Veeco Dektak 6M Profilometer

Similar documents
Dektak II SOP Revision 1 05/30/12 Page 1 of 5. NRF Dektak II SOP

Dektak Step by Step Instructions:

DektakXT Profilometer. Standard Operating Procedure

Approved by Principal Investigator Date: Approved by Super User: Date:

Cover Page for Lab Report Group Portion. Boundary Layer Measurements

Figure 1. MFP-3D software tray

SPM Training Manual Veeco Bioscope II NIFTI-NUANCE Center Northwestern University

E X P E R I M E N T 1

Scanning Electron Microscopy (FEI Versa 3D Dual Beam)

Standard Operating Procedure of nanoir2-s

Processing data with Mestrelab Mnova

KRAMER ELECTRONICS LTD. USER MANUAL

Operation Procedure for Phillips XL30 ESEM

Standard AFM Modes User s Manual

Tutorial 3 Normalize step-cycles, average waveform amplitude and the Layout program

MestReNova A quick Guide. Adjust signal intensity Use scroll wheel. Zoomen Z

2.1. Log on to the TUMI system (you cannot proceed further until this is done).

PHI 5000 VersaProbe TM Operator s Guide

System Requirements SA0314 Spectrum analyzer:

Linkage 3.6. User s Guide

University of Minnesota Minnesota Nano Center Standard Operating Procedure

MestReNova Manual for Chem 201/202. October, 2015.

Data Acquisition Using LabVIEW

PYROPTIX TM IMAGE PROCESSING SOFTWARE

BitWise (V2.1 and later) includes features for determining AP240 settings and measuring the Single Ion Area.

LUMIGEN INSTRUMENT CENTER X-RAY CRYSTALLOGRAPHIC LABORATORY: WAYNE STATE UNIVERSITY

Getting started with

GS122-2L. About the speakers:

SEM- EDS Instruction Manual

TT AFM LongBeach Procedures and Protocols V2.1

SNG-2150C User s Guide

STANDARD OPERATING PROCEDURE: ASYLUM MFP-3D AFM

Data Collection Using APEX3. March 30, Chemical Crystallography Laboratory

Quick Start ATOMIC FORCE MICROSCOPE West Campus Imaging Core

BEAMAGE 3.0 KEY FEATURES BEAM DIAGNOSTICS PRELIMINARY AVAILABLE MODEL MAIN FUNCTIONS. CMOS Beam Profiling Camera

Preparing for remote data collection at NE-CAT

Quick Start Bruker Dimension Icon AFM

The PHI VersaProbe operates with two essential software programs: PHI Summitt and Vacuum Watcher. A third program, MultiPak, handles data reduction.

Wafer defects can t hide from

Contacts: English Department Office 238 Moreland Media Services 109 Kidder

USING MEDIACAST FOR LIVE EVENTS THERE ARE THREE SEPARATE KEYS NEEDED TO ACCESS THE FRONT AND BACK OF THE VIDEO CART.

READ THIS FIRST. Morphologi G3. Quick Start Guide. MAN0412 Issue1.1

Grid Interactive Advanced Features Self-Supply Mode

Camtasia for Mac. Create Your First Video Guide. Release 1.1. February TechSmith Corporation. All rights reserved.

USER MANUAL FOR DDT 2D. Introduction. Installation. Getting Started. Danley Design Tool 2d. Welcome to the Danley Design Tool 2D program.

Table of Contents Introduction

Topic: Instructional David G. Thomas December 23, 2015

PHY221 Lab 1 Discovering Motion: Introduction to Logger Pro and the Motion Detector; Motion with Constant Velocity

EDL8 Race Dash Manual Engine Management Systems

POL-200 Semiautomatic Polarimeter. Instruction Manual BANTE INSTRUMENTS CO., LTD

SIDRA INTERSECTION 8.0 UPDATE HISTORY

Transmitter Interface Program

Asylum MFP-3D Standard Operating Procedures

The BAT WAVE ANALYZER project

Analyzing Numerical Data: Using Ratios I.B Student Activity Sheet 4: Ratios in the Media

PS User Guide Series Seismic-Data Display

SOUND LABORATORY LING123: SOUND AND COMMUNICATION

Motion Amplification

TechNote: MuraTool CA: 1 2/9/00. Figure 1: High contrast fringe ring mura on a microdisplay

Magnetic Force Microscope (MFM)

Practicum 3, Fall 2010

AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode)

K Service Source. Apple High-Res Monochrome Monitor

A-ATF (1) PictureGear Pocket. Operating Instructions Version 2.0

Class Notes for Cite While You Write Basics. EndNote Training

Operating Guide. ViewClix offers a revolutionary experience for seniors and their families and friends.

Durham Magneto Optics Ltd. NanoMOKE 3 Wafer Mapper. Specifications

HV/PHA Adjustment (PB) Part

K Service Source. Apple High-Res Monochrome Monitor

Connection for filtered air

TL-2900 AMMONIA & NITRATE ANALYZER DUAL CHANNEL

ME EN 363 ELEMENTARY INSTRUMENTATION Lab: Basic Lab Instruments and Data Acquisition

Asylum MFP-3D Standard Operating Procedures

Precautions and Disclaimers What You Can Do with Geometry Manager Pro Check Your Computer System requirements...

ebars (Electronic Barcoded Assets Resource System) ebars: ANNUAL PHYSICAL EQUIPMENT INVENTORY INSTRUCTION MANUAL

Source/Receiver (SR) Setup

LedSet User s Manual V Official website: 1 /

9070 Smart Vibration Meter Instruction Manual

MIE 402: WORKSHOP ON DATA ACQUISITION AND SIGNAL PROCESSING Spring 2003

Maintenance Service Bulletin

User Manual. English. Sequencer Control Option BE3200. I en HBM: public

Import and quantification of a micro titer plate image

Reason Overview3. Reason Overview

Spinner- an exercise in UI development. Spin a record Clicking

MODFLOW - Grid Approach

S-Series Server Setup Quiz

BME 3512 Biomedical Laboratory Equipment List

University of Minnesota College of Science and Engineering Characterization Facility Ganesha SAXSLAB User manual

using the Scott A Speakman, Ph.D Center for Materials Science and Engineering at MIT

McIDAS-V Tutorial Using HYDRA to Interrogate Hyperspectral Data updated September 2015 (software version 1.5)

v. 8.0 GMS 8.0 Tutorial MODFLOW Grid Approach Build a MODFLOW model on a 3D grid Prerequisite Tutorials None Time minutes

Matrox PowerStream Plus

Event recording (or logging) with a Fluke 287/289 Digital Multimeter

imso-104 Manual Revised August 5, 2011

Precision DeEsser Users Guide

Manual Version Ver 1.0

127566, Россия, Москва, Алтуфьевское шоссе, дом 48, корпус 1 Телефон: +7 (499) (800) (бесплатно на территории России)

DS-575W User's Guide

Using DICTION. Some Basics. Importing Files. Analyzing Texts

Guide to Analysing Full Spectrum/Frequency Division Bat Calls with Audacity (v.2.0.5) by Thomas Foxley

Transcription:

Veeco Dektak 6M Profilometer System Ranges/Resolutions Range (Å) Resolution (Å) 50 (5nm) to 65K 1 0.5K to 655K 10 2K to 2620K 40 8K to 10000K (1mm) 160 Maximum sample thickness: 31.75mm Scan range: 50 to 30,000um (30mm) Scan time: 3 to 100 seconds Data points: 300 pts/sec Note: This unit uses a 12.5 micron diamond stylus, which will scratch softer surfaces. The stylus is also very expensive, please be careful.

Under no circumstances should any user attempt any repairs or modifications to the machine. Please see a staff member if you have any questions, or desire any special processing not covered in these instructions. 1. Starting up: A. Log on to the instrument using the equipment log on computer B. Sign into the Dektak computer as user Dektak6M, using the password listed on the computer monitor. C. Click on the Dektak32.exe icon to start the program. Wait for 1 minute while the instrument is initialized. 2. Setting up scan parameters: A. Go to Window > Scan Routines. Click on the Scan Parameters box to change the scan variables. B. Variables depend on your sample and preferences: 1. ID: Arbitrary number used to identify your sample, if desired. 2. Length: Scan length desired, in microns (50-30,000 microns). Longer baseline measurements are usually more accurate. 3. Duration: Time taken to run the scan (3-100 s). Values of 10-30 seconds are typical, and longer durations lead to increased accuracy in measurements.

4. Sample: Number of data points taken per scan (300 points per second duration). 5. Resolution: Dependent upon Length and Duration. 6. Stylus Force: Force exerted by stylus (1-15 mg). Photoresist measurements should be taken with a stylus force of 1-3 mg to avoid damaging the photoresist. 7. Measurement Range: Describes the vertical resolution of the sample. (650 nm to 1 mm) 655 kǻ is typical. Higher values are appropriate for very deep or rough structures. 8. Profile: Select Hills, Valleys, or Hills and Valleys depending on the topography of your sample. Select Hills and Valleys if you are unsure. C. When the variables have been set, click OK, then the Sample Positioning button. 3. Loading a sample: A. Click on Window Sample Positioning to open the control screen, or use the sample positioning button on the left side of the screen. B. Verify that the tower is in the up position before loading your sample, and whenever you are removing or touching the sample. The stylus needle should be about 1 inch from the stage. If not, press the Tower Up button. C. Load your sample onto the stage roughly in position under the stylus (you will be able to see your sample in the monitor to fine tune the position later). TE: The sample should be arranged such that a front-to-back scan yields the desired information. The image on the computer monitor is rotated 90 degrees counter-clockwise relative to the view as you face the instrument.

Profilometer Controls 4. Taking a measurement: A. When your sample has been roughly positioned, click the Tower Down button. The stylus will slowly move towards your sample, touch the sample briefly, and rise back up a small (~1mm) amount. If the stylus does not retract, click Release stylus. B. A video image of the needle and stage should be visible. If not, adjust the lighting using the Increase Light Intensity Intensity (shaded) button. button or Decrease Light C. Zoom and focus can be adjusted using the lower and upper focus rings on the microscope, respectively. D. The green crosshairs show approximately where the stylus will touch down on the sample. Adjust the stage using the knobs in front of the stage and/or rotating the top of the stage until the sample is located under the crosshairs.

(The scan moves from left to right on the screen, front to back on the sample from the perspective of the user.) E. When the sample is in position, lower the lid and click the Scan button. F. If there is a problem with the scan, or the stylus is about to run off of the sample, click the red ABORT button on the left side of the screen. If after aborting the stylus is touching the sample, press the Stylus Up button. G. The scan will display in real-time, then it will automatically redraw/rescale it and trace analysis screen will come up. H. To manually level the stage (when the plot goes diagonally off the screen): 1. Turn the large horizontal black dial at the front of the stage (between the knobs) as you scan a flat portion of your sample. 2. Try to level the trace, turn Clockwise to raise the trace, turn Counter- Clockwise to lower the trace. Note: moving the wheel 1cm will raise the trace about 3Å per micron traveled i.e., for a 2000 micron trace, moving the wheel by 1cm will raise or lower the trace by about 0.6 microns. 5. Data trace adjustments: A. The stylus trace will appear in the dataplot window, and parameters can be adjusted after about 30 seconds (while the stylus returns to its original position).

B. The cursors labeled R and M refer to Reference and Measurement, respectively. The sampling width of the cursors can be adjusted by clicking Plot Band Width. The values in this window are in microns. C. To Level the graph to your substrate: 1. Select the R cursor (lower right panel), move it with the arrows keys, click on the F button to switch to fine tuning mode; or enter a um value directly in the display box (lower center panel) to select a point on your reference surface. 2. Select the M cursor, move it to the other side of your feature on the same reference surface. 3. Click the Plot Level button.

D. To Zero the graph to your substrate: 1. Select the R cursor, move it with the arrow keys to your reference surface. 2. Go to Plot > Zero to redraw the graph. E. To Zoom In on a portion of the graph: 1. Click on the graph and drag a box around the area to be magnified. 2. Go to Plot > Replot to redraw the graph. 3. To Zoom Out to the original settings, click on Plot > Replot again. 6. Analytic data functions: A. The height of a step can be determined by moving R and M to the desired locations and looking at the vertical distance between them (Vert_D at the bottom of the screen). A more accurate determination can be made by averaging over a section at the top and bottom of a step. To do this: a. Click Analysis Analytical Functions. b. In the menu, select the box labeled ASH (Average Step Height) c. Choose the length of the trace you wish to average over, for R and M, in the pop-up box, and click OK. d. Select the Measure box at the bottom of the menu, and click Compute to calculate the step height. R and M can be further adjusted if desired. e. Note: Each time you Compute, it will add the new value(s) to the Analytic Results box. B. Other measurement features: 1. Area under the curve between the R and M cursors with respect to the zero grid line. a. Position the R and M cursors. b. Go to Analysis Analytical Functions select the box labeled Area from the menu, and click Compute. 2. R a (Average Roughness) of the surface between the R and M cursors with respect to a mean line drawn through the surface trace. a. Position the R and M cursors. b. Go to Analysis Analytical Functions, select the box labeled Ra from the menu, and click Compute. 3. AvgHt (Average Height) of the surface between the R and M cursors with respect to the zero grid line. a. Position the R and M cursors. b. Go to Analysis Analytical Functions, select the box labeled AvgHt from the menu, and click Compute.

4. Other measurements can be made from the Analytical Functions menu. See the Help menu for detailed instructions. (Relevant Help sections that may be useful are listed under the Index tab: Roughness Parameters, Waviness Parameters, Height Parameters, Geometry Parameters. ) 7. Exporting Data and/or Photos: A. Click File Export to save the settings and data to a *.txt file. This file can then be opened in Excel for further processing. B. Click on the Sample Positioning button, click File > Save Video Image to save the image as a *.jpg, *.bpm, *.tif, or *.tga image file. C. Take a screen shot: Ctrl+Print Screen, then paste the copied screen shot into MS Paint. 8. Taking another measurement: A. To return to the sample to make another measurement or remove the sample, click on the Sample Positioning button. If the wafer needs to be moved, click the Stylus Up button and wait for the stylus to move completely upwards. 9. Removing the sample: A. Click the Tower Up button and wait for the tower to move completely upwards. B. Remove the sample, and return the stage to a location roughly centered beneath the stylus using the X and Y knobs. 10. Shutting down: A. Make sure that: a. the tower is UP. b. the Dektak32.exe program is closed. i. When asked Do you wish to save your current changes to the current automation program? answer! ii. When asked Do you wish to save the current scan data and analytical function results? answer YES if desired (and you haven t already), and follow the instructions. Save your file as *.txt if you want to open it in EXCEL. c. you are logged off of the instrument at the central LogOn computer.

Under no circumstances should any user attempt any repairs or modifications to the machine. Please see a staff member if you have any questions, or desire any special processing not covered in these instructions.

Sample *.JPG Image and Excel Data 2000um Scan 25000 Surface Profile Angstroms 20000 15000 10000 5000 0-5000 0 200 400 600 800 1000 1200 1400 1600 1800 2000 um

2009-09-24-10:52:22 DATA FILE: Default.001 Date: 9/24/2009 10:41 AUTO PROG: Default.mp Units: SCI ScID: test5 Sclen: 2000 Stitch: #Stchs: 1 Speed: Select Mrange: 65kÅ Prof: H & V Tspeed: High Force:4.44659e-3231073741824 Drnge-: 0 Drnge+: 0 Rcur: 1109 Rwidth: 50 Mcur: 1917 Mwidth: 25 RVTI: 14029 MVTI: 14169 Disp: Raw Smthed: Wavfed: Ruffed: Smdeg: 1 Smband: 0 HPF: 0 LPF: 0 BP: Calib: VSF: 1.00E+00 VCF: 9.97E-01 Levpt: 7.50E+01 Levslp: -1.11E+01 Zeropt: 3.39E+04 Horzpt: 0.00E+00 NumPts: 18000 Hsf: 1.11E-01 Function Result R.Cur M.Cur R.Width M.Width Ash 14224 851 1148 50 25 Ra* 376 1109 1917 50 25 Maxdev* 5160 1109 1917 50 25 Ash 67 1109 1917 50 25 SCALED DATA: Surface Profile 27-4 -130-197 -138-55 -73-163 -177-42 124 146 20-87 -75-23 -65-175 -195-52 126 171 66-39 -37 18 16-56 -99-41 56 74-10 -100-102 -49-38 -100-156 -124-26 44 50 45 75 123 118 51-11 -5