Comparison of SONY ILX511B CCD and Hamamatsu S10420 BT-CCD for VIS Spectroscopy

Similar documents
Compact multichannel MEMS based spectrometer for FBG sensing

Measurement of Microdisplays at NPL

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

111 Highland Drive Putnam, CT USA PHONE (860) FAX (860) SM32Pro SDK

Mini-spectrometers. TM series. High sensitivity type (integrated with backthinned type CCD image sensor) C10082CA/C10083CA series

Mini-spectrometers. RC series. Compact and low cost C11009MA, C11010MA: for installation into measurement equipment.

TM series. Trigger function included

FT series. Spectral response range 790 to 920 nm Spectral resolution Typ. 0.4 nm

MS series. Parameter Min. Typ. Max. Unit Driving voltage V Power consumption mw Video rate khz

Appendix B: An Automated Spectroscope

LCOS-SLM (Liquid Crystal on Silicon - Spatial Light Modulator)

Mini-spectrometers. TG series. High sensitivity type (integrated with backthinned type CCD image sensor) Optical characteristics

Advanced Sensor Technologies

Mini-spectrometer. TG series. Enhanced near infrared sensitivity type. C9405CB. High sensitivity in near infrared region

Wavelength selective electro-optic flip-flop

Mini-spectrometer. TG series. For near IR, integrating optical system, image sensor and circuit. C9406GC, C9913GC, C9914GB

ModBox-1310nm-1550nm-NRZ 1310nm & 1550 nm, 28 Gb/s, 44 Gb/s Reference Transmitters

TR-SES-200 High Resolution VUV/UV Transmission and Reflection Evaluation System Manual

Applications. l Image input devices l Optical sensing devices

Reduction of Device Damage During Dry Etching of Advanced MMIC Devices Using Optical Emission Spectroscopy

TM series TM-UV/VIS-MOS

Module 11 : Link Design

CAEN Tools for Discovery

Sodern recent development in the design and verification of the passive polarization scramblers for space applications

specification display applications cronus spectrometer & colorimeter

Agilent 86120B, 86120C, 86122A Multi-Wavelength Meters Technical Specifications

Practicality and Performance

CCD 143A 2048-Element High Speed Linear Image Sensor

ModBox-850nm-NRZ-series

Selected Problems of Display and Projection Color Measurement

GMOS CCD Upgrade Options S. Kleinman, J. Jensen 26Sep08

ModBox-1310nm-1550nm-28Gbaud-PAM nm & 1550 nm, 28 Gbaud PAM-4 Reference Transmitter

ISOMET. Compensation look-up-table (LUT) and Scan Uniformity

Mini-spectrometer. TM series. Trigger function and a high sensitivity CMOS image sensor included. C11697MB. Optical characteristics

Mini-spectrometer. TG series. Enhanced near infrared sensitivity type. C9405CB. High sensitivity in near infrared region

Photodetector Testing Facilities at Nevis Labs & Barnard College. Reshmi Mukherjee Barnard College, Columbia University

CCD Element Linear Image Sensor CCD Element Line Scan Image Sensor

Agilent Agilent 86120B, 86120C, 86122A Multi-Wavelength Meters Data Sheet

Mini-spectrometer. TF series. Compact and thin, built-in high-sensitivity CMOS image sensor for Raman spectroscopy C14214MA. Applications.

Study of Timing and Efficiency Properties of Multi-Anode Photomultipliers

Beam test of the QMB6 calibration board and HBU0 prototype

Agilent 81600B Tunable Laser Source Family

Standard Operating Procedure of nanoir2-s

ABSTRACT 1. INTRODUCTION 2. EXPERIMENTS. Corresponding author: +1 (518) ;

mini-spectrometer TG series Long-wavelength type (to 2.55 μm) nearinfrared C11118GA Optical characteristics (Ta=25 C)

USA Price List. January or 8-channel LIBSCAN system (100 mj, 1064 nm laser) Alternative LIBSCAN configurations

Abstract. Keywords INTRODUCTION. Electron beam has been increasingly used for defect inspection in IC chip

LaserPXIe Series. Tunable Laser Source PRELIMINARY SPEC SHEET

DPD80 Visible Datasheet

User Manual OVP Raman

Screen investigations for low energetic electron beams at PITZ

Experiment 9 Analog/Digital Conversion

Fluorescence Spectrophotometer

Agilent 81600B Tunable Laser Source Family Technical Specifications August New model: nm, low SSE output!

Spatial Response of Photon Detectors used in the Focusing DIRC prototype

CESR BPM System Calibration

DPD80 Infrared Datasheet

ModBox-CBand-NRZ series C-Band, 28 Gb/s, 44 Gb/s, 50 Gb/s Reference Transmitters

Silicon PhotoMultiplier Kits

Transmissive XBPM developments at PSF/BESSY. Martin R. Fuchs

CCD220 Back Illuminated L3Vision Sensor Electron Multiplying Adaptive Optics CCD

Teaching Plasma Nanotechnologies Based on Remote Access

IMAGING GROUP. * With dual port readout at 16MHz/port Detector shown with a C-mount nose and lens, sold separately

CBF500 High resolution Streak camera

Non-Invasive Energy Spread Monitoring for the JLAB Experimental Program via Synchrotron Light Interferometers

Power saving in LCD panels

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

3 EXPERIMENTAL INVESTIGATIONS Caroline Robson. 3.1 Aims and Objectives. 3.2 Experimental Method Set Up of the Test Stand

TF series. Spectral response range 340 to 830 nm Spectral resolution Typ. 2.3 nm

Designed for back-thinned CCD area image sensor

DESIGN OF VISIBLE LIGHT COMMUNICATION SYSTEM

??? Introduction. Learning Objectives. On completion of this chapter you will be able to:

Operating Instructions for PMT Tube P3

SPATIAL LIGHT MODULATORS

Monitor QA Management i model

Agilent 86120B, 86120C, 86122B Multi-Wavelength Meters. Data Sheet

Very High QE bialkali PMTs

New Medical Light Source using NTT s Communication Laser Technology

Mini-spectrometers. RC series. Compact and low cost C11009MA, C11010MA: for installation into measurement equipment.

OSA20 KEY FEATURES SPEC SHEET OPTICAL SPECTRUM ANALYZER

Studies of large dynamic range silicon photomultipliers for the CMS HCAL upgrade

Novel persistent phosphors of lanthanide-chromium co-doped. yttrium aluminum gallium garnet: design concept with vacuum

Online Detection and Characterization of MPB Wood furnish to Optimize OSB Mill Processing Efficiency

DATASHEET. Intematix ChromaLit. Remote Phosphor Light Source. Features & Benefits. Applications and Uses. Unprecedented design freedom for solid state

Cree Engineering Services Testing Laboratory (CESTL) Photometric Testing and Evaluation Report

o-microgigacn Data Sheet Revision Channel Optical Transceiver Module Part Number: Module: FPD-010R008-0E Patch Cord: FOC-CC****

Photon counting module

Development of an Abort Gap Monitor for High-Energy Proton Rings *

Projection Displays Second Edition

High Efficiency White OLEDs for Lighting

2.5 Gb/s Transponder with Mux/Demux (1310 and 1550 nm) 54TR Series

Mahdad Manavi LOTS Technology, Inc.

HP 70950B OPTICAL SPECTRUM ANALYZER

25W 9xxnm Uncooled Multimode Laser Diode Module

OPTICAL MEASURING INSTRUMENTS. MS9710C 600 to 1750 nm OPTICAL SPECTRUM ANALYZER GPIB. High Performance for DWDM Optical Communications

Modulation transfer function of a liquid crystal spatial light modulator

Scope: Using Wave Division Multiplexing (WDM) for the Protection Data Interface (PDI) of the 7SD5 / 7SD61.

APPLICATION NOTE. Fiber Alignment Now Achievable with Commercial Software

CMOS linear image sensors

NMOS linear image sensor

Transcription:

Comparison of SONY ILX511B CCD and Hamamatsu S10420 BT-CCD for VIS Spectroscopy Technical Note Thomas Rasmussen VP Business Development, Sales, and Marketing Publication Version: March 16 th, 2013-1 -

Introduction One of the key elements in a diffractive spectrometer is the line array detector used to measure intensity versus wavelength. Two of the more popular detector arrays are the low cost SONY ILX511B front illuminated CCD and the higher cost Hamamatsu S10420 Back-Thinned CCD (BT-CCD). It is generally a difficult task to make a direct comparison of detector arrays based purely on the data sheet information. Therefore, Ibsen Photonics have done comparative measurements on these two detector arrays when used inside an actual spectrometer unit, and in this Technical Note we show results for sensitivity and Signal-to-Noise Ratio (SNR) for the two detectors. See Figure 1 for a schematic diagram of the diode array spectrometer used for the tests. It is important to note that Ibsen Photonics has no preference for any of the two detectors and we sell spectrometers using both. The purpose of this technical note is merely to provide some objective information that we think is useful for our customers in deciding which detector type to choose. Also, while we focus on sensitivity and SNR in this technical note, it is important to note that there are other parameters (like linearity, built-in dark pixels, cost, etc.) that may also influence the choice of detector for a given application. Figure 1: Example of diode array diffractive spectrometer. - 2 -

Sensitivity The sensitivity of a spectrometer is determined by many parameters (diffraction efficiency of the grating, reflectivity of mirror coatings, size of slits and apertures, detector sensitivity etc.). In order to compare the sensitivity of the two detectors alone we measured the spectral response of the same spectrometer equipped with either the SONY ILX511B or the Hamamatsu S10420-1004-01 detector. The ILX511B was with-out a window and was coated with a Lumigen coating while the S10420-1004-01 detector had a standard quartz window. The same control electronics were used for both spectrometers. Integration time was 2 msec and the input light source was a calibrated Tungsten-halogen lamp. The gain in the electronics was adjusted to ensure that the dynamic range of the detector matched the dynamic range of the 16 bit A/D converter. The results are shown on Figure 2 and clearly show that the ILX511B detector provides higher sensitivity than the S10420 in the 480 730 nm. Below 480 nm and above 730 nm the S10420 has higher sensitivity than the ILX511B. Figure 3 shows the sensitivity ratio as a function of wavelength for the two detectors. As an example, if you want to adjust your integration time to obtain a certain signal level (in counts), the integration time needs to be up to 2x longer for the S10420 than for the ILX511B. Alternatively, you must use a lamp with twice the intensity for your spectroscopy with the S10420 than with the ILX511B. However, sensitivity is only telling us how much signal level we can get. We also need to consider the noise level of the two detectors. Figure 2: Measured sensitivity of a VIS spectrometer using either the ILX511B or the S10420 detector. - 3 -

Figure 3: Sensitivity ratio of S10420 over ILX511B. Signal-to-Noise Ratio (SNR) The RMS noise level of a detector depends on the optical signal level on the detector so, generally speaking, there is not a single number characterizing SNR for a spectrometer. The measured RMS noise as a function of signal level is shown on Figure 4. As can be seen the RMS noise is more than 2x lower for the S10420 than for the ILX511B. Spectrometers are normally characterized by the SNR at full signal (65,535 counts for a 16 bit A/D converter) Dynamic range (Full signal divided by Dark noise or noise at 0 counts). Table 1 lists these parameters for the two detectors and Figure 5 shows the SNR as a function of signal level. SNR Dynamic range (at full signal) ILX511B 380 1,000 S10420 820 5,000 Table 1: SNR and Dynamic range comparison - 4 -

Figure 4: Noise as a function of signal level (16 bit A/D converter). Figure 5: SNR as a function of signal level (16 bit A/D converter). Using Figure 3 and Figure 5 it is possible to estimate the SNR difference at any signal level, wavelength and integration time (as long as the integration time is kept below ~500 msec). As an example, let us assume we adjusted the integration time of a spectrometer with the ILX511B to 50,000 counts at the peak wavelength (~550 nm). This will provide a SNR of 320 read from Figure 5. If we use the same spectrometer with the S10420 and keep the same integration time, we will get a signal level of 25,000 counts because of the lower sensitivity of the S10420 and therefore the SNR obtained at 550 nm will now be 500. - 5 -

As this example shows, the SNR is still better for the S10420 than for the ILX511B when we take the difference in sensitivity into account. Averaging considerations One way to improve the SNR is to perform averaging over a number of consecutive measurements. By doing so, the noise level is improved by a factor of the Square Root of N, where N is the number of measurements. So, since the ILX511B can reach the same signal level as the S10420 in half the time, we can use this to perform twice as many measurements with the ILX511B compared to the S10420. As an example, let us assume we adjust the integration time of a spectrometer using the S10420 to 50,000 counts at 550 nm. Without averaging, this provides a SNR of 710. For a spectrometer with the ILX511B we can obtain the 50,000 counts in half the integration time. So, we will do two measurements and calculate the average of the two measurements. This will give us a SNR using the ILX511B of sqrt(2)*320 = 450. This example is just another way of confirming that the S10420 provides a better SNR than the ILX511B. Cost considerations As written in the introduction the ILX511B detector array is lower cost than the S10420. However, for many applications the ILX511B has to have its cover window removed and AR-coated to reduce oscillations in the spectral response due to a Silica passivation layer on the chip. This AR-coating obviously adds extra cost to the ILX511B, but it is still lower cost than the S10420 by at least a factor of 2. Also, the electronics interface to the S10420 BT-CCD detector is more complex and requires more electrical components than the ILX511B. This causes the electronics for the S10420 to be more costly than the electronics for the ILX511B. Summary In summary, we have made comparative measurement of two highly sensitive CCD detector arrays commonly used in spectroscopy the SONY ILX511B and the Hamamatsu S10420. Both detectors have their strong and weak points, so the final choice of which detector best matches your application depends on the following: If technical performance i.e the best Signal-to-Noise ratio for the available light input is most important, you should opt for the Hamamatsu S10420. If low cost is most important, you should opt for the Sony ILX511B. We hope you find this technical note useful and are happy to receive feedback and suggestions for improvement as well as ideas for future technical notes. - 6 -