CHAPTER 4: HIGH ENERGY X-RAY GENERATORS: LINEAR ACCELERATORS. Jason Matney, MS, PhD

Similar documents
RF Power Generation II

Lecture 17 Microwave Tubes: Part I

Low-Energy Electron Linacs and Their Applications in Cargo Inspection

Detailed Design Report

CHAPTER 4 OSCILLOSCOPES

Design and Simulation of High Power RF Modulated Triode Electron Gun. A. Poursaleh

III. Proton-therapytherapy. Rome SB - 3/5 1

Dark current and multipacting trajectories simulations for the RF Photo Gun at PITZ

4.4 Injector Linear Accelerator

Klystron Tubes. Two forms of such a device, also called linear beam klystron, are given in the following figure.

UNIT-3 Part A. 2. What is radio sonde? [ N/D-16]

DEVELOPMENT OF A 10 MW SHEET BEAM KLYSTRON FOR THE ILC*

RF Power Klystrons & 20 Year Look. R. Nelson 7/15/15

Tutorial: Trak design of an electron injector for a coupled-cavity linear accelerator

Design, Fabrication and Testing of Gun-Collector Test Module for 6 MW Peak, 24 kw Average Power, S-Band Klystron

CHAPTER 3 OSCILLOSCOPES AND SIGNAL GENERATOR

PEP II Design Outline

CATHODE-RAY OSCILLOSCOPE (CRO)

These are used for producing a narrow and sharply focus beam of electrons.

The FLASH objective: SASE between 60 and 13 nm

The Cathode Ray Tube

Teltron Delection Tube D

Particle Beam Production - A Synchrotron-Based System - Prof. Dr. Thomas Haberer Scientific-technical Director Heidelberg Iontherapy Center

Commissioning the TAMUTRAP RFQ cooler/buncher. E. Bennett, R. Burch, B. Fenker, M. Mehlman, D. Melconian, and P.D. Shidling

A HIGH POWER LONG PULSE HIGH EFFICIENCY MULTI BEAM KLYSTRON

Elements of a Television System

DOSE DELIVERY SYSTEM OF THE VARIAN PROBEAM SYSTEM WITH CONTINUOUS BEAM

Development of Multiple Beam Guns for High Power RF Sources for Accelerators and Colliders

Available online Journal of Scientific and Engineering Research, 2018, 5(5): Research Article

THE OPERATION OF A CATHODE RAY TUBE

DEPARTMENT OF ELECTRONICS AND COMMUNICATION ENGINEERING QUESTION BANK

CATHODE RAY OSCILLOSCOPE (CRO)

TECHNICAL SPECIFICATION Multi-beam S-band Klystron type BT267

Production of quasi-monochromatic MeV photon in a synchrotron radiation facility

Linac 4 Instrumentation K.Hanke CERN

INSTRUMENT CATHODE-RAY TUBE

DESIGN AND TECHNOLOGICAL ASPECTS OF KLYSTRON DEVELOPMENT

THE OPERATION OF A CATHODE RAY TUBE

Low Frequency Gyrotrons for Fusion

RF Solutions for Science.

INFN School on Electron Accelerators. RF Power Sources and Distribution

Development of High Power Vacuum Tubes for Accelerators and Plasma Heating

Development of high power gyrotron and EC technologies for ITER

CATHODE RAY OSCILLOSCOPE. Basic block diagrams Principle of operation Measurement of voltage, current and frequency

IOT RF Power Sources for Pulsed and CW Linacs

Recent ITER-Relevant Gyrotron Tests

DESIGN AND PERFORMANCE OF L-BAND AND S-BAND MULTI BEAM KLYSTRONS

These tests will be repeated for different anode positions. Radiofrequency interaction measurements will be made subsequently. A.

Pulsed Klystrons for Next Generation Neutron Sources Edward L. Eisen - CPI, Inc. Palo Alto, CA, USA

Low-Noise, High-Efficiency and High-Quality Magnetron for Microwave Oven

Current status of XFEL/SPring-8 project and SCSS test accelerator

3 cerl. 3-1 cerl Overview. 3-2 High-brightness DC Photocathode Gun and Gun Test Beamline

Present Status and Future Upgrade of KEKB Injector Linac

3B SCIENTIFIC PHYSICS

Television Projects of Fernseh GmbH (Int. Dr. Möller)

High Brightness Injector Development and ERL Planning at Cornell. Charlie Sinclair Cornell University Laboratory for Elementary-Particle Physics

Implementing a Proton Beam Scanning System within an Operating Clinical Facility

The field cage for a large TPC prototype

KLYSTRON GUN ARCING AND MODULATOR PROTECTION

The Elettra Storage Ring and Top-Up Operation

SLAC-PUB-2380 August 1979 (A)

Ride the Lightning. A Review of Radiographic Physics

L14 - Video. L14: Spring 2005 Introductory Digital Systems Laboratory

OSCILLOSCOPE AND DIGITAL MULTIMETER

CPI Gyrotrons For Fusion EC Heating

Experimental Results of the Coaxial Multipactor Experiment. T.P. Graves, B. LaBombard, S.J. Wukitch, I.H. Hutchinson PSFC-MIT

PHGN 480 Laser Physics Lab 4: HeNe resonator mode properties 1. Observation of higher-order modes:

LASERTRON SIMULATION WITH A TWO-GAP OUTPUT CAVITY*

Jefferson Lab Experience with Beam Halo, Beam Loss, etc.

The PEFP 20-MeV Proton Linear Accelerator

Pseudospark-sourced Micro-sized Electron Beams for High Frequency klystron Applications

Basic rules for the design of RF Controls in High Intensity Proton Linacs. Particularities of proton linacs wrt electron linacs

TITLE PAGE. Title of paper: PUSH-PULL FEL, A NEW ERL CONCEPT Author: Andrew Hutton. Author Affiliation: Jefferson Lab. Requested Proceedings:

Activities on FEL Development and Application at Kyoto University

RF plans for ESS. Morten Jensen. ESLS-RF 2013 Berlin

Brief Description of Circuit Functions. The brief ckt. description of V20 107E5 17 Monitor

Evaluation of Performance, Reliability, and Risk for High Peak Power RF Sources from S-band through X-band for Advanced Accelerator Applications

North Damping Ring RF

IOT OPERATIONAL EXPERIENCE ON ALICE AND EMMA AT DARESBURY LABORATORY

Studies on an S-band bunching system with hybrid buncher

PEP-I1 RF Feedback System Simulation

LCLS RF Reference and Control R. Akre Last Update Sector 0 RF and Timing Systems

DEVELOPMENT OF X-BAND KLYSTRON TECHNOLOGY AT SLAC

Operating Experience and Reliability Improvements on the 5 kw CW Klystron at Jefferson Lab

INSTRUMENT CATHODE-RAY TUBE

SRS and ERLP developments. Andrew moss

4.9 BEAM BLANKING AND PULSING OPTIONS

14 GHz, 2.2 kw KLYSTRON GENERATOR GKP 22KP 14GHz WR62 3x400V

TOSHIBA Industrial Magnetron E3328

FIR Center Report. Development of Feedback Control Scheme for the Stabilization of Gyrotron Output Power

K800 RF AMPLIFIER TUBE UPGRADE

Experience with the Cornell ERL Injector SRF Cryomodule during High Beam Current Operation

Screen investigations for low energetic electron beams at PITZ

Defense Technical Information Center Compilation Part Notice

Svetlana 3CX10,000A7/8160

TWO BUNCHES WITH NS-SEPARATION WITH LCLS*

Brown, A., Merkert, J., & Wilson, R. (2014). Build your own particle accelerator. Science in School, (30),

Investigation of Radio Frequency Breakdown in Fusion Experiments

RECENT PROGRESS IN UPGRADE OF THE HIGH INTENSITY THzzz zz-fel AT OzSAKzA UNIVERSITYzzzz

DELIVERY RECORD. Location: Ibaraki, Japan

Transcription:

CHAPTER 4: HIGH ENERGY X-RAY GENERATORS: LINEAR ACCELERATORS Jason Matney, MS, PhD

Objectives Medical electron linear accelerators (often shortened to LINAC) The Basics Power Supply Magnetron/Klystron Accelerator structure Standing-Wave/Traveling Wave X-Ray vs. e- Therapy Modes

Remember Bremsstrahlung Production? Electrons are released from cathode and travel to the anode, accelerated in route and attain K.E. as they drop through the potential difference. As the e- passes in the vicinity of positively charged target all or part of the electron s energy is dissociated from it and propagates in space as EM radiation X-ray Energy ranges from 0-KvP (max accelerating potential)

Why do we need a LINAC? Range of X-ray Energy 0-KvP (max accelerating potential). Khan, Figure 3.9 If you want higher energy X-rays You need to hit the target with faster electrons This is the purpose of the accelerator

Depth-Dose Relationship for Different Energy Photons 60 Co (d max =0.5cm) 1.17 & 1.33 MeV % DD = D D d d max For lower energies PDD is close to 100% at surface 100 As increase energy 100% is at deeper depths Increase Energy, Increase skin sparing kev -> MeV energy: even more Skin Sparing <30kV Grentz Rays orthovoltage 100-500 kv Superficial 10-100 kv Contact Therapy

Depth-Dose Relationship for Different Energy Photons kev to MEV: Increase Energy, Increase skin sparing, deeper penetration % DD = D D d d max 100

Cutaway Diagram of Varian Linac Khan, Figure 4.8C

Microwave Power = Photons! E=hv h=6.626*10-34 J*s v = frequency (1/s) Microwaves have very High Frequencies High Energy ev = 1.6*10 19 J c = λν

Resonance Resonance causes an object to move back and forth or up and down. This motion is generally called oscillation. When the force application frequency matches the natural frequency of an object it will begin to resonate. The forcing function adds energy at just the right moment during the oscillation cycle so that the oscillation is reinforced. This makes the oscillation's amplitude grow larger and larger. It s like pushing a swing lots of little pushes at the right time

Resonance In Radiation therapy microwave devices make extensive use of resonant microwave cavities: Magnetrons Klystrons Accelerator Structure The resonance Phenomenon occurs at 3000MHz (corresponding to a 10cm wavelength), which is determined by the dimensions of the cavity. Cavities are formed of copper for high electrical and thermal conductivity.

Typical Medical LINAC Khan, Figure 4.5 Our Goal To understand all parts of this LINAC diagram!

LINAC Components: Varian Clinac Karzmark and Morton, Figure 15

Basic LINAC Components 1. Modulator Simultaneously provides high voltage DC pulses to the magnetron or klystron and e - gun. 2. Magnetron/Klystron provides high frequency microwaves. 3. Electron Gun cathode that provides source of electrons injected into accelerator structure. 4. Wave Guide Carries microwave power from magnetron or klystron through the accelerator structure. 5. Accelerator Structure Accelerates e - s from an electron gun using microwave power from magnetron or klystron. 6. Treatment Head Directs, collimates, shapes, and monitors the treatment beam

Typical Medical LINAC Khan, Figure 4.5 Magnetron/Klystron provides high frequency microwaves.

The Magnetron Device that PRODUCES microwaves Cylindrical construction w/central cathode and outer anode w/resonant cavities Khan, Figure 4.6

Magnetron The Steps to Generate Microwaves 1. Cathode heated by inner filament. e- released by thermionic emission. 2. Static magnetic field applied perpendicular to the plane of cavities crosssection. 3. Pulsed DC electric field applied between anode and cathode. accelerating power Khan, Figure 4.6c

Magnetron The Results The e-s released from cathode are accelerated toward the anode by the action of pulsed DC electric field. e-s move in complex spirals toward the resonant cavities (due to influence of magnetic field), radiating energy in the form of microwaves. Microwaves are lead to accelerator structure via waveguide. Resonant Frequency 3000MHz

Klystron Klystron = Microwave Amplifier driven by low power microwave oscillator. Different from magnetron, does NOT generate microwaves.

Klystron The steps of Microwave Amplification 1. Cathode heated by hot wire e-s released via thermionic emission. 2. Low level microwaves injected into buncher cavity. Setup alternating electric field across the gap between left and right cavity walls 3. Velocity Modulation Let s talk about this step in detail!!

Klystron-The steps of microwave amplification Step 3. Velocity Modulation More Details a b c - Electric Field Accelerates e - s + Electric Field Decelerates e - s Fast electrons that arrive in buncher cavity early, between points a and b, encounter a retarding Electric Field Slowed Electrons that arrive at time b, when Electric Field=0 velocity unaffected Slower e-s that arrive at buncher cavity later, between b and c encounter accelerating Electric Field Accelerated Net Effect e- stream forms into bunches groups of e - s traveling at same velocity.

Klystron The Rest of the Steps for Microwave Amplification 4. Drift tube - Distance along which electrons moving at different velocities merge into discrete bunches. 5. Catcher Cavity - As e-s leave drift tube and traverse catcher cavity gap, they generate a retarding electric field (like charges repel) at the ends of the cavity, initiating energy conversion process. the e-s kinetic energy is converted to EM radiation microwaves.

Klystron The Rest of the Steps for Microwave Amplification 6. Microwaves are lead to accelerator structure via wave guide. 7. Collector: Residual beam energy that is not converted to microwave power is dissipated as heat in collector cavity (there are also some x- rays, thus encased in shielding structure).

Klystron Klystron = Microwave Amplifier driven by low power microwave oscillator. Different from magnetron, does NOT generate microwaves.

Magnetron verses Klystron Magnetron 1. Used in Elekta 2. Circular Geometry 3. Typical Peak power 2MW (5MW in newer references) 4. Produces high frequency microwaves 5. Sends high frequency microwaves to accelerator tube via wave guide Klystron 1. Used in Varian and Siemens 2. Linear Geometry 3. Higher Peak Power 5MW (7MW in newer references) 4. Amplifies Low frequency microwaves, resulting in high frequency microwaves 5. Sends high frequency microwaves to accelerator tube via wave guide Same Result High Frequency microwaves to Accelerator Tube

Typical Medical LINAC Let s discuss the accelerator tube in more detail.

Accelerator Tube Once we have microwave accelerating power (from either magnatron or klystron), we need a place to accelerate electrons: Accelerator Tube (Two Types) Traveling Wave Standing Wave Don t lose site of the goal: Accelerator Structure Accelerates electrons from an electron gun using microwave power from magnetron or klystron

Traveling Wave-The Basics High frequency microwaves are transmitted down evacuated tube through accelerating cavities ( 3000 MHz). A pre-buncher is used to reduce the velocity of EM wave to correspond to speed of injected e - s. e - s travel at crest of wave and undergo acceleration. EM waves are absorbed at end of accelerator to prevent backward reflected wave (which would interact with incoming waves).

Traveling Wave Surfer Analogy Karzmark and Morton, Figure 28 Traveling Wave Principle a. A boy surfing on water wave, accelerates to the right. b. e-s occupying a similar position on an advancing negative E field. c. Associated charge distribution that pushes (- charge) and pulls (+ charge) the e- bunches along the cylinder 4 cavities per l

Standing Wave-The Basics 1. Microwave power is fed into the structure via input wave guide at proximal end (e- gun end). 2. Incident wave reflected backward. 3. Now we have 2 waves: Incident wave Reflected wave 4. The 2 waves are reflected back and forth from one end of the accelerator tube to the other many times. 5. The effective electric field is the sum of the forward and backward waves. Magnitude of E fields is additive when the forward and backward waves are in same direction. Magnitude of E field cancels out when forward and backward waves are in opposing directions.

Creating a Standing Wave Standing Wave Produced when 2 traveling waves of equal period and amplitude travel through waveguide in opposite directions! In accelerator the forward and backward waves exist simultaneously, shown separately here for clarity. Karzmark and Morton, Figure 29

Standing Wave Standing wave E field patterns in an accelerator structure for combined forward and backward waves @ three instances in time. Every other cavity zero electric field essential in transporting microwave power, but do not contribute to e- acceleration. Role=transfer and couple power bet cavities Can be moved off axis to shorten length of cavity One in 4 cavities negative. Karzmark and Morton, Figure 30

Bimodal (Side-Coupled) Accelerator Optimize cavities along the beam axis for acceleration Optimize off-axis cavities for microwave power transport Every other cavity zero electric fieldessential in transporting microwave power, but do not contribute to e- acceleration Role=transfer and couple power bet cavities Can be moved off axis to shorten length of cavity Result shorter Accelerator Structure, less space required in Rx room.

Accelerator Tube Example Elekta linac accelerator tube

Traveling Wave vs Standing Wave Standing Wave Traveling Wave More stable Less stable More expensive Shorter accelerator structure Due to bimodal configuration Manufacturers Varian Siemens Less expensive Longer accelerator structure Manufacturers Elekta Usually used with Magnetron Usually used with Klystron Note: Single Energy (6X) standing wave accelerators often use Magnetrons

What about LINACs with more than one beam energy? Changing the beam energy will be different for traveling-wave and standing-wave accelerators.

Dual X-ray Energy Mode A Summary Standing-wave 1. Change ratio of microwaves fed into first and second portions of cavity Energy Switch: High-power Microwave Circuit 2. Broad band buncher. Traveling Wave 1. Beam Loading Karzmark, pg 26-28 Increasing injected beam current from gun and keeping magnetron or klystron power constant Remember: Wave travels only in forward direction Amplitude of the E field in the second portion of accelerator structure can be changed without significant effect on first portion (capture and buncher)

One more step.. Where are we aiming the beam?

First things first Where is the beam aimed Low energy, 6MV (standing wave) LINACs Relatively short accelerator tube is vertically mounted, electrons proceed straight from accelerator tube to target. High energy 6 MV (standing Wave) LINACs Accelerator tube too long to vertically mount, so horizontally mount the accelerator tube. But now we have a problem we have horizontal beam, BUT we want vertical beam!! The solution = Bending Magnets

Bending Magnet 270 degrees Deflects beam in 270 o loop. Lower energy e - s are deflected through smaller radius loop. Higher energy e - s are deflected through larger radius loop. All components brought back together at same position, angle, and beam cross section as when they left accelerator structure

Bending Magnet Slalom Style Figure is from: Tim Waldron The Theory and Operation of Computer-Controlled Medical linear Accelerators". 2002 AAPM Annual Meeting, Refresher Course. MO-A-517A-01 7/15/02

Now that we have that high energy e - Beam, How do we make it useful for Rx? What are we going to hit with it? Do we want to treat with photons or electrons?

LINAC in X-Ray Mode Khan, Figure 4.8A X-Ray Mode Target 1 st collimation fixed, non-variable Flattening Filter Monitor ionization chambers Field defining Light 2 nd Collimator System Multi Leaf collimators

LINAC in X-Ray Mode Khan, Figure 4.8A X-Ray Mode Target

Flattening Filter e- High energy x-rays are forward peaked, max intensity located centrally with decreased intensity at periphery. Flattening filter thick at center thin at edges Target Forward peaked x- ray beam Attenuates more photons from center less from edges: Flatter beam, covers more area Flattened X-Ray Beam

LINAC in X-Ray Mode Khan, Figure 4.8A X-Ray Mode Monitor ionization chambers

Monitor Ionization Chamber Dose Rate of accelerator beam may very unpredictable. Therefore, cannot to rely on elapsed time to control dose delivered to patient. Radiation leaving target or scattering foil passes through monitor ionization chamber. Radiation produces ionization current proportional to beam Intensity. Current is converted to monitor units or MU Purpose of monitor ionization chamber: Dose to patient controlled by programming accelerator to deliver prescribed # MU. Monitor field symmetry.

LINAC Light Field To help see the projected field on our patients Mirror slides into place for visual verification, moves out of beam path when in use

Multileaf collimators Devices using thick leaves of High Z material that help shape the exiting photon field laterally.

Multileaf collimators

LINAC in e- Mode Khan, Figure 4.8B Electron Mode Scattering Foil Monitor ionization chambers Field defining Light Collimator System 2 nd e - collimation = cone, downstream of photon collimators placed close to skin to dec. e- scatter in air 3 rd e - collimation = cutout downstream of cone custom shape field to match Rx site Note: Beam flatness depends on design of cone or applicator. e - mode NO TARGET!!

LINAC: e- Mode Scattering Foils Electron beam exits as narrow pencil beam with a small diameter. Need to spread it out to be clinically useful. Effect of a single foil Gaussian distribution Unusable beam Dual Foil System 2 High Z Foils separated by air gap 2nd foil thicker at center to even distribution that results after beam passes through 1 st foil Note: thin foil, causes electrons to scatter i.e. spread out, thicker foils would result Brems. x-rays.

Electron Applicators or Cones Electrons are more likely to scatter in air and require additional, closer collimation Final field shaping of electron field by applicator and special insert Lead insert custom shaped to specific treatment field size Cerrobend: 50% Bi, 27% Pb, 13% Sn, 10% Cd

Beam Current Photon verses Electron Mode Beam current - number of electrons ejected from electron gun. Beam current is much higher in photon mode. Recall: Bremsstrahlung production is inefficient. Some of the photons are attenuated in flattening filter.

Cutaway Diagram of Varian Linac Khan, Figure 4.8C

Practice Questions If that was not too painful

Question 1: The purpose of a scattering foil in the electron mode of LINAC is to : A. Absorb scattered electrons. B. Change x-ray beam into electrons. C. Shield the ion chambers. D. Absorb excess RF energy E. Broaden and flatten the beam. Correct Answer: E. Changes narrow beam into a broader beam by forcing the electron path to spread due to collisions in the foil!

Question 1: The purpose of a scattering foil in the electron mode of LINAC is to : A. Absorb scattered electrons. B. Change x-ray beam into electrons. C. Shield the ion chambers. D. Absorb excess RF energy E. Broaden and flatten the beam. Correct Answer: E. Changes narrow beam into a broader beam by forcing the electron path to spread due to collisions in the foil!

Question 2 Which of the following does NOT occur when LINAC is changed from x-ray mode to electron mode? (excluding units with scanned electron beams). A. The target is removed. B. Scattering foil is placed in the beam. C. The monitor chamber is removed. D. An electron applicator is attached. E. The beam current decreases. Correct Answer: C. Monitor chamber NOT removed must monitor beam output for all modalities

Question 2 Which of the following does NOT occur when LINAC is changed from x-ray mode to electron mode? (excluding units with scanned electron beams). A. The target is removed. B. Scattering foil is placed in the beam. C. The monitor chamber is removed. D. An electron applicator is attached. E. The beam current decreases. Correct Answer: C. Monitor chamber NOT removed must monitor beam output for all modalities

Question 3 Klystrons and Magnetrons are: A. Devices used to bend beams of electrons. B. Located in treatment head of the LINAC. C. Beam focusing devices. D. Sources of microwave power. E. Part of the LINAC timer circuit. Correct Answer: D. Sources of microwave power (Magnetrons generate microwaves, klystrons amplify microwaves)

Question 3 Klystrons and Magnetrons are: A. Devices used to bend beams of electrons. B. Located in treatment head of the LINAC. C. Beam focusing devices. D. Sources of microwave power. E. Part of the LINAC timer circuit. Correct Answer: D. Sources of microwave power (Magnetrons generate microwaves, klystrons amplify microwaves)

Question 4 With regard to the production of electron beams by LINACs, which of the following is TRUE? 1. The Beam current is much higher in electron mode than in photon mode No photon production inefficient, need more current 2. Beam flatness depends on design of cone or applicator. TRUE 3. The bending magnet is rotated out of the beam when electrons are selected. No required to point electrons down toward isocenter. 4. Thick scattering foils can reduce bremsstrahlung. No thicker foils result in more interactions and produce more x-rays! 5. All of the above

Question 4 With regard to the production of electron beams by LINACs, which of the following is TRUE? 1. The Beam current is much higher in electron mode than in photon mode No photon production inefficient, need more current 2. Beam flatness depends on design of cone or applicator. TRUE 3. The bending magnet is rotated out of the beam when electrons are selected. No required to point electrons down toward isocenter. 4. Thick scattering foils can reduce bremsstrahlung. No thicker foils result in more interactions and produce more x-rays! 5. All of the above

References 1. The Physics of Radiation Therapy 4 th Ed., Faiz Khan 2. A Primer on Theory and Operation of Linear Accelerators in Radiation Therapy, C.J. Karzmark and Robert Morton 3. Linear Accelerators for Radiation Therapy, 2 nd edition, D. Greene and P.C. Williams 4. Radiation Therapy Physics, 2 nd edition, William Hende and Geoffrey Ibbott 5. Theory and Operation of Computer Controlled Medical Linear Accelerators, Handout form 2002 AAPM Refresher course, Tim Waldron