NI 9238 Datasheet DATASHEET. 4 AI, ±500 mv, 24 Bit, 50 ks/s/ch Simultaneous

Similar documents
NI 9239 DATASHEET. 4 AI, ±10 V, 24 Bit, 50 ks/s/ch Simultaneous

USB Contents. Definitions SPECIFICATIONS. RF Power Sensor Device

NI 9244 Datasheet DATASHEET. 3 AI/1 Neutral, 400 Vrms L-N/690 Vrms L-L, 24 Bit, 50 ks/s/ch Simultaneous

Note In this document, the NI 9202 with spring terminal and the NI 9202 with DSUB are referred to inclusively as the NI 9202.

NI 9230 DATASHEET. 3 AI, ±30 V, 24 Bit, 12.8 ks/s/ch Simultaneous

NI 9251 with mini XLR

NI 9260 DATASHEET. 2 AO, 3 Vrms, 24 Bit, 51.2 ks/s/ch Simultaneous

Topology 1-wire 16 Bank 8 1, 1-wire Octal 16 1, 1-wire Quad 32 1, 1-wire Dual 64 1, 1-wire 128 1

NI 9246 DATASHEET. 3-Channel, 20 Arms Continuous, 24-Bit, Analog Input Module

This document lists specifications for the PXI-5691 RF amplifier.

This document lists specifications for the NI PXIe-1491 high-definition multimedia interface (HDMI) analyzer.

Specifications describe the warranted, traceable product performance over ambient temperature ranges of 0 C to 55 C, unless otherwise noted.

NI PXI/PXIe-2543 Specifications

R Series for PXI Express Digital RIO with Kintex-7 325T FPGA. ni.com/manuals. may impair the protection the NI PXIe-7822R provides.

2.7 GHz RF Vector Signal Analyzer with Digital Downconversion. This document lists specifications for the NI PXI-5661 RF vector signal analyzer.

Thermocouple and RTD Modules for Compact FieldPoint

NI R Series Multifunction RIO Specifications

Digital Input Modules for Compact FieldPoint

FP-AI-112 and cfp-ai-112

USER MANUAL. Inverted Pendulum Experiment CAPTIVATE. MOTIVATE. GRADUATE.

Used for writing unscaled or raw data and covers the range from negative full scale (0) to positive full scale (65,535).

Specification Value Comments

Sound and Vibration Data Acquisition

Sound and Vibration Data Acquisition

BNC-2120 INSTALLATION GUIDE. Connector Accessory for Multifunction DAQ Devices

Software. Documentation. Test Equipment

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

USER MANUAL. IP02 Base Unit Experiment. Set Up and Configuration CAPTIVATE. MOTIVATE. GRADUATE.

PMG. Vibration Energy Harvesters

Specifications. Reference Documentation. Performance Conditions

MAP110 AC-DC Series Data Sheet

Analog Input Module HART Ex i / I.S. Inputs, 8 Channels Type 9461/

STBART0200K standard analog input kit STB - 2 I - 15 bits + sign

Supplement. TDS1000 and TDS2000 Series User Manual Specifications Update

02/11/2015

PicoScope 3000 Series Automotive User guide

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

WVR500 Waveform/Vector Monitor

Agilent 86120B, 86120C, 86122A Multi-Wavelength Meters Technical Specifications

Datasheet. LED Transformers. LED Transformer 120W 24V V. LED Transformers

PSI-MOS-RS232/FO 850 E Serial to Fiber Converter

Bluetooth Tester CBT. Specifications. Specifications. Version January 2006

PicoScope 4000 Automotive PC Oscilloscopes

Single output models feature wide-range output adjustability to meet a wide variety of standard and user-specific output voltage requirements.

Advanced Test Equipment Rentals ATEC (2832)

Analog Input Module HART Ex n Inputs, 8 Channels Series 9461/15

32 Channel CPCI Board User Manual

DATA SHEET. The Advantys model STBDAI5260 Digital Input Module provides 2 isolated discrete input points that operate on a 115 VAC power source.

NI USRP-2950R/2952R/2953R/ 2954R

2 MHz Lock-In Amplifier

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

BNC-2090A QUICK START GUIDE. Rack-Mount Connector Accessory for E/M Series DAQ Devices

3500/42M Proximitor*/Seismic Monitor

MKH 8060 Directional studio microphone

3500/42E Vibration Monitor

FLIR CM82: 600 Amp Power Clamp Meter with VFD Mode

FREQUENCY CONVERTER 1/3 RACK-MOUNTED BLOCK CONVERTER. Narda-MITEQ FEATURES OPTIONS. Unit shown with Option 17. Unit shown without Option 17

Technical Data. HF Tuner WJ-9119 WATKINS-JOHNSON. Features

ThinkRF D GHz RF Downconverter

PicoScope 2000 Series PC Oscilloscopes

Instructions. P MHz 1X/10X Passive Probe

FLIR CM82: 600 Amp Power Clamp Meter with VFD Mode

Advantys STB DATA SHEET. Advantys STB Output Module. Description. Features

National Park Service Photo. Utah 400 Series 1. Digital Routing Switcher.

DT9857E. Key Features: Dynamic Signal Analyzer for Sound and Vibration Analysis Expandable to 64 Channels

Operating instructions

R&S TS-ISC In-System Calibration Kit On-site calibration solution for R&S CompactTSVP

R&S RT-Zxx High-Bandwidth Probes Specifications

SEL-3405 High-Accuracy IRIG-B Fiber-Optic Transceiver

PicoScope 4000 Series PC Oscilloscopes

Digital Lock-In Amplifiers SR850 DSP lock-in amplifier with graphical display

R&S ZN-Z85 Switch Matrix Specifications

4 MHz Lock-In Amplifier

3500/42M Proximitor * /Seismic Monitor Bently Nevada* Asset Condition Monitoring

100 MHz, 100 MS/s, 14-Bit Digitizer

BNC-2110 DESKTOP AND DIN RAIL-MOUNTABLE BNC ADAPTER

SIL-2 8-Ch Analog Input Series Thermocouple, High Level, Low Level

Agilent 87405C 100 MHz to 18 GHz Preamplifier

MULTIBAND 1/3 RACK-MOUNTED

PACSystems* RX3i Thermocouple Input Module, 12 Channels, IC695ALG412-CB

R&S ZN-Z154 Calibration Unit Specifications

ArmorPOINT RTD and Thermocouple Input Modules

PicoScope 6407 Digitizer

FP-QUAD-510. Features. Power Requirement OPERATING INSTRUCTIONS. 4-Axis, Quadrature Input Module

1/3 RACK-MOUNTED BLOCK CONVERTERS

NI GSM/EDGE+ Toolkit Specifications

PicoScope 6407 Digitizer

45 Hz to 500 Hz 1.0% Hz to 1 khz 2.0% + 3. Accuracy 0.09% + 2. Range / resolution. Range / resolution. Accuracy 1% + 2

R&S RT-Zxx High-Voltage and Current Probes Specifications

LXM23DU10M3X motion servo drive - Lexium 23 - single phase V - 1 kw - I/O

PACSystems* RX3i. Isolated Thermocouple Input Module, 6 Channels, IC695ALG306-EB Isolated Thermocouple Input Module, 12 Channels, IC695ALG312-EB

LC1D80F7 TeSys D contactor - 3P(3 NO) - AC-3 - <= 440 V 80 A V AC 50/60 Hz coil

ni.com Digital Signal Processing for Every Application

Process Transmitter RMA 422

USER GUIDE. AXIS T8640 Ethernet Over Coax Adaptor PoE+ ENGLISH

POWER. DA10-M Series. 10 Watts. Data Sheet. Total Power: 10 Watts Input Voltage: Vac # of Outputs: Single, 5 Volts SPECIAL FEATURES

Smart Cameras for Embedded Machine Vision

R&S ZN-Z151/-Z152/-Z153 Calibration Unit Specifications

RK-2 ENVIRONMENTAL DATA CONTACTLESS MAGNETOSTRICTIVE LINEAR POSITION TRANSDUCER WITH FLANGED HEAD. Main characteristics

Data Pattern Generator

Transcription:

DATASHEET NI 9238 Datasheet 4 AI, ±500 mv, 24 Bit, 50 ks/s/ch Simultaneous Screw-terminal connectivity Anti-alias filters 250 Vrms, CAT II, channel-to-channel isolation The NI 9238 is a 4-channel analog input module for NI CompactDAQ and NI CompactRIO chassis. The measurement range of 500 mv can be used with many low-voltage sensors such as current shunts and current transducers in high-current applications. Kit Contents NI 9238 NI 9238 Getting Started Guide Accessories NI 9971 Backshell Connector Kit

C SERIES LOW VOLTAGE ANALOG INPUT MODULE COMPARISON Product Name Signal Levels Channels Sample Rate Simultaneous Connectivity Resolution Isolation NI 9205 ±200 mv, ±1 V, ±5 V, ±10 V 32 single-ended, 16 differential 250 ks/s No Spring Terminal, DSUB 16-bit 250 Vrms Ch-Earth (Spring), 60 VDC Ch-Earth (DSUB) NI 9206 ±200 mv, ±1 V, ±5 V, ±10 V 32 single-ended, 16 differential 250 ks/s No Spring Terminal 16-bit 600 VDC Ch-Earth NI 9238 ±500 mv 4 differential 50 ks/s Yes Screw Terminal 24-bit 250 Vrms Ch-Ch NI C Series Overview NI provides more than 100 C Series modules for measurement, control, and communication applications. C Series modules can connect to any sensor or bus and allow for high-accuracy measurements that meet the demands of advanced data acquisition and control applications. Measurement-specific signal conditioning that connects to an array of sensors and signals Isolation options such as bank-to-bank, channel-to-channel, and channel-to-earth ground -40 C to 70 C temperature range to meet a variety of application and environmental needs Hot-swappable The majority of C Series modules are supported in both CompactRIO and CompactDAQ platforms and you can move modules from one platform to the other with no modification. CompactRIO CompactRIO combines an open-embedded architecture with small size, extreme ruggedness, and C Series modules in a platform powered by the NI LabVIEW reconfigurable I/O (RIO) architecture. Each system contains an FPGA for custom timing, triggering, and processing with a wide array of available modular I/O to meet any embedded application requirement. 2 ni.com NI 9238 Datasheet

CompactDAQ CompactDAQ is a portable, rugged data acquisition platform that integrates connectivity, data acquisition, and signal conditioning into modular I/O for directly interfacing to any sensor or signal. Using CompactDAQ with LabVIEW, you can easily customize how you acquire, analyze, visualize, and manage your measurement data. Software LabVIEW Professional Development System for Windows Use advanced software tools for large project development Generate code automatically using DAQ Assistant and Instrument I/O Assistant Use advanced measurement analysis and digital signal processing Take advantage of open connectivity with DLLs, ActiveX, and.net objects Build DLLs, executables, and MSI installers NI LabVIEW FPGA Module Design FPGA applications for NI RIO hardware Program with the same graphical environment used for desktop and real-time applications Execute control algorithms with loop rates up to 300 MHz Implement custom timing and triggering logic, digital protocols, and DSP algorithms Incorporate existing HDL code and third-party IP including Xilinx IP generator functions Purchase as part of the LabVIEW Embedded Control and Monitoring Suite NI 9238 Datasheet National Instruments 3

NI LabVIEW Real-Time Module Design deterministic real-time applications with LabVIEW graphical programming Download to dedicated NI or third-party hardware for reliable execution and a wide selection of I/O Take advantage of built-in PID control, signal processing, and analysis functions Automatically take advantage of multicore CPUs or set processor affinity manually Take advantage of real-time OS, development and debugging support, and board support Purchase individually or as part of a LabVIEW suite NI 9238 Circuitry AI0+ Overvoltage Protection ADC Amplifier Prefilter AI3+ AI0- AI3- Overvoltage Protection Channel-to-Channel Isolation ADC Amplifier Prefilter NI 9238 Input signals on each channel are buffered, conditioned, and then sampled by an ADC. Each AI channel provides an independent signal path and ADC, enabling you to sample all channels simultaneously. Filtering The NI 9238 uses a combination of analog and digital filtering to provide an accurate representation of in-band signals and reject out-of-band signals. The filters discriminate between signals based on the frequency range, or bandwidth, of the signal. The three important bandwidths to consider are the passband, the stopband, and the anti-imaging bandwidth. The NI 9238 represents signals within the passband, as quantified primarily by passband ripple and phase nonlinearity. All signals that appear in the alias-free bandwidth are either unaliased signals or signals that have been filtered by at least the amount of the stopband rejection. 4 ni.com NI 9238 Datasheet

Passband The signals within the passband have frequency-dependent gain or attenuation. The small amount of variation in gain with respect to frequency is called the passband flatness. The digital filters of the NI 9238 adjust the frequency range of the passband to match the data rate. Therefore, the amount of gain or attenuation at a given frequency depends on the data rate. Figure 1. Typical Passband Response for the NI 9238 0.025 Gain (db) 0.0 0.025 0.050 0.0 0.1 0.2 0.3 Frequency/Data Rate (Hz) 0.4 0.5 Stopband The filter significantly attenuates all signals above the stopband frequency. The primary goal of the filter is to prevent aliasing. Therefore, the stopband frequency scales precisely with the data rate. The stopband rejection is the minimum amount of attenuation applied by the filter to all signals with frequencies within the stopband. Alias-Free Bandwidth Any signals that appear in the alias-free bandwidth are not aliased artifacts of signals at a higher frequency. The alias-free bandwidth is defined by the ability of the filter to reject frequencies above the stopband frequency. The alias-free bandwidth is equal to the data rate minus the stopband frequency. Data Rates The frequency of a master timebase (f M ) controls the data rate (f s ) of the NI 9238. The NI 9238 includes an internal master timebase with a frequency of 12.8 MHz, but the module also can accept an external master timebase or export its own master timebase. To synchronize the data rate of an NI 9238 with other modules that use master timebases to control sampling, all of the modules must share a single master timebase source. The following equation provides the available data rates of the NI 9238: NI 9238 Datasheet National Instruments 5

256 = where n is any integer from 1 to 31. However, the data rate must remain within the appropriate data rate range. When using the internal master timebase of 12.8 MHz, the result is data rates of 50 ks/s, 25 ks/s, 16.667 ks/s, and so on down to 1.613 ks/s depending on the value of n. When using an external timebase with a frequency other than 12.8 MHz, the NI 9238 has a different set of data rates. Note The NI 9151 R Series Expansion chassis does not support sharing timebases between modules. NI 9238 Specifications The following specifications are typical for the range -40 C to 70 C unless otherwise noted. All voltages are relative to the AI- signal on each channel unless otherwise noted. Caution Do not operate the NI 9238 in a manner not specified in this document. Product misuse can result in a hazard. You can compromise the safety protection built into the product if the product is damaged in any way. If the product is damaged, return it to NI for repair. Input Characteristics Number of channels ADC resolution Type of ADC Sampling mode Internal master timebase (f M ) Frequency Accuracy Data rate range (f s ) using internal master timebase Minimum Maximum Data rate range (f s ) using external master timebase Minimum Maximum 4 analog input channels 24 bits Delta-Sigma (with analog prefiltering) Simultaneous 12.8 MHz ±100 ppm maximum 1.613 ks/s 50 ks/s 390.625 S/s 51.36 ks/s Data rates (f s ) 256, n = 1, 2,, 31 6 ni.com NI 9238 Datasheet

Input voltage range (AI+ to AI-) Nominal Minimum Typical scaling coefficient Overvoltage protection Input coupling Input impedance (AI+ to AI-) ±0.5 V ±0.496 V 74.506 nv/lsb ±30 V DC >1 GΩ Measurement Conditions Table 1. Accuracy Percent of Reading (Gain Error) Percent of Range 1 (Offset Error) Calibrated Maximum (-40 C to 70 C) ±0.20% ±0.06% Typical (23 C ±5 C) ±0.07% ±0.005% Uncalibrated2 Maximum (-40 C to 70 C) ±1.8% ±0.3% Typical (23 C ±5 C) ±0.7% ±0.1% Input noise Stability Gain drift Offset drift Post calibration gain match (ch-to-ch, 20 khz) Phase mismatch (ch-to-ch, 20 khz) Phase mismatch (module-to-module, maximum) Phase nonlinearity (f s = 50 ks/s) Input delay Passband Frequency Flatness (f s = 50 ks/s) 3.9 μvrms ±7 ppm/ C 1.3 μv/ C 100 mdb maximum 0.13 /khz maximum (0.13 /khz f in ) + (360 f in /f M ) 0.12 maximum 40 5 512 /f s + 3.5 μs 0.453 f s 0.11 db peak-to-peak maximum 1 Range equals 0.5 V. 2 Uncalibrated accuracy refers to the accuracy achieved when acquiring in raw or unscaled modes where the calibration constants stored in the module are not applied to the data. NI 9238 Datasheet National Instruments 7

Stopband Frequency Rejection Alias-free bandwidth 0.547 f s 100 db 0.453 f s -3 db bandwidth (f s = 50 ks/s) 24.6 khz Crosstalk (0 khz to 24.6 khz) CMRR (f in = 60 Hz) Spurious-Free Dynamic Range (SFDR) 1 khz, -60 db FS Total Harmonic Distortion (THD) 1 khz, -1 db FS Power Requirements -115 db 140 db 110 db -85 db Power consumption from chassis Active mode Sleep mode Thermal dissipation Active mode Sleep mode 730 mw maximum 50 μw maximum 1.48 W maximum 0.5 W maximum Physical Characteristics If you need to clean the module, wipe it with a dry towel. Tip For two-dimensional drawings and three-dimensional models of the C Series module and connectors, visit ni.com/dimensions and search by module number. Screw-terminal wiring Gauge Wire strip length Temperature rating Torque for screw terminals Wires per screw terminal 0.05 mm 2 to 1.5 mm 2 (30 AWG to 14 AWG) copper conductor wire 6 mm (0.24 in.) of insulation stripped from the end 90 C minimum 0.22 N m to 0.25 N m (1.95 lb in. to 2.21 lb in.) One wire per screw terminal; two wires per screw terminal using a 2-wire ferrule Ferrules 0.25 mm 2 to 1.5 mm 2 8 ni.com NI 9238 Datasheet

Connector securement Securement type Torque for screw flanges Weight Screw flanges provided 0.2 N m (1.80 lb in.) 146 g (5.15 oz) Safety Voltages Connect only voltages that are within the following limits. Channel-to-channel isolation Up to 2,000 m altitude Continuous Withstand 2,001 m to 5,000 m altitude Continuous Withstand Channel-to-earth ground isolation Up to 2,000 m altitude Continuous Withstand 2,001 m to 5,000 m altitude Continuous Withstand Division 2 and Zone 2 hazardous locations applications (Channel-to-channel and channel-to-earth ground) 250 Vrms, Measurement Category II 1,500 Vrms, verified by a 5 s dielectric test 60 VDC, Measurement Category I 1,000 VDC 250 Vrms, Measurement Category II 3,000 Vrms, verified by a 5 s dielectric test 60 VDC, Measurement Category I 1,000 VDC 60 VDC, Measurement Category I Measurement Category I is for measurements performed on circuits not directly connected to the electrical distribution system referred to as MAINS voltage. MAINS is a hazardous live electrical supply system that powers equipment. This category is for measurements of voltages from specially protected secondary circuits. Such voltage measurements include signal levels, special equipment, limited-energy parts of equipment, circuits powered by regulated lowvoltage sources, and electronics. Caution If using in Division 2 or Zone 2 hazardous locations applications, do not connect the NI 9238 to signals or use for measurements within Measurement Categories II, III, or IV. Measurement Category II is for measurements performed on circuits directly connected to the electrical distribution system. This category refers to local-level electrical distribution, such as that provided by a standard wall outlet, for example, 115 V for U.S. or 230 V for Europe. NI 9238 Datasheet National Instruments 9

Caution Do not connect the NI 9238 to signals or use for measurements within Measurement Categories III or IV. Note Measurement Categories CAT I and CAT O are equivalent. These test and measurement circuits are not intended for direct connection to the MAINS building installations of Measurement Categories CAT II, CAT III, or CAT IV. Hazardous Locations U.S. (UL) Class I, Division 2, Groups A, B, C, D, T4; Class I, Zone 2, AEx na IIC T4 Canada (C-UL) Class I, Division 2, Groups A, B, C, D, T4; Class I, Zone 2, Ex na IIC T4 Europe (ATEX) and International (IECEx) Ex na IIC T4 Gc Safety and Hazardous Locations Standards This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use: IEC 61010-1, EN 61010-1 UL 61010-1, CSA 61010-1 EN 60079-0:2012, EN 60079-15:2010 IEC 60079-0: Ed 6, IEC 60079-15; Ed 4 UL 60079-0; Ed 5, UL 60079-15; Ed 3 CSA 60079-0:2011, CSA 60079-15:2012 Note For UL and other safety certifications, refer to the product label or the Online Product Certification section. Electromagnetic Compatibility This product meets the requirements of the following EMC standards for sensitive electrical equipment for measurement, control, and laboratory use: EN 61326-1 (IEC 61326-1): Class A emissions; Industrial immunity EN 55011 (CISPR 11): Group 1, Class A emissions EN 55022 (CISPR 22): Class A emissions EN 55024 (CISPR 24): Immunity AS/NZS CISPR 11: Group 1, Class A emissions AS/NZS CISPR 22: Class A emissions FCC 47 CFR Part 15B: Class A emissions ICES-001: Class A emissions Note In the United States (per FCC 47 CFR), Class A equipment is intended for use in commercial, light-industrial, and heavy-industrial locations. In Europe, 10 ni.com NI 9238 Datasheet

Canada, Australia and New Zealand (per CISPR 11) Class A equipment is intended for use only in heavy-industrial locations. Note Group 1 equipment (per CISPR 11) is any industrial, scientific, or medical equipment that does not intentionally generate radio frequency energy for the treatment of material or inspection/analysis purposes. Note For EMC declarations and certifications, and additional information, refer to the Online Product Certification section. CE Compliance This product meets the essential requirements of applicable European Directives, as follows: 2014/35/EU; Low-Voltage Directive (safety) 2014/30/EU; Electromagnetic Compatibility Directive (EMC) 94/9/EC; Potentially Explosive Atmospheres (ATEX) Online Product Certification Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for this product, visit ni.com/ certification, search by model number or product line, and click the appropriate link in the Certification column. Shock and Vibration To meet these specifications, you must panel mount the system. Operating vibration Random (IEC 60068-2-64) Sinusoidal (IEC 60068-2-6) Operating shock (IEC 60068-2-27) 5 g rms, 10 Hz to 500 Hz 5 g, 10 Hz to 500 Hz 30 g, 11 ms half sine; 50 g, 3 ms half sine; 18 shocks at 6 orientations Environmental Refer to the manual for the chassis you are using for more information about meeting these specifications. Operating temperature (IEC 60068-2-1, IEC 60068-2-2) Storage temperature (IEC 60068-2-1, IEC 60068-2-2) Ingress protection Operating humidity (IEC 60068-2-78) -40 C to 70 C -40 C to 85 C IP40 10% RH to 90% RH, noncondensing NI 9238 Datasheet National Instruments 11

Storage humidity (IEC 60068-2-78) Pollution Degree 2 Maximum altitude 3 Indoor use only. 5% RH to 95% RH, noncondensing 5,000 m Environmental Management NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers. For additional environmental information, refer to the Minimize Our Environmental Impact web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document. Waste Electrical and Electronic Equipment (WEEE) EU Customers At the end of the product life cycle, all NI products must be disposed of according to local laws and regulations. For more information about how to recycle NI products in your region, visit ni.com/environment/weee. 电子信息产品污染控制管理办法 ( 中国 RoHS) Calibration 中国客户 National Instruments 符合中国电子信息产品中限制使用某些有害物质指令 (RoHS) 关于 National Instruments 中国 RoHS 合规性信息, 请登录 ni.com/environment/rohs_china (For information about China RoHS compliance, go to ni.com/environment/rohs_china.) You can obtain the calibration certificate and information about calibration services for the NI 9238 at ni.com/calibration. Calibration interval 1 year 3 Refer to the Safety Voltages section for more information about altitude. Refer to the NI Trademarks and Logo Guidelines at ni.com/trademarks for information on NI trademarks. Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering NI products/technology, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your media, or the National Instruments Patent Notice at ni.com/patents. You can find information about end-user license agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer to the Export Compliance Information at ni.com/ legal/export-compliance for the NI global trade compliance policy and how to obtain relevant HTS codes, ECCNs, and other import/export data. NI MAKES NO EXPRESS OR IMPLIED WARRANTIES AS TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS. U.S. Government Customers: The data contained in this manual was developed at private expense and is subject to the applicable limited rights and restricted data rights as set forth in FAR 52.227-14, DFAR 252.227-7014, and DFAR 252.227-7015. 2016 National Instruments. All rights reserved. 376138A-02 Feb16