1 Real-time ph and Conductivity Data Acquisition with the ACQUITY UPLC H-Class Bio System, Auto Blend Plus, and the GE Healthcare Monitor ph/c-900 INTRODUCTION The ACQUITY UPLC H-Class Bio System with Auto Blend Plus Technology can deliver buffers with accurate ph and salt content using the familiar vocabulary of ph and salt concentration, rather than requiring the calculation of pump percentages. Constant levels, steps, or gradients for both the ph and salt concentrations of mobile phases are independently controlled and can be delivered simultaneously. The performance of the system has been extensively tested and validated. The algorithms used in Auto Blend Plus, coupled with the performance of the ACQUITY UPLC H-Class Bio System, accurately and precisely deliver the programmed ph and salt concentrations. For some applications, it can be useful to supplement control functions with real-time ph and conductivity monitoring. This can be helpful during methods development to identify the conditions where a particular protein elutes. Monitoring has also revealed ways in which the delivered ph is altered while passing through an ion exchange column. It is also a convenient way to determine when a column is equilibrated with the buffered mobile phase. Finally, observation of ph and conductivity can be one of the system suitability tests that effectively measures if the system is functioning properly in the mechanical sense and that stocks of buffers and salts have been correctly prepared. Monitoring is not mandatory in any of these examples, but it can occasionally be beneficial. Waters laboratories have extensive experience with the Amersham Pharmacia (GE Healthcare) Biotech Monitor ph/c-900. This system allows the acquisition of real-time ph and conductivity data using Empower 2 or Empower 3 Software. The mv analog output from the meter can be digitally converted by a Waters esat/in module and collected as data channels in Empower. Creating a derived channel in Empower can further facilitate the visualization of this mv data as the actual ph and conductivity levels (µs/cm). Herein, we describe a simple procedure to organize the hardware and software of the ACQUITY UPLC H-Class Bio System, the esat/in module, and the Monitor ph/c-900. Some suggested settings are included. Directions for the creation of a channel derived from Empower and examples of the output possible from this configuration are also provided. 1

2 MAT E R IA L S System configuration Waters ACQUITY UPLC H-Class Bio System Waters Empower 2 or Empower 3 Software Waters esat/in (P/N ) with accompanying data cables GE Healthcare Monitor ph/c-900 with the accompanying analog cables (2), GE Healthcare conductivity flow cell (P/N ) and GE Healthcare ph electrode (P/N ) One Ethernet cable of appropriate length One I/O injection start cable A dual-row barrier strip with at least four positions (for example RadioShack P/N ) P RO C E DU R E Before beginning this procedure, all units must be unplugged and powered down. This procedure assumes that both the Monitor ph/c-900 and the ACQUITY UPLC H-Class Bio System have been set up and calibrated as directed by the manufacturer. 1. At the rear of the esat/in module, connect both the power and the Ethernet cables. The Ethernet cable is plugged into the connector marked Data. Connect the other end of the Ethernet cable to any position of the Ethernet switch located on the rear of the ACQUITY UPLC H-Class Bio System as shown in Figure Connect the two esat/in input cables at the front of the esat/in module into the Channel 1 and Channel 2 receptacles displayed in Figure Attach the inject start cable (I/O) to the inject start connector of the esat/in. Wire the plug for positions 1 and 2 as seen in Figure The other end of the I/O cable is connected to the event-in connector on the rear of the system s biosample Manager-FTN as shown in Figure On the rear panel of the Monitor ph/c-900, connect the analog signal cables to the plugs labeled ph and Cond as shown in Figure The individual wires at the ends of each of the ph/c-900 analog signal cables are plastic-capped, colored, and bear a plastic number label. Uncap wires #1 (brown) and #2 (red) for each cable. The ends of the input cable for the esat/in are a ground, a white (+), and a black (-) wire as shown in Figure Connect the ph/c-900 analog signal cables to the esat/in data cables using a dual-row barrier strip. Specifically, connect the wires as shown in Figure 7; the brown (#1) to the white (+), and the red (#2) to the black (-). The barrier strip makes this very simple. For the purpose of this set-up, make sure that the ph analog signal cable is connected to the Channel 1 data cable, and the conductivity analog signal cable is attached to the Channel 2 data cable. 8. Plug in and power on the units to complete the physical set-up. Figure 1. Rear of esat/in. Power, Ethernet connection. Corresponding Ethernet connection to biosample Manager-FTN. 2

3 Figure 2. esat/in front panel. Data cables installed in the positions for channels 1 and 2. Figure 3. esat/in front panel. Inject start cable (I/O) installed. Figure 4. Rear of biosample Manager-FTN. Inject start cable installed. Figure 5. Rear panel of ph/c-900. ph and Cond analog signal cables installed. Figure 6. The (1) capped wires of the ph/c-900 analog signal cable, and (2) the end of the esat/in input cable. The set up will require two identical cable sets: one for ph and one for conductivity. 3

4 ph Analog Cable esat/in channel 1 Conductivity Analog Cable esat/in channel 2 Figure 7. The completed connection between the esat/in and the ph/c-900 using a 10-position barrier strip. Setting the analog ph range of the Monitor ph/c-900 The Monitor ph/c-900 has a feature that allows the user to define the range of the analog out signal of the meter, specifying mv/ph units. This setting is independent of that of the calibration ph range. Setting the analog range wide or narrow will change the mv/ph unit relationship of the analog output of the meter. Table 1 gives examples of the difference in mv/ph for two analog ranges. Note: Using a derived channel in Empower 2 will necessitate a ph 0 to 14 analog range setting. Analog ph range mv/ph unit 0 to to to to Table 1. Analog ph range settings and the corresponding mv/ph units. 4

5 Derived channels for ph and Conductivity in Empower 2 Software The direct analog output from both the ph and conductivity meters of the ph/c-900 will be in millivolts (mv). While this is useful, it is more desirable to convert this data into meaningful units of ph and ms/cm values that are familiar to users. This can be accomplished by creating a derived data channel in Empower 2 or Empower 3 Software. To create derived channels, follow this procedure: 1. With all units powered and recognized by Empower, enter the project view and open a new method set. Do not enter the method set editor wizard. 2. Right click the Derived Channels in the navigation tree and select New, then Derived Channel as shown in Figure The Edit Derived Channel window is now visible as shown in Figure From the drop-down Channel menu, select esat/in Channel 1 as displayed in Figure In the Constant box, enter the conversion factor to convert the analog mv output to ph. The conversion factor is a single number that will convert the mv output from the meter to ph using the inverse of the mv/ph unit determined for the analog range set by the user. As an example, for the analog range ph 0 to 14, the mv/ph unit was 71 so the conversion factor is 1/71 or ph/mv as shown in Figure Click OK and then name the channel as shown in Figure Repeat the procedure for esat/in Channel 2, entering the conversion constant for the conductivity data, and naming the channel appropriately. The conductivity flow cell was calibrated following the procedure in the manufacturer s user manual. For a 1.00 M NaCl solution, the full scale conductivity was ms/cm. The analog output range is from 0 to 1000 mv. The conductivity unit (ms/cm) per mv for the flow cell is then ms/cm/1000 or ms/cm/mv. This is the conversion constant for the mv output of the conductivity flow cell. 8. An example of a completed method set editor window with two derived channels can be seen in Figure 13. This method set can then be used to process the multiple channel mv data from the meter, yielding results in the appropriate ph and conductivity values. Figure 8. Creating a derived channel. 5

6 Figure 9. The Edit Derived Channel window. Figure 10. esat/in channel selected. Figure 11. Conversion factor for mv to ph entered. Figure 12. Naming the derived channel. Figure 13. Example of a final Method Set editor window with the two, named derived channels. 6

7 Examples Following the procedure outlined here will allow the real-time acquisition of ph and conductivity data as channels in Empower. This data can be visualized as the direct mv output of the meter or as ph and conductivity levels in a derived channel. Two examples of this mv output and the respective derived channels are shown for ph in Figure 14, and conductivity in Figure 15. This capability also allows the simultaneous collection and visualization of more complex data. For example, a ph gradient comparison demonstrates the derived channel conversion from mv output to ph as shown in Figure 16. More complex data can also be collected, such as the acquisition of ph and conductivity channels simultaneously with the output of an optical detector yielding meaningful and valuable information as shown in Figure 17. mv ph A B Figure 14. A linear ph gradient using Auto Blend Plus as (A) mv data acquired directly from the meter, and (B) as the more meaningful ph values from the derived channel. The ph gradient ranged from 5.26 to 9.61, beginning with a 10-minute hold, then the gradient from 10 to 40 minutes, followed by a hold from 40 to 60 minutes. No column was used. A flow restrictor was in place to ensure proper gradient delivery mv A B Figure 15. A linear NaCl gradient using Auto Blend Plus as (A) mv data acquired directly from the meter, and (B) as the more meaningful µs/cm values from the Empower-derived channel. The NaCl gradient was from 0 to 250 mm NaCl, beginning with a 10-minute hold, then the gradient from 10 to 40 minutes, followed by a hold from 40 to 60 minutes. No column was used. A flow restrictor was in place to ensure proper gradient delivery µs/cm

8 CONCLUSION ph Figure 16. Linear ph gradient profile comparison between a Waters Protein-Pak Hi Res CM cation exchange column, 4.6 x 100 mm, 7 µm (blue) and no column (black). The ph gradient for both ranged from 5.26 to 9.61, beginning with a 10-minute hold, then the gradient from 10 to 40 minutes, followed by a hold from 40 to 60 minutes. The ACQUITY UPLC H-Class Bio System with Auto Blend Plus Technology can deliver ph and salinity mobile phases within the limits of the buffer stocks used. The ph levels and saline concentrations can be independently controlled and simultaneously delivered. Connecting a Monitor ph/c-900 meter through an esat/in module enables the acquisition of real-time ph and conductivity data in Empower Software. Utilizing the derived channel feature of Empower, this real-time data can be visualized as the correct ph and conductivity. This type of real-time monitoring is not required but can be very beneficial AU Figure 17. Example of a chromatogram showing the separation of a protein mixture with three types of data: UV absorbance at 280 nm (black), ph profile (blue), and conductivity (green). The method was 0 to 20 minute(s) simultaneous ph 5.26 to 9.61, and 0 to 250 mm NaCl linear gradients, followed by a 20 to 24 minute hold; then from 24 to 25 minutes a step down to ph 5.26 and up to 500 mm NaCl, a hold from 25 to 27 minutes; then from 27 to 50 minutes a return to initial conditions and re-equilibration. The sample was analyzed on a Protein-Pak Hi Res CM cation exchange column, 4.6 x 100 mm, 7 µm. Waters and ACQUITY UPLC are registered trademarks of Waters Corporation. The Science of What s Possible, Auto Blend Plus, Protein-Pak, and Empower are trademarks of Waters Corporation. All other trademarks are the property of their respective owners Waters Corporation. Produced in the U.S.A. December EN LB-PDF Waters Corporation 34 Maple Street Milford, MA U.S.A. T: F:

AIM INTRODUCTION SIMPLIFIED WORKFLOW

AIM INTRODUCTION SIMPLIFIED WORKFLOW CD-R PATENT PROTECTION: BRAND CHARACTERIZATION TO IDENTIFY COUNTERFEIT GOODS USING SIMPLIFIED MASS SPECTROMETRY James Morphet and Eleanor Riches Waters Corporation, Manchester, UK AIM To provide an easy-to-use

More information

Integration of the Zetasizer µv and OmniFACE into an SEC system

Integration of the Zetasizer µv and OmniFACE into an SEC system Integration of the Zetasizer µv and OmniFACE into an SEC system PARTICLE SIZE MOLECULAR WEIGHT Introduction The Zetasizer µv is a light scattering system capable of simultaneous dynamic (DLS) and static

More information

This document courtesy of:

This document courtesy of: This document courtesy of: Data Weighing Systems, Inc. Contact Us For immediate assistance call 1-800-750-6842 Operation Manual Sartorius Basic Meter PB-11 98648-012-08 Contents 4 General View 6 Warning

More information

ICS-3000 System Installation Qualification

ICS-3000 System Installation Qualification IC System IQ ICS-3000 System Installation Qualification 2006 Dionex Corporation Document No. 065080 Revision 02 September 2006 Doc. 065080-02 2006, Dionex Corporation All rights reserved worldwide. Printed

More information

WQ Series Water Quality Bench top Meters

WQ Series Water Quality Bench top Meters USER GUIDE WQ Series Water Quality Bench top Meters Model WQ500 ph, ORP and Temperature Model WQ510 ph, ORP, Conductivity, TDS, Salinity, & Temperature Model WQ530 ph, ORP, Conductivity, TDS, Salinity,

More information

Creating and fully utilizing the retention time standard processing method for a more automated IC workflow

Creating and fully utilizing the retention time standard processing method for a more automated IC workflow TECHNICAL NOTE Creating and fully utilizing the retention time standard processing method for a more automated IC workflow No. 178 Jay Lorch, Thermo Fisher Scientific, Sunnyvale, CA Keywords Retention

More information

OMNISTAR GX2. GX2-LM1000E Series 1310 nm Broadcast Transmitter DATA SHEET BENEFITS. 1 GHz bandwidth

OMNISTAR GX2. GX2-LM1000E Series 1310 nm Broadcast Transmitter DATA SHEET BENEFITS. 1 GHz bandwidth DATA SHEET BENEFITS OMNISTAR GX2 GX2-LM1000E Series 1310 nm Broadcast Transmitter 1 GHz bandwidth High module density up to 16 transmitter modules in a 4 RU housing High performance: Advanced predistortion

More information

Sartorius Professional Meter

Sartorius Professional Meter Operation Manual Sartorius Professional Meter PP-15, PP-20, PP-25, PP-50 98648-013-07 Contents 4 Warning and Safety Information 5 Quick Start Guide for ph Measurement 7 Introduction 8 Direct Menu Keys,

More information

Achieving More Efficient Data Review with OpenLAB CDS

Achieving More Efficient Data Review with OpenLAB CDS Achieving More Efficient Data Review with OpenLAB CDS Technical Note Introduction Separations have enjoyed major advances that have significantly reduced chromatographic run times. Ultra high performance

More information

TL-2900 AMMONIA & NITRATE ANALYZER DUAL CHANNEL

TL-2900 AMMONIA & NITRATE ANALYZER DUAL CHANNEL TL-2900 AMMONIA & NITRATE ANALYZER DUAL CHANNEL DATA ACQUISITION SYSTEM V.15.4 INSTRUCTION MANUAL Timberline Instruments, LLC 1880 S. Flatiron Ct., Unit I Boulder, Colorado 80301 Ph: (303) 440-8779 Fx:

More information

Cablecast SX. Setup Guide. c Tightrope Media Systems For Cablecast version Build 206

Cablecast SX. Setup Guide. c Tightrope Media Systems For Cablecast version Build 206 Cablecast SX Setup Guide c Tightrope Media Systems For Cablecast version 5.2.11 Build 206 Printed June 5, 2015 1 Cablecast SX Setup 1.1 Prerequisites 1.2 Overview of Setup Thank you for purchasing a Cablecast

More information

TMT data acquisition on the LTQ-Orbitrap XL Mass Spectrometer

TMT data acquisition on the LTQ-Orbitrap XL Mass Spectrometer TECH TIP # 70 TMT data acquisition on the LTQ-Orbitrap XL Mass Spectrometer Introduction TR0070.0 Successful acquisition of quantitative tandem mass tag (TMT*) data with the Thermo Scientific LTQ-Orbitrap

More information

HPLC Tips and Tricks. Tiziana Ladisa Sales Support Specialist for Chromatography Italy Thermo Fisher Scientific, Rodano (MI)

HPLC Tips and Tricks. Tiziana Ladisa Sales Support Specialist for Chromatography Italy Thermo Fisher Scientific, Rodano (MI) HPLC Tips and Tricks Tiziana Ladisa Sales Support Specialist for Chromatography Italy Thermo Fisher Scientific, Rodano (MI) The world leader in serving science Overview Thermo Scientific UltiMate 3000

More information

Scan Converter Installation Guide

Scan Converter Installation Guide Scan Converter Installation Guide Software on supplied disks Please note: The software included with your scan converter is OPTIONAL. It is not needed to make the scan converter work properly. This software

More information

The Syscal family of resistivity meters. Designed for the surveys you do.

The Syscal family of resistivity meters. Designed for the surveys you do. The Syscal family of resistivity meters. Designed for the surveys you do. Resistivity meters may conveniently be broken down into several categories according to their capabilities and applications. The

More information

ChromQuest 5.0 Chromatography Data System

ChromQuest 5.0 Chromatography Data System ChromQuest 5.0 Chromatography Data System User Guide for the SpectraSYSTEM LC CHROM-97256 Revision A March 2008 2008 Thermo Fisher Scientific Inc. All rights reserved. ChromQuest is a trademark of Thermo

More information

Instrument Control and Data Acquisition with Chromeleon

Instrument Control and Data Acquisition with Chromeleon Technical Note 67 Instrument Control and Data Acquisition with Chromeleon Introduction Laboratories generally have a combination of chromatography devices from different manufacturers and, therefore, face

More information

Liquid Chromatography- Mass Spectrometer Manual

Liquid Chromatography- Mass Spectrometer Manual Liquid Chromatography- Mass Spectrometer Manual Joshua Willis, Elizabeth Sattely Department of Chemical Engineering Stanford University November 6, 2014 Abstract This manual will explain the LC/MS, its

More information

Shad-o-Box X-Ray Camera Hardware Manual

Shad-o-Box X-Ray Camera Hardware Manual Shad-o-Box X-Ray Camera Hardware Manual P/N 1015 Rev. 08 Shad-o-Box, RadEye and ShadoCam are trademarks of Rad-icon Imaging Corp. All other brand and product names are trademarks or registered trademarks

More information

Installation / Set-up of Autoread Camera System to DS1000/DS1200 Inserters

Installation / Set-up of Autoread Camera System to DS1000/DS1200 Inserters Installation / Set-up of Autoread Camera System to DS1000/DS1200 Inserters Written By: Colin Langridge Issue: Draft Date: 03 rd July 2008 1 Date: 29 th July 2008 2 Date: 20 th August 2008 3 Date: 02 nd

More information

Data Acquisition Using LabVIEW

Data Acquisition Using LabVIEW Experiment-0 Data Acquisition Using LabVIEW Introduction The objectives of this experiment are to become acquainted with using computer-conrolled instrumentation for data acquisition. LabVIEW, a program

More information

ENGR 1000, Introduction to Engineering Design

ENGR 1000, Introduction to Engineering Design ENGR 1000, Introduction to Engineering Design Unit 2: Data Acquisition and Control Technology Lesson 2.4: Programming Digital Ports Hardware: 12 VDC power supply Several lengths of wire NI-USB 6008 Device

More information

CytoFLEX Flow Cytometer Quick Start Guide

CytoFLEX Flow Cytometer Quick Start Guide Sheath Waste CLASS 1 LASER PRODUCT COMPLIES WITH 21 CFR 1040.10 AND 1040.11 EXCEPT FOR DEVIATIONS PURSUANT TO LASER NOTICE NO. 50 DATED JUNE 24, 2007 MANUFACTURED Sheath B49008AC February 2015 CytoFLEX

More information

Application Note Using Buffered Outputs and Patch Panels with the SETPOINT Machinery Protection System

Application Note Using Buffered Outputs and Patch Panels with the SETPOINT Machinery Protection System Application Note Using Buffered Outputs and Patch Panels with the SETPOINT Machinery Protection System Doc 1446106 Page 1 of 8 Overview The SETPOINT Machinery Protection System provides three separate

More information

PACSystems* RX3i. Isolated Thermocouple Input Module, 6 Channels, IC695ALG306-EB Isolated Thermocouple Input Module, 12 Channels, IC695ALG312-EB

PACSystems* RX3i. Isolated Thermocouple Input Module, 6 Channels, IC695ALG306-EB Isolated Thermocouple Input Module, 12 Channels, IC695ALG312-EB September 2013 PACSystems* RX3i Isolated Thermocouple Input Module, 6 Channels, IC695ALG306-EB Isolated Thermocouple Input Module, 12 Channels, IC695ALG312-EB Isolated +24 VDC Power Isolated Thermocouple

More information

STB Front Panel User s Guide

STB Front Panel User s Guide S ET-TOP BOX FRONT PANEL USER S GUIDE 1. Introduction The Set-Top Box (STB) Front Panel has the following demonstration capabilities: Pressing 1 of the 8 capacitive sensing pads lights up that pad s corresponding

More information

MAP Optical Power Meter Module (mopm-b1)

MAP Optical Power Meter Module (mopm-b1) COMMUNICATIONS TEST & MEASUREMENT SOLUTIONS MAP Optical Power Meter Module (mopm-b1) Key Features Panel mount or remote head configuration Single, dual, or quad channel configurations available 250 khz

More information

ivw-fd122 Video Wall Controller MODEL: ivw-fd122 Video Wall Controller Supports 2 x 2 Video Wall Array User Manual Page i Rev. 1.

ivw-fd122 Video Wall Controller MODEL: ivw-fd122 Video Wall Controller Supports 2 x 2 Video Wall Array User Manual Page i Rev. 1. MODEL: ivw-fd122 Video Wall Controller Supports 2 x 2 Video Wall Array User Manual Rev. 1.01 Page i Copyright COPYRIGHT NOTICE The information in this document is subject to change without prior notice

More information

LM/TM-30xx, 31xx Series LCD Monitor User s Manual Rev. A0

LM/TM-30xx, 31xx Series LCD Monitor User s Manual Rev. A0 LM/TM-30xx, 31xx Series LCD Monitor User s Manual Rev. A0 FCC NOTICE This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instructions

More information

Cablecast Server. Setup Guide. c Tightrope Media Systems For Cablecast version Build 74

Cablecast Server. Setup Guide. c Tightrope Media Systems For Cablecast version Build 74 Cablecast Server Setup Guide c Tightrope Media Systems For Cablecast version 6.1.2 Build 74 Printed July 22, 2016 1 Cablecast Server Setup 1.1 Prerequisites 1.2 Overview of Setup Thank you for purchasing

More information

OmniStar GX2 Headend Optics Platform GX2 LM1000E Series

OmniStar GX2 Headend Optics Platform GX2 LM1000E Series arris.com OmniStar GX2 Headend Optics Platform GX2 LM1000E Series 1310 nm Broadcast Transmitter FEATURES High density optical platform with up to 16 transmitters per chassis for headend space optimization

More information

MTI-2100 FOTONIC SENSOR. High resolution, non-contact. measurement of vibration. and displacement

MTI-2100 FOTONIC SENSOR. High resolution, non-contact. measurement of vibration. and displacement A worldwide leader in precision measurement solutions MTI-2100 FOTONIC SENSOR High resolution, non-contact measurement of vibration and displacement MTI-2100 Fotonic TM Sensor Unmatched Resolution and

More information

LAUREL ELECTRONICS, INC.

LAUREL ELECTRONICS, INC. LAUREL ELECTRONICS, INC. Laureate Digital Panel Meter for Process, Strain & Potentiometer Follower Signals Features Selectable ±0.2, ±2, ±20, ±200, ±300 & ±600 Vdc voltage ranges Selectable ±2, ±20, ±200

More information

QIP7232 P2. Hybrid QAM/IP High-definition Set-top. Quick Start Guide

QIP7232 P2. Hybrid QAM/IP High-definition Set-top. Quick Start Guide QIP7232 P2 Hybrid QAM/IP High-definition Set-top Quick Start Guide Before You Begin Introduction Congratulations on receiving a Motorola QIP7232 Hybrid QAM/IP High-definition Set-top. This document will

More information

Operating Instructions

Operating Instructions CNTX Contrast sensor Operating Instructions CAUTIONS AND WARNINGS SET-UP DISTANCE ADJUSTMENT: As a general rule, the sensor should be fixed at a 15 to 20 angle from directly perpendicular to the target

More information

DM1624, DM1612, DM812

DM1624, DM1612, DM812 Installation Guide Hardware and Software DM Series Digital Processors models DM1624, DM1612, DM812 LECTROSONICS, INC. 1 Installation Specific Information Only This guide covers only installation related

More information

15 Inch CGA EGA VGA to XGA LCD Wide Viewing Angle Panel ID# 833

15 Inch CGA EGA VGA to XGA LCD Wide Viewing Angle Panel ID# 833 15 Inch CGA EGA VGA to XGA LCD Wide Viewing Angle Panel ID# 833 Operation Manual Introduction This monitor is an open frame LCD Panel monitor. It features the VESA plug & play system which allows the monitor

More information

IQORX30 / IQORX31. Single Mode Fiber Optic Receivers for 3G/HD/SD-SDI Signals

IQORX30 / IQORX31. Single Mode Fiber Optic Receivers for 3G/HD/SD-SDI Signals IQORX30 / IQORX3 Single Mode Fiber Optic Receivers for 3G/HD/SD-SDI Signals Operator s Manual May 009 Snell & Wilcox Ltd., Southleigh Park House, Eastleigh Road, Havant, Hants, PO9 PE, United Kingdom.

More information

4 MHz Lock-In Amplifier

4 MHz Lock-In Amplifier 4 MHz Lock-In Amplifier SR865A 4 MHz dual phase lock-in amplifier SR865A 4 MHz Lock-In Amplifier 1 mhz to 4 MHz frequency range Low-noise current and voltage inputs Touchscreen data display - large numeric

More information

CLIPSTER. 3D LUT File Generation with the Kodak Display Manager. Supplement

CLIPSTER. 3D LUT File Generation with the Kodak Display Manager. Supplement Supplement: CLIPSTER 3D LUT File Generation with the Kodak Display Manager (Version 1.0) CLIPSTER 3D LUT File Generation with the Kodak Display Manager Supplement Supplement for the CLIPSTER Documentation:

More information

Scan Converter Quick Installation Guide

Scan Converter Quick Installation Guide Scan Converter Quick Installation Guide Software Note: No software is required to use your scan converter. Please complete the hardware installation and system setup before you determine the need to the

More information

Dell Wyse 5030 PCoIP Zero Client

Dell Wyse 5030 PCoIP Zero Client Dell Wyse 5030 PCoIP Zero Client User Guide Regulatory Model: PxN Regulatory Type: PxN001 Notes, cautions, and warnings NOTE: A NOTE indicates important information that helps you make better use of your

More information

Overview. Shipped in the Venue Vizion Package: Simplified Integration Process. Installation consists of 6 easy steps:

Overview. Shipped in the Venue Vizion Package: Simplified Integration Process. Installation consists of 6 easy steps: Overview Shipped in the Venue Vizion Package: Four two-channel QMOD Encoder-Modulators Each unit can accept up to 2 inputs, providing up to 8 channels in the system. One ICE-HE-DXL Display Control Center

More information

EGM Einthoven Goldberger Module Type 701

EGM Einthoven Goldberger Module Type 701 Operating Instructions for the PLUGSYS Module EGM Einthoven Goldberger Module Type 701 ECG amplifier for bipolar extremity leads after Einthoven and unipolar extremity leads after Goldberger (Version:

More information

Gazer VI700A-SYNC/IN and VI700W- SYNC/IN INSTALLATION MANUAL

Gazer VI700A-SYNC/IN and VI700W- SYNC/IN INSTALLATION MANUAL Gazer VI700A-SYNC/IN and VI700W- SYNC/IN INSTALLATION MANUAL Contents List of compatible cars... 3 Package contents... 4 Special information... 6 Car interior disassembly and connection guide for Ford

More information

Yamaha LC4 Installation Guide

Yamaha LC4 Installation Guide Installation Guide Yamaha LC4 Installation Guide Contents LC4 Installation Page List of LC4 Controller Parts................. 1 Step 1: Positioning of Instruments in Class...... 2 Step 2: Standard Installation

More information

Fiber-coupled light sources

Fiber-coupled light sources Optogenetics catalog 7.4 - Fiber-coupled light sources 9 Fiber-coupled light sources The fiber optic circuits are driven by light and hence the need to couple the light sources into the optical fiber.

More information

Honeywell HomMed Wireless Scale Quick Reference Guide

Honeywell HomMed Wireless Scale Quick Reference Guide Honeywell HomMed Wireless Scale Quick Reference Guide Wireless Scale Bluetooth Enabled (Optional Peripheral) Introduction This Quick Reference Guide together with the Genesis DM Manual are both training

More information

LAUREL. Laureate Digital Panel Meter for Load Cell & Microvolt Input ELECTRONICS, INC. Features. Description

LAUREL. Laureate Digital Panel Meter for Load Cell & Microvolt Input ELECTRONICS, INC. Features. Description Description LAUREL ELECTRONICS, INC. Features Laureate Digital Panel Meter for Load Cell & Microvolt Input 20, 50, 100, 250 & 500 mv ranges Span adjust from 0 to ±99,999, zero adjust from -99,999 to +99,999

More information

SwiftMix Automation Safety

SwiftMix Automation Safety Operations Manual SwiftMix TM Automation Thank you for your purchase of the SwiftMix automation for the 5088 console. Everyone at Rupert Neve Designs hopes you enjoy using this tool as much as we have

More information

Setup Guide. Dolby PRM Rev. 1.1

Setup Guide. Dolby PRM Rev. 1.1 Setup Guide Dolby PRM-4200 Rev. 1.1 Introduction CalMAN Required Software Version CalMAN Display Calibration Software can automatically create an optimized 3D calibration LUT for the Dolby PRM-4200 monitor.

More information

User Guide. HDMI Fiber Optic Extender. DVI-7350a

User Guide. HDMI Fiber Optic Extender. DVI-7350a User Guide HDMI Fiber Optic Extender DVI-7350a Table of Contents Section Page Product Safety.................................... 1 1.0 Introduction...2 2.0 Specifications...3 3.0 Package Contents...3 4.0

More information

013-RD

013-RD Engineering Note Topic: Product Affected: JAZ-PX Lamp Module Jaz Date Issued: 08/27/2010 Description The Jaz PX lamp is a pulsed, short arc xenon lamp for UV-VIS applications such as absorbance, bioreflectance,

More information

Triple RTD. On-board Digital Signal Processor. Linearization RTDs 20 Hz averaged outputs 16-bit precision comparator function.

Triple RTD. On-board Digital Signal Processor. Linearization RTDs 20 Hz averaged outputs 16-bit precision comparator function. Triple RTD SMART INPUT MODULE State-of-the-art Electromagnetic Noise Suppression Circuitry. Ensures signal integrity even in harsh EMC environments. On-board Digital Signal Processor. Linearization RTDs

More information

Release Notes for LAS AF version 1.8.0

Release Notes for LAS AF version 1.8.0 October 1 st, 2007 Release Notes for LAS AF version 1.8.0 1. General Information A new structure of the online help is being implemented. The focus is on the description of the dialogs of the LAS AF. Configuration

More information

Quick Setup Guide for IntelliAg Model CTA

Quick Setup Guide for IntelliAg Model CTA STEP 3: Auto Configuration (identifies sensors connected to each module) Auto config is performed at the factory, but may need to be done in the field as changes are made to the system or if options are

More information

Vasudevan Agilent 1100 Series HPLC w/ DAD & FLD Detector (nonbuffer

Vasudevan Agilent 1100 Series HPLC w/ DAD & FLD Detector (nonbuffer Vasudevan Agilent 1100 Series HPLC w/ DAD & FLD Detector (nonbuffer solvents) Updated November 14, 2017 Instrument instructions can be found at: http://academic.bowdoin.edu/chemistry/resources/instructions.shtml

More information

Model KBF 115 Constant climate chambers with large temperature / humidity range

Model KBF 115 Constant climate chambers with large temperature / humidity range Model KBF 115 Constant climate chambers with large temperature / humidity range The BINDER KBF is the specialist for unconditionally reliable stability testing and precise maintenance of constant climate

More information

GALILEO Timing Receiver

GALILEO Timing Receiver GALILEO Timing Receiver The Space Technology GALILEO Timing Receiver is a triple carrier single channel high tracking performances Navigation receiver, specialized for Time and Frequency transfer application.

More information

MadiXtreme / Alpha-Link XLogic I/O system for PC and Mac Setup Guide V1.0 XLogic. This is SSL.

MadiXtreme / Alpha-Link XLogic I/O system for PC and Mac Setup Guide V1.0 XLogic. This is SSL. www.solidstatelogic.com MadiXtreme / Alpha-Link XLogic I/O system for PC and Mac Setup Guide V1.0 XLogic. This is SSL. Document History 82BSA101A March 2012 V1.0 Initial Release Contents Introduction 1

More information

Using an oscilloscope - The Hameg 203-6

Using an oscilloscope - The Hameg 203-6 Using an oscilloscope - The Hameg 203-6 What does an oscilloscope do? Setting up How does an oscilloscope work? Other oscilloscope controls Connecting a function generator Microphones audio signals and

More information

Compact live fiber identifier with integrated optical power meter

Compact live fiber identifier with integrated optical power meter FI-Series Live Fiber Identifier Compact live fiber identifier with integrated optical power meter User Manual ZP-PKG-0535 REV 1 Notice Copyright Trademarks Every effort was made to ensure that the information

More information

12.1 Inch CGA EGA VGA SVGA LCD Panel - ID #492

12.1 Inch CGA EGA VGA SVGA LCD Panel - ID #492 12.1 Inch CGA EGA VGA SVGA LCD Panel - ID #492 Operation Manual Introduction This monitor is an open frame LCD Panel monitor. It features the VESA plug & play system which allows the monitor to automatically

More information

FS1-X. Quick Start Guide. Overview. Frame Rate Conversion Option. Two Video Processors. Two Operating Modes

FS1-X. Quick Start Guide. Overview. Frame Rate Conversion Option. Two Video Processors. Two Operating Modes FS1-X Quick Start Guide Overview Matching up and synchronizing disparate video and audio formats is a critical part of any broadcast, mobile or post-production environment. Within its compact 1RU chassis,

More information

Model 4700 Photodiode Characterizer

Model 4700 Photodiode Characterizer Model 4700 Photodiode Characterizer Complete PD Measurement system The 4700 Photodiode Characterizer is a complete photodiode test system. It will characterize PDs or APDs (upcoming) without the need for

More information

EXT-3GSDI-FOSM User Manual

EXT-3GSDI-FOSM User Manual EXT-3GSDI-FOSM User Manual www.gefen.com ASKING FOR ASSISTANCE Technical Support: Telephone (818) 772-9100 (800) 545-6900 Fax (818) 772-9120 Technical Support Hours: 8:00 AM to 5:00 PM PST Monday through

More information

1. Add Capture Antibody diluted in coating buffer. 4. Add standards, controls and samples to the plate in duplicate.

1. Add Capture Antibody diluted in coating buffer. 4. Add standards, controls and samples to the plate in duplicate. Matched antibody pair kits include a capture and biotinylated detector antibody pair and a calibrated protein standard. Kits are available in two sizes, with enough reagents for either 2 or 10 x 96-well

More information

High intensity led illuminator ILED42A FEATURES: DESCRIPTION: ELECTRICAL CHARACTERISTICS:

High intensity led illuminator ILED42A FEATURES: DESCRIPTION: ELECTRICAL CHARACTERISTICS: High intensity led illuminator May 2010 Rev 0.3 ILED42A FEATURES: High luminous flux Low cost Fully protected Wide input voltage range Available with different led colours and lenses DESCRIPTION: The IDP

More information

Gigabit Multi-mode SX to Single Mode LX Converter. User s Manual NGF-728 Series. Warning COPYRIGHT

Gigabit Multi-mode SX to Single Mode LX Converter. User s Manual NGF-728 Series. Warning COPYRIGHT COPYRIGHT Gigabit Multi-mode SX to Single Mode LX Converter User s Manual NGF-728 Series All rights reserved. No part of this publication may be reproduced, stored in a retrieval system, or transmitted

More information

Installation Note. Agilent 89441A Option AYC Installation Kit

Installation Note. Agilent 89441A Option AYC Installation Kit Installation Note Agilent 89441A Option AYC Installation Kit Part Number: 89441-90074 Printed in USA February 2002 Notice. The information contained in this document is subject to change without notice.

More information

Revision 1.2d

Revision 1.2d Specifications subject to change without notice 0 of 16 Universal Encoder Checker Universal Encoder Checker...1 Description...2 Components...2 Encoder Checker and Adapter Connections...2 Warning: High

More information

Altman Lighting Spectra Cyc 50 Specification

Altman Lighting Spectra Cyc 50 Specification 1.01 CYCLORAMA LIGHTING A. General 1. The fixture shall be a compact, lightweight color-mixing LED asymmetrical wash fixture with 8 or 16 bit DMX control of intensity and color. The fixture shall be the

More information

ivw-fd133 Video Wall Controller MODEL: ivw-fd133 Video Wall Controller Supports 3 x 3 and 2 x 2 Video Wall Array User Manual Page i Rev. 1.

ivw-fd133 Video Wall Controller MODEL: ivw-fd133 Video Wall Controller Supports 3 x 3 and 2 x 2 Video Wall Array User Manual Page i Rev. 1. MODEL: ivw-fd133 Video Wall Controller Supports 3 x 3 and 2 x 2 Video Wall Array User Manual Rev. 1.01 Page i Copyright COPYRIGHT NOTICE The information in this document is subject to change without prior

More information

Model KBF P 240 Humidity test chambers with ICH-compliant light source

Model KBF P 240 Humidity test chambers with ICH-compliant light source Model KBF P 240 Humidity test chambers with ICH-compliant light source The KBF P humidity test chamber, equipped with ICH-compliant light source, is an expert when it comes to photostability tests and

More information

MS2540 Current Loop Receiver with RS485 Communication

MS2540 Current Loop Receiver with RS485 Communication MS2540 Current Loop Receiver with RS485 Communication User Manual Metal Samples Company A Division of Alabama Specialty Products, Inc. 152 Metal Samples Rd., Munford, AL 36268 Phone: (256) 358 4202 Fax:

More information

Gazer VI700A-SYNC2 and VI700W- SYNC2 INSTALLATION MANUAL

Gazer VI700A-SYNC2 and VI700W- SYNC2 INSTALLATION MANUAL Gazer VI700A-SYNC2 and VI700W- SYNC2 INSTALLATION MANUAL Contents List of compatible cars... 3 Package contents... 4 Special information... 6 Car interior disassembly and connection guide for Ford Focus...

More information

J R Sky, Inc. tel: fax:

J R Sky, Inc.  tel: fax: STEREO OPTICAL RECORDING SYSTEM N UOPTIX STEREO OPTICAL RECORDING MONITOR LEFT SYSTEM MODE PREVIEW RECORD BIAS RECORD REV SETUP TEST RIGHT INPUT SETUP INPUT BIAS SETUP BIAS INPUT STEREO AUX MONO DIRECT

More information

MRF-300/RFX150 INSTALLATION MANUAL

MRF-300/RFX150 INSTALLATION MANUAL MRF-300/RFX150 INSTALLATION MANUAL TM Multi-Zone RF Base Station for the MX-3000, the Aurora, TM TM TM the Aeros, the Omega and the Osiris remote controls. MRF-300 Installation Manual 2005 Universal Remote

More information

2013, 2014 Hewlett-Packard Development Company, L.P.

2013, 2014 Hewlett-Packard Development Company, L.P. User Guide 2013, 2014 Hewlett-Packard Development Company, L.P. The only warranties for HP products and services are set forth in the express warranty statements accompanying such products and services.

More information

Warning and Safety Information. FCC Information

Warning and Safety Information. FCC Information Installation Manual Warning and Safety Information FCC Information This device complies with FCC Rules Part 15 Operation and is subject to the following two conditions: (1) This device may not cause harmful

More information

Flat-Bed Module Recorders

Flat-Bed Module Recorders Flat-Bed Module Recorders Model No. 08376-50 08376-55 08376-60 0115-0192 4/28/00 Table of Contents Introduction...3 Power Requirements...3 Chart Paper Installation...3 Pen Installation...5 Grounding...5

More information

Speed Performance Reliability. Medicinal Chemistry Natural Products Peptides & Polymers Organic Synthesis Purifications

Speed Performance Reliability. Medicinal Chemistry Natural Products Peptides & Polymers Organic Synthesis Purifications Automated Flash Chromatography Systems Medicinal Chemistry Natural Products Peptides & Polymers Organic Synthesis Purifications Speed Performance Reliability CombiFlash Rf - Making Fl Improve Your Productivity

More information

BNC-2110 DESKTOP AND DIN RAIL-MOUNTABLE BNC ADAPTER

BNC-2110 DESKTOP AND DIN RAIL-MOUNTABLE BNC ADAPTER INSTALLATION GUIDE BNC-2110 DESKTOP AND DIN RAIL-MOUNTABLE BNC ADAPTER Introduction This installation guide describes how to install and configure your BNC-2110 accessory with an E Series or waveform generation

More information

SC26 Magnetic Field Cancelling System

SC26 Magnetic Field Cancelling System SPICER CONSULTING SYSTEM SC26 SC26 Magnetic Field Cancelling System Makes the ambient magnetic field OK for electron beam tools in 300 mm wafer fabs Real time, wideband cancelling from DC to > 9 khz fields

More information

ADVANCED OPTICAL FIBER SOLUTIONS

ADVANCED OPTICAL FIBER SOLUTIONS Fiber Laser Building Blocks Fiber Laser Cavities and All-Fiber Beam Combiners A Furukawa Company www.ofsoptics.com ADVANCED OPTICAL FIBER SOLUTIONS for Your Next Multi-Kilowatt Fiber Laser Applications

More information

26 Inch CGA/EGA/VGA/DVI to WXGA/1080p LCD - ID#703

26 Inch CGA/EGA/VGA/DVI to WXGA/1080p LCD - ID#703 26 Inch CGA/EGA/VGA/DVI to WXGA/1080p LCD - ID#703 Operation Manual Introduction This monitor is an open frame LCD Panel monitor. It features the VESA plug & play system which allows the monitor to automatically

More information

with the Field-IQ Crop Input Control System

with the Field-IQ Crop Input Control System with the Field-IQ Crop Input Control System Quick Reference Card CONNECTING THE SYSTEM Ag25 GNSS antenna (P/N 68040-OOS) TNC/TNC right-angle cable (P/N 50449) Cable assembly, display to Field-IQ (P/N 50449)

More information

Camera Control Unit 55D-BS VOLUME OPEN CABLE MIC OFF GND 2W CAMERA CABLE TALLY. Camera Adapter 55D-CA 9.39" 7.40" 5.2"

Camera Control Unit 55D-BS VOLUME OPEN CABLE MIC OFF GND 2W CAMERA CABLE TALLY. Camera Adapter 55D-CA 9.39 7.40 5.2 The Telemetrics Coax/Fiber Link is an affordable camera control system with increased operating distance. Using frequency multiplexing the following signals are transmitted over a single coaxial cable:

More information

Quick Setup Guide for IntelliAg Model NTA

Quick Setup Guide for IntelliAg Model NTA STEP 3: Auto Configuration (identifies sensors connected to each module) Auto config is performed at the factory, but may need to be done in the field as changes are made to the system or if options are

More information

Industrial Diode Laser (IDL) System IDL Series

Industrial Diode Laser (IDL) System IDL Series COMMERCIAL LASERS Industrial Diode Laser (IDL) System IDL Series Key Features Round, top-hat beam profile for uniform power distribution Warranted for full rated power in either pulsed or continuous wave

More information

FS3. Quick Start Guide. Overview. FS3 Control

FS3. Quick Start Guide. Overview. FS3 Control FS3 Quick Start Guide Overview The new FS3 combines AJA's industry-proven frame synchronization with high-quality 4K up-conversion technology to seamlessly integrate SD and HD signals into 4K workflows.

More information

PACSystems* RX3i Thermocouple Input Module, 12 Channels, IC695ALG412-CB

PACSystems* RX3i Thermocouple Input Module, 12 Channels, IC695ALG412-CB September 2013 PACSystems* RX3i Thermocouple Input Module, 12 Channels, IC695ALG412-CB The PACSystems * Thermocouple Input module IC695ALG412 provides twelve isolated differential thermocouple input channels.

More information

Advanced Setup Guide

Advanced Setup Guide Advanced Setup Guide The only warranties for Hewlett-Packard products and services are set forth in the express statements accompanying such products and services. Nothing herein should be construed as

More information

Model CB 160 CO 2 incubators with hot air sterilization and heat sterilizable CO 2 sensor

Model CB 160 CO 2 incubators with hot air sterilization and heat sterilizable CO 2 sensor Model CB 160 CO 2 incubators with hot air sterilization and heat sterilizable CO 2 sensor The BINDER CB series CO 2 incubator is the premium class among the CO 2 incubators. It is suitable for all sensitive

More information

1.2 GHz GS7000 Node RF Split Upgrade Application Note

1.2 GHz GS7000 Node RF Split Upgrade Application Note 1.2 GHz GS7000 Node RF Split Upgrade Application Note Overview Introduction Cable operators have experienced an exponential rise in the requirement for more reverse path bandwidth due to the popularity

More information

Installation and Users Guide Addendum. Software Mixer Reference and Application. Macintosh OSX Version

Installation and Users Guide Addendum. Software Mixer Reference and Application. Macintosh OSX Version Installation and Users Guide Addendum Software Mixer eference and Application Macintosh OSX Version ynx Studio Technology Inc. www.lynxstudio.com support@lynxstudio.com Copyright 2004, All ights eserved,

More information

Prisma Optical Networks Ancillary Modules

Prisma Optical Networks Ancillary Modules Optoelectronics Prisma Optical Networks Ancillary Modules Description The Prisma platform is capable of utilizing a combination of modules which address a variety of revenue generating applications. The

More information

DMC550 Technical Reference

DMC550 Technical Reference DMC550 Technical Reference 2002 DSP Development Systems DMC550 Technical Reference 504815-0001 Rev. B September 2002 SPECTRUM DIGITAL, INC. 12502 Exchange Drive, Suite 440 Stafford, TX. 77477 Tel: 281.494.4505

More information

DATA CENTER OPTICAL ASSEMBLIES TO SUPPORT QSFP, CFP AND CXP INDUSTRY STANDARD PINOUT CONFIGURATIONS

DATA CENTER OPTICAL ASSEMBLIES TO SUPPORT QSFP, CFP AND CXP INDUSTRY STANDARD PINOUT CONFIGURATIONS DATA CENTER OPTICAL ASSEMBLIES TO SUPPORT QSFP, CFP AND CXP INDUSTRY STANDARD PINOUT CONFIGURATIONS Data Center Optical Assemblies to Support QSFP, CFP and CXP Industry Standard Pinout Configurations QSFP,

More information

S8000 Series Modular ph & ORP Systems

S8000 Series Modular ph & ORP Systems S8000 Series Modular ph & ORP Systems Modular Flat Surface ph/orp Electrodes Configurable for In-line or Submersion Use Optional plug-in Electronic Interfaces Enhanced Sensorex ERP Reference System REDOX,

More information