Guidance for Quality Assurance of PACS Diagnostic Display Devices

Size: px
Start display at page:

Download "Guidance for Quality Assurance of PACS Diagnostic Display Devices"

Transcription

1 Guidance for Quality Assurance of PACS Diagnostic Display Programme NPFIT DOCUMENT RECORD ID KEY Sub-Prog / PACS Project NPFIT-PAC-DES Prog. Director Max Jones Owner Jerry Norman Version 1.0 Author Peter Simpson Version Date 03/08/06 Status Approved Guidance for Quality Assurance of PACS Diagnostic Display Crown Copyright 2006 Page 1 of 15

2 Amendment History: Version Date Amendment History /08/05 First draft for comment /09/05 Authority and PACSnet comments. This version for approval /10/05 Redrafting of Purpose section to indicate how this document is expected to be utilised /03/06 Feedback from PACSnet and RCR /05/06 Update for pixel defects Forecast Changes: Anticipated Change None. When Reviewers: This document must be reviewed by the following. Indicate any delegation for sign off. Name Signature Title / Responsibility Date Version NHS CFH PACS Central Technical Team PACS National Clinical Reference Panel (NCRP) NHS CFH Cluster PACS CTA PACS National Implementatio n Reference Group (NIRG) Approvals: This document requires the following approvals: Name Signature Title / Responsibility Date Version Jerry Norman PACS Technical Team Lead Max Jones PACS Programme Crown Copyright 2006 Page 2 of 15

3 PACS National Clinical Reference Panel Director Distribution: As Reviewers and Approvals above plus:- PACS Cluster CTA s LSP PACS Crown Copyright 2006 Page 3 of 15

4 Document Status: This is a controlled document. This document version is only valid at the time it is retrieved from controlled filestore, after which a new approved version will replace it. On receipt of a new issue, please destroy all previous issues (unless a specified earlier issue is baselined for use throughout the programme). Related Documents: These documents will provide additional information. Ref no Doc Reference Number Title Version 1 NPFIT-SHR-QMS-PRP-0015 Glossary of Terms Consolidated.doc 11 See References, Section 5. Glossary of Terms: List any new terms created in this document. Mail the NPO Quality Manager to have these included in the master glossary above [1]. Term Acronym Definition Crown Copyright 2006 Page 4 of 15

5 Contents 1. Purpose Scope Introduction Working Environment Rooms Staffing requirements for display QA Room illumination and arrangement of displays Displays General display condition Image quality checks Contrast ratio Geometry Distance measurement Angle measurement Spatial resolution Contrast resolution Luminance uniformity Veiling Glare Reflectivity Maximum luminance DICOM GSDF conformance Noise Crosstalk Focus References ACR-NEMA DICOM Standard (PS 3) DIN V ISO 9241 and VESA Flat Panel Display Measurements (FPDM) Standard The American Association of Physicists in Medicine (AAPM) The Institute of Physics in Engineering (IPEM) The Royal College of Radiologists (RCR) Appendix A QA Specification and Testing Appendix B Additional Tests Crown Copyright 2006 Page 5 of 15

6 1. Purpose The purpose of this document is to provide outline guidance to NHS Connecting for Health (CfH) Cluster PACS Programmes (and their associated clinical reference groups) to enable the development and implementation of locally defined procedures for the Quality Assurance (QA) of PACS display devices. This document is not intended to be a statement of requirements for the LSP PACS to comply with. Each LSP PACS is responsible for delivering appropriate quality control measures as detailed in the Authority s Requirements, for example, requirement detailed below The Contractor shall ensure that all the following display devices shall have a facility for regular quality control and shall be able to be calibrated to DICOM grey scale display function: mammography reporting; orthopaedic reporting; cardiology reporting; radiology reporting; mammography review; orthopaedic review; cardiology review; and Accident & Emergency review Scope Displays Displays devices may be classified as Primary or Secondary. (They may also be referred to by the alternative names Diagnostic or Review respectively). Primary display devices are used for the clinical interpretation of medical images. Secondary display devices are used for review of medical images, usually in conjunction with the report. This document deals with the QA of primary diagnostic display devices used for clinical image interpretation. Where images are reviewed without a requirement for clinical interpretation, the image quality is considered to be of secondary importance. The quality of display in these circumstances should be considered locally, depending on the purpose of the review. 2. Introduction The QA of displays is essential to maintain both the performance of the display and the consistent presentation of images throughout an imaging department and beyond. Crown Copyright 2006 Page 6 of 15

7 Manufacturers of many primary class medical displays supply tools to support QA, principally for luminance and greyscale calibration. However there is no recommendation as to the frequency of these checks, only that they should be performed regularly. Many displays include luminance probes built in to the displays which continuously monitor and maintain the display s luminance and can be used to predict the lifetime of the backlight. Other performance parameters are less easily automated and rely on a degree of user intervention or on examination by appropriately qualified personnel. These tests are detailed below. The first section deals with the general working environment in which the displays are used; subsequent sections are concerned with the various checks that are to be performed on the displays themselves. Finally, there is a list of standards and other documentation that may be of assistance in running a QA programme. Recommended QA specifications and testing frequencies are listed in the appendix at the end of this document. 3. Working Environment 3.1. Rooms The lighting conditions of the areas used for the displays, and the locations of the displays within these areas, should be inspected Staffing requirements for display QA General daily QA checks that require viewing of the screen without use of specialist equipment are best done by those that routinely use the displays, i.e. radiographers, radiologists and other clinicians. The more in-depth checks that do require use of specialist equipment are best done by technical support staff, who should also be responsible for maintaining records of the QA checks Room illumination and arrangement of displays Check general room lighting conditions and arrangement of displays, including window blinds, room lighting, positioning of displays, adjustability of display positioning, desk height, chairs. Room illumination should be checked every 6 months Window Blinds Window blinds should allow for the complete blocking of external illumination (primarily daylight) Room Lighting There should be no direct lighting falling on the displays during their normal use. It may be necessary to have extra lighting available when the room is in use for purposes other than use of the displays (e.g. cleaning, meetings). Dimmer switches should be used to enable the overhead lighting to be lowered to an appropriate level. Supplementary task lighting, e.g. desk lamp, should be considered to allow reading of associated papers (e.g. request forms), though care must be taken with their positioning and luminance. Room lighting levels can be measured using an illuminance meter. Crown Copyright 2006 Page 7 of 15

8 Display Positioning Displays should be positioned such that any reflections are minimised, i.e. they should be side-on to any sources of illumination such as doors and should not be positioned directly opposite each other. Adjustability of display positioning, desk height and chairs: It should be possible for users to adjust the equipment so that it can be used comfortably and safely. 4. Displays Appendix A details a QA testing list together with the suggested testing frequency for diagnostic and review displays. These tests are further described in this section. The Appendix also notes whether tests should be performed as part of acceptance testing, and suggests action limits for the results of the tests General display condition Check the cleanliness and general condition of the display, keyboard and any associated wiring and base computer. Note: cleaning materials used for displays 4.2. Image quality checks Checks on image quality can be divided into several areas, listed below. Note that the listed equipment required is in addition to appropriate, contractor supplied, test images, which will be required for all tests. Note: In the interest of consistency, a standard set of images (e.g. SMPTE, AAPM, etc.) should be agreed across all Clusters Contrast ratio This is a measurement of the difference in luminance between the brightest areas of a display screen and the darkest areas. Equipment required: luminance meter Geometry A test to check that the geometry of the display is acceptable, i.e. to check that the straight lines are straight, any circles are circular and any square are square. Equipment required: ruler 4.5. Distance measurement A test to check that any distance measurements read from the screen are correct. This may also involve checking that measurements displayed by workstation software are correct Angle measurement A test to check that any distance measurements read from the screen are correct. This may also involve checking that measurements displayed by workstation software are correct. Crown Copyright 2006 Page 8 of 15

9 4.7. Spatial resolution Spatial resolution is a measure of the capability of the display to present information on two closely-spaced objects in the image such that the two objects can be discriminated. Spatial resolution is usually expressed as the number of black and white lines within one millimetre (lp/mm) that can be distinguished Contrast resolution Contrast resolution is a measure of the ability of the display to present areas of different image intensity such that there is a difference in luminance when displayed (i.e. it is the resolution of different grey levels). Equipment required: luminance meter 4.9. Luminance uniformity Displays should demonstrate equal luminance across the display screen and between all display screens attached to the same workstation. Equipment required: luminance meter Veiling Glare Veiling glare is a measure of how much a dark area in the image is brightened by surrounding bright areas. The effect of veiling glare is to reduce the contrast. In practice, this test is only meaningful for CRT displays. Equipment required: Luminance meter, screen mask (i.e. a mask used to cover the display screen apart from the area used to make the measurements required for this test. Alternatively, a funnel or cone may be used) Reflectivity Display screens should be manufactured so that there is a minimum of reflection from the screen. Reflections can prove distracting to the observer and can also serve to reduce the contrast resolution. Equipment required: luminance meter Maximum luminance It is important to measure the maximum luminance of a display regularly. A decline in maximum luminance leads to a commensurate decline in contrast resolution. Maximum luminance measurements can also act as a useful predictor for the lifetime of the backlight in liquid crystal displays. Equipment required: luminance meter DICOM GSDF conformance Part 14 of the DICOM standard describes a Grayscale Standard Display Function, which aims to ensure that images displayed on different devices have a similar appearance, provided that the devices are calibrated to the DICOM GSDF. Calibration to the GSDF should be checked regularly. Equipment required: luminance meter. Crown Copyright 2006 Page 9 of 15

10 4.14. Noise Noise in a displayed image is the random variation of luminance over the image. Noise in an image can potentially obscure diagnostic information and can also increase the fatigue of the user of the display. A visual impression of the noise in an image should be made Crosstalk Crosstalk is the unwanted leakage of electronic signal from one channel to an adjacent channel. In liquid crystal display systems this can lead to the intended luminance of a pixel (i.e. the intended voltage across the liquid crystal cell that forms the pixel) being modified away from the intended value. A visual impression of the presence of crosstalk in an image should be made. Note: Use of SMPTE test image is acceptable, as is the use of TG18-QC or indeed any image with appropriate bar patterns Focus A visual impression of the focus of CRT displays should be made and adjustments made when necessary. Crown Copyright 2006 Page 10 of 15

11 5. References The following references have informed the guidance contained in this document ACR-NEMA DICOM Standard (PS 3) In 1984, the American College of Radiology (ACR) and the National Electrical Manufacturers Association (NEMA) formed a committee that produced and currently maintains the Digital Imaging and Communications in Medicine (DICOM) standard. The committee produced a document, Grayscale Standard Display Function (NEMA 2000), which specified a standardized display function known as the Grayscale Standard Display Function (GSDF) for grayscale images. The intent of the standard was to allow images transferred using the DICOM standard to be displayed on any DICOMcompatible display device with a consistent grayscale appearance DIN V The German standards institution, Deutsches Institut für Normung e.v. (DIN), standard 6868 part 57, Image Quality Assurance in X-Ray Diagnostics, Acceptance Testing for Image Display (DIN 2001) was developed as an acceptance testing standard addressing the requirements for display systems. The standard recognizes the use of DICOM GSDF ISO 9241 and The International Standards Organization (ISO) standard, ISO :1992, Ergonomic Requirements for Office Work with Visual Display Terminals (VDTs) Part 3: Visual Display Requirements (ISO 1992), gave test methods and performance requirements primarily for displays used for text based terminals. ISO :2001, Ergonomic Requirements for Work with Visual Displays Based on Flat Panels. Part 2: Ergonomic Requirements for Flat Panel Displays (ISO 2001) VESA Flat Panel Display Measurements (FPDM) Standard This standard was introduced to specify measurement methods and procedures for both CRT and flat panel display devices. However this does not recommend performance or compliance criteria for display applications The American Association of Physicists in Medicine (AAPM) The American Association of Physicists in Medicine (AAPM) has produced probably the most thorough document on the assessment of medical displays, Assessment of display performance for medical imaging systems. This describes methods and tools for the assessment of displays and suggests regimes for acceptance testing and for daily, monthly/quarterly and annual quality control tests. Crown Copyright 2006 Page 11 of 15

12 5.6. The Institute of Physics in Engineering (IPEM) In the UK the Institute of Physics in Engineering (IPEM) has recently published Report 91 Recommended Standards for the Routine Performance Testing of Diagnostic X-Ray Systems. Report 91 replaces IPEM report 77 and provides essential guidance for anyone responsible for diagnostic X-Ray equipment. This includes a section on quality assurance of diagnostic displays The Royal College of Radiologists (RCR) The UK Royal College of Radiologists (RCR) PACS & Teleradiology special interest group has produced guidelines on the specification and quality assurance of primary diagnostic display devices. This guidance includes a minimum and recommended screen display size and resolution for diagnostic image interpretation, in conjunction with best practice guidance on using diagnostic workstations. Crown Copyright 2006 Page 12 of 15

13 6. Appendix A QA Specification and Testing Table 1 Summary of Image Display Device Guidance from IPEM 2005 Recommended Standards for Routine Performance Testing of Diagnostic X-ray Imaging Systems, IPEM Report 91 Physical parameter Frequency Remedial level Image display monitor condition Greyscale contrast luminance ratio Distance and Angle Calibration Daily to weekly Image display devices should be clean, and the perceived contrast of the test pattern should be consistent between display devices connected to the same workstation. Verify that the 5% and 95% patches are visible. 3 monthly Ratio white to black (maximum to minimum luminance) < monthly ± 5 mm ± 3 Spatial Resolution 3 monthly Grade AAPM TG18-QC resolution patterns according to the reference score (CX > 4) Greyscale drift 6 to 12 monthly Black baseline ± 25% White baseline ± 20% DICOM greyscale calibration Luminance Uniformity Variation between monitors 6 to 12 monthly GSDF ± 10% 6 to 12 monthly U% > 30% 6 to 12 monthly Black baseline > 30% White baseline > 30% Room illumination 6 to 12 monthly > 15 lux Crown Copyright 2006 Page 13 of 15

14 Table 2 Summary of the RCR PACS & Teleradiology special interest group minimum and recommended specification and QA acceptance testing for primary diagnostic display devices used for clinical image interpretation. This guidance applies to all workstations where CR, DR, fluoroscopy, ultrasound, CT, MR, nuclear medicine and PET images are viewed (excluding mammography). Minimum [1] Recommended [1] Screen Resolution [2] (Native Pixel Array) Screen Size (Viewable Diagonal) [5] Maximum Luminance >= 1280 x 1024 [3] (~ 1.3 megapixels) >= 1500 x 2000 [4] (~ 3 megapixels) >= 42 cm (~17 ) >= 50 cm (~ 20 ) > 170 cd/m2 [6] >= 500 cd/m2 [7] Luminance Contrast Ratio >= 250:1 [6] [8] >= 500:1 (Maximum/Minimum) Grey-scale Calibration Within 10% GSDF [8] Calibrated to GSDF [5] >= 10 bit grey-scale Grey-scale bit depth 8 bit grey-scale (24 bit colour) [9] Video Display Interface Digital-analogue Digital video interface (DVI) Pixel Defects ISO classification [10] Class 2 (2 per million) Class 1 (0 defects) Notes [1] The minimum and recommended specifications for diagnostic display devices are only appropriate if clinical image viewing is performed according to image viewing guidelines. All diagnostic image interpretation should be performed on DICOM images making use of the application software zoom, pan, magnification, and windowing tools to optimise spatial and contrast resolution. [2] LCD devices should be run at their native resolution to ensure there is a 1:1 match between screen pixels and screen resolution, and therefore no loss of image quality due to screen interpolation. CRT displays can be run at a variety of resolutions with no loss of display quality; however care should be taken that the correct aspect ratio is maintained to avoid distortion of the image. [3] Where the majority of reporting performed on a diagnostic workstation is of cross-sectional imaging, lower resolution landscape style displays (>= 1.3 megapixels) are considered adequate, providing larger images are interpreted with the aid of systematic magnification. [4] High fidelity (>= 3 megapixels) portrait style displays are recommended in radiology and other areas where large numbers of plain radiographic images are reported to reduce requirements for systematic magnification, and thus reduce image interpretation and reporting times. [5] Display devices may be set initially to operate at a fraction of the maximum luminance in the manufacturer s specification. This can be adjusted to compensate for the decline in performance of the back-light over time whist maintaining grey-scale calibration. [6] AAPM TG18 recommendation [7] High luminance displays can increase the number of perceivable grey-scale levels (JND index steps) but may have a detrimental effect in user performance through fatigue and the Crown Copyright 2006 Page 14 of 15

15 human visual adaptation response. The optimum operating luminance level may vary between users. [8] IPEM 91 recommendation [9] 24-bit and 32-bit colour are equivalent to 8-bit monochrome grey-scale. Colour display devices are recommended for displaying colour images, but they generally perform less well than monochrome display devices in terms of maximum luminance and contrast ratio. [10] The number of permissible pixel defects per million is defined by the ISO standard. Class 1 panels should have no defects. Class 2 panels should be replaced if they have > 2 whole pixel defects per million. Appropriate use of application software zoom, pan and magnification tools can negate the effect of pixel defects in clinical practice. 7. Appendix B Additional Tests In addition to the tests listed in Appendix A, the following tests are recommended by PACSnet as being important parts of a QA programme. Quarterly tests: - Room inspection Check that window blinds, dimmer switches and lighting are in place and working appropriately. - Noise Check that the noise level in the displayed image is acceptable by using the test image and test method supplied in the AAPM Assessment of Display Performance for Medical Imaging Systems. (ref:1.25 AAPM) - Geometry of displayed image Check that straight lines in a test image are displayed correctly (ref: 1.25, AAPM) - Crosstalk and other video image artefacts Use AAPM TG18-QC, SMPTE or other similar test image to check that levels of crosstalk and other artefacts are acceptable (ref: 1.25, AAPM) Acceptance testing: - Veiling glare A new CRT display (or a single display from a set bought at the same time) should be tested for veiling glare (ref: 1.25, AAPM) - Reflectivity A new display (or a single display from a set bought at the same time) should be tested for reflectivity (ref: 1.25, AAPM) - Variation of luminance and contrast ratio with viewing angle A new display (or a single display from a set bought at the same time) should be tested for variation of luminance and contrast ratio with viewing angle (ref: 1.25, AAPM) Crown Copyright 2006 Page 15 of 15

Guidelines for Assuring Softcopy Image Quality

Guidelines for Assuring Softcopy Image Quality Guidelines for Assuring Softcopy Image Quality What s inside? Quality Control Guidelines Softcopy QA testing and frequencies Danny Deroo Product and R&D Manager QA Products ABSTRACT To ensure diagnostic

More information

Display Quality Assurance: Considerations When Establishing a Display QA Program. Mike Silosky, M.S. 8/3/2017

Display Quality Assurance: Considerations When Establishing a Display QA Program. Mike Silosky, M.S. 8/3/2017 Display Quality Assurance: Considerations When Establishing a Display QA Program Mike Silosky, M.S. 8/3/2017 Objectives and Outline Why, Who, What, When, Where? Discuss the resources that may be needed

More information

3/2/2016. Medical Display Performance and Evaluation. Objectives. Outline

3/2/2016. Medical Display Performance and Evaluation. Objectives. Outline Medical Display Performance and Evaluation Mike Silosky, MS University of Colorado, School of Medicine Dept. of Radiology 1 Objectives Review display function, QA metrics, procedures, and guidance provided

More information

Display Quality Assurance: Recommendations from AAPM TG270 for Tests, Tools, Patterns, and Performance Criteria

Display Quality Assurance: Recommendations from AAPM TG270 for Tests, Tools, Patterns, and Performance Criteria Display Quality Assurance: Recommendations from AAPM TG270 for Tests, Tools, Patterns, and Performance Criteria Nicholas B. Bevins, Ph.D. TG270 Co-chair Display Check 2 TG270 Goals Provide an update to

More information

Display Quality Assurance: Recommendations from AAPM TG270 for Tests, Tools, Patterns, and Performance Criteria

Display Quality Assurance: Recommendations from AAPM TG270 for Tests, Tools, Patterns, and Performance Criteria Display Quality Assurance: Recommendations from AAPM TG270 for Tests, Tools, Patterns, and Performance Criteria Nicholas B. Bevins, Ph.D. TG270 Co-chair Display Check 2 1 TG270 Goals Provide an update

More information

Evaluation report. Eizo RadiForce G33-N 3MP greyscale flat panel liquid crystal display (LCD) CEP 07003

Evaluation report. Eizo RadiForce G33-N 3MP greyscale flat panel liquid crystal display (LCD) CEP 07003 Evaluation report Eizo RadiForce G33-N 3MP greyscale flat panel liquid crystal display (LCD) CEP 07003 August 2007 Contents 2 Executive summary... 3 Introduction... 5 Product description... 6 Evaluation

More information

Equipment Quality Control for Primary Displays June 5, Imaging Physics CancerCare Manitoba

Equipment Quality Control for Primary Displays June 5, Imaging Physics CancerCare Manitoba Equipment Quality Control for Primary Displays June 5, 2018 Imaging Physics CancerCare Manitoba Purpose An equipment quality control (QC) program establishes baseline performance levels, tracks system

More information

Equipment Quality Control for Digital Radiography February 22, Imaging Physics CancerCare Manitoba

Equipment Quality Control for Digital Radiography February 22, Imaging Physics CancerCare Manitoba Equipment Quality Control for Digital Radiography February 22, 2018 Imaging Physics CancerCare Manitoba Purpose An equipment quality control (QC) program establishes baseline performance levels, tracks

More information

Coronis 5MP Mammo. The standard of care for digital mammography

Coronis 5MP Mammo. The standard of care for digital mammography Coronis 5MP Mammo The standard of care for digital mammography The standard of care For thousands of women every day, details make all the difference. This understanding, along with many years of commitment

More information

MX215. Your advantages. 2MP Medical-Display

MX215. Your advantages. 2MP Medical-Display MX215 Your advantages The EIZO MX215 features factory-preset DICOM tonal values and displays greyscale tones of radiological images in accordance with the requirements of the object. The advantage here

More information

Nio. Industry-standard diagnostic display systems

Nio. Industry-standard diagnostic display systems Nio Industry-standard diagnostic display systems Diagnostic confidence in grayscale With the Nio diagnostic display system, Barco brings dependable diagnostic imaging to its true potential. Presenting

More information

Monitor QA Management i model

Monitor QA Management i model Monitor QA Management i model 1/10 Monitor QA Management i model Table of Contents 1. Preface ------------------------------------------------------------------------------------------------------- 3 2.

More information

MX215. Your advantages. 2MP Medical-Display

MX215. Your advantages. 2MP Medical-Display MX215 Your advantages The EIZO MX215 features factory-preset DICOM tonal values and displays greyscale tones of radiological images in accordance with the requirements of the object. The advantage here

More information

+ Human method is pattern recognition based upon multiple exposure to known samples.

+ Human method is pattern recognition based upon multiple exposure to known samples. Main content + Segmentation + Computer-aided detection + Data compression + Image facilities design + Human method is pattern recognition based upon multiple exposure to known samples. + We build up mental

More information

Nio. Industry-standard diagnostic display systems

Nio. Industry-standard diagnostic display systems Nio Industry-standard diagnostic display systems Diagnostic confidence in grayscale With the Nio diagnostic display system, Barco brings dependable diagnostic imaging to its true potential. Presenting

More information

White Paper. Uniform Luminance Technology. What s inside? What is non-uniformity and noise in LCDs? Why is it a problem? How is it solved?

White Paper. Uniform Luminance Technology. What s inside? What is non-uniformity and noise in LCDs? Why is it a problem? How is it solved? White Paper Uniform Luminance Technology What s inside? What is non-uniformity and noise in LCDs? Why is it a problem? How is it solved? Tom Kimpe Manager Technology & Innovation Group Barco Medical Imaging

More information

Selection and Quality Assurance of Monitors for Mammogram Diagnosis

Selection and Quality Assurance of Monitors for Mammogram Diagnosis 924-8566 153 1 108-6105 2-15-2 B 5F 2006 2 20 Selection and Quality Assurance of Monitors for Mammogram Diagnosis Noriyuki HASHIMOTO and Kiyotaka HIRA R&D, Visual Products, EIZO NANAO CORPORATION 153 Shimokashiwano,

More information

Quality Assurance (QA) Guidelines for Medical Imaging Display Systems

Quality Assurance (QA) Guidelines for Medical Imaging Display Systems 41-1 Japan Industries Association of Radiological Systems Standards (JESRA X-0093*A -2010 ) -English version- Enacted August 8, 2005 Revised June 11, 2010 Quality Assurance (QA) Guidelines for Medical

More information

FFDM Quality Control in Canada - a Vendor Neutral Approach

FFDM Quality Control in Canada - a Vendor Neutral Approach FFDM Quality Control in Canada - a Vendor Neutral Approach Rasika Rajapakshe, PhD, FCCPM BC Cancer Agency-Center for the Southern Interior Kelowna, BC, Canada 2011 Joint AAPM/COMP Meeting August 4, 2011

More information

MX194. Your advantages. 1MP Medical-Display

MX194. Your advantages. 1MP Medical-Display MX194 Your advantages The 19-inch MX194 with a traditional 5:4 aspect ratio stands out thanks to its factory-preset DICOM characteristic curve, which guarantees faithful reproduction of radiological images

More information

White Paper. Diagnostic Color Displays. What s inside?

White Paper. Diagnostic Color Displays. What s inside? White Paper Diagnostic Color Displays What s inside? Why is color becoming more important for diagnostic imaging? How does a color display work? What are the requirements of a diagnostic color display?

More information

This talk covers currently available display technology.

This talk covers currently available display technology. Introduction to Current Display Technologies for Medical Image Viewing Perspectives for the TG270 Update on Display Quality Control Alisa Walz-Flannigan, PhD (DABR) Mayo Clinic, Rochester, Minnesota AAPM

More information

MQSA Quality Control Manual

MQSA Quality Control Manual MQSA Quality Control Manual For The MX50T, MX50N, & CX50N 5 Mega-Pixel LCD Monitors (Document Number: OMI-1035) WIDE Corporation Document No. OMI-1035 Rev. No. 1.2 Page 1 CONTENTS Revision History 4 Documents

More information

Hospital Wide. Healthcare Display Solutions DICOM Displays, Large Screen Displays and Projectors

Hospital Wide. Healthcare Display Solutions DICOM Displays, Large Screen Displays and Projectors Hospital Wide Healthcare Display Solutions DICOM Displays, Large Screen Displays and Projectors THE WIDEST RANGE OF DISPLAY SOLUTIONS For Complete Hospital Wide Installations NEC offers a wide range of

More information

systems that are found throughout the large acute hospital sector so it would not be unusual to find a range of commercialoff-the-shelf

systems that are found throughout the large acute hospital sector so it would not be unusual to find a range of commercialoff-the-shelf Quality of commercial-offthe-shelf (COTS) monitors displaying dental radiographs D. J. McIlgorm,* 1 C. Lawinski, 2 S. Ng 3 and J. P. McNulty 4 IN BRIEF Reports that commercial-off-the-shelf (COTS) computer

More information

Mammo Tomosynthesis 5MP

Mammo Tomosynthesis 5MP Mammo Tomosynthesis 5MP Display system for digital breast imaging Approved for TOMOSYNTHESIS MAMMOGRAPHY The standard of care For thousands of women every day, details make all the difference. This understanding

More information

Role of Color in Telemedicine Applications. Elizabeth A. Krupinski, PhD

Role of Color in Telemedicine Applications. Elizabeth A. Krupinski, PhD Role of Color in Telemedicine Applications Elizabeth A. Krupinski, PhD Background Color displays common clinical practice Radiology growing acceptance & use Other ologies & telemed routinely used No validated

More information

RX850. Your advantages. 8MP Medical-Display

RX850. Your advantages. 8MP Medical-Display RX850 Your advantages The RadiForce RX850 can effectively replace conventional multiscreen solutions with a single eight-megapixel screen. It clearly and simultaneously displays a wide variety of images,

More information

BUREAU OF ENERGY EFFICIENCY

BUREAU OF ENERGY EFFICIENCY Date: 26 th May, 2016 Schedule No.: 11 Color Televisions 1. Scope This schedule specifies the energy labeling requirements for color televisions with native resolution upto 1920 X 1080 pixels, of CRT,

More information

Coronis Uniti (MDMC-12133) 12MP diagnostic display system for PACS and breast imaging

Coronis Uniti (MDMC-12133) 12MP diagnostic display system for PACS and breast imaging 12MP diagnostic display system for PACS and breast imaging PACS and breast imaging on one display In grayscale and color (featuring unique color calibration!) With touch pad for fast control The only display

More information

CLINICALLY PROVEN DESIGNS

CLINICALLY PROVEN DESIGNS CLINICALLY PROVEN DESIGNS Radiology & PACS WWW.TLGC.COM Rely on Optik View high definition diagnostic and reference displays to provide your Customer s with industry s highest quality medical visualization.

More information

White Paper. Missing Pixels in Medical Grayscale Flat Panel Displays. Geert Carrein Director Product Management. W hat s inside?

White Paper. Missing Pixels in Medical Grayscale Flat Panel Displays. Geert Carrein Director Product Management. W hat s inside? White Paper Missing Pixels in Medical Grayscale Flat Panel Displays W hat s inside? What are missing pixels? International Standards to characterize LCD quality Missing Pixels, overview for different LCD

More information

Overview. ACR Accreditation Update in Mammography. ACR Topics. Requirements Today. What s Coming For Tomorrow

Overview. ACR Accreditation Update in Mammography. ACR Topics. Requirements Today. What s Coming For Tomorrow ACR Accreditation Update in Mammography Eric Berns, PhD University of Colorado Hospital Denver Health Medical Center Denver, CO *No financial disclosures to report Overview ACR Topics Requirements Today

More information

2017 Computed Tomography

2017 Computed Tomography 2017 Computed Tomography QUALITY CONTROL MANUAL Radiologist s Section Radiologic Technologist s Section Qualified Medical Physicist s Section 2017 Computed Tomography QUALITY CONTROL MANUAL Radiologist

More information

Performance characteristics and quality assurance considerations for displays used in interventional radiology and cardiac catheterization facilities

Performance characteristics and quality assurance considerations for displays used in interventional radiology and cardiac catheterization facilities Received: 2 January 2018 Revised: 7 March 2018 Accepted: 7 December 2018 DOI: 10.1002/acm2.12433 MEDICAL IMAGING Performance characteristics and quality assurance considerations for displays used in interventional

More information

White Paper. In Plane Switching Pro technology for medical imaging. Geert Carrein Director Product Management. What s inside?

White Paper. In Plane Switching Pro technology for medical imaging. Geert Carrein Director Product Management. What s inside? White Paper In Plane Switching Pro technology for medical imaging What s inside? An introduction to LCD technologies A primer on IPS-Pro Why is IPS-Pro LCD technology important for medical imaging? What

More information

Optimizing the Workflow of Radiologists

Optimizing the Workflow of Radiologists Clinical Performance That Matters Optimizing the Workflow of Radiologists www.barcomedical.com Meeting the Challenges of Today s Reading Rooms Radiology has long been at the epicenter of healthcare, with

More information

Australian Dental Journal

Australian Dental Journal Australian Dental Journal The official journal of the Australian Dental Association Australian Dental Journal 2012; 57:(1 Suppl): 16 23 doi: 10.1111/j.1834-7819.2011.01660.x The impact of computer display

More information

1 Your computer screen

1 Your computer screen U.S.T.H.B / C.E.I.L Unit 7 Computer science L2 (S2) 1 Your computer screen Discuss the following questions. 1 What type of display do you have? 2 What size is the screen? 3 Can you watch TV on your PC

More information

Coronis Fusion 6MP DL. 6 MegaPixel wide-screen diagnostic color display system

Coronis Fusion 6MP DL. 6 MegaPixel wide-screen diagnostic color display system Coronis Fusion 6MP DL 6 MegaPixel wide-screen diagnostic color display system Live free Smart ideas can make a big difference. Sometimes, they can even change your view... That is exactly what Coronis

More information

Clarity and confidence. Raising the bar in diagnostic imaging with Barco radiology and mammography displays

Clarity and confidence. Raising the bar in diagnostic imaging with Barco radiology and mammography displays Clarity and confidence Raising the bar in diagnostic imaging with Barco radiology and mammography displays Breaking new ground in healthcare imaging 1986 Barco s first medical display for ultrasound, MCD-10

More information

QUALITY CONTROL AND PATIENT DOSES FROM X-RAY EXAMINATIONS IN SOME HOSPITALS IN THAILAND

QUALITY CONTROL AND PATIENT DOSES FROM X-RAY EXAMINATIONS IN SOME HOSPITALS IN THAILAND QUALITY CONTROL AND PATIENT DOSES FROM X-RAY EXAMINATIONS IN SOME HOSPITALS IN THAILAND P. Plainoi, W. Diswath, N. Manatrakul Ministry of Public Health, Nonthaburi, Thailand XA0101612 Abstract Quality

More information

Coronis Fusion multi-modality displays. The ultimate in diagnostic flexibility

Coronis Fusion multi-modality displays. The ultimate in diagnostic flexibility Coronis Fusion multi-modality displays The ultimate in diagnostic flexibility The ultimate in diagnostic flexibility Radiology reading rooms are fast becoming one of the busiest departments in the healthcare

More information

Scope: All CT staff technologist

Scope: All CT staff technologist APPROVED BY: Radiology Technical Director Page 1 of 6 Purpose: The QC program assesses relative changes in system performance as determined by the technologist, service engineer, qualified medical physicist,

More information

PROLITE B2791HSU-B1 HAS + PIVOT FLICKER FREE + BLUE LIGHT

PROLITE B2791HSU-B1 HAS + PIVOT FLICKER FREE + BLUE LIGHT PROLITE B2791HSU-B1 27 solid Full HD monitor your e will love Looking at a screen for long hours can be quite tiring for the e. But it doesn t have to. The new 27 iiyama ProLite B2791HSU is big enough

More information

Coronis. A new generation of premium diagnostic display systems

Coronis. A new generation of premium diagnostic display systems Coronis A new generation of premium diagnostic display systems Diagnostic precision redefi ned Building on years of experience in state-of-the-art visualization, Barco has advanced medical imaging with

More information

2012 Computed Tomography

2012 Computed Tomography 2012 Computed Tomography QUALITY CONTROL MANUAL Radiologist s Section Radiologic Technologist s Section Medical Physicist s Section 2012 Computed Tomography QUALITY CONTROL MANUAL Radiologist s Section

More information

RX350. Your advantages. 3MP Medical-Display

RX350. Your advantages. 3MP Medical-Display RX350 Your advantages RadiForce RX350 features 3-megapixel resolution and high luminance perfect for precisely displaying radiological images. The device provides high image quality, advantageous for displaying

More information

Global Trade Medical Supplies

Global Trade Medical Supplies Features: Achieve Clarity True to the Source Data A medical monitor needs to be capable of high brightness in order to meet performance standards. However, in order to achieve high brightness in an LCD

More information

DCI Requirements Image - Dynamics

DCI Requirements Image - Dynamics DCI Requirements Image - Dynamics Matt Cowan Entertainment Technology Consultants www.etconsult.com Gamma 2.6 12 bit Luminance Coding Black level coding Post Production Implications Measurement Processes

More information

What to consider when choosing a mammography display

What to consider when choosing a mammography display What to consider when choosing a mammography display Screen size and resolution In digital breast imaging, the quality of the medical display has a direct impact on the decisions you make. Next to display

More information

TechNote: MuraTool CA: 1 2/9/00. Figure 1: High contrast fringe ring mura on a microdisplay

TechNote: MuraTool CA: 1 2/9/00. Figure 1: High contrast fringe ring mura on a microdisplay Mura: The Japanese word for blemish has been widely adopted by the display industry to describe almost all irregular luminosity variation defects in liquid crystal displays. Mura defects are caused by

More information

Medical Imaging Working Group

Medical Imaging Working Group Medical Imaging Working Group Czech Academy of Sciences Národní 3 117 20 Prague 1, Czech Republic 28 June 2017 Craig Revie, MIWG chair, opened the meeting at 10:45 and introduced the agenda as follows:

More information

DirectView Elite CR System. Improve workflow, productivity, and patient throughput.

DirectView Elite CR System. Improve workflow, productivity, and patient throughput. Improve workflow, productivity, and patient throughput. DirectView DirectView Compact, single cassette CR systems designed to improve workflow, productivity, and patient throughput. The is small, easy

More information

Chapter 10. Lighting Lighting of Indoor Workplaces 180

Chapter 10. Lighting Lighting of Indoor Workplaces 180 Chapter 10 Lighting 10.1 Lighting of Indoor Workplaces 180 10 10 Lighting 10.1 Lighting of Indoor Workplaces In March 2003, the German version of the European Standard EN 12464-1 Lighting of workplaces,

More information

ASSESSMENT OF DISPLAY PERFORMANCE FOR MEDICAL IMAGING SYSTEMS

ASSESSMENT OF DISPLAY PERFORMANCE FOR MEDICAL IMAGING SYSTEMS DISCLAIMER Medical physicists, investigators, vendors, or other users can utilize the authentic copyrighted TG18 patterns supplied in conjunction with this report for any professional, investigational,

More information

CARESTREAM DIRECTVIEW Elite CR System

CARESTREAM DIRECTVIEW Elite CR System CARESTREAM DIRECTVIEW Elite CR System Improve workflow, productivity, and patient throughput. The CARESTREAM DIRECTVIEW Elite CR System is small, easy to install and easy to use. This powerful distributed

More information

CARESTREAM DIRECTVIEW Elite CR System

CARESTREAM DIRECTVIEW Elite CR System CARESTREAM DIRECTVIEW Elite CR System Improve workflow, productivity, and patient throughput. The CARESTREAM DIRECTVIEW Elite CR System is small, easy to install and easy to use. This powerful distributed

More information

General viewing conditions for subjective assessment of quality of SDTV and HDTV television pictures on flat panel displays

General viewing conditions for subjective assessment of quality of SDTV and HDTV television pictures on flat panel displays Recommendation ITU-R BT.2022 (08/2012) General viewing conditions for subjective assessment of quality of SDTV and HDTV television pictures on flat panel displays BT Series Broadcasting service (television)

More information

Essentials of Digital Imaging

Essentials of Digital Imaging Essentials of Digital Imaging Module 3 Transcript 2016 ASRT. All rights reserved. Essentials of Digital Imaging Module 3 Display 1. ASRT Animation 2. Welcome Welcome to the Essentials of Digital Imaging

More information

Chapter Medical Table of Contents

Chapter Medical Table of Contents Chapter Table of Contents Replacement / Patient Monitoring / Modality LCD Series... 57 Colour Displays in Wide Format... 58 42 Surgical LCD for Operating Theaters... 59 24 Endoscopy LCD... 60 19 Endoscopy

More information

HONEYWELL VIDEO SYSTEMS HIGH-RESOLUTION COLOR DOME CAMERA

HONEYWELL VIDEO SYSTEMS HIGH-RESOLUTION COLOR DOME CAMERA Section 00000 SECURITY ACCESS AND SURVEILLANCE HONEYWELL VIDEO SYSTEMS HIGH-RESOLUTION COLOR DOME CAMERA PART 1 GENERAL 1.01 SUMMARY The intent of this document is to specify the minimum criteria for the

More information

ENERGY STAR Program Requirements Product Specification for Televisions. Eligibility Criteria Version 5.3

ENERGY STAR Program Requirements Product Specification for Televisions. Eligibility Criteria Version 5.3 ENERGY STAR Program Requirements Product Specification for Televisions Eligibility Criteria Version 5.3 Following is the Version 5.3 ENERGY STAR Product Specification for Televisions. A product shall meet

More information

GUIDELINES ON THE USE OF DISPLAY SCREEN EQUIPMENT

GUIDELINES ON THE USE OF DISPLAY SCREEN EQUIPMENT GUIDELINES ON THE USE OF DISPLAY SCREEN EQUIPMENT 1. Preamble In view of the legislative requirements and growing concern on the health effects related to the use of display screen equipment (DSE), the

More information

224. Acceptability of screen reflections: lighting strategies for improving quality of the visual environment in the Classrooms of the Future

224. Acceptability of screen reflections: lighting strategies for improving quality of the visual environment in the Classrooms of the Future 224. Acceptability of screen reflections: lighting strategies for improving quality of the visual environment in the Classrooms of the Future Tharinee Ramasoot 1 * and Steve Fotios 1 School of Architecture,

More information

MEDICAL PRODUCT OVERVIEW

MEDICAL PRODUCT OVERVIEW NEC Display Solutions February 2011 MEDICAL PRODUCT OVERVIEW DICOM DISPLAY SOLUTIONS Hospital Wide Display Solutions Desktop, Projector and Large Format H E A L T H C A R E NEC Display Solutions Hospital

More information

-Technical Specifications-

-Technical Specifications- Annex I to Contract 108733 NL-Petten: the delivery, installation, warranty and maintenance of one (1) X-ray computed tomography system at the JRC-IET -Technical Specifications- INTRODUCTION In the 7th

More information

DRAFT. Proposal to modify International Standard IEC

DRAFT. Proposal to modify International Standard IEC Imaging & Color Science Research & Product Development 2528 Waunona Way, Madison, WI 53713 (608) 222-0378 www.lumita.com Proposal to modify International Standard IEC 61947-1 Electronic projection Measurement

More information

Medical Monitor Solutions

Medical Monitor Solutions Medical Monitor Solutions Worldwide, more and more hospitals and other health care institutions are rapidly moving towards a filmless environment. In line with this trend, medical image display monitors

More information

Operating room monitor Medical-grade 4K UHD monitor

Operating room monitor Medical-grade 4K UHD monitor Operating room monitor Medical-grade 4K UHD monitor EJ-ML432Z Perfection in detail Crystal-clear image reproduction Moist disinfection Energy-efficient *Actual resolution 3.840 2.160p EJ-ML432Z PAGE 2

More information

Barco Healthcare. Product catalog

Barco Healthcare. Product catalog Barco Healthcare Product catalog Supporting healthcare professionals per year More than a billion times each year, healthcare professionals rely on Barco to make life-critical decisions. Every day, Barco

More information

How to Match the Color Brightness of Automotive TFT-LCD Panels

How to Match the Color Brightness of Automotive TFT-LCD Panels Relative Luminance How to Match the Color Brightness of Automotive TFT-LCD Panels Introduction The need for gamma correction originated with the invention of CRT TV displays. The CRT uses an electron beam

More information

Coronis Uniti. See what you haven t seen before

Coronis Uniti. See what you haven t seen before Coronis Uniti See what you haven t seen before The facts The MarkeTech Group surveyed over 200 radiologists in Europe and North America to ask about the challenges in radiology. The key findings? About

More information

Is image manipulation necessary to interpret digital mammographic images efficiently?

Is image manipulation necessary to interpret digital mammographic images efficiently? Loughborough University Institutional Repository Is image manipulation necessary to interpret digital mammographic images efficiently? This item was submitted to Loughborough University's Institutional

More information

MDSC-2232 Full HD 32-inch surgical display

MDSC-2232 Full HD 32-inch surgical display Full HD 32-inch surgical display For the best hand-eye coordination Allows for easy cleaning and disinfection Works with Nexxis for video-over-ip integration Barco s is a 32" near-patient surgical display

More information

VeriLUM 5.2. Video Display Calibration And Conformance Tracking. IMAGE Smiths, Inc. P.O. Box 30928, Bethesda, MD USA

VeriLUM 5.2. Video Display Calibration And Conformance Tracking. IMAGE Smiths, Inc. P.O. Box 30928, Bethesda, MD USA VeriLUM 5.2 Video Display Calibration And Conformance Tracking IMAGE Smiths, Inc. P.O. Box 30928, Bethesda, MD 20824 USA Voice: 240-395-1600 Fax: 240-395-1601 Web: www.image-smiths.com Technical Support

More information

More Info at Open Access Database Process Control for Computed Tomography using Digital Detector Arrays

More Info at Open Access Database  Process Control for Computed Tomography using Digital Detector Arrays Digital Industrial Radiology and Computed Tomography (DIR 2015) 22-25 June 2015, Belgium, Ghent - www.ndt.net/app.dir2015 More Info at Open Access Database www.ndt.net/?id=18082 Process Control for Computed

More information

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD INTERNATIONAL STANDARD IEC 61947-1 First edition 2002-08 Electronic projection Measurement and documentation of key performance criteria Part 1: Fixed resolution projectors IEC 2002 Copyright - all rights

More information

Christie MicroTiles. Technical Frequently Asked Questions (FAQs)

Christie MicroTiles. Technical Frequently Asked Questions (FAQs) Christie MicroTiles Technical Frequently Asked Questions (FAQs) December 11, 2009 Index 1 Size and physical installation... 3 1.1 How many tiles will I need to fit a physical space?... 3 1.2 What is the

More information

BRIGHT BRIGHTER BRIGHTEST ONE ILLUMINATOR DESIGN THREE LIGHT SOURCES. featuring the EvenVue Reflector System

BRIGHT BRIGHTER BRIGHTEST ONE ILLUMINATOR DESIGN THREE LIGHT SOURCES. featuring the EvenVue Reflector System BRIGHT BRIGHTER BRIGHTEST ONE ILLUMINATOR DESIGN THREE LIGHT SOURCES featuring the EvenVue Reflector System DVC (DIAGNOSTIC VIEWING CENTER) The features and benefits of the TECHLINE can be combined with

More information

Intuitive Workflow by Barco. Designed for the way you work, naturally.

Intuitive Workflow by Barco. Designed for the way you work, naturally. Intuitive Workflow by Barco Designed for the way you work, naturally. As the volume and complexity of patient exams continue to grow, radiologists face increasing demands to boost their productivity. Many

More information

American National Standard for Lamp Ballasts High Frequency Fluorescent Lamp Ballasts

American National Standard for Lamp Ballasts High Frequency Fluorescent Lamp Ballasts American National Standard for Lamp Ballasts High Frequency Fluorescent Lamp Ballasts Secretariat: National Electrical Manufacturers Association Approved: January 23, 2017 American National Standards Institute,

More information

Health Professions Council Education & Training Panel 5 July 2007 NORDOFF ROBBINS MUSIC THERAPY CENTRE - MA MUSIC THERAPY

Health Professions Council Education & Training Panel 5 July 2007 NORDOFF ROBBINS MUSIC THERAPY CENTRE - MA MUSIC THERAPY Health Professions Council Education & Training Panel 5 July 2007 NORDOFF ROBBINS MUSIC THERAPY CENTRE - MA MUSIC THERAPY Executive Summary and Recommendations Introduction The visitors report for the

More information

The University of Texas at Austin September 30, 2011

The University of Texas at Austin September 30, 2011 SECTION 27 08 20 COPPER TESTING PART 1 - GENERAL 1.1 SUMMARY A. Test measurements shall be taken for all balanced-twisted pair cabling, including horizontal and backbone copper cables and wall-to-rack

More information

ENERGY STAR Program Requirements Product Specification for Televisions. Eligibility Criteria Draft 1 Version 6.0

ENERGY STAR Program Requirements Product Specification for Televisions. Eligibility Criteria Draft 1 Version 6.0 ENERGY STAR rogram Requirements roduct Specification for Televisions Eligibility Criteria Draft 1 Version 6.0 1 2 Following is the Draft 1 Version 6.0 ENERGY STAR roduct Specification for Televisions.

More information

Radiography and Mammography Viewing Systems

Radiography and Mammography Viewing Systems Radiography and Mammography Viewing Systems For assistance contact your authorized RADX Dealer, or call S&S Technology at 800-231-1747 E-mail salessupport@radx.com www.ssxray.com Mammography Digital Viewing

More information

Colour Matching Technology

Colour Matching Technology Colour Matching Technology For BVM-L Master Monitors www.sonybiz.net/monitors Colour Matching Technology BVM-L420/BVM-L230 LCD Master Monitors LCD Displays have come a long way from when they were first

More information

multi-function meters

multi-function meters multi-function meters eclipse 2 eclipse 7 installation and operating manual 1 GENERAL DESCRIPTION 2 INSTALLATION 3 WIRING INFORMATION 4 2 ECLIPSE 2 METERS 2.1 PROGRAMMING THE METER 5 2.2 INFORMATION 6

More information

ENERGY STAR Program Requirements Product Specification for Televisions. Draft Test Method

ENERGY STAR Program Requirements Product Specification for Televisions. Draft Test Method ENERGY STAR Program Requirements Product Specification for Televisions Draft Test Method Note: EPA is committed to supporting and adopting the television test procedure currently under development by the

More information

CAPTURE CAPTURE. VERSiON 1.2. Specialists in Medical. Digital Imaging Solutions

CAPTURE CAPTURE. VERSiON 1.2. Specialists in Medical. Digital Imaging Solutions CAPTURE Add-on module for capturing images from analog video sources CAPTURE anything to DICOm VERSiON 1.2 Specialists in Medical Digital Imaging Solutions iq-capture Connect any medical device to your

More information

Measurement of Microdisplays at NPL

Measurement of Microdisplays at NPL Conference on Microdisplays Measurement of Microdisplays at NPL Christine Wall, Dr Julie Taylor, Colin Campbell 14 th Sept 2001 Overview Displays measurement at NPL Why measure microdisplays? Measurement

More information

INTERNATIONAL STANDARD

INTERNATIONAL STANDARD INTERNATIONAL STANDARD IEC 61947-2 First edition 2001-09 Electronic projection Measurement and documentation of key performance criteria Part 2: Variable resolution projectors IEC 2001 Copyright - all

More information

Table 1. Factors affecting display quality and associated tradeoff factors.

Table 1. Factors affecting display quality and associated tradeoff factors. UDC 548-144: 681.327.17 LCD Technology VAkio Sotokawa VOsamu Ishibashi (Manuscript received July 28,1998) This paper looks at the various factors that affect the display quality of liquid crystal displays

More information

Coronis 5MP Mammo. Online User Guide

Coronis 5MP Mammo. Online User Guide Coronis 5MP Mammo Online User Guide (This page intentionally left blank.) 2 Table of contents Table of contents Using the online User Guide...4 Sources of information... 4 User interface... 4 Graphic board

More information

Solution for Nonuniformities and Spatial Noise in Medical LCD Displays by Using Pixel-Based Correction

Solution for Nonuniformities and Spatial Noise in Medical LCD Displays by Using Pixel-Based Correction Solution for Nonuniformities and Spatial Noise in Medical LCD Displays by Using Pixel-Based Correction Tom Kimpe, Albert Xthona, Paul Matthijs, and Lode De Paepe Liquid crystal displays (LCD) are rapidly

More information

Understanding PQR, DMOS, and PSNR Measurements

Understanding PQR, DMOS, and PSNR Measurements Understanding PQR, DMOS, and PSNR Measurements Introduction Compression systems and other video processing devices impact picture quality in various ways. Consumers quality expectations continue to rise

More information

Australian/New Zealand Standard

Australian/New Zealand Standard AS/NZS 3821:1998 IEC 60406:1997 Australian/New Zealand Standard Cassettes for medical X-ray diagnosis Radiographic cassettes and mammographic cassettes AS/NZS 3821:1998 This Joint Australian/New Zealand

More information

Guidelines for Quality Assurance in Radiation Protection for Diagnostic X-Ray Facilities: Large X-Ray Facilities

Guidelines for Quality Assurance in Radiation Protection for Diagnostic X-Ray Facilities: Large X-Ray Facilities Guidelines for Quality Assurance in Radiation Protection for Diagnostic X-Ray Facilities: Large X-Ray Facilities J L Poletti 1995/1 NRL Report GUIDELINES FOR QUALITY ASSURANCE IN RADIATION PROTECTION FOR

More information

Request for Proposals

Request for Proposals Request for Proposals Reference: ERDFIAI2012-4006A Caring First Ltd, is a limited liability company providing healthcare services. The company is currently commissioning a private new-build hospital complete

More information

AU OPTRONICS CORPORATION. Specification for Approval. INCOMING INSPECTION STANDARD FOR A201SN02 TFT-LCD MODULES (A- Grade)

AU OPTRONICS CORPORATION. Specification for Approval. INCOMING INSPECTION STANDARD FOR A201SN02 TFT-LCD MODULES (A- Grade) AU OPTRONICS CORPORATION Specification for Approval INCOMING INSPECTION STANDARD FOR A201SN02 TFT-LCD MODULES (A- Grade) The content of this technical information is subject to change without notice. Please

More information