Evaluation report. Eizo RadiForce G33-N 3MP greyscale flat panel liquid crystal display (LCD) CEP 07003

Similar documents
Guidance for Quality Assurance of PACS Diagnostic Display Devices

Display Quality Assurance: Considerations When Establishing a Display QA Program. Mike Silosky, M.S. 8/3/2017

3/2/2016. Medical Display Performance and Evaluation. Objectives. Outline

MX215. Your advantages. 2MP Medical-Display

Medical Monitor Solutions

MX215. Your advantages. 2MP Medical-Display

Global Trade Medical Supplies

MX194. Your advantages. 1MP Medical-Display

Monitor QA Management i model

RX850. Your advantages. 8MP Medical-Display

White Paper. Uniform Luminance Technology. What s inside? What is non-uniformity and noise in LCDs? Why is it a problem? How is it solved?

Nio. Industry-standard diagnostic display systems

Nio. Industry-standard diagnostic display systems

+ Human method is pattern recognition based upon multiple exposure to known samples.

Display Quality Assurance: Recommendations from AAPM TG270 for Tests, Tools, Patterns, and Performance Criteria

Display Quality Assurance: Recommendations from AAPM TG270 for Tests, Tools, Patterns, and Performance Criteria

Guidelines for Assuring Softcopy Image Quality

This talk covers currently available display technology.

RX350. Your advantages. 3MP Medical-Display

P XGA TFT Monitor. User s Manual

Retired. 1. Power On/Off Button 4.Minus (-) Button 2. Power Indicator LED 5.Menu Select Button 3. Plus (+) Button 6.

Coronis Uniti (MDMC-12133) 12MP diagnostic display system for PACS and breast imaging

White Paper. Missing Pixels in Medical Grayscale Flat Panel Displays. Geert Carrein Director Product Management. W hat s inside?

Coronis 5MP Mammo. The standard of care for digital mammography

CAPTURE CAPTURE. VERSiON 1.2. Specialists in Medical. Digital Imaging Solutions

MDSC-2232 Full HD 32-inch surgical display

Selection and Quality Assurance of Monitors for Mammogram Diagnosis

1 Your computer screen

Lecture 14: Computer Peripherals

Medical Monitor Solutions

Chapter Medical Table of Contents

Diamond View DV180. Colour LCD Monitor (45.9 cm) LCD Panel Size. User s Manual

CORONIS. Medical grayscale flat panel display system

LCD1510+, MultiSync LCD1810. XtraView technology for accurate viewing from any angle. Flat screen liquid crystal display with a super-slim profile

Medical Monitor Solutions

White Paper. In Plane Switching Pro technology for medical imaging. Geert Carrein Director Product Management. What s inside?

LA1500R USER S GUIDE.

Medical Monitor Solutions

Advantech Proudly Presents Professional Monitor Series

BeneHeart R3. Electrocardiograph

Please read PRECAUTIONS, this User s Manual, and Setup Manual (separate volume) carefully to familiarize yourself with safe and effective usage.

Durable and Reliable Design for 24/7 Use. Overview. Narrow Bezel Optimized for Video Walls. 42 Narrow bezel commercial LED display

Quality Assurance (QA) Guidelines for Medical Imaging Display Systems

CARESTREAM DIRECTVIEW Elite CR System

CARESTREAM DIRECTVIEW Elite CR System

Christie MicroTiles. Technical Frequently Asked Questions (FAQs)

General Items: Reading Materials: Miscellaneous: Lecture 8 / Chapter 6 COSC1300/ITSC 1401/BCIS /19/2004. Tests? Questions? Anything?

PMCL300 Series TFT LCD Monitor 17-INCH, 19-INCH, OR 19-INCH WIDE, WITH MULTIMODE FUNCTIONALITY

32" Full HD Large Format LED Display, Ideal for Commercial Use CDE3200-L

2013, 2014 Hewlett-Packard Development Company, L.P.

1. Opens, selects or exits the OSD menu. 4. Selects video input, DVI-I#1 or DVI-I#2.

55 (54.6 viewable) Full HD Direct-lit LED Commercial Display CDE5502 Direct LED backlight technology for superior color performance

TFT LCD MONITOR PRODUCT SPECIFICATION MODEL: KTS270DPE02 ISSUE DATE: Prepared by KORTEK R&D CENTER KORTEK CORPORATION

HP Engage One 10.1 Display (Black) HP Engage One 10.1 Display (White)

Acer AL1917 Service Guide. Service guide files and updates are available on the CSD web: for more information, Please refer to http: csd.acer.com.

V9A01 Solution Specification V0.1

F250. Advanced algorithm enables ultra high speed and maximum flexibility. High-performance Vision Sensor. Features

Mammo Tomosynthesis 5MP

Fine Pitch LED Display System

These are used for producing a narrow and sharply focus beam of electrons.

CLINICALLY PROVEN DESIGNS

Reading. Display Devices. Light Gathering. The human retina

DirectView Elite CR System. Improve workflow, productivity, and patient throughput.

-Technical Specifications-

QuickSpecs. Models RB146AA#ABA Standard Configuration RB146AT#ABA Promotional Part Number (SmartBuy) HP L5006tm 15-inch LCD Touchscreen Monitor

HP EliteDisplay LED Backlit Monitors. User Guide

A Perfect Solution for Digital Signage in Commercial Areas CDE4600-L

Compact Size Perfect for rack mount router and other applications with space limitations.

Types of CRT Display Devices. DVST-Direct View Storage Tube

28 To 48 Rugged UltraWide Industrial Displays

Hospital Wide. Healthcare Display Solutions DICOM Displays, Large Screen Displays and Projectors

DRAFT. Proposal to modify International Standard IEC

PT-LW330 S P E C F I L E. LCD Projectors. As of April Specifications and appearance are subject to change without notice. 1/9.

Hitachi Europe Ltd. ISSUE : app084/1.0 APPLICATION NOTE DATE : 28/04/99

SmartSwitch TM. Wide View LCD 36 x 24 Pushbutton DISTINCTIVE CHARACTERISTICS PART NUMBER & DESCRIPTION

Coronis Fusion 6MP DL. 6 MegaPixel wide-screen diagnostic color display system

Digital High Resolution Display Technology. A New Way of Seeing Things.

2.2. VIDEO DISPLAY DEVICES

QuickSpecs. HP N220h 21.5-inch Monitor. Overview. 1. Menu/OK 3. Plus ( + ) 5. Power 2. Minus ( - ) 4. Exit/Back

Product General Specification

PT-LB300 S P E C F I L E. LCD Projectors. As of April Specifications and appearance are subject to change without notice. 1/8.

QuickSpecs. HP DreamColor Z27x G2 Studio Display. Technical Specifications. Front. HP DreamColor Z27x G2 Studio Display

Tech Paper. HMI Display Readability During Sinusoidal Vibration

NEC MultiSync PA Series with SpectraView II

TFT COLOR LCD MODULE NL6448AC30-12

Specifications SMART Board 6075 interactive flat panel with iq Model SPNL-6275

Using Low-Cost Plasma Displays As Reference Monitors. Peter Putman, CTS, ISF President, ROAM Consulting LLC Editor/Publisher, HDTVexpert.

Television Analyser - TVA97

LMT X1 - a 12.1" VHB

PRODUCT NO.: PT-L735 PRODUCT NAME: Ultra Portable LCD Projector

How to Match the Color Brightness of Automotive TFT-LCD Panels

Flat Display Systems for Medical Imaging. model

H SERIES INDOOR FINE PIXEL PITCH LED DISPLAY. Yaham Optoelectronics Co., Ltd

7.5ft Outdoor Double-Sided Interactive LCD Sign, Transit Center Kiosk Model: NISDK-550PCT, 55 Double-Sided

The measurements are stored in non-volatile memory, which retains data even when the power down.

Click to Print This Page. 9,500 ANSI / 10,500 ISO Lumens Contrast Ratio: 10,000:1 (Dynamic Black) 1,000:1 native Part Number:

F l a t D i sp l ay Sys t e m s for Me d ica l Im a g i n g. model

LCD Display Wall Narrow Bezel Series

FFDM Quality Control in Canada - a Vendor Neutral Approach

PMCL500 Series Narrow Bezel LCD Monitors 42-, 46-, AND 55-INCH HIGH RESOLUTION MONITORS

Transcription:

Evaluation report Eizo RadiForce G33-N 3MP greyscale flat panel liquid crystal display (LCD) CEP 07003 August 2007

Contents 2 Executive summary... 3 Introduction... 5 Product description... 6 Evaluation method... 8 Technical performance... 9 Purchasing... 17 Acknowledgements... 18 References... 19 Author and report information... 20

Executive summary 3 The product The Eizo RadiForce G33-N 3MP greyscale flat panel liquid crystal display (LCD) is one of a number of competing electronic display products on the UK market. It utilises a thin film transistor (TFT) active matrix liquid crystal display (AMLCD) with a three megapixel (3 MP) display matrix. Field of use With continuing investment in digital technologies for diagnostic imaging in the NHS, it has become increasingly common for clinicians to view medical images on electronic display devices, rather than X-ray film. This has many advantages, not least the cost savings in not printing film and managing its storage. Advances in the technology employed in electronic display devices for medical applications have resulted in cathode ray tube (CRT) displays nearing obsolescence, with liquid crystal displays (LCD) now the market leaders. Display devices purchased for use in medical imaging should be fit for the specific intended purpose, whether this is for review of images, or for primary diagnosis. Further, the minimum specification of display required depends on the source modality of the images to be viewed. Primary diagnosis of mammography images will require higher specification displays than will primary diagnosis of magnetic resonance imaging (MRI) images. The Eizo RadiForce G33-N is intended for use in diagnostic medical imaging for primary reporting of monochrome images. Applications of the product include: computed radiography (CR), direct digital radiography (DDR), computed tomography (CT), MRI and angiography. It is not intended for the diagnostic interpretation of mammography images, for which higher resolution displays are recommended. National guidance Guidelines on the specification of display devices for particular uses have been produced by the Royal College of Radiologists (RCR) UK PACS & Teleradiology Group and these should be referred to when making purchasing decisions. Purchasing guidance NHS Connecting for Health (CfH) is tasked with delivering the National Programme for IT (NPfIT), which is intended to create a coherent IT infrastructure across the NHS. This infrastructure will include picture archiving and communications systems (PACS), key components of which are display devices. The NPfIT is being delivered throughout England within five regions, each working with a single local service provider (LSP). Display devices should be available to purchase through the NPfIT via the LSP s product catalogue. However, not all commercially available display devices will necessarily be included in each LSP s catalogue, since each LSP tends to conduct their own checks and balances, as such items directly affect their service level agreements (SLAs). It is not compulsory to purchase products from the LSP catalogue and purchasers are free to source display devices in whichever way is most convenient and economic for them.

Executive summary 4 As of 1 st April 2007, accountability for the delivery of the NPfIT has been transferred to Strategic Health Authorities (SHAs) as part of the NPfIT Local Ownership Programme (NLOP). Therefore, in the first instance, you may have to go to your local SHA for assistance and advice with the procurement of your display devices as part of the NPfIT. Evaluation method The evaluation methods were based on the American Association of Physicists in Medicine (AAPM) On-line Report No. 03 Assessment of Display Performance for Medical Imaging Systems (April 2005). Measurements were made of: geometric distortion, luminance response, luminance stability, luminance uniformity, variation of luminance with viewing angle, contrast resolution and spatial resolution. The display was evaluated in combination with a Matrox MED3MP-DVI 3MP graphics card and BIOS 1.1 17 EC display driver package, recommended by Eizo as being appropriate for use with this type of display and together with the display, loaned by them for the period of the evaluation. The display was evaluated in a darkroom. No attempt was made to distinguish between the performance of the display and that of the supplied graphics controller card. CEP verdict The measured performance of the Eizo RadiForce G33-N indicates that it is suitable for use in diagnostic medical imaging for primary reporting of monochrome images. The display has a high contrast ratio, little luminance variation across the display and good luminance stability over time. As expected of liquid crystal display (LCD) technology, there is no geometric distortion in the displayed image.

Introduction 5 With the advent of digital imaging systems, it has become increasingly common to view medical images on electronic display devices rather than film. This has many advantages, not least the cost-saving involved in not printing film and managing its storage. Liquid crystal displays (LCDs) offer a number of advantages over the cathode ray tube (CRT) displays that they increasingly replace in medical imaging. LCDs are not affected by screen flicker, geometric distortion or focus problems, all of which commonly afflict CRTs. Furthermore, they take up much less space and consume less power. They do however have some minor disadvantages compared with CRTs, including a relatively low maximum viewing angle (affecting the number of people who can reliably view an image on the same screen), lower refresh rate (which can cause blur on fast-moving images), and lower contrast. However, these shortcomings are becoming less significant as LCD technology matures. Guidelines on the specification of display devices for particular uses have been produced by the Royal College of Radiologists (RCR) UK PACS & Teleradiology Group and these should be referred to when making purchasing decisions. Displays used for primary reporting should be calibrated to the digital imaging and communications in medicine (DICOM) Part 14 Greyscale Standard Display Function. Many current LCD devices on the UK market allow for automated calibration to the DICOM curve and some devices can send results of their self-calibration over a network for remote quality assurance. The Eizo RadiForce G33-N, as supplied, did not have any self-calibration facilities. The Eizo RadiForce G33 (not tested here) does include self-calibration facilities. The evaluation methods were based on the American Association of Physicists in Medicine (AAPM) On-line Report No. 03 Assessment of Display Performance for Medical Imaging Systems (April 2005). The display was evaluated in a darkroom. No attempt was made to distinguish between the performance of the display and that of the supplied graphics controller card.

Product description 6 The Eizo RadiForce G33-N greyscale flat-panel liquid crystal display utilises a thin film transistor (TFT) active matrix liquid crystal display (AMLCD) with a three megapixel (3 MP) display matrix and uses the Quad extended Graphics Array (QXGA) display standard. The stand allows adjustment of height and tilt and rotation between portrait and landscape orientation. Display supplied Eizo Nanao RadiForce G33-N 3 megapixel display, serial number 10012095. Graphics card supplied Matrox MED3MP-DVI, serial number BB96759. Display driver package Graphics BIOS 1.1 17 EC. The Matrox PowerDesk software provides features additional to the standard Microsoft Windows display control, including the ability to adjust screen orientation, the option to use one display or multiple displays and choice of display mode (stretched or independent, i.e. whether an image can spread over multiple screens or fill one screen at most). Figure 1. Eizo RadiForce G33-N 3 MP display

Product description 7 Product specifications Table 1. Eizo G33-N specifications (from manufacturer's literature) Dimensions Screen size Height (with stand) Width Depth Viewable image Weight Power supply Power consumption Minimum Maximum Power saving mode Operating conditions Temperature Humidity Pressure Resolution Native resolution Pixel size Pixel spacing Response time Scan frequency Horizontal scan Vertical scan Dot clock 318 x 423.9 mm (W x H) (portrait mode) 515.5-597.5 mm (portrait mode) 368 mm 209 mm 529 mm (20.8 ) 10.0 kg 120 / 200-240 VAC +/- 10% 50 / 60 Hz 1.0-0.8 / 0.5-0.4 A 90 W 100 W (with USB hub) < 3 W (USB hub not connected) 0-35 C 30-80 % relative, non-condensing 70-106 kpa 1536 x 2048 pixels Not quoted 0.207 x 0.207 mm 50 ms 31-100 khz 48-71.5 Hz 165 MHz (max)

Evaluation method 8 The evaluation methods were based on the American Association of Physicists in Medicine (AAPM) On-line Report No. 03 Assessment of Display Performance for Medical Imaging Systems (April 2005). The display was evaluated in a darkroom. No attempt was made to distinguish between the performance of the display and that of the supplied graphics controller card. The device was tested for geometric distortion, luminance response, luminance stability and uniformity, variation of luminance with viewing angle, contrast resolution and spatial resolution. The supplied device was tested attached to a Dell Dimension 8100 PC (Pentium 4, 1.5 GHz; 256 MB RAM; 74.5 GB HDD) owned by PACSnet, running Windows 2000 Professional v5.00.2195 SP 3.

Technical performance 9 Geometric distortion In an LCD, pixel positions are fixed and are determined by the position of the individual cells. Therefore, in a properly designed and manufactured display, there should be no geometric distortion apparent. If images are rescaled for display, this may introduce minor distortion. To establish that there was no geometric distortion, a test pattern of squares (AAPM TG18- QC) was displayed at 1:1 display pixel: image pixel and the size of the squares in the test image measured. One square: Two squares: Four squares: 21 mm x 21 mm 42 mm x 42 mm 84 mm x 84 mm There was no measurable geometric distortion in the displayed test image. No distortion was seen in any other test images used during the evaluation. Display reflections No formal measurement of the reflectance of the display screen was performed. Subjectively, the display reflectance was similar to that of other displays in its class. Luminance response No calibration software was supplied with the display. However, a number of options were available via the display s firmware. These were a choice of curve and a choice of maximum luminance. Curve options were: 1-DICOM 2-NATIVE 3-CAL Maximum luminance could be adjusted in a range from 0 % to 120 % in 1 % steps. The device was supplied set to the DICOM curve and with a luminance setting of 72%. These settings gave appropriate values for the expected use of this display (i.e. a maximum luminance of 450 cd m-2) and were therefore used for the majority of the tests.

Technical performance 10 To check the response of the display to changes in input values, the display s luminance was measured across the available range of input values. Results of the measurements are shown in figure 2. Figure 2. Luminance response to varying input value 500 450 400 350 Luminance (cd/sq m) 300 250 200 150 100 50 0 0 50 100 150 200 250 300 Input value (8-bit) To check conformance with the DICOM Greyscale Display Function (GSDF), a number of points were measured for contrast difference and compared to the contrast difference specified by the DICOM GSDF. The results of this test are shown in figure 3, which also shows the target values, together with 10% and 20% variation curves. The graph shows that the measured values are well within the 10% variation curve.

Technical performance 11 Figure 3. DICOM conformance 0.03 0.025 0.02 JND difference 0.015 0.01 Measured +20% -20% +10% -10% D(GSDF) 0.005 0-100 200 300 400 500 600 700 800 JND Luminance uniformity An LCD panel may not be uniformly illuminated by the backlight, resulting in uneven intensity and shading over the screen. The screen glass itself may also contribute to non-uniformity in the displayed image. Luminance uniformity was measured by displaying a plain white image over the entire screen and making measurements of the luminance at 9 points over the screen, in a 3 x 3 matrix. Mean results of these measurements (in cd m -2 ) using the Eizo luminance probe were: 419 426 425 429 432 442 407 417 418 These results are illustrated in figure 4.

Technical performance 12 Figure 4. Luminance uniformity measured using the Eizo luminance probe 200 150 Luminance (cd/sq m) 100 50 T 0 L C B C Vertical Position Horizontal Position R The AAPM TG18 Display Assessment protocol states that the maximum luminance variation should be less than 30%, this luminance variation being calculated by the formula: L var = 200 * (L max L min )/(L max + L min ). In the above case, the maximum variation is L var = 200 * (442 407)/(442+407) = 4.1 % The luminance variation over the display screen falls below the upper limit of 30% recommended by the AAPM and is therefore acceptable.

Technical performance 13 Luminance stability The stability of the luminance of the display was measured, both short-term luminance changes following switching on the device and longer-term drift over several hours. Luminance changes following switching on the device To measure the luminance changes following switching on the device (the warm-up time), the system was set up to display a white square at peak luminance at the centre of the display. After leaving the display switched off overnight, the display was then powered on and luminance readings taken every few seconds until the display luminance had stabilised. Results are shown in figure 5. This shows the luminance of the device for the first 300 seconds (5 minutes) following switch-on. Figure 5. Warm-up time 600 500 Luminance (cd/sq m) 400 300 200 100 0 0 50 100 150 200 250 300 350 Time (s) The display reached its target luminance and remained there after less than one minute. In normal use, this would be well within the time required to start up the base PC and for the user to log in. Luminance drift To measure any drift in luminance, the display was switched on and a test image displayed in the same manner as for the warm-up measurements above. Measurements were taken for a significantly longer period of 480 minutes (8 hours, approximately one working day) and the results are shown in figure 6.

Technical performance 14 Figure 6. Luminance stability 500 450 400 350 Luminance (cd/sq m) 300 250 200 150 100 50 0 0 60 120 180 240 300 360 420 480 Time (minutes) There was no significant drift in the peak luminance value over the measurement period. Contrast ratio The contrast ratio was measured using the method described by the American National Standards Institute (ANSI). This requires the display of a 4 x 4 checkerboard pattern, with eight areas at peak white and eight areas at black. The luminance of each of these areas is measured and the luminance ratio is calculated by dividing the mean luminance of the white areas by the mean luminance of the black areas. The display is masked, apart from an aperture for the area being measured, during this test in order to reduce the effect of scattered light on the measurements. Results: Mean white luminance: 423.4 cd m -2 Mean black luminance: 0.854 cd m -2 Calculated contrast ratio: 496 The calculated value for the contrast ratio exceeds the minimum value of 400 recommended by the AAPM and is therefore acceptable.

Technical performance 15 Spatial resolution The resolution of an LCD display is fixed at the time of manufacture. At the native resolution of the panel, the resolution is exact. Should it be required to display images at different resolutions, the image data will need to be rescaled before display. This rescaling can introduce distortion in the displayed image. The display resolution was evaluated using the AAPM TG18 test image. As expected, the limiting factor for the resolution was the size of the pixels. With test images displayed at 1:1 image pixel: display pixel (i.e. at 1536 x 2048 pixels image size) there was no loss of spatial resolution. Display noise Display noise was not measured but appeared to be minimal. Angular dependency of luminance The apparent brightness of displays can vary considerably with viewing angle. This is especially true for LCD flat-panel displays. This variation of luminance can affect not only the perceived brightness of the displayed image, but also its contrast. The same image viewed from different angles can convey quite different information. LCD devices should be viewed from directly in front of the display. Figure 7 shows the variation of peak luminance as viewing angle varies, with the display in the portrait orientation. There is a decline in luminance as viewing angle increases. Figure 7. Variation of luminance with viewing angle 500 450 400 Luminance (cd/sq m) 350 300 250 200 150 vertical horizontal 100 50 0-90 -60-30 0 30 60 90 Viewing angle (degrees)

Technical performance 16 Figure 8 shows the variation of contrast ratio as viewing angle varies, with the display in the portrait orientation. Again, it can be seen that there is a decline in luminance as viewing angle increases. Figure 8. Variation of contrast ratio with viewing angle 600 500 Luminance (cd/sq m) 400 300 200 vertical horizontal 100 0-90 -60-30 0 30 60 90 Viewing angle (degrees) The decline in luminance with viewing angle in both portrait and landscape orientation is due to the structure of the display screen, in particular the properties of the polarisers used on the liquid crystal cells and the length of the light path taken through the cells, both of which vary with the angle of incidence of the light passing through. Slightly different effects are seen as screen orientation changes due to the shape of the liquid crystal cells, which are not square in this device. This is typical for this type of display. Electronic cross-talk Electronic cross-talk is the effect of signal on one channel affecting the signal on a nearby channel. In a display, this can mean the luminance of one pixel being affected by signal intended for another pixel. There was no effect from cross-talk apparent on a visual inspection of the displayed image. Pixel defects LCDs can have stuck pixels, i.e. pixels which are permanently on or off. Some pixels may be improperly connected to adjoining pixels, rows or columns. There were no pixel defects apparent on visual inspection of the display.

Purchasing 17 NHS Connecting for Health and the National Programme for IT (NPfIT) NHS Connecting for Health (CfH) is tasked with delivering the National Programme for IT (NPfIT), which is intended to create a coherent IT infrastructure across the NHS. This infrastructure will include picture archiving and communications systems (PACS), key components of which are display devices. The NPfIT is being delivered throughout England within five regions, each working with a single local service provider (LSP). Display devices should be available to purchase through the NPfIT via the LSP s product catalogue. However, not all commercially available display devices will necessarily be included in each LSP s catalogue, since each LSP tends to conduct their own checks and balances, as such items directly affect their service level agreements (SLAs). It is not compulsory to purchase products from the LSP catalogue and purchasers are free to source display devices in whichever way is most convenient and economic for them. Accountability for the delivery of the NPfIT was transferred on 1 April 2007 to the Strategic Health Authorities (SHAs) as part of the NPfIT Local Ownership Programme (NLOP). CfH is working closely with the 10 SHAs in ensuring activities with LSPs continue effectively have established three geographic areas linking the LSPs and SHAs together; North, Midlands and East Programme for IT (NMEPfIT) with six SHAs and Computer Sciences Corporation (CSC) as the LSP; London Programme for IT (LPfIT) with one SHA and BT as the LSP Southern Programme for IT (SPfIT) with three SHAs and Fujitsu as the LSP. Prospective purchasers of display devices and related products might wish to seek guidance from their local SHAs on their available options.

Acknowledgements 18 PACSnet would like to thank the Eizo Nanao Corporation and their UK distributors, Avnet Applied Computing for their assistance: Avnet Applied Computing Ltd. Genesis Business Park Albert Drive Sheerwater Woking Surrey GU21 5RW Tel: 01483 719500 Fax: 01483 719560 www.eizo.co.uk

References 19 Assessment of Display Performance for Medical Imaging Systems, American Association of Physicists in Medicine (AAPM) On-line Report No. 03, April 2005 Digital Imaging and Communications in Medicine (DICOM) Part 14: Greyscale Standard Display Function, National Electrical Manufacturers Association, 2003 Data Projection Equipment and Large Screen Data Displays -- Test Methods and Performance Characteristics, American National Standards Institute (ANSI) IT7.215-1992, 1992. Electronic Display Measurement: Concepts, techniques and instrumentation. Keller, Peter A., John Wiley and Sons, New York, 1997 Royal College of Radiologists (RCR) PACS & Teleradiology Group. Guidelines on Diagnostic Display Devices. Version 2.0 (Approved), 1 st June 2006: http://www.pacsgroup.org.uk/forum/messages/419/17843.html?0 (web URL accurate at time of publication)

Author and report information 20 Eizo RadiForce G-33N 3MP greyscale flat panel liquid crystal display (LCD) PACSnet Bence Jones Offices St George s Hospital Blackshaw Road Tooting London SW17 0QT Tel: 020 8725 3315 Email: query@pacsnet.org.uk www.pacsnet.org.uk About CEP The Centre for Evidence-based Purchasing (CEP) is part of the Policy and Innovation Directorate of the NHS Purchasing and Supply Agency. We underpin purchasing decisions by providing objective evidence to support the uptake of useful, safe and innovative products and related procedures in health and social care. We are here to help you make informed purchasing decisions by gathering evidence globally to support the use of innovative technologies, assess value and cost effectiveness of products, and develop nationally agreed protocols. Sign up to our email alert service All our publications since 2002 are available in full colour to download from our website. To sign up to our email alert service and receive new publications straight to your mailbox contact: Centre for Evidence-based Purchasing Room 152C Skipton House 80 London Road SE1 6HL Tel: 020 7972 6080 Fax: 020 7975 5795 Email: cep@pasa.nhs.uk www.pasa.nhs.uk/cep Crown Copyright 2007