Introduction to S1210 EMI Pre-Compliance Test Software Edward Pan edward.pan@rigol.com RIGOL TECHNOLOGIES,INC. Rev.1.0 2017-OCT
S1210 EMI Test Solution S1210 EMI Pre-Compliance Test Software
Product Highlights Implement EMI Pre-Compliance testing to find root causes of circuitry problem. Simplify the Pre-Compliance testing procedure based on complicated testing environment. Predict Pass/Fail result in front of different kinds of EMI testing standards before sending to formal compliance testing. Save your constrained budget under highly competitive market.
Product Features Amplitude Calibration Function Standard Limited Line Setting Function Sweep Function Peak Value Search Function Test Report Generation Graphic Type Operation Peak and Marker Table Multi-Display Mode Support Demo Mode
Exclusive Advantages Pre-define and preview the calibration data and curve Self-edit limited line and support two limited line simultaneously Support inconsecutive segment sweep function Support both conducted and radiated inspection Support peak table and marker table User-friendly testing report generation Support two traces graphical comparison Demo Mode offering
User Benefits Easily observe the testing data via PC Laptop Intuitive testing and calibration concepts to implement different projects. Fit multiple EMI international standards via self-edited limit line Clear testing result and logical report Easily self-adjusted graphical environment to fit users observation Apparent and detailed analysis logistics and methodologies.
Demo Procedure
Required Equipment List 1. DSA875/832/832E/815/710/705 Spectrum Analyzer 2. EMI-DSA800 Option 3. Ultra Sigma Software 4. S1210 EMI Pre-Compliance Test Software 5. NFP-3 Near Field Probe ( For Radiated Test Only ) 6. USB A Type to B Type Cable x 1 7. Personal Computer ( With WIN10/8/7/Vista System ) 8. EMI DUT Source
Spectrum Analyzer Pre-Setup Use A to B Type USB cable to connect Spectrum Analyzer and PC A Type Connector B Type Connector
Spectrum Analyzer Pre-Setup Use NFP-3 Near Field Probe and connect to DSA800 Spectrum Analyzer NFP-3 Near Field Probe Set Connect to DSA800
Spectrum Analyzer Setup 1. Press Preset on DSA800 firstly to original settings
Spectrum Analyzer Setup 2. Press FREQ and Setup Start (30MHz), Stop (1GHz) frequency on DSA800 3. Press BW/Det and adjust the RBW value to 100KHz
Spectrum Analyzer Setup 4. Press AMPT on DSA800, and Set the Vertical Axis Units to dbuv
Spectrum Analyzer Setup 5. Start the S1210 EMI Software 6. Press File and press New to open a project test_0
Spectrum Analyzer Setup 7. Select test_0 and go to Device to connect PC to Spectrum, as you can see the green USB Address will be shown on the bottom side of the screen
Spectrum Analyzer Setup 8. Go to Scan Config and select the Limited Line 1 file ( EN55011_CAG1_Radiated_Qpeak_10m.lim ) 9. Set the detector as Peak, RBW value to 100KHz, and select Clear/Write when complete 8 9
Spectrum Analyzer Setup 10. Go to Segment Config and Setup the frequency sweeping segment Start from 30MHz to 1GHz, RBW to 100KHz, Sweep time to 0 sec.
Spectrum Analyzer Setup 11. Use NFP-3 Near Field Probe to approach the left side of DSA800 LCD Display. ( To inspect the EMI source from CCFL Adaptor of LCD Back light module)
Spectrum Analyzer Setup 12. Go to Measure and implement the Pre-Scan 13. The EMI Spectral Lines are shown on the screen
Back up
S1210 EMI Test Solution Double click the mouse on the graphic and the markers will shown on the screen and the marker table as well.
S1210 EMI Test Solution Go to Set Reference and users can do two trace comparison
S1210 EMI Test Solution Users can Zoom In the dedicated peak and observe inside
S1210 EMI Test Solution The Zoom In Spectral Lines
How to edit the Limit Line?
How to edit the limit line? 1. Press the Data Manager at the top of the display
How to edit the limit line? 2. Select the limit.lim file
How to edit the limit line? 3. Edit the limit line parameters at right side window.
How to edit the limit line? 4. Export and save the file
How to edit the limit line? 5. Select one folder to export and save file
How to edit the limit line? 6. User can see the limit line file inside and folder and can import to S1210 Software by Scan Config setup item.
THANK YOU