RADIATION MONITOR RADEX RD1706

Size: px
Start display at page:

Download "RADIATION MONITOR RADEX RD1706"

Transcription

1 RADIATION MONITOR RADEX RD1706 OPERATING MANUAL

2 RADEX and РАДЭКС are the registered brands of QUARTA-RAD Ltd. 2

3 Radiation monitor RADEX RD1706, hereinafter referred to as the device, is designed for detection and evaluation of the level of ionizing radiation hereinafter referred to as irradiation. The device evaluates radiation environment by the magnitude of the ambient equivalent of dose rate Н*(10) of gamma radiation (hereinafter referred as a dose rate) with taking into account gamma radiation and the pollution of objects by sources of beta particles. The device obtains the following distinctive features: * * * * * * * period of observation shortens with the increase of a dose rate; increase in the number of observation cycles improves the reliability of indications; smooth threshold signal changing in a wide range; displaying the value of a background dose rate; displaying a dose rate exceeding over a background dose rate; vibra-call signal as the additional alarming function; possibility to function with one battery of «ААА» type. The monitor is used for evaluation of the radiation level afield, indoors and for estimation of a contamination level of materials and products. Operating conditions: at the temperature ranging from -20 C to +50 C with a relative humidity no more than 80 % at the temperature of +25 C. The results obtained with this device can not be used for official decision on radiation environment and degree of pollution. 3

4 TABLE OF CONTENTS Precaution measures... Physical configuration... Display format in monitoring and background modes... Operating mode... Monitoring mode... Menu mode... LEVELS... BACKGROUND... SETUP... SERVICE... Background mode... Preparations for using... Device using... Radiation control in quarters and public buildings... Marking and sealing... Packing... Transportation and storage... Technical service... Maintenance and updating... Troubleshooting... Performance data... Technical specifications

5 Precautionary measures Do not use, place or storage the device for a long time under the impact of intense direct solar light or height temperature, for example on an instrument board or in a trunk. The impact of the sunlight or temperature can lead to electrolyte leakage from power elements, their overheating or explosion and therefore to arson, burn and other injuries. High temperature can also cause malformation of the device's frame. Do not storage the device in damp and dusty places. It can cause arson, electrical shock and other damages. Protect the unit from shocks and heavy mechanical stress that can cause damages of the device. This device is a high-precision technology. Avoid its falling and mechanical stress. The device is not water-proof; it cannot be used in the environment with a high humidity or in water. If the product wets, immediately switch off the device and apply to «QUARTA-RAD» Ltd. If a small amount of water has reached its surface or it was exposed to salty air it is necessary to switch off a product, to wipe it with dry fabric, to place it in a warm dry premise and to dry it before a complete evaporation of moisture from its internal. Keep the device away from devices such as electric or magnetic motors which generate strong magnetic fields. Do not use and place the monitor close to areas where high electromagnetic signals are generated, such as transmitting towers. Superhigh frequencies can lead to malfunctioning. Do not attempt yourself to dismantle or fix the device. Do not place the unit in a microwave and do not carry out any monitoring with ionizers and ozonizers turned on. Do not let particles penetrate through a perforated hole inside the unit. Do not touch electrical links of a battery block. It can cause corrosion of links and affect normal functioning. Condensate can occur on internal parts of the device in case of a rapid temperature decrease from higher to lower level. To prevent this place the device inside a plastic bag. Keep it there till its temperature becomes of the environment. 5

6 In order to prevent damages to the unit do not use it if a condensate has appeared. If it is so extract power batteries from the unit and wait for a while till a condensate evaporates. It is possible to use the device after a complete vaporization of a condensate only. Extract power batteries if the device is not being used for a long time and keep the unit in dry and cool conditions. It is not recommended to keep the unit in places like labs where chemical agents potentially may cause corrosion. It is necessary to check working capacity of the unit before usage if it was kept in storage for a long time. * * * * * * * * * Precautionary measures for LCD LCD possesses a high-precision technology. Nevertheless a disabled pixel can appear on LCD in the form of a permanent black dot on it, and it is not considered as a malfunction and does not impact on the image. Do not hit and press on the LCD as it can cause fractures and damages of the display. In case of LCD surface contamination, switch off the devise and wipe it gently with a soft tissue that does not scratch it. Do not leave the device under the impact of a direct solar and fluorescent light for a long time. Do not let deleterious chemical substances, such as acids, alkali, and solvents etc. fall on the display and do not keep it in places where these substances are located. Response time of the LCD increases at low temperatures and display can darken at high temperatures. Standard features of display will restore at room temperature. Precautionary measures for power elements Keep power elements out of reach of children. Its chemical agents represent danger in case of casual swallowing. In this case apply to a doctor immediately. Do not hold a power supply with metallic tools such as pliers for it can cause a short circuit. Do not heat or dismantle power elements for it can fracture. 6

7 Physical configuration The unit is designed as a portable and handheld device with a self-contained power supply LCD 2. «MENU» button and its function icon (on display). The button responds to five functions: - «MENU» responds in «Monitoring» and «Background» modes; - «SELECT» responds in «Menu» mode; - «CHANGE» responds in «Menu» mode; - «START» responds in «Menu» mode; - shifting the cursor of an altering figure in «Menu» mode. 3. «CURSOR» button and its function icon (on display). The button responds to two functions: - switches on LCD backlight in «Monitoring» and «Background» modes; moves cursor and alters numbers in «Menu» mode. - 7

8 4. «OFF» button and its function icon. The button has six functions: - switches on the device; - «RETURN» sends you one level back in «Menu» mode; - «EXIT» - transfers you from «Menu» to «Monitoring» or «Background» modes; - «END» responds in «Menu» mode; -«SAVE» - responds in «Menu» mode; - «OFF» - switches off the devise in «Monitoring» and «Background» modes. The icons suggest to user the functions of buttons making it easier to operate. Further in the text only icons of buttons will be pointed out. Instruction to press the button with a corresponding icon means clicking the applicable button of the device. The pressed button of the unit has animation, meaning that clicking any active button leads to temporary altering the shape of an icon on display from to and back to. In all displays shown below the animation of pressed buttons is not depicted. The cover for a battery bay is disposed on the backside of the unit. battery bay cover for a battery bay power supply 8

9 Display format in «Monitoring» and «Background» modes After activating the unit displays a screen of RD1706 with button icons and individual or manufacture settings. Location of 14 fields and possible icons are shown on two display pictures below Icon of battery charge: - complete battery charge; - partly discharged battery; - completely discharged battery 2.Conventional symbol of the unit icon of quantum registration. 4.Icon of «MENU» button. This field contains one of the icons displaying the current «MENU» button function: «MENU» «SELECT» «CHANGE» «START» or 9

10 5.Icon of «CURSOR» button (page 7). This field contains one of the following icons: - switching on backlight in «Monitoring» and «Background» modes; - moving cursor and altering value of a threshold in «Menu» mode. 6. Icon of «OFF» button (page 7). This field contains one of the following icons: RETURN» - responds in «Menu» mode, EXIT - while transferring from «Menu» to «Monitoring» and «Background» modes, END - responds in «Menu» mode, SAVE» - responds in «Menu» mode, OFF - responds in «Monitoring» mode. « « 7. Vibra-call icon: - vibra-call is on. When the vibra-call is off the icon is lacking; 8.Bell icon: - the audio signal is on. When the bell is off the icon is lacking; 9. Threshold signal icon, for example - signaling starts when indications reach 0,30 μsv/h (threshold signal can be set in a range from 0.10 to 0.90 μsv/h) or - when threshold signal is off. 10.Units icon: - microsievert per hour. 10

11 After a short cycle of observation the screen with read outs appears Digital readings of a dose rate in μsv/h; 12.Icon of the number observations performed: - corresponds to the short (half of the complete) cycle of observation - corresponds to one complete cycle of observation; - corresponds to two complete cycles of observation; - corresponds to three complete cycles of observation; - corresponds to four and more complete cycles of observation. 13.The value of a dose rate «NOISE ХХХ», where ХХХ is the value of a dose rate in μsv/h; Icon showing that the value of a dose rate is less than the value of a background dose rate. At the same time «0,00» value displays on the screen. - Icon showing that the value of a dose rate is over 999μSv/h. At the same time «999» value displays on the screen. 11

12 Operating mode There are three operating modes in the unit: «Monitoring», «Background» and «Menu». The «Monitoring» mode is set automatically at start of unit's functioning. This mode gives estimation of a dose rate and indications output to the LCD. The time of a dose rate estimation varies from 24 sec when a dose rate is less than 3,5 μsv/h till 1 sec when a dose rate exceeds 72,0 μsv/h. The button «CURSOR» in «Monitoring» mode (page 7) fulfills the function of turning on/off backlight of the display. A short term click turns on backlight for two seconds. It allows seeing indications in a twilight or darkness. The importance of a backlight during a bright illumination is not remarkable. Note. It is important to remember that a frequent switching of backlight shortens strongly the time of continuous operation of the device. «Background» mode (page 25) switches on in «BACKGROUND» section (page 18) in «Menu» mode. Just as in «Monitoring» mode the dose rate estimation is performed but it gives out two indications simultaneously; they are the exceeding of a dose rate over background dose rate and additionally a value of a dose rate of background. This mode is convenient for examining premises, when it is important to know the difference between indications indoors and those on an open air and how to determine correctly the value of a dose rate outdoors. The button «CURSOR» (page 7) turns on backlight in «Background» mode. «Menu» mode turns on by clicking «MENU» button (page 7) in case if altering of manufacture's set up is needed. In «Menu» mode the estimation of a dose rate is not being conducted. 12

13 Monitoring mode This mode gives the value of a dose rate displaying indications on the LCD. Every registered quantum is displayed on LCD as icon. Blinking frequency of icon is proportional to the strength of a radiation dose rate. A short cycle of observation is introduced in order to reduce a waiting time of the first reading. The indications of a short cycle and its icon (page 11, screen 12) are displayed twice faster than in a complete cycle, but the readings of a short cycle are approximate and must be precised during the next complete cycle. Readings received in this cycle are useful for the initial dose rate estimation. After a complete cycle of observation the readings and icon are displayed. The icon showing the number of observations performed allow determining visually the number of averaging cycles. It is useful for detecting a place of a strong variation in a dose rate (local radioactive contamination) for in this case the icon changes to and it is easy to remark. At the same time the estimation of a dose rate starts all over again from the first cycle, so averaging indication of the previous cycles are not taking into account. The period of observation depends on a dose rate value and varies from 26 sec to 1 sec. The period of observation equals to 26 sec when a dose rate value is in the range from 0,05 μsv/h to 3,5 μsv/h and when a dose rate value exceeds 3,5 μsv/h the time of observation gradually shortens and reaches its minimum of 1 sec at a dose rate of 72,0 μsv/h. Further a dose rate increase does not shorten the time of observation (it remains 1 sec). 13

14 Menu mode In «Menu» mode you can arrange the magnitude of a threshold signal, turn on/off alarm and vibra-call, enable «Background» mode, estimate the background radiation and also receive reference information. Switching to «Menu» mode ceases monitoring of a dose rate. In order to switch from «Monitoring» to «Menu» mode press and release «MENU» button (page 7). The main menu displays on the screen. At start the icon is always situated on the upper line of the menu, like in this case it is opposite to «LEVELS» section. Note. On the screen a pointer demonstrates which button it is necessary to click in order to switch from the present screen to the following one. «CURSOR» button shown on the screen as moves the pointer in «Menu» section. Clicking the button moves the cursor down only. After reaching the lowest position it transfers to the upper line. Choosing the section needed of the menu is implemented by pressing «SELECT» button. Returning to the previous section is implemented by pressing «RETURN» button. Switching from «Menu» to «Monitoring» modes is implemented by pressing «EXIT» button. This switching to «Monitoring» mode causes the estimation of irradiation to start from the first cycle. All individual settings are saved after turning off the devise as well. 14

15 «LEVELS» section Exceeding of a threshold dose rate value which can be set in «LEVELS» section provokes to alarm signal actuation. A threshold signal should be switched on when it is necessary to receive alarm or vibra signals in case of a fixed radiation dose rate exceeding. User himself sets a value of a threshold signal (range from 0,1 up to 99,0 μsv/h). Choosing the type of a signal (audio or vibra signal) is made in «SETUP» section (page 22). Thus, if a dose rate is registered during one observation exceeds the installed threshold level than turns on one of the two or both signals: a) when AUDIO signal is activated than with every quantum captured the sound signal will occur. If the selected threshold exceeds 3,0 μsv/h than a single sound signal is heard; b) when vibra signal is on it will function till the end of the cycle of observation. When a threshold signal is off, audio signal actuates with every registered quantum and a vibra-call does not function at all. Example of changing a level of a threshold signal from 0,30 μsv/h to 45, 0 μsv/h is presented below. 15

16 To change the level of a threshold switch from «Monitoring» to «Menu» mode (screen 1). The main menu displays on the screen (screen 2)

17 Press «SELECT» button and the screen 3 will appear. Adouble-click on button (screens 4 and 5) moves the cursor to «0,30 μsv/h». Press «CHANGE» button and the icon will be placed opposite to «0,30 μsv/h» and a digit 3 becomes underlined (screen 6); A double-click on button establishes «0,30 μsv/h» value (screens 7 and 8). Press button and the underlined figure moves one position left (screen 9). Press four times button and the required value of «4,50 μsv/h» will be set (screens 10 and 11). Press «SAVE» button and icon places opposite to «ON», icon stands opposite to «OFF» (screen 12) To return to the main menu screen press «RETURN» button (screen 13). To return to «Monitoring» mode press «EXIT» button. The screen «RD15706» appears and the icon of the new selected threshold () will be in the right upper corner. «Monitoring» mode stars functioning at the same time. The installed threshold signal is not activated if icon stands opposite to «OFF» and the device operates according to settings which are in «SETUP» section (page 22). For example, the audio beep signals with every registered quantum if the icon is placed only opposite to «AUDIO». If the icon is placed only opposite to «VIBRA-CALL», a vibra signal will not actuate because a vibra-call responds with a turned on threshold level only. 17

18 «BACKGROUND» section «Background» mode can be turned on in this section. It calculates an average value of a background dose rate. To activate «Background» mode switch from «Monitoring» mode to «Menu» mode (screen 1). Amain menu displays (screen 2). Press button and cursor places opposite to «BACKGROUND» section (screen 3). Press «SELECT» button and screen 4 appears. Double-click on button and cursor places opposite to «BACKGROUND» (screens 5 and 6). Press «CHANGE» button. Screen 7 appears where digits «1», «2», «3», «4», «5» determine a number of sighting points, which are necessary for conducting evaluation of a background radiation. Select the first sighting point according to recommendations described on page 29. Press «START» button, screen 8 appears on display and a cycle of estimation of a background starts. Approximately within 100 sec icon appears on the display and the cycle of background estimation in the first sighting point ends (screen 9). Screen 10 with indications in the first line which stand for the value of background dose rate in the first sighting point appears on the display. Note. A cycle of estimation of a background can be interrupted. To do so press «RETURN» button, thus a cycle of estimation will be interrupted and screen 4 revealed. 18

19

20 Press «START» button and wait till completion of the cycle of «BACKGROUND» estimation (screen 13). Avalue in the second line is the evaluation of a dose rate in the second sighting point (screen 14). In order to receive accurate estimation of a background it is necessary to conduct similar observations in 5 sighting points (screen 15). Note. Anumber of sighting points can be reduced by clicking «END» button without waiting for ending of the fifth cycle of estimation. But it reduces integrity of the reading. After 5 indications have been made press «END» button and screen 15 appears. To return to the main screen in «Menu» mode press «RETURN» button and screen 16 appears. Press «EXIT» button, the screen «RD1706» (screen 17) appears and «Background» mode starts operating (page 25). After a short cycle of observation two indications will be displayed (screen 18): a) a dose rate difference between the average indication and background; b) background dose rate, selected from indications of 5 sighting points. In order to terminate «Background» mode and to return to «Monitoring» mode it is necessary to switch to «Menu» mode (screens 1 and 2) and to select «BACKGROUND» section (screen 3) where icon should be placed opposite to «OFF» (screen 5). To return to the main menu screen of «Menu» mode press «RETURN» button, screen 6 appears. Press «EXIT» button. Screen «RD1706» reveals (screen 7) and «Monitoring» mode starts functioning. After short cycle of observation only one read out of a dose rate displays. 20

21

22 «SETUP» section Two types of signal can be chosen in «SETUP» section: «AUDIO» and «VIBRA-CALL» which can be turned on/off by placing icon opposite to a correspondent choice. Example of activating a vibra signal instead of an audio one is demonstrated below. To activate a vibra signal switch from «Monitoring» to «Menu» mode by pressing «MENU» button (screen 1). Press button two times, the cursor will be displaced opposite to «SETUP» section (screen 2). Press «SELECT» button, screen 3 appears. Press «CHANGE» button, icon opposite to «AUDIO» disappears (screen 4). Press button, the cursor is displaced opposite to «VIBRA-CALL» (screen 5). Press «CHANGE» button, icon opposite to «VIBRA- CALL» appears, meaning activation of a vibra signal (screen 6). To return to the main menu screen press «RETURN» button, screen 6 will appear. To start «Monitoring» mode press «EXIT» button, icon is presenting while icon is lacking on the main screen «RD1706». 22

23 From this moment a vibra signal turns on if the value of a dose rate exceeds the installed threshold of alert when a threshold function is not turned off (page 15). A vibra signal starts functioning in a pulse-mode till the end of an observation cycle. Note. Using a vibra call shortens sharply the device's non stop operating time. Setting an audio signal is similar to modifying a vibra call. In «Monitoring» and «Background» modes the installed setting are revealed with icons on the main screen «RD1706»: - audio signal is on, - vibra call is on. When audio and/or vibra signals are off the icons are lacking. Note. 1.A sound signal is usually used in noisy premises or on the street. 2.In quiet places such as offices and apartments it is recommended to use vibra call or sound alarm with a threshold installed. 23

24 «SERVICE» section In «SERVICE» section the following is mentioned: * phone number of an organization where a radiation monitor RADEX RD1706 can be acquired, * phone number of the QUARTA-RAD Ltd. Company, which sells wholesale a radiation monitor RADEX RD1706; * internet address where you can find news about a radioactive situation in Russia, get acquainted with radioactivity safety norms and other normative documents. Place a pointer opposite to «SERVICE» with «CURSOR» button. Press «SELECT» button. To return to the main menu press «RETURN». Pressing «EXIT» transfers to «Monitoring» mode. 24

25 «Background» mode This mode is convenient for examining inside the buildings. In «Background» mode the estimation of the radiation dose rate is similar to the one in «Monitoring» mode (page 13), yet the display gives not just one but two indications. One stands for exceeding of a dose rate over background dose rate, the second is for a background dose rate. This mode is convenient for examining inside buildings, when it is necessary to know how indications indoors differ from the ones outdoors. The actuation of the «Background» mode and evaluating background dose rate are given in «BACKGROUND» section (page 18). In «Background» mode screen RD1706 has the same view as in «Monitoring» mode but after the end of estimation cycle two values are displayed: 1.- exceeding of a dose rate over background dose rate (in the example it is 0,07 μsv/h) 2.- value of a background dose rate (in the example it is 0,10 μsv/h) 1 2 That a dose rate value in this example equals to 0,07 + 0,10 = 0,17 μsv/h. When the value of the radiation dose rate is less than of a background, the value «0,00» and the icon are displayed. Switching the device from «Background» to «Monitoring» mode is done in «BACKGROUND» section (page 18) by placing icon opposite to «OFF». The rules and sequence of defining the average value of background dose rate are covered on page

26 Preparations for using Before actuating the device read the operation manual and functions of operating controls carefully (page 7). Installation of power elements 1.Take off the cover of a battery compartment. 2.Install into the battery compartment one or two power elements size " AAA" abiding polarity of contacts. 3. Install the cover of a battery compartment. When conducting a long-term examination we recommend you to install two power elements, for a short-term just one as it is possible to use one power element. Do not mix the used and new power elements. We recommend altering individual settings before conducting the evaluation (page 22). Recommendations on examining the objects Remember that the ionizing radiation has a static random character therefore indications of the monitor in identical conditions will not remain strict. For a precise definition of a dose rate it is necessary to conduct not less than 4 cycles of monitoring without turning off the device. For defining radioactivity of the food, household items etc. approximate the monitor to the object of examination on a distance from 5 to 10 mm and turn it on. For defining radioactivity of fluids, the evaluation of a dose rate is conducted above the unclosed surface of a fluid. For the device protection it is recommended to use a polyethylene package, but no more than in one layer. Do not let the moisture to get inside the device. The estimation output exceeding the natural background that is ordinary for the given location is the evidence of the radioactive impurity of the inspected object. The "Background" mode is presented for convenience of defining such objects (page 25). Detecting the location of an ionizing radiation source it is necessary to move a working device above the surface of examined object, being oriented on frequency increase of a sound signal (in menu settings: levels - off, audio - on). Remember, that the frequency of signals as approaching the source will increase sharply, and in accordance with moving away will decrease in the same manner. 26

27 Device using Switching on the device To switch on the device press «OFF» button (page 7) and the screen «RD1706» will be displayed. The sequence of survey After switching on the device the examination of the radioactive environment starts. During the time of observations every registered quantum of radiation is accompanied by a displayed presentation icon (page 9) and a short audible signal, if the sound is switched on and the threshold is off. The frequency of occurrence of the icon on display is proportional to a dose rate. After 13 seconds of unit's functioning, the first result of a short cycle* and its icon are displayed. * The short cycle of observation is equal to a half of a complete cycle and made for prompting the deriving of preliminary results. The first reliable result depends on a dose rate and is displayed after 1-26 seconds of observation and designated by icon. 27

28 Icons - corresponds to one complete cycle of observation; - corresponds to two complete cycles of observation; - corresponds to three complete cycles of observation; -corresponds to four and more complete cycles of observation. The first results of observation is displayed as an average value of two short cycles, the second - as average value of the two complete cycles of observation, the third - as average value of three complete cycles of observation and each following result - as average value of four previous observations. The device analyzes a diversion of the current value regarding the result of a previous observation and averages it. If the difference exceeds a defined value, the current result is displayed instead of the average one. For example, results of three observations the average result is 0,20μSv/h and the value 0,80 μsv/h is registered in the fourth cycle, than the result of the fourth observation will not be averaged and we shall see 0,80μSv/h and icon. This function of the article allows detecting sharp changes of a dose rate. If you want to find a source of radiation turn off a threshold signal, switch on audio signal and pay attention not to digital readouts only but to a sound frequency and frequency of appearances of icon. Its appearance frequency is proportional to a dose rate thus the higher the frequency the closer you are to a source. Switching off the device To switch off the device press " OFF" button (page 7) and hold it till disappearance of the messages from the display. 28

29 Radiation control in quarters and public buildings. According to modern ideas about danger of small doses of radiation, the defensive measures of ngabited buildings should be conducted if a dose rate of gamma radiation inside a building exceeds a dose rate outside for more than 0,20μSv/h. The estimation of a dose rate outside (background) is carried out near to the inspected building not less than in 5 sighting points standing from 30 to 100 m apart from buildings and facilities and not closer than 20 m from each other. The sighting points of estimation should be selected on location with a natural soil without any man-caused modifications (road metal, sand, asphalt) and radioactive contamination. The unit should be at 1 m high above the earth surface. The total time of estimation of a background dose rate will take minutes but if you want to receive authentic results it is necessary to conduct a complete term examination. For your convenience «Background» mode (page 25) is supplemented to the device which makes it easier to conduct examination. 29

30 Marking and Sealing On the frame of the device are marked: - trade mark of the manufacturer; - trade mark - RADEX. Conventional designation of the monitor - RD1706 appears on display at its turning on. The assembly number of the device is on the main menu screen. The device is not sealed up by the manufacturer. Packing Packing ensures the safety of the device during the transportation. The packing of device must be carried out in closed aerated premises at the temperature from +15 to +40 C and humidity up to 80 %, and without aggressive admixtures and dash in environment. Transportation and storage The device in the manufacturer's packaging may be transported by any transport at any distance. Arrangement and restraint of the container with the packed product in vehicles should provide a steady condition the container and exclude it's moving during transportation. At transportation of the device it is necessary to ensure its protection from the atmospheric precipitates. The conditions of transportation of the device in the package must correspond to: - temperature span.... from -20 to +40 C - relative humidity at +25 C, not higher than % Before putting the device into operation it should be kept in the warehouse in a package of the manufacturer at temperature of from +5 to +40 C and RH up to 80 % at temperature of +25 C. It is not allowed to keep the device without packing. The device at temperature below 0 C should be kept for an hour in the premise at a room temperature before opening and putting into operation. 30

31 Technical service The maintenance of the unit involves: 1) dusting off the unit from an outboard surface; 2) power elements should be replaced if the message displays; 3) power elements should be expelled from battery compartment in case of a long term of interruption (more than one month); 4) wipe the display only with a soft tissue. During cleaning the device should be off. Do not let smaller articles penetrate the device through a perforation. Maintenance and updating All repairs of the unit are made at the manufacturer'spremises QUARTA-RAD Ltd. Company. Address: Quarta-Rad Ltd., 2 Building, 3, Podol'skikh Kursantov, Moscow, , Russia phone/fax +7 (495) , phone +7 (495) , quarta@quarta-rad.ru http: // Troubleshooting Possible malfunction Possible cause of malfunction Way of clearance There is no information on display after turning on. The power elements are not installed or the power elements are installed disregarding the polarity. The messagethe power elements are discharged is displayed. below the possible level. Install the power elements in accordance with the polarity Replace the power elements. 31

32 Actions in extreme conditions CAUTION! IF THE ITEM HAS DISPLAYEDADOSE RATE MORE THAN 1.20 μsv/h IT IS NECESSARYTO LEAVE THIS ZONE URGENTLYANDAPPLYTO ASTATE SANITARY - EPIDEMIOLOGICAL OFFICE FOR CONDUCTING A DETAILED RADIATIVE EXAMINATION. Performance data The radioactivity monitor RADEX RD1706, evaluates radiation environment by the magnitude of the ambient equivalent of dose rate Н*(10) of gamma radiation with taking into account gamma radiation and the pollution of objects by sources of beta particles. Individual settings: - switches on/off alarm signal; - switches on/off vibra-call signal; - choosing a threshold level which exceeding turns on alarm and vibra-call signals; - evaluation of a background dose rate; - displaying a value of a background dose rate; - displaying a dose rate exceeding over a background dose rate; - possibility to function with one battery of «ААА» type although it reduces the time of a non-stop performance. 32

33 Technical specifications Range of dose rate indications, μsv/h. 0, 05 to 999 Range of registered energy: 1) gamma radiation, MeV... 0,1 to 1,25 2) X-ray radiation, MeV ,03 to 3,0 3) beta radiation, MeV from 0,25 to 3,5 Reproducibility of indications (at confidential probability 0.95),% /Р P is a doze rate in μsv/h. Threshold levels, μsv/h , 1 to 99,0 Time of calculation, s to 26 Time of indication..continuously Power elements, size «AAA»....one or two Time of continuous work of the device, not less than, hours 500 Overall dimensions, height x breadth x depth, mm, not more than x60x26 Weight (without power elements), kg, not more than.. 0,09 Notes: 1 Period of observation shortens with the increase of a dose rate. The time of observation shortens with the value higher than 3,0 μsv/h. 2 The increase in the number of performed cycles improves the reliability of indications. 3 The continuous operating of the device is calculated on the basis of manufactures settings and two batteries with a capacity of 1350 mah. These elements have 24AU abbreviation and others. 4 Manufactures settings: threshold.... «0,30 μsv/h», background mode -... OFF, settings: alarm signal-... ON, vibra-call signal... OFF. 33

34 Memo

35

36

TRF STEP-DOWN TRANSFORMER USER MANUAL

TRF STEP-DOWN TRANSFORMER USER MANUAL TRF STEP-DOWN TRANSFORMER USER MANUA www.ventilation-system.com 2013 ! WARNING The present operation manual consisting of the technical details, operating instructions and technical specification applies

More information

Technical description and user manual. Survey Meter SM 8 D. Sensortechnik und Elektronik Pockau GmbH. Siedlungsstraße 5-7 D Pockau-Lengefeld

Technical description and user manual. Survey Meter SM 8 D. Sensortechnik und Elektronik Pockau GmbH. Siedlungsstraße 5-7 D Pockau-Lengefeld Technical description and user manual Survey Meter SM 8 D Sensortechnik und Elektronik Pockau GmbH Siedlungsstraße 5-7 D 09509 Pockau-Lengefeld www.step-sensor.de Germany STEP-SM8D BD-EN-20170619-2 - state

More information

AU OPTRONICS CORPORATION. Specification for Approval. INCOMING INSPECTION STANDARD FOR A201SN02 TFT-LCD MODULES (A- Grade)

AU OPTRONICS CORPORATION. Specification for Approval. INCOMING INSPECTION STANDARD FOR A201SN02 TFT-LCD MODULES (A- Grade) AU OPTRONICS CORPORATION Specification for Approval INCOMING INSPECTION STANDARD FOR A201SN02 TFT-LCD MODULES (A- Grade) The content of this technical information is subject to change without notice. Please

More information

User Manual CC DC 24 V 5A. Universal Control Unit UC-1-E. General Information SET. Universal Control Unit UC-1 Of Central Lubrication PAUSE CONTACT

User Manual CC DC 24 V 5A. Universal Control Unit UC-1-E. General Information SET. Universal Control Unit UC-1 Of Central Lubrication PAUSE CONTACT Universal Control Unit UC-1-E User Manual General Information Universal Control Unit UC-1 Of Central Lubrication CC DC 24 V 5A / M 15 SL /MK 31 M Z 30 General Information Contents Universal Control Unit

More information

Color Star Advanced Talking Color Identifier Light Detector and Light Analyzer

Color Star Advanced Talking Color Identifier Light Detector and Light Analyzer USER MANUAL Color Star Advanced Talking Color Identifier Light Detector and Light Analyzer EN TABLE OF CONTENTS 1. INTRODUCTION 2 2. DESCRIPTION OF THE DEVICE 3 3. CHARGING THE BATTERY 4 4. HOW TO TURN

More information

DH551C/DH550C/DL550C Double Sided Display User Manual

DH551C/DH550C/DL550C Double Sided Display User Manual DH551C/DH550C/DL550C Double Sided Display User Manual Disclaimer BenQ Corporation makes no representations or warranties, either expressed or implied, with respect to the contents of this document. BenQ

More information

Digital Indicating ORP Meter AER-101-ORP. Instruction Manual

Digital Indicating ORP Meter AER-101-ORP. Instruction Manual Digital Indicating ORP Meter AER-101-ORP Instruction Manual Preface Thank you for purchasing our AER-101-ORP, Digital Indicating ORP Meter. This manual contains instructions for the mounting, functions,

More information

VMA573 and VMA or 7 Wide Screen Color LCD Monitor. Owner s Manual. Installation Guide

VMA573 and VMA or 7 Wide Screen Color LCD Monitor. Owner s Manual. Installation Guide VMA573 and VMA773 5.6 or 7 Wide Screen Color LCD Monitor Owner s Manual Installation Guide 7 headrest / stand alone wide monitor 5.6 headrest / stand alone wide monitor 2 WARINING! The Clarion VMA 573

More information

User s Manual. LG Industrial Systems G3F AT4A. LG Programmable Logic Controller

User s Manual. LG Industrial Systems G3F AT4A. LG Programmable Logic Controller User s Manual LG Programmable Logic Controller G3F AT4A GLOFA G4F AT3A LG Industrial Systems CONTENTS Chapter 1. INTRODUCTION 1.1 Features 1-1 Chapter 2. SYSTEM CONFIGURATION 2.1 Example of System Configuration

More information

CHIMEI INNOLUX DISPLAY CORPORATION

CHIMEI INNOLUX DISPLAY CORPORATION DISPLAY CORPORATION LCD MODULE SPECIFICATION Customer: Model Name: AT043TN20 Date: 2010/05/10 Version: 01 Preliminary Specification Final Specification Remark 4.3 FOG (FPC:44.05mm) For Customer s Acceptance

More information

USER MANUAL. EMF Meter. Model EMF510. Additional User Manual Translations available at

USER MANUAL. EMF Meter. Model EMF510. Additional User Manual Translations available at USER MANUAL EMF Meter Model EMF510 Additional User Manual Translations available at www.extech.com Introduction Thank you for selecting the Extech EMF510 EMF Meter. This instrument measures electromagnetic

More information

Video Door Entry System Kits 2 Wires Installation

Video Door Entry System Kits 2 Wires Installation Video Door Entry System Kits 2 Wires Installation Safety & Operating Instructions Réf. NVM-902/NVC-9102B1+NPS-17 2014-10-20 These instructions are for your safety. Please read through them thoroughly before

More information

28 & 32 & 40 & 55 & 65 & 84-INCH TFT-LCD 4K MONITOR

28 & 32 & 40 & 55 & 65 & 84-INCH TFT-LCD 4K MONITOR 28 & 32 & 40 & 55 & 65 & 84-INCH TFT-LCD 4K MONITOR INSTRUCTION MANUAL Please read this manual thoroughly before use, and keep it handy for future reference. TABLE OF CONTENTS 1, General information...

More information

CITIZEN QUARTZ DOUBLE FACE. Model No. JQ4 Cal. No. C820

CITIZEN QUARTZ DOUBLE FACE. Model No. JQ4 Cal. No. C820 CITIZEN QUARTZ DOUBLE FACE Model No. JQ4 Cal. No. C820 1. Main Features This is a combination quartz watch with a full range of features such as Time, Calendar, Alarm, Chronograph, and Timer, plus the

More information

User Manual. Color video door phone CDV-70P PM0770P Printed In Korea /

User Manual. Color video door phone CDV-70P PM0770P Printed In Korea / User Manual Color video door phone CDV-70P 513-11, Sangdaewon-dong, Jungwon-gu, Seongnam-si, Gyeonggi-do, Korea 513-11, Sangdaewon-dong, Jungwon-gu, Seongnam-si, Gyeonggi-do, Korea Business Dept. : +82-31-7393-540~550

More information

OWNER S MANUAL MOTORIZED 7 WIDE TFT LCD COLOR MONITOR CNT-701

OWNER S MANUAL MOTORIZED 7 WIDE TFT LCD COLOR MONITOR CNT-701 OWNER S MANUAL PW MOTORIZED 7 WIDE TFT LCD COLOR MONITOR CNT-701 ANY CHANGES OR MODIFICATIONS IN CONSTRUCTION OF THIS UNIT DEVICE WHICH IS NOT APPROVED BY THE PARTY RESPONSIBLE FOR COMPLIACE COULD VOID

More information

MONOPRICE. 27" UHD IPS 4K Ultra Slim Aluminum Monitor. Quick User's Guide P/N 24658

MONOPRICE. 27 UHD IPS 4K Ultra Slim Aluminum Monitor. Quick User's Guide P/N 24658 MONOPRICE 27" UHD IPS 4K Ultra Slim Aluminum Monitor P/N 24658 Quick User's Guide SAFETY WARNINGS AND GUIDELINES Please read this entire manual before using this device, paying extra attention to these

More information

SM-10 SAT Level Meter User s Manual

SM-10 SAT Level Meter User s Manual SM-10 SAT Level Meter User s Manual DAGATRONICS CORPORATION 263-1 DUCKIDONG, ILSAN, KOYANG, KYUNGKIDO, KOREA TEL: +82-31-916-8005 FAX: +82-31-916-8080 Email: dagatron@dagatron.com Website: www.dagatron.com

More information

CONTENTS 8 ACCESSORIES 13 9 TROUBLE SHOOTING AND ADVICE SPECIFICATIONS BATTERY USAGE CAUTION 13

CONTENTS 8 ACCESSORIES 13 9 TROUBLE SHOOTING AND ADVICE SPECIFICATIONS BATTERY USAGE CAUTION 13 CONTENTS 1 PREFACE 2 2 SAFETY PRECAUTIONS 2 3 FRONT & BACK VIEW, REAR CONNECTION OF LCD 3 4 TV INSTALLATION DRAWINGS 4 4.1 ANTENNA CONNECTION 4 4.2 AV1 INPUT CONNECTION 4 4.3 AV2 INPUT CONNECTION 4 4.4

More information

INNOLUX DISPLAY CORPORATION LCD MODULE SPECIFICATION

INNOLUX DISPLAY CORPORATION LCD MODULE SPECIFICATION INNOLUX DISPLAY CORPORATION LCD MODULE SPECIFICATION Customer: Model Name: AT050TN33 V.1 SPEC NO.: A050-33-TT-11 Date: 2008/11/25 Version: 01 Preliminary Specification Final Specification For Customer

More information

Weekly Time Switch. Rated time Time setting range Time division 24 hrs x 7 days 00:00 to 23:59 1min

Weekly Time Switch. Rated time Time setting range Time division 24 hrs x 7 days 00:00 to 23:59 1min Weekly Time Switch Easy Programming with Large LCD Display and Interactive Functions Programming for 24 hrs x 7 days using just five switches. Sixteen program steps available. Power supply freely selectable

More information

MANUAL ENGLISH Core Club Ordercode: D2314

MANUAL ENGLISH Core Club Ordercode: D2314 MANUAL ENGLISH Core Club Ordercode: Highlite International B.V. Vestastraat 2 6468 EX Kerkrade the Netherlands Table of contents Warning... 2 Unpacking Instructions... 2 Safety Instructions... 2 Operating

More information

LED Spot Light User Manual Model LS90 Big Dipper Laser Science And Technology Co., Ltd Please read this manual carefully before operating

LED Spot Light User Manual Model LS90 Big Dipper Laser Science And Technology Co., Ltd Please read this manual carefully before operating LED Spot Light User Manual Model:LS90 Big Dipper Laser Science And Technology Co., Ltd REV: 1.01 Please read this manual carefully before operating Warnings: Please check the package whether any damage

More information

GT-1C7-L/GT-1C7 GT-1M3-L/GT-1M3

GT-1C7-L/GT-1C7 GT-1M3-L/GT-1M3 SERVICE MANUAL Residential/Tenant Station OPERATION MANUAL Thank you for selecting Aiphone for your communication needs. * Please read this manual for safe and correct use of the system, and keep this

More information

Thank you for purchasing SEIKO SHOOTING TIMER KT-401. Before using your SEIKO SHOOTING TIMER, please read this manual carefully for its proper use

Thank you for purchasing SEIKO SHOOTING TIMER KT-401. Before using your SEIKO SHOOTING TIMER, please read this manual carefully for its proper use 紙 Thank you for purchasing SEIKO SHOOTING TIMER KT-401. Before using your SEIKO SHOOTING TIMER, please read this manual carefully for its proper use and care. Keep this manual handy for ready reference.

More information

HOME GUARD USER MANUAL

HOME GUARD USER MANUAL HOME GUARD USER MANUAL CONTENTS 1. SAFETY PRECAUTIONS...2 2. INTRODUCTION...3 3. FEATURES...4 4. ACCESSORIES...5 5. INSTALLATION...6 6. NAME and FUNCTION of EACH PART...7 6.1 Front Pannel...7 6.2 Monitoring

More information

USER MANUAL. 22" Class Slim HD Widescreen Monitor L215DS

USER MANUAL. 22 Class Slim HD Widescreen Monitor L215DS USER MANUAL 22" Class Slim HD Widescreen Monitor L215DS TABLE OF CONTENTS 1 Getting Started Package Includes Installation 2 Control Panel / Back Panel Control Panel Back Panel 3 On Screen Display 4 Technical

More information

4K LED MONITOR 430LED4K & 550LED4K INSTRUCTION MANUAL. Please read this manual thoroughly before use, and keep it handy for future reference.

4K LED MONITOR 430LED4K & 550LED4K INSTRUCTION MANUAL. Please read this manual thoroughly before use, and keep it handy for future reference. 4K LED MONITOR 430LED4K & 550LED4K INSTRUCTION MANUAL Please read this manual thoroughly before use, and keep it handy for future reference. CONTENTS 1. GENERAL INFORMATION...3 2. 3. 1.1. 1.2. 1.3. 2.1.

More information

Real-Time PCR System TC

Real-Time PCR System TC Detecting amplifier Real-Time PCR System TC 9443-007 - 46482062-2004 TECHNICAL PASSPORT ЛТОК 170604.00 DNA-Technology research-and-production close corporation Moscow, 2004 Contents 1. General information...

More information

LCD Thermometer / Clock S No. 1253

LCD Thermometer / Clock S No. 1253 Installation and Operating Manual LCD Thermometer / Clock S No. 1253 The 3 fold thermometer with crystal clock is purpose build for the mounting in caravans, boats and intervention vehicles. Please read

More information

Ambient Weather WS-01T-C Color Changing Wireless Temperature Night Light with Ambient Backlight (Console Only) User Manual

Ambient Weather WS-01T-C Color Changing Wireless Temperature Night Light with Ambient Backlight (Console Only) User Manual Ambient Weather WS-01T-C Color Changing Wireless Temperature Night Light with Ambient Backlight (Console Only) User Manual Table of Contents 1 Introduction... 2 2 Getting Started... 2 2.1 Parts List...

More information

BEAM MOVING HEAD YODN Lamp. 132W 2R MOVING HEAD Brightness / Stability USER MANUAL. Version 1.0 beta

BEAM MOVING HEAD YODN Lamp. 132W 2R MOVING HEAD Brightness / Stability USER MANUAL. Version 1.0 beta BEAM MOVING HEAD YODN Lamp W R MOVING HEAD Brightness / Stability This product manual contains important information about the safe installation and use of this projector. Please read and follow these

More information

28 4K LED monitor. User Manual M284K

28 4K LED monitor. User Manual M284K 28 4K LED monitor User Manual M284K CONTENTS Safety Information... 2 What s included..... 4 Getting Started....... 8 Troubleshooting.... 14 Specification.... 15 2 of 15 SAFETY INFORMATION Read these instructions

More information

PLL1920M LED LCD Monitor

PLL1920M LED LCD Monitor PLL1920M LED LCD Monitor USER'S GUIDE www.planar.com Content Operation Instructions...1 Safety Precautions...2 First Setup...3 Front View of the Product...4 Rear View of the Product...5 Installation...6

More information

USER MANUAL. 27 Full HD Widescreen LED Monitor L27ADS

USER MANUAL. 27 Full HD Widescreen LED Monitor L27ADS USER MANUAL 27 Full HD Widescreen LED Monitor L27ADS TABLE OF CONTENTS 1 Getting Started 2 Control Panel/ Back Panel 3 On Screen Display 4 Technical Specs 5 Care & Maintenance 6 Troubleshooting 7 Safety

More information

OFI Optical Fiber Identifier User s Guide

OFI Optical Fiber Identifier User s Guide OFI Optical Fiber Identifier User s Guide OFI Optical Fiber Identifier User s Guide 2002-2008, AFL Telecommunications, all rights reserved. OFI1-11-1000 Revision A, 05.02.08 Specifications are subject

More information

CLOVER DISPLAY LTD. LCD MODULE SPECIFICATION. Model: CV9162E _

CLOVER DISPLAY LTD. LCD MODULE SPECIFICATION. Model: CV9162E _ LCD MODULE SPECIFICATION Model: - - - - _ Revision 10 Engineering Timmy Kwan Date 11 August 2010 Our Reference 9016 ADDRESS : ROOM 1006, 10/F WESTIN CENTRE, 26 HUNG TO ROAD, KWUN TONG, KOWLOON, HONG KONG.

More information

LED Thunder S-150 Code 1097

LED Thunder S-150 Code 1097 LED Thunder S-150 Code 1097 User Manual 1 1 SAFETY INSTRUCTIONS This device has left the factory in perfect condition. In order to maintain this condition and to ensure a safe operation, it is absolutely

More information

PLL2210MW LED Monitor

PLL2210MW LED Monitor PLL2210MW LED Monitor USER'S GUIDE www.planar.com Content Operation Instructions...1 Safety Precautions...2 First Setup...3 Front View of the Product...4 Rear View of the Product...5 Quick Installation...6

More information

User Manual PS-684. HDBaseT Extender Kit 70m. All Rights Reserved. Version: UHBT70P_2016V1.2

User Manual PS-684. HDBaseT Extender Kit 70m. All Rights Reserved. Version: UHBT70P_2016V1.2 User Manual PS-684 All Rights Reserved Version: UHBT70P_2016V1.2 Preface Read this user manual carefully before using this product. Pictures shown in this manual is for reference only, different model

More information

DATAGAUSS XL HARD DRIVE DEGAUSSER

DATAGAUSS XL HARD DRIVE DEGAUSSER TECHNICAL MANUAL Operating and Maintenance Instructions for DATAGAUSS XL HARD DRIVE DEGAUSSER ZZ001211 / ZZ001212 VS SECURITY PRODUCTS LTD DATAGAUSS XL Hard drive degausser OPERATING MANUAL PRODUCTION

More information

CITIZEN QUARTZ Windsurfing. Model No. MC6XXX Cal. No. D28A/D28B INSTRUCTION MANUAL CTZ-B Name and Functions of Buttons and Displays

CITIZEN QUARTZ Windsurfing. Model No. MC6XXX Cal. No. D28A/D28B INSTRUCTION MANUAL CTZ-B Name and Functions of Buttons and Displays CITIZEN QUARTZ Windsurfing 4. Name and Functions of Buttons and Displays Model No. MC6XXX Cal. No. D28A/D28B INSTRUCTION MANUAL CTZ-B6910 1. Explanation of Displays and Buttons Although the display may

More information

USER MANUAL. 27 Full HD Widescreen LED Monitor L270E

USER MANUAL. 27 Full HD Widescreen LED Monitor L270E USER MANUAL 27 Full HD Widescreen LED Monitor L270E TABLE OF CONTENTS 1 Getting Started 2 Control Panel/ Back Panel 3 On Screen Display 4 Technical Specs 5 Care & Maintenance 6 Troubleshooting 7 Safety

More information

Part names (continued) Remote control

Part names (continued) Remote control Introduction Part names (continued) Remote control (1) STANDBY ( 25) (1) (2) ON ( 25) (3) (3) ID - 1 / 2 / 3 / 4 s ( 18) (4) (4) COMPUTER 1 ( 27) (7) (5) COMPUTER 2 * (8) (6) COMPUTER 3 * (10) (13) (7)

More information

CITIZEN QUARTZ SUPER CHRONOGRAPH With Electroluminescence. Model No. JN6XXX Cal.No.C40 INSTRUCTION MANUAL CTZ-B6767

CITIZEN QUARTZ SUPER CHRONOGRAPH With Electroluminescence. Model No. JN6XXX Cal.No.C40 INSTRUCTION MANUAL CTZ-B6767 CITIZEN QUARTZ SUPER CHRONOGRAPH With Electroluminescence Adjusting the Analog Time At the time adjustment position, turn the crown to adjust the time. Model No. JN6XXX Cal.No.C40 INSTRUCTION MANUAL CTZ-B6767

More information

/ PAT01 USER MANUAL

/   PAT01 USER MANUAL PONOVO POWER CO., LTD No. 139 Jinghai Third Road, BDA, Beijing, China, 100176 Office TEL. +86 (10) 59089666 E-Mail Website Info@relaytest.com www.relaytest.com / www.ponovo.com.cn PAT01 USER MANUAL VERSION:

More information

User Manual Color video door phone

User Manual Color video door phone User Manual Color video door phone CDV-43Q Thank Thank you you for for purchasing purchasing COMMAX COMMAX products. products. Please Please carefully carefully read read this this User s User s Guide

More information

Single cable multiswich programmer PC102W

Single cable multiswich programmer PC102W Single cable multiswich programmer PC102W 1. Product description The PC102W - single cable multiswich programmer (in the text - programmer) is useful instrument while configuring and troubleshooting SAT

More information

INNOLUX. Record of Revision. Version Revise Date Page Content /12/26 Initial Release

INNOLUX. Record of Revision. Version Revise Date Page Content /12/26 Initial Release InnoLux copyright 2004 All rights reserved, Copying forbidden. Record of Revision Version Revise Date Page Content 1 2005/12/26 Initial Release Contents 1. General Specifications...1 2. Pin Assignment...2

More information

User Manual TL-2X1-HDV 2x1 HDMI & VGA Switcher All Rights Reserved Version: TL-2X1-HDV_160630

User Manual TL-2X1-HDV 2x1 HDMI & VGA Switcher All Rights Reserved Version: TL-2X1-HDV_160630 User Manual TL-2X1-HDV 2x1 HDMI & VGA Switcher All Rights Reserved Version: TL-2X1-HDV_160630 Preface Read this user manual carefully before using this product. Pictures shown in this manual are for reference

More information

INSTRUCTION MANUAL. Heart Rate Monitor. English. Model HR-100CAN ST/SP SET MODE TIME

INSTRUCTION MANUAL. Heart Rate Monitor. English. Model HR-100CAN ST/SP SET MODE TIME INSTRUCTION MANUAL Heart Rate Monitor Model HR-100CAN ST/SP SET MODE TIME English TABLE OF CONTENTS Before Using the Monitor Introduction...3 Components and Functions...4 Your Personal Data Maximum Heart

More information

LITE-ON TECHNOLOGY CORPORATION

LITE-ON TECHNOLOGY CORPORATION Features * Lead (Pb) free product RoHS compliant. * Low power consumption. * High efficiency. * Versatile mounting on p.c. board or panel. * I.C. compatible/low current requirement. * Popular T-1 diameter.

More information

PanelView 1400e CRT Maintenance

PanelView 1400e CRT Maintenance Release Note PanelView 1400e CRT Maintenance Maximizing the life of your PanelView 1400e, CRT Terminals To maximize the life of a CRT, the following is strongly recommended: Adjust the external brightness

More information

PXL2760MW LED LCD Monitor

PXL2760MW LED LCD Monitor PXL2760MW LED LCD Monitor USER'S GUIDE www.planar.com Content Operation Instructions...1 Safety Precautions...2 Package Overview...3 First Setup...4 Front View of the Product...5 Rear View of the Product...6

More information

P XGA TFT Monitor. User s Manual

P XGA TFT Monitor. User s Manual P6151 15 XGA TFT Monitor User s Manual Disclaimers This manual has been carefully checked and believed to contain accurate information. Axiomtek Co., Ltd. assumes no responsibility for any infringements

More information

CITIZEN QUARTZ Basketball Timer. Model No. ME2XXX Cal. No. D310 INSTRUCTION MANUAL CTZ-B EXPLANATION OF DISPLAY AND BUTTONS

CITIZEN QUARTZ Basketball Timer. Model No. ME2XXX Cal. No. D310 INSTRUCTION MANUAL CTZ-B EXPLANATION OF DISPLAY AND BUTTONS CITIZEN QUARTZ Basketball Timer 3. EXPLANATION OF DISPLAY AND BUTTONS Model No. ME2XXX Cal. No. D310 INSTRUCTION MANUAL CTZ-B8003 When reading this instruction manual please keep the watch diagram in view.

More information

Copyright 2018 Xi an NovaStar Tech Co., Ltd. All Rights Reserved. No part of this document may be copied, reproduced, extracted or transmitted in any

Copyright 2018 Xi an NovaStar Tech Co., Ltd. All Rights Reserved. No part of this document may be copied, reproduced, extracted or transmitted in any MRV366 Receiving Card Document Version: Document Number: V1.0.1 NS110100658 Copyright 2018 Xi an NovaStar Tech Co., Ltd. All Rights Reserved. No part of this document may be copied, reproduced, extracted

More information

User Manual TL-2X1-HDVC 2x1 HDMI & VGA Switcher with Control All Rights Reserved Version: TL-2X1-HDVC_160630

User Manual TL-2X1-HDVC 2x1 HDMI & VGA Switcher with Control All Rights Reserved Version: TL-2X1-HDVC_160630 User Manual TL-2X1-HDVC 2x1 HDMI & VGA Switcher with Control All Rights Reserved Version: TL-2X1-HDVC_160630 Preface Read this user manual carefully before using this product. Pictures shown in this manual

More information

PL2410W LCD Monitor USER'S GUIDE.

PL2410W LCD Monitor USER'S GUIDE. PL2410W LCD Monitor USER'S GUIDE www.planar.com Content Operation Instructions...1 Safety Precautions...2 First Setup...3 Front View of the Product...4 Rear View of the Product...5 Quick Installation...6

More information

INSTRUCTION & INSTALLATION GUIDE

INSTRUCTION & INSTALLATION GUIDE Digital electric radiator INSTRUCTION & INSTALLATION GUIDE 2 Introduction harmoni digital electric radiators are programmable with an exclusive electronic temperature programmer and manufactured with quality

More information

Winmate Communication INC.

Winmate Communication INC. 20.1 Military Grade Display Model: R20L100-RKA2ML User s Manual Winmate Communication INC. May, 2011 1 IMPORTANT SAFETY INSTRUCTIONS Please read these instructions carefully before using the product and

More information

Hip Hop Activity Box. Instruction Manual. Hip Hop Activity Box 22875, October 2018 Copyright ROMPA Ltd

Hip Hop Activity Box. Instruction Manual. Hip Hop Activity Box 22875, October 2018 Copyright ROMPA Ltd Hip Hop Activity Box Instruction Manual Hip Hop Activity Box 22875, October 2018 Copyright ROMPA Ltd Contents 3 Panel Overview 5 Colour 5 Description 5 Technical Specification 6 Dimensions 7 Contents 8

More information

SquareLED - Aura Bar & Matrix Beam Light 100

SquareLED - Aura Bar & Matrix Beam Light 100 SquareLED - Aura Bar & Matrix Beam Light 100 1. SAFETY INSTRUCTIONS Please read these instructions carefully they include the important information about the installation usage and maintenance of this

More information

Ambient Weather WS-01T Color Changing Wireless Temperature Night Light with Ambient Backlight User Manual

Ambient Weather WS-01T Color Changing Wireless Temperature Night Light with Ambient Backlight User Manual Ambient Weather WS-01T Color Changing Wireless Temperature Night Light with Ambient Backlight User Manual Table of Contents 1 Introduction... 2 2 Getting Started... 2 2.1 Parts List... 2 2.2 Recommend

More information

User Manual Color video door phone

User Manual Color video door phone User Manual Color video door phone CDV-70AR3 Thank you for purchasing COMMAX products. Please carefully read this User s Guide (in particular, precautions for safety) before using a product and follow

More information

How to use and maintain survey meters and a personal digital dosimeter - Materials for explaining radiation measurement equipment and devices -

How to use and maintain survey meters and a personal digital dosimeter - Materials for explaining radiation measurement equipment and devices - Nuclear Disaster Response Basic Training Session How to use and maintain survey meters and a personal digital dosimeter - Materials for explaining radiation measurement equipment and devices - (English

More information

Spectra Batten (Order code: LEDJ95)

Spectra Batten (Order code: LEDJ95) www.prolight.co.uk Spectra Batten (Order code: LEDJ95) Safety WARNING FOR YOUR OWN SAFETY, PLEASE READ THIS USER MANUAL CAREFULLY BEFORE YOUR INITIAL START-UP! CAUTION! Keep this equipment away from rain,

More information

5.0 inch TFT LCD SPECIFICATION

5.0 inch TFT LCD SPECIFICATION 5.0 inch TFT LCD Without Touch Panel SPECIFICATION MODEL NAME: LMCRA050Z H N1 Date: 2013 / 02 / 18 Customer Signature Customer Approved Date Approved By Reviewed By Record of Revision Version Revise Date

More information

USER MANUAL Version 1.5.2

USER MANUAL Version 1.5.2 USER MANUAL Version 1.5.2 CONTEÚDO Introduction... 1 Technical Specifications... 1 Power supply... 1 Dimensions (LxWxH)... 1 Package Contents... 1 Installation Instructions... 2 Mounting on the TV... 2

More information

OFI-400 Series Optical Fiber Identifiers. OFI-400 Series Models

OFI-400 Series Optical Fiber Identifiers. OFI-400 Series Models OFI-400 Series Optical Fiber Identifiers Quick Reference Guide MODEL FIBER SIZE DESCRIPTION AND FUNCTION OFI-400 SM fibers: 250 µm coated 900 µm coated Ribbon 2 mm jacketed 3 mm jacketed Designed for use

More information

CM-S23349SV. Vari-Focal IR Bullet Camera

CM-S23349SV. Vari-Focal IR Bullet Camera Vari-Focal IR Bullet Camera User s Guide CM-S23349SV SAFETY PRECAUTIONS WARNING 1. Be sure to use only the standard adapter that is specified in the specification sheet. Using any other adapter could cause

More information

User Manual Color Video Door Phone

User Manual Color Video Door Phone User Manual Color Video Door Phone CMV-43A Thank you for purchasing COMMAX products. Please carefully read this User s Guide (in particular, precautions for safety) before using a product and follow instructions

More information

SV-LCD50. Installation and User Guide. Thin-Film Transistor (TFT) Liquid Crystal Display (LCD) Color Rear Vision Monitor. Version 1.

SV-LCD50. Installation and User Guide. Thin-Film Transistor (TFT) Liquid Crystal Display (LCD) Color Rear Vision Monitor. Version 1. SV-LCD50 Installation and User Guide Thin-Film Transistor (TFT) Liquid Crystal Display (LCD) Color Rear Vision Monitor Version 1.00 August 2004 SV-LCD50 Installation and User Guide TFT LCD Color Rear Vision

More information

Outdoor IR Audio Camera

Outdoor IR Audio Camera Outdoor IR Audio Camera User s Guide CM-S22326BW-AD SAFETY PRECAUTIONS WARNING 1. Be sure to use only the standard adapter that is specified in the specification sheet. Using any other adapter could cause

More information

NT-9600 Wireless Barcode Scanner. Introduction

NT-9600 Wireless Barcode Scanner. Introduction Guangzhou Netum Electronic Technology Co., Ltd TEL: +86 20 82679969*816 FAX: +86 20 82684887 E-mail: scottchiu@gzxlscan.com Address: Unit137, the Pacific Industry Area, Xintang Town, Zengcheng District,

More information

Table of Contents Precautions... 2

Table of Contents Precautions... 2 Table of Contents Precautions... 2 Special notes on LCD monitors... 2 Package contents... 3 Installation instructions... 3 Assembling the monitor... 3 Adjusting the viewing angle... 4 Connecting the devices...

More information

SE-SMD0603-PW65 SMD 0603 purweiss, wasserklar

SE-SMD0603-PW65 SMD 0603 purweiss, wasserklar 1. Features: Lighting Color: Ultra Bright Pure White Lens Color: Water Clear Flat Mold 2. (Package Dimensions): Notes: All dimensions are in millimeters. Tolerance is ± 0.10mm unless otherwise noted 03.05.2017

More information

Users Manual. Stroboscope. September 2014

Users Manual. Stroboscope. September 2014 820-2 Stroboscope Users Manual September 2014 2014 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All product names are trademarks of their respective companies.

More information

Ambient Weather WS-01 Intelligent Color Changing Temperature Night Light with Ambient Backlight User Manual

Ambient Weather WS-01 Intelligent Color Changing Temperature Night Light with Ambient Backlight User Manual Ambient Weather WS-01 Intelligent Color Changing Temperature Night Light with Ambient Backlight User Manual Table of Contents 1 Introduction... 1 2 Warnings... 2 3 Getting Started... 2 3.1 Parts List...

More information

INSTALLATION MANUAL. ST-CVTMD420-WPIR-W Covert Motion Detection Color Camera. v1.3 8/11/11 1

INSTALLATION MANUAL. ST-CVTMD420-WPIR-W Covert Motion Detection Color Camera. v1.3 8/11/11 1 INSTALLATION MANUAL ST-CVTMD420-WPIR-W Covert Motion Detection Color Camera v1.3 8/11/11 1 PACKAGE CONTENTS This package contains: One ST-CVTMD420-WPIR-W covert motion detection camera One installation

More information

Quick Start. RSHS1000 Series Handheld Digital Oscilloscope

Quick Start. RSHS1000 Series Handheld Digital Oscilloscope Quick Start RSHS1000 Series Handheld Digital Oscilloscope General Safety Summary Carefully read the following safety precautions to avoid personal injury and prevent damage to the instrument or any products

More information

Operation Manual VMS 3.0 Video System

Operation Manual VMS 3.0 Video System Operation Manual VMS 3.0 Video System for the AlterG Anti-Gravity Treadmill 1 This manual covers operation procedures for the following AlterG products: AlterG Video System model VMS 3.0 NOTE: The following

More information

INSTRUCTION & OPERATION MANUAL. MODEL MMW-05 (5kW) MUROMACHI MICROWAVE FIXATION SYSTEM

INSTRUCTION & OPERATION MANUAL. MODEL MMW-05 (5kW) MUROMACHI MICROWAVE FIXATION SYSTEM INSTRUCTION & OPERATION MANUAL MODEL MMW-05 (5kW) MUROMACHI MICROWAVE FIXATION SYSTEM MUROMACHI KIKAI CO., LTD. TOKYO PRECAUTIONS The Microwave Fixation System is developed to rapidly halt brain chemical

More information

DEEPFRAME BASIC KIT- USER MANUAL VERSION ORIGINAL USER MANUAL

DEEPFRAME BASIC KIT- USER MANUAL VERSION ORIGINAL USER MANUAL DEEPFRAME BASIC KIT- USER MANUAL VERSION 1.0 - ORIGINAL USER MANUAL new type of mixed reality display that enables digital content to appear as a hologram on top of reality seen 1 Content Security precautions

More information

ACUBRITE 23 SS. Manual. Stainless Steel Chassis 23" LCD Display. Content

ACUBRITE 23 SS. Manual. Stainless Steel Chassis 23 LCD Display. Content ACUBRITE 23 SS Stainless Steel Chassis 23" LCD Display Manual Introduction... 2 Hardware Installation... 2 The Display Timing... 5 The Display Outline Dimensions... 6 The Display Controls... 7 The Screen

More information

17 19 PROFESSIONAL LCD COLOUR MONITOR ART

17 19 PROFESSIONAL LCD COLOUR MONITOR ART 17 19 PROFESSIONAL LCD COLOUR MONITOR ART. 41657-41659 Via Don Arrigoni, 5 24020 Rovetta S. Lorenzo (Bergamo) http://www.comelit.eu e-mail:export.department@comelit.it WARNING: TO REDUCE THE RISK OF FIRE

More information

TABLE OF CONTENTS Important Safety Instructions Package Content Setting Up the Display Trouble shooting Specifications Product Dimensions

TABLE OF CONTENTS Important Safety Instructions Package Content Setting Up the Display Trouble shooting Specifications Product Dimensions TABLE OF CONTENTS Important Safety Instructions...1 1.1 Safety precautions and maintenance....1 1.2 Use.......4 1.3 Installation Notes.......7 Package Content...9 2.1 Unpacking...9 2.2 Accessories......10

More information

May result in personal injury or property loss. Indicate the behaviors that must be prohibited

May result in personal injury or property loss. Indicate the behaviors that must be prohibited User Manual VXM143U Please read this manual carefully for proper usage before lnstallatlon and use. It is recommended to keep this manual for future check. The pictures in this manual are only for reference.

More information

PLL2710W LED LCD Monitor

PLL2710W LED LCD Monitor PLL2710W LED LCD Monitor USER'S GUIDE www.planar.com Content Operation Instructions...1 Safety Precautions...2 Package Overview...3 First Setup...4 Front View of the Product...5 Rear View of the Product...6

More information

v1.0.0 January AlphaLab, Inc. All rights reserved. TriField EMF Meter Owner s Manual

v1.0.0 January AlphaLab, Inc. All rights reserved. TriField EMF Meter Owner s Manual v1.0.0 January 2018 2018 AlphaLab, Inc. All rights reserved. TriField EMF Meter Owner s Manual TABLE OF CONTENTS Overview... 1 Introduction... 1 Features... 1 Applications... 1 Using the TriField EMF Meter...

More information

VITEK VTM-TLM191 VTM-TLM240

VITEK VTM-TLM191 VTM-TLM240 VTM-TLM191 VTM-TLM240 19 & 24 Professional LED Monitors with HDMI, VGA, and Looping BNC VITEK FEATURES 19 & 24 Wide Screen LED Display Panel HDMI, VGA, and Looping BNC Composite Video Inputs & Stereo Audio

More information

42 Freestanding Infrared Multi Touch Screen Kiosk User s Manual

42 Freestanding Infrared Multi Touch Screen Kiosk User s Manual 42 Freestanding Infrared Multi Touch Screen Kiosk User s Manual Manual Version L42HD-T2.2 Safety Instructions Please keep the display away from any heat sources such as radiators or direct sunlight. Place

More information

This device is a lighting effect for professional use on stages, in discotheques, theatres, etc..

This device is a lighting effect for professional use on stages, in discotheques, theatres, etc.. At first, thank you for purchasing QUADRO 24. For the sake of safety and regular operation of this projector, please read this manual carefully before using and operating it, lest incur any personal injury

More information

TFT LCD MONITOR USER MANUAL. L80AP and L101AP

TFT LCD MONITOR USER MANUAL. L80AP and L101AP TFT LCD MONITOR USER MANUAL L80AP - 8.0 and L101AP - 10.1 Table Of Contents Table of contents/ Warning.... 2 Precautions...3 About this user manual and products / Items included in the delivery..... 4

More information

SLS-50 Stabilized Laser Source. User s Manual. Shineway Technologies, Inc. All rights reserved.

SLS-50 Stabilized Laser Source. User s Manual. Shineway Technologies, Inc. All rights reserved. SLS-50 Stabilized Laser Source User s Manual Shineway Technologies, Inc. All rights reserved. Safety Instructions The WARNING sign denotes a hazard. It calls attention to a procedure, practice, or the

More information

Littmann Electronic Stethoscope

Littmann Electronic Stethoscope Littmann Electronic Stethoscope Model 4000 Superior acoustics Amplification up to 18 times greater than best conventional Littmann stethoscopes Three frequency response modes: bell, diaphragm, and extended

More information

TE-4HCM-S User Manual. l ::: J TECH SUPPORT. MetraDealer.com

TE-4HCM-S User Manual. l ::: J TECH SUPPORT. MetraDealer.com TE-4HCM-S User Manual ibeamusa.com MetraDealer.com techsupport@metra-autosound.com TECH SUPPORT l ::: J 800-253-8324 Contents 1. Precautions...3 2. Accessories... 5 3. Connections... 6 4. Menu operation

More information

Instruction Manual. Electronic Level - Inclinometer Model No

Instruction Manual. Electronic Level - Inclinometer Model No Electronic Level - Inclinometer Model No. 40-6060 Instruction Manual Congratulations on your choice of this Electronic Level - Inclinometer. We suggest you read this instruction manual thoroughly before

More information

DMX LED light effect with 4 lenses

DMX LED light effect with 4 lenses DMX LED light effect with 4 lenses User manual 1 Safety precautions WARNING: This unit may cause serious injury to the eyes when used incorrectly. It is therefore strongly advised to read this user manual

More information

Tablet Oscilloscope Quick Guide

Tablet Oscilloscope Quick Guide Tablet Oscilloscope Quick Guide For tbook Series Shenzhen Micsig Instruments Co., Ltd. Copyright Copyright Shenzhen Micsig Instruments Co., Ltd. All Rights Reserved. Version Version: MKX2014-001; Product

More information