUMIGEN INSTRUMENT CENTER X-RAY CRYSTALLOGRAPHIC LABORATORY: WAYNE STATE UNIVERSITY

Dave Jones Design Phone: (607) Lake St., Owego, NY USA

Using the Agilent for Single Crystal Work

1. Check the accelerating voltage, must be at 200 kv (right screen), HT (µa) (left panel) at and Emission (left panel) at

Preface... Specification...14

Working with a Tektronix TDS 3012B Oscilloscope EE 310: ELECTRONIC CIRCUIT DESIGN I

Characterization of Diamond Samples CHESS Run Summer 2013

Manual placement system MPL3100. for BGA, CSP and Fine-Pitch components

Subj: General MTI2000 and Piezo Task Setup and Operation. High-Voltage Displacement Meter (HDVM) Configuration.

Installation and User Guide 458/CTR8 8-Channel Ballast Controller Module

Booya16 SDR Datasheet

Please feel free to download the Demo application software from analogarts.com to help you follow this seminar.

How to Optimize Ad-Detective

Quick Start. RSHS1000 Series Handheld Digital Oscilloscope

Preface. The information in this document is subject to change without notice and does not represent a commitment on the part of NT-MDT.

Indiana Center for Biological Microscopy. Zeiss LSM-510. Confocal Microscope

OM2000N INSTALLATION MANUAL

15 Inch CGA EGA VGA to XGA LCD Wide Viewing Angle Panel ID# 833

26 Inch CGA/EGA/VGA/DVI to WXGA/1080p LCD - ID#703

Electrical connection

A-ATF (1) PictureGear Pocket. Operating Instructions Version 2.0

Fully ly Automaticti. Motorised Satellite t TV System. User s manual REV

ivw-ud322 / ivw-ud322f

Transcription:

AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode) 1. Log into the Log Usage system on the SMIF web site 2. Open Nanoscope 6.14r1 software by double clicking on the Nanoscope 6.14r1 desktop icon 3. Select the desired mode of operation by choosing the appropriate workspace in the File/Open Workspace menu. The workspaces are found in the directory D:V614r1/Workspaces. The following workspaces are good starting points and can be used for most samples: Tapping Mode: smif_default_tapping_d.wks Contact Mode: smif_default_contact_d.wks 4. Click on Scan in the Workspace window to connect the software to the instrument 5. Mount the appropriate AFM tip into the cantilever holder. Be sure that the tip is in contact with the back end and left side of the groove per the following picture. TESP Tip properly positioned in cantilever holder prior to clamping TESP Tip properly positioned in cantilever holder after clamping You may use your own AFM tip, or purchase one of the following from SMIF: Tapping Mode: TESP silicon probe Contact Mode: ORC8-10 silicon nitride probe 6. Mount the cantilever holder onto the end of the scanner head a. Tighten the set screw to unclamp the scanner head from the dove tail and remove the scanner head from the cradle. Be sure not to pull on the cables that connect the scanner head to the electronics box b. Mount the cantilever holder by gently fitting it onto the pins at the end of the scanner head. DO NOT use excessive force as this may damage the scanner head c. Replace the scanner head in the cradle and loosen the set screw to clamp the head assembly in the dove tail 7. Align the laser a. Turn the illumination off by entering a 0 in the Illumination box of the Other controls menu in the Scan ParmList window. b. Place the blank side of a business card face up on the granite stage so that the laser is visible on the card c. Open the Meter window by clicking on Meter in the Workspace window Page 1 of 6 Revision 7

d. Using the two knobs at the top of the scanner head, follow the diagram below to adjust the laser position so that the laser is visually aligned with is the tip of the cantilever on the AFM tip. 1. Move the laser in the x direction to find AFM tip edge. Then move slightly back from the edge. 2. Move the laser in the y direction to find the cantilever. 3. Move the laser in the x direction to the right to find the cantilever edge. Now make fine adjustments to the x and y laser position knobs to maximize the signal sum in the meter window as shown below Red dot centered for detector alignment Signal sum maximized for laser alignment e. Remove the business card from the granite stage 8. Align the photodetector a. Use the two knobs on the left side of the scanner head to adjust the photodetector position 1. Coarse adjustment: Center the position of the red dot in the window on the scanner head 2. Fine adjustment: Center the position of the red dot in the cross-hairs of the meter display screen as shown above Tapping Mode: Set the Vertical Deflection to 0.0 volts ( 0.5V) Contact Mode: Set the Vertical Deflection to about 2.0 Volts b. Close the Meter window Page 2 of 6 Revision 7

9. Load sample a. Place sample on stage b. Select Navigate window and move the Z-position up using the Z-motor up arrow button in the Navigate window. (This is a precaution to prevent the sample from hitting the head assembly) c. Increase illumination from zero to 100 d. Move the sample under the probe using the X-Y stage movement arrows in the Navigate window. 10. Locate and focus on tip a. Click on Locate Tip in the Navigate window b. Zoom all the way out by clicking on Zoom Out c. Center the tip of the cantilever under the cross hairs using the two adjustment knobs located to the left of the optical objective on the microscope d. Focus on the tip end of the cantilever using the Optics arrows keys in the Navigate window e. Click on Zoom in to zoom all the way in or use the up arrow to zoom partially in on the tip and re-focus and re-center the tip if necessary. f. Click OK (The software will automatically change to the focus on sample surface menu) 11. Focus on the sample surface a. Zoom out to the maximum position to get the largest field of view possible b. Use the Z-motor down arrow button in the Navigate window to visually close the gap between the sample and the scanner head to about 2-3mm. c. Now look at image on screen and keep adjusting the Z-motor position until the sample surface is in focus. If the sample is smooth and has no surface features to focus on, then change the Focus On setting to Tip Reflection and adjust the Z-position until the reflection of the AFM tip is in focus. Page 3 of 6 Revision 7

d. Move to the desired measurement area of the sample under the cross hairs using the X-Y arrow keys in the Navigate window e. Zoom in on the area of interest and fine focus using the Z-position if necessary 12. Tune Cantilever (Tapping Mode AFM only) a. Click on the Tuning Fork Icon b. Set the Start and End frequency to bracket the resonant frequency value given on the box that the cantilever probe came from. (Typically 0kHz 500kHz for TESP silicon probes) c. Set the target amplitude to 2 volts d. Set the peak offset to 5% e. Click on Auto Tune button f. Note: If the tool does not auto tune properly and the probe tip is in good condition, try resetting the controller communication connection by selecting Tools/Reset Controller and then repeating the auto tune. g. When completed, click the Exit button 13. Set Initial Scan Parameters a. In the Scan Controls panel, enter the following settings: 1. Scan Size: Set to desired measurement area (< 70 m) 2. Aspect Ratio 1:1 3. X and Y offsets: 0 4. Scan angle: 0 5. Scan rate: Tapping mode: 1 Hz Contact mode: 2 Hz 6. Samples/line: 256, 512, or 1024 7. Lines: 256, 512, or 1024 8. Slow scan enabled 9. XY closed loop Lower setting: Lower resolution, faster speed Higher setting: Higher resolution, slower speed Off for scan size of < 1 m On for scan size of > 1 m b. For Tapping Mode AFM, in the Feedback Controls Panel, enter the following settings: 1. Integral Gain: 0.5 2. Proportional Gain: 0.7 3. Leave other settings alone c. For Contact Mode AFM, in the Feedback Controls Panel, enter the following settings: 1. Setpoint: 0 volts (2 volts difference from value set in Step 8) 2. Integral Gain: 2.0 3. Proportional Gain: 3.0 14. Close the instrument cover 15. Click on the Engage Icon Engage Withdraw Page 4 of 6 Revision 7

16. Optimize Scan Parameters a. Tapping Mode AFM 1. If the line in the Z-Center position moves far to the extended end (indicator turns red), the tip has false engaged. Disengage the probe tip and execute the engage command again 2. Check to see if Trace and Retrace are tracking each other well. If they are tracking, the lines should look the same, but they will not necessarily exactly overlap each other. 3. If they are not tracking well, adjust the Setpoint and/or Gains, to improve the tracking (See training notebook pages 36-41). It is often easier to disable the slow scan axis while making adjustments, so that the scanned feature stays in the field of view. Typical adjustments are as follows: Lower the setpoint to improve tracking between the trace and re-trace Lower the gains to reduce the noise b. Contact Mode AFM 1. Once engaged increase the Setpoint by using the right arrow key and observe the Z-Center Position. If the line in the Z-Center position moves far to the extended end (indicator turns red), the tip has false engaged. Disengage the probe tip, increase the setpoint by +2 volts and execute the engage command again. If the Z-Center position does not change greatly, you are on the sample surface and can go to the next step. 2. Check to see if Trace and Retrace are tracking each other well. If they are tracking, the lines should look the same, but they will not necessarily exactly overlap each other. 3. If they are not tracking well, adjust the Setpoint, Gains, and/or Scan Rate to improve the tracking (See training notebook pages and 36-41). It is often easier to disable the slow scan axis while making adjustments, so that the scanned feature stays in the field of view. 4. Remember that in Contact Mode, increasing the Setpoint increases the force on the sample. Minimizing the imaging force is recommended to prevent damage to the sample and the probe tip. 17. Set desired Scan Size, Scan Angle, and Offsets Page 5 of 6 Revision 7

Image Capturing 1. Choose Capture Filename from the RealTime pull down menu 2. Choose your directory to store the image, and enter a filename for the image. Note: directories must be created in Windows outside of the AFM software. Click OK. 3. Click on the camera icon to begin capturing an image. The image will begin capturing at the beginning of a scan frame. To move the probe to the beginning of a scan frame, click on the frame up or frame down icon 4. When capturing, the status bar at the bottom of the screen will indicate Capture: On, and when the capture is completed, it will indicate Capture: Done When you are finished: 1. Click on the Withdraw icon to stop scanning Engage Withdraw 2. Open the instrument cover 3. Select the Navigate window a. Raise the Probe head away from the sample using the Z-motor up arrow in the Navigate window b. Move the stage to the load position (completely out from under the scanner head) using the X-Y stage arrows in the Navigate window 4. Unload your sample 5. Remove the cantilever holder from the end of the scanner head and unmount your probe (Store your probe in your personal Gel-Pak storage box) 6. Close the instrument cover 7. If you wish to save changes to the workspace, Choose File/Save Workspace As and give the file a new workspace name. DO NOT save changes to the SMIF default workspaces 8. Exit Nanoscope software 9. Logout of the Log Usage system on the SMIF web site Page 6 of 6 Revision 7