R&S Power Viewer Plus. Software Manual. Version 9.0. Printed in Germany. Manual

Size: px
Start display at page:

Download "R&S Power Viewer Plus. Software Manual. Version 9.0. Printed in Germany. Manual"

Transcription

1 Software R&S Power Viewer Plus Version 9.0 Printed in Germany 1

2 Dear Customer, R&S is a registered trademark of Rohde & Schwarz GmbH & Co. KG. R&S is referred to as R&S throughout this manual. Mac and Mac OS are trademarks of Apple Inc., registered in the U.S. and other countries. Microsoft and Windows are trademarks of Microsoft Corporation in the United States and/or other countries. Trade names are trademarks of their respective owners Rohde & Schwarz GmbH & Co. KG Munich, Germany Printed in Germany Subject to change Data without tolerance limits is not binding. 2

3 1 Contents Contents 1 Contents 3 2 Overview 7 3 Key Software Features 8 4 Power Sensor Technologies Thermal Power Sensors CW Power Sensors Multi-Path Diode Power Sensors Average Power Sensors Wideband Diode Power Sensors Sensor Modules (OTA) Uncertainty Calculation Measurements at 10 dbm Measurements at 50 dbm The Influence of Mismatch Software Installation System Requirements Installation on Windows-Based Systems Installation on Mac OS X Sensor Firmware Requirements Supported R&S NRPZ Sensors Running Multiple Instances Connecting Sensors to the PC Using Multiple Sensors Configuring the Application Data Format Drawing Performance Hardcopy Settings Timeout-Related Settings USB-Related Settings USB Device Tree Debug Options Setting the Application Colors First Steps Numeric Entry Fields Using USB Extension Hardware R&S NRP-Z3 Active USB Adapter R&S NRP-Z4 Passive USB Adapter R&S NRP-Z5 Sensor Hub Third-Party Products... The Menu Bar

4 10 Contents File Sensor Info Measurement Data Processing Window Help... The Toolbar Selecting a Sensor General Measurement Settings Hardcopy Features Print Report Copy to Clipboard Save Graphics to File The Message Log Channel Assignment Measurements Overview Continuous Power Measurement Settings Numerical Data View Negative Power Readings Accuracy of Peak Power Measurements Relative Measurements Analog Meter Trend Chart Histogram Display Q-Q-Plot OTA Continuous Power Measurement Settings Trace Measurements Measurement Settings Custom Settings Graphical Trace View Context Menu Trace Representations View Modes Lines Reference Trace Using Markers Pulse Width Measurements Pulse Rise-Time Measurements Zooming Time Zoom Level Zoom Automatic Pulse Measurement

5 Contents Measurement Results Common Measurement Tasks OTA Trace Measurement Settings Statistics Settings Graphical Data View PDF Mode PDF Background Information CDF Mode CCDF Mode Timeslot Mode Settings Graphical Data View Multi-Channel Power Measurements Channel Configuration Measurement Settings Mathematical Expressions OTA Multi-Channel Measurement Settings Script-Based Measurement Script Syntax Printing to the Script Output Window Printing to the Message Window Waiting Periods Timers Power Conversion File Operations Defining Devices Communication Timeout Suppressing SCPI Log Messages SCPI Commands and Queries Simulated Queries for Use With NRP-Z sensors Reading Results from New Generation Sensors Waiting for Measurement Completion Checking for Errors Displaying Scalar Results Displaying Array Data Numerically Displaying Array Data as a Bar Graph Displaying Array Data as a Trace Displaying Array Data as a Time Series Plot Displaying Text Sending Data to Processing Panels

6 Contents Examples Triggered Average Power Measurement Burst Power Measurement Buffered-Mode Measurements Connecting to the R&S NRP2 Base Unit Connecting to the R&S NRT Base Unit Data Processing Panels 23.1 The Data Log Settings The Context Menu The Time Line Indicator Zooming Limit Monitoring Settings Configuring the Server Client Connections Statistical Analysis Histogram Display Q-Q-Plot The Context Menu Analysis-Panel Settings Gauges Panel Analog Meter The Context Menu Gauges-Panel Settings Updating NRP and NRP-Z Sensor Firmware Recovering from a Failed Update Using S-Parameter Profiles 26 VXI PnP Programming Guide Sensor Resource Strings Numeric Results Opening and Closing the Sensor Connection Multiple Sensors Error Handling Zeroing Identifying a Sensor (*IDN?) Continuous Average Power Measurement Trace Measurements Single-Shot Events Peak Trace Data Automatic Pulse Measurement Customer Support Appendix Command Line Options Setting the Application Style

7 2 Overview Overview The new R&S NRP power sensors from Rohde & Schwarz represent the latest in power measurement technology. They offer all the functionality of conventional power meters, and more, within the small housing of a power sensor. The USB interface on an R&S NRP sensor enables operation with an R&S NRP power meter, or with a PC running under either Microsoft Windows, Mac OS X, or Linux. is an easy-to-use, feature-packed software package that offers capabilities beyond those of a regular power meter. It simplifies measurement tasks, such as average-power, timeslot, statistics and trace measurements. In addition, up to 16 sensors can be utilized for measuring average power simultaneously. Results, such as the reflection coefficient or gain, can be computed from the measured values. Particularly the capabilities for use with a desktop or laptop PC make an R&S NRP sensor an ideal and cost-effective solution for lab testing or for automated systems. The rugged design is suitable for use in the field for performing such tasks as servicing antenna systems. This manual describes the installation and use of the Power Viewer Plus software. This application is available free of charge from the Rohde & Schwarz website. As a prerequisite it requires the installation of the R&S NRP Toolkit. This toolkit contains drivers and small applications. It is also available free of charge from the Rohde & Schwarz website. To enable integration of the sensor into custom ATE systems, a versatile and powerful VXI PnP driver is available for Microsoft Windows, Mac OS X, and Linux-based systems. The new R&S NRP power sensor family also supports SCPI remote control commands via USB-TCM. Coding examples can be found in the application note 1GP69. 7

8 3 Key Software Features Key Software Features Power Viewer Plus is powerful PC software that simplifies many measurement tasks. This software is part of the R&S NRP Toolkit and is available free of charge. The following overview lists some of the key features that Power Viewer Plus offers. Measuring the average and peak powers and viewing the numeric results. Optionally, adding an analog bar display, a trend chart or statistical analysis. Fig. 3.1: Viewing measurement results. Viewing the RF power envelope down to a resolution of 5 ns/div; measuring pulse parameters with automatic pulse analysis; using markers, and measuring within time gates. Fig. 3.2: Pulse measurement. 8

9 Key Software Features Measuring the average power of up to 16 sensors, and optionally computing results from the measured values. Fig. 3.3: Measuring average power. Viewing the average and peak powers for up to 16 consecutive timeslots in a bar chart. Fig. 3.4: Viewing average and peak powers. 9

10 Key Software Features Performing statistical CCDF, CDF, or PDF analysis of the envelope power. Fig. 3.5: Statistical analysis. Recording up to 4 channels of any measured data to memory and/or to a file. Fig. 3.6: Recording measured data. 10

11 Key Software Features Performing 4-channel statistical analysis on any measured data. Fig. 3.7: Four-channel statistical analysis. Configuring 16-channel limit monitoring for any measured data; optionally, sending limit violations to a remote host via a TCP/IP server. Fig. 3.8: Limit monitoring. 11

12 Key Software Features Creating custom measurements in script mode. Fig. 3.9: Script mode for custom measurements. Configurable analog and digital gauges. Fig. 3.10: Two gauges showing measurement values. Please note that some of the features listed above are only supported by certain R&S NRP power sensors. For example, thermal sensors do not provide statistical signal analysis or trace measurements. 12

13 4 Power Sensor Technologies Power Sensor Technologies Rohde & Schwarz offers a large choice of USB power sensors that use different technologies and cover a wide range of frequencies and power levels. This chapter briefly outlines the differences between the sensor technologies and indicates which sensor would best fit certain measurement tasks. All R&S NRP power sensors are standalone instruments that contain the detector, the analog circuitry, and the digital signal processing in a single housing. The entire instrument is fully characterized during the production process, which eliminates the need for later calibration using a reference power source. Zeroing is generally only required for measuring low power levels, in which case the zeroing offset, noise, and drift stated in the specification sheet contribute to the overall accuracy. 4.1 Thermal Power Sensors Thermal sensors use a load resistor for converting the RF power into heat. The temperature difference between this resistor and the surrounding area is measured by thermocouples. The resulting DC voltage is proportional to the RF power. coaxial RF feeder co-planar line thermal transducer 1 mm tapered transmisssion Fig : Detector design in thermal power sensor. The Rohde & Schwarz thermal power sensors can be used from DC up to their specified upper frequency limit. The dynamic range is typically in the order of 55 db and starts at power levels of around 35 dbm. Thermal power sensors provide the highest accuracy and linearity of all power sensors on the market. Their measurements are not influenced by the modulation or harmonics, and the results always represent the average signal power. However, the nature of the underlying sensor technology limits the dynamic range. Furthermore, the measurement speed is generally slower than that of diode sensors. Thermal sensors cannot measure the envelope of an RF signal. 4.2 CW Power Sensors CW sensors are simple diode sensors that contain a half-wave or full-wave rectifier as the detector element. At power levels below -20 dbm, the diode characteristic provides an almost linear relationship between the detector output voltage and the RF power. This power range is referred to as the square-law region of the detector diode. CW sensors typically use the diode at power levels beyond the square-law region, and the software must compensate for the resulting 13

14 Power Sensor Technologies non-linearities. With CW signals, this compensation is possible, and the sensor provides correct readings of the average RF power. Modulated or pulsed signals, as well as signals containing harmonics, may lead to large measurement errors at levels that exceed the detector's square-law region. Due to these limitations, the Rohde & Schwarz NRP product range does not include CW sensors. 4.3 Multi-Path Diode Power Sensors The Rohde & Schwarz multi-path diode sensors use up to three independent full-wave diode detectors. These detectors, along with their analog and digital signal processing, are referred to as paths. Each path is designed for operation in a separate power range, with a 6 db overlap between the paths. Fig : Multi-path diode sensor design. The data from all paths is processed in parallel. For each power level within the specified sensor limits, at least one path operates within the detector's square-law region and delivers an output signal that is proportional to the RF energy. The sensor software automatically determines the path that best fits the incident RF power. As a result, these sensors exhibit little sensitivity to modulation and harmonics. The sensors always measure the average signal power at a performance level that is close to that of thermal sensors. Due to the ease of use and excellent performance offered by these sensors, Rohde & Schwarz calls these devices R&S Universal Power Sensors. The universal power sensors' dynamic range and measurement speed are higher than can be achieved with thermal sensors. For most signals and measurement tasks, universal power sensors are ideal devices. These sensors also allow measurement of the RF envelope, but the sampling rate of about 150 khz must be considered as a limiting factor in such cases. 4.4 Average Power Sensors The Rohde & Schwarz Average Power Sensors also use three diode paths. Unlike the universal power sensors, the detector design used for average power sensors allows an RF frequency as low as 9 khz. Due to this detector design, the bandwidth is lower. Consequently, this sensor is only intended for performing average power measurements. 4.5 Wideband Diode Power Sensors Wideband diode sensors use a single full-wave diode detector and operate it across the entire useful power range. The detector's bandwidth is much higher than with CW sensors, and the sample rate is in the order of 80 MHz. 14

15 Power Sensor Technologies Fig : Wideband power sensor design. Similar to CW sensors, the wideband diode sensor's digital signal processing circuitry compensates the non-linear diode characteristic in real-time. Due to the wider bandwidth and fast sampling rate, this is even possible for fast amplitude changes (AM) of the RF envelope. Wideband diode sensors are ideal when the RF envelope should be measured, e.g. for the analysis of pulsed signals. Additionally, these devices can measure the signal statistics, such as the PDF, CDF, CCDF, and average power for modulated signals. The following chart shows the relationship between power levels and applications that generally fit a wideband diode sensor. Average Power, Av. Power, fmod < BW fmod > BW Peak Power Envelope (Averaged) Envelope (Real Time) -60 dbm -20 dbm +23 dbm Fig : Wideband power sensor applications. At power levels below 20 dbm (square-law region), these sensors exhibit little sensitivity to modulation and harmonics. Average power measurements are possible down to a level of about 60 dbm. For higher power levels, care must be taken when the RF envelope is amplitude modulated at frequencies that exceed the detector's analog bandwidth. In such cases, it is no longer possible to compensate the RF envelope in real-time, and measurement errors in the order of several percent may occur. The wider bandwidth used by these sensors generally implies a higher noise floor. Average power measurements overcome this issue by using averaging techniques. When taking single-shot measurements, however, the higher noise floor must be considered. This is especially the case with peak power measurements. Please see the chapter 14.4, "Continuous Power Measurements Accuracy of Peak Power Measurements" in this document for more details. It must also be noted that triggering is always a real-time process that is based on samples that have not yet been subject to averaging. As a result, power levels in the order of 20 dbm or higher are required when 15

16 Power Sensor Technologies using the sensor's internal trigger feature. Decreasing the sensor's bandwidth decreases the noise floor and, therefore, also decreases the lower trigger-level limit. 4.6 Sensor Modules (OTA) The sensor modules provide reliable over the air (OTA) measurement of RF power on up to three antennas simultaneously. The sensor modules do not contain internal RF detector diodes. Instead they provide an interface for connecting up to three antennas with integrated power detector. Fig : OTA antenna. The power sensor is designed for measuring power in high frequency bands, used in modern high-performance wireless system standard, e.g. IEEE802.11ad and 5G. This sensor module and the specific antennas have been developed to provide measurements of radiated RF power. 5 Uncertainty Calculation This chapter briefly explains how to calculate the measurement uncertainty based on the figures provided in the sensor's specifications. The data sheet lists the absolute uncertainty for power measurements in db depending on the power level and frequency. Other contributors, such as zero offset or noise, are provided in watts and can be converted into db using the following equation. e=10 db log P P P This equation uses P as the power level of interest and ΔP as the relative error. The result is the error e in db. Uncertainties are statistical measures, and they must be added by summing up the squared uncertainties and then calculating the square root: U = U 1 U 2 U This equation can be used for uncertainties in logarithmic scale (db) or in percent (%). Uncertainties are commonly provided in db, but the following equation permits conversion into percent: 16

17 Uncertainty Calculation U db U %=100% db 1 To gain a simple approximation, the following formula can be used: U % 10 ln 10 U db=23 U db 5.1 Measurements at 10 dbm The power level range from 10 dbm to 0 dbm is widely used. Therefore, our first example here calculates the absolute uncertainty for the R&S NRP-Z11 when measuring a CW signal at 2 GHz and at a power level of 10 dbm. The temperature shall be 30 ºC. All values marked with an arrow ( ) are taken from the R&S NRPZxx Power Sensor Specifications that are available on the Rohde & Schwarz website. Power level in W Used path1 100 µw 2 Uncertainty for absolute power measurements Zero Offset2 Zero Drift 47 nw 3 nw Measurement noise Multiplier for 40 ms integration Time is sqrt(10.24s/t) 6.3 nw Total expanded uncertainty x 16 = nw db N/A db db db db % The example shows that the influence of zero offset and drift is negligible. Consequently, zeroing of the sensor is not required when performing practical measurement tasks. The integration time can be set to a very short value of 40 ms. This means that an averaging count of one, combined with two chopper cycles and a measurement window (aperture) of 20 ms, is sufficient. The total integration time is twice the aperture time multiplied by the averaging filter count. 1 2 Automatic path selection, see Transition regions in data sheet. Without zeroing 17

18 5.2 Uncertainty Calculation Measurements at 50 dbm This example calculates the absolute uncertainty for the R&S NRPZ11 when used for measuring a CW signal at 2 GHz and at a very low power level of 50 dbm. The temperature shall be 30 ºC. All values marked with an arrow ( ) are taken from the R&S NRPZxx Power Sensor Specifications that are available from the Rohde & Schwarz website. Power level in W Used path 10 nw 1 Uncertainty for absolute power measurements Zero offset after zeroing Zero drift after zeroing Measurement noise Multiplier for 1.28 s integration time is sqrt(10.24s/t) x pw 35 pw db db 65 pw = 182 pw Total expanded uncertainty db db db % After zeroing, the absolute accuracy is 0.12 db when using an integration time of 1.28 s. This integration time can be achieved with an average filter count of 32 and a measurement window of 20 ms. Further improvement of the uncertainty is possible by increasing the averaging filter count. The total integration time is twice the aperture time multiplied by the averaging filter count. 5.3 The Influence of Mismatch Power sensors are always calibrated to measure the power of the incident RF wave. This means that the sensor corrects the reading for the internal losses and reflections. As a result, different power sensors that were connected to an ideal 50 ohm source would all show exactly the same result. In the real world, however, neither the power sensor nor the source match an impedance of 50 ohms exactly. The reflection that is caused by the power sensor itself is specified by the standing wave ratio (SWR), which is typically around 1.2. This means that a small portion of the RF wave is reflected back towards the source as a return wave. An ideal source would absorb this return wave entirely. Since the power sensor is calibrated to measure the incident wave and compensates for its own reflections, the reading is correct. Real signal sources are not ideal either. They also reflect a portion of the return wave back to the power sensor. This portion adds to the incident RF wave and influences the measurement result. The uncertainty calculations in the previous chapter did not include the error caused by mismatch. 18

19 Uncertainty Calculation The following equation shows the minimum and maximum possible incident power based on the reflection coefficient of the source and the load: P GZ0 1 r G r L P i 2 PGZ0 1 r G r L 2 PGZ0: Power from signal source Pi: Incident power to power sensor rg: Generator reflection coefficient rl: Load reflection coefficient Depending on the phase angle, the incident power varies between the left and right term of the equation. The following equations can approximate the maximum relative deviation εmax between the source power PGZ0 and the incident power Pi: max % 200 % r G r L max db 8.7 db r G r L for max 20 % for max 1 db Uncertainty calculations use statistical figures instead of the εmax errors from the equations above. The following equation shows the relationship between the expanded uncertainty (k = 2) and the error. max db U db=2 2 This shows that the expanded uncertainty used for the uncertainty calculation is higher than the maximum error. Data sheets often express the impedance matching of a device as a standing wave ratio (SWR). The relationship between the SWR and the reflection coefficient is expressed by the following equations: s= 1 r L 1 r L r L= s 1 s 1 The example below demonstrates the influence of mismatch caused by a signal source that is directly connected to a power sensor: Load: Source: R&S NRPZ11 R&S SMBV100A SWR = 1.2 SWR = 1.6 rl = 0.09 rg = 0.23 UdB = ± 8.7 db = ± 0.25 db 19

20 6 Software Installation Software Installation The following section outlines the process for installing Power Viewer Plus on various platforms. 6.1 System Requirements The following hardware and software prerequisites must be fulfilled for running the Power Viewer Plus application. Hardware requirements Standard desktop PC or laptop, or an Intel-based Apple Mac 1024 x 768 screen resolution USB 2.0 interface 100 Mbit LAN R&S NRP-Z3, R&S NRP-Z4, or R&S NRP-Z5 adapter (required for R&S NRP-Z power sensors) Operating systems (choice of) Microsoft Windows 7, Microsoft Windows 8, Microsoft Windows 10 Mac OS X 10.6 or later 32/64-Bit Linux distribution with kernel 2.6.x (e.g. Ubuntu 10.4 LTS x86, 11.4 x86) Software packages R&S NRP Toolkit V4.10 or later. The toolkit provides the required USB and low level drivers for all R&S NRP-Z power sensors. R&S VISA or National Instruments VISA required for all new R&S NRP power sensors. 20

21 6.2 Software Installation Installation on Windows-Based Systems Since fall 2015 the Power Viewer Plus is not part of the R&S NRP Toolkit anymore. Instead it is now a separate installer package. Please note that the installation of the R&S NRP Toolkit is required before installing the Power Viewer Plus application. 1. Disconnect all NRP power sensors from the PC. 2. Install or update the R&S NRP Toolkit on your PC. 3. Start the installer Fig : installer. 4. Please follow the installer steps until the process completed. After the installation has completed, the sensors can be connected to the PC. If the USB drivers were updated or newly installed, recognizing the sensor may take more time when it is plugged in for the very first time. The installer writes data to the following location on your system: Application Files %COMMONPROGRAMFILES(x86)%\Rohde-Schwarz\Power Viewer Plus Example Scripts %PUBLIC%\Documents\Rohde-Schwarz\Power Viewer Plus 21

22 6.3 Software Installation Installation on Mac OS X In contrast to the Windows R&S NRP Toolkit, the Mac OS X toolkit installer contains the following components: RsNrpLib.framework (low-level driver) RsNrpz.framework (VXI PnP driver) HTML help files for the VXI PnP driver Power Viewer Plus and documentation Example programs for use with the VXI PnP driver The toolkit comes as a.pkg installer. Double-click this file from any folder on your desktop. Please follow all instructions provided by the installer. After successful installation, the Power Viewer Plus application can be started from the Rohde-Schwarz / Power Viewer folder that was created in the Mac OS X application directory. Fig : The "Power Viewer" folder. The folder Scripts contains examples for the Power Viewer Plus script-based measurement mode. Please see the chapter Script-Based Measurement in this book for details. The current.pkg installer cleans up any existing Power Viewer Plus installation from the previous disk image based installers. In addition to the files placed in the /Application folder the installer also adds two frameworks to the /Frameworks folder. The NrpLib framework is the low level driver for all USB communication to the NRP-Z power sensors. It can also be used to communicatate in legacy mode with the newer NRP sensors. The RsNrpZ driver is the VXI PnP driver. It provides a C interface and can be used for own applications. Power Viewer Plus communicates through the NrpLib driver with all NRP-Z sensors and it uses VISA for all new NRP power sensors. 22

23 6.4 Software Installation Sensor Firmware Requirements Power Viewer Plus may require newer firmware versions on certain power sensors. Please see the firmware update section in this manual for more details on updating the sensor firmware. The latest firmware files are available free of charge from the Rohde & Schwarz website. NRP-Z Power Sensors R&S NRP-Z8x R&S NRP-Z1x R&S NRP-Z2x R&S NRP-Z3x R&S NRP-Z5x 1.20 or later 4.08 or later 4.08 or later 4.08 or later 4.08 or later New generation NRP Power Sensors R&S NRPxxS(N) R&S NRPxxA(N) R&S NRPxxT(N) R&S NRPM or later or later or later or later 23

24 6.5 Software Installation Supported R&S NRPZ Sensors The following table provides an overview of the sensors that are supported in Power Viewer Plus. The vendor ID for all R&S NRP sensors is 0x0AAD. Sensor USB ID NRP-Z11 0x0C Cont Supported Measurement Trace Timeslot Statistics NRP-Z21 0x03 NRP-Z211 0xA6 NRP-Z22 0x13 NRP-Z221 0xA7 NRP-Z23 0x14 NRP-Z24 0x15 NRP-Z31 0x2C NRP-Z41 0x96 NRP-Z51 0x16 NRP-Z52 0x17 NRP-Z55 0x18 NRP-Z56 0x19 NRP-Z57 0x70 NRP-Z58 0xA8 NRP-Z91 0x21 NRP-Z81 0x23 NRP-Z85 0x83 NRP-Z86 0x95 NRP-Z27 0x2F NRP-Z28 0x51 NRP-Z37 0x2D NRP-Z92 0x62 NRP-Z98 0x52 NRPC33 0xB6 NRPC40 0x8F NRPC50 0x90 NRPC33-B1 0xC2 NRPC40-B1 0xC3 NRPC50-B1 0xC4 FSH-Z1 0x0B FSH-Z18 0x1A 24

25 Software Installation The following tables list the new generation NRP power sensors introduced by the end of These sensors are available with USB interface and LAN interface. The LAN sensors use the N (Network) suffix in their device name. The USB interface supports both, legacy USB communication using the existing VXI PnP drivers as well as USB-TMC with VISA. 3-Path Diode Sensors Sensor USB ID NRP8S 0x00E2 Cont Supported Measurement Trace Timeslot Statistics NRP8SN 0x0137 NRP18S 0x0138 NRP18SN 0x0139 NRP33S 0x0145 NRP33SN 0x0146 NRP40S 0x015F NRP40SN 0x0160 NRP50S 0x0161 NRP50SN 0x0162 NRP33SN-V 0x0168 Sensor NRP18T USB ID 0x0150 Cont NRP18TN 0x0151 NRP33T 0x0152 NRP33TN 0x0153 NRP40T 0x0154 NRP40TN 0x0155 NRP50T 0x0156 NRP50TN 0x0157 Thermal Sensors Supported Measurement Trace Timeslot Statistics Average Sensors for EMC Applications Sensor USB ID NRP6A 0x0178 Cont NRP6AN 0x0179 NRP18A 0x014E NRP18AN 0x014F Supported Measurement Trace Timeslot Statistics 25

26 Software Installation Sensor Modules (OTA) Sensor USB ID NRPM3 0x0195 Cont Supported Measurement Trace Timeslot Statistics If no sensor is detected, Power Viewer Plus automatically activates a simulated sensor called NRP-Z Running Multiple Instances Since fall 2015 the low level drivers can handle multiple applications using the drivers simultaneously. The drivers grant exclusive access to the application opening a specific power sensor. Trying to open the same sensor from another application generates a 'sensor in use' error message. Starting with V8.0 Power Viewer Plus supports this new driver feature. Power Viewer Plus checks to see if any other instance is already running on the system. If so, a warning message appears. Fig : Warning message indicating that an application is already running. 26

27 7 Connecting Sensors to the PC Connecting Sensors to the PC Please see your R&S NRP power sensor's manual for information on how to put the sensor into operation. Follow these instructions to prevent damage to the sensor, particularly if you are putting it into operation for the first time. The following section provides additional information that is related to the USB interface or to operating multiple sensors simultaneously. 7.1 Using Multiple Sensors If multiple sensors need to be connected to a single computer, check to ensure that the overall current requirements for operating all sensors can be met. Each single sensor draws between 300 ma and 500 ma, depending on the sensor type. Example: The R&S NRP-Z81 sensor is rated at up to 500 ma supply current. Using four sensors simultaneously on one hub requires a total current of at least two amperes. Many consumer hubs cannot provide this current over a long period of time, even if they are rated for this value. 27

28 Connecting Sensors to the PC 7.2 Using USB Extension Hardware R&S NRP-Z3 Active USB Adapter The figure shows the configuration with the R&S NRP-Z3 active USB adapter, which also makes it possible to feed in a trigger signal for the timeslot and trace modes. The order in which the cables are connected is not critical. Fig : Configuration with the active USB adapter R&S NRP-Z4 Passive USB Adapter The figure below is a schematic of the measurement setup. The order in which the cables are connected is not critical. Fig : Configuration with the passive USB adapter 28

29 7.2.3 Connecting Sensors to the PC R&S NRP-Z5 Sensor Hub The R&S NRP-Z5 sensor hub allows up to four power sensors to be operated on one PC. It combines the following functions: 4-port USB 2.0 hub with Multi-TT architecture Power supply Through-wired trigger bus Trigger input and trigger output via BNC sockets It is possible to cascade several R&S NRP-Z5 sensor hubs by connecting the R&S Instrument port of an R&S NRP-Z5 to one of the sensor ports of another R&S NRP-Z5. However, external triggering and the use of the Trigger Master function are then not possible. Instead, it is recommended that you connect all R&S NRP-Z5 hubs individually to the USB host or to an interposed USB hub. Then feed the external trigger signal to all R&S NRP-Z5 hubs via their trigger inputs. Fig : Connecting the USB hub. 29

30 7.2.4 Connecting Sensors to the PC Third-Party Products This section lists devices that are manufactured by other vendors and have been used successfully with R&S NRP-Zxx power sensors. Rohde & Schwarz cannot provide a continuous guarantee that these products will work with R&S NRP-Zxx sensors, because technical changes or newer versions of these products are not retested: Icron ( offers the USB Ranger 110/410 products that are compliant with the USB 1.1 specification and can be used to cover a distance of up to 100 meters by using standard Cat 5 UTP cabling. Icron ( offers the USB Ranger 2224 product that is compliant with the USB 2.0 specification and can be used to cover a distance of up to 500 meters by using a multi-mode optical fiber. When large distances between the control PC and the sensor(s) are required, a combination of the USB Ranger 2224 and the R&S NRP-Z5 has demonstrated reliable operation. Fig : Setup with 100 m optical fibre. Digi ( makes the AnywhereUSB Network-enabled USB hub. This product is used to access a USB device over a TCP/IP network. 30

31 8 Configuring the Application Configuring the Application Power Viewer Plus provides a settings dialog that can be accessed by selecting Configure Options from the main menu. This dialog box is structured using separate tabs for drawing operations, timeouts, hard copies, USB, and debugging. 8.1 Data Format Fig : Data format settings..csv Files The list separator is the character that separates columns in a.csv file. The default character is the semicolon. Negative Power Readings The power sensor zeroing procedure sets the zero power reference level to the current A/D converter reading. Due to the Gaussian nature of sample noise readings below this reference value may occur. The sensor returns these readings as negative power values in Watts. Mathematically it is not possible to convert negative numbers to a logarithmic scale. This settings allows the user to select between two methods for the conversion. By default the software uses the absolute reading for the calculation of the logarithmic value. This method produces naturally looking noise but undershoots following a falling edge may appear as overshoots or ringing. The other clipping method sets all power readings that fall below the sensor lower measurement range to exactly this power level. Noise does not look quite as natural anymore but it is clearly visible which numbers fall below the zero reference level. 31

32 8.2 Configuring the Application Drawing Performance The drawing performance can be adjusted to accommodate slow PCs. Activating these features lowers CPU load or adds additional idle time. Fig : Drawing performance settings. Disable Transparency Effects Lowers CPU consumption by avoiding semi-transparent drawing operations. Transparent drawing is used, for example, for the grid lines in the trace mode, because it makes it possible to see trace points that fall exactly onto a grid line. Number of Video Points for Traces Set to 500 by default, this number provides a good compromise between measurement speed and resolution. The higher the number of video points, the higher the CPU load and acquisition time. On lowperformance PCs, it may be desirable to lower this number. Set Additional Video Update Time Adds idle time between two measurements. This reduces CPU load and provides resources to other applications. The default idle time between two measurements is in the order of 100 ms. Disable LCD Background Replaces the blue color gradient used in all LCD displays with a simple gray color. This option is useful for increasing the display contrast and for reducing CPU usage. Disable Trace Anti-Aliasing Turns off anti-aliasing in all trace and statistics measurement panels. Turning anti-aliasing off speeds up drawing operations and reduces CPU usage. 32

33 8.3 Configuring the Application Hardcopy Settings Power Viewer Plus creates print reports or copies measurement results to the system clipboard. This greatly simplifies documentation tasks. Please see the "Hardcopy Features" and "Copy to Clipboard" sections for additional details. Fig : Hardcopy settings. Do Not Invert Colors By default, the application uses printer-friendly colors when copying data to the system clipboard. This feature can be turned off by choosing not to invert the screen colors. Use Custom Size... The Copy to Clipboard function always creates a bitmap of a fixed size. This simplifies documentation tasks, since any display resolution may be used, and you do not need to specifically rescale captured images. 33

34 8.4 Configuring the Application Timeout-Related Settings The Timeout tab is shown below and is mainly used for connections across USB extenders or USB-to-LAN interfaces. These devices often introduce large turnaround times that need to be taken care of. Fig : Timeout settings. Set USB Communication By default, this value is set internally to 5 seconds. Connections across the Internet (e.g. using the Digi AnywhereUSB device, may require values of up to 15 seconds. Set Measurement Timeout This function is used internally to set the time between the point when a measurement is initiated and the maximum waiting time for the result. Normally, the internal time of 5 seconds should be sufficient. However, very slow connections may make it necessary to increase this time. 34

35 8.5 Configuring the Application USB-Related Settings The USB tab is shown below and is used for altering USB interface related settings on Microsoft Windows-based operating systems. Fig : USB settings. Long Distance Mode This mode is only available for Windows-based operating systems. It reduces the number of simultaneous read processes, which lowers USB resource allocation in the operation systems dramatically. AnywhereUSB connections, for example, require activation of the Long Distance Connection mode. Selective Suspend Mode Windows can turn off unused USB hubs or unused ports on USB hubs. This is the default setting on most fresh installations. In some situations this mechanism does not work properly and can leave a hub turned off or in an undefined state. Disabling selective suspend turns this power saving mechanism off for all hubs and subsequently requires a system reboot. The selective suspend should only be turned off if USB devices do not get activated after they were plugged into a USB port. 35

36 8.6 Configuring the Application USB Device Tree The Sensors tab is shown below and is used for analyzing the USB device tree on Microsoft Windows-based operating systems. The tree is mainly intended for diagnostic purposes because some sensor / hub configurations have shown poor performance. These configurations are highlighted with a yellow exclamation mark. Fig : USB device tree. The following USB configurations should be avoided: NRP-Z sensors that are directly connected to Single-TT USB hubs. NRP-Z sensors that are directly connected to bus powered USB hubs. NRP-Z sensors that are directly connected to the PC's USB port (root hub). Rohde & Schwarz generally recommends to operate NRP-Z power sensors with Multi-TT USB hubs. The hub should be equipped with a power supply that is rated for the total current of all connected sensors. Each individual hub port should be capable of delivering up to 500 ma to the USB device. 36

37 8.7 Configuring the Application Debug Options The debug options are mainly intended for debugging purposes. The following list contains debug options that may be used with certain measurements: contav.fastmode=1 multi.fastmode=1 This option increases the measurement rate in the continuous power or multi-channel measurement mode and is explained in more detail in the related section in this manual. trace.thick=1 This option draws bold traces in the trace measurement instead of using thin lines. Combined with a low trace point count, this setting is useful for outdoor service applications. trace.meastime=1 When this option is enabled, the Power Viewer software displays the total trace measurement time in the trace window. This time is the period starting at the initiation of the measurement and ending when all data is received by the host. tsl.peak=0 When this setting is disabled, the Power Viewer software omits peak readings in the timeslot measurement mode. Please note that peak measurements are subject to higher noise content, and the readings are only useful for levels greater than 5 dbm. contav.cmd=<cmd_list> trace.cmd=<cmd_list> multi.cmd<ch>=<cmd_list> If set accordingly, Power Viewer Plus appends the SCPI commands provided in the command list <cmd_list> at the end of the measurement configuration. The command list can either be a single SCPI command or a list of commands separated by a semicolon (;). For the multi-channel measurement mode, the channel number must also be provided. Using these commands is risky, because it may leave the sensor and user interface in different states. trace.noinfo=1 This option suppresses the Measure information box in the trace window. 37

38 8.8 Configuring the Application Setting the Application Colors Power Viewer Plus provides a color settings dialog box that can be accessed by selecting Configure Colors from the main menu. All color changes are applied immediately in all application windows. Therefore, it is possible to open windows, such as the trace measurement, and observe the color changes directly. Fig : Color settings dialog box. Preset The entire application can be set to one of the predefined color schemes or to a user-defined color set. Save As This button saves the user color scheme to a file. Load This button loads a color scheme from a file and replaces the current user color set. Brightness The brightness control changes the brightness for the entire application. Changing the brightness setting does not affect any of the user's color definitions. Contrast The contrast control changes the contrast setting for the entire application. Increasing the contrast reduces the brightness of background colors and increases the brightness of foreground colors. Changing the contrast does not affect any of the user's color definitions. Color tiles The small colored tiles represent the color of the individual elements. 38

39 Configuring the Application One of these tiles can be selected for editing using the HSV color controls. HSV color control The application uses the HSV color model to define the application colors. This color model uses hue, saturation and value instead of red, green and blue components. The hue represents the angle on the color wheel between 0 and 360. This value is meaningless for non-chromatic colors, such as gray. The saturation is set in the range between 0 and 255; it defines how strong the color is. Grayish colors have very low saturation, whereas strong colors use high saturation values. The value defines the lightness; this parameter is also set between 0 and 255. The brighter the color is, the higher the value is. 39

40 9 First Steps First Steps The main application window is divided into three major sections. The measurement window area The settings panel on the right side The upper and lower toolbars Only one measurement can be active at a time, but it is possible to tile multiple measurement windows and switch from one to the other. All measurement windows have the same sensor assigned. If the settings panel is enabled, it is always located on the right side. Its content changes with the currently activated measurement window. Measurement window Measurement-related settings Measurement selection Sensor selection Data log running indicator Measurement running indicator General parameters Fig. 9.1: The main application window. 40

41 9.1 First Steps Numeric Entry Fields The Power Viewer Software uses custom entry fields for most numeric data. These entry fields are closely related to regular text entry boxes that allow the user to enter any text. Custom entry fields differ from regular entry fields in that they format and validate the user input when the enter button is pressed, or the field looses the focus. Numbers are entered with or without their SI prefix. The SI prefix can be one of the following letters: G M k m n u p f Giga Mega kilo milli nano micro pico femto The entry fields also provide a "tooltip" help function that shows the minimum and maximum permissible input value. Additionally, a step size can be defined to increase or decrease the value when the mouse wheel is turned. Fig Example of the help tooltip function. The step value is defined by the current cursor position within the entry field. 41

42 First Steps 9.2 The Menu Bar File Fig : File settings. File Open Project Loads a previously saved configuration. These settings affect all measurements and fully restore the state of the entire application, including window positions. File Save Project As Saves the configuration of the entire application to a file. This file may later be used to restore a measurement configuration. Measurement data is not saved as part of the settings file. File Exit Aborts all running measurements, disconnects from the power sensor, and subsequently ends the application. All settings are saved Sensor Fig : Sensor menu. Sensor Scan for Sensors Starts the process of detecting available R&S NRP USB sensors. Activating this menu item repopulates the sensor selection control. If no sensor is detected, a sensor simulation function (R&S NRP-Z00) will be available. Scanning is also performed automatically when new sensors are connected to the PC or sensors are removed. The automatic scanning capabilities are inhibited while measurements are running, and they resume after all measurements have been stopped. 42

43 First Steps Sensor List Used USB Sensors Opens a window that shows which application is using an R&S NRP-Z Power Sensor in legacy mode. Fig : Sensor usage by application. The window only lists NRP-Z Power Sensors that are in use by the low level drivers in legacy mode. All newer R&S NRP Power Sensors are controlled via the VISA interface and are not listed here. Sensor Channel Assignment Displays a panel that allows the user to assign alias names to each sensor. This simplifies working with multiple sensors. Alias names are only valid within Power Viewer Plus. Sensor LAN Sensor Management Displays a panel that allows the user to manage connected R&S NRP LAN sensors. These sensors must be manually configured for use with the application. Sensor LAN Sensor Configuration Displays a panel that allows the user to configure the network settings of the currently selected LAN sensor. Sensor Update Firmware Opens the sensor firmware update dialog. Please see the firmware update section for detailed information about the update process. Sensor S-Parameter Opens the s-parameter profile editing dialog. This dialog is used to download s-parameter sets of two-port devices to the sensor. 43

44 9.2.3 First Steps Info Fig : Info menu. Info Get Properties Displays a panel that contains a set of important sensor properties, such as the frequency and power range, as well as the firmware version. Info Get Extended Information Reads all available information from the selected sensor. This menu option is only available when no measurements are running. Info Run Self Test Performs a self-test on the selected sensor and returns the results as text message. The detector's noise level is measured as part of the sensor self-test routines. This only works when no RF signal is applied to the sensor's input while the test is running Measurement Fig : Measurement menu. Measurement Start Starts the measurement in the window that is currently active. This button is disabled when another measurement window is already running. Please note that some sensors may not support all measurement modes. In such cases, the start button is disabled, even if the measurement window is open and no measurement is running. 44

45 First Steps Measurement Stop Stops the currently active measurement. To add a level of protection, a measurement can only be stopped when its window is active and selected. This prevents unintentional stopping of a measurement. Measurement Reset to Defaults Reset the currently active measurement settings back to default parameters. Measurement Continuous Opens the panel for the continuous measurement mode. In this mode, the power sensors perform asynchronous measurements on the signal power over a definable time interval (aperture time). Measurement Trace Opens the panel for the Trace measurement mode. The panel displays the envelope power versus time. Measurement Statistics Opens the panel for the Statistics measurement mode. In this mode, the signals CDF, CCDF, or PDF can be measured. Measurement Timeslot Opens the panel for the Timeslot measurement mode. This mode measures the average and peak power of a definable number of successive timeslots. Measurement Multi Channel Opens a panel that can display continuous power readings for up to 16 sensors. Measurement Scripting Opens the scripting window. The scripting measurement module is used to execute SCPI scripts or to define custom measurements. Please see the scripting section for additional details Data Processing Fig : Data Processing menu. This menu contains functions that do not perform measurements but receive measured values for evaluation. Most data processing functions must be started manually after the measurement has begun. Typically, the data processing automatically finishes when the measurement stops. Data Processing Data Log The data log captures up to four different values over a definable period. The data is captured in two ways: The default method stores the readings in up to data bins (memory). This data can be viewed 45

46 First Steps and exported to a file. The second method writes the captured values directly to a file while the measurement is running. There is no limitation on the number of data points when writing to a file. Data Processing Analysis The analysis window evaluates up to four measurands statistically. In the default configuration, the Power Viewer Plus creates a histogram view in each analysis channel. Data Processing Gauges The gauges window displays up to four numerical values on a large display. Each gauge can be switched between a digital and an analog display. The gauges panel is useful if values need to be read from a greater distance. Data Processing Limit Monitor The limit monitor module compares up to 16 measurands against upper and lower warning and error limits. It can send limit violations to a remote host via its internal TCP/IP server Window Fig.9.2.7: Window menu. Window Copy Graphics to Clipboard Sends the content of the currently activated measurement window to the system clipboard. This option is only available for measurements that display their results in graphical form (such as trace, statistics, timeslot and data log measurements). The copy-to-clipboard function simplifies documentation tasks, because the graphics can simply be pasted into other applications. Please see chapter 10.2 "Copy to Clipboard" for a detailed description. Window Save Graphics to File This function is similar to the above menu option, but it creates a.png file on the user's desktop that contains the screen shot. Window Print Report Creates a printout of the measurement that is currently activated. The printout is a one-page document that contains the measurement and all important sensor settings. Colors are inverted where necessary to avoid a black background. This option is only available for measurements that display their result as graphics (such as trace, statistics, timeslot, and data log measurements). Please see chapter 10.1 "Print Report" for additional details. 46

47 First Steps Window Save Measurement Data Saves measurement data from the currently active window to a.csv file. This extension stands for comma-separated values. Files in this format list data in columns that are separated by a single comma. This option is only available for measurements such as the trace, statistics or data log measurements. Comma-separated value lists can easily be imported into most applications, such as Microsoft Excel or Open Office. Window Show Tool Bar Enables or disables the upper tool bar. Disabling the tool bar is useful if the application shall be used with screen resolutions of 800 x 600 pixels or less. Window Toggle Settings Panel Enables or disables the settings panel on the right side of the application window. Removing the settings panel frees some display space and can be useful if the screen resolution is limited, e.g. 640x480 pixels. Window Message Log The log window captures all internal messages generated by the application. By default only information and error messages are captured. Starting the application with the command line argument --debug also enables log messages. The log message window is automatically activated in case of an error. This menu item is used to bring up the log window manually Help Fig : Help menu option. Help About This Software Displays program information, such as the software version number and licensing information. 47

48 9.3 First Steps The Toolbar The application provides a main toolbar that is located at the top of the main program window. This toolbar hosts shortcuts to commonly used functions and measurements. Stop measurement Start measurement Toggle settings Save measurement data Save graphics to file Copy graphics to clipboard Print measurement report Save project file Load project file Fig : The main toolbar. 9.4 Selecting a Sensor A second toolbar is located at the lower border. It is used for sensor selection and for general settings. This toolbar is divided into two sections: The left side provides measurement and data-log running indicators as well as a control for sensor selection. If no sensor was detected during the last USB bus scan, only the sensor simulation function (NRP-Z00) is available. This simulation capability can be used for basic demonstration and testing of the program's functionality. Fig : Second toolbar with sensor selection. The application remembers the last sensor selection and tries to reuse this device if it was detected during a USB scan. If the last sensor that was used is no longer detected, the first detected sensor is used instead. Please note that changing the sensor type may affect measurement settings. Power Viewer Plus double-checks measurement settings before a measurement is started and corrects values if necessary. 48

49 9.5 First Steps General Measurement Settings The right toolbar section provides general settings for defining the signal frequency, level offset, or gamma correction settings, or for selecting the use of an S-parameter set. Fig : Second toolbar with general settings. Please note that the general settings are applicable to all measurement functions, except for multi-channel measurements. The multi-channel measurement function provides individual settings for each sensor. Signal Frequency This frequency is used to correct measurement results in various ways. It is essential that the current carrier frequency be set. Otherwise, non-linearities or temperature dependencies considerably greater than those stated in the data sheet can arise. Level Offset The offset accounts for external losses. If, for example, a 60 db directional coupler is used to sense power from a DVB-T transmitter, the coupling loss can be used as the offset. Power Viewer Plus sets up the sensor accordingly and displays the corrected power measurements. Corrections Dialog The Corrections dialog contains advanced correction options that are less frequently used. The dialog settings apply to all power measurements, except for multi-channel mode and the script mode. Fig : Corrections dialog. 49

50 First Steps S-Parameters (embedding) This check box activates one of the S-parameter correction data sets stored in the sensor. S-parameter correction is used to compensate for a component (attenuator, directional coupler) connected ahead of the sensor by means of its S-parameter data set. Using S-parameters instead of a fixed offset allows more precise measurements, because the interaction between the sensor and the component can be taken into account. The S-parameter controls may be grayed out if the s-parameter state is locked in the sensor. Please see the S-Parameters section in this manual for detailed information on how to lock or unlock S-parameters. Further information about practical applications in which S-parameter correction is useful can be found in the application note 1GP70 on the Rohde&Schwarz web site. Gamma Correction The gamma correction value sets the source's complex reflection coefficient. A magnitude value of zero corresponds to an ideally matched source, and a value of one to total reflection. The phase angle can be set between and degrees. 50

51 10 Hardcopy Features Hardcopy Features Power Viewer Plus provides two features that greatly simplify documentation tasks. With a simple mouse click, it is possible to create a print report for the trace, statistics, or data log panel. Additionally, the current graphics can be copied to the system clipboard and pasted into any other application Print Report The print button in the toolbar automatically creates a one-page measurement report from the current data. Colors are inverted for printer friendliness. The picture below shows an example of the generated form. Fig : Example of a printed report. On Linux-based systems, the printer selection dialog offers printing directly to a.pdf file, in which case a PDF document is created without the use of any third-party software. 51

52 10.2 Hardcopy Features Copy to Clipboard The copy-to-clipboard function creates a bitmap of fixed size from the current measurement and subsequently places the bitmap into the system clipboard. By default, colors are inverted, and a resolution of 800 x 600 pixels is used. If this is not acceptable, these parameters can be changed in the settings dialog box. The figure below shows a captured measurement at a resolution of 800 x 600 pixels. Fig : Graphics copied to clipboard Save Graphics to File The save graphics to file function creates a bitmap of fixed size from the current measurement and subsequently creates a.png file on the desktop or in the user's home directory. By default, colors are inverted, and a resolution of 800 x 600 pixels is used. If this is not acceptable, these parameters can be changed in the settings dialog box. 52

53 11 The Message Log The Message Log The Message Log window can be activated from the Window menu. This window lists text messages, warnings, and errors that are generated by the application or by the VXI PnP driver. Fig. 11.1: The message log window. Clear Clears all of the window's content. Copy Copies the window content as text to the system clipboard. This text may then be pasted into other applications, such as clients. Dealing with unexpected behavior If the program or sensor displays unexpected behavior, it is advisable to forward a detailed problem description along with system information (such as the sensor type, serial number and firmware version string) to the R&S customer support: customersupport@rohde-schwarz.com 53

54 12 Channel Assignment Channel Assignment Power Viewer Plus maintains a list of alias names that can be assigned to sensors. Each R&S NRP sensor can have an individual name assigned to it, which is displayed throughout the application as an additional piece of information. If no alias name is set for a sensor, the application only displays its type and serial number in all sensor selection controls. The Channel Assignment dialog uses the placeholder <name> if no alias name has been defined. Double-clicking the name field allows the user to edit the entry. Fig. 12.1: The channel assignment dialog. Using alias names simplifies measurement tasks that involve multiple sensors. For example, calculating an amplifier gain requires measurement of the input and output power. Alias names, such as input or output, may be assigned to the sensors connected to these ports. Sensors that are detected during a scan are indicated by illuminated light bulbs, whereas unavailable devices appear as gray bulbs. 54

55 13 Measurements Overview Measurements Overview The following list briefly describes all measurements available in Power Viewer Plus. The availability of the individual measurements depends on the selected sensor. Not all sensors support all measurement modes, e.g. a thermal power sensor does not offer a trace measurement. Continuous Average Continuous average power measurement with one single sensor. OTA (Continuous Average) Continuous average power measurement with all antenna elements connected to the OTA sensor. Timeslot Average power measurement in consecutive timeslots. Statistics Measure the power distribution such as CCDF or PDF. Trace Measure the RF power envelope. OTA (Trace) Measure the RF power envelope with all antennas connected to the OTA sensor. Multi Channel Measure the average power with multiple sensors simultaneously. OTA (Multi Channel) Measure the average power with multiple OTA sensors simultaneously. Script Use Java Script for custom measurement tasks. Please see the following chapters for a detailed description of these measurements. 55

56 14 Measurements Overview Continuous Power Measurement In this mode, the measurement signal's average power is measured asynchronously within definable time intervals. This time interval is referred to as the sampling window or aperture time. The width of a sampling window is preset to a length that is optimal for the selected sensor, but it can be changed to other values. The measurements are performed with chopper stabilization to obtain more accurate results with reduced noise and zero offset. Therefore, a measurement is always performed over two sampling windows, with the polarity of the detector output signal being reversed for the second window. Taking the difference between the output signals minimizes the video path's influence on noise and on zero drift. When the averaging function is activated, the averaging factor determines how often the described measurement cycle is repeated Settings Fig : Settings for continuous power measurements. Zero This option is only available while the measurement is running. It starts the zeroing sequence for the current sensor. For this purpose, the RF signal must be switched off, or the sensor must be disconnected from the signal source. The sensor automatically detects the presence of any significant power, which causes zeroing to be aborted and an error to be generated. The LED next to the zeroing button changes it's color depending on the current zeroing state. Grey Zeroing was not performed since the measurement got started. Blue Zeroing is in progress. The measurement results are not updated during this time. Green Zeroing finished successfully. Red Zeroing failed. In this case the LED initially flashes and 56

57 Continuous Power Measurement then turns red continuously. The zeroing process may take more then 8 seconds to complete and varies with the sensor model. Generally, it is possible to run the sensor zeroing with a small signal (such as broadband noise) applied to the sensor. This makes it possible to compensate for this signal in later measurements. Averaging The averaging mode can be set to either Auto or. In manual mode, the sensor uses an averaging factor that is set by the user between 1 (no averaging) and In auto mode, the sensor determines the optimum average filter count based on a resolution of 0.01 db. Count This is the number of measured values that have to be averaged to generate the measurement results. Raising the averaging factor reduces fluctuation in measured values and lengthens the amount of time required to complete the measurement. Duty Cycle The duty cycle can be set as a percentage when pulse-modulated signals are corrected. With correction activated, the sensor calculates pulse power from the duty cycle and the average power. Aperture Time (Sampling Window) The sampling window (aperture time) is the time period that is used to form one sample. The Power Viewer Plus software automatically uses a default window that best fits the selected sensor. Wider sampling windows may be required if the measurement result exhibits fluctuations due to modulation. In this case, it is beneficial to set the sampling window length to a value equal to the modulation period. Smoothing The smoothing filter is a steep-slope digital lowpass filter used to suppress result variations due to modulation. Smoothing should be activated to reduce result variations due to modulation when the size of the sampling window cannot, or should not, be set to exactly equal the modulation period. If the selected sampling window is 5 to 9 times larger than a modulation period, the display variations are usually sufficiently reduced. With smoothing deactivated, 300 to 3000 periods are required to obtain the same effect. When smoothing is deactivated, the sampling values are considered to be equivalent, and they are averaged in a sampling window, which means that the measuring instrument acts as an integrator. As described above, optimum suppression of result variations is obtained when the size of the sampling window exactly equals the modulation period. Otherwise, modulation can have a considerable influence, even if the sampling window is much larger than the modulation period. The response can be improved considerably by weighting samples, which is equivalent to video filtering. This is exactly what happens when smoothing is activated. Since the smoothing filter increases the sensor's inherent noise by approx. 20%, it should always be deactivated when it is not required. Debug settings Debug settings are entered in the debug options field in the program settings dialog. Open this dialog from the program menu by selecting Configure Options Debug. contav.fastmode=0 1 57

58 Continuous Power Measurement In normal measurement mode, the sensor measurements are initiated at a lower rate. All measurements are forwarded to the data processing panels. Only one measurement within a time period of 100 ms is sent to the numeric display. This ensures a convenient display update rate while simultaneously allowing for fast settling of the moving average filter. In applications in which CPU consumption is not critical, it is possible to increase the measurement rate to capture more sensor readings in data processing panels, such as the data log Numerical Data View Power Viewer Plus displays average power readings in numerical and graphical form. The numerical display shows the main measurement as well as additional information, such as the averaging count, measurement mode, and measurement window (aperture). Icons Averaging filter count Sampling window (aperture) Relative indicator Peak power Peak-to-average ratio Average power Fig : The numerical data view. The units can be switched between dbm, dbw, dbµv, and W. A context menu can be activated by right-clicking in the display area. This menu sets the display resolution to 0.001, 0.01, or 0.1. The display resolution setting does not affect the measurement itself, it only limits the number of visible digits. Both linear and logarithmic power readings are average measurements based on the current average filter setting. If the sensor provides peak power data, these readings are displayed below the main reading. 58

59 Continuous Power Measurement In the upper left display corner, a set of icons informs the user about the measurement state. Duty-cycle correction This icon is displayed in the display's upper left corner when duty-cycle correction is active. S-parameter device This icon is displayed in the display's upper left corner when Sparameter correction is active. Offset This icon is displayed in the display's upper left corner when a level offset is set. Over-range warning This icon is displayed in the display's upper left corner when the power level approaches the sensor destruction limit Negative Power Readings When a noisy signal is measured close to the power level at which the sensor was zeroed, negative power readings may occur in the linear scale. The logarithmic scale ignores the polarity and always uses the linear power reading's absolute value. In rare cases, the reading may be exactly zero. Since it is impossible to convert zero into a logarithmic scale, a value that is 20 db below the minimum specified measurement level is used instead Accuracy of Peak Power Measurements Please note that care must be taken to ensure the accuracy of peak power readings. Chopper stabilization or averaging techniques cannot be used with peak measurements. As a result, the measurement noise level is substantially higher. The following section discusses the influence of the higher noise level in great detail. Noise needs to be looked at as a statistical process which can be described by a normal distribution. The figure below shows the shape of this distribution. The dark area marks values that are less than one standard deviation away from the mean value µ. For a normal distribution, 68.2 % of all values fall into this range. Fig : Normal distribution. 59

60 Continuous Power Measurement Peak power measurements record the maximum power that was detected within the observation period. The longer this observation period is, the more likely it is that a higher power value will occur. The following discussion explains the influence of noise on the peak-power measurement accuracy. This example uses the technical data for the R&S NRP-Z81 wideband power sensor. An observation time of 500 µs at full bandwidth (a 12.5 ns sampling time) results in samples that are to be evaluated. The sample noise specified for the NRP-Z81 sensor running at full bandwidth is typically 2 µw (for 2 standard deviations). This means that 34.1 % of all values are less than 1 µw above the average value. The probability for one sample out of the given values is 1 / = The associated σ value for this probability for the normal distribution is about 4.5 according to the following equation: x = 1 2 e 1 2 x 2 The peak sampling noise that needs to be expected based on the normal distribution is, therefore: 1 µw * 4.5 = 4.5 µw. Due to the detector's non-linear response, additional noise multiplying factors need to be considered depending on the signal level applied to the sensor. These factors are provided in the sensor's technical specifications. For a signal level of 1 mw at room temperature, the noise multiplication factor is 3.8. The overall noise power that needs to be expected calculates to: 4.5 µw * 3.8 = 17.1 µw. The total error based on the above assumption for a 0 dbm signal would then be: 17.1 µw 100 % =1.71 % 1 mw Power Viewer Plus removes the numeric peak and peak-to-average ratio for peak readings below 3 dbm Relative Measurements Relative measurements display the current reading relative to a previously set reference power level. This measurement mode is useful when the measurement task requires analysis of a power reading's stability or drift. When the measurement mode is changed to relative, Power Viewer Plus saves the current reading as a reference value. Subsequent mode changes then no longer alter this reference power level, and a new level is set using the REF button. 60

61 14.6 Continuous Power Measurement Analog Meter Power Viewer Plus also displays power readings graphically as an analog bar graph. The bar graph shows the average power level as a blue arrow and the peak power level as a red arrow. Each arrow holds the maximum value for a time period of about 5 seconds. Maximum readings are indicated by the smaller and darker arrows. Average power reading with 5 seconds max. hold Peak power reading Fig : Analog meter in absolute measurement mode. The analog meter can also be used in the relative measurement mode, in which case it displays the change in the measured value relative to a previously set value. A context menu can be invoked by a right mouse click and used to change the display range between a 20 db, 10 db, 2 db, and 1 db full scale. Fig : Analog meter in relative measurement mode. Please note that some sensors only provide average power measurements. In such cases, the red arrow is not visible. 61

62 14.7 Continuous Power Measurement Trend Chart For analysis of power readings over time, Power Viewer Plus provides an additional statistics panel. This panel supports multiple views, such as a trend chart, a histogram, or the Q-Q-plot. The view mode is selected via the panel's context menu. The trend display shows past measured values over time. New values are appended on the right side of the chart, and they move to the left side with time. Average power Peak power Fig : Trend chart in absolute measurement mode. Analogously to the analog meter, the blue trace indicates average power, whereas the red trace is used for peak power readings. The trend chart does not provide a time scale, because the time varies depending on the filter and measurement-window settings. Please note that Power Viewer Plus sets the sensor to a moving average filter mode. This ensures a constant measurement rate regardless of the averaging filter count or sampling window length. As a result, fast level changes do not appear as a step in the trend chart. Instead, they exhibit a smooth transition from one level to the other. Fig : Step response for a measurement taken with an NRP-Z81 on a 3GPP signal with the manual averaging filter set to 256 and the aperture set to 10 ms. 62

63 14.8 Continuous Power Measurement Histogram Display The histogram sorts the measured values into categories (data bins) that are evenly distributed between the minimum and maximum readings. The results are displayed as a bar chart in which the height of a bar indicates how many measurements fall into each category. The number of samples that are used for evaluation can be 250, 1000 or Statistics Count Average power Standard deviation Fig : Histogram display. The minimum (Min) and maximum (Max) power readings are displayed in the upper left corner. The count (Cnt) indicates how many readings were accumulated for the analysis. The count remains at a constant value as soon as the set number of readings has been reached. The average (Avg) power of all accumulated readings and the sample standard deviation (s) is displayed in the panel's upper right corner. The following formulas are used to calculate these two parameters: 1 x = N N i =1 xi N 1 s= xi x 2 N 1 i=1 The terms "L95" and "U95" are used for the 95 % confidence intervals for the average power level: L95 x 2 s N U95 x 2 s N 63

64 14.9 Continuous Power Measurement Q-Q-Plot The Q-Q-Plot (Quantile-Quantile-Plot) is a graphical method for comparing two probability distributions. The Power Viewer Plus software provides a normal probability plot that compares the probability distribution of the measured values against an ideal normal distribution. If the measured values are distributed normally, all graph points are located on a straight line. Departures from this straight line indicate that the normal distribution model is a poor fit for the distribution of the measured values. Fig : Q-Q-Plot display. Two additional red lines mark the 95 % confidence band. If all graph points are located within this band, the measured values have a normal distribution at a 95 % confidence level. The diagram is vertically and horizontally scaled to σ. Therefore, each grid line represents a step size of 1 σ. The Q-Normal-Plot is used as a graphical test for normal distribution. 64

65 15 OTA Continuous Power Measurement OTA Continuous Power Measurement The OTA continuous average power measurement is very similar to the regular continuous average power measurement. The main difference is that OTA sensors deliver three individual power readings from their individual antenna elements. The OTA measurement therefore displays three continuous average power readings simultaneously. This chapter only describes the differences to the regular continuous average power measurement Settings Fig : Settings for OTA continuous power measurements. Channels The three channels of the OTA sensor can be individually enabled or disabled. Channels without an antenna connection must be disabled to avoid error messages. 65

66 16 Trace Measurements Trace Measurements In the trace mode, the envelope power can be recorded as a function of time. This is done by sampling power over a time interval that can be specified by the user. The power values are assigned to a number of pixels (video points) that each contain data, such as the average power, the maximum power and a randomly sampled value Measurement Settings Fig : Zeroing controls. Zero This option is only available while the measurement is running. It starts the zeroing sequence for the current sensor. For this purpose, the RF signal must be switched off, or the sensor must be disconnected from the signal source. The sensor automatically detects the presence of any significant power, which causes zeroing to be aborted and an error to be generated. The LED next to the zeroing button changes it's color depending on the current zeroing state. Grey Zeroing was not performed since the measurement got started. Blue Zeroing is in progress. The measurement results are not updated during this time. Green Zeroing finished successfully. Red Zeroing failed. In this case the LED initially flashes and then turns red continuously. The zeroing process may take more then 8 seconds to complete and varies with the sensor model. Generally, it is possible to run the sensor zeroing with a small signal (such as broadband noise) applied to the sensor. This makes it possible to compensate for this signal in later measurements. 66

67 Trace Measurements The power scale is defined by two parameters: the reference level and the level step per division. Both values can be changed in steps by pressing the plus and minus buttons on the settings panel. In addition, values can be entered manually in a logarithmic or linear scale. Fig : Setting the vertical scale. Reference level (Ref) Specifies the upper limit of the trace view area. This setting only affects the graphical data representation in the application. It has no influence on the measurement or sensor configuration. /div Sets the scaling of the level axis. Zooming operations keep the reference level constant and adjust the lower level accordingly. Auto Set Tries to adjust the level scaling, trigger level and timing to match the applied signal. All other parameters are set back to defaults. If the auto set process fails, all settings are left untouched. Fig : Setting the horizontal scale. Power Viewer Plus uses a fixed grid of 10 divisions for the time scale. The time resolution is set per division with the lowest possible value being 5 ns/div. Please note that not all sensors support the same time resolution. Additional information can be found in the sensor data sheets. The two buttons (plus and minus) increase or decrease the time per division value in fixed steps. These steps follow the order or vice versa. The appropriate unit is automatically added to the numeric value. Trigger Position This setting defines the trigger point's position within the trace view area. The arrow buttons move the trace back or forth by one division. The trace position setting allows the user to view the signal at times before the physical trigger point. Please note that this time interval (pre-trigger) depends on the sensor hardware used. Power Viewer Plus automatically corrects invalid ranges for the current sensor. 67

68 Trace Measurements Delay The trigger delay creates a delayed trigger point that is not identical to the physical trigger point. Power Viewer Plus uses the delayed trigger point as the zero position for the time axis. The delay setting can be used to compensate for signal delays caused by long cabling and external trigger source. Fig : Setting the averaging count. Averaging Count This value sets the number of traces to be evaluated to form one measurement result. Averaging reduces the noise level of the average trace but increases measurement time. Changing the averaging count does not have any effect on traces that represent random sampling data. A peak trace representation usually increases slightly in level with an increase in the averaging count. It should also be noted that using trace averaging requires a stable trigger event. Otherwise, traces with different timings are averaged, which leads to erratic results. The averaging count can be set to Real Time, in which case all averaging is turned off entirely. This is useful for measuring single-shot events. When the averaging function is deactivated, measurements are performed without chopper stabilization, meaning that a measurement then consists of a single sampling sequence activated by a trigger event. Otherwise, the detector s output-voltage polarity is reversed automatically for alternate sampling sequences. This suppresses lowfrequency noise and increases the accuracy with which the average power is measured at each pixel. Averaging has no effect on the randomly selected samples; the largest values for each averaging sequence are output as peak values. Fig : Setting the trigger. Mode The trigger mode can be set to Free Run, Auto, Normal, or Single mode. 68

69 Trace Measurements In free run mode, the sensor does not wait for trigger events and continuously acquires waveforms. The waveform display is not synchronized, and the waveforms typically roll across the display. Averaging is disabled in this mode. In auto trigger mode, the sensor acquires data, even if no trigger event occurs. Auto mode uses a timer that starts when a trigger occurs. After 1 s has elapsed, the software forces a trigger release and restarts its timer. In the absence of valid trigger events, the acquired waveforms are not synchronized. Average filter settings do not apply in this case. When valid triggers exist, the waveforms become synchronized and averaging can be used. In normal mode, the sensor only acquires a waveform when it is triggered by an internal or external trigger event. If no trigger event occurs over a period of about 2 seconds, the Trig? indication is shown on the user interface. The last waveform acquired remains on the display, and the sensor keeps waiting for the next trigger event. If no waveform has been acquired, the display remains blank. In single mode, the sensor arms its trigger unit when the ARM button is pressed. Subsequently, it performs one acquisition as soon as the trigger condition is met. If no trigger event occurs over a period of about 2 seconds, the Trig? indication is shown on the user interface. The waveform acquisition includes averaging, which means that 2*N trigger events must occur before the acquisition completes. If single-shot events are to be analyzed, the averaging filter count must be set to Real Time. Source This setting establishes the trigger signal's source. Internal means that the trigger event is generated by the applied RF signal and by the set trigger level. The External setting uses the BNC input signal from the R&S NRP-Z3 or R&S NRP-Z5 adapter. When an external source is used, the trigger level and hysteresis functions are not effective. The option 'External SMB' activates the SMB trigger input that is available on all new generation NRP power sensors. In the case the user can select between an input impedance of 10 kohms or 50 Ohms. Output The new R&S NRP Power Sensors can use their built in SMB connector as trigger output. In this case a rising edge is generated when the power measurement starts. Please see the power sensor user manual for details on how to use this feature. Slope The trigger slope can be set to either the positive or negative edge. This setting is available for all trigger sources. Level This level setting establishes the trigger threshold for internal triggering derived from the test signal. In order to achieve stable triggering conditions, a trigger level above 40 dbm is advisable. Holdoff The holdoff setting suppresses trigger events within the set holdoff time (in seconds), starting from the time of the last successful triggering. The holdoff time must be larger than the total trace time. Dropout This setting establishes the dropout time in microseconds. With a positive (or alternatively: negative) trigger slope, the dropout time is the minimum time for which the signal must be below (above) the trigger power level before triggering can occur again. As with the holdoff parameter, unwanted trigger events can be excluded. The set dropout 69

70 Trace Measurements time only affects the internal trigger source. Trigger level Fig : Holdoff time (red) and dropout time (blue). The figure above shows the various times related to a trigger event. The green diamonds mark the trigger point in time, located on the rising edge of a pulse. The red arrow marks the holdoff time. This time must elapse before the trigger system is rearmed and the system can retrigger. The blue arrows mark the dropout time periods. The signal level must remain below the trigger threshold before the system can retrigger. Using the dropout time is useful in many applications: Triggering on the first pulse of a pulse train. Triggering on bursts that contain intra-pulse modulation with AM content. Triggering on levels that are close to the pulse top. Fig : Additional measurement settings. Pulse This button opens the Auto Pulse Settings panel. Use this panel to configure all parameters that are relevant to the automatic pulse measurement. The check-box is used to enable or disable the automatic pulse measurement globally. Marker This button opens the Advanced Marker Settings panel. Use this panel to configure all parameters that are relevant to the marker measurements. The check-box is used to enable or disable the marker measurements globally. 70

71 16.2 Trace Measurements Custom Settings The Power Viewer Plus software provides a mechanism for adding custom settings to the trace measurement configuration. For this purpose, SCPI commands can be sent at the end of the trace measurement configuration. Adding these commands is generally not required, but it may be useful for special applications or for debugging. The SCPI commands can be specified on the debug page for the program settings dialog. This dialog is reached from the menu bar by selecting Configure Options Debug. The syntax for adding custom commands is: trace.cmd=<command> Multiple lines may be appended after each other. Lines starting with a hash mark (#) are treated as a comment and ignored. Disabling the NRP-Z81 equivalent sampling mode: trace.cmd=sens:trac:esam:auto OFF NRP-Z81 trace measurements at reduced bandwidth: trace.cmd=sens:bwid:vid "5 MHz" Using a specific sensor path with the 3-path diode sensors: trace.cmd=sens:rang:auto OFF trace.cmd=sens:rang 2 71

72 16.3 Trace Measurements Graphical Trace View The graphical trace view contains the information shown below. Reference level Legend Delayed trigger Physical trigger Scaling Level lines Trigger level Number of samples / Trigger Measurement Time gate area Time at right screen border Time at left screen border Fig : The graphical trace view. 72

73 Trace Measurements If the trigger level is located outside of the visible area, the green trigger level line disappears. In its place, a little arrow next to the T indicates the direction in which the trigger level is located. Fig : Trigger level located above the visible area. If the trigger position (time) is located outside the display area, the green arrow rotates and indicates the direction in which the physical trigger position is located. Fig : Trigger position to the left of the visible area Context Menu The graphical trace view provides a context menu that can be activated by right-clicking the mouse. This context menu contains all functions directly related to the graphical data representation, and its settings do not affect sensor settings. Fig : Context menu for the graphical trace view. 73

74 Trace Measurements Log / Lin Trace data can be viewed in a linear scale (watts) or in a logarithmic scale (dbm). The menu switches back and forth between these two views. However, when switching from linear to log mode, negative y-values are truncated, and the lowest dbm level set to 60 dbm Trace Representations Some power sensors (e.g. the R&S NRP-Z8x) return multiple trace data representations. Due to the high sampling rate that these sensors use, the average, random and peak power information is available for each video point. The difference between these representations is outlined below. Average This representation averages the power values of identical measurement points, i.e. points at the same distance from the trigger point. This reduces noise, since the noise reduction is proportional to the square root of the averaging factor. This is the default representation, and it is available with all sensors. Peak The maximum of all samples taken at identical measurement points, i.e. points at the same distance from the trigger point, are found and output. Random The random values are obtained from the first measurement cycle. Repetition of the measurement cycle has no effect on the result. 74

75 16.6 Trace Measurements View Modes The view modes affect the way that trace data is rendered on the display. View modes do not affect the sensor configuration. Therefore, any view mode can be combined with any sensor configuration. Normal In normal mode, each trace captured by the sensor is rendered on the screen. Newer traces replace older ones. This is the default view mode, and it is useful for most applications. Envelope In the envelope mode, trace data is accumulated over up to 256 traces, and the area between the minimum and maximum readings is filled with a semi-transparent color. The current trace is highlighted within this area. This mode is very useful for finding glitches, transients, or other random events. The screen shot below shows an EDGE burst in the envelope view mode, and all traces average, random and peak activated. Fig : EDGE burst in envelope view mode. 75

76 Trace Measurements Persistent Dots In the persistent-dots mode, the software collects data on up to 32 trace measurements. These trace data points are displayed without interconnecting lines. This mode is useful for analyzing waveform anomalies. The example below shows a jittered AM signal that is captured in realtime mode (no averaging). Fig : Jittered AM signal in persistent dots view mode. Moving Average and Exp. Moving Average The moving average modes provide additional video filtering of up to 512 trace measurements. The regular moving-average mode displays the average trace data of the last 512 measurements. All traces are weighted equally. This filter provides the best noise reduction. The exponential moving average filter uses an exponential weighting of the last 512 measurements. The last measurement is weighted the most, whereas older measurements only have little influence on the displayed result. This filter responds faster to signal changes but provides less noise reduction. Since the view modes do not affect the measurement itself, it is generally possible to combine these video averaging filters with un-triggered measurements, e.g. noise-like signals. The result is the average power of uncorrelated signal portions. 76

77 16.7 Trace Measurements Lines Fig : Measurements related to level lines and the time gate. Time-gate and level lines may be activated for simple measurement tasks. Both line sets can be dragged by holding down the left mouse button. Text boxes at the lower border show all related readings, such as delta values and power readings within the gated area. Time Gate 1 The position of line 1 relative to the trigger position 2 The position of line 2 relative to the trigger position Δ The time difference between both lines Gated Meas. Av The average power between the time-gate lines Pk The peak power between the time-gate lines Δ The ratio of the peak and average powers Level Lines 1 The level at which line 1 is positioned 2 The level at which line 2 is positioned Δ The ratio of both levels Measure Mod. Depth The level lines can optionally be used for measuring a signal's modulation depth according to the following equation: P 1 P 2 P 1 P 2 Depth %= Reference Trace Trace data can be saved to memory and shown as a reference curve. The To Ref context menu entry saves all active traces to memory and displays them in a slightly darker color together with the measured traces. Reference trace data gets scaled or moved when the x- or y-scaling is changed. Clicking Clear Ref removes all reference trace data from memory. 77

78 16.9 Trace Measurements Using Markers Markers can be tied to trace points for performing automated measurements. The context menu is used to activate the Marker Configuration dialog shown below. Fig : Marker configuration dialog. Power Viewer Plus offers one reference and up to four delta markers. Each of the markers can be assigned to the average, random, or peak trace. This assignment is independent of whether the trace is enabled for viewing. The reference marker is set to a point in time relative to the trigger point. Alternatively, it may be set to automatic peak-tracking mode, which positions the marker to the maximum power value within the visible trace area. Fig : Markers in trace display. The reference marker is indicated with a small purple arrow that points downward to the trace that it is associated with. Delta markers can be defined relative to the reference marker or relative to other delta markers. This allows the user to build chains of markers and perform measurements between marker points. Each delta marker can be operated in different operating modes. These modes are outlined below. Fixed Time This mode is available for the reference and the delta markers. The marker is positioned at a fixed point in time. This time is relative to the trigger point, whereas the delta marker time is related to the marker that it is dependent on. Fixed time markers can be dragged using the mouse. Auto Peak This option is only available for the reference marker. If selected, this marker automatically tracks the peak-power value within the visible trace area. Δ Level This option is only available for the delta markers. If selected, this marker is automatically positioned at the desired signal level by 78

79 Trace Measurements searching to the right or left. The level is entered as a value that is relative to the power level upon which the marker is dependent. An additional exclude time can be set to inhibit searching within this period of time. Next Peak This option is only available for the delta markers. If selected, this marker automatically searches for the next signal peak. The level value defines a relative threshold for the peak search. It should be set so that noise peaks are not accidentally evaluated as signal peaks. An additional exclude time can be set to inhibit searching within this period Pulse Width Measurements An automatic pulse width measurement can be performed using three markers. The reference marker is set to a fixed point in time at which its level exactly matches the pulse top power. The first delta marker is set to search a level that is 3 db down from the reference marker to the left side. This positions marker one at the rising edge's 50 % power level. The second delta marker is set to search the same power level to the right side. It is related to the first delta marker and thus marks positions at the same power level on the falling edge. The result is shown in the figure below. Fig : Markers in a pulse-width measurement. The pulse width can be read from the ΔT measurement for delta marker two. Fig : Marker readings for the pulse width measurement. 79

80 Trace Measurements Pulse Rise-Time Measurements The rise or fall time of a pulse can be measured using a set of two delta markers and the reference marker. The reference marker is set to a fixed point in time at which its level exactly matches the pulse top power. The first delta marker is set to search a level that is 1 db down to the left side. It is related to the reference marker. The second delta marker is set to search a level that is 19 db down to the left side. It is related to the first delta marker. The result is shown in the figure below. Fig : Rise time measurement using markers. The pulse rise time can be read from the ΔT measurement for delta marker two. Fig : Marker readings for the pulse rise-time measurement. 80

81 16.10 Trace Measurements Zooming The Power Viewer Software supports time and level zooming in the trace panel. Both zoom operations are controlled using the left mouse key and the mouse wheel. Time zoom operations always reconfigure the sensor, whereas level zooms only adjust the view port Time Zoom The time zoom mode is entered by holding the left mouse key down while dragging the mouse cursor horizontally. A blue area highlights the minimum area that will be shown when the zoom operation completes. When the left mouse key is released, two horizontal arrows indicate that the zoom operation can now be performed using the mouse wheel. Fig : Time zoom in a trace measurement. Rolling the mouse (scroll) wheel forward or clicking into the marked area zooms into the trace and magnifies the marked area. The time resolution is rounded to the next useful value on a scale. Rolling the mouse wheel backwards zooms out. The center point for the zoom operation is the center point of the marked area. The time resolution is rounded to the next useful value on a scale. After the zoom operation has completed, the marked area and the arrows disappear. The zoom operation can be aborted by clicking the right or left mouse key instead of rolling the mouse wheel. Undo When no zoom area is selected, the mouse wheel serves as an undo function for the time zoom. Rolling the mouse wheel backwards restores the previous timing. 81

82 Trace Measurements Level Zoom The level zoom mode is activated by clicking and releasing the left mouse key in the trace window. A set of vertical arrows marks the zoom level and indicates that the zoom operation can now be completed using the mouse wheel. Fig : Level zoom in a trace measurement. Rolling the mouse wheel forward magnifies the trace and centers it around the selected zoom point. Rolling the mouse wheel backwards undoes the last level-zoom operation. The level-zoom mode is automatically aborted if the mouse cursor is moved without rolling the wheel. This prevents users from remaining in level zoom-mode unintentionally. 82

83 16.11 Trace Measurements Automatic Pulse Measurement Pulses are measured in line with the international IEC 469 standard. The algorithm in the power sensor first determines the pulse amplitude of the measured trace by establishing the distance between the top power and base power. This pulse amplitude then serves as the basis for the percentage values for the three reference levels that can be preset high, mid, and low. These three levels are used to measure the pulse. Typical values are 90 %, 50 %, and 10 %. While the low and high reference levels are only employed to determine the pulse s rise and fall times, the "mid" reference level is used to determine all other timing parameters. Automatic pulse measurement can be activated in the trace settings if the sensor supports this feature (e.g. R&S NRP Z-8x). The measurement is performed inside the sensor, and data is provided for each captured trace. Pulse measurement data can also be sent to the data processing panels. The following figure explains the terms used with the automatic pulse measurement. Peak power Top power High ref. level Mid ref. level Pulse width Low ref. level Base power Rise time Fall time Fig : Automatic pulse measurement. Base power The base power is the average power for the period without any signal content. Typically, this is the average noise power. The base power is referred to as the 0 % level. 83

84 Trace Measurements Top power The top power is the pulse power excluding any overshot at the beginning of the pulse. This power level is referred to as the 100 % level. Peak power The peak power is the maximum power value captured in the trace. The low, mid, and high reference levels The three reference power levels are specified by the user as a percentage of the pulse amplitude. The amplitude is the power difference between the pulse's top power and its base power. The three reference power levels are used to determine all pulse timing related values, such as the rise time and fall time or the pulse width and pulse repetition time. Please note that automatic pulse measurements are based on the average trace data returned from the sensor. It is important that stable trigger conditions exist and that the average count is set high enough in order to run the automatic pulse measurement function. Enabling automatic pulse measurement slightly slows down the measurement rate. This is normal and is due to the sensor's internal processing routines. The automatic pulse measurement settings are configured in a separate dialog that is accessible from the trace measurement configuration dialog box. Fig : Automatic pulse measurement settings. Algorithm This control is used to select the analysis algorithm for determining the pulse-top and pulse-base powers. These two power levels are fundamental for all further signal analysis. The default setting, Histogram, determines the pulse-top and pulse- 84

85 Trace Measurements base power levels by evaluating the probability density of all the values for one recorded trace. This algorithm is recommended for most pulses. The Integration algorithm approximates the pulse by an ideal signal with the same energy content, pulse duration, and pulse period. Use this algorithms for pulses with fast rise times and fall times as well as for pulses with amplitude variations (e.g. modulation). The Peak algorithm is not available on all of the wideband power sensor's firmware versions. Please consult the sensor firmware's release notes to see if this feature is supported. The algorithm uses the overall peak power as the reference and is useful for Gaussian pulse shapes. Prefer High Res. This check box enables high-resolution pulse analysis if the sensor firmware supports this feature. High-resolution pulse analysis uses an equivalent sampling technique and provides more precise measurement results. Measurement time increases when highresolution pulse analysis is used. When the sensor uses high-resolution pulse analysis, an additional line in the results table displays the equivalent sampling time. Ref Level The low, mid, and high reference-level values define the levels used for determining the pulse timing. All values are specified as a percentage of the pulse amplitude (the difference between the top and base powers). Levels are related to power readings in watts. The following table compares levels expressed in volts, watts and dbw. Level (% V) Level (% W) Level (dbw)

86 Trace Measurements Measurement Results These check boxes select which results are shown in the graphical trace view. Low-level and high-level values are marked by small horizontal lines for the rising and falling edge that the sensor evaluates. The positions of the rising and falling edge are marked with small diamonds. Fig : Automatic pulse measurement in trace view. Algorithm The analysis algorithm is used to determine the pulse-top and pulsebase powers. These two power levels are fundamental for the pulse analysis, because they are required for calculating the pulse amplitude. The reference levels are specified as a percentage of the pulse amplitude, and they are required for measuring the entire pulse timing. Equiv Sampling This entry is only displayed if the high-resolution pulse analysis is used by the sensor firmware and shows the equivalent sampling time. This time provides a good measure of the accuracy that is achieved with the automatic pulse measurements. The following example demonstrates the difference in resolution: Without equivalent sampling: video points < 400 ns / sample With equivalent sampling: Signal dependent 2.5 ns High-resolution pulse analysis is useful when pulses with large duty cycles need to be measured (for example, in radar applications). 86

87 Trace Measurements Unit The reference level values are specified as a percentage of the pulse amplitude, and they can be voltage or power related. Level The low, mid, and high reference levels that are used for the automatic pulse measurement. The levels are set as a percentage of the pulse amplitude. Pulse Duration The time between the pulse's first rising edge and subsequent falling edge. If the duration cannot be determined, "---" is displayed. Pulse Period The time between two consecutive edges of the same type. If the period cannot be determined, "---" is displayed. Rise Time The time of the first rising edge. The rise time is defined as the period between the point in time when the signal exceeds the lower reference level and the point in time when it exceeds the upper reference level. If no rising edge can be detected, "---" is displayed. Rising Edge The point in time when the first rising edge occurs, i.e. the point in time at which the signal exceeds the average reference level. The rising edge position is based on the delayed trigger point. Rising Edge To Phys. Trig. Unlike the standard rising edge time, this time is related to the physical trigger. If no trigger delay is set, this measurement is omitted. Fall Time The time of the first falling edge. The fall time is defined as the period between the point in time when the signal falls below the upper reference level and the point in time when it falls below the lower reference level. If no falling edge can be detected, "---" is displayed. Falling Edge The point in time when the first falling edge occurs, i.e. the point in time at which the signal falls below the average reference level. The fallingedge position is referenced to the delayed trigger point. Falling Edge To Phys. Trig. In contrast to the standard falling-edge time, this time is related to the physical trigger. If no trigger delay is set, this measurement is omitted. Peak Power The maximum power value for the entire trace. Top Power The pulse's top power. This power level is determined by the analysis algorithm and is fundamental for the entire pulse analysis. 87

88 16.12 Trace Measurements Common Measurement Tasks The trace panel can be used for many different measurement tasks. The following overview provides advice on how to configure the trace measurement for common signals. Repetitive pulsed signals without jitter Trigger Normal Averaging 2 View Normal This is the most commonly used setting. It measures the signal's envelope power continuously. Repetitive pulsed signals with jitter Trigger Normal Averaging Real Time (OFF) View Envelope, Persistent Dots Continuously measures the envelope power. Averaging should not be used if the jitter needs to be observed. CW-like signals, e.g. WCDMA without trigger Trigger Free Run Averaging N/A View Moving Average Continuously measures the average power level of asynchronous trace measurements. Single-shot signals Trigger Averaging View Single ARM Real Time (OFF) Normal The averaging should be set to real time, because the sensor otherwise requires multiple trigger events to generate one measurement result. 88

89 17 OTA Trace Measurement OTA Trace Measurement The OTA trace measurement is very similar to the regular trace measurement. The main difference is that OTA sensors deliver three individual power traces from their individual antenna elements. The OTA measurement therefore displays three average power traces rather than an average power, peak power and random sample trace. This chapter only describes the differences to the regular trace measurement Settings Fig : Settings for OTA trace measurements. Channels The three channels of the OTA sensor can be individually enabled or disabled. Channels without an antenna connection must be disabled to avoid error messages. 89

90 18 Statistics Statistics The Statistics mode makes it possible to measure the envelope power's complementary cumulative distribution function (CCDF) or probability density function (PDF) Settings Fig : Settings for the statistics mode. Zero This option is only available while the measurement is running. It starts the zeroing sequence for the current sensor. For this purpose, the RF signal must be switched off, or the sensor must be disconnected from the signal source. The sensor automatically detects the presence of any significant power, which causes zeroing to be aborted and an error to be generated. The LED next to the zeroing button changes it's color depending on the current zeroing state. Grey Zeroing was not performed since the measurement got started. Blue Zeroing is in progress. The measurement results are not updated during this time. Green Zeroing finished successfully. Red Zeroing failed. In this case the LED initially flashes and then turns red continuously. The zeroing process may take more then 8 seconds to complete and varies with the sensor model. Generally, it is possible to run the sensor zeroing with a small signal (such as broadband noise) applied to the sensor. This makes it possible to compensate for this signal in later measurements. Distribution Function Power Viewer Plus displays the CDF, CCDF, or PDF for a signal. 90

91 Statistics The CCDF shows how often a measured power value is above a certain level. For this purpose, the x-axis uses a logarithmic power scale. The y-coordinate is scaled logarithmically and in probabilities from 100 % down to 10-4 %. The CCDF can also be viewed as a relative graph, in which case the x-axis is scaled to the signal's average power. Total Samples The number of samples determines the length of the evaluation window. The overall window length is calculated from the sampling rate (80 MHz, 12.5 ns interval) and the sample count. The default evaluation length is one million samples, and the maximum permissible value is 768 million samples. The application uses an asynchronous measurement mode for the statistics measurement. Thus, a measurement is started without waiting for any trigger event and stopped when the set evaluation period has elapsed. Sample counts larger than 24 million points are divided into up to 32 sets of measurements, each of which evaluates up to 24 million points. At full bandwidth, the time required for one measurement with 24 million samples is 0.3 seconds. Therefore, larger sample counts require more time for the measurement to stabilize. Reference Curves The current curve can be saved as a reference curve and used for comparisons with later measurements. Any reference curve data is only stored locally in the graphics view. It cannot be saved to a file. The statistics view supports up to four reference curves in addition to live measurement data. Show AWGN Reference Curve When this function is enabled, the system plots an ideal AWGN curve. This reference curve can be used with all statistics modes. 91

92 Statistics 18.2 Graphical Data View The graphical data view may contain the information shown below. The level and probability lines are activated with a right mouse click into the graphics area. Average power indicator Level line Legend Numeric readings Probability line Maximum level Probability line values Level line values Minimum level Fig : The graphical data view. The application supports absolute or relative CCDF displays as well as the PDF and CDF displays. If numeric results are required, the display can be tiled to show the table with numeric results on the right side. Fig : Tiling the display to show the table with numeric results. Table mode This mode is activated by calling up the context menu (by right clicking 92

93 Statistics into the graphics area) and selecting Table View PDF Mode PDF stands for probability density function. It shows the probability at which readings occur at a given level. For this purpose, the logarithmic level scale is divided into sections, which are referred to as bins. Each measurement that falls within a bin increases the bin counter by one. At the end of the measurement cycle, all bin counts are normalized by the linear scale's bin boundaries. Fig : The PDF curve of a 3GPP signal with an AWGN reference curve at an average power level of 0 dbm. The figure above shows a 3GPP signal that has a power distribution similar to that of AWGN. At low signal levels (left side) the resolution of the sensor's A/D converter might become visible. 93

94 Statistics PDF Background Information This section outlines the relationship between the displayed PDF curve and various signals, such as 3GPP or WCDMA. For a (gauss) noise-like signal, it can be assumed that the signal is being generated by I/Q modulation of an RF carrier. Both the I-signal and the Q-signal are normally distributed with a mean value of zero and a standard deviation of one. The resulting carrier envelope power is proportional to the sum of I2 and Q2: 2 P I Q 2 The figure below shows the resulting vector and RF signal based on the two normally distributed baseband signals. Q I Fig : Vector resulting from I/Q modulation and a modulated RF carrier. The PDF of a normal distribution can be described by the following formula: f,, x = 1 x 2 exp If a new random variable is created from k independent, normally distributed random variables with a mean of zero (µ=0) and a variance of one ( σ=1 ), the resulting distribution is the chi-square distribution: { k /2 1/2 x k / 2 1 e x/ 2 x 0 f x, k = k /2 0 x 0 } 94

95 Statistics In the case of I/Q modulation, the degree of freedom is two (k=2), which simplifies the PDF according to the following equation: 1 1 x f x,2 = e 2 2 This equation demonstrates that the AWGN signal produces a PDF that can be approximated by the exponential function. This is the AWGN reference curve that the Power Viewer Software adds as the reference. The statistics become more complex when the signal consists of two components, such as a constant RF carrier with a modulated signal. Q I Fig : Vector resulting from a Fig : Constant signal with added I/Q modulation and PDF. In this case, the mean values of the I- and Q-signals are not zero (µ 0) and the chi-square distribution cannot be used for the resulting envelope power. Please note that the assumptions above are valid for power levels that are significantly higher than the power-sensor noise level. 95

96 18.4 Statistics CDF Mode Fig : The CDF curve of a 3GPP signal with an AWGN reference curve at an average power level of 0 dbm. CDF stands for cumulative distribution function and is the integral of the PDF curve. The advantage of the CDF curve is that it emphasizes minimum power values. The ideal AWGN reference signal remains the exponential function, because the integral of ex remains ex. 96

97 18.5 Statistics CCDF Mode Fig The CCDF curve of a 3GPP signal with an AWGN reference curve at an average power level of 0 dbm. CCDF stands for complementary cumulative distribution function and is the complement of the CDF curve. Its curve is calculated using the following equation: CCDF =1 CDF The advantage of the CCDF curve is that it emphasizes peak power values. That is the main reason why the CCDF plays an important role in many applications. The ideal AWGN reference signal is described by the exponential function. 97

98 19 Timeslot Mode Timeslot Mode This mode measures the average power of a definable number (up to 16) of successive timeslots within a frame structure with equal spacing. When the averaging function is activated (averaging factor of two or more), measurements are performed with chopper stabilization to obtain more accurate results with reduced noise and zero offset. Chopper stabilization involves reversing the polarity of the detector output signal from frame to frame. Taking the difference of the output signals minimizes the video path's effect on noise and zero drift Settings Fig : Zeroing controls. Zero This option is only available while the measurement is running. It starts the zeroing sequence for the current sensor. For this purpose, the RF signal must be switched off, or the sensor must be disconnected from the signal source. The sensor automatically detects the presence of any significant power, which causes zeroing to be aborted and an error to be generated. The LED next to the zeroing button changes it's color depending on the current zeroing state. Grey Zeroing was not performed since the measurement got started. Blue Zeroing is in progress. The measurement results are not updated during this time. Green Zeroing finished successfully. Red Zeroing failed. In this case the LED initially flashes and then turns red continuously. The zeroing process may take more then 8 seconds to complete and varies with the sensor model. Generally, it is possible to run the sensor zeroing with a small signal (such as broadband noise) applied to the sensor. This makes it possible to compensate for this signal in later measurements. The power scale section sets the y-axis for the timeslot display. The numbers do not affect the measurement itself. Fig : Setting the power scale. Reference Sets the maximum power level for the timeslot display. 98

99 Timeslot Mode /div Defines the scaling between 0.1 db/div and 20 db/div. Fig : Setting the timeslot structure. The timing section sets all parameters that are required to precisely define the timeslot structure for the signal to be analyzed. To obtain stable and reliable results, it is essential that these parameters match the signal exactly. Number Defines the number of timeslots that belong to a single frame. The permissible range is from one to 16. Width Sets the width of a single timeslot within the frame structure. Each timeslot has exactly the same width. Exclude Defines the time gaps at the beginning or at the end of a timeslot that is not evaluated for the measurement. These parameters are used to define the spacing between adjacent timeslots. Delay Specifies the time between the physical trigger event and the start of the first timeslot. The averaging filter reduces the noise level significantly. Fig : Setting the filter count for the averaging filter. Count The filter count sets the number of frames that are to be evaluated to form one measurement result. Timeslot measurements always require stable and reliable trigger conditions. This is particularly important when averaging is enabled and multiple measurements are used for one result. 99

100 Timeslot Mode Fig : Trigger settings. Source The trigger source can be either internal or external. When an external source is used (R&S NRP-Z3 or R&S NRP-Z5 adapter), only the settings for polarity and holdoff time are available. The new generation NRP power sensors provide a second SMB trigger input. This option is available as External SMB. Slope The trigger slope can be set to either the positive or negative edge. This setting is available for all trigger sources. Level Sets the trigger threshold for internal triggering derived from the test signal. In order to achieve stable trigger conditions, it is advisable to use a trigger level above 20 dbm. Holdoff Suppresses trigger events within the set holdoff time (in seconds), starting from the time of the last successful triggering. Dropout Sets the dropout time in microseconds. With a positive (or alternatively: negative) trigger slope, the dropout time is the minimum time for which the signal must be below (above) the trigger power level before triggering can reoccur. As with the holdoff parameter, unwanted trigger events can be excluded. The set dropout time only affects the internal trigger source. 100

101 19.2 Timeslot Mode Graphical Data View The graphical timeslot power view contains the information shown in the figure below. Reference level Level scaling Trigger level Timeslot width Marker measurement Average filter count Fig : The timeslot panel. All bars are updated continuously. The update rate depends mainly on the setting for the average filter count. The higher the filter count, the lower the update rate and the noise level. Markers A maximum of four markers can be placed on any of the 16 timeslots to achieve precise average power readings. These readings are displayed below the bar graph area. Fig : Context menu in the timeslot average view. 101

102 Timeslot Mode A context menu can be activated within the bargraph view by rightclicking with the mouse. This menu is used for selecting the active marker (one to four). As a next step, markers can be dragged by holding down the left mouse button within the graphics area. A thin line displays the column on which the marker is to be placed. Moving the mouse to the very left or very right border disables the marker. 102

103 20 Multi-Channel Power Measurements Multi-Channel Power Measurements Many power measurement tasks require simultaneous measurement of multiple channels. The multi-channel measurement panel provides up to 16 parallel power measurements. In addition, four results can be computed using mathematical expressions. Channel number and sensor Averaging count Sensor status Measured value Mathematical configuration Sampling window Computed value Fig. 20.1: Multi-channel measurement. Each of the 16 channels use a separate sensor configuration. Therefore, assigning the same sensor to multiple channels is not permitted. Data Processing All measured values are automatically forwarded to the data processing panels. 103

104 20.1 Multi-Channel Power Measurements Channel Configuration Power Viewer Plus continuously measures the average power of up to 16 sensors. These measurements are referred to as channels (Ch 1 through Ch 16). Each channel must be either turned OFF or assigned to an R&S NRP sensor. Fig : Sensor zeroing settings. Zero This button starts the sensor zeroing process for the currently selected channel only. None of the other channels are affected. For sensor zeroing to work, no RF power may be applied to the sensor and no measurement may be running. All Zero Starts the sensor zeroing process for all configured sensors. Only sensors that are marked by a green LED can be zeroed. Left and right arrow These buttons switch the system to the previous or next channel. Sensor selection This control lists all sensors that were discovered during the last scan and the sensors required by the currently loaded project. Available devices are marked with a green LED. Devices that are required by the currently loaded project but appear to be unavailable are marked with a yellow LED. Sensors that are already assigned to a measurement channel display their channel numbers in square brackets. These devices cannot be assigned again to a different channel. Each sensor can be configured individually. This accounts for the use of different sensor technologies or individual path losses or measurement frequencies within one setup. If an alias name has been assigned to a sensor, this name is displayed first followed by the sensor type and serial number in parentheses. 104

105 20.2 Multi-Channel Power Measurements Measurement Settings The measurement settings are similar to the ones available for the continuous power average mode. Please refer to the chapter 14, "Continuous Power Measurements," for a more detailed discussion of these parameters. Fig : Multi-channel measurement settings. The general correction settings, such as the frequency, level offset, or S-parameter correction, are not used in multi-channel measurement mode. Instead, each channel provides its own settings Mathematical Expressions Power Viewer Plus provides a set containing the most commonly required mathematical operations that can be computed from two measured values. These four computed values are referred to as channels M1, M2, M3 and M4. Fig : Mathematical settings for multi-channel measurements. Sum Two measured power values are added in linear scale (watts). The result is displayed in watts or dbm. Diff Two measured values are subtracted from each other in linear scale (watts). The result is displayed in watts or dbm. Ratio The ratio is calculated by dividing one measurement by the other one. The result is displayed without a unit or is converted to db. 105

106 Multi-Channel Power Measurements RCoeff The reflection coefficient is computed from two measured values in logarithmic scale using the following equation: RC 10 P1 P2 20 SWR The standing wave ratio is computed from two measured values using the following equation: SWR 1 RC 1 RC 106

107 21 OTA Multi-Channel Measurement OTA Multi-Channel Measurement The OTA multi-channel measurement is very similar to the regular multichannel measurement. The main difference is that OTA sensors deliver three individual power values from their individual antenna elements. The OTA measurement therefore displays three average power values per sensor. This chapter only describes the differences to the regular multi-channel measurement Settings Fig : Settings for OTA multi-channel measurement. Channels The three channels of the OTA sensor can be individually enabled or disabled. Channels without an antenna connection must be disabled to avoid error messages. 107

108 22 Script-Based Measurement Script-Based Measurement The script-based measurement function is useful for implementing custom measurement tasks that cannot be covered by the measurement panels provided in Power Viewer Plus. For this purpose, a script language is used. This language is based on the ECMAScript scripting language, as defined in the standard ECMA-262. Microsoft's JScript, and Netscape's JavaScript are also based on the ECMAScript standard. Fig. 22.1: The script window. The scripting panel provides three pages. The first page shows the measurement results and status messages during script execution. This page is automatically activated when the script gets started. Data can be viewed numerically or graphically as trace or bar chart. The layout of this panel is designed by the UI editor which is located on the third page. The second page contains the script editor itself. 108

109 22.1 Script-Based Measurement Script Syntax The detailed description of the script syntax is out of scope of this user manual. Please refer to the script standard, that can be found under the following link: Alternatively, any tutorial or documentation on JavaScript could be used. In addition to the script standard the global object pvp exists within the Power Viewer script context. This global object extends the script functionality. This section mainly describes the pvp script extension Printing to the Script Output Window Any text or string variable can be printed to the script output window. The print method is used to do this. Non-string variables must be converted into strings before using them with this method. pvp.print( 'My Text' ) var dval = 1e-3 pvp.print( 'Measurement = ' + dval.tostring() ) 22.3 Printing to the Message Window Three methods exists to output text messages to the application's message log. By default only information and error messages are shown on the message log. Debug output can be enabled by starting the application with the command line argument --debug. pvp.info( 'This is an info text.' ) pvp.debug( 'This is a debug message.' ) pvp.error( 'This is an error message.' ) 22.4 Waiting Periods The sleep method is used to wait for a definable amount of time. The waiting period is specified in milliseconds, and script execution is suspended for at least this amount of time. var itimems = 500 pvp.sleep( itimems ) 22.5 Timers The script extension provides built in timers. These timers can be used to measure the execution time of measurement tasks. The following method returns the current timer resolution in seconds. var dresolution = pvp.timerresolution() This time method returns the time since script start in seconds. var dcurrtime = pvp.time() The following method returns the time of the last SCPI execution in seconds. 109

110 Script-Based Measurement var dlastexectime = pvp.lastexectime() This usertime method returns the elapsed time since the last call to this method in seconds. usertime() // do something double delapsed = usertime() 22.6 Power Conversion Two conversion methods exist for the commonly required task to convert from linear power in Watts to logarithmic power levels. The todbm method converts a power value from Watts to dbm. var dpowdbm = pvp.todbm( 1e-3 ) pvp.print( 'Power Level in dbm is ' + dpowdbm.tostring() ) The todbw method converts a power value from Watts to dbw. var dpowdbw = pvp.todbw( 1e-3 ) pvp.print( 'Power Level in dbw is ' + dpowdbw.tostring() ) 22.7 File Operations The script language contains three functions related to file-based output operations. All file operations should be used with great care because they provide full access to the entire file system using the current user account of the application. The following function call removes a file from the hard drive. pvp.removefile( 'C:/Temp/MyFile.tmp' ) Writing text data to a file is possible using the following two functions. pvp.writetofile( 'C:/Temp/MyFile.tmp', 'String to write' ) pvp.appendtofile( 'C:/Temp/MyFile.tmp', 'String to append' ) Both functions write a single line of text. The first function removes the content of an existing file and then writes the text string. The second function appends the string to the end of the file. 110

111 22.8 Script-Based Measurement Defining Devices The primary task for the script language is to remote control power sensors or other measurement equipment via SCPI commands. Before sending remote control commands to an instrument it is required to define this instrument as a device. The currently selected sensor is available as device with the index zero. Other sensors may be defined for use with the script language in which case device index values starting at one will be assigned. The definedevice method defines a sensor ID and also opens the sensor connection. It must always be used before attempting to send SCPI commands to a sensor. All sensor connections are automatically closed when the script finishes. If this is not desired an infinite loop must be added to the end of the script. The definedevice method may also be used to define VISA interface connections to Rohde&Schwarz signal generators. In this case the argument is interpreted as the VISA resource string. Prepare the primary power sensor for use with the script. If device index 0 is used the second argument is ignored. The method automatically uses the resource string of the currently selected power sensor. pvp.definedevice( 0, '' ) It is also possible to open other R&S instruments for use in the script. In this case the second argument sets the VISA resource string of the instrument. pvp.definedevice( 1, 'TCPIP:: ::INSTR' ) 22.9 Communication Timeout The setdevicetimeout method sets the timeout for all instrument communication. The first parameter is the device ID used with the definedevice method. The second parameter is the timeout value in seconds. pvp.setdevicetimeout( 1, 10 ) Suppressing SCPI Log Messages By default all SCPI communication is logged to the applications message log. The logging can be turned on or off with the setechostate method. The logging state affects all devices. pvp.setechostate( 0 )... pvp.setechostate( 1 ) SCPI Commands and Queries SCPI commands are directly send to the instrument. The list of available commands can be found in the instrument's user manual. It is important to note that the capital letters used in the commands are the short form, whereas the lower case letters describe the complete command. Command in user manual: SENSe:AVERage:COUNt 32 Possible use: SENS:AVER:COUN

112 Script-Based Measurement SENSE:AVERAGE:COUNT 32 Queries are used similarly to SCPI commands, but they end with a question mark. The scpi method requires the device index as the first argument. The device index 0 is used for the primary power sensor. Before using the scpi method the sensor must be initialized by a call to definedevice. pvp.definedevice( 0, '' ) pvp.scpi( 0, '*RST' ) var sinfo = pvp.scpi( 0, 'SYST:INFO?' ) pvp.print( 'Info = ' + sinfo ) var sserial = pvp.scpi( 0, 'SYST:INFO? Serial ' ) pvp.print( 'Serial = ' + sserial ) var sfreq = pvp.scpi( 0, 'SENS:FREQ?' ) pvp.print( 'Correction Frequency = ' + sfreq ) The scpi method sends the command text provided as second argument to the instrument. If the command contains a question mark the method subsequently reads the instrument's response. The method can therefore be used for write and read operations. The returned data is always a string. Use the JavaScript's parsefloat() function to convert a string to a floating point value Simulated Queries for Use With NRP-Z sensors The following queries conform to the SCPI language but cannot be found in the NRP-Z sensor's user manual, because the sensors do not directly support them. Instead, these sensors send their measured values in binary format to the driver layer. The driver caches this information and provide them to the application using the driver. In order to access this cached information a set of queries are implemented in the scpi method. Reading scalar measurement data is achieved by using the READ? or FETCH? queries. Command Syntax: READ[:AUX]? FETCH[:AUX]? The READ? query initiates a measurement, waits for up to 5 seconds, and then fetches one single float-type result. The READ:AUX? query initiates a measurement, waits for up to 5 seconds, and then fetches a set of three scalar results. The READ:AUX? query can only be used if the sensor is configured for AUX mode. The FETCH? queries are similar to the READ? queries, but they do not initiate a measurement. Instead, the sensor is immediately queried for a result. An error is generated if no result is available. Four additional queries read or fetch buffer (array) measurement data from the sensor. Syntax: READ:BUFF[:AUX]? FETCH:BUFF[:AUX]? 112

113 Script-Based Measurement In contrast to the READ? and FETCH? queries, which read scalar results, these queries return an entire array of floating-point values. If the returned data is saved to a variable, this variable is automatically set to the floating-point array data type Reading Results from New Generation Sensors The new generation NRP power sensors support the SCPI language internally. In contrast to the SCPI standard the FETCH? query automatically waits for completion. This simplifies synchronization. pvp.scpi( 'INIT:IMM' ) var sresult = pvp.scpi( 'FETCH?' ) The result string contains either a single value or a list of values separated by comma. This depends on the measurement mode Waiting for Measurement Completion The sync method waits for the completion of a task or until the definable timeout has expired. This is useful when it is necessary to synchronize to the measurement completion state. During the waiting period, the user can stop the script using the Stop button. // start ARB in signal generator pvp.scpi( 1, 'BB:ARB:STAT ON' ) pvp.sync( 1, 5 ) // continue... Note: The time granularity of the instrument polling depends on the operating system. On Windows installations the minimum polling interval is in the order of 10 to 15 ms. For time critical applications it is therefore recommended to use the READ? query instead. This method is generally not required for NRP-Z power sensors. Instead use the READ? Query with these sensors. The new generation NRP power sensors do not provide the READ? query. Instead these sensors use a combination of INIT and FETCH? to read measurement results Checking for Errors The checkerror method executes the SYST:ERR? query repeatedly until no more errors are read from the instrument. The return value of this method is true if no errors were read from the instrument. The argument is the instrument's device index. if( pvp.checkerror( 0 ) == 0 ) { pvp.print( 'Errors occurred' ) } 113

114 22.14 Script-Based Measurement Displaying Scalar Results Scalar variable data can be displayed in one of the three numeric displays using the sendscalartonumericfield method. The first parameter is the value to be displayed. The second parameter is the reference name of the numeric display. The third parameter is optional and sets the unit. GUI Editor Measurement Panel var dlevel = pvp.todbm( 1e-3 ) pvp.sendscalartonumericfield( dlevel, 'Num', 'dbm' ) Displaying Array Data Numerically Array data can be displayed numerically in a table field using the sendarraytotable method. The array data is passed to the bar graph as comma-separated values in a single text string. The second parameter is the reference name of the table display. var svalues = '1e-3,1e-4,1e-5,1e-6' pvp.sendarraytotable( svalues, 'Table' ) Displaying Array Data as a Bar Graph Array data of up to 32 elements can be displayed as bar graphs using the sendarraytobars method. The data representation for the bar graph always uses a logarithmic scale (dbm). All input data must be provided as linear power values in Watts. The array data is passed to the bar graph as comma-separated values in a single string. The second parameter is the reference name of the bar graph display. var svalues = '1e-3,1e-4,1e-5,1e-6' pvp.sendarraytobars( svalues, 'Bar' ) Displaying Array Data as a Trace Array data of any length can be displayed as trace graphs using the sendarraytotrace method. The data representation for the trace always uses a logarithmic scale (dbm). All input data must be provided as linear power values in Watts. The array data is passed to the bar graph as comma-separated values in a single string. The second parameter is the reference name of the trace graph display. var SValues = '0' for( var i = 1; i < 360; i++ ) { svalues = svalues.concat( ',' ) dval = 1e-3+5e-4*Math.sin( 2*3.14*i/90 ) svalues = svalues.concat( dval.tostring() ) } pvp.sendarraytotrace( svalues, 'Trace' ); 114

115 22.18 Script-Based Measurement Displaying Array Data as a Time Series Plot Arrays data of any length can be displayed as time-series plots using the sendarraytotimeseries method. The plot's Y-axis is automatically scaled based on the maximum value in the array. The data representation for the trace always uses a logarithmic scale (dbm). All input data must be provided in linear power values in Watts. The array data is passed to the bar graph as commaseparated values in a single string. The second parameter is the reference name of the time series graph display.... pvp.sendarraytotimeseries( svalues, 'TSer' ); Displaying Text String variables can be sent to a text element on the UI. Use the sendstringtotextfield method to do this. GUI Editor Measurement Panel var stext = 'Test\nMultiLine\nText' pvp.sendstringtotextfield( stext, 'SInfo' ) Sending Data to Processing Panels Scalar variable data can be forwarded to all data sinks using the sendscalartosink instruction. The channel number must be in the range between one and four. It addresses the channel in the data-processing panels. var dvalue = 1e-3 var ichannel = 1 pvp.sendscalartosink( dvalue, ichannel ) 115

116 22.21 Script-Based Measurement Examples Triggered Average Power Measurement The triggered continuous average measurement measures a single scalar power value once a trigger event occurs. In the example below, a pulse with a duration of 50 ms occurs once every second. The pulse's power level varies slowly. The sensor should provide an individual reading for each pulse. The example script uses a measurement window of 10 ms. Each result is generated from two internal measurements (with chopper enabled) within the sensor. var dfreq = 1e9 var dtriglev = 1e-3 // connect to primary sensor pvp.definedevice( 0, '' ) // configure power measurement pvp.scpi( 0, '*RST' ) pvp.scpi( 0, 'SENS:FUNC "POW:AVG"' ) pvp.scpi( 0, 'SENS:FREQ ' + dfreq.tostring() ) pvp.scpi( 0, 'TRIG:LEV ' + dtriglev.tostring() ) pvp.scpi( 0, 'TRIG:SOUR INT' ) pvp.scpi( 0, 'TRIG:SLOP POS' ) pvp.scpi( 0, 'SENS:AVER:STAT ON' ) pvp.scpi( 0, 'SENS:AVER:COUN 1' ) pvp.scpi( 0, 'SENS:POW:AVG:APER 10e-3' ) // infinite loop do { pvp.scpi( 0, 'INIT' ) // new sensor generation var savpwr = pvp.scpi( 0, 'FETCH?' ) var davpwr = parsefloat( savpwr ) var dlevel = pvp.todbm( davpwr ) pvp.sendscalartonumericfield( dlevel, 'Av Power', 'dbm' ) pvp.sendscalartosink( davpwr, 1 ) pvp.sleep( 10 ) } while( 1 ) The screen shot below shows the readings as they were captured in the data log window: Fig : Triggered average power measurement readings. 116

117 Script-Based Measurement Burst Power Measurement The average burst measurement measures the average burst power. The burst timing is determined automatically based on the trigger threshold and the exclude times. The following signal is used with this measurement: Fig : Signal for the average burst measurement. This example script measures the average burst power in dbm. var dfreq = 1e9 var dtriglev = 0.1e-3 // connect to primary sensor pvp.definedevice( 0, '' ) // configure pvp.scpi( 0, pvp.scpi( 0, pvp.scpi( 0, pvp.scpi( 0, pvp.scpi( 0, pvp.scpi( 0, pvp.scpi( 0, pvp.scpi( 0, pvp.scpi( 0, pvp.scpi( 0, pvp.scpi( 0, power measurement '*RST' ) 'SENS:FUNC "POW:BURST:AVG"') 'SENS:FREQ ' + dfreq.tostring() ) 'TRIG:LEV ' + dtriglev.tostring() + ' W' ) 'TRIG:HYST 3 DB' ) 'SENS:POW:BURST:DTOL 1e-6' ) 'SENS:TIM:EXCL:START 10e-6' ) 'SENS:TIM:EXCL:STOP 10e-6' ) 'SENS:AVER:COUN 4' ) 'SENS:AVER:STAT ON' ) 'SENS:AVER:TCON REP' ) // infinite loop do { pvp.scpi( 0, 'INIT' ) // new sensor generation var savpwr = pvp.scpi( 0, 'FETCH?' ) var davpwr = parsefloat( savpwr ) var dlevel = pvp.todbm( davpwr ) pvp.sendscalartonumericfield( dlevel, 'Burst Av', 'dbm' ) pvp.sendscalartosink( davpwr, 1 ) pvp.sleep( 100 ) } while( 1 ) 117

118 Script-Based Measurement Buffered-Mode Measurements In buffered mode, the sensor captures a predefined number of measurements and then sends all results back to the PC. In the example below, a single-shot sequence of 20 bursts is measured using an individual trigger for each burst. Fig : Buffered-mode measurement of a 20-burst sequence. The INIT command starts the measurement cycle. The sensor evaluates 20 trigger events and sets a flag once the measurement has been completed. The FETCH? query synchronizes to the measurement completion state. var dfreq = 1e9 var dtriglev = 1.0e-5 // -20 dbm // connect to primary sensor pvp.definedevice( 0, '' ) // configure pvp.scpi( 0, pvp.scpi( 0, pvp.scpi( 0, pvp.scpi( 0, pvp.scpi( 0, pvp.scpi( 0, pvp.scpi( 0, pvp.scpi( 0, pvp.scpi( 0, pvp.scpi( 0, pvp.scpi( 0, pvp.scpi( 0, pvp.scpi( 0, power measurement '*RST' ) 'SENS:FUNC "POW:AVG"' ) 'SENS:FREQ ' + dfreq.tostring() ) 'TRIG:LEV ' + dtriglev.tostring() ) 'TRIG:SOUR INT' ) 'TRIG:SLOP POS' ) 'TRIG:DELAY 0.1e-3' ) 'TRIG:COUNT 16' ) 'SENS:AVER:STAT ON' ) 'SENS:AVER:COUN 2' ) 'SENS:POW:AVG:APER 2e-3' ) 'SENS:POW:AVG:BUFF:SIZE 16' ) 'SENS:POW:AVG:BUFF:STATE ON' ) pvp.scpi( 0, 'INIT' ) var svalues = pvp.scpi( 0, 'FETCH?' ) // new sensor generation pvp.print( 'Result: ' + svalues ) pvp.sendarraytobars( svalues, 'Readings' ) 118

119 Script-Based Measurement Connecting to the R&S NRP2 Base Unit The script may also be used to access other Rohde&Schwarz instruments such as the NRP2 base unit. The following example code demonstrates how to read numeric data from channel A and B. The results are sent to numeric displays. // NRP2 base unit connected via LAN pvp.definedevice( 1, 'TCPIP:: ::INSTR' ) pvp.scpi( 1, '*RST' ) pvp.scpi( 1, '*CLS' ) pvp.scpi( pvp.scpi( pvp.scpi( pvp.scpi( pvp.scpi( 1, 1, 1, 1, 1, 'SYST:SPEED FAST' ) 'SENS1:AVER:COUNT:AUTO OFF; :SENS2:AVER:COUNT:AUTO OFF' ) 'SENS1:AVER:STAT OFF; :SENS2:AVER:STAT OFF' ) 'INIT1:CONT OFF; :INIT2:CONT OFF' ) 'SENS1:FREQ 1 GHz; :SENS2:FREQ 1 GHz' ) if( pvp.checkerror( 1 ) ) { do { pvp.scpi( 1, 'INIT1; :INIT2' ) var sresults = pvp.scpi( 1, 'FETCH1?; FETCH2?' ) var sresultlist = sresults.split( ';' ) dval1 = parsefloat( sresultlist[0] ) dval2 = parsefloat( sresultlist[1] ) pvp.sendscalartonumericfield( dval1, 'P1', 'dbm' ) pvp.sendscalartonumericfield( dval1, 'P2', 'dbm' ) pvp.sleep( 200 ) } while( 1 ) } 119

120 Script-Based Measurement Connecting to the R&S NRT Base Unit The script may also be used to access other Rohde&Schwarz instruments such as the NRT base unit. The following example code demonstrates how to read numeric data from this base unit. The results are printed to the script output window. pvp.definedevice( 0, 'GPIB0::28::INSTR' ) pvp.scpi( '*RST' ) pvp.print( pvp.scpi( 1, '*IDN?' ) ) pvp.print( pvp.scpi( 1, 'TEST:SENS?' ) ) pvp.scpi( pvp.scpi( pvp.scpi( pvp.scpi( pvp.scpi( 1, 1, 1, 1, 1, 'SENS1:FREQ 1e9' ) 'SENS1:FUNC:CONC OFF' ) 'UNIT1:POW:REL:STAT OFF' ) 'UNIT1:POW W' ) 'SENS1:FUNC "POW:FORW:AVER"' ) // // // // no concurrent meas. no relative meas. read watts mode pvp.scpi( 1, 'TRIG; *OPC?' ) sresult = pvp.scpi( 1, 'SENS1:DATA? "POW:FORW:AVER"') pvp.print( 'Result = ' + sresult ) 120

121 23 Data Processing Panels Data Processing Panels Power Viewer Plus distinguishes between measurements and data processing panels. Measurements (continuous, trace, etc.) generate data, whereas the data processing panels only receive these measurands. Internally, all measurands are supplemented with an exact time stamp and sent to the data processing panels. These panels feature input filters and sort out values of interest. The figure below shows the principle architecture implemented in Power Viewer Plus. Processing Measurements Continuous Bar Chart Trace Data Log Statistics Value Timestamp Unit Timeslot Limit Analysis Multi-Ch. Gauges Scripting Fig. 23.1: Power Viewer Plus data flow. Each data processing panel can be configured individually. Therefore, different data-processing panels could evaluate different data simultaneously. The input filter of each data processing panel is configured by the following parameters. Source Measurand 1 4 The Source denominates the measurement from which the panel may receive data. Depending on which measurement this is, there is a corresponding set of measurands that can be assigned to the dataprocessing panel's input channels. 121

122 Data Processing Panels The table below lists data that is available to the data-processing panels for each measurement. Cont. Multi Ch. Timeslot Statistics Trace Script Pav Ppk Pav, rel Pav, rel Average Ch M Mkr 1 Mkr 2 Mkr 3 Mkr 3 Pav n% Gate Pav Gate Ppk Mkr Ref P Mkr Dn dp Mkr Dn dt Rise Time Fall Time Pulse Width Period Time Pulse Ptop Pulse Ppk Ch W W 1 % 1 W W W W W W db W W W 1 s s s s s dbm dbm Average power Peak power Average power, relative Average power, relative Average filter count Measurement channel Math channel Marker 1 reading Marker 2 reading Marker 3 reading Marker 4 reading Average power Reading at n % Gated measurement, average Gated measurement, peak Reference marker, power Marker n, power ratio Marker n, time difference Pulse rise time Pulse fall time Pulse width Pulse period time Pulse top power Pulse peak power Channel

123 23.1 Data Processing Panels The Data Log The Data Log panel captures up to four scalar measurements over a period of several days. A maximum of data points is available for memory recordings in the log panel. Writing captured data to a file places no restrictions on the number of recorded samples (other than file size restrictions imposed by the underlying operating system). Memory recording and file recording are independent and can be configured separately. The log panel shows a preview of all data points. If multiple measured values fall within the time period covered by one video point, statistics containing minimum, maximum, and average values are automatically generated. X-zoom area Y-unit Leveling mode View mode Time-line data Level lines Channel name Write indicator Time-line indicator Measurement and time information Fig : Data log panel. 123

124 Data Processing Panels Settings The data-log panel has its own settings dialog similar to the one for measurements. Unlike measurements, the data log does not generate data but rather processes measurands from active measurements. Fig : State controls. Start / Stop The data log process does not start automatically when a measurement is started; it needs to be activated separately. This allows the user to setup the measurement first and then start the log. Starting the log erases all previous data from the log memory. The data log's capturing state is indicated by a hard disk icon in the lower left corner of the application window. The data log is not active. The data log captures data. Fig : Feed definitions. Feeds First, set the Source to select the measurement that feeds data into the log panel. Second, select up to eight measurands for the eight data-log channels. Power values are normally received in watts but may be converted into dbm. 124

125 Data Processing Panels Fig : Capture time settings. Max. Capture Time The maximum capture time can be set to a period of up to seven days. The data points in the panel's internal buffer are spaced evenly across the capture time. Once the capture time has elapsed, the data log is automatically deactivated, and no more data is recorded. Please note that the data recorder captures up to data points to memory. Additionally, live file writing can be activated with no limit for the number of data points. These two processes are independent of each other. The data-log panel's captured data may also be saved to a file with the save button from the application's tool bar. The output format is ASCII.csv, which saves data in individual columns that are separated by commas. Fig : Live file output settings. Log Data to File The data log can be configured for writing measured data to a file while the measurement is running. Writing takes place instantaneously and may, therefore, slow down the measurement. Interval The interval sets the time period for collecting sensor readings. At the end of each time interval the software writes a record to the log file. This record contains the minimum, maximum, and average value of the accumulated readings. The time interval should be set such that the file size remains within acceptable limits. File Name The file name specifies the file to which data is appended during the log process. The file format is.csv. Besides regular file names, there is a set of place holders that can be used within the file name. $t $h %m %d %t The temporary directory The user's home directory The name of the measurement The day when the data log was started The time when the data log was started Example: $t\pvp-%m-%d-%t.csv 125

126 Data Processing Panels This creates log files with the name pvp-continuous csv if used with the "continuous" measurement mode. The storage location is the temporary folder set in the operating system, e.g. C:\TEMP The Context Menu The data-log panel provides a context menu that can be activated by right-clicking in the graphics area. When multiple traces are visible, the context menu is invoked for the trace that is located at the current mouse position. Fig : Context menu for the data-log panel Auto Level This This option sets the current trace back to full auto leveling. This is useful when previous zoom actions were used to magnify trace details. Entire Time This menu item sets all traces back to the entire log time. This option is useful when previous zoom actions were used to magnify trace details. The time axis is always common for all traces. Therefore, this action involves all traces. Auto Scale All This option combines both of the above options. It sets the x-axis back to the entire log time and returns all traces back to full auto leveling. Show Statistics The data log always uses data points evenly spaced across the total log time. Each data point represents a bin that may contain multiple measured values. To prevent the loss of any information, each bin is represented by the average, minimum, and maximum value that was received. By default, the envelope of the recorded data is displayed (minimum and maximum). If this is not desired, the statistics view can be disabled so that it only shows average data. 126

127 Data Processing Panels Lines This menu item activates level lines and time lines. The functionality of these lines is similar to the ones described in the chapter 15.1, "Trace Measurements." Time Lines to Window Places the time lines back into the viewing area at the 10 % and 90 % points. Clear All Data This menu entry stops the data log and removes all data from the log. To prevent unintentional deletions, the user is queried before the log is stopped and data is erased The Time Line Indicator The time-line indicator is located below the data log traces. It shows the overall log time and the currently visible area in case a trace section is magnified using the zoom operations. Fig : The time-line indicator. Additionally, the time-line indicator displays a small marker at the current write position as well as the position of the two time lines Zooming Zooming can be performed in the x- or y-direction separately by using the left mouse button and the mouse wheel. Y-direction Select the center point for the y-zoom and click once using the left mouse button. A zoom-point indicator is placed at this point. Turn the mouse wheel forward to zoom in. Turn the mouse wheel backwards to zoom out. Press the left mouse key again to disable zoom mode and remove the zoom-point indicator. X-direction Position the mouse at the beginning of the x-range that should be magnified. Press the left mouse key and hold it while you move the mouse to the right side. Release the left mouse key at the right end of the x-zoom area. A semi-transparent area marks the zoom range. Turn the mouse wheel forward to zoom into the highlighted area. The the system remembers the x-zoom ranges. Turning the mouse wheel backwards restores the last display range. 127

128 23.2 Data Processing Panels Limit Monitoring The limit monitoring panel receives up to 16 scalar measurands and compares them against warning and error thresholds. The limit monitoring panel does not take measurements itself; instead, it evaluates data that is generated by the measurement panels (continuous, trace, timeslot, etc.). The screen shot below shows the limit-monitoring panel receiving data from a continuous power measurement. Fig : Limit monitoring panel. The limit monitor continuously compares incoming data against high and low warning thresholds and error thresholds. Each time a measurand passes one of these thresholds, the software generates an entry in the incident log. The capacity of this log is limited, and it only displays the most recent entries. Besides playing an acoustic warning signal, the limit monitor can also log the incidents in a file or send them via TCP/IP to a remote host. 128

129 Data Processing Panels Settings Fig : The limit monitor settings. Enable Enables or disables limit monitoring globally. Source / Channels These settings define the origin of the measurands that are to be monitored for limit violations. Alias Name Each channel can use an alias name instead of the combination of measurement and measurand. Fig : Settings for warning and error levels. 129

130 Data Processing Panels Warning Thresholds / Error Thresholds When these functions are enabled, the system compares the measurand against the upper and lower limit values. Each time the measurand passes one of these thresholds, an entry is generated in the incident log. Power values can be entered in watts (default) or in dbm. Fig : Action settings for responses to threshold incidents. Actions By default, incidents are logged to the incident log. The capacity of this log is limited, and it only displays the most recent entries. The action settings are used to enable additional features, such as writing to a log file or sending incidents to a remote host. Log to File A file entry is generated each time a limit violation occurs or ceases. New incidents are appended to the end of the log file. Writing occurs immediately. Consequently, this can slow down the measurement rate. TCP/IP Server The limit monitor can be configured to start a TCP/IP server process for as long as a measurement is active. Remote applications can connect to this server and receive incidents. Play Sound An acoustic signal is generated each time a limit violation occurs or ceases Configuring the Server The server process allows for one TCP/IP connection from a remote host. Once this connection is established, all incidents are sent to the remote host as text messages. The limit monitor starts the server when the measurement is started and stops it when the measurement is stopped. Only one client connection is allowed at a time. Note: Running the server process may present a security risk to your IT network. By default, the server is configured to only allow connections from the local host ( ). It may also be required to configure your local firewall software to allow for inbound traffic on the port used by the server. 130

131 Data Processing Panels Fig : The server settings dialog. Host IP Sets the IP address for the network interface at which the server is listening for incoming connections. By default, the IP address is used for the local machine. Port Sets the port at which the server listens for incoming connections. It must be ensured that the port is not already being used by another service. The command: netstat -an find /i listening provides a list of all ports that the system is currently listening to. Access Rules The access rules define which clients are allowed to connect to the limit monitoring server. For this purpose, two lists of IP addresses define hosts that are denied and hosts that are allowed to connect. The order in which these two lists are evaluated can be selected. The second list to be evaluated takes precedence over the list that is evaluated first. Each list may contain the keyword "All" or a comma-separated list of IP addresses, or fractions of IP addresses. Example: Order: Deny: Allow: Deny, Allow All In the example above, the deny list is evaluated first. The keyword "All" indicates that any incoming request will be rejected. Second, the allow list is evaluated. All servers that match x.x are allowed to connect. Require Authentication When this authentications setting is enabled, the server prompts a 131

132 Data Processing Panels connecting client for a login and a password. Server status messages are shown on the application's message-log panel. This window can be activated from the menu bar by selecting Window Message Log Client Connections The telnet and netcat tools can be used as clients for connecting to the limit-monitoring server. The client is invoked from the command line using one of the following commands: telnet <ip-address> <port> nc <ip-address> <port> If the server can be reached, a welcome message is printed and the user may be requested to enter the authentication data. Fig : Client communication. Note: Netcat is a very versatile tool, and it can also be used to install malicious "backdoors" on the host PC. For this reason, most virus protection software packages classify nc.exe as a threat and disable its execution. The limit monitoring server uses port by default. This port is not typically used by other applications. The default telnet port is

133 23.3 Data Processing Panels Statistical Analysis The analysis panel receives up to four scalar measurands and performs a statistical analysis on this data. Similar to the data log, the analysis panel does not take measurements itself but merely evaluates data that is generated in the measurement panels (continuous, trace, timeslot, etc.). The screen shot below shows the analysis panel receiving data from a trace measurement (pulse analysis). Fig : The analysis panel. Each of the four views within the analysis panel can be configured to a histogram display or a quantile plot. The four analysis views operate independently and can be freely assigned to any available measurand. By default, each view is configured to the histogram display and for evaluation of the last 1000 measurements. These settings can be changed using the context menu for the individual view. 133

134 Data Processing Panels Histogram Display The histogram sorts the measured values into categories (data bins) that are evenly distributed between the minimum and maximum reading. The result is displayed as a bar chart, with the height of the bars indicating how many measurements fall into each category. The number of samples that are used for evaluation can be selected to be 250, 1000, or Min and max power Statistics Count Average power Standard deviation Fig : Histogram display. The minimum (Min) and maximum (Max) power readings are displayed in the upper left corner. The count (Cnt) indicates how many readings were accumulated for the analysis. The count remains at a constant value as soon as the set number of readings has been reached. The average (Avg) power of all accumulated readings and the sample standard deviation (s) is displayed in the panel's upper right corner. The following formulas are used to calculate these two parameters: 1 x = N s= N i =1 xi N 1 x x 2 N 1 i=1 i The terms L95 and U95 are used for the 95 % confidence intervals for the average power level: L95 x 2 s N U95 x 2 s N 134

135 Data Processing Panels Q-Q-Plot The Q-Q-Plot (Quantile-Quantile-Plot) is a graphical method for comparing two probability distributions. The Power Viewer Plus software provides a normal probability plot that compares the probability distribution of the measured values against a normal distribution. The Q-Normal-Plot is, therefore, used as a graphical test for normal distribution. Fig : Q-Q-Plot display. The outer pair of red lines marks the 95 % confidence band. If all graph points are located within this band, the normal probability test is positive at a 95 % confidence level. The diagram is vertically and horizontally scaled to steps of one σ. When all measured values are distributed normally, all graph points are located on a straight line. Departures from this straight line indicate that the normal distribution model is a poor fit for the distribution of the measured values. Fig : Example distribution. Fig shows a non-ideal normal distribution. The number of measured quantiles for higher values (on the right side in the histogram) is larger than it should be for normally distributed data. The Q-Q-Plot shows this deviation in the upper right corner. 135

136 Data Processing Panels The Context Menu Each view has its own context menu that is used to configure the data representation in this view. Fig : Context menu for a view. Points The view always captures the last 5000 measured values, but the number of evaluated samples can be set to the last 250 or 1000 values or to all 5000 of them. Changing the number of evaluated samples does not erase any data. Histogram / Q-Q-Plot These settings determine the view to the histogram or the quantiles display. Both representations are calculated from the same data, and changing the representation does not erase any data. Clear Clears all data in this view. After the data is cleared, the message Waiting for data... appears in the quantile plot representation until 50 samples have been collected. In the histogram representation, the histogram is not painted for less than 50 samples. 136

137 Data Processing Panels Analysis-Panel Settings The analysis panel has its own settings dialog similar to the measurements. Unlike measurements, the analysis panel does not generate any data but merely captures readings from active measurements. Fig : Analysis-panel settings. Start / Stop The analysis process does not start automatically when a measurement is started. Instead, it needs to be activated separately. This allows the user to setup the measurement before starting the analysis. Starting the analysis erases all previous data from the views. Auto Start / Stop The auto start/stop option automatically starts the analysis process when a measurement is started and valid data is received. The analysis is topped when the measurement stops. Clear All This button clears the data in all analysis views. Default All This button sets all views back to the histogram display of the last 1000 measurements. It also deletes the data in all views. Feeds First, set the Source to select the measurement that feeds data into the log panel. Second, select up to four measurands for the four data log channels. 137

138 23.4 Data Processing Panels Gauges Panel The gauges panel receives up to four scalar measurands and displays the value on a sizable gauge. Similar to the analysis, the gauges panel does not take measurements itself but merely evaluates data that is generated in the measurement panels (continuous, trace, timeslot, etc.). The screen shot below shows the gauges panel receiving data from a trace measurement (pulse analysis). Fig : Gauges-panel. Each of the four views within the gauges panel can be configured to a digital or analog display. The four gauge views operate independently and can be freely assigned to any available measurand. By default, each view is configured to the digital display. This setting can be changed using the context menu for the individual view Analog Meter The needle of the analog meter displays the filtered measurand. The filter is an exponential moving average filter with 8 bins. Fig : Context menu for a view. The blue and red dots show the minimum and maximum value in the moving average filter buffer. The text AUTO in the upper right corner indicates that the meter is currently set to auto scaling. This works in most cases but it may also be desirable to set a specific scale. Setting the scale is possible by using the muse wheel and the Ctrl key. 138

139 Data Processing Panels Mouse Wheel Changes the upper border of the meter. This option is only available for logarithmic meters (dbm scale). Ctrl + Mouse Wheel Pressing and holding the Ctrl key replaces the unit text in the center by the currently selected measurement range. Turning the mouse wheel while the Ctrl key is pressed increases or decreases the measurement range The Context Menu Each view has its own context menu that is used to configure the data representation in this view. Fig : Context menu for a view. Digital The view shows a numeric value on a digital display. The values are displayed as they are send by the measurement. Absolute power values in Watts are automatically converted to dbm. Analog Turns the view into the analog meter mode. Historic Enables a digital display using ancient Nixie tubes Gauges-Panel Settings The gauges panel are identical with the analysis panel settings. 139

140 24 Updating NRP and NRP-Z Sensor Firmware Updating NRP and NRP-Z Sensor Firmware Power Viewer Plus provides a dialog for updating the R&S NRP and R&S NRP-Z sensor firmware. You can call up the dialog for this by selecting Sensor Firmware Update from the menu bar. Please ensure that the following requirements are met before attempting to update the sensor firmware. Download the firmware file from the Rohde & Schwarz website and store it on a local hard drive. Do not use a network storage location. The file extension is.nrp for all NRP-Z sensors and.rsu for all NRP sensors. Ensure that the selected firmware file fits the sensor model. Do not interrupt the update process. Wait until you either get an error message or the Done message and the update is completed. Do not run or start any other software accessing the sensor during the firmware update process. Do not start or use any other applications that may cause a high CPU load during the firmware update process. USB Sensors Make sure that you are able to operate the sensor without getting any USB interface errors. Test your PC's USB interface with the sensor before you attempt to start the update process. Ensure that the USB port that the sensor is connected to provides enough power. Do not use simple office-quality USB hubs, because they often use thermal fuses that could be tripped when too much current is consumed. Instead, use a direct USB port on your PC. Rohde & Schwarz recommends having only one sensor connected during the update process. Remove any unneeded USB hardware and any other R&S USB sensors before starting the update process. This is not a requirement for a regular firmware update process, but it lowers the risk of overloading the USB interface by drawing too much current. Do not use laptops for the update, and ensure that power saving is disabled. If you need to use a laptop and you encounter errors (timeout messages) during the update, it is advisable to connect the sensor via a USB hub that uses its own power supply. This is required, because some laptop computers use USB hardware that issues regular suspend commands that could interrupt the firmware update process. Do not use any USB extenders (such as USB LAN adapters, etc). 140

141 Updating NRP and NRP-Z Sensor Firmware Once all the above requirements are met, the firmware update process can be started. Rohde & Schwarz recommends following the steps below: Setup the sensor connection and wait for the operating system to update the USB drivers (if required). Start Power Viewer Plus. Verify that the sensor was detected during the initial scan. Do not start any measurements, but open the firmware update dialog from the menu bar. Ensure that the correct sensor is selected. Ensure that the correct sensor information is listed. Verify the sensor type and serial number. Select the sensor firmware file (.nrp or.rsu). Press the Start button and wait until the update sequence has finished. Fig. 24.1: Dialog for updating firmware Recovering from a Failed Update After an unsuccessful update, the sensor is usually still accessible via USB, but it will no longer operate normally. In this case, the update process needs to be repeated. 141

142 25 Using S-Parameter Profiles Using S-Parameter Profiles This functionality is currently only available in the Windows-based version of Power Viewer Plus. S-parameter correction compensates for the losses and reflections introduced by a component such as an attenuator, directional coupler, or matching pad that is attached to a power sensor. Using S-parameters instead of a fixed offset increases measurement accuracy by accounting for the interaction between the sensor and the component. As a result, the sensor's reference plane shifts from the sensor's RF connector to the input of the device being applied externally. This procedure is also referred to as embedding a device. Fig. 25.1: Schematic view of embedding a device S-parameter data sets are stored in the sensor's internal calibration data structure. As a result this data is not lost during a device reset or a power cycle. In addition, S-parameter data can be configured such that it becomes active automatically when the sensor is powered up. This feature is useful if a two-port device such as an attenuator pad is permanently attached to a sensor. Fig. 25.2: Schematic view of scattering parameters The equation below includes the S-parameter device, the power sensor's reflection coefficient, and the source reflection coefficient. The sensor firmware uses this equation when the source reflection coefficient is also specified (when gamma correction is enabled). bg =b 2 [ (1 s 22 Γ S )(1 s 11 Γ G ) ΓG Γ S s 12 s21 ] The measurement error caused by using a simple offset table instead of the s-parameter correction can be calculated by the following equation: Error % ±100 % [( 2 (1 s 22 Γ S )(1 s 11 Γ G ))2 1 ] 142

143 Using S-Parameter Profiles Power Viewer Plus provides a dialog for updating s-parameter profiles in R&S NRP and R&S NRP-Z power sensors. You can call up the dialog for this by selecting Sensor S-Parameter from the menu bar. S-parameter data is stored permanently in the sensor flash memory. All s-parameter data is organized in profiles. Each profile consists of the complex correction values, a mnemonic, a level range and additional flags. All R&S NRP and R&S NRP-Z power sensors support s-parameter correction (single profile) but some models also support switching between different profiles. In all R&S NRP-Z sensors the s-parameter profiles are part of the factory calibration data set. The new generation R&S NRP sensors contain a factory calibration data set and a user data set. For safety reasons Power Viewer Plus only allows access to the user data set in these sensors. Fig. 25.3: S-Parameter Profile Editor Initialize from Sensor Reads the entire calibration data structure from the sensor flash memory and extracts the s-parameter device data. S-parameter sets are always stored as part of the sensor non-volatile calibration data. Reading the calibration data set does not alter any data in the sensor. Load Backup File Power Viewer Plus always creates a backup copy of the calibration data before attempting to update the sensor flash memory. If the write operation fails, e.g. due to a power loss during the write operation the data can simply be recovered from the backup file. Power Viewer Plus 143

144 Using S-Parameter Profiles compares the sensor type and serial number in the backup file with the selected sensor. If the numbers disagree the backup file cannot be loaded. Write to Sensor Updated s-parameter data sets must be written back to the sensor flash memory in order to become effective. Before the flashing process starts Power Viewer Plus creates a backup copy of the existing calibration data set on the local hard drive. Flags: Lock s-parameter state and device Locking the s-parameter state activates or deactivates the s-parameter state permanently. The user cannot change the state and the related controls are grayed out in the Power Viewer Plus application. The central section of this dialog contains all the controls required for the s-parameter management. A prerequisite for these controls to become active is that a data set was successfully loaded from a sensor or a matching backup file. Flags: Use flags from USER data set All new generation R&S NRP power sensors contain two data sets that may contain s-parameter profiles. The factory data set contains sensor calibration data and is written during production. This data set may contain s-parameter profile information, for example if sensors are shipped with a permanently attached attenuator. The second data set is the user data set. Power Viewer Plus only allows access to this user data set for all new generation R&S NRP power sensors. Setting this flag overrides the Lock and Default flag from the factory data set. Use this flag if the s-parameter state is locked in the factory data set and you need to use other s-parameter profiles than the one stored in the factory data set. A locked s-parameter data set typically disables all related controls in the correction dialog. S-parameter profiles residing in the user data set start at index 1001 whereas profiles from the factory data set start at index one. Current Profile This control contains all s-parameter profiles that are currently available in the calibration data set. New empty records can be added to the list by using the New button. Existing records can be removed from the list with the Delete button. These changes do not become effective immediately in the sensor. Instead all modifications are applied to a memory copy of the calibration data set. Once changes are made a sensor update is required and can be initiated by clicking the Write to Sensor button. 144

145 Using S-Parameter Profiles Mnemonic The mnemonic defines a name that is used to name an s-parameter device. The mnemonic is also stored as part of the sensor s-parameter profile. Load / Save Loads or saves s-parameter data from or to a file. If data is loaded all existing s-parameter data in the selected profile is lost. Saving data does not alter the s-parameter profile. View Starts the viewer dialog and displays all s-parameter data of the currently selected profile. Fig. 25.4: S-Parameter Viewer Frequency Range This line displays the frequency range covered by the s-parameter set. S21 Range This line displays the minimum and maximum s21 value in the s-parameter set. Power Limits Depending on the loss or gain introduced by the s-parameter device the sensor's measurement power range changes. These settings define the upper and lower sensor power limit if the s-parameter profile is selected during a measurement. Power Viewer Plus populates both entry fields with recommended values when s-parameter data is loaded from a file. Active by Default This option automatically enables the selected s-parameter profile when the sensor is powered up. This option is useful if components such as attenuator pads remain attached to a sensor. 145

146 26 VXI PnP Programming Guide VXI PnP Programming Guide The Power Viewer Plus software is based entirely on the Rohde & Schwarz VXI PnP drivers for the R&S NRP power sensors. This driver provides a C function interface and is recommended for all user applications. Please see the NRP software download section for the latest version of the VXI PnP driver. Custom applications must include the rsnrpz.h file when using the VXI PnP driver. This driver uses VISA data types, but if VISA is not installed, the required data types are defined in the header file itself. It should be noted that multiple applications cannot access the NRP driver simultaneously. The minimum requirement for using the VXI PnP driver functions is installation of the R&S NRP Toolkit. The toolkit package contains the USB drivers as well as the NrpControl2 low-level driver DLL. Generally, it is possible to build applications with the rsnrpz.c and rsnrpz.h files directly compiled into the application. In this case, the application only depends on NrpControl2.lib. Alternatively, the application can include rsnrpz.h and link against rsnrpz.lib. The following diagram shows the NRP-Z driver architecture and possible application options. Application Application (rsnrpz.h / NrpControl2.lib) (rsnrpz.h / rsnrpz.lib) rsnrpz_32.dll, rsnrpz_64.dll rsnrpz.c NrpControl2.dll, NrpControl2_64.dll USB Driver Fig. 26.1: Driver architecture on Windows-based systems Sensor Resource Strings The power sensors are identified by a unique VISA resource string. This string is passed on to the rsnrpz_init() function in order to open the sensor connection. The resource string has the following format: USB::0x0AAD::<usb-id>::<serial> The value 0x0AAD is the Rohde & Schwarz vendor ID, and it cannot be changed. The USB ID is unique for each sensor type. A list of USB device ID numbers is provided in chapter 6.5, "Supported R&S NRPZ Sensors." The serial number is the serial number for the individual sensor. 146

147 26.2 VXI PnP Programming Guide Numeric Results All numeric power readings are provided in watts. The reading includes a level offset that is either set by the user or provided by the Sparameter device. In many applications, however, it is desirable to show power readings on a logarithmic scale. Conversion can be performed by taking the absolute value of the power reading, adding a very small offset, and then using the log10() function: double dvaldbm = * log10( fabs( dvalw ) + 1e-32 ) The small offset ensures that the log10 function's argument can never be exactly zero Opening and Closing the Sensor Connection The first step is to open the sensor connection. The following code lines demonstrate how this is done using the driver functions. The example sets the USB timeout to 5 seconds and does not reset the sensor. The return value is the USB session that must be used in all further communication with the sensor: ViStatus lerr ViSession ulusbsession rsnrpz_settimeout( 5000 ) lerr = rsnrpz_init ( <USB resource string>, false, false, &ulusbsession ) if( lerr!=0 )...error handling... If the sensor is no longer needed for further measurements, the connection should be closed. The following lines demonstrate how to do this. After the sensor is closed, the session number must not be used anymore: lerr = rsnrpz_close( ulusbsession ) ulusbsession = Multiple Sensors Multiple sensors may be opened simultaneously within one application. The rsnrpz_init() function must be called once for each sensor using the sensor resource string. The USB session ID numbers returned from this function are then used to access the individual sensors. 147

148 26.5 VXI PnP Programming Guide Error Handling Most rsnrpz_ driver functions return an error code. If the function call is successful, the return value is zero. If an error was returned, further information may be obtained from the rsnrpz_error_message() function. This function translates the error code into a human-readable text message: char szmessage[256] rsnrpz_error_message( ulusbsession, lerr, szmessage ) Additionally, errors that arise inside the sensor can be queried from the sensor error queue. The rsnrpz_error_query() should be called as long as the lerr2 return variable is not equal to zero and the return code lerr is zero: lerr = rsnrpz_error_query( ulusbsession, &lerr2, szmessage ) Please note that both functions require a valid session number. If the rsnrpz_init() function fails and no valid session number is available (session is zero), these functions cannot be used. In most cases, this indicates that the sensor has already been opened by another application, or that the sensor firmware is outdated Zeroing Zeroing the sensor might be required if very low signal levels need to be measured. The time required for the zeroing procedure varies from sensor to sensor. It must also be noted that the zero offset value is not permanent. A sensor reset command does not clear the zero offset, but a power loss requires re-zeroing. The following example starts the zeroing process with the function rsnrpz_chan_zero(). This function returns immediately. The following call to rsnrpz_chan_iszerocomplete() determines the completion state of the zeroing process. The function should be called repeatedly in a loop, but it must be ensured that there is enough CPU time available for the driver process ( Sleep(), SwitchToThread() ). Using a short sleep that gives the CPU away to the background thread is recommended. If the zeroing fails, an error code that is not equal to zero is returned: lerr = rsnrpz_chan_zero( ulusbsession, 1 ) unsigned short usmeascompleted lerr = rsnrpz_chan_iszerocomplete( ulusbsession, 1, &usmeascompleted ) 148

149 26.7 VXI PnP Programming Guide Identifying a Sensor (*IDN?) It is often necessary to identify a sensor and retrieve its type and serial number. The *IDN? command is available in the SCPI language for this task. The C-based VXIPnP driver provides functions that read extended sensor information and can be used for this purpose: char szmanuf[256]; lerr = rsnrpz_chan_info( ulusbsession, 1, "Manufacturer", 256, szmanuf ); char sztype[256]; lerr = rsnrpz_chan_info( ulusbsession, 1, "Type", 256, sztype ); char szserial[256]; lerr = rsnrpz_chan_info( ulusbsession, 1, "Serial", 256, szserial ); printf( "%s,%s,%s", szmanuf, sztype, szserial ); 26.8 Continuous Average Power Measurement This example demonstrates how to implement a simple continuous average power measurement. First, the sensor's operation mode needs to be set. This step only needs to be performed once when multiple average power measurements are required: lerr = rsnrpz_chan_mode( ulusbsession, 1, RSNRPZ_SENSOR_MODE_CONTAV ) Second, the carrier frequency must be set. Setting the carrier frequency is always required for precise measurements: lerr = rsnrpz_chan_setcorrectionfrequency( ulusbsession, 1, dcarrierhz ) 149

150 VXI PnP Programming Guide In many cases, the power sensor is not directly connected to the DUT and compensation must be made for additional cable loss. The two functions described below enable and set the level offset. Please note that further level-related commands expect levels that include the correction factor: lerr = rsnrpz_corr_setoffset( ulusbsession, 1, doffsetdb ) lerr = rsnrpz_corr_setoffsetenabled( ulusbsession, 1, true ) The aperture time is the time for which the sensor integrates the signal in order to generate a a single sample. Normally, the sensor uses a default aperture that is best for noise and measurement speed. When measuring AM modulated signals with a known period time, it is advisable to set the aperture time to multiples of the period time. In such cases, using a low averaging filter count, such as two, can deliver stable measurement results: lerr = rsnrpz_chan_setcontavaperture( ulusbsession, 1, dwindows ) The averaging filter can be configured to either manual mode or automatic mode. The auto mode example below sets the filter to a 0.01 db resolution: lerr = rsnrpz_avg_configureavgauto( ulusbsession, 1, 3 ) Alternatively, the averaging filter mode can be set to a fixed value, such as 2, 4, 8, or 16. This setting is best if a constant measurement time is required and the signal level does not change much: lerr = rsnrpz_avg_configureavg( ulusbsession, 1, ulavcnt ) The averaging filter can be configured to repeating mode or moving filter mode. In repeating mode, each measurement cycle initially clears the filter and then accumulates measurements until the filter is entirely filled: lerr = rsnrpz_avg_setterminalcontrol( ulusbsession, 1, RSNRPZ_TERMINAL_CONTROL_REPEAT ) 150

151 VXI PnP Programming Guide The rsnrpz_chan_initiate() function starts one measurement cycle. The function returns immediately. Therefore, the application must subsequently poll the sensor for measurement completion: lerr = rsnrpz_chan_initiate( ulusbsession, 1) The completion state should be polled in a loop, but it must be ensured that the CPU is made available to the driver thread between subsequent polls: Sleep( 100 ) or SwitchToThread() ViBoolean bmeascompleted lerr = rsnrpz_chan_ismeasurementcomplete( ulusbsession, 1, &bmeascompleted ) When the measurement cycle has completed successfully, the result can be read, and a new measurement cycle may be started: ViReal64 fmeasresult lerr = rsnrpz_meass_fetchmeasurement( ulusbsession, 1, &fmeasresult ) 151

152 26.9 VXI PnP Programming Guide Trace Measurements This example demonstrates how to implement a trace measurement for a repeating signal that provides a stable trigger condition. First, the sensor's operation mode needs to be set. This step is only required initially: lerr = rsnrpz_chan_mode( ulusbsession, 1, RSNRPZ_SENSOR_MODE_SCOPE ) Second, the carrier frequency must be set. Setting the carrier frequency is required for precise power measurements: lerr = rsnrpz_chan_setcorrectionfrequency( ulusbsession, 1, dcarrierhz ) In many cases, the power sensor is not directly connected to the DUT, and compensation must be made for additional cable loss. The following two functions enable and set the level offset: lerr = rsnrpz_corr_setoffset( ulusbsession, 1, doffsetdb ) lerr = rsnrpz_corr_setoffsetenabled( ulusbsession, 1, benoffset ) The function below configures the measurement bandwidth. Using a lower bandwidth decreases measurement noise and increases trigger sensitivity. The list of available bandwidth IDs can be obtained using rsnrpz_bandwidth_getbwlist(): lerr = rsnrpz_bandwidth_setbw( ulusbsession, 1, 0 ) The number of video points for the trace measurement is set using rsnrpz_scope_setpoints(). Using 500 points usually represents a good compromise between USB transfer speed and resolution. The trace data's transfer time increases with the number of video points: lerr = rsnrpz_scope_setpoints( ulusbsession, 1, ivideopoints ) 152

153 VXI PnP Programming Guide The trace time sets the overall capture time for one trace measurement. Each video point represents the time period resulting from the trace time divided by the number of video points: lerr = rsnrpz_scope_settime( ulusbsession, 1, dtracetime ) The offset time is used to capture signal portions before the trigger point. The valid time range depends on the sensor and must be looked up in the sensor manual. The function call is not required if this feature is not needed. An offset time of zero starts trace capturing at the trigger position: lerr = rsnrpz_scope_setoffsettime( ulusbsession, 1, doffsettime ) Configuring the trigger condition is crucial for all trace measurements. The following lines configure the trigger system to internal triggering on a positive slope. The hysteresis should be set to a small value (e.g. 1 db) to allow for stable triggering. The dropout time can be set optionally and requires the signal to fall below the trigger threshold for the defined period of time before the trigger system rearms again. Please note that the trigger level is set in linear units. lerr = rsnrpz_trigger_setsource( ulusbsession, 1, RSNRPZ_TRIGGER_SOURCE_INTERNAL ) lerr = rsnrpz_trigger_setslope( ulusbsession, 1, RSNRPZ_SLOPE_POSITIVE ) lerr = rsnrpz_trigger_setdropouttime( ulusbsession, 1, ddropouttime ) lerr = rsnrpz_trigger_sethysteresis( ulusbsession, 1, 1 ) lerr = rsnrpz_trigger_setlevel( ulusbsession, 1, dtriglevelw ) Setting an averaging filter is, in most cases, desired when trace data is to be measured. Averaging reduces the noise dramatically and therefore increases the dynamic range: lerr = rsnrpz_scope_setaveragecount( ulusbsession, 1, iaveragecount ) lerr = rsnrpz_scope_setaverageenabled( ulusbsession, 1, true ) The averaging filter can be operated in either repeating or moving mode. In repeating mode, the filter content is cleared at the beginning of the measurement cycle. Once the filter is entirely filled, the measurement terminates and the result can be read: lerr = rsnrpz_scope_setaverageterminalcontrol( ulusbsession, 1, RSNRPZ_TERMINAL_CONTROL_REPEAT ) The rsnrpz_chan_initiate() function call starts the measurement cycle and returns immediately. lerr = rsnrpz_chan_initiate( ulusbsession, 1) Before any data can be read from the sensor, the measurement status must be polled repeatedly. This polling must be implemented in such a way that the CPU becomes available to the driver thread periodically: 153

154 VXI PnP Programming Guide Sleep( 100 ) or SwitchToThread() ViBoolean bmeascompleted lerr = rsnrpz_chan_ismeasurementcomplete( ulusbsession, 1, &bmeascompleted ) After the measurement has completed, the data array can be read using the rsnrpz_meass_fetchbuffermeasurement() function. The values that are returned are in linear units and include any offset configured using rsnrpz_corr_setoffset(). The number of trace points must match the number of video points set with rsnrpz_scope_setpoints(): ViReal64 pdmeasav[itracepoints] ViInt32 ireadcount lerr = rsnrpz_meass_fetchbuffermeasurement( ulusbsession, 1, itracepoints, pdmeasav, &ireadcount ) The settings from the above example return the average trace representation. Based on the averaging filter settings and the trace time, the sensor captures multiple samples and calculates the average trace data. The return data is provided in linear units. However, many applications require power values on a logarithmic scale. The conversion can be done using the following equation: Plog = 10 log10( Plin ) dbm + 30 dbm Care must be taken if signal portions close to the noise floor must be converted. Depending on the zero reference point for the internal analog-to-digital converter, negative power readings may occur. This is normal behavior, and in most cases, it is possible to simply use the linear power reading's absolute value for the log10() function. The low power values do not typically contribute to any measurement. In very rare cases, a power value that is exactly zero may arise. Zero cannot be converted into the logarithmic scale and must, therefore, be replaced by another value. The Power Viewer Plus software uses the closest valid point in such cases Single-Shot Events Measuring single-shot events requires slightly different averaging filter settings. Please note that disabling averaging also reduces the sensor's dynamic range. The average filter count is set to one, and the filter is disabled: lerr = rsnrpz_scope_setaveragecount( ulusbsession, 1, 1 ) lerr = rsnrpz_scope_setaverageenabled( ulusbsession, 1, false ) Additionally, non-z81 sensors require enabling of the real-time mode. In this mode, the chopper is turned off and only one single trace is processed. (The Z81 sensor does not require this command.): rsnrpz_scope_setrealtimeenabled( ulusbsession, 1, true ) 154

155 VXI PnP Programming Guide Peak Trace Data Wideband sensors, such as the NRP-Z81, provide multiple trace data representations. The AVERAGE trace representation is always sent and cannot be deselected. Alternatively, the sensor can be switched to auxiliary mode, and it can send two additional representations, such as the RANDOM and MAXIMUM trace data: lerr = rsnrpz_chan_setauxiliary( ulusbsession, 1, RSNRPZ_AUX_RNDMAX ) When auxiliary data is enabled, the trace data must be read from the driver cache using the rsnrpz_meass_fetchbuffermeasurementaux() function. In a way similar to the regular fetch function, all data is provided in linear units and contains the level offset: ViReal64 pdmeasav[itracepoints] ViReal64 pdmeasrnd[itracepoints] ViReal64 pdmeaspeak[itracepoints] ViInt32 ireadcount lerr = rsnrpz_meass_fetchbuffermeasurementaux( ulusbsession, 1, 0, itracepoints, pdmeasav, pdmeasrnd, pdmeaspeak, &ireadcount ) 155

156 VXI PnP Programming Guide Automatic Pulse Measurement Wideband sensors, such as the NRP-Z81, can perform automatic pulse measurements in trace mode. Enabling the automated pulse measurement increases the measurement and processing time inside the sensor. The following two functions enable the automatic pulse measurement and set the algorithm to the histogram type. lerr = rsnrpz_scope_meas_setmeasenabled( ulusbsession, 1, true ) lerr = rsnrpz_scope_meas_setmeasalgorithm( ulusbsession, 1, RSNRPZ_SCOPE_MEAS_ALG_HIST ) The only three configuration parameters required by the automatic pulse analysis are the low, mid, and high thresholds as a percentage of the pulse top power. lerr = rsnrpz_scope_meas_setlevelthresholds( ulusbsession, 1, dlevmidpercent, dlevlowpercent, dlevhighpercent ) After completion of the trace measurement, the pulse measurement results can be read from the driver cache using the functions listed below. Values that cannot be determined are indicated using a quiet NaN. #ifdef LINUX #define isnan(_x) (fpclassify((float)_x)==fp_nan) #else #define isnan(_x) ((_X)!=(_X)) #endif The general pulse timing can be read using the rsnrpz_scope_meas_getpulsetimes() function. The duty cycle and period time require at least two pulses to fall into the trace window. The measurement of the pulse width requires at least a rising and falling edge. rsnrpz_scope_meas_getpulsetimes( ulusbsession, 1, &ddutycycle, &dpulsewidth, &dperiodtime ) 156

157 VXI PnP Programming Guide The rising and falling edge times are read using the same function twice: rsnrpz_scope_meas_getpulsetransition( ulusbsession, 1, RSNRPZ_SLOPE_POSITIVE, &drisetime, &driseposition, &driseovershot ) rsnrpz_scope_meas_getpulsetransition( ulusbsession, 1, RSNRPZ_SLOPE_NEGATIVE, &dfalltime, &dfallposition, &dfallovershot ) There is a series of functions available for the measuring the different pulse-power levels. The pulse peak power and the pulse top power are typically of greater interest: rsnrpz_scope_meas_getpulsepower( ulusbsession, 1, &daveragepower, &dminpeak, &dmaxpeak ) rsnrpz_scope_meas_getpulselevels( ulusbsession, 1, &dtoppower, &dbasepower ) rsnrpz_scope_meas_getpulsereferencelevels( ulusbsession, 1, &dlowreflevel, &dhighreflevel, &dmidreflevel ) Please note that all power readings are in linear units, and they contain any level offset that was previously set. 157

158 27 Customer Support Customer Support Technical support where and when you need it For quick, expert help with any Rohde & Schwarz equipment, contact one of our Customer Support Centers. A team of highly qualified engineers provides telephone support and will work with you to find a solution to your query on any aspect of the operation, programming or application of Rohde & Schwarz equipment. Up-to-date information and upgrades To keep your instrument up-to-date and to be informed about new application notes related to your instrument, please send an to the Customer Support Center stating your instrument and your wish. We will take care to ensure that you get the right information. Europe, Africa, Middle East Phone customersupport@rohde-schwarz.com North America Phone TEST-RSA ( ) customer.support@rsa.rohde-schwarz.com Latin America Phone customersupport.la@rohde-schwarz.com Asia/Pacific Phone customersupport.asia@rohde-schwarz.com China Phone / customersupport.china@rohde-schwarz.com 158

159 Appendix 28 Appendix 28.1 Command Line Options The Power Viewer Plus software supports a set of command line options that affect the application's look and feel as well its startup behavior: --native The user interface look is left as native as possible. --classic-pv This option starts Power Viewer Plus in a mode in which it only displays the continuous power measurement window. This is similar to the classic Power Viewer application: Disables all features but the continuous power measurement. Always starts with a fixed application window size. Continuous power measurement is activated. The analog bar and trend display are not available. The measurement starts automatically if a sensor is detected. --no-splash This option omits the initial splash screen and speeds up the application startup. --project <file> This option loads a specific project file at startup. If the application is available, the default project file is written. If the specified project file is not available, the default settings are applied. --sensor <sensor> This option includes no-splash and omits the initial sensor scanning. Instead, the specified sensor is made available regardless of its physical availability. The sensor must be defined by the sensor type and by its serial number (for example: Z11, ). --no-multi Disables the multi-channel measurement mode. 159

160 Appendix --no-flash Disables the firmware flash dialog. --no-timeslot Disables the timeslot measurement mode. --no-statistics Disables the statistics measurement mode. --no-trace Disables the trace measurement mode. --no-scripting Disables the scripting measurement mode. --no-ota Disables all OTA measurements (NRPM3 sensor module). --no-datalog Disables the data log window. --no-analysis Disables the data analysis window. --no-monitor Disables the limit monitoring window. --no-gauges Disables the gauges window. --no-barchart Disables the bar chart window. --no-sparam Disables the s-parameter configuration dialog. --debug Writes additional log messages to the message log window. This may be useful for debugging software problems. --visa=<visa vendor> Selects a specific VISA driver. Permissible arguments are RS, NI and Agilent. At startup the Power Viewer Plus application looks for a file named.pvoptions.txt in the user's home directory. If the file is present then all lines not starting with the # character will be added to the command line arguments. Thus, the file can be utilized to add command line options without creating a special desktop link. Example: #.pvoptions.txt # Permanent command line options for Power Viewer Plus # --debug --visa=rs --no-splash --no-ota 160

161 28.2 Appendix Setting the Application Style The style of the Power Viewer Plus user interface can be changed using the -style command-line option. Changing the style might be useful if the application should use the operating system's look and feel. By default, Power Viewer Plus uses an internal style that is independent of the underlying operating system. -style <style> <style> User Interface Example Plastique Cleanlooks Windows Motif WindowsXP 161

Using R&S NRP Series Power Sensors with Android TM Handheld Devices. Application Note. Products: R&S NRP Series. R&S NRP-Zxx Series

Using R&S NRP Series Power Sensors with Android TM Handheld Devices. Application Note. Products: R&S NRP Series. R&S NRP-Zxx Series Using R&S NRP Series Power Sensors with Android TM Handheld Devices Products: R&S NRP Series R&S NRP-Zxx Series This application note describes how to connect and use the highly popular R&S NRP family

More information

R&S NRP-Z Power Sensor Programming Guide Application Note

R&S NRP-Z Power Sensor Programming Guide Application Note R&S NRP-Z Power Sensor Programming Guide Application Note Products: R&S NRP-Z Power Sensors The R&S NRP-Z power sensors from Rohde & Schwarz represent the latest in power measurement technology. They offer

More information

Advanced Test Equipment Rentals ATEC (2832)

Advanced Test Equipment Rentals ATEC (2832) E stablished 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) Technical Datasheet Scalar Network Analyzer Model 8003-10 MHz to 40 GHz The Giga-tronics Model 8003 Precision Scalar

More information

Boonton 4540 Remote Operation Modes

Boonton 4540 Remote Operation Modes Application Note Boonton 4540 Remote Operation Modes Mazumder Alam Product Marketing Manager, Boonton Electronics Abstract Boonton 4540 series power meters are among the leading edge instruments for most

More information

LadyBug Technologies, LLC LB5908A True-RMS Power Sensor

LadyBug Technologies, LLC LB5908A True-RMS Power Sensor LadyBug Technologies, LLC LB5908A True-RMS Power Sensor LB5908ARev8 LadyBug Technologies www.ladybug-tech.com Telephone: 707-546-1050 Page 1 LB5908A Data Sheet Key PowerSensor+ TM Specifications Frequency

More information

Fast. Accurate. USB-capable. Power sensors from Rohde & Schwarz

Fast. Accurate. USB-capable. Power sensors from Rohde & Schwarz Fast. Accurate. USB-capable. Power sensors from Rohde & Schwarz Power_sensors_bro_en_3606-7147-32.indd 1 22.05.2014 14:08:59 Fast. Accurate. USB-capable. Power sensors from Rohde & Schwarz The most important

More information

Fast. Accurate. USB-capable. Power sensors from Rohde & Schwarz

Fast. Accurate. USB-capable. Power sensors from Rohde & Schwarz Fast. Accurate. USB-capable. Power from Rohde & Schwarz Power bro_en_3606-7147-32.indd 1 01.08.2013 16:57:35 Fast. Accurate. USB-capable. Power from Rohde & Schwarz The most important features for accurate

More information

8500A. Advanced Test Equipment Rentals ATEC (2832) channel capability. For tests on pulse mod- SERIES PEAK POWER METERS

8500A. Advanced Test Equipment Rentals ATEC (2832) channel capability. For tests on pulse mod- SERIES PEAK POWER METERS Established 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) Page 1 of 5 8500A The Giga-tronics 8500A Series Peak Power Meters THE ACCURACY STANDARD channel capability. For tests

More information

Model 7330 Signal Source Analyzer Dedicated Phase Noise Test System V1.02

Model 7330 Signal Source Analyzer Dedicated Phase Noise Test System V1.02 Model 7330 Signal Source Analyzer Dedicated Phase Noise Test System V1.02 A fully integrated high-performance cross-correlation signal source analyzer from 5 MHz to 33+ GHz Key Features Complete broadband

More information

E4416A/E4417A EPM-P Series Power Meters and E-Series E9320 Peak and Average Power Sensors DATA SHEET

E4416A/E4417A EPM-P Series Power Meters and E-Series E9320 Peak and Average Power Sensors DATA SHEET E4416A/E4417A EPM-P Series Power Meters and E-Series E9320 Peak and Average Power Sensors DATA SHEET EPM-P Power Meter Specifications Specifications describe the instrument s warranted performance and

More information

USB Mini Spectrum Analyzer User s Guide TSA5G35

USB Mini Spectrum Analyzer User s Guide TSA5G35 USB Mini Spectrum Analyzer User s Guide TSA5G35 Triarchy Technologies, Corp. Page 1 of 21 USB Mini Spectrum Analyzer User s Guide Copyright Notice Copyright 2011 Triarchy Technologies, Corp. All rights

More information

Basic RF Amplifier Measurements using the R&S ZNB Vector Network Analyzer and SMARTerCal. Application Note

Basic RF Amplifier Measurements using the R&S ZNB Vector Network Analyzer and SMARTerCal. Application Note Basic RF Amplifier Measurements using a R&S ZNB Analyzer and SMARTerCal Mark Bailey 2013-03-05, 1ES, Version 1.0 Basic RF Amplifier Measurements using the R&S ZNB Vector Network Analyzer and SMARTerCal.

More information

USB Mini Spectrum Analyzer User Manual TSA Program for PC TSA4G1 TSA6G1 TSA8G1

USB Mini Spectrum Analyzer User Manual TSA Program for PC TSA4G1 TSA6G1 TSA8G1 USB Mini Spectrum Analyzer User Manual TSA Program for PC TSA4G1 TSA6G1 TSA8G1 Triarchy Technologies Corp. Page 1 of 17 USB Mini Spectrum Analyzer User Manual Copyright Notice Copyright 2013 Triarchy Technologies,

More information

Yellow Frog. Manual Version 1.1

Yellow Frog. Manual Version 1.1 Yellow Frog Manual Version 1.1 1 YellowFrog Contents PC Requirements...... 2 YellowFrog Power Meter Measurement.... 3 YellowFrog PC Software..... 3 Main Screen....... 4 Input Overload....... 5 Battery

More information

USB Mini Spectrum Analyzer User Manual PC program TSA For TSA5G35 TSA4G1 TSA6G1 TSA12G5

USB Mini Spectrum Analyzer User Manual PC program TSA For TSA5G35 TSA4G1 TSA6G1 TSA12G5 USB Mini Spectrum Analyzer User Manual PC program TSA For TSA5G35 TSA4G1 TSA6G1 TSA12G5 Triarchy Technologies, Corp. Page 1 of 17 USB Mini Spectrum Analyzer User Manual Copyright Notice Copyright 2013

More information

E4416A EPM-P Series Single Channel Power Meter

E4416A EPM-P Series Single Channel Power Meter United States Home >... > Power Meters & Power Sensors > EPM and EPM-P Series Power Meters > E4416A EPM-P Series Single Channel Power Meter Product Status: Currently Orderable Currently Supported Key Specifications

More information

FCPM-6000RC. Mini-Circuits P.O. Box , Brooklyn, NY (718)

FCPM-6000RC. Mini-Circuits  P.O. Box , Brooklyn, NY (718) USB / Ethernet Integrated Frequency Counter & Power Meter 50Ω -30 dbm to +20 dbm, 1 MHz to 6000 MHz The Big Deal Automatically synchronized power & frequency measurements USB and Ethernet control Includes

More information

Application Note #63 Field Analyzers in EMC Radiated Immunity Testing

Application Note #63 Field Analyzers in EMC Radiated Immunity Testing Application Note #63 Field Analyzers in EMC Radiated Immunity Testing By Jason Galluppi, Supervisor Systems Control Software In radiated immunity testing, it is common practice to utilize a radio frequency

More information

PRELIMINARY INFORMATION. Professional Signal Generation and Monitoring Options for RIFEforLIFE Research Equipment

PRELIMINARY INFORMATION. Professional Signal Generation and Monitoring Options for RIFEforLIFE Research Equipment Integrated Component Options Professional Signal Generation and Monitoring Options for RIFEforLIFE Research Equipment PRELIMINARY INFORMATION SquareGENpro is the latest and most versatile of the frequency

More information

8650A SERIES UNIVERSAL POWER METERS

8650A SERIES UNIVERSAL POWER METERS 8650A SERIES UNIVERSAL METERS 8650A SERIES UNIVERSAL METERS The Capabilities to Test Today s Sophisticated Communications Systems TIME BURST AVERAGE The Giga-tronics 8650A Series Universal Power Meters

More information

7000 Series Signal Source Analyzer & Dedicated Phase Noise Test System

7000 Series Signal Source Analyzer & Dedicated Phase Noise Test System 7000 Series Signal Source Analyzer & Dedicated Phase Noise Test System A fully integrated high-performance cross-correlation signal source analyzer with platforms from 5MHz to 7GHz, 26GHz, and 40GHz Key

More information

E X P E R I M E N T 1

E X P E R I M E N T 1 E X P E R I M E N T 1 Getting to Know Data Studio Produced by the Physics Staff at Collin College Copyright Collin College Physics Department. All Rights Reserved. University Physics, Exp 1: Getting to

More information

Transmitter Interface Program

Transmitter Interface Program Transmitter Interface Program Operational Manual Version 3.0.4 1 Overview The transmitter interface software allows you to adjust configuration settings of your Max solid state transmitters. The following

More information

R&S NRP Power Sensor Family Specifications

R&S NRP Power Sensor Family Specifications NRP-Family_dat-sw_en_3607-0852-22_Cover.indd 1 Data Sheet 01.01 Test & Measurement R&S NRP Power Sensor Family Specifications 08.12.2014 09:29:11 CONTENTS Definitions... 3 Overview of the R&S NRPxxS(N)

More information

Analyze Frequency Response (Bode Plots) with R&S Oscilloscopes Application Note

Analyze Frequency Response (Bode Plots) with R&S Oscilloscopes Application Note Analyze Frequency Response (Bode Plots) with R&S Oscilloscopes Application Note Products: R&S RTO2002 R&S RTO2004 R&S RTO2012 R&S RTO2014 R&S RTO2022 R&S RTO2024 R&S RTO2044 R&S RTO2064 This application

More information

PulseCounter Neutron & Gamma Spectrometry Software Manual

PulseCounter Neutron & Gamma Spectrometry Software Manual PulseCounter Neutron & Gamma Spectrometry Software Manual MAXIMUS ENERGY CORPORATION Written by Dr. Max I. Fomitchev-Zamilov Web: maximus.energy TABLE OF CONTENTS 0. GENERAL INFORMATION 1. DEFAULT SCREEN

More information

Please feel free to download the Demo application software from analogarts.com to help you follow this seminar.

Please feel free to download the Demo application software from analogarts.com to help you follow this seminar. Hello, welcome to Analog Arts spectrum analyzer tutorial. Please feel free to download the Demo application software from analogarts.com to help you follow this seminar. For this presentation, we use a

More information

Spectrum Analyzer 1.6 GHz 3 GHz R&S HMS-X

Spectrum Analyzer 1.6 GHz 3 GHz R&S HMS-X HMS-X_bro_de-en_3607-0181-3X_v0200.indd 1 Product Brochure 02.00 Test & Measurement Spectrum Analyzer 1.6 GHz 3 GHz R&S HMS-X 15.03.2016 15:24:06 1 Basic Unit + 3 Options Key facts Frequency range: 100

More information

R&S RT-Zxx High-Voltage and Current Probes Specifications

R&S RT-Zxx High-Voltage and Current Probes Specifications R&S RT-Zxx High-Voltage and Current Probes Specifications Test & Measurement Data Sheet 14.00 CONTENTS Definitions... 3 Probe/oscilloscope chart... 4 R&S RT-ZH10/-ZH11 high-voltage probes... 5 R&S RT-ZD01

More information

R&S BCDRIVE R&S ETC-K930 Broadcast Drive Test Manual

R&S BCDRIVE R&S ETC-K930 Broadcast Drive Test Manual R&S BCDRIVE R&S ETC-K930 Broadcast Drive Test Manual 2115.1347.02 05 Broadcast and Media Manual The Manual describes the following R&S Broadcast Drive Test software. 2115.1360.02 2115.1360.03 2116.5146.02

More information

Operating Instructions

Operating Instructions Operating Instructions HAEFELY TEST AG KIT Measurement Software Version 1.0 KIT / En Date Version Responsable Changes / Reasons February 2015 1.0 Initial version WARNING Introduction i Before operating

More information

R&S RSC Step Attenuator Where precise signal levels count

R&S RSC Step Attenuator Where precise signal levels count Test & Measurement Product Brochure 01.00 Step Attenuator Where precise signal levels count Step Attenuator At a glance The is a switchable, mechanical step attenuator. It is available in various models

More information

Calibrating the CMD Output Level for BER

Calibrating the CMD Output Level for BER Ref Lvl -32 dbm -40-50 Marker 1 [T1] -35.50 dbm 5.000000 ms RBW 300 khz RF Att 0 db VBW 3 MHz SWT 5 ms Unit dbm 1 SUMMARY [T1] A RMS AVG -85.86 dbm SGL TRG -60-70 1RM -80 EXT -90-100 -110-120 -130 T1 T2

More information

Technical Datasheet. 8650B Series Universal Power Meters 10 MHz to 50 GHz. 8651B Single Channel 8652B Dual Channel Rev.

Technical Datasheet. 8650B Series Universal Power Meters 10 MHz to 50 GHz. 8651B Single Channel 8652B Dual Channel Rev. Technical Datasheet 8650B Series Universal Power Meters 10 MHz to 50 GHz 8651B Single Channel 8652B Dual Channel 34946 Rev. B / US040813 8650B Series Universal Power Meters The Giga-tronics 8650B Series

More information

R&S FSW-K54 EMI Measurement Application Detecting and eliminating electromagnetic

R&S FSW-K54 EMI Measurement Application Detecting and eliminating electromagnetic R&S FSW-K54 EMI Measurement Application Detecting and eliminating electromagnetic interference Test & Measurement Product Brochure 01.00 R&S FSW-K54 EMI Measurement Application At a glance The R&S FSW-K54

More information

Agilent E4416A/E4417A EPM-P Series Power Meters and E-Series E9320 Peak and Average Power Sensors

Agilent E4416A/E4417A EPM-P Series Power Meters and E-Series E9320 Peak and Average Power Sensors View at www.testequipmentdepot.com Agilent E4416A/E4417A EPM-P Series Power Meters and E-Series E9320 Peak and Average Power Sensors Data Sheet Test Equipment Depot - 800.517.8431-99 Washington Street

More information

Portable Performance for Debug and Validation

Portable Performance for Debug and Validation WaveJet 300A Oscilloscopes 100 MHz 500 MHz Portable Performance for Debug and Validation A UNIQUE TOOLSET FOR PORTABLE OSCILLOSCOPES Key Features 100 MHz, 200 MHz, 350 MHz and 500 MHz bandwidths Sample

More information

D-Lab & D-Lab Control Plan. Measure. Analyse. User Manual

D-Lab & D-Lab Control Plan. Measure. Analyse. User Manual D-Lab & D-Lab Control Plan. Measure. Analyse User Manual Valid for D-Lab Versions 2.0 and 2.1 September 2011 Contents Contents 1 Initial Steps... 6 1.1 Scope of Supply... 6 1.1.1 Optional Upgrades... 6

More information

Expect to Make Waves.

Expect to Make Waves. Expect to Make Waves. The New Oscilloscope Large 10.4" LCD touch screen Long capture time Extensive communication capabilities www.lecroy.com The New Oscillos From its large 10.4" LCD touch screen to its

More information

AMIQ-K2 Program for Transferring Various-Format I/Q Data to AMIQ. Products: AMIQ, SMIQ

AMIQ-K2 Program for Transferring Various-Format I/Q Data to AMIQ. Products: AMIQ, SMIQ Products: AMIQ, SMIQ AMIQ-K2 Program for Transferring Various-Format I/Q Data to AMIQ The software AMIQ-K2 enables you to read, convert, and transfer various-format I/Q data files to AMIQ format. AMIQ-K2

More information

R&S Cable Rider ZPH Cable and Antenna Analyzer Specifications

R&S Cable Rider ZPH Cable and Antenna Analyzer Specifications R&S Cable Rider ZPH Cable and Antenna Analyzer Specifications year Data Sheet Version 01.01 CONTENTS Definitions... 3 Specifications of the R&S Cable Rider ZPH Cable and Antenna Analyzer... 4 Frequency...

More information

R&S ZVA110 Vector Network Analyzer Specifications

R&S ZVA110 Vector Network Analyzer Specifications ZVA110_dat-sw_en_5214-4813-22_cover.indd 1 Data Sheet 04.00 Test & Measurement R&S ZVA110 Vector Network Analyzer Specifications 15.11.2013 14:42:28 CONTENTS Definitions... 3 Specifications... 4 Overview...

More information

USB Smart Power Sensor

USB Smart Power Sensor 50Ω -30 dbm to +20 dbm, 1 MHz to 8000 MHz The Big Deal Fast measurement speed, 10 msec USB HID device compatible with 32/64 Bit operating systems Includes Measurement Application GUI (Graphical User Interface)

More information

Application Note DT-AN DTU-315 Verification of Specifications

Application Note DT-AN DTU-315 Verification of Specifications DTU-315 Verification of Specifications APPLICATION NOTE January 2018 Table of Contents 1. Introduction... 3 General Description of the DTU-315... 3 Purpose of this Application Note... 3 2. Measurements...

More information

Extension kit for R&S Vector Network Analysers

Extension kit for R&S Vector Network Analysers NM300 Data Sheet September 2012 Extension kit for R&S Vector Network Analysers Characterisation of Nonlinear RF/HF Components in Time and Frequency domain Extension kit for R&S Vector Network Analysers

More information

Product Brochure Version R&S RSC Step Attenuator Where precise signal levels count

Product Brochure Version R&S RSC Step Attenuator Where precise signal levels count Product Brochure Version 02.00 Step Attenuator Where precise signal levels count RSC_bro_en_5214-4413-12_v0200.indd 1 07.09.2018 10:36:40 Step Attenuator At a glance The is a switchable, mechanical step

More information

Datasheet SHF A

Datasheet SHF A SHF Communication Technologies AG Wilhelm-von-Siemens-Str. 23D 12277 Berlin Germany Phone +49 30 772051-0 Fax ++49 30 7531078 E-Mail: sales@shf.de Web: http://www.shf.de Datasheet SHF 19120 A 2.85 GSa/s

More information

Agilent 5345A Universal Counter, 500 MHz

Agilent 5345A Universal Counter, 500 MHz Agilent 5345A Universal Counter, 500 MHz Data Sheet Product Specifications Input Specifications (pulse and CW mode) 5356C Frequency Range 1.5-40 GHz Sensitivity (0-50 deg. C): 0.4-1.5 GHz -- 1.5-12.4 GHz

More information

NanoGiant Oscilloscope/Function-Generator Program. Getting Started

NanoGiant Oscilloscope/Function-Generator Program. Getting Started Getting Started Page 1 of 17 NanoGiant Oscilloscope/Function-Generator Program Getting Started This NanoGiant Oscilloscope program gives you a small impression of the capabilities of the NanoGiant multi-purpose

More information

Concise NFC Demo Guide using R&S Test Equipment Application Note

Concise NFC Demo Guide using R&S Test Equipment Application Note Concise NFC Demo Guide using R&S Test Equipment Application Note Products: R&S SMBV100A R&S SMBV-K89 R&S FS-K112PC R&S RTO R&S RTO-K11 R&S CSNFC-B8 R&S FSL R&S FSV R&S FSW R&S ZVL This concise NFC Demo

More information

Signal Stability Analyser

Signal Stability Analyser Signal Stability Analyser o Real Time Phase or Frequency Display o Real Time Data, Allan Variance and Phase Noise Plots o 1MHz to 65MHz medium resolution (12.5ps) o 5MHz and 10MHz high resolution (50fs)

More information

Manual of Operation for WaveNode Model WN-2m. Revision 1.0

Manual of Operation for WaveNode Model WN-2m. Revision 1.0 Manual of Operation for WaveNode Model WN-2m. Revision 1.0 TABLE OF CONTENTS 1. Description of Operation 2. Features 3. Installation and Checkout 4. Graphical Menus 5. Information for Software Expansion

More information

R&S ZVA-Zxx Millimeter-Wave Converters Specifications

R&S ZVA-Zxx Millimeter-Wave Converters Specifications ZVA-Zxx_dat-sw_en_5214.2033.22_umschlag.indd 1 Data Sheet 13.00 Test & Measurement R&S ZVA-Zxx Millimeter-Wave Converters Specifications 28.01.2013 15:08:06 CONTENTS General information... 3 Definitions...

More information

EDL8 Race Dash Manual Engine Management Systems

EDL8 Race Dash Manual Engine Management Systems Engine Management Systems EDL8 Race Dash Manual Engine Management Systems Page 1 EDL8 Race Dash Page 2 EMS Computers Pty Ltd Unit 9 / 171 Power St Glendenning NSW, 2761 Australia Phone.: +612 9675 1414

More information

USB-TG124A Tracking Generator User Manual

USB-TG124A Tracking Generator User Manual USB-TG124A Tracking Generator User Manual Signal Hound USB-TG124A User Manual 2017, Signal Hound, Inc. 35707 NE 86th Ave La Center, WA 98629 USA Phone 360.263.5006 Fax 360.263.5007 This information is

More information

R&S ZN-Z103 Calibration Unit Specifications. Data Sheet V02.01

R&S ZN-Z103 Calibration Unit Specifications. Data Sheet V02.01 R&S ZN-Z103 Calibration Unit Specifications Data Sheet V02.01 CONTENTS Definitions... 3 Measurement range... 5 Effective system data... 5 General data... 6 Ordering information... 7 2 Rohde & Schwarz R&S

More information

Agilent E4416A/E4417A EPM-P Series Power Meters and E-Series E9320 Peak and Average Power Sensors. Data Sheet

Agilent E4416A/E4417A EPM-P Series Power Meters and E-Series E9320 Peak and Average Power Sensors. Data Sheet Agilent E4416A/E4417A EPM-P Series Power Meters and E-Series E9320 Peak and Average Power Sensors Data Sheet EPM-P power meter specifications Specifications describe the instrument s warranted performance

More information

Calibrating Measuring Microphones and Sound Sources for Acoustic Measurements with Audio Analyzer R&S UPV

Calibrating Measuring Microphones and Sound Sources for Acoustic Measurements with Audio Analyzer R&S UPV Product: R&S UPV Calibrating Measuring Microphones and Sound Sources for Acoustic Measurements with Audio Analyzer R&S UPV Application Note 1GA47_0E This application note explains how to use acoustic calibrations

More information

Dynamic re-referencing Microvolt-level measurements with the R&S RTO oscilloscopes

Dynamic re-referencing Microvolt-level measurements with the R&S RTO oscilloscopes RTO_app-bro_3607-2855-92_v0100.indd 1 Microvolt-level measurements with the R&S RTO Test & Measurement Application Brochure 01.00 Dynamic re-referencing Microvolt-level measurements with the R&S RTO oscilloscopes

More information

R&S ZVA-Zxx Millimeter-Wave Converters Specifications

R&S ZVA-Zxx Millimeter-Wave Converters Specifications R&S ZVA-Zxx Millimeter-Wave Converters Specifications Data Sheet Version 19.00 CONTENTS Definitions... 3 General information... 4 Specifications... 5 Test port... 5 Source input (RF IN)... 5 Local oscillator

More information

Part No. ENC-LAB01 Users Manual Introduction EncoderLAB

Part No. ENC-LAB01 Users Manual Introduction EncoderLAB PCA Incremental Encoder Laboratory For Testing and Simulating Incremental Encoder signals Part No. ENC-LAB01 Users Manual The Encoder Laboratory combines into the one housing and updates two separate encoder

More information

NOTICE: This document is for use only at UNSW. No copies can be made of this document without the permission of the authors.

NOTICE: This document is for use only at UNSW. No copies can be made of this document without the permission of the authors. Brüel & Kjær Pulse Primer University of New South Wales School of Mechanical and Manufacturing Engineering September 2005 Prepared by Michael Skeen and Geoff Lucas NOTICE: This document is for use only

More information

User Guide & Reference Manual

User Guide & Reference Manual TSA3300 TELEPHONE SIGNAL ANALYZER User Guide & Reference Manual Release 2.1 June 2000 Copyright 2000 by Advent Instruments Inc. TSA3300 TELEPHONE SIGNAL ANALYZER ii Overview SECTION 1 INSTALLATION & SETUP

More information

Digital Delay / Pulse Generator DG535 Digital delay and pulse generator (4-channel)

Digital Delay / Pulse Generator DG535 Digital delay and pulse generator (4-channel) Digital Delay / Pulse Generator Digital delay and pulse generator (4-channel) Digital Delay/Pulse Generator Four independent delay channels Two fully defined pulse channels 5 ps delay resolution 50 ps

More information

TransitHound Cellphone Detector User Manual Version 1.3

TransitHound Cellphone Detector User Manual Version 1.3 TransitHound Cellphone Detector User Manual Version 1.3 RF3 RF2 Table of Contents Introduction...3 PC Requirements...3 Unit Description...3 Electrical Interfaces...4 Interface Cable...5 USB to Serial Interface

More information

R&S FSV-K40 Phase Noise Measurement Application Specifications

R&S FSV-K40 Phase Noise Measurement Application Specifications FSV-K40_dat-sw_en_5213-9705-22_cover.indd 1 Data Sheet 02.00 Test & Measurement R&S FSV-K40 Phase Noise Measurement Application Specifications 06.10.2014 14:51:49 CONTENTS Specifications... 3 Ordering

More information

Advanced Techniques for Spurious Measurements with R&S FSW-K50 White Paper

Advanced Techniques for Spurious Measurements with R&S FSW-K50 White Paper Advanced Techniques for Spurious Measurements with R&S FSW-K50 White Paper Products: ı ı R&S FSW R&S FSW-K50 Spurious emission search with spectrum analyzers is one of the most demanding measurements in

More information

Spectrum Analyser Basics

Spectrum Analyser Basics Hands-On Learning Spectrum Analyser Basics Peter D. Hiscocks Syscomp Electronic Design Limited Email: phiscock@ee.ryerson.ca June 28, 2014 Introduction Figure 1: GUI Startup Screen In a previous exercise,

More information

Application Note DT-AN-2115B-1. DTA-2115B Verification of Specifations

Application Note DT-AN-2115B-1. DTA-2115B Verification of Specifations DTA-2115B Verification of Specifations APPLICATION NOTE January 2018 Table of Contents 1. Introduction... 3 General Description of the DTA-2115B... 3 Purpose of this Application Note... 3 2. Measurements...

More information

R&S Cable Rider ZPH Cable and Antenna Analyzer Specifications

R&S Cable Rider ZPH Cable and Antenna Analyzer Specifications R&S Cable Rider ZPH Cable and Antenna Analyzer Specifications year Data Sheet Version 02.02 CONTENTS Definitions... 3 Specifications... 4 Frequency... 4 Measurements... 4 Channel power meter (R&S ZPH-K19

More information

Vector Network Analyzer TTR503A/TTR506A USB Vector Network Analyzer Preliminary Datasheet. Subject to change.

Vector Network Analyzer TTR503A/TTR506A USB Vector Network Analyzer Preliminary Datasheet. Subject to change. Vector Network Analyzer TTR503A/TTR506A USB Vector Network Analyzer Preliminary Datasheet. Subject to change. Applications Academic/Education Design, development and manufacturing of passive and active

More information

National Park Service Photo. Utah 400 Series 1. Digital Routing Switcher.

National Park Service Photo. Utah 400 Series 1. Digital Routing Switcher. National Park Service Photo Utah 400 Series 1 Digital Routing Switcher Utah Scientific has been involved in the design and manufacture of routing switchers for audio and video signals for over thirty years.

More information

R&S ZN-Z151/-Z152/-Z153 Calibration Unit Specifications

R&S ZN-Z151/-Z152/-Z153 Calibration Unit Specifications ZN-Z151_152_153_dat-sw_en_3607-0881-22_v0100_cover.indd 1 Data Sheet 01.00 Test & Measurement R&S ZN-Z151/-Z152/-Z153 Calibration Unit Specifications 07.10.2014 11:35:47 CONTENTS Definitions... 3 Measurement

More information

SNG-2150C User s Guide

SNG-2150C User s Guide SNG-2150C User s Guide Avcom of Virginia SNG-2150C User s Guide 7730 Whitepine Road Revision 001 Richmond, VA 23237 USA GENERAL SAFETY If one or more components of your earth station are connected to 120

More information

VIDEO GRABBER. DisplayPort. User Manual

VIDEO GRABBER. DisplayPort. User Manual VIDEO GRABBER DisplayPort User Manual Version Date Description Author 1.0 2016.03.02 New document MM 1.1 2016.11.02 Revised to match 1.5 device firmware version MM 1.2 2019.11.28 Drawings changes MM 2

More information

RF Explorer RackPRO. User Manual. Introduction. Greetings fellow traveler on the RF spectrum.

RF Explorer RackPRO. User Manual. Introduction. Greetings fellow traveler on the RF spectrum. RF Explorer RackPRO User Manual Introduction Greetings fellow traveler on the RF spectrum. The RF Explorer RackPRO (referred to in this document in shorthand as RackPRO ) has been designed to be intuitive

More information

EASY-MCS. Multichannel Scaler. Profiling Counting Rates up to 150 MHz with 15 ppm Time Resolution.

EASY-MCS. Multichannel Scaler. Profiling Counting Rates up to 150 MHz with 15 ppm Time Resolution. Multichannel Scaler Profiling Counting Rates up to 150 MHz with 15 ppm Time Resolution. The ideal solution for: Time-resolved single-photon counting Phosphorescence lifetime spectrometry Atmospheric and

More information

medlab One Channel ECG OEM Module EG 01000

medlab One Channel ECG OEM Module EG 01000 medlab One Channel ECG OEM Module EG 01000 Technical Manual Copyright Medlab 2012 Version 2.4 11.06.2012 1 Version 2.4 11.06.2012 Revision: 2.0 Completely revised the document 03.10.2007 2.1 Corrected

More information

Agilent Agilent 86120B, 86120C, 86122A Multi-Wavelength Meters Data Sheet

Agilent Agilent 86120B, 86120C, 86122A Multi-Wavelength Meters Data Sheet Agilent Agilent 86120B, 86120C, 86122A Multi-Wavelength Meters Data Sheet Agilent multi-wavelength meters are Michelson interferometer-based instruments that measure wavelength and optical power of laser

More information

News from Rohde&Schwarz Number 195 (2008/I)

News from Rohde&Schwarz Number 195 (2008/I) BROADCASTING TV analyzers 45120-2 48 R&S ETL TV Analyzer The all-purpose instrument for all major digital and analog TV standards Transmitter production, installation, and service require measuring equipment

More information

R&S RT-Zxx High-Bandwidth Probes Specifications

R&S RT-Zxx High-Bandwidth Probes Specifications R&S RT-Zxx High-Bandwidth Probes Specifications Test & Measurement Data Sheet 14.00 CONTENTS Definitions... 3 Probe/oscilloscope chart... 4 R&S RT-ZZ80 transmission line probe... 5 R&S RT-ZS10/-ZS10E/-ZS20/-ZS30

More information

R&S FS-Z60/75/90/110 Harmonic Mixers for the R&S FSP/FSU/ FSQ/FSUP/FSV

R&S FS-Z60/75/90/110 Harmonic Mixers for the R&S FSP/FSU/ FSQ/FSUP/FSV Test & Measurement Data Sheet 04.00 R&S FS-Z60/75/90/110 Harmonic Mixers for the R&S FSP/FSU/ FSQ/FSUP/FSV R&S FS-Z60/75/ 90/110 Harmonic Mixers At a glance The R&S FS-Z60/-Z75/-Z90/-Z110 harmonic mixers

More information

R&S NRPM Over-the-Air (OTA) Power Measurement Solution Specifications

R&S NRPM Over-the-Air (OTA) Power Measurement Solution Specifications R&S NRPM Over-the-Air (OTA) Power Measurement Solution Specifications year Data Sheet Version 04.00 CONTENTS Definitions... 3 R&S NRPM3 three-channel sensor module... 4 R&S NRPM-A66 single-polarized antenna

More information

BitWise (V2.1 and later) includes features for determining AP240 settings and measuring the Single Ion Area.

BitWise (V2.1 and later) includes features for determining AP240 settings and measuring the Single Ion Area. BitWise. Instructions for New Features in ToF-AMS DAQ V2.1 Prepared by Joel Kimmel University of Colorado at Boulder & Aerodyne Research Inc. Last Revised 15-Jun-07 BitWise (V2.1 and later) includes features

More information

Full-featured CW Microwave Counters for Field, Factory or Lab

Full-featured CW Microwave Counters for Field, Factory or Lab Full-featured CW Microwave Counters for Field, Factory or Lab Product Overview Agilent 53150A 20 GHz Counter Agilent 53151A 26.5 GHz Counter 46 GHz Counter High performance microwave counters: at home,

More information

RF Level Test System +20 dbm to 130 dbm

RF Level Test System +20 dbm to 130 dbm NRVD Power Meter optional Therm. Sensor A B Power: >-15 dbm DUT (Signal Generator, Communication Tester) 1 MHz - 3.5/6 GHz +20 dbm... -130 dbm Diode Sensor Z4 Power: -15 to -40 dbm 6 db Power =< -40 dbm

More information

R&S ZN-Z85 Switch Matrix Specifications

R&S ZN-Z85 Switch Matrix Specifications R&S ZN-Z85 Switch Matrix Specifications Data Sheet Version 01.02 CONTENTS Definitions... 3 Block diagrams... 4 Specifications... 5 General features... 5 Performance data... 5 Remote control... 5 Switching

More information

R&S ETH Handheld TV Analyzer Portable DVB-T/H signal analysis up to 3.6/8 GHz

R&S ETH Handheld TV Analyzer Portable DVB-T/H signal analysis up to 3.6/8 GHz R&S ETH Handheld TV Analyzer Portable DVB-T/H signal analysis up to 3.6/8 GHz Broadcast Product Brochure 02.00 R&S ETH Handheld TV Analyzer At a glance The R&S ETH handheld TV analyzer was specially designed

More information

The Measurement Tools and What They Do

The Measurement Tools and What They Do 2 The Measurement Tools The Measurement Tools and What They Do JITTERWIZARD The JitterWizard is a unique capability of the JitterPro package that performs the requisite scope setup chores while simplifying

More information

R&S VSE Vector Signal Explorer Base Software Specifications

R&S VSE Vector Signal Explorer Base Software Specifications Data Sheet Version 10.00 R&S VSE Vector Signal Explorer Base Software Specifications VSE_dat-sw_en_3607-1371-22_v1000_cover.indd 1 26.04.2018 10:05:34 CONTENTS Definitions... 3 Specifications... 4 Minimum

More information

R&S ZN-Z154 Calibration Unit Specifications

R&S ZN-Z154 Calibration Unit Specifications ZN-Z154_dat-sw_en_3607-0481-22_v0101_cover.indd 1 Data Sheet 01.01 Test & Measurement R&S ZN-Z154 Calibration Unit Specifications 25.06.2014 10:27:09 CONTENTS Definitions... 3 Measurement range... 4 Effective

More information

Getting started with Spike Recorder on PC/Mac/Linux

Getting started with Spike Recorder on PC/Mac/Linux Getting started with Spike Recorder on PC/Mac/Linux You can connect your SpikerBox to your computer using either the blue laptop cable, or the green smartphone cable. How do I connect SpikerBox to computer

More information

Getting Started with the LabVIEW Sound and Vibration Toolkit

Getting Started with the LabVIEW Sound and Vibration Toolkit 1 Getting Started with the LabVIEW Sound and Vibration Toolkit This tutorial is designed to introduce you to some of the sound and vibration analysis capabilities in the industry-leading software tool

More information

Microwave Counter, Power Meter and DVM in One Portable Package

Microwave Counter, Power Meter and DVM in One Portable Package Agilent 53140 Series Microwave Counter, Power Meter and DVM in One Portable Package Product Overview Everything you need for the installation and maintenance of microwave links: A choice of frequency counter

More information

R&S HA-Z24E External Preamplifier 1 GHz to 85 GHz Specifications

R&S HA-Z24E External Preamplifier 1 GHz to 85 GHz Specifications R&S HA-Z24E External Preamplifier 1 GHz to 85 GHz Specifications Data Sheet Version 01.01 Definitions General Product data applies under the following conditions: Three hours storage at ambient temperature

More information

Spectrum Analyzer 1.6 GHz 3 GHz HMS-X

Spectrum Analyzer 1.6 GHz 3 GHz HMS-X Spectrum Analyzer 1.6 GHz 3 GHz 1 Basic Unit + 3 Options Your Spectrum Analyzer Key facts Frequency range: 100 khz to 1.6 GHz/3 GHz* 1 Spectral purity greater than -100 dbc/hz (at 100 khz) SWEEP from 20

More information

R&S FSW-B512R Real-Time Spectrum Analyzer 512 MHz Specifications

R&S FSW-B512R Real-Time Spectrum Analyzer 512 MHz Specifications R&S FSW-B512R Real-Time Spectrum Analyzer 512 MHz Specifications Data Sheet Version 02.00 CONTENTS Definitions... 3 Specifications... 4 Level... 5 Result display... 6 Trigger... 7 Ordering information...

More information

PYROPTIX TM IMAGE PROCESSING SOFTWARE

PYROPTIX TM IMAGE PROCESSING SOFTWARE Innovative Technologies for Maximum Efficiency PYROPTIX TM IMAGE PROCESSING SOFTWARE V1.0 SOFTWARE GUIDE 2017 Enertechnix Inc. PyrOptix Image Processing Software v1.0 Section Index 1. Software Overview...

More information

User s Manual. Log Scale (/LG) GX10/GX20/GP10/GP20/GM10 IM 04L51B01-06EN. 3rd Edition

User s Manual. Log Scale (/LG) GX10/GX20/GP10/GP20/GM10 IM 04L51B01-06EN. 3rd Edition User s Manual Model GX10/GX20/GP10/GP20/GM10 Log Scale (/LG) 3rd Edition Introduction Thank you for purchasing the SMARTDAC+ Series GX10/GX20/GP10/GP20/GM10 (hereafter referred to as the recorder, GX,

More information

Dragon. manual version 1.6

Dragon. manual version 1.6 Dragon manual version 1.6 Contents DRAGON TOP PANEL... 2 DRAGON STARTUP... 2 DRAGON STARTUP SCREEN... 2 DRAGON INFO SCREEN... 3 DRAGON MAIN SCREEN... 3 TURNING ON A TRANSMITTER... 4 CHANGING MAIN SCREEN

More information