1 Guidelines for Quality Assurance in Radiation Protection for Diagnostic X-Ray Facilities: Large X-Ray Facilities J L Poletti 1995/1 NRL Report

2 GUIDELINES FOR QUALITY ASSURANCE IN RADIATION PROTECTION FOR DIAGNOSTIC X-RAY FACILITIES: LARGE X-RAY FACILITIES NATIONAL RADIATION LABORATORY MINISTRY OF HEALTH CHRISTCHURCH NEW ZEALAND

3 CONTENTS Page ABSTRACT 1 INTRODUCTION 1 2 IMPORTANT CONCEPTS FOR QA PROGRAMMES 3 3 SPECIFIC REQUIREMENTS FOR LARGE FACILITIES Automatic film processors X-ray generators X-ray tubes Automatic exposure controls (AEC) Light beam diaphragms (LBD) X-ray cassettes X-ray image intensifier systems Image intensifiers Fluoroscopic television chains Fast film changers Cine systems and small format cameras Digital subtraction imaging (DSI) systems Computed tomography scanners Mammography machines Tomography machines Mobile radiographic equipment Mobile image intensifier equipment Grids Protective equipment Darkrooms Viewboxes Technique charts Dose measurements Approval of the QA programme 17 4 AN OUTLINE QA PROGRAMME FOR LARGE FACILITIES 17. BIBLIOGRAPHY 21

4

5 ABSTRACT The Code of safe practice for the use of x-rays in medical diagnosis (NRL C5 1 ) requires that each x-ray facility has an appropriate quality assurance (QA) programme in radiation protection. The objective of the quality assurance programme is to ensure accurate diagnosis and to ensure that doses are kept as low as reasonably achievable. In addition the quality assurance programme should ensure compliance with NRL C5 1 at all times. This requires an inhouse system of regular checks and procedures. These guidelines are intended to assist x-ray facilities to comply with the NRL C5 requirement, by outlining the features considered to be appropriate for a QA programme. The concepts involved in QA programmes are described, and details are given of the types of tests to be performed. An indication of QA equipment required is given, and suggested test frequencies are outlined.

6

7 1 INTRODUCTION For the purposes of these guidelines, a large x-ray facility is defined as having at least one of the following (in addition to general radiography rooms): - More than one fluoroscopy room - More than one film processor - A mammography machine - A digital fluoroscopy system - A CT scanner The use of ionizing radiation in New Zealand is controlled by the Radiation Protection Act (1965). Licences under this Act may be granted for a number of purposes, including medical diagnosis. All licences for the use of x-rays for medical diagnosis include a condition that the requirements of the Code of safe practice for the use of x-rays in medical diagnosis (NRL C5 1 ) are met. Among the requirements of NRL C5 are a quality assurance programme in radiation protection. The objective of the quality assurance programme is to ensure accurate diagnosis and to ensure that doses are kept as low as reasonably achievable. In addition the quality assurance programme should ensure compliance with NRL C5 1 at all times. This requires an in-house system of regular checks and procedures as detailed in these guidelines. In addition, and completely independent from the quality assurance programme, each facility is required to have a complete radiation protection survey performed at least once every four years. For those facilities with image intensifier systems, a CT scanner or mammography machine, these must be done every two years. The radiation protection survey is intended to focus on radiation safety and checks for compliance with the appropriate requirements of NRL C5. As part of this, the radiation protection survey acts as an external independent audit of the quality assurance programme. The tests and measurements made during a radiation protection survey cannot be considered to be part of the quality assurance programme. The radiation protection survey is currently performed free of charge to the facility by NRL staff. A qualified health physicist is also permitted to do radiation protection surveys, provided that the protocol and equipment used are acceptable to NRL. A comprehensive radiation protection quality assurance programme requiring some test equipment is appropriate for large facilities (as defined above). The general requirements for a quality assurance programme in radiation protection, as given in NRL C5 and of relevance to large x-ray facilities, are summarised as follows: (Note that should and shall have specified meanings within NRL C5, but not elsewhere in these guidelines.) 1 The principal licensee for any facility that uses x-rays for medical diagnosis shall ensure that a suitable programme of quality assurance (with respect to radiation protection), is instituted and maintained. 1

8 2 The programme shall ensure as a primary goal, accurate and timely diagnosis. As secondary goals the programme shall ensure minimisation of radiation exposure and risk and of discomfort and cost to patient and community. These secondary goals shall always be balanced against the primary goal. 3 The programme shall comprise such routine checks and procedures as are required to give reasonable confidence in the continuing compliance with this Code of Practice. The programme shall be approved by a qualified health physicist, to ensure that the quality control procedures are sufficient to ensure compliance with this Code. The programme shall include quality control of x-ray film processing facilities. Note: A programme is not to be confused with a radiation protection survey. 4 There shall be a well-defined responsibility and reporting structure, appropriate to the size and scope of the facility. Each staff member shall routinely review the results of checks for which they are responsible and report summary results to their superior. Any anomalous check shall be reported immediately. Each staff member shall be responsible for the maintenance of the programme by any personnel under his/her control. 5 Procedures should be standardised and set down in protocols or local rules (a quality assurance manual) wherever possible. 6 All equipment shall be checked at suitable regular intervals to ensure it is operating within suitable tolerances of accuracy and consistency. The tests performed and their frequency shall be approved by a qualified health physicist. All measurements and maintenance shall be recorded in an equipment log. As well as routine tests any faults or breakdowns shall be logged and reported to superiors. 7 Acceptance tests shall be performed on all new equipment to (a) (b) (c) ensure that it meets the manufacturer's specifications; ensure that it complies with this Code; establish baseline data for subsequent quality assurance. 8 Control charts shall be established for all parameters measured. Control limits shall be established for all parameters. If a measured value of any parameter exceeds a control limit, action shall be taken to correct the parameter. 9 A retake analysis shall be performed at regular intervals to monitor the effectiveness of the programme. 10 The frequency with which a particular parameter is tested should be determined by both the likelihood and the consequences of an error beyond the acceptable tolerances. 2

9 11 The programme should conform to the procedures and tolerances given in NCRP report 99 (National Council on Radiation Protection and Measurements, 1988) 2 or Assurance of quality in the diagnostic x-ray department The programme for CT facilities should include the recommendations given in IEC It should be noted that a properly implemented QA programme will result in significant benefits. These will include reduced costs due to reduced repeat rates, increased accuracy of diagnosis, reduced equipment down-time and increased morale and job satisfaction for staff 5,6,7,8. 2 IMPORTANT CONCEPTS FOR QA PROGRAMMES Acceptance testing Whenever a new piece of equipment is installed, acceptance tests should be performed. This may take a few days for complex equipment 9. The purposes of the acceptance tests are three-fold. - First it is important to check that the equipment meets the specifications set out in the purchase contract with the supplier. (For this reason it is important that purchase contracts clearly state the requirements for the equipment.) - Second, the acceptance tests establish baseline values for the parameters that are to be monitored during the QA programme. - Third, the acceptance tests will show whether the machine complies with the requirements of NRL C5, at installation. Control charts Once baseline values for the QA parameters have been established a system has to be set up to ensure that these parameters are maintained to within acceptable tolerances. Control charts are an essential part of the system 7. A control chart is a plot of the measured parameter with time. A typical control chart is shown in figure 1. The control chart has three horizontal lines, giving the desired value of the parameter and the allowable limits. The desired value is determined from the acceptance tests, or from tests made at the start of the QA programme (first ensuring that the equipment is performing correctly). Allowable limits may come from many sources, such as published protocols, regulatory requirements, or the effects on other parameters that may be affected. 3

10 Figure 1. An example of a control chart Upper limit 96 Value 94 Control value /2/94 2/2/94 Lower limit 3/2/94 4/2/94 5/2/94 8/2/94 9/2/94 10/2/94 11/2/94 12/2/94 15/2/94 16/2/94 17/2/94 18/2/94 19/2/94 22/2/94 23/2/94 24/2/94 25/2/94 26/2/94 Date Each time a measurement is made the result is plotted on the control chart. Whenever a parameter goes outside the control limit, the measurement should be repeated. If no mistake has been made and the parameter is still out of control, then immediate action must be taken to correct the parameter. If this is not done, then the entire QA programme is a waste of time. It is sometimes possible to observe trends in the data that suggest that a parameter will become out of control in the near future (as in the example above). It is advisable to take corrective action at this stage, rather than to wait until the parameter is out of control. Reject/retake analysis A reject is defined as any film rejected by the department as scrap for any reason 7,10,11. A retake is a patient film that has to be retaken because of an error. Retakes are only a part of all rejected films. Rejects may be due to three main causes. 1 Retakes 2 Films wasted due to other causes, such as fogging, equipment breakdowns, etc 3 Trial films to establish exposure settings that are not viewed by a radiologist as part of the diagnosis. (For example, some films in a tomographic series.) Analysis of the rejected films over a period of time will enable causes of rejects to be determined and improvements to be made. The more detailed the analysis performed, then the more information will be obtained. The simplest approach is just to collect all rejects and to express the rate as a percentage of all films used. This gives a baseline value for the reject rate for future reference. 4

11 This may be done continually, or for shorter periods on a regular basis. The minimum period should be at least six weeks. The first two weeks' results are generally discarded to eliminate the "startup effect". If the rejects are sorted by work area or room, then corrective action may be directed to where it is most needed. If they are categorised by reason for rejection, then effort may be directed at reducing the most common errors. Further subdivision by anatomical region may help determine which examinations are causing the most difficulty. Finally, analysis by staff member may be used to direct training to staff with difficulties in particular areas. Equipment logs Each x-ray room should have an equipment log 7. To be recorded in this log are the dates and details of all QA corrective action, breakdowns and routine servicing. This should include, if possible, an estimate of the downtime. Responsibility for the QA programme In every department, a particular person should be assigned the overall responsibility for the QA programme 7. Clearly, this person should be familiar with all the x-ray equipment and with the principles of radiology quality assurance. Where appropriate, it is well worthwhile to establish a committee to oversee the programme. This committee should include representatives of all the areas involved, MRTs, radiologists, medical physicists, service personnel and management. (Note that NRL C5 requires that the QA programme be approved by a qualified health physicist or directly by NRL.) The QA manual The procedures involved for the entire QA programme must be recorded in a manual 12. This is to ensure that all tests are carried out in a consistent and reproducible manner. If not, then parameters may appear to be out of control, where in fact the change was due to differences in the measurement technique. Clearly the manual should be readily available to all personnel concerned. Staff training At the start of the QA programme, staff will need to be given sufficient training to carry out the QA procedures for which they are responsible. Periodic refresher training will be required also during the programme to keep staff up to date with development of the programme and to provide feedback on its effectiveness and acceptability to staff. 3 SPECIFIC REQUIREMENTS FOR LARGE FACILITIES All equipment and accessories need to be included in the QA programme, although the frequencies may be very different for each item. This chapter describes the parameters that need to be tested for each item of equipment and gives the suggested frequency of tests. Where appropriate, recommendations for 5

12 test equipment are given. Specific step-by-step descriptions and tolerance values of all the tests are not given. Such detailed information may be found in many of the publications listed in the bibliography. In many cases the tolerances required may be determined from the requirements of NRL C5. In any case the qualified health physicist may choose appropriate frequencies where specific guidance is not given elsewhere. Note that QA measurement instruments or systems are available from a number of manufacturers. These generally combine kvp, dose and time measurement capabilities, and in some cases are computer interfaced and have QA reporting software. The cost of such instrumentation is small compared to the cost of the x-ray equipment and is well justified. It is essential that all test equipment be calibrated on a regular basis. NRL may be consulted if necessary for details of equipment calibration. 3.1 Automatic film processors Of all equipment in x-ray departments, film processors cause the greatest proportion of rejects 7,10,13,14. Consequently, QA of film processors will give the greatest improvement in reject rate. Therefore, film processor QA should be the first priority of any QA programme and will probably form a major share of the work of the programme. Processor QA must be performed daily to be fully effective. The principles of processor QA are simple and well described in many documents 2,3,15,16,17,18,19,20. The first step is to ensure that the processor is operating correctly at the start of the programme, usually by cleaning, replacing chemicals and checking the replenishment rates and developer temperature. A light sensitometer should then be used daily to expose test films that are processed without delay 20,21. Measurements of the densities with a densitometer are then made to give base+fog, mid density (speed) and contrast indexes which are plotted on the control charts. Systems that automate much of this procedure are commercially available. (Note that special measures are required for modern lowcrossover films, or films with different emulsion speeds on each side.) If any parameter is found to be out of control, corrective action must be taken. The appropriate action may often be determined by consulting trouble-shooting charts. (Charts are generally supplied by x-ray film or film processor manufacturers.) These show the action to be taken, for example, if contrast and speed are down but fog is up. Control charts and test films should be stored for future reference, and a log book should be kept for each processor. Two further items are important for processor QA. The first is that the test films must always come from the same box of film, to eliminate variations between film batches. Second, when the box is almost empty, a new box should be started in parallel, to check for differences. It may be necessary to adjust the control chart values if the new batch is slightly different or even to reject the batch of film, if it is too far out of control. (It has happened!) 6

13 3.2 X-ray generators Peak kilovoltage The most important parameter to monitor is the peak kilovoltage, since small drifts in the kvp can significantly alter the film density. Large x-ray facilities should have some form of kvp instrument. Either an inferential digital kvp meter or an Ardran and Crooks type penetrameter 22,23 may be used. Alternatively, a qualified health physicist may be contracted to make kvp and other measurements requiring expensive equipment. Depending on the age and stability of the generator, kvp measurements should be made at least annually, and in addition, after servicing, and if a drift in kvp is suspected for any reason. Linearity with ma/mas Good linearity will ensure that the same film blackening can be obtained for the same mas, regardless of the ma/time combination used. To assess linearity a dosemeter is required 7, although results of reduced accuracy may be obtained using film 24. Depending on the age and stability of the generator, linearity measurements should be made at least annually, after servicing, and if unpredictable results are being achieved with changes in ma. Reproducibility of mas It is clearly important that the film blackening should always be the same for a given machine setting. Good reproducibility is therefore essential for consistent radiography. Reproducibility may be assessed using a dosemeter 7, although results of reduced accuracy may be obtained using film 24. The standard deviation should not exceed 5% of the mean dose 1. Exposure timer accuracy and reproducibility These contribute to mas linearity and reproducibility. Many instruments are available with time measurement capabilities, while a spinning top may be used for one- and two-pulse machines 7,17. Waveform monitoring Increasing numbers of test devices are capable of displaying the x-ray output and/or the kvp waveform. It is strongly recommended that the waveform be checked at a range of generator settings whenever the full set of tests is done 25. Regular x-ray generator tests All of the above tests for kvp, linearity, reproducibility and exposure timer require the use of test equipment and in many cases will require the services of a qualified health physicist 26. Therefore, in order to provide a quick check on the entire radiographic system that is easily performed by radiography staff, it is recommended that a stepwedge exposure test be performed periodically. Although such a test is not likely to be able to provide a diagnosis for any equipment problems, it will show whether there have been any changes since the previous test. The stepwedge exposure frequency should be as determined by a qualified health physicist, depending on the age and stability of the equipment. 7

14 Weekly tests may be appropriate for older machines, while monthly to quarterly may be more appropriate for modern equipment. The design of a suitable stepwedge is given in figure 2. This may easily be constructed from layers of 2.5 mm aluminium. (NRL may be able to supply suitable wedges should there be sufficient demand.) This should be radiographed using an mas for which the image of the thickest part of the wedge is just discernible above the base+fog density. For single phase machines with 2.5 to 3.0 mm total filtration, 80 kvp should be used, while for three-phase and medium frequency machines 70 kvp should be used. An 18 x 24 or 24 x 30 cm cassette may be used. (If a higher kvp is typically used for a particular room, then a 1 mm Cu plate added to the wedge may be appropriate.) The cassette, each side of the wedge, must be shielded with lead or lead rubber. The kvp, mas and FFD required for this should be recorded in the QA manual and should be used for all subsequent tests, unless there is some change to the x-ray machine or film processing. In this case a new mas should be determined for subsequent tests. The first image should be kept in a safe place and used as a reference image to compare with the subsequent images. While the images may be assessed by eye the use of a densitometer is preferable. Figure 2. Design for a stepwedge suitable for QA of x-ray machines 220 mm 50 mm Top view 3.0 mm Aluminium Optional 1 mm Cu plate Side view End view In the event that the image differs markedly from the reference image, then the reason should be sought. If there has been no change in processing conditions, as determined by the daily processor QA, then there could be a fault in the x-ray machine. 3.3 X-ray tubes Filtration The filtration must comply with the NRL 1 requirement for greater than 2.5 mm Al equivalent in the primary beam. To measure the total filtration, a set of high purity aluminium filters and a dosemeter are required. The half value layer should be measured at a known kvp and the total filtration inferred from an appropriate 8

15 chart 27. It may be possible to inspect the tube assembly to establish that filtration complies, by adding up the equivalent filtration of each component, plus any added filtration. Filtration needs to be measured at acceptance testing and then only after servicing or modification to the tube assembly. Labelling NRL C5 requires that the focal spot position be marked, to enable radiation protection survey measurements to be made accurately and consistently. A label is also required, giving the specifications of the tube. This should be checked at acceptance testing and after major servicing. 3.4 Automatic exposure controls (AEC) Significant reductions in the retake rate may be achieved through consistent use of AEC for all exposures 28. However, this can only be achieved if the AEC device itself is correctly adjusted and is included in the QA programme. A simple device for streamlining the AEC QA was described recently 29. QA involves taking films for various phantom thicknesses, kvp and ma settings, and for each combination of AEC detectors. A measurement protocol appropriate to the type of AEC system should be determined by a qualified health physicist. 3.5 Light beam diaphragms (LBD) There are four aspects to LBD performance to be considered 7,17. These are accuracy, delineation (centring), brightness and edge sharpness. Requirements for these are given in NRL C5. LBDs need to be checked regularly as they are prone to being knocked out of alignment. They should be checked at least twice yearly, and more often if found to be necessary. The LBD may be tested using a commercial test tool or the "eight pennies method" 17. It is important also to check the alignment of the x-ray beam with the grid and the image receptor for wall and table buckys. 3.6 X-ray cassettes X-ray cassettes must be kept in good condition. They should be cleaned and checked regularly. All cassettes of the same type should be of the same speed. This should be checked periodically 7,17,24. Film/screen contact should be checked using a wire mesh or similar test object. Light leaks should be checked for, by processing a film that has been kept in a cassette exposed to light for an hour or more without x-ray exposure. Intensifying screens must be replaced when they are worn out, or if they become damaged. 3.7 X-ray image intensifier systems General considerations Image intensifier systems are generally complex, with a number of separate components that all have to be in good condition and correctly adjusted for image quality and patient dose to be acceptable 30. The image intensifier insert is one of the few components in diagnostic radiology that has a clearly limited life, and 9

16 which deteriorates in a relatively predictable manner during its life. Because of this deterioration, the dose rate to the patient required for satisfactory image quality will increase during the life of the II, to two or even three times the value when new. A properly budgeted replacement programme for II inserts is essential for all x- ray departments with fluoroscopy equipment. A large department with ten image intensifiers should budget to replace (for example) one or two inserts each year on average. Regular maintenance and adjustment of all components are essential for all image intensifier systems, while a full assessment by a qualified health physicist should be made at least annually. A full radiation protection survey is required by NRL C5 at two year intervals. Maximum entrance surface dose rate The maximum dose rate permitted by NRL C5 at the position of the patient's skin is 50 mgy per minute. To test this, a dosemeter is required. For systems with automatic brightness control, sufficient lead (>2 mm) should be placed on the II face to drive the technique factors to maximum. For systems with manual control, the maximum technique factors should be set. This test should be performed at least annually and after servicing. Note that for modern systems it may be more appropriate to use about 5.0 mm of copper to represent the largest patient likely to be examined, to check the maximum dose rate likely to be reached in practice. Average patient dose rate Because the performance of the II insert deteriorates with time, the dose to patients should be regularly monitored 30,31. At acceptance testing and radiation protection surveys the dose rate for a patient equivalent phantom (see below) should be determined, and the technique factors (kvp, ma, focus-ii distance and TV monitor brightness and contrast settings) should be recorded in the QA manual. For systems with automatic brightness control, the phantom should be screened regularly and the technique factors checked. If these have changed significantly, the reason should be sought, and the system serviced if necessary. For manual systems, the technique factors recorded in the QA manual should be set, and the image of the phantom should be checked. If the brightness seems to have changed then again the reason should be sought, and the system serviced if necessary. Patient equivalent phantom Each facility must have a patient equivalent phantom to enable QA tests to be made. The actual phantom used is not critical 31. However, the phantom must be reproducible. The phantom should measure at least 300 x 300 mm. The thickness will depend on the material used. Suitable phantoms could consist of 250 mm of water in a container, 200 mm of perspex, 250 mm of oil tempered hardboard, 2.5 mm of copper or 45 mm of aluminium. Note that thick phantoms that are approximately tissue equivalent are preferable in most circumstances, as the exit beam quality and scatter levels are more typical of those that occur in actual examinations. 10

17 Automatic brightness control (ABC) The ABC system is responsible for maintaining constant brightness at the TV monitor, regardless of patient attenuation. ABC systems operate in many different ways, including control of some or all of the kvp, ma and video gain. It is crucial that the ABC system be well maintained and adjusted at all times, as ABC system faults may cause excessive dose rates and/or poor image quality. In the worst case, an ABC fault may cause the generator to run at maximum kvp and ma. The parameter that the ABC system ultimately controls is the II input dose rate. Therefore ABC system performance can be monitored by regular measurement of the II input dose rate. This should be done with a range of phantom thicknesses, at least twice per year 7,17, Image intensifiers II input dose rate The II input dose rate is a key parameter to monitor, as it gives a guide to the condition of the image intensifier and it can indicate the presence of other problems with the system 30,32. NRL C5 gives limits for the II input dose rate. In general most intensifier systems should give satisfactory image quality at dose rates much lower than the limits. To measure the II input dose rate a relatively sensitive electrometer is required, fitted with a pancake ionization chamber of at least 50 cc volume. The grid (if fitted) should be removed for this test, or alternatively the chamber should be positioned between the grid and the II face. If the grid cannot be removed then an allowance must be made for the grid attenuation, typically 50% for primary beam (no scatter). If the II input dose rate has changed significantly since the previous tests then adjustments to the system may be required. Any II that is found to require input dose rates greater than the limits in NRL C5 should be replaced. Image contrast and resolution At acceptance testing, a set of measurements should be made using approved II test objects 32,33. These tests should be repeated at least annually. Where approved test objects are not available for routine testing, an image quality QA phantom should be devised 31. This will set threshold values for contrast and resolution. This should be used regularly, to check that no deterioration in the image quality has occurred. Conversion factor The best guide to the condition of an image intensifier is obtained by a measurement of its conversion factor. This requires specialised equipment and generally can only be done by a qualified health physicist or by the manufacturer. For some systems it is not practicable to do it at all. Where the conversion factor can be measured, replacement of the II insert is generally recommended when the conversion factor has reduced to a third of the value when the II was new. The conversion factor should be measured annually where possible. 11

18 Where the conversion factor cannot be measured, indirect methods need to be applied to assess the condition of the II insert, as described above 30. Focus The focus controls for the II should be regularly checked and adjusted if necessary, using a line pair test plate or fine grid. The best compromise between central and peripheral resolution should be achieved 7, Fluoroscopic television chains Generally, television chains have a number of adjustable parameters that all have to be within tolerance for optimum image quality. For example, these parameters may include peak white and black level voltages, blanking circle diameter, II vignetting correction, etc. The adjustments provided and the optimum voltage values depend on the individual manufacturer. Therefore, the manufacturer's test protocols and recommended test frequencies should be adhered to. Assessment of most of the TV system parameters may be made using a test object placed on the II face 30, and a storage oscilloscope with single TV line facility. The measured parameters should be compared to the manufacturer's specifications, although general recommendations that apply to the majority of systems may also be applied 30. Among the most important parameters is the noise produced by the TV system. The peak-to-peak TV camera noise should be less than the quantum noise at normal II input dose rates and should be less than 10% of the peak white voltage. 3.8 Fast film changers Angiographic film changers require several tests to be made periodically to ensure that the film transport is synchronised to the x-ray exposures, that the film/screen contact is satisfactory and that the screens are clean and in good condition 34. To properly check the film transport and synchronisation, a dual trace oscilloscope should be used 34. Film/screen contact can be tested using a mesh as for conventional x-ray cassettes. The grid should be removed from the film changer for this test. The intensifying screens should be carefully cleaned and checked for damage. 3.9 Cine systems and small format cameras The crucial parameter for both cine systems and small format cameras 35,36 is the dose per frame at the II input face. Generally a dose of 0.1 to 0.2 µgy per frame is recommended for cine depending on the II field size. For small format cameras, the dose for correct operation should be determined at acceptance testing. The dose per frame should be checked using a dosemeter at regular intervals, at least twice per year, and more often for intensively used systems. 12

19 For cine systems, jitter of the camera and or projector can be a problem. The Society of Motion Picture and Television Engineers (SMPTE) test films can be used to check for projector jitter. If the projector is found to be OK, and jitter is observed in the clinical films, then the cine camera is at fault 3. The cine camera, spot film camera and cine projector optics should be carefully cleaned regularly, and inspected for deterioration. Any optical elements in poor condition must be replaced Digital subtraction imaging (DSI) systems Before any quality assurance measurements can be made on DSI systems, the x- ray and image intensifier systems must be tested as described above. Most measurements on DSI systems require specialised phantoms and test equipment, and should be made by a qualified health physicist 37,38,39. The tests should be carried out every three months. A typical series of tests would follow the protocol provided with the DSI test objects supplied by Leeds University 37. The discussion below is based on the use of the Leeds test objects. II input dose per frame Fundamental to the tests is a measurement of the dose per frame at the entrance surface position of the patient and at the input face of the image intensifier. A sensitive electrometer and a large (>50 cc) pancake chamber are required for the latter measurement. The dose per frame at the patient entrance surface is recorded for comparison purposes. The II input dose per frame is needed to determine that the system is correctly adjusted and that the image intensifier is still in satisfactory condition for DSI requirements. Operating headroom It is important for digital subtraction systems, that the dose rate (dose per frame) at the image intensifier input face be set sufficiently low to prevent saturation of the video signal in the brightest parts of the image. If saturation should occur then clinical information may be lost in the subtracted images. Saturation is avoided by setting the dose rate at a level that gives a video signal that is less than the maximum. The percentage that the operating level is set below the maximum is termed the headroom. However, if the headroom is too great, then the actual useable part of the video signal (the dynamic range) will be too small. The automatic dose rate/dose per frame is therefore a compromise between dynamic range and headroom. The Leeds "jig test object" is designed to measure the headroom for peak sensing systems. It gives a subtracted image of a stepwedge. The headroom is assessed by counting the number of steps of this stepwedge that are visible. As each step corresponds to 4% above the operating video level, the headroom is the number of steps visible, times 4%. There is now available an analogous Leeds test object designed to be used with average sensing systems. 13

20 The jig test object also contains a low contrast square. If this square is not visible, then the electronic noise of the system is excessive. Dynamic range A further test of dynamic range is made using the Leeds "quadrant test object". This has four sets of low contrast details. Each set is in a sector of differing x-ray intensity, obtained by different thicknesses of copper filter. The relative x-ray intensities of the four sectors (quadrants) are 100%, 33%, 10% and 3.3%. Clearly, as the x-ray intensity is reduced the details will become more difficult to see. Detectability of the details in each quadrant is therefore a test of the dynamic range. Contrast-detail performance The Leeds "contrast-detail test object" is very similar to the NRL CD test object used for conventional IIs, as reported above. However, the contrasts for each detail size are lower, reflecting the greater capability of DSA systems. A contrastdetail assessment should be included in the quarterly tests. Misregistration The Leeds "Dalmatian" test object may be used to assess the accuracy of registration of the mask and subtraction images. This has a large number of 11 mm details spaced in a uniform matrix over the test object. The test object remains in place for both the mask and subtraction images and therefore should disappear in the subtracted image. Any misregistration results in halos in the subtracted images. Monitors and hard copy devices The viewing monitors and the hard copy film device for DSI need to be checked at least weekly, as drifts in performance can significantly reduce image quality 40, Computed tomography scanners CT scanners are generally covered by a service contract that includes calibrations and some form of regular preventative maintenance (often monthly). This is essential to ensure satisfactory calibration of the machine at all times. Generally the sorts of tests made are invasive and can only be made by the service agent. However, a maintenance contract does not relieve the facility from the responsibility for regular quality assurance measurements 42,43. At the very least the CT number of air and water and the noise level at the centre of a uniform phantom should be assessed weekly. Regular checks of the modulation of a bar pattern phantom should be done at least monthly (using the ROI standard deviation function) to check the resolution 43. The CTDI in air should be measured annually and after major servicing (such as x-ray tube replacement), in order to monitor patient dose levels. The multi format camera and the video display units must be checked regularly, at least weekly 40,41. 14

21 3.12 Mammography machines Mammography is the area of radiology requiring the greatest effort in quality assurance. Mammography can only be effective with a comprehensive quality assurance programme 44. A complete programme for mammography QA would require more extensive description than is appropriate here. Only the types of tests to be performed are outlined here, while the recommended frequencies are included in section 4. There are several excellent published QA programmes 45,46,47. It is strongly recommended that all mammography facilities obtain at least one of these publications. The greatest priority for mammography QA should be the film processor. Careful daily quality control of film processing is a pre-requisite for satisfactory mammography. All of the relevant items for general x-ray machines and automatic exposure controls should also be monitored regularly. These generally need to be monitored more often than in general radiology and to have tighter control limits. In addition, specialist features of mammography should be checked. These include the focal spot size, beam quality, compression device, regular assessments of an imaging phantom, daily exposure of a perspex phantom to check the AEC device, and a regular check on the dose level Tomography machines The x-ray generator and x-ray tube should be checked in accordance with the recommendations above before tests on the tomographic system are applied. Items to check for tomography are stability, resolution, layer selection and angle of swing 7,17. BIR 3 give a good explanation of the tests to be used. Stability is assessed visually by watching the tomographic motion during an exposure from a distance of about 2 m. This should be done without x-ray exposure if possible, or with the LBD closed. (Otherwise the operator should wear a lead apron.) The movement should be continuous, smooth and at a regular speed. Smoothness of swing may also be tested using a pinhole in a lead sheet to produce an image of the tube movement. The image should be regular and of even density. Resolution is tested using a wire mesh test object, with a selection of meshes from 1 up to 2 holes mm -1. Resolution should be at least 1.2 mm -1. Layer thickness and layer height should be assessed using an angulated scale. These tests should be done at least annually, and more often if found to be necessary Mobile radiographic equipment The same requirements apply to mobile equipment as to equivalent fixed machines. However, because of the harsher treatment often afforded to mobile 15

22 machines, the quality assurance tests may need to be done more frequently 7. In particular the LBD alignment should be checked often. Additional items to check include the mechanical features, such as tube locks, wheel brakes, safety of electrical cables, etc Mobile image intensifier equipment Mobile image intensifier systems should meet the same requirements as fixed II systems. However, because of the harsher treatment often afforded to mobile machines, the quality assurance tests may need to be done more frequently. In particular the alignment of the x-ray tube with the II should be regularly checked to ensure that the primary x-ray beam is completely intercepted by the II Grids Grids should periodically be checked for warping and damage and should be radiographed to check for uniformity Protective equipment Lead aprons and lead gloves should be thoroughly checked at least annually for any signs of wear or damage. If they appear to be suspect, then they should be tested using x-ray film or a fluoroscopy machine Darkrooms A well laid out darkroom that is clean and free from light leaks is essential for satisfactory radiography. The entrance should be light tight, either a maze or well-sealed door. Safelights should be the correct colour for the film type being used and should be no more than 25 watts. (Note that "novelty lamps" are not suitable.) A check for light leaks should be made at least annually. A film fog test should also be done at least annually 1,7, Viewboxes The viewbox(es) should be in an area shielded from direct sunlight or bright artificial light. It should be possible to dim the lighting. The viewbox should be cleaned inside and out at least annually. The fluorescent lamps should be replaced if they become too dim. The lamps should all be the same colour and the same wattage 49. Inexpensive light meters can be obtained, to simplify QA of viewboxes. NRL C5 1 gives brightness values for viewboxes. Locally derived tolerances for uniformity should also be applied. Note that the brightness of viewboxes tends to increase by 20 to 30% as they warm up and also note that the brightness of the fluorescent tubes reduces with age. There is evidence that the colour temperature of viewboxes is important also 49. This should be checked periodically as appropriate. 16

23 3.20 Technique charts It is essential that all radiographic technique factors are recorded and displayed on a technique chart 28,50. This should include the kvp, ma and exposure time to set (or AEC detector, density and speed settings) the cassette type and size to use, the focus to film distance, and any important centring points and angulations. These technique charts must be kept up to date and must be clearly legible Dose measurements As part of the QA programme, NRL C5 requires that all x-ray facilities periodically assess the dose to patients for a number of common examinations (skull, chest, thoracic spine, abdomen, lumbar spine and pelvis and the more complex procedures barium enema, barium meal and IVU). Patient dose estimates generally require the assistance of a qualified health physicist or NRL. Three methods may be used for these dose assessments. All three methods rely on the use of organ dose data obtained from computer simulation of x-ray examinations of an average patient 51,52. The simplest method is to calculate the doses from the technique factors used for average patients, using x-ray output calibration data for each x-ray tube. The second method is to use thermoluminescent dosimeters to measure the entrance surface doses for a number of average patients. Finally a dose area product meter can be fitted to the x-ray machine for a number of examinations. This is the most practical method for fluoroscopy examinations Approval of the QA programme NRL C5 requires that the QA programme be approved by a qualified health physicist. For large x-ray facilities, a programme following the guidelines herein should be satisfactory. However, NRL or a qualified health physicist must give overall approval to the programme and should be consulted concerning any details of the programme that are in doubt. All documentation for the QA programme should be kept in a safe place, so that the details may be checked during NRL radiation protection surveys. 4 AN OUTLINE QA PROGRAMME FOR LARGE FACILITIES Each x-ray facility is required by NRL C5 to institute a QA programme in radiation protection that is appropriate to its size and scope. In essence the QA programme involves the implementation of procedures to ensure that all of the items of equipment are tested at the appropriate frequency, and that corrections are made when parameters are found to be outside the permitted tolerance. Besides the actual QA measurements, essential components of the QA programme are acceptance tests and reject/repeat analysis. The programme should be set down in a QA manual. Responsibility for the QA programme should be assigned to one person and where appropriate, should be overseen by a QA committee. These concepts are discussed in section 2 above. 17

24 Notes It is intended in the programme below that for each category, the tests in the previous category should be included. For example the weekly tests would include all of the daily tests, plus those to be done weekly. The frequencies are those typically required for average equipment. The frequencies may be modified in the light of experience, to be more frequent or less frequent as necessary. Clearly, it is assumed below that corrective action will be taken immediately, should any tests reveal that QA parameters are out of control. Equipment To perform a satisfactory QA programme, each facility should have (or have access to) the following test equipment. A light sensitometer, with single/double emulsion and blue/green capability. A densitometer. A basic QA dosemeter. A kvp meter, or kvp penetrameter. An inexpensive light meter (Luxmeter). A good quality thermometer. An aluminium stepwedge. A set of aluminium (type 1100) filters. Some form of resolution test object or set of meshes. Basic II image quality phantom. Patient equivalent phantom. Specialised phantoms for DSA, CT, mammography, etc as appropriate. Some equipment may require specialised test jigs, attachment or imaging objects. (The total cost of this equipment would be in the range $10,000 to $15,000, not including specialised DSA/CT/mammography phantoms.) Daily tests For each film processor in the department, a film exposed to a sensitometer should be processed, the densities read and the results posted on the control chart. This test should be done at the same time each day, after the processor has been in use for an hour or so. The mammography machine AEC should be checked with a standard (40 mm) perspex phantom, and the mas should be recorded. Video displays and hard copy devices should be checked. 18

25 Weekly tests A stepwedge should be radiographed for each x-ray tube in the department and the density parameters should be plotted on a control chart. If any measured parameters are found to be out of control, then corrective action must be taken. (This may be done monthly if weekly tests show no variations.) A mammography QA phantom should be imaged and the details detected should be recorded. If not done daily, video displays and hard copy devices should be checked. CT scanners should have the CT numbers of air and water and the noise at the centre of the water phantom measured. The modulation of a bar pattern phantom could also be checked using the ROI software. For fluoroscopy systems, the patient equivalent phantom should be screened, using a standardised machine setup, and the technique factors noted. Monthly tests The intensifying screens should be carefully inspected each month and cleaned if necessary. A reduced cleaning frequency may be possible in the light of experience. They should be cleaned at least every six months. The mammography machine AEC should be thoroughly tested. For mobile machines fitted with an LBD, the alignment should be checked. Wall and table buckys should have the beam alignment checked. A stepwedge radiograph should be produced for all of the x-ray tubes at the facility, if not required weekly. These should be compared to the reference stepwedge for each tube. Any differences should be investigated and corrective action taken if necessary. Checks of patient dose rate and image contrast and resolution as described above should be made at facilities with image intensifiers. Quarterly tests DSA machines should be tested quarterly as described above. For mammography machines the mean glandular dose should be checked, by comparing the mas required to image a 40 mm perspex phantom. For machines fitted with an LBD, the alignment should be checked. Annual tests Protective aprons and gloves should be given a careful visual inspection. If suspect, then they should be referred for more careful checking. 19

QUALITY CONTROL AND PATIENT DOSES FROM X-RAY EXAMINATIONS IN SOME HOSPITALS IN THAILAND

QUALITY CONTROL AND PATIENT DOSES FROM X-RAY EXAMINATIONS IN SOME HOSPITALS IN THAILAND QUALITY CONTROL AND PATIENT DOSES FROM X-RAY EXAMINATIONS IN SOME HOSPITALS IN THAILAND P. Plainoi, W. Diswath, N. Manatrakul Ministry of Public Health, Nonthaburi, Thailand XA0101612 Abstract Quality

More information

Equipment Quality Control for Digital Radiography February 22, Imaging Physics CancerCare Manitoba

Equipment Quality Control for Digital Radiography February 22, Imaging Physics CancerCare Manitoba Equipment Quality Control for Digital Radiography February 22, 2018 Imaging Physics CancerCare Manitoba Purpose An equipment quality control (QC) program establishes baseline performance levels, tracks

More information

The American College of Radiology Mammography Accreditation Program Screen-Film: Frequently Asked Questions (Updated: March 16, 2018)

The American College of Radiology Mammography Accreditation Program Screen-Film: Frequently Asked Questions (Updated: March 16, 2018) The American College of Radiology Mammography Accreditation Program Screen-Film: Frequently Asked Questions (Updated: March 16, 2018) Table of Contents ACR Mammography Accreditation... 1 The 1999 ACR Mammography

More information

Ride the Lightning. A Review of Radiographic Physics

Ride the Lightning. A Review of Radiographic Physics Ride the Lightning A Review of Radiographic Physics Overview Basic principles of x-ray production Imaging equipment Quality Control Technical Factors Image QA Preparing for the registry The ARRT just wants

More information

Australian/New Zealand Standard

Australian/New Zealand Standard AS/NZS 3821:1998 IEC 60406:1997 Australian/New Zealand Standard Cassettes for medical X-ray diagnosis Radiographic cassettes and mammographic cassettes AS/NZS 3821:1998 This Joint Australian/New Zealand

More information

Scope: All CT staff technologist

Scope: All CT staff technologist APPROVED BY: Radiology Technical Director Page 1 of 6 Purpose: The QC program assesses relative changes in system performance as determined by the technologist, service engineer, qualified medical physicist,

More information

2012 Computed Tomography

2012 Computed Tomography 2012 Computed Tomography QUALITY CONTROL MANUAL Radiologist s Section Radiologic Technologist s Section Medical Physicist s Section 2012 Computed Tomography QUALITY CONTROL MANUAL Radiologist s Section

More information

Ergonomic design packed with powerful performance MULTIMOBIL 5C. Answers for life.

Ergonomic design packed with powerful performance MULTIMOBIL 5C. Answers for life. Ergonomic design packed with powerful performance MULTIMOBIL 5C Answers for life. MULTIMOBIL 5C Mobile C-arm image intensifier system with high frequency technology Siemens has led the way in development

More information

FFDM Quality Control in Canada - a Vendor Neutral Approach

FFDM Quality Control in Canada - a Vendor Neutral Approach FFDM Quality Control in Canada - a Vendor Neutral Approach Rasika Rajapakshe, PhD, FCCPM BC Cancer Agency-Center for the Southern Interior Kelowna, BC, Canada 2011 Joint AAPM/COMP Meeting August 4, 2011

More information

More Info at Open Access Database Process Control for Computed Tomography using Digital Detector Arrays

More Info at Open Access Database  Process Control for Computed Tomography using Digital Detector Arrays Digital Industrial Radiology and Computed Tomography (DIR 2015) 22-25 June 2015, Belgium, Ghent - www.ndt.net/app.dir2015 More Info at Open Access Database www.ndt.net/?id=18082 Process Control for Computed

More information

Guidelines for Assuring Softcopy Image Quality

Guidelines for Assuring Softcopy Image Quality Guidelines for Assuring Softcopy Image Quality What s inside? Quality Control Guidelines Softcopy QA testing and frequencies Danny Deroo Product and R&D Manager QA Products ABSTRACT To ensure diagnostic

More information

2017 Computed Tomography

2017 Computed Tomography 2017 Computed Tomography QUALITY CONTROL MANUAL Radiologist s Section Radiologic Technologist s Section Qualified Medical Physicist s Section 2017 Computed Tomography QUALITY CONTROL MANUAL Radiologist

More information

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009 2033-17 Joint ICTP/ Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for CT 3: Practical Experiences Claire-Louise Chapple Freeman Hospital

More information

JEFFERSON COLLEGE. Image Intensification & Equipment

JEFFERSON COLLEGE. Image Intensification & Equipment JEFFERSON COLLEGE COURSE SYLLABUS RAD175 Image Intensification & Equipment 3 Credit Hours Revised by: Janet E. Akers BS RT (R)(M) Date: October 3, 2013 Kenny Wilson, Director, Health Occupation Programs

More information

Intra-oral X-ray unit with high frequency DC generator. Product Data

Intra-oral X-ray unit with high frequency DC generator. Product Data Intra-oral X-ray unit with high frequency DC generator Tubehead Features Product Data High voltage circuit type High frequency (DC) High voltage frequency 30 khz Rated output voltage 60-65 - 70 kvp ± 8%

More information

NEMA XR 25 COMPUTED TOMOGRAPHY DOSE CHECK

NEMA XR 25 COMPUTED TOMOGRAPHY DOSE CHECK NEMA XR 25 COMPUTED TOMOGRAPHY DOSE CHECK NEMA Standards Publication XR 25-2010 Computed Tomography Dose Check Published by: National Electrical Manufacturers Association 1300 North 17th Street, Suite

More information

-Technical Specifications-

-Technical Specifications- Annex I to Contract 108733 NL-Petten: the delivery, installation, warranty and maintenance of one (1) X-ray computed tomography system at the JRC-IET -Technical Specifications- INTRODUCTION In the 7th

More information

Fieldbus Testing with Online Physical Layer Diagnostics

Fieldbus Testing with Online Physical Layer Diagnostics Technical White Paper Fieldbus Testing with Online Physical Layer Diagnostics The significant benefits realized by the latest fully automated fieldbus construction & pre-commissioning hardware, software

More information

BUREAU OF ENERGY EFFICIENCY

BUREAU OF ENERGY EFFICIENCY Date: 26 th May, 2016 Schedule No.: 11 Color Televisions 1. Scope This schedule specifies the energy labeling requirements for color televisions with native resolution upto 1920 X 1080 pixels, of CRT,

More information

Monitor QA Management i model

Monitor QA Management i model Monitor QA Management i model 1/10 Monitor QA Management i model Table of Contents 1. Preface ------------------------------------------------------------------------------------------------------- 3 2.

More information

STX Stairs lighting controller.

STX Stairs lighting controller. Stairs lighting controller STX-1795 The STX-1795 controller serves for a dynamic control of the lighting of stairs. The lighting is switched on for consecutive steps, upwards or downwards, depending on

More information

Laser Beam Analyser Laser Diagnos c System. If you can measure it, you can control it!

Laser Beam Analyser Laser Diagnos c System. If you can measure it, you can control it! Laser Beam Analyser Laser Diagnos c System If you can measure it, you can control it! Introduc on to Laser Beam Analysis In industrial -, medical - and laboratory applications using CO 2 and YAG lasers,

More information

High Frequency X-Ray Technology. Wide Screen Control with Touch-Technology

High Frequency X-Ray Technology. Wide Screen Control with Touch-Technology High Frequency X-Ray Technology Wide Screen Control with Touch-Technology The Q-VISION HF Series TM The Q-VISION Series TM integrates innovative high frequency design along with superb functionality, resulting

More information

BitWise (V2.1 and later) includes features for determining AP240 settings and measuring the Single Ion Area.

BitWise (V2.1 and later) includes features for determining AP240 settings and measuring the Single Ion Area. BitWise. Instructions for New Features in ToF-AMS DAQ V2.1 Prepared by Joel Kimmel University of Colorado at Boulder & Aerodyne Research Inc. Last Revised 15-Jun-07 BitWise (V2.1 and later) includes features

More information

Understanding PQR, DMOS, and PSNR Measurements

Understanding PQR, DMOS, and PSNR Measurements Understanding PQR, DMOS, and PSNR Measurements Introduction Compression systems and other video processing devices impact picture quality in various ways. Consumers quality expectations continue to rise

More information

World First Slim Cassette Type Digital Mammo. Upgrade Solution

World First Slim Cassette Type Digital Mammo. Upgrade Solution World First Slim Cassette Type Digital Mammo. Upgrade Solution Still Analog? Slim Cassette Type Digital Mammo. Upgrade Solution Contents 3 P. 5 P. 7 P. 9 P. 11 P. 13 P. 1824C & 2430C, easy digitalization

More information

Vascular. Development of Trinias FPD-Equipped Angiography System. 1. Introduction. MEDICAL NOW No.73 (2013.2) Yoshiaki Miura

Vascular. Development of Trinias FPD-Equipped Angiography System. 1. Introduction. MEDICAL NOW No.73 (2013.2) Yoshiaki Miura Vascular Development of Trinias FPD-Equipped Angiography System Medical Systems Division, Shimadzu Corporation Yoshiaki Miura 1. Introduction Shimadzu has developed Trinias (one ceiling-mounted type C12

More information

CARESTREAM VITA/VITA LE/VITA SE CR System Long Length Imaging User Guide

CARESTREAM VITA/VITA LE/VITA SE CR System Long Length Imaging User Guide CARESTREAM VITA/VITA LE/VITA SE CR System Long Length Imaging User Guide Use of the Guide Carestream CR Systems are designed to meet international safety and performance standards. Personnel operating

More information

Display Quality Assurance: Considerations When Establishing a Display QA Program. Mike Silosky, M.S. 8/3/2017

Display Quality Assurance: Considerations When Establishing a Display QA Program. Mike Silosky, M.S. 8/3/2017 Display Quality Assurance: Considerations When Establishing a Display QA Program Mike Silosky, M.S. 8/3/2017 Objectives and Outline Why, Who, What, When, Where? Discuss the resources that may be needed

More information

Overview. ACR Accreditation Update in Mammography. ACR Topics. Requirements Today. What s Coming For Tomorrow

Overview. ACR Accreditation Update in Mammography. ACR Topics. Requirements Today. What s Coming For Tomorrow ACR Accreditation Update in Mammography Eric Berns, PhD University of Colorado Hospital Denver Health Medical Center Denver, CO *No financial disclosures to report Overview ACR Topics Requirements Today

More information

Intelligent Pendulum Hardness Tester BEVS 1306 User Manual

Intelligent Pendulum Hardness Tester BEVS 1306 User Manual Intelligent Pendulum Hardness Tester BEVS 1306 User Manual Please read the user manual before operation. PAGE 1 Content 1. Company Profile... 3 2. Product Introduction... 3 3. Operation Instruction...

More information

UPU 5000 Inline Delamination Detection and Process Trending with the Ultrasonic Measuring System

UPU 5000 Inline Delamination Detection and Process Trending with the Ultrasonic Measuring System UPU 5000 Inline Delamination Detection and Process Trending with the Ultrasonic Measuring System Delamination Detection and Process Trending with the GreCon Ultrasonic Measuring System Exploit all reserves

More information

Guidance for Quality Assurance of PACS Diagnostic Display Devices

Guidance for Quality Assurance of PACS Diagnostic Display Devices Guidance for Quality Assurance of PACS Diagnostic Display Programme NPFIT DOCUMENT RECORD ID KEY Sub-Prog / PACS Project NPFIT-PAC-DES-0042.06 Prog. Director Max Jones Owner Jerry Norman Version 1.0 Author

More information

X-ray Tube Housing Assembly H1076X H1076Y Print No.HA Release Date:

X-ray Tube Housing Assembly H1076X H1076Y Print No.HA Release Date: H1076X 0.6 1.5 H1076Y 0.6 1.5 Print No.HA1006 1 Release Date: 2013 07 15 Description X-ray Tube Housing Assembly The tube unit is so constructed that an X-ray tube is sealed in a diagnostic type protective

More information

NEW ACHIEVEMENTS IN THE FIELD OF DEVELOPMENT OF MOBILE X-RAY TV- SYSTEM

NEW ACHIEVEMENTS IN THE FIELD OF DEVELOPMENT OF MOBILE X-RAY TV- SYSTEM NEW ACHIEVEMENTS IN THE FIELD OF DEVELOPMENT OF MOBILE X-RAY TV- SYSTEM A.V. Kovalev 1, A.V. Fedotov 1, Yu.A. Khnykov 1, A.A. Kovalev 1 1 JSC Spectrum RII, Moscow, Russia Abstract: Among various directions

More information

Rev. 01 February 2014 X- RAY SAFETY MANUAL

Rev. 01 February 2014 X- RAY SAFETY MANUAL X- RAY SAFETY MANUAL 1 I. INTRODUCTION The possession and use of radiation producing equipment at Clemson University is authorized by the South Carolina Department of Health and Environmental Control (S.C.

More information

Request for Proposals

Request for Proposals Request for Proposals Reference: ERDFIAI2012-4006A Caring First Ltd, is a limited liability company providing healthcare services. The company is currently commissioning a private new-build hospital complete

More information

Operating instructions Retro-reflective sensor. OJ50xx laser / / 2010

Operating instructions Retro-reflective sensor. OJ50xx laser / / 2010 Operating instructions Retro-reflective sensor OJ50xx laser 70481 / 00 05 / 010 Contents 1 Preliminary note3 1.1 Symbols used 3 Safety instructions 3 4 Installation 4 4.1 Installation of the supplied mounting

More information

Standard Definition. Commercial File Delivery. Technical Specifications

Standard Definition. Commercial File Delivery. Technical Specifications Standard Definition Commercial File Delivery Technical Specifications (NTSC) May 2015 This document provides technical specifications for those producing standard definition interstitial content (commercial

More information

Mechanical aspects, FEA validation and geometry optimization

Mechanical aspects, FEA validation and geometry optimization RF Fingers for the new ESRF-EBS EBS storage ring The ESRF-EBS storage ring features new vacuum chamber profiles with reduced aperture. RF fingers are a key component to ensure good vacuum conditions and

More information

Operating instructions Through-beam sensor. OJ51xx laser / / 2010

Operating instructions Through-beam sensor. OJ51xx laser / / 2010 Operating instructions Through-beam sensor OJ5xx laser 70480 / 00 05 / 200 Contents Preliminary note. Symbols used 2 Safety instructions Functions and features 4 4 Installation 4 5 Electrical connection5

More information

Assessment of the Impact of Dark Signal on Image Quality in Computerized Mammography. Introduction

Assessment of the Impact of Dark Signal on Image Quality in Computerized Mammography. Introduction Assessment of the Impact of Dark Signal on Image Quality in Computerized Mammography Alexia Valasi 1, Maria Argyrou 1, Ioannis Floros 1, Elisavet Molyvda 2, Anastasios Siountas 2, Stella Synefia 1, Ioannis

More information

Imaging solutions for any size patient and any size budget.

Imaging solutions for any size patient and any size budget. Imaging solutions for any size patient and any size budget. VetTek V e t e r i n a r y X - R a y S y s t e m s VetTek We See Things Differently & It Shows In Your Image Imagine a veterinary x-ray system

More information

IAEA RER/9/135 COURSE ON OPTIMIZATION IN COMPUTED TOMOGRAPHY Sofia, Bulgaria, Tube current modulation and dose reduction : How TCM works

IAEA RER/9/135 COURSE ON OPTIMIZATION IN COMPUTED TOMOGRAPHY Sofia, Bulgaria, Tube current modulation and dose reduction : How TCM works IAEA RER/9/135 COURSE ON OPTIMIZATION IN COMPUTED TOMOGRAPHY Sofia, Bulgaria, 2017 Tube current modulation and dose reduction : How TCM works Dean Pekarovič UMC Ljubljana, Institute of Radiology Quality

More information

3/2/2016. Medical Display Performance and Evaluation. Objectives. Outline

3/2/2016. Medical Display Performance and Evaluation. Objectives. Outline Medical Display Performance and Evaluation Mike Silosky, MS University of Colorado, School of Medicine Dept. of Radiology 1 Objectives Review display function, QA metrics, procedures, and guidance provided

More information

CHAPTER 4 OSCILLOSCOPES

CHAPTER 4 OSCILLOSCOPES CHAPTER 4 OSCILLOSCOPES 4.1 Introduction The cathode ray oscilloscope generally referred to as the oscilloscope, is probably the most versatile electrical measuring instrument available. Some of electrical

More information

Comed Medical Systems Co., Ltd. Office 707, Woolim Lion s Valley I, 311-3, Sangdaewon-dong, Seongnam-si, Gyeonggi-do, Korea Tel:

Comed Medical Systems Co., Ltd. Office 707, Woolim Lion s Valley I, 311-3, Sangdaewon-dong, Seongnam-si, Gyeonggi-do, Korea Tel: Comed Medical Systems Co., Ltd. Office 707, Woolim Lion s Valley I, 311-3, Sangdaewon-dong, Seongnam-si, Gyeonggi-do, Korea Tel:+82-31-737-2211 Fax:+82-31-737-2210 Factory 236-4 Sangdaewon-dong, Seongnam-si,

More information

Dental Line. High frequency intraoral x-ray unit. radiology ahead

Dental Line. High frequency intraoral x-ray unit. radiology ahead Dental Line High frequency intraoral x-ray unit radiology ahead enhance your operating process Hi-frequency generator, for shorter exposures and reduced dose Wireless X-ray pushbutton, available for ultimate

More information

Elements of a Television System

Elements of a Television System 1 Elements of a Television System 1 Elements of a Television System The fundamental aim of a television system is to extend the sense of sight beyond its natural limits, along with the sound associated

More information

Approved by Principal Investigator Date: Approved by Super User: Date:

Approved by Principal Investigator Date: Approved by Super User: Date: Approved by Principal Investigator Date: Approved by Super User: Date: Standard Operating Procedure BNC Dektak 3030 Stylus Profilometer Version 2011 May 16 I. Purpose This Standard Operating Procedure

More information

CATHODE RAY OSCILLOSCOPE. Basic block diagrams Principle of operation Measurement of voltage, current and frequency

CATHODE RAY OSCILLOSCOPE. Basic block diagrams Principle of operation Measurement of voltage, current and frequency CATHODE RAY OSCILLOSCOPE Basic block diagrams Principle of operation Measurement of voltage, current and frequency 103 INTRODUCTION: The cathode-ray oscilloscope (CRO) is a multipurpose display instrument

More information

G635. APPLIED SCIENCE Working waves ADVANCED GCE. Wednesday 27 January 2010 Morning. Duration: 1 hour 30 minutes

G635. APPLIED SCIENCE Working waves ADVANCED GCE. Wednesday 27 January 2010 Morning. Duration: 1 hour 30 minutes ADVANCED GCE APPLIED SCIENCE Working waves G635 * OCE / 11697* Candidates answer on the Question Paper OCR Supplied Materials: None Other Materials Required: Electronic calculator Ruler (cm/mm) Wednesday

More information

MONDO VIDEOSCREENS AND LED SOLUTIONS

MONDO VIDEOSCREENS AND LED SOLUTIONS MONDO VIDEOSCREENS AND LED SOLUTIONS DESCRIPTION A MONDO video screen is not only a product but a full package that includes personalized service done at different stages, from before to after the production

More information

FREE TV AUSTRALIA OPERATIONAL PRACTICE OP- 59 Measurement and Management of Loudness in Soundtracks for Television Broadcasting

FREE TV AUSTRALIA OPERATIONAL PRACTICE OP- 59 Measurement and Management of Loudness in Soundtracks for Television Broadcasting Page 1 of 10 1. SCOPE This Operational Practice is recommended by Free TV Australia and refers to the measurement of audio loudness as distinct from audio level. It sets out guidelines for measuring and

More information

PRJTPFL inch PRJTPFL inch PRJTPFL inch. Fixed Wall Mount Projector Screen. Universal Home/Office Projector Viewing Display

PRJTPFL inch PRJTPFL inch PRJTPFL inch. Fixed Wall Mount Projector Screen. Universal Home/Office Projector Viewing Display PRJTPFL102-100 - inch PRJTPFL112-110 - inch PRJTPFL122-120 - inch Fixed Wall Mount Projector Screen Universal Home/Office Projector Viewing Display Be sure to read this manual before use so you will know

More information

DISCLAIMER. This document is current at the date of downloading. Hunter Water may update this document at any time.

DISCLAIMER. This document is current at the date of downloading. Hunter Water may update this document at any time. DISCLAIMER This Standard Technical Specification was developed by Hunter Water to be used for construction or maintenance of water and/or sewerage works that are to become the property of Hunter Water.

More information

RSNA 2006 November 26 to December 1 Chicago. Guest author for ImPACT Dr. Koos Geleijns, Medical Physicist, Leiden University Medical Center.

RSNA 2006 November 26 to December 1 Chicago. Guest author for ImPACT Dr. Koos Geleijns, Medical Physicist, Leiden University Medical Center. RSNA 2006 November 26 to December 1 Chicago Guest author for ImPACT Dr. Koos Geleijns, Medical Physicist, Leiden University Medical Center. Once again, more than 60,000 participants (including professional

More information

PicoScope 3000 Series Automotive User guide

PicoScope 3000 Series Automotive User guide PicoScope 3000 Series Automotive User guide PS3000A044 v1.0 I PicoScope 3000 Series Automotive PC Oscilloscopes Table of Contents 1 Introduction...2...2 1 Overview...2 2 Minimum PC requirements...2 3 Installation

More information

Standard. Substitute Test or Procedure. Required Test or. 1 Scan Increment Accuracy. Initially and Annually Initially and Annually

Standard. Substitute Test or Procedure. Required Test or. 1 Scan Increment Accuracy. Initially and Annually Initially and Annually Manufacturer s Recommendations for Alternate Dental CBCT QA Program VaTech: Model PAX i3d, i3d Green, i3d Smart Table 6 Medical Physicist s Computed Tomography QC Survey Required Test or Item Procedure

More information

Linatron - M9 & M9A. Modular high-energy X-ray source. 2.0 Performance

Linatron - M9 & M9A. Modular high-energy X-ray source. 2.0 Performance The Linatron -M is a modular system. The control console, modulator, and RF unit are common to all model configurations. Only the X-ray head changes to match the application. The Linatron - M is designed

More information

Contents. 1. General Information. 2. Contents. 3. Operating Instruction. 4. Program update. 5. Trouble Shooting. 6. Specifications

Contents. 1. General Information. 2. Contents. 3. Operating Instruction. 4. Program update. 5. Trouble Shooting. 6. Specifications Contents 1. General Information 1-1. Introduction 1-2. Proper use and operation 1-3. Safety Notes 2. Contents 2-1. Accessory Include 2-2. Name of parts 3. Operating Instruction 3-1. Connection Diagram

More information

ADS Basic Automation solutions for the lighting industry

ADS Basic Automation solutions for the lighting industry ADS Basic Automation solutions for the lighting industry Rethinking productivity means continuously making full use of all opportunities. The increasing intensity of the competition, saturated markets,

More information

Colour Explosion Proof Video Camera USER MANUAL VID-C

Colour Explosion Proof Video Camera USER MANUAL VID-C Colour Explosion Proof Video Camera USER MANUAL VID-C Part Number: MAN-0036-00 Rev 4 Copyright 2002 Net Safety Monitoring Inc. Printed in Canada This manual is provided for informational purposes only.

More information

Peak Atlas IT. RJ45 Network Cable Analyser Model UTP05. Designed and manufactured with pride in the UK. User Guide

Peak Atlas IT. RJ45 Network Cable Analyser Model UTP05. Designed and manufactured with pride in the UK. User Guide GB05-7 Peak Atlas IT RJ45 Network Cable Analyser Model UTP05 Designed and manufactured with pride in the UK User Guide Peak Electronic Design Limited 2001/2013 In the interests of development, information

More information

WELDING CONTROL UNIT: TE 450 USER MANUAL

WELDING CONTROL UNIT: TE 450 USER MANUAL j WELDING CONTROL UNIT: TE 450 USER MANUAL RELEASE SOFTWARE No. 1.50 DOCUMENT NUMBER: MAN 4097 EDITION: MARCH 1998 This page is left blank intentionally. 2 / 34 TABLE OF CONTENTS SUBJECTS PAGE WELDING

More information

LED Spot 300W. Please read this user manual before your operation

LED Spot 300W. Please read this user manual before your operation LED Spot 300W Please read this user manual before your operation 1. Introduction 2. General Guideline 3. Safety Instructions 4. Cleaning and Maintenance 5. Technical Parameters 6. DMX Channels 7. Remark

More information

A simple anthropomorphic phantom used to demonstrate the effectiveness of CT dose modulation functions.

A simple anthropomorphic phantom used to demonstrate the effectiveness of CT dose modulation functions. A simple anthropomorphic phantom used to demonstrate the effectiveness of CT dose modulation functions. Lynn Bateman & Peter Hiles North Wales Medical Physics How can we test CT dose modulation? Lynn Bateman

More information

Interface Practices Subcommittee SCTE STANDARD SCTE Measurement Procedure for Noise Power Ratio

Interface Practices Subcommittee SCTE STANDARD SCTE Measurement Procedure for Noise Power Ratio Interface Practices Subcommittee SCTE STANDARD SCTE 119 2018 Measurement Procedure for Noise Power Ratio NOTICE The Society of Cable Telecommunications Engineers (SCTE) / International Society of Broadband

More information

Hospital Wide. Healthcare Display Solutions DICOM Displays, Large Screen Displays and Projectors

Hospital Wide. Healthcare Display Solutions DICOM Displays, Large Screen Displays and Projectors Hospital Wide Healthcare Display Solutions DICOM Displays, Large Screen Displays and Projectors THE WIDEST RANGE OF DISPLAY SOLUTIONS For Complete Hospital Wide Installations NEC offers a wide range of

More information

Mammo Tomosynthesis 5MP

Mammo Tomosynthesis 5MP Mammo Tomosynthesis 5MP Display system for digital breast imaging Approved for TOMOSYNTHESIS MAMMOGRAPHY The standard of care For thousands of women every day, details make all the difference. This understanding

More information

THE NEW LASER FAMILY FOR FINE WELDING FROM FIBER LASERS TO PULSED YAG LASERS

THE NEW LASER FAMILY FOR FINE WELDING FROM FIBER LASERS TO PULSED YAG LASERS FOCUS ON FINE SOLUTIONS THE NEW LASER FAMILY FOR FINE WELDING FROM FIBER LASERS TO PULSED YAG LASERS Welding lasers from ROFIN ROFIN s laser sources for welding satisfy all criteria for the optimized laser

More information

FDA MAMMOSCAN FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM MAMMOGRAPHY

FDA MAMMOSCAN FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM MAMMOGRAPHY MAMMOGRAPHY MAMMOSCAN FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM Biopsy Attachment џ MAMMOSCAN is a digital scanning technology development taking a special place in ADANI Medical X-ray product line. џ Its

More information

COMPOSITE VIDEO LUMINANCE METER MODEL VLM-40 LUMINANCE MODEL VLM-40 NTSC TECHNICAL INSTRUCTION MANUAL

COMPOSITE VIDEO LUMINANCE METER MODEL VLM-40 LUMINANCE MODEL VLM-40 NTSC TECHNICAL INSTRUCTION MANUAL COMPOSITE VIDEO METER MODEL VLM- COMPOSITE VIDEO METER MODEL VLM- NTSC TECHNICAL INSTRUCTION MANUAL VLM- NTSC TECHNICAL INSTRUCTION MANUAL INTRODUCTION EASY-TO-USE VIDEO LEVEL METER... SIMULTANEOUS DISPLAY...

More information

Fully ly Automaticti. Motorised Satellite t TV System. User s manual REV

Fully ly Automaticti. Motorised Satellite t TV System. User s manual REV REV. 1.0 Fully ly Automaticti Motorised Satellite t TV System User s manual Customer Help Line: 1300 139 255 Support Email: support@satkingpromax.com.au Website: www.satkingpromax.com.au www.satkingpromax.com.au

More information

CytoFLEX Flow Cytometer Quick Start Guide

CytoFLEX Flow Cytometer Quick Start Guide Sheath Waste CLASS 1 LASER PRODUCT COMPLIES WITH 21 CFR 1040.10 AND 1040.11 EXCEPT FOR DEVIATIONS PURSUANT TO LASER NOTICE NO. 50 DATED JUNE 24, 2007 MANUFACTURED Sheath B49008AC February 2015 CytoFLEX

More information

Innovative Rotary Encoders Deliver Durability and Precision without Tradeoffs. By: Jeff Smoot, CUI Inc

Innovative Rotary Encoders Deliver Durability and Precision without Tradeoffs. By: Jeff Smoot, CUI Inc Innovative Rotary Encoders Deliver Durability and Precision without Tradeoffs By: Jeff Smoot, CUI Inc Rotary encoders provide critical information about the position of motor shafts and thus also their

More information

High Value-Added IT Display - Technical Development and Actual Products

High Value-Added IT Display - Technical Development and Actual Products High Value-Added IT Display - Technical Development and Actual Products ITAKURA Naoki, ITO Tadayuki, OOKOSHI Yoichiro, KANDA Satoshi, MUTO Hideaki Abstract The multi-display expands the desktop area to

More information

Operating Instructions

Operating Instructions CNTX Contrast sensor Operating Instructions CAUTIONS AND WARNINGS SET-UP DISTANCE ADJUSTMENT: As a general rule, the sensor should be fixed at a 15 to 20 angle from directly perpendicular to the target

More information

CARESTREAM DIRECTVIEW Elite CR System

CARESTREAM DIRECTVIEW Elite CR System CARESTREAM DIRECTVIEW Elite CR System Improve workflow, productivity, and patient throughput. The CARESTREAM DIRECTVIEW Elite CR System is small, easy to install and easy to use. This powerful distributed

More information

THE EFFECT OF PERFORMANCE STAGES ON SUBWOOFER POLAR AND FREQUENCY RESPONSES

THE EFFECT OF PERFORMANCE STAGES ON SUBWOOFER POLAR AND FREQUENCY RESPONSES THE EFFECT OF PERFORMANCE STAGES ON SUBWOOFER POLAR AND FREQUENCY RESPONSES AJ Hill Department of Electronics, Computing & Mathematics, University of Derby, UK J Paul Department of Electronics, Computing

More information

CARESTREAM DIRECTVIEW Elite CR System

CARESTREAM DIRECTVIEW Elite CR System CARESTREAM DIRECTVIEW Elite CR System Improve workflow, productivity, and patient throughput. The CARESTREAM DIRECTVIEW Elite CR System is small, easy to install and easy to use. This powerful distributed

More information

HDMI Demystified April 2011

HDMI Demystified April 2011 HDMI Demystified April 2011 What is HDMI? High-Definition Multimedia Interface, or HDMI, is a digital audio, video and control signal format defined by seven of the largest consumer electronics manufacturers.

More information

Handling and storage of cinematographic film

Handling and storage of cinematographic film Rec. ITU-R BR.1219 1 RECOMMENDATION ITU-R BR.1219* Rec. ITU-R BR.1219 HANDLING AND STORAGE OF CINEMATOGRAPHIC FILM RECORDING (Question ITU-R 109/11) (1995) The ITU Radiocommunication Assembly, considering

More information

INSTRUMENT CATHODE-RAY TUBE

INSTRUMENT CATHODE-RAY TUBE INSTRUMENT CATHODE-RAY TUBE 14 cm diagonal rectangular flat face domed mesh post-deflection acceleration improved spot quality for character readout high precision by internal permanent magnetic correction

More information

What do we mean by workload? Number of scans Type of scans/mix of scans? Total mas Total dip

What do we mean by workload? Number of scans Type of scans/mix of scans? Total mas Total dip CT & Workload (BIR) David Sutton What do we mean by workload? Number of scans Type of scans/mix of scans? Total mas Total dip What do we do with it? Need to relate workload to total scatter dose in the

More information

TRANSCENSION 6-CHANNEL DMX DIMMER PACK (order code: BOTE40) USER MANUAL

TRANSCENSION 6-CHANNEL DMX DIMMER PACK (order code: BOTE40) USER MANUAL www.prolight.co.uk TRANSCENSION 6-CHANNEL PACK (order code: BOTE40) USER MANUAL SAFETY WARNING FOR YOUR OWN SAFETY, PLEASE READ THIS USER MANUAL CAREFULLY BEFORE YOUR INITIAL START-UP! CAUTION! Keep this

More information

INSTRUCTION & INSTALLATION GUIDE

INSTRUCTION & INSTALLATION GUIDE Digital electric radiator INSTRUCTION & INSTALLATION GUIDE 2 Introduction harmoni digital electric radiators are programmable with an exclusive electronic temperature programmer and manufactured with quality

More information

Official Journal L 191, 23/07/2009 P

Official Journal L 191, 23/07/2009 P Commission Regulation (EC) No 642/2009 of 22 July 2009 implementing Directive 2005/32/EC of the European Parliament and of the Council with regard to ecodesign requirements for televisions Text with EEA

More information

Technologist Quality Control Procedures

Technologist Quality Control Procedures Technologist Quality Control Procedures The specific procedures for the Technologist Quality Control Program are those specified in the most current ACR MRI QC Manual. WEEKLY MRI EQUIPMENT QUALITY CONTROL

More information

This document is meant purely as a documentation tool and the institutions do not assume any liability for its contents

This document is meant purely as a documentation tool and the institutions do not assume any liability for its contents 2009R0642 EN 12.09.2013 001.001 1 This document is meant purely as a documentation tool and the institutions do not assume any liability for its contents B COMMISSION REGULATION (EC) No 642/2009 of 22

More information

Dimensions. Model Number. Electrical connection. Features. Pinout Product information. Indicators/operating means. LGS25 Serie.

Dimensions. Model Number. Electrical connection. Features. Pinout Product information. Indicators/operating means. LGS25 Serie. Q Dimensions Transmitter Detection field + 9. Detection field 7.8 n Beam. Beam Receiver III 0. 0 Fix H H H Fn Model Number Light grid with fixed cable with -pin, M x connector, and fixed cable with 8-pin,

More information

PicoScope 4000 Automotive PC Oscilloscopes

PicoScope 4000 Automotive PC Oscilloscopes PicoScope 4000 Automotive PC Oscilloscopes User's Manual ps4000a.en-1 Copyright 2008 Pico Technology Ltd. All rights reserved. Contents I Contents 1 Introduction...1 1 Overview...1...1 2 Minimum PC requirements...2

More information

INSTRUMENT CATHODE-RAY TUBE

INSTRUMENT CATHODE-RAY TUBE Instrument cathode-ray tube D14-363GY/123 INSTRUMENT CATHODE-RAY TUBE mono accelerator 14 cm diagonal rectangular flat face internal graticule low power quick heating cathode high brightness, long-life

More information

Drift Tubes as Muon Detectors for ILC

Drift Tubes as Muon Detectors for ILC Drift Tubes as Muon Detectors for ILC Dmitri Denisov Fermilab Major specifications for muon detectors D0 muon system tracking detectors Advantages and disadvantages of drift chambers as muon detectors

More information

Coronis 5MP Mammo. The standard of care for digital mammography

Coronis 5MP Mammo. The standard of care for digital mammography Coronis 5MP Mammo The standard of care for digital mammography The standard of care For thousands of women every day, details make all the difference. This understanding, along with many years of commitment

More information

BRIGHT BRIGHTER BRIGHTEST ONE ILLUMINATOR DESIGN THREE LIGHT SOURCES. featuring the EvenVue Reflector System

BRIGHT BRIGHTER BRIGHTEST ONE ILLUMINATOR DESIGN THREE LIGHT SOURCES. featuring the EvenVue Reflector System BRIGHT BRIGHTER BRIGHTEST ONE ILLUMINATOR DESIGN THREE LIGHT SOURCES featuring the EvenVue Reflector System DVC (DIAGNOSTIC VIEWING CENTER) The features and benefits of the TECHLINE can be combined with

More information

Guidelines for Specification of LED Lighting Products 2010

Guidelines for Specification of LED Lighting Products 2010 Guidelines for Specification of LED Lighting Products 2010 September 2010 Introduction With LED s emerging as a new functional light source there is a need to ensure performance claims are made in a consistent

More information

High Resolution Multicolor Contrast Scanner. Dimensioned drawing

High Resolution Multicolor Contrast Scanner. Dimensioned drawing Specifications and description KRTM 20 High Resolution Multicolor Contrast Scanner Dimensioned drawing en 01-2011/06 50116669 12mm 20mm 50mm 12-30 V DC 50 / 25 khz We reserve the right to make changes

More information

Creating room. for your visions

Creating room. for your visions BuckyDiagnost Radiography solutions Creating room for your visions 2 What makes a Rad Room simply perfect? Room for all your ideas Efficient, powerful, individual and innovative. This sums up just some

More information

Advanced Test Equipment Rentals ATEC (2832)

Advanced Test Equipment Rentals ATEC (2832) E stablished 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) Technical Datasheet Scalar Network Analyzer Model 8003-10 MHz to 40 GHz The Giga-tronics Model 8003 Precision Scalar

More information