Scanning Electron Microscopy (FEI Versa 3D Dual Beam)

Size: px
Start display at page:

Download "Scanning Electron Microscopy (FEI Versa 3D Dual Beam)"

Transcription

1 Scanning Electron Microscopy (FEI Versa 3D Dual Beam) This operating procedure intends to provide guidance for basic measurements on a standard sample with FEI Versa 3D SEM. For more advanced techniques or imaging in environmental SEM mode, please contact NCC staff members. Log On to the System 1. Please use your RCS ID to logon to the computer. 2. You will see a counter on the screen, minimize this screen. Do not forget to logout when your session ends. Starting the System 1. Type the username and password on the xtm Log On screen. The application status window will appear, click on hide. 2. The screen now shows the previous imaging parameters. Each of the quadrants corresponds to: E-beam, I-beam, Nav-Cam, CCD camera. Note: Operations are made on the active quad highlighted in blue. (Example: First quadrant, e-beam in Figure 1) Figure 1: Main screen, each of the quadrants corresponds to: Quadrant 1 (E-beam), Quadrant 2 (Ibeam), Quadrant 3 (Navigation Camera (Nav-Cam)), Quadrant 4 (CCD camera), Top panel and <Tabs> in the Right panel is to adjust parameters 1

2 Sample Loading 1. To load the sample, vent the chamber first. VENT button is located on the first tab on the right of the screen,, on the <Beam Control> tab,, on the right panel. Once hit, the button will turn into yellow and the venting will take 3min. In addition, the diagram on the bottom right turns into orange,, indicating that it is being vented. Wait until the venting process is completed. 2. Open the chamber door, (i) place the stub, (ii) using the hexagonal screw driver, finger tight the stub to the base, (iii) Then click on PUMP on the <Beam Control> tab,, on the right panel. (Figure 2) Figure 2. Loading the sample (i) place the stub, (ii) tighten to the base (iii) support the chamber door and click on PUMP 3. The button will turn into yellow, and the chamber on the diagram will turn orange,, wait until the chamber pressure is stabilized and diagram turns green,. Figure 3. Setting up the high-vacuum in the chamber Moving the stage 1. Hide the Application Status bar. Live sample movements can be observed in the fourth quadrant (CCD camera). To move the stage: 2

3 a. Either type in on the <Navigation> tab,, second from the left, the red arrow right next to Z indicates that the stage is not linked, the reference distance is the base of the sample chamber. Start typing 30mm, which raises the stage, then type in 40mm, then raise the stage gradually until you reach 5 mm below 10mm working distance line (yellow line on the CCD camera) (Figure 4A) b. Other option is to use the mouse, click and hold on the wheel, and slowly drag the mouse. Yellow bar and an arrow will appear on the fourth quadrant. (Figure 4B) Figure 4. (A) Raising the stage by typing on the <Navigation> tab, (B) Raising the stage by mouse click 2. Go to <Stage> menu at the top of the window and click on <take a nav-cam photo>. (Figure 5) This will bring the sample under the navigation camera, take a photo (appears on the third quadrant (Nav-Cam)) and bring the sample to original position. Figure 5. (A) Click on the take a nav-cam photo on the top menu (B) The sample appears on the third quadrant (Nav-Cam) (C) The stage comes back to original position 3

4 3. Navigate on the sample on the third quad, to record a specific position, doubleclick on a point on a point of interest and add this position using the ADD button on the <Navigation> tab,. Starting the Beams 1. To start the beams, activate the corresponding window and on the <Beam Control> tab, click on BEAM ON button. Yellow BEAM ON button means, it is activated. (Do not forget the click on BEAM ON button before leaving) <Beam Control> tab BEAM ON button 2. If you are planning to make Pt deposition, it should be heated up. To heat up Pt, <Patterning> tab,,on the right panel, and double click on Cold. It will turn into yellow as it heats up, be advised that you should wait for 10 min to stabilize. Linking the Stage 1. Turn on the e-beam. Click on the unpause button on the first quadrant (e-beam),. 2. Raise the stage to working distance to 10mm by mouse (click and hold the wheel and drag on the fourth quad until the sample surface reached WD = 10mm, yellow bar). 3. Adjust the focus (coarse and fine knobs), brightness and contrast (B/C) using the knobs on the control bar. (Figure 6) 4

5 Figure 6. Manual Control Bar 4. Adjust the magnification to 2000X and re-focus. (Figure 7) 5. Link the stage, using the fourth button at the top bar,. The red arrow on the <Navigation> tab should turn into gray, once the stage is linked. (Figure 7) 6. Once the stage is linked, re-link the stage as much as you wish, especially after making large stage movements. If you have multiple samples with different heights, relink the stage as you move. Figure 7. Linking the stage (i) Raise the stage to WD = 10mm, (ii) Focus on a feature at 2000X, adjust brightness and contrast using the knobs on the control bar, (iii) Link the stage (iv) Make sure the red arrow on the <Navigation> tab turns into gray. 5

6 High Resolution E-beam Imaging 1. Make sure first quadrant (e-beam) is active window, then check the detector from the <Detectors> tab, the right panel as ETD Secondary Electrons., on 2. On the point of interest in the first quadrant, increase the magnification by the knobs on the control bar. Navigation on the sample surface is made by double clicking on a point, or click and hold mouse wheel and drag. 3. Starting parameters (Figure 8) a. Select an appropriate voltage (10kV or 20kV) and adjust the spot size to 5. b. Every time you change the spot size, the B/C will change, you may re-adjust them by the knobs, if you get lost, try using the auto brightness/contrast button,. c. If you prefer to work on a reduced area,, (the top third button from the left), or you may change the dwell time to 300 ns and the resolution. Figure 8. Parameters Setting buttons in the top panel 4. Adjust focus, brightness/contrast, stigmation and lens modulator iteratively. (Figure 9) a. Focus: Using the knobs (coarse and fine) on the control bar, or use the mouse: hold the right click and drag left and right. b. Brightness and contrast (B/C): Using the knobs on the control bar, or use the bar on the <Beam Control> tab,, on the right panel. c. Stigmatism: Using the knobs (X and Y stigmator) on the control bar, or use the mouse: hold (shift+right click) and drag to left-right (X stigmator) and up-down (Ystigmator) or use the panel on the <Beam Control> tab,, on the right panel. d. Lens alignment: Using the Direct Adjustment button at the top panel,, click on lens modulator. Working on a reduced area, amplitude if needed., is easier; change the 6

7 Figure 9. (i) Working in a reduced area, (ii) Adjusting B/C and stigmators on the right panel, (ii) Lens Modulator adjustment via Direct Adjustments Saving Image 1. Click on the arrow on Image Acquisition button at the top of the window, and take a photo. This will save the image on the pre-set quality and scanning parameters, (dwell time:10 us, Resolution: 3072x2048) 2. You may measure the features by clicking the Direct measurements button at the top of the window, delete them by keyboard. 3. Save the image with the measurement on the save as window: a. Save image with Databar, saves the information on the screen as a databar b. Save image with overlaid graphics saves the image with the measurement and text. Please save your results to Network>SPC-VE52>Shared Data to a folder by your full name and surname. 7

8 Back-Scattered Electron Imaging 1. On the second (I-beam) quad, left click on the i-beam icon on the bottom left corner and change it to e-beam icon. 2. Highlight the first (e-beam) quad and select your point of interest and focus on the image and link the stage (this is very important!) 3. Highlight the second quad, then go to <Detectors> tab,, and select Detector as CBS (Concentric Back Scattered Electron Detector) and then click on INSERT button. 4. The incoming CBS detector interferes with the CCD camera, the live image will be paused, unpause the fourth quadrant (CCD camera). 5. Once the CBS detector is in, you may see it coming from left. The CBS detector will change the B/C of your image on e-beam quad. While working with both ETD and CBS detectors, adjusting focus will affect both images, only brightness and contrast adjustments can be made separately. Auto-brightness button adjusts both quads. 6. To combine secondary and back-scattered imaging, change the third quadrant (Nav-Cam) to e-beam, by clicking the icon on the bottom left, this will combine the secondary and back-scattered images in a single window (third quadrant). The detector is now MIX, where you may adjust the colors of the ETD and CBS by sliding the bar. 7. You may separately collect images from the ETD and CBS detectors by Image Acquisition button. 8. Once finished, highlight the second quad, go to <Detectors> tab and retract the CBS detector. Then, go to fourth quadrant (CCD cam) and unpause it and receive live image from your sample. 8

9 Setting Eucentric Position At the eucentric position, the stage tilt and the beam axes intersect. When the stage is tilted or rotated in any direction, this point remains focused and almost does not shift. 1. To check the eucentric point, make sure you performed all the steps in moving and linking the stage and high resolution e-beam imaging. Follow the instructions below (Figure 10): i. Set the working distance to 10mm by typing in the Z on the <Navigation> tab, ii. pick and center a feature at 10kX focus on the image in the first quad and iii. link the stage. iv. Then type the tilt angle as 5. Check the mark and the feature: if it moves up then use the slider on the right to bring it down or vice versa. Figure 10. Setting the eucentric position: (i) set the working distance as 10mm (ii) increase the magnification to 10kX, focus on an identifiable feature, (iii) link the stage, (iv) change the tilt angle gradually, center the feature using the slider 2. Iterate this procedure for 10, 20, 30 and 52 (Figure 11). Then tilt the stage back to 0, the feature should come back to its original position. 9

10 Figure 11. Stage tilt and presentation of a feature High Resolution I-Beam Imaging 1. Check the eucentric height (working distance = 10mm), tilt the stage to On the first (e-beam) quadrant, focus on the center of interest at 5kX magnification. 3. On the second (i-beam) quadrant, set the voltage to 30kV and current to 30pA. 4. Ion beam is damaging to your sample, you may follow these steps to focus on your sample: i. Instead of continuous imaging, highlight the second (i-beam) quad and use take snapshot option under Image Acquisition button at the top panel. Adjust the brightness/contrast and focus and take snapshot between iterations. ii. Or start with an area nearby of the point of interest and use the reduced area for adjustments. Set the reduced area on a feature, unpause the i-beam and adjust focus, brightness and contrast in the reduced area. Act fast! iii. Or change the direction of the beam, using beam shift panel, on the <Beam Control> tab, shift the beam, do the adjustments and bring to original beam position by right clicking on the panel and zero. 5. When you take a snapshot, if the magnifications are different in e-beam and i- beam screens, you may couple magnifications on the Beam Control tab, this will take the active screen magnification as the basis. 10

11 Pt-Deposition with e-beam 1. Before deposition Pt on a feature, make sure the GIS is on. 2. Pick the feature, set magnification (example: 3500X) and focus 3. Change the standard mode to analytical mode. Wait until it settles and link the stage (in first (e-beam ) quadrant). Analytical mode: In this mode the largest (1000 µm) aperture is automatically selected and the beam current is limited to a range. It is used for, imaging and mapping at currents above 4.0 na, and precise beam current control and high beam current stability. 4. Insert the Pt needle by clicking the green square on <Patterning> tab,. This will insert the Pt-needle, avoid making z-movements when the needle is in, and small movements on x and y directions. Inserting the needle will change the B/C. 5. On the Patterning tab (Figure 12): i. select the pattern as rectangular and ii. draw the rectangle (yellow at first), iii. from the applications menu, select Pt e-dep structures make sure the pattern turns into green. iv. Adjust the size of the area by typing, set the z to 100nm, the time required to complete the process is listed on this tab. 11

12 Figure 12. Pt-deposition in the Analytical Mode: (i) On the patterning tab, select the rectangular pattern, (ii) draw the area to be deposited, (iii) select the appropriate application and adjust the size. 6. Click on the start patterning button at the top panel. During the process, the active quad (e-beam) turns into green. 7. When the deposition is completed, first take the needle out by clicking the green square on the Gas Injection. 8. Delete the pattern and switch to Standard Mode. 9. Unpause the beam, you may see the Pt-deposited surface. Pt-Deposition with i-beam The Pt-deposition with e-beam Analytical mode is to protect the feature even during the Pt-deposition with i-beam. The i-beam is very damaging to surface, hence Pt-deposition with e-beam adds a protective cover. Follow the instructions below: 12

13 1. Focus on the point of interest, adjust magnification and link the stage on the first quad (e-beam), activate the second quad (i-beam) and adjust the accelerating voltage and current to 30 kv and 30 pa, respectively. 2. Check the eucentric point (on the point of interest), and tilt the stage to Insert the Pt needle by clicking the green square on <Patterning> tab,. This will insert the Pt-needle, avoid making z-movements when the needle is in, and small movements on x and y directions. Inserting the needle will change the B/C. 4. Activate the i-beam quad and take a snapshot by Image Acquisition button, you may even see the Pt-needle inserted. Select rectangular pattern and draw. Then select the Pt-dep from the applications menu, adjust the size of the pattern. (Figure 13) Figure 13. Pt-deposition: (i) On the patterning tab, select the rectangular pattern, (ii) draw the area to be deposited, (iii) select the appropriate application and adjust the size. 5. The time given is calculated based on the accelerating voltage and current settings (30 kv and 30pA), although these setting might be good for imaging purposes, 13

14 not sufficient for Pt-deposition. Increasing the current will yield faster deposition, however mills/damages the surface as well. Rule of thumb: The trade off between the milling and deposition: Calculate the area (x = 20 µm, y = 10 µm) as 200 µm2, multiply this by 6, 1200 µm2, this corresponds to the maximum current that should be used in terms of pa. In this case, the maximum current should be 1200 pa = 1.2 na. 6. Change the current to 1 na, control the time required. This will shift the image, so take a snapshot, if necessary correctly place your pattern. 7. Simultaneously, you may watch and record the deposition process via e-beam imaging. This feature ispi- is available at the <Patterning> tab,. Depending on the time interval, the e-beam quad will be refreshed. (Figure 14) Figure 14. Pt-deposition and simultaneous imaging with e-beam 8. When the deposition is completed, first take the needle out by clicking the green square on the Gas Injection and delete the pattern. 9. Switch to e-beam quad for imaging. Cross-Sectioning Cross sections are cut in a stair step fashion to allow the exposed layers to be seen with the electron beam when the stage is tilted to 52. To section a sample, you may protect the feature by Pt-deposition, or you may directly cross section the sample. 1. Focus on the point of interest, adjust magnification and link the stage on the first quad (e-beam), adjust the accelerating voltage and spot size to 30 kv and 5, respectively. 14

15 2. Check the eucentricity of the feature. The working distance = 10mm is the eucentric point. 3. Activate the second quadrant (i-beam), adjust the accelerating voltage and current to 30 kv and 30 pa, respectively. 4. If needed, deposit Pt on the surface, refer to previous sections for instructions. (F2 F3 F4) Figure 15. The progress of Pt-deposition on the surface 5. Focus on the point of interest in the second quadrant (i-beam) and go to <Patterning> tab,. From the drop-down list, select the regular cross section. And draw the pattern on the point of interest. Select the application as Si-multipass New, (or choose the application close to your sample from the list). Adjust the size of the area to be sectioned. Make sure the y-size is at least 2 times higher than the x-size. Set the z to 5 um. 6. Notice the total time, which is based on the current, 30 pa. Adjust the current to 15 na and check the time. Take a snapshot to check the final position of the pattern. 7. Set the imaging parameters in the first quadrant (e-beam), set ispi parameters on the <Patterning> tab. 8. Start the process with the Start Pattern button,. 15

16 Figure 16. Cross-sectioning process (i) ispi allows to view the progress, (ii) The sectioning on the second quadrant starts from the bottom and moves to top, its size is adjusted 9. Once the process in finished, change the current back to imaging current, 10 pa. Delete the pattern. 10. You may now activate the first quadrant (e-beam) and save images. 11. To image the cross-section with the i-beam, tilt the stage back to Rotate the stage 180 (compucentric rotation): a. Use the short cut key F12 b. Type in the rotation on the <Navigation> tab. 13. Change the scan rotation by 180, using the short cut key Shift-F12. Cleaning Cross-section 1. To minimize the damage and expose the cross-section, one fine-tuning is to clean the cross-section. 2. On the same spot, increase the magnification. On the <Patterning> tab, select the Cleaning Cross Section pattern and place the pattern, just to cover the wall of the cross-section. 3. Select Si-ccs New as application, adjust the size of the pattern and set the depth as the same you used in cross-sectioning. 4. The purpose is to clean the wall while minimizing the damage on the surface, therefore set the current to 1/5 (or less) of the current you used for cross-sectioning. (1nA) 5. You may iterate the procedure for lower currents for multiple times and increase the tilt angle to 53. (0.5 na, and 100 pa in this example) 6. After the process is completed, change the current to imaging current. 7. You may now activate the first quad (e-beam) and save images. 8. To image the cross-section with the i-beam, tilt the stage back to Rotate the stage 180 : a. Use the short cut key F12 16

17 b. Type in the rotation on the <Navigation> tab. 10. Change the scan rotation by 180, using the short cut key Shift-F12. System Shut-Down 1. Retract any inserted detectors. 2. Turn off the beams, on the Beam Control tab, either click on SLEEP button (or click on BEAM ON button for the beams you have used). 3. Return the stage to home position. The home stage option is under Stage > Home stage on the top menu. 4. Click on VENT button to remove your sample, wait until the chamber is completely vented (the pump will stop). 5. After removing the sample, hit PUMP again and leave the chamber in vacuum. 6. Log Off the system after the chamber vacuum is reached and the icon is fully green. 7. Log out from the user tracker.. 8. Record your time and experiment specifications in the logbook. 17

2.1. Log on to the TUMI system (you cannot proceed further until this is done).

2.1. Log on to the TUMI system (you cannot proceed further until this is done). FEI DB235 ex-situ lift out TEM sample preparation procedure Nicholas G Rudawski ngr@ufledu (805) 252-4916 Last updated: 06/19/15 DISCLAIMER: this procedure describes one specific method for preparing ex-situ

More information

Standard Operating Procedure for FEI Helios 660 NanoLab Part I: SEM Version

Standard Operating Procedure for FEI Helios 660 NanoLab Part I: SEM Version Standard Operating Procedure for FEI Helios 660 NanoLab Part I: SEM Version Helios reservations may be made online using the NERCF website. Note: Always wear gloves when venting the system and exchanging

More information

University of Minnesota Nano Fabrication Center Standard Operating Procedure

University of Minnesota Nano Fabrication Center Standard Operating Procedure Equipment Name: Focused Ion Beam (FIB) Coral Name: fib Revision Number: 2 Model: FEI Quanta 200 3D Revisionist: Kevin Roberts Location: Area 3 Date: 9/17/2013 1 Description The Quanta 200 3D is a DualBeam

More information

FEI FIB Focused Ion Beam

FEI FIB Focused Ion Beam Operating Manual Part 1 FEI FIB Focused Ion Beam IF IN DOUBT, ASK 1.00 IF ANYTHING UNUSUAL HAPPENS, OR IF THERE IS ANYTHING YOU ARE UNSURE ABOUT, STOP AND CONTACT ME! DO NOT PROCEED OR ATTEMPT TO FIX THE

More information

FEI Strata Dual-beam FIB

FEI Strata Dual-beam FIB FEI Strata Dual-beam FIB Quick start Guide Compiled by Mat t hew Hughes and Tony Chen Page 0 Purpose of this guide INTRODUCTION This quick start guide is meant to provide cursory operational knowledge

More information

FIB Operating Procedure. Effective Date: 08/14/2012 Author(s): Jiong Hua Phone:

FIB Operating Procedure. Effective Date: 08/14/2012 Author(s): Jiong Hua Phone: FIB Operating Procedure Effective Date: 08/14/2012 Author(s): Jiong Hua Phone: 402-472-3773 Email: jhua2@unl.edu 1 1 Introduction 1.1 Key Words Focused Ion Beam (FIB), FEI Strata 201, Ion milling 1.2 Purpose

More information

Operation Procedure for Phillips XL30 ESEM

Operation Procedure for Phillips XL30 ESEM Operation Procedure for Phillips XL30 ESEM The ESEM will be left in the ON state when not in use. The chamber will be at high vacuum, filament on, stage at home position, VAC and HT buttons lit, and monitor

More information

CHECKLIST FOR VERIOS OPERATION 1. GENERAL The SEM lab is used assuming "operating room" cleanliness, i.e., the SEM lab is a high visibility lab and

CHECKLIST FOR VERIOS OPERATION 1. GENERAL The SEM lab is used assuming operating room cleanliness, i.e., the SEM lab is a high visibility lab and CHECKLIST FOR VERIOS OPERATION 1. GENERAL The SEM lab is used assuming "operating room" cleanliness, i.e., the SEM lab is a high visibility lab and must be kept clean and neat so clean up behind yourself

More information

Description of task... 2 Sample preparation... 2 Sample insertion and SEM settings... 3 Eucentricity... 4

Description of task... 2 Sample preparation... 2 Sample insertion and SEM settings... 3 Eucentricity... 4 These are simple notes taken by Charlie Sanabria while learning how to use the Focused Ion Beam (FIB) in the Scanning Electron Microscope (SEM) at the Applied Superconductivity Center (ASC). The instructions

More information

The Quanta 200 3D Simplified Operation Manual

The Quanta 200 3D Simplified Operation Manual The Quanta 200 3D Simplified Operation Manual CONTENTS 1. SYSTEM OVERVIEW... 4 2. VACUUM SYSTEM... 5 3. SOFTWARE CONTROL... 6 Pages and Modules... 7 4. BASIC OPERATION OF SEM AND FIB... 12 Specimen Preparation

More information

Dektak Step by Step Instructions:

Dektak Step by Step Instructions: Dektak Step by Step Instructions: Before Using the Equipment SIGN IN THE LOG BOOK Part 1: Setup 1. Turn on the switch at the back of the dektak machine. Then start up the computer. 2. Place the sample

More information

Standard Operating Procedure II: EDS (Bruker Flat-Quad)

Standard Operating Procedure II: EDS (Bruker Flat-Quad) Standard Operating Procedure II: EDS (Bruker Flat-Quad) ywcmatsci.yale.edu ESC II, Room A119F 810 West Campus Drive West Haven, CT 06516 Version 1.1, October 2018 1 > FOLLOW the SOP strictly to keep the

More information

1. Check the accelerating voltage, must be at 200 kv (right screen), HT (µa) (left panel) at and Emission (left panel) at

1. Check the accelerating voltage, must be at 200 kv (right screen), HT (µa) (left panel) at and Emission (left panel) at JEOL 2010F MANUAL Quick check list 1. Check the accelerating voltage, must be at 200 kv (right screen), HT (µa) (left panel) at 0.96-0.97 and Emission (left panel) at 155-160. 2. Check the vacuum sequence

More information

Technical Procedure for Scanning Electron Microscope/ Energy Dispersive X-Ray System (SEM/EDX) for non-gsr Casework

Technical Procedure for Scanning Electron Microscope/ Energy Dispersive X-Ray System (SEM/EDX) for non-gsr Casework Technical Procedure for Scanning Electron Microscope/ Energy Dispersive X-Ray System (SEM/EDX) for non-gsr Casework 1.0 Purpose This technical procedure shall be followed for the operation of the Scanning

More information

PHI 5000 VersaProbe TM Operator s Guide

PHI 5000 VersaProbe TM Operator s Guide PHI 5000 VersaProbe TM Operator s Guide Part No. 705921 Rev. A Copyright 2006 ULVAC-PHI, INC. 370 Enzo, Chigasaki, JAPAN The PHI logo ( ) is a registered trademark of ULVAC-PHI, INC. Physical Electronics,

More information

Figure 1. MFP-3D software tray

Figure 1. MFP-3D software tray Asylum MFP-3D AFM SOP January 2017 Purpose of this Instrument: To obtain 3D surface topography at sub-nanometer scale resolution, measure contact and friction forces between surfaces in contact, measure

More information

1.2 Universiti Teknologi Brunei (UTB) reserves the right to award the tender in part or in full.

1.2 Universiti Teknologi Brunei (UTB) reserves the right to award the tender in part or in full. TENDER SPECIFICATIONS FOR THE SUPPLY, DELIVERY, INSTALLATION AND COMMISSIONING OF ONE UNIT OF VARIABLE PRESSURE ENVIRONMENTAL SCANNING ELECTRON MICROSCOPE (SEM) CUM ENERGY DISPERSIVE SPECTROSCOPY (EDS)

More information

Tender Notification for the procurement of a "Dual beam (FIB - FE SEM) system" at IISc (Last Date for submission of tenders: 31st March 2016)

Tender Notification for the procurement of a Dual beam (FIB - FE SEM) system at IISc (Last Date for submission of tenders: 31st March 2016) Tender Notification for the procurement of a "Dual beam (FIB - FE SEM) system" at IISc (Last Date for submission of tenders: 31st March 2016) Dear Sir/Madam, Kindly send your best quotation for the following

More information

The PHI VersaProbe operates with two essential software programs: PHI Summitt and Vacuum Watcher. A third program, MultiPak, handles data reduction.

The PHI VersaProbe operates with two essential software programs: PHI Summitt and Vacuum Watcher. A third program, MultiPak, handles data reduction. PHI VersaProbe Scanning XPS System I. Overview The PHI VersaProbe operates with two essential software programs: PHI Summitt and Vacuum Watcher. A third program, MultiPak, handles data reduction. PHI Summitt

More information

DektakXT Profilometer. Standard Operating Procedure

DektakXT Profilometer. Standard Operating Procedure DektakXT Profilometer Standard Operating Procedure 1. System startup and sample loading: a. Ensure system is powered on by looking at the controller to the left of the computer.(it is an online software,

More information

Kratos AXIS Ultra DLD X-ray Photoelectron Spectrometer Instructions

Kratos AXIS Ultra DLD X-ray Photoelectron Spectrometer Instructions Kratos AXIS Ultra DLD X-ray Photoelectron Spectrometer Instructions Note: Enter your complete name, CMRF user code, and date into the record book. You must have previously filled out a project information

More information

Standard Operating Procedure of nanoir2-s

Standard Operating Procedure of nanoir2-s Standard Operating Procedure of nanoir2-s The Anasys nanoir2 system is the AFM-based nanoscale infrared (IR) spectrometer, which has a patented technique based on photothermal induced resonance (PTIR),

More information

AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode)

AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode) AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode) 1. Log into the Log Usage system on the SMIF web site 2. Open Nanoscope 6.14r1 software by double clicking on the Nanoscope 6.14r1 desktop

More information

JEM 3010 Manual (2011 Edition)

JEM 3010 Manual (2011 Edition) JEM 3010 Manual (2011 Edition) Basic Alignment Instructions Check the vacuum (power supply closet) and write on the log book: Gun pressure SIP2 < 2x10-5 Pa (usually 0.8-1.0x10-5 Pa) Column pressure SIP1

More information

Data Collection Using APEX3. March 30, Chemical Crystallography Laboratory

Data Collection Using APEX3. March 30, Chemical Crystallography Laboratory Data Collection Using APEX3 Page 1 of 10 Data Collection Using APEX3 March 30, 2017 Chemical Crystallography Laboratory Author: Douglas R. Powell Data Collection Using APEX3 Page 2 of 10 Distribution Douglas

More information

JEM 2100 Manual. Operation and Basic Alignment Instructions. Check the vacuum levels (power supply closet) Column (blue scale): < 2.

JEM 2100 Manual. Operation and Basic Alignment Instructions. Check the vacuum levels (power supply closet) Column (blue scale): < 2. 1. Verify the vacuum: JEM 2100 Manual Operation and Basic Alignment Instructions Check the vacuum levels (power supply closet) Column (blue scale): < 2.5x10-5 Pa If vacuum is not good enough, contact somebody

More information

Mini Micro Pulse Lidar System

Mini Micro Pulse Lidar System Mini Micro Pulse Lidar System MiniMPL-532-C Sensor Suite Operations Manual Version: June 2016 THIS PAGE INTENTIONALLY LEFT BLANK 2 Table of Contents MINIMPL SENSOR SUITE SYSTEM: RECORD OF PURCHASE... 4

More information

SPM Training Manual Veeco Bioscope II NIFTI-NUANCE Center Northwestern University

SPM Training Manual Veeco Bioscope II NIFTI-NUANCE Center Northwestern University SPM Training Manual Veeco Bioscope II NIFTI-NUANCE Center Northwestern University Introduction: Scanning Probe Microscopy (SPM) is a general term referring to surface characterization techniques that utilize

More information

Nova NanoSEM Superior Imaging and Analytical Performance

Nova NanoSEM Superior Imaging and Analytical Performance Nova NanoSEM Superior Imaging and Analytical Performance FEI Nova NanoSEM scanning electron microscopes combine best-in-class imaging with superb analytical performance in one easy-to-use instrument.

More information

Backside Circuit Edit on Full-Thickness Silicon Devices

Backside Circuit Edit on Full-Thickness Silicon Devices Backside Circuit Edit on Full-Thickness Silicon Devices Presentation Title Line 1 Title Line Two Can I really skip the global thinning step?! Date Presenter Name Chad Rue FEI Company, Hillsboro, OR, USA

More information

Preparing for remote data collection at NE-CAT

Preparing for remote data collection at NE-CAT Preparing for remote data collection at NE-CAT Important Note: The beamtime and remote login privileges are intended just for you and your group. You are not allowed to share these with any other person

More information

F7000NV ROBOT VISION OPERATING MANUAL

F7000NV ROBOT VISION OPERATING MANUAL Rev. C Feb 2012 F7000NV ROBOT VISION OPERATING MANUAL Rev. C Feb 2012 This page has intentionally been left blank. Contents Contents Chapter 1. Getting Started... 5 1. Preface... 5 2. Manuals... 5 3. Setting

More information

E X P E R I M E N T 1

E X P E R I M E N T 1 E X P E R I M E N T 1 Getting to Know Data Studio Produced by the Physics Staff at Collin College Copyright Collin College Physics Department. All Rights Reserved. University Physics, Exp 1: Getting to

More information

INCA ENERGY EDS TRAINING. System Block Diagram. INCA Energy Software. Xiang Yang EM SMU. Navigators. Point & ID Navigator.

INCA ENERGY EDS TRAINING. System Block Diagram. INCA Energy Software. Xiang Yang EM SMU. Navigators. Point & ID Navigator. INCA ENERGY EDS TRAINING Xiang Yang EM Center @ SMU System Block Diagram INCA Energy Software Navigators Point & ID Navigator Data tree 1 Spectrum Export Bitmap TIFF JPEG Metafile EMSA ISIS Optimum Conditions

More information

STANDARD OPERATING PROCEDURE: ASYLUM MFP-3D AFM

STANDARD OPERATING PROCEDURE: ASYLUM MFP-3D AFM STANDARD OPERATING PROCEDURE: ASYLUM MFP-3D AFM Purpose of this Instrument: To obtain 3D surface topography at sub-nanometer scale resolution, measure contact and friction forces between surfaces in contact,

More information

Focused Ion Beam System MI4050

Focused Ion Beam System MI4050 SCIENTIFIC INSTRUMENT NEWS 2016 Vol. 7 SEPTEMBER Technical magazine of Electron Microscope and Analytical Instruments. Technical Explanation Focused Ion Beam System MI4050 Yasushi Kuroda *1, Yoshihisa

More information

Veeco Dektak 6M Profilometer

Veeco Dektak 6M Profilometer Veeco Dektak 6M Profilometer System Ranges/Resolutions Range (Å) Resolution (Å) 50 (5nm) to 65K 1 0.5K to 655K 10 2K to 2620K 40 8K to 10000K (1mm) 160 Maximum sample thickness: 31.75mm Scan range: 50

More information

Dektak II SOP Revision 1 05/30/12 Page 1 of 5. NRF Dektak II SOP

Dektak II SOP Revision 1 05/30/12 Page 1 of 5. NRF Dektak II SOP Page 1 of 5 NRF Dektak II SOP The Dektak II-A is a sensitive stylus profilometer. A diamond-tipped stylus is moved laterally across the surface while in contact and measures deflections of the tip. It

More information

FACSAria I Standard Operation Protocol Basic Operation

FACSAria I Standard Operation Protocol Basic Operation FACSAria I Standard Operation Protocol Basic Operation 1. Checking Lasers Status a. Please check the ON / OFF of the lasers. Sufficient time (~30 minutes) need to be given to allow the laser(s) to warm

More information

CHA EVAPORATOR. User guidelines. p.1. by Carlos Manzanedo. Last revised: 10/05/2000

CHA EVAPORATOR. User guidelines. p.1. by Carlos Manzanedo. Last revised: 10/05/2000 CHA EVAPORATOR User guidelines by Carlos Manzanedo. Last revised: 10/05/2000 p.1 Introduction. The CHA evaporator has the following characteristics: 1. 2 independent resistive power supplies. 2. 1 Electron

More information

Personal Protective Equipment Wear nitrile gloves, lab coat, and safety glasses as a minimum protection, unless otherwise indicated.

Personal Protective Equipment Wear nitrile gloves, lab coat, and safety glasses as a minimum protection, unless otherwise indicated. 4pt Bending, Mouse This protocol is for standard Jepsen 4pt bending of adult mouse bone. Safety considerations Please reference the Jepsen laboratory when using this protocol. This protocol is subject

More information

SEM- EDS Instruction Manual

SEM- EDS Instruction Manual SEM- EDS Instruction Manual Double-click on the Spirit icon ( ) on the desktop to start the software program. I. X-ray Functions Access the basic X-ray acquisition, display and analysis functions through

More information

Quick Start ATOMIC FORCE MICROSCOPE West Campus Imaging Core

Quick Start ATOMIC FORCE MICROSCOPE West Campus Imaging Core Quick Start ATOMIC FORCE MICROSCOPE West Campus Imaging Core 1 Turn On the laser power 2 Open enclosure: - lift the door latch and open the enclosure door. 3 2 1 1.Unlock scanner: Lift the lever to the

More information

Husky Stadium CLUB HUSKY Seat Selection Instruction Manual

Husky Stadium CLUB HUSKY Seat Selection Instruction Manual Husky Stadium 2013 CLUB HUSKY 1 Husky Athletics is very excited to share this state-of-the-art 3D technology with you. You will have the ability to view and select the best available seats according to

More information

Quick Start Bruker Dimension Icon AFM

Quick Start Bruker Dimension Icon AFM Do not remove Quick Start Bruker Dimension Icon AFM March 3, 2015 GLA Contacts Harold Fu (hfu@caltech.edu) Weilai Yu (wyyu@caltech.edu) Bruker Tech Support (AFMSupport@bruker-nano.com 800-873-9750) Watch

More information

Linkage 3.6. User s Guide

Linkage 3.6. User s Guide Linkage 3.6 User s Guide David Rector Friday, December 01, 2017 Table of Contents Table of Contents... 2 Release Notes (Recently New and Changed Stuff)... 3 Installation... 3 Running the Linkage Program...

More information

HV/PHA Adjustment (PB) Part

HV/PHA Adjustment (PB) Part HV/PHA Adjustment (PB) Part Contents Contents 1. How to set Part conditions...1 1.1 Setting conditions... 1 2. HV/PHA adjustment sequence...7 3. How to use this Part...9 HV/PHA Adjustment (PB) Part i

More information

Cover Page for Lab Report Group Portion. Boundary Layer Measurements

Cover Page for Lab Report Group Portion. Boundary Layer Measurements Cover Page for Lab Report Group Portion Boundary Layer Measurements Prepared by Professor J. M. Cimbala, Penn State University Latest revision: 23 February 2017 Name 1: Name 2: Name 3: [Name 4: ] Date:

More information

Revision 1.2d

Revision 1.2d Specifications subject to change without notice 0 of 16 Universal Encoder Checker Universal Encoder Checker...1 Description...2 Components...2 Encoder Checker and Adapter Connections...2 Warning: High

More information

CHA EVAPORATOR Short Reference User guidelines

CHA EVAPORATOR Short Reference User guidelines CHA EVAPORATOR Short Reference User guidelines by Carlos Manzanedo. Last revised: 10/05/2000 p.1 Procedure: The CHA evaporator is divided into 16 different panels. These panels are labeled on the machine

More information

LedSet User s Manual V Official website: 1 /

LedSet User s Manual V Official website:   1 / LedSet User s Manual V2.6.1 1 / 42 20171123 Contents 1. Interface... 3 1.1. Option Menu... 4 1.1.1. Screen Configuration... 4 1.1.1.1. Instruction to Sender/ Receiver/ Display Connection... 4 1.1.1.2.

More information

Standard AFM Modes User s Manual

Standard AFM Modes User s Manual Standard AFM Modes User s Manual Part #00-0018-01 Issued March 2014 2014 by Anasys Instruments Inc, 325 Chapala St, Santa Barbara, CA 93101 Page 1 of 29 Table of contents Chapter 1. AFM Theory 3 1.1 Detection

More information

PRODUCT NEWS FEI LAUNCHES APREO HIGH- PERFORMANCE SEM RENISHAW OFFERS CONFOCAL RAMAN MICROSCOPE

PRODUCT NEWS FEI LAUNCHES APREO HIGH- PERFORMANCE SEM RENISHAW OFFERS CONFOCAL RAMAN MICROSCOPE 40 PRODUCT NEWS Larry Wagner, LWSN Consulting Inc. lwagner10@verizon.net FEI LAUNCHES APREO HIGH- PERFORMANCE SEM FEI (Hillsboro, Ore.) announced the new Apreo scanning electron microscope (SEM), offering

More information

First Question: Camera head. Lighting unit. Shooting stage

First Question: Camera head. Lighting unit. Shooting stage Elmo P30 Visualiser First Question: Q. Is everyone familiar with exactly what a visualiser is? A. A visualiser is effectively a camera on an arm, usually with a shooting stage and its own lighting source.

More information

Liquid Chromatography- Mass Spectrometer Manual

Liquid Chromatography- Mass Spectrometer Manual Liquid Chromatography- Mass Spectrometer Manual Joshua Willis, Elizabeth Sattely Department of Chemical Engineering Stanford University November 6, 2014 Abstract This manual will explain the LC/MS, its

More information

Index. - Registration assistant of momit Home 1 - Start of session/registration 2 - Registration of devices. - momit Home App 1.

Index. - Registration assistant of momit Home 1 - Start of session/registration 2 - Registration of devices. - momit Home App 1. Index - Registration assistant of momit Home 1 - Start of session/registration 2 - Registration of devices - momit Home App 1. Start 2. Device 2.1 Control panel a. State of operation b. Temperature b1.

More information

Checking the fluids of the cart

Checking the fluids of the cart 1 Preparing FACSAria for sterile sorting Prepare and autoclave these fluids at least 2 days in advance: Fluids Remarks Prepare 5L (for 1 period) or 10L (for 2 Each stainless-steel-sheath-tank has a periods)

More information

SYNCHRONOUS LECTURE HALL. i n s t r u c t o r g u i d e

SYNCHRONOUS LECTURE HALL. i n s t r u c t o r g u i d e SYNCHRONOUS LECTURE HALL i n s t r u c t o r g u i d e New Brunswick Digital Classroom Services Welcome! The Synchronous Lecture Halls connecting the Busch and Cook/Douglass campuses are part of President

More information

Getting started with

Getting started with Getting started with Electricity consumption monitoring single phase for homes and some smaller light commercial premises OVERVIEW: The OWL Intuition-e electricity monitoring system comprises of three

More information

TL-2900 AMMONIA & NITRATE ANALYZER DUAL CHANNEL

TL-2900 AMMONIA & NITRATE ANALYZER DUAL CHANNEL TL-2900 AMMONIA & NITRATE ANALYZER DUAL CHANNEL DATA ACQUISITION SYSTEM V.15.4 INSTRUCTION MANUAL Timberline Instruments, LLC 1880 S. Flatiron Ct., Unit I Boulder, Colorado 80301 Ph: (303) 440-8779 Fx:

More information

TT AFM LongBeach Procedures and Protocols V2.1

TT AFM LongBeach Procedures and Protocols V2.1 TT AFM LongBeach Procedures and Protocols V2.1 1. Startup Procedure 1. Turn on PC: Allow to boot to Windows. Turn on monitor. Password is afm 2. Turn on second PC that controls the video camera. 3. Turn

More information

Toshiba TLP-B2 Training Module Introduction

Toshiba TLP-B2 Training Module Introduction Toshiba TLP-B2 Training Module Introduction Toshiba TLP-B2 Introduction Welcome! Welcome to the Toshiba TLP-B2 training module offered by UT Career Services! It is highly recommended that all faculty &

More information

A6 OPERATING INSTRUCTIONS

A6 OPERATING INSTRUCTIONS Amerec s A6 control for the AX steamer is a touch screen control intended to be mounted on a wall, generally in or near the steam bath. It may be mounted directly on the wall surface or, using an optional

More information

Tender Notification for the procurement of a Scanning Electron Microscope" at IISc (Last Date for submission of tenders: 3 rd October 2018)

Tender Notification for the procurement of a Scanning Electron Microscope at IISc (Last Date for submission of tenders: 3 rd October 2018) Tender Notification for the procurement of a Scanning Electron Microscope" at IISc (Last Date for submission of tenders: 3 rd October 2018) Dear Sir/Madam, We are looking for a high-resolution scanning

More information

WAVES Cobalt Saphira. User Guide

WAVES Cobalt Saphira. User Guide WAVES Cobalt Saphira TABLE OF CONTENTS Chapter 1 Introduction... 3 1.1 Welcome... 3 1.2 Product Overview... 3 1.3 Components... 5 Chapter 2 Quick Start Guide... 6 Chapter 3 Interface and Controls... 7

More information

Training Module Style Guide

Training Module Style Guide Training Module Style Guide December 2017 Glossary of Terms Ensco when referring to the company ENSCO when referring to a rig name (e.g. ENSCO DS-10) PayZone when referring to the Ensco Intranet jackup

More information

... read The Art of Tap Tuning by Roger H. Siminoff (Hal Leonard Publishing).

... read The Art of Tap Tuning by Roger H. Siminoff (Hal Leonard Publishing). ... PO Box 2992 Atascadero, CA 93423 USA siminoff@siminoff.net www.siminoff.net 805.365.7111 Instruction Manual and Set-up Strobosoft v2.0 for tap tuning Rev: 11/25 /13 Pt# n/a StroboSoft is a software

More information

WJ-HD616K/716K Quick Reference Guide

WJ-HD616K/716K Quick Reference Guide WJ-HD616K/716K Quick Reference Guide Remote Operation Using Internet Explorer For a local operation quick guide refer to Local Quick Guide available for download http://panasonic.ca/english/customercare/operatinginstructions/query.asp

More information

VISSIM TUTORIALS This document includes tutorials that provide help in using VISSIM to accomplish the six tasks listed in the table below.

VISSIM TUTORIALS This document includes tutorials that provide help in using VISSIM to accomplish the six tasks listed in the table below. VISSIM TUTORIALS This document includes tutorials that provide help in using VISSIM to accomplish the six tasks listed in the table below. Number Title Page Number 1 Adding actuated signal control to an

More information

Stark Spectroscopy Deanna s Experimental Procedure NWU Hupp Lab Fall 2003

Stark Spectroscopy Deanna s Experimental Procedure NWU Hupp Lab Fall 2003 Stark Spectroscopy Deanna s Experimental Procedure NWU Hupp Lab Fall 2003 1. Generate mixed-valent state of compound check in 1mm cell. Ideally want Abs 1. 2. Setting up the instrument New Dewar i) Approx.

More information

LEGO MINDSTORMS PROGRAMMING CAMP. Robotics Programming 101 Camp Curriculum

LEGO MINDSTORMS PROGRAMMING CAMP. Robotics Programming 101 Camp Curriculum LEGO MINDSTORMS PROGRAMMING CAMP Robotics Programming 101 Camp Curriculum 2 Instructor Notes Every day of camp, we started with a short video showing FLL robots, real robots or something relevant to the

More information

University of Minnesota College of Science and Engineering Characterization Facility Ganesha SAXSLAB User manual

University of Minnesota College of Science and Engineering Characterization Facility Ganesha SAXSLAB User manual University of Minnesota College of Science and Engineering Characterization Facility Ganesha SAXSLAB User manual # Section Pg # 1 Planning a SAXS experiment 1 2 Initial Setup 2 3 Sample mounting and loading

More information

Vasudevan Agilent 1100 Series HPLC w/ DAD & FLD Detector (nonbuffer

Vasudevan Agilent 1100 Series HPLC w/ DAD & FLD Detector (nonbuffer Vasudevan Agilent 1100 Series HPLC w/ DAD & FLD Detector (nonbuffer solvents) Updated November 14, 2017 Instrument instructions can be found at: http://academic.bowdoin.edu/chemistry/resources/instructions.shtml

More information

Operating Procedures for RECO1 & RECO2

Operating Procedures for RECO1 & RECO2 Operating Procedures for RECO1 & RECO2 Reconstruction of Data 1) Open CT Pro 3D a. Remember, when the computer you are working on is receiving files from the scanner, it cannot reconstruct data b. You

More information

INSTALATION PROCEDURE

INSTALATION PROCEDURE INSTALLATION PROCEDURE Overview The most difficult part of an installation is in knowing where to start and the most important part is starting in the proper start. There are a few very important items

More information

Commissioning the TAMUTRAP RFQ cooler/buncher. E. Bennett, R. Burch, B. Fenker, M. Mehlman, D. Melconian, and P.D. Shidling

Commissioning the TAMUTRAP RFQ cooler/buncher. E. Bennett, R. Burch, B. Fenker, M. Mehlman, D. Melconian, and P.D. Shidling Commissioning the TAMUTRAP RFQ cooler/buncher E. Bennett, R. Burch, B. Fenker, M. Mehlman, D. Melconian, and P.D. Shidling In order to efficiently load ions into a Penning trap, the ion beam should be

More information

AEROTRAK PORTABLE AIRBORNE PARTICLE COUNTER MODEL 9110 QUICK START GUIDE

AEROTRAK PORTABLE AIRBORNE PARTICLE COUNTER MODEL 9110 QUICK START GUIDE AEROTRAK PORTABLE AIRBORNE PARTICLE COUNTER MODEL 9110 QUICK START GUIDE Thank you for purchasing a TSI AeroTrak Model 9110 Portable Airborne Particle Counter (particle counter). This guide will help you

More information

February 2007 Edition /A. Getting Started Guide for the VSX Series Version 8.5.3

February 2007 Edition /A. Getting Started Guide for the VSX Series Version 8.5.3 February 2007 Edition 3725-21286-009/A Getting Started Guide for the VSX Series Version 8.5.3 GETTING STARTED GUIDE FOR THE VSX SERIES Trademark Information Polycom, the Polycom logo design, and ViewStation

More information

Transient Stability Events & Actions

Transient Stability Events & Actions ETAP TIP No. 009 Transient Stability Events & Actions Applicable ETAP Versions: 5.5.0, 5.5.5, 5.5.6 (For lower versions, some of the descriptions and procedures below may differ in some ways) Event is

More information

ViewCommander- NVR Version 3. User s Guide

ViewCommander- NVR Version 3. User s Guide ViewCommander- NVR Version 3 User s Guide The information in this manual is subject to change without notice. Internet Video & Imaging, Inc. assumes no responsibility or liability for any errors, inaccuracies,

More information

Operation of CEBAF photoguns at average beam current > 1 ma

Operation of CEBAF photoguns at average beam current > 1 ma Operation of CEBAF photoguns at average beam current > 1 ma M. Poelker, J. Grames, P. Adderley, J. Brittian, J. Clark, J. Hansknecht, M. Stutzman Can we improve charge lifetime by merely increasing the

More information

University of Minnesota Minnesota Nano Center Standard Operating Procedure

University of Minnesota Minnesota Nano Center Standard Operating Procedure Equipment Name: CHA Evaporator Coral Name: ebevap-cha Revision Number: 12 Model: SEC 600 Revisionist: L. von Dissen Location: PAN Bay 3 Date: 04/13/2018 1 Description The CHA Evaporator is a single source

More information

This guide gives a brief description of the ims4 functions, how to use this GUI and concludes with a number of examples.

This guide gives a brief description of the ims4 functions, how to use this GUI and concludes with a number of examples. Quick Start Guide: Isomet ims Studio Isomet ims Studio v1.40 is the first release of the Windows graphic user interface for the ims4- series of 4 channel synthezisers, build level rev A and rev B. This

More information

May 2006 Edition /A. Getting Started Guide for the VSX Series Version 8.5

May 2006 Edition /A. Getting Started Guide for the VSX Series Version 8.5 May 2006 Edition 3725-21286-008/A Getting Started Guide for the VSX Series Version 8.5 GETTING STARTED GUIDE FOR THE VSX SERIES Trademark Information Polycom, the Polycom logo design, and ViewStation are

More information

invr User s Guide Rev 1.4 (Aug. 2004)

invr User s Guide Rev 1.4 (Aug. 2004) Contents Contents... 2 1. Program Installation... 4 2. Overview... 4 3. Top Level Menu... 4 3.1 Display Window... 9 3.1.1 Channel Status Indicator Area... 9 3.1.2. Quick Control Menu... 10 4. Detailed

More information

Vision Sensor Short Manual

Vision Sensor Short Manual Vision Sensor FQ Short Manual Cat. No. Z306-E1-02A Table of Contents 1. Introduction 1-1 FQ-series Vision Sensors....................................... 4 1-2 Measurement Process.........................................

More information

JS007WQK HEAVY DUTY WIRELESS REVERSING KIT 7 LCD DIGITAL QUAD RECORDING MONITOR with WATERPROOF CCD CAMERA

JS007WQK HEAVY DUTY WIRELESS REVERSING KIT 7 LCD DIGITAL QUAD RECORDING MONITOR with WATERPROOF CCD CAMERA JS007WQK HEAVY DUTY WIRELESS REVERSING KIT 7 LCD DIGITAL QUAD RECORDING MONITOR with WATERPROOF CCD CAMERA The JS007WQK is loaded with userfriendly features and is ideal for use in heavy duty vehicles.

More information

HOW TO USE THE POLYCOM REMOTE CONTROL... 2 MAKING A CALL FROM THE ADDRESS BOOK... 3 ANSWERING A CALL... 4 HANGING UP A CALL... 4 REDIALING A CALL...

HOW TO USE THE POLYCOM REMOTE CONTROL... 2 MAKING A CALL FROM THE ADDRESS BOOK... 3 ANSWERING A CALL... 4 HANGING UP A CALL... 4 REDIALING A CALL... HOW TO USE THE POLYCOM REMOTE CONTROL... 2 MAKING A CALL FROM THE ADDRESS BOOK... 3 ANSWERING A CALL... 4 HANGING UP A CALL... 4 REDIALING A CALL... 4 CAMERA CONTROLS... 5 AUDIO CONTROLS... 5 VIDEO INPUT...

More information

ME EN 363 ELEMENTARY INSTRUMENTATION Lab: Basic Lab Instruments and Data Acquisition

ME EN 363 ELEMENTARY INSTRUMENTATION Lab: Basic Lab Instruments and Data Acquisition ME EN 363 ELEMENTARY INSTRUMENTATION Lab: Basic Lab Instruments and Data Acquisition INTRODUCTION Many sensors produce continuous voltage signals. In this lab, you will learn about some common methods

More information

BitWise (V2.1 and later) includes features for determining AP240 settings and measuring the Single Ion Area.

BitWise (V2.1 and later) includes features for determining AP240 settings and measuring the Single Ion Area. BitWise. Instructions for New Features in ToF-AMS DAQ V2.1 Prepared by Joel Kimmel University of Colorado at Boulder & Aerodyne Research Inc. Last Revised 15-Jun-07 BitWise (V2.1 and later) includes features

More information

MIE 402: WORKSHOP ON DATA ACQUISITION AND SIGNAL PROCESSING Spring 2003

MIE 402: WORKSHOP ON DATA ACQUISITION AND SIGNAL PROCESSING Spring 2003 MIE 402: WORKSHOP ON DATA ACQUISITION AND SIGNAL PROCESSING Spring 2003 OBJECTIVE To become familiar with state-of-the-art digital data acquisition hardware and software. To explore common data acquisition

More information

4.9 BEAM BLANKING AND PULSING OPTIONS

4.9 BEAM BLANKING AND PULSING OPTIONS 4.9 BEAM BLANKING AND PULSING OPTIONS Beam Blanker BNC DESCRIPTION OF BLANKER CONTROLS Beam Blanker assembly Electron Gun Controls Blanker BNC: An input BNC on one of the 1⅓ CF flanges on the Flange Multiplexer

More information

Contacts: English Department Office 238 Moreland Media Services 109 Kidder

Contacts: English Department Office 238 Moreland Media Services 109 Kidder Contacts: English Department Office 238 Moreland 7-3244 Media Services 109 Kidder September 2006 Welcome...3 Starting Out...5 Unlocking the Station...5 Touch Panel Operation...5 Projector...6 Selecting

More information

PRODUCT SPECIFICATIONS

PRODUCT SPECIFICATIONS FOS Beam 150W 1 PRODUCT SPECIFICATIONS TECHNICAL SPECIFICATION Voltage: 100/240 Volt AC, 50/60 Hz. Power Consumption: 250 Watt. Light Source: 150 Watt LED, 6800K Beam angle: 2 Colors: 11 dichroic color

More information

VSX Series Getting Started Guide

VSX Series Getting Started Guide VSX Series Getting Started Guide September, 2003 Edition 3725-21286-002/A VSX SERIES GETTING STARTED GUIDE Trademark Information Patent Information Polycom and the Polycom logo design are registered trademarks

More information

Troubleshooting Guide for E-Poll Book

Troubleshooting Guide for E-Poll Book Troubleshooting Guide for E-Poll Book CHANGING USERS ON THE E-POLL BOOK Changing Users on the E-poll Book 1. Tap Return to Main button on the voter search screen. 2. Tap on the Manage Polls tab in the

More information

XNET-NVR User s Guide

XNET-NVR User s Guide XNET-NVR User s Guide Ver. 1.0 (070918) 1 of 39 Table of Contents 1. Program Installation... 4 2. Overview... 4 3. Top Level Menu... 5 3.1. Display Window... 9 3.1.1. Channel Status Indicator Area... 9

More information

GS122-2L. About the speakers:

GS122-2L. About the speakers: Dan Leighton DL Consulting Andrea Bell GS122-2L A growing number of utilities are adapting Autodesk Utility Design (AUD) as their primary design tool for electrical utilities. You will learn the basics

More information

Manual placement system MPL3100. for BGA, CSP and Fine-Pitch components

Manual placement system MPL3100. for BGA, CSP and Fine-Pitch components Manual placement system MPL3100 for BGA, CSP and Fine-Pitch components Part No: MPL3100BA1.0e Issue Date: 02/2001 You have opted for an ESSEMTEC MPL3100 pick and place system. We thank you for this decision

More information

Achieving More Efficient Data Review with OpenLAB CDS

Achieving More Efficient Data Review with OpenLAB CDS Achieving More Efficient Data Review with OpenLAB CDS Technical Note Introduction Separations have enjoyed major advances that have significantly reduced chromatographic run times. Ultra high performance

More information