1 4pt Bending, Mouse This protocol is for standard Jepsen 4pt bending of adult mouse bone. Safety considerations Please reference the Jepsen laboratory when using this protocol. This protocol is subject to change. We hold no responsibility for errors in the protocol or user error that lead to personal or property damage. Please follow all federal, local, and institutional rules and regulations when using this protocol. Personal Protective Equipment Wear nitrile gloves, lab coat, and safety glasses as a minimum protection, unless otherwise indicated. Potential Hazards Hydraulic testing systems have the potential to do great bodily harm if used improperly. Never have any part of your body below the actuator when the hydraulics are on. Materials Several cotton swabs (to use dowels for practice test) 50lb Sensotec load cell (S/N ) External LVDT (050 MHR) Small frame with extra set of springs Jepsen 4pt mouse fixtures Squeeze bottle of 1X Phosphate Buffered Saline (PBS) Glass petri dish Small pair pickups 4x4 gauze Method Equipment Set-Up 1. Set-up small frame with a set of extra springs 2. Mount the load cell and connect to the proper load cell cable 3. Mount the external LVDT and connect to the DTR-451 input cable 4. Make sure the points are aligned properly. See the image below for properly aligned points. a. Set the brass point first. b. While holding one of the lower points, loosen the screw (barely) for that point. If you don t hold the point that you are loosening, it will rotate, which causes alignment issues. Hold the point in position as you tighten it! You want the gap between the two bottom points to be consistent from front to back. c. If you are having issues with positioning, spin the bottom two points around. They are easier to align from one direction. d. Make sure all 4 screws in the image are tight before testing begins. 4pt Bending, Mouse Page 1 of 6 Revision 1

2 MTS System Set-Up 1. Turn on computer 2. Turn on the Teststar box (Switch is located on the back-top right) 3. Turn on DTR-451 (Digital Transducer Readout) 4. Turn on main hydraulics in pump room (turn to low first, then to high) 5. Log-in to computer (user: ORL user; password: trabeculae) 6. Double Click Station Manager Icon 7. Select a configuration file (ex. External_lvdt.cfg) 8. Select a parameter set (ex. ext_lvdt) 9. Click Open 10. Select MPT icon from left task menu 11. Display Signal Auto Offset by selecting the first button under the Station Controls (this field box contains the input signals for the parameter set you selected and allows you to zero any of the inputs by selecting the button in between the current value and offset boxes) 12. Create a new output folder for the day in the directory C:\MTS_Specimen_Data (every user should create their own folder under their name or the study name and then organize data by date) 13. Display the MPT Specimen Editor by clicking the third button from the left on the MPT control panel (it looks like 3 dumbbells) and set the Specimen Directory Path to the folder for the day that you just created in step 12 (ex. C:\MTS_Specimen_Data\User\StudyName\TodaysDate). 14. Close Specimen Editor 15. Reset hydraulic interlocks on the Station control panel. Select the HSM Low button, then the High button. This can also be done on the Remote Station Controller. 16. Select Enable Manual Contol on the Remote Station Controller and find the center of the actuator s traveling distance (+5 mm). Zero the axial displacement on the Signal Auto Offset panel. 4pt Bending, Mouse Page 2 of 6 Revision 1

3 17. Move the MTS crosshead to within your working distance and then tighten down all the bolts. 18. Record the PID values of the system. Go into Station Setup, Channels-Axial. The PID values should be in the first tab. System Warm-Up 1. System gets warmed up at the beginning of each test day to get rid of transient changes in temperature, performance, etc. Warm-up consists of running a cyclic program for an hour. 2. Open the warmuplarge Procedure by selecting the open procedure button, which is the first from the left on the MPT control panel, and select the warmuplarge file 3. Create a new specimen called warmup. To create a new specimen, click on the second button from the left on the MPT control panel (it looks like a single dumbbell). No data will be written to this warmup file, but the system needs a specimen to proceed 4. Do not have any fixtures below the actuator during warmup. 5. If Enable Manual Command is selected, deselect it (On the Remote Station Controller). 6. Select the Play button on the MPT Station Control panel and wait an hour. LVDT Calibration 1. While the system is warming up, is the most time efficient to run the LVDT calibration. 2. Select the LVDT in which you want to use (Jepsen lab uses 050 MHR.) 3. Mount the LVDT rod and cylinder into an axial micrometer stage. 4. Disconnect the Teststar cable from the DTR-451 output and connect it to a volmeter. 5. Make sure the LVDT cable is not connected to the DTR-451 and turn the Zero adjustment on the DTR-451 to reach a 0.00 reading on the voltmeter 6. Reconnect the LVDT cable to the DTR-451 and move the LVDT core with the micrometer stage so that the readout is as close to 0.00 as possible on the voltmeter. 7. Zero the micrometer stage display and select the mm increment button to work in metric units. (This is the null position of the LVDT from which the plus and minus displacements are usually measured) 8. Move the LVDT core from the null position to the LVDT plus full scale position in the positive direction (ex. +1mm) using the micrometer stage adjustment. (You should set this range based on your test design.) 9. Turn the Span adjustment to reach a display reading of approximately +10 Volts on the voltmeter. (This should correspond to approximately 10,000 on the DTR-451 display) 10. You may need to adjust the Phase control next to obtain a maximum display reading of +10 Volts, otherwise skip this step. 11. Displace the LVDT core back to the precise null position with the micrometer stage adjustment. If necessary adjust the Zero control until the voltmeter display reads 0.00 Volts. 12. Displace the LVDT core to the negative full scale position (ex. -1mm) using the micrometer stage adjustment. 13. Turn the Span adjustment to reach a display reading of approximately -10 Volts on the voltmeter. (This should correspond to approximately 10,000 on the DTR-451 display) 14. Repeat Steps 8-13 as necessary until the plus and negative full range positions coordinate as closely as possible to the +\- 10 Volt display on the voltmeter. 15. If the directionality of the core appears reverse it is okay as the directionality is based on which direction you have the rod relative to the cylinder. 4pt Bending, Mouse Page 3 of 6 Revision 1

4 16. Move the micrometer in equal increments over the full range (positive and negative) and record the voltage output on the voltmeter at each increment. Make sure to have at least 10 data points. 17. Use excel to record and graph the data points so that you can calculate the calibration factor in mm/volts (Based on your range you should be able to determine that your calibration factor is close to the ideal calibration factor) For example, if your range was +/-1mm your ideal calibration factor is (2 mm / 20 Volts =.1 mm/volts). 18. Mount the LVDT rod and cylinder into your test set-up. 19. Reconnect DTR-451 output cable to Teststar cable Standard Testing 1. Several small dowels should be tested at the beginning of each day to confirm good data collection and point alignment. 2. Open the SpecialK procedure from the MPT panel. This program ramps down to -3 mm displacement at a rate of 0.05 mm/sec. Data is acquired at a frequency of 25 Hz and is sent to an output file (ex. Specimen.dat) in your directory. Open the Procedure Editor in the MPT panel to confirm these settings haven t changed. 3. Open the Scope window. Ensure that it is displaying load and the external LVDT. 4. Zero the load readout by selecting the Auto Offset button in the Signal Auto Offset window. 5. Move the actuator down (make sure you select Enable Manual Control on the Remote Station Controller) until the upper points overlap the lower points. 6. Create a new specimen in the MPT panel. It is important that you always create a new specimen for each specimen or else data may get combined into one output file. 7. Load the dowel between the points ensuring contact with all 4 points. 8. Slowly move the actuator up/down to get good contact. In rare cases, adjusting the brass point on the top may be necessary. Make sure that the dowel is snug in place, without trying to apply any pre-load. 9. Once the dowel is in position, disable the Manual Control on the Remote Station Controller. 10. Adjust the LVDT height until you are in the proper range. Make sure that you are starting in the correct positive or negative position ensuring that you will stay in range throughout your test. 11. Zero both the axial displacement and external LVDT readings on the Signal Auto Offset window. 12. Select the Play button on the Station Control panel. This begins the data acquisition. 13. A Start button will appear in the bottom of the MPT panel. Pressing this will begin the actuator movement. 14. Repeat for all standards. If anything appears odd in the scope window, make any adjustments as needed. Specimen Testing 1. Open the SpecialK procedure file from the MPT panel 2. Zero the load readout from Signal Auto Offset window. 3. Create a new specimen using the identification numbers of your testing samples. 4. Load your sample a. Place the anterior side of the bone in tension. b. Center the mid-diaphysis of the bone between the upper two points. c. Bring the actuator down slowly so that the bone is in contact with all 4 points. In rare cases, you may need to loosen the brass points at the top. d. The bone should be snug, without having a pre-load on it. 4pt Bending, Mouse Page 4 of 6 Revision 1

5 e. If having difficulties positioning the sample, remove it, place in PBS, and try again. It is critical to keep the bone hydrated, as a drying bone will have different mechanical properties. f. If there is still a lot of soft tissue on the bone, gently remove it with gauze. Rehydrate before loading the bone into the fixture. 5. Check that the external LVDT is still reading in an appropriate range as the dowels can be different diameters from the bone samples. As the LVDT has a ±1mm range, make sure the DTR- 451 box is reading at the minimum, so that as loading commences, it will move towards Place the bone sample back in PBS, and confirm that the LVDT is reading in range. 6. Secure your sample into the test set-up 7. Disable the Manual Control on the Remote Station Controller. 8. Zero both the axial displacement and external LVDT readings on the Signal Auto Offset window. 9. Select the Play button on the Station Control panel. This begins the data acquisition. 10. A Start button will appear in the bottom of the MPT panel. Pressing this will begin the actuator movement. 11. Repeat for all samples 12. Record the position of the break for every sample. If anything outside of the ordinary happens during a test (rotation of a bone sample, slipping, etc) note it next to the test. MPT Procedure Editor Any test can be set-up with the MPT Procedure Editor wizard which is found by selecting the Procedure Editor button on the MPT panel (it is the fourth from the right). The process palette (the button that looks like a painter s palette) associated with this window allows you to drag and drop different types of test commands, data acquisition inputs, events, etc. If you double click on the button associated with the procedure type it will pop up a new window for you to enter the test parameters. Make sure to always include a data limit detector that will interrupt your test command if the limit is reached. Base your limit on the type of load cell you are using. Make sure that if you start with a previously existing procedure and edit it that you save it as a NEW procedure and DO NOT save over the existing procedure unless you are the person who created it. Shutting Down 1. Before turning off the hydraulics, move the actuator up and remove the small testing frame from below the actuator. Then turn off the hydraulics. The actuator drifts after the pressure is turned off, so you want to make sure it doesn t run in to anything. 2. Close the Signal Auto Offset Window and Station Manager. 3. Answer the series of questions as follows: Do you really wish to quit? YES You have changed the station parameters.do you want to save these changes? NO Save the current test parameters? NO 4. Turn off the Teststar box and DTR Log-Off Certification 4pt Bending, Mouse Page 5 of 6 Revision 1

6 I have read and understand the above SOP. I agree to contact my Supervisor or Lab Manager if I plan to modify this procedure. Name Signature Trainer s signature Date Principal Investigator / Safety Officer Date (signature) Printed name: Designated Trainer Date (signature) Printed name: 4pt Bending, Mouse Page 6 of 6 Revision 1

Data Collection Using APEX3. March 30, Chemical Crystallography Laboratory

Data Collection Using APEX3. March 30, Chemical Crystallography Laboratory Data Collection Using APEX3 Page 1 of 10 Data Collection Using APEX3 March 30, 2017 Chemical Crystallography Laboratory Author: Douglas R. Powell Data Collection Using APEX3 Page 2 of 10 Distribution Douglas

More information

AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode)

AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode) AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode) 1. Log into the Log Usage system on the SMIF web site 2. Open Nanoscope 6.14r1 software by double clicking on the Nanoscope 6.14r1 desktop

More information

FEI FIB Focused Ion Beam

FEI FIB Focused Ion Beam Operating Manual Part 1 FEI FIB Focused Ion Beam IF IN DOUBT, ASK 1.00 IF ANYTHING UNUSUAL HAPPENS, OR IF THERE IS ANYTHING YOU ARE UNSURE ABOUT, STOP AND CONTACT ME! DO NOT PROCEED OR ATTEMPT TO FIX THE

More information

TL-2900 AMMONIA & NITRATE ANALYZER DUAL CHANNEL

TL-2900 AMMONIA & NITRATE ANALYZER DUAL CHANNEL TL-2900 AMMONIA & NITRATE ANALYZER DUAL CHANNEL DATA ACQUISITION SYSTEM V.15.4 INSTRUCTION MANUAL Timberline Instruments, LLC 1880 S. Flatiron Ct., Unit I Boulder, Colorado 80301 Ph: (303) 440-8779 Fx:

More information

LUMIGEN INSTRUMENT CENTER X-RAY CRYSTALLOGRAPHIC LABORATORY: WAYNE STATE UNIVERSITY

LUMIGEN INSTRUMENT CENTER X-RAY CRYSTALLOGRAPHIC LABORATORY: WAYNE STATE UNIVERSITY Standard Operating Procedure for the Bruker X8 APEX II Single-Crystal X- Ray Diffractometer Contact Manager: Dr. Cassie Ward ward@wayne.edu Office room 061 Chemistry (313) 577-2587 LIC Lab: (313) 577-0518

More information

Figure 1. MFP-3D software tray

Figure 1. MFP-3D software tray Asylum MFP-3D AFM SOP January 2017 Purpose of this Instrument: To obtain 3D surface topography at sub-nanometer scale resolution, measure contact and friction forces between surfaces in contact, measure

More information

MTL Software. Overview

MTL Software. Overview MTL Software Overview MTL Windows Control software requires a 2350 controller and together - offer a highly integrated solution to the needs of mechanical tensile, compression and fatigue testing. MTL

More information

Topic: Instructional David G. Thomas December 23, 2015

Topic: Instructional David G. Thomas December 23, 2015 Procedure to Setup a 3ɸ Linear Motor This is a guide to configure a 3ɸ linear motor using either analog or digital encoder feedback with an Elmo Gold Line drive. Topic: Instructional David G. Thomas December

More information

E X P E R I M E N T 1

E X P E R I M E N T 1 E X P E R I M E N T 1 Getting to Know Data Studio Produced by the Physics Staff at Collin College Copyright Collin College Physics Department. All Rights Reserved. University Physics, Exp 1: Getting to

More information

Scanning Electron Microscopy (FEI Versa 3D Dual Beam)

Scanning Electron Microscopy (FEI Versa 3D Dual Beam) Scanning Electron Microscopy (FEI Versa 3D Dual Beam) This operating procedure intends to provide guidance for basic measurements on a standard sample with FEI Versa 3D SEM. For more advanced techniques

More information

Checking the fluids of the cart

Checking the fluids of the cart 1 Preparing FACSAria for sterile sorting Prepare and autoclave these fluids at least 2 days in advance: Fluids Remarks Prepare 5L (for 1 period) or 10L (for 2 Each stainless-steel-sheath-tank has a periods)

More information

Series 1100 ColorTS Servo Manual Registration System

Series 1100 ColorTS Servo Manual Registration System Series 1100 ColorTS Servo Manual Registration System 400 Oser Avenue Hauppauge NY 11788 631 434 3700 phone 631 434 3718 fax www.empregister.com December 14, 2005 TABLE OF CONTENTS TABLE OF CONTENTS...

More information

Part No. ENC-LAB01 Users Manual Introduction EncoderLAB

Part No. ENC-LAB01 Users Manual Introduction EncoderLAB PCA Incremental Encoder Laboratory For Testing and Simulating Incremental Encoder signals Part No. ENC-LAB01 Users Manual The Encoder Laboratory combines into the one housing and updates two separate encoder

More information

Manual Supplement. This supplement contains information necessary to ensure the accuracy of the above manual.

Manual Supplement. This supplement contains information necessary to ensure the accuracy of the above manual. Manual Title: 744 Users Supplement Issue: 6 Part Number: 691287 Issue Date: 4/06 Print Date: September 1998 Page Count: 8 Revision/Date: 1, 2/99 This supplement contains information necessary to ensure

More information

NOTICE: This document is for use only at UNSW. No copies can be made of this document without the permission of the authors.

NOTICE: This document is for use only at UNSW. No copies can be made of this document without the permission of the authors. Brüel & Kjær Pulse Primer University of New South Wales School of Mechanical and Manufacturing Engineering September 2005 Prepared by Michael Skeen and Geoff Lucas NOTICE: This document is for use only

More information

Cover Page for Lab Report Group Portion. Boundary Layer Measurements

Cover Page for Lab Report Group Portion. Boundary Layer Measurements Cover Page for Lab Report Group Portion Boundary Layer Measurements Prepared by Professor J. M. Cimbala, Penn State University Latest revision: 23 February 2017 Name 1: Name 2: Name 3: [Name 4: ] Date:

More information

RADIO FREQUENCY SYSTEMS

RADIO FREQUENCY SYSTEMS RADIO FREQUENCY SYSTEMS Optimizer RT FAQ s Q. What information is require before running the software? The Serial Number of each ACU MUST be recorded with the Model number of the antenna that it is attached

More information

Activity P27: Speed of Sound in Air (Sound Sensor)

Activity P27: Speed of Sound in Air (Sound Sensor) Activity P27: Speed of Sound in Air (Sound Sensor) Concept DataStudio ScienceWorkshop (Mac) ScienceWorkshop (Win) Speed of sound P27 Speed of Sound 1.DS (See end of activity) (See end of activity) Equipment

More information

PHI 5000 VersaProbe TM Operator s Guide

PHI 5000 VersaProbe TM Operator s Guide PHI 5000 VersaProbe TM Operator s Guide Part No. 705921 Rev. A Copyright 2006 ULVAC-PHI, INC. 370 Enzo, Chigasaki, JAPAN The PHI logo ( ) is a registered trademark of ULVAC-PHI, INC. Physical Electronics,

More information

Intelligent Pendulum Hardness Tester BEVS 1306 User Manual

Intelligent Pendulum Hardness Tester BEVS 1306 User Manual Intelligent Pendulum Hardness Tester BEVS 1306 User Manual Please read the user manual before operation. PAGE 1 Content 1. Company Profile... 3 2. Product Introduction... 3 3. Operation Instruction...

More information

Transmitter Interface Program

Transmitter Interface Program Transmitter Interface Program Operational Manual Version 3.0.4 1 Overview The transmitter interface software allows you to adjust configuration settings of your Max solid state transmitters. The following

More information

MTI-2100 FOTONIC SENSOR. High resolution, non-contact. measurement of vibration. and displacement

MTI-2100 FOTONIC SENSOR. High resolution, non-contact. measurement of vibration. and displacement A worldwide leader in precision measurement solutions MTI-2100 FOTONIC SENSOR High resolution, non-contact measurement of vibration and displacement MTI-2100 Fotonic TM Sensor Unmatched Resolution and

More information

Vasudevan Agilent 1100 Series HPLC w/ DAD & FLD Detector (nonbuffer

Vasudevan Agilent 1100 Series HPLC w/ DAD & FLD Detector (nonbuffer Vasudevan Agilent 1100 Series HPLC w/ DAD & FLD Detector (nonbuffer solvents) Updated November 14, 2017 Instrument instructions can be found at: http://academic.bowdoin.edu/chemistry/resources/instructions.shtml

More information

SC24 Magnetic Field Cancelling System

SC24 Magnetic Field Cancelling System SPICER CONSULTING SYSTEM SC24 SC24 Magnetic Field Cancelling System Makes the ambient magnetic field OK for the electron microscope Adapts to field changes within 100 µs Touch screen intelligent user interface

More information

Manual placement system MPL3100. for BGA, CSP and Fine-Pitch components

Manual placement system MPL3100. for BGA, CSP and Fine-Pitch components Manual placement system MPL3100 for BGA, CSP and Fine-Pitch components Part No: MPL3100BA1.0e Issue Date: 02/2001 You have opted for an ESSEMTEC MPL3100 pick and place system. We thank you for this decision

More information

Fig. 1. The Front Panel (Graphical User Interface)

Fig. 1. The Front Panel (Graphical User Interface) ME 4710 Motion and Control Data Acquisition Software for Step Excitation Introduction o These notes describe LabVIEW software that can be used for data acquisition. The overall software characteristics

More information

SC24 Magnetic Field Cancelling System

SC24 Magnetic Field Cancelling System SPICER CONSULTING SYSTEM SC24 SC24 Magnetic Field Cancelling System Makes the ambient magnetic field OK for the electron microscope Adapts to field changes within 100 µs Touch screen intelligent user interface

More information

SC26 Magnetic Field Cancelling System

SC26 Magnetic Field Cancelling System SPICER CONSULTING SYSTEM SC26 SC26 Magnetic Field Cancelling System Makes the ambient magnetic field OK for electron beam tools in 300 mm wafer fabs Real time, wideband cancelling from DC to > 9 khz fields

More information

Manual Supplement. This supplement contains information necessary to ensure the accuracy of the above manual.

Manual Supplement. This supplement contains information necessary to ensure the accuracy of the above manual. Manual Title: 9500B Users Supplement Issue: 2 Part Number: 1625019 Issue Date: 9/06 Print Date: October 2005 Page Count: 6 Version 11 This supplement contains information necessary to ensure the accuracy

More information

2.1. Log on to the TUMI system (you cannot proceed further until this is done).

2.1. Log on to the TUMI system (you cannot proceed further until this is done). FEI DB235 ex-situ lift out TEM sample preparation procedure Nicholas G Rudawski ngr@ufledu (805) 252-4916 Last updated: 06/19/15 DISCLAIMER: this procedure describes one specific method for preparing ex-situ

More information

Standard Operating Procedure of nanoir2-s

Standard Operating Procedure of nanoir2-s Standard Operating Procedure of nanoir2-s The Anasys nanoir2 system is the AFM-based nanoscale infrared (IR) spectrometer, which has a patented technique based on photothermal induced resonance (PTIR),

More information

Application Note AN-708 Vibration Measurements with the Vibration Synchronization Module

Application Note AN-708 Vibration Measurements with the Vibration Synchronization Module Application Note AN-708 Vibration Measurements with the Vibration Synchronization Module Introduction The vibration module allows complete analysis of cyclical events using low-speed cameras. This is accomplished

More information

ME EN 363 ELEMENTARY INSTRUMENTATION Lab: Basic Lab Instruments and Data Acquisition

ME EN 363 ELEMENTARY INSTRUMENTATION Lab: Basic Lab Instruments and Data Acquisition ME EN 363 ELEMENTARY INSTRUMENTATION Lab: Basic Lab Instruments and Data Acquisition INTRODUCTION Many sensors produce continuous voltage signals. In this lab, you will learn about some common methods

More information

Guide to the FACS Calibur

Guide to the FACS Calibur Guide to the FACS Calibur Turn on the pump. Check to see if the waste is full, if so replace it with an empty box (label it waste) of FACSflow sheath fluid and replace the sheath with a new box (I will

More information

Operating Instructions

Operating Instructions Operating Instructions HAEFELY TEST AG KIT Measurement Software Version 1.0 KIT / En Date Version Responsable Changes / Reasons February 2015 1.0 Initial version WARNING Introduction i Before operating

More information

PulseCounter Neutron & Gamma Spectrometry Software Manual

PulseCounter Neutron & Gamma Spectrometry Software Manual PulseCounter Neutron & Gamma Spectrometry Software Manual MAXIMUS ENERGY CORPORATION Written by Dr. Max I. Fomitchev-Zamilov Web: maximus.energy TABLE OF CONTENTS 0. GENERAL INFORMATION 1. DEFAULT SCREEN

More information

Standard Operating Procedure II: EDS (Bruker Flat-Quad)

Standard Operating Procedure II: EDS (Bruker Flat-Quad) Standard Operating Procedure II: EDS (Bruker Flat-Quad) ywcmatsci.yale.edu ESC II, Room A119F 810 West Campus Drive West Haven, CT 06516 Version 1.1, October 2018 1 > FOLLOW the SOP strictly to keep the

More information

Three Axis Digital Readout System

Three Axis Digital Readout System NEWALL MEASUREMENT SYSTEMS C80 Three Axis Digital Readout System CONTENTS 2 SPECIFICATIONS 3 CONNECTIONS 4 MOUNTING 4 Arm Mounting (Non-adjustable) 4 Arm Mounting (Adjustable) 5 Face Mounting (Adjustable)

More information

1-Touch Vibratory Sieve Shaker SS-10

1-Touch Vibratory Sieve Shaker SS-10 1-Touch Vibratory Sieve Shaker SS-10 Safety Instructions WARNING!! This machine operates on electric current. Improper operation could result in electrical shock, electrocution, or an explosion! 1. ALWAYS

More information

Approved by Principal Investigator Date: Approved by Super User: Date:

Approved by Principal Investigator Date: Approved by Super User: Date: Approved by Principal Investigator Date: Approved by Super User: Date: Standard Operating Procedure BNC Dektak 3030 Stylus Profilometer Version 2011 May 16 I. Purpose This Standard Operating Procedure

More information

ivw-fd122 Video Wall Controller MODEL: ivw-fd122 Video Wall Controller Supports 2 x 2 Video Wall Array User Manual Page i Rev. 1.

ivw-fd122 Video Wall Controller MODEL: ivw-fd122 Video Wall Controller Supports 2 x 2 Video Wall Array User Manual Page i Rev. 1. MODEL: ivw-fd122 Video Wall Controller Supports 2 x 2 Video Wall Array User Manual Rev. 1.01 Page i Copyright COPYRIGHT NOTICE The information in this document is subject to change without prior notice

More information

Approved by: / / R. Battaglia 12/16/2016

Approved by: / / R. Battaglia 12/16/2016 Fabrication Laboratory Revision: H Rev Date: 12/16/16 Approved by: Process Engineer / / R. Battaglia 12/16/2016 Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Varian

More information

PSC300 Operation Manual

PSC300 Operation Manual PSC300 Operation Manual Version 9.10 General information Prior to any attempt to operate this Columbia PSC 300, operator should read and understand the complete operation of the cubing system. It is very

More information

Mini Micro Pulse Lidar System

Mini Micro Pulse Lidar System Mini Micro Pulse Lidar System MiniMPL-532-C Sensor Suite Operations Manual Version: June 2016 THIS PAGE INTENTIONALLY LEFT BLANK 2 Table of Contents MINIMPL SENSOR SUITE SYSTEM: RECORD OF PURCHASE... 4

More information

Simple and highly effective technology to communicate your brand s distinctive character

Simple and highly effective technology to communicate your brand s distinctive character . . . Advantages 4 Simple and highly effective technology to communicate your brand s distinctive character COST EFFECTIVE No need to print graphics, you can change your message every day! No media player

More information

Fully ly Automaticti. Motorised Satellite t TV System. User s manual REV

Fully ly Automaticti. Motorised Satellite t TV System. User s manual REV REV. 1.0 Fully ly Automaticti Motorised Satellite t TV System User s manual Customer Help Line: 1300 139 255 Support Email: support@satkingpromax.com.au Website: www.satkingpromax.com.au www.satkingpromax.com.au

More information

GEOSATpro GS120. DiSEqC 1.2 Motorized H-H H Motor

GEOSATpro GS120. DiSEqC 1.2 Motorized H-H H Motor DiSEqC 1.2 Motorized H-H H Motor GEOSATpro GS120 Compatible with DiSEqC 1.2 & USALS Receivers Adjustable Hardware Limiters for 140 Degree Coverage Goto X Preprogrammed for North American Satellites LED

More information

TECHNICAL MANUAL DRAW-WIRE DISPLACEMENT TRANSDUCER TYPE DWT

TECHNICAL MANUAL DRAW-WIRE DISPLACEMENT TRANSDUCER TYPE DWT RDP Customer Document TECHNICAL MANUAL DRAW-WIRE DISPLACEMENT TRANSDUCER TYPE DWT Doc. Ref CD1004L BS EN ISO 9001 Certificate No. FM13141 Affirmed by Declaration of Conformity USA & Canada RDP Electrosense

More information

TE 86 MULTI-STATION HIP JOINT SIMULATOR

TE 86 MULTI-STATION HIP JOINT SIMULATOR TE 86 MULTI-STATION HIP JOINT SIMULATOR Introduction The TE 86 Multi-station Hip Joint Simulator is manufactured under licence from Dr Vesa Saikko, Laboratory of Machine Design, Department of Mechanical

More information

HRXRD Analysis of Epitaxial Thin Films

HRXRD Analysis of Epitaxial Thin Films HRXRD Analysis of Epitaxial Thin Films on the Rigaku Smartlab Multipurpose Diffractometer Scott A Speakman, Ph.D Center for Materials Science and Engineering at MIT For assistance in the X-ray lab, contact

More information

INSTALATION PROCEDURE

INSTALATION PROCEDURE INSTALLATION PROCEDURE Overview The most difficult part of an installation is in knowing where to start and the most important part is starting in the proper start. There are a few very important items

More information

ORM0022 EHPC210 Universal Controller Operation Manual Revision 1. EHPC210 Universal Controller. Operation Manual

ORM0022 EHPC210 Universal Controller Operation Manual Revision 1. EHPC210 Universal Controller. Operation Manual ORM0022 EHPC210 Universal Controller Operation Manual Revision 1 EHPC210 Universal Controller Operation Manual Associated Documentation... 4 Electrical Interface... 4 Power Supply... 4 Solenoid Outputs...

More information

CytoFLEX Flow Cytometer Quick Start Guide

CytoFLEX Flow Cytometer Quick Start Guide Sheath Waste CLASS 1 LASER PRODUCT COMPLIES WITH 21 CFR 1040.10 AND 1040.11 EXCEPT FOR DEVIATIONS PURSUANT TO LASER NOTICE NO. 50 DATED JUNE 24, 2007 MANUFACTURED Sheath B49008AC February 2015 CytoFLEX

More information

University of Minnesota Minnesota Nano Center Standard Operating Procedure

University of Minnesota Minnesota Nano Center Standard Operating Procedure Equipment Name: CHA Evaporator Coral Name: ebevap-cha Revision Number: 12 Model: SEC 600 Revisionist: L. von Dissen Location: PAN Bay 3 Date: 04/13/2018 1 Description The CHA Evaporator is a single source

More information

TOMELLERI ENGINEERING MEASURING SYSTEMS. TUBO Version 7.2 Software Manual rev.0

TOMELLERI ENGINEERING MEASURING SYSTEMS. TUBO Version 7.2 Software Manual rev.0 TOMELLERI ENGINEERING MEASURING SYSTEMS TUBO Version 7.2 Software Manual rev.0 Index 1. Overview... 3 2. Basic information... 4 2.1. Main window / Diagnosis... 5 2.2. Settings Window... 6 2.3. Serial transmission

More information

The PHI VersaProbe operates with two essential software programs: PHI Summitt and Vacuum Watcher. A third program, MultiPak, handles data reduction.

The PHI VersaProbe operates with two essential software programs: PHI Summitt and Vacuum Watcher. A third program, MultiPak, handles data reduction. PHI VersaProbe Scanning XPS System I. Overview The PHI VersaProbe operates with two essential software programs: PHI Summitt and Vacuum Watcher. A third program, MultiPak, handles data reduction. PHI Summitt

More information

Operation Procedure for Phillips XL30 ESEM

Operation Procedure for Phillips XL30 ESEM Operation Procedure for Phillips XL30 ESEM The ESEM will be left in the ON state when not in use. The chamber will be at high vacuum, filament on, stage at home position, VAC and HT buttons lit, and monitor

More information

PHY221 Lab 1 Discovering Motion: Introduction to Logger Pro and the Motion Detector; Motion with Constant Velocity

PHY221 Lab 1 Discovering Motion: Introduction to Logger Pro and the Motion Detector; Motion with Constant Velocity PHY221 Lab 1 Discovering Motion: Introduction to Logger Pro and the Motion Detector; Motion with Constant Velocity Print Your Name Print Your Partners' Names Instructions August 31, 2016 Before lab, read

More information

12months. on-site warranty. DZE ELECTRONIC PRESSURE SWITCH for detection of overload per EN 81 2 featuring two adjustable switching points

12months. on-site warranty. DZE ELECTRONIC PRESSURE SWITCH for detection of overload per EN 81 2 featuring two adjustable switching points BUCHER PRODUCTS AVAILABLE FROM HYDRATEC DZE ELECTRONIC PRESSURE SWITCH for detection of overload per EN 81 2 featuring two adjustable switching points 12months on-site warranty All our work comes with

More information

BROTHER, BROTHER & SONS

BROTHER, BROTHER & SONS AREA SOFT BROTHER, BROTHER & SONS ApS INTRO Brother, Brother & Sons ApS The information in this manual is subject to change without notice. Go to www.brothers-sons.dk for the latest version. Brother, Brother

More information

D-901 PC SOFTWARE Version 3

D-901 PC SOFTWARE Version 3 INSTRUCTION MANUAL D-901 PC SOFTWARE Version 3 Please follow the instructions in this manual to obtain the optimum results from this unit. We also recommend that you keep this manual handy for future reference.

More information

INSTRUCTION AND MAINTENANCE VIDEO MOTORE PROJECTION SCREEN

INSTRUCTION AND MAINTENANCE VIDEO MOTORE PROJECTION SCREEN INSTRUCTION AND MAINTENANCE VIDEO MOTORE PROJECTION SCREEN 1.1 TECHNICAL DATA PRODUCT DESCRIPTION Screens made with the following projection fabrics: SOFT WHITE, MATT WHITE SOFT, SOFT WHITE TRANSOUND,

More information

PogoStick and Research Pogo App Operator s Guide

PogoStick and Research Pogo App Operator s Guide PogoStick and Research Pogo App Operator s Guide Research POGO App Overview Getting Started with the Research POGO App When you first start the Research POGO App, you will be asked to log into your Precision

More information

180w led gobo moving USER MANUAL

180w led gobo moving USER MANUAL 180w led gobo moving USER MANUAL 1. BEFORE YOU BEGIN What is included 1 x Light 1 x Mounting bracket and screws 1 x Power cable 1 x User Manual Unpacking Instructions Congratulations on purchasing ART-TECH

More information

ASE 369 K Measurements and Instrumentation. LAB #9: Impulse-Force Hammer; Vibration of Beams

ASE 369 K Measurements and Instrumentation. LAB #9: Impulse-Force Hammer; Vibration of Beams ASE 369 K Measurements and Instrumentation LAB #9: Impulse-Force Hammer; Vibration of Beams Equipment: Dell Optiplex computer with National Instruments PCI-MIO-16E-4 data-acquisition board and the Virtual

More information

Casambi App User Guide

Casambi App User Guide Casambi App User Guide Version 1.5.4 2.1.2017 Casambi Technologies Oy Table of contents 1 of 28 Table of contents 1 Smart & Connected 2 Using the Casambi App 3 First time use 3 Taking luminaires into use:

More information

Stage Wash 7x 10W LED Moving Head (RGBW)

Stage Wash 7x 10W LED Moving Head (RGBW) Stage Wash 7x 10W LED Moving Head (RGBW) P/N 612870 User's Manual CONTENTS SAFETY WARNINGS AND GUIDELINES... 3 INTRODUCTION... 4 FEATURES... 5 CUSTOMER SERVICE... 5 PACKAGE CONTENTS... 6 PRODUCT OVERVIEW...

More information

Import and quantification of a micro titer plate image

Import and quantification of a micro titer plate image BioNumerics Tutorial: Import and quantification of a micro titer plate image 1 Aims BioNumerics can import character type data from TIFF images. This happens by quantification of the color intensity and/or

More information

Introduction to the FTA2000

Introduction to the FTA2000 Introduction to the FTA2000 This document will help you get started with your new FTA2000. The FTA2000 is a highly automated instrument, which may seem complex at first, but each of its individual functions

More information

Digital Video Recorder

Digital Video Recorder Digital Video Recorder Quick Operation Guide UD.6L0202B0067A02 Thank you for purchasing our product. If there is any question or request, please do not hesitate to contact dealer. This manual is applicable

More information

Working with CSWin32 Software

Working with CSWin32 Software Working with CSWin32 Software CSWin32 provides a PC interface for Coiltek s ultrasonic control products. The software expands the palette of control features of the CS-5000 and CS-6100 series controls;

More information

Operating Manual. Automated Gear. Apollo Design Technology, Inc Fourier Drive Fort Wayne, IN USA

Operating Manual. Automated Gear. Apollo Design Technology, Inc Fourier Drive Fort Wayne, IN USA Operating Manual Automated Gear Apollo Design Technology, Inc. 4130 Fourier Drive Fort Wayne, IN 46818 USA PH: +01(260)497-9191 FX: +01(260)497-9192 www.apollodesign.net 11-25-09 5-6 POWERING UP THE RIGHT

More information

Unicable II Programmer. IDLU-PROG01-OOOOO-OPP Item: Installation & User Guide

Unicable II Programmer. IDLU-PROG01-OOOOO-OPP Item: Installation & User Guide Unicable II Programmer IDLU-PROG01-OOOOO-OPP Item: 5273 Installation & User Guide 1 2 Thank you for purchasing Inverto s advanced Unicable II programmer. Before installing and using the programmer, please

More information

USER MANUAL FOR THE ANALOGIC GAUGE FIRMWARE VERSION 1.1

USER MANUAL FOR THE ANALOGIC GAUGE FIRMWARE VERSION 1.1 by USER MANUAL FOR THE ANALOGIC GAUGE FIRMWARE VERSION 1.1 www.aeroforcetech.com Made in the USA! WARNING Vehicle operator should focus primary attention to the road while using the Interceptor. The information

More information

Owner's Manual. TOUCH SCREEN CONTROLLER for Air Conditioning Control System. Model BMS-CT5120UL. English

Owner's Manual. TOUCH SCREEN CONTROLLER for Air Conditioning Control System. Model BMS-CT5120UL. English TOUCH SCREEN CONTROLLER for Air Conditioning Control System Model BMS-CT5120UL English Contents 1 Precautions for safety.................................................. 5 2 Main functions........................................................

More information

INSTALLATION INSTRUCTIONS RGB / RGBW 24V LED TAPE

INSTALLATION INSTRUCTIONS RGB / RGBW 24V LED TAPE LLI-LCC4.4W LLI-LCCW4.4W LLI-LCCW5.8W **DANGER: Prior to installation, disconnect power at the source.** Make sure you have all components listed below: LED tape 24VDC Power Supply Connectors (if creating

More information

ST350 Radiation Counter

ST350 Radiation Counter SPECTECH ST350 Radiation Counter (For Macintosh) Operating and Service Manual March 2003 Spectrum Techniques Page 1 Model ST350 RADIATION COUNTER The ST350 Radiation Counter brings new dimensions to Nuclear

More information

DektakXT Profilometer. Standard Operating Procedure

DektakXT Profilometer. Standard Operating Procedure DektakXT Profilometer Standard Operating Procedure 1. System startup and sample loading: a. Ensure system is powered on by looking at the controller to the left of the computer.(it is an online software,

More information

DS-7200HVI/HFI-SH Series DVR Quick Operation Guide

DS-7200HVI/HFI-SH Series DVR Quick Operation Guide DS-7200HVI/HFI-SH Series DVR Quick Operation Guide UD.6L0202B0019A01 Thank you for purchasing our product. If there is any question or request, please do not hesitate to contact dealer. This manual is

More information

EVD-L04/100A1-960, EVD-L08/200A1-960 and. EVD-L16/400A1-960 DVRs. Quick Operation Guide

EVD-L04/100A1-960, EVD-L08/200A1-960 and. EVD-L16/400A1-960 DVRs. Quick Operation Guide EVD-L04/100A1-960, EVD-L08/200A1-960 and EVD-L16/400A1-960 DVRs Quick Operation Guide Thank you for purchasing our product. If there is any question or request, please do not hesitate to contact dealer.

More information

VISSIM TUTORIALS This document includes tutorials that provide help in using VISSIM to accomplish the six tasks listed in the table below.

VISSIM TUTORIALS This document includes tutorials that provide help in using VISSIM to accomplish the six tasks listed in the table below. VISSIM TUTORIALS This document includes tutorials that provide help in using VISSIM to accomplish the six tasks listed in the table below. Number Title Page Number 1 Adding actuated signal control to an

More information

Revision 1.2d

Revision 1.2d Specifications subject to change without notice 0 of 16 Universal Encoder Checker Universal Encoder Checker...1 Description...2 Components...2 Encoder Checker and Adapter Connections...2 Warning: High

More information

Transfer Switch. OTECA (Spec A) OTECB (Spec A) OTECC (Spec A) OTECD (Spec A) Amperes. English Original Instructions (Issue 5)

Transfer Switch. OTECA (Spec A) OTECB (Spec A) OTECC (Spec A) OTECD (Spec A) Amperes. English Original Instructions (Issue 5) Operator Manual Transfer Switch 40-1000 Amperes OTECA (Spec A) OTECB (Spec A) OTECC (Spec A) OTECD (Spec A) English Original Instructions 10-2015 962-0131 (Issue 5) Table of Contents 1. SAFETY PRECAUTIONS...

More information

Noise Detector ND-1 Operating Manual

Noise Detector ND-1 Operating Manual Noise Detector ND-1 Operating Manual SPECTRADYNAMICS, INC 1849 Cherry St. Unit 2 Louisville, CO 80027 Phone: (303) 665-1852 Fax: (303) 604-6088 Table of Contents ND-1 Description...... 3 Safety and Preparation

More information

Figure 1: Device components

Figure 1: Device components Order No. : 2860 10 Order No. : 2830 10 Operation- and Assembly Instructions 1 Safety instructions Electrical equipment may only be installed and fitted by electrically skilled persons. Failure to observe

More information

USER MANUAL FOR THE ANALOGIC GAUGE FIRMWARE VERSION 1.0

USER MANUAL FOR THE ANALOGIC GAUGE FIRMWARE VERSION 1.0 by USER MANUAL FOR THE ANALOGIC GAUGE FIRMWARE VERSION 1.0 www.aeroforcetech.com Made in the USA! WARNING Vehicle operator should focus primary attention to the road while using the Interceptor. The information

More information

Operating Instructions

Operating Instructions CNTX Contrast sensor Operating Instructions CAUTIONS AND WARNINGS SET-UP DISTANCE ADJUSTMENT: As a general rule, the sensor should be fixed at a 15 to 20 angle from directly perpendicular to the target

More information

Vibratory Deck Sieves 15 in. (380 mm)

Vibratory Deck Sieves 15 in. (380 mm) Instruction Sheet P/N 1604433-01 Vibratory Deck Sieves 15 in. (380 mm) Introduction This instruction sheet covers the vibratory deck sieves listed in the following tables. 15-Inch Deck Sieves with 2.5

More information

Quick Operation Guide of LTN7700/7600 Series NVR

Quick Operation Guide of LTN7700/7600 Series NVR Quick Operation Guide of LTN7700/7600 Series NVR UD.6L0202B0042A02 Thank you for purchasing our product. If there is any question or request, please do not hesitate to contact dealer. This manual is applicable

More information

EDL8 Race Dash Manual Engine Management Systems

EDL8 Race Dash Manual Engine Management Systems Engine Management Systems EDL8 Race Dash Manual Engine Management Systems Page 1 EDL8 Race Dash Page 2 EMS Computers Pty Ltd Unit 9 / 171 Power St Glendenning NSW, 2761 Australia Phone.: +612 9675 1414

More information

Troubleshooting the CTS 1100

Troubleshooting the CTS 1100 CHAPTER 7 Troubleshooting the CTS 1100 Revised: November 2011, Contents You may want to periodically test system components using the hardware and software tests available in the Cisco TelePresence System

More information

RK-2 ENVIRONMENTAL DATA CONTACTLESS MAGNETOSTRICTIVE LINEAR POSITION TRANSDUCER WITH FLANGED HEAD. Main characteristics

RK-2 ENVIRONMENTAL DATA CONTACTLESS MAGNETOSTRICTIVE LINEAR POSITION TRANSDUCER WITH FLANGED HEAD. Main characteristics RK-2 CONTACTLESS MAGNETOSTRICTIVE LINEAR POSITION TRANSDUCER WITH FLANGED HEAD (ANALOG OR START/STOP OUTPUT) Main characteristics Absolute transducer Strokes from 50 to 4000mm (RK-2- -N/E/S) Digital output

More information

Installation and Setting up Instructions for the 990 Signal Conditioning Instrument

Installation and Setting up Instructions for the 990 Signal Conditioning Instrument Installation and Setting up Instructions for the 990 Signal Conditioning Instrument Contents Page 1.0 Overview... 3 2.0 Installation.. 4 2.1 Electrical connections... 4 2.2 Cable selection... 4 2.3 Electrical

More information

Table of Contents. Versa TILE & Versa DRIVE D2 Quick Start Manual

Table of Contents. Versa TILE & Versa DRIVE D2 Quick Start Manual Versa TILE & Versa DRIVE D2 Table of Contents 1. Introduction 2 2. Versa TILE Assembly & Cabling 2 3. Computer & Versa DRIVE Connections 2 4. RasterMAPPER 3 5. Buffer Board 4 6. Testing the System 5 7.

More information

Overview. Signal Averaged ECG

Overview. Signal Averaged ECG Updated 06.09.11 : Signal Averaged ECG Overview Signal Averaged ECG The Biopac Student Lab System can be used to amplify and enhance the ECG signal using a clinical diagnosis tool referred to as the Signal

More information

800 Displaying Series Flowmeter

800 Displaying Series Flowmeter TECHNICAL PRODUCT INSTRUCTION SHEET 800 Displaying Series Flowmeter OVERVIEW The principle of operation is very simple. A jet of liquid is directed at a free running Pelton wheel turbine in a specially

More information

USER MANUEL. SNIPE 2 Ref R13

USER MANUEL. SNIPE 2 Ref R13 USER MANUEL SNIPE 2 Ref. 0141317R13 Contents 1. General Information 1-1. Introduction 1-2. Proper use and operation 1-3. Safety notes......... 2 3 3 2. Contents 2-1. Accessory included 2-2. Name of parts......

More information

Technology Control Technology

Technology Control Technology L e a v i n g C e r t i f i c a t e Technology Control Technology P I C A X E 1 8 X Prog. 1.SOUND Output Prog. 3 OUTPUT & WAIT Prog. 6 LOOP Prog. 7...Seven Segment Display Prog. 8...Single Traffic Light

More information

HD-CM HORIZON DIGITAL CABLE METER

HD-CM HORIZON DIGITAL CABLE METER HD-CM OFF! Max RF i/p = +17dBm 75Ω Max AC/DC i/p = 120Vrms MENU INPUT ON HORIZON DIGITAL CABLE METER Horizon Global Electronics Ltd. Unit 3, West Side Flex Meadow Harlow, Essex CM19 5SR Phone: +44(0) 1279

More information

ENGINEERING COMMITTEE Interface Practices Subcommittee AMERICAN NATIONAL STANDARD ANSI/SCTE

ENGINEERING COMMITTEE Interface Practices Subcommittee AMERICAN NATIONAL STANDARD ANSI/SCTE ENGINEERING COMMITTEE Interface Practices Subcommittee AMERICAN NATIONAL STANDARD ANSI/SCTE 102 2010 Cable Retention Force Testing of Trunk & Distribution Connectors NOTICE The Society of Cable Telecommunications

More information