HDS1021M Handheld Digital Storage. Oscilloscope & Multimeter. User Manual

Size: px
Start display at page:

Download "HDS1021M Handheld Digital Storage. Oscilloscope & Multimeter. User Manual"

Transcription

1 HDS1021M Handheld Digital Storage Oscilloscope & Multimeter User Manual

2 Dec edition Copy Right in this Manual Lilliput Company. All rights have been reserved. The Lilliput s products are under the protection of the patent rights in America and other countries, including ones which have already obtained the patent rights and those which are applying for. The information in this manual will replace all that in the materials published originally. The information in this manual was correct at the time of printing. However, OWON will continue to improve products and reserves the rights to changes specification at any time without notice. OWON is the registered trademark of the Lilliput Company. Xiamen Lilliput Technology Co.,Ltd.: the 5th floor, B Area, Chuangxin Mansion, Software Park,ZhenZhuWan, Huandao RD, Xiamen,Fujian,China Fujian Lilliput Optoelectronics Technology Co.,Ltd.:The mansion of optoelectronics hengsan road, Lantian industrial zone,zhangzhou,fujian,china Tel: Fax: Mail:sales@owon.com.cn

3 General Warranty The Lilliput warrants that the product will be free from defects in materials and workmanship for a period of three years from the date of purchase of the product by the original purchaser from the Lilliput company. And the warranty period of accessories such as probe, battery, adapter is one year. This warranty only applies to the original purchaser and is not transferable to the third party. If the product proves defective during the warranty period, Lilliput either will repair the defective product without charge for parts and labor, or will provide a replacement in exchange for the defective product. Parts, modules and replacement products used by Lilliput for warranty work may be new or reconditioned to like new performance.all replaced parts, modules and products become the property of Lilliput. In order to obtain service under this warranty, Customer must notify Lilliput of the defect before the expiration of the warranty period. Customer shall be responsible for packaging and shipping the defective product to the service center designated by Lilliput, and with a copy of customer proof of purchase. This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate maintenance and care. Lilliput shall not be obligated to furnish service under this warranty a) to repair damage resulting from attempts by personnel other than Lilliput representatives to install, repair or service the product; b) to repair damage resulting from improper use or connection to incompatible equipment; c) to repair any damage or malfunction caused by the use of non-lilliput supplies; or d) to service a product that has been modified or integrated with other products when the effect of such modification or integration increases the time or difficulty of servicing the product. Please contact the nearest Lilliput s Sales and Service Offices for services or a complete copy of the warranty statement. Excepting the after-sales services provided in this summary or the applicable warranty statements, Lilliput will not offer any guarantee for maintenance definitely declared or hinted, including but not limited to the implied guarantee for marketability and special-purpose acceptability. Lilliput should not take any responsibilities for any indirect, special or consequent damages.

4 Contents Table of Contents Chapter Page 1.DECLARATION OF CONFORMITY Declaration of Conformity Open the Oscilloscope Parts Box,You Can See the Parts Below (see the picture below) 1 2.SAFETY INFORMATION Safety Terms and Symbols Specific Warning and Caution Terms that Appear throughout the Manual Terms Used on the Product Symbols Used on the Product General Safety Information 3 3. GENERAL CHARACTERISTICS 5 4.PERFORMING THE GENERAL INSPECTION Performing the General Inspection Check whether there is any Damage on it Due to Transportation Make a Check on Accessories Make a Check on the Complete Instrument 6 5.INPUT CONNECTIONS Input Connections Description for Front Panel and Keys 8 6.USING THE SCOPE About this Chapter Power-Up the Oscilloscope 10 1

5 Contents 6.3 Oscilloscope Operation Window Navigating a Menu Manually Setting the Vertical System, Horizontal System and Trigger Position Resetting the Oscilloscope Input Connections Displaying an Unknown Signal with Auto Set Automatic Zero-returning of Trigger Horizontal Position and 50% setting of Trigger Level Position Automatic Measurements Freezing the Screen Using Average for Smoothing Waveforms Using Persistence to Display Waveforms Using Peak Detection to Display Glitches Selecting AC-coupling Reversing the Polarity of the Displayed Waveform 21 7.USING THE MULTIMETER About this Chapter Making Meter Connections Multimeter Operation Window Making Multimeter Measurements Measuring Resistance Values Making a Diode Measurement On-off Test Making a Capacitance Measurement Making a DC Voltage Measurement Making a AC Voltage Measurement Making a DC Current Measurement Making an AC Current Measurement 29 2

6 Contents 7.5 Freezing the Readings Taking a Relative Measurement Selecting Automatic/ManualRange Adjustment 32 8.ADVANCED FUNCTION OF OSCILLOSCOPE About this Chapter Setting the Channel Setting the Channel Coupling Make Open and Close Settings on Channel Adjusting the Probe Scale Setting of Inverted Waveform Setting the Trigger System Triggering Control Edge Triggering Video Triggering Acquiring Mode Setting Display Setting Display Style Persistence Cymometer Waveform Saving Setups Function Setting Menu Making Automatic Measurements Setting the Cursor Measurements Autoscale System State Menu Setting of Time Base Mode Data Transmission 51 9.TROUBLE SHOOTING 53 3

7 Contents 10.APPENDIX Appendix A: Specifications Oscilloscope Meter General Specifications Appendix B: Maintenance and Cleaning Common Maintenance Storage of Oscilloscope Replacing the Lithium Battery Unit 59 4

8 1-Declaration of Conformity 1.Declaration of Conformity 1.1 Declaration of Conformity HDS1021M Handheld Digital Storage Oscilloscope & Multimeter Manufacturer: Fujian Lilliput Optoelectronic Technology Co., Ltd. Statement of Conformity Base on the result using appropriate standards,the product is in conformity with Electromagnetic Compatibility Directive 89/336/EEC Low Voltage Directive 73/23/EEC Sample tests Standards used: EN :2001(2nd edition) Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory use-part 1:General Requirements EN61326:1997+A1:1998+A2:2001+A3:2003 Electrical Equipment for Measurement, Control and Laboratory use-emc Requirements-part 1. General Requirements EN :2000+A2:2005 Electromagnetic Compatibility(EMC)-part 3:Limits-Section 2:Limits for Harmonic Current Emissions (Equipment Input Current less than/equal to 16A per phase) EN :1995+A1:2001 Electromagnetic Compatibility (EMC)-part 3:Limits-Section 3 Limitation of Voltage Fluctuations and Flicker in Low-Voltage Supply systems for Equipment with Rated Current less than or equal to 16A The tests have been performed in typical configuration. This conformity is indicated by the symbol CE, i.e."conformite Europeenne". 1

9 1-Declaration of Conformity 1.2 Open the Oscilloscope Parts Box,You Can See the Parts Below (see the picture below) # Description Standard Optional 1 Oscilloscope and Battery 2 Power adapter 3 Oscilloscope Probe x 1 (grey) 4 Multimeter test lead x 2 (black and red) 5 Extension module for big current measurement 6 Module for small capacitance measurement 7 The output termeral of 1KHz/5v signal 8 Probe adjustment tools 9 USB communication cable 10 User Manual 11 CD-ROM (software) 12 Soft carrying case figure 1: Digital Oscilloscope Parts 1

10 2-Safety Information 2.Safety Information In order to ensure the correct using and the best efficient service, please carefully read this user s manual. 2.1 Safety Terms and Symbols Specific Warning and Caution Terms that Appear throughout the Manual Warning: Warning identifies conditions and actions that pose hazards to the users. Caution: Caution identifies conditions and actions that may damage the product or other properties Terms Used on the Product The following terms appears possibly on the product: Danger: Warning: Notice: The term Danger is used in this manual to indicate that when you read this mark, personal injury may be caused to you immediately. The term Warning is used in this manual to indicate that when you read this mark, personal injury may not be caused to you immediately, but you need to be cautionary. The term Notice is used in this manual to indicate that damages may be caused on this product or other properties Symbols Used on the Product. The following symbols appear possibly on the product. High Voltage Please Consult the Manual. Protective Ground End Earth End on the Shell Grounding End for Measurement 2

11 2-Safety Information 2.2 General Safety Information Carefully read the following safety information in order to avoid any personal injury and damage on this product or any products connected with it. This product can only be used in the specified applications to prevent any possible dangers. Warning: To avoid fire or electrical shock. Please use proper power adapter. Use only the power adapter appointed by the manufacturer and subject to approval of being used in the user s country. Warning: To avoid fire or electrical shock if a test tool input is connected to more 42V peak (30Vrms) or on circuits of more than 4800VA: Use only insulated voltage probes, test leads and adapter supplied with the test tool, or indicated by OWON as suitable for the Oscilloscope & Multimeter. Before use, inspect voltage probes, test leads and accessories for mechanical damage and replace when damaged. Remove all probes, test leads and accessories (power adapter, USB line )that are not in use. Do not apply voltages that differ more than 400 V from earth ground to any input when measuring in a CAT Ⅱ environment. Do not apply voltages that differ more than 400 V from each other to the isolated inputs when measuring in a CAT Ⅱ environment. Do not apply input voltages above the rating of the instrument Use caution when using 1:1 test leads because the probe tip voltage will be directly transmitted to the Oscilloscope & Multimeter. Do not use exposed metal BNC or banana plug connectors. Do not insert metal objects into connectors. Always use the Oscilloscope & Multimeter only in the manner specified. Voltage ratings that are mentioned in the warning, are giver as limits for working voltage. The represent V ac rms (50-60Hz) for AC sine wave applications and as V dc for DC applications. Overvoltage Category Ⅱ refers to local level, CAT is the prefix refers to level II, Class II is a low-voltage high-energy level, which is applicable for appliances and portable equipment. Only qualified technical personnel are permitted to perform maintenance. Always connect the power adapter first to the AC outlet before connecting it to the 3

12 2-Safety Information Oscilloscope & Multimeter. Pay attention to the nominal values of all terminals:to avoid fire or electric shock, please keep a watchful eye on all nominal values and marks specified for this product. Before any connection performed on this product, carefully read the user s manual of the product for further information of nominal values. No operation is allowed without the instrument cover plate: If the cover plate or panel has been removed, do not perform any operation on this product. No touch is allowed on bare conductors:when the product is powered on, do not touch any bare joints or parts of the scope meter. Operation is prohibited in case of any undetermined failure:when in doubt any damage on this product, consult the qualified personnel for checking on it. Keep ventilation in good condition:refer to the user manual for detail installation instructions in order to fix this product correctly and provide it with good ventilation conditions. No operation is allowed under a humid environment. No operation is allowed under an explosive environment. Keep clean and dry on the product surface. The method which stipulated according to the factory does not use the equipment, possibly can damage the protection which the equipment provides. 4

13 3-General Characteristics 3. General Characteristics Oscilloscope Record length of 6,000 points; Reading-out with the cursor; Eighteen automatic measurement functions; Autoscale function; Color liquid crystal display of high resolution and high contrast; Storage and call-out of waveforms; Automatic setting function provided capable of fast setting; Implementation of detecting the average and peak values of the waveform; Edge and video triggering function; USB communication port; Multiple Language User Interface. Multimeter 3 3/4 digits; Volts,Amps,Ohms,Diode,Capacitance,Continuity measurement; 20A maximum amplitude; Isolated inputs between oscilloscope and multimeter. 5

14 4-Performing the General Inspection 4.Performing the General Inspection 4.1 Performing the General Inspection When you have got a new HDS1021M series oscilloscope, it is suggested that you should perform a general inspection on the instrument according to the following steps Check whether there is any Damage on it Due to Transportation If the packing boxes or foam cushions are found in serous damage, keep them in a proper place till the complete instrument and accessories have passed the electrical and mechanical tests Make a Check on Accessories The accessory list has been described in the picture of Digital Oscilloscope Parts of this manual. You can make a check and find whether there is any accessory loss with reference to the Appendix. In case of any accessory loss or damage, consult the OWON dealer responsible for such a business or the local office of OWON Make a Check on the Complete Instrument If the instrument is damaged in its appearance or it fails in normal operation or performance test, consult the OWON dealer responsible for such a business or the local office of OWON. If the instrument is damaged due to transportation, keep the packing in a proper place and consult the transportation department and the OWON dealer responsible for such business, who will provide an instrument replacement or maintenance. 6

15 5-Input connections 5.Input Connections 5.1 Input Connections See the following figure 2: Description: figure 2 1. The power adapter is supplied for AC power supply and battery recharging. 2. Multimeter test lead. 3. Multimeter input jacks, including three circular banana jacks and two square jacks. The three circular jacks are used for voltage, current and resistance inputs, while the two square jacks are used for capacitance inputs. 4. Oscilloscope probe. 5. Oscilloscope channel input. 6. The output terminal of 1KHz/5V square-wave test signal. 7

16 5-Input connections 5.2 Description for Front Panel and Keys See the following figure 3: Description: figure 3 1. Power adapter jack. 2. USB jack. 3. The output port of 1KHz/5V square-wave test signal. 4. : Power switch. 5. A: Multimeter current measurement key. 6. V: Multimeter voltage measurement key. 7. R: Multimeter resistance, triode, On/Off and capacitance measurement key. 8. OSC : Oscilloscope left-direction adjustment key. 9. OSC : Oscilloscope right-direction adjustment key. 10. OSC OPTION: Oscilloscope setting key. 8

17 5-Input connections With the combination application of the four keys OSC, OSC, OSC and OSC, the users can make the following settings circularly by pressing OSC OPTION. The settings include: Voltage Unit Scale of Channel; Primary Time base (TIME BASE), zero point position of channel, trigger horizontal position (TIME) and trigger level position (TRIG). In cursor measurement mode, the users can adjust the positions of Cursor 1(V1 or T1) and Cursor 2 (V2 or T2). 11. OSC : Oscilloscope display downward adjustment key. 12. OSC : Oscilloscope display upward adjustment key. 13. OSC/DMM: Operation mode switching key between oscilloscope and multimeter. 14. AUTO SET: Under the Multimeter Mode, when performing the current or voltage measurement, you can make a measurement switch between AC and DC with this key pressed; when performing the resistance measurement, you can select resistance, diode, On/Off or capacitance measurement circularly with this key. While this key is used for auto setting under the oscilloscope operation mode. 15. RUN/STOP: key for running or stopping the operation. 16. MENU : Choose the lower item on the menu list. 17. MENU: Show / Hide the menu. 18. MENU : Choose the upper item on the menu list. 19. F1~F5: Switch or Adjust options for each menu. 9

18 6-Using the scope 6.Using the Scope 6.1 About this Chapter This chapter provides a step-by-step introduction to the scope functions. The introduction does not cover all of the capabilities of the scope functions but gives basic examples to show how to use the menus and perform basic operations. 6.2 Power-Up the Oscilloscope Connect oscilloscope to AC power via a power adapter as shown in Figure 1. (The oscilloscope may still work with built-in Li-ion battery even without AC power supply) Turn the oscilloscope on by pressing down the power on/off key. The instrument then performs selfchecking after power on. A greeting window and a sentence press any key to continue will display on the screen when the system finishes selfchecking. The users can press any key to enter the measuring function. The oscilloscope is powered up in its last setup configuration. 6.3 Oscilloscope Operation Window See the following figure 4: Description: figure 4: Oscilloscope Operation Window 1. Battery electric quantity indicating symbols, including,, and. 10

19 6-Using the scope 2. Auto measurement window 1, in which f means frequency, T means cycle, V means the average value, Vp the peak-peak value, Vk the root-mean-square value,. Ma the maximum amplitude value, Mi the minimum amplitude value, Vt the Voltage value of the waveform s flat top value, Vb the Voltage value of the waveform s flat base, Va the amplitude value, Os the overshoot value, Ps the Preshoot value, RT the rise time value, FT the fall time value, PW the +width value, NW the -Width value, +D the +Duty value, -D the -Duty value. 3. Auto measurement window The pointer indicates the horizontal triggering position. 5. This reading gives the Time Difference between the horizontal triggering position and the screen centerline. It reads zero when the pointer is in the center of the screen. 6. The trigger state indicates the following information. Auto: The oscilloscope is working in the automatic mode and displaying the waveform under the non-trigger state. Trig d: The oscilloscope has detected a trigger and collecting the information generated after the trigger. Ready: All pre-triggered data have been captured and the oscilloscope has been ready to receive trigger signals. Scan: The oscilloscope can gather and display the waveform data continuously in scanning mode. Stop: The oscilloscope has stopped collecting the waveform data. 7. The green pointer shows the trigger voltage level. 8. A hidden-style menu: With the MENU key pressed, you can view or hide the menu. 9. It reads the value of trigger voltage level. 10. Menu setting options: There are different setting options for different menus. 11. The reading gives the value of primary time base. 12. These graphics present the coupling modes of the channel. The graphic ~ indicates AC, the graphic - indicates DC,the graphic indicates GND. 13. This reading shows the vertical Voltage Unit Scale of the channel. 14. OSC OPTION operation prompt: There are different prompts for different OSC OPTION operations. 15. The red pointer gives the grounding datum point of the waveform, which is the zero position. No display of this pointer indicates that the channel has not been opened. 16. Waveform of the input singal. 11

20 6-Using the scope 17. Waveform display area. 6.4 Navigating a Menu The following example shows how to use the tool s menus to select a function, as shown in the following figure. 1. Press the MENU key to display the Function Menu on the right of the screen and the corresponding optional settings on the bottom. Press MENU again will hide the Function Menu. 2. Press the MENU or MENU key to select different function menus. 3. Choose one key from F1 to F5 and press it to change function setting. See the following figure 5: figure 5: the Tool s Menus 6.5 Manually Setting the Vertical System, Horizontal System and Trigger Position With the combination application of the four keys OSC, OSC, OSC and OSC, the users can make the following settings circularly by pressing OSC OPTION. The settings include: Voltage Unit Scale of the Channel; Primary Time base (TIME BASE), zero point position of channel, trigger horizontal position (TIME) and trigger level position (TRIG). The following example shows how to use OSC OPTION key to make a setting. 1. Press once the OSC OPTION key; the following is displayed at the bottom left side of the screen, as shown in the figure below. / Time Base / Volts/Div 12

21 6-Using the scope See the following figure 6: figure 6: Voltage Unit Scale 2. Press the key OSC or OSC to adjust the vertical scale and press OSC or OSC to adjust the horizontal time scale. 3. Press the OSC OPTION key one more time, and the following display is visible at the bottom left side of the screen, shown as the following figure. / Time / Zero See the following figure 8: figure 7: Zero Point Position 6. Press OSC or OSC key to adjust the zero position in vertical direction and press OSC or OSC key to adjust the horizontal position. 9. Press OSC OPTION key once more and the following appears at the bottom left of the screen, shown as the following figure. 13

22 6-Using the scope / Time / Trig See the following figure 8: figure 8: Trigger Level Position 10. Press the OSC or OSC key to adjust the trigger position and press OSC or OSC key to adjust the horizontal position. 11. Press the OSC OPTION key again and return back to step 1. Term interpretation Vertical scale factor: It stands for the voltage amplitude represented by a division in the vertical direction of the display area, through the adjustment of which you can amplify or attenuate the signal and thus regulate the signal amplitude into the expected measurement range. Vertical zero position: It is referred to as the grounding datum point, through the adjustment of which you can regulate the display position of the waveform on the screen. Primary time base: It means the time values represented by a division in the horizontal direction of the display area. Trigger horizontal position: It means the time deviation between the actual trigger point and the screen central line, which will be displayed as 0 at the center point of the screen. Trigger level position: It represents the voltage deviation between the actual trigger level and the zero position of the triggering signal source channel. 6.6 Resetting the Oscilloscope If you want to reset the Oscilloscope to the factory settings, do the following: 1. Press MENU key and the function menu appears on the right side of the screen 14

23 6-Using the scope 2. Press the MENU or MENU key to select FUNCTION setting and three options are visible at the bottom of the screen. 3. Press F1 key to select the factory settings. The oscilloscope is set to be the factory settings. See the following figure 9: figure 9: Reset the Oscilloscope 6.7 Input Connections See figure 2. Look at the bottom and the right of the Oscilloscope. The Oscilloscope has six signal inputs: one safety BNC jack inputs for scope measurements, three safety 4-mm banana jack inputs for Multimeter R, V and A measurements, and two quadratic jack inputs for Multimeter capacitance measurements. Isolated input allows independent floating measurements between Multimeters and Scopes. 6.8 Displaying an Unknown Signal with Auto Set The Auto-Set feature lets the Oscilloscope display and measure unknown signals automatically. This function optimizes the position, range, time base, triggering and assures a stable display of virtually any waveform. This feature is especially useful for quickly checking several signals. To enable the Auto-Set feature, do the following: 1. Connect the test probe to the tested signals. 2. Press the AUTO SET key and the Oscilloscope is under the automatic measurement condition. The tested signals appear on the screen. 15

24 6-Using the scope 6.9 Automatic Zero-returning of Trigger Horizontal Position and 50% setting of Trigger Level Position When we adjust the trigger horizontal position and trigger level position to be maximal to make it off the screen center remotely, then we perform the following steps to make trigger horizontal position return to zero and trigger level position set to 50% automatically. 1. Press V key, the trigger horizontal position automatically returns to zero. 2. Press R key, the trigger level position automatically sets to 50% Automatic Measurements The Oscilloscope offers 18 types of automatic scope measurements. Your can display two numeric readings: measurement 1 and measurement 2. These readings are selectable independently, and the measurements can be done on the input waveform. To choose a average value measurement for measurement 1, do the following: 1. Press MENU key and the function menu appears on the right side of the screen. 2. Press MENU or MENU key to select MEAS SET1. Five items selectable are visible at the bottom of the screen. 3. Press F3 key to select Mean. The measurement 1 window shows the average value--v. To choose a Peak-Peak measurement for measurement 2, do the following: 1. Press MENU key and the function menu is displayed on the right side of the screen. 2. Press MENU or MENU key and select MEAS SET2, with 5 items selectable displayed at the bottom of the screen. 3. Press F4 key to select PK-PK. The measurement 2 window shows the peak-peak value--vp. See the following figure 10: 16

25 6-Using the scope figure 10: Automatic Scope Measurements 6.11 Freezing the Screen You can freeze the screen (all readings and waveforms) 1. Press the RUN/STOP key to freeze the screen and STOP appears at top right side of the screen. 2. Press the RUN/STOP key once more to resume your measurement. See the following figure 11: figure 11: Freezing the Screen 6.12 Using Average for Smoothing Waveforms To smooth the waveform, do the following: 1. Press the MENU key and the function menu appears on the right side of the screen. 2. Press MENU or MENU key to select ACQU MODE, with four items selectable 17

26 6-Using the scope displayed at the bottom of the screen. 3. Press the F3 key to select Average, then, press F4 key to jump to Averaging 16 item. This averages the outcomes of 16 acquisitions and shows the final averaging result on the screen, shown as the following figures. See the following figure 12: figure 12: Average Factor Sampling Mode 6.13 Using Persistence to Display Waveforms You can use Persistence to observe dynamic signals. 1. Press MENU key and the function menu appears on the right side of the screen. 2. Press MENU or MENU key to select DISP SET. Four items selectable are displayed at the bottom of the screen. 3. Press F2 key to select Persist 1 sec, 2 sec, and 5 sec, infinite or close. In this case, jump to Infinite and the observed dynamic is kept on the screen continuously. When the item Close is selected, the Persistence function is closed. Look at the display, a screen like the following figure 13 can be shown. 18

27 6-Using the scope figure 13: Persistence to Observe Dynamic Signals 6.14 Using Peak Detection to Display Glitches You can use this function to display events (glitches or other asynchronous waveforms) of 50 ns or wider. 1. Press MENU key and the function menu appears at the right side of the screen. 2. Press MENU or MENU key to select the ACQU MODE. Four items selectable are displayed at the bottom of the screen. 3. Press F2 key and jump to Peak Detect. In this case, you can test the glitch. Now, you can see a screen that looks like the following figure 14. Term interpretation figure 14: Peak Detection Collecting mode: The oscilloscope transforms the collected analog data into a digital form after 19

28 6-Using the scope they are gathered in the following three different modes, such as sampling, peak value detection and averaging values. Sampling: The oscilloscope takes samples from the signal at a equal time interval to reconstruct the waveform in this mode, by which the analog signal can be expressed correctly in most cases, yet, the rapid changes can not be collected between two sampling time intervals, causing the confusion and loss the narrow pulse in the signal probably. Peak value detection: The oscilloscope takes samples from the maximum and minimum of signals in each sampling interval and shows the waveform with the sampled data in this mode, thus, by which the oscilloscope may collect the possibly lost narrow pulse in the sampling mode but the noise is obvious. Averaging values: the oscilloscope collects several waveforms and average over them, and displays the averaged waveform in this mode, by which the random noise can be reduced. Duration time: When a new waveform is displayed, the previous waveform shown on the screen does not disappear immediately only to be displayed for a period of time, that is, the duration time, by setting which, the waveform can be displayed more continuously and thus a display similar to that shown by the analog oscilloscope can be achieved. Roll scan mode: The oscilloscope updates the waveform sampling points by scrolling display through a screen from left to right in this mode, which is only applicable to the primary time base setting of above 50ms Selecting AC-coupling After a reset, the Oscilloscope is dc-coupled so that ac and dc voltages appear on the screen, Use ac-coupling when you wish to observe a small ac signal that rides on a dc signal. To select ac-coupling, do the following: 1. Press MENU key and the function menu appears at the right side of the screen. 2. Press MENU or MENU key to select the CH SETUP. Four items selectable are visible at the bottom of the screen. 3. Press the F1 key and jump to AC. The bottom left side of the screen displays the ac-coupling icon. Now, you can see a screen that looks like the following figure

29 6-Using the scope figure 15: AC-Coupling 6.16 Reversing the Polarity of the Displayed Waveform To invert the input signal waveform, do the following: 1. Press the MENU key and the function menu appears at the right side of the screen. 2. Press the MENU or MENU key to select CH SETUP. Four items selectable are displayed at the bottom of the screen. 3. Press F4 key to jump to Inverted. The inverted waveform of the channel is displayed on the screen. Now, you can see a screen that looks like the following figure 16. figure 16: Inverted On 21

30 7-Using the Multimeter 7.Using the Multimeter 7.1 About this Chapter This chapter provides a step-by-step introduction to the multi-meter functions of the test tool hereafter. The introduction gives basic examples to show how to use the menus and perform basic operations. 7.2 Making Meter Connections Use the three 4-mm safety banana jack inputs for the Meter functions: COM, V/Ω, ma. Two quadratic capacitance jacks: CX See figure 2 for the connections. 7.3 Multimeter Operation Window figure 17: Multimeter Operation Window Description 1. Battery electric quantity indictor. 2. Manual/Auto range indictors, among which the MANUAL means measuring range in manual 22

31 7-Using the Multimeter operation mode and Auto refers to the measuring range in automatic operation mode. 3. Measurement mode indicators: DCV: Direct voltage measurement ACV: Alternating voltage measurement DCA: Direct current measurement ACA: Alternating current measurement R: Resistance measurement : Diode measurement : On/Off measurement C: Capacitance measurement 4. The relative magnitude measurement indicator. 5. Running state indicators, among which RUN expresses continuous update and STOP represents the screen locking. 6. The reference value of the relative magnitude measurement. 7. The multiplying power of the dial indication. To multiply the reading of dial pointer by multiplying power will get the measurement result. 8. The mail reading of measurement 9. Automatic control measuring range. 10. Absolute/ relative magnitude measuring control: The sign expresses the absolute magnitude measuring control and represents the relative magnitude measuring control. Manually measuring range control. 11. Manual measurement control. 12. Test lead indicated the scale of test reading, different test modes display different colors. 7.4 Making Multimeter Measurements Press DMM/OSC key, the oscilloscope will switch to the multimeter measure, the screen will display the multimeter windows, at the same time, prompt to correctly insert testing pen of the multimeter, at this time, then press any key to enter into multimeter measure. 23

32 7-Using the Multimeter Measuring Resistance Values To measure a resistance, do the following: 1. Press the R key and R appear at the top of the screen. 2. Insert the black lead into the COM banana jack input and the red lead into the V/Ω banana jack input. 3. Connect the red and black test leads to the resistor. The resistor value readings are shown on the screen in Ohm. Now, you can see a screen that looks like the following figure 18. figure 18: Resistance Measurement Making a Diode Measurement To make a measurement on the diode, do the following: 1. Press the R key and R appear at the top of the screen. 2. Press AUTO SET key till the following is displayed on the screen. 3. Insert the black lead into the COM banana jack input and the red lead into the V/Ω banana jack input. 4. Connect the red and black leads to the resistor and the diode resistor readings are displayed on the screen in V. Now, you can see a screen that looks like the following figure

33 7-Using the Multimeter figure 19: Diode Measurement On-off Test To perform an On-off test, do the following: 1. Press the R key and R appears on the top of the screen. 2. Press the AUTO SET key till the following is shown on the screen. 3. Insert the black lead into the COM banana jack input and the red lead into the V/Ω banana jack input. 4. Connect the red and black leads to the test point. If the resistance value of the tested point is less than 50Ω, you will hear beep sound from the test tool. Now, you can see a screen that looks like the following figure 20. figure 20: On-off Test 25

34 7-Using the Multimeter Making a Capacitance Measurement To measure a capacitance, do the following: 1. Press the R key and R appears on the top of the screen 2. Press the AUTO SET key till C appears at the top of the screen. 3. Insert the measured capacitance into the quadratic jack and the screen shows the capacitance reading. Notice: when measured value is less than 5 nf capacitance, please use small capacitance measurer of this multimeter and use relative value measuring mode to improve measuring precision. It will take about 30seconds if capacitance measurement is large than 40uF. Now, you can see a screen that looks like the following figure 21. figure 21: Capacitance Measurement Making a DC Voltage Measurement To measure a DC voltage, do the following: 1. Press the V key and DCV appears at the top of the screen. 2. Insert the black lead into the COM banana jack input and the red lead into the V/Ω banana jack input. 3. Connect the red and black leads to the measured point and the measured point voltage value is displayed on the screen. Now, you can see a screen that looks like the following figure

35 7-Using the Multimeter figure 22: DC Voltage Measurement Making a AC Voltage Measurement To measure the AC voltage, do the following: 1. Press the V key and DCV appears at the top of the screen. 2. Press the AUTO SET key and ACV appears at the top of the screen. 3. Insert the black lead into the COM banana jack input and the red lead into the V/Ω banana jack input. 4. Connect the red and black leads to the measured points and the AC voltage values of measured points will be displayed on the screen. Look at the display, you can see a screen that looks like the following figure 23. figure 23: AC Voltage Measurement 27

36 7-Using the Multimeter Making a DC Current Measurement To measure a DC current which is less than 400 ma, do the following: 1. Press the A key and DCA appears at the top of the screen. The unit on the main reading screen is ma. ma and 20A will display on the right bottom of screen, press F4 or F5 to switch the measurement between ma and 20A. 400mA is acquiescently. 2. Insert the black lead into the COM banana jack input and the red lead into the ma banana jack input. 3. Connect the red and black leads to the measured points and the DC current values of measured points will be displayed on the screen. Look at the display, you can see a screen that looks like the following figure 24. figure 24: DC Current Measurement for 400 ma To measure a DC current which is larger than 400 ma, do the following: 1. Press the A key and DCA appears at the top of the screen. The unit on the main reading screen is ma.. 2. Press F5 key change to 20A measurement, the unit on the main reading screen is A. 3. Plug current extended module in current measure jack, then plug the probe in the module 4. Connect the red and black leads to the measured point and the DC current value of the measured point will be displayed on the screen. 5. Press F4 return to 400 ma measure. Look at the display, you can see a screen that looks like the following figure

37 7-Using the Multimeter figure 25: DC Current Measurement for 20A Making an AC Current Measurement To measure an AC current which is less than 400 ma, do the following: 1. Press the A key and DCA appears at the top of the screen. The unit on the main reading screen is ma. ma and 20A will display on the right bottom of screen, press F4 or F5 to switch the measurement between ma and 20A. 400mA is acquiescently. 2. Press the AUTO SET key once and ACA is visible at the top of the screen.. 3. Insert the black lead into the COM banana jack input and the red lead into the ma banana jack input. 4. Connect the red and black leads to the measured point and the AC current value of the measured point will be displayed on the screen. Look at the display, you can see a screen that looks like the following figure 26. figure 26: AC Current Measurement for 400 ma 29

38 7-Using the Multimeter To measure an AC current which is larger than 400 ma, do the following: 1. Press the A key once and DCA is visible on the top of the screen. 2. Press F5 to select 20A measure, the unit of main reading window is A. 3. Press the AUTO SET key once and ACA is visible on the top of the screen. 4. Plug current extended module in current measure jack, then plug the probe in the module. 5. Connect the red and black leads to the measured point and the AC current value of the measured point will be displayed on the screen. 6. Press F4 return to 400mA measure. Look at the display, you can see a screen that looks like the following figure 27. figure 27: AC Current Measurement for 20A 7.5 Freezing the Readings You can freeze the displayed readings at any time. 1. Press the RUN /STOP key to freeze the screen and STOP will be displayed at the top right of the screen. 2. Again, press the RUN /STOP key, you can resume your measurement. Look at the display, you can see a screen that looks like the following figure

39 7-Using the Multimeter figure 28: Freezing the Readings 7.6 Taking a Relative Measurement A currently measured result relative to the defined reference value is displayed in a relative measurement. The following example shows how to take a relative measurement. At first, it is required to acquire a reference value. 1. Press R key and R is displayed on the top side of the screen. 2. Press the AUTO SET key till C appears at the top of the screen. 3. Plug capacitance extended module in capacitance measure jack. 4. When the reading leveling off, press F2 key and is displayed on the top side of the screen. The saved reference value is displayed below. 5. Plug capacitance, the displayed major reading on the screen is actual the capacitance value. Look at the display, you can see a screen that looks like the following figure 29. figure 29: Relative Measurement 31

40 7-Using the Multimeter 7.7 Selecting Automatic/ManualRange Adjustment The defaulted range mode of the instrument is automatic range. To switch to the manual range, perform the following steps: 1. Press F1 key and MANUAL is displayed on the top left side of the screen to enter the manual range mode. 2. Under the manual range mode, the measuring range is increased by a stage when pressing F1 key each time, and when reaching the highest stage, it jumps to the lowest stage by pressing F1 key once again. To multiply the reading of dial pointer by multiplying power and the unit of main reading on the screen will get the measurement result. 3. Press F3 key and AUTO is displayed on the top left side of the screen to switch back to the automatic range mode. Look at the display, you can see a screen that looks like the following figure 30. figure 30: Automatic/Manual Range Adjustment Attention:capacitance measurement without manual range mode. 32

41 8-Advanced Function of Oscilloscope 8.Advanced Function of Oscilloscope 8.1 About this Chapter This chapter will detail the oscilloscope function of the test tool. 8.2 Setting the Channel Each channel has its own independent vertical menu and each item can be set respectively based on the specific channel. To make channel settings, do the following: 1. Press the MENU key and the function menu appears at the right of the screen. 2. Press the MENU or MENU key to jump to CH SETUP and 4 options appears at the bottom of the screen. 3. Select and press key from F1 through F4 keys to make different settings. Now, you can find a screen that looks like the following figure 31. figure 31: Setting the Vertical The following Table describes the Vertical Channel menu: 33

42 8-Advanced Function of Oscilloscope Function menu Setting Description Coupling AC DC Ground The DC component in the input signal is blocked. The AC and DC components of the input signal are allowed. Input signal is interrupted. Channel OFF Close the channel. ON Open the channel. 1X Probe 10X Select one according the probe attenuation factor to ensure a 100X correct vertical scale reading. 1000X Inverted OFF Waveform is displayed normally. ON Open the Invert function of the waveform setting Setting the Channel Coupling For example, the measured signal is a square wave signal containing a DC offset. Press F1 Coupling first and then AC to make an ac coupling setting. The DC component contained in the tested signal is blocked. Press F1 Coupling first and then DC to make a DC coupling setting. Both dc and ac components contained in the tested signal are permitted. Press F1 Coupling first and then Ground to make a Ground coupling setting. Input signal is interrupted. The waveform is displayed as the following figure 32, figure 33, figure 34. figure 32: AC Coupling 34

43 8-Advanced Function of Oscilloscope figure 33: DC Coupling figure 34: Ground Coupling Make Open and Close Settings on Channel Press F2 Channel key first, then select OFF to make a Close setting on channel. Press F2 Channel key first, then select ON to make an Open setting on channel Adjusting the Probe Scale It is necessary to adjust the probe attenuation scale factor correspondingly in the channel operation menu in order to comply with the probe attenuation scale. If it is a 10:1 probe, the scale of the input channel of the oscilloscope should be selected as 10X to avoid any error occurring in the displayed scale factor information and tested data. Press F3 Probe to jump to the relative probe. Table: Probe attenuation factor and the corresponding menu setting 35

44 8-Advanced Function of Oscilloscope Probe attenuation factor Corresponding Menu Setting 1:1 1X 10:1 10X 100:1 100X 1000:1 1000X Setting of Inverted Waveform Inverted waveform: The displayed signal reverses 180 degrees relatively to the ground potential. Press F4 Inverted to start Invert; again press F4 Inverted to close Invert. 8.3 Setting the Trigger System The Trigger defines the time when the acquisition of data and display of waveform start. If it is set correctly, the trigger can turn an unstable display into a significant waveform. When starting the acquisition of data, the oscilloscope collects sufficient data to draw the waveform at the left side of the triggering point. With waiting for the triggering condition, the oscilloscope is gathering data continuously. After a trigger is detected, the oscilloscope gathers enough data continuously to draw the waveform at the right side of the triggering point. To make a trigger mode setting, do the following: 1. Press the MENU key and the function menu appears at the right of the screen. 2. Press the MENU or MENU key to select TRIG MODE and five items selectable are displayed at the bottom of the screen. 3. Select and press one from F1 through F5 key to make a different setting. 4. Press the OSC OPTION key and the following is shown on the screen: / Time / Trig 5. Press the OSC or OSC key to adjust the trigger level position. Now, look at the display, you can see a screen in the following figure35. 36

45 8-Advanced Function of Oscilloscope figure 35: Edge Trigger 8.4 Triggering Control There are two triggering modes including Edge triggering and Video triggering. Each trigger mode is set by different function menu. Edge triggering: It occurs when the trigger input passes through a given level along the specified direction. Video triggering: Perform video field trigger, line trigger, odd field trigger, even field trigger or line NO. trigger on the standard video signals. The following describes the Edge triggering and Video triggering menus respectively Edge Triggering The Edge triggering is a mode by which trigger occurs at the triggering threshold value of the input signal edge. With the Edge triggering selected, the trigger happens on the rise or fall edge of the input signal, shown as figure 35. The Edge triggering menu is described in the following table. Function menu Settings Description Slope Rising Triggering on the rise edge of the signal. Falling Triggering on the fall edge of the signal. Auto Acquisition of waveforms is possible even if there is no triggering condition detected. Mode Normal Acquisition of waveforms can only be done when the triggering condition is satisfied. Single The sampling is performed on a waveform when one trigger is detected, then stop sampling. Sens 0.2~1.0div Set trigger sensitivity. Holdoff To go to holdoff menu 37

46 8-Advanced Function of Oscilloscope Video Triggering With Video triggering selected, the oscilloscope performs the NTSC, PAL or SECAM standard video signals field, line,odd field,even field or line num trigger. Figure 36 is the Video Line Trigger, figure 37,figure 38 is the Video Line trigger. figure 36: Video Line Trigger figure 37: Video Line NO. trigger(page1) figure 38: Video Line NO. trigger(page2) when the sync is Line,Field,Odd Field,Even Field,the menu is shown as bellow. Function menu Settings Description Line Make a video line trigger synchronization setting. Sync Field Make a video field trigger synchronization setting. Odd Field Make a video odd field trigger synchronization setting. Even Field Make a video even field trigger synchronization setting. MODU Holdoff NTSC PAL/SECAM Video format setting Go to the holdoff menu. when the sync is Designed Line,the menu is shown as bellow. 38

47 8-Advanced Function of Oscilloscope Function menu Settings Description Sync Designed Line Make a video designed line synchronization setting. MODU Holdoff NTSC PAL/SECAM Video format setting Go to the holdoff menu. Go to the next menu. Line Line No. increase decrease Set the line value to increase. Set the line value to decrease. Set and Show the line valve. Back up to previous menu. When you go to the holdoff menu,you can see a screen in the following figure39. Figure39:Trigger Holdoff the Holdoff menu is described in the following table: Function menu Settings Description Off Time 100ns~10s Set time slot before another trigger event. Time Reset Off Time Back increase decrease Set the off time to increase. Set the off time to decrease. Reset Holdoff time to100ns. Back up to previous menu. Note: Trigger Holdoff can stabilize complex waveform,such as the pulse range.holdoff time is the oscilloscope s waiting period before starting a new trigger.during Holdoff,oscilloscope will not 39

48 8-Advanced Function of Oscilloscope trigger until Holdoff ends. Term interpretation Trigger modes: There are three kinds of trigger modes available for this oscilloscope, they are auto, normal and single shot. Automatic trigger mode: The oscilloscope can acquire the waveform without any triggering condition detected in this mode, in which it will be triggered compulsively when waiting for a specified period of time without any triggering condition ignited When an invalid trigger is enforced,the oscilloscope can not keep the waveform in phase. Normal trigger mode: In this mode, the oscilloscope cannot acquire the waveform till it is triggered. When there is not any trigger, the oscilloscope will display the origin waveform without new waveforms captured. Single shot mode: In this mode, the oscilloscope will detect a trigger and capture a waveform at each time when the customer presses the RUN/STOP key. Holdoff: Set time slot before another trigger event. 8.5 Acquiring Mode Setting The Acquiring Mode menu is described in the list shown as below. Function menu Settings Description Sample Normal sampling mode. Peak Detect Used to detect the jamming glitch and reduce the possible blurring. Average Used to reduce the random and unrelated noises. Several average factors are available for being selected. Averages 4, 16, 64 Select the average factor. or Display Setting The DISP SET menu is described in the following table. Function menu Settings Description Type Vectors Dots The vector is filled up spaces between neighboring sampling points in the display. Only sampling points are displayed. Persist OFF 1s 2s 5s Infinite Setting persistence time for each sampling point. 40

OWON HDS1022M Handheld Digital Storage Oscilloscope & Multimeter User s Manual

OWON HDS1022M Handheld Digital Storage Oscilloscope & Multimeter User s Manual OWON HDS1022M Handheld Digital Storage Oscilloscope & Multimeter User s Manual WWW.OWON.COM.CN Contents CONTENTS 1. SAFETY INFORMATION...5 1.1. SAFETY TERMS AND SYMBOLS...5 1.1.1. Specific warning and

More information

OWON HDS2062M Handheld Digital Storage Oscilloscope & Multimeter User s Manual

OWON HDS2062M Handheld Digital Storage Oscilloscope & Multimeter User s Manual OWON HDS2062M Handheld Digital Storage Oscilloscope & Multimeter User s Manual 1 LIMITED WARRANTY & LIMITATON OF LIABILITY Each OWON product is warranted to be free from defects in material and workmanship

More information

OS Series Handheld Digital Storage. Oscilloscope & Multimeter. User Manual

OS Series Handheld Digital Storage. Oscilloscope & Multimeter. User Manual OS Series Handheld Digital Storage Oscilloscope & Multimeter User Manual OS-1022 OS-2062 20MHz 60 MHz General Warranty We warrants that the product will be free from defects in materials and workmanship

More information

Model P2016. Dual Channel Handheld Digital Storage. Oscilloscope & Multimeter. Quick Scope Guide

Model P2016. Dual Channel Handheld Digital Storage. Oscilloscope & Multimeter. Quick Scope Guide Model P2016 Dual Channel Handheld Digital Storage Oscilloscope & Multimeter Quick Scope Guide General Warranty BNC warrants that the product will be free from defects in materials and workmanship for 3

More information

OS-1022 Handheld Digital Storage. Oscilloscope & Multimeter. User Manual

OS-1022 Handheld Digital Storage. Oscilloscope & Multimeter. User Manual OS-1022 Handheld Digital Storage Oscilloscope & Multimeter User Manual General Warranty We warrants that the product will be free from defects in materials and workmanship for a period of three years from

More information

OWON. HDS2062M-N Handheld Digital Storage Oscilloscope & Multimeter User s Manual

OWON. HDS2062M-N Handheld Digital Storage Oscilloscope & Multimeter User s Manual OWON HDS2062M-N Handheld Digital Storage Oscilloscope & Multimeter User s Manual WWW.OWON.COM.CN 1 WARRANTY Lilliput warrants that this OWON brand product will be free from defects in materials and workmanship

More information

HDS-N Dual Channel Series Handheld Digital Storage. Oscilloscope & Multimeter. User Manual

HDS-N Dual Channel Series Handheld Digital Storage. Oscilloscope & Multimeter. User Manual HDS-N Dual Channel Series Handheld Digital Storage Oscilloscope & Multimeter User Manual HDS1022M-N HDS1022M-I HDS2062M-N HDS3102M-N HDS4202M-N Note: "I" indicates that the two channels are isolated electrically.

More information

72-8 Series Handheld Digital Storage. Oscilloscope & Multimeter. User Manual

72-8 Series Handheld Digital Storage. Oscilloscope & Multimeter. User Manual 72-8 Series Handheld Digital Storage Oscilloscope & Multimeter User Manual 72-8470 72-8474 www.tenma.com 1 WARRANTY Tenma Test Equipment warrants that this product will be free from defects in materials

More information

Model P2016. Dual Channel Handheld Digital Storage. Oscilloscope & Multimeter. User Manual

Model P2016. Dual Channel Handheld Digital Storage. Oscilloscope & Multimeter. User Manual Model P2016 Dual Channel Handheld Digital Storage Oscilloscope & Multimeter User Manual April 2018 Edition, Version 1.6.3BNC Copyright. All rights reserved. The Berkeley Nucleonics products are under the

More information

PeakTech 1195 / Operation manual. Digital Storage Oscilloscopes/DMM

PeakTech 1195 / Operation manual. Digital Storage Oscilloscopes/DMM PeakTech 1195 / 1205 Operation manual Digital Storage Oscilloscopes/DMM Package Contents (see the picture below) Description 1. PeakTech 1195/1205 incl. Accu 2. Power Adaptor 3. Oscilloscope Probes 4.

More information

Quick Start. RSHS1000 Series Handheld Digital Oscilloscope

Quick Start. RSHS1000 Series Handheld Digital Oscilloscope Quick Start RSHS1000 Series Handheld Digital Oscilloscope General Safety Summary Carefully read the following safety precautions to avoid personal injury and prevent damage to the instrument or any products

More information

Instructions. P MHz 1X/10X Passive Probe

Instructions. P MHz 1X/10X Passive Probe Instructions P2100 100 MHz 1X/10X Passive Probe 071-0774-01 071077401 Copyright Tektronix, Inc. All rights reserved. Tektronix products are covered by U.S. and foreign patents, issued and pending. Information

More information

Quick Start. SHS1000 Series Handheld Digital Oscilloscope QS03010-E02B 2015 SIGLENT TECHNOLOGIES CO., LTD

Quick Start. SHS1000 Series Handheld Digital Oscilloscope QS03010-E02B 2015 SIGLENT TECHNOLOGIES CO., LTD Quick Start SHS1000 Series Handheld Digital Oscilloscope QS03010-E02B 2015 SIGLENT TECHNOLOGIES CO., LTD Guaranty and Declaration Copyright SIGLENT TECHNOLOGIES CO., LTD. All Rights Reserved. Trademark

More information

RIGOL. Data Sheet. DS1000B Series Digital Oscilloscopes DS1074B, DS1104B, DS1204B. Product Overview. Easy to Use Design. Applications.

RIGOL. Data Sheet. DS1000B Series Digital Oscilloscopes DS1074B, DS1104B, DS1204B. Product Overview. Easy to Use Design. Applications. RIGOL Data Sheet Product Overview DS1000B series oscilloscopes are designed with four analog channels and 1 external trigger channel, which can capture multi-channel signal simultaneously and meet industrial

More information

Digital Storage Oscilloscopes 2550 Series

Digital Storage Oscilloscopes 2550 Series Data Sheet Digital Storage Oscilloscopes 2550 Series The 2550 series digital storage oscilloscopes provide high performance and value in 2-channel and 4-channel configurations. With bandwidth from 70 MHz

More information

Fluke 196C/199C. ScopeMeter. Users Manual

Fluke 196C/199C. ScopeMeter. Users Manual Fluke 196C/199C ScopeMeter Users Manual 4822 872 30482 October 2001 2001 Fluke Corporation. All rights reserved. Printed in the Netherlands. All product names are trademarks of their respective companies.

More information

KHT 1000C HV-Probe Calibrator. Instruction Manual

KHT 1000C HV-Probe Calibrator. Instruction Manual KHT 1000C HV-Probe Calibrator Instruction Manual Copyright 2015 PMK GmbH All rights reserved. Information in this publication supersedes that in all previously published material. Specifications are subject

More information

2016 RIGOL TECHNOLOGIES, INC.

2016 RIGOL TECHNOLOGIES, INC. RIGOL Data Sheet Product Overview DS1000B series oscilloscopes are designed with four analog channels and 1 external trigger channel, which can capture multi-channel signal simultaneously and meet industrial

More information

TDS Series. User Manual

TDS Series. User Manual 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com TDS Series Touchscreen Digital Storage Oscilloscopes User Manual TDS7104 TDS8104

More information

Burlington County College INSTRUCTION GUIDE. for the. Hewlett Packard. FUNCTION GENERATOR Model #33120A. and. Tektronix

Burlington County College INSTRUCTION GUIDE. for the. Hewlett Packard. FUNCTION GENERATOR Model #33120A. and. Tektronix v1.2 Burlington County College INSTRUCTION GUIDE for the Hewlett Packard FUNCTION GENERATOR Model #33120A and Tektronix OSCILLOSCOPE Model #MSO2004B Summer 2014 Pg. 2 Scope-Gen Handout_pgs1-8_v1.2_SU14.doc

More information

Tablet Oscilloscope Quick Guide

Tablet Oscilloscope Quick Guide Tablet Oscilloscope Quick Guide For tbook Series Shenzhen Micsig Instruments Co., Ltd. Copyright Copyright Shenzhen Micsig Instruments Co., Ltd. All Rights Reserved. Version Version: MKX2014-001; Product

More information

ScopeMeter 190 Series Specifications

ScopeMeter 190 Series Specifications Seite 1 von 7 ScopeMeter 190 Series Specifications Product Home Features Specifications Models, Options & Accessories Oscilloscope Mode Meter Mode Recorder Mode General Specifications Oscilloscope Mode

More information

Technical Reference. TDS 684A, TDS 744A, & TDS 784A Digitizing Oscilloscope Performance Verification and Specifications

Technical Reference. TDS 684A, TDS 744A, & TDS 784A Digitizing Oscilloscope Performance Verification and Specifications Technical Reference TDS 684A, TDS 744A, & TDS 784A Digitizing Oscilloscope Performance Verification and Specifications 070-8990-04 Please check for change information at the rear of this manual. Fifth

More information

INSTRUCTIONAL MANUAL FOR LCD ZOOM MICROSCOPE

INSTRUCTIONAL MANUAL FOR LCD ZOOM MICROSCOPE INSTRUCTIONAL MANUAL FOR LCD ZOOM MICROSCOPE ? 8 LCD Screen? 10.4 LCD Screen LCD Zoom Microscope Instruction Manual Please read the Instruction Manual carefully before installation and keep it for future

More information

User Manual. Digital Storage Oscilloscopes Models 2534, 2540 & 2542

User Manual. Digital Storage Oscilloscopes Models 2534, 2540 & 2542 User Manual Digital Storage Oscilloscopes Models 2534, 2540 & 2542 General Safety Summary General Safety Summary Review the following safety precautions to avoid injury and prevent damage to this product

More information

User Manual. Digital Storage Oscilloscopes Models 2534, 2540 & General Safety Summary. Version 1.03

User Manual. Digital Storage Oscilloscopes Models 2534, 2540 & General Safety Summary. Version 1.03 General Safety Summary General Safety Summary User Manual Digital Storage Oscilloscopes Models 2534, 2540 & 2542 Review the following safety precautions to avoid injury and prevent damage to this product

More information

Operating Instructions

Operating Instructions Operating Instructions HAEFELY TEST AG KIT Measurement Software Version 1.0 KIT / En Date Version Responsable Changes / Reasons February 2015 1.0 Initial version WARNING Introduction i Before operating

More information

EGM Einthoven Goldberger Module Type 701

EGM Einthoven Goldberger Module Type 701 Operating Instructions for the PLUGSYS Module EGM Einthoven Goldberger Module Type 701 ECG amplifier for bipolar extremity leads after Einthoven and unipolar extremity leads after Goldberger (Version:

More information

Electrical and Electronic Laboratory Faculty of Engineering Chulalongkorn University. Cathode-Ray Oscilloscope (CRO)

Electrical and Electronic Laboratory Faculty of Engineering Chulalongkorn University. Cathode-Ray Oscilloscope (CRO) 2141274 Electrical and Electronic Laboratory Faculty of Engineering Chulalongkorn University Cathode-Ray Oscilloscope (CRO) Objectives You will be able to use an oscilloscope to measure voltage, frequency

More information

Advanced Test Equipment Rentals ATEC (2832)

Advanced Test Equipment Rentals ATEC (2832) E stablished 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) Technical Datasheet Scalar Network Analyzer Model 8003-10 MHz to 40 GHz The Giga-tronics Model 8003 Precision Scalar

More information

4CH/2CH Digital Storage Oscilloscopes

4CH/2CH Digital Storage Oscilloscopes Model: 2553, 2555, 2556, 2557, 2558, 2559 4CH/2CH Digital Storage Oscilloscopes USER MANUAL Safety Summary The following safety precautions apply to both operating and maintenance personnel and must be

More information

User Guide. Single-Link DVI Active Cable Extender. DVI-7171c

User Guide. Single-Link DVI Active Cable Extender. DVI-7171c User Guide Single-Link DVI Active Cable Extender DVI-7171c TABLE OF CONTENTS SECTION PAGE PRODUCT SAFETY...1 PRODUCT LIABILITY...1 1.0 INTRODUCTION...2 2.0 SPECIFICATIONS...3 3.0 PACKAGE CONTENTS...4 4.0

More information

PSM-003. Micro Polarization Controller/Scrambler. User Guide

PSM-003. Micro Polarization Controller/Scrambler. User Guide PSM-003 Micro Polarization Controller/Scrambler User Guide Version: 1.0 Date: August 23, 2012 General Photonics, Incorporated is located in Chino California. For more information visit the company's website

More information

Agilent Technologies 54522A

Agilent Technologies 54522A Agilent Technologies 54522A Data Sheet Product Specifications General Specifications Maximum Sample Rate 54522A 2 GSa/s Number of Channels (all are simultaneous acquisition) 54522A: 2 Record Length 32,768

More information

TDS 520B, TDS 540B, TDS 620B, TDS 644B, TDS 680B, TDS 684B, TDS 724A, TDS 744A, & TDS 784A

TDS 520B, TDS 540B, TDS 620B, TDS 644B, TDS 680B, TDS 684B, TDS 724A, TDS 744A, & TDS 784A Technical Reference TDS 520B, TDS 540B, TDS 620B, TDS 644B, TDS 680B, TDS 684B, TDS 724A, TDS 744A, & TDS 784A Digitizing Oscilloscopes Performance Verification and Specifications 070-9384-01 Copyright

More information

SDS1000C Specifications

SDS1000C Specifications SDS1000C Specifications File Version V1.2 Siglent technology Co., Ltd CHARACTERISTIC: The highest Single real-time sampling rate can be up to 500MHzsa/s; Equivalent sampling rate is up to 10GSa/s. Memory

More information

Fluke /

Fluke / Fluke 190-104/190-204 ScopeMeter 190 Series II Service Manual PN 4822 872 05405 March 2011 2011 Fluke Corporation, All rights reserved. Printed in the Netherlands All product names are trademarks of their

More information

17 19 PROFESSIONAL LCD COLOUR MONITOR ART

17 19 PROFESSIONAL LCD COLOUR MONITOR ART 17 19 PROFESSIONAL LCD COLOUR MONITOR ART. 41657-41659 Via Don Arrigoni, 5 24020 Rovetta S. Lorenzo (Bergamo) http://www.comelit.eu e-mail:export.department@comelit.it WARNING: TO REDUCE THE RISK OF FIRE

More information

Selection Table. Ordering Information

Selection Table.  Ordering Information Selection Table Fluke 199 Fluke 196 Fluke 192 Fluke 123 Bandwidth 200 MHz 100 MHz 60 MHz 20 MHz Max. real-time sample rate 2.5 GS/s 1 GS/s 500 MS/s 25 MS/s Inputs and digitizers 2 plus external / DMM input

More information

LeCroy Digital Oscilloscopes

LeCroy Digital Oscilloscopes LeCroy Digital Oscilloscopes Get the Complete Picture Quick Reference Guide QUICKSTART TO SIGNAL VIEWING Quickly display a signal View with Analog Persistence 1. Connect your signal. When you use a probe,

More information

Oscilloscopes for field applications

Oscilloscopes for field applications Oscilloscopes for field applications ScopeMeter 123 and 190 Series 20 MHz to 200 MHz bandwidth Up to 2.5 GS/s real-time sampling Up to 5 hours battery operating time Easy to use with Connect-and-View triggering

More information

Model: S-4904T/R. Wireless HD Transmission System. User Manual. Please read this User Manual throughout before using.

Model: S-4904T/R. Wireless HD Transmission System. User Manual. Please read this User Manual throughout before using. Model: S-4904T/R Wireless HD Transmission System User Manual Please read this User Manual throughout before using. Preface Congratulations on your purchase of this product. Please read this user manual

More information

Industriefunkuhren. Technical Manual. IRIG-B Generator-Module for analogue / digital Signals of Type: IRIG-B / IEEE C / AFNOR NF S87-500

Industriefunkuhren. Technical Manual. IRIG-B Generator-Module for analogue / digital Signals of Type: IRIG-B / IEEE C / AFNOR NF S87-500 Industriefunkuhren Technical Manual IRIG-B Generator-Module for analogue / digital Signals of Type: IRIG-B / IEEE C37.118 / AFNOR NF S87-500 Module 7628 ENGLISH Version: 02.01-06.03.2013 2 / 20 7628 IRIG-B

More information

PicoScope 6407 Digitizer

PicoScope 6407 Digitizer YE AR PicoScope 6407 Digitizer HIGH PERFORMANCE USB DIGITIZER Programmable and Powerful 1 GHz bandwidth 1 GS buffer size 5 GS/s real-time sampling Advanced digital triggers Built-in function generator

More information

TVAC20000 User manual

TVAC20000 User manual TVAC20000 User manual Version 01/2010 Original English user manual. Keep for future use. 10 Introduction Dear Customer, Thank you for purchasing this product. This product meets the requirements of the

More information

HD Digital Set-Top Box Quick Start Guide

HD Digital Set-Top Box Quick Start Guide HD Digital Set-Top Box Quick Start Guide Eagle Communications HD Digital Set-Top Box Important Safety Instructions WARNING TO REDUCE THE RISK OF FIRE OR ELECTRIC SHOCK, DO NOT EXPOSE THIS PRODUCT TO RAIN

More information

DA IN 1-OUT LINE DRIVER WITH EQUALIZATION + AUDIO USER S GUIDE

DA IN 1-OUT LINE DRIVER WITH EQUALIZATION + AUDIO USER S GUIDE MANUAL PART NUMBER: 400-0430-001 1-IN 1-OUT LINE DRIVER WITH UALIZATION + AUDIO USER S GUIDE TABLE OF CONTENTS Page PRECAUTIONS / SAFETY WARNINGS... 2 GENERAL...2 GUIDELINES FOR RACK-MOUNTING...2 INSTALLATION...2

More information

Evolution Digital HD Set-Top Box Important Safety Instructions

Evolution Digital HD Set-Top Box Important Safety Instructions Evolution Digital HD Set-Top Box Important Safety Instructions 1. Read these instructions. 2. Keep these instructions. 3. Heed all warnings. 4. Follow all instructions. 5. Do not use this apparatus near

More information

LIGHT COPILOT II. elationlighting.com Internet:

LIGHT COPILOT II.   elationlighting.com Internet: LIGHT COPILOT II E-mail: info@ elationlighting.com Internet: http://www.elationlighting.com 1 Introduction Thank you for your purchase of the LIGHT COPILOT II. The LIGHT COPILOT II is an intelligent lighting

More information

HOME THEATER. HDMI Selector Switches. Vanco Part Numbers (5x1) (3x1) Technical Support

HOME THEATER. HDMI Selector Switches. Vanco Part Numbers (5x1) (3x1) Technical Support HOME THEATER HDMI Selector Switches Vanco Part Numbers 280710 (5x1) 280711 (3x1) Technical Support www.vanco1.com info@vanco1.com 800-626-6445 DEAR CUSTOMER Thank you for purchasing this product. For optimum

More information

Oscilloscopes for field applications

Oscilloscopes for field applications Oscilloscopes for field applications ScopeMeter Test Tools provide: From 20 to 200 MHz bandwidth and up to 2.5 GS/s real-time sampling Large, high-resolution screen Digital Persistence and fast display

More information

25 MHz Digital Storage Oscilloscope

25 MHz Digital Storage Oscilloscope 99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com Model: 2530B 25 MHz Digital Storage Oscilloscope USER MANUAL Safety Summary The

More information

User Manual. TDS3SDI 601 Digital Video Application Module

User Manual. TDS3SDI 601 Digital Video Application Module User Manual TDS3SDI 601 Digital Video Application Module 071-0787-00 071078700 Copyright Tektronix, Inc. All rights reserved. Tektronix products are covered by U.S. and foreign patents, issued and pending.

More information

PicoScope PC Oscilloscopes. User's Guide. ps2203.en r4 Copyright Pico Technology Limited. All rights reserved.

PicoScope PC Oscilloscopes. User's Guide. ps2203.en r4 Copyright Pico Technology Limited. All rights reserved. PicoScope 2203 PC Oscilloscopes User's Guide PicoScope 2203 User's Guide I Contents 1 Welcome...1 2 Introduction...2 1 Using this guide 2 Safety symbols 3 Safety warning 4 FCC notice 5 CE notice...2...2...3...4...4...5

More information

MINI PC SCOPE PCSU01. User manual. test leads software download USB cable design enclosure

MINI PC SCOPE PCSU01. User manual. test leads software download USB cable design enclosure MINI PC SCOPE PCSU01 User manual Features test leads software download USB cable design enclosure Specifications oscilloscope: o bandwidth: DC to 200 khz ± 3 db o input impedance: 100 ko / 20 pf o maximum

More information

2.4 GHz WIRELESS VIDEO SENDER SYSTEM MODEL: VS6234

2.4 GHz WIRELESS VIDEO SENDER SYSTEM MODEL: VS6234 2.4 GHz WIRELESS VIDEO SENDER SYSTEM MODEL: VS6234 Please read this manual thoroughly before operating this system OPERATING INSTRUCTIONS 03/02 1 SAFETY INSTRUCTIONS CAUTION! RISK OF ELECTRIC SHOCK. DO

More information

HOME GUARD USER MANUAL

HOME GUARD USER MANUAL HOME GUARD USER MANUAL CONTENTS 1. SAFETY PRECAUTIONS...2 2. INTRODUCTION...3 3. FEATURES...4 4. ACCESSORIES...5 5. INSTALLATION...6 6. NAME and FUNCTION of EACH PART...7 6.1 Front Pannel...7 6.2 Monitoring

More information

Reference. TDS7000 Series Digital Phosphor Oscilloscopes

Reference. TDS7000 Series Digital Phosphor Oscilloscopes Reference TDS7000 Series Digital Phosphor Oscilloscopes 07-070-00 0707000 To Use the Front Panel You can use the dedicated, front-panel knobs and buttons to do the most common operations. Turn INTENSITY

More information

2510 Series Handheld Digital Storage Oscilloscopes

2510 Series Handheld Digital Storage Oscilloscopes Model: 2511, 2512, 2515, 2516 2510 Series Handheld Digital Storage Oscilloscopes USER MANUAL Safety Summary The following safety precautions apply to both operating and maintenance personnel and must be

More information

Oscilloscope Guide Tektronix TDS3034B & TDS3052B

Oscilloscope Guide Tektronix TDS3034B & TDS3052B Tektronix TDS3034B & TDS3052B Version 2008-Jan-1 Dept. of Electrical & Computer Engineering Portland State University Copyright 2008 Portland State University 1 Basic Information This guide provides basic

More information

User Guide. HDMI Fiber Optic Extender. DVI-7350a

User Guide. HDMI Fiber Optic Extender. DVI-7350a User Guide HDMI Fiber Optic Extender DVI-7350a Table of Contents Section Page Product Safety.................................... 1 1.0 Introduction...2 2.0 Specifications...3 3.0 Package Contents...3 4.0

More information

User Manual PS-684. HDBaseT Extender Kit 70m. All Rights Reserved. Version: UHBT70P_2016V1.2

User Manual PS-684. HDBaseT Extender Kit 70m. All Rights Reserved. Version: UHBT70P_2016V1.2 User Manual PS-684 All Rights Reserved Version: UHBT70P_2016V1.2 Preface Read this user manual carefully before using this product. Pictures shown in this manual is for reference only, different model

More information

INSTRUCTION GUIDE. for the. Keysight (formally Agilent; Hewlett-Packard) Model #33509B. and. Tektronix. Fall 2017

INSTRUCTION GUIDE. for the. Keysight (formally Agilent; Hewlett-Packard) Model #33509B. and. Tektronix. Fall 2017 v3 INSTRUCTION GUIDE for the Keysight (formally Agilent; Hewlett-Packard) WAVEFORM GENERATOR Model #33509B 33500 Series and Tektronix OSCILLOSCOPE Model #MSO2004B Fall 2017 Pg. 2 Scope-Gen Handout_part

More information

PicoScope 3000 Series Automotive User guide

PicoScope 3000 Series Automotive User guide PicoScope 3000 Series Automotive User guide PS3000A044 v1.0 I PicoScope 3000 Series Automotive PC Oscilloscopes Table of Contents 1 Introduction...2...2 1 Overview...2 2 Minimum PC requirements...2 3 Installation

More information

Colour Explosion Proof Video Camera USER MANUAL VID-C

Colour Explosion Proof Video Camera USER MANUAL VID-C Colour Explosion Proof Video Camera USER MANUAL VID-C Part Number: MAN-0036-00 Rev 4 Copyright 2002 Net Safety Monitoring Inc. Printed in Canada This manual is provided for informational purposes only.

More information

PRODUCT MANUAL : 1x8 Splitter/Extender : Receiver (Sold Separately)

PRODUCT MANUAL : 1x8 Splitter/Extender : Receiver (Sold Separately) PRODUCT MANUAL Product Name: S-VGA Splitter/Extenders over Category 5e/Cat6 Cable Part Numbers: 90-12116: 1x2 Splitter/Extender 90-12112: 1x8 Splitter/Extender 90-12113: Receiver (Sold Separately) WARNING:

More information

SNG-2150C User s Guide

SNG-2150C User s Guide SNG-2150C User s Guide Avcom of Virginia SNG-2150C User s Guide 7730 Whitepine Road Revision 001 Richmond, VA 23237 USA GENERAL SAFETY If one or more components of your earth station are connected to 120

More information

Expect to Make Waves.

Expect to Make Waves. Expect to Make Waves. The New Oscilloscope Large 10.4" LCD touch screen Long capture time Extensive communication capabilities www.lecroy.com The New Oscillos From its large 10.4" LCD touch screen to its

More information

LCD VALUE SERIES (32 inches)

LCD VALUE SERIES (32 inches) LCD VALUE SERIES (32 inches) http://www.orionimages.com All contents of this document may change without prior notice, and actual product appearance may differ from that depicted herein 1. SAFETY INSTRUCTION

More information

USB-TG124A Tracking Generator User Manual

USB-TG124A Tracking Generator User Manual USB-TG124A Tracking Generator User Manual Signal Hound USB-TG124A User Manual 2017, Signal Hound, Inc. 35707 NE 86th Ave La Center, WA 98629 USA Phone 360.263.5006 Fax 360.263.5007 This information is

More information

PicoScope 4000 Automotive PC Oscilloscopes

PicoScope 4000 Automotive PC Oscilloscopes PicoScope 4000 Automotive PC Oscilloscopes User's Manual ps4000a.en-1 Copyright 2008 Pico Technology Ltd. All rights reserved. Contents I Contents 1 Introduction...1 1 Overview...1...1 2 Minimum PC requirements...2

More information

S op o e p C on o t n rol o s L arni n n i g n g O bj b e j ctiv i e v s

S op o e p C on o t n rol o s L arni n n i g n g O bj b e j ctiv i e v s ET 150 Scope Controls Learning Objectives In this lesson you will: learn the location and function of oscilloscope controls. see block diagrams of analog and digital oscilloscopes. see how different input

More information

Kramer Electronics, Ltd. USER MANUAL. Model: FC Standards Converter / TBC

Kramer Electronics, Ltd. USER MANUAL. Model: FC Standards Converter / TBC Kramer Electronics, Ltd. USER MANUAL Model: FC-4000 Standards Converter / TBC Contents Contents 1 Introduction 1 2 Getting Started 1 3 Overview 1 4 Your Standards Converter / TBC 2 4.1 Connecting the

More information

Winmate Communication INC.

Winmate Communication INC. 20.1 Military Grade Display Model: R20L100-RKA2ML User s Manual Winmate Communication INC. May, 2011 1 IMPORTANT SAFETY INSTRUCTIONS Please read these instructions carefully before using the product and

More information

PicoScope 6407 Digitizer

PicoScope 6407 Digitizer YE AR HIGH PERFORMANCE USB DIGITIZER Programmable and Powerful 1 GHz bandwidth 1 GS buffer size 5 GS/s real-time sampling Advanced digital triggers Built-in function generator USB-connected Signals Analysis

More information

ScopeMeter 190 Series II

ScopeMeter 190 Series II ScopeMeter 190 Series II Fluke 190-062, -102, -104, -202, -204, -502 Users Manual May 2011 2011 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All product

More information

Choosing an Oscilloscope

Choosing an Oscilloscope Choosing an Oscilloscope By Alan Lowne CEO Saelig Company (www.saelig.com) Post comments on this article at www.nutsvolts.com/ magazine/article/october2016_choosing-oscilloscopes. All sorts of questions

More information

User s Manual. TDO1000/TDO2000 Series Oscilloscopes

User s Manual. TDO1000/TDO2000 Series Oscilloscopes User s Manual TDO1000/TDO2000 Series Oscilloscopes Manual Print History The manual print history shown below lists all the printing dates and editions. The printing date changes when a new edition is released.

More information

User Guide. HDMI Active Cable Extender. DVI-7370c

User Guide. HDMI Active Cable Extender. DVI-7370c User Guide HDMI Active Cable Extender DVI-7370c TABLE OF CONTENTS SECTION PAGE PRODUCT SAFETY...1 PRODUCT LIABILITY STATEMENT........................ 1 1.0 INTRODUCTION...2 2.0 SPECIFICATIONS...3 3.0 PACKAGE

More information

USERS MANUAL GR ONE 300 GR ONE 700 GR ONE 1400

USERS MANUAL GR ONE 300 GR ONE 700 GR ONE 1400 USERS MANUAL GR ONE 300 GR ONE 700 GR ONE 1400 PURE SOUND USER MANUAL PURPOSE This user manual is dedicated to the use of the bass amplifiers GR-ONE 700, GR-ONE700x2 and GR-ONE 300. The manual includes

More information

Cryoelectronics. MS-FLL User s Manual. Mr. SQUID Flux-Locked Loop. STAR Cryoelectronics 25 Bisbee Court, Suite A Santa Fe, NM U. S. A.

Cryoelectronics. MS-FLL User s Manual. Mr. SQUID Flux-Locked Loop. STAR Cryoelectronics 25 Bisbee Court, Suite A Santa Fe, NM U. S. A. Cryoelectronics MS-FLL User s Manual Mr. SQUID Flux-Locked Loop STAR Cryoelectronics 25 Bisbee Court, Suite A Santa Fe, NM 87508 U. S. A. STAR Cryoelectronics, LLC ii Table of Contents Revision Record...

More information

imso-104 Manual Revised August 5, 2011

imso-104 Manual Revised August 5, 2011 imso-104 Manual Revised August 5, 2011 Section 1 Getting Started SAFETY 1.10 Quickstart Guide 1.20 SAFETY 1.30 Compatibility 1.31 Hardware 1.32 Software Section 2 How it works 2.10 Menus 2.20 Analog Channel

More information

32 Channel CPCI Board User Manual

32 Channel CPCI Board User Manual 0 Sections Page 1.0 Introduction 1 2.0 Unpacking and Inspection 1 3.0 Hardware Configuration 1 4.0 Board Installation 5 5.0 I/O Connections and the Front Panel 5 5.1 Front Panel Layout 5 5.2 Input and

More information

PRO-HDMI2HD. HDMI to SDI/3G-HD-SD Converter. User Manual. Made in Taiwan

PRO-HDMI2HD. HDMI to SDI/3G-HD-SD Converter. User Manual. Made in Taiwan PRO-HDMI2HD HDMI to SDI/3G-HD-SD Converter User Manual Made in Taiwan rev.1008 103 Quality Circle, Suite 210 Huntsville, Alabama 35806 Tel: (256) 726-9222 Fax: (256) 726-9268 Email: service@pesa.com Safety

More information

HDBaseT RECEIVER B-520-RX-230-IR INSTALLATION MANUAL

HDBaseT RECEIVER B-520-RX-230-IR INSTALLATION MANUAL HDBaseT RECEIVER B-520-RX-230-IR INSTALLATION MANUAL IMPORTANT SAFETY INSTRUCTIONS To reduce the risk of fire or electric shock, read and follow all instructions and warnings in this manual. Keep this

More information

INSTALLATION MANUAL FT-FOTR-8VD-ST-S. 8-Channel Digital Duplex Baseband Video Transmitter and Receiver With Reverse Data Transmission for PTZ Cameras

INSTALLATION MANUAL FT-FOTR-8VD-ST-S. 8-Channel Digital Duplex Baseband Video Transmitter and Receiver With Reverse Data Transmission for PTZ Cameras INSTALLATION MANUAL FT-FOTR-8VD-ST-S 8-Channel Digital Duplex Baseband Transmitter and Receiver With Reverse Transmission for PTZ Cameras v1.0 4/5/11 1 PACKAGE CONTENTS This package contains: One each

More information

Kramer Electronics, Ltd. USER MANUAL. Models: TR-1YC, s-video Isolation Transformer TR-2YC, s-video Dual Isolation Transformers

Kramer Electronics, Ltd. USER MANUAL. Models: TR-1YC, s-video Isolation Transformer TR-2YC, s-video Dual Isolation Transformers Kramer Electronics, Ltd. USER MANUAL Models: TR-1YC, s-video Isolation Transformer TR-2YC, s-video Dual Isolation Transformers Contents Contents 1 Introduction 1 2 Getting Started 1 2.1 Quick Start 1 3

More information

User Manual. TDS3VID Extended Video Application Module

User Manual. TDS3VID Extended Video Application Module User Manual TDS3VID Extended Video Application Module 071-0328-02 071032802 Copyright Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers,

More information

OPERATING INSTRUCTIONS TOM-0431IP

OPERATING INSTRUCTIONS TOM-0431IP OPERATING INSTRUCTIONS TOM-0431IP Table of Contents FCC Information -------------------------------------------------------------------- 2 Safety and Environmental Precautions ------------------------------------------------

More information

English. User Manual sub8 Subwoofer SUBWOOFER. Supporting your digital lifestyle

English. User Manual sub8 Subwoofer SUBWOOFER. Supporting your digital lifestyle English User Manual sub8 Subwoofer U SUBWOOFER Supporting your digital lifestyle Table of Contents Important Safety Precautions........ 2 Introduction / What s in the Box?...... 3 Front & Rear Panels............

More information

2x1 HDMI SWITCH. with Multiview and PIP Vanco Part Number: EVSW21MV

2x1 HDMI SWITCH. with Multiview and PIP Vanco Part Number: EVSW21MV 2x1 HDMI SWITCH with Multiview and PIP Vanco Part Number: EVSW21MV 2x1 HDMI Switch with Multiview and Picture-in-Picture www.vanco1.com 800.626.6445 DEAR CUSTOMER Thank you for purchasing this product.

More information

USER MANUAL. 27 Full HD Widescreen LED Monitor L27ADS

USER MANUAL. 27 Full HD Widescreen LED Monitor L27ADS USER MANUAL 27 Full HD Widescreen LED Monitor L27ADS TABLE OF CONTENTS 1 Getting Started 2 Control Panel/ Back Panel 3 On Screen Display 4 Technical Specs 5 Care & Maintenance 6 Troubleshooting 7 Safety

More information

WAVEJET 300 SERIES OSCILLOSCOPES. New Cover to Come. Unmatched Performance, Portability, and Value

WAVEJET 300 SERIES OSCILLOSCOPES. New Cover to Come. Unmatched Performance, Portability, and Value WAVEJET 300 SERIES OSCILLOSCOPES New Cover to Come Unmatched Performance, Portability, and Value ALL THE TOOLS YOU NEED Automatic Measurements Save time making measurements on your signals by using the

More information

SATRI AMPLIFIER AMP-51R. Owner s Manual

SATRI AMPLIFIER AMP-51R. Owner s Manual SATRI AMPLIFIER AMP-51R Owner s Manual contents SAFETY INSTRUCTIONS 4 INTRODUCTION 6 OVERVIEW (FRONT PANEL) 8 OVERVIEW (REAR PANEL) 9 OVERVIEW (REMOTE CONTROL) 1 1 OPERATION 12 TROUBLESHOOTING 13 SPECIFICATION

More information

HDBaseT RECEIVER B-540-RX-330-IR INSTALLATION MANUAL

HDBaseT RECEIVER B-540-RX-330-IR INSTALLATION MANUAL HDBaseT RECEIVER B-540-RX-330-IR INSTALLATION MANUAL IMPORTANT SAFETY INSTRUCTIONS To reduce the risk of fire or electric shock, read and follow all instructions and warnings in this manual. Keep this

More information

Tube Roller Shakers. User Guide. Version 1.2

Tube Roller Shakers. User Guide. Version 1.2 Tube Roller Shakers User Guide Version 1.2 Control panel Rollers Side retaining panels Analog models LED display Drip tray (not visible) Digital models Power On/Off and control dial Roller retaining panel

More information

Kramer Electronics, Ltd. USER MANUAL. Models: 6808, SDI-AES De-Embedder 6809, SDI-AES Embedder

Kramer Electronics, Ltd. USER MANUAL. Models: 6808, SDI-AES De-Embedder 6809, SDI-AES Embedder Kramer Electronics, Ltd. USER MANUAL Models: 6808, SDI-AES De-Embedder 6809, SDI-AES Embedder Contents Contents 1 Introduction 1 2 Getting Started 1 3 Overview 1 3.1 About the 6808 SDI-AES De-Embedder

More information

USER MANUAL. 27 Full HD Widescreen LED Monitor L270E

USER MANUAL. 27 Full HD Widescreen LED Monitor L270E USER MANUAL 27 Full HD Widescreen LED Monitor L270E TABLE OF CONTENTS 1 Getting Started 2 Control Panel/ Back Panel 3 On Screen Display 4 Technical Specs 5 Care & Maintenance 6 Troubleshooting 7 Safety

More information

HDBaseT RECEIVER B-520-RX-330-IR INSTALLATION MANUAL

HDBaseT RECEIVER B-520-RX-330-IR INSTALLATION MANUAL HDBaseT RECEIVER B-520-RX-330- INSTALLATION MANUAL IMPORTANT SAFETY INSTRUCTIONS To reduce the risk of fire or electric shock, read and follow all instructions and warnings in this manual. Keep this manual

More information

PX1710M LCD Monitor USER S GUIDE

PX1710M LCD Monitor USER S GUIDE PX1710M LCD Monitor USER S GUIDE www.planar.com The information contained in this document is subject to change without notice. This document contains proprietary information that is protected by copyright.

More information