QUALITY CONTROL AND PATIENT DOSES FROM X-RAY EXAMINATIONS IN SOME HOSPITALS IN THAILAND

Size: px
Start display at page:

Download "QUALITY CONTROL AND PATIENT DOSES FROM X-RAY EXAMINATIONS IN SOME HOSPITALS IN THAILAND"

Transcription

1 QUALITY CONTROL AND PATIENT DOSES FROM X-RAY EXAMINATIONS IN SOME HOSPITALS IN THAILAND P. Plainoi, W. Diswath, N. Manatrakul Ministry of Public Health, Nonthaburi, Thailand XA Abstract Quality control measurements on 203 diagnostic X-ray units were carried out in 126 hospitals in the central region of Thailand during The measurements consisted of tube voltage, half-value layer (HVL), exposure time, radiation output, beam alignment, light beam diaphragm and entrance surface dose (ESD) in four common radiographic procedures namely adult chest PA, adult mass chest PA in mobile bus unit, abdomen AP and mammography (cephalo caudal view). ESD measurements of 320 examinations were performed using parallel plate ionization chamber and Keithley model 35050A Dosimeter on 192 X-ray units (conventional and mobile) and 11 mammography units. The analysis of test results showed that: 1. 92% had X-ray tube voltage within the tolerance limit of 10% and HVL 3.03 mmal ± 0.55 SD at 80 kvp measured 2. 90% had exposure time within tolerance limit of 10% 3. 86% and 98% had acceptable beam alignment and light beam diaphragm 4. 9 had radiation output > 25 p,gy at 1 m. for true 80 KVp 5. It was found that the ESD values were adult chest (PA) varied from 0.18 mgy to 1.17 mgy (mean ± SD). adult mass chest (PA) varied from mgy to 1.03 mgy (mean ± SD). abdomen (AP) varied from mgy to 6.30 mgy (mean ± SD) mammography (cephalo caudal view) varied from 3.49 mgy to mgy (mean SD) Further surveys are necessary and are being done to include measurements of image quality and for propagation of quality assurance activities in Thailand so as to reduce patient doses while maintaining the image quality. 1. Introduction Diagnostic radiology is the main contributor to the man-made exposure of general population. Since Quality Assurance (QA) programmes ensure high quality diagnostic images with the least possible radiation dose to the patient, all countries have been recommended to introduce programmes for their radiological facilities. In Thailand periodic quality control of all X ray equipment in use is required by legislation. For monitoring purposes radiation leakage, total beam filtration, exposure time, tube voltage (kvp), radiation output, accuracy of beam limiting devices and adequacy of room shielding design were measured. Where available, tolerance limits established by FDA, NCRP, AAPM and ACR were used as a reference [1-4]. According to a programme of patient dose measurements was introduced as part of the quality assurance service already provided for X-ray departments throughout many countries. The emphasis that each patient exposure should be as low as reasonably practicable and encouragement to X ray departments to formulate a strategy for dose reduction [5]. We, therefore, initiated to add the patient dose measurement of entrance surface dose in four common radiographic procedures into our routine protocol as a pilot study. The performance characteristics of 203 diagnostic X-ray units were carried out in 126 hospitals during These diagnostic units were located in the central region of Thailand in government and private hospitals; they covered the whole range of commercial equipment in Thailand. They represent 8% of the total diagnostic units of central region and 3% of the country. 99

2 2. Materials and methods The quality control measurements were carried out on 203 X ray units. The measured units consisted of conventional and mammographic X ray units. Performance measurements were focused on tube voltage, beam quality (half - value layer), exposure time, radiation output, beam alignment and accuracy of beam limiting devices, and entrance surface dose (ESD) in four common radiographic procedures namely adult chest PA, adult mass chest PA in mobile bus unit, abdomen AP and mammography (cephalo caudal view). The parameters which were measured are listed in Table 1 and Table 2 for conventional and mammography X ray units respectively, as well as the acceptability criteria [1-4]. The X ray tube potential and exposure time were measured at 60, 70, 80, 90 and 100 kv using Keithley dosimeter/kvp readout model 35050A and kvp divider model 35080A at 56 cm. X ray source to table top. The X ray beam quality, radiation output and radiation output reproducibility were performed at 80 kv actual beam for 20 mas at 40 cm. from the X ray tube focus to chamber using Keithley dosimeter/kvp readout model 35050A, ionization chamber model B and 100 cm. aluminuim sheets (No. 1100) were insert between the tube and ionization chamber in case of beam quality measurement. The linearity of the radiation output (variation of the output as a function of mas) was checked between 5 mas and 64 mas. The correspondance between the light beam and the actual X ray beam, and X ray beam alignment were tested using the collimator and beam alignment test tools by RMI at 100 cm. from the X ray tube focus to the test tool which was placed on the table. The patient doses for conventional X ray units, were calculated measuring the X ray output and using the tabulated exposure parameters of each X ray examination used in the hospital in question. While the surface dose of the mammographic X ray units were measured on RMI acryric phantom as required by the American College of Radiology (ACR) with the Keithley dosimeter/kvp readout model A, ionization chamber model B using the local examination technique in each place. Table 1. Acceptability criteria of the measured parameters for conventional X ray units kv accuracy kv reproducibility Filtration at 80 kv actual beam Tube output (80 kv at 1 m from the focus) Reproducibility of the tube output Linearity of the tube output Exposure time Accuracy Reproducibility Correspondance light beam and actual X ray beam Leakeage radation at lm from the focus in any direction > 2.3 mmal >25 ugy(ma.s) ~ l < 2% of FID <1 u.gy.s 4 100

3 Table 2. Acceptability criteria of the measured parameters for mammography X ray units. Results kv accuracy kv reproducibility Filtration at 30 kv actual beam Reproducibility of the tube output Linearity of the tube output Automatic exposure cells reproducibility Correspondance light beam and actual X ray beam Leakage radiation at 1 m from the focus in any direction ± 2% > 0.3 mmal ^ 1% of FID <1 ugy.s"' The analysis of X ray units test results showed that 92% had acceptable deviation between nominal and measured values of X ray tube voltage with in the tolerance limit of 10% and HVL 3.03 mm.al ± 0.55 SD at 80 kvp actual beam whereas 90% of the X ray units meet the recommended limits of exposure time (deviation < 10%). Of the X ray generators assessed, 84% had tolerable (ugy/mas) linearity. For the light beam and actual X ray beam 2% of the X ray units showed a difference of more than 2% of FID (focus to image distance). Moreover, in 86% of the X ray tubes, the beam was properly aligned. Measurements on the X ray tubes showed that 96% had adequate beam filtration (HVL > 2.3 mm.al, 80 kvp) and 9 had radiation output less than 25 j,gy at 1 m. for true 80 kvp whereas in 98% exposure reproducibility had acceptable variation within the tolerance limit of. The entrance surface doses for three main types of examinations (some variety of chest such as adult chest PA nonbucky, adult chest PA with bucky and adult mass chest PA in mobile bus units, abdomen, and mammogram) which were measured during the years , are presented in Table 3. Figures 1 to 5 show the entrance surface dose of three categories of chest techniques, abdomen and mammogram respectively in term of mode (frequency of data) and median values for entrance surface dose of each technique. Table 3. for three main types of examinations in different years Examination Adult Chest PA (without bucky) Adult chest PA (with bucky) Adult mass chest PA (mobile bus unit) Abdomen Mammogram mean ESD (mgy)

4 10 ' J V 1 1 ZL r (mgy) Figure 1. of adult chest PA nonbucky Figure 2. of chest PA with bucky stand 3 2 -TT ).050.0eD.070.0SD.100.mil0.mi ).lSO.ia) D.lS0.2C0.2C e ).3ID.32D.33) ).3«.33).3E n 0.' Figure 3. of adult mass chest PA in mobile bus units 102

5 15 T * % C Figure 4. of abdomen Figure 5. of MAMMOGRAM 3. Discussion The results presented here confirm that the acceptable level of the quality control in diagnostic X ray equipment have led to the improvement and reduction in patient doses. In addition, the procedure which was adopted has been well received by the radiologists, commercial firms and public services. This encouraged us to extend the inspection to technical parameters such as focus, image quality, image recording system, television monitors, film storage conditions, and film development methods. The implemention of Quality Assurance in Thailand may take time. Staff training and administrative actions of the radiation protection authorities will improve the existing situation for the benefit of patients and staff. References [1] Title 21, Code of Federal Regulations, Part , FDA, U.S. Government Printing Office, Washington, D.C., 1991 and revisions. [2] National Council on Radiation Protection and Measurements: Quality Assurance for Diagnostic Imaging Equipment. NCRP 99, NCRP, Bethesda, Meryland, [3] AAPM report No. 31, Standardized Methods for Measuring Diagnostic X-Ray Exposures, AAPM Diagnostic X-Ray Imaging Committee TG #8 (1991). [4] American College of Radiology, Mammography Quality Control Radiologist's manual, Radiologic Technologist's Manual and Medical Physicist's Manual, [5] Guidance Note PM77 from the Health and Safety Executive. Fitness of Equipment used for Medical Exposure to Ionizing Radiation (1992). 103

Equipment Quality Control for Digital Radiography February 22, Imaging Physics CancerCare Manitoba

Equipment Quality Control for Digital Radiography February 22, Imaging Physics CancerCare Manitoba Equipment Quality Control for Digital Radiography February 22, 2018 Imaging Physics CancerCare Manitoba Purpose An equipment quality control (QC) program establishes baseline performance levels, tracks

More information

FFDM Quality Control in Canada - a Vendor Neutral Approach

FFDM Quality Control in Canada - a Vendor Neutral Approach FFDM Quality Control in Canada - a Vendor Neutral Approach Rasika Rajapakshe, PhD, FCCPM BC Cancer Agency-Center for the Southern Interior Kelowna, BC, Canada 2011 Joint AAPM/COMP Meeting August 4, 2011

More information

Intra-oral X-ray unit with high frequency DC generator. Product Data

Intra-oral X-ray unit with high frequency DC generator. Product Data Intra-oral X-ray unit with high frequency DC generator Tubehead Features Product Data High voltage circuit type High frequency (DC) High voltage frequency 30 khz Rated output voltage 60-65 - 70 kvp ± 8%

More information

Guidelines for Quality Assurance in Radiation Protection for Diagnostic X-Ray Facilities: Large X-Ray Facilities

Guidelines for Quality Assurance in Radiation Protection for Diagnostic X-Ray Facilities: Large X-Ray Facilities Guidelines for Quality Assurance in Radiation Protection for Diagnostic X-Ray Facilities: Large X-Ray Facilities J L Poletti 1995/1 NRL Report GUIDELINES FOR QUALITY ASSURANCE IN RADIATION PROTECTION FOR

More information

Overview. ACR Accreditation Update in Mammography. ACR Topics. Requirements Today. What s Coming For Tomorrow

Overview. ACR Accreditation Update in Mammography. ACR Topics. Requirements Today. What s Coming For Tomorrow ACR Accreditation Update in Mammography Eric Berns, PhD University of Colorado Hospital Denver Health Medical Center Denver, CO *No financial disclosures to report Overview ACR Topics Requirements Today

More information

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009 2033-17 Joint ICTP/ Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for CT 3: Practical Experiences Claire-Louise Chapple Freeman Hospital

More information

2017 Computed Tomography

2017 Computed Tomography 2017 Computed Tomography QUALITY CONTROL MANUAL Radiologist s Section Radiologic Technologist s Section Qualified Medical Physicist s Section 2017 Computed Tomography QUALITY CONTROL MANUAL Radiologist

More information

Staff dose and good prac/ces in CT guided IR procedures. Jonas Andersson Medical Physicist, Ph.D.

Staff dose and good prac/ces in CT guided IR procedures. Jonas Andersson Medical Physicist, Ph.D. Staff dose and good prac/ces in CT guided IR procedures Jonas Andersson Medical Physicist, Ph.D. jonas.s.andersson@vll.se Contents Good prac/ces for staff dose management Staff dose reported in the literature

More information

Rev. 01 February 2014 X- RAY SAFETY MANUAL

Rev. 01 February 2014 X- RAY SAFETY MANUAL X- RAY SAFETY MANUAL 1 I. INTRODUCTION The possession and use of radiation producing equipment at Clemson University is authorized by the South Carolina Department of Health and Environmental Control (S.C.

More information

2012 Computed Tomography

2012 Computed Tomography 2012 Computed Tomography QUALITY CONTROL MANUAL Radiologist s Section Radiologic Technologist s Section Medical Physicist s Section 2012 Computed Tomography QUALITY CONTROL MANUAL Radiologist s Section

More information

Guidelines for Assuring Softcopy Image Quality

Guidelines for Assuring Softcopy Image Quality Guidelines for Assuring Softcopy Image Quality What s inside? Quality Control Guidelines Softcopy QA testing and frequencies Danny Deroo Product and R&D Manager QA Products ABSTRACT To ensure diagnostic

More information

Madero Ote. 686, Centro Histórico, C.P Morelia, Michoacán México. #300, Col. Cuauhtémoc, C.P Morelia, Michoacán, México

Madero Ote. 686, Centro Histórico, C.P Morelia, Michoacán México. #300, Col. Cuauhtémoc, C.P Morelia, Michoacán, México Verification of and DLP values for a head tomography reported by the manufacturers of the CT scanners, using a CT dose profiler probe, a head phantom and a piranha electrometer Edith Castillo Corona 1,2,

More information

Collimation and Light Field Brightness

Collimation and Light Field Brightness Collimation and Light Field Brightness Physics and Reasoning Principle 1: Primary X Ray beam is blocked by breast support table. Principle 2: Need to properly imageas as much breast tissue as possible.

More information

X-ray Tube Housing Assembly H1076X H1076Y Print No.HA Release Date:

X-ray Tube Housing Assembly H1076X H1076Y Print No.HA Release Date: H1076X 0.6 1.5 H1076Y 0.6 1.5 Print No.HA1006 1 Release Date: 2013 07 15 Description X-ray Tube Housing Assembly The tube unit is so constructed that an X-ray tube is sealed in a diagnostic type protective

More information

Australian/New Zealand Standard

Australian/New Zealand Standard AS/NZS 3821:1998 IEC 60406:1997 Australian/New Zealand Standard Cassettes for medical X-ray diagnosis Radiographic cassettes and mammographic cassettes AS/NZS 3821:1998 This Joint Australian/New Zealand

More information

Assessment of the Impact of Dark Signal on Image Quality in Computerized Mammography. Introduction

Assessment of the Impact of Dark Signal on Image Quality in Computerized Mammography. Introduction Assessment of the Impact of Dark Signal on Image Quality in Computerized Mammography Alexia Valasi 1, Maria Argyrou 1, Ioannis Floros 1, Elisavet Molyvda 2, Anastasios Siountas 2, Stella Synefia 1, Ioannis

More information

Selection and Quality Assurance of Monitors for Mammogram Diagnosis

Selection and Quality Assurance of Monitors for Mammogram Diagnosis 924-8566 153 1 108-6105 2-15-2 B 5F 2006 2 20 Selection and Quality Assurance of Monitors for Mammogram Diagnosis Noriyuki HASHIMOTO and Kiyotaka HIRA R&D, Visual Products, EIZO NANAO CORPORATION 153 Shimokashiwano,

More information

Display Quality Assurance: Considerations When Establishing a Display QA Program. Mike Silosky, M.S. 8/3/2017

Display Quality Assurance: Considerations When Establishing a Display QA Program. Mike Silosky, M.S. 8/3/2017 Display Quality Assurance: Considerations When Establishing a Display QA Program Mike Silosky, M.S. 8/3/2017 Objectives and Outline Why, Who, What, When, Where? Discuss the resources that may be needed

More information

IAEA RER/9/135 COURSE ON OPTIMIZATION IN COMPUTED TOMOGRAPHY Sofia, Bulgaria, Tube current modulation and dose reduction : How TCM works

IAEA RER/9/135 COURSE ON OPTIMIZATION IN COMPUTED TOMOGRAPHY Sofia, Bulgaria, Tube current modulation and dose reduction : How TCM works IAEA RER/9/135 COURSE ON OPTIMIZATION IN COMPUTED TOMOGRAPHY Sofia, Bulgaria, 2017 Tube current modulation and dose reduction : How TCM works Dean Pekarovič UMC Ljubljana, Institute of Radiology Quality

More information

FDA MAMMOSCAN FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM MAMMOGRAPHY

FDA MAMMOSCAN FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM MAMMOGRAPHY MAMMOGRAPHY MAMMOSCAN FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM Biopsy Attachment џ MAMMOSCAN is a digital scanning technology development taking a special place in ADANI Medical X-ray product line. џ Its

More information

A simple anthropomorphic phantom used to demonstrate the effectiveness of CT dose modulation functions.

A simple anthropomorphic phantom used to demonstrate the effectiveness of CT dose modulation functions. A simple anthropomorphic phantom used to demonstrate the effectiveness of CT dose modulation functions. Lynn Bateman & Peter Hiles North Wales Medical Physics How can we test CT dose modulation? Lynn Bateman

More information

Imaging solutions for any size patient and any size budget.

Imaging solutions for any size patient and any size budget. Imaging solutions for any size patient and any size budget. VetTek V e t e r i n a r y X - R a y S y s t e m s VetTek We See Things Differently & It Shows In Your Image Imagine a veterinary x-ray system

More information

What do we mean by workload? Number of scans Type of scans/mix of scans? Total mas Total dip

What do we mean by workload? Number of scans Type of scans/mix of scans? Total mas Total dip CT & Workload (BIR) David Sutton What do we mean by workload? Number of scans Type of scans/mix of scans? Total mas Total dip What do we do with it? Need to relate workload to total scatter dose in the

More information

Breast Imaging Essentials

Breast Imaging Essentials Breast Imaging Essentials Module 10 Transcript 2016 ASRT. All rights reserved. Breast Imaging Essentials Module 10 Digital Breast Tomosynthesis 1. ASRT Animation 2. Welcome Welcome to Module 10 of Breast

More information

World First Slim Cassette Type Digital Mammo. Upgrade Solution

World First Slim Cassette Type Digital Mammo. Upgrade Solution World First Slim Cassette Type Digital Mammo. Upgrade Solution Still Analog? Slim Cassette Type Digital Mammo. Upgrade Solution Contents 3 P. 5 P. 7 P. 9 P. 11 P. 13 P. 1824C & 2430C, easy digitalization

More information

Equipment Quality Control for Primary Displays June 5, Imaging Physics CancerCare Manitoba

Equipment Quality Control for Primary Displays June 5, Imaging Physics CancerCare Manitoba Equipment Quality Control for Primary Displays June 5, 2018 Imaging Physics CancerCare Manitoba Purpose An equipment quality control (QC) program establishes baseline performance levels, tracks system

More information

What to consider when choosing a mammography display

What to consider when choosing a mammography display What to consider when choosing a mammography display Screen size and resolution In digital breast imaging, the quality of the medical display has a direct impact on the decisions you make. Next to display

More information

The American College of Radiology Mammography Accreditation Program Screen-Film: Frequently Asked Questions (Updated: March 16, 2018)

The American College of Radiology Mammography Accreditation Program Screen-Film: Frequently Asked Questions (Updated: March 16, 2018) The American College of Radiology Mammography Accreditation Program Screen-Film: Frequently Asked Questions (Updated: March 16, 2018) Table of Contents ACR Mammography Accreditation... 1 The 1999 ACR Mammography

More information

Understanding CT image quality

Understanding CT image quality IAEA RER/9/135 COURSE ON OPTIMIZATION IN COMPUTED TOMOGRAPHY Sofia, Bulgaria, 2017 Understanding CT image quality Dean Pekarovič UMC Ljubljana, Institute of Radiology Quality and Safety office Role of

More information

Coronis 5MP Mammo. The standard of care for digital mammography

Coronis 5MP Mammo. The standard of care for digital mammography Coronis 5MP Mammo The standard of care for digital mammography The standard of care For thousands of women every day, details make all the difference. This understanding, along with many years of commitment

More information

Dental Line. High frequency intraoral x-ray unit. radiology ahead

Dental Line. High frequency intraoral x-ray unit. radiology ahead Dental Line High frequency intraoral x-ray unit radiology ahead enhance your operating process Hi-frequency generator, for shorter exposures and reduced dose Wireless X-ray pushbutton, available for ultimate

More information

3/2/2016. Medical Display Performance and Evaluation. Objectives. Outline

3/2/2016. Medical Display Performance and Evaluation. Objectives. Outline Medical Display Performance and Evaluation Mike Silosky, MS University of Colorado, School of Medicine Dept. of Radiology 1 Objectives Review display function, QA metrics, procedures, and guidance provided

More information

Guidance for Quality Assurance of PACS Diagnostic Display Devices

Guidance for Quality Assurance of PACS Diagnostic Display Devices Guidance for Quality Assurance of PACS Diagnostic Display Programme NPFIT DOCUMENT RECORD ID KEY Sub-Prog / PACS Project NPFIT-PAC-DES-0042.06 Prog. Director Max Jones Owner Jerry Norman Version 1.0 Author

More information

Scope: All CT staff technologist

Scope: All CT staff technologist APPROVED BY: Radiology Technical Director Page 1 of 6 Purpose: The QC program assesses relative changes in system performance as determined by the technologist, service engineer, qualified medical physicist,

More information

XR 3000 Real-Time X-Ray Inspection System

XR 3000 Real-Time X-Ray Inspection System XR 3000 Real-Time X-Ray Inspection System Operation Manual Rev A, 4/2002BMS P/N 5050-0511 PACE USA PACE Europe 9893 Brewers Court Sherbourne House, Laurel, MD 20723 Sherbourne Drive USA Tilbrook, Milton

More information

Ride the Lightning. A Review of Radiographic Physics

Ride the Lightning. A Review of Radiographic Physics Ride the Lightning A Review of Radiographic Physics Overview Basic principles of x-ray production Imaging equipment Quality Control Technical Factors Image QA Preparing for the registry The ARRT just wants

More information

Comparison of Measured Values of CTDI and DPL with Standard Reference values of Different CT Scanners for dose Management

Comparison of Measured Values of CTDI and DPL with Standard Reference values of Different CT Scanners for dose Management 2017 IJSRST Volume 3 Issue 1 Print ISSN: 2395-6011 Online ISSN: 2395-602X Themed Section: Science and Technology Comparison of Measured Values of CTDI and DPL with Standard Reference values of Different

More information

The Black Piranha Easy & Fast X-ray Quality Control

The Black Piranha Easy & Fast X-ray Quality Control From Radiation to Information The Black Piranha Easy & Fast X-ray Quality Control The Black Piranha with accessories makes X-ray QA easy and fast. Connection is automatic just plug n play. The Quick Check

More information

Monitor QA Management i model

Monitor QA Management i model Monitor QA Management i model 1/10 Monitor QA Management i model Table of Contents 1. Preface ------------------------------------------------------------------------------------------------------- 3 2.

More information

In recent years, CT technology has undergone profound

In recent years, CT technology has undergone profound PATIENT SAFETY Y.T. Niu M.E. Olszewski Y.X. Zhang Y.F. Liu J.F. Xian Z.C. Wang Experimental Study and Optimization of Scan Parameters That Influence Radiation Dose in Temporal Bone High-Resolution Multidetector

More information

JEFFERSON COLLEGE. Image Intensification & Equipment

JEFFERSON COLLEGE. Image Intensification & Equipment JEFFERSON COLLEGE COURSE SYLLABUS RAD175 Image Intensification & Equipment 3 Credit Hours Revised by: Janet E. Akers BS RT (R)(M) Date: October 3, 2013 Kenny Wilson, Director, Health Occupation Programs

More information

NEMA XR 25 COMPUTED TOMOGRAPHY DOSE CHECK

NEMA XR 25 COMPUTED TOMOGRAPHY DOSE CHECK NEMA XR 25 COMPUTED TOMOGRAPHY DOSE CHECK NEMA Standards Publication XR 25-2010 Computed Tomography Dose Check Published by: National Electrical Manufacturers Association 1300 North 17th Street, Suite

More information

Display Quality Assurance: Recommendations from AAPM TG270 for Tests, Tools, Patterns, and Performance Criteria

Display Quality Assurance: Recommendations from AAPM TG270 for Tests, Tools, Patterns, and Performance Criteria Display Quality Assurance: Recommendations from AAPM TG270 for Tests, Tools, Patterns, and Performance Criteria Nicholas B. Bevins, Ph.D. TG270 Co-chair Display Check 2 1 TG270 Goals Provide an update

More information

Display Quality Assurance: Recommendations from AAPM TG270 for Tests, Tools, Patterns, and Performance Criteria

Display Quality Assurance: Recommendations from AAPM TG270 for Tests, Tools, Patterns, and Performance Criteria Display Quality Assurance: Recommendations from AAPM TG270 for Tests, Tools, Patterns, and Performance Criteria Nicholas B. Bevins, Ph.D. TG270 Co-chair Display Check 2 TG270 Goals Provide an update to

More information

Quality Assurance Implementation at the Roberts Proton Therapy Center. James McDonough 3 August 2013

Quality Assurance Implementation at the Roberts Proton Therapy Center. James McDonough 3 August 2013 Quality Assurance Implementation at the Roberts Proton Therapy Center James McDonough 3 August 2013 1 Roberts Proton Therapy Center Machine configuration and layout 4 gantries, 1 fixed beam line, 1 research

More information

Practix. Mobile Radiography. All a Bucky station needs plus mobility. Practix 400

Practix. Mobile Radiography. All a Bucky station needs plus mobility. Practix 400 Practix 400 Specifications Practix Mobile Radiography All a Bucky station needs plus mobility Practix 400 Practix 400 meets the most demanding requirements of medical care, combining the virtues of mobile

More information

A method for calculating the dose length product from CT DICOM images

A method for calculating the dose length product from CT DICOM images The British Journal of Radiology, 84 (2011), 236 243 A method for calculating the dose length product from CT DICOM images 1 I A TSALAFOUTAS, PhD and 2 S I METALLIDIS, MSc 1 Medical Physics Department,

More information

GUIDELINES ON THE USE OF DISPLAY SCREEN EQUIPMENT

GUIDELINES ON THE USE OF DISPLAY SCREEN EQUIPMENT GUIDELINES ON THE USE OF DISPLAY SCREEN EQUIPMENT 1. Preamble In view of the legislative requirements and growing concern on the health effects related to the use of display screen equipment (DSE), the

More information

User Manual. Portable Dental X-ray System Xray2go Model: MINIX-S. (Rev )

User Manual. Portable Dental X-ray System Xray2go Model: MINIX-S. (Rev ) User Manual Portable Dental X-ray System Xray2go Model: MINIX-S (Rev. 3.6.4) 2018.03 DO NOT OPERATE THIS DEVICE UNTIL YOU HAVE READ THIS MANUAL. This manual contains installation, operation and maintenance

More information

Mammo Tomosynthesis 5MP

Mammo Tomosynthesis 5MP Mammo Tomosynthesis 5MP Display system for digital breast imaging Approved for TOMOSYNTHESIS MAMMOGRAPHY The standard of care For thousands of women every day, details make all the difference. This understanding

More information

Essentials of Digital Imaging

Essentials of Digital Imaging Essentials of Digital Imaging Module 3 Transcript 2016 ASRT. All rights reserved. Essentials of Digital Imaging Module 3 Display 1. ASRT Animation 2. Welcome Welcome to the Essentials of Digital Imaging

More information

CARESTREAM DIRECTVIEW Elite CR System

CARESTREAM DIRECTVIEW Elite CR System CARESTREAM DIRECTVIEW Elite CR System Improve workflow, productivity, and patient throughput. The CARESTREAM DIRECTVIEW Elite CR System is small, easy to install and easy to use. This powerful distributed

More information

Standard. Substitute Test or Procedure. Required Test or. 1 Scan Increment Accuracy. Initially and Annually Initially and Annually

Standard. Substitute Test or Procedure. Required Test or. 1 Scan Increment Accuracy. Initially and Annually Initially and Annually Manufacturer s Recommendations for Alternate Dental CBCT QA Program VaTech: Model PAX i3d, i3d Green, i3d Smart Table 6 Medical Physicist s Computed Tomography QC Survey Required Test or Item Procedure

More information

MPKOMNP. kéï=~ë=çñw== d^ifiblp. j~áåíéå~ååé=fåëíêìåíáçåë=d^ifiblp=`çãñçêí=mirp. båöäáëü. Cover page

MPKOMNP. kéï=~ë=çñw== d^ifiblp. j~áåíéå~ååé=fåëíêìåíáçåë=d^ifiblp=`çãñçêí=mirp. båöäáëü. Cover page kéï=~ë=çñw== MPKOMNP d^ifiblp j~áåíéå~ååé=fåëíêìåíáçåë=d^ifiblp=`çãñçêí=mirp båöäáëü Cover page Sirona Dental Systems GmbH Table of contents 1 Safety... 4 2 Operation notes... 5 3 Auxiliary devices required...

More information

User Manual. Portable Dental X-ray System Xray2go Model: MINIX-S. (Rev v3) 1806

User Manual. Portable Dental X-ray System Xray2go Model: MINIX-S. (Rev v3) 1806 User Manual Portable Dental X-ray System Xray2go Model: MINIX-S (Rev. 3.6.5v3) 1806 DO NOT OPERATE THIS DEVICE UNTIL YOU HAVE READ THIS MANUAL. This manual contains installation, operation and maintenance

More information

CARESTREAM DIRECTVIEW Elite CR System

CARESTREAM DIRECTVIEW Elite CR System CARESTREAM DIRECTVIEW Elite CR System Improve workflow, productivity, and patient throughput. The CARESTREAM DIRECTVIEW Elite CR System is small, easy to install and easy to use. This powerful distributed

More information

Linatron - M9 & M9A. Modular high-energy X-ray source. 2.0 Performance

Linatron - M9 & M9A. Modular high-energy X-ray source. 2.0 Performance The Linatron -M is a modular system. The control console, modulator, and RF unit are common to all model configurations. Only the X-ray head changes to match the application. The Linatron - M is designed

More information

High Frequency X-Ray Technology. Wide Screen Control with Touch-Technology

High Frequency X-Ray Technology. Wide Screen Control with Touch-Technology High Frequency X-Ray Technology Wide Screen Control with Touch-Technology The Q-VISION HF Series TM The Q-VISION Series TM integrates innovative high frequency design along with superb functionality, resulting

More information

2

2 1 2 3 4 5 6 7 1. Bauhs, J. A., Vrieze, T. J., Primak, A. N., Bruesewitz, M. R., & McCollough, C. H. (2008). CT Dosimetry: Comparison of Measurement Techniques and Devices1. Radiographics, 28(1), 245-253.

More information

UArm. Series. Analog or Digital U-Arm System

UArm. Series. Analog or Digital U-Arm System UArm Analog or Digital U-Arm System Series A world class direct digital imaging system with positioning flexibility and ease of use for all procedures. Eighty Years of Excellence UArm Series POSITIONING

More information

DirectView Elite CR System. Improve workflow, productivity, and patient throughput.

DirectView Elite CR System. Improve workflow, productivity, and patient throughput. Improve workflow, productivity, and patient throughput. DirectView DirectView Compact, single cassette CR systems designed to improve workflow, productivity, and patient throughput. The is small, easy

More information

RSNA 2006 November 26 to December 1 Chicago. Guest author for ImPACT Dr. Koos Geleijns, Medical Physicist, Leiden University Medical Center.

RSNA 2006 November 26 to December 1 Chicago. Guest author for ImPACT Dr. Koos Geleijns, Medical Physicist, Leiden University Medical Center. RSNA 2006 November 26 to December 1 Chicago Guest author for ImPACT Dr. Koos Geleijns, Medical Physicist, Leiden University Medical Center. Once again, more than 60,000 participants (including professional

More information

Corrigendum-II. Up to 30 th October 2018 till 17:00 Hrs. 31 st October 2018 till 17:00 Hrs. 2 nd November 2018 till 14:00 Hrs.

Corrigendum-II. Up to 30 th October 2018 till 17:00 Hrs. 31 st October 2018 till 17:00 Hrs. 2 nd November 2018 till 14:00 Hrs. Bihar Medical Services & Infrastructure Corporation Limited 4 th floor State Building Construction Corporation Limited. Hospital Road, Shastri Nagar, Patna 800023, Phone/Fax: +91612 2283287,+ 91612 2283288

More information

Classic. Best Practice QA Solutions for All Rotational Therapies & IMRT For International Use Only. Classic. RIT113 Technical Specifications

Classic. Best Practice QA Solutions for All Rotational Therapies & IMRT For International Use Only. Classic. RIT113 Technical Specifications Technical Specifications Best Practice QA Solutions for All Rotational Therapies & IMRT For International Use Only for detailed Technical Specifications visit www.radimage.com Can I get just the right

More information

Ergonomic design packed with powerful performance MULTIMOBIL 5C. Answers for life.

Ergonomic design packed with powerful performance MULTIMOBIL 5C. Answers for life. Ergonomic design packed with powerful performance MULTIMOBIL 5C Answers for life. MULTIMOBIL 5C Mobile C-arm image intensifier system with high frequency technology Siemens has led the way in development

More information

Hospital Wide. Healthcare Display Solutions DICOM Displays, Large Screen Displays and Projectors

Hospital Wide. Healthcare Display Solutions DICOM Displays, Large Screen Displays and Projectors Hospital Wide Healthcare Display Solutions DICOM Displays, Large Screen Displays and Projectors THE WIDEST RANGE OF DISPLAY SOLUTIONS For Complete Hospital Wide Installations NEC offers a wide range of

More information

VetPro DC Animal Health Dental X-ray System. User Manual Rev. H ECN: P3415

VetPro DC Animal Health Dental X-ray System. User Manual Rev. H ECN: P3415 Animal Health Dental X-ray System User Manual 00-02-1601 Rev. H ECN: P3415 Midmark Corporation 675 Heathrow Drive Lincolnshire, Illinois 60069 U.S.A. Phone: 800-MIDMARK Fax: (847) 415-9801 midmarkanimalhealth.com

More information

Vascular. Development of Trinias FPD-Equipped Angiography System. 1. Introduction. MEDICAL NOW No.73 (2013.2) Yoshiaki Miura

Vascular. Development of Trinias FPD-Equipped Angiography System. 1. Introduction. MEDICAL NOW No.73 (2013.2) Yoshiaki Miura Vascular Development of Trinias FPD-Equipped Angiography System Medical Systems Division, Shimadzu Corporation Yoshiaki Miura 1. Introduction Shimadzu has developed Trinias (one ceiling-mounted type C12

More information

I. FILL IN THE BLANKS

I. FILL IN THE BLANKS 1 BHARAT SEVAK SAMAJ NATIONAL DEVELOPMENT AGENCY, PROMOTED BY GOVERNMENT OF INDIA CENTRAL BOARD OF EXAMINATIONS BSS NATIONAL VOCATIONAL EDUCATION MISSION AHE004-X-RAY TECHNICIAN ONE YEAR EXAMINATION 2011-12

More information

Request for Proposals

Request for Proposals Request for Proposals Reference: ERDFIAI2012-4006A Caring First Ltd, is a limited liability company providing healthcare services. The company is currently commissioning a private new-build hospital complete

More information

MR Accreditation Programs - E. Jackson

MR Accreditation Programs - E. Jackson MRI Accreditation Programs: An Overview of Each and Specifics of One Edward F. Jackson, PhD Department of Imaging Physics 1 Diagnostic - MRI Safety and Accreditation Educational Objectives At the conclusion

More information

Case 1 8/1/2017. Ring Artifact? James M. Kofler, Ph.D. Mayo Clinic Rochester. AAPM Annual Meeting TU-B Artifacts: CT 2

Case 1 8/1/2017. Ring Artifact? James M. Kofler, Ph.D. Mayo Clinic Rochester. AAPM Annual Meeting TU-B Artifacts: CT 2 James M. Kofler, Ph.D. Mayo Clinic Rochester AAPM Annual Meeting 2017 1 Case 1 Ring Artifact? 2 3 1 Star Pattern caused by metal earrings! 4 Same star artifact commonly caused by Dental Amalgam Metal Implants

More information

Radiation Safety System for Stanford Synchrotron Radiation Laboratory*

Radiation Safety System for Stanford Synchrotron Radiation Laboratory* SLAC PUB-8817 April 16, 2001 Radiation Safety System for Stanford Synchrotron Radiation Laboratory* James C. Liu, N. E. Ipe and R. Yotam Stanford Linear Accelerator Center, P. O. Box 4349, Stanford, CA

More information

Creating room. for your visions

Creating room. for your visions BuckyDiagnost Radiography solutions Creating room for your visions 2 What makes a Rad Room simply perfect? Room for all your ideas Efficient, powerful, individual and innovative. This sums up just some

More information

How to use and maintain survey meters and a personal digital dosimeter - Materials for explaining radiation measurement equipment and devices -

How to use and maintain survey meters and a personal digital dosimeter - Materials for explaining radiation measurement equipment and devices - Nuclear Disaster Response Basic Training Session How to use and maintain survey meters and a personal digital dosimeter - Materials for explaining radiation measurement equipment and devices - (English

More information

The Digital Debate. Competing Technologies. Up and Coming 1/5

The Digital Debate. Competing Technologies. Up and Coming 1/5 The Digital Debate axisimagingnews.com/2001/04/the-digital-debate/ The long-awaited transition from film-based radiography to electronic review of medical images has begun. There are a number of factors

More information

GammaSafe System. Exertus Selen Circa Projectors. Emergency Source Retrieval

GammaSafe System. Exertus Selen Circa Projectors. Emergency Source Retrieval GammaSafe System Exertus Selen Circa Projectors Emergency Source Retrieval The Exertus Selen Circa series of Se.75 Projectors consists of the Circa 40 / 80 / 120 models which are all Tungsten shielded

More information

Influence of display quality on radiologists performance in the detection of lung nodules on radiographs

Influence of display quality on radiologists performance in the detection of lung nodules on radiographs The British Journal of Radiology, 80 (2007), 738 743 Influence of display quality on radiologists performance in the detection of lung nodules on radiographs N BULS, MSc, W SHABANA, MD, PhD, P VERBEEK,

More information

G635. APPLIED SCIENCE Working waves ADVANCED GCE. Wednesday 27 January 2010 Morning. Duration: 1 hour 30 minutes

G635. APPLIED SCIENCE Working waves ADVANCED GCE. Wednesday 27 January 2010 Morning. Duration: 1 hour 30 minutes ADVANCED GCE APPLIED SCIENCE Working waves G635 * OCE / 11697* Candidates answer on the Question Paper OCR Supplied Materials: None Other Materials Required: Electronic calculator Ruler (cm/mm) Wednesday

More information

MQSA Quality Control Manual

MQSA Quality Control Manual MQSA Quality Control Manual For The MX50T, MX50N, & CX50N 5 Mega-Pixel LCD Monitors (Document Number: OMI-1035) WIDE Corporation Document No. OMI-1035 Rev. No. 1.2 Page 1 CONTENTS Revision History 4 Documents

More information

Nio. Industry-standard diagnostic display systems

Nio. Industry-standard diagnostic display systems Nio Industry-standard diagnostic display systems Diagnostic confidence in grayscale With the Nio diagnostic display system, Barco brings dependable diagnostic imaging to its true potential. Presenting

More information

CHAPTER 4: HIGH ENERGY X-RAY GENERATORS: LINEAR ACCELERATORS. Jason Matney, MS, PhD

CHAPTER 4: HIGH ENERGY X-RAY GENERATORS: LINEAR ACCELERATORS. Jason Matney, MS, PhD CHAPTER 4: HIGH ENERGY X-RAY GENERATORS: LINEAR ACCELERATORS Jason Matney, MS, PhD Objectives Medical electron linear accelerators (often shortened to LINAC) The Basics Power Supply Magnetron/Klystron

More information

CARESTREAM VITA/VITA LE/VITA SE CR System Long Length Imaging User Guide

CARESTREAM VITA/VITA LE/VITA SE CR System Long Length Imaging User Guide CARESTREAM VITA/VITA LE/VITA SE CR System Long Length Imaging User Guide Use of the Guide Carestream CR Systems are designed to meet international safety and performance standards. Personnel operating

More information

The PEFP 20-MeV Proton Linear Accelerator

The PEFP 20-MeV Proton Linear Accelerator Journal of the Korean Physical Society, Vol. 52, No. 3, March 2008, pp. 721726 Review Articles The PEFP 20-MeV Proton Linear Accelerator Y. S. Cho, H. J. Kwon, J. H. Jang, H. S. Kim, K. T. Seol, D. I.

More information

The Alice Silicon Pixel Detector (SPD) Peter Chochula for the Alice Pixel Collaboration

The Alice Silicon Pixel Detector (SPD) Peter Chochula for the Alice Pixel Collaboration The Alice Silicon Pixel Detector (SPD) Peter Chochula for the Alice Pixel Collaboration The Alice Pixel Detector R 1 =3.9 cm R 2 =7.6 cm Main Physics Goal Heavy Flavour Physics D 0 K π+ 15 days Pb-Pb data

More information

Non-Destructive Examination Benches and Analysis Laboratories in support to the Experimental Irradiation Process in the Future Jules Horowitz MTR

Non-Destructive Examination Benches and Analysis Laboratories in support to the Experimental Irradiation Process in the Future Jules Horowitz MTR Non-Destructive Examination Benches and Analysis Laboratories in support to the Experimental Irradiation Process in the Future Jules Horowitz MTR D. Parrat 1, P. Kotiluoto 2, T. Jäppinen 2, C. Roure 1,

More information

NEW ACHIEVEMENTS IN THE FIELD OF DEVELOPMENT OF MOBILE X-RAY TV- SYSTEM

NEW ACHIEVEMENTS IN THE FIELD OF DEVELOPMENT OF MOBILE X-RAY TV- SYSTEM NEW ACHIEVEMENTS IN THE FIELD OF DEVELOPMENT OF MOBILE X-RAY TV- SYSTEM A.V. Kovalev 1, A.V. Fedotov 1, Yu.A. Khnykov 1, A.A. Kovalev 1 1 JSC Spectrum RII, Moscow, Russia Abstract: Among various directions

More information

arxiv: v1 [physics.ins-det] 1 Nov 2015

arxiv: v1 [physics.ins-det] 1 Nov 2015 DPF2015-288 November 3, 2015 The CMS Beam Halo Monitor Detector System arxiv:1511.00264v1 [physics.ins-det] 1 Nov 2015 Kelly Stifter On behalf of the CMS collaboration University of Minnesota, Minneapolis,

More information

Is image manipulation necessary to interpret digital mammographic images efficiently?

Is image manipulation necessary to interpret digital mammographic images efficiently? Loughborough University Institutional Repository Is image manipulation necessary to interpret digital mammographic images efficiently? This item was submitted to Loughborough University's Institutional

More information

"CHARM Access and operational Procedure

CHARM Access and operational Procedure EDMS NO. REV. VALIDITY 1390490 0.1 DRAFT CERN CH1211 Geneva 23 Switzerland REFERENCE XXXX Date: 2015-09-04 OPERATING PROCEDURE "CHARM Access and operational Procedure ABSTRACT : The aim of this procedure

More information

WM2013 Conference, February 24 28, 2013, Phoenix, Arizona USA

WM2013 Conference, February 24 28, 2013, Phoenix, Arizona USA ABSTRACT Unique Regulatory Approach for Licensing the Port Hope Remediation Project in Canada 13315 M. Kostova, D. Howard and P. Elder Directorate of Nuclear Cycle and Facilities Regulation Canadian Nuclear

More information

Revised Textbook Edition Digital Radiography & PACS

Revised Textbook Edition Digital Radiography & PACS Fleitz Continuing Education Jeana Fleitz, M.E.D., RT(R)(M) The X-Ray Lady 6511 Glenridge Park Place, Suite 6 Louisville, KY 40222 Telephone (502) 425-0651 Fax (502) 327-7921 Web address www.x-raylady.com

More information

Radiography and Mammography Viewing Systems

Radiography and Mammography Viewing Systems Radiography and Mammography Viewing Systems For assistance contact your authorized RADX Dealer, or call S&S Technology at 800-231-1747 E-mail salessupport@radx.com www.ssxray.com Mammography Digital Viewing

More information

Data Collection Using APEX3. March 30, Chemical Crystallography Laboratory

Data Collection Using APEX3. March 30, Chemical Crystallography Laboratory Data Collection Using APEX3 Page 1 of 10 Data Collection Using APEX3 March 30, 2017 Chemical Crystallography Laboratory Author: Douglas R. Powell Data Collection Using APEX3 Page 2 of 10 Distribution Douglas

More information

35 cm x 35 cm) technique visit each colliery on a. scheme radiographic technique and film quality have

35 cm x 35 cm) technique visit each colliery on a. scheme radiographic technique and film quality have British Journal of Industrial Medicine, 1973, 3, 365-374 A comparison of conventional and grid techniques for chest radiography in field surveys J. S. WASHINGTON, J. A. DICK, M. JACOBSEN1, and W. M. PRENTICE2

More information

Initial Experience With Automatic Image Transmission to an Intensive Care Unit Using Picture Archiving and Communications System Technology

Initial Experience With Automatic Image Transmission to an Intensive Care Unit Using Picture Archiving and Communications System Technology Initial Experience With Automatic Image Transmission to an Intensive Care Unit Using Picture Archiving and Communications System Technology Carl E. Ravin This report describes a refinement of the interaction

More information

GUARD BAGAŽO TIKRINIMO VARTAI

GUARD BAGAŽO TIKRINIMO VARTAI GUARD BAGAŽO TIKRINIMO VARTAI Eil. Nr. 1. XJ5030 Aprašymas Suitable for all areas where small suitcases and packages are allowed to be inserted in The highest penetration level to the steel in its class

More information

Journal of Diagnostic Radiography and Imaging

Journal of Diagnostic Radiography and Imaging Journal of Diagnostic Radiography and Imaging Instructions to Authors Original articles, editorial correspondence, case reports, notices, suggestions for review articles, product reviews and so forth should

More information

Physics Requirements for the CXI Ion Time-of-Flight

Physics Requirements for the CXI Ion Time-of-Flight PHYSICS REQUIREMENT DOCUMENT (PRD) Doc. No. SP-391-000-30 R0 LUSI SUB-SYSTEM CXI Physics Requirements for the CXI Ion Time-of-Flight Sébastien Boutet CXI Scientist, Author Paul Montanez CXI Lead Engineer

More information

WiFi RF Measurements and Compliance with the FCC RF Safety Limit

WiFi RF Measurements and Compliance with the FCC RF Safety Limit Pinnacle Telecom Group Professional and Technical Services WiFi RF Measurements and Compliance with the FCC RF Safety Limit Ridgewood Public Schools April 14, 2015 14 Ridgedale Avenue - Suite 209 Cedar

More information