Madero Ote. 686, Centro Histórico, C.P Morelia, Michoacán México. #300, Col. Cuauhtémoc, C.P Morelia, Michoacán, México

Similar documents
2

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Comparison of Measured Values of CTDI and DPL with Standard Reference values of Different CT Scanners for dose Management

NEMA XR 25 COMPUTED TOMOGRAPHY DOSE CHECK

QUALITY CONTROL AND PATIENT DOSES FROM X-RAY EXAMINATIONS IN SOME HOSPITALS IN THAILAND

Standard. Substitute Test or Procedure. Required Test or. 1 Scan Increment Accuracy. Initially and Annually Initially and Annually

2012 Computed Tomography

IAEA RER/9/135 COURSE ON OPTIMIZATION IN COMPUTED TOMOGRAPHY Sofia, Bulgaria, Tube current modulation and dose reduction : How TCM works

Understanding CT image quality

2017 Computed Tomography

A method for calculating the dose length product from CT DICOM images

Request for Proposals

What do we mean by workload? Number of scans Type of scans/mix of scans? Total mas Total dip

A simple anthropomorphic phantom used to demonstrate the effectiveness of CT dose modulation functions.

Quality Assurance Implementation at the Roberts Proton Therapy Center. James McDonough 3 August 2013

In recent years, CT technology has undergone profound

Equipment Quality Control for Digital Radiography February 22, Imaging Physics CancerCare Manitoba

RSNA 2006 November 26 to December 1 Chicago. Guest author for ImPACT Dr. Koos Geleijns, Medical Physicist, Leiden University Medical Center.

The Black Piranha Easy & Fast X-ray Quality Control

Iterative Reconstruction with Philips idose Characterising Image Quality in Attempting to Realise its Potential

Scope: All CT staff technologist

Assessment of the Impact of Dark Signal on Image Quality in Computerized Mammography. Introduction

3/2/2016. Medical Display Performance and Evaluation. Objectives. Outline

Staff dose and good prac/ces in CT guided IR procedures. Jonas Andersson Medical Physicist, Ph.D.

X-ray Tube Housing Assembly H1076X H1076Y Print No.HA Release Date:

Brilliance CT 64-channel configuration

Intra-oral X-ray unit with high frequency DC generator. Product Data

FFDM Quality Control in Canada - a Vendor Neutral Approach

Promises and Perils of Proton Therapy Beam Delivery (Implications) or Towards Cost Effective Particle Therapy

G. Pittá(*), S. Braccini TERA Foundation, Novara, Italy (*) Corresponding author.

-Technical Specifications-

Implementing a Proton Beam Scanning System within an Operating Clinical Facility

DOSE DELIVERY SYSTEM OF THE VARIAN PROBEAM SYSTEM WITH CONTINUOUS BEAM

Overview. ACR Accreditation Update in Mammography. ACR Topics. Requirements Today. What s Coming For Tomorrow

arxiv:hep-ex/ v1 27 Nov 2003

FDA MAMMOSCAN FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM MAMMOGRAPHY

MPKOMNP. kéï=~ë=çñw== d^ifiblp. j~áåíéå~ååé=fåëíêìåíáçåë=d^ifiblp=`çãñçêí=mirp. båöäáëü. Cover page

A new shielding calculation method for X-ray computed tomography regarding scattered radiation

Introduction. Edge Enhancement (SEE( Advantages of Scalable SEE) Lijun Yin. Scalable Enhancement and Optimization. Case Study:

Non-Destructive Examination Benches and Analysis Laboratories in support to the Experimental Irradiation Process in the Future Jules Horowitz MTR

World First Slim Cassette Type Digital Mammo. Upgrade Solution

CHAPTER 4: HIGH ENERGY X-RAY GENERATORS: LINEAR ACCELERATORS. Jason Matney, MS, PhD

Display Quality Assurance: Considerations When Establishing a Display QA Program. Mike Silosky, M.S. 8/3/2017

Dental Line. High frequency intraoral x-ray unit. radiology ahead

G635. APPLIED SCIENCE Working waves ADVANCED GCE. Wednesday 27 January 2010 Morning. Duration: 1 hour 30 minutes

BEAMAGE 3.0 KEY FEATURES BEAM DIAGNOSTICS PRELIMINARY AVAILABLE MODEL MAIN FUNCTIONS. CMOS Beam Profiling Camera

Precision, power, and productivity

True comfort and flexibility with the power of 3T.

Monitor QA Management i model

National Institute of Radiological Sciences. Naoya Saotome

Imaging diagnostico in Sanità Stato attuale e prospettive

A Comparison of DPM and VMC++ Mont Carlo codes applied to heterogeneous media

Guidelines for Quality Assurance in Radiation Protection for Diagnostic X-Ray Facilities: Large X-Ray Facilities

MODE FIELD DIAMETER AND EFFECTIVE AREA MEASUREMENT OF DISPERSION COMPENSATION OPTICAL DEVICES

Biograph Vision See a whole new world of precision

New Medical Light Source using NTT s Communication Laser Technology

Data Collection Using APEX3. March 30, Chemical Crystallography Laboratory

Department of Medical Physics, Graduate School of Nuclear and Allied Science, University of Ghana, Accra - Ghana, West Africa

DICOM Correction Proposal

Display Quality Assurance: Recommendations from AAPM TG270 for Tests, Tools, Patterns, and Performance Criteria

Display Quality Assurance: Recommendations from AAPM TG270 for Tests, Tools, Patterns, and Performance Criteria

NEW ACHIEVEMENTS IN THE FIELD OF DEVELOPMENT OF MOBILE X-RAY TV- SYSTEM

JOURNAL OF PHARMACEUTICAL RESEARCH AND EDUCATION AUTHOR GUIDELINES

Case 1 8/1/2017. Ring Artifact? James M. Kofler, Ph.D. Mayo Clinic Rochester. AAPM Annual Meeting TU-B Artifacts: CT 2

Mechanical aspects, FEA validation and geometry optimization

AAMD Writing Competition Guidelines for Submission

A dedicated data acquisition system for ion velocity measurements of laser produced plasmas

Collimation and Light Field Brightness

Impact of DMD-SLMs errors on reconstructed Fourier holograms quality

Technical description and user manual. Survey Meter SM 8 D. Sensortechnik und Elektronik Pockau GmbH. Siedlungsstraße 5-7 D Pockau-Lengefeld

Cosmos2 The Economical X-ray Solution with the Best Image Quality and High Reliability

How Quadrupole Size Affects Research

Classic. Best Practice QA Solutions for All Rotational Therapies & IMRT For International Use Only. Classic. RIT113 Technical Specifications

The Digital Debate. Competing Technologies. Up and Coming 1/5

Comparison of Mixed-Effects Model, Pattern-Mixture Model, and Selection Model in Estimating Treatment Effect Using PRO Data in Clinical Trials

PBS Products from Pyramid

ISO/IEC INTERNATIONAL STANDARD

BUREAU OF ENERGY EFFICIENCY

Vascular. Development of Trinias FPD-Equipped Angiography System. 1. Introduction. MEDICAL NOW No.73 (2013.2) Yoshiaki Miura

Durham Magneto Optics Ltd. NanoMOKE 3 Wafer Mapper. Specifications

A NOVEL GANTRY FOR PROTON THERAPY AT THE PAUL SCHERRER INSTITUTE

Nuclear Associates and

Laser Beam Analyser Laser Diagnos c System. If you can measure it, you can control it!

JEFFERSON COLLEGE. Image Intensification & Equipment

Academic and Research Staff. Prof. John G. King Dr. John W. Coleman Dr. Edward H. Jacobsen. Norman D. Wittels

VISION. Instructions to Authors PAN-AMERICA 23 GENERAL INSTRUCTIONS FOR ONLINE SUBMISSIONS DOWNLOADABLE FORMS FOR AUTHORS

ISO INTERNATIONAL STANDARD

Development of New Carbon Therapy Facility and Future Plan of HIMAC

FS3 Series Tri-Lens for SEOUL SEMICONDUCTOR Z-POWER P4 TM LEDs

arxiv: v1 [physics.ins-det] 1 Nov 2015

Guidelines for Assuring Softcopy Image Quality

Linatron - M9 & M9A. Modular high-energy X-ray source. 2.0 Performance

Commissioning the TAMUTRAP RFQ cooler/buncher. E. Bennett, R. Burch, B. Fenker, M. Mehlman, D. Melconian, and P.D. Shidling

Optical Technologies Micro Motion Absolute, Technology Overview & Programming

Assessing the variability of seismic response analysis of a tall guyed telecommunication tower with ambient vibration measurements

TORCH a large-area detector for high resolution time-of-flight

Technologist Quality Control Procedures

Recent APS Storage Ring Instrumentation Developments. Glenn Decker Advanced Photon Source Beam Diagnostics March 1, 2010

CARESTREAM VITA/VITA LE/VITA SE CR System Long Length Imaging User Guide

Characterisation of the far field pattern for plastic optical fibres

Transcription:

Verification of and DLP values for a head tomography reported by the manufacturers of the CT scanners, using a CT dose profiler probe, a head phantom and a piranha electrometer Edith Castillo Corona 1,2, Ix-Berenice García Ferreira 2, Jesús García Herrera 2, Sergio Román López 2, Omar Salmerón Covarrubias 3 1 Comisión Estatal de Protección contra Riesgos Sanitarios. Servicios de Salud de Michoacán. Av. Madero Ote. 686, Centro Histórico, C.P. 58000 Morelia, Michoacán México 2 Centro Estatal de Atención Oncológica. Servicios de Salud de Michoacán, Gertrudis Bocanegra #300, Col. Cuauhtémoc, C.P. 58020. Morelia, Michoacán, México 3 Hospital General Dr. Miguel Silva. Servicios de Salud de Michoacán. Calle Isidro Huarte SN, Centro Histórico, C.P. 58000 Morelia, Michoacán México 1 Address Correspondence to Edith Castillo Corona (e-mail: edithcastillocorona@gmail.com) Abstract The extensive use of Computed Tomography (CT) and the associated increase in patient dose calls for an accurate dose evaluation technique. The CT contributes up to 70% of the total dose given to patients during X-ray examinations. The rapid advancements in CT technology are placing new demands on the methods and equipment that are used for quality assurance. The wide beam widths found in CT scanners with multiple beam apertures make it impossible to use existing CT ionization chambers to measure the total dose given to the patient. Using a standard 10 cm CT ionization chamber may result in inaccurate measurements due to underestimation of the dose profile for wide beams. The use a CT dose profiler based on solid-state technology and the Piranha electrometer from RTI electronics provides a potential solution to the arising concerns over patient dose. This study intend to evaluate the feasibility and accuracy of CT Dose Index () and Dose Length Product (DLP) values for a head tomography reported by the manufacturers of the CT scanners at each study. Keywords:, DLP, CT dose profiler, Piranha. 1. - Introduction. Modern CT scanners provide two dose parameters that both became available by the scanner manufacturers around 2001: the Volume ( vol ) measured in mgy, and the dose-length product (DLP) measured in mgy-cm. vol is a measure of the average dose within the scan volume to a standardized phantom. The total amount of radiation delivered to a standardized phantom is represented by the DLP, which is the product of vol and the scan length. In 1981, Shope et al., introduced the as a metric to quantify the radiation output from a CT examination consisting of multiple contiguous CT scans (ie, multiple adjacent transverse rotations of the x-ray tube along the patient longitudinal axis). This means that 426

the is defined as the integral of the single scan radiation dose profile along the z-axis, normalized to the thickness of the imaged section ( slice thickness ). In the same reference, showed that with corrections for scan spacing, the can estimate Multiple Scan Average Dose (MSAD) in a standardized, convenient manner. is usually measured with a pencil-shaped ionization chamber, although methods have been developed that use alternative detectors, including an optically stimulated luminescence probe and a solid-state real-time detector. 1.1.- Theoretical basis. Because CT is an X-ray based imaging technique that increasingly gives a significant contribution to the patient dose, it is of major importance to have Quality Assurance (QA) procedures with high validity is updated with the growing development of new techniques in CT. In CT the most common parameter for estimating the radiation dose is the. is the integral of air-kerma along the rotational symmetry axis for the X-ray tube, here denoted z, divided with the number of simultaneously acquired slices, N, of nominal thickness T: L 1 2 L NT 2 D( z) dz (1) Where the L in eq. (1) defines the length over which the integral is made, ideally the length of integration should be equal or longer than the actual physical air-kerma profile width. The quantity can be interpreted as the radiation energy deposited in a slice with a thickness corresponding to the nominal beam collimation thickness. The dose inside the slice is the and the dose outside the slice is excluded (see figure 1). Figure 1.- The dose inside the slice is the and the dose outside the slice is excluded. There are a number of different quantities related to. The most common are summarized in the table below: (see the CT Dose Profiler-Users Manual-English-6.2A) 427

1 D( z) dz T Pitch 1 2 w 100( center) 100( peripheral) 3 3 d NT 428 Symbol Quantity Remarks CT Dose Index General dose description for CT MSAD Multiple Scan Average Dose As but corrected for pitch 100 (100) Current definition of w Weighted Main descriptor of local dose vol Volume As w but corrected for pitch DLP Dose Length Product Includes the irradiated volume and represents the overall exposure for an examination In single slice CT (figure 1) the expression for is defined (in mgy) as: where T is the nominal beam collimation thickness in mm and D(z) is the dose profile. On the y-axis the quantity is relative dose. 100 is acquired by reducing the integral to go between -50 and 50. For Multiple Slide CT (MSCT), is defined in equation (1), where N is the number of detectors and T is the width of a detector. w represents an average value of the 100 inside a phantom (this requires five measurements, one in each hole): a) In the case of single slice CT, the slice thickness is determined by the width of the detector. b) The slice thickness in MSCT is determined by the number of detectors and the widths of the detectors. In spiral CT there is an additional factor called the CT pitch factor. It is defined as the table movement per gantry rotation: where d is the distance in mm that the couch moves between consecutive serial scans or per 360º rotation in helical scanning, N is the number of detectors and T is the detector thickness in mm. vol is the same as w but with respect to the pitch factor in helical (spiral) scanning:

vol Pitch DLP L w The dose-length product, DLP, includes the irradiated volume and represents the overall exposure for an examination and is calculated as following: vol where L is the scan length of a certain examination. vol is conceptually similar to the MSAD but is standardized with respect to the integration limits (±50 mm) and the factor used to convert the exposure or air kerma measurement to dose in air. vol represents the dose within the scan volume from a particular scan protocol for a standardized phantom. It is a measure of the scanner output and not patient dose. International Electrotechnical Commisions (IEC) standard 60601-2-44 requires that the vol be displayed on the console of modern CT scanners before the scan is initiated (See IEC. (2002). Medical electrical equipment, part 2-44). The purpose of this work was to evaluate the feasibility and accuracy of and DLP values for a head tomography reported by the manufacturers of the CT scanners at each study using a CT dose profiler probe based on solid-state technology and the Piranha electrometer from RTI electronics. 2.- MATERIALS AND METHODS Clinical protocols were investigated for head CT examinations using typical scan ranges. The CT protocols investigated were in helical simulation mode on General Electric CT Light Speed RT scanner, kv max: 140 kv, I max: 380 ma. Comparing the DLP and values shown in the studies investigated with those obtained with the experimental arrangement shown in figure 2. Figure 2.- Experimental arrangement The equipments used in the experimental arrangement are: 429

1. A General Electric CT Light Speed RT scanner. 2. A CT Dose Profiler (CTDP) probe. 3. The Piranha electrometer from RTI Electronics and the Ocean Software. 4. A PMMA head phantom The CT Dose Profiler (CTDP) probe is a highly advanced point dose probe designed to fit into the standard phantoms to evaluate computed tomography systems. There is no limit to the slice width that users can measure with the CTDP. When using this probe for measurements, the traditional five axial scans with an ion chamber are replaced with one helical (spiral) scan with the CTDP probe in the center hole of the phantom head (Figure 3). The CT Dose Profiler replaces the conventional TLD and OSL methods or film for dose profile measurements. The CT Dose Profiler probe is designed to be used with the Piranha X-ray electrometer and a PC running the Ocean 2014 software. Figure 3.- The CT head phantom on the head support and the CTDP probe in the center hole with the connector pointing towards the couch and connected with the Piranha detector. A CT Dose Profiler (RTI Electronics, Mölndal, Sweden) equipped with 2 2 0.3 mm solid-state detector chip was used for measurement of absorbed dose. The output signal of the detector probe was connected to an X-ray electrometer (Piranha, RTI Electronics) and dosimetry information stored on a laptop computer running Ocean software (RTI Electronics) (Figure 4). 430

Figure 4.- Dosimetry information in Ocean software. 3.- RESULTS The data of the analysis to some clinical protocols are shown in table 1: Table 1.- Clinical Scan Parameters for the CT Examinations Studied. Standard parameters: 120 kvp 10 ma Type Scan range vol DLP [mm] [mgy] [mgy-cm] Helical I91.750-S155.750 22.4 596.19 Helical I150.00-S185.00 24.93 886.8 Helical I252.00-S168.00 14.14 660.23 Helical I160.500-S99.500 22.44 629.85 The results of measurements with the experimental arrangement are shown below: Real time display This report was created with Ocean. Test date: 22/05/2015 431

432

433

Test equipment used Meter(s): Electrometer(s): Piranha S/N CB2-13010824 PiranhaCTDP S/N DP2-13040016 4.- DISCUSSION The values of vol and DLP displayed by the manufacturer of the CT in the analyzed clinical studies differ from the values obtained with Piranha and the CTDP, however, to obtain the averages of both parameters observe that the values of vol are very similar, and the values of DLP differ by more than 100%; this last is explained by the way to perform the image study of head since physician always ask that the scan become a little larger than the head, which affects the calculation of the DLP that performs the CT. The values of DLP obtained with the experimental arrangement use exactly the size of the phantom head, from there these results. If in analyzed clinical studies will conform strictly to the size of the head, the results would be similar. A quicker way to perform quality assurance has been introduced by using the CT Dose Profiler probe and the Ocean 2014 software. To be able to use it, the k-factor must be known for the CT-unit and the type of phantom that is used for the measurement. A number of k-factors for common CT-units are used by the software and listed in the Appendix k- factors (see CT Dose Profiler User's Manual 2014-06-23/6.2A). 5.- CONCLUSIONS The displayed vol given by a manufacturer may be a representative figure for that model and not the value measured on the particular CT scanner (see International Standard IEC 60601-2-44. Particular requirements for the safety of x-ray equipment for computed tomography, International Standard, Geneva, Switzerland, International Electrotechnical Commission, 2002). To measure the 100 with the CT Dose Profiler in the center hole of a head phantom with one helical scan exposure and then multiply it with the k-factor to get w and vol and then calculate DLP is, of course, faster than doing the five exposures with the pencil ion chamber. With the CT Dose Profiler you can also see a visible image of the dose profile that will tell you if something is wrong with the system. vol provides a very useful way to compare the doses delivered by various scan protocols or to achieve a specific level of image quality for a specific size patient. vol can be used to prescribe the right dose for a specific patient size and diagnostic task. However, vol and DLP values cannot be used as a surrogate for patient dose, either in epidemiologic assessments of potential late effects or for potential deterministic effects. Neither vol nor its derivative, dose-length product DLP, should be used to estimate effective dose or potential cancer risk for any individual patient. In conclusion, it is imperative that measures of the radiation output of a CT system can be easily and practically measured in a consistent and robust form. vol and DLP meet these criteria. The vol and DLP parameters tells the medical physicist precisely how the machine was operated, and it can be used, in 434

435 conjunction with information regarding patient size and the scanned anatomy, to estimate patient dose. The and DLP values are not, however, patient dose estimator. Acknowledgments We want to thank COFEPRIS for their support to the State Program on Radiation Protection. We are very grateful for the support of the company "Electrónica y Medicina, S. A." (EYMSA), for the presentation of this work in. REFERENCES AAPM Report 96, Report of AAPM Task Group 23: CT Dosimetry. 2008 Cynthia H. McCollough, Shuai Leng, Lifeng Yu, Dianna D. Cody, John M. Boone, Michael F. McNitt-Gray (2011). CT Dose Index and Patient Dose : They Are Not the Same Thing. Radiology: Volume 259: Number 2 May 2011. FDA. [online]. What are the Radiation Risks from CT? http://www.fda.gov/radiation- EmittingProducts/RadiationEmittingProductsandProcedures/MedicalImaging/Medica lx-rays/ucm115329.htm. [Reviewed on March 2015]. (Unless otherwise noted, the contents of the FDA Web site (www.fda.gov)--both text and graphics--are not copyrighted. They are in the public domain and may be republished, reprinted and otherwise used freely by anyone without the need to obtain permission from FDA. Credit to the U.S. Food and Drug Administration as the source is appreciated but not required.) International Electrotechnical Commission. (2002). Medical electrical equipment, part 2-44: particular requirements for the safety of x-ray equipment for computed tomography. IEC publication no. 60601-2-44. Ed 2.1. Geneva, Switzerland. Jodie A. Christner, James M. Kofler, Cynthia H. McCollough (2010). Estimating Effective Dose for CT Using Dose Length Product Compared With Using Organ Doses: Consequences of Adopting International Commission on Radiological Protection Publication 103 or Dual-Energy Scanning. AJR 2010; 194:881 889 Lewis Maria (2005). Radiation dose issues in multi-slice CT scanning. MHRA ImPACT technology update no. 3. January 2005. RTI Electronics AB (2014). CT Dose Profiler User's Manual English- Version 6.2A. RTI article number: 9630512-00 RTI Electronics AB (2014). Piranha & QABrowser User's Manual 2014-06/5.5C. RTI article number: 9629051-10 RTI Electronics AB (2012). Ocean Reference Manual 2012-10-01/2.3A. RTI article number: 9604550-00 Shope TB, Gagne RM, Johnson GC (1981). A method for describing the doses delivered by transmission x-ray computed tomography. Med Phys 8 (4): 488 495