Calibrating Measuring Microphones and Sound Sources for Acoustic Measurements with Audio Analyzer R&S UPV
|
|
- Beatrix Little
- 1 years ago
- Views:
Transcription
1 Product: R&S UPV Calibrating Measuring Microphones and Sound Sources for Acoustic Measurements with Audio Analyzer R&S UPV Application Note 1GA47_0E This application note explains how to use acoustic calibrations in order to be able to enter or read sound pressure level directly in the Audio Analyzer R&S UPV. An application program is described which performs the necessary measurements for the calibration of microphones and loudspeakers, manages the calibration values and enters them into any appropriate user defined setup. Subject to change GA47_0E
2 Contents 1 Overview Introduction Use of Reference Values in the UPV Installation of the Application Program Operation of the Application Program... 7 Microphone Calibration... 7 Loudspeaker Calibration... 8 Using Calibration Values Hardware and Software Requirements Appendix: Using the Quick Launch Buttons Ordering information GA47-0E 2 Rohde & Schwarz
3 1 Overview This application note explains how to use acoustic calibrations in order to be able to enter or read sound pressure level directly in the Audio Analyzer R&S UPV. An application program is described which performs the necessary measurements for the calibration of microphones and loudspeakers, manages the calibration values and enters them into any appropriate user defined setup. The Audio Analyzer R&S UPV is abbreviated to UPV for the remainder of this Application Note. 2 Introduction Acoustic measurements involve the measurement of sound pressure level or the generation of a sound field, or both. Typical examples are noise measurements, loudspeaker measurements, microphone measurements and measurements on systems like hearing aids and mobile phones. A sound level meter, for example, is supposed to directly display the sound pressure level in dbspl (db referred to 20 µpa sound pressure). If an audio analyzer like UPV, which measures voltage, is used for this purpose, the sensitivity of the microphone s ~ VM = p M ~ has to be determined, wherein V ~ M is the rms value of the microphone output voltage and ~ p is the rms value of the sound pressure which produced the output voltage. The value of the sound pressure is obtained by dividing the measured voltage by this sensitivity value. The measurement with which the value of the microphone sensitivity is obtained is commonly called microphone calibration. The sound pressure level for the microphone calibration is generated with a so-called acoustic calibrator which generates a sinewave signal with defined frequency (commonly 1 khz) and defined sound pressure (usually 1 Pa or 10 Pa). As measurement microphones are small and have a well-defined mechanical structure, the sensitivity is frequency-independent within a certain frequency range. Therefore calibration at a single frequency is sufficient. Certain measurements require the generation of a defined sound pressure at a certain point. In order to be able to set a desired sound pressure level in the generator of the UPV, the sensitivity of the loudspeaker ~ p = L s L ~ VG has to be determined, wherein p~ L is the sound pressure generated by the loudspeaker at the pre-defined measurement point, and generator output voltage which produced the sound pressure. V ~ G is the 1GA47-0E 3 Rohde & Schwarz
4 As a loudspeaker has a far more complex mechanical structure than a measuring microphone, and radiation effects additionally influence the loudspeaker sensitivity depending on the frequency and on the location of the measurement point, the sensitivity of a loudspeaker usually is frequency dependent. For this reason the calibration of a loudspeaker consists of two steps: 1. Measurement of the absolute sensitivity at one frequency 2. Measurement of the frequency response relative to this frequency. In order to generate a defined sound pressure at the measurement point, the generator output voltage has to be set to the desired sound pressure divided by the loudspeaker sensitivity, and to be corrected by the inverse frequency response (equalization). It has to be observed that the loudspeaker calibration is only valid for the point of the calibration, due to the propagation properties of the sound wave. Note that an equalization in the sense of a frequency dependent amplitude correction is only possible with signals which are defined in the frequency domain, like (swept) sinewave and multisine signals. For the equalization of a complex signal defined in the time domain like speech, a filter is required. 3 Use of Reference Values in the UPV For numerical function results, the microphone sensitivity can be entered in the Function Config Panel as Reference Value: If the value entered as reference value is the microphone output produced by 1 Pa sound pressure, the resulting display value is dbpa (db referred to 1 Pa). If it is desired to display dbspl (db referenced to 20 µpa), the microphone output voltage for 20 µpa has to be entered as reference value. It is obtained by subtracting 94 db from the voltage level measured at 1 Pa, or by dividing this voltage by GA47-0E 4 Rohde & Schwarz
5 For a level display, the unit has to be set to dbr. For display of sound pressure the unit has to be set to V/Vr. The labeling in the display can be adapted to the desired unit if the checkbox Unit Auto is unchecked (see example above). For sweeps, the reference value can be entered in the Sweep Graph Config Panel. Alternatively, Reference in the sweep graph config panel can be set to Meas Panel. This copies the reference value entered in the Function Config Panel to the sweep. The same applies respectively to FFT spectra (FFT Graph Config Panel) and 1/n octave spectra (Bar Graph Config Panel). For the input monitor and the level monitor the reference can be set according to the numeric function result in the Input Config Panel or Level Mon Config Panel, respectively. The only difference is that there is one common reference value for both channels. On the generator side the sensitivity is entered as reference value in the Generator Config Panel. The loudspeaker frequency response can be stored from the Sweep Graph Config Panel as equalization list: The checkbox Invert has to be ticked in order to invert the measured frequency response for the equalization. Norm Freq has to be set to the frequency at which the absolute sensitivity had been measured. The stored equalization list can then be opened as equalization file in frequency-defined generator signals: 1GA47-0E 5 Rohde & Schwarz
6 For each frequency the amplitude is individually corrected by the factor given in the equalization file, whereby frequencies are interpolated between those present in the file. 4 Installation of the Application Program The use of the application program requires option UPV-K1 to be installed. The application program can be installed as executable program or opened as a project in the IDE of Visual Basic.NET for modification and new compilation. For just using the executable, copy the installer files CalibrationTool.msi, Config.ini, dotnetfx.exe, Settings.ini and Setup.exe to the hard disk of your UPV and start Setup.exe by double-clicking. If the DotNet runtime framework is already present on the UPV (e.g. when the VB.NET IDE is installed or other DotNet application programs had been previously installed), it is sufficient to copy and start CalibrationTool.msi. The program is stored in the program files folder under Rohde&Schwarz\CalibrationTool\CalibrationTool.exe. The installer also creates a shortcut on the desktop of the UPV. Furthermore, the installer copies the necessary setup files to the program folder, and the program installs a folder D:\UPV\Calibrations automatically at its first startup. 1GA47-0E 6 Rohde & Schwarz
7 If the program is compiled and started from the IDE, the folder "Calibrations" in D:\UPV has to be created in advance, and the two setups MicCal.set and SourceCal.set have to be copied to it. For editing and compiling the source code, unzip the project folder and copy it to the hard disk of the UPV (e.g. in D:\Visual Studio Projects ). Click Open Project on the start page of the Visual Basic IDE, browse to the folder and open AcousticCalibration1.sln. 5 Operation of the Application Program The use of the application program requires the respective measurement setup to be prepared for the entry and application of the reference values. For details please see section 3 above. The application note comprises also an example setup SweepExample.set for calibration of a loudspeaker and two different microphones. In particular, channels for which a microphone calibration is to be entered by the tool, have to be active in the setup. The references of graphs should be set to Meas Panel. In order to enable proper loudspeaker calibration, Equalizer should be checked in the Generator Function Panel. Furthermore dbr amplitude values in the Generator Function Panel should be set in the setup sufficiently low not to overrange the Generator. Any illegal operation condition caused by the entry of the calibration values produces an error window and closes the application program. Starting AcousticCalibration1.exe opens the following entry window on the UPV: The window is divided into three panels: One for the loudspeaker calibration, one for the microphone calibration and one for updating stored calibration values to the UPV setup. Microphone Calibration The microphone calibration is available for each of the two input channels. Mic Calibration starts the calibration routine. Once the calibration routine has successfully finished, a file can be specified to store the results. This file name is simultaneously entered into the Calibration File text box which specifies the file from which the calibration value to be entered into the UPV setup is taken. Calibrate and Update automatically enters the value in the 1GA47-0E 7 Rohde & Schwarz
8 Function Config Panel for the respective channel. To select a calibration value from a file without perfoming the calibration routine, use Open File. To calibrate a microphone, connect it to the desired analyzer input. Insert the microphone into an acoustic calibrator and switch on the calibrator. After clicking the Mic Calibration button of the respective channel a window opens which queries the SPL produced by the calibrator. If Selective is activated, the calibration measurement is performed with a fast 3 rd octave band pass filter. Therefore in this case also the frequency of the signal produced by the calibrator has to be entered. After that a series of voltage measurements is started. If the single results are sufficiently consistent, the sensitivity of the microphone is displayed, and a file selector box is opened for entering the file name to store the value. Loudspeaker Calibration Loudspeaker calibration requires a valid microphone calibration to exist for the analyzer channel used for the speaker calibration measurement. The used channel has to be selected in Analyzer Channel. Lower Limit, Upper Limit and Point Count specify the parameters for the frequency response sweep done during the calibration measurement. Selective toggles between wideband and selective loudspeaker measurement. To calibrate a loudspeaker, connect it to one of the UPV outputs. If required, a power amplifier can be inserted. Connect a microphone to one of the analyzer inputs and calibrate the microphone or load a valid calibration for this microphone from a file. Select the correct analyzer channel for the loudspeaker calibration. Place the microphone at the point in front of the loudspeaker for which the calibration should be valid. After clicking Speaker Calibration, the SPL for the calibration has to be entered. Due to nonlinearities of loudspeakers there may be deviations in sensitivity and frequency response between calibrations executed at different levels. It is also possible to perform calibrations for a set of different levels and store them under different file names. After the level has been confirmed, the calibration starts with first determining the absolute sensitivity at 1 khz, then measuring the relative frequency response. In a second run the obtained equalization is checked and corrected before a third sweep is done to verify the result of the calibration. After completion of the measurement, a file name has to be specified for storing the calibration values. Subsequently the UPV is returned to local control and the calibration tool window is minimized in order to show the sweeps: 1GA47-0E 8 Rohde & Schwarz
9 After the Source Sensitivity window has been closed by clicking ok, the calibration tool window re-appears and the previous setup of the UPV is restored. It is essential that the test configuration (gain of the power amplifier, positioning of loudspeaker and device to be measured) is not altered between the calibration and the measurement in which the calibration is to be used. Using Calibration Values Update Calibrations enters the calibration values from the files specified in the three text boxes as reference values for the ticked generator and analyzer channels. An update to input and monitor display is only possible in single channel mode. It can be chosen whether the dbr values in the setup and displays should be interpreted as dbspl (relative to 20 upa) or as dbpa (relative to 1 Pa). Load Setup and Update Calibrations allows to load a UPV setup and automatically enter the activated calibration values as reference values into the loaded setup. The application program returns the UPV firmware to local control after termination of each action. Therefore it can be kept open during the use of the UPV. However, it is not possible to run another remote control application while the calibration tool window is open. The window is minimized after loading and updating the setup and can be retrieved from the Windows task bar. 6 Hardware and Software Requirements The Audio Analyzer UPV needs firmware version or higher installed. For the use of the application program, option UPV-K1 (Sequence Control) is required. 7 Appendix: Using the Quick Launch Buttons The UPV offers 8 quick launch buttons which can be utilized either for loading pre-defined setups or for starting pre-defined macros. The assignment of the buttons can be defined in Utilities Quick Launch Config : Fig. 1 Starting the Quick Launch Config Panel This menu opens the following configuration panel: 1GA47-0E 9 Rohde & Schwarz
10 Fig. 2 Quick Launch Config Panel Use the buttons on the right side to browse to the setup file or exe file which should be called by the respective button. The buttons on the left side can be utilized to specify icons on the buttons. Text entered in the text box on the left side will be displayed as tool tip text for the respective button when the mouse cursor is staying on the area of the button. 8 Ordering information UPV Audio Analyzer DC 250 khz UPV-K1 Sequence Controller For additional information about audio analyzers and other measurement equipment, see the Rohde & Schwarz website ROHDE & SCHWARZ GmbH & Co. KG. Mühldorfstraße 15. D München. Postfach D München. Tel (089) Fax (089) Internet: This application note and the supplied programs may only be used subject to the conditions of use set forth in the download area of the Rohde & Schwarz website. 1GA47-0E 10 Rohde & Schwarz
Testing Speech Quality of Mobile Phones in a Live Network Application Note
Testing Speech Quality of Mobile Phones in a Live Network Application Note Products: R&S UPV R&S UPV-K63 R&S UPV-K1 This application note provides an application program and instructions for performing
Generating the Noise Field for Ambient Noise Rejection Tests Application Note
Generating the Noise Field for Ambient Noise Rejection Tests Application Note Products: R&S UPV R&S UPV-K9 R&S UPV-K91 This document describes how to generate the noise field for ambient noise rejection
Basic RF Amplifier Measurements using the R&S ZNB Vector Network Analyzer and SMARTerCal. Application Note
Basic RF Amplifier Measurements using a R&S ZNB Analyzer and SMARTerCal Mark Bailey 2013-03-05, 1ES, Version 1.0 Basic RF Amplifier Measurements using the R&S ZNB Vector Network Analyzer and SMARTerCal.
Getting Started with the LabVIEW Sound and Vibration Toolkit
1 Getting Started with the LabVIEW Sound and Vibration Toolkit This tutorial is designed to introduce you to some of the sound and vibration analysis capabilities in the industry-leading software tool
Using the BHM binaural head microphone
11/17 Using the binaural head microphone Introduction 1 Recording with a binaural head microphone 2 Equalization of a recording 2 Individual equalization curves 5 Using the equalization curves 5 Post-processing
Using R&S NRP Series Power Sensors with Android TM Handheld Devices. Application Note. Products: R&S NRP Series. R&S NRP-Zxx Series
Using R&S NRP Series Power Sensors with Android TM Handheld Devices Products: R&S NRP Series R&S NRP-Zxx Series This application note describes how to connect and use the highly popular R&S NRP family
EMI Automotive Band Evaluation
EMI Automotive Band Evaluation Hartmut A. Maier 07.2014-01.00 EMI Automotive Band Evaluation Application Note Products: R&S EMC32 R&S EMC32-K51 EMI Automotive Band Evaluation extension is an on-board EMI
Features/Specifications
Introduction Thank you for purchasing the DD Audio DSI-1(Digital Signal Integrator). The DSI-1 is a feature rich audio signal processor that will allow you to precisely tune the acoustics of your car audio
Easy Analysis and Precise Optimization of ATSC or ATSC Mobile DTV Single-Frequency Networks Application Note
Easy Analysis and Precise Optimization of ATSC or ATSC Mobile DTV Single-Frequency Networks Application Note Products: R&S ETL Keeping an ATSC (or ATSC Mobile DTV) single-frequency network (SFN) running
Pre-5G-NR Signal Generation and Analysis Application Note
Pre-5G-NR Signal Generation and Analysis Application Note Products: R&S SMW200A R&S VSE R&S SMW-K114 R&S VSE-K96 R&S FSW R&S FSVA R&S FPS This application note shows how to use Rohde & Schwarz signal generators
Out of Band Spurious Measurement for Bluetooth Modules
Products: Signal Analyser FSIQ26/FSP13/FSU8/FSQ26 Out of Band Spurious Measurement for Bluetooth Modules This application notes describes the out of band Spurious emission measurement for Bluetooth modules
SigPlay User s Guide
SigPlay User s Guide . . SigPlay32 User's Guide? Version 3.4 Copyright? 2001 TDT. All rights reserved. No part of this manual may be reproduced or transmitted in any form or by any means, electronic or
Correlated Receiver Diversity Simulations with R&S SFU
Application Note Marius Schipper 10.2012-7BM76_2E Correlated Receiver Diversity Simulations with R&S SFU Application Note Products: R&S SFU R&S SFE R&S SFE100 R&S SFC R&S SMU200A Receiver diversity improves
Concise NFC Demo Guide using R&S Test Equipment Application Note
Concise NFC Demo Guide using R&S Test Equipment Application Note Products: R&S SMBV100A R&S SMBV-K89 R&S FS-K112PC R&S RTO R&S RTO-K11 R&S CSNFC-B8 R&S FSL R&S FSV R&S FSW R&S ZVL This concise NFC Demo
NanoGiant Oscilloscope/Function-Generator Program. Getting Started
Getting Started Page 1 of 17 NanoGiant Oscilloscope/Function-Generator Program Getting Started This NanoGiant Oscilloscope program gives you a small impression of the capabilities of the NanoGiant multi-purpose
Transmitter Measurements on Bluetooth TM Modules with FSP
Products: FSP Transmitter Measurements on Bluetooth TM Modules with FSP This application note shows how to perform measurements on transmitter modules largely in conformance with the Bluetooth test specifications
System Requirements SA0314 Spectrum analyzer:
System Requirements SA0314 Spectrum analyzer: System requirements Windows XP, 7, Vista or 8: 1 GHz or faster 32-bit or 64-bit processor 1 GB RAM 10 MB hard disk space \ 1. Getting Started Insert DVD into
T80-K1 V2.22 Release Note
T80-K1 V2.22 Release Note Products: R&S T80-K1 This document gives an overview of the additional features and improvements that are implemented with version 2.22 Release Note -
Binaural Measurement, Analysis and Playback
11/17 Introduction 1 Locating sound sources 1 Direction-dependent and direction-independent changes of the sound field 2 Recordings with an artificial head measurement system 3 Equalization of an artificial
Transmitter Measurements on Bluetooth TM Modules
Products: FSE, FSIQ Transmitter Measurements on Bluetooth TM Modules This application note describes how to perform measurements on transmitter modules largely in conformance with the Bluetooth test specifications
Cable Calibration Function for the 2400B/C and 2500A/B Series Microwave Signal Generators. Technical Brief
Cable Calibration Function for the 2400B/C and 2500A/B Series Microwave Signal Generators Technical Brief Quickly and easily apply a level correction table to compensate for external losses or power variations
NPR - Noise Power Ratio Signal Generation and Measurement
Products: AMIQ, SMIQ, SMR, FSEx, FSIQ, FSP, FSU NPR - Noise Power Ratio Signal Generation and Measurement Noise Power Ratio (NPR) is an add-on tool for WinIQSim to generate noise power ratio stimulus signals
USB Mini Spectrum Analyzer User s Guide TSA5G35
USB Mini Spectrum Analyzer User s Guide TSA5G35 Triarchy Technologies, Corp. Page 1 of 21 USB Mini Spectrum Analyzer User s Guide Copyright Notice Copyright 2011 Triarchy Technologies, Corp. All rights
SC24 Magnetic Field Cancelling System
SPICER CONSULTING SYSTEM SC24 SC24 Magnetic Field Cancelling System Makes the ambient magnetic field OK for the electron microscope Adapts to field changes within 100 µs Touch screen intelligent user interface
Diamond Cut Productions / Application Notes AN-2
Diamond Cut Productions / Application Notes AN-2 Using DC5 or Live5 Forensics to Measure Sound Card Performance without External Test Equipment Diamond Cuts DC5 and Live5 Forensics offers a broad suite
USB Mini Spectrum Analyzer User Manual TSA Program for PC TSA4G1 TSA6G1 TSA8G1
USB Mini Spectrum Analyzer User Manual TSA Program for PC TSA4G1 TSA6G1 TSA8G1 Triarchy Technologies Corp. Page 1 of 17 USB Mini Spectrum Analyzer User Manual Copyright Notice Copyright 2013 Triarchy Technologies,
Manual for the sound card oscilloscope V1.41 C. Zeitnitz english translation by P. van Gemmeren, K. Grady and C. Zeitnitz
Manual for the sound card oscilloscope V1.41 C. Zeitnitz english translation by P. van Gemmeren, K. Grady and C. Zeitnitz C. Zeitnitz 12/2012 This Software and all previous versions are NO Freeware! The
EASERA SysTune. Developed by. AFMG Ahnert Feistel Media Group. The creators of EASE and EASERA.
EASERA SysTune - Developed by AFMG Ahnert Feistel Media Group The creators of EASE and EASERA www.afmg.eu Software Manual,, December 2007 Copyright 2006-2007 SDA Software Design Ahnert GmbH Contents -
Quick Start for TrueRTA (v3.5) on Windows XP (and earlier)
Skip directly to the section that covers your version of Windows (XP and earlier, Vista or Windows 7) Quick Start for TrueRTA (v3.5) on Windows XP (and earlier) Here are step-by-step instructions to get
AcoustiSoft RPlusD ver
AcoustiSoft RPlusD ver 1.2.03 Feb 20 2007 Doug Plumb doug@etfacoustic.com http://www.etfacoustic.com/rplusdsite/index.html Software Overview RPlusD is designed to provide all necessary function to both
Manual for TV software. TT-Viewer version Figure: TT-budget S2-3200
Manual for TV software TT-Viewer version 2.7.0 Figure: TT-budget S2-3200 Index Manual TT-Viewer 3 1. Starting TT-Viewer software 3 2. General settings 5 3. Assignment of hardware 6 3.1 Unicable 7 4. Renderer
R&S FSV-K40 Phase Noise Measurement Application Specifications
FSV-K40_dat-sw_en_5213-9705-22_cover.indd 1 Data Sheet 02.00 Test & Measurement R&S FSV-K40 Phase Noise Measurement Application Specifications 06.10.2014 14:51:49 CONTENTS Specifications... 3 Ordering
Speeding up Spectrum Analyzer Measurements Application Note
Speeding up Spectrum Analyzer Measurements Application Note Products: R&S FPS R&S FSW R&S FSV R&S SGT100A R&S SMW200A R&S SMBV100A Test time is a critical parameter when it comes to evaluating the cost
Grid Interactive Advanced Features Self-Supply Mode
Grid Interactive Advanced Features Self-Supply Mode CHILICON POWER MICROINVERTERS AND CP-100 CORTEX GATEWAY SYSTEM OVERVIEW Chilicon Power Advanced Inverter functions Power Factor Control: -0.8 to +0.8
NOTICE: This document is for use only at UNSW. No copies can be made of this document without the permission of the authors.
Brüel & Kjær Pulse Primer University of New South Wales School of Mechanical and Manufacturing Engineering September 2005 Prepared by Michael Skeen and Geoff Lucas NOTICE: This document is for use only
Tests on 3G-Base Stations to TS with FSIQ and SMIQ
Products: FSIQ, SMIQ Tests on 3G-Base Stations to TS 25.141 with FSIQ and SMIQ This application note describes how to measure the various WCDMA signals which are used for transmitter tests on FDD base
E X P E R I M E N T 1
E X P E R I M E N T 1 Getting to Know Data Studio Produced by the Physics Staff at Collin College Copyright Collin College Physics Department. All Rights Reserved. University Physics, Exp 1: Getting to
Five Reasons to Upgrade from Legacy VNAs to a R&S ZNB Vector Network Analyzer
Five Reasons to Upgrade from Legacy VNAs to a R&S ZNB Vector Network Analyzer Summary: Vector Network Analyzers have evolved from complicated specialist tools to easy-to-use, yet even more powerful RF
FOUR CHANNEL USB RECORDER PCRU01. User manual. Table of Contents
FOUR CHANNEL USB RECORDER PCRU01 User manual Table of Contents Features 2 Specifications 2 hardware 2 software: 2 system requirements 2 Software installation 2 SAFETY and WARNINGS 3 Warranty 3 Connections
LTE Bitstream Verification. Application Note. Products: R&S SMW200A R&S SMU200A R&S SMx-K55 R&S SMx-K81 R&S FS-K10xPC R&S FSW R&S FSQ R&S FSV R&S FPS
Application Note Bernhard Schulz, Fabian Liebl 01.2015-1MA161_1e LTE Bitstream Verification Application Note Products: R&S SMW200A R&S SMU200A R&S SMx-K55 R&S SMx-K81 R&S FS-K10xPC R&S FSW R&S FSQ R&S
Agilent MOI for HDMI 1.4b Cable Assembly Test Revision Jul 2012
Revision 1.11 19-Jul 2012 Agilent Method of Implementation (MOI) for HDMI 1.4b Cable Assembly Test Using Agilent E5071C ENA Network Analyzer Option TDR 1 Table of Contents 1. Modification Record... 4 2.
SC26 Magnetic Field Cancelling System
SPICER CONSULTING SYSTEM SC26 SC26 Magnetic Field Cancelling System Makes the ambient magnetic field OK for electron beam tools in 300 mm wafer fabs Real time, wideband cancelling from DC to > 9 khz fields
Measurement of Beamforming Antenna in Transmit Mode Application Note
Measurement of Beamforming Antenna in Transmit Mode Application Note Products: R&S ZVA R&S FSV R&S NRP R&S Quickstep Rapid and accurate verification of beamforming weights is required for example, for
Application Note #63 Field Analyzers in EMC Radiated Immunity Testing
Application Note #63 Field Analyzers in EMC Radiated Immunity Testing By Jason Galluppi, Supervisor Systems Control Software In radiated immunity testing, it is common practice to utilize a radio frequency
Quick Start Guide. Multidimensional Imaging
Quick Start Guide Multidimensional Imaging Printed 11/2012 Multidimensional Imaging Content Quick Start Guide Content 1 Introduction 4 2 Set up multi-channel experiments 5 2.1 Set up a new experiment
R&S ZVA110 Vector Network Analyzer Specifications
ZVA110_dat-sw_en_5214-4813-22_cover.indd 1 Data Sheet 04.00 Test & Measurement R&S ZVA110 Vector Network Analyzer Specifications 15.11.2013 14:42:28 CONTENTS Definitions... 3 Specifications... 4 Overview...
Benefits of the R&S RTO Oscilloscope's Digital Trigger. <Application Note> Products: R&S RTO Digital Oscilloscope
Benefits of the R&S RTO Oscilloscope's Digital Trigger Application Note Products: R&S RTO Digital Oscilloscope The trigger is a key element of an oscilloscope. It captures specific signal events for detailed
MOTOTRBO. Tech Notes. Auto-Tuning MOTOTRBO Mobiles with the Aeroflex 8800SX. Technical Enablement Team. Place your image on top of this gray box.
Auto-Tuning MOTOTRBO Mobiles with the Aeroflex 8800SX MOTOTRBO Place your image on top of this gray box. Tech Notes If no graphic is applicable, MTN-13 delete gray box and notch-out behind gray box, from
USB Mini Spectrum Analyzer User Manual PC program TSA For TSA5G35 TSA4G1 TSA6G1 TSA12G5
USB Mini Spectrum Analyzer User Manual PC program TSA For TSA5G35 TSA4G1 TSA6G1 TSA12G5 Triarchy Technologies, Corp. Page 1 of 17 USB Mini Spectrum Analyzer User Manual Copyright Notice Copyright 2013
GSM Mobile Tests under Conditions of Fading
GSM Mobile Tests under Conditions of Fading Application Note 1MA02_0E Subject to change Detlev Liebl 97-09 Products: Digital Radiocommunication Test Set CRTP02 / CRTC02 Signal Generator SMIQ Contents 1.
OPERATING MANUAL MINIRATOR MR2 / MR-PRO
OPERATING MANUAL MINIRATOR MR2 / MR-PRO NTI Contact Info NTI AG Im alten Riet 102 9494 Schaan Liechtenstein, Europe Tel.: +423-239 6060 Fax: +423-239 6089 E-Mail: info@nti-audio.com Web: www.nti-audio.com
Using the Forum Application for Remote Control Application Note. Forum is a free scripting tool for remote control of Rohde & Schwarz instruments.
Using the Forum Application for Remote Control Application Note Forum is a free scripting tool for remote control of Rohde & Schwarz instruments. Application Note Fabian Liebl November 2011-1MA196_1e Table
R&S ZNBT Vector Network Analyzer Specifications
R&S ZNBT Vector Network Analyzer Specifications year Data Sheet Version 05.02 CONTENTS Definitions... 3 Measurement range... 4 Measurement speed... 6 Measurement accuracy of the R&S ZNBT8... 8 Measurement
R&S CMW500 Digital IQ with CADENCE Emulator Application Note
R&S CMW500 Digital IQ with CADENCE Emulator Application Note Products: R&S CMW500 R&S EX-IQ-BOX R&S FSQ R&S FSV R&S EXBOX-Z3 This application note explains how to bring a CADENCE system which is attached
Quick Operation Guide of LTN7700/7600 Series NVR
Quick Operation Guide of LTN7700/7600 Series NVR UD.6L0202B0042A02 Thank you for purchasing our product. If there is any question or request, please do not hesitate to contact dealer. This manual is applicable
Dynamic re-referencing Microvolt-level measurements with the R&S RTO oscilloscopes
RTO_app-bro_3607-2855-92_v0100.indd 1 Microvolt-level measurements with the R&S RTO Test & Measurement Application Brochure 01.00 Dynamic re-referencing Microvolt-level measurements with the R&S RTO oscilloscopes
JD725A Cable and Antenna Analyzer - Dual Port
COMMUNICATIONS TEST & MEASUREMENT SOLUTIONS JD725A Cable and Antenna Analyzer - Dual Port Key Features Portable and lightweight handheld instrument Built-in wireless frequency bands as well as the most
USB-TG124A Tracking Generator User Manual
USB-TG124A Tracking Generator User Manual Signal Hound USB-TG124A User Manual 2017, Signal Hound, Inc. 35707 NE 86th Ave La Center, WA 98629 USA Phone 360.263.5006 Fax 360.263.5007 This information is
Electrical and Electronic Laboratory Faculty of Engineering Chulalongkorn University. Cathode-Ray Oscilloscope (CRO)
2141274 Electrical and Electronic Laboratory Faculty of Engineering Chulalongkorn University Cathode-Ray Oscilloscope (CRO) Objectives You will be able to use an oscilloscope to measure voltage, frequency
DS-7200HVI/HFI-SH Series DVR Quick Operation Guide
DS-7200HVI/HFI-SH Series DVR Quick Operation Guide UD.6L0202B0019A01 Thank you for purchasing our product. If there is any question or request, please do not hesitate to contact dealer. This manual is
Advanced Techniques for Spurious Measurements with R&S FSW-K50 White Paper
Advanced Techniques for Spurious Measurements with R&S FSW-K50 White Paper Products: ı ı R&S FSW R&S FSW-K50 Spurious emission search with spectrum analyzers is one of the most demanding measurements in
Versatile RF Fading Simulator With R&S FSQ/FSG/FSV and R&S SMU Application Note
Versatile RF Fading Simulator With R&S FSQ/FSG/FSV and R&S SMU Application Note Products: R&S SMU200A R&S SMU-B17 R&S SMU-B14 R&S FSQ R&S FSG R&S FSQ-B17 R&S FSV R&S FSV-B17 R&S FSV-B70 Fading in the baseband
Generation and Measurement of Burst Digital Audio Signals with Audio Analyzer UPD
Generation and Measurement of Burst Digital Audio Signals with Audio Analyzer UPD Application Note GA8_0L Klaus Schiffner, Tilman Betz, 7/97 Subject to change Product: Audio Analyzer UPD . Introduction
MINI PC SCOPE PCSU01. User manual. test leads software download USB cable design enclosure
MINI PC SCOPE PCSU01 User manual Features test leads software download USB cable design enclosure Specifications oscilloscope: o bandwidth: DC to 200 khz ± 3 db o input impedance: 100 ko / 20 pf o maximum
R&S FSV-K73 3G FDD UE (UL) Measurements incl. HSUPA Specifications
FSV-K73_dat-sw_en_5214-0976-22_cover.indd 1 Data Sheet 02.00 Test & Measurement R&S FSV-K73 3G FDD UE (UL) Measurements incl. HSUPA Specifications 01.08.2013 17:36:27 CONTENTS Specifications... 3 Frequency...
INSTALATION AND OPERATION MANUAL ABYSSAL OS Overlay Module Version 1.3
INSTALATION AND OPERATION MANUAL ABYSSAL OS Overlay Module Version 1.3 Thank you for purchasing the Abyssal OS Overlay Module for your ROV. This instruction manual contains all the information you ll need
SynthiaPC User's Guide
Always There to Beautifully Play Your Favorite Hymns and Church Music SynthiaPC User's Guide A Product Of Suncoast Systems, Inc 6001 South Highway 99 Walnut Hill, Florida 32568 (850) 478-6477 Table Of
QUICK START GUIDE FOR DEMONSTRATION CIRCUIT /12/14 BIT 10 TO 65 MSPS DUAL ADC
LTC2286, LTC2287, LTC2288, LTC2290, LTC2291, LTC2292, LTC2293, LTC2294, LTC2295, LTC2296, LTC2297, LTC2298 or LTC2299 DESCRIPTION Demonstration circuit 816 supports a family of s. Each assembly features
R&S RSC Step Attenuator Where precise signal levels count
Test & Measurement Product Brochure 01.00 Step Attenuator Where precise signal levels count Step Attenuator At a glance The is a switchable, mechanical step attenuator. It is available in various models
QUICK START GUIDE FOR DEMONSTRATION CIRCUIT /12/14 BIT 10 TO 105 MSPS ADC
LTC2280, LTC2282, LTC2284, LTC2286, LTC2287, LTC2288 LTC2289, LTC2290, LTC2291, LTC2292, LTC2293, LTC2294, LTC2295, LTC2296, LTC2297, LTC2298 or LTC2299 DESCRIPTION Demonstration circuit 851 supports a
Zoom Pro Classroom Facilitator and User Guide
Zoom Pro Classroom Facilitator and User Guide Version 1 08/21/2017 Tom M c Kellips, DM 1 Table of Contents Introduction...3 Technical Support...3 Starting a video conference...3 Sharing a screen...4 Recording...4
Manual of Operation for WaveNode Model WN-2m. Revision 1.0
Manual of Operation for WaveNode Model WN-2m. Revision 1.0 TABLE OF CONTENTS 1. Description of Operation 2. Features 3. Installation and Checkout 4. Graphical Menus 5. Information for Software Expansion
R&S GU221 Filter Control Unit Specifications
R&S GU221 Filter Control Unit Specifications Secure Communications Data Sheet 01.00 Definitions General Product data applies under the following conditions: Three hours storage at ambient temperature followed
A-Line LOUDSPEAKER SYSTEM. Mobile Audio Concert Sound Fixed Installation Pro Entertainment. English
A-Line LOUDSPEAKER SYSTEM Mobile Audio Concert Sound Fixed Installation Pro Entertainment English A-LINE FAMILY A Perfect Match THE A-LINE IS YOUR ENTRY TO PROFESSIONAL DYNACORD LIVE SOUND Powered loudspeakers
RF Level Test System +20 dbm to 130 dbm
NRVD Power Meter optional Therm. Sensor A B Power: >-15 dbm DUT (Signal Generator, Communication Tester) 1 MHz - 3.5/6 GHz +20 dbm... -130 dbm Diode Sensor Z4 Power: -15 to -40 dbm 6 db Power =< -40 dbm
SISII User Guide SISII. Sound Editor STC-S521. User Guide
SISII SISII Sound Editor STC-S521 Note to Customer NOTE TO CUSTOMER Thank you for choosing our product. We hope you will find STC software useful as it will help you to resolve your tasks. Before getting
dbtechnologies QUICK REFERENCE
dbtechnologies QUICK REFERENCE 1 DVA Composer Ver3.1 dbtechnologies What s new in version 3.1 COMPOSER WINDOW - DVA T8 line array module now available in the System Models window. - Adding modules in the
News from Rohde&Schwarz Number 195 (2008/I)
BROADCASTING TV analyzers 45120-2 48 R&S ETL TV Analyzer The all-purpose instrument for all major digital and analog TV standards Transmitter production, installation, and service require measuring equipment
Oscilloscope Guide Tektronix TDS3034B & TDS3052B
Tektronix TDS3034B & TDS3052B Version 2008-Jan-1 Dept. of Electrical & Computer Engineering Portland State University Copyright 2008 Portland State University 1 Basic Information This guide provides basic
Converting MediaFLO Waveform Files to R&S SFU / SFE / SFE100 ARB Format Using IQWIZARD/WinIQSIM TM for R&S SFx-K35 ARB
Products: R&S SFU Broadcast Test System, R&S SFE Broadcast Tester, R&S SFE100 Test Transmitter Converting MediaFLO Waveform Files to R&S SFU / SFE / SFE100 ARB Format Using IQWIZARD/WinIQSIM TM for R&S
R&S FSW-K144 5G NR Measurement Application Specifications
R&S FSW-K144 5G NR Measurement Application Specifications Data Sheet Version 01.00 CONTENTS Definitions... 3 Specifications... 4 Overview... 4 Assignment of option numbers to link modes... 4 Supported
Introduction to EndNote Online
Introduction to EndNote Online Creating an EndNote Online account Go to EndNote Online. Click on the Access EndNote Online button and, if prompted, enter your Warwick username and password to confirm you
Aurora Grid-Tie Installation Instructions (Model Number: PVI-3.0-OUTD-US-W) Revision 4.1
Aurora Grid-Tie Installation Instructions (Model Number: PVI-3.0-OUTD-US-W) Revision 4.1 Contents 1) Grid-Tie Installation Block Diagram... 3 2) Installation Steps.... 4 2.1) Initial Setup.... 4 2.1.1)
Digiline 8 All-in-One Audio Matrix Verstärker Digiline 8 All-in-One 8x8 Audio Matrix Verstärker mit 1300W
Digiline 8 All-in-One Audio Matrix Verstärker Digiline 8 All-in-One 8x8 Audio Matrix Verstärker mit 1300W Mayrwiesstr. 25-5300 Hallwang Tel.: +43 (0)662-66 15 56 - Fax: +43 (0)662-66 09 63 satec@satec.at
Tutorial FITMASTER Tutorial
Tutorial 2.20 FITMASTER Tutorial HEKA Elektronik Phone +49 (0) 6325 / 95 53-0 Dr. Schulze GmbH Fax +49 (0) 6325 / 95 53-50 Wiesenstrasse 71 Web Site www.heka.com D-67466 Lambrecht/Pfalz Email sales@heka.com
R&S MDS-21 Absorbing Clamp Measurement of disturbance power and screening effectiveness on cables
MDS-21_bro_en_3607-5319-12_v0101.indd 1 Product Brochure 01.01 Test & Measurement Measurement of disturbance power and screening effectiveness on cables 17.11.2016 15:21:31 At a glance The absorbing clamp
R&S RT-ZF20 Power Deskew Fixture User Manual
R&S RT-ZF20 Power Deskew Fixture User Manual (B00X2) User Manual Test & Measurement 1800.0040.02 04 This manual describes the following R&S RT-ZF models: R&S RT-ZF20 (1800.0004.01) 2016 Rohde & Schwarz
Mobile Tests for GSM 900 / 1800 / 1900 under Fading Conditions
FADING 900 1800 1900 Products: CRTP02, CRTC02, SMIQ Mobile Tests for GSM 900 / 1800 / 1900 under Fading Conditions This application note describes how the most important fading tests for mobile phonesin
Supplement to the Operating Instructions. PRemote V 1.2.x. Dallmeier electronic GmbH. DK GB / Rev /
Supplement to the Operating Instructions PRemote V 1.2.x 1 DK 180.000.000 GB / Rev. 1.2.3 / 030416 PRemote V 1.2.x Copyright All rights reserved. This document may not be copied, photocopied, reproduced,
R&S SMBV-Z1 Reference Frequency Converter Specifications
Test & Measurement Data Sheet 01.01 R&S SMBV-Z1 Reference Frequency Converter Specifications Version 01.01, July 2011 CONTENTS Definitions... 3 Introduction... 4 Specifications... 4 Input signal...4 Output
TL-2900 AMMONIA & NITRATE ANALYZER DUAL CHANNEL
TL-2900 AMMONIA & NITRATE ANALYZER DUAL CHANNEL DATA ACQUISITION SYSTEM V.15.4 INSTRUCTION MANUAL Timberline Instruments, LLC 1880 S. Flatiron Ct., Unit I Boulder, Colorado 80301 Ph: (303) 440-8779 Fx:
DVB-T USB SET-TOP BOX
DVB-T USB SET-TOP BOX User Manual Version: 1.0 (February 2005) TRANSYSTEM INC. No.1-2 Li-Hsin Rd.I Science-Based Industrial Park, Hsinchu, Taiwan Tel:+886-3-5780393 Fax:+886-3-5784111 e-mail: sales@transystem.com.tw
VideoMate U3 Digital Terrestrial USB 2.0 TV Box Start Up Guide
VideoMate U3 Digital Terrestrial USB 2.0 TV Box Start Up Guide Compro Technology, Inc. www.comprousa.com Copyright 2001-2005. Compro Technology, Inc. No part of this document may be copied or reproduced
J.M. Stewart Corporation 2201 Cantu Ct., Suite 218 Sarasota, FL Stewartsigns.com
DataMax INDOOR LED MESSAGE CENTER OWNER S MANUAL QUICK START J.M. Stewart Corporation 2201 Cantu Ct., Suite 218 Sarasota, FL 34232 800-237-3928 Stewartsigns.com J.M. Stewart Corporation Indoor LED Message
Keysight Method of Implementation (MOI) for VESA DisplayPort (DP) Standard Version 1.3 Cable-Connector Compliance Tests Using E5071C ENA Option TDR
Revision 1.00 February 27, 2015 Keysight Method of Implementation (MOI) for VESA DisplayPort (DP) Standard Version 1.3 Cable-Connector Compliance Tests Using E5071C ENA Option TDR 1 Table of Contents 1.
PHY221 Lab 1 Discovering Motion: Introduction to Logger Pro and the Motion Detector; Motion with Constant Velocity
PHY221 Lab 1 Discovering Motion: Introduction to Logger Pro and the Motion Detector; Motion with Constant Velocity Print Your Name Print Your Partners' Names Instructions August 31, 2016 Before lab, read
Using Spectrum Laboratory (Spec Lab) for Precise Audio Frequency Measurements
Using Spectrum Laboratory (Spec Lab) for Precise Audio Frequency Measurements Ver 1.15 Nov 2009 Jacques Audet VE2AZX ve2azx@amsat.org WEB: ve2azx.net NOTE: SpecLab version V2.7 b18 has some problems with
EMI Software R&S ES-K1
EMI Software R&S ES-K1 for fast, accurate and reproducible measurements User-friendly EMI test software under Windows EMI measurements to commercial and military standards such as CISPR, VDE, FCC, EACL,
UB22z Specifications. 2-WAY COMPACT FULL-RANGE See NOTES TABULAR DATA for details CONFIGURATION Subsystem DESCRIPTION
DESCRIPTION Ultra-compact 2-way system Wide projection pattern LF on angled baffles to maintain a wide upper/midrange beamwidth High output, high definition sound DESCRIPTION The UB22z is engineered for
R&S SFD DOCSIS Signal Generator Signal generator for DOCSIS 3.1 downstream and upstream
R&S SFD DOCSIS Signal Generator Signal generator for DOCSIS 3.1 downstream and upstream SFD_bro_en_3607-3739-12_v0100.indd 1 Product Brochure 01.00 Test & Measurement Broadcast & Media year 24.05.2016