1 CT & Workload (BIR) David Sutton
2 What do we mean by workload? Number of scans Type of scans/mix of scans? Total mas Total dip
3 What do we do with it? Need to relate workload to total scatter dose in the room. Manufacturer s data generally an isodose plot showing scatter distribution, typically in ngy per mas for a given scan protocol. Manufacturer s data typically for a head phantom and a body phantom and in two perpendicular planes.
4 Typical Isodose Plot
5 Critical Direction Isodose plot not symmetrical, ti and dh has critical directions. Shielding should be specified according to these worst case scenarios A H
6
7
8 /(distance)2 A B C D E F G H Dose per slice (microgray y)
9 120 kvp; 250 mas
10 CT Calculation (G) Distance: (4.0/cos 36) = 4.9 m Weekly workload: 3340 x 250 mas slices Distance to 2 µgy isodose: 1.7 m Transmission: 7.2 x 10-3 Concrete thickness: 100 mm Thickness at 36 : 100 x (1 + cos36)/2 = 90 mm Dose per week: 2 x 3340 x (1.7/4.9) 2 = 800 µgy
11 Scatter from walls in CT Protection against radiation scattered from adjacent walls or the ceiling is usually required It is likely to be necessary: 1) to take shielding to ceiling height 2) to have a fully enclosed operator s cubicle with door
12 Relating plot to dose Isodose plot not symmetrical, and has critical directions. Shielding should be specified according to these worst case scenarios Ideally then, one needs to determine the total mas for the two generic types of phantom, head and body.
13 Getting the data Information available from a scanner can include predicted mas predicted dlp, CTDI vol True dlp mas. It all depends on the scanner being used and its age.
14 Getting the data - What data? Which scans are being done? Which scans are going to be to be done? Crucial questions to answer How many of them there are going to be?
15 How to get the data The first thing to do is to find out what the scanner is going to be used for. This means speaking to a human being! Or even more than one!
16 Data The length of the working day The number of patients per day The scans to be performed The relative number of different types of scan. Crystal ball gazing g
17 Data For all of the major scan types you need some idea of the scan protocol e.g. 400 ma max on automa 120 kvp Pitch 1.375:1 Projected dip
18 Getting the data Information available from a scanner can include predicted mas predicted dlp, CTDI vol True dlp mas. It all depends on the scanner being g used and its age. Getting the data - What data? Which scans are being g done? Which scans are going to be to be done? Crucial questions to answer? How many of them there are going to be? Data The length of the working day The number of patients per day The scans to be performed The relative number of different types of scan. Crystal ball gazing Data For all of the major scan types you need some idea of the scan protocol e.g. 400 ma max on automa 120 kvp Pitch 1.375:1 Projected dip Not a trivial task!
19 Exisiting Data You may be able to base your decision on existing data, for example this may be a replacement room, or a room which will be used for a similar purpose as one elsewhere. Conceptually this should ease the process since it is very unusual not to have access to a RIS system. To prepare and to give some examples of the process we chose 5 scanners and looked at data at between 750 and 1000 sequential patients on each.
20 Clean up data
21 Cleaned
22 Dundee
23 Edinburgh
24 Relative Contributrions Head Body Other N NWIP 52% 40% 8% 1201 NWOP 24% 69% 7% 757 WGH 38% 58% 4% 1110 STJ 40% 54% 6% 866 DCN 85% 6% 9% 968
25 ? Others Others WGH CT Angiogram Aorta CT Angiogram Renal/Abdominal CT Guided Ablation CT Guided Aspiration CT Guided Biopsy CT Guided Drainage CT Hip Rt CT Liver Triple Phase Others - NW CAICN CANKL CANKR CART CBIOPB CCALR CCHAP CCHES CCHESC CCSPN CDRAID CELBL CFACE CFACI
26 Options Your scanner gives you an mas reading / you have mas workload data. You want to use the isodose method You want to use the dlp method? Your scanner gives you a dlp reading / you have dlp workload data You want to use the dlp method You want to use the isodose method
27 You want to use the isodose CT Abdomen and Pelvis method CT Abdomen b With Anatomical Groups / Protocol Types Contrast CT Abdomen/Pelvis With Contrast CT Angiogram Aorta CT Angiogram Pulmonary CT Angiogram CT Chest CT Chest and Biopsy CT Chest High Resolution CT Chest/Abdo CT Chest/Abdo/Pelvis/ With Contrast CT Colonoscopy (Virtual) CT Head CT KUB CT Urogram Head Abdomen Chest Chest & Pelvis Angiography
28 Workload Determination Chest Abdo Pelvis An Example
29 1) Determine Scan Length
30 2) N Months Workload DLP (Smart ma) from RIS
31 3) Data nctdiw = 8.3 mgy/100 mas Pitch = nctdivol= 6 mgy/100 mas Scan Length = 75 cm (from phantom) ndlp= 450 mgycm/100mas
32 4) Workload Mean dlp = 930 mgycm Mean mas per CAP = 100*930/ 450 mas =200 mas (using ma modulation)
33 Options Your scanner gives you an mas reading / you have mas workload data. You want to use the isodose method You want to use the dlp method? Your scanner gives you a dlp reading / you have dlp workload data You want to use the dlp method You want to use the isodose method
34 Example from 2006 (i) Workload- Patients & Protocols Scatter diagrams state the dose per rotation at given exposure parameters. To calculate annual dose at each barrier, need to: (i) correct dose for exam exposure parameters (ii) multiply by total number of rotations per year Estimations are therefore required for typical exam parameters (ma, rotation time, kvp, exam time & beam width) and number of examinations per year.
35 Example from 2006 (ii) Number of Examinations Based on Hospital A workload (2002) and amended by CT radiographer to anticipated demand. Estimated daily workload: 7 head exams, 7 AP body exams, 8 CAP body exams. Cardiac session half day per week, 8 patients. On remaining half day assume 3 head exams, 3 AP body exams, 4 CAP body exams Assume scanner working 5 days a week, 52 weeks per year. Annual workload: 1612 head exams, 1612 AP body exams, 1872 CAP body exams, 416 cardiac exams
36 Example from 2006 (iii) Exam Parameters Four standard protocols are head abdo-pelvis chest-abdo-pelvis cardiac ac CTA. Protocols have been derived form number of sources. Parameters have been corrected for Toshiba Aquilion 64 MSCT scanner.
37 Example from 2006 (iv) Annual Dose Estimate Scatter diagrams are for Toshiba Aquilion 64 CT scanner. Dose is given per rotation at standard scanning parameters: EG Body: 120 kvp, 200 ma, 1 sec, 32mm beam Once the dose at a given position is calculated it must be scaled for the actual scan parameters estimated for Aquilion 64 head, body & cardiac protocols (see table below). Assume linear scaling factors for slice thickness, mas and beam width. Scaling factor for energy is taken as kvp 2.
38 Example from 2006 (v) Annual Dose is given by: number of exams x rotations/exam x scaling factors x dose/rotation Sc aling Factors Ex am Type & No of Rot s/ kvp mas Sl ice Para met ers Ex am Ex am [120] [100] [32 mm] s Head (PF) k Vp, 330 mas, 4mm Head (HS) k Vp, 330 mas, 8mm Bo dy (AP) k Vp, 200 mas, 32 mm Bo dy (CA P) k Vp, 200 mas, 32 mm Cardiac k Vp, 200 mas, 32 mm
39 Example from 2006 (vi) Annual Doses Corridor Reporting Outside Corner of Ctrl Ceiling above Ceiling above Corner of Wall Wall Room Corridor and Corridor Exam room Exam room and Ceiling Body Dose AB BB CB DB EB FB Per Rot ugy Body (AP) Body (CAP) Cardiac Corridor Reporting Outside Corner of Ctrl Ceiling above Ceiling above Corner of Wall Total Annual Wall Room Corridor and Corridor Exam room Exam room and Ceiling Dose mgy
40 Totally New Facility Identifying proposed workload may be tricky. Understand what the unit is intended to be used df for.
41 Any Facility Beware of using published reference levels as the basis of your calculation. There is a significant time lag between data collection and publication.
42 Reminder The use of CT increases inexorably The indications for CT increase inexorably 0.25 head of population Per Year England USA
43 Why The Increase? Not just CT replacing other techniques Multi Slice CT and slice wars Changes in practice Changes in protocol / acquisition modes Differing volumes Differing numbers of slices Hypertestosis Catch all protocols Standards, What Standards?.
44 Any Facility Beware of using published reference levels as the basis of your calculation. There is a significant time lag between data collection and publication. Reference levels l are not necessarily indicative of current practice Need to know protocols before designing shielding.
45 Thank You The Scary McLarey Family
A simple anthropomorphic phantom used to demonstrate the effectiveness of CT dose modulation functions.
A simple anthropomorphic phantom used to demonstrate the effectiveness of CT dose modulation functions. Lynn Bateman & Peter Hiles North Wales Medical Physics How can we test CT dose modulation? Lynn Bateman
More informationJoint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009
2033-17 Joint ICTP/ Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for CT 3: Practical Experiences Claire-Louise Chapple Freeman Hospital
More informationUnderstanding CT image quality
IAEA RER/9/135 COURSE ON OPTIMIZATION IN COMPUTED TOMOGRAPHY Sofia, Bulgaria, 2017 Understanding CT image quality Dean Pekarovič UMC Ljubljana, Institute of Radiology Quality and Safety office Role of
More informationStaff dose and good prac/ces in CT guided IR procedures. Jonas Andersson Medical Physicist, Ph.D.
Staff dose and good prac/ces in CT guided IR procedures Jonas Andersson Medical Physicist, Ph.D. jonas.s.andersson@vll.se Contents Good prac/ces for staff dose management Staff dose reported in the literature
More informationIterative Reconstruction with Philips idose Characterising Image Quality in Attempting to Realise its Potential
Iterative Reconstruction with Philips idose Characterising Image Quality in Attempting to Realise its Potential Julie Smyth & Philip Doyle Regional Medical Physics Service Outline Preamble Image Quality
More informationIAEA RER/9/135 COURSE ON OPTIMIZATION IN COMPUTED TOMOGRAPHY Sofia, Bulgaria, Tube current modulation and dose reduction : How TCM works
IAEA RER/9/135 COURSE ON OPTIMIZATION IN COMPUTED TOMOGRAPHY Sofia, Bulgaria, 2017 Tube current modulation and dose reduction : How TCM works Dean Pekarovič UMC Ljubljana, Institute of Radiology Quality
More information2
1 2 3 4 5 6 7 1. Bauhs, J. A., Vrieze, T. J., Primak, A. N., Bruesewitz, M. R., & McCollough, C. H. (2008). CT Dosimetry: Comparison of Measurement Techniques and Devices1. Radiographics, 28(1), 245-253.
More informationA method for calculating the dose length product from CT DICOM images
The British Journal of Radiology, 84 (2011), 236 243 A method for calculating the dose length product from CT DICOM images 1 I A TSALAFOUTAS, PhD and 2 S I METALLIDIS, MSc 1 Medical Physics Department,
More informationStandard. Substitute Test or Procedure. Required Test or. 1 Scan Increment Accuracy. Initially and Annually Initially and Annually
Manufacturer s Recommendations for Alternate Dental CBCT QA Program VaTech: Model PAX i3d, i3d Green, i3d Smart Table 6 Medical Physicist s Computed Tomography QC Survey Required Test or Item Procedure
More informationQUALITY CONTROL AND PATIENT DOSES FROM X-RAY EXAMINATIONS IN SOME HOSPITALS IN THAILAND
QUALITY CONTROL AND PATIENT DOSES FROM X-RAY EXAMINATIONS IN SOME HOSPITALS IN THAILAND P. Plainoi, W. Diswath, N. Manatrakul Ministry of Public Health, Nonthaburi, Thailand XA0101612 Abstract Quality
More informationMadero Ote. 686, Centro Histórico, C.P Morelia, Michoacán México. #300, Col. Cuauhtémoc, C.P Morelia, Michoacán, México
Verification of and DLP values for a head tomography reported by the manufacturers of the CT scanners, using a CT dose profiler probe, a head phantom and a piranha electrometer Edith Castillo Corona 1,2,
More informationRequest for Proposals
Request for Proposals Reference: ERDFIAI2012-4006A Caring First Ltd, is a limited liability company providing healthcare services. The company is currently commissioning a private new-build hospital complete
More informationComparison of Measured Values of CTDI and DPL with Standard Reference values of Different CT Scanners for dose Management
2017 IJSRST Volume 3 Issue 1 Print ISSN: 2395-6011 Online ISSN: 2395-602X Themed Section: Science and Technology Comparison of Measured Values of CTDI and DPL with Standard Reference values of Different
More informationNEMA XR 25 COMPUTED TOMOGRAPHY DOSE CHECK
NEMA XR 25 COMPUTED TOMOGRAPHY DOSE CHECK NEMA Standards Publication XR 25-2010 Computed Tomography Dose Check Published by: National Electrical Manufacturers Association 1300 North 17th Street, Suite
More informationRSNA 2006 November 26 to December 1 Chicago. Guest author for ImPACT Dr. Koos Geleijns, Medical Physicist, Leiden University Medical Center.
RSNA 2006 November 26 to December 1 Chicago Guest author for ImPACT Dr. Koos Geleijns, Medical Physicist, Leiden University Medical Center. Once again, more than 60,000 participants (including professional
More informationTrue comfort and flexibility with the power of 3T.
True comfort and flexibility with the power of 3T. With a large 71 cm aperture and the quietest exams in the industry, the Vantage Titan 3T is the most comfortable 3T MRI system for all of your patients.
More informationIn recent years, CT technology has undergone profound
PATIENT SAFETY Y.T. Niu M.E. Olszewski Y.X. Zhang Y.F. Liu J.F. Xian Z.C. Wang Experimental Study and Optimization of Scan Parameters That Influence Radiation Dose in Temporal Bone High-Resolution Multidetector
More informationBrilliance CT 64-channel configuration
Brilliance CT 64-channel configuration with Essence technology The Brilliance CT 64-channel configuration is designed to help you conduct the most advanced multislice CT studies possible. These systems
More informationEquipment Quality Control for Digital Radiography February 22, Imaging Physics CancerCare Manitoba
Equipment Quality Control for Digital Radiography February 22, 2018 Imaging Physics CancerCare Manitoba Purpose An equipment quality control (QC) program establishes baseline performance levels, tracks
More informationFDA MAMMOSCAN FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM MAMMOGRAPHY
MAMMOGRAPHY MAMMOSCAN FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM Biopsy Attachment џ MAMMOSCAN is a digital scanning technology development taking a special place in ADANI Medical X-ray product line. џ Its
More informationPrecision, power, and productivity
Precision, power, and productivity Philips Brilliance CT Big Bore oncology configuration Philips Healthcare continues its tradition of innovation in oncology with a CT solution that specifically addresses
More informationLightSpeed 16 With Xtream CT Scanner System
Page 1 Introduction The CT Scanner with Xtream TM technology is the next step in GE s CT Continuum and its awardwinning LightSpeed CT platform. The CT Scanner routinely acquires 16 sub-millimeter slices
More informationTechnologist Quality Control Procedures
Technologist Quality Control Procedures The specific procedures for the Technologist Quality Control Program are those specified in the most current ACR MRI QC Manual. WEEKLY MRI EQUIPMENT QUALITY CONTROL
More informationEM1. Transmissive Optical Encoder Module Page 1 of 8. Description. Features
Description Page 1 of 8 The EM1 is a transmissive optical encoder module. This module is designed to detect rotary or linear position when used together with a codewheel or linear strip. The EM1 consists
More informationEM1. Transmissive Optical Encoder Module Page 1 of 8. Description. Features
Description Page 1 of 8 The EM1 is a transmissive optical encoder module designed to be an improved replacement for the HEDS-9000 series encoder module. This module is designed to detect rotary or linear
More informationGuidelines for Quality Assurance in Radiation Protection for Diagnostic X-Ray Facilities: Large X-Ray Facilities
Guidelines for Quality Assurance in Radiation Protection for Diagnostic X-Ray Facilities: Large X-Ray Facilities J L Poletti 1995/1 NRL Report GUIDELINES FOR QUALITY ASSURANCE IN RADIATION PROTECTION FOR
More informationCT Numbers: Think of a number, double it, add 20, divide by 4. Jane Edwards Royal Free Hospital, London
CT Numbers: Think of a number, double it, add 20, divide by 4 Jane Edwards Royal Free Hospital, London Background 2 different manufacturers CT scanners available on site: GE Lightspeed Plus; Philips Brilliance
More informationMultiparametric MRI Prostate Imaging Protocol November 2015 Full Acquisition Protocol with Parameters GE 3T Magnet with Software Version DV25
3Plane Loc SSFSE Multiparametric MRI Prostate Imaging Protocol November 2015 Full Acquisition Protocol with Parameters GE 3T Magnet with Software Version DV25 Save Series Scan After acquisition, scroll
More information2012 Computed Tomography
2012 Computed Tomography QUALITY CONTROL MANUAL Radiologist s Section Radiologic Technologist s Section Medical Physicist s Section 2012 Computed Tomography QUALITY CONTROL MANUAL Radiologist s Section
More informationQuality Assurance Implementation at the Roberts Proton Therapy Center. James McDonough 3 August 2013
Quality Assurance Implementation at the Roberts Proton Therapy Center James McDonough 3 August 2013 1 Roberts Proton Therapy Center Machine configuration and layout 4 gantries, 1 fixed beam line, 1 research
More informationRide the Lightning. A Review of Radiographic Physics
Ride the Lightning A Review of Radiographic Physics Overview Basic principles of x-ray production Imaging equipment Quality Control Technical Factors Image QA Preparing for the registry The ARRT just wants
More informationPractical Application of the Phased-Array Technology with Paint-Brush Evaluation for Seamless-Tube Testing
ECNDT 2006 - Th.1.1.4 Practical Application of the Phased-Array Technology with Paint-Brush Evaluation for Seamless-Tube Testing R.H. PAWELLETZ, E. EUFRASIO, Vallourec & Mannesmann do Brazil, Belo Horizonte,
More informationMore Info at Open Access Database Process Control for Computed Tomography using Digital Detector Arrays
Digital Industrial Radiology and Computed Tomography (DIR 2015) 22-25 June 2015, Belgium, Ghent - www.ndt.net/app.dir2015 More Info at Open Access Database www.ndt.net/?id=18082 Process Control for Computed
More informationOptima CT660 Optima CT660
Product Data Sheet Rev 2 Page 1/18 Optima CT660 Optima CT660 Milwaukee, USA - Fax: 1 262 544 3384 Tokyo, Japan - Fax: 81 425 85 5490 Paris, France - Fax: 33 1 30 70 94 35 Optima TM CT660 Product Description
More informationEM1. Transmissive Optical Encoder Module Page 1 of 9. Description. Features
Description Page 1 of 9 The EM1 is a transmissive optical encoder module designed to be an improved replacement for the HEDS-9000 series encoder module. This module is designed to detect rotary or linear
More informationDICOM Correction Proposal
DICOM Correction Proposal STATUS Final Text Date of Last Update 2015/11/16 Person Assigned Submitter Name Ulrich Busch (ulrich.busch@varian.com) Michael Moyers Submission Date 2011/05/19 Correction Number
More informationFFDM Quality Control in Canada - a Vendor Neutral Approach
FFDM Quality Control in Canada - a Vendor Neutral Approach Rasika Rajapakshe, PhD, FCCPM BC Cancer Agency-Center for the Southern Interior Kelowna, BC, Canada 2011 Joint AAPM/COMP Meeting August 4, 2011
More informationHigh Frequency X-Ray Technology. Wide Screen Control with Touch-Technology
High Frequency X-Ray Technology Wide Screen Control with Touch-Technology The Q-VISION HF Series TM The Q-VISION Series TM integrates innovative high frequency design along with superb functionality, resulting
More information-Technical Specifications-
Annex I to Contract 108733 NL-Petten: the delivery, installation, warranty and maintenance of one (1) X-ray computed tomography system at the JRC-IET -Technical Specifications- INTRODUCTION In the 7th
More informationSHF Communication Technologies AG. Wilhelm-von-Siemens-Str. 23D Berlin Germany. Phone Fax
SHF Communication Technologies AG Wilhelm-von-Siemens-Str. 23D 12277 Berlin Germany Phone +49 30 772051-0 Fax +49 30 7531078 E-Mail: sales@shf.de Web: http://www.shf.de Datasheet SHF 11220 A 60 GBaud PAM4
More informationBuilt to do More. Brivo CT slice CT product data sheet
GE Healthcare Built to do More. Brivo CT385 16-slice CT product data sheet Built for your Confidence. BIG Performance, small footprint Smart CT Desk Energy Saving Mode Simply Advance Hilight Scintillator
More informationNEC Display Solutions of America, Inc. U300X Installation Guide Desktop and Ceiling Mount v1.3
Contents Product Description, Lens Specs, Notes and Formulas Page 1 Diagrams and Distance Charts Ceiling Mounted Installation Page 2 Desktop Setup Page 3 Cabinet Dimensions Top, Front and Right Side Page
More informationJEFFERSON COLLEGE. Image Intensification & Equipment
JEFFERSON COLLEGE COURSE SYLLABUS RAD175 Image Intensification & Equipment 3 Credit Hours Revised by: Janet E. Akers BS RT (R)(M) Date: October 3, 2013 Kenny Wilson, Director, Health Occupation Programs
More informationA new shielding calculation method for X-ray computed tomography regarding scattered radiation
Radiol Phys Technol (2017) 10:213 226 DOI 10.1007/s12194-016-0387-9 A new shielding calculation method for X-ray computed tomography regarding scattered radiation Hiroshi Watanabe 1,2 Kimiya Noto 3 Tomokazu
More informationLesson 07: Ultrasound Transducers. This lesson contains 62 slides plus 16 multiple-choice questions.
Lesson 07: Ultrasound Transducers This lesson contains 62 slides plus 16 multiple-choice questions. Accompanying text for the slides in this lesson can be found on pages 33 through 42 in the textbook:
More informationScope: All CT staff technologist
APPROVED BY: Radiology Technical Director Page 1 of 6 Purpose: The QC program assesses relative changes in system performance as determined by the technologist, service engineer, qualified medical physicist,
More informationDepartment of Medical Physics, Graduate School of Nuclear and Allied Science, University of Ghana, Accra - Ghana, West Africa
2018 IJSRST Volume 4 Issue 5 Print ISSN: 2395-6011 Online ISSN: 2395-602X Themed Section: Science and Technology Assessment of Adult Patient Radiation Dose and Image Quality in CT Examinations Performed
More information21 rue La Noue Bras de Fer Nantes - France Phone : +33 (0) website :
21 rue La Noue Bras de Fer - 44200 Nantes - France Phone : +33 (0) 240 180 916 - email : info@systemplus.fr - website : www.systemplus.fr 2012 September - Version 1 Written by: Maher Sahmimi DISCLAIMER
More information2017 Computed Tomography
2017 Computed Tomography QUALITY CONTROL MANUAL Radiologist s Section Radiologic Technologist s Section Qualified Medical Physicist s Section 2017 Computed Tomography QUALITY CONTROL MANUAL Radiologist
More informationThe Black Piranha Easy & Fast X-ray Quality Control
From Radiation to Information The Black Piranha Easy & Fast X-ray Quality Control The Black Piranha with accessories makes X-ray QA easy and fast. Connection is automatic just plug n play. The Quick Check
More informationabc Mark Scheme Mathematics 4302 Specification B General Certificate of Secondary Education Module 5 Paper 1 Tier F 43005/1F
Version : 1.0: 0608 abc General Certificate of Secondary Education Mathematics 4302 Specification B Module 5 Paper 1 Tier F 43005/1F Mark Scheme 2008 examination - June series Mark schemes are prepared
More informationImpact of DMD-SLMs errors on reconstructed Fourier holograms quality
Journal of Physics: Conference Series PAPER OPEN ACCESS Impact of DMD-SLMs errors on reconstructed Fourier holograms quality To cite this article: D Yu Molodtsov et al 2016 J. Phys.: Conf. Ser. 737 012074
More informationUArm. Series. Analog or Digital U-Arm System
UArm Analog or Digital U-Arm System Series A world class direct digital imaging system with positioning flexibility and ease of use for all procedures. Eighty Years of Excellence UArm Series POSITIONING
More informationPhantom Test Guidance for Use of the Small MRI Phantom for the MRI Accreditation Program
Phantom Test Guidance for Use of the Small MRI Phantom for the MRI Accreditation Program 1 Contents 0.0 INTRODUCTION 4 0.1 Overview and Purpose 4 0.2 The Phantom 4 0.3 The Required Images 5 0.4 The Image
More informationX-Ray Machines, CT Scanners, MRIs: The Pivotal Role of the GE Research and Development Center
CT Scanner Protoype at UCSF Medical Center 1976 X-Ray Machines, CT Scanners, MRIs: The Pivotal Role of the GE Research and Development Center by Walter L. Robb Early Years In 1895, Professor Wilhelm Conrad
More informationPerformance Evaluation of Industrial Computed Radiography Image Display System
Performance Evaluation of Industrial Computed Radiography Image Display System More info about this article: http://www.ndt.net/?id=21169 Lakshminarayana Yenumula *, Rajesh V Acharya, Umesh Kumar, and
More informationApplication Note AN-708 Vibration Measurements with the Vibration Synchronization Module
Application Note AN-708 Vibration Measurements with the Vibration Synchronization Module Introduction The vibration module allows complete analysis of cyclical events using low-speed cameras. This is accomplished
More informationHDMI Extender via 4 Single-mode fibers 4 x LC Connector Extends HDMI + RS232 full-duplex up to 1500 meters
Description AMRT-FD-05K-4LC extender enables PC HDMI and RS232 link to far end display monitor through fibers, and the maximum communication distance is up to 1500 meters. AMRT-FD-05K-4LC is pure hardware
More informationCONTENTS 1. GENERAL INFORMATION INTRODUCTION PRODUCT INFORMATION DESCRIPTION OF TESTS CHANNEL BANDWIDTH...
Test Report Number : GETEC-E3-09-083 Page 2 / 33 CONTENTS. GENERAL INFORMATION...4 2. INTRODUCTION...5 3. PRODUCT INFORMATION...6 3. DESCRIPTION OF EUT...6 3.2 SUPPORT EQUIPMENT / CABLES USED...7 3.3 MODIFICATION
More informationCollimation and Light Field Brightness
Collimation and Light Field Brightness Physics and Reasoning Principle 1: Primary X Ray beam is blocked by breast support table. Principle 2: Need to properly imageas as much breast tissue as possible.
More informationInstallation Evaluation Form
265 S. Federal Hwy Unit 163 Deerfield Beach, FL 33441-4146 Phone (954) 425-0199 Fax (954) 827-2408 Toll-Free 1-866-SCAN-SND info@scansound.com www.scansound.com Please have someone at the MRI imaging center
More informationMechanical aspects, FEA validation and geometry optimization
RF Fingers for the new ESRF-EBS EBS storage ring The ESRF-EBS storage ring features new vacuum chamber profiles with reduced aperture. RF fingers are a key component to ensure good vacuum conditions and
More informationPhotodiode Detector with Signal Amplification
107 Bonaventura Dr., San Jose, CA 95134 Tel: +1 408 432 9888 Fax: +1 408 432 9889 www.x-scanimaging.com Linear X-Ray Photodiode Detector Array with Signal Amplification XB8801R Series An X-Scan Imaging
More informationELECTRICAL PERFORMANCE REPORT
CIRCUITS & DESIGN ELECTRICAL PERFORMANCE REPORT DENSIPAC 4 ROW Date: 06-12-2006 Circuits & Design EMEA Circuits & Design 1/21 06/12/2006 1 INTRODUCTION... 3 2 CONNECTORS, TEST BOARDS AND TEST EQUIPMENT...
More informationMini Micro Pulse Lidar System
Mini Micro Pulse Lidar System MiniMPL-532-C Sensor Suite Operations Manual Version: June 2016 THIS PAGE INTENTIONALLY LEFT BLANK 2 Table of Contents MINIMPL SENSOR SUITE SYSTEM: RECORD OF PURCHASE... 4
More informationLTC 113x & LTC123x FlexiDome Series Fixed Dome Cameras
LTC 113x & LTC123x FlexiDome Series Fixed Dome Cameras Eng Installation Instructions F D E NL I IMPORTANT SAFEGUARDS 1. Read Instructions All the safety and operating instructions should be read before
More informationPart 1: Introduction to computer graphics 1. Describe Each of the following: a. Computer Graphics. b. Computer Graphics API. c. CG s can be used in
Part 1: Introduction to computer graphics 1. Describe Each of the following: a. Computer Graphics. b. Computer Graphics API. c. CG s can be used in solving Problems. d. Graphics Pipeline. e. Video Memory.
More informationNuclear Associates and
Nuclear Associates 76-907 and 76-908 AAPM MRI Phantoms Users Manual March 2005 Manual No. 38616 Rev. 3 2003, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks
More informationVascular. Development of Trinias FPD-Equipped Angiography System. 1. Introduction. MEDICAL NOW No.73 (2013.2) Yoshiaki Miura
Vascular Development of Trinias FPD-Equipped Angiography System Medical Systems Division, Shimadzu Corporation Yoshiaki Miura 1. Introduction Shimadzu has developed Trinias (one ceiling-mounted type C12
More informationINSTRUMENT CATHODE-RAY TUBE
Instrument cathode-ray tube D14-363GY/123 INSTRUMENT CATHODE-RAY TUBE mono accelerator 14 cm diagonal rectangular flat face internal graticule low power quick heating cathode high brightness, long-life
More informationNon-Destructive Examination Benches and Analysis Laboratories in support to the Experimental Irradiation Process in the Future Jules Horowitz MTR
Non-Destructive Examination Benches and Analysis Laboratories in support to the Experimental Irradiation Process in the Future Jules Horowitz MTR D. Parrat 1, P. Kotiluoto 2, T. Jäppinen 2, C. Roure 1,
More informationFunctional overview. System configuration FAQ
Vdc 206-13-40 100/110/115/12 110/125V D C MDE IN JPN RSM100 (Relay Setting and Monitoring System) The PC interface software RSM100 allows users to access Toshiba GR-series relays from a local or remote
More informationOperating Procedures for RECO1 & RECO2
Operating Procedures for RECO1 & RECO2 Reconstruction of Data 1) Open CT Pro 3D a. Remember, when the computer you are working on is receiving files from the scanner, it cannot reconstruct data b. You
More informationAPPLICATION NOTE VACUUM FLUORESCENT DISPLAY MODULE
AN-E-3237A APPLICATION NOTE VACUUM FLUORESCENT DISPLAY MODULE GRAPIC DISPLAY MODULE GP92A1A GENERAL DESCRIPTION FUTABA GP92A1A is a graphic display module using a FUTABA 128 64 VFD. Consisting of a VFD,
More informationAcoustical design of Shenzhen Concert Hall, Shenzhen China
Acoustical design of Shenzhen Concert Hall, Shenzhen China K. Oguchi and Y. Toyota Nagata Acoustics Inc., 2130 Sawtelle Blvd., Suite 307A, Los Angeles, CA 90025, USA oguchi@nagata.co.jp 321 The Shenzhen
More informationCharacterization and improvement of unpatterned wafer defect review on SEMs
Characterization and improvement of unpatterned wafer defect review on SEMs Alan S. Parkes *, Zane Marek ** JEOL USA, Inc. 11 Dearborn Road, Peabody, MA 01960 ABSTRACT Defect Scatter Analysis (DSA) provides
More informationExam 6. / / Month Day Year. Lung CT With Contrast Protocol: FORCE Comple on Form. Acros c: Par cipant ID #: Date: Technician ID: SECTION A: Screening
SECTION A: Screening 1. Did par cipant consent to the lung CT? Par cipant ID #: Technician ID: Date: Acros c: / / Month Day Year STOP. Do not perform any CT scan. Refused (a er consen ng) STOP. Do not
More informationMSK Imaging Fundamentals
MSK Imaging Fundamentals Goals Improve image quality Provide best possible product for our customers Patient Referring clinician Radiologist Reduce number of callback cases Goal reduction of 50% in 3 months
More informationPreparing for remote data collection at NE-CAT
Preparing for remote data collection at NE-CAT Important Note: The beamtime and remote login privileges are intended just for you and your group. You are not allowed to share these with any other person
More informationBUREAU OF ENERGY EFFICIENCY
Date: 26 th May, 2016 Schedule No.: 11 Color Televisions 1. Scope This schedule specifies the energy labeling requirements for color televisions with native resolution upto 1920 X 1080 pixels, of CRT,
More informationScanning System S-2100
2D laser measurement system The fastest 2D laser measurement system in the world 119 m range Scan rate >1 million points/sec 360 vertical field of view System description The PENTAX Scanning System S-2100
More informationGaAs MMIC Double Balanced Mixer
Page 1 The is a passive double balanced MMIC mixer. It features excellent conversion loss, superior isolations and spurious performance across a broad bandwidth, in a highly miniaturized form factor. Low
More informationThe field cage for a large TPC prototype
EUDET The field cage for a large TPC prototype T.Behnke, L. Hallermann, P. Schade, R. Diener December 7, 2006 Abstract Within the EUDET Programme, the FLC TPC Group at DESY in collaboration with the Department
More informationMRI in Clinical Practice
MRI in Clinical Practice MRI in Clinical Practice Gary Liney With 62 Figures Gary Liney, PhD MRI Lecturer University of Hull Centre for MR Investigations Hull Royal Infirmary Hull UK British Library Cataloguing
More informationCARESTREAM DIRECTVIEW Elite CR System
CARESTREAM DIRECTVIEW Elite CR System Improve workflow, productivity, and patient throughput. The CARESTREAM DIRECTVIEW Elite CR System is small, easy to install and easy to use. This powerful distributed
More informationPT-CW230 S P E C F I L E. As of November Specifications and appearance are subject to change without notice.
Product Number : Product Name : As of vember 2012. Specifications and appearance are subject to change without notice. 1 / 1 3 Specifications Main unit Power supply Power consumption DLP chip Panel size
More informationCARESTREAM DIRECTVIEW Elite CR System
CARESTREAM DIRECTVIEW Elite CR System Improve workflow, productivity, and patient throughput. The CARESTREAM DIRECTVIEW Elite CR System is small, easy to install and easy to use. This powerful distributed
More informationIntra-oral X-ray unit with high frequency DC generator. Product Data
Intra-oral X-ray unit with high frequency DC generator Tubehead Features Product Data High voltage circuit type High frequency (DC) High voltage frequency 30 khz Rated output voltage 60-65 - 70 kvp ± 8%
More informationSynchronization circuit with synchronized vertical divider system for 60 Hz TDA2579C
FEATURES Synchronization and horizontal part Horizontal sync separator and noise inverter Horizontal oscillator Horizontal output stage Horizontal phase detector (sync to oscillator) Triple current source
More informationI. FILL IN THE BLANKS
1 BHARAT SEVAK SAMAJ NATIONAL DEVELOPMENT AGENCY, PROMOTED BY GOVERNMENT OF INDIA CENTRAL BOARD OF EXAMINATIONS BSS NATIONAL VOCATIONAL EDUCATION MISSION AHE004-X-RAY TECHNICIAN ONE YEAR EXAMINATION 2011-12
More informationEssentials of Digital Imaging
Essentials of Digital Imaging Module 3 Transcript 2016 ASRT. All rights reserved. Essentials of Digital Imaging Module 3 Display 1. ASRT Animation 2. Welcome Welcome to the Essentials of Digital Imaging
More information45 MW, 22.8 GHz Second-Harmonic Multiplier for High-Gradient Tests*
US High Gradient Research Collaboration Workshop. SLAC, May 23-25, 2007 45 MW, 22.8 GHz Second-Harmonic Multiplier for High-Gradient Tests* V.P. Yakovlev 1, S.Yu. Kazakov 1,2, and J.L. Hirshfield 1,3 1
More informationLIGHT PROTON THERAPY PROJECT
17 th of MAY 2018 LIGHT PROTON THERAPY PROJECT Yevgeniy Ivanisenko on behalf of ADAM team FORM-01040-A AVO-ADAM Advanced Oncotherapy (AVO) is a public company ADAM is R&D center of AVO ~ 100 employees
More informationQuadrupoles have become the most widely used
ARTICLES A Novel Tandem Quadrupole Mass Analyzer Zhaohui Du and D. J. Douglas Department of Chemistry, University of British Columbia, Vancouver, B. C., Canada A new tandem mass analyzer is described.
More informationPracticum 3, Fall 2010
A. F. Miller 2010 T1 Measurement 1 Practicum 3, Fall 2010 Measuring the longitudinal relaxation time: T1. Strychnine, dissolved CDCl3 The T1 is the characteristic time of relaxation of Z magnetization
More informationModeling Microwave Waveguide Components: The Tuned Stub
Modeling Microwave Waveguide Components: The Tuned Stub Roger W. Pryor, Ph.D. 1 1 Pryor Knowledge Systems *Corresponding author: 4918 Malibu Drive, Bloomfield Hills, MI, 48302-2253, rwpryor@pksez1.com
More informationIn PS 3.3, Section C RT Ion Beams Session Record Module, add the following attributes and make the changes indicated:
DICOM Correction Proposal STATUS Letter Ballot Date of Last Update 2016/03/18 Person Assigned Ulrich Busch Submitter Name Olivier Vierlinck Submission Date 2015/02/03 Correction Number CP-1013 Log Summary:
More informationProcedures for conducting User QA on the scanner
Procedures for conducting User QA on the scanner Sample setup: The User QA phantom is clearly labeled and is stored on one of the shelves to the side of the magnet. Note proper orientation of the bottle,
More informationDICOM Conformance Statement
Philips Medical Systems DICOM Conformance Statement Integris Systems with High Speed DICOM Image Interface MCV 2974 MCV 3621 Cardiac DICOM XA MCV 3761 Vascular DICOM XA and DICOM RIS Interface MCV 3031
More informationECE 4220 Real Time Embedded Systems Final Project Spectrum Analyzer
ECE 4220 Real Time Embedded Systems Final Project Spectrum Analyzer by: Matt Mazzola 12222670 Abstract The design of a spectrum analyzer on an embedded device is presented. The device achieves minimum
More information