h c HUNG CHANG D IBM PC 586, Pcntium Software Drive : Windows 95 Dual Clmnel Sampling per channel : 100 MS/s

Similar documents
Oscilloscope Guide Tektronix TDS3034B & TDS3052B

PRELIMINARY INFORMATION. Professional Signal Generation and Monitoring Options for RIFEforLIFE Research Equipment

Expect to Make Waves.

Digital Storage Oscilloscopes 2550 Series

MSO-28 Oscilloscope, Logic Analyzer, Spectrum Analyzer

ScopeMeter 190 Series Specifications

RIGOL. Data Sheet. DS1000B Series Digital Oscilloscopes DS1074B, DS1104B, DS1204B. Product Overview. Easy to Use Design. Applications.

MINI PC SCOPE PCSU01. User manual. test leads software download USB cable design enclosure

S op o e p C on o t n rol o s L arni n n i g n g O bj b e j ctiv i e v s

Quick Start. RSHS1000 Series Handheld Digital Oscilloscope

Burlington County College INSTRUCTION GUIDE. for the. Hewlett Packard. FUNCTION GENERATOR Model #33120A. and. Tektronix

SMART Trigger modes like Glitch, Window and Dropout allow you to capture precisely the events of interest.

WAVEJET 300 SERIES OSCILLOSCOPES. New Cover to Come. Unmatched Performance, Portability, and Value

Please feel free to download the Demo application software from analogarts.com to help you follow this seminar.

Embest DSO2300. Feature. General Description:

2016 RIGOL TECHNOLOGIES, INC.

WAVEJET 300 SERIES OSCILLOSCOPES. Unmatched Performance, Portability, and Value

PicoScope 6407 Digitizer

Reference. TDS7000 Series Digital Phosphor Oscilloscopes

DIGITAL OSCILLOSCOPES & DIGITAL SCOPES

SDS1000C Specifications

Technical Reference. TDS 684A, TDS 744A, & TDS 784A Digitizing Oscilloscope Performance Verification and Specifications

User s Manual. GAO-DSO25216A Digital Storage Oscilloscope with Logic Analyzer, FFT Spectrum and Counter. Revision Software Win98/me/2000/xp Version

PicoScope 6407 Digitizer

SX-METRO A software package to meet all your needs

Manual Supplement. This supplement contains information necessary to ensure the accuracy of the above manual.

~.Yr / 12. Offset range: ±2 V (2 mv/div to 200 mv/div), ±SO V (> 200 mv/div to 5 V/div) "digital storage oscilloscope" 13.

Analog Arts SA985 SA975 SA935 SA915 Product Specifications [1]

Spectrum Analyser Basics

PC Oscilloscope Spectrum Analyzer Logic Analyzer

Portable Performance for Debug and Validation

PicoScope PC Oscilloscopes. User's Guide. ps2203.en r4 Copyright Pico Technology Limited. All rights reserved.

DataSheet. SDS1000DL Series Digital Oscilloscope

Analog Arts SA985 SA975 SA935 SA915 Product Specifications

NanoGiant Oscilloscope/Function-Generator Program. Getting Started

imso-104 Manual Revised July 19, 2012

University of Utah Electrical & Computer Engineering Department ECE1050/1060 Oscilloscope

Overview. Know Your Oscilloscope. Front Panel. Rear Panel. Sharing Agilent s Resources with Engineering Educators

Beginners How to Test DSO138mini

Tablet Oscilloscope Quick Guide

EMS DATA ACQUISITION AND MANAGEMENT (LVDAM-EMS) MODEL 9062-C

PicoScope 4000 Series PC Oscilloscopes

FOUR CHANNEL USB RECORDER PCRU01. User manual. Table of Contents

TDS 520B, TDS 540B, TDS 620B, TDS 644B, TDS 680B, TDS 684B, TDS 724A, TDS 744A, & TDS 784A

50 MHz Digital Storage Oscilloscope

ATS MS/s 8-Bit PCI Digitizer

Experiment 13 Sampling and reconstruction

PicoScope 2000 Series PC Oscilloscopes

ME EN 363 ELEMENTARY INSTRUMENTATION Lab: Basic Lab Instruments and Data Acquisition

Agilent Technologies 54522A

imso-104 Manual Revised August 5, 2011

DataSheet SDS1000CML Series Digital Oscilloscope

DataSheet SDS1000CFL Series Digital Oscilloscope

INSTRUCTION GUIDE. for the. Keysight (formally Agilent; Hewlett-Packard) Model #33509B. and. Tektronix. Fall 2017

Working with a Tektronix TDS 3012B Oscilloscope EE 310: ELECTRONIC CIRCUIT DESIGN I

ADS1000C, CAL / CML Series

MIE 402: WORKSHOP ON DATA ACQUISITION AND SIGNAL PROCESSING Spring 2003

PC Scope PCS64i. PCS64i POWER DC ON AC

Experiment # 4 Counters and Logic Analyzer

Choosing an Oscilloscope

Electrical and Electronic Laboratory Faculty of Engineering Chulalongkorn University. Cathode-Ray Oscilloscope (CRO)

Quick Start. SHS1000 Series Handheld Digital Oscilloscope QS03010-E02B 2015 SIGLENT TECHNOLOGIES CO., LTD

PORTABLE OSCILLOSCOPE

Analog Discovery Scope and Waveform Generator Edited 11/15/2016 by Eric Scotti & DGH

PC Scopes PCS500 PCS100 / K8031. CHECK THE Pc-Lab2000 TM GETTING STARTED GUIDE FOR SOFTWARE INSTALLATION

What to look for when choosing an oscilloscope

4CH/2CH Digital Storage Oscilloscopes

PicoScope. User guide. Copyright 2005 Pico Technology Limited. All rights reserved. PSW044 v1.5

PicoScope 4000 Automotive PC Oscilloscopes

Instrument Firmware Release Notes

SigPlay User s Guide

SNG-2150C User s Guide

2 MHz Lock-In Amplifier

Model P2016. Dual Channel Handheld Digital Storage. Oscilloscope & Multimeter. Quick Scope Guide

PicoScope 3000 Series Automotive User guide

M570 series oscilloscope User s Guide Version 1.02

25 MHz Digital Storage Oscilloscope

Analyze Frequency Response (Bode Plots) with R&S Oscilloscopes Application Note

Logic Analyzer Auto Run / Stop Channels / trigger / Measuring Tools Axis control panel Status Display

Lab 2: A/D, D/A, and Sampling Theorem

University of Utah Electrical Engineering Department EE1050/1060 Oscilloscope. Name:, Lab TA:

LeCroy Digital Oscilloscopes

SDS 1072/1074CFL / SDS 1102/1104CFL SDS 1202/1204CFL / SDS 1302/1304CFL Digital Storage Oscilloscopes 70MHz / 100MHz / 200MHz /300MHz

User Manual. Digital Oscilloscope V1.9

1 scope channel. 2 scope channels* 200 MSa/s 4 MB memory/ch. 200 MSa/s 2 MB memory/ch. 200 MSa/s 2 MB memory/ch

Fluke /

User Manual. GPS-1000B+ series. Digital Oscilloscope V 1.1

QRF5000 MDU ENCODER. Data Sheet

Analog Storage Oscilloscope TS-81000/ Newly developed CCD(Charge-coupled device) scan converter tube

Digital Oscilloscope Waveform Generator DC Power Supply Digital Multimeter Spectrum Analyzer Handheld Oscilloscope Probes & Accessories

OSCILLOSCOPE AND DIGITAL MULTIMETER

PC Oscilloscope Spectrum Analyzer Logic Analyzer

Introduction to the oscilloscope and digital data acquisition

User Manual. Digital Storage Oscilloscopes Models 2534, 2540 & General Safety Summary. Version 1.03

Two Channel PC Oscilloscopes with Arbitrary Waveform Generator OMSP-2000 Series

Advanced Test Equipment Rentals ATEC (2832)

PicoScope 6 Training Manual

What is oscilloscope? What an oscilloscope looks like. Oscilloscopes, accessories, applications Ján Šaliga

User Manual SDS1000CML+/SDS1000DL+ Digital Oscilloscope UM0101A-E01A SIGLENT TECHNOLOGIES CO,.LTD

DAC Express Release 3.4 (VT9801B)

Transcription:

h c HUNG CHANG http://www-hhvngchang.com Hongjc Bldg. 301-2, Hongje-dong. Seodaemun-ku, Seoul, Korea TEL : 82-2-395-8611-19 FAX : 82-2-395-5381184 D IBM PC 586, Pcntium Software Drive : Windows 95 Dual Clmnel Sampling per channel : 100 MS/s m Auto setup D Auto Calibration Memory Depth : 32K byte/cl-l

PC BASED DSO 2100 How To.... Installing software Installing hardware Auto Setup Guide to operations Screen Auto Calibration Probe Calibration FFT XY plot Hardware Specifications Commands File menu View menu Help menu

Installing software 1. Insert the distribution disk into CD drive. 2. Run Windows. 3. Select File menu. 4. Select Run option. 5. Enter file to run CD:\Install.exe 6. Follow the on screen instructions.

Installing hardware Please follow these instructions for installing the PC BASED DSO 2 100. 1) Connect the DSO to the provided cable. 2) Connect the cable to the parallel port of your computer. 3) Connect the Oscilloscope probe to the BNC connector (CH.A) and connect the probe tip to the CAL terminal on the back of the DSO 4) Connect the power cable to the DSO. 5) Plug the power adapter into the wall. 6) After checking all connections, turn on the computer and peripherals. You are now ready to install the software. 7) After installing the software, calibrating the scope probes is a good idea. See the section below on hardware, the Calibration Procedure

Auto setup Auto setup operates on the analog channel that is currently defined as the trigger channel. It will attempt to find setting that provide a stable and consistent acquisition. Auto setup will<<auto setup>> on the following settings Voltage/Division for trigger channel. When you dual mode, Auto setup is selected to CH.A

Guide to operations When making measurements with the PC BASED DSO 2 100 Digital Storage Oscilloscope, meaningful data can only be captured with some prior knowledge of the characteristics of the circuit under test. Before initiating any capture cycles, the PC BASED DSO 2 100 must be configured using the control program. See the software section later in the manual for instructions on these procedures. To connect the PC BASED DSO 2 100 to the test circuit, there are two standard BNC probes, one for each Analog input channel The scope probes have removable hook clips on their ends and an attached alligator clip for the signal ground connection. Make sure that there is a common ground between the PC BASED DSO 2100 and the circuit under test. Each Analog channel probe has a calibration adjustment. It is important that this calibration be made at least twice a year. See Auto Calibration for more information.

Screen. Main Screen Zero Point Tick Mark Cursors A and B Horizontal Scroll Bar Display Time A Tireger Menu Parmater Display

Main Screen The main display is made up of areas: - On the right side of the screen is the settings/parameter display. - In the middle are the data displays. - The top is the analog waveform display. - Above the analog display is a Time A graph representing the data buffer. This also shows the location of the cursors and which part of the buffer is being displayed. - On the right side of the screen you will see the settings for one of the analog channels. You can select the active channel with the select button. - Just to the left of the analog waveform display are markers for the ground points. CH-A markings are closest to the display, than CH-B. Time A The Time A Graph display represents the data buffer. The graph shows the locations of Cursor A, Cursor B, and the Trigger cursor, and also the current portion of the buffer displayed on the screen. Each item is displayed in its color. If Thumbnails are turned on a miniature waveform is displayed, otherwise a horizontal line is displayed for each channel.

Zero Point Tick Marks Located to the right of the Analog Display. The Zero Point Tick Marks are > shaped. These display the ground points of each analog channel. Zero Point Tick Marks associated with Channel A are left most and Channel B through Display are successively further to the left. These Tick Marks can be moved by click of mouse by grabbing and dragging with a pointing device

Cursors A and B The Cursors provide an easy means of voltage or time-and-frequency measurements. For a selected channel, the voltage or time-and-frequency difference between the two cursors is shown in the Parameters Display area. Cursor A and Cursor B can be moved by: - Grabbing and dragging the cursors with a pointing device. - Selecting the Cursor by clicking on the Cursor A or Cursor B button. - Selecting the cursor from the mouse.

Display Time A Shows or hides the display of the analog waveform in the Time A Graph area. A Time A display is a compressed version of the main display that shows a complete view of the data buffer. With Time A off, a line is displayed with a bar at each end that depicts relative position of each displayed channel. A box is drawn to depict the area that is displayed by the full-size window, that box is the same color as the grid. The cursor positions are also displayed. Note: Turning this off will speed up the initial display of data.

Horizontal Scroll Bar This scroll bar is used in conjunction with a selected waveform. The Horizontal Scroll Bar will move a selected waveform or cursor left or right in the display area.

Trigger menu Trigger settings Source Slope Couple Mode menu commands. - - -..-- ~..

Trigger settings All trigger and capture parameters are displayed in this box and can be altered in it as well. These parameters are also accessible through the pull down menus. Source Set trigger channel. Slope Set trigger slope. Mode Set trigger mode. See also Trigger Basics.

. Source Trigger CH- 1 Trigger on channel 1. Trigger CH-2 Trigger on channel 2. Trigger EXT Trigger on channel EXT

Trigger Slope -- When the trigger channel is set to one of the analog channels. Trigger Slope can be set to positive(rising) or negative(falling) edge.

Couple Mode menu commands The trigger mode can be set in the trigger dialog box to the following: Normal Trigger on a rising or falling edge at a certain level AC TV Horizontal Trigger on Horizontal TV signal. TV Vertical Trigger on Vertical TV signal.

Parameter Display The Parameters Display shows the following: Coupling (for selected channel) V/Div (for selected channel) Probe settings ( for selected x10, xl ) Cursor measurements Time Voltage Frequency Clicking on the Select button will change which channel is selected. Clicking on the display channel button on the toolbar will change which channel is selected. Clicking on the Edit button allows the user to change settings for the selected channel.

Probes Calibration For the most accurate measurements possible, the probes used must be calibrated. There is a built in calibration signal point on the CAL terminal connector. Follow this procedure to calibrate each probe: 1) 1) Verify that the probe attenuation setting matches the actual probe 2) Connect the probe (BNC end) to the channel A or channel B inputs. 3) Connect the scope probe tip (pointy end) to the CAL terminal connector by holding the probe s tip against the calibration point. A Square wave signal should appear on the screen 4) Adjust the probe calibration(a screw adjustment near tip) until a true square wave is shown on the screen, noting the comers of the waveform which should be sharp and square, not rounded over or peaky Auto Calibration Auto Calibration of the PC BASED DSO 2 100 hardware is performed by the software each time it is started up. This ensures that the hardware is fnnctioning correctly before each use. While the tests are being performed, the following window showing a list of each test is displayed.

FFT The FFT dialog allows control and display of FFT s. The following controls are available: Horizontal zoom Select horizontal zoom ratio. Vertical scale Select vertical scaling factor.

XY Plot XY Plot will display the buffer data with one channel of data as the X axis data and another channel of data as the Y axis data. This mode duplicates the XY Plots commonly found on analog oscilloscopes. The radio buttons on the right side of this dialog box select which channel of data will be used for the X axis and Y axis.

. Hardware Specifications CLOCK SPECIFICATIONS Internal Frequency : 1 Hz to 30 MHz Timebase: 50ns to 320s / Division displayable, Scroll mode to 1s ANALOG INPUT SPECIFICATIONS Input of Channels : CH.A, CH.B Input Voltage Sensitivity: Probe 1:l lomv/div to 5Vldiv Probe 10: 1 100 mv / div to 50 V / div Input Impedance analog channels : 1 M ohm shunted by 25 pf Absolute Max. Input Voltage: OV continuous and 250 V (DC + AC < 10 KHz) Transient. Usable Input Voltage Range : Trigger: Horizontal mode: 5ov Positive or Negative Slope adjustable level. Auto Normal TV horizontal TV vertical Single Normal Auto MEMORY Data Acquisition Memory Modes. Memory Size: 32WCH. Minimum required: a minimum of 4 Mbytes RAM is necessary to use the DSO control program. DISPLAY Cursors: There are two cursors. Cursor-A and Cursor-B they are both time and voltage or frequency cursors. They can be moved using the horizontal and vertical scroll bars or by grabbing and dragging them. Differences are automatically calculated and displayed on the screen.

File menu commands The File menu offers the following: Data Load Load file. Data, settings. Select which channels to load. Data Save Save file. Data, settings. Graph Save to picture Graph copy clipboard Print Prints data. Exit Exit PC BASED DSO 2 100 software

View menu commands The View menu offers the following: Grid Show or hide grid on analog display. Cursor Quickly moves Cursors onto display window. Independent Scrolling Select channel for independent movement. Lines Displays connections between analog data points only (fastest). Dots Displays the analog data points only. Time A Shows a compressed version of the entire buffer directly below the toolbar.. FFT Display FFT window. XY Plot Display XY plot..

Help menu commands The Help menu offers the following commands, which provide you assistance with this application: Using Help Provides general instructions on using help. About Displays the version number of this application.

fised DSO 2100 PC BASED DIGITAL STORAGE OSCM LOSCOPE Iyg HUNG CHANG http://www. hungchangxam

This chapter introduces you to the PC BASED DSO 2100 program and provides specifications for the PC BASED DSO 2100. It is divided into two sections as follows; Introduction Suyportcd Hardware Provides a general produrt dcscriytion Prozridcs spwyficatinns for the follozuirzg scope cnrd 21 Oil Digital Storage Oscilloscope. PC BASED DSO 2100 is a powerful software package for controlling the PC BASED DSO 2100 IBM PC-based oscilloscope. The program allows multiple boards to operate together to perform simultaneous data acquisition on up to 2 different channels as rate as high as 10OMS/s. The user has complete control of all capture and display feature. As many as 2 channels can be viewed together on the I C screen, and signals and system setups can be saved to disk for future use. PC BASED DSO 2100 requires a VGA monitor and graphics adapter and will make use a numeric co-prosessor, if present. Display screens can be printed to the printer that supported to the windows 95

This instrument is installation category( Over Voltage Category) : II If the equipment is used in a manner not specified by the manufacturer the protection provided by the equipment may be impaired. Power off : Remove the power cable from DC input jack Symbol Guide A : Caution ( refer to accompanying documents ) zz : Direct current h : Frame or chassis terminal

Minimum System : IBM PC 586, I entium or compatible with 8MB of RAM Requirements (Recommend 16MB); VGA graphics card and monitor; and at least 10 MB of free hard disk space. Scope card board must be present in the system to acquire data. At windows 95 software. Print Port : ECI? EPI? Nomal. Mode. PC Connection : Parallel Power : DC -+I- 8 V, (+/-lo%) 0.8/0.2 A Environmental Conditions : 0 C - 40 C indoor use, Altitude up to 2000 m 50% -8O%RH Time Range Cursors Vertical Mode Horizontal Mode Grid Zoom X-Y Mode Coupling Input Range Scroll Mode 5 ns/div to 1 Hour/div l-2-5 -step. (Scroll mode to 1s) av and AT CH A/CH B/ DUAVADD/SUB Auto/Nor&Single on/off Voluntary Screen AC, TV-V/H 10 mv/div to 5 VfDIV l-2-5 9- Step 1 s/div--1 Hour/div

Inputs per scope Sample Rate Memory Bandwidth Resolution Amplitude Auto Setup Connector Impedance Auto CAL External Depth calibration card 2 CH A : 100 MS/s, CF I B : 100 MS/s, DUAL : 100 MS/s 32 KB/CH 30 MHz 8 bits Max 250 V + pk 50 Hz to 30 MHz BNC lmr,25pf Gain Accuracy :i 3% lkhz2v Max 100 V+pk Trigger Source Trigger Slope Trigger Type CHlICH2JEXT Positive or Negative Analog

Channel 2 Spectral range 0 Hz-50 MHz Horizontal Zoom Select Horizontal zoom ratio Vertical Scale Select vertical scaliry factor Hard copy Materials Supplied The printer that supported to the windows 95 PC BASED DSO 2100 Installation CD Software Drives Available as a DLL for the windows 95 environment for Visual C and Visual Basic Dimensions Weight 225(H)x125(W)x35mm(D) 400 g approx Warrants that this product will be free from defects in materials and workmanship for a period of one(l) year from the lute of shipment.

PC BASED DSO AC Adapter PC Interface Cable 1 Set KNV-1615, AC 230 V, 50 Hz, 150 ma (Output : DC +/-8 V, l/ia) HYOSUNG, 1 Eh 1 EA Probe 2EA [I; the surface of the case is dirty rub the affected area lightly with a soft cloth soaked in light natural detergent To avoid electric shock, No operator serviceable parts inside refer servicing qualified service personnel.

This chapter describes the installation procedure for the PC BASED DSO 2100 program and for the PC BASED DSO 2100. You will perform the following steps; 1) Connect the DSO to the provided cable 2) Connect the cable to the parallel port of your computer. 3) Connect the oscilloscope probe to the BNC connector(c1-i.a) and connect the probe tip to the cal terminal on the back of the DSO. 4) Connect the power cable to the DSO. 5) Plug the power adapter into the well 6) This instrument is designed for natural-aucooling There should be space at least 3Cm from left, right side of the instrument To install the PC BASED DSO 2100 program on your PC, select the appropriate CD for your computer s CD drive.. Inscrf fire &tribrrfion CD info driac CD. fl Rirf~ zuidozus. H Select File rrrerrrl. m Select RWI 0pfion.. Eflfcrfile to run [ CD drive ] : setup. exe Follom fire 011 screcrr insfrzrcfion.

ADAPTER PC