H1076X 0.6 1.5 H1076Y 0.6 1.5 Print No.HA1006 1 Release Date: 2013 07 15
Description X-ray Tube Housing Assembly The tube unit is so constructed that an X-ray tube is sealed in a diagnostic type protective tube housing of electric shockproof, radiation protection, etc X-ray tube assembly with double-focus rotating anode X-ray tube H1076 is intended to be used for all routine diagnostic examinations with conventional or digital radiographic and fluoroscopy workstations of OEM (Original Equipment Manufacturer). The designation is composed as follows: H1 076 X Housing one KL76 tube High voltage socket with 90 degree direction H1 076 Y Housing one KL76 tube High voltage socket with 270 degree direction Kailong product version adheres to international standards, e.g. CE, FDA, SFDA. Special features Long-term high dose yield Focal spots are superimposed Compact tube assembly housing - with optional fan cooling system for high loads Print No.HA1006 2 Release Date: 2013 07 15
Technical Data Property Specification Standard Nominal input power(s) of the anode F 1 F 2 IEC 60613 18kW(50Hz) 19kW(60Hz) 50kW(50Hz) 51kW(60Hz) Anode heat storage capacity 160 k IEC 60613 (220kHU) Maximum cooling capacity of the anode 560W Heat storage capacity 900kJ IEC 60613 Max. continuous heat dissipation 180W IEC 60613 without Air-circular Anode material Anode top coating material Rhenium-Tungsten-TZM(RTM) Rhenium-Tungsten-(RT) Target angle (Ref: reference axis) 14 IEC 60788 Total filtration = inherent filtration + additional filtration 2.5 mm Al / 75kV 1.5 mm + (2x 0.5) mm Al IEC 60601-1-3 IEC 60522 Focal spot nominal value(s) F1(small focus) F2(large focus) IEC 60336 0.6 1.5 X-ray tube nominal voltage 125kV IEC 60613 Max. Tube Current Large Focus 800mA Small Focus 340mA Data on cathode heating /AC, < 20 khz Max. current Max voltage Leakage radiation at 125 kv / 3mA in 1m distance X-ray tube assembly weight X-ray Tube Housing Assembly F1 F 2 5.4 A 5.4 A 13 V 18 V 0,8 mgy/h Approx. 18 kg IEC60601-1-3 Print No.HA1006 3 Release Date: 2013 07 15
H1074X,H1074Y 1.0 2.0 Conditions for operation, storage and transport Operation Limits Operation Transport and storage Ambient temperature From 10 C to 40 C From - 20 C to + 70 C Relative humidity From 30 % to 75 % From 5% to 95% Barometric pressure From 700 hpa to 1060hPa From 500 hpa to 1060 hpa Stator key values 1-phase stator Test point C-P C S Winding resistance 18.0... 22,0 Ω 45.0... 55,0 Ω Max.permissible operating voltage(run-up) 230V±10% Recommend operating voltage(run-up) 160V±10% Braking voltage 70VDC Run-on voltage in exposure 80 V rms Run-on voltage in fluoroscopy 20V -40V rms Run-up time(depending on starter system) 1.2s Print No.HA1006 4 Release Date: 2013 07 15
Heating and cooling curve of anode Heating and cooling curves of X-ray tube assembly Print No.HA1006 5 Release Date: 2013 07 15
Emission curves of the cathode Print No.HA1006 6 Release Date: 2013 07 15
Single and series loading Print No.HA1006 7 Release Date: 2013 07 15
Maximum radiation field The field coverage depends on the source-image distance (SIO) and the anode angle. For example field coverage of 48cm x 48cm can be achieved at 1 m SIO with this tube assembly (14 anode angle) Print No.HA1006 8 Release Date: 2013 07 15
X-ray tube assembly dimensional drawings X-ray Tube Housing Assembly Print No.HA1006 9 Release Date: 2013 07 15
Notes!!! In the interest of complying with legal requirements concerning the environmental compatibility of products (protection of natural resources, avoidance of waste) we endeavor to reuse components and to return them to the production cycle. We guarantee the reliability and functionality, quality and life of these components by taking extensive quality assurance measures including all test procedures, just as for brand new components. Reused parts are qualified-as- good-as-new. The requirements of IEC 62309 are fulfilled. Kailong Medical Solution has performed a conformity assessment in accordance with the MDD for this component. The tube assembly is not a sterile product. The housing surfaces are designed for easy cleaning disinfection (damp cloth). X-ray Tube Housing Assembly Applications with contacts to animal or human tissue or in body cavities are not intended. Use the original packaging for returning the tube assembly to the manufacturer. With increased tube assembly load one or two fans can be used in addition. In single fan operation the fan must be provided basically on the anode side. The fans are operated at an alternating voltage of 24 V (50/60 Hz) and a power of approx. 20 W. The installation and the connection of the fans are described in the installation instructions for the fan. Fans must be cleaned once annually. Your customer service can do this correctly. Unless otherwise indicated, all load data specified apply to six-pulse, twelve-pulse, multi-pulse or DC voltage generators. Caution! Handle with Care. Print No.HA1006 10 Release Date: 2013 07 15
Caution At high housing temperature take care to avoid contact when the protective cover is missing. Prior to an examination it is your responsibility as the user to ensure that all safety features are functional and that the product is ready for operation. Caution, operation of the tube assembly only with housing covers fitted! To ensure the safety of both the user and the patient, the customer service must be notified immediately when a defect has been detected, and the operation of the tube assembly concerned must be terminated. Collimators with a weight up to 300N may be fastened to the collimator flange of the X-ray tube assembly. The described product (s) is/are not commercially available. Content serves as preliminary information. The Kailong Medical Instrument Co Ltd., is ISO 9001 certified, manufactures in accordance with the Quality System Regulations (QSR) as defined by the Food and Drug Administration (FDA) and endeavors to comply with legal requirements concerning the environmental compatibility of its products. The reproduction, transmission or use of this document or its contents is not permitted without express written consent. Offenders will be liable for damages. Kailong reserves the right to modify the design and specifications contained herein without prior notice. All rights reserved, particularly in connection with patent applications or registrations of utility model or design. Kailong Medical Instrument Co Ltd., For further information, please contact: Kailong Medical Instrument Co Ltd., Addres:6, No. 6 Road, Dongzhou Industrial Zone, Fuyang, Hangzhou, China 311401 Fax: +86571 87195003 http://www.kailongx-ray.com Print No.HA1006 11 Release Date: 2013 07 15