Installation, Operation and Maintenance Manual. Sample Sequencer 5

Size: px
Start display at page:

Download "Installation, Operation and Maintenance Manual. Sample Sequencer 5"

Transcription

1 Installation, Operation and Maintenance Manual Sample Sequencer 5 SPD Rev. 0 11/11 Original Instructions SENTRY EQUIPMENT CORP 966 Blue Ribbon Circle N Oconomowoc, WI USA Phone: Fax: sales@sentry-equip.com

2

3 Declaration of Conformity Sentry Equipment Corp. declares under our sole responsibility that our Sample Sequencer 5 to which this declaration relates is in conformity with the following standards: Product Safety: Certified to CAN/CSA C22.2 No (Certificate # XXXXXXX) Low Voltage Directive 2006/95/EC Required Standards include: EN :2001/IEC :2001 Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1: General requirements Electromagnetic Compatibility: EMC Directive 2004/108/EC: Required Standards include: EN :2006 / IEC 61326:1997 w\a1:1998 & A2:2001 Electrical equipment for measurement, control, and laboratory use EMC requirements - Part 1: General requirements Immunity: Required tests include: IEC ESD IEC Radiated Susceptibility IEC EFT IEC Surges IEC RF Cond. IEC Magnetic Susceptibility IEC Volt dips, interrupts Emissions: Required tests include: EN Class A limits EN Harmonic EN Voltage Fluctuation (Flicker) FCC Compliance: FCC Part & Canadian Compliance: ICES 003 RoHS Compliance: Directive 2002/95/EC This product complies with the RoHS Directive. Type of Equipment: Electrical Equipment for Measurement, Control, and Laboratory Use Part Number: F Year CE Mark Affixed to Product: 2011 SPD i

4 TABLE OF CONTENTS 1 SPECIFICATIONS SAFETY PRECAUTIONS USE OF HAZARD INFORMATION DEFINITIONS QUICK START GUIDE ELECTRICAL CONNECTIONS CONTROL SCREEN STARTING AND STOPPING THE SAMPLE SEQUENCER WIDGETS SCREEN SETTING THE SEQUENCE PLACING SAMPLES IN SAMPLE/BYPASS SETTING SAMPLE TIMERS SETTING THE MODE EDITING THE SAMPLE NAMES EDITING ANALYZER NAMES GENERAL DESCRIPTION MODES OF OPERATION INPUTS AND OUTPUTS OPTIONAL MANIFOLD SAMPLING VALVE OPERATION DESCRIPTION OF CONTROLS SCREEN DESCRIPTIONS OPERATING MODES VALVE OUTPUTS INSTALLATION UNPACKING MOUNTING ELECTRICAL CONNECTIONS SEQUENCER BOARD CONNECTIONS ALTERNATE SEQUENCER BOARD CONNECTIONS STARTING UP THE SAMPLE SEQUENCER MAINTENANCE CLEANING RECHARGEABLE BATTERY DISASSEMBLY CALIBRATION TROUBLESHOOTING GUIDE FACTORY ASSISTANCE SPARE PARTS LIST ii SPD

5 11 ACCESSORIES WARRANTY APPENDIX A - RECOMMENDED PIPING APPENDIX B USING A SEQUENCER WITH HACH SERIES 5000 ANALYZER APPENDIX C MODBUS REGISTERS SPD iii

6

7 1 SPECIFICATIONS Specifications are subject to change without notice. Applications: The Sample Sequencer is designed to support sample stream switching in the batch or continuous modes. Maximum Number of Samples: 8 Maximum Number of Analyzers: 4 Power: Input Voltage: Vac Frequency Range: 50/60 Hz Input Current: 1.2 A / 115 V 0.72 A / 230 V Inputs from Analyzer: Four analog inputs: 4-20 ma (100 load) Four digital inputs for end-of-analysis indication. Four digital inputs for calibration indication. Four digital inputs for analyzer system alarm indication. Outputs: 10base-T Ethernet supports MODBUS TCP/IP communication and embedded web capability. Valve outputs (12 VDC) to control a maximum of eight solenoid valves for sample stream switching. Individual outputs capable of handle up to 2A solenoid load. Total solenoid load current not to exceed 12VDC (Please consult factory if individual solenoid load is greater than 12VDC). 12 digital outputs for point number indication and replicated Analyzer Alarms. 0-40V,60mA sinking outputs Current Outputs: Eight 4 20 ma outputs. Maximum Load: 550 Ω Nominal Accuracy: Analog Inputs: ± 0.2% full scale Analog Outputs: ± 0.2% full scale Resolution: Analog Inputs: 10 µa/lsb Analog Outputs: 10 µa/lsb Data Logging: Records stored on removable SD card. Storage limations based on size of SD card. SPD

8 User Interface: 4.3 color touch screen, 65,000 colors. Enclosure: 304 Stainless steel, 16 GA. Designed to meet NEMA 4X (indoor), IPX0 Dimensions: 7.00 (177.8 mm) H x 8.00 (203.2 mm) W x 7.00 (177.8 mm) D Mounting: Panel or surface. Ambient Temperature: C ( F); Relative Humidity 5-95% Environmental: Indoor use only Operating Altitude: m ( ft.) (Consult factory for other altitudes.) Mains supply fluctuation: -10% and +10% Transient overvoltage: 1500Vac Pollution degree: 2 Agency Approvals: Shipping Weight: 15 lbs (6.80 kg) 2 SPD

9 2 SAFETY PRECAUTIONS Please read the entire manual before attempting to unpack, set up or operate this product. Pay careful attention to all warnings, cautions and notes. Failure to do so could result in serious personal injury or equipment damage. 2.1 Use of Hazard Information If multiple hazards exist, the signal word corresponding to the greatest hazard shall be used. 2.2 Definitions Indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury. Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury. Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury. CAUTION used without the safety alert symbol indicates a potentially hazardous situation which, if not avoided, may result in property damage. NOTE Information that requires special emphasis SHALL This word understood to be mandatory SHOULD This word understood to be advisory SPD

10 Indique une situation imminente et dangereuse qui doit être évité immédiatement afin d éviter tout risque de blessures sérieuses et qui pourrait même entraîner la mort. AVERTISSEMENT Indique une situation potentiellement dangereuse et qui doit être évité immédiatement afin d éviter tout risque de blessures sérieuses et qui pourrait même entraîner la mort. PRÉCAUTION Indique une situation potentiellement dangereuse et qui doit être évité immédiatement afin d éviter tout risque de blessures mineures. PRÉCAUTION ATTENTION une utilisation sans le symbole de sécurité indique une situation potentiellement dangereuse et que si elle n est pas évité, pourrait entrainer des dommages à la propriété. NOTE Informations requérant une attention spéciale DOIT Ce mot indique une obligation POURRAIT Ce mot indique un avertissement 4 SPD

11 3 QUICK START GUIDE The following guide is intended to help you get started using the Sentry Sample Sequencer. This guide does not include detailed explanations and should be used in conjunction with the remainder of this manual to build a full understanding of the Sample Sequencer. 3.1 Electrical Connections Connect all electrical connections necessary for the application. If the analyzer being shared is a batch type analyzer, an end of analysis signal must be connected to the Sample Sequencer End of Batch input. Without this signal, the Sample Sequencer will never advance to the next sample. 3.2 Control Screen With the display showing the Operation Screen (see Figure 6), touch anywhere on the screen excluding the left and right arrow areas. If prompted, enter password. This will take you to the Control Screen (see Figure 8). 3.3 Starting and Stopping the Sample Sequencer From the Control Screen (see Figure 8), you can start and stop any active sequence by touching the red indicator on the screen. Once touched, the indicator will change and the sequencer state will update. 3.4 Widgets Screen From the Control Screen, touch the wrench button in the lower right corner of the screen. This will take you to the first Widgets Screen (see Figure 9). 3.5 Setting the Sequence Navigate to the first Widgets Screen. Touch the Sequences widget to take you to the Sequences Screen (see Figure 29). Using the buttons on the screen, create the sequences desired. Changes are saved by touching the save button in the lower right hand corner. To exit without saving changes, touch the escape button in the lower left hand corner. 3.6 Placing Samples in Sample/Bypass Navigate to the first Widgets Screen. Touch the Bypass widget to access the Bypass Line Screen (see Figure 12). Touch the indicator light next to the desired sample line you wish to switch. The state of the sample line will update once indicator has been touched. The Sample state indicates the sample point will be sampled. The Bypass state indicates the sample point will be bypassed the next time its position is up in the sequence. Changes are saved by touching the save button in the lower right hand corner. To exit without saving changes, touch the escape button in the lower left hand corner. 3.7 Setting Sample Timers Navigate to the first Widgets Screen. Touch the Flush Times or Sample Times widget to access the appropriate screen (see Figures 10 and 11). Touching the time on the SPD

12 sample line you would like to update will take you to an Enter Time screen. Enter the time desired. When finished, touch the enter button in the lower right corner. This will bring you back to the times screen and the time you just entered will be updated. Do this for each time you would like to update. Save changes by touching the save button in the lower right hand corner of the screen. To exit without saving changes, touch the escape button in the lower left hand corner. 3.8 Setting the Mode Navigate to the first Widgets Screen. Touch the Mode widget to access the Mode Screen (see Figure 13). Touch an indicator to toggle the sequence between the two modes, Batch or Continuous. Make desired changes. Save changes by touching the save button in the lower right hand corner. To exit without saving changes, touch the escape button in the lower left hand corner. 3.9 Editing the Sample Names Navigate to the first Widgets Screen. Touch the Sample Names widget to access the Sample Names Screen (see Figure 30). Touch the sample name you would like to change to bring up the Sample Names Keypad Screen. When finished, touch the enter button in the lower right corner. This will update the Sample Names Screen with the name you just entered. Make any additional name changes. Save changes by touching the save button in the lower right hand corner. To exit without saving changes, touch the escape button in the lower left hand corner Editing Analyzer Names Navigate to the first Widgets Screen. Touch the Analyzer Names widget to access the Analyzer Names Screen (see Figure 31). Touch the analyzer name you would like to change to bring up the Analyze Names Keypad Screen (see Figure 32). When finished, touch the enter button in the lower right corner. This will update the Analyzer Names Screen with the name you just entered. Make any additional name changes. Save changes by touching the save button in the lower right hand corner. To exit without saving changes, touch the escape button in the lower left hand corner. 6 SPD

13 4 GENERAL DESCRIPTION The Sentry Sample Sequencer is a microprocessor-based instrument capable of electronically switching up to eight sample streams one at a time to an analyzer and of providing signals to recorders. If a total of eight sample streams are not needed for one process, the remaining points may be assigned to additional analyzers, up to four in total. A Sentry manifold sampling valve specially designed to obtain optimum representative samples is available for sample stream switching with the Sample Sequencer. The Sample Sequencer reduces the number of required analyzers while maintaining equivalent data accuracy. Reliable sample analysis comparison is obtained by using a Sample Sequencer with one dedicated analyzer. In addition, significant operating cost savings are obtained by not having to install, maintain, and service multiple analyzers. The Sample Sequencer is capable of operating in the continuous mode for continuous analyzers or in the batch mode for batch analyzers. In addition, the Sample Sequencer is capable of measuring the analyzer analog output signal and converting the analog signal to digital data to be sent out using MODBUS TCP/IP or Embedded web using the Ethernet communication port. Using the MODBUS capability, one can connect to a PC or system network via the Ethernet cable to one or more Sequencers for remote monitoring, control, and data acquisition. Track and Hold outputs are available to provide independent outputs for each sample point. Digital outputs are also available to provide channel number indication as well as replicating analyzer alarms. Figures 1-3 illustrate a typical system arrangement for sample stream switching with the Sample Sequencer. Customer-supplied equipment is shown for clarification. 4.1 Modes of Operation The Sample Sequencer has two programmable modes of operation: BATCH Used for analyzers that take a sample "bite" (or batch), produce an analytical result, and then provide a contact signal output, in typically 1-20 minutes CONTINUOUS Used for analyzers that continuously produce an analytical result. NOTE If using the Sample Sequencer with a Hach analyzer, see Appendix B. SPD

14 Figure 1: Sample Sequencer with Continuous Analyzer 8 SPD

15 Figure 2: Sample Sequencer with Batch Analyzer SPD

16 Figure 3: Sample Sequencer with two Analyzers 10 SPD

17 4.2 Inputs and Outputs A variety of inputs and outputs are available on the Sample Sequencer Analog Inputs: Four analog inputs are located on the Sample Sequencer board for reading a 4 20 ma signal from up to four different analyzers. The Sample Sequencer converts this signal into Engineering Units for display and outputs the same signal for the Track and Hold analog outputs Analyzer Alarm Inputs: Four analyzer alarm inputs are located on the Sample Sequencer board. These inputs are typically used to indicate a loss of sample flow to the analyzer. An analyzer alarm will cause the analyzer description to flash on the operation screen. An analyzer alarm will also cause the digital alarm output for that sequencer to be set. When the Sample Sequencer is used in the Continuous mode, an analyzer alarm will immediately close the active sample valve and open the next sample valve. That sample valve will remain open until the analyzer alarm clears, one minute elapses, or if the sum of the sample and flush times is less than one minute when the timer expires. After one minute, the Sample Sequencer will open the next sample valve. The Sample Sequencer will continue trying to establish flow every minute until the analyzer alarm clears. When the Sample Sequencer is used in the Batch mode, an analyzer alarm will not immediately close the active sample valve. Instead, when the analyzer alarm is received, the Sample Sequencer will start a timeout timer. If the Sample Sequencer does not receive an End of Batch indication from the analyzer or the alarm does not clear before that timeout timer, the Sample Sequencer will close the current sample valve and open the next valve in the sequence. If the End of Batch indication is received before the timeout timer, the Sample Sequencer assumes the alarm is not a loss of flow and continues with normal operation. The operator can adjust the timeout timer by setting each sample s Sample Timer Calibration Hold Inputs: Four calibration hold inputs are located on the Sample Sequencer board. These inputs are used to indicate that the analyzer is performing a calibration. The calibration hold inputs do not perform a function when the Sample Sequencer is in the Batch mode. In the Continuous mode, the Sample Sequencer will wait at the active sample with the valve open when a calibration indication is received. Once the calibration indication clears, the Sample Sequencer will resume operation entering the Flush state for the sample that was active before the calibration hold indication was received End of Batch Inputs: Four End of Batch inputs are located on the Sample Sequencer board. These inputs are used to indicate when a batch type analyzer is finished taking a measurement. Without this input connected, the Sample Sequencer will never advance to the next sample when the Sample Sequencer is in the Batch mode. The End of Batch inputs are not used in the Continuous mode. SPD

18 4.2.5 Valve Outputs: Eight valve outputs are located on the Sample Sequencer board. These outputs supply 12 VDC to operate the sample valves. The eight valves may be divided between four analyzers in any combination Track and Hold Outputs: Eight analog outputs are provided on the Sample Sequencer board. These outputs are 4-20 ma outputs. When a valid reading is taken from the analyzer for a given sample, the Sample Sequencer updates ( Tracks ) the corresponding signal on the appropriate output. When the sample is not currently being analyzed, the Sample Sequencer Holds the corresponding signal on the appropriate output at its last known value Digital Outputs: A total of twelve digital outputs are available on the Sample Sequencer board. Four of these outputs are used to replicate the four-analyzer alarm inputs while the remaining eight are used for sample point indication. An on output indicates the point is currently being sampled and the analyzer reading corresponds to that sample point Ethernet Communications: The Sample Sequencer is capable of connecting to a network for communicating to a PC or control system via MODBUS TCP/IP. In addition, the sequencer provides the option to view settings via embedded web. For more information on these features, refer to Appendix D. 4.3 Optional Manifold Sampling Valve Although the Sample Sequencer may be used to control most manifold valve arrangements, the Sentry Manifold Sample Valve offers a convenient optional accessory to obtain a representative sample. The patented design incorporates block and bleed design and provides continuously flowing samples with no areas for cross contamination to occur. The valve is available in either four or eight sample line models. Contact Sentry for more information about this special sampling valve. 12 SPD

19 5 OPERATION To ensure the protection provided by this equipment is not impaired, this equipment must not be installed or used in any manner other than that which is specified in this manual. AVERTISSEMENT Afin de s assurer que la protection prévue par cette pièce d équipement n est pas défaillante, cet équipement ne doit pas être installé ou utilisé de manière autre que spécifié dans le manuel. 5.1 Description of Controls The following table and figure describe controls on the Sample Sequencer touch screen display: Table 1: Control Descriptions DESCRIPTION FUNCTION Widgets Enter button Escape button Alpha keypad Numeric keypad Save button Touching a widget will take you to the corresponding function screen Various screens will have an enter button in the lower right corner. Touching this button will take the value you have entered and update the previous screen with that value. Various screens will have an escape button in the lower left corner. Touching this button will return you to the previous screen and any values will not be saved. Various screens will have an alpha keypad for entering data. Use this keypad to enter necessary data. Press the Enter button when done using keypad to return to the previous screen with the data you have entered. Various screens will have a numeric keypad for entering data. Use this keypad to enter necessary data. Press the Enter button when done using keypad to return to the previous screen with the data you have entered. Various screens will have a save button in the lower right corner. Touching this button will save entered values and return you to the previous screen. SPD

20 Figure 4: Sample Sequencer Front Panel 14 SPD

21 5.2 Screen Descriptions Figure 5: Screen Navigation Operation Password Control Widget Flush Times Sample Times Bypass Mode Sequencer Active Password Date/Time Communications IO Data L i Sequences Sample Names Analog In - View Analog In - Cal Analyzer Names Units Preference Info Analog Out - Test Analog Out - Cal Digital In - View Digital In - Test Sol. Valves - Test Analog In - Scale Touch Screen - Cal SPD

22 5.2.1 Operation Screen The Operation Screen is the main screen on the Sample Sequencer. This screen contains all the information indicating the state of the Sample Sequencer. No system changes can be made on this screen. Sequence control is done on the Control Screen. Figure 6: Operation Screen Status Open Sample Valve Name Sequence Sample Line Names Arrow for Next Sequencer Analyzer Description Most Recent Active Value Depending on the settings, there are one or two sequences displayed per Operation Screen. Touching the left or right arrows on the screen will take you to additional sequences. Sequence sample line names: Displayed to the left of the analyzer with the appropriate sample name indicated in the sample line tubing. Status: In Continuous mode, this displays the state of the Sample Sequencer, time remaining, and measurement value. The three available states are FLUSH, SAMPLE, and OFF. The time displayed is the time remaining in the current state. The measurement value is only displayed when the Sample Sequencer is in the SAMPLE state. In Batch mode, this indicates the state of the Sequencer, OFF or SAMPLE. Open Sample Valve Name: Displays the sample description for the currently open valve. Analyzer Description (Continuous and Batch modes): User definable field that describes the analyzer being sequenced. The maximum number of characters is four. When an analyzer alarm or calibration signal is received, this description will flash. Most Recent Active Value: In Continuous mode, the current measurement value is displayed. In Batch mode, this displays the description of the previously sampled point and its measurement value. The measurement value on the analyzer corresponds to this sample point. 16 SPD

23 5.2.2 Password Entry Screen Figure 7: Password Entry Screen If password protection is enabled, the Password Entry Screen will be displayed when the user attempts to enter the Control Screen. Enter the correct password to gain access to change the setting on the Sample Sequencer. The password can be changed by touching the password widget on the Widget Screen. If you do not know the password, please contact Sentry Equipment Corp Control Screen The Control Screen is used for changing the state of the sequences. This screen contains the name and the state of each active sequence. Figure 8: Control Screen Depending on your settings, there can be one or two sequences displayed on the Control Screen. Touching the indicator next to each sequence will turn the sequence on or off. Touching the wrench button in the lower right hand corner of the screen will take you to the Widgets. SPD

24 5.2.4 Widgets Figure 9: Widgets The Widgets are reached by touching the wrench button in the lower right hand corner of the Control Screen. Selecting a widget will take you to its corresponding screen. Selecting the cancel button in the lower left corner will return you to the previous screen. Selecting the arrow button in the lower right corner will take you to additional widgets Flush Times Use this widget to update Flush Times. In Continuous mode, Flush Time is the amount of time the analyzer is flushed out before the analyzer reading is considered valid. The valve for the sample point is open during both the Sample Time and the Flush Time. In Batch mode, Flush Time is not used. Figure 10: Flush Times Screen To change Flush Time for a sample, touch the time that you would like to change. The Enter Time Screen will be shown. Enter the time desired using this screen; when finished, touch the enter button to update the value in the Flush Times Screen, or press the escape button to return to the Flush Times Screen without entering the value. Time values are limited to 59 minutes and 59 seconds. Once you have updated all the Flush Times, press the save button to save all times on the screen, or touch the escape button to leave the screen without saving the updated values. 18 SPD

25 5.2.6 Sample Times Use this widget to update Sample Times. Sample Time is the amount of time the analyzer is allowed to analyze the sample while the reading is considered valid. The Track and Hold value and Point Identification are active during the Sample Time. The valve for this sample point is open during both the Sample Time and the Flush Time. In Batch mode, Sample Time serves as a timeout timer that operates only when the analyzer is in alarm. While in alarm, if an End-of-Batch signal is not received before Sample Time times out, the Sample Sequencer will close the current sample point and proceed with the next sample point. Since the Analyzer Alarm input may serve to indicate more than a loss of sample flow, Sample Time should be set to give the analyzer sufficient time to attempt to complete an analysis and output an End-of-Batch indication. Figure 11: Sample Times Screen To change a Sample Time for a sample, touch the time that you would like to change. The Enter Time Screen will appear. Enter the time desired using this screen; when finished, touch the enter button to enter the value in the Sample Times Screen, or press the escape button to return to the Sample Times screen without entering the value. Time values are limited to 59 minutes and 59 seconds. Once you have updated all the Sample Times, press the save button to save all times on the screen, or press the escape button to leave the screen without saving updated values. SPD

26 5.2.7 Bypass Sample streams may be temporarily skipped in the sequence by placing the sample point number in Bypass. If the Sample Sequencer is currently on a sample point that has been placed in Bypass, the Sample Sequencer will immediately skip to the next point in the sequence. Occasionally, it is necessary to hold an analysis on one sample point. To accomplish this, simply place all the undesired sample points in Bypass. Sample points are placed in either Sample or Bypass by using the Bypass Screen. Figure 12: Bypass Screen Touch the indicator for the sample line to be switched. This will toggle the sample line between Sample and Bypass. When a point that is placed in Bypass is next to be sequenced, the point will be skipped and the next point that is in Sample will be sequenced. Press the save button to save the changes made on this screen. To return to the previous screen without saving, press the escape button Mode The Sample Sequencer is capable of operating in two different modes, depending on the type of analyzer being used. Continuous mode is for continuous output analyzers such as sodium, conductivity, ph, O 2, some colorimetric, and many other analyzers. Batch mode is for analyzers that take a "bite" or batch of sample and then spend some time analyzing it before outputting a new fixed analog signal. These typically include silica, phosphate, and other analyzers. (Consult specific analyzer documentation to determine the type of analyzer.) The mode is changed from the Mode Screen. 20 SPD

27 Figure 13: Mode Screen Touch the indicator next to the analyzer to be changed. Touching the indicator toggles the analyzer between Continuous and Batch modes. Press the save button to save the changes made on this screen. To return to the previous screen without saving, press the escape button Sequences Active The Sample Sequencer is capable of sequencing up to four different analyzers. Select the number of analyzers to be sequenced on the Sample Sequencer. Figure 14: Sequences Active Screen Touch the number of sequences active using this screen. Press the save button to save the changes made on this screen. To return to the previous screen without saving, press the escape button. SPD

28 Password Setup If desired, the Sample Sequencer can be password protected. Password settings are changed on the Password Setup Screen Figure 15: Password Setup Screen Select the blue button to toggle between Enabled and Disabled for password protection. When Enabled is selected, the current password will be displayed. To change the password, simply touch the password, and a keypad will appear. Use the keypad to enter a new password. Select the enter button when finished to return to the Password Setup Screen. Press the save button to save the changes made on this screen. To return to the previous screen without saving, press the escape button Clock The Sample Sequencer contains a real time clock that provides the time and date stamp for the data logger. The real time clock is set from the Set Date Screen. Figure 16: Set Date Screen The top time displayed is the current time on the system. To change the time, change the Set Time to: settings by pressing the area you wish to change. For example, to change the month, touch the month area of Set Time to. Once you have the correct time in the Set Time to area, 22 SPD

29 press the save button to update the time. Touching the escape button will return to the previous screen without saving Communications The Sample Sequencer has the capability to communicate via Modbus TCP/IP or embedded web using the 10base-T Ethernet port. To set up these options, enter the correct IP, Subnet, and Gateway values. If using Modbus, set desired address. Figure 17: Communications Screen IO Use the IO widget to view and update input and output settings for the Sample Sequencer. Analog and digital inputs and outputs can be viewed, tested, and calibrated from this screen. Also, the analog input scale can be set, the solenoid valve outputs can be tested, and the touch screen can be calibrated. Figure 18: I/O Screen Select the IO function you would like to perform by touching the black button to the right of the description. SPD

30 IO Analog inputs -View To view analog inputs, use the Analog Inputs View Screen. Figure 19: Analog Inputs - View Screen This screen displays the 4-20mA values for the analog inputs. Press the escape button when finished viewing to return to the previous screen. No changes can be made from this screen IO Analog inputs - Cal To calibrate analog inputs, use the Analog Inputs Cal Screen. Figure 20: Analog Inputs - Cal Screen Use this screen when you wish to calibrate the 4-20mA values for the analog inputs. To calibrate analog inputs: 1. Supply 4.00 ma from a calibrated source to the desired input. 2. Touch the Press to Cal area of the screen below the input you wish to calibrate. 24 SPD

31 3. The status area will indicate that the input is calibrating. 4. When finished, the input will read Enter ma. 5. Supply 20 ma from a calibrated source to the input. 6. Touch the Press to Cal area to begin. 7. The status area will then indicate that the input is calibrating. 8. Once the input has finished calibrating, the status area will be blank and calibration will be complete. 9. Repeat process for any additional analog inputs you wish to calibrate. 10. When finished, touch the save button to save all calibration values, or the escape button to return to the previous screen without saving IO Analog outputs - Test To test analog outputs, use the Analog Outputs Test screen. Figure 21: Analog Outputs - Test Screen 1. Use this screen when you wish to test the 4-20mA values for the analog outputs. To test an analog output: 2. Choose the output you wish to test. 3. Connect a calibrated multimeter to the desired output. By default, the output should be reading 4 ma. 4. To output a value other than 4 ma, use the up and down arrows next to the output you are testing. This will step the output between 4 and 20 ma. The 5 available output levels are 0%, 25%, 50%, 75%, and 100% of full scale. 5. Verify reading with your multimeter. 6. When finished, touch the escape button to return to the previous screen. SPD

32 IO Analog Outputs - Cal To calibrate analog outputs, use the Analog Outputs Cal Screen. Figure 22: Analog Outputs - Cal Screen Use this screen when you wish to calibrate the 4-20mA values for the analog outputs. To calibrate analog outputs: 1. Connect a calibrated multimeter to the desired output. 2. As displayed on the screen, the output should be reading 20 ma. If the multimeter isn t reading 20.0 ma, use the up and down arrows on the screen until the multimeter is reading 20.0 ma. 3. After you have set the 20 ma output, touch the 20.0 ma area on the screen. This will update the output to 4 ma. If the multimeter isn t reading 4.0 ma, use the up and down arrows on the screen until the multimeter is reading 4.0 ma. 4. After you have set the 4 ma output, touch the 4.0 ma area on the screen. 5. Repeat the above steps for any other outputs you would like to calibrate. 6. When finished, touch the save button to save all calibration values, or the escape button to return to the previous screen without saving. 26 SPD

33 IO Digital Inputs - View To view digital inputs, use the Digital Inputs View Screen. Figure 23: Digital Inputs - View Screen Use this screen when you wish to view the digital inputs. When the indicator is illuminated, the digital input is on. When finished, use the escape button to return to the previous screen. No changes can be made from this screen IO Digital Outputs - Test To test digital outputs, use the Digital Outputs Test Screen. Figure 24: Digital Outputs - Test Screen Use this screen when you wish to test the digital outputs. Touching the desired output will toggle the output on/off. When the indicator is illuminated, the digital output is on. When finished, use the escape button to return to the previous screen. SPD

34 IO Solenoid Valves - Test To test solenoid valves, use the Solenoid Valve Test Screen. Figure 25: Solenoid Valve Outputs - Test Screen Use this screen when you wish to test the solenoid valve outputs. Touching the desired output will toggle the output on/off. When the indicator is illuminated, the solenoid valve output is on. When finished, use the escape button to return to the previous screen IO Analog Input Scale To set the scale for the analog inputs, use the Analog Input Scale Screen. This screen allows the analog input value to be displayed in engineering units. To display engineering units, the scale should be set to the same range as the analyzer. Figure 26: Analog Input Scale Screen When you wish to change the scale of an analog input: 1. Touch the scale setting you wish to change. This will bring up a numeric keypad. 2. Enter the desired value. 3. When finished, touch the enter button in the lower right hand corner of the numeric keypad screen. This will enter your value in the appropriate field on the Analog Input Scale Screen. 28 SPD

35 4. Repeat for any other scale values you would like to adjust. 5. When the indicator is illuminated, the solenoid valve output is on. 6. When finished, touch the save button to save all scale values, or touch the escape button to return to the previous screen without saving IO Touchscreen Cal To calibrate the touchscreen, use the Touchscreen - Cal Screen. Figure 27: Touchscreen - Cal Screen Use this screen when you wish to recalibrate the touchscreen. This screen also displays each time the unit is powered up. Follow the prompts on the screen to perform the touchscreen calibration. For better results, it is recommended to use a stylus for this calibration. This calibration is timed; failure to follow the prompts in a timely manner will return you to the main screen without saving the calibration values. The screen will indicate the calibration is complete if the prompts were followed correctly. SPD

36 Data Logging To view logged data, use the Data Logging Screen. Figure 28: Data logging Screen This screen displays data that has been logged to the SD card. Use the arrows to scroll through the data. The data will include the sample name, a time stamp, and the value. The Date Stamp and Time Stamp are taken the moment the data is logged. To exit, use the escape key. No changes can be made from this screen Sequences To sequence the sample lines for the appropriate analyzer, use the Sequences Screen. Figure 29: Sequences Screen The order that the Sample Sequencer opens sample points can be customized. The Sequences Screen displays the active sequences. Seq x corresponds to the analyzer to which the samples are being sent. To select the desired sequence, touch the Seq x area of the sequence you wish to change. Any unused sample line 30 SPD

37 can be sequenced. Each sequence is limited to a total of 16 sequence points. At the end of the sequence, the sequence will wrap to the first point. Not all 16 points need to be filled. When finished, touch the save button to save all sequence values, or touch the escape button to return to the previous screen without saving Sample Names To name each sample line, use the Sample Names Screen. The factory default for each sample point description is in the form of SAMPLE x NAME. Figure 30: Sample Names Screen To name a sample line: 1. Select the sample line you wish to update by touching the corresponding sample name area. 2. An alphanumeric keypad will appear where the sample name can be entered. 3. When you have completed the sample name, touch the enter button in the lower right hand corner. This will enter the sample name you have created on the Sample Names Screen. 4. Repeat the above process for any sample name you would like to change. 5. When finished, touch the save button to save all sample line names, or touch the escape button to return to the previous screen without saving. SPD

38 Analyzer Names To name each analyzer, use the Analyzer Names Screen. Figure 31: Analyzer Names Screen Figure 32: Analyzer Names Keypad To update analyzer names: 7. Touch the name area of the analyzer you wish to update. 8. Enter the analyzer name using the keypad that appears. 9. When you have completed the analyzer name, touch the enter button in the lower right hand corner. This will enter the analyzer name you have created on the Analyzer Names Screen. 10. Repeat this process for each analyzer name you would like to change. 11. When finished, touch the save button to save all analyzer names, or touch the escape button to return to the previous screen without saving. 32 SPD

39 Units To set the units to be displayed for each sequence, use the Units Screen. Figure 33: Units Screen The units for the active sequences are displayed on this screen. Units x corresponds to the units for sequence x. To update the units for an analyzer: 12. Touch the unit area you wish to update. 13. Use the keypad that appears to enter the unit name. 14. When you have completed the unit name, touch the enter button in the lower right hand corner. This will enter the unit name you have created on the Units Screen. 15. Repeat this process for each unit name you would like to change. 16. When finished, touch the save button to save all unit names, or touch the escape button to return to the previous screen without saving. SPD

40 Preferences To set the preferences for the Sample Sequencer, use the Preferences Screen. Figure 34: Preferences Screen From this screen you can select the number of sequences you would like displayed on the main screen. When finished, touch the save button to save all preferences, or touch the escape button to return to the previous screen without saving Information To view information about the Sample Sequencer, use the Information Screen. Figure 35: Information Screen This screen shows revision and contact information for the Sample Sequencer. When finished, touch the escape button to return to the previous screen. 34 SPD

41 5.3 Operating Modes Continuous Mode This mode is for continuous output analyzers such as sodium, conductivity, ph, O 2, some colorimetric, and many others. Timing Considerations for the Continuous Mode Figure 36 illustrates the timing diagram for the continuous mode. The following considerations should be made when determining the timing for your system. Analyzer Alarm: An analyzer alarm will immediately close the active sample valve and open the next sample valve. That sample valve will remain open until the analyzer alarm clears or one minute elapses. After one minute, the Sample Sequencer will open the next sample valve. The Sample Sequencer will continue trying to establish flow every minute until the analyzer alarm clears. Calibration Hold: A calibration hold signal will cause the Sample Sequencer to wait at the active sample with the valve open. Once the calibration indication clears, the Sample Sequencer will resume operation entering the Flush state for the sample that was active before the calibration hold indication was received. Flush Time: After switching to a successive stream, there will be a settling time for the analyzer to stabilize to the new sample. The analyzer analog output signal is ignored during this time. The necessary Flush time depends on the valve manifold design, sample line length from the manifold arrangement, sample velocity and analyzer response time. Sample Time: Sample Time is the time period during which a valid analyzer signal is read. Analog outputs are updated and point identification relay contacts are closed during this time. The necessary Sample time should be set long enough to recognize stable readings and to provide recording devices time to capture the data. SPD

42 Figure 36: Continuous Mode Timing Diagram 36 SPD

43 5.3.2 Batch Mode This mode is for analyzers that take a "bite" or batch of sample and then spend some time analyzing it before outputting a new fixed analog signal. Typical types are reagent additive, chromatographs, etc. When these analyzers have completed an analysis there is just one output value until the analyzer's next cycle is completed. An "End-of-Batch" contact closure signal must be provided at the conclusion of a batch by the batch type analyzers. The Hach Series 5000 analyzer operates in this manner. See Appendix B for instructions about connecting a Hach 5000 analyzer to the Sample Sequencer. Timing Considerations for the Batch Mode Figure 37 illustrates the timing diagram for the batch mode, showing how each sample stream is switched when the analyzer outputs an end-of-batch signal to the Sample Sequencer. If the Sample Sequencer does not receive an end-of-batch signal, the Sample Sequencer will never advance to the next sample. Upon receiving an end-of-batch signal from the analyzer, the Sample Sequencer reads the analyzer value, updates the Track and Hold output, and closes the point identification relay. Note: The updated Track and Hold output and point identification relay are for the previous sample point. The open valve number will never match the closed relay point identification in the batch mode. The following considerations should be made when determining the timing for your batch mode system: End-of-Batch: The Sample Sequencer receives this signal when the analysis is completed and the analyzer is ready to begin a new analysis. Analyzer Alarm: The Sample Sequencer receives this signal when the analyzer is in alarm, such as a loss of sample flow condition. Once the analyzer alarm signal is received the Sample Sequencer will start the Sample timer, and replicate the analyzer alarm for use by external recorders and control systems. Sample Time: The Sample time in the batch mode is a timeout timer that operates only when the analyzer is in alarm. While in alarm, if an End-of-Batch signal is not received before the Sample time times out, the Sample Sequencer will close the current sample point and proceed with the next sample point. Since the Analyzer Alarm input may serve to indicate more than a loss of sample flow, the Sample time should be set to give the analyzer sufficient time to attempt to complete an analysis and output an End-of-Batch indication. SPD

44 Figure 37: Batch Mode Timing Diagram 5.4 Valve Outputs The Sample Sequencer switches sample streams to an analyzer using the valve outputs on the board. These outputs provide 12 VDC to control solenoid valves. Sentry offers a manifold sample valve for use with the Sample Sequencer. With additional hardware, other solenoid valve operating voltages may be used. 38 SPD

45 6 INSTALLATION This instrument should be installed by qualified technical personnel to ensure adherence to all applicable electrical codes. AVERTISSEMENT Cet instrument doit être installé par un technicien certifié afin d assurer que toutes les normes au code électrique sont respectées. To ensure the protection provided by this equipment is not impaired, this equipment must not be installed or used in any manner other than that which is specified in this manual. AVERTISSEMENT Afin de s assurer que la protection prévue par cette pièce d équipement n est pas défaillante, cet équipement ne doit pas être installé ou utilisé de manière autre que spécifié dans le manuel. 6.1 Unpacking After carefully removing the Sample Sequencer from its shipping carton, inspect it for any damage. Report damages to the shipper at once. 6.2 Mounting It is recommended the Sample Sequencer be mounted close to the analyzer to permit viewing of both devices at the same time. This allows the operator to control the Sample Sequencer while viewing the analyzer. The Sample Sequencer can mount easily to a panel or wall. All mounting methods must be capable of supporting at least 24 pounds (10.9 kg). SPD

46 Figure 38: Enclosure Dimensions: Bottom View Figure 39: Enclosure Dimensions: Side View 40 SPD

47 6.2.1 Wall Mounting 1. Use the dimension information shown in Figure 40 to mark positions on the wall for four screw holes. Position the device such that a minimum of two (2) mounting screws mount to a stud. 2. Drill pilot holes in the marked positions for four ¼ -inch (M6) screws. 3. Use four ¼-inch (M6), hex-head or pan-head screws to secure the device to the wall. Figure 40: Mounting Dimensions Figure 41: Wall Mounting the Sample Sequencer SPD

48 6.2.2 Panel Mount 1. Turn off all power to the Sample Sequencer and disconnect all cables if necessary. 2. Measure the panel hole using the dimension information given in Figure x. 3. Cut out the panel hole. 4. Place supplied panel mount gasket on the rear facing lip of the Sample Sequencer. 5. Fit the Sample Sequencer into the panel hole and secure it with 4 bolts and nuts that were supplied with the Sample Sequencer 5 6. Make all wiring connections as described in the following sections. Figure 42: Panel Mount Dimensions 42 SPD

49 Figure 43: Panel Mounting 6.3 Electrical Connections Electrical Shock Hazard All input and output wiring should be installed in such a manner that access to bare copper is prohibited during normal operation. AVERTISSEMENT Avertissement de choc électrique- Tout le filage devra être installé de manière à éviter que toute partie de fil soit dénudé (sur le cuivre) durant l utilisation normale Access holes sized for ½ NPT (15 mm) and ¾ NPT (20 mm) electrical conduit fittings are provided on the bottom of the enclosure. Sealing type conduit fittings are required to maintain the enclosure s rating. SPD

50 6.3.1 Power Connection The Sample Sequencer contains a universal power supply. The power supply is capable of accepting an input voltage range of VAC and frequency of 50 or 60 Hz. The Sample Sequencer will automatically operate with power supplied in this range. A power supply lead of 16 AWG (1.3 mm 2 ) copper wire is recommended. It is recommended the Sample Sequencer be connected to an electrical outlet or outlet strip with surge suppression and filtering. A suitable external over-current protection device such as a fuse (Class CC, 2A, Time delay) or circuit breaker (C trip curve, 2 A) is recommended for both the hot (L) and neutral (N) leads. Using this method, the plug will serve as the disconnect device. A 16 AWG, 60 o C, VW-1, 600V, type SJT cord is recommended for non-detachable connections. For permanent connections the use of a flexible S0W cable or better must be used. Sizing must be in accordance to the CEC. The mains and output terminal are rated within the following condition: Mains terminals rated up to 300V, 10A, AWG, PA insulating material, Inflammability class V0, Type: Plug, 3 position, and screw connection. Output terminals rated up to 300V, 8A, 14-30AWG, PA insulating material, Inflammability class V0, Type: Plug, 6 or 8 position, and screw connection. For permanently connected installations: - A switch or circuit breaker shall be included in the building installation. - It shall be in close proximity to the equipment and within easy reach of the OPERATOR. - It shall be marked as the disconnecting device for the equipment. The cord and plug shall comply with the following standards: Table 2: Power Cord Requirements Location Cord Plug United States ANSI/UL 817 ANSI/UL 498 Canada CSA C22.2 No. 21 or CSA C22.2 No. 21 and C22.2 No. 42 C22.2 No. 49 European Community IEC or Appropriate IEC standard If desired, the Sample Sequencer may also be hard wired to instrument-quality power using appropriate certified conduit, fittings, and wiring. A suitable external over-current protection device, such as a fuse (Class CC, 2A, Time delay) or circuit breaker (C trip curve, 2 A), and disconnect device is recommended. The over-current protection and disconnect devices shall be installed on both the hot (L) and neutral (N) leads. The disconnect device shall be located near the equipment and marked with appropriate ON(1) OFF(O) markings as specified by local codes. For fixed wiring methods, the ground conductor shall have green with a yellow stripe as the insulation. During connection of the power wire the grounding lead must be suitably routed and long enough so that it is the last conductor to take the strain. Installation should be performed by qualified personal in accordance with local codes and procedures. Figure 44 shows the power connections on the board. 44 SPD

51 Figure 44: Grounding I/O Cable Grounding and EMC Considerations To maintain CSA and CE ratings for EMC, all I/O cables must be shielded and grounded according to Figure 44. In addition, I/O cables must use ferrites to reduce EMI. To properly ground the Sequencer, a ring terminal ground lug should be connected to your ground wire and secured to the enclosure ground stud, which is marked with a ground symbol and located next to the customer power connection. For an insulated ground conductor, use a green or green with yellow stripe conductor which is the same size as the mains conductor. SPD

52 6.4 Sequencer Board Connections The Sample Sequencer provides digital outputs to indicate the current active channel. These connections may be used with a recorder or in situations where analog I/O at the receiving device is limited. In addition to these point identification contacts, replicated analyzer alarms are also located on the board. Figure 45 shows typical connections. Figure 45: Sequencer board connections Valve Output Connections The valve outputs of the Sample Sequencer output 12 VDC, intended for use with the Sentry Manifold Sample Valve. Other manifold valves may be used; however additional hardware such as interposing relays may be required. Typical valve connections are shown in Figure TCP/IP MODBUS andembedded Web Connections TCP/IP MODBUS and Embedded Web connections are made using the Ethernet port. For more information on using the TCP/IP MODBUS and Embedded Web features, refer to Appendix C. 46 SPD

53 Figure 46: Electrical Wiring Diagram 6.5 Alternate Sequencer Board Connections Aside from a power source, the Sample Sequencer 5 can alternatively be connected to the above IO using the quick connects ports and corresponding cables to a customer supplied terminal block. These ports are located on the bottom of the sequencer behind the sealed fitting connections and labeled S5L and S5R (see Figure 47). SPD

54 Figure 47: Alternate Electrical Connections S5L S5R Figure 48: Alternate Electrical Wiring Diagram S5L S5R 48 SPD

55 Figure 49: Alternate Connection Pin-Out Chart SPD

56 6.6 Starting Up the Sample Sequencer 1. Ensure the Sample Sequencer is securely mounted. 2. Wire the Sample Sequencer to the analyzer, sample valves, and other devices as described in the Installation section of this manual. 3. Apply power. 4. Setup the Sample Sequencer as required for your application. This includes setting the sample timers, operating mode, analog input ranges, and custom sequence. Refer to the Quick Start Guide and Operation sections of this manual. 50 SPD

Installation, Operation & Maintenance Manual

Installation, Operation & Maintenance Manual Original Instructions Installation, Operation & Maintenance Manual Sentry Sample Sequencer 5 Controller Sample Conditioning Automation S-SW-IOM-00270-5 1-17 Do not install, maintain, or operate this equipment

More information

AES-402 Automatic Digital Audio Switcher/DA/Digital to Analog Converter

AES-402 Automatic Digital Audio Switcher/DA/Digital to Analog Converter Broadcast Devices, Inc. AES-402 Automatic Digital Audio Switcher/DA/Digital to Analog Converter Technical Reference Manual Broadcast Devices, Inc. Tel. (914) 737-5032 Fax. (914) 736-6916 World Wide Web:

More information

Advanced Digital Melting Point Apparatus

Advanced Digital Melting Point Apparatus Advanced Digital Melting Point Apparatus User Guide Version 1.1 Figure 1: Front view Sample heating block Power on/off (at rear) Printer output (at rear) LCD screen Viewer Capillary storage Control panel

More information

AES-404 Digital Audio Switcher/DA/Digital to Analog Converter

AES-404 Digital Audio Switcher/DA/Digital to Analog Converter Broadcast Devices, Inc. AES-404 Digital Audio Switcher/DA/Digital to Analog Converter Technical Reference Manual Broadcast Devices, Inc. Tel. (914) 737-5032 Fax. (914) 736-6916 World Wide Web: www.broadcast-devices.com

More information

User Manual. Model 1372A and 1374A HDMI Switchers. 1T-SX-632 Model 1372A 2X1 Switcher. v1.3 2x1 SWITCHER. v1.3 INPUT ENHANCE POWER

User Manual. Model 1372A and 1374A HDMI Switchers. 1T-SX-632 Model 1372A 2X1 Switcher. v1.3 2x1 SWITCHER. v1.3 INPUT ENHANCE POWER User Manual 1T-SX-632 Model 1372A 2X1 Switcher v1.3 v1.3 2x1 SWITCHER 1 2 INPUT ENHANCE POWER 1 2 INPUT ENHANCE POWER Model 1372A and 1374A HDMI Switchers Table Of Contents 1.0 Introduction.......................

More information

Tube Rotator. User Guide. Version 1.2

Tube Rotator. User Guide. Version 1.2 Tube Rotator User Guide Version 1.2 Figure 1: Fixed Speed Model Tube holder spindle Tilt adjustment wheel IEC power inlet socket (at rear) Power on/off switch Figure 2: Variable Speed Model Tube holder

More information

Tube Roller Shakers. User Guide. Version 1.2

Tube Roller Shakers. User Guide. Version 1.2 Tube Roller Shakers User Guide Version 1.2 Control panel Rollers Side retaining panels Analog models LED display Drip tray (not visible) Digital models Power On/Off and control dial Roller retaining panel

More information

SQM40/41 Actuators for air and gas dampers

SQM40/41 Actuators for air and gas dampers SQM40/41 Actuators for air and gas dampers Description SQM40/41 actuators are used for the positioning of flow control valves, butterfly valves, dampers or any application requiring rotary motion. The

More information

4 Wiring Brochure Wiring and installation of specific control

4 Wiring Brochure Wiring and installation of specific control - Wiring Brochure tekmarnet 4 User Switch 480 W 480 09/09 1 Information Brochure Choose controls to match application 2 Application Brochure Design your mechanical applications 3 Rough In Wiring Rough-in

More information

Installation Manual SaVi Note Underwater LED Light

Installation Manual SaVi Note Underwater LED Light Installation Manual SaVi Note Underwater LED Light Model Numbers SAVI-NOTE7, SAVI-NOTE0 Table of Contents Safety Precautions...2 SaVi Note Install Instructions...3- M Instructions...- Warnings READ AND

More information

Power Injector 1520 Series

Power Injector 1520 Series Power Injector 1520 Series Technical Specifications Input voltage 100 to 240 VAC Output voltage 56.0 VDC Voltage range tolerance 54 VDC to 57 VDC Maximum current 1.43 A No load current 15 ma 56VDC@0.71A

More information

Model 1421 Distribution Amplifier

Model 1421 Distribution Amplifier Model 1421 Distribution Amplifier Installation and Operating Instructions The 1421 Distribution Amplifier provides four independent, wide bandwidth outputs from one video input. The unit is color compatible

More information

HD Digital Set-Top Box Quick Start Guide

HD Digital Set-Top Box Quick Start Guide HD Digital Set-Top Box Quick Start Guide Eagle Communications HD Digital Set-Top Box Important Safety Instructions WARNING TO REDUCE THE RISK OF FIRE OR ELECTRIC SHOCK, DO NOT EXPOSE THIS PRODUCT TO RAIN

More information

Installation and User Guide 458/CTR8 8-Channel Ballast Controller Module

Installation and User Guide 458/CTR8 8-Channel Ballast Controller Module Installation and User Guide 458/CTR8 8-Channel Ballast Controller Module Helvar Data is subject to change without notice. www.helvar.com i Contents Section Page Introduction 1 Installation 2 1. Attach

More information

TEK-LCD 7801A. NEMA 4X Loop-Powered Feet and Inches Level Meter. ACCESSORIES. Technology Solutions

TEK-LCD 7801A. NEMA 4X Loop-Powered Feet and Inches Level Meter.   ACCESSORIES. Technology Solutions Technology Solutions TEK-LCD 7801A NEMA 4X Loop-Powered Feet and Inches Level Meter ACCESSORIES www.tek-trol.com Flow Level Temperature Pressure Valves Analyzers Accessories TekValSys Introduction The

More information

TeamWork Kits Installation Guide

TeamWork Kits Installation Guide TX 0 RX COM +5V APARATUS US TeamWork Kits Installation Guide TeamWork 400 and TeamWork 600 Kits The TeamWork 400 and TeamWork 600 kits consist of an HDMI switcher, system controller, Cable Cubby, and cables

More information

MS2540 Current Loop Receiver with RS485 Communication

MS2540 Current Loop Receiver with RS485 Communication MS2540 Current Loop Receiver with RS485 Communication User Manual Metal Samples Company A Division of Alabama Specialty Products, Inc. 152 Metal Samples Rd., Munford, AL 36268 Phone: (256) 358 4202 Fax:

More information

Sprite TL Quick Start Guide

Sprite TL Quick Start Guide Sprite TL Quick Start Guide with 115 VAC Power Cord and 4-Conductor Signal Cable Reference Manual Sprite TL Online and downloadable Product Manuals and Quick Start Guides are available at www.hydrosystemsco.com

More information

Colour Explosion Proof Video Camera USER MANUAL VID-C

Colour Explosion Proof Video Camera USER MANUAL VID-C Colour Explosion Proof Video Camera USER MANUAL VID-C Part Number: MAN-0036-00 Rev 4 Copyright 2002 Net Safety Monitoring Inc. Printed in Canada This manual is provided for informational purposes only.

More information

PAM-1840 Preamplifier Operation Manual

PAM-1840 Preamplifier Operation Manual PAM-1840 Preamplifier Operation Manual 1 TABLE OF CONTENTS INTRODUCTION 3 GENERAL INFORMATION 4 SPECIFICATIONS 4 OPERATING INSTRUCTIONS 5 MAINTENANCE 6 2 INTRODUCTION BEFORE APPLYING POWER Review this

More information

LTC 113x & LTC123x FlexiDome Series Fixed Dome Cameras

LTC 113x & LTC123x FlexiDome Series Fixed Dome Cameras LTC 113x & LTC123x FlexiDome Series Fixed Dome Cameras Eng Installation Instructions F D E NL I IMPORTANT SAFEGUARDS 1. Read Instructions All the safety and operating instructions should be read before

More information

NS-3 RF Noise Source Operation Manual

NS-3 RF Noise Source Operation Manual RF Noise Source Operation Manual Version 2.04 June 3, 2016 SPECIFICATIONS Frequency... Maximum output level... Output flatness... (at max output level) Impedance... Displayed level... Repeatability...

More information

DIGITAL TIME CLOCK. used with the Meal-Time Control Panel. AGRI-TIME TM Control Instruction Page 1

DIGITAL TIME CLOCK. used with the Meal-Time Control Panel. AGRI-TIME TM Control Instruction Page 1 TM AGRI- DIGITAL CLOCK used with the 34385 Meal-Time Control Panel MF1116A12 May 1995 Page 1 Warranty Information Chore-Time Equipment warrants each new product manufactured by it to be free from defects

More information

Long Range Ethernet Extender

Long Range Ethernet Extender CopperLink Model 2160 Series Long Range Ethernet Extender Quick Start Guide Part Number: 07M2160-QS, Rev. B Revised: February 24, 2012 Sales Office: +1 (301) 975-1000 Technical Support: +1 (301) 975-1007

More information

UDC100 Universal Digital Controller. Specification. Overview. Features. Features, continued /99 Page 1 of 4

UDC100 Universal Digital Controller. Specification. Overview. Features. Features, continued /99 Page 1 of 4 UDC100 Universal Digital Controller 51-52-03-29 11/99 Page 1 of 4 Specification Overview The UDC100 Universal Digital Controller is a microprocessor-based 1/4 DIN low cost temperature controller. It combines

More information

MOONRING MULTIPLES INSTALLATION INSTRUCTIONS MOONRING MULTISTAK TM STACKED CANOPY. A nd Avenue, Unit 1 Oakland,

MOONRING MULTIPLES INSTALLATION INSTRUCTIONS MOONRING MULTISTAK TM STACKED CANOPY. A nd Avenue, Unit 1 Oakland, INSTALLATION INSTRUCTIONS MOONRING MULTIPLES MOONRING MULTISTAK TM STACKED CANOPY A 1035 22nd Avenue, Unit 1 Oakland, CA 94606 P 510.489.2530 E TalkToUs@alwusa.com W alwusa.com MR-MultiStak - Safety &

More information

DTV742 Installation Quick Start Guide

DTV742 Installation Quick Start Guide DTV742 Installation Quick Start Guide The Wegener Model DTV742 8VSB 4-Channel Multiplexor receives VHF/UHF broadcast HDTV signals and provides output ASI streams for connection to cable system QAM modulators.

More information

AS-300D Series Smart Bench Scales Owner s Manual

AS-300D Series Smart Bench Scales Owner s Manual DETE CTO A Division of Cardinal Scale Manufacturing Co. AS-300D Series Smart Bench Scales Owner s Manual 8527-M214-O1 Rev B 01/03 PO BOX 151 WEBB CITY, MO 64870 417-673-4631 Printed in USA INTRODUCTION

More information

AEROTRAK PORTABLE AIRBORNE PARTICLE COUNTER MODEL 9110 QUICK START GUIDE

AEROTRAK PORTABLE AIRBORNE PARTICLE COUNTER MODEL 9110 QUICK START GUIDE AEROTRAK PORTABLE AIRBORNE PARTICLE COUNTER MODEL 9110 QUICK START GUIDE Thank you for purchasing a TSI AeroTrak Model 9110 Portable Airborne Particle Counter (particle counter). This guide will help you

More information

Operating Manual. Automated Gear. Apollo Design Technology, Inc Fourier Drive Fort Wayne, IN USA

Operating Manual. Automated Gear. Apollo Design Technology, Inc Fourier Drive Fort Wayne, IN USA Operating Manual Automated Gear Apollo Design Technology, Inc. 4130 Fourier Drive Fort Wayne, IN 46818 USA PH: +01(260)497-9191 FX: +01(260)497-9192 www.apollodesign.net 11-25-09 5-6 POWERING UP THE RIGHT

More information

HD Digital MPEG2 Encoder / QAM Modulator

HD Digital MPEG2 Encoder / QAM Modulator HD Digital MPEG2 Encoder / QAM Modulator HDMI In QAM Out series Get Going Guide ZvPro 800 Series is a one or two-channel unencrypted HDMI-to-QAM MPEG 2 Encoder / QAM Modulator, all in a compact package

More information

Modular Lube Lubrication Systems System Controls

Modular Lube Lubrication Systems System Controls Model 84501 Program Timer Solid State Designed to control the lubrication cycle frequency of air-operated single-stroke pumps. Timer turns pump on/off at programmed intervals via a 3-way or 4-way air solenoid

More information

Operating Instructions

Operating Instructions CNTX Contrast sensor Operating Instructions CAUTIONS AND WARNINGS SET-UP DISTANCE ADJUSTMENT: As a general rule, the sensor should be fixed at a 15 to 20 angle from directly perpendicular to the target

More information

Evolution Digital HD Set-Top Box Important Safety Instructions

Evolution Digital HD Set-Top Box Important Safety Instructions Evolution Digital HD Set-Top Box Important Safety Instructions 1. Read these instructions. 2. Keep these instructions. 3. Heed all warnings. 4. Follow all instructions. 5. Do not use this apparatus near

More information

TeamWork 601 Kit Installation Guide

TeamWork 601 Kit Installation Guide C G G TX RX COM +V APARATUS US 0 TeamWork 0 Kit Installation Guide TeamWork 0 Kit The TeamWork 0 kit consists of an analog and digital video switcher, system controller, Cable Cubby, and cables packaged

More information

Sentinel I24 Digital Input and Output Configuration

Sentinel I24 Digital Input and Output Configuration Application Bulletin: #155 Date: October 19, 2007 Sentinel I24 Digital Input and Output Configuration The Sentinel I24 can communicate with external hardware using digital inputs and outputs. There are

More information

Digital Input Modules for Compact FieldPoint

Digital Input Modules for Compact FieldPoint Digital Modules for Compact FieldPoint NI cfp-di-300, NI cfp-di-301, NI cfp-di-304, NI 8-,16-, or 32-channel inputs 24 VDC inputs 4 to 250 VDC inputs 15 to 250 VAC inputs (50/60 Hz AC) 3 to 250 VAC inputs

More information

VF-STB IP Appliance User Guide

VF-STB IP Appliance User Guide VF-STB IP Appliance User Guide Copyright 2007 Video Furnace, Inc. All Rights Reserved. Copyright 2002-2007 Video Furnace, Inc. All Rights Reserved This document refers to the current published version

More information

Operating Instructions

Operating Instructions Operating Instructions SDI Input board Model No. AV-HS04M1 РУССКИЙ FRANÇAIS DEUTSCH ENGLISH ESPAÑOL ITALIANO Before operating this product, please read the instructions carefully and save this manual for

More information

Trusted 40 Channel 120 Vac Digital Input FTA

Trusted 40 Channel 120 Vac Digital Input FTA PD-T8824 Trusted Trusted 40 Channel 120 Vac Digital Input FTA Product Overview The Trusted 40 Channel 120 Vac Digital Input Field Termination Assembly (FTA) T8824 is designed to act as the main interface

More information

Litile34 OPERATION MANUAL

Litile34 OPERATION MANUAL Litile34 OPERATION MANUAL Seamless Tiled Panel Wall Solution for Large Area Digital Signage Display (1st Edition 3/25/2009) All information is subject to change without notice. Approved by Checked by Prepared

More information

7 Day Digital Time Switch

7 Day Digital Time Switch 7 Day Digital Time Switch Model: NTT08 Installation & Operating Instructions 1 1. General Information These instructions should be read carefully and retained for further reference and maintenance. 2.

More information

Installation Manual Original Instructions - IW4001

Installation Manual Original Instructions - IW4001 Installation Manual Original Instructions - IW4001 Installation Manual 1 General Operator and Supervisor Information Signal Word Definition Signal Word Panel Table of Contents Operator and Supervisor Information

More information

Installation. SAPTF33xx-1xx in the Network. Standard Configuration

Installation. SAPTF33xx-1xx in the Network. Standard Configuration SAPTF33xx-1xx in the Network Standard Configuration One Unit A device (SAPTF33xx-100) and one device () are required for the standard configuration. The Unit A device is connected to the while the device

More information

FIELD MOUNTED RATE TOTALISER MODEL 202D

FIELD MOUNTED RATE TOTALISER MODEL 202D FIELD MOUNTED RATE TOTALISER MODEL 202D 16 June 2003 Introduction 1 1. INTRODUCTION The Model 202Di Rate Totaliser is a microprocessor based instrument which accepts a frequency or pulse input from

More information

ST8-WiFi Timer. Installation Guide and Operations Manual. English MIN M D YYYY

ST8-WiFi Timer. Installation Guide and Operations Manual. English MIN M D YYYY ST8-WiFi Timer Installation Guide and Operations Manual AM M D YYYY English Contents ST8-WiFi Timer Installation Guide and Operations Manual Introduction Welcome to Rain Bird... 1 Timer Features... 1 Controls

More information

FCC Compliance Statement

FCC Compliance Statement FCC Compliance Statement This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: 1. This device may not cause harmful interference, and 2. This device

More information

ipump 622 Streaming Media Decoder Installation Quick Start Guide

ipump 622 Streaming Media Decoder Installation Quick Start Guide ipump 622 Streaming Media Decoder Installation Quick Start Guide General Information Unpacking and Inspection The Wegener ipump 622 Enterprise Media Server is an integrated satellite receiver and video

More information

Documentation on all Paxton products can be found on our web site -

Documentation on all Paxton products can be found on our web site - 11/05/2012 Ins-30202-US Net2 Entry - Monitor Paxton Technical Support 1.800.672.7298 supportus@paxton-access.com Technical help is available: Monday - Friday from 02:00 AM - 8:00 PM (EST) Documentation

More information

RA-RS232, RB-RS232. Setup and Installation Guide Addendum For RadioRA RS232 Interface

RA-RS232, RB-RS232. Setup and Installation Guide Addendum For RadioRA RS232 Interface RA-RS232, RB-RS232 Setup and Installation Guide Addendum For RadioRA RS232 Interface A Comprehensive Step-by-Step Guide for Programming and Operating the Lutron RadioRA RS232 Interface Note: Please leave

More information

19 / 20.1 / 22 WIDE SCREEN TFT-LCD MONITOR

19 / 20.1 / 22 WIDE SCREEN TFT-LCD MONITOR 19 / 20.1 / 22 WIDE SCREEN TFT-LCD MONITOR V193/ V220 Series V202 Series USER MANUAL www.viewera.com Rev. 2.0 Table of Contents EMC Compliance......1 Important Precautions...2 1. Package contents....3

More information

AES Channel Digital/Analog Audio Switcher/DA/Digital to Analog Converter

AES Channel Digital/Analog Audio Switcher/DA/Digital to Analog Converter Broadcast Devices, Inc. AES-408 8 Channel Digital/Analog Audio Switcher/DA/Digital to Analog Converter Technical Reference Manual Broadcast Devices, Inc. Tel. (914) 737-5032 Fax. (914) 736-6916 World Wide

More information

HD-One DX/DX500 HDMI Extender

HD-One DX/DX500 HDMI Extender HD-One DX/DX500 HDMI Extender User Manual Version 2.0 tvone 2791 Circleport Drive, Erlanger, KY 41018, USA. Americas: 859-282-7303 EMEA: +44 (0) 1843 873322 Email: tech.usa@tvone.com www.tvone.com CSG-DX-500

More information

Transfer Switch. OTECA (Spec A) OTECB (Spec A) OTECC (Spec A) OTECD (Spec A) Amperes. English Original Instructions (Issue 5)

Transfer Switch. OTECA (Spec A) OTECB (Spec A) OTECC (Spec A) OTECD (Spec A) Amperes. English Original Instructions (Issue 5) Operator Manual Transfer Switch 40-1000 Amperes OTECA (Spec A) OTECB (Spec A) OTECC (Spec A) OTECD (Spec A) English Original Instructions 10-2015 962-0131 (Issue 5) Table of Contents 1. SAFETY PRECAUTIONS...

More information

CAUTION RISK OF ELECTRIC SHOCK NO NOT OPEN

CAUTION RISK OF ELECTRIC SHOCK NO NOT OPEN Evolution Digital HD Set-Top Box Important Safety Instructions 1. Read these instructions. 2. Keep these instructions. 3. Heed all warnings. 4. Follow all instructions. 5. Do not use this apparatus near

More information

Scoreboard Operator s Instructions MPCX Volleyball Control

Scoreboard Operator s Instructions MPCX Volleyball Control Scoreboard Operator s Instructions MPCX Volleyball Control Since 1934 Retain this manual in your permanent files Rev. 2/3/2012 135-0137 These Instructions are for the Following Models: LED models: Incandescent

More information

MASTR II BASE STATION 12/24V POWER SUPPLY 19A149979P1-120 VOLT/60 Hz 19A149979P2-230 VOLT/50 Hz

MASTR II BASE STATION 12/24V POWER SUPPLY 19A149979P1-120 VOLT/60 Hz 19A149979P2-230 VOLT/50 Hz Mobile Communications MASTR II BASE STATION 12/24V POWER SUPPLY 19A149979P1-120 VOLT/60 Hz 19A149979P2-230 VOLT/50 Hz CAUTION THESE SERVICING INSTRUCTIONS ARE FOR USE BY QUALI- FIED PERSONNEL ONLY. TO

More information

TS2.8 Sub OWNER S MANUAL

TS2.8 Sub OWNER S MANUAL TS2.8 Sub OWNER S MANUAL TS2.8 Sub CONTENTS IMPORTANT SAFETY INSTRUCTIONS 03 WARNINGS 03 FUSE PROTECTION 04 WARNING: STRONG MAGNETIC FIELD 04 EMC / EMI 04 ECODESIGN STANDBY POWER CONSUMPTION 04 WARRANTY

More information

Process Transmitter RMA 422

Process Transmitter RMA 422 Technical Information TI 072R/24/ae Process Transmitter RMA 422 Multi-functional 1-2 channel top hat DIN rail unit with loop power supply, alarm set point monitoring, mathematics function and 1-2 analog

More information

Multifunction Digital Timer

Multifunction Digital Timer Multifunction Digital Timer 72 x72 mm Timer with Easy-to-use Functions Nine output modes accommodate a wide variety of applications. All parameters set by scroll-through menus accessed from the front panel.

More information

GE Interlogix Fiber Options S700V & S702V. Instruction Manual FIBER-OPTIC VIDEO TRANSMISSION SYSTEM

GE Interlogix Fiber Options S700V & S702V. Instruction Manual FIBER-OPTIC VIDEO TRANSMISSION SYSTEM g GE Interlogix Fiber Options Instruction Manual S700V & S702V FIBER-OPTIC VIDEO TRANSMISSION SYSTEM Federal Communications Commission and Industry Canada Radio Frequency Interference Statements This equipment

More information

HD Digital MPEG2 Encoder / QAM Modulator

HD Digital MPEG2 Encoder / QAM Modulator HD Digital MPEG2 Encoder / QAM Modulator YPrPb VGA In QAM Out series Get Going Guide ZvPro 600 Series is a one or two-channel Component or VGA-to-QAM MPEG 2 Encoder/ Modulator, all in a compact package

More information

VPM2. Operation Manual

VPM2. Operation Manual VPM2 Operation Manual Whip Corporation 361 Farmington Ave. P.O. Box 17183 Louisville, KY 40217-0183 USA 502-637-1451 800-626-5651 Fax 502-634-4512 www.whipmix.com LISTED Features The Whip VPM2 is designed

More information

Guide for installers. METTLER TOLEDO MultiRange System solution analog Ex1. Hazardous area. Safe area

Guide for installers. METTLER TOLEDO MultiRange System solution analog Ex1. Hazardous area. Safe area Guide for installers METTLER TOLEDO MultiRange System solution analog Ex1 Hazardous area Safe area System solution analog Ex1 Contents Contents Page 1 Safety precautions... 2 2 System overview... 3 2.1

More information

Operating Instructions 07/2007 Edition. SINAMICS G130/G150 Line harmonics filter. sinamics

Operating Instructions 07/2007 Edition. SINAMICS G130/G150 Line harmonics filter. sinamics Operating Instructions 07/2007 Edition SINAMICS G130/G150 Line harmonics filter sinamics s Safety information 1 General 2 SINAMICS SINAMICS G130/G150 Operating Instructions Mechanical installation 3 Electrical

More information

Digital Time Switch. Installation & Operating Instructions

Digital Time Switch. Installation & Operating Instructions Digital Time Switch Model: NTT06 24 Hour General Purpose Digital Timer Model: NTT07 7 Day General Purpose Digital Timer Installation & Operating Instructions 1 1. General Information These instructions

More information

Model 1476-C SuperQuad HR

Model 1476-C SuperQuad HR Model 1476-C SuperQuad HR Installation and Operating Instructions Table of Contents Page Table of Content... 2 System Description... 3 Features... 3 Installation... 4 Internal Setups... 4 Connections...

More information

User s Manual TOS7200. Insulation Resistance Tester DANGER. 1. Preface Preparation Safety Precautions during Testing 19

User s Manual TOS7200. Insulation Resistance Tester DANGER. 1. Preface Preparation Safety Precautions during Testing 19 No. IB028911 Nov. 2016 User s Manual Insulation Resistance Tester TOS7200 1. Preface 11 2. Preparation 14 3. Safety Precautions during Testing 19 4. Basic Operation 23 5. Using Terminals and Connectors

More information

Electronic converter for level transmitters MT03L Instructions manual

Electronic converter for level transmitters MT03L Instructions manual Electronic converter for level transmitters MT03L Instructions manual R-MI-MT03L Rev.: 1 English version PREFACE Thank you for choosing the MT03L converter from MT03 series of Tecfluid S.A. This instruction

More information

VPM2. Operator's Manual

VPM2. Operator's Manual VPM2 Operator's Manual Whip Mix Corporation 361 Farmington Ave. P.O. Box 17183 Louisville, KY 40217-0183 USA 502-637-1451 800-626-5651 Fax 502-634-4512 www.whipmix.com Features The Whip Mix VPM2 is designed

More information

Active UTP Transceiver Hub Installation Guide

Active UTP Transceiver Hub Installation Guide Active UTP Transceiver Hub Installation Guide Models Include: HubWayEX16 - Sixteen (16) Channels - Thirty-two (32) Channels Rev. 011810 More than just power. TM Overview: Altronix HubWayEX16/EX32 Active

More information

Intelligent Security and Fire Ltd

Intelligent Security and Fire Ltd User Manual Product ranges covered by this manual Vi-P14 Vi-P14A Document Reference Date Firmware Vi-Q4C1 Viq601a.doc 26/11/2009 From Viq001a21 Videoswitch Telephone 01252-851510 Ocean House, Redfields

More information

User Guide. Centrex Recording Interface

User Guide. Centrex Recording Interface User Guide Centrex Recording Interface Table of Contents Introduction... 2 The Meridian Business Set... 3 Key Numbering Plan (18 button add-on)... 4 Key Numbering Plan (36 button add-on)... 5 Key Numbering

More information

Trusted 40 Channel Analogue Input FTA

Trusted 40 Channel Analogue Input FTA PD-T8830 Trusted Trusted 40 Channel Analogue Input FTA Product Overview The Trusted 40 Channel Analogue Input Field Termination Assembly (FTA) T8830 is designed to act as the main interface between a field

More information

Panel cutout required: 1.772" x 3.622" (45mm x 92mm) 1.76" (45mm) 2.45" (62mm) 3.20" (81mm) 3.60" (91mm) 0.59" (15mm) Special Features

Panel cutout required: 1.772 x 3.622 (45mm x 92mm) 1.76 (45mm) 2.45 (62mm) 3.20 (81mm) 3.60 (91mm) 0.59 (15mm) Special Features NEMA4X, IP65 Front Bezel Meter with Relays Option RELAY2 RELAY1 24V OUT POWER 4 3 2 1 6 5 2 1 2 1 NO NC COM NO NC COM RTD TC P P RTD 3 4 1 2 5 TC 6 SWITCH Rear View Gasket APM765 Panel Meter Description

More information

ORDERING Page 6 BASLER RELAY STANDARDS, DIMENSIONS, ACCESSORIES Request bulletin SDA

ORDERING Page 6 BASLER RELAY STANDARDS, DIMENSIONS, ACCESSORIES Request bulletin SDA BE1-59NC CAPACITOR NEUTRAL OVERVOLTAGE RELAY The BE1-59NC Capacitor Neutral Overvoltage Relay provides sensitive protection for capacitor banks. ADDITIONAL INFORMATION INSTRUCTION MANUAL ADVANTAGES Helps

More information

SignalOn Series. L-Band Power Inserter Module INSTALLATION & OPERATION MANUAL. 1.2 GHz. D3.

SignalOn Series. L-Band Power Inserter Module INSTALLATION & OPERATION MANUAL. 1.2 GHz.     D3. SignalOn Series D3.1/CCAP Compliant 1.2 GHz L-Band Power Inserter Module INSTALLATION & OPERATION MANUAL www.atxnetworks.com www.atxnetworks.com Although every effort has been taken to ensure the accuracy

More information

Scoreboard Operator s Instructions MPCX SCD / DGT / Pitch Time Control

Scoreboard Operator s Instructions MPCX SCD / DGT / Pitch Time Control Scoreboard Operator s Instructions MPCX SCD / DGT / Pitch Time Control Since 1934 Retain this manual in your permanent files Rev. 2/3/2012 135-0136 These Instructions are for the Following Models: LED

More information

Analog Input Module HART Ex i / I.S. Inputs, 8 Channels Type 9461/

Analog Input Module HART Ex i / I.S. Inputs, 8 Channels Type 9461/ > 8 channels for 2-wire HART transmitters > Intrinsically safe inputs Ex ia IIC > Galvanic separation between inputs and system > Open-circuit and short-circuit monitoring for each field circuit > Module

More information

Model Number Structure

Model Number Structure Cycle Control Units CSM DS_E_7_1 Refer to Safety Precautions for All Power Controllers. Used in Combination with the to Enable High-precision Temperature Control Use cycle control to achieve power control

More information

Trusted 40 Channel 120 Vac Digital Input FTA

Trusted 40 Channel 120 Vac Digital Input FTA ICSTT-RM290F-EN-P (PD-T8824) Trusted Product Overview The Trusted 40 Channel 120 Vac Digital Input Field Termination Assembly (FTA) T8824 is designed to act as the main interface between a field device

More information

FiliBuster. User Manual

FiliBuster. User Manual FiliBuster User Manual Unit 1A 126 Great North Road Hatfield, Hertfordshire AL9 5JN UK Tel: +44 (0) 1462 600101 Email: moreinfo@interspaceind.com www.interspaceind.com Cueing and Presentation Control Specialists

More information

SINAMICS G130 / G150. Line harmonics filter. Operating Instructions 05/2010 SINAMICS

SINAMICS G130 / G150. Line harmonics filter. Operating Instructions 05/2010 SINAMICS SINAMICS G130 / G150 Line harmonics filter Operating Instructions 05/2010 SINAMICS s Safety information 1 General 2 SINAMICS SINAMICS G130 / G150 Operating Instructions Mechanical installation 3 Electrical

More information

www.greenelectricalsupply.com MaxLite 6 & 8 Commercial Downlight Retrofit General Safety Information To reduce the risk of death, personal injury or property damage from fire, electric shock, falling parts,

More information

Model#: IN-MDRI3MF. Hardware User Manual. 3MP Indoor Mini Dome with Basic WDR, Fixed lens. (PoE) Ver. 2013/02/04

Model#: IN-MDRI3MF. Hardware User Manual. 3MP Indoor Mini Dome with Basic WDR, Fixed lens. (PoE) Ver. 2013/02/04 Model#: IN-MDRI3MF 3MP Indoor Mini Dome with Basic WDR, Fixed lens Hardware User Manual (PoE) Ver. 2013/02/04 Table of Contents 0. Precautions 3 1. Introduction 4 Package Contents... 4 Features and Benefits...

More information

User Guide. HDMI Fiber Optic Extender. DVI-7350a

User Guide. HDMI Fiber Optic Extender. DVI-7350a User Guide HDMI Fiber Optic Extender DVI-7350a Table of Contents Section Page Product Safety.................................... 1 1.0 Introduction...2 2.0 Specifications...3 3.0 Package Contents...3 4.0

More information

USER S MANUAL

USER S MANUAL 612720 USER S MANUAL SAFETY WARNINGS AND GUIDELINES Prior to operation, check the unit and power cord for physical damage. Do not use if physical damage has occurred. Before plugging the unit into a power

More information

NI PXI/PXIe-2543 Specifications

NI PXI/PXIe-2543 Specifications NI PXI/PXIe-2543 Specifications 6.6 GHz Dual 4 1 Terminated Solid State Multiplexer (Dual SP4T) This document lists specifications for the NI PXI/PXIe-2543 (NI 2543) multiplexer module. All specifications

More information

USER MANUAL. 27 Full HD Widescreen LED Monitor L270E

USER MANUAL. 27 Full HD Widescreen LED Monitor L270E USER MANUAL 27 Full HD Widescreen LED Monitor L270E TABLE OF CONTENTS 1 Getting Started 2 Control Panel/ Back Panel 3 On Screen Display 4 Technical Specs 5 Care & Maintenance 6 Troubleshooting 7 Safety

More information

FD Trinitron Colour Television

FD Trinitron Colour Television R 4-205-569-32(1) FD Trinitron Television Instruction Manual GB KV-14LM1U 2000 by Sony Corporation NOTICE FOR CUSTOMERS IN THE UNITED KINGDOM A moulded plug complying with BS1363 is fitted to this equipment

More information

SAFETY WARNINGS AND GUIDELINES INTRODUCTION CUSTOMER SERVICE

SAFETY WARNINGS AND GUIDELINES INTRODUCTION CUSTOMER SERVICE SAFETY WARNINGS AND GUIDELINES Prior to operation, check the unit and power cord for physical damage. Do not use if physical damage has occurred. Before plugging the unit into a power outlet, ensure that

More information

Single output models feature wide-range output adjustability to meet a wide variety of standard and user-specific output voltage requirements.

Single output models feature wide-range output adjustability to meet a wide variety of standard and user-specific output voltage requirements. RoHS Lead-Solder-Exemption Compliant New 3.3 V and 5 V Output Models Universal Input 85-264 VAC Industry-Standard Footprint: 7.0" x 4.3" x 1.97" (177.8 x 109.2 x 50.0 mm) Input Transient & ESD Compliance

More information

BE3-GPR GENERATOR PROTECTIVE RELAY

BE3-GPR GENERATOR PROTECTIVE RELAY BE3-GPR GENERATOR PROTECTIVE RELAY Behind-the-Panel Mounting Semi-flush Mounting Basler Electric s BE3-GPR generator protective relay offers multiple protective features in a single package. Its microprocessor-based

More information

VARIABLE SPEED USER MANUAL. Premium Efficiency Variable Speed Motor

VARIABLE SPEED USER MANUAL. Premium Efficiency Variable Speed Motor 165 VARIABLE SPEED USER MANUAL Premium Efficiency Variable Speed Motor COPYRIGHT Copyright 2013, Regal Beloit America, Inc. Tipp City, Ohio. All rights reserved. TRADEMARKS All trademarks and registered

More information

DV6819 Quick Reference Guide V1.0. Smart TV Box. Quick Reference Guide. Please do read user manual before you operate the TV box.

DV6819 Quick Reference Guide V1.0. Smart TV Box. Quick Reference Guide. Please do read user manual before you operate the TV box. DV6819 Quick Reference Guide V1.0 Smart TV Box Quick Reference Guide Please do read user manual before you operate the TV box. ~ 1 ~ DV6819 Quick Reference Guide V1.0 Safety instruction Please keep the

More information

User Guide. Single-Link DVI Active Cable Extender. DVI-7171c

User Guide. Single-Link DVI Active Cable Extender. DVI-7171c User Guide Single-Link DVI Active Cable Extender DVI-7171c TABLE OF CONTENTS SECTION PAGE PRODUCT SAFETY...1 PRODUCT LIABILITY...1 1.0 INTRODUCTION...2 2.0 SPECIFICATIONS...3 3.0 PACKAGE CONTENTS...4 4.0

More information

TeamWork Installation Guide

TeamWork Installation Guide C G G 00-0V/ A MAX TX RX +V APARATUS US 0 TeamWork Installation Guide TeamWork TeamWork is a fully customizable collaboration system comprised of an switcher, Show Me cables, a control processor, and a

More information

3500/42E Vibration Monitor

3500/42E Vibration Monitor 3500/42E Vibration Monitor Bently Nevada* Asset Condition Monitoring Description The 3500/42E Vibration Monitor is a 4-channel monitor that accepts input from proximity and seismic transducers, conditions

More information

Instruction Manual Fixed Speed Vortex Mixer Analog Vortex Mixer Digital Vortex Mixer Pulsing Vortex Mixer

Instruction Manual Fixed Speed Vortex Mixer Analog Vortex Mixer Digital Vortex Mixer Pulsing Vortex Mixer Instruction Manual Fixed Speed Vortex Mixer Analog Vortex Mixer Digital Vortex Mixer Pulsing Vortex Mixer Table of Contents Package Contents............ 1 Warranty............ 1 Installation............

More information

Site Installation Model MP-8433

Site Installation Model MP-8433 Site Installation Model MP- Rev. //0 SCOREBOARD SITE INSTALLATION INSTRUCTIONS CAUTION: All American Scoreboards (AAS) recommends the sign be installed by a licensed contractor, and must meet all local

More information