Delmar Reynolds CardioDirect 12-S with CardioCollect Interface Tango M2 Interface Notes

Similar documents
Schiller AT-10 plus Tango M2 Interface Notes

GE CardioSoft (Version 6.01 or higher) Tango M2 Interface Notes

Philips StressVue Tango M2 Interface Notes

Burdick Quest Tango M2 Interface Notes

Pulse Biomedical QRS-Card Tango M2 Interface Notes

Cardioline Cube Stress Tango M2 Interface Notes

Medset FLASHLIGHT ERGO Tango M2 Interface Notes

Quinton Q-Stress Version 6.x.x Tango M2 Interface Notes

Marquette Case 12, Case 15, Case 16, and Centra Stress Systems Tango M2 Interface Notes

Mortara XScribe Versions 3.xx and 5.xx Tango M2 Interface Notes

COSMED Quark CPET metabolic cart Tango M2 Interface Notes

Mortara X-Scribe Tango+ Interface Notes

Welch Allyn CardioPerfect Workstation Tango+ Interface Notes

Honeywell HomMed Wireless Scale Quick Reference Guide

Absolute volume control is also possible with SP Controls option Audio Follow Video Preamplifer.

INTRODUCTION This procedure should only be performed if the instrument fails to meet the Performance Check tests for Output Zero or Offset Accuracy

M5-H002. Multiview T-35. DVB-T to PAL / 5 channels on all TV s

22" Touchscreen LED Monitor USER'S GUIDE

R Series Quick Reference Guide

PXL2760MW LED LCD Monitor

PXL2470MW LED LCD Monitor

1 Synchronising Xsens with the Delsys Trigno EMG System

USER MANUAL. 27 Full HD Widescreen LED Monitor L27ADS

IMPORTANT! This instruction guide explains how to install your CCTV system.

Overview. Know Your Oscilloscope. Front Panel. Rear Panel. Sharing Agilent s Resources with Engineering Educators

USER MANUAL. 22" Class Slim HD Widescreen Monitor L215DS

2 Connect your video source

PLL2210MW LED Monitor

PLL2710W LED LCD Monitor

WQ Series Water Quality Bench top Meters

June/July 2016 Med Tips

Wall Ball Setup / Calibration

PL2410W LCD Monitor USER'S GUIDE.

USER MANUAL Full HD Widescreen LED Monitor L236VA

Activity P32: Variation of Light Intensity (Light Sensor)

USER MANUAL Full HD Widescreen LED Monitor L215ADS

USER MANUAL Full HD Widescreen LED Monitor L215IPS

Configuring the Stack ST8961 VS Module when used in conjunction with a Stack ST81xx series display.

IRIG-B PTP Clock Converter Output Module Hardware Installation Manual

PLL1920M LED LCD Monitor

Modbus for SKF IMx and Analyst

Data Acquisition Instructions

J.M. Stewart Corporation 2201 Cantu Ct., Suite 218 Sarasota, FL Stewartsigns.com

User Manual MODEL: KK1500-TR. Touch Display LCD Monitor. Installation Guide. 15 Resistive Touch LCD Monitor

GWL/ Power Group Technology Solutions Stay Powered for the Future

Sibelmed Sibelsound. XS PT Test Strips. Save $ *Minimum buy 4boxes. 35 Ellen Street, WOLLONGONG NSW 2500 Fax orders to:

User Manual 15" LCD Open frame SAW Touch Monitor KOT-0150US-SA4W. Table of Contents

USER MANUAL. 27 Full HD Widescreen LED Monitor L270E

USER MANUAL. 28" 4K Ultra HD Monitor L28TN4K

TFT LCD USER MANUAL TFT-15 TFT-17 TFT-19

Quick Setup 1 Unpack the projector

Video to PC Adapter. Model: Operation Manual

2 Connect your video source

Table of Contents. Versa TILE & Versa DRIVE D2 Quick Start Manual

RQflex 10. Manual English

2 Connect your video source

EdgeConnect Module Quick Start Guide ITERIS INNOVATION FOR BETTER MOBILITY

Optical Design Manufacturing, Inc. Sprint 2.5. Alcatel-Lucent (ALU) Fiber Inspection and Testing Procedures. Technical Support Document

ASSEMBLY AND CALIBRATION

Quick Start Guide Building and Scheduling a Show

Amino. Digital Television. Sun City & Hilton Head Island

Cisco MX200/MX300/EX90 User Guide

SmartCrystal Cinema Neo

TeamWork Installation Guide

Installation Note. Agilent 89441A Option AYC Installation Kit

TeamWork Kits Installation Guide

APM CALIBRATION PROCEDURE Rev. A June 3, 2015

USER MANUAL. 27" 2K QHD LED Monitor L27HAS2K

SmartCrystal Cinema Neo

SC-HD-2A HDMI Scaler & Audio Embedder / Extractor

Cisco/Scientific Atlanta 330 & 430. HD Television. Sun City & Hilton Head Island. Q u i c k TELEPHONE TELEVISION INTERNET WIRELESS SECURITY

Exercise 1: Muscles in Face used for Smiling and Frowning Aim: To study the EMG activity in muscles of the face that work to smile or frown.

Integration of the Zetasizer µv and OmniFACE into an SEC system

Features of the 745T-20C: Applications of the 745T-20C: Model 745T-20C 20 Channel Digital Delay Generator

Camera Setup Instructions

P-2 Installing the monitor (continued) Carry out as necessary

Med Tips. PRODUCT INFORMATION FreeStyle Battery Warranty. Q: What is the warranty for my FreeStyle Battery?

Intelligent Pendulum Hardness Tester BEVS 1306 User Manual

Using LightSpace CMS for User Created Calibration LUTs

Video Conferencing. Participant Orientation Guide

Video Server SED-2100R/S. Quick Installation Guide

Total Chlorine Sensor

TELEDYNE HASTINGS INSTRUMENTS

AcoustiSoft RPlusD ver

Variwrap Controller Manual

PRECAUTIONS CONTENTS: Please follow these precautions:

HDMI over CAT5 HDBaseT Extender - RS232 - IR - Ultra HD 4K ft (100m)

RF Solution for LED Display Screen

MTS/T-BERD 8000 Platform Optical Spectrum Analyzer Modules

J6 User Manual. User Manual. Multi-Screen Splicing Processor J6. Xi an NovaStar Tech Co., Ltd. Rev1.0.1 NS

QUICK START GUIDE. IP Camera & NVR Bundle ENGLISH

Cisco 9865H Signal Decoder Troubleshooting Guide

Electrocardiogram (ECG) Version Date: 2017-FEB-15 Document Effective Date: 2017-MAY-15

MENU EXECUTE Shiloh Road Alpharetta, Georgia (770) FAX (770) Toll Free

Instructions for Use: Video Inspection Scope with Display

Displays Open Frame Monitor Model Number: AND-TFT-150Bxx

Optical Receiver Manual. Transmitter OP-OR212JSE. Shenzhen Optostar Optoelectronics Co., Ltd (Version 2)

Users Manual. Stroboscope. September 2014

Cable Calibration Function for the 2400B/C and 2500A/B Series Microwave Signal Generators. Technical Brief

Four-way Return Optical Receiver Manual. Transmitter OP-OR124RJ. Shenzhen Optostar Optoelectronics Co., Ltd (Version 1)

Transcription:

Delmar Reynolds CardioDirect 12-S with CardioCollect Interface Tango M2 Interface Notes To setup Tango M2 with the CardioDirect stress system, simply follow the directions below. 1. Verify Correct RS-232 and ECG Trigger Cables RS-232 Cable used to communicate with the CardioDirect. This cable enables the CardioDirect to prompt Tango M2 when it needs a BP measurement, and allows the Tango M2 BP reading to be transferred to the CardioDirect display and reports. o SunTech Part Number = 91-0013-01 o Connection Ends: CardioDirect Connection 9 pin female Tango M2 Connection 9 pin female ECG Trigger Cable provides the ECG signal from the CardioDirect to the Tango M2. o SunTech Part Number = 91-0066-01 o Connection Ends: CardioDirect Connection 25 pin male Tango M2 Connection BNC male Delmar Reynolds CardioDirect with CaridoCollect Tango M2 Interface Notes Page 1 of 5

Tango M2 Monitor RS-232 Connection ECG Trigger Connection Part # 91-0066-01 2. Connect the RS-232 Cable From: the RS-232 connection on the rear panel of the Tango M2. To: the PC port on the back of the CardioDirect system. 3. Connect the ECG Trigger Cable From: the BNC External ECG connection on the rear panel of the Tango M2. To: the port on the CardioCollect device. 3. Tango M2 Monitor Setup a. When the operating screen is displayed, press the SELECT button once. This will bring up the MAIN MENU screen. b. Using the UP or DOWN arrows, highlight MONITOR SET UP and press the SELECT c. Using the UP or DOWN arrows, highlight STRESS SYSTEM and press the SELECT d. Using the UP or DOWN arrows, scroll through the list until CardioDirect is highlighted and press the SELECT button to confirm the choice. e. Using the UP or DOWN arrows, select EXIT twice to return to the operating screen. 5. CardioDirect System Setup a. Start the CardioNavigator on your PC by clicking Start > Programs > Cardionavigator > Cardionavigator. b. Select File Options. c. Select Interface Blood Pressure. d. From there, select Tango. e. Then select the appropriate communication port. Delmar Reynolds CardioDirect with CaridoCollect Tango M2 Interface Notes Page 2 of 5

f. In Setup for Tango, set the Time for NIBP measurement(s) to 50 seconds. g. Check that the CardioNavigator software can communicate with the Tango M2 by using the Test h. You can use the Reynolds Medical CardioCollect device (v1.05 or later) as the ECG interface. In CardioCollect, select the following settings: QRS Trigger: on Polarity: negative Pulse width: 50 ms Delay: 2 ms FAQs for Tango M2 Q. The Tango M2 displays a Status Message. What does it mean and what do I do? A. See the elibrary within your Tango M2 for quick troubleshooting tips. You can find the elibrary under the Main Menu; select Monitor Setup E-Library Status Messages. You can also look in the Tango M2 User Manual under the Status Messages & Alarms section for details on each Status Message and solution. Q. The Tango M2 monitor returns a result of 0/0 after blood pressure (BP) measurements. What do I need to do to get a BP reading? A. There are certain noisy conditions where the Tango M2 cannot accurately measure BP. When the Tango M2 encounters these situations, it returns a reading of 0/0. Microphone placement is critical for reliable operation of the Tango M2; there are many places to find cuff placement help. a. See the elibrary within your Tango M2 for quick cuff placement tutorials. You can find the elibrary under the Main Menu; select Monitor Setup E-Library Tutorials. b. Look in the Tango M2 User Manual under the Using Tango M2 during a Stress Test section for details on each type of cuff; the Orbit-K and the Single Patient Use (SPU) kit. c. Follow the instructions in the Cuff Tutorial (located on the SunTech Medical website under Support Customer Technical Support Video Tutorials) for correct microphone placement. Q. Can I use a heart rate or blood pressure simulator to test whether the Tango M2 is working correctly with my stress system? A. You cannot use a heart rate or blood pressure simulator to test whether the Tango M2 is working with your stress system. The Tango M2 monitor requires that the ECG signal and the Korotkoff sounds, collected by the microphone in the cuff, originate from the same source, meaning the patient. Q. How can I adjust the brightness of the Tango M2 display? A. You can adjust the contrast of the Tango M2 display by following these steps: 1. When the operating screen is displayed, press the SELECT button once. This will bring up the Main Menu screen. 2. Using the UP or DOWN arrows, highlight Monitor Setup and press the SELECT 3. Using the UP or DOWN arrows, highlight Brightness and press the SELECT Delmar Reynolds CardioDirect with CaridoCollect Tango M2 Interface Notes Page 3 of 5

4. Using the UP or DOWN arrows, modify the contrast of the screen. When you are finished, and press the SELECT button to confirm the choice. 5. Using the UP or DOWN arrows, select EXIT twice to return to the operating screen. Q. How can I get MAP on my Tango M2? A. Register your Tango M2 Monitor online or through mail to receive MAP as a function on your Tango M2 monitor. Please note that due to FDA Regulations MAP is not available to the United States market. On the web go to SunTechMed.com Support Sales Support Product Registration Q. How do I clean the Orbit-K cuff after a stress test? A. You can do either of the following: 1. Use a medical grade mild disinfectant wipe on the cuff or spray a cleaning solution onto a cloth and wipe the cuff. Afterwards, lay flat or line dry. 2. Remove the bladder and microphone from the outer shell of the Orbit-K cuff. Machine wash the shell in warm water with a mild detergent (50-140 F or 10-60 C). Lay flat or line dry the cuff. Do not place the cuff in a dryer. Q. My Tango M2 displays a message, Please VERIFY CALIBRATION or Equipment Maintenance and Calibration Required. What do I do? A. Verification of Pressure Calibration is required every year to maintain the accuracy of the Tango M2 s BP readings. You will need to contact your local SunTech Medical Service Center for help. In addition, the following items will be needed to verify calibration. Equipment Required: 1. Calibrated electronic manometer or equivalent. 2. 500mL volume or the Orbit-K Adult Plus cuff wrapped around something that will not break or crush (no glass). 3. Hand Inflation Bulb with bleed valve. 4. Tubing, Tee pieces, and miscellaneous connectors or you can order the T-Tube Kit (SunTech Part # 98-0030-00). Service Centers: For customers in the Americas SunTech Medical, Inc. 507 Airport Boulevard Suite 117 Morrisville, NC 27560-8200 USA Phone: +1.919.654.2300 +1.800.421.8626 (toll free within US) Fax: +1.919.654.2301 Email: CustomerService@SunTechMed.com For customers in Europe, the Middle East, Africa, Asia, and the Pacific SunTech Medical, Ltd. Oakfield Industrial Estate Delmar Reynolds CardioDirect with CaridoCollect Tango M2 Interface Notes Page 4 of 5

Eynsham, Oxfordshire OX29 4TS UK England Phone: +44 (0) 1865.884.234 Fax: +44 (0) 1865.884.235 Email: CustomerService@SunTechMed.com Delmar Reynolds CardioDirect with CaridoCollect Tango M2 Interface Notes Page 5 of 5