This document is a reference document that shows the menus in the 5500sc, 9610sc and 9611sc analyzers. There are 3 top-level menus:

Similar documents
ORM0022 EHPC210 Universal Controller Operation Manual Revision 1. EHPC210 Universal Controller. Operation Manual

Quick Start Operating Instructions

Click Here To Start Demo

Quick Setup Guide for IntelliAg Model NTA

Quick Setup Guide for IntelliAg Model CTA

Sample BD Tech Concepts LLC

EDL8 Race Dash Manual Engine Management Systems

Sample BD Tech Concepts LLC

DCP100 Digital Control Programmer Specifications

OPERATING AND SAFETY INSTRUCTIONS for DIGITAL TEMPERATURE CONTROLS (PLSM SERIES)

Blown Film Extrusion. Line Control. 5 Layers CD Series IBC Temperature USER MANUAL

SINGLE ZONE CLIMATE ZONING SYSTEM. Technical Manual. Polyaire Pty Ltd

TUSKIN Equipment Corporation

Panel cutout required: 1.772" x 3.622" (45mm x 92mm) 1.76" (45mm) 2.45" (62mm) 3.20" (81mm) 3.60" (91mm) 0.59" (15mm) Special Features

DOC AS950 Controller. 06/2016, Edition 8. Operations

All real signals have scale factor 10. Integer, Index and Logic has always scale factor 1.

FRQM-2 Frequency Counter & RF Multimeter

Software Manual Control Panel for Professional Single Booster Units Models: MM3 BW3

SPECIFICATION NO Model 207 Automatic GTAW Welding System

B. The specified product shall be manufactured by a firm whose quality system is in compliance with the I.S./ISO 9001/EN 29001, QUALITY SYSTEM.

Condition Monitoring Custom Products

AEROTRAK PORTABLE AIRBORNE PARTICLE COUNTER MODEL 9110 QUICK START GUIDE

UNIGAS Hand-held combustion gas analyzer with 3 sensors

ТМ-04 Multichannel long range UHF alarm transmitter Operating manual. Sectron Ltd.

Comfort System T-32-P Universal Thermostat. Operation Manual

SCALE & WEIGHT DISPLAYS

Control Unit CU 2.1 User Manual V6850A HPLC

USER & ENGINEER INSTRUCTION MANUAL

Transmitter Interface Program

bio bas.1 USER S MANUAL

WELDING CONTROL UNIT: TE 450 USER MANUAL

802DN Series A DeviceNet Limit Switch Parameter List

Model KT 53 Cooling incubators with thermoelectric cooling

SPECIFICATION NO NOTE

LAUREL ELECTRONICS, INC.

PLC Control Unit for a CSM-E Electrical Compact Clean Steam Generator

User interface. Abbreviations / Meanings

Chapter 2: Scanner Operations NOTE: Install the software cartridge Power the Scanner Select the software title Identify the vehicle

Ambient NOx monitor APNA-370 Operation Manual CODE:GZ D

MAMOS PC SOFTWARE. User manual. Version: /2017 Software version:

dissolved oxygen transmitter

CARLO GAVAZZI Automation Components

Tracer AdaptiView Display

Quick-Start for READ30

OPERATION MANUAL OF MULTIHEAD WEIGHER

REMOTE I/O R30 SERIES. PC CONFIGURATOR SOFTWARE Model: R30CFG Ver Users Manual

MF-PFT. Programmable differential pressure transmitter for pressure or flow measurement and control. DIL1 Output/Utsignal

Application Note: Using the Turner Designs Model 10-AU Fluorometer to Perform Flow Measurements in Sanitary Sewers by Dye Dilution

DOC TU /2017, Edition 3. Basic User Manual

Model KT 115 Cooling incubators with thermoelectric cooling

013-RD

Model KT 53 Refrigerated incubators with thermoelectric refrigeration

OPERATING AND SAFETY INSTRUCTIONS for DIGITAL TEMPERATURE CONTROLS (PLSM SERIES)

TL-2900 AMMONIA & NITRATE ANALYZER DUAL CHANNEL

Ford AMS Test Bench Operating Instructions

VLC-3 USER'S MANUAL. Light Program Controller. M rev. 04 K rev. 00 & ( ( 5, 352*5$0 1 : $ 2 ' 6(77,1*6 )81&7,216

Sartorius Professional Meter

Vorne Industries. 87/719 Analog Input Module User's Manual Industrial Drive Itasca, IL (630) Telefax (630)

Compatibility mode. rev. 0.1, 10/09/2015. Compatibility mode Sx tool manual 1

PROGRAMMING CODE SHEET

Ambient NH 3 monitor APNA-370 Instruction Manual CODE:GZ

RWL11X2 GREENHOUSE CONTROLLER. Operation Manual

TRIMBLE GPS / 10MHz REFERENCE MONITOR DISPLAY V January 2015

with the Field-IQ Crop Input Control System

SPG700 Multiformat Reference Sync Generator Release Notes

Biolyte 2000 Quick User Manual

SAUTER flexotron RDT405 Manual

User Guide. Color Touchscreen Programmable Thermostat. ComfortSense Model: 13H /2017 Supersedes

Model KT 170 Cooling incubators with thermoelectric cooling

Variwrap Controller Manual

POL-200 Semiautomatic Polarimeter. Instruction Manual BANTE INSTRUMENTS CO., LTD

ZIRCONIA OXYGEN ANALYZER CONVERTER

Ambiflex MF620 USER GUIDE

263 Series LED Bargraph Indicator and Controllers

Oxymax W COS61. Technical Information. Dissolved oxygen measurement Optical sensor acc. to the fluorescence quenching principle

Registers type ILR7XX and ILR7XXT

Manual Supplement. This supplement contains information necessary to ensure the accuracy of the above manual.

Professional 4-channel Thermocouple Thermometer plus Infrared Thermometer

MSC+ Controller. Operation Manual

MG-XV operating instruction. Measuring of norm signals, 4-8-digit. Panel instrument type MG-BV Construction instrument type MG-AV

UNIVERSAL DIGITAL METER DC Volts and Amps AC RMS Volts and Amps Thermocouples and RTDs Process Signals Strain Gauge and Load Cell

Ford AMS Test Bench Operating Instructions

Portable Data Recorder HMG Operating Manual (Original Manual)

Ambient SO 2 /H 2 S monitor APSA-H370 Instruction Manual

A summary of all menu lines available in self-diagnostic mode is shown below.

DCT 532. Industrial Pressure Transmitter with i²c interface. Stainless Steel Sensor

Programming instructions Software Version 5

ComfortChoice Touch Thermostat. Designed for ZigBee R Wireless Technology USER GUIDE

Capacitance Limit Detection nivotester FTC 420/421/422

Introduction to the FTA2000

USER MANUAL FOR THE ANALOGIC GAUGE FIRMWARE VERSION 1.1

VOB - data over Video Overlay Box

Tekna evo. innovation technology future 03

Sprite TL Quick Start Guide

Part No. ENC-LAB01 Users Manual Introduction EncoderLAB

OPERATOR S. MODEL Computron 8000 PRESSURE FRYER REGISTER WARRANTY ONLINE AT

Model 7600 HD/SD Embedder/ Disembedder Data Pack

pc-based controller user guide for vers software

Chapter 23 Dimmer monitoring

Pacific Sun Triton R2 Bali Revision 9channel LED aquarium lamp

Transcription:

Controller menus 5500sc, 9610sc and 9611sc analyzers DOC273.53.80566 Introduction This document is a reference document that shows the menus in the 5500sc, 9610sc and 9611sc analyzers. There are 3 top-level menus: Main menu on page 1 Cal menu on page 6 Diag menu on page 6 Main menu START ANALYZER or STOP ANALYZER Starts or puts the analyzer back into operation. If in operation mode, puts the analyzer in shutdown mode. REAGENTS/STANDARDS GRAB SAMPLE SETUP SYSTEM VIEW DATA LINK2SC SD CARD SETUP (conditional) Sets and resets reagent, standard and cleaning solution levels. Primes the reagents. Sets the blank value (if applicable). Measures an external sample or dispenses a sample for external analysis. Sets the main configuration settings for the analyzer. Shows measurement data, analyzer data and data logs. Lets the user share data between analyzers, process probes and compatible laboratory instruments. Shows when an SD card is installed. Updates the firmware and software. Downloads the event and data logs. > REAGENTS/STANDARDS SET REAGENT LEVEL SET STANDARD LEVEL SET CLEANING LEVEL RESET REAGENT LEVELS RESET STANDARD LEVELS RESET CLEANING SOLUTION LEVELS Sets the volume of a specific reagent in the reagent bottle to an approximate value. Range: 1 100%. Sets the volume of a specific standard solution in the standard bottle to an approximate value. Range: 1 100%. Sets the volume of the cleaning solution in the cleaning solution bottle to an approximate value. Range: 1 100%. Sets the volume of reagent in the reagent bottles to 100% full. Sets the volume of standard solution in the standard bottle to 100% full. Sets the volume of the cleaning solution in the cleaning solution bottle to 100% full. 1

PRIME REAGENTS ENTER BLANK VALUE Starts the flow of all reagents through the tube and valve system. IMPORTANT: Prime the reagents each time the reagents are changed to remove bubbles and push the new reagents through the system. Applicable only to silica analyzers with Hach-prepared reagents: Sets the reagent blank value to the value that is printed on the R1 bottle label. > GRAB SAMPLE GRAB SAMPLE IN GRAB SAMPLE OUT Analyzes an external sample or standard. Removes a sample directly from a sample line for external analysis. MEAS MODE EDIT INTERVAL (conditional) MEAS UNITS > SETUP SYSTEM Changes the mode of the measurement cycle. s: interval or continuous (default). The continuous mode measures approximately every 9 minutes. Changes the interval time when MEAS MODE is set to interval. s: 10 to 240 minutes (default = 15 minutes). Changes the measurement units that are shown on the display and in the data log. s: ppb (default), ppm, mg/l, µg/l. SIGNAL AVERAGE Selects the number of measurements that are used to calculate an average measurement (1-5). The signal average reduces variability in measurements (default = 1, no averaging). EDIT ANALYZER NAME EDIT CHANNEL NAME CONFIGURE SEQUENCER (optional) SAMPLE MISSING SET DATE & TIME DISPLAY SETUP DISABLE REMINDERS MANAGE DEVICES INSTRUMENT INFORMATION CALCULATION SETUP OUTPUTS SETUP NETWORK (conditional) Changes the name that is shown on the top of the measurement screen (maximum of 16 characters). Changes the name of the sample source that is shown on the measurement screen (maximum of 10 characters). Starts or stops measurements for individual sample sources. Sets the measurement order of the sample sources when there is more than one sample source. Sets the operation that occurs if no sample is detected. s: ON DELAY (default) or OFF DELAY. ON DELAY: the instrument waits the remainder of the measurement cycle, then goes to the next sample in the sequence. OFF DELAY: the instrument waits 10 seconds, then goes to the next sample in the sequence. Sets the analyzer time and date. Changes the analyzer language. Adjusts the order of the measurements on the screen. Adjusts the contrast of the display. Stops scheduled maintenance alerts for individual components. s: tubing, stir bar, colorimeter cell, air filter, stir motor, air relief valve, pinch valve, air compressor, reagent valves, sample valves, standard valves, colorimeter LED, fan filter, air check valve. Installs or removes input modules. Shows the analyzer information. Sets up variables, parameters, units and formulas for the analyzer. Selects and configures the 4 20 ma setup, the relay setup and the error hold mode. Shows only if a network card is installed. The network card supports the Modbus, Profibus and HART communication protocols. 2

SECURITY SETUP AIR PURGE RESET DEFAULTS Enables or disables the pass code (default = HACH55). Enables the air purge for use with an external air supply. s: on or off (default). Off: an external air supply is not used. The vent fan is enabled and the air filter is installed. On: an external air supply is connected to the instrument. The vent fan is disabled. The air filter is replaced with a fan filter plug. To use this feature, make sure that the fan filter plug is installed. Refer to the documentation that is supplied with the air purge kit. Sets the configuration to the factory defaults. ACTIVATION > SETUP SYSTEM > SETUP OUTPUTS > 4 20 ma SETUP Sets the output parameters such as high and low values. The available options change with the selected function in the SET FUNCTION menu. SELECT SOURCE Selects the output. s: None (if the relay is not configured), the analyzer name or calculation (if a calculation formula has been configured). SET PARAMETER Selects the measurement channel from the list. SET FUNCTION SET TRANSFER Selects a function. Other options will change based on which function is selected. LINEAR CONTROL Signal is linear. PID CONTROL Signal works as a PID (Proportional, Integral, Derivative) controller. LOGARITHMIC Signal is logarithmic within the process variable range. BILINEAR Signal is given as two linear segments within the process variable range. Sets the transfer value when TRANSFER is or will be selected as the system ERROR HOLD MODE. Range: 3.0 to 23.0 ma (default = 4.000). SET FILTER Enters a time-average filter value of 0 to 120 seconds (default = 0). SCALE 0mA/4mA Selects the scale (0 20 ma or 4 20 ma). ACTIVATION > SETUP SYSTEM > SETUP OUTPUTS > RELAY SETUP Sets the output parameters such as high and low alarm values. The available options change with the selected function in the SET FUNCTION menu. SELECT SOURCE Selects the output. s: None (if the relay is not configured), the analyzer name or calculation (if a calculation formula has been configured). SET FUNCTION SET TRANSFER FAIL SAFE Selects a function. ALARM The relay starts when the upper or lower alarm value is triggered. FEEDER CONTROL The relay shows if a process value is larger or falls below a setpoint. EVENT CONTROL The relay toggles if a process value reaches an upper or lower limit. SCHEDULER The relay switches at times independently of any process value. WARNING The relay shows warning and error conditions in probes. PROCESS EVENT The relay switches when the analyzer does a specified operation. Selects active or inactive. Selects yes or no. 3

> VIEW DATA ANALYZER DATA Shows the analyzer status information. MEASUREMENT DATA Shows the measurement information. LOG DATA Selects the data log and/or the event log. DATA LOG Shows the measurement values. EVENT LOG Shows all analyzer information (e.g., alarms, warnings, configuration changes, and so on). Select the start time, number of hours and/or number of readings. Element > VIEW DATA > ANALYZER DATA Definition CELL TEMP Shows the temperature of the colorimeter cell block heater (ideally 49.8 C to 50.2 C (121.64 F to 122.36 F)). REAGENT TEMP AMBIENT TEMP SAMPLE TEMP AIR PRESS LED DUTY CYCLE HEATER DUTY CYCLE SAMPLE FLOW SAMPLE PRESS 1 SAMPLE PRESS 2 REAGENT 1 REAGENT 2 REAGENT 3 REAGENT 4 (if applicable) STD SOLUTION (if applicable) STANDARD 1 (if applicable) STANDARD 2 (if applicable) FAN SPEED LEAK COUNTS Shows the reagent temperature before it enters the colorimeter. Shows the air temperature inside the electronics area. Shows the sample pre-heater block temperature (typically 45 C to 55 C (113 F to 131 F) but could reach 58 C (136.4 F)). Shows the reagent air pressure in the reagent bottles (ideally 3.95 to 4.10 psi). Depends on condition of colorimeter cell and age of analyzer (typically 7,200 to 40,000 counts). Shows the percentage of time the colorimeter heater is on to keep a constant temperature of 50 C (122 F). Shows the approximate sample flow into the colorimeter, measured during the flush cycle. Shows the sample pressure before the pre-heater block (ideally 2 to 4.5 psi based on incoming sample pressure). Shows the sample pressure after the sample pre-heater, which is used to calculate the sample flow. The sample pressure is almost zero when the flush is set to off and is approximately 0.2 psi when the flush is on (depends on inlet pressure and flow). Shows the level of the reagent in the reagent 1 bottle. Shows the level of the reagent in the reagent 2 bottle. Shows the level of the reagent in the reagent 3 bottle. Shows the level of the reagent in the reagent 4 bottle. Shows the level of the standard solution in the standard solution bottle. Shows the level of the standard solution in the standard solution 1 bottle. Shows the level of the standard solution in the standard solution 2 bottle. Shows the speed of the ventilation fan. Shows the leak counts (range 0 to 1023). A count of more than 511 indicates a fluid leak. 4

Element > VIEW DATA > MEASUREMENT DATA Definition LAST MEAS CHANNEL Shows the last channel that was measured. LAST MEAS TIME LAST ABS LAST CONC NEXT MEAS TIME DARK REF SAMPLE DARK STD DEV REF STD DEV SAMPLE STD DEV SAMPLE VOLUME REAGENT 1 REAGENT 2 REAGENT 3 Shows the time of the last measurement. Shows the last absorbance reading. Shows the concentration of the last measurement. Shows the time of the next measurement. Shows the number of A/D counts that were measured when the LED was off. Shows the reference A/D count that is used to compensate for natural color and turbidity. Shows the A/D counts measurement (after color development) that is used to determine the concentration of the sample. Shows the standard deviation of the dark counts from six readings. Shows the standard deviation of the reference counts from six readings. Shows the standard deviation of the sample counts from 6 readings. Shows the total sample flush volume through the colorimeter for the measurement cycle. Shows the calculated reagent delivery time to the sample based on temperature, pressure and viscosity. Shows the calculated reagent delivery time to the sample based on temperature, pressure and viscosity. Shows the calculated reagent delivery time to the sample based on temperature, pressure and viscosity. > LINK2SC CREATE A NEW JOB Starts the grab sample operation for the measurement value exchange between the analyzer and the laboratory. JOB LIST JOB ID MIN JOB ID MAX Selects the job file to send the job to the laboratory or deletes the job. JOB TO LAB the analyzer data is sent to the SD card as a job file. ERASE JOB deletes the data. Specifies the minimum value for the ID number range. Specifies the maximum value for the ID number range. > SD CARD SETUP UPGRADE SOFTWARE (conditional) SAVE LOGS Shows when an upgrade file is available on the SD card. Select the specific device for the upgrade. Selects the device for the download data and to save the logs for last day, last week, last month or all. 5

MANAGE CONFIGURATION Saves and restores the backup settings, restore settings and/or transfer settings between instruments. WORK WITH DEVICES READ DEVICE FILES Selects the data for each device to save on the SD card. s: sensor diag, measurement data (curve data for a measurement cycle), cal history, cal data and/or test script. WRITE DEVICE FILES-shows when an upgrade file is available for a new measurement cycle script. Cal menu START MANUAL CAL SET AUTO CALIBRATION CALIBRATION DATA Starts a calibration immediately or at the end of the current measurement cycle. Sets a schedule for calibration at regular intervals. Shows data from the last calibration and the due date and time of the next scheduled calibration. OVERRIDE CALIBRATION Enters a new slope and/or zero value (reagent blank). When OVERRIDE CALIBRATION is selected, the automatic calibration is disabled for the selected option. OUTPUT CALIBRATION RESET DEFAULT CAL Selects a 4 20 ma output and sets the output values to be sent. Resets the calibration data to the default values and disables the automatic calibration. When complete, calibrate the analyzer. Diag menu DIAGNOSTICS PROGNOSYS Shows the errors and warnings for the instrument or the installed modules. The active warnings or reminders show until they are acknowledged or reset. Then, the color of the display background goes to white. Shows the variables that trigger the service indicator and the measurement health indicator on the display. CURRENT STATUS Shows the current instrument statuses that follow: OPERATION Current measurement mode. SAMPLE CHANNEL Current sample channel. STEP STATUS Current step in the measurement cycle. STEP TIME Step time remaining. MINUTES LEFT Minutes left in the current step. COMPLETION the percent completion of the current measurement cycle. ANALYZER HELP PERFORM TEST OUTPUTS VIEW LED MODBUS STATS SERVICE Shows all possible errors, warnings and reminders with troubleshooting hints. Runs a diagnostic test for individual analyzer components. Shows the current status of the 4 20 ma and relay outputs with the options to examine, hold and simulate the outputs. The LED shines in the colorimeter cell for improved viewing during troubleshooting. The LED can stay on for 1 to 999 seconds. Shows the status of the Modbus ports: sensor, controller, network and service. Shows the number of good and bad transmissions. Shows the service parts information and the history. SERVICE PART Shows the last and the next service date and the remaining days. PART INFORMATION Shows the replaced part and the current run time. UPCOMING SERVICE Shows the next part which needs to be replaced. SERVICE HISTORY Shows the date and time of the replaced parts. 6

SYSTEM DATA I2C DATA Shows the system information. TEMPERATURE Shows the measured temperature of the A/D device in Celsius (C). POWER SOURCE FREQUENCY Shows the line power frequency (Hz). POWER SOURCE VOLTAGE Shows the line power voltage (V). 12 V VOLTAGE Shows the measured power supply voltage (V DC). 3.3 V VOLTAGE Shows the measured regulated 3.3 V supply (V DC). 12 V CURRENT Shows the measured 12 V power supply current (Amps). Shows the display information (I2C) and the version number. OVERFEED RESET Resets the overfeed timer. > PERFORM TEST REAGENT DELIVERY SAMPLE DELIVERY CAL SOL. DELIVERY MIXER Opens each reagent valve for a specified time (50 milliseconds to 65 seconds) or a specified volume (20 to 9,999 µl). Opens each sample valve for a specified time (1 to 9999 seconds). Opens the calibration standard valve for a specified time (1 to 9999 seconds). Starts the mixer in a clockwise or counterclockwise (CCW/CW) direction at a specified speed (10 to 500 rpm) for a specified time (1 to 9999 seconds). COLORIMETER HEATER Sets the colorimeter heater to a specified temperature (20 to 60 C / 68 to 140 F). The measured temperature is shown. SAMPLE HEATER COLORIMETER STATUS LED A2D AIR PUMP FAN ANALYZER TYPE SELECT SCRIPT SET CHANNELS Sets the sample heater to a specified temperature (20 to 60 C / 68 to 140 F). The measured temperature is shown. Increases the colorimeter LED duty cycle in 5% increments from 0% until saturation. The A2D counts are shown for 0%, a percent between 0% and saturation and the first saturation value. Operates each of the LED indicators on the front panel (off, red, green, yellow). The test operates continuously until stopped. Sets the colorimeter LED intensity to examine the cell transmittance for the A2D output. Changes and controls the air pressure. SET SETPOINT Range: 1 to 9.99 psi. LOW and HIGH DEADBAND Range: 0 to 1 psi. SET LOW and HIGH VALUE Range: 5 99.99 psi. START Starts the air pump with the entered settings. Adjusts the fan duty cycle. For use by technical support only. Toggles between normal instrument script and test script. For use by technical support only. 7

HACH COMPANY World Headquarters P.O. Box 389, Loveland, CO 80539-0389 U.S.A. Tel. (970) 669-3050 (800) 227-4224 (U.S.A. only) Fax (970) 669-2932 orders@hach.com www.hach.com HACH LANGE GMBH Willstätterstraße 11 D-40549 Düsseldorf, Germany Tel. +49 (0) 2 11 52 88-320 Fax +49 (0) 2 11 52 88-210 info-de@hach.com www.de.hach.com HACH LANGE Sàrl 6, route de Compois 1222 Vésenaz SWITZERLAND Tel. +41 22 594 6400 Fax +41 22 594 6499 Hach Company/Hach Lange GmbH, 2016. All rights reserved. 11/2016, Edition 1 *DOC273.53.80566*