Overview. Signal Averaged ECG
|
|
- Paul Roberts
- 5 years ago
- Views:
Transcription
1 Updated : Signal Averaged ECG Overview Signal Averaged ECG The Biopac Student Lab System can be used to amplify and enhance the ECG signal using a clinical diagnosis tool referred to as the Signal Averaged ECG. The technique allows you to identify small electrical currents, called ventricular (ven-trik'u-ler) late potentials. This test is used to help identify people at risk of a dangerous rhythm in the heart's lower (pumping) chambers. This is called ventricular arrhythmia (ah-rith'me-ah), and it can lead to sudden cardiac death. For years, cardiologists examined paper charts of ECG signals to create an averaged ECG complex. They took each individual cardiac cycle, measured each part of the complex with a ruler, wrote down each value, and calculated the means to derive an averaged ECG complex. As you can imagine (or have experienced), this task took a great deal of time and patience. When the need arose to analyze two or more different scenarios (e.g., baseline response), the task had to be performed even more carefully for accurate comparison of the ECGs. The BSL PRO software has a powerful averaging feature that automatically derives an averaged signal from a specific number of cycles (averaging is found in the Find Peak dialogue). And, you can easily compare two, three, or more averaged ECG signals with just a few mouse clicks. NOTE: This lesson uses L05 Electrocardiography I as source data. If you have already recorded the ECG data you want to average, click here to jump to the Analysis section. Page 1 of 7
2 Objectives 1. Optional: Record multiple segments of ECG data under different conditions. 2. Provide a Signal Averaged ECG for each of the recording segments. 3. Compare the Signal Averaged ECG data for different conditions. Equipment BIOPAC electrode set (SS2L) BIOPAC disposable vinyl electrodes (EL503) three electrodes per subject BIOPAC electrode gel (GEL1) BIOPAC abrasive pad (ELPAD) or skin cleanser or alcohol prep Cot or lab table and pillow Computer running Windows 7/VistaXP or Mac OS X Biopac Student Lab PRO software BIOPAC Data Acquisition Unit (MP36/35/30/45) Setup Hardware 1. Make sure that the MP3X unit is off. 2. Turn on the computer. 3. Plug the electrode lead (SS2L) into CH 2. Subject 1. Remove all substantial metal jewelry from the Subject and make sure that the subject is not touching any metal (metal pipes, chairs etc.). 2. Place three electrodes on the subject as the table below indicates. Electrode Right arm Right leg Left leg Position Centered on the anterior wrist (same side as the palm of the hand), about " down from the palm Medial, just above the ankle, flat on skin (not over bone) Medial, just above the ankle, flat on skin (not over bone). Lead Color White Black Red 3. Attach the electrode leads by color as shown in the table above. The pinch connectors work like small clothes pins; however, they will attach to the electrode only on one side. You may have to rotate the pin to make sure the metal on the inside of the clip is connected, touching and clamped onto the electrode at the base of the nipple. Clip connector cables to subject's clothing, or place so that there is no strain on the electrode clips or the cable wires at any point in the setup. 4. Turn the MP3X unit on (assuming the AC100A power adapter has already been connected). 5. Wait 5 minutes after the electrodes have been attached to the skin to begin recording to give the gel time to settle and maximize conductivity. Page 2 of 7
3 Software 1. Launch the BSL Lessons software on the host computer. If the BSL Lessons software is not available, click here to set up BSL PRO. 2. Choose Lesson 5: L05-ECG-1 3. Save as the desired file name when prompted. Calibration Automatic in BSL Lesson 5; not required if using BSL PRO. Recording Follow BSL Lesson 5 instructions to record the following segments and label markers as indicated: Segment Duration Marker Label Condition 1 60 sec Supine Subject should be lying down, and not moving or talking at any time while recording is taking place sec Seated Subject should sit up quickly, then start recording. Do not record while subject is sitting up or making any type of movements sec Deep breathing sec After exercise Subject should take about 5 deep breaths over the recording period. Subject should exercise for a few minutes, such as pushups, sit-ups, jumping jacks, etc., to dramatically increase heart rate. Connectors can be removed and reconnected to subject, however the electrodes should not be taken off. Subject should then relax and record the 60-second period after exercise has stopped, but while the heart rate is still increased. Note: Any movement of the body while recording will cause small electrical impulses that will be detected by the electrodes, and thus recorded as inaccurate data. It is very important that the subject not move at anytime during recording. Page 3 of 7
4 Analysis Set up the Averaging Function To use the averaging function for ECG in BSL PRO, first determine the typical length of each ECG cycle. BSL PRO identifies the peak of each R-wave as a reference point. Because of this, set a window around the R- wave to look backward at the P and Q waves and forward to examine the S and T components. 1. Launch the BSL PRO software. 2. Open the ECG data file in BSL PRO. 3. Set a measurement channel to SC (selected channel) and the measurement to Delta T. 4. Zoom in on one representative ECG complex from the first segment. 5. With the cursor on the peak of the R-wave, click and drag backward past the P-wave to the midpoint of the P-wave and previous T-wave (see below). 6. Write down the delta T value (remember to note negative time); this time period represents the averaging period before the R-wave. 7. Click on the peak of the R-wave, and drag the cursor past the T wave. 8. Write down the delta T value (positive time); this time period represents the averaging period after the R-wave. 9. Use the markers to identify segments and highlight the area that you are averaging, as shown below. 10. Select Transform > Find Peak. Page 4 of 7
5 11. Check the Off-line Averaging box. 12. Set the cursors: o Enter the negative delta T measurement into the Set first cursor to box. o Enter the positive delta T measurement into the Set second cursor to box. 13. Check that the Threshold Level is not zero. o The value of the Threshold Level in the Peak Detector box determines how the software will read each R-wave. The value displayed in the Threshold Level box was determined by the selected area you chose. If the value is 0, cancel and reselect the area you want to view. 14. Click the Setup Averaging box. 15. Set the Averaging Range to Selected Area and click OK. 16. Click the Ave Start button. o The system will calculate the averaged ECG Complex. The signal will look similar to this: Page 5 of 7
6 Comparing Multiple Waveforms While this averaged ECG signal by itself provides an excellent diagnostic tool, the ability to compare other scenarios will be an even more valuable tool. The following graph shows the averaged ECG complex for the segments of data recorded when the subject is at rest, after sitting up, breathing deeply, and immediately after exercise. 1. Repeat above procedure for all segments, using the same peak parameters for each segment. IMPORTANT! Do not recalculate or change the time before and after the peak (the cursor settings in the Find Peak dialog) for each segment of data. You must use the same time frame for all segments to center all the averaged data around the R-wave so that when all the averaged graphs are compared they will align with respect to the R-wave. 2. View all analyses in one graph window. a. Select one of the averaged wave forms and choose Edit > Select All. b. Copy the averaged complex. c. Choose another R-waveform box (the one you wish to paste the copied waveform in to). d. Choose Edit>Insert Waveform. All your R-waveforms should be in one graph window. 3. Label each channel individually: On the first averaged waveform, click the Channel Label box (far left of the channel), and change the name of the channel. 4. Change the colors of the waveforms to provide more contrast in the final presentation: Right-click on the waveform, choose Color, select a color from the options provided. 5. Set all channels to the same Scale: Click on the Vertical Axis (right side of each channel) and click all three of the "Apply to All Channels" buttons in the Set Screen Vertical Axis dialog. NOTE: You must use the "Apply to All Channels" command. If you do not, the channels will not be uniform in Scope mode, and you will not get an accurate comparison. 6. Click the Scope Mode icon in the Tool Bar (top left corner) to change the graphs from Chart mode to Scope mode. Chart mode (the default) displays each individual averaged waveform, while Scope mode provides an overlapped comparison of each averaged ECG cycle. Chart Mode Page 6 of 7
7 Scope mode If Scope mode does not display all the waveforms in the window: Two very similar R-waves may be overlapping one another. All of the waveforms may not be set to the same scale. Make sure to select the "Apply to All Channels" commands in the Set Vertical Scale box. If Apply to All Channels is not selected, the channels will not be uniform in Scope mode and will not provide an accurate comparison. Recording with BSL PRO if BSL Lesson 5 is not available 1. Launch the BSL PRO software on the host computer. 2. From the File > Open menu, select Files of type: Graph Template (*.gtl) and open the H23 Signal Averaging ECG.gtl file. 3. Follow the Recording instructions. IMPORTANT! Remember to type your comments into the Marker field when commencing a new recording segment. This will allow you to correlate the body position with the data when performing the averaging process. Appendix GRAPH TEMPLATE SETTINGS Click here to open a PDF of the graph template file settings. The BSL PRO Graph Template file automatically establishes the settings shown in the table. Page 7 of 7
Lesson 14 BIOFEEDBACK Relaxation and Arousal
Physiology Lessons for use with the Biopac Student Lab Lesson 14 BIOFEEDBACK Relaxation and Arousal Manual Revision 3.7.3 090308 EDA/GSR Richard Pflanzer, Ph.D. Associate Professor Indiana University School
More informationLesson 10 Manual Revision BIOPAC Systems, Inc.
Physiology Lessons for use with the Lesson 10 ELECTROOCULOGRAM (EOG) I Eye Movement Saccades and Fixation during Reading Manual Revision 3.7.3 061808 Vertical J.C. Uyehara, Ph.D. Biologist BIOPAC Systems,
More informationLesson 1 EMG 1 Electromyography: Motor Unit Recruitment
Physiology Lessons for use with the Biopac Science Lab MP40 Lesson 1 EMG 1 Electromyography: Motor Unit Recruitment PC running Windows XP or Mac OS X 10.3-10.4 Lesson Revision 1.20.2006 BIOPAC Systems,
More information#PS168 - Analysis of Intraventricular Pressure Wave Data (LVP Analysis)
BIOPAC Systems, Inc. 42 Aero Camino Goleta, Ca 93117 Ph (805)685-0066 Fax (805)685-0067 www.biopac.com info@biopac.com #PS168 - Analysis of Intraventricular Pressure Wave Data (LVP Analysis) The Biopac
More informationUser Guide EMG. This user guide has been created to educate and inform the reader about doing EMG measurements
User Guide EMG This user guide has been created to educate and inform the reader about doing EMG measurements For more information about NeXus, our BioTrace+ software, please visit our website or contact
More informationHeart Rate Variability Preparing Data for Analysis Using AcqKnowledge
APPLICATION NOTE 42 Aero Camino, Goleta, CA 93117 Tel (805) 685-0066 Fax (805) 685-0067 info@biopac.com www.biopac.com 01.06.2016 Application Note 233 Heart Rate Variability Preparing Data for Analysis
More informationBASIC TUTORIAL. Jocelyn Mariah Kremer, Mike Mullins Documentation BIOPAC Systems, Inc. William McMullen Vice President BIOPAC Systems, Inc.
BASIC TUTORIAL Windows 7, Vista, or XP Mac OS X 10.5-10.7 Jocelyn Mariah Kremer, Mike Mullins Documentation BIOPAC Systems, Inc. William McMullen Vice President BIOPAC Systems, Inc. BIOPAC Systems, Inc.
More informationExercise 1: Muscles in Face used for Smiling and Frowning Aim: To study the EMG activity in muscles of the face that work to smile or frown.
Experiment HP-9: Facial Electromyograms (EMG) and Emotion Exercise 1: Muscles in Face used for Smiling and Frowning Aim: To study the EMG activity in muscles of the face that work to smile or frown. Procedure
More informationiworx Sample Lab Experiment HM-3: The Electrogastrogram (EGG) and the Growling Stomach
Experiment HM-3: The Electrogastrogram (EGG) and the Growling Stomach Background Do you ever wonder why your stomach growls, that funny sound it makes when you are really hungry? Stomach growling is the
More informationLesson 9. Physiology Lessons for use with the Biopac Student Lab GALVANIC SKIN RESPONSE & THE POLYGRAPH
Physiology Lessons for use with the Biopac Student Lab PC under Windows 95/98/NT 4.0/2000 or Macintosh Lesson 9 GALVANIC SKIN RESPONSE & THE POLYGRAPH Manual Revision 08232001.PL3.6.6-ML3.0.3 Richard Pflanzer,
More informationBiopac Student Lab PRO Manual. Professional Version 3.7.7
Biopac Student Lab PRO Manual Professional Version 3.7.7 Reference Manual for Biopac Student Lab PRO Software and MP36/35/30 or MP45 Hardware 42 Aero Camino, Goleta, CA 93117 (805) 685-0066, FAX (805)
More informationExperiment HM-8: The Electrogastrogram (EGG) and the Growling Stomach
Experiment HM-8: The Electrogastrogram (EGG) and the Growling Stomach Background Do you ever wonder why your stomach growls, that funny sound it makes when you are really hungry? Stomach growling is the
More informationActivity P27: Speed of Sound in Air (Sound Sensor)
Activity P27: Speed of Sound in Air (Sound Sensor) Concept DataStudio ScienceWorkshop (Mac) ScienceWorkshop (Win) Speed of sound P27 Speed of Sound 1.DS (See end of activity) (See end of activity) Equipment
More informationElectrocardiogram (ECG) Version Date: 2017-FEB-15 Document Effective Date: 2017-MAY-15
Title: Electrocardiogram (ECG) Version Date: 2017-FEB-15 Document Effective Date: 2017-MAY-15 Number: Data Collection Site (DCS) Version: 2.4 Number of Pages: SOP_DCS_0009 5 1.0 Purpose: The purpose of
More informationPseudorandom Stimuli Following Stimulus Presentation
BIOPAC Systems, Inc. 42 Aero Camino Goleta, CA 93117 Ph (805) 685-0066 Fax (805) 685-0067 www.biopac.com info@biopac.com Application Note AS-222 05.06.05 Pseudorandom Stimuli Following Stimulus Presentation
More informationPlease feel free to download the Demo application software from analogarts.com to help you follow this seminar.
Hello, welcome to Analog Arts spectrum analyzer tutorial. Please feel free to download the Demo application software from analogarts.com to help you follow this seminar. For this presentation, we use a
More informationGetting Started. Connect green audio output of SpikerBox/SpikerShield using green cable to your headphones input on iphone/ipad.
Getting Started First thing you should do is to connect your iphone or ipad to SpikerBox with a green smartphone cable. Green cable comes with designators on each end of the cable ( Smartphone and SpikerBox
More informationAnalyzing and Saving a Signal
Analyzing and Saving a Signal Approximate Time You can complete this exercise in approximately 45 minutes. Background LabVIEW includes a set of Express VIs that help you analyze signals. This chapter teaches
More informationBiopac Student Lab Lesson 9 ELECTRODERMAL ACTIVITY & POLYGRAPH Analysis Procedure. Rev
42 Aero Camino, Goleta, CA 93117 www.biopac.com Biopac Student Lab Lesson 9 ELECTRODERMAL ACTIVITY & POLYGRAPH Analysis Procedure Rev. 12292017 Richard Pflanzer, Ph.D. Associate Professor Emeritus Indiana
More informationME EN 363 ELEMENTARY INSTRUMENTATION Lab: Basic Lab Instruments and Data Acquisition
ME EN 363 ELEMENTARY INSTRUMENTATION Lab: Basic Lab Instruments and Data Acquisition INTRODUCTION Many sensors produce continuous voltage signals. In this lab, you will learn about some common methods
More informationTL-2900 AMMONIA & NITRATE ANALYZER DUAL CHANNEL
TL-2900 AMMONIA & NITRATE ANALYZER DUAL CHANNEL DATA ACQUISITION SYSTEM V.15.4 INSTRUCTION MANUAL Timberline Instruments, LLC 1880 S. Flatiron Ct., Unit I Boulder, Colorado 80301 Ph: (303) 440-8779 Fx:
More informationGame Show Physiology
iworx Physiology Lab Experiment Experiment HN-8: Game Show Physiology Game Show Physiology Note: The lab presented here is intended for evaluation purposes only. iworx users should refer to the User Area
More informationExperiment P32: Sound Waves (Sound Sensor)
PASCO scientific Vol. 2 Physics Lab Manual P32-1 Experiment P32: (Sound Sensor) Concept Time SW Interface Macintosh file Windows file waves 45 m 700 P32 P32_SOUN.SWS EQUIPMENT NEEDED Interface musical
More informationQuick Reference Manual
Quick Reference Manual V1.0 1 Contents 1.0 PRODUCT INTRODUCTION...3 2.0 SYSTEM REQUIREMENTS...5 3.0 INSTALLING PDF-D FLEXRAY PROTOCOL ANALYSIS SOFTWARE...5 4.0 CONNECTING TO AN OSCILLOSCOPE...6 5.0 CONFIGURE
More informationmedlab One Channel ECG OEM Module EG 01000
medlab One Channel ECG OEM Module EG 01000 Technical Manual Copyright Medlab 2012 Version 2.4 11.06.2012 1 Version 2.4 11.06.2012 Revision: 2.0 Completely revised the document 03.10.2007 2.1 Corrected
More informationE X P E R I M E N T 1
E X P E R I M E N T 1 Getting to Know Data Studio Produced by the Physics Staff at Collin College Copyright Collin College Physics Department. All Rights Reserved. University Physics, Exp 1: Getting to
More informationActivity P32: Variation of Light Intensity (Light Sensor)
Activity P32: Variation of Light Intensity (Light Sensor) Concept DataStudio ScienceWorkshop (Mac) ScienceWorkshop (Win) Illuminance P32 Vary Light.DS P54 Light Bulb Intensity P54_BULB.SWS Equipment Needed
More informationProcessing data with Mestrelab Mnova
Processing data with Mestrelab Mnova This exercise has three parts: a 1D 1 H spectrum to baseline correct, integrate, peak-pick, and plot; a 2D spectrum to plot with a 1 H spectrum as a projection; and
More informationDektak Step by Step Instructions:
Dektak Step by Step Instructions: Before Using the Equipment SIGN IN THE LOG BOOK Part 1: Setup 1. Turn on the switch at the back of the dektak machine. Then start up the computer. 2. Place the sample
More informationTutorial 3 Normalize step-cycles, average waveform amplitude and the Layout program
Tutorial 3 Normalize step-cycles, average waveform amplitude and the Layout program Step cycles are defined usually by choosing a recorded ENG waveform that shows long lasting, continuos, consistently
More informationWhat s New in Raven May 2006 This document briefly summarizes the new features that have been added to Raven since the release of Raven
What s New in Raven 1.3 16 May 2006 This document briefly summarizes the new features that have been added to Raven since the release of Raven 1.2.1. Extensible multi-channel audio input device support
More informationNIHR Southampton Biomedical Research Centre
The NIHR Southampton (BRC) has a tight quality assurance system for the writing, reviewing and updating of Standard Operating Procedures. As such, version-controlled and QA authorised Standard Operating
More informationSpectrum Analyser Basics
Hands-On Learning Spectrum Analyser Basics Peter D. Hiscocks Syscomp Electronic Design Limited Email: phiscock@ee.ryerson.ca June 28, 2014 Introduction Figure 1: GUI Startup Screen In a previous exercise,
More informationHelp Manual. Monitoring Devices. For the following devices: SJM Confirm Implantable Cardiac Monitor Confirm Rx Insertable Cardiac Monitor
Monitoring Devices Merlin Patient Care System Help Manual For the following devices: SJM Confirm Implantable Cardiac Monitor Confirm Rx Insertable Cardiac Monitor Unless otherwise noted, indicates that
More informationDIGITAL PERSONAL STUDIO Version 1.30 Addendum
DIGITAL PERSONAL STUDIO Version 1.30 Addendum Contents V1.30 FEATURES...1 AK.SYS TRACKVIEW...2 INSTALLING AK.SYS TRACKVIEW...2 USING AK.SYS TRACKVIEW...3 METERS...4 IN / OUT TIMES...5 TIMECODE DISPLAY...5
More informationSession 1 Introduction to Data Acquisition and Real-Time Control
EE-371 CONTROL SYSTEMS LABORATORY Session 1 Introduction to Data Acquisition and Real-Time Control Purpose The objectives of this session are To gain familiarity with the MultiQ3 board and WinCon software.
More informationCalibrating the timecode signal input
Chapter 5 Calibrating the timecode signal input Computer hardware can introduce an offset between the timecode signal and the video signal, which causes the timecode and video to be offset when they are
More informationBioGraph Infiniti Physiology Suite
Thought Technology Ltd. 2180 Belgrave Avenue, Montreal, QC H4A 2L8 Canada Tel: (800) 361-3651 ٠ (514) 489-8251 Fax: (514) 489-8255 E-mail: mail@thoughttechnology.com Webpage: http://www.thoughttechnology.com
More informationTopic: Instructional David G. Thomas December 23, 2015
Procedure to Setup a 3ɸ Linear Motor This is a guide to configure a 3ɸ linear motor using either analog or digital encoder feedback with an Elmo Gold Line drive. Topic: Instructional David G. Thomas December
More informationSigPlay User s Guide
SigPlay User s Guide . . SigPlay32 User's Guide? Version 3.4 Copyright? 2001 TDT. All rights reserved. No part of this manual may be reproduced or transmitted in any form or by any means, electronic or
More informationiworx Sample Lab Experiment AN-13: Crayfish Motor Nerve
Experiment AN-13: Crayfish Motor Nerve Background The purpose of this experiment is to record the extracellular action potentials of crayfish motor axons. These spontaneously generated action potentials
More informationUSB Mini Spectrum Analyzer User s Guide TSA5G35
USB Mini Spectrum Analyzer User s Guide TSA5G35 Triarchy Technologies, Corp. Page 1 of 21 USB Mini Spectrum Analyzer User s Guide Copyright Notice Copyright 2011 Triarchy Technologies, Corp. All rights
More informationReason Overview3. Reason Overview
Reason Overview3 In this chapter we ll take a quick look around the Reason interface and get an overview of what working in Reason will be like. If Reason is your first music studio, chances are the interface
More informationEAN-Performance and Latency
EAN-Performance and Latency PN: EAN-Performance-and-Latency 6/4/2018 SightLine Applications, Inc. Contact: Web: sightlineapplications.com Sales: sales@sightlineapplications.com Support: support@sightlineapplications.com
More informationCable Calibration Function for the 2400B/C and 2500A/B Series Microwave Signal Generators. Technical Brief
Cable Calibration Function for the 2400B/C and 2500A/B Series Microwave Signal Generators Technical Brief Quickly and easily apply a level correction table to compensate for external losses or power variations
More informationThe Measurement Tools and What They Do
2 The Measurement Tools The Measurement Tools and What They Do JITTERWIZARD The JitterWizard is a unique capability of the JitterPro package that performs the requisite scope setup chores while simplifying
More informationStandard Operating Procedure of nanoir2-s
Standard Operating Procedure of nanoir2-s The Anasys nanoir2 system is the AFM-based nanoscale infrared (IR) spectrometer, which has a patented technique based on photothermal induced resonance (PTIR),
More informationSIDRA INTERSECTION 8.0 UPDATE HISTORY
Akcelik & Associates Pty Ltd PO Box 1075G, Greythorn, Vic 3104 AUSTRALIA ABN 79 088 889 687 For all technical support, sales support and general enquiries: support.sidrasolutions.com SIDRA INTERSECTION
More informationThought Technology Ltd Belgrave Avenue, Montreal, QC H4A 2L8 Canada
Thought Technology Ltd. 2180 Belgrave Avenue, Montreal, QC H4A 2L8 Canada Tel: (800) 361-3651 ٠ (514) 489-8251 Fax: (514) 489-8255 E-mail: _Hmail@thoughttechnology.com Webpage: _Hhttp://www.thoughttechnology.com
More informationCapstone Experiment Setups & Procedures PHYS 1111L/2211L
Capstone Experiment Setups & Procedures PHYS 1111L/2211L Picket Fence 1. Plug the photogate into port 1 of DIGITAL INPUTS on the 850 interface box. Setup icon. the 850 box. Click on the port 1 plug in
More informationIntroduction To LabVIEW and the DSP Board
EE-289, DIGITAL SIGNAL PROCESSING LAB November 2005 Introduction To LabVIEW and the DSP Board 1 Overview The purpose of this lab is to familiarize you with the DSP development system by looking at sampling,
More informationPHY221 Lab 1 Discovering Motion: Introduction to Logger Pro and the Motion Detector; Motion with Constant Velocity
PHY221 Lab 1 Discovering Motion: Introduction to Logger Pro and the Motion Detector; Motion with Constant Velocity Print Your Name Print Your Partners' Names Instructions August 31, 2016 Before lab, read
More informationGetting started with Spike Recorder on PC/Mac/Linux
Getting started with Spike Recorder on PC/Mac/Linux You can connect your SpikerBox to your computer using either the blue laptop cable, or the green smartphone cable. How do I connect SpikerBox to computer
More informationPrecision DeEsser Users Guide
Precision DeEsser Users Guide Metric Halo $Revision: 1670 $ Publication date $Date: 2012-05-01 13:50:00-0400 (Tue, 01 May 2012) $ Copyright 2012 Metric Halo. MH Production Bundle, ChannelStrip 3, Character,
More informationMackie Control and Cubase SX/SL
Mackie Control and Cubase SX/SL - 1 - The information in this document is subject to change without notice and does not represent a commitment on the part of Steinberg Media Technologies AG. The software
More informationHow to Set Up Continuous EEG (CEEG)
How to Set Up Continuous EEG (CEEG) OBTAIN SUPPLIES 1. EEG module (yellow) 2. EEG cable with wires 3. NuPrep cream and a face cloth 4. Paediatric electrodes (use new package) STORING Location All supplies
More informationDDA-UG-E Rev E ISSUED: December 1999 ²
7LPHEDVH0RGHVDQG6HWXS 7LPHEDVH6DPSOLQJ0RGHV Depending on the timebase, you may choose from three sampling modes: Single-Shot, RIS (Random Interleaved Sampling), or Roll mode. Furthermore, for timebases
More informationISCEV SINGLE CHANNEL ERG PROTOCOL DESIGN
ISCEV SINGLE CHANNEL ERG PROTOCOL DESIGN This spreadsheet has been created to help design a protocol before actually entering the parameters into the Espion software. It details all the protocol parameters
More informationLab experience 1: Introduction to LabView
Lab experience 1: Introduction to LabView LabView is software for the real-time acquisition, processing and visualization of measured data. A LabView program is called a Virtual Instrument (VI) because
More informationHV/PHA Adjustment (PB) Part
HV/PHA Adjustment (PB) Part Contents Contents 1. How to set Part conditions...1 1.1 Setting conditions... 1 2. HV/PHA adjustment sequence...7 3. How to use this Part...9 HV/PHA Adjustment (PB) Part i
More informationTransitHound Cellphone Detector User Manual Version 1.3
TransitHound Cellphone Detector User Manual Version 1.3 RF3 RF2 Table of Contents Introduction...3 PC Requirements...3 Unit Description...3 Electrical Interfaces...4 Interface Cable...5 USB to Serial Interface
More informationAppendix D. UW DigiScope User s Manual. Willis J. Tompkins and Annie Foong
Appendix D UW DigiScope User s Manual Willis J. Tompkins and Annie Foong UW DigiScope is a program that gives the user a range of basic functions typical of a digital oscilloscope. Included are such features
More informationHBI Database. Version 2 (User Manual)
HBI Database Version 2 (User Manual) St-Petersburg, Russia 2007 2 1. INTRODUCTION...3 2. RECORDING CONDITIONS...6 2.1. EYE OPENED AND EYE CLOSED CONDITION....6 2.2. VISUAL CONTINUOUS PERFORMANCE TASK...6
More informationBiometrics Ltd LABORATORY DATA ACQUISITION SYSTEM OPERATING MANUAL. TYPE NOS. LS900 Including DataLINK DLK900 Software version 5.0
Biometrics Ltd LABORATORY DATA ACQUISITION SYSTEM OPERATING MANUAL TYPE NOS. LS900 Including DataLINK DLK900 Software version 5.0 This manual should be used in conjunction with Biometrics Ltd Goniometer
More informationPRODUCT SHEET
PULSE OXIMETRY OXY100E Module (18-321 BPM) OXY200 Module (for veterinary use only, 18-450 BPM) TSD124 Series SPO2 Transducers for OXY100E TSD270 Series SPO2 Transducers for OXY200 These modules measure
More informationExperiment BI-1: ECG Signal Conditioning
Experiment BI-1: ECG Signal Conditioning Equipment Required PC Mac Computer IX-TA-220 data acquisition unit and power supply USB cale ire-b3g ECG cale and electrode lead wires C-DIN-BB: Din to Breadoard
More information0.01 (amplitude) The amplitude at 0.01s from QRS loop (angle) The angle at 0.01s from QRS loop
Operation Instructions for Resting ECG 8 8 font ring 0.01 (amplitude) The amplitude at 0.01s from QRS loop 0.01 (angle) The angle at 0.01s from QRS loop 0.02 (amplitude) The amplitude at 0.02s from QRS
More informationPRODUCT SHEET
PULSE OXIMETRY OXY100E Module (18-321 BPM) OXY200 Module (for veterinary use only, 18-450 BPM) TSD124 Series SPO2 Transducers for OXY100E or OXYSSH-SYS TSD270 Series SPO2 Transducers for OXY200 These modules
More informationUSB Mini Spectrum Analyzer User Manual TSA Program for PC TSA4G1 TSA6G1 TSA8G1
USB Mini Spectrum Analyzer User Manual TSA Program for PC TSA4G1 TSA6G1 TSA8G1 Triarchy Technologies Corp. Page 1 of 17 USB Mini Spectrum Analyzer User Manual Copyright Notice Copyright 2013 Triarchy Technologies,
More informationMTL Software. Overview
MTL Software Overview MTL Windows Control software requires a 2350 controller and together - offer a highly integrated solution to the needs of mechanical tensile, compression and fatigue testing. MTL
More informationQuick Start. RSHS1000 Series Handheld Digital Oscilloscope
Quick Start RSHS1000 Series Handheld Digital Oscilloscope General Safety Summary Carefully read the following safety precautions to avoid personal injury and prevent damage to the instrument or any products
More informationIntroduction to the oscilloscope and digital data acquisition
Introduction to the oscilloscope and digital data acquisition Eric D. Black California Institute of Technology v1.1 There are a certain number of essential tools that are so widely used that every aspiring
More informationAgilent DSO5014A Oscilloscope Tutorial
Contents UNIVERSITY OF CALIFORNIA AT BERKELEY College of Engineering Department of Electrical Engineering and Computer Sciences EE105 Lab Experiments Agilent DSO5014A Oscilloscope Tutorial 1 Introduction
More informationLogic Analyzer Auto Run / Stop Channels / trigger / Measuring Tools Axis control panel Status Display
Logic Analyzer The graphical user interface of the Logic Analyzer fits well into the overall design of the Red Pitaya applications providing the same operating concept. The Logic Analyzer user interface
More informationAnalysis of AP/axon classes and PSP on the basis of AP amplitude
Analysis of AP/axon classes and PSP on the basis of AP amplitude In this analysis manual, we aim to measure and analyze AP amplitudes recorded with a suction electrode and synaptic potentials recorded
More informationExperiment PP-1: Electroencephalogram (EEG) Activity
Experiment PP-1: Electroencephalogram (EEG) Activity Exercise 1: Common EEG Artifacts Aim: To learn how to record an EEG and to become familiar with identifying EEG artifacts, especially those related
More informationOperating Guide. ViewClix offers a revolutionary experience for seniors and their families and friends.
ViewClix Mini TM Operating Guide ViewClix offers a revolutionary experience for seniors and their families and friends. To make using ViewClix an easy and fun experience for you and your loved ones, we
More informationA low noise multi electrode array system for in vitro electrophysiology. Mobius Tutorial AMPLIFIER TYPE SU-MED640
A low noise multi electrode array system for in vitro electrophysiology Mobius Tutorial AMPLIFIER TYPE SU-MED640 Information in this document is subject to change without notice.no part of this document
More informationSystem Requirements SA0314 Spectrum analyzer:
System Requirements SA0314 Spectrum analyzer: System requirements Windows XP, 7, Vista or 8: 1 GHz or faster 32-bit or 64-bit processor 1 GB RAM 10 MB hard disk space \ 1. Getting Started Insert DVD into
More informationTransmitter Interface Program
Transmitter Interface Program Operational Manual Version 3.0.4 1 Overview The transmitter interface software allows you to adjust configuration settings of your Max solid state transmitters. The following
More informationCHEMISTRY SEMESTER ONE
APPENDIX A USING THE SPECTROMETER FOR AN EMISSION SPECTROSCOPY NANSLO REMOTE WEB-BASED SCIENCE LAB ACTIVITY The following provides information how to use the spectrometer controls for the Emission Spectroscopy
More informationApplication Note AN-708 Vibration Measurements with the Vibration Synchronization Module
Application Note AN-708 Vibration Measurements with the Vibration Synchronization Module Introduction The vibration module allows complete analysis of cyclical events using low-speed cameras. This is accomplished
More informationStart with our emedia Catalog Click My Help! Slide 3. Slide 4 Select Read ebooks
Slide 1 Learn to Learn how to download ebooks and transfer them to your ereader. Presented by Lauren Stokes, Virtual Library Manager. Contact information for additional assistance: 702.507.6300 or via
More informationNew DSP Family CM Sweep Feature
Introduction Application Note The purpose of this document is to provide instruction on the proper use of the CM Sweep feature, included on the 1G DSP, and optional on the 360 DSP and 720 DSP meters. The
More informationDave Jones Design Phone: (607) Lake St., Owego, NY USA
Manual v1.00a June 1, 2016 for firmware vers. 2.00 Dave Jones Design Phone: (607) 687-5740 34 Lake St., Owego, NY 13827 USA www.jonesvideo.com O Tool Plus - User Manual Main mode NOTE: New modules are
More informationiworx Sample Lab Experiment AM-9: Crayfish Gut Pharmacology
Experiment AM-9: Crayfish Gut Pharmacology The Dissection 1. In a dissecting pan, cover a crayfish with ice for 5 minutes. 2. With scissors, remove the entire crayfish abdomen (tail). 3. Place the crayfish
More informationColor Correction in Final Cut Studio Introduction to Color
Color Correction in Final Cut Studio Introduction to Color Part 1: Getting Started with Color Part 2: Managing and Applying Grades upart 3: Using the Scopes and Auto Balanceo Part 4: Copying from One Clip
More informationNOTICE: This document is for use only at UNSW. No copies can be made of this document without the permission of the authors.
Brüel & Kjær Pulse Primer University of New South Wales School of Mechanical and Manufacturing Engineering September 2005 Prepared by Michael Skeen and Geoff Lucas NOTICE: This document is for use only
More informationMIE 402: WORKSHOP ON DATA ACQUISITION AND SIGNAL PROCESSING Spring 2003
MIE 402: WORKSHOP ON DATA ACQUISITION AND SIGNAL PROCESSING Spring 2003 OBJECTIVE To become familiar with state-of-the-art digital data acquisition hardware and software. To explore common data acquisition
More informationUSB AUDIO INTERFACE I T
USB AUDIO INTERFACE EN DE FR ES IT JA Contents Introduction...3 Contents in this Operation Manual... 3 Features... 3 Panel Controls and Terminals (Details)...4 Rear Panel... 4 Front Panel... 6 Panel Controls
More informationimso-104 Manual Revised August 5, 2011
imso-104 Manual Revised August 5, 2011 Section 1 Getting Started SAFETY 1.10 Quickstart Guide 1.20 SAFETY 1.30 Compatibility 1.31 Hardware 1.32 Software Section 2 How it works 2.10 Menus 2.20 Analog Channel
More informationProcedures Guide. Tektronix. HDMI Sink Instruments Differential Impedance Measurement
Procedures Guide Tektronix HDMI Sink Instruments Differential Impedance Measurement Rev. 1.1: October 13, 2010 2 Measurement Procedures Equipment Required Table 1 lists the equipment required to perform
More informationPersonal Protective Equipment Wear nitrile gloves, lab coat, and safety glasses as a minimum protection, unless otherwise indicated.
4pt Bending, Mouse This protocol is for standard Jepsen 4pt bending of adult mouse bone. Safety considerations Please reference the Jepsen laboratory when using this protocol. This protocol is subject
More informationPLEN Motion Editor Operation Procedures Temporary File
1 PLEN Motion Editor Operation Procedures Temporary File 2 Table of Contents I. Introduction 3 1. Overview of MotionEditor 3 2. Operating Environment. 3 3. Installing the MotionEditor... 3 4. PLEN's Preparation
More informationAnalyze Frequency Response (Bode Plots) with R&S Oscilloscopes Application Note
Analyze Frequency Response (Bode Plots) with R&S Oscilloscopes Application Note Products: R&S RTO2002 R&S RTO2004 R&S RTO2012 R&S RTO2014 R&S RTO2022 R&S RTO2024 R&S RTO2044 R&S RTO2064 This application
More informationSTB Front Panel User s Guide
S ET-TOP BOX FRONT PANEL USER S GUIDE 1. Introduction The Set-Top Box (STB) Front Panel has the following demonstration capabilities: Pressing 1 of the 8 capacitive sensing pads lights up that pad s corresponding
More informationSOUND LABORATORY LING123: SOUND AND COMMUNICATION
SOUND LABORATORY LING123: SOUND AND COMMUNICATION In this assignment you will be using the Praat program to analyze two recordings: (1) the advertisement call of the North American bullfrog; and (2) the
More informationLinrad On-Screen Controls K1JT
Linrad On-Screen Controls K1JT Main (Startup) Menu A = Weak signal CW B = Normal CW C = Meteor scatter CW D = SSB E = FM F = AM G = QRSS CW H = TX test I = Soundcard test mode J = Analog hardware tune
More informationHardware Setup. HP Dual TV Tuner/Digital Video Recorder. Document Part Number:
Hardware Setup HP Dual TV Tuner/Digital Video Recorder Document Part Number: 374787-001 November 2004 This guide provides steps to help you set up your HP Dual TV Tuner/Digital Video Recorder hardware
More informationB2 Spice A/D Tutorial Author: B. Mealy revised: July 27, 2006
B2 Spice A/D Tutorial Author: B. Mealy revised: July 27, 2006 The B 2 Spice A/D software allows for the simulation of digital, analog, and hybrid circuits. CPE 169, however, is only concerned with the
More information