1 A.- F. Miller 2012 T1&T2 Measurement 1 Practicum 3, Fall 2012 Measuring the longitudinal relaxation time: T1. Strychnine, dissolved CDCl3 The T1 is the characteristic time of relaxation of Z- magnetization, also known as the longitudinal relaxation time. Knowledge of T1 enables you to make an informed choice of the duration of delay between scans. If too short a relaxation delay is chosen, then the peaks of slowly recovering protons will be smaller than they should be, and it will not be possible to compare peak areas to learn about proton numbers. Also, we use the T1s our first estimate for choosing mix times in NOESY spectra, and 3xT1 as our relaxation delay in NOESY spectra. Hence the need to know T1. Start with a beautiful 1d spectrum, and a calibrated pw90 (recall that this varies with the choice of power). (It turns out that the default value in the probe file is a good compromise for people in a rush or not bothered about perfection.) Fine point: phase up the spectrum and then type setlp0. This sets up the delay between the end of the read pulse and the start of data collection so that no zero- order phase correction is needed and all spectra will be positive absorptive. Recollect to embed the new delay in your parameter set. Under the Experiments choose 'convert current parameters to do... > Relax. Measurements > T1' This will retain your current sw, tof and pw90, but it will add a few features to the experiment to support measurement of the magnetization's recovery after inversion. To see the pulse sequence click Sequence in Figure 1. The first pulse is a 180 pulse (p1=2*pw90) that will invert magnetization. This places it in a high- energy state with respect to the field, so it will 'decay' back to the resting state. Time is allowed for this process, in the form of a delay called d2. So far all magnetization is still along Z, but its amplitude is recovering with time. To determine what the amplitude is after a given time d2, we use a 90 pulse (pw) [4] to 'read' our magnetization by rotating it into the XY plane where our receiver coils detect it [5]. We cannot monitor recovery continuously, instead we repeat the experiment 6 or more times, waiting a different amount of time d2 in each repeat before executing the read pulse. Thus we end up with a series of 1d spectra, each representing a different duration of recovery after the inversion pulse. In Acquire>Defaults, a set of three boxes invites you to enter first- guesses of the T1s you sample may have, and Agilent uses these to set the delays in the experiment to perform a measurement. Set T1 Mode to inversion recovery [6] (ignore the NOE choice unless you are measuring 13 C T1s. [7]), then set Min T1 to the shortest T1 you expect to be present among your resonances (say 0.5 sec.: enter 0.5) [8], Max T1 to the longest T1 you anticipate (err on the long side here, say 5 (seconds)) [9] and the time you are willing to devote to the experiment (eg. 0.1 (hours)) [10].
2 A.- F. Miller 2012 T1&T2 Measurement 2 The Min. and especially the Max times will be used to choose an array of d2 values ranging from very short to something on the order of your maximum estimate. You can see these by clicking on 'Arrays'. [11]. The experiment takes a long time in part due to the long relaxation time d1 allowed between scans [12]. This is chosen to be 5 times your estimated longest T1 you guessed ([9]). The duration of the experiment also increases in proportion to the number of scans collected for each value of the delay d2. Although the number of scans appears in this panel, you are not allowed to change it to something smaller than 4. For a good value of T1 it is wise to use multiples of 4 scans, but for a first quick check on a strong sample one scan will do. Therefore, go to Acquire>Acquisition and change the number of scans to 1. Make sure that gain is a number, not autogain [13]. You may also want to go into the 'Array' panel and decrease the number of values of d2 used (Click [11] in Figure 1). Because we are expecting an exponential recovery, we anticipate that an exponential array will be a better choice than a linear array (Figure 2). Thus in this case successive values of d2 are 2x the previous value, rather than being some value plus the previous value. The increment is the ln(2). If you want to triple each value to get the next one, use an increment of ln(3)=1.1. If I am rushing I choose an array size of 6 (the minimum allowed) and an increment 1.1. The resulting array covers a factor of approximately 700. Thus for a longest estimate of 3 seconds I want a shortest value of I go back and make adjustments to the first value and increment as needed to get the last value I want. As always, close the Array box, check Sequence, Show Time and then Acquire. The result should be a series of spectra in which we see that spins are initially inverted and then gradually recover as the delay d2 is made longer (Figure 3). In the LHS (left- hand side) panel, with the 'Arrayed Spectra' tab active [0], choose to display spectra stacked vertically (activate the radio button ( ). Check the 'values' box to show the duration of d2 corresponding to each spectrum. Change values in 'Chart Dimensions' to alter the vertical spacing and heights of the individual spectra. For Vert. Pos. use 20 (from 90) Vert Height 300 [4] (from 900). Under Offset Vertical change this interactively (step size of 10) [5] with +10, it gives 17.7 (a good display) and in "Show". You see that different peaks recover in different times. The null represents approx T1 *0.7. Divide null time by 0.7 to estimate T1 1 sec/.7 = 1.4 sec for peak at 6 ppm. 0.5 s /.7 = 0.7 S for aliphatics. In Process>T1 Analysis (Figure 4 ), click on Display Last Spectrum. It should be upright as shown. If it is not, you will have to phase it manually (after activating the additional options on the right- hand- side of the graphical display window ). Figure 4 shows a case where the software picked a ridiculous number of lines.
3 A.- F. Miller 2012 T1&T2 Measurement 3 Activate the action buttons on the RHS of the graphics display window and activate threshold (Figure 5 ). Move the threshold up to be more selective. You can also use the cursors to restrict your analysis to one region of the spectrum by placing cursors around it and clicking the '+' magnifying glass. Then click on 'Do T1 Analysis' [4]. All the lines the instrument picked will be in the table in the RHS box [5]. First, all the picked peaks are listed and each is assigned an 'index' number, which will be used below to provide the T1 for the corresponding peak. Next there is a summary of the analysis in which each peak's T1 is provided along with the associated error. Finally a detailed analysis follows for each picked peak (just beginning to appear in the box [5]). You can see the recovery of each peak after inversion by clicking "Display all Peaks" [6], this is valuable because it allows you to confirm that the recoveries do in fact conform to single- exponential recoveries. However there is often too much data to be able to distinguish different build- ups. You can also display the data for selected resonances only. Using the chemical shifts provided with the spectrum (Figure 6 ) in conjunction with the numerical output telling you which index number corresponds to each chemical shift value you can select a few lines and plot the peak amplitude vs. duration of recovery d2 for them. Next to 'Display Selected Fits' enter the line numbers for the lines you would like information on, for example "1, 10, 36". Then <return>. A time course of magnetization recovery is produced for the peaks number 1, 10 and 36 in the line list (Figure 7). Don't believe T1 values derived from curves that don't fit the data well. Don't believe any values whose errors are more than 10% the value (read the numerical output in the summary of analysis section). Don't believe a T1 that are more than one third of the recovery time you used between scans (=d1+at). If the T1s you get are longer than (d1+at)/3, repeat the experiment with a longer d1, roughly 5 times the longest T1 you got. To get a copy of the graphical output, use the drag- down menu at the top of the screen File>Print Screen. In the window that opens (Figure 8) select PSland, ( ) file, give name with path (eg. che555/data/yourname/yourfile (omit the.ps), POSTSCRIPT, Mono/Color... Save, then Close. A digital copy of the graphics window will be saved for you to place in documents, presentations etc. If you want a white background, be sure to change the graphics window in advance (and change it back afterwards) by dragging down under Edit to Display. It is often helpful to be able to take the peak amplitudes out of the Agilent software for analysis in a spread- sheet or analysis software of your choice. To do this, select all the text you want in the text output window (Figure 9, ). Copy it (Ctrl C). Then go to the Linux drag- down menu called 'Applications' and select Accessories, then Text Editor. In the editor, Paste (Ctrl V), and the text you selected should appear. Save this document in your own folder in CHE555/data/YOUR- INITIALS.
4 A.- F. Miller 2012 T1&T2 Measurement 4 As usual, you can save your data set as a.fid file using the diskette icon in upper left of the VNMR window. Please save all your things in the directory with your own initials, in CHE555/data. Illustration of why the relaxation delay must be longer than T 1 If you do not allow sufficient time between scans for resonances to recover, they will be smaller in subsequent scans. If the total recycle time (d1+at) is shorter than the T1s of some resonances but longer than the T1s of others, then the former resonances will be shrunken more than the others, in your spectra. Thus, peak areas will no longer be proportional to the numbers of protons associated with each peak. Figure 11 compares the results of using four different d1 values all in combination with an at of 0.5. Thus, the total recycle times were 11.5, 3.8, 1.5 and 0.8 s (for d1 values of 11, 3.3, 1.0 and 0.3 s). The results show that the 0.8 s recycle time causes aromatic resonances, in particular, to be proportionaltely smaller (Figure 12). Note that the TMS line is also much smaller compared to the other lines when recycling is fast. The above partial saturation of slow- relaxing resonances can be prevented either by choosing a longer (d1 + at) or by using a smaller tip- angle pulse. If you use a 45 pulse then you do not have to wait as long between scans, because magnetization does not need to recover as much. Using a smaller tip angle in combination with a shorter recycle time allows you to get the best data per hour of spectrometer time. This is invaluable for weak spectra such as 13 C 1ds, or samples that are not stable. Once you know T1 the software can calculate the optimal pulse width and tip angle given a choice of d1. In the command line type 'ernst(3.4, 13.6)' for the example of a case where the T1 is 3.4 s. and the pw90 is 13.6 us. Thus the general form of the command is 'ernst(t1_est, pw90)'. This causes the software will update the pulse width and tip angle of the experiment for maximal sensitivity per hour of spectrometer time, given the d1 value in your parameter set. Measuring the transverse relaxation time T2 As for T1 measurement, begin with a beautiful 1d spectrum and a calibrated pw90. For best results phase up a nice 1d spectrum and type setlp0 to obtain delays that automatically produce an upright spectrum, then recollect to embed the new delay in your parameter set. Although the NMR experiment is a different one, the workflow is very similar to that employed to measure T1. But first... have a look at the FID from your beautiful 1d.
5 A.- F. Miller 2012 T1&T2 Measurement 5 Figure 12 shows that the amplitude, whether positive or negative, decays with what looks like a single exponential (roughly). We can estimate the characteristic time of the decay by looking for the time at which the amplitude has dropped to half of its starting value, in this case 0.2 s. This gives an estimate for T2, called T2* of T2*=τ1/2/0.7 = 0.2/ s. However this is not the actual T2 because the FID amplitude decays not only to spin randomization but also due to magnetic field inhomogeneity, J- coupling, and other coherent effects. To measure the actual T2s of the spins, and thus get at a number that informs us about chemical exchange and other dynamics, we need to cancel out the other effects (more in lecture). To do this, we use a Carr- Purcell- Meiboom- Gill (CPMG) refocussing sequence. Obtain the NMR experiment employing the CPMG, use the draw- down menu Experiments>Convert current param. to do...>relax. Measurements>T2 Meas. Figure 13 shows the NMR experiment. To measure T2 (rather than T1) the first thing we do is employ a 90 pulse to put magnetization into the XY plane. Next we want to let it relax (transverse relaxation) and we will collect a whole series of spectra each resulting from a different duration of relaxation. During the relaxation we have the CPMG pulses going and the pulse sequence shows this train of pulses as the contents of the red parentheses (think of the drumming that runs in the background of rock songs). The number above and to the left of the red parentheses tells us how many times the parenthetical pulse train is repeated. Because there needs to be an integer number of repeats, the delay d2 can only be multiples of the duration of one repeat. Software takes care of this for us. In Acquire>Defaults [4], we enter our best guesses of the shortest T2 we expect [5] and the longest T2 [6], we also enter the amount of time we are willing to spend varies depending on the accuracy desired. For 1 H work we can ignore the NOE choices [7] (check in Acquire>Channels that this is true and see the Observe/Decouple statement). Then click "Array Relaxation Delay" [8], and the software will set up an array with a series of d2 values. Click on the Arrays button to see the d2 values suggested and make changes if you want (see above). Also change the number of scans as needed, in Acquire>Acquisition. Close the Array box, check Sequence, Show Time and then Acquire. When the experiment finishes it will automatically perform the analysis based on its choice of threshold etc. However we can look behind the scenes to see the spectra collected (Figure 14). In this display (produced using "arrayed Spectra" ) the result of a short relaxation delay is on the bottom and results of increasingly long delays are the spectra are stacked vertically above (showing the delay duration used for each spectrum ). Note that different peaks decay at different rates, in that the resonances on the left- hand side are essentially gone after 6.4 seconds whereas the three peaks on the right have decreased by a factor of 4 but are still visible. Figure 15 shows the spectrum that will appear when the experiment completes (Assumes that Acquire>Future Actions contains "process" as the "when experiment
6 A.- F. Miller 2012 T1&T2 Measurement 6 finishes".) As often, the software has chosen a reasonable threshold for peak detection, but it may not produce the results you want. In Process>T2 Analysis, you may need to go through the ppm peak list to identify the indices of the peaks you care about so that you can look up the T2s of those values in the large list of T2s for all peaks [4]. It is helpful when sorting through long lists of peaks, to have the spectrum itself be labeled. You can type dpf in the command line [5] to get ppm labels on the peaks, as shown here. If you want more peaks you will need to set a threshold yourself. As shown in Figure 16, click on "Display First Spectrum", activate the spectral analysis tools that normally appear to the right of the spectrum by clicking on the top icon, this will cause the rest of the icons to appear, including the one that allows you to set a threshold. After setting the threshold, click on "Do T2 Analysis" [4]. In this example, I raised the threshold so that only 6 peaks were analyzed [5]. In the text output box I have scrolled down a little [6] to show that beneath the list of peaks, and the list of their T2s and associated errors, the software provides details of the quality of the fits in the form of the calculated amplitude for each relaxation time for comparison with the actual amplitude [7]. As described for T1 analysis, you can select all of this text, copy it and paste it into a text file that you can copy to a thumb- drive and then read into spread sheet of mathematical software for more detailed analysis. Figure 17 shows the same text data output a little higher up, where the summary of the T2 results appears. Note the modest differences among the T2s of different resonances. In molecules where a proton is exchanging with solvent, or a portion of the molecule undergoes intermediate time- scale motions, you can see T2 values that are 10- fold shorter (or more) than the T2s of rigid non- exchanging Hs. Upon clicking "Display All Fits" (Figure 18) we see that the model (single exponential decay) provides a good description of the data, indicating that the T2 values have value, this along with the relatively low errors make these values trustworthy. Knowledge of the T2s is helpful when setting up 2D experiments that require resonances to persist in the XY plane for some delay within the pulse sequences. Examples include COSY and TOCSY.
7 [4] [5] T1 and T2, Figure 1 [6] [8] [9] [7] [10] [11] [12] [13] T1 and T2, Figure 2
8 [0] T1 and T2, Figure 3 [4] [5] T1 and T2, Figure 4
9 T1 and T2, Figure 5 [5] [6] [4] T1 and T2, Figure 6
10 T1 and T2, Figure 7 T1 and T2, Figure 8
11 T1 and T2, Figure 9 T1 and T2, Figure 10
12 T1 and T2, Figure 11 Recovery delay (s) Often the relevant time is d1 + at. T1 and T2, Figure 12
13 T1 and T2, Figure 13 [4] [5] [7] [6] [8] T1 and T2, Figure 14
14 [5] T1 and T2, Figure 15 [4] T1 and T2, Figure 16 [5] [6] [8] [4] [7]
15 T1 and T2, Figure 17 T1 and T2, Figure 18
Practicum 3, Fall 2010
A. F. Miller 2010 T1 Measurement 1 Practicum 3, Fall 2010 Measuring the longitudinal relaxation time: T1. Strychnine, dissolved CDCl3 The T1 is the characteristic time of relaxation of Z magnetization
More informationProcessing data with Mestrelab Mnova
Processing data with Mestrelab Mnova This exercise has three parts: a 1D 1 H spectrum to baseline correct, integrate, peak-pick, and plot; a 2D spectrum to plot with a 1 H spectrum as a projection; and
More informationBasic 13 C Acquisition and Processing 4
Chapter Basic 13 C Acquisition and Processing 4 Goto Introduction 4.1 This chapter describes the acquisition and processing of a 13 C spectrum acquired with a simple one-pulse sequence with and without
More informationProcedure for Acquiring 2D Inverse Detection Spectra HMQC, HMBC and HSQC
Procedure for Acquiring 2D Inverse Detection Spectra HMQC, HMBC and HSQC This protocol is for qualified users operating Bruker Avance 400 in IBC only. Dr. Jao STOP accepts no responsibility for actions
More informationIMSERC NMR MANUAL 05: Manual Operation of Agilent NMR Spectrometers (Chem350 Interface)
IMSERC NMR MANUAL 05: Manual Operation of Agilent NMR Spectrometers (Chem350 Interface) Last updated: October 12, 2011 by Josh Kurutz THIS PAGE = QUICK START GUIDE 0) At the computer, make sure VNMRJ is
More informationMestReNova A quick Guide. Adjust signal intensity Use scroll wheel. Zoomen Z
MestReNova A quick Guide page 1 MNova is a program to analyze 1D- and 2D NMR data. Start of MNova Start All Programs Chemie NMR MNova The MNova Menu 1. 2. Create expanded regions Adjust signal intensity
More informationIntroduction to 1D & 2D NMR using VnmrJ
Introduction to 1D & 2D NMR using VnmrJ Scott R. Burt September 12, 2011 This document is an introductory tutorial for collecting the most important NMR spectra in organic chemistry: (1D) 1 H, 13 C, NOESY,
More information1 H HOMODECOUPLING EXPERIMENT
1 H HOMODECOUPLING EXPERIMENT This is a simple and quick means of determining if two resonances are coupled. The HOMODEC experiment is most effective for relatively simple spectra where the couplings are,
More informationMestReNova Manual for Chem 201/202. October, 2015.
1. Introduction to 1-D NMR Data Processing with MestReNova The MestReNova program can do all of the routine NMR data processing needed for Chem 201 and 202 and will be available through the Reed downloads
More informationcyclenoe 1D NOE Difference Experiment
cyclenoe 1D NOE Difference Experiment NOTE: If you are interested in qualitative or long-range NOE s, you should use the NOESY1D experiment (see NOESY1D handout for experimental details). However, you
More informationPINMRF. Checkout Quiz - Varian Inova-300 Version
PINMRF Checkout Quiz - Varian Inova-300 Version Please carefully read every question and select the best answer(s). Some questions may have more than one correct answer. You must select all the correct
More informationPulseCounter Neutron & Gamma Spectrometry Software Manual
PulseCounter Neutron & Gamma Spectrometry Software Manual MAXIMUS ENERGY CORPORATION Written by Dr. Max I. Fomitchev-Zamilov Web: maximus.energy TABLE OF CONTENTS 0. GENERAL INFORMATION 1. DEFAULT SCREEN
More informationNMR. picospin. Maintenance Guide
NMR picospin Maintenance Guide 269-302600 Revision A January 2013 2013 Thermo Fisher Scientific Inc. All rights reserved. For U.S. Technical Support, please contact: Thermo Fisher Scientific 5225 Verona
More informationStandard Operating Procedure of nanoir2-s
Standard Operating Procedure of nanoir2-s The Anasys nanoir2 system is the AFM-based nanoscale infrared (IR) spectrometer, which has a patented technique based on photothermal induced resonance (PTIR),
More informationVnmrj Training Guide. Vnmrj overview
Vnmrj Training Guide Eugenio Alvarado. Department of Chemistry, University of Michigan. 10/29/10, 05/05/11 This manual describes the most basic operation of our Varian NMR spectrometers with Vnmrj 2.2D
More informationPS User Guide Series Seismic-Data Display
PS User Guide Series 2015 Seismic-Data Display Prepared By Choon B. Park, Ph.D. January 2015 Table of Contents Page 1. File 2 2. Data 2 2.1 Resample 3 3. Edit 4 3.1 Export Data 4 3.2 Cut/Append Records
More informationFall 2014 Chem 636 Lab #4
latest update: 140922 (cgf) Fall 2014 Chem 636 Lab #4 Assignment due at beginning of lab, week 5 (Sept 30 Oct 2). Use Athena (AC-300) for section I, and Persephone (AVANCE-500) for Sections II and III.
More informationBitWise (V2.1 and later) includes features for determining AP240 settings and measuring the Single Ion Area.
BitWise. Instructions for New Features in ToF-AMS DAQ V2.1 Prepared by Joel Kimmel University of Colorado at Boulder & Aerodyne Research Inc. Last Revised 15-Jun-07 BitWise (V2.1 and later) includes features
More informationRAD 465 (MRI) Lecture one (Pulse Sequences) Ruba Khushaim MSc
RAD 465 (MRI) Lecture one (Pulse Sequences) Ruba Khushaim MSc Outline : Spine echo pulse sequence SE Fast spin echo pulse sequence FSE Inversion recovery pulse sequence IR Gradient pulse sequence GS Pulse
More informationThe Measurement Tools and What They Do
2 The Measurement Tools The Measurement Tools and What They Do JITTERWIZARD The JitterWizard is a unique capability of the JitterPro package that performs the requisite scope setup chores while simplifying
More informationTips and Tricks Part-II: Bits and pieces
Tips and Tricks Part-II: Bits and pieces Clemens Anklin Pittsburgh April 2016 Innovation with Integrity Overview 1H -13C correlation experiments Which ones to choose and a few pitfalls. Non-uniform or
More informationSystem Requirements SA0314 Spectrum analyzer:
System Requirements SA0314 Spectrum analyzer: System requirements Windows XP, 7, Vista or 8: 1 GHz or faster 32-bit or 64-bit processor 1 GB RAM 10 MB hard disk space \ 1. Getting Started Insert DVD into
More informationNMR TRAINING MANUAL LINUX-BASED SPECTROMETERS. And VnmrJ 2.x TEXAS A&M UNIVERSITY DEPARTMENT OF CHEMISTRY
NMR TRAINING MANUAL LINUX-BASED SPECTROMETERS And VnmrJ 2.x TEXAS A&M UNIVERSITY DEPARTMENT OF CHEMISTRY November 9, 2009 TABLE OF CONTENTS 1. OVERVIEW 1 1.1 WHAT IS FTNMR? 1 1.2 MULTITASKING 1 1.3 DUAL
More informationE X P E R I M E N T 1
E X P E R I M E N T 1 Getting to Know Data Studio Produced by the Physics Staff at Collin College Copyright Collin College Physics Department. All Rights Reserved. University Physics, Exp 1: Getting to
More informationLinrad On-Screen Controls K1JT
Linrad On-Screen Controls K1JT Main (Startup) Menu A = Weak signal CW B = Normal CW C = Meteor scatter CW D = SSB E = FM F = AM G = QRSS CW H = TX test I = Soundcard test mode J = Analog hardware tune
More informationPre-processing of revolution speed data in ArtemiS SUITE 1
03/18 in ArtemiS SUITE 1 Introduction 1 TTL logic 2 Sources of error in pulse data acquisition 3 Processing of trigger signals 5 Revolution speed acquisition with complex pulse patterns 7 Introduction
More informationSEM- EDS Instruction Manual
SEM- EDS Instruction Manual Double-click on the Spirit icon ( ) on the desktop to start the software program. I. X-ray Functions Access the basic X-ray acquisition, display and analysis functions through
More informationMIE 402: WORKSHOP ON DATA ACQUISITION AND SIGNAL PROCESSING Spring 2003
MIE 402: WORKSHOP ON DATA ACQUISITION AND SIGNAL PROCESSING Spring 2003 OBJECTIVE To become familiar with state-of-the-art digital data acquisition hardware and software. To explore common data acquisition
More informationClassical simulations with PJNMR BCMB/CHEM LAB 01/25/12
Classical simulations with PJNMR BCMB/CHEM LAB 01/25/12 The program PJNMR was written by Paul- Jean Letourneau as a joint project between the PENCE & NANUC projects in Canada. It rotates vectors representing
More informationFigure 1. MFP-3D software tray
Asylum MFP-3D AFM SOP January 2017 Purpose of this Instrument: To obtain 3D surface topography at sub-nanometer scale resolution, measure contact and friction forces between surfaces in contact, measure
More informationFor the SIA. Applications of Propagation Delay & Skew tool. Introduction. Theory of Operation. Propagation Delay & Skew Tool
For the SIA Applications of Propagation Delay & Skew tool Determine signal propagation delay time Detect skewing between channels on rising or falling edges Create histograms of different edge relationships
More informationUser Manual VM700T Video Measurement Set Option 30 Component Measurements
User Manual VM700T Video Measurement Set Option 30 Component Measurements 070-9654-01 Test Equipment Depot - 800.517.8431-99 Washington Street Melrose, MA 02176 - FAX 781.665.0780 - TestEquipmentDepot.com
More informationWhat s New in Raven May 2006 This document briefly summarizes the new features that have been added to Raven since the release of Raven
What s New in Raven 1.3 16 May 2006 This document briefly summarizes the new features that have been added to Raven since the release of Raven 1.2.1. Extensible multi-channel audio input device support
More informationPCIe: EYE DIAGRAM ANALYSIS IN HYPERLYNX
PCIe: EYE DIAGRAM ANALYSIS IN HYPERLYNX w w w. m e n t o r. c o m PCIe: Eye Diagram Analysis in HyperLynx PCI Express Tutorial This PCI Express tutorial will walk you through time-domain eye diagram analysis
More informationGS Bloch Equations Simulator 1. GS Introduction to Medical Physics IV Exercise 1: Discrete Subjects
GS02-1193 Bloch Equations Simulator 1 GS02-1193 Introduction to Medical Physics IV Exercise 1: Discrete Subjects Once SpinWright is running, select the Subject tab. The GUI display toward the top of the
More informationNanoTrack Cell and Particle Tracking Primer
NanoTrack Cell and Particle Tracking Primer The NanoTrack Pnode allows the user to track single cells and particles with nanometer precision at very fast tracking speeds. The speed of the tracking is dependent
More informationSigPlay User s Guide
SigPlay User s Guide . . SigPlay32 User's Guide? Version 3.4 Copyright? 2001 TDT. All rights reserved. No part of this manual may be reproduced or transmitted in any form or by any means, electronic or
More information#PS168 - Analysis of Intraventricular Pressure Wave Data (LVP Analysis)
BIOPAC Systems, Inc. 42 Aero Camino Goleta, Ca 93117 Ph (805)685-0066 Fax (805)685-0067 www.biopac.com info@biopac.com #PS168 - Analysis of Intraventricular Pressure Wave Data (LVP Analysis) The Biopac
More informationBrainMaster tm System Type 2E Module & BMT Software for Windows tm. Display Screens for Master.exe
BrainMaster tm System Type 2E Module & BMT Software for Windows tm Display Screens for Master.exe 1995-2004 BrainMaster Technologies, Inc., All Rights Reserved BrainMaster and From the Decade of the Brain
More informationStaMPS Persistent Scatterer Practical
StaMPS Persistent Scatterer Practical ESA Land Training Course, Leicester, 10-14 th September, 2018 Andy Hooper, University of Leeds a.hooper@leeds.ac.uk This practical exercise consists of working through
More informationAgilent DSO5014A Oscilloscope Tutorial
Contents UNIVERSITY OF CALIFORNIA AT BERKELEY College of Engineering Department of Electrical Engineering and Computer Sciences EE105 Lab Experiments Agilent DSO5014A Oscilloscope Tutorial 1 Introduction
More informationReference. TDS7000 Series Digital Phosphor Oscilloscopes
Reference TDS7000 Series Digital Phosphor Oscilloscopes 07-070-00 0707000 To Use the Front Panel You can use the dedicated, front-panel knobs and buttons to do the most common operations. Turn INTENSITY
More information013-RD
Engineering Note Topic: Product Affected: JAZ-PX Lamp Module Jaz Date Issued: 08/27/2010 Description The Jaz PX lamp is a pulsed, short arc xenon lamp for UV-VIS applications such as absorbance, bioreflectance,
More informationBurlington County College INSTRUCTION GUIDE. for the. Hewlett Packard. FUNCTION GENERATOR Model #33120A. and. Tektronix
v1.2 Burlington County College INSTRUCTION GUIDE for the Hewlett Packard FUNCTION GENERATOR Model #33120A and Tektronix OSCILLOSCOPE Model #MSO2004B Summer 2014 Pg. 2 Scope-Gen Handout_pgs1-8_v1.2_SU14.doc
More informationISOMET. Compensation look-up-table (LUT) and How to Generate. Isomet: Contents:
Compensation look-up-table (LUT) and How to Generate Contents: Description Background theory Basic LUT pg 2 Creating a LUT pg 3 Using the LUT pg 7 Comment pg 9 The compensation look-up-table (LUT) contains
More informationTopic: Instructional David G. Thomas December 23, 2015
Procedure to Setup a 3ɸ Linear Motor This is a guide to configure a 3ɸ linear motor using either analog or digital encoder feedback with an Elmo Gold Line drive. Topic: Instructional David G. Thomas December
More informationApplication Note AN-708 Vibration Measurements with the Vibration Synchronization Module
Application Note AN-708 Vibration Measurements with the Vibration Synchronization Module Introduction The vibration module allows complete analysis of cyclical events using low-speed cameras. This is accomplished
More informationDigital Storage Oscilloscopes 2550 Series
Data Sheet Digital Storage Oscilloscopes 2550 Series The 2550 series digital storage oscilloscopes provide high performance and value in 2-channel and 4-channel configurations. With bandwidth from 70 MHz
More informationA-ATF (1) PictureGear Pocket. Operating Instructions Version 2.0
A-ATF-200-11(1) PictureGear Pocket Operating Instructions Version 2.0 Introduction PictureGear Pocket What is PictureGear Pocket? What is PictureGear Pocket? PictureGear Pocket is a picture album application
More informationCMRRpack2. a collection of pulse sequences from the. University of Minnesota, Center for Magnetic Resonance Research (CMRR) Version: 2.
CMRRpack2 a collection of pulse sequences from the University of Minnesota, Center for Magnetic Resonance Research (CMRR) Version: 2.63_package This document generated: Mon Dec 16 2013 15:55:46 Contents
More informationGoals of tutorial. Introduce NMRbox platform
Introduce NMRbox platform Goals of tutorial Showcase NMRbox with NUS tools A dozen different NUS processing tools installed and configured more coming. Demonstrate potential of NMRbox Now that the platform
More informationNoise. CHEM 411L Instrumental Analysis Laboratory Revision 2.0
CHEM 411L Instrumental Analysis Laboratory Revision 2.0 Noise In this laboratory exercise we will determine the Signal-to-Noise (S/N) ratio for an IR spectrum of Air using a Thermo Nicolet Avatar 360 Fourier
More informationSPM Training Manual Veeco Bioscope II NIFTI-NUANCE Center Northwestern University
SPM Training Manual Veeco Bioscope II NIFTI-NUANCE Center Northwestern University Introduction: Scanning Probe Microscopy (SPM) is a general term referring to surface characterization techniques that utilize
More informationAgilent 86120B, 86120C, 86122A Multi-Wavelength Meters Technical Specifications
Agilent 86120B, 86120C, 86122A Multi-Wavelength Meters Technical Specifications March 2006 Agilent multi-wavelength meters are Michelson interferometer-based instruments that measure wavelength and optical
More informationThe BAT WAVE ANALYZER project
The BAT WAVE ANALYZER project Conditions of Use The Bat Wave Analyzer program is free for personal use and can be redistributed provided it is not changed in any way, and no fee is requested. The Bat Wave
More informationDektakXT Profilometer. Standard Operating Procedure
DektakXT Profilometer Standard Operating Procedure 1. System startup and sample loading: a. Ensure system is powered on by looking at the controller to the left of the computer.(it is an online software,
More informationGS122-2L. About the speakers:
Dan Leighton DL Consulting Andrea Bell GS122-2L A growing number of utilities are adapting Autodesk Utility Design (AUD) as their primary design tool for electrical utilities. You will learn the basics
More informationLab experience 1: Introduction to LabView
Lab experience 1: Introduction to LabView LabView is software for the real-time acquisition, processing and visualization of measured data. A LabView program is called a Virtual Instrument (VI) because
More informationWELDING CONTROL UNIT: TE 450 USER MANUAL
j WELDING CONTROL UNIT: TE 450 USER MANUAL RELEASE SOFTWARE No. 1.50 DOCUMENT NUMBER: MAN 4097 EDITION: MARCH 1998 This page is left blank intentionally. 2 / 34 TABLE OF CONTENTS SUBJECTS PAGE WELDING
More informationDurham Magneto Optics Ltd. NanoMOKE 3 Wafer Mapper. Specifications
Durham Magneto Optics Ltd NanoMOKE 3 Wafer Mapper Specifications Overview The NanoMOKE 3 Wafer Mapper is an ultrahigh sensitivity Kerr effect magnetometer specially configured for measuring magnetic hysteresis
More informationAgilent Agilent 86120B, 86120C, 86122A Multi-Wavelength Meters Data Sheet
Agilent Agilent 86120B, 86120C, 86122A Multi-Wavelength Meters Data Sheet Agilent multi-wavelength meters are Michelson interferometer-based instruments that measure wavelength and optical power of laser
More informationspiff manual version 1.0 oeksound spiff adaptive transient processor User Manual
oeksound spiff adaptive transient processor User Manual 1 of 9 Thank you for using spiff! spiff is an adaptive transient tool that cuts or boosts only the frequencies that make up the transient material,
More informationPractical Application of the Phased-Array Technology with Paint-Brush Evaluation for Seamless-Tube Testing
ECNDT 2006 - Th.1.1.4 Practical Application of the Phased-Array Technology with Paint-Brush Evaluation for Seamless-Tube Testing R.H. PAWELLETZ, E. EUFRASIO, Vallourec & Mannesmann do Brazil, Belo Horizonte,
More informationEASY-MCS. Multichannel Scaler. Profiling Counting Rates up to 150 MHz with 15 ppm Time Resolution.
Multichannel Scaler Profiling Counting Rates up to 150 MHz with 15 ppm Time Resolution. The ideal solution for: Time-resolved single-photon counting Phosphorescence lifetime spectrometry Atmospheric and
More informationimso-104 Manual Revised August 5, 2011
imso-104 Manual Revised August 5, 2011 Section 1 Getting Started SAFETY 1.10 Quickstart Guide 1.20 SAFETY 1.30 Compatibility 1.31 Hardware 1.32 Software Section 2 How it works 2.10 Menus 2.20 Analog Channel
More informationTorsional vibration analysis in ArtemiS SUITE 1
02/18 in ArtemiS SUITE 1 Introduction 1 Revolution speed information as a separate analog channel 1 Revolution speed information as a digital pulse channel 2 Proceeding and general notes 3 Application
More information127566, Россия, Москва, Алтуфьевское шоссе, дом 48, корпус 1 Телефон: +7 (499) (800) (бесплатно на территории России)
127566, Россия, Москва, Алтуфьевское шоссе, дом 48, корпус 1 Телефон: +7 (499) 322-99-34 +7 (800) 200-74-93 (бесплатно на территории России) E-mail: info@awt.ru, web:www.awt.ru Contents 1 Introduction...2
More informationUSB Mini Spectrum Analyzer User Manual PC program TSA For TSA5G35 TSA4G1 TSA6G1 TSA12G5
USB Mini Spectrum Analyzer User Manual PC program TSA For TSA5G35 TSA4G1 TSA6G1 TSA12G5 Triarchy Technologies, Corp. Page 1 of 17 USB Mini Spectrum Analyzer User Manual Copyright Notice Copyright 2013
More informationThe Kaffeine Handbook. Jürgen Kofler Christophe Thommeret Mauro Carvalho Chehab
Jürgen Kofler Christophe Thommeret Mauro Carvalho Chehab 2 Contents 1 Kaffeine Player 5 1.1 The Start Window...................................... 5 1.2 Play a File..........................................
More informationKJM Organic NMR Spectroscopy. Laboratory Experiments - part 1
KJM5250 - Organic NMR Spectroscopy 1 Laboratory Experiments - part 1 Pulse width, T1, Decoupling, DEPT, NOE Difference Spectroscopy (Xwinnmr on PC, version 3.5) on DPX 200 and DPX 300 Department of Chemistry
More informationimso-104 Manual Revised July 19, 2012
imso-104 Manual Section 1 Getting Started SAFETY 1.10 Quickstart Guide 1.20 SAFETY 1.30 Compatibility 1.31 Hardware 1.32 Software Section 2 How it works 2.10 Menus 2.20 Analog Channel 2.21 On / Off 2.22
More informationSOUND LABORATORY LING123: SOUND AND COMMUNICATION
SOUND LABORATORY LING123: SOUND AND COMMUNICATION In this assignment you will be using the Praat program to analyze two recordings: (1) the advertisement call of the North American bullfrog; and (2) the
More informationMAutoPitch. Presets button. Left arrow button. Right arrow button. Randomize button. Save button. Panic button. Settings button
MAutoPitch Presets button Presets button shows a window with all available presets. A preset can be loaded from the preset window by double-clicking on it, using the arrow buttons or by using a combination
More informationHow to Optimize Ad-Detective
How to Optimize Ad-Detective Ad-Detective technology is based upon black level detection. There are several important criteria to consider: 1. Does the video have black frames to detect? Are there any
More informationGuide to the 600 and 800 Note means enter key
Guide to the 600 and 800 Note means enter key 1. Check that no one else is already using the spectrometer! a. Is the Fid Flash panel flashing? b. Does the Acquisition Information panel says no acquisition
More informationISOMET. Compensation look-up-table (LUT) and Scan Uniformity
Compensation look-up-table (LUT) and Scan Uniformity The compensation look-up-table (LUT) contains both phase and amplitude data. This is automatically applied to the Image data to maximize diffraction
More informationProcesses for the Intersection
7 Timing Processes for the Intersection In Chapter 6, you studied the operation of one intersection approach and determined the value of the vehicle extension time that would extend the green for as long
More informationGuide to Analysing Full Spectrum/Frequency Division Bat Calls with Audacity (v.2.0.5) by Thomas Foxley
Guide to Analysing Full Spectrum/Frequency Division Bat Calls with Audacity (v.2.0.5) by Thomas Foxley Contents Getting Started Setting Up the Sound File Noise Removal Finding All the Bat Calls Call Analysis
More informationUSB Mini Spectrum Analyzer User s Guide TSA5G35
USB Mini Spectrum Analyzer User s Guide TSA5G35 Triarchy Technologies, Corp. Page 1 of 21 USB Mini Spectrum Analyzer User s Guide Copyright Notice Copyright 2011 Triarchy Technologies, Corp. All rights
More informationNOTICE: This document is for use only at UNSW. No copies can be made of this document without the permission of the authors.
Brüel & Kjær Pulse Primer University of New South Wales School of Mechanical and Manufacturing Engineering September 2005 Prepared by Michael Skeen and Geoff Lucas NOTICE: This document is for use only
More informationUSB Mini Spectrum Analyzer User Manual TSA Program for PC TSA4G1 TSA6G1 TSA8G1
USB Mini Spectrum Analyzer User Manual TSA Program for PC TSA4G1 TSA6G1 TSA8G1 Triarchy Technologies Corp. Page 1 of 17 USB Mini Spectrum Analyzer User Manual Copyright Notice Copyright 2013 Triarchy Technologies,
More informationSource/Receiver (SR) Setup
PS User Guide Series 2015 Source/Receiver (SR) Setup For 1-D and 2-D Vs Profiling Prepared By Choon B. Park, Ph.D. January 2015 Table of Contents Page 1. Overview 2 2. Source/Receiver (SR) Setup Main Menu
More information1.0 ThermoNicolet Nexus 670 FTIR Spectrometer Instructions
1.0 ThermoNicolet Nexus 670 FTIR Spectrometer Instructions 1.1 Click on the OMNIC icon to open the software. 1.2 Check that a signal is being measured by opening the Experiment Setup menu under the Collect
More informationThis guide gives a brief description of the ims4 functions, how to use this GUI and concludes with a number of examples.
Quick Start Guide: Isomet ims Studio Isomet ims Studio v1.40 is the first release of the Windows graphic user interface for the ims4- series of 4 channel synthezisers, build level rev A and rev B. This
More informationComputer Graphics Hardware
Computer Graphics Hardware Kenneth H. Carpenter Department of Electrical and Computer Engineering Kansas State University January 26, 2001 - February 5, 2004 1 The CRT display The most commonly used type
More informationExperiment PP-1: Electroencephalogram (EEG) Activity
Experiment PP-1: Electroencephalogram (EEG) Activity Exercise 1: Common EEG Artifacts Aim: To learn how to record an EEG and to become familiar with identifying EEG artifacts, especially those related
More informationLogic Analyzer Auto Run / Stop Channels / trigger / Measuring Tools Axis control panel Status Display
Logic Analyzer The graphical user interface of the Logic Analyzer fits well into the overall design of the Red Pitaya applications providing the same operating concept. The Logic Analyzer user interface
More informationPrecise Digital Integration of Fast Analogue Signals using a 12-bit Oscilloscope
EUROPEAN ORGANIZATION FOR NUCLEAR RESEARCH CERN BEAMS DEPARTMENT CERN-BE-2014-002 BI Precise Digital Integration of Fast Analogue Signals using a 12-bit Oscilloscope M. Gasior; M. Krupa CERN Geneva/CH
More information9070 Smart Vibration Meter Instruction Manual
9070 Smart Vibration Meter Instruction Manual Overall machine and bearing conditions: vibration values are displayed with color coded alarm levels for ISO values and Bearing Damage (BDU). Easy vibration
More informationPHI 5000 VersaProbe TM Operator s Guide
PHI 5000 VersaProbe TM Operator s Guide Part No. 705921 Rev. A Copyright 2006 ULVAC-PHI, INC. 370 Enzo, Chigasaki, JAPAN The PHI logo ( ) is a registered trademark of ULVAC-PHI, INC. Physical Electronics,
More informationQ Light Controller+ Positions and EFX explained
Q Light Controller+ Positions and EFX explained February 13 th, 2015 Author: Massimo Callegari 1.Introduction When a QLC+ project includes several moving heads or scanners, it is necessary to have the
More informationPQ-Box 100 Quick Start Instructions
PQ-Box 100 Quick Start Instructions These instructions are provided for the purpose on providing a quick start to PQ-Box 100 installation and operation. Please refer to the user handbook for full details.
More informationQuick Start Bruker Dimension Icon AFM
Do not remove Quick Start Bruker Dimension Icon AFM March 3, 2015 GLA Contacts Harold Fu (hfu@caltech.edu) Weilai Yu (wyyu@caltech.edu) Bruker Tech Support (AFMSupport@bruker-nano.com 800-873-9750) Watch
More informationAFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode)
AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode) 1. Log into the Log Usage system on the SMIF web site 2. Open Nanoscope 6.14r1 software by double clicking on the Nanoscope 6.14r1 desktop
More informationThe Cocktail Party Effect. Binaural Masking. The Precedence Effect. Music 175: Time and Space
The Cocktail Party Effect Music 175: Time and Space Tamara Smyth, trsmyth@ucsd.edu Department of Music, University of California, San Diego (UCSD) April 20, 2017 Cocktail Party Effect: ability to follow
More informationSIDRA INTERSECTION 8.0 UPDATE HISTORY
Akcelik & Associates Pty Ltd PO Box 1075G, Greythorn, Vic 3104 AUSTRALIA ABN 79 088 889 687 For all technical support, sales support and general enquiries: support.sidrasolutions.com SIDRA INTERSECTION
More informationData Collection Using APEX3. March 30, Chemical Crystallography Laboratory
Data Collection Using APEX3 Page 1 of 10 Data Collection Using APEX3 March 30, 2017 Chemical Crystallography Laboratory Author: Douglas R. Powell Data Collection Using APEX3 Page 2 of 10 Distribution Douglas
More informationPicoScope. User guide. Copyright 2005 Pico Technology Limited. All rights reserved. PSW044 v1.5
PicoScope User guide I PicoScope User Guide Table of Contents 1 Introduction...3...3 1 What is PicoScope?...3 2 Why use PicoScope?...4 3 Screen layout...4 4 Display area...5 5 Customisation...5 6 Exporting
More information1 Ver.mob Brief guide
1 Ver.mob 14.02.2017 Brief guide 2 Contents Introduction... 3 Main features... 3 Hardware and software requirements... 3 The installation of the program... 3 Description of the main Windows of the program...
More informationZONE PLATE SIGNALS 525 Lines Standard M/NTSC
Application Note ZONE PLATE SIGNALS 525 Lines Standard M/NTSC Products: CCVS+COMPONENT GENERATOR CCVS GENERATOR SAF SFF 7BM23_0E ZONE PLATE SIGNALS 525 lines M/NTSC Back in the early days of television
More information