Tech Note: How to measure additive phase noise of amplifiers using the 7000 Series

Similar documents
Agilent E5500 Series Phase Noise Measurement Solutions Product Overview

7000 Series Signal Source Analyzer & Dedicated Phase Noise Test System

Model 7330 Signal Source Analyzer Dedicated Phase Noise Test System V1.02

Noise Detector ND-1 Operating Manual

Mastering Phase Noise Measurements (Part 3)

INTEGRATED ASSEMBLIES MICROWAVE SOLUTIONS FROM TELEDYNE COUGAR

R-1580A Microwave Downconverter. Product Brochure

Broadcast Television Measurements

Application Note 5098

User s Manual 2.4. SERIES 7000 Models MODEL 7070, MODEL 7300 SERIES 7000

Keysight Technologies E5500 Series Phase Noise Measurement Solutions

Quartzlock Model A7-MX Close-in Phase Noise Measurement & Ultra Low Noise Allan Variance, Phase/Frequency Comparison

Application Note DT-AN-2115B-1. DTA-2115B Verification of Specifations

Analyze Frequency Response (Bode Plots) with R&S Oscilloscopes Application Note

Advanced Test Equipment Rentals ATEC (2832)

Application Note DT-AN DTU-315 Verification of Specifications

USB-TG124A Tracking Generator User Manual

HP 71910A and 71910P Wide Bandwidth Receiver Technical Specifications

Model RTSA7550 Specification v1.1

SYNTHESIZED COMMUNICATION CONVERTERS

RF Measurements You Didn't Know Your Oscilloscope Could Make

MP5000 Wireless Test Station

High-Value 100 Series (Outdoor)

R&S FSV-K40 Phase Noise Measurement Application Specifications

SHF Communication Technologies AG,

SIDC-5009 Series VHF/UHF WIDEBAND TUNER/CONVERTER. FREQUENCY RANGE: 20 to 3000 MHz

High Performance Outdoor Test Translators

MILLIMETER WAVE VNA MODULE BROCHURE

ANSI/SCTE 40 Conformance Testing Using the R&S SFU, R&S SFE and R&S SFE100

Product Brochure. MP5000 Wireless Test Station

Agilent CSA Spectrum Analyzer N1996A

The high-end network analyzers from Rohde & Schwarz now include an option for pulse profile measurements plus, the new R&S ZVA 40 covers the

MAXTECH, Inc. BRC-1000 Series. C-Band Redundant LNB Systems. Technology for Communications. System Block Diagrams

USB Mini Spectrum Analyzer User Manual TSA Program for PC TSA4G1 TSA6G1 TSA8G1

Calibrate, Characterize and Emulate Systems Using RFXpress in AWG Series

Spectrum Analyzer 1.6 GHz 3 GHz R&S HMS-X

R&S RSC Step Attenuator Where precise signal levels count

Features. PFD Output Voltage 2000 mv, Pk - Pk. PFD Gain Gain = Vpp / 2π Rad khz 100 MHz Square Wave Ref.

Technical Data. HF Tuner WJ-9119 WATKINS-JOHNSON. Features

RF Characterization Report

Modular Block Converter Systems

L-Band Block Upconverter MKT-74 Rev B JULY 2017 Page 1 of 7

R3267/3273 Spectrum Analyzers

SIDC-5004 VHF/UHF WIDEBAND TUNER/CONVERTER. FREQUENCY RANGE: 20 to 3000 MHz

SIDC-6003 MICROWAVE WIDEBAND DOWNCONVERTER / TUNER UP TO

VXI RF Measurement Analyzer

Keysight Technologies High Power Ampliier Measurements Using Nonlinear Vector Network Analyzer. Application Note

SIDC-6005 MICROWAVE WIDEBAND DOWNCONVERTER / TUNER UP TO

Switching Solutions for Multi-Channel High Speed Serial Port Testing

ENGINEERING COMMITTEE Interface Practices Subcommittee AMERICAN NATIONAL STANDARD ANSI/SCTE Composite Distortion Measurements (CSO & CTB)

The Distortion Magnifier

R&S SGMA Product Family Compact fast reliable

MODEL Band Channels Up/Down IF In/Out RF In/Out Step Size Remote M&C Price

Interface Practices Subcommittee SCTE STANDARD SCTE Measurement Procedure for Noise Power Ratio

FREQUENCY CONVERTER 1/3 RACK-MOUNTED BLOCK CONVERTER. Narda-MITEQ FEATURES OPTIONS. Unit shown with Option 17. Unit shown without Option 17

Product Brochure Version R&S RSC Step Attenuator Where precise signal levels count

Ku-Band Redundant LNB Systems. 1:1 System RF IN (WR75) TEST IN -40 db OFFLINE IN CONTROLLER. 1:2 System POL 1 IN (WR75) TEST IN -40 db POL 2 IN

Instrumentation Grade RF & Microwave Subsystems

R-1550A Tempest Wide Range Receiver

Supplemental Measurements of System Background Noise in 10GBASE-T Systems

Agilent 87075C Multiport Test Set Product Overview

Extension kit for R&S Vector Network Analysers

DVO700 P FIBRE OPTIC TRANSMITTER

Unicable II Programmable/cascadable multiswitch with 32 User Bands, 4 Universal/Wideband Satellite inputs & 1 Terrestrial input

Spectrum Analyzer 1.6 GHz 3 GHz HMS-X

USB Mini Spectrum Analyzer User Manual PC program TSA For TSA5G35 TSA4G1 TSA6G1 TSA12G5

R&S ETH Handheld TV Analyzer Portable DVB-T/H signal analysis up to 3.6/8 GHz

Signal Stability Analyser

SIR MICROWAVE WIDEBAND DSP RECEIVERS UP TO 26.5 GHz. WIDE FREQUENCY RANGE: GHz

Digital Lock-In Amplifiers SR850 DSP lock-in amplifier with graphical display

ThinkRF D GHz RF Downconverter

DA MHz Series of Narrowband or Wideband Distribution Amplifiers

RF Level Test System +20 dbm to 130 dbm

SNG-2150C User s Guide

DA : Series of Narrowband or Wideband Distribution Amplifiers

Introduction to the new software V230 for the minivna

Keysight Technologies High-Power Measurements Using the E5072A ENA Series Network Analyzer. Application Note

MULTIBAND 1/3 RACK-MOUNTED

SSA3000X Series Spectrum Analyzer

DA E: Series of Narrowband or Wideband Distribution Amplifiers

SHF Communication Technologies AG,

Transmitter Interface Program

SC24 Magnetic Field Cancelling System

MAMX Sub-Harmonic Pumped Mixer GHz Rev. V1. Functional Schematic. Features. Description. Pin Configuration 1

Prisma Optical Networks Ancillary Modules

How To Demonstrate Improved ACLR Dynamic Range With FSU and Noise Correction

PCIe: EYE DIAGRAM ANALYSIS IN HYPERLYNX

Evaluation Board For ADF Integrated VCO & Frequency Synthesizer

CHAPTER 3 SEPARATION OF CONDUCTED EMI

Namaste Project 3.4 GHz Interference Study Preliminary document - Work in Progress updated

Agilent ESA Series Spectrum Analyzers

ENGINEERING COMMITTEE

HMC7056. Block Upconverters / HPA's. Typical Applications. General Description. Features. Functional Block Diagram

OBSOLETE HMC7056. Block Upconverters / HPA's. Typical Applications. General Description. Features. Functional Block Diagram

SC24 Magnetic Field Cancelling System

Interface Practices Subcommittee SCTE STANDARD SCTE Composite Distortion Measurements (CSO & CTB)

1/3 RACK-MOUNTED BLOCK CONVERTERS

Sources of Error in Time Interval Measurements

MP7200. RF Recorder / Player.

FRQM-2 Frequency Counter & RF Multimeter

Transcription:

Berkeley Nucleonics Corporation Tech Note: How to measure additive phase noise of amplifiers using the 7000 Series Additive phase noise, also known as residual phase noise, is the self phase noise of a component that adds to an existing signal as the signal passes through it. Hence, the additive (residual) phase noise measurement is a valuable technique used to identify the phase noise contribution of a single component as part of a system design. This application note focuses on the single-channel additive phase noise measurement of an RF amplifier. Introduction One benefit of the 7000 Series is its capability to measure additive phase noise (also known as residual phase noise) of non-self-oscillating devices under test (DUT) such amplifiers, transmitters, mixers or prescalers. Although this can be done on a single channel; using a two-channel cross-correlation achieves the lowest phase noise. Theory The single channel additive phase noise measurement works as shown below in the schematic diagram (Figure 1). A stimulus source is set to the desired frequency. The signal is split into two paths. One path is fed into the DUT input and the other path is fed into the REF input. In one signal path a phase shifter is inserted to allow phase adjustments between REF and DUT port. The 7000 SERIES compares the phase of the DUT and REF paths using a low noise phase detector. Using the phase shifter, the two signal paths are set to phase quadrature (90 degrees apart) at the phase detector. It can be shown that under perfect quadrature the correlated noise in the two signal paths (in particular phase noise from the stimulus source) is cancelled out. Without any DUT, this measurement can be used to determine the setup phase noise floor. Adding the DUT into one signal path will add the DUT s phase noise and will be detected by the phase detector as deviation from quadrature (because it is only appearing in one signal path). Note that in a single channel measurement, the phase shifter and the DUT can be located in either the DUT or REF signal paths. This gives some additional flexibility in the measurement setup. Figure 1: Schematic of the Measurement Setup Berkeley Nucleonis Corporation - 2955 Kerner Blvd, San Rafael CA 94901 USA Tel: 415-453-9955 Fax: 415-453-9956 E: info@berkeleynucleonics.com Chat:www.berkeleynucleonics.com

Additive noise measurements are typically at a lower noise level and are more difficult to perform than absolute phase noise measurements, requiring careful measurement setup and calibration. While the setup involves traditional microwave plumbing, the calibration can be tedious and prone to significant error. Berkeley Nucleonics s 7000 Series of phase noise analyzers addresses this problem by completely automating the calibration. Measurement setup and procedure The hardware setup is shown below in Figure 2. In addition to the 7000 SERIES, a low phase noise signal source (in this case, the Model 845 20GHz signal generator is used), a power splitter and an electrically tunable or mechanical phase shifter are required. Figure 2: The Test Setup 2

Figure 3: GUI Screenshot of the "Additive Phase Noise" tab The measurement procedure can be split into two steps: A. Determine the measurement noise floor B. Perform the actual measurement on DUT Determining the setup noise floor (step A) is especially important to ensure that the DUT noise is properly defined and is not influenced by other noise contributors. This step may be omitted if instrument noise floor is already known for the particular setup. In step B, the DUT is added into one signal path and the measurement is repeated this time including the DUT. Both steps require a full measurement to be performed; the only difference is that the DUT is replaced with a cable to measure the noise floor. The GUI Setup Starting the GUI, first the connection to the 7000 SERIES is established and the Additive Phase Noise tab is selected (see Figure 3). The next steps are: 1. As in the absolute phase noise tab, select offset range and number of correlations. If additional signal attenuation or amplification is desired, then use the input attenuator/low noise amplifier stage. 3

Figure 4: Rotating Phase at the Phase Detector Using the Phase Shifter Determines Phase Detector Constant of the Measurement Setup Note: Care must be taken, that the DUT and the REF ports are operated at appropriate power levels: The DUT input power should be between +3 and +10 dbm. Above +10 dbm, the DUT input attenuator should be adjusted. Below +3 dbm, the input low noise amplifier may help to improve the measurement noise floor. The REF input should be operated between +9 and +17 dbm. Below +9 dbm the REF input low noise amplifier may be switched in to improve the noise floor. It is recommended to adjust the measurement setup if possible before using the buffer or attenuation options. For instance, the DUT can be placed either on the DUT or on the REF side which will affect the power levels at the DUT and RED input. To measure the input power of the signals, use the DUT input and press the "auto" button to start the frequency and power measurement of the attached signal. 2. Select the One Channel measurement. The phase shifter can be chosen to be either a manually controlled phase shifter (as in this example), a voltage controlled phase shifter (via the 7000 SERIES tuning port), or a remote controlled digital phase shifter from Colby Instruments (Colby compatibility not yet implemented, ask Berkeley Nucleonics support for status). 4

Figure 5: Adjusting to Phase Quadrature Using Phase Shifter 3. Once the setup is completed, the calibration procedure is started by clicking on the first Measure button. Using the phase shifter, the phase is now rotated continuously over 360 degrees to determine the phase detector constant (see Figure 4). Complete the calibration step by clicking on the corresponding "Confirm button. 4. In a next step, the signals at the phase detector are adjusted to quadrature (90 phase shifted in respect to each other) using the phase shifter (refer to Figure 5 above and Figure 6 below). Clicking on the second "Measure" button will continuously show the current phase error which has to be reduced to 0 by phase shifting the reference. Once the phase error is minimized, complete the calibration step by clicking on the second Confirm button. 5. Once the phase detector has near quadrature phases and is confirmed by the user, the Start button is enabled and the actual measurement can be performed (see Error! Reference source not found.) by clicking on the green "Start" button. The GUI will automatically switch to the phase noise plot and show the result after the measurement is finished. 5

Figure 6: Quadrature Achieved within Accepted Tolerance Range Analyzing the Result Figure 7 below shows the measured instrument noise floor (as measured in step A) of the particular setup as the blue curve. Using a stimulus source of +16 dbm at 2 GHz, the setup noise floor is lower than -175 dbc/hz for offsets above 10 khz and below -140 dbc/hz at 10 Hz offset. Using two channel cross-correlation, noise floors below -190 dbc/hz can be achieved. The measurement including the DUT amplifier is shown as the green curve. Figure 7: Phase Noise of the DUT Amplifier 6

In Figure 7, the measurement including the DUT amplifier is shown as the green curve. For offsets up 4 MHz, the DUT phase noise dominates the setup noise floor (i.e. shows a higher phase noise value), indicating that the measurement is dominated by the DUT noise. At higher offsets, the green trace reaches the setup noise floor. Only a cross-correlation measurement enables lower level measurements. Conclusion As shown in this application note, a simple procedure using the 7000 Series allows measurement of amplifier phase noise at very low phase noise levels. For lowest phase noise floors, the setup must be duplicated on a second measurement channel and cross-correlation must be applied. 7