GC/LC-MS: data acquisition rate and peak reconstruction

Size: px
Start display at page:

Download "GC/LC-MS: data acquisition rate and peak reconstruction"

Transcription

1 GC/LC-MS: data acquisition rate and peak reconstruction Nyquist (Shannon-Kotelnikov-Whittaker) theorem Signal sampling does not involve any loss of information as long as the sampling frequency is at least twice as high as the highest frequency in the signal. Six points (at least) are required for an unbiased sampling of a perfect Gaussian peak, representing an ideal chromatographic peak. In a real chromatographic trace: the signal is noisy (noise must be adequately sampled) peak tailing/fronting occur (difficult detection of start/end points of the peak)

2 Overlaid fast GC-MS extracted ion chromatograms of chlorpyrifos-methyl (m/z 125), measured at different acquisition rates/number of data points obtained, show the dramatic effects of an inadequate signal sampling: Scanned m/z range: Hz 3.18 Hz 0.84 Hz 5.98 Hz Hz Note that the peak width at the baseline is about 0.02 min!

3 As shown by the following data, obtained by a GC-MS instrument equipped with a quadrupole analyzer, 7-10 points across a peak represent a good value in terms of peak area/height reproducibility: peak height peak area

4 Data acquisition rate: quadrupole MS vs. other GC detectors detector FID data acquisition rate Up to 200 Hz (very low dead volume; no extracolumn band broadening) Micro ECD Up to Hz (some peak broadening due to a still too large cell volume) Quadrupole MS From few Hz (old, slow scanning instruments) up to Hz (new, fast scanning analyzers) depending on the mass range scanned

5 Signal acquisition in MS full scan mode During a MS scan cycle multiple readings are taken at each m/z step and then averaged/summed to give one data point. For instance 2 N (N = 0,1,2,., 7, selectable by the operator) readings are performed by the Agilent 5973N MSD; the number of readings and the scanned mass range determine the acquisition rate: N N Acquisition rate (scans/s) ( mass range) On a quadrupole instrument (with unit resolution) MS peaks are usually 1 u wide and 7-8 points are typically needed per peak in order to characterize it well enough for a detector to consistently determine peak centroid value and position.

6 The following parameters are commonly considered: data acquisition rate = number of spectra acquired per second scan time = time spent on scanning the specified mass range inter-scan delay = time elapsing between one scan ending and the next scan starting scan speed: up to u/s for a new, fast scanning quadrupole ca u/s for an old, slow scanning quadrupole

7 In these tables, acquisition parameters are calculated under different operating conditions. The minimum detectable peak width is based on a 7 point peak definition.

8 Scan speed limitations A systematic investigation performed by P. Korytar and coworkers in 2005 (J. Chromatogr. A 1067 (2005) ) has led to the following conclusions about the effect of scan speed: the quality of the averaged mass spectra is found to be essentially constant up to 7000 u/s; increasing scan speed reduces resolution, which results in a significant loss (nominal resolution > 1 u) above 9000 u/s; at scan speeds higher than u/s the loss of resolution is so significant that the averaged mass spectra become distorted.

9 Comparison of averaged mass spectra (with no background subtraction) obtained for 3,3,4,4,5- Pentachlorobiphenyl at different scan speeds. Scanned mass range: 350 u (m/z ); inter-scan delay: 0.1 s. Spectral details are clearly lost at higher scan speed (although the effect is not so evident if peak centroids are reported).

10 Effect of inter-scan delay time If the delay time is too short proper processing and recording of the acquired spectrum cannot be achieved: The comparison of averaged mass spectra of octachloro-undecane acquired at different inter-scan delays shows loss of information at the low m/z range for shorter delays when the mass scan occurs from high to low m/z values.

11 Effect of mass resolving power As mass resolving power increases the transmission of ions through the analyzer and then towards the ion detector decreases, resulting in a loss of sensitivity, i.e. a lower signal-to-noise response: S/N response for the GC- MS analysis of 0.02 g/ml fenbuconazole performed with a quadrupole mass spectrometer operated at variable mass resolving power. Mass resolving power (log scale)

12 Signal acquisition in Selected Ion Monitoring (SIM) mode Pros Higher duty cycle (duty cycle decreases as the number of monitored ions increases) Excellent sensitivity Cons No Mass Spectrum available; the information obtained is limited only to target compounds

13 When multiple SIM is adopted, i.e. the signal from several m/z windows is acquired consecutively, the data acquisition rate is given by the following equation: where n is the number of monitored ions.

14 Increasing the number of monitored ions (i.e. the amount of information available) leads to a decrease in data acquisition rate, thus limiting the minimum detectable width for a chromatographic peak: Nonetheless, even for baseline widths of 300 ms, up to four different ions can be still monitored without sacrificing the seven-data pointsper-peak rule.

15 Ion traps: effect of scan speed on sampling rate A single ion trap microscan consists in an AGC prescan and an analytical scan: RF microscan Note that ion injection time is the most relevant part of the total scan time in this case.

16 Several microscans are summed in a single scan cycle, averaged and saved as one data point of the chromatographic trace:

17 The sampling rate across a chromatographic peak when ion trap-ms is adopted for detection can be increased by: reducing the number of microscans per data point reducing the injection time, which can be achieved in two ways: 1 - setting a lower maximum injection time, which reduces the number of ions in the trap, hence sensitivity; 2 - increase the efficiency of the ion source and ion optics, to improve the transmission of ions, thus filling the trap to the same level in a shorter period of time.

18 Microscans Scan event time (s) Range Range ,28 0,42 4 0,51 0,75 6 0,73 1,09 8 0,96 1, ,19 1,76

19 Sampling rate issues with non scanning analyzers: ToF The Time-of-Flight is a typical example of fast, non scanning analyzer. Since ToF resolution is increased with time of flight, a clear contrast can be observed between speed and resolution in a ToF device: High speed TOF High Res TOF Spectral acquisition rate (spectra/s) Upper mass limit (u) Mass Resolution (FWHM) Mass accuracy (ppm) Typical application at m/z at m/z 614 < 5 Fast GC, GC GC Note that these data are referred to GC-MS applications, thus the upper mass limit is not much higher than 1000 u

20 An example of a GC-MS system based on a high resolution ToF is the Waters-Micromass GCT Premier spectrometer, in which orthogonal acceleration and a single reflectron are adopted: The main features of this instrument are: Resolution: 7000 (FWHM); Accuracy: < 5 ppm; ion beam sampling rate: up to 25 KHz; spectral acquisition rate: up to 20 Hz.

21 The following scheme is referred to a GC-MS system equipped with a high speed ToF (Pegasus HT TOFMS from Leco). The maximum acquisition rate in this case is 500 Hz :

22 Peak reconstruction: scanning vs non scanning analyzers When very fast GC separations are considered, the acquisition rate provided by a non scanning analyzer, like ToF, enables a huge improvement in peak reconstruction, compared to a scanning analyzer: Non scanning (ToF) Scanning (quadrupole)

23 Signal-to-noise ratio: scanning versus non scanning analyzers The increase of mass range has a negative effect on the S/N ratio of a scanning analyzer, like a quadrupole, since a greater portion of ion population is lost when longer scans are required. The phenomenon is absent in the case of non scanning analyzers, like ToF. The effect is apparent when a comparison is made between the S/N response obtained for 0.02 g/ml of fenbuconazole using a time-of-flight (non scanning) or a quadrupole (scanning) LC-MS instrument:

24 Spectral skewing Spectral skewing is an alteration of peak intensity ratios in the MS spectrum, resulting from a non symmetric variation of the analyte concentration and a relatively low spectral acquisition rate. The spectral intensity alteration will depend on the type of peak asymmetry: scan

25 Once all the spectra acquired under the chromatographic peak are averaged the skewing effect will be almost canceled. However, if the peak is not symmetric skewing will affect the final spectrum, as shown in the following case (scan occurring from high to low m/z ratios):

26 Skewing-free spectra are important to perform correctly deconvolution processes on complex chromatographic bands, i.e. to resolve several overlapping peaks on the basis of increasing/decreasing intensity of ions in mass spectra acquired during band elution. A powerful deconvolutive algorithm, based on the systematic extraction of ion currents from spectra underlying a complex band, is embedded in the PeakFind software (Leco):

27 The algorithm needs a high number of acquired spectra per second to work properly, as shown in the following example, related to the GC-MS separation of 9 pesticides in 4 s:

28 Comparison between GC-MS separations with different speed SEF is the Speed Enhancement Factor, a factor used to normalize GC separations to that of a standard capillary column (30 m length, 0.25 mm I.D., 0.25 m thickness of stationary phase), operated at typical gas flow rate (Helium, 1 ml/min = 34 cm/s): The collection frequencies are calculated to give 5 points across full peak width.

29 Comparison between different mass anayzers used in GC-MS in terms of acquisition rate Analyzer Upper mass limit (u) Spectral acquisition rate Resolution Note that the Sector (i.e. a electrostatic/magnetic analyzer) has, by far, the highest resolution but also the lowest acquisition rate.

30 Practical approaches (and compromises) to achieve speed, sensitivity or selectivity in GC-MS

31 Speed (more points per GC peak) How to achieve it Decreasing the number of raw spectra (microscans) to be averaged in full scan mode decreasing the time spent per ion in SIM mode (dwell time, isolation time) Reducing the scan range in full scan mode or the number of monitored ions in SIM Decreasing the number of transients to be summed Analyzer Q, IT Sector Q, IT Sector TOF What is sacrificed Reproducibility of spectra and/or ion ratios Selectivity (ability to identify /confirm) Sensitivity, Selectivity (resolution) Decreasing resolution Sector Selectivity

32 Sensitivity (increased S/N, i.e. decreased LOD if not limited by chemical noise) How to achieve it Using SIM Decreasing spectrum storage rate, i.e. increasing the number of transients to be summed Increasing ion injection time Decreasing resolution Analyzer Q Sector TOF IT Sector What is sacrificed Analytical scope (targeted analysis only) Speed Speed Selectivity (resolution)

33 Selectivity (decreased chemical noise, i.e. potentially decreased LOD, depending on the extent of sample interference) How to achieve it Analyzer What is sacrificed Increasing the resolution Sector Speed, sensitivity Using MS n Using high-resolution IT, triple Q Q-TOF TOF Speed, sensitivity Speed

34 High Resolution TOF-MS: selectivity of detection The following GC High Res TOF-MS traces, obtained for a wastewater extract (spiked with pesticides at 50 pg/ml concentration), show the difference in extracted ion chromatograms for atrazine (windows A and B; m/z ) and chlorpyriphos (windows C and D; m/z ) when changing the extraction window width from 1 to 0.02 m/z units: atrazine chlorpyriphos A fold LOD improvement is achieved using a narrower window

35 Advantages and limitations of High Resolution TOF MS Low spectral acquisition rates, typical of high resolution ToF-MS, provide higher sensitivity but result unavoidably in rather poor peak shapes: Phosalone (50 pg injected) Peak width: ca 1 s J. Chromatogr. A 1058 (2004)

36 Adapting MS detection to GC GC Comprehensive two-dimensional GC (GC GC) provides three-dimensional chromatographic information, characterized by remarkably higher resolution and capacity, compared to one dimensional GC:

37 Comprehensive two-dimensional GC (GC GC) is based on the selection and transfer of all the fractions arising from a conventional GC separation (10-30 m column) to a fast GC separation, accomplished on a shorter column (1-2 m) and at higher speed: a) injector b) first column c) connectors d) modulator e) second column f) detector g) second oven (optional) A fraction eluting from the first column is parked inside the modulator until the previous fraction has completed its further separation in the second column.

38 Possible approaches for fraction transfer A fraction eluting from the first column is alternatively condensed into the downstream (D) and upstream (U) section of the modulator using liquid CO 2 jets. When the jet is shut-off revaporization (followed by transfer downwards) occurs. A fraction eluting from the first column is loaded into a conventional 6-port injection valve. When the valve is switched to the inject position the fraction can be transferred to the second column by an auxiliary gas flow.

39 Combination of two columns having a non polar (D1) and a polar (D2) stationary phase, respectively, introduces a second degree of freedom in analyte separation, i.e. polarity, besides volatility: D2 ( polar column) Low BP, high polarity Low BP, Medium polarity Medium BP, high polarity Medium BP, medium polarity High BP, High polarity High BP, Medium polarity Low BP, low polarity Medium BP, low polarity High BP, low polarity Low BP Medium BP High BP D1 (non polar column)

40 Detection and data processing in GC GC-MS The high-speed conditions of the second column result in very sharp peaks (~200 ms basewidth on average) that require detectors with fast response times (~20 Hz minimum). Flame ionization detectors (FID) and more selective detectors (e.g., electron capture, nitrogen chemiluminescence) that have been developed for high-speed GC ( > 50 Hz data acquisition rates) have successfully been applied also to GC GC. Quadrupole and magnetic sector high-resolution mass spectrometers also have been interfaced, but the slow nature of their operation limits their use to the selective mode, in order to speed up detector response time. TOF MS has been recently introduced by Leco as detector for GC GC with an additional dimension of resolution, i.e. deconvolution of complex bands.

41 Total ion chromatogram contour plot of a congested region in a GC GC chromatogram. Deconvolved slice. chromatographic Each peak marker indicates the presence of a sample component with a unique mass spectrum, as identified by the MS deconvolution software. Each colored trace represents a unique product ion

42 In this example, the PeakFind algorithm, developed by Leco, was applied to a complex band resulting from the co-elution of 4 hydrocarbons, belonging to a Naphta sample: Note that product ion with m/z 43 (fragmentation is not negligible in this case, as EI is performed) is shared by two analytes, the other ones are unique for the three remaining species. The software is able to extract automatically interference-free mass spectra, in which signals from ions shared between co-eluting analytes are accurately proportioned.

43 Species related to Peaks 17 and 18 were detected and accurately identified by library search against the NIST (National Institute for Standards and Technology) database, even though they were separated by only 0.07 seconds. Peaks 19 and 20 were accurately identified despite the high number of shared signals (shown in red) observed in their mass spectra.

44 Advantages of GC GC TOF MS Capability to fulfil demanding analytical tasks Enhanced separation of target analytes from matrix co-extracts Improved detectability: LOD (at S/N = 5) as low as 0.2 to 30 pg injected, detectability enhancement (compared to onedimensional GC TOF-MS) typically fold Reliable identification: full mass spectral information by time-of-flight mass spectrometry, integrated with the deconvolution capability of special software, allows for reliable identification of trace analytes in complex matrices.

GCxGC Comprehensive Two-Dimensional Gas Chromatography

GCxGC Comprehensive Two-Dimensional Gas Chromatography GCxGC Comprehensive Two-Dimensional Gas Chromatography Delivering the Right Results GCxGC Comprehensive Two-Dimensional Gas Chromatography Take your chromatography to the next level LECO comprehensive

More information

GCxGC Comprehensive Two-Dimensional Gas Chromatography

GCxGC Comprehensive Two-Dimensional Gas Chromatography GCxGC Comprehensive Two-Dimensional Gas Chromatography Delivering the Right Results LECO s GCxGC Comprehensive Two-Dimensional Gas Chromatography Take your chromatography to the next level LECO comprehensive

More information

How Quadrupole Size Affects Research

How Quadrupole Size Affects Research Quadrupole Size Comparison APPLICATION NOTE NOTE How Quadrupole Size Affects Research Mass Spectrometry is one of the most widely used analytical techniques and the most common mass analyzer is the quadrupole.

More information

BitWise (V2.1 and later) includes features for determining AP240 settings and measuring the Single Ion Area.

BitWise (V2.1 and later) includes features for determining AP240 settings and measuring the Single Ion Area. BitWise. Instructions for New Features in ToF-AMS DAQ V2.1 Prepared by Joel Kimmel University of Colorado at Boulder & Aerodyne Research Inc. Last Revised 15-Jun-07 BitWise (V2.1 and later) includes features

More information

Supporting Information for Effectiveness of Global, Low-Degree Polynomial Transformations for GCxGC Data Alignment

Supporting Information for Effectiveness of Global, Low-Degree Polynomial Transformations for GCxGC Data Alignment Supporting Information for Effectiveness of Global, Low-Degree Polynomial Transformations for GCxGC Data Alignment Davis W. Rempe, * Stephen E. Reichenbach, * Qingping Tao, Chiara Cordero, Wayne E. Rathbun,

More information

Agilent 7000 Series Triple Quadrupole GC/MS System

Agilent 7000 Series Triple Quadrupole GC/MS System Agilent 7000 Series Triple Quadrupole GC/MS System Concepts Guide The Big Picture Agilent Technologies Notices Agilent Technologies, Inc. 2011 No part of this manual may be reproduced in any form or by

More information

Liquid Chromatography- Mass Spectrometer Manual

Liquid Chromatography- Mass Spectrometer Manual Liquid Chromatography- Mass Spectrometer Manual Joshua Willis, Elizabeth Sattely Department of Chemical Engineering Stanford University November 6, 2014 Abstract This manual will explain the LC/MS, its

More information

Achieving More Efficient Data Review with OpenLAB CDS

Achieving More Efficient Data Review with OpenLAB CDS Achieving More Efficient Data Review with OpenLAB CDS Technical Note Introduction Separations have enjoyed major advances that have significantly reduced chromatographic run times. Ultra high performance

More information

Noise. CHEM 411L Instrumental Analysis Laboratory Revision 2.0

Noise. CHEM 411L Instrumental Analysis Laboratory Revision 2.0 CHEM 411L Instrumental Analysis Laboratory Revision 2.0 Noise In this laboratory exercise we will determine the Signal-to-Noise (S/N) ratio for an IR spectrum of Air using a Thermo Nicolet Avatar 360 Fourier

More information

HPLC Tips and Tricks. Tiziana Ladisa Sales Support Specialist for Chromatography Italy Thermo Fisher Scientific, Rodano (MI)

HPLC Tips and Tricks. Tiziana Ladisa Sales Support Specialist for Chromatography Italy Thermo Fisher Scientific, Rodano (MI) HPLC Tips and Tricks Tiziana Ladisa Sales Support Specialist for Chromatography Italy Thermo Fisher Scientific, Rodano (MI) The world leader in serving science Overview Thermo Scientific UltiMate 3000

More information

TMT data acquisition on the LTQ-Orbitrap XL Mass Spectrometer

TMT data acquisition on the LTQ-Orbitrap XL Mass Spectrometer TECH TIP # 70 TMT data acquisition on the LTQ-Orbitrap XL Mass Spectrometer Introduction TR0070.0 Successful acquisition of quantitative tandem mass tag (TMT*) data with the Thermo Scientific LTQ-Orbitrap

More information

Waters GCT GCMS Training Manual (Nominal Mass GCMS version) 02/06/2012 S.V.

Waters GCT GCMS Training Manual (Nominal Mass GCMS version) 02/06/2012 S.V. 1234 Hach Hall 515-294-5805 www.cif.iastate.edu Waters GCT GCMS Training Manual (Nominal Mass GCMS version) 02/06/2012 S.V. Location: Contact: 1238 Hach Hall Steve Veysey, 1234 Hach Hall; Kamel Harrata,

More information

Bruce Quimby. Agilent Technologies October 28, For Forensic Use. Page 1

Bruce Quimby. Agilent Technologies October 28, For Forensic Use. Page 1 Bruce Quimby Application Scientist Agilent Technologies October 28, 2009 Page 1 Tools For Improving GC/MS Analysis Retention Time Locking (RTL) Method Translation Synchronous SIM/Scan Column effluent split

More information

Electrospray-MS Charge Deconvolutions without Compromise an Enhanced Data Reconstruction Algorithm utilising Variable Peak Modelling

Electrospray-MS Charge Deconvolutions without Compromise an Enhanced Data Reconstruction Algorithm utilising Variable Peak Modelling Electrospray-MS Charge Deconvolutions without Compromise an Enhanced Data Reconstruction Algorithm utilising Variable Peak Modelling Overview A.Ferrige1, S.Ray1, R.Alecio1, S.Ye2 and K.Waddell2 1 PPL,

More information

SEM- EDS Instruction Manual

SEM- EDS Instruction Manual SEM- EDS Instruction Manual Double-click on the Spirit icon ( ) on the desktop to start the software program. I. X-ray Functions Access the basic X-ray acquisition, display and analysis functions through

More information

UWPR Advancing Proteomics

UWPR Advancing Proteomics How to setup a PRM assay on the QExactive Plus (SW v 2.7) Inclusion List First you need a list of peptides representing the proteins you want to monitor/quantify. You can use prior DDA data, databases

More information

Quadrupoles have become the most widely used

Quadrupoles have become the most widely used ARTICLES A Novel Tandem Quadrupole Mass Analyzer Zhaohui Du and D. J. Douglas Department of Chemistry, University of British Columbia, Vancouver, B. C., Canada A new tandem mass analyzer is described.

More information

New Filling Pattern for SLS-FEMTO

New Filling Pattern for SLS-FEMTO SLS-TME-TA-2009-0317 July 14, 2009 New Filling Pattern for SLS-FEMTO Natalia Prado de Abreu, Paul Beaud, Gerhard Ingold and Andreas Streun Paul Scherrer Institut, CH-5232 Villigen PSI, Switzerland A new

More information

Magnum GCMS Training and Operating Instructions 02/07/2012 S.V.

Magnum GCMS Training and Operating Instructions 02/07/2012 S.V. Magnum GCMS Training and Operating Instructions 02/07/2012 S.V. 1234 Hach Hall 515-294-4805 www.cif.iastate.edu Location: Contact: 1238 Hach Hall Steve Veysey, 1234 Hach Hall; Kamel Harrata, 1236 Hach

More information

Staying Connected in GC

Staying Connected in GC Staying Connected in GC Daron Decker / Eric Pavlich Chromatography Technical Specialists Page 1 Leaks (Oxygen) Damages GC Columns Repair of the GC column is impossible Replacements means re-calibration

More information

m RSC Chromatographie Integration Methods Second Edition CHROMATOGRAPHY MONOGRAPHS Norman Dyson Dyson Instruments Ltd., UK

m RSC Chromatographie Integration Methods Second Edition CHROMATOGRAPHY MONOGRAPHS Norman Dyson Dyson Instruments Ltd., UK m RSC CHROMATOGRAPHY MONOGRAPHS Chromatographie Integration Methods Second Edition Norman Dyson Dyson Instruments Ltd., UK THE ROYAL SOCIETY OF CHEMISTRY Chapter 1 Measurements and Models The Basic Measurements

More information

AIM INTRODUCTION SIMPLIFIED WORKFLOW

AIM INTRODUCTION SIMPLIFIED WORKFLOW CD-R PATENT PROTECTION: BRAND CHARACTERIZATION TO IDENTIFY COUNTERFEIT GOODS USING SIMPLIFIED MASS SPECTROMETRY James Morphet and Eleanor Riches Waters Corporation, Manchester, UK AIM To provide an easy-to-use

More information

RGA13, 12/10/17 Ultra High Resolution 20mm Quadrupole with Dual Zone operation

RGA13, 12/10/17 Ultra High Resolution 20mm Quadrupole with Dual Zone operation RGA13, 12/10/17 Ultra High Resolution 20mm Quadrupole with Dual Zone operation The DLS-20 Hiden s 20mm Triple Filter Quadrupole By comparison, 6mm Triple Filter Quadrupole Quadrupole High resolution Quadrupoles

More information

PulseCounter Neutron & Gamma Spectrometry Software Manual

PulseCounter Neutron & Gamma Spectrometry Software Manual PulseCounter Neutron & Gamma Spectrometry Software Manual MAXIMUS ENERGY CORPORATION Written by Dr. Max I. Fomitchev-Zamilov Web: maximus.energy TABLE OF CONTENTS 0. GENERAL INFORMATION 1. DEFAULT SCREEN

More information

2. AN INTROSPECTION OF THE MORPHING PROCESS

2. AN INTROSPECTION OF THE MORPHING PROCESS 1. INTRODUCTION Voice morphing means the transition of one speech signal into another. Like image morphing, speech morphing aims to preserve the shared characteristics of the starting and final signals,

More information

Introduction to System Suitability Tests Michael Wright Bioanalytical Services LGC Fordham, UK

Introduction to System Suitability Tests Michael Wright Bioanalytical Services LGC Fordham, UK Introduction to System Suitability Tests Michael Wright Bioanalytical Services LGC Fordham, UK Science for a safer world 13/10/2016 2 2 What is a System Suitability Test (SST)? Also referred to as a System

More information

Customer Responsibilities. Important Customer Information Infinity LC/1260 Infinity LC Site Preparation Checklist

Customer Responsibilities. Important Customer Information Infinity LC/1260 Infinity LC Site Preparation Checklist 1290 Site Preparation Infinity LC/1260 Checklist Infinity LC Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this

More information

FASTFLIGHT-2 Digital Signal Averager. Exceptionally fast LC/TOF-MS or GC/TOF-MS data acquisition... with a simple USB-2 connection to your computer!

FASTFLIGHT-2 Digital Signal Averager. Exceptionally fast LC/TOF-MS or GC/TOF-MS data acquisition... with a simple USB-2 connection to your computer! SIGNAL RECOVERY Acquire up to 100 Spectra/second with the 4 GHz FASTFLIGHT-2 Exceptionally fast LC/TOF-MS or GC/TOF-MS data acquisition... with a simple USB-2 connection to your computer! 250 ps interleaved

More information

REPORT DOCUMENTATION PAGE

REPORT DOCUMENTATION PAGE REPORT DOCUMENTATION PAGE Form Approved OMB No. 0704-0188 Public reporting burden for this collection of information is estimated to average 1 hour per response, including the time for reviewing instructions,

More information

MestReNova A quick Guide. Adjust signal intensity Use scroll wheel. Zoomen Z

MestReNova A quick Guide. Adjust signal intensity Use scroll wheel. Zoomen Z MestReNova A quick Guide page 1 MNova is a program to analyze 1D- and 2D NMR data. Start of MNova Start All Programs Chemie NMR MNova The MNova Menu 1. 2. Create expanded regions Adjust signal intensity

More information

Commissioning the TAMUTRAP RFQ cooler/buncher. E. Bennett, R. Burch, B. Fenker, M. Mehlman, D. Melconian, and P.D. Shidling

Commissioning the TAMUTRAP RFQ cooler/buncher. E. Bennett, R. Burch, B. Fenker, M. Mehlman, D. Melconian, and P.D. Shidling Commissioning the TAMUTRAP RFQ cooler/buncher E. Bennett, R. Burch, B. Fenker, M. Mehlman, D. Melconian, and P.D. Shidling In order to efficiently load ions into a Penning trap, the ion beam should be

More information

Durham Magneto Optics Ltd. NanoMOKE 3 Wafer Mapper. Specifications

Durham Magneto Optics Ltd. NanoMOKE 3 Wafer Mapper. Specifications Durham Magneto Optics Ltd NanoMOKE 3 Wafer Mapper Specifications Overview The NanoMOKE 3 Wafer Mapper is an ultrahigh sensitivity Kerr effect magnetometer specially configured for measuring magnetic hysteresis

More information

Software Operation Manual

Software Operation Manual Varian, Inc. 2700 Mitchell Drive Walnut Creek, CA 94598-1675/USA 4000 GC/MS Version 6 Software Operation Manual Varian, Inc. 2004-2009 Printed in U.S.A. 03-914999-00:Rev. 8 Trademark Acknowledgments Microsoft,

More information

GG450 4/12/2010. Today s material comes from p in the text book. Please read and understand all of this material!

GG450 4/12/2010. Today s material comes from p in the text book. Please read and understand all of this material! GG450 April 13, 2010 Seismic Reflection III Data Processing Today s material comes from p. 163-198 in the text book. Please read and understand all of this material! Reflection Processing We've been talking

More information

Precise Digital Integration of Fast Analogue Signals using a 12-bit Oscilloscope

Precise Digital Integration of Fast Analogue Signals using a 12-bit Oscilloscope EUROPEAN ORGANIZATION FOR NUCLEAR RESEARCH CERN BEAMS DEPARTMENT CERN-BE-2014-002 BI Precise Digital Integration of Fast Analogue Signals using a 12-bit Oscilloscope M. Gasior; M. Krupa CERN Geneva/CH

More information

Sample Analysis Design. Element2 - Basic Software Concepts (cont d)

Sample Analysis Design. Element2 - Basic Software Concepts (cont d) Sample Analysis Design Element2 - Basic Software Concepts (cont d) Samples per Peak In order to establish a minimum level of precision, the ion signal (peak) must be measured several times during the scan

More information

Techniques for Extending Real-Time Oscilloscope Bandwidth

Techniques for Extending Real-Time Oscilloscope Bandwidth Techniques for Extending Real-Time Oscilloscope Bandwidth Over the past decade, data communication rates have increased by a factor well over 10X. Data rates that were once 1Gb/sec and below are now routinely

More information

Technical Note. High-speed Agilent 1100 Series diode-array detector SL for optimization of resolution, sensitivity, spectral sensitivity and linearity

Technical Note. High-speed Agilent 1100 Series diode-array detector SL for optimization of resolution, sensitivity, spectral sensitivity and linearity High-speed Agilent Series diode-array detector SL for optimization of resolution, sensitivity, spectral sensitivity and linearity Technical Note Abstract This application has been verified using an Agilent

More information

Assessing and Measuring VCR Playback Image Quality, Part 1. Leo Backman/DigiOmmel & Co.

Assessing and Measuring VCR Playback Image Quality, Part 1. Leo Backman/DigiOmmel & Co. Assessing and Measuring VCR Playback Image Quality, Part 1. Leo Backman/DigiOmmel & Co. Assessing analog VCR image quality and stability requires dedicated measuring instruments. Still, standard metrics

More information

Reduction of Device Damage During Dry Etching of Advanced MMIC Devices Using Optical Emission Spectroscopy

Reduction of Device Damage During Dry Etching of Advanced MMIC Devices Using Optical Emission Spectroscopy Reduction of Device Damage During Dry Etching of Advanced MMIC Devices Using Optical Emission Spectroscopy D. Johnson, R. Westerman, M. DeVre, Y. Lee, J. Sasserath Unaxis USA, Inc. 10050 16 th Street North

More information

PAL LSI - Robust and Reliable Sample Handling

PAL LSI - Robust and Reliable Sample Handling PAL LSI - Robust and Reliable Sample Handling The PAL System, ingeniously reliable, flexible and productive PAL LSI: Innovative features giving real benefits Easy maintenance and adaptation: The different

More information

set. Important Note: the system must be calibrated before meaningful quant results can be obtained!

set. Important Note: the system must be calibrated before meaningful quant results can be obtained! Quant Initialization Before any quantitative results can be obtained, the detector parameters and the analysis parameters must be set. Important Note: the system must be calibrated before meaningful quant

More information

MODE FIELD DIAMETER AND EFFECTIVE AREA MEASUREMENT OF DISPERSION COMPENSATION OPTICAL DEVICES

MODE FIELD DIAMETER AND EFFECTIVE AREA MEASUREMENT OF DISPERSION COMPENSATION OPTICAL DEVICES MODE FIELD DIAMETER AND EFFECTIVE AREA MEASUREMENT OF DISPERSION COMPENSATION OPTICAL DEVICES Hale R. Farley, Jeffrey L. Guttman, Razvan Chirita and Carmen D. Pâlsan Photon inc. 6860 Santa Teresa Blvd

More information

Practicum 3, Fall 2010

Practicum 3, Fall 2010 A. F. Miller 2010 T1 Measurement 1 Practicum 3, Fall 2010 Measuring the longitudinal relaxation time: T1. Strychnine, dissolved CDCl3 The T1 is the characteristic time of relaxation of Z magnetization

More information

Signal processing in the Philips 'VLP' system

Signal processing in the Philips 'VLP' system Philips tech. Rev. 33, 181-185, 1973, No. 7 181 Signal processing in the Philips 'VLP' system W. van den Bussche, A. H. Hoogendijk and J. H. Wessels On the 'YLP' record there is a single information track

More information

ECE 4220 Real Time Embedded Systems Final Project Spectrum Analyzer

ECE 4220 Real Time Embedded Systems Final Project Spectrum Analyzer ECE 4220 Real Time Embedded Systems Final Project Spectrum Analyzer by: Matt Mazzola 12222670 Abstract The design of a spectrum analyzer on an embedded device is presented. The device achieves minimum

More information

STANDARD OPERATING PROCEDURE:

STANDARD OPERATING PROCEDURE: Q-EXACTIVE STANDARD OPERATING PROCEDURE: Purpose of this Instrument: Essential tool for molecular weight identification for small molecules, peptides, proteins, ect. Location: WVU Chemistry Research Laboratory

More information

Agilent 86120B, 86120C, 86122A Multi-Wavelength Meters Technical Specifications

Agilent 86120B, 86120C, 86122A Multi-Wavelength Meters Technical Specifications Agilent 86120B, 86120C, 86122A Multi-Wavelength Meters Technical Specifications March 2006 Agilent multi-wavelength meters are Michelson interferometer-based instruments that measure wavelength and optical

More information

Linrad On-Screen Controls K1JT

Linrad On-Screen Controls K1JT Linrad On-Screen Controls K1JT Main (Startup) Menu A = Weak signal CW B = Normal CW C = Meteor scatter CW D = SSB E = FM F = AM G = QRSS CW H = TX test I = Soundcard test mode J = Analog hardware tune

More information

Peak Detection with Varian MS Workstation for Varian 220 and 240 GCMS

Peak Detection with Varian MS Workstation for Varian 220 and 240 GCMS Peak Detection with Varian MS Workstation for Varian 220 and 240 GCMS by: Randall Bramston-Cook Lotus Consulting 5781 Campo Walk Long Beach, Ca 90803 310/569-0128 ebramstoncook@msn.com February 12, 2010

More information

Agilent 5345A Universal Counter, 500 MHz

Agilent 5345A Universal Counter, 500 MHz Agilent 5345A Universal Counter, 500 MHz Data Sheet Product Specifications Input Specifications (pulse and CW mode) 5356C Frequency Range 1.5-40 GHz Sensitivity (0-50 deg. C): 0.4-1.5 GHz -- 1.5-12.4 GHz

More information

Interface Practices Subcommittee SCTE STANDARD SCTE Measurement Procedure for Noise Power Ratio

Interface Practices Subcommittee SCTE STANDARD SCTE Measurement Procedure for Noise Power Ratio Interface Practices Subcommittee SCTE STANDARD SCTE 119 2018 Measurement Procedure for Noise Power Ratio NOTICE The Society of Cable Telecommunications Engineers (SCTE) / International Society of Broadband

More information

Agilent PN Time-Capture Capabilities of the Agilent Series Vector Signal Analyzers Product Note

Agilent PN Time-Capture Capabilities of the Agilent Series Vector Signal Analyzers Product Note Agilent PN 89400-10 Time-Capture Capabilities of the Agilent 89400 Series Vector Signal Analyzers Product Note Figure 1. Simplified block diagram showing basic signal flow in the Agilent 89400 Series VSAs

More information

PAL LSI - Robust and Reliable Sample Handling

PAL LSI - Robust and Reliable Sample Handling PAL LSI - Robust and Reliable Sample Handling PAL LSI: Enhanced productivity due to innovative technology Large Sample Capacity: 162 x 2 ml-vials can be stored on one tray holder. The standard length PAL

More information

CAEN Tools for Discovery

CAEN Tools for Discovery Viareggio March 28, 2011 Introduction: what is the SiPM? The Silicon PhotoMultiplier (SiPM) consists of a high density (up to ~10 3 /mm 2 ) matrix of diodes connected in parallel on a common Si substrate.

More information

Digital Signal. Continuous. Continuous. amplitude. amplitude. Discrete-time Signal. Analog Signal. Discrete. Continuous. time. time.

Digital Signal. Continuous. Continuous. amplitude. amplitude. Discrete-time Signal. Analog Signal. Discrete. Continuous. time. time. Discrete amplitude Continuous amplitude Continuous amplitude Digital Signal Analog Signal Discrete-time Signal Continuous time Discrete time Digital Signal Discrete time 1 Digital Signal contd. Analog

More information

Using the new psychoacoustic tonality analyses Tonality (Hearing Model) 1

Using the new psychoacoustic tonality analyses Tonality (Hearing Model) 1 02/18 Using the new psychoacoustic tonality analyses 1 As of ArtemiS SUITE 9.2, a very important new fully psychoacoustic approach to the measurement of tonalities is now available., based on the Hearing

More information

User Manual OVP Raman

User Manual OVP Raman Version 6 User Manual OVP Raman 2006 BRUKER OPTIK GmbH, Rudolf-Plank-Straße 27, D-76275 Ettlingen, www.brukeroptics.com All rights reserved. No part of this manual may be reproduced or transmitted in any

More information

NanoGiant Oscilloscope/Function-Generator Program. Getting Started

NanoGiant Oscilloscope/Function-Generator Program. Getting Started Getting Started Page 1 of 17 NanoGiant Oscilloscope/Function-Generator Program Getting Started This NanoGiant Oscilloscope program gives you a small impression of the capabilities of the NanoGiant multi-purpose

More information

MTS/T-BERD 8000 Platform Optical Spectrum Analyzer Modules

MTS/T-BERD 8000 Platform Optical Spectrum Analyzer Modules COMMUNICATIONS TEST & MEASUREMENT SOLUTIONS MTS/T-BERD 8000 Platform Optical Spectrum Analyzer Modules MTS/T-BERD platform Applications Provisioning and maintenance of ROADM networks Commissioning of DWDM

More information

OPTICAL MEASURING INSTRUMENTS. MS9710C 600 to 1750 nm OPTICAL SPECTRUM ANALYZER GPIB. High Performance for DWDM Optical Communications

OPTICAL MEASURING INSTRUMENTS. MS9710C 600 to 1750 nm OPTICAL SPECTRUM ANALYZER GPIB. High Performance for DWDM Optical Communications OPTICAL SPECTRUM ANALYZER 600 to 750 nm GPIB High Performance for DWDM Optical Communications The is a diffraction-grating spectrum analyzer for analyzing optical spectra in the 600 to 750 nm wavelength

More information

If you have questions or problems in providing anything described as Customer

If you have questions or problems in providing anything described as Customer Agilent Site Preparation 1260 Infinity Checklist Purification Solution Thank you for purchasing Agilent 1260 Infinity Purification Solution. To get you started and to assure a successful and timely installation,

More information

Real-time QC in HCHP seismic acquisition Ning Hongxiao, Wei Guowei and Wang Qiucheng, BGP, CNPC

Real-time QC in HCHP seismic acquisition Ning Hongxiao, Wei Guowei and Wang Qiucheng, BGP, CNPC Chengdu China Ning Hongxiao, Wei Guowei and Wang Qiucheng, BGP, CNPC Summary High channel count and high productivity bring huge challenges to the QC activities in the high-density and high-productivity

More information

CM3106 Solutions. Do not turn this page over until instructed to do so by the Senior Invigilator.

CM3106 Solutions. Do not turn this page over until instructed to do so by the Senior Invigilator. CARDIFF UNIVERSITY EXAMINATION PAPER Academic Year: 2013/2014 Examination Period: Examination Paper Number: Examination Paper Title: Duration: Autumn CM3106 Solutions Multimedia 2 hours Do not turn this

More information

Synthesized Clock Generator

Synthesized Clock Generator Synthesized Clock Generator CG635 DC to 2.05 GHz low-jitter clock generator Clocks from DC to 2.05 GHz Random jitter

More information

Performance. Reliability. Productivity. Automated Flash Chromatography Systems

Performance. Reliability. Productivity. Automated Flash Chromatography Systems Performance Reliability Productivity Automated Flash Chromatography Systems CombiFlash Rf+ Family of Purification Systems Flash chromatography is the science of refinement Teledyne Isco continues to refine

More information

DAT335 Music Perception and Cognition Cogswell Polytechnical College Spring Week 6 Class Notes

DAT335 Music Perception and Cognition Cogswell Polytechnical College Spring Week 6 Class Notes DAT335 Music Perception and Cognition Cogswell Polytechnical College Spring 2009 Week 6 Class Notes Pitch Perception Introduction Pitch may be described as that attribute of auditory sensation in terms

More information

Application Note AN-708 Vibration Measurements with the Vibration Synchronization Module

Application Note AN-708 Vibration Measurements with the Vibration Synchronization Module Application Note AN-708 Vibration Measurements with the Vibration Synchronization Module Introduction The vibration module allows complete analysis of cyclical events using low-speed cameras. This is accomplished

More information

SPECIFICATION NO Model 207 Automatic GTAW Welding System

SPECIFICATION NO Model 207 Automatic GTAW Welding System 1.0 Introduction The Model 207 is a completely self-contained Gas Tungsten Arc Welding (GTAW) System requiring only input power, inert gas and AMI Welding Head (or manual torch) for operation. Its small

More information

For the SIA. Applications of Propagation Delay & Skew tool. Introduction. Theory of Operation. Propagation Delay & Skew Tool

For the SIA. Applications of Propagation Delay & Skew tool. Introduction. Theory of Operation. Propagation Delay & Skew Tool For the SIA Applications of Propagation Delay & Skew tool Determine signal propagation delay time Detect skewing between channels on rising or falling edges Create histograms of different edge relationships

More information

MTS/T-BERD 8000 Platform

MTS/T-BERD 8000 Platform Key Features New optical design for field applications 50% reduction in size and weight for true OSNR measurements in ROADM networks Full spectral range of 1250 to 1650 nm for DWDM and CWDM testing High-resolution

More information

R&S FSW-B512R Real-Time Spectrum Analyzer 512 MHz Specifications

R&S FSW-B512R Real-Time Spectrum Analyzer 512 MHz Specifications R&S FSW-B512R Real-Time Spectrum Analyzer 512 MHz Specifications Data Sheet Version 02.00 CONTENTS Definitions... 3 Specifications... 4 Level... 5 Result display... 6 Trigger... 7 Ordering information...

More information

PS User Guide Series Seismic-Data Display

PS User Guide Series Seismic-Data Display PS User Guide Series 2015 Seismic-Data Display Prepared By Choon B. Park, Ph.D. January 2015 Table of Contents Page 1. File 2 2. Data 2 2.1 Resample 3 3. Edit 4 3.1 Export Data 4 3.2 Cut/Append Records

More information

Agilent 81600B Tunable Laser Source Family

Agilent 81600B Tunable Laser Source Family Agilent 81600B Tunable Laser Source Family Technical Specifications August 2007 The Agilent 81600B Tunable Laser Source Family offers the full wavelength range from 1260 nm to 1640 nm with the minimum

More information

Precision testing methods of Event Timer A032-ET

Precision testing methods of Event Timer A032-ET Precision testing methods of Event Timer A032-ET Event Timer A032-ET provides extreme precision. Therefore exact determination of its characteristics in commonly accepted way is impossible or, at least,

More information

Model 7330 Signal Source Analyzer Dedicated Phase Noise Test System V1.02

Model 7330 Signal Source Analyzer Dedicated Phase Noise Test System V1.02 Model 7330 Signal Source Analyzer Dedicated Phase Noise Test System V1.02 A fully integrated high-performance cross-correlation signal source analyzer from 5 MHz to 33+ GHz Key Features Complete broadband

More information

UNIT-3 Part A. 2. What is radio sonde? [ N/D-16]

UNIT-3 Part A. 2. What is radio sonde? [ N/D-16] UNIT-3 Part A 1. What is CFAR loss? [ N/D-16] Constant false alarm rate (CFAR) is a property of threshold or gain control devices that maintain an approximately constant rate of false target detections

More information

Hands-on session on timing analysis

Hands-on session on timing analysis Amsterdam 2010 Hands-on session on timing analysis Introduction During this session, we ll approach some basic tasks in timing analysis of x-ray time series, with particular emphasis on the typical signals

More information

Fast Orbit Feedback at the SLS. Outline

Fast Orbit Feedback at the SLS. Outline Fast Orbit Feedback at the SLS 2nd Workshop on Beam Orbit Stabilisation (December4-6, 2002, SPring-8) T. Schilcher Outline Noise Sources at SLS Stability / System Requirements Fast Orbit Feedback Implementation

More information

HEAD. HEAD VISOR (Code 7500ff) Overview. Features. System for online localization of sound sources in real time

HEAD. HEAD VISOR (Code 7500ff) Overview. Features. System for online localization of sound sources in real time HEAD Ebertstraße 30a 52134 Herzogenrath Tel.: +49 2407 577-0 Fax: +49 2407 577-99 email: info@head-acoustics.de Web: www.head-acoustics.de Data Datenblatt Sheet HEAD VISOR (Code 7500ff) System for online

More information

DELTA MODULATION AND DPCM CODING OF COLOR SIGNALS

DELTA MODULATION AND DPCM CODING OF COLOR SIGNALS DELTA MODULATION AND DPCM CODING OF COLOR SIGNALS Item Type text; Proceedings Authors Habibi, A. Publisher International Foundation for Telemetering Journal International Telemetering Conference Proceedings

More information

1. General principles for injection of beam into the LHC

1. General principles for injection of beam into the LHC LHC Project Note 287 2002-03-01 Jorg.Wenninger@cern.ch LHC Injection Scenarios Author(s) / Div-Group: R. Schmidt / AC, J. Wenninger / SL-OP Keywords: injection, interlocks, operation, protection Summary

More information

5G New Radio Technology and Performance. Amitava Ghosh Nokia Bell Labs July 20 th, 2017

5G New Radio Technology and Performance. Amitava Ghosh Nokia Bell Labs July 20 th, 2017 5G New Radio Technology and Performance Amitava Ghosh Nokia Bell Labs July 20 th, 2017 1 Performance : NR @ sub 6 GHz 2 Motivation: Why 5G New Radio @ sub 6GHz Ubiquitous coverage for mmtc and URLLC Access

More information

Signal Stability Analyser

Signal Stability Analyser Signal Stability Analyser o Real Time Phase or Frequency Display o Real Time Data, Allan Variance and Phase Noise Plots o 1MHz to 65MHz medium resolution (12.5ps) o 5MHz and 10MHz high resolution (50fs)

More information

FAST MOBILITY PARTICLE SIZER SPECTROMETER MODEL 3091

FAST MOBILITY PARTICLE SIZER SPECTROMETER MODEL 3091 FAST MOBILITY PARTICLE SIZER SPECTROMETER MODEL 3091 MEASURES SIZE DISTRIBUTION AND NUMBER CONCENTRATION OF RAPIDLY CHANGING SUBMICROMETER AEROSOL PARTICLES IN REAL-TIME UNDERSTANDING, ACCELERATED IDEAL

More information

Agilent 81600B Tunable Laser Source Family Technical Specifications August New model: nm, low SSE output!

Agilent 81600B Tunable Laser Source Family Technical Specifications August New model: nm, low SSE output! New model: 1260 1375 nm, low SSE output! Agilent Tunable Laser Source Family Technical Specifications August 2004 The Agilent Tunable Laser Source Family offers the from 1260 nm to 1640 nm with the minimum

More information

Development of Simple-Matrix LCD Module for Motion Picture

Development of Simple-Matrix LCD Module for Motion Picture Development of Simple-Matrix LCD Module for Motion Picture Kunihiko Yamamoto* Shinya Takahashi* Kouki Taniguchi* * A1203 Project Team Abstract A simple-matrix LCD module (12.1-in. SVGA) has been developed

More information

SPEAR 3: Operations Update and Impact of Top-Off Injection

SPEAR 3: Operations Update and Impact of Top-Off Injection SPEAR 3: Operations Update and Impact of Top-Off Injection R. Hettel for the SSRL ASD 2005 SSRL Users Meeting October 18, 2005 SPEAR 3 Operations Update and Development Plans Highlights of 2005 SPEAR 3

More information

Realizing Waveform Characteristics up to a Digitizer s Full Bandwidth Increasing the effective sampling rate when measuring repetitive signals

Realizing Waveform Characteristics up to a Digitizer s Full Bandwidth Increasing the effective sampling rate when measuring repetitive signals Realizing Waveform Characteristics up to a Digitizer s Full Bandwidth Increasing the effective sampling rate when measuring repetitive signals By Jean Dassonville Agilent Technologies Introduction The

More information

Practical Application of the Phased-Array Technology with Paint-Brush Evaluation for Seamless-Tube Testing

Practical Application of the Phased-Array Technology with Paint-Brush Evaluation for Seamless-Tube Testing ECNDT 2006 - Th.1.1.4 Practical Application of the Phased-Array Technology with Paint-Brush Evaluation for Seamless-Tube Testing R.H. PAWELLETZ, E. EUFRASIO, Vallourec & Mannesmann do Brazil, Belo Horizonte,

More information

EASY-MCS. Multichannel Scaler. Profiling Counting Rates up to 150 MHz with 15 ppm Time Resolution.

EASY-MCS. Multichannel Scaler. Profiling Counting Rates up to 150 MHz with 15 ppm Time Resolution. Multichannel Scaler Profiling Counting Rates up to 150 MHz with 15 ppm Time Resolution. The ideal solution for: Time-resolved single-photon counting Phosphorescence lifetime spectrometry Atmospheric and

More information

Part 1: Introduction to Computer Graphics

Part 1: Introduction to Computer Graphics Part 1: Introduction to Computer Graphics 1. Define computer graphics? The branch of science and technology concerned with methods and techniques for converting data to or from visual presentation using

More information

7000 Series Signal Source Analyzer & Dedicated Phase Noise Test System

7000 Series Signal Source Analyzer & Dedicated Phase Noise Test System 7000 Series Signal Source Analyzer & Dedicated Phase Noise Test System A fully integrated high-performance cross-correlation signal source analyzer with platforms from 5MHz to 7GHz, 26GHz, and 40GHz Key

More information

VivoSense. User Manual Galvanic Skin Response (GSR) Analysis Module. VivoSense, Inc. Newport Beach, CA, USA Tel. (858) , Fax.

VivoSense. User Manual Galvanic Skin Response (GSR) Analysis Module. VivoSense, Inc. Newport Beach, CA, USA Tel. (858) , Fax. VivoSense User Manual Galvanic Skin Response (GSR) Analysis VivoSense Version 3.1 VivoSense, Inc. Newport Beach, CA, USA Tel. (858) 876-8486, Fax. (248) 692-0980 Email: info@vivosense.com; Web: www.vivosense.com

More information

Advanced Sensor Technologies

Advanced Sensor Technologies Advanced Sensor Technologies Jörg Amelung Fraunhofer Institute for Photonics Microsystems Name of presenter date Sensors as core element for IoT Next phase of market grow New/Advanced Requirements based

More information

Peak Detection with Varian Star Workstation for Varian 3800 and 450 Gas Chromatographs

Peak Detection with Varian Star Workstation for Varian 3800 and 450 Gas Chromatographs Peak Detection with Varian Star Workstation for Varian 3800 and 450 Gas Chromatographs by: Randall Bramston-Cook Lotus Consulting 5781 Campo Walk Long Beach, Ca 90803 310/569-0128 ebramstoncook@msn.com

More information

TESLA FEL-Report

TESLA FEL-Report Determination of the Longitudinal Phase Space Distribution produced with the TTF Photo Injector M. Geitz a,s.schreiber a,g.von Walter b, D. Sertore a;1, M. Bernard c, B. Leblond c a Deutsches Elektronen-Synchrotron,

More information

Customer Responsibilities. Important Customer Information. Agilent InfinityLab LC Series Site Preparation Checklist

Customer Responsibilities. Important Customer Information. Agilent InfinityLab LC Series Site Preparation Checklist Agilent Site Preparation InfinityLab Checklist LC Series Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification

More information

Overview and Interpretation of D7900/D7169 Merge Analysis

Overview and Interpretation of D7900/D7169 Merge Analysis Overview and Interpretation of D7900/D7169 Merge Analysis Crude Oil Quality Association New Orleans, LA March 14, 2019 Value of Merged Simdis Analysis Requires very little sample (10-50 mls) Much faster

More information

NMR. picospin. Maintenance Guide

NMR. picospin. Maintenance Guide NMR picospin Maintenance Guide 269-302600 Revision A January 2013 2013 Thermo Fisher Scientific Inc. All rights reserved. For U.S. Technical Support, please contact: Thermo Fisher Scientific 5225 Verona

More information