IMSERC NMR MANUAL 05: Manual Operation of Agilent NMR Spectrometers (Chem350 Interface)

Similar documents
Processing data with Mestrelab Mnova

PINMRF. Checkout Quiz - Varian Inova-300 Version

Vnmrj Training Guide. Vnmrj overview

MestReNova Manual for Chem 201/202. October, 2015.

Introduction to 1D & 2D NMR using VnmrJ

MestReNova A quick Guide. Adjust signal intensity Use scroll wheel. Zoomen Z

Procedure for Acquiring 2D Inverse Detection Spectra HMQC, HMBC and HSQC

Basic 13 C Acquisition and Processing 4

Practicum 3, Fall 2012

Standard Operating Procedure of nanoir2-s

Practicum 3, Fall 2010

Quick Start Bruker Dimension Icon AFM

Class Notes for Cite While You Write Basics. EndNote Training

GS122-2L. About the speakers:

1 H HOMODECOUPLING EXPERIMENT

Analyzing and Saving a Signal

Lab experience 1: Introduction to LabView

Dektak Step by Step Instructions:

Character Users Guide

Defining and Labeling Circuits and Electrical Phasing in PLS-CADD

USB Mini Spectrum Analyzer User s Guide TSA5G35

NMR TRAINING MANUAL LINUX-BASED SPECTROMETERS. And VnmrJ 2.x TEXAS A&M UNIVERSITY DEPARTMENT OF CHEMISTRY

Figure 1. MFP-3D software tray

Introduction to EndNote Desktop

BME 3512 Biomedical Laboratory Equipment List

ME EN 363 ELEMENTARY INSTRUMENTATION Lab: Basic Lab Instruments and Data Acquisition

CMRRpack2. a collection of pulse sequences from the. University of Minnesota, Center for Magnetic Resonance Research (CMRR) Version: 2.

1.0 ThermoNicolet Nexus 670 FTIR Spectrometer Instructions

Using the Agilent for Single Crystal Work

AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode)

Contacts: English Department Office 238 Moreland Media Services 109 Kidder

Classical simulations with PJNMR BCMB/CHEM LAB 01/25/12

PHI 5000 VersaProbe TM Operator s Guide

NMR picospin User Guide

PHY221 Lab 1 Discovering Motion: Introduction to Logger Pro and the Motion Detector; Motion with Constant Velocity

NMR. picospin. Maintenance Guide

McIDAS-V Tutorial Using HYDRA to Interrogate Hyperspectral Data updated September 2015 (software version 1.5)

UCI NMR MANUAL DRX400 GN500 CRYO500 AVANCE600

System Requirements SA0314 Spectrum analyzer:

Operating Instructions for Scanmark insight 4ES Scanner with QuickScore

MIE 402: WORKSHOP ON DATA ACQUISITION AND SIGNAL PROCESSING Spring 2003

imso-104 Manual Revised August 5, 2011

iworx Sample Lab Experiment HM-3: The Electrogastrogram (EGG) and the Growling Stomach

How to Build A Table of Authorities in Word * By: Morgan Otway

Reference. TDS7000 Series Digital Phosphor Oscilloscopes

EndNote for Mac. User Guide. UTS Library University of Technology Sydney UTS CRICOS PROVIDER CODE 00099F

3 CHOPS - LIP SYNCHING

Standard Operating Procedure II: EDS (Bruker Flat-Quad)

Linrad On-Screen Controls K1JT

FACSAria I Standard Operation Protocol Basic Operation

Import and quantification of a micro titer plate image

Vasudevan Agilent 1100 Series HPLC w/ DAD & FLD Detector (nonbuffer

Cisco Spectrum Expert Software Overview

Topic: Instructional David G. Thomas December 23, 2015

Game Show Physiology

Experiment HM-8: The Electrogastrogram (EGG) and the Growling Stomach

Introduction to EndNote Online

Please feel free to download the Demo application software from analogarts.com to help you follow this seminar.

SPM Training Manual Veeco Bioscope II NIFTI-NUANCE Center Northwestern University

FIB Operating Procedure. Effective Date: 08/14/2012 Author(s): Jiong Hua Phone:

Swinburne University of Technology

TL-2900 AMMONIA & NITRATE ANALYZER DUAL CHANNEL

The BAT WAVE ANALYZER project

Amazon Account 3/14 TM 1

Universal Mode. Survey Settings. The Universal Toolbar. First Run - Initial Configuration via the Setup Button

Waters GCT GCMS Training Manual (Nominal Mass GCMS version) 02/06/2012 S.V.

Kratos AXIS Ultra DLD X-ray Photoelectron Spectrometer Instructions

Getting Started with the LabVIEW Sound and Vibration Toolkit

USB Mini Spectrum Analyzer User Manual PC program TSA For TSA5G35 TSA4G1 TSA6G1 TSA12G5

Exercise #1: Create and Revise a Smart Group

TT AFM LongBeach Procedures and Protocols V2.1

SR-D8-M, SR-D8-S. (Ver ) SOFTWARE INSTRUCTIONS

Fall 2014 Chem 636 Lab #4

EndNote Tutorial Handout Table of Contents

Reason Overview3. Reason Overview

***Please be aware that there are some issues of compatibility between all current versions of EndNote and macos Sierra (version 10.12).

Effective Test Procedures for Installing and Maintaining RF Transmitter Sites

MultiQ Digital signage template system for widescreen monitors

Application Note AN-708 Vibration Measurements with the Vibration Synchronization Module

cyclenoe 1D NOE Difference Experiment

Getting started with EndNote X7

Ultra 4K Tool Box. Version Release Note

DektakXT Profilometer. Standard Operating Procedure

Gustavus Adolphus College. Some Scientific Software of Interest

Optiflex Interactive Video System

USB Mini Spectrum Analyzer User Manual TSA Program for PC TSA4G1 TSA6G1 TSA8G1

Word Tutorial 2: Editing and Formatting a Document

SmartScore Quick Tour

STANDARD OPERATING PROCEDURE: ASYLUM MFP-3D AFM

Library ebooks and Your Sony ereader

The Kaffeine Handbook. Jürgen Kofler Christophe Thommeret Mauro Carvalho Chehab

EndNote for Mac. EndNote for PC. User Guide. UTS Library University of Technology Sydney UTS CRICOS PROVIDER CODE 00099F

Procedures for conducting User QA on the scanner

Precision DeEsser Users Guide

ExtIO Plugin User Guide

SEM- EDS Instruction Manual

SOUND LABORATORY LING123: SOUND AND COMMUNICATION

Using EndNote Online: health

Manual for TV software. TT-Viewer version Figure: TT-budget S2-3200

Operating Guide. ViewClix offers a revolutionary experience for seniors and their families and friends.

Transcription:

IMSERC NMR MANUAL 05: Manual Operation of Agilent NMR Spectrometers (Chem350 Interface) Last updated: October 12, 2011 by Josh Kurutz THIS PAGE = QUICK START GUIDE 0) At the computer, make sure VNMRJ is running under the account named northrup, and that the instrument is locked & not acquiring data 1) Load a new PROTON experiment by pulling down Experiments Proton from the main menu bar 2) Eject the CDCl 3 ( lock ) standard sample from the Start/Standard panel by clicking Eject 3) Remove the sample from the magnet 4) Carefully load your sample into a spinner using the sample positioner 5) Place your sample on top of the magnet so it floats on the (loud) eject air. On Au400, you will need to push it past the silicone grabbers, holding the spinner, not the sample tube. 6) Insert your sample by clicking Insert on the Start/Standard panel 7) Select your solvent in Start/Standard pulldown menu 8) Retrieve the standard shims by clicking the light blue retrieve default shims button. 9) Lock on the sample with the manual locking procedure in the Start/Lock tab. 10) Shim on the sample using the manual shimming procedure. 11) Make sure your sample is locked and that the lock level is between ~15 and 90. 12) Move to the Acquire/Default H1 panel by clicking on the Acquire panel s tab and then the Default panel title. 13) Set the number of scans to 1 and acquire a preliminary 1D 1 H spectrum by clicking the green Acquire button. 14) Evaluate the spectrum examine spectral width, phasing, linewidth, referencing, etc., and make necessary changes. 15) If necessary, acquire more scans; in the Acquire/Default H1 panel, pull down the number of scans pulldown to the desired number of scans a. NB: Keep in mind that S/N improves with the square root of the number of scans; increasing them from 4 to 8 will increase S/N by a factor of ~1.4, but from 4 to 16 will improve it by a factor of 2. 16) Save your spectrum by clicking of the floppy disk save button 17) Continue with further experiments and other samples if you have them 18) When you are done with your last sample, eject it using the Eject button 19) Insert the lock standard using the Insert button in the Start/Standard panel 20) Sign out from the pen-and-paper logbook, noting any instrument problems 21) Process your spectrum further and perhaps print it out using an offline processing computer 1

2

Manual Operation of Agilent NMR Spectrometers (Chem350 Interface) Setup and Acquisition of a 1D 1 H NMR spectrum Step 0) At the computer, make sure VNMRJ is running, and that the instrument is locked & not acquiring data If necessary, wake up the computer by moving the mouse or typing a key on the keyboard (preferably the spacebar). If you are unfamiliar with the layout of the VNMRJ window, you should take a minute to orient yourself: Figure 1. Orientation to the VnmrJ window First, navigate to the Start/Standard panel, indicated in the lower half of the VNMRJ window. 1

Toward the upper left corner of the Start/Standard panel, locate the Operator designation. For Chem350, it should read northrup. To determine whether the instrument is locked, look in the lower-left corner of the VNMRJ panel (near the trash can) and read the number under the word Lock. If the number is blue and above ~15, the system is locked. Figure 2. Operator designation & lock status * PROBLEM? If the instrument is not locked on the lock standard when you walk up to the instrument, either the lock circuit is not engaged, there is no sample in the magnet, or the shims are too poor to support locking. Proceed normally, but when you try to eject the sample that should be in the magnet, make sure that a sample emerges. If no sample comes out, contact staff immediately. Usually when you walk up to the instrument, the VNMRJ program is running. If it is not, you ll just see the underlying Linux desktop for the login account northrup (Figure 3): Figure 3. Essential parts of the linux desktop If VNMRJ is not running, start it by clicking on the VNMRJ icon on the Linux desktop (Figure 3). 2

Step 1) Load a new PROTON experiment by pulling down Experiments Proton from the VNMRJ main menu. You can see that the spectrometer offers a large number of experiments, but the Proton experiment is the choice on top. Figure 4. Start a new 1H 1D experiment using fresh parameters by pulling down the VnmrJ Manin Menu's pulldown labeled "Experiments" to the "proton" choice. 3

After selecting the Proton experiment, your main display window should show you a diagram of the pulse sequence for the PROTON spectrum. Make sure you are viewing the correct panel in the lower part of the display; click on the Start tab and the Standard panel label. Figure 5. The Proton pulse sequence and parameter interface. Step 2) Eject the sample from the Start/Standard panel by clicking Eject Figure 6. The eject button You should see the lock standard floating on top of the magnet. The lock standard is simply a sample of CDCl 3 in a flamed-sealed tube with a red-painted top. IF NOTHING COMES OUT OF THE MAGNET WHEN YOU CLICK EJECT, GET HELP FROM FACILITY STAFF IMMEDIATELY. 4 Figure 7. The Lock Standard floating atop the magnet

Step 3) Remove the lock standard from the magnet. Be careful to lift it all the way out before bringing it toward you. The most common way for people to break samples in the magnet it to snap off the tube below the spinner as they remove the sample from the magnet. Figure 8. Carefully removing the sample from atop the magnet. IF YOU EVER BREAK A SAMPLE IN OR NEAR THE MAGNET, YOU MUST ALERT IMSERC STAFF RIGHT AWAY! IF YOU BREAK A SAMPLE IN THE MAGNET BETWEEN 8 AM AND 10 PM, YOU MUST IMMEDIATELY CONTACT JOSH OR YUYANG IN PERSON OR BY PHONE AND EMAIL AND PLACE A STOP SIGN FROM (obtained from the front of the lab, near the login computers) ON THE KEYBOARD! BEWTWEEN 10 PM AND 8 AM, YOU MUST PLACE A STOP SIGN ON THE INSTRUMENTʼS KEYBOARD AND EMAIL JOSH AND YUYANG AND SUBMIT A PROBLEM REPORT WHEN LOGGING OUT OF THE FOM. YOU WILL NOT BE CHARGED OR PENALIZED FOR BREAKING A SAMPLE!!! SAMPLE BREAKAGE DOES NOT NECESSARILY CAUSE MUCH DAMAGE TO THE INSTRUMENT, BUT IF THE NEXT SAMPLE IS INSERTED OVER BROKEN GLASS, IT WILL BREAK THE PROBE, CAUSING THOUSANDS OF DOLLARS WORTH OF DAMAGE AND POTENTIALLY SEVERAL WEEKS OF INSTRUMENT DOWNTIME. 5

Step 4) Carefully load your sample into a spinner using the sample positioner Figure 9. Position the sample in the spinner. Step 5) Step 6) Place your sample on top of the magnet so it floats on the (loud) eject air. On Au400, you may need to push it past the silicone grabbers; simply hold the spinner not the probe and push it past the grabbers. Insert your sample by clicking Insert on the Start/Standard panel Figure 10. The Insert button. 6

Step 7) Select your solvent in Start/Standard pulldown menu Figure 11. Selecting the sample's solvent. Step 8) Retrieve the standard shims by clicking the light blue read default shims button in the Start/Standard panel. Figure 12. Reading the default shims 7

Step 9) Lock on the sample with the manual locking procedure in the Start/Lock panel. First, navigate to the Start/Lock panel. Second, turn the Lock off by clicking the OFF button in the lock panel. Third, begin observing your 2 H lock signal by clicking the Lock Scan button: Figure 13. The Lock panel The signal you see is the first part of the 2 H FID. Any waves you observe arise from the beat frequency difference between the fixed spectrometer frequency and the current resonance frequency of your solvent deuterium atoms. Your goal is to adjust the resonance frequency of your solvent so that it matches the transmitter frequency of the spectrometer s lock channel. You make those adjustments by changing the magnetic field strength, which you accomplish by changing the strength of a homogeneous corrective field called Z0 named because its field is the same across the length of the sample, thus has a zeroth-order dependence on Z. Now adjust the magnetic field by changing the value of Z0. Large-scale changes can be effected using the slider bar, but they usually aren t necessary. Left-click to make Z0 lower, right-click to make it higher, and center-click to toggle through the amounts Z0 will change (±1 ±10 ±100 ±1 ). You generally just pick a direction, change Z0, then observe whether the number of beats in your signal increases or decreases. If it decreases, continue changing Z0 in that direction; if it increases, turn around because you re going the wrong way. Figure 14. Making Z0 adjustments 8

You should see changes in your 2 H lock signal: Figure 15. Lock signal when Z0 is incorrectly set If you make adjustments to Z0 and do not observe any signals, you may need to increase your signal strength. Do this first by amplifying the lock signal more by increasing the Lock Gain; accomplish this by right-clicking on the Lock gain button. If your signal is too noisy or you ve reached the maximum gain, try applying a stronger lock RF field to increase your 2 H magnetization; increase your lock RF power by right-clicking the Lock power button. Figure 16. Adjustments for lock gain and power You should ultimately get a signal that looks like the one below, or at least one that looks close. When you ve done this, click the On button on the Lock panel to engage the lock feedback loop. The lock status should then read Regulated. Figure 17. G00d-looking lock signal, regulated. 9

Step 10) Shim on the sample using the manual shimming procedure. First, navigate to the Start/Shim panel: Figure 18. The Start/Shim panel. The purpose of shimming is to make the magnetic field across the sample more uniform, but we do not generally assess field homogeneity directly. Instead, we use the lock level to gauge shim quality. In the figure above, the lock level reads 13.5 (on a 0-to-100 scale), and on the screen will probably flicker between 13.3 and 13.7. To shim we adjust the current running through various non-superconducting shim coils in the magnet. We accomplish this by adjusting the setting of various buttons, primarily Z1 and Z2, using the left-, right-, and center-buttons on the mouse just as was done when adjusting Z0 to establish the lock. Left-click to make each shim lower, rightclick to make it higher, and center-click to toggle through the amounts the shim will change (±1 ±10 ±100 ±1 ). One does not know which way to adjust each shim at first it is an empirical process. For each shim, choose one direction of adjustment and assess whether the lock level rises or falls. If it falls, switch to the other direction. If it rises, continue in that direction until it falls, then back up and stay with the shim value that corresponds to the maximum lock level. Then move on to a different shim. Here we start by adjusting Z1 from -2828 to -3028 to reach a higher lock level (29.8): Figure 19. Adjusting Z1 improves the lock level. 10

Then adjusting Z2 from 1458 to 1428 yielded a further gain in lock level (to 35.2). Figure 20. Adjusting Z2 improves the lock level further. Z2 and Z1 shims are dependent on one another, so after adjusting Z2, go back and adjust Z1. If you want further improvement, you may wish to adjust Z3, X, XZ, Y, and YZ. One can spend all day shimming! To avoid that, you probably want to stop when no large improvements can be made. At this point, you should acquire a spectrum and assess whether the observed lineshape is good enough for your purposes. Step 11) Make sure your sample is locked and that the lock level is between ~15 and 90. If you need to adjust it, first change your lock gain (probably setting it lower), then change your lock power. Below a lock level of 15, you risk losing the lock control. Above that, and you may have trouble with your lock level going above 100 if you improve the shims further. There s no magic lock level number you re aiming for when shimming. Be careful to not increase the lock power too much; you may induce saturation of the 2 H signal, in which the lock level fluctuates or is unreliable. Step 12) Move to the Acquire/Default H1 panel by clicking on the Acquire panel s tab and then the Default panel title. Figure 21. The Acquire/Default H1 panel. Step 13) Set the number of scans to 1 and acquire a preliminary 1D 1 H spectrum by clicking the green Acquire button. Under Acquisition Options, pull down the Number of Scans menu to 1. Then click Acquire. 11 Figure 22. Selecting the number of scans.

Step 14) Evaluate the spectrum examine spectral width, phasing, linewidth, referencing, etc Figure 23. Spectrum viewing tools. Step 14a) Examine some of the prominent lines in your spectrum more closely by zooming in on them. Place your left cursor to the left of the peak with your left mouse button : Figure 24. The full 1D 1H spectrum, starting to zoom in. Next, place your right cursor to the right of the peaks by right-clicking : Figure 25. Full 1H 1D spectrum, including the right-cursor for zooming. 12

Next, click the zoom in button (magnifying glass with + sign, ): Figure 26. The spectrum, zoomed in upon. If you wish, increase the vertical intensity of the spectrum by center-clicking above a peak. This will bring the point in the spectrum directly underneath your cursor right up to the cursor. You can also use the scroller wheel to adjust intensity. Figure 27. Spectrum portion with vertical scale adjusted. Does that look OK to you? It may be hard to tell unless you ve investigated that spectrum before. It s often better to look at the solvent peak, since that shouldn t change significantly from sample to sample. Step 14b) Start gauging your spectrum quality by examining the linewidth of the residual solvent peak (aka, the solvent peak, but which is really the protiated portion of the deuterated solvent. Remember that 99.9%-deuterated CDCl 3 will still have 0.1% CHCl 3, which you observe and can even use as a chemical shift reference.) This is easier for solvents that present single lines, like CDCl 3, than solvents with multiple lines arising from 2 H- 1 H coupling, like dmso-d6. 13

Get a better look at your residual solvent line by zooming in on it; you may need to zoom out to the full spectrum first, which you accomplish using the zoom out button:. Figure 28. Zoom in on the solvent line. Next, you need to place your cursor precisely on the solvent peak. You can accomplish this in two steps: first by placing the (left) cursor somewhere on the peak. Figure 29. Place you cursor on the peak. Then, in the Process/Weighting panel, (1) click find nearest line, then (2) display linewidth. Your linewidth will read out in the display windows. Your solvent line should be below 2.0 Hz for routine data, and ~1.0 or below for publication quality. Figure 30. Find nearest line, display linewidth. 14

Step 14c) While your cursor is on the tip of the solvent peak, you may also wish to ensure your spectrum is accurately referenced with respect to frequency in ppm. The most definitive way to reference your spectrum is to include a small amount of tetramethyl silane (TMS) in your sample, and ensure that its 1 H peak resides at 0.000 ppm; this is the IUPAC-approved method for organic samples. If you have no TMS in your sample, which is very common, you can reference the spectrum to the chemical shift of the residual 1 H signal from the solvent. To confer the correct chemical shift on the solvent peak, leave the cursor on top of the peak, navigate to the Process/Default panel, and under the Reference section locate the Reference cursor to box. In that box, type the correct chemical shift of the solvent; for CHCl 3, you would type 7.24 in that box and hit the return key. This changes the spectrum scale so that the cursor s position reads 7.24 ppm. Figure 31. Manually referencing a spectrum by defining the frequency of a solvent peak. Step 14d) Continue evaluating your spectrum by increasing the vertical scale to see if there are any low-lying peaks that you may wish to observe. If so, you may wish to take another spectrum with a larger number of scans to make these peaks more clear. Step 15) If necessary, acquire more scans; in the Acquire/Default H1 panel, pull down the number of scans pulldown to the desired number of scans a. NB: Keep in mind that S/N improves with the square root of the number of scans; increasing them from 4 to 16 will increase S/N by a factor of ~1.4, but from 4 to 16 will improve it by a factor of 2 15

Step 16) Save your spectrum by clicking of the floppy disk save button A dialog window will pop up that will show you the files in the location where the last user saved their spectra. Navigate from here to your own folder by A) Clicking on the folder selector pulldown, which opens the file browser dialog B) Clicking the walkon account s folder C) Clicking (or creating) your folder D) Typing in a legal file name in the File Name box; legal = alphanumerics only, upper/lowercase OK, no spaces or special characters like single or double quotes. Step 17) Continue with further experiments other samples if you have them. Step 18) When you are done with your last sample, eject it using the Eject button in the Start/Standard panel 16

Step 19) Insert the lock standard using the Insert button in the Start/Standard panel, and establish lock on it. Step 20) Sign out from the pen-and-paper logbook, noting any instrument problems Step 21) Process your spectrum further and perhaps print it out using an offline processing computer 17