Flow Cytometry Histograms: Transformations, Resolution, and Display

Size: px
Start display at page:

Download "Flow Cytometry Histograms: Transformations, Resolution, and Display"

Transcription

1 Review Article Flow Cytometry Histograms: Transformations, Resolution, and Display David Novo, 1 * James Wood 2 1 De Novo Software, 3250 Wilshire Blvd. Suite 803, Los Angeles, California 2 Department of Cancer Biology, Wake Forest University School of Medicine, Winston Salem North Carolina Received 25 March 2008; Accepted 6 May 2008 *Correspondence to: David Novo, 3250 Wilshire Blvd. Suite 803, Los Angeles, CA 90010, USA david.novo@denovosoftware.com Published online 8 July 2008 in Wiley InterScience ( wiley.com) DOI: /cyto.a International Society for Advancement of Cytometry Abstract Flow cytometry data analysis routinely includes the use of one- or two-parameter histograms to visualize the data. These histograms have traditionally been plotted with either a linear or logarithmic scale. However, the recent trend of performing the logarithmic conversion in software has made apparent some limitations of the traditional visual presentation of logarithmic data. This review discusses the mathematics of presenting data on a histogram and emphasizes the difference between scaling and binning. The review introduces the concept of an effective resolution to describe how the bin width changes in a variable bin-width histogram. The change in effective resolution is used to explain the commonly observed valley and picket fencing artifacts. These result from the effective resolution of the display histogram being too high for the data being presented. Recently, several different binning transformations have been described that are becoming more popular because they allow one to view a large dynamic range of data on a single plot, while allowing the display of negative data values. While each of the transforms is based upon different equations, they all exhibit very similar properties. All of the transforms bin the data logarithmically at high channel values and linearly at low channel values. The linear scaling of the lower channels serves to limit the effective resolution of the histogram, thus minimizing the valley and picket fencing artifacts. The newer transformations are not without their own limitations and recommendations for the appropriate manner of presenting flow cytometry data using these newer transformations are discussed. ' 2008 International Society for Advancement of Cytometry Key terms histogram; data display; transformation; binning; scaling FLOW cytometry data consist of multiparameter optical (e.g. light scatter, fluorescence intensities), or electronic (e.g. Coulter volume) measurements from many thousands (or millions) of cells. These data are almost always graphed on one- or two-parameter histograms. The goal of visualizing data of a measured parameter of a sample population is to determine the actual probability function (APF) of the property being investigated. In various statistical literature, the APF is also referred to as the actual probability distribution, actual frequency distribution or actual probably density distribution. A histogram is a direct tabulation of the frequencies of the measured values obtained by measuring a specific parameter of a sample population (1). Flow cytometry data histograms have been universally interpreted as being an accurate visual representation of the APF of the property being measured. They are traditionally displayed on either linear or logarithmically scaled axes. The logarithmic transformation has been found to be useful because it allows for the visualization of at least four decades of dynamic range on a single graph, and it makes log-normal distributions, common to biological systems, appear more symmetrical. Some instruments perform the logarithmic transformation in hardware, using logarithmic amplifiers, while other instruments only save linearly scaled data, and the logarithmic transformation Cytometry Part A 73A: , 2008

2 occurs in software (2). More recently, several different authors have proposed alternatives to the logarithmic function for the display of flow cytometry data (3 7) to overcome some of the logarithmic transformation s perceived shortcomings. This review will discuss the following: (i) the difference between binning and scaling of flow cytometry data; (ii) the display of data in uniform and variable bin-width histograms; (iii) visual artifacts associated with variable bin-width histograms; (iv) how these new transformations attempt to display the flow cytometry data distributions so as to accurately reflect the underlying distribution of the cell populations in the sample; and (v) caveats of the new transformations. SCALING VS. BINNING Flow cytometry histograms are a direct tabulation of the frequencies of measured values in a fixed number of channels or bins. They are often described as being displayed in a particular scale without reference to the underlying process of binning. However, scaling and binning are separate procedures that impact how data are visualized in a histogram. Scaling refers to the scale associated with an axis and does not describe the width of the bins in the histogram. The difference between scaling and binning can be clearly appreciated in Figure 1. Figures 1B and 1C show data from the same APF as the data displayed in Figure 1A, on an apparent logarithmic display. The logarithmic display was either accomplished by simply changing the scale in Microsoft Excel (Fig. 1B) or by simulating the digitization of a logarithmically amplified signal as it would be if performed in the flow cytometer (Fig. 1C). The plot in Figure 1B preserves the peak heights and the overall shape of the curve, whereas the plot in Figure 1C did not. Figures 1D 1F show data from 100,000 cells simulated from a uniform APF, i.e., the measured values encompassed all possible values. Figure 1D shows the data on a linear scale, where the APF is clearly understood. The data is displayed on a logarithmic axis either by changing the scale in Microsoft Excel (Fig. 1E) or by simulating a logarithmically amplified signal (Fig. 1F). The population in Figure 1F appears skewed, with many more events having high data values than lower ones. Even these simplistic data sets show there is a clear difference between the different mechanisms of displaying data on a logarithmic scale. Scaling of an axis refers to changing the relative position of coordinates along the axis. Simply changing the scale preserves the number of peaks and the peak heights, but can make peaks appear wider, narrower, skewed, or can change their relative positions, depending on the scaling used. Visually comparing peaks in different regions of the nonlinear scale can be nonintuitive, since the peaks may look dramatically different e.g., larger or smaller apparent area solely due to the nonlinear axis scaling. For example, the logarithmic scale will make distributions in the lower portion of the scale look disproportionately larger in area than if an identical distribution was displayed in the upper portion of the scale. Binning is a mathematical process that also affects how an APF appears in a histogram. All data displayed on flow cytometry histograms are binned data. Binning is the process of assigning data to discrete adjoining categories, generally referred to as channel values or bins. The appearance of the histogram is affected by the width and number of bins into which the APF is distributed, and whether the bins are all the same or have variable widths. Binning is initially performed by the flow cytometer electronics itself, when an analog-to-digital converter (ADC) samples an analog signal and assigns it to a discrete channel. The ADC resolution determines the number of possible channels to which a given analog signal can be assigned. Most of the older flow cytometers in use today have 10 bit ADCs (1,024 channels), while newer cytometers have up to 24 bit ADCs (16,777,216 channels). It may not be commonly appreciated that the data shown on flow cytometry plots is generally further binned by the flow cytometry data analysis software at a lower resolution than it was digitized. The decrease in resolution serves to emphasize the underlying shape of the distribution without being distracted by noise that is often evident in the higher resolution data. The calculation of the optimal number of bins needed to represent an APF has been a common subject in statistics literature (8). For instance, if 5000 events are acquired on an 18 bit resolution cytometer (262,144 possible channels), it is likely that there will be no more than a few events in any particular channel (Fig. 2A). Reducing the resolution to 10 bits (Fig. 2B) allows for a more accurate representation of the underlying APF. The simplest binning algorithm is linear binning, shown in Eq. (1), where O(y) is the underlying APF, R is the range of the original data and b is the number of bins. BðxÞ ¼ Z rðxþ1þ rx OðyÞdy; where r ¼ R b : ð1þ Linear binning has the characteristic that each individual channel in the binned dataset always represents the same range of the original dataset. Linear binning to a lower resolution will always increase the number of counts in the binned channels as compared to the original and can result in a decrease in the number of peaks if the resolution is lowered sufficiently. Another way to bin the data is logarithmically, according to the formula shown in Eq. (2), where a is the base and c ¼ log a ðrþ b. Z a cðxþ1þ BðxÞ ¼ OðyÞdy: ð2þ a cx Logarithmic binning can result from (i) the analog signal passing through a logarithmic amplifier and an analog to digital converter, or (ii) by software transforming linearly binned data that was saved by the flow cytometer. In contrast to linear binning, no two logarithmically binned channels represent the same absolute range in the original function. For example, channel 1023 of an ideal 4 decade logarithmically amplified histogram samples from 10,000 times as much of the original voltage signal range as channel 1. In fact, the range of the original signal represented by each channel in the logarithmic 686 Flow Cytometry Histograms

3 Figure 1. Comparison of Logarithmic displays obtained by binning and scaling. Populations were created from 5,000 simulated voltage measurements ranging from to 10 V. Linear and logarithmically amplified histograms were constructed by assigning the simulated measurement to channel based on the formulas: Linear Channel 5 (x/10) 3 (1024); where x is a simulated measurement. Log Channel 5 [log (x) 1 3) 3 256; where x is a simulated measurement. (A C) 5,000 cells were simulated to have a mean and standard deviation of 1 and 0.8 V, respectively. (A) Data were graphed as if signal was acquired through a 10 bit ADC. Channel 0 contains 573 events. (B) Data from Figure 1A was graphed on a log scale using Microsoft Excel. (C) Data were graphed as if signal was acquired using an ideal logarithmic amplifier. Channel 0 contains 540 events. Median value is 991. (D F) 100,000 cells were simulated from a uniform distribution. (D) Data were graphed as if signal was acquired through a 10 bit ADC. (E) Data from Figure 1A were graphed on a log scale using Microsoft Excel. (F) Data were graphed as if signal was acquired using an ideal logarithmic amplifier. [Color figure can be viewed in the online issue, which is available at histogram increases exponentially with increasing channel number. Each channel width increases by the same relative fraction as the previous one. Thus in contrast to a linearly binned histogram in which the relative width of the histogram channels decreases and the absolute width stays constant, the relative width of the logarithmic histogram channels remains Cytometry Part A 73A: ,

4 Figure 2. Effect of lowering the resolution on the ability to visualize the APF. Lowering the resolution of the histogram is often critical for visualizing the structure of the APF. Five thousand cells were simulated and channel values assigned using a simulated 18 bit ADC converter as described in Figure 1. Data were plotted on histograms with 18 bit (A) or 10 bit (B) resolution. constant and the absolute width increases exponentially. Thus logarithmic histograms have the advantage that the width of two populations that have the same coefficient of variation will be nearly identical, irrespective of the position on the histogram. This allows for easy visual comparison of the number of cells in different peaks anywhere along the axis range, provided they have the same CVs. While the end result of logarithmic binning may be to fit multiple decades of dynamic range onto an axis, it is not the same as applying a logarithmic scale to the axis. While displaying data on a logarithmic axis can be useful for biological data, care must be taken to appreciate that for a typical flow cytometry plot the logarithmic axis is a byproduct of the nonuniform bin widths and can introduce distortions along the axis that can lead to misinterpretation of the APF. In contrast, because true logarithmic scaling only affects the horizontal axis, the shape of the original APF is usually more recognizable (Fig. 1). EFFECTIVE RESOLUTION As discussed previously, when data is binned logarithmically, no two binned channels represent data from the same range as the original data. Thus, it is convenient to define the effective resolution (ER) of a particular channel in the logarithmic binned histogram as the number of channels there would be if the original data was binned linearly with the same equivalent bin width as the particular log channel. The topmost channel in a 1,023 resolution log scaled histogram samples from 0.9% of full scale. Dividing the histogram into channels with widths equal to 0.9% of full scale results in 111 channels, or 6.79 bit ER. Conversely, the ER of channel 0 on a logarithmically binned histogram is greater than 20 bits. Figure 4 shows how the ER changes across the entire range of a 10 bit 4 and 5.41 decade histogram, respectively. In both cases, the maximum effective resolution (MER) is far greater than the nominal resolution of 10 bit ADC converter itself, which was the original intent of the logarithmic conversion. The change in effective resolution along the histogram gives rise to two visual artifacts that have been previously noted in the literature. The first will be termed the valley artifact, which has been described previously in Refs. 3, 6 and 9. Briefly, as the gain on a particular channel was decreased, the population moved towards the origin, as expected. However, this only happened up to a point. Once the gain was decreased further, the visible peak appeared to decrease in size and large numbers of cells began appearing on the axis. Figure 3 shows how, as the mean of a distribution with a given standard deviation (SD) is decreased, the distribution mode approaches the value of the SD as the mean approaches and is decreased below the value of the SD. As the mean decreases below the SD, the mode stays fixed and the peak height at the mode decreases. The accumulation of events on the axis is dependent on several factors including the acquisition electronics. For the duration of the paper the term valley artifact will refer to the decrease in the count values close to channel zero. Cells were never visible in what appeared to be a valley between the peak Figure 3. As the mean of a gaussian distribution with a specific SD is decreased, the value of the mode of the distribution tracks the value of the mean until the value of the mean approaches the value of the SD. As the value of the mean becomes less than the value of the SD, the mode remains fixed at a value close to the value of the SD. This is commonly observed when applying spectral compensation to flow cytometry data. As the compensation is increased, the mean and mode will initially decrease together with the mode finally becoming stationary. At this point the valley artifact becomes evident as the area between the mode and the histogram origin. 688 Flow Cytometry Histograms

5 and the axis (3). The second visual artifact is the picket fencing artifact, shown in Figure 5. Figure 5A shows the population displayed on a linear axis, where it is clear that there is only a single population. However, when this data is graphed on a logarithmic axis (Fig. 5B), a picketed pattern appears near the origin, making it appear as if there are many discrete populations. Pickets can also be observed in two-dimensional plots. The change in ER from the low end to the high end of the histogram explains both the valley and picket fencing artifacts. These artifacts result from the interplay between the resolution of the data and the ER of the histogram on which the data is being plotted. The bulk of the remainder of this paper will explore this relationship, which is critical for understanding both the source of the artifacts and why they are not seen with alternative binning functions. EXPLANATION OF VALLEY ARTIFACT The valley artifact is most evident when the standard deviation of the population approaches the value of the population mean. Thus a significant portion of the APF lies below the first histogram channel. A simple explanation for the valley artifact is that channels at the low end of the logarithmically binned histogram sample from such a small range of the original signal that the probability of an event resulting in a signal that falls into that range becomes very small. This is analogous to the phenomenon described in Figure 2A; however, in that case the bins were too small along the entire range of the data. In the case of the logarithmic histogram, the bins only become too small towards the origin. This almost always includes at least the first decade of the logarithmic histogram. From a mathematical point of view, the ER of the histogram increases exponentially with decreasing channel. If the dynamic range of the log converter would be increased (i.e. the low end of the scale were to be extended), it would only serve to increase the apparent width of the valley. The valley artifact will be evident for any peak where the SD of the peak approaches the mean, i.e., where a significant proportion of the events in the peak lie on the axis (Fig. 3). In these cases, the frequency distribution displayed by the logarithmically binned histogram does not provide an easily recognizable representation of the APF. It is possible to correct for the variable bin width by dividing the frequency distribution histogram values by the channel width so that the frequency per unit channel is plotted. This is the definition of a binned probability density histogram and histograms constructed in this manner appear similar to the histograms displayed in Figures 1B and 1E. In fact given a sufficient enough number of events per channel, it is possible to recover the shape of the APF from the logarithmic histogram by normalizing the data to the histogram binwidth to reflect the relative ER of each bin (10,11). Figure 4. Effective resolution of a histogram changes along the axis. Note that Y-axis is a base 2 logarithmic scale. Dashed Line: Effective resolution of a 10 bit; 4 decade histogram Solid line: Effective resolution of a 10 bit; 5.41 decade histogram. EXPLANATION OF PICKET FENCING The picket fencing artifact results from improper graphing of the variable sized channels at the low end of the logarithmic histogram. Picket fencing occurs when data are improperly binned into a histogram whose MER is greater than that of the data itself. Figure 4A shows that a 10 bit, 4 decade log histogram has a MER of 20 bits. Logarithmic amplifiers sample data directly from the analog signal, the resolution of which will always exceed the MER of whatever histogram into which it is eventually binned. The lack of picket fencing in instruments with logarithmic amplifiers is not caused by so called nonlogarithmic behavior, amplifier noise, or random digitization errors associated with the conversion of small signals. It is simply that the original analog signal has sufficient resolution to be graphed on a histogram with a 20 bit MER. However, in newer digital instruments, the logarithmic conversion is performed in software on data that has already been binned to a finite resolution by an ADC. Incorrectly graphing data in the region of a histogram where the ER exceeds the resolution of the ADC will result in the picket fencing artifact. This can be appreciated visually in Figure 6A, which is simply a magnified view of the histogram in Figure 5B. Notice that the width of the individual peaks (or pickets) representing a channel is constant, whereas the width of the actual channel changes along the X-axis. To fill up the entire first channel on the logarithmic histogram with peaks as narrow as shown, the original digitized data would need many fractional values between 1 and 2, which do not exist. Placing all the events from channel 1 of the original data at the leftmost border of channel 1 on the logarithmic histogram gives the impression that the resolution of the original data is higher than it is, i.e., it appears that all of the original data is really at A more visually accurate graphing of the data would spread the data from the original channel over the entire range it represents on the logarithmic scale. This is why smoothing or dithering works to reduce picket fencing. Just as many linear channels are binned into a single log channel at the high end of the log scale, at the low end of the logarithmic histo- Cytometry Part A 73A: ,

6 Figure 5. Picket fencing example. (A) Five thousand cells from a simulated normally distributed population with a mean in channel 30 and a 33% CV on an 18 bit linear scale (only first 100 channels are shown). (B) The linear data was transformed to log in software and graphed on a log scale. gram a single linear channel should be spread over many logarithmic channels. Figures 6B and 6C show the data properly graphed on a log scale. The individual peaks in Figure 6B are blockier than in Figure 5B, which accurately reflects the varying widths of the channels. Notice that the counts in Figure 6C are less than the corresponding counts in Figure 6A because the data is spread over multiple logarithmic channels. The blockiness is also apparent when data are properly graphed on 2D plots (Fig. 7A), however, if the data are graphed on a density plot, with the background color matching the color used for low density events, then much of the undesirable blockiness blends into the background (Fig. 7B). the raw data will ensure that picket fencing artifacts do not appear since the ER of the plot never exceeds the resolution of the original data. However, it will not necessarily eliminate the CHARACTERISTICS OF NEW HISTOGRAM SCALE TRANSFORMATIONS Recently, several different binning transformations have been described (3 7) that are becoming more popular because they allow one to view a large dynamic range of data on a single plot, while allowing the display of negative data values, and minimizing valley and picket fencing artifacts. While each of the transforms is based upon different equations, they all exhibit very similar properties. All of the transforms bin the data logarithmically at high channel values and linearly at low channel values. All of the transforms allow for the adjustment of a transition point between the logarithmic and linear binning, first derivative matched (4,6), or more gradual like the biexponential (5,7) and hyperlog (3). We use the generic term transition point to refer to the w parameter of the biexponential transformation (5) and the b parameter hyperlog transformation (3). The primary effect of adjusting the transition point is to limit the MER of the transformed plot. This ensures that the width of the binned channels never gets so small that the chance of an event being assigned to that channel becomes negligible. Figures 8A and 8B show the effective resolution of all the transformations, with the parameters adjusted so that the resolution does not exceed 18 and 14 bits, respectively. It can be seen that the main difference between the transformations is the way that they transition from logarithmic to linear binning. Adjusting the parameters of the transformation such that the MER of the histogram does not exceed the resolution of Figure 6. Correct graphing of data where MER of histogram exceeds resolution of the data. (A) Magnified view of same histogram as displayed in Figure 5B. (B) Same data as in Figure 5B and histogram is generated using the interpolation algorithm described in the text. (C) First decade of data from Figure 6B. 690 Flow Cytometry Histograms

7 strongly dependent on the choice of the transition point. Because the transition point is not usually displayed with the histogram, the investigator needs to verify that any two histograms are truly comparable. Figure 9 shows an example of two similar datasets displayed with differing transition points that give distinctly different visual impressions. The data in Figures 9A and 9B are essentially identical, except that the data shown in Figure 9B contains a few cells with higher negative X-axis (egfp) values than in Figure 9A. This caused the software to calculate a different transition value for the biexponential transformation in Figures 9A and 9B. Since the X-axis in Figure 9B displays a much wider range of negative values, the data on the plots appear visually different, which can be misleading unless care is taken to inspect the axis labels. Figure 9C shows the data in 9B plotted on the same X-axis scale (T 5 262,144, w 5 50) as 9A. CONCLUSIONS To summarize, there is a distinct difference between simply scaling data by changing the relative positions of the axis coordinates and scaling the data by changing the histogram bin-widths. The valley and picket fencing artifact associated with the logarithmic binning both result from the effective resolution of the logarithmic histogram increasing as the chan- Figure 7. 2D plots graphed with corrections to eliminate picket fencing. (A) Standard dot plot. Large blocks near the axis result from the large size of the channels near the origin. (B) Density plots eliminate much of the undesired visual effects associated with the large channels near the axis. valley artifact. Depending on the CVs of the populations and the number of events in the low end of the data, it may be necessary to further decrease the MER so that the lower channels of the histogram have enough events to see some structure in the data. In fact, transition points suggested by the authors (3,5) result in histograms with MERs that are far lower than the resolution of the data of most modern flow cytometers. These new transformations introduce a new caveat when displaying histogram data. The placement of the transition can have profound impact on the display of the data. Firstly, if the transition point occurs within a population distribution, it has not been demonstrated that the distribution would not be distorted by the transition from logarithmic to linear behavior. Secondly, when visually comparing histograms between data sets and sometimes within data sets, it is important that the transition point is the same since the scaling of the data is Figure 8. Effective resolution of different display scale transformations. Transition parameters were adjusted to such that the MER is either 18 (A) or14(b) bit. Solid black, logarithmic (4 decade, 10 bit); solid gray, derivative matched; dashed gray, hyperlog; dashed black, biexponential. Cytometry Part A 73A: ,

8 channel is too low to see any structure. The picket fencing results from improper graphing of the data in the region where the histogram resolution exceeds the resolution of the instrument s ADC s resolution. New transformations that transition between linear and logarithmic binning help to minimize these artifacts and provide a way to display negative data values. The primary motivation of these transformations is to make the data appear more visually intuitive. The transition parameters of these transformations should be held constant whenever data is compared to avoid plots that are visually misleading solely as a result of differing axis scales. MATHEMATICAL MODELING AND GRAPHICS METHODS Simulations were simulated using Microsoft Excel or custom software written in Delphi (Code Gear, Scotts Valley, CA). Unless otherwise specified the simulation results and flow cytometry data were plotted with FCS Express Version 3 (De Novo Software, Los Angeles, CA). ACKNOWLEDGMENTS The authors thank Howard Shapiro, John Nolan, Ralph Rossi, and Alessio Palini for their insightful comments when reviewing the manuscript. They thank Ger van den Engh for his suggestions that resulted in Figure 7B. They also thank Nicole Beauchamp of Wake Forest University for the data used in Figure 9. Figure 9. The effect of changing the transition parameter on data display. Data in panels A and B are almost identical, however, the BD FACSDiva TM software automatically calculated a transition value for the biexponential function that binned the data differently, resulting in radically different appearance of the data. Panel C shows the same data as Panel B, with the biexponential parameters adjusted in FCS Express TM such that the binning was similar to that shown in Panel A. The differing scales on the X-axis make it difficult to interpret the data. [Color figure can be viewed in the online issue, which is available at nel number decreases. The valley artifact results from the resolution increasing so dramatically that the widths of logarithmic channels become so small that the number of events per LITERATURE CITED 1. Boas ML. Mathematical Methods in the Physical Sciences. New York: Wiley; pp Shapiro HM. Practical Flow Cytometry, 4th ed. New Jersey: Wiley; pp Bagwell CB. Hyperlog-a flexible log-like transform for negative, zero, and positive valued data. Cytometry A 2005;64A: Battye FL. A mathematical simple alternative to the logarithmic transform for flow cytometric fluorescence data displays. In: 2005 ISAC Samuel A. Latt Conference, Queensland, Australia, November, Parks DR, Roederer M, Moore WA. A new Logicle display method avoids deceptive effects of logarithmic scaling for low signals and compensated data. Cytometry A 2006;69A: Wood JCS. Techniques to compress the scale of flow cytometry data: Benefits, artifacts and solutions. Cytometry A 2004;59A: Herzenberg LA, Tung J, Moore WA, Herzenberg LA, Parks DR. Interpreting flow cytometry data: A guide for the perplexed. Nat Immunol 2006;7: Scott DW. On optimal and data-based histograms. Biometrika 1979;66: Roederer M. Spectral compensation for flow cytometry: Visualization artifacts, limitations, and caveats. Cytometry 2001;45: Wood JCS. Fundamental flow cytometer properties governing sensitivity and resolution. Cytometry 1998;33: Wood JCS, Quintana J. Using basic physical parameters to predict instrument performance at low fluorescence light levels. Cytometry 1998;34: Flow Cytometry Histograms

DELTA MODULATION AND DPCM CODING OF COLOR SIGNALS

DELTA MODULATION AND DPCM CODING OF COLOR SIGNALS DELTA MODULATION AND DPCM CODING OF COLOR SIGNALS Item Type text; Proceedings Authors Habibi, A. Publisher International Foundation for Telemetering Journal International Telemetering Conference Proceedings

More information

Lecture 2 Video Formation and Representation

Lecture 2 Video Formation and Representation 2013 Spring Term 1 Lecture 2 Video Formation and Representation Wen-Hsiao Peng ( 彭文孝 ) Multimedia Architecture and Processing Lab (MAPL) Department of Computer Science National Chiao Tung University 1

More information

MODE FIELD DIAMETER AND EFFECTIVE AREA MEASUREMENT OF DISPERSION COMPENSATION OPTICAL DEVICES

MODE FIELD DIAMETER AND EFFECTIVE AREA MEASUREMENT OF DISPERSION COMPENSATION OPTICAL DEVICES MODE FIELD DIAMETER AND EFFECTIVE AREA MEASUREMENT OF DISPERSION COMPENSATION OPTICAL DEVICES Hale R. Farley, Jeffrey L. Guttman, Razvan Chirita and Carmen D. Pâlsan Photon inc. 6860 Santa Teresa Blvd

More information

Frequencies. Chapter 2. Descriptive statistics and charts

Frequencies. Chapter 2. Descriptive statistics and charts An analyst usually does not concentrate on each individual data values but would like to have a whole picture of how the variables distributed. In this chapter, we will introduce some tools to tabulate

More information

BitWise (V2.1 and later) includes features for determining AP240 settings and measuring the Single Ion Area.

BitWise (V2.1 and later) includes features for determining AP240 settings and measuring the Single Ion Area. BitWise. Instructions for New Features in ToF-AMS DAQ V2.1 Prepared by Joel Kimmel University of Colorado at Boulder & Aerodyne Research Inc. Last Revised 15-Jun-07 BitWise (V2.1 and later) includes features

More information

Interface Practices Subcommittee SCTE STANDARD SCTE Measurement Procedure for Noise Power Ratio

Interface Practices Subcommittee SCTE STANDARD SCTE Measurement Procedure for Noise Power Ratio Interface Practices Subcommittee SCTE STANDARD SCTE 119 2018 Measurement Procedure for Noise Power Ratio NOTICE The Society of Cable Telecommunications Engineers (SCTE) / International Society of Broadband

More information

Spectrum Analyser Basics

Spectrum Analyser Basics Hands-On Learning Spectrum Analyser Basics Peter D. Hiscocks Syscomp Electronic Design Limited Email: phiscock@ee.ryerson.ca June 28, 2014 Introduction Figure 1: GUI Startup Screen In a previous exercise,

More information

Guide to the FACS Calibur

Guide to the FACS Calibur Guide to the FACS Calibur Turn on the pump. Check to see if the waste is full, if so replace it with an empty box (label it waste) of FACSflow sheath fluid and replace the sheath with a new box (I will

More information

For the SIA. Applications of Propagation Delay & Skew tool. Introduction. Theory of Operation. Propagation Delay & Skew Tool

For the SIA. Applications of Propagation Delay & Skew tool. Introduction. Theory of Operation. Propagation Delay & Skew Tool For the SIA Applications of Propagation Delay & Skew tool Determine signal propagation delay time Detect skewing between channels on rising or falling edges Create histograms of different edge relationships

More information

What is Statistics? 13.1 What is Statistics? Statistics

What is Statistics? 13.1 What is Statistics? Statistics 13.1 What is Statistics? What is Statistics? The collection of all outcomes, responses, measurements, or counts that are of interest. A portion or subset of the population. Statistics Is the science of

More information

Chapter 27. Inferences for Regression. Remembering Regression. An Example: Body Fat and Waist Size. Remembering Regression (cont.)

Chapter 27. Inferences for Regression. Remembering Regression. An Example: Body Fat and Waist Size. Remembering Regression (cont.) Chapter 27 Inferences for Regression Copyright 2007 Pearson Education, Inc. Publishing as Pearson Addison-Wesley Slide 27-1 Copyright 2007 Pearson Education, Inc. Publishing as Pearson Addison-Wesley An

More information

PulseCounter Neutron & Gamma Spectrometry Software Manual

PulseCounter Neutron & Gamma Spectrometry Software Manual PulseCounter Neutron & Gamma Spectrometry Software Manual MAXIMUS ENERGY CORPORATION Written by Dr. Max I. Fomitchev-Zamilov Web: maximus.energy TABLE OF CONTENTS 0. GENERAL INFORMATION 1. DEFAULT SCREEN

More information

The Measurement Tools and What They Do

The Measurement Tools and What They Do 2 The Measurement Tools The Measurement Tools and What They Do JITTERWIZARD The JitterWizard is a unique capability of the JitterPro package that performs the requisite scope setup chores while simplifying

More information

Chapter 7. Scanner Controls

Chapter 7. Scanner Controls Chapter 7 Scanner Controls Gain Compensation Echoes created by similar acoustic mismatches at interfaces deeper in the body return to the transducer with weaker amplitude than those closer because of the

More information

LabView Exercises: Part II

LabView Exercises: Part II Physics 3100 Electronics, Fall 2008, Digital Circuits 1 LabView Exercises: Part II The working VIs should be handed in to the TA at the end of the lab. Using LabView for Calculations and Simulations LabView

More information

FACSAria I Standard Operation Protocol Basic Operation

FACSAria I Standard Operation Protocol Basic Operation FACSAria I Standard Operation Protocol Basic Operation 1. Checking Lasers Status a. Please check the ON / OFF of the lasers. Sufficient time (~30 minutes) need to be given to allow the laser(s) to warm

More information

On Figure of Merit in PAM4 Optical Transmitter Evaluation, Particularly TDECQ

On Figure of Merit in PAM4 Optical Transmitter Evaluation, Particularly TDECQ On Figure of Merit in PAM4 Optical Transmitter Evaluation, Particularly TDECQ Pavel Zivny, Tektronix V1.0 On Figure of Merit in PAM4 Optical Transmitter Evaluation, Particularly TDECQ A brief presentation

More information

Introduction. Edge Enhancement (SEE( Advantages of Scalable SEE) Lijun Yin. Scalable Enhancement and Optimization. Case Study:

Introduction. Edge Enhancement (SEE( Advantages of Scalable SEE) Lijun Yin. Scalable Enhancement and Optimization. Case Study: Case Study: Scalable Edge Enhancement Introduction Edge enhancement is a post processing for displaying radiologic images on the monitor to achieve as good visual quality as the film printing does. Edges

More information

DCI Requirements Image - Dynamics

DCI Requirements Image - Dynamics DCI Requirements Image - Dynamics Matt Cowan Entertainment Technology Consultants www.etconsult.com Gamma 2.6 12 bit Luminance Coding Black level coding Post Production Implications Measurement Processes

More information

Statistics for Engineers

Statistics for Engineers Statistics for Engineers ChE 4C3 and 6C3 Kevin Dunn, 2013 kevin.dunn@mcmaster.ca http://learnche.mcmaster.ca/4c3 Overall revision number: 19 (January 2013) 1 Copyright, sharing, and attribution notice

More information

Chapter 6. Normal Distributions

Chapter 6. Normal Distributions Chapter 6 Normal Distributions Understandable Statistics Ninth Edition By Brase and Brase Prepared by Yixun Shi Bloomsburg University of Pennsylvania Edited by José Neville Díaz Caraballo University of

More information

E X P E R I M E N T 1

E X P E R I M E N T 1 E X P E R I M E N T 1 Getting to Know Data Studio Produced by the Physics Staff at Collin College Copyright Collin College Physics Department. All Rights Reserved. University Physics, Exp 1: Getting to

More information

Lab P-6: Synthesis of Sinusoidal Signals A Music Illusion. A k cos.! k t C k / (1)

Lab P-6: Synthesis of Sinusoidal Signals A Music Illusion. A k cos.! k t C k / (1) DSP First, 2e Signal Processing First Lab P-6: Synthesis of Sinusoidal Signals A Music Illusion Pre-Lab: Read the Pre-Lab and do all the exercises in the Pre-Lab section prior to attending lab. Verification:

More information

homework solutions for: Homework #4: Signal-to-Noise Ratio Estimation submitted to: Dr. Joseph Picone ECE 8993 Fundamentals of Speech Recognition

homework solutions for: Homework #4: Signal-to-Noise Ratio Estimation submitted to: Dr. Joseph Picone ECE 8993 Fundamentals of Speech Recognition INSTITUTE FOR SIGNAL AND INFORMATION PROCESSING homework solutions for: Homework #4: Signal-to-Noise Ratio Estimation submitted to: Dr. Joseph Picone ECE 8993 Fundamentals of Speech Recognition May 3,

More information

Algebra I Module 2 Lessons 1 19

Algebra I Module 2 Lessons 1 19 Eureka Math 2015 2016 Algebra I Module 2 Lessons 1 19 Eureka Math, Published by the non-profit Great Minds. Copyright 2015 Great Minds. No part of this work may be reproduced, distributed, modified, sold,

More information

Distribution of Data and the Empirical Rule

Distribution of Data and the Empirical Rule 302360_File_B.qxd 7/7/03 7:18 AM Page 1 Distribution of Data and the Empirical Rule 1 Distribution of Data and the Empirical Rule Stem-and-Leaf Diagrams Frequency Distributions and Histograms Normal Distributions

More information

Iterative Direct DPD White Paper

Iterative Direct DPD White Paper Iterative Direct DPD White Paper Products: ı ı R&S FSW-K18D R&S FPS-K18D Digital pre-distortion (DPD) is a common method to linearize the output signal of a power amplifier (PA), which is being operated

More information

Elasticity Imaging with Ultrasound JEE 4980 Final Report. George Michaels and Mary Watts

Elasticity Imaging with Ultrasound JEE 4980 Final Report. George Michaels and Mary Watts Elasticity Imaging with Ultrasound JEE 4980 Final Report George Michaels and Mary Watts University of Missouri, St. Louis Washington University Joint Engineering Undergraduate Program St. Louis, Missouri

More information

ZONE PLATE SIGNALS 525 Lines Standard M/NTSC

ZONE PLATE SIGNALS 525 Lines Standard M/NTSC Application Note ZONE PLATE SIGNALS 525 Lines Standard M/NTSC Products: CCVS+COMPONENT GENERATOR CCVS GENERATOR SAF SFF 7BM23_0E ZONE PLATE SIGNALS 525 lines M/NTSC Back in the early days of television

More information

Blueline, Linefree, Accuracy Ratio, & Moving Absolute Mean Ratio Charts

Blueline, Linefree, Accuracy Ratio, & Moving Absolute Mean Ratio Charts INTRODUCTION This instruction manual describes for users of the Excel Standard Celeration Template(s) the features of each page or worksheet in the template, allowing the user to set up and generate charts

More information

Reconstruction of Ca 2+ dynamics from low frame rate Ca 2+ imaging data CS229 final project. Submitted by: Limor Bursztyn

Reconstruction of Ca 2+ dynamics from low frame rate Ca 2+ imaging data CS229 final project. Submitted by: Limor Bursztyn Reconstruction of Ca 2+ dynamics from low frame rate Ca 2+ imaging data CS229 final project. Submitted by: Limor Bursztyn Introduction Active neurons communicate by action potential firing (spikes), accompanied

More information

Chapter 1. Introduction to Digital Signal Processing

Chapter 1. Introduction to Digital Signal Processing Chapter 1 Introduction to Digital Signal Processing 1. Introduction Signal processing is a discipline concerned with the acquisition, representation, manipulation, and transformation of signals required

More information

UNIVERSAL SPATIAL UP-SCALER WITH NONLINEAR EDGE ENHANCEMENT

UNIVERSAL SPATIAL UP-SCALER WITH NONLINEAR EDGE ENHANCEMENT UNIVERSAL SPATIAL UP-SCALER WITH NONLINEAR EDGE ENHANCEMENT Stefan Schiemenz, Christian Hentschel Brandenburg University of Technology, Cottbus, Germany ABSTRACT Spatial image resizing is an important

More information

Measurement of overtone frequencies of a toy piano and perception of its pitch

Measurement of overtone frequencies of a toy piano and perception of its pitch Measurement of overtone frequencies of a toy piano and perception of its pitch PACS: 43.75.Mn ABSTRACT Akira Nishimura Department of Media and Cultural Studies, Tokyo University of Information Sciences,

More information

Dither Explained. An explanation and proof of the benefit of dither. for the audio engineer. By Nika Aldrich. April 25, 2002

Dither Explained. An explanation and proof of the benefit of dither. for the audio engineer. By Nika Aldrich. April 25, 2002 Dither Explained An explanation and proof of the benefit of dither for the audio engineer By Nika Aldrich April 25, 2002 Several people have asked me to explain this, and I have to admit it was one of

More information

An Empirical Analysis of Macroscopic Fundamental Diagrams for Sendai Road Networks

An Empirical Analysis of Macroscopic Fundamental Diagrams for Sendai Road Networks Interdisciplinary Information Sciences Vol. 21, No. 1 (2015) 49 61 #Graduate School of Information Sciences, Tohoku University ISSN 1340-9050 print/1347-6157 online DOI 10.4036/iis.2015.49 An Empirical

More information

Getting Started. Connect green audio output of SpikerBox/SpikerShield using green cable to your headphones input on iphone/ipad.

Getting Started. Connect green audio output of SpikerBox/SpikerShield using green cable to your headphones input on iphone/ipad. Getting Started First thing you should do is to connect your iphone or ipad to SpikerBox with a green smartphone cable. Green cable comes with designators on each end of the cable ( Smartphone and SpikerBox

More information

Technical Bulletin. Standardizing Application Setup Across Multiple Flow Cytometers Using BD FACSDiva Version 6 Software. Abstract

Technical Bulletin. Standardizing Application Setup Across Multiple Flow Cytometers Using BD FACSDiva Version 6 Software. Abstract arch 212 Standardizing Application Setup Across ultiple Flow Cytometers Using BD FACSDiva Version 6 Ellen einelt, ervi Reunanen, ark Edinger, aria Jaimes, Alan Stall, Dennis Sasaki, Joe Trotter Contents

More information

MIE 402: WORKSHOP ON DATA ACQUISITION AND SIGNAL PROCESSING Spring 2003

MIE 402: WORKSHOP ON DATA ACQUISITION AND SIGNAL PROCESSING Spring 2003 MIE 402: WORKSHOP ON DATA ACQUISITION AND SIGNAL PROCESSING Spring 2003 OBJECTIVE To become familiar with state-of-the-art digital data acquisition hardware and software. To explore common data acquisition

More information

Signal Stability Analyser

Signal Stability Analyser Signal Stability Analyser o Real Time Phase or Frequency Display o Real Time Data, Allan Variance and Phase Noise Plots o 1MHz to 65MHz medium resolution (12.5ps) o 5MHz and 10MHz high resolution (50fs)

More information

LCD and Plasma display technologies are promising solutions for large-format

LCD and Plasma display technologies are promising solutions for large-format Chapter 4 4. LCD and Plasma Display Characterization 4. Overview LCD and Plasma display technologies are promising solutions for large-format color displays. As these devices become more popular, display

More information

The Extron MGP 464 is a powerful, highly effective tool for advanced A/V communications and presentations. It has the

The Extron MGP 464 is a powerful, highly effective tool for advanced A/V communications and presentations. It has the MGP 464: How to Get the Most from the MGP 464 for Successful Presentations The Extron MGP 464 is a powerful, highly effective tool for advanced A/V communications and presentations. It has the ability

More information

CFlow User Guide. Science is hard. Flow cytometry should be easy.

CFlow User Guide. Science is hard. Flow cytometry should be easy. CFlow User Guide Science is hard. Flow cytometry should be easy. Table of Contents Chapter 1 The Basics - Operating the Accuri C6 Flow Cytometer... 4 Introduction to CFlow and Running the C6... 4 CFlow

More information

FPGA Hardware Resource Specific Optimal Design for FIR Filters

FPGA Hardware Resource Specific Optimal Design for FIR Filters International Journal of Computer Engineering and Information Technology VOL. 8, NO. 11, November 2016, 203 207 Available online at: www.ijceit.org E-ISSN 2412-8856 (Online) FPGA Hardware Resource Specific

More information

m RSC Chromatographie Integration Methods Second Edition CHROMATOGRAPHY MONOGRAPHS Norman Dyson Dyson Instruments Ltd., UK

m RSC Chromatographie Integration Methods Second Edition CHROMATOGRAPHY MONOGRAPHS Norman Dyson Dyson Instruments Ltd., UK m RSC CHROMATOGRAPHY MONOGRAPHS Chromatographie Integration Methods Second Edition Norman Dyson Dyson Instruments Ltd., UK THE ROYAL SOCIETY OF CHEMISTRY Chapter 1 Measurements and Models The Basic Measurements

More information

Removing the Pattern Noise from all STIS Side-2 CCD data

Removing the Pattern Noise from all STIS Side-2 CCD data The 2010 STScI Calibration Workshop Space Telescope Science Institute, 2010 Susana Deustua and Cristina Oliveira, eds. Removing the Pattern Noise from all STIS Side-2 CCD data Rolf A. Jansen, Rogier Windhorst,

More information

Why t? TEACHER NOTES MATH NSPIRED. Math Objectives. Vocabulary. About the Lesson

Why t? TEACHER NOTES MATH NSPIRED. Math Objectives. Vocabulary. About the Lesson Math Objectives Students will recognize that when the population standard deviation is unknown, it must be estimated from the sample in order to calculate a standardized test statistic. Students will recognize

More information

User s Manual. Log Scale (/LG) GX10/GX20/GP10/GP20/GM10 IM 04L51B01-06EN. 2nd Edition

User s Manual. Log Scale (/LG) GX10/GX20/GP10/GP20/GM10 IM 04L51B01-06EN. 2nd Edition User s Manual Model GX10/GX20/GP10/GP20/GM10 Log Scale (/LG) User s Manual 2nd Edition Introduction Notes Trademarks Thank you for purchasing the SMARTDAC+ Series GX10/GX20/GP10/GP20/GM10 (hereafter referred

More information

CS229 Project Report Polyphonic Piano Transcription

CS229 Project Report Polyphonic Piano Transcription CS229 Project Report Polyphonic Piano Transcription Mohammad Sadegh Ebrahimi Stanford University Jean-Baptiste Boin Stanford University sadegh@stanford.edu jbboin@stanford.edu 1. Introduction In this project

More information

Erasing 9840 and 9940 tapes

Erasing 9840 and 9940 tapes Erasing 9840 and 9940 tapes Erasing data tapes was fairly simple in the past. Bulk erasers, also known as degausers, did a good job of demagnetizing the tapes and erasing all data. With newer tapes, such

More information

Evaluating Oscilloscope Mask Testing for Six Sigma Quality Standards

Evaluating Oscilloscope Mask Testing for Six Sigma Quality Standards Evaluating Oscilloscope Mask Testing for Six Sigma Quality Standards Application Note Introduction Engineers use oscilloscopes to measure and evaluate a variety of signals from a range of sources. Oscilloscopes

More information

Realizing Waveform Characteristics up to a Digitizer s Full Bandwidth Increasing the effective sampling rate when measuring repetitive signals

Realizing Waveform Characteristics up to a Digitizer s Full Bandwidth Increasing the effective sampling rate when measuring repetitive signals Realizing Waveform Characteristics up to a Digitizer s Full Bandwidth Increasing the effective sampling rate when measuring repetitive signals By Jean Dassonville Agilent Technologies Introduction The

More information

User s Manual. Log Scale (/LG) GX10/GX20/GP10/GP20/GM10 IM 04L51B01-06EN. 3rd Edition

User s Manual. Log Scale (/LG) GX10/GX20/GP10/GP20/GM10 IM 04L51B01-06EN. 3rd Edition User s Manual Model GX10/GX20/GP10/GP20/GM10 Log Scale (/LG) 3rd Edition Introduction Thank you for purchasing the SMARTDAC+ Series GX10/GX20/GP10/GP20/GM10 (hereafter referred to as the recorder, GX,

More information

2. AN INTROSPECTION OF THE MORPHING PROCESS

2. AN INTROSPECTION OF THE MORPHING PROCESS 1. INTRODUCTION Voice morphing means the transition of one speech signal into another. Like image morphing, speech morphing aims to preserve the shared characteristics of the starting and final signals,

More information

HEBS: Histogram Equalization for Backlight Scaling

HEBS: Histogram Equalization for Backlight Scaling HEBS: Histogram Equalization for Backlight Scaling Ali Iranli, Hanif Fatemi, Massoud Pedram University of Southern California Los Angeles CA March 2005 Motivation 10% 1% 11% 12% 12% 12% 6% 35% 1% 3% 16%

More information

Visual Encoding Design

Visual Encoding Design CSE 442 - Data Visualization Visual Encoding Design Jeffrey Heer University of Washington A Design Space of Visual Encodings Mapping Data to Visual Variables Assign data fields (e.g., with N, O, Q types)

More information

CytoFLEX Flow Cytometer Quick Start Guide

CytoFLEX Flow Cytometer Quick Start Guide Sheath Waste CLASS 1 LASER PRODUCT COMPLIES WITH 21 CFR 1040.10 AND 1040.11 EXCEPT FOR DEVIATIONS PURSUANT TO LASER NOTICE NO. 50 DATED JUNE 24, 2007 MANUFACTURED Sheath B49008AC February 2015 CytoFLEX

More information

SEM- EDS Instruction Manual

SEM- EDS Instruction Manual SEM- EDS Instruction Manual Double-click on the Spirit icon ( ) on the desktop to start the software program. I. X-ray Functions Access the basic X-ray acquisition, display and analysis functions through

More information

How to Manage Color in Telemedicine

How to Manage Color in Telemedicine [ Document Identification Number : DIN01022816 ] Digital Color Imaging in Biomedicine, 7-13, 2001.02.28 Yasuhiro TAKAHASHI *1 *1 CANON INC. Office

More information

STAT 113: Statistics and Society Ellen Gundlach, Purdue University. (Chapters refer to Moore and Notz, Statistics: Concepts and Controversies, 8e)

STAT 113: Statistics and Society Ellen Gundlach, Purdue University. (Chapters refer to Moore and Notz, Statistics: Concepts and Controversies, 8e) STAT 113: Statistics and Society Ellen Gundlach, Purdue University (Chapters refer to Moore and Notz, Statistics: Concepts and Controversies, 8e) Learning Objectives for Exam 1: Unit 1, Part 1: Population

More information

Pre-processing of revolution speed data in ArtemiS SUITE 1

Pre-processing of revolution speed data in ArtemiS SUITE 1 03/18 in ArtemiS SUITE 1 Introduction 1 TTL logic 2 Sources of error in pulse data acquisition 3 Processing of trigger signals 5 Revolution speed acquisition with complex pulse patterns 7 Introduction

More information

Color Spaces in Digital Video

Color Spaces in Digital Video UCRL-JC-127331 PREPRINT Color Spaces in Digital Video R. Gaunt This paper was prepared for submittal to the Association for Computing Machinery Special Interest Group on Computer Graphics (SIGGRAPH) '97

More information

The Definition of 'db' and 'dbm'

The Definition of 'db' and 'dbm' P a g e 1 Handout 1 EE442 Spring Semester The Definition of 'db' and 'dbm' A decibel (db) in electrical engineering is defined as 10 times the base-10 logarithm of a ratio between two power levels; e.g.,

More information

Sample Analysis Design. Element2 - Basic Software Concepts (cont d)

Sample Analysis Design. Element2 - Basic Software Concepts (cont d) Sample Analysis Design Element2 - Basic Software Concepts (cont d) Samples per Peak In order to establish a minimum level of precision, the ion signal (peak) must be measured several times during the scan

More information

Processing data with Mestrelab Mnova

Processing data with Mestrelab Mnova Processing data with Mestrelab Mnova This exercise has three parts: a 1D 1 H spectrum to baseline correct, integrate, peak-pick, and plot; a 2D spectrum to plot with a 1 H spectrum as a projection; and

More information

Keysight Technologies Understanding and Improving Network Analyzer Dynamic Range. Application Note

Keysight Technologies Understanding and Improving Network Analyzer Dynamic Range. Application Note Keysight Technologies Understanding and Improving Network Analyzer Dynamic Range Application Note Introduction Achieving the highest possible network analyzer dynamic range is extremely important when

More information

Getting Started with the LabVIEW Sound and Vibration Toolkit

Getting Started with the LabVIEW Sound and Vibration Toolkit 1 Getting Started with the LabVIEW Sound and Vibration Toolkit This tutorial is designed to introduce you to some of the sound and vibration analysis capabilities in the industry-leading software tool

More information

KONRAD JĘDRZEJEWSKI 1, ANATOLIY A. PLATONOV 1,2

KONRAD JĘDRZEJEWSKI 1, ANATOLIY A. PLATONOV 1,2 KONRAD JĘDRZEJEWSKI 1, ANATOLIY A. PLATONOV 1, 1 Warsaw University of Technology Faculty of Electronics and Information Technology, Poland e-mail: ala@ise.pw.edu.pl Moscow Institute of Electronics and

More information

Chapter 5. Describing Distributions Numerically. Finding the Center: The Median. Spread: Home on the Range. Finding the Center: The Median (cont.

Chapter 5. Describing Distributions Numerically. Finding the Center: The Median. Spread: Home on the Range. Finding the Center: The Median (cont. Chapter 5 Describing Distributions Numerically Copyright 2007 Pearson Education, Inc. Publishing as Pearson Addison-Wesley Copyright 2007 Pearson Education, Inc. Publishing as Pearson Addison-Wesley Slide

More information

Machine Vision System for Color Sorting Wood Edge-Glued Panel Parts

Machine Vision System for Color Sorting Wood Edge-Glued Panel Parts Machine Vision System for Color Sorting Wood Edge-Glued Panel Parts Q. Lu, S. Srikanteswara, W. King, T. Drayer, R. Conners, E. Kline* The Bradley Department of Electrical and Computer Eng. *Department

More information

Human Hair Studies: II Scale Counts

Human Hair Studies: II Scale Counts Journal of Criminal Law and Criminology Volume 31 Issue 5 January-February Article 11 Winter 1941 Human Hair Studies: II Scale Counts Lucy H. Gamble Paul L. Kirk Follow this and additional works at: https://scholarlycommons.law.northwestern.edu/jclc

More information

Precise Digital Integration of Fast Analogue Signals using a 12-bit Oscilloscope

Precise Digital Integration of Fast Analogue Signals using a 12-bit Oscilloscope EUROPEAN ORGANIZATION FOR NUCLEAR RESEARCH CERN BEAMS DEPARTMENT CERN-BE-2014-002 BI Precise Digital Integration of Fast Analogue Signals using a 12-bit Oscilloscope M. Gasior; M. Krupa CERN Geneva/CH

More information

Bootstrap Methods in Regression Questions Have you had a chance to try any of this? Any of the review questions?

Bootstrap Methods in Regression Questions Have you had a chance to try any of this? Any of the review questions? ICPSR Blalock Lectures, 2003 Bootstrap Resampling Robert Stine Lecture 3 Bootstrap Methods in Regression Questions Have you had a chance to try any of this? Any of the review questions? Getting class notes

More information

Laser Beam Analyser Laser Diagnos c System. If you can measure it, you can control it!

Laser Beam Analyser Laser Diagnos c System. If you can measure it, you can control it! Laser Beam Analyser Laser Diagnos c System If you can measure it, you can control it! Introduc on to Laser Beam Analysis In industrial -, medical - and laboratory applications using CO 2 and YAG lasers,

More information

Supplemental Material: Color Compatibility From Large Datasets

Supplemental Material: Color Compatibility From Large Datasets Supplemental Material: Color Compatibility From Large Datasets Peter O Donovan, Aseem Agarwala, and Aaron Hertzmann Project URL: www.dgp.toronto.edu/ donovan/color/ 1 Unmixing color preferences In the

More information

CAEN Tools for Discovery

CAEN Tools for Discovery Viareggio March 28, 2011 Introduction: what is the SiPM? The Silicon PhotoMultiplier (SiPM) consists of a high density (up to ~10 3 /mm 2 ) matrix of diodes connected in parallel on a common Si substrate.

More information

Investigation of Digital Signal Processing of High-speed DACs Signals for Settling Time Testing

Investigation of Digital Signal Processing of High-speed DACs Signals for Settling Time Testing Universal Journal of Electrical and Electronic Engineering 4(2): 67-72, 2016 DOI: 10.13189/ujeee.2016.040204 http://www.hrpub.org Investigation of Digital Signal Processing of High-speed DACs Signals for

More information

Design of an Error Output Feedback Digital Delta Sigma Modulator with In Stage Dithering for Spur Free Output Spectrum

Design of an Error Output Feedback Digital Delta Sigma Modulator with In Stage Dithering for Spur Free Output Spectrum Vol. 9, No. 9, 208 Design of an Error Output Feedback Digital Delta Sigma odulator with In Stage Dithering for Spur Free Output Spectrum Sohail Imran Saeed Department of Electrical Engineering Iqra National

More information

A NEW LOOK AT FREQUENCY RESOLUTION IN POWER SPECTRAL DENSITY ESTIMATION. Sudeshna Pal, Soosan Beheshti

A NEW LOOK AT FREQUENCY RESOLUTION IN POWER SPECTRAL DENSITY ESTIMATION. Sudeshna Pal, Soosan Beheshti A NEW LOOK AT FREQUENCY RESOLUTION IN POWER SPECTRAL DENSITY ESTIMATION Sudeshna Pal, Soosan Beheshti Electrical and Computer Engineering Department, Ryerson University, Toronto, Canada spal@ee.ryerson.ca

More information

SPP-100 Module for use with the FSSP Operator Manual

SPP-100 Module for use with the FSSP Operator Manual ` Particle Analysis and Display System (PADS): SPP-100 Module for use with the FSSP Operator Manual DOC-0199 A; PADS 2.8.2 SPP-100 Module 2.8.2 2545 Central Avenue Boulder, CO 80301 USA C O P Y R I G H

More information

Analysis and sorting of cells with FACSAria II flow cytometer Tiina Pessa-Morikawa / Revised

Analysis and sorting of cells with FACSAria II flow cytometer Tiina Pessa-Morikawa / Revised Analysis and sorting of cells with FACSAria II flow cytometer 10.6.2010 Tiina Pessa-Morikawa / Revised 16-03-17 This instruction refers to the operating manual (FAUG, i.e. FACSAria II Users Guide and page

More information

LAB 1: Plotting a GM Plateau and Introduction to Statistical Distribution. A. Plotting a GM Plateau. This lab will have two sections, A and B.

LAB 1: Plotting a GM Plateau and Introduction to Statistical Distribution. A. Plotting a GM Plateau. This lab will have two sections, A and B. LAB 1: Plotting a GM Plateau and Introduction to Statistical Distribution This lab will have two sections, A and B. Students are supposed to write separate lab reports on section A and B, and submit the

More information

Extraction Methods of Watermarks from Linearly-Distorted Images to Maximize Signal-to-Noise Ratio. Brandon Migdal. Advisors: Carl Salvaggio

Extraction Methods of Watermarks from Linearly-Distorted Images to Maximize Signal-to-Noise Ratio. Brandon Migdal. Advisors: Carl Salvaggio Extraction Methods of Watermarks from Linearly-Distorted Images to Maximize Signal-to-Noise Ratio By Brandon Migdal Advisors: Carl Salvaggio Chris Honsinger A senior project submitted in partial fulfillment

More information

Intra-frame JPEG-2000 vs. Inter-frame Compression Comparison: The benefits and trade-offs for very high quality, high resolution sequences

Intra-frame JPEG-2000 vs. Inter-frame Compression Comparison: The benefits and trade-offs for very high quality, high resolution sequences Intra-frame JPEG-2000 vs. Inter-frame Compression Comparison: The benefits and trade-offs for very high quality, high resolution sequences Michael Smith and John Villasenor For the past several decades,

More information

Environmental Controls Laboratory

Environmental Controls Laboratory (Electro-Oculography Application) Introduction Spinal cord injury, cerebral palsy, and stroke are some examples of clinical problems which can have a large effect on upper extremity motor control for afflicted

More information

Circular Statistics Applied to Colour Images

Circular Statistics Applied to Colour Images Circular Statistics pplied to Colour Images llan Hanbury PRIP, TU Wien, Favoritenstraße 9/183, -1040 Vienna, ustria hanbury@prip.tuwien.ac.at bstract Three methods for summarising the characteristics of

More information

Dithering in Analog-to-digital Conversion

Dithering in Analog-to-digital Conversion Application Note 1. Introduction 2. What is Dither High-speed ADCs today offer higher dynamic performances and every effort is made to push these state-of-the art performances through design improvements

More information

MestReNova A quick Guide. Adjust signal intensity Use scroll wheel. Zoomen Z

MestReNova A quick Guide. Adjust signal intensity Use scroll wheel. Zoomen Z MestReNova A quick Guide page 1 MNova is a program to analyze 1D- and 2D NMR data. Start of MNova Start All Programs Chemie NMR MNova The MNova Menu 1. 2. Create expanded regions Adjust signal intensity

More information

Study of White Gaussian Noise with Varying Signal to Noise Ratio in Speech Signal using Wavelet

Study of White Gaussian Noise with Varying Signal to Noise Ratio in Speech Signal using Wavelet American International Journal of Research in Science, Technology, Engineering & Mathematics Available online at http://www.iasir.net ISSN (Print): 2328-3491, ISSN (Online): 2328-3580, ISSN (CD-ROM): 2328-3629

More information

Results of the June 2000 NICMOS+NCS EMI Test

Results of the June 2000 NICMOS+NCS EMI Test Results of the June 2 NICMOS+NCS EMI Test S. T. Holfeltz & Torsten Böker September 28, 2 ABSTRACT We summarize the findings of the NICMOS+NCS EMI Tests conducted at Goddard Space Flight Center in June

More information

White Paper. Uniform Luminance Technology. What s inside? What is non-uniformity and noise in LCDs? Why is it a problem? How is it solved?

White Paper. Uniform Luminance Technology. What s inside? What is non-uniformity and noise in LCDs? Why is it a problem? How is it solved? White Paper Uniform Luminance Technology What s inside? What is non-uniformity and noise in LCDs? Why is it a problem? How is it solved? Tom Kimpe Manager Technology & Innovation Group Barco Medical Imaging

More information

Analysis of WFS Measurements from first half of 2004

Analysis of WFS Measurements from first half of 2004 Analysis of WFS Measurements from first half of 24 (Report4) Graham Cox August 19, 24 1 Abstract Described in this report is the results of wavefront sensor measurements taken during the first seven months

More information

HIGH QUALITY GEOMETRY DISTORTION TOOL FOR USE WITH LCD AND DLP PROJECTORS

HIGH QUALITY GEOMETRY DISTORTION TOOL FOR USE WITH LCD AND DLP PROJECTORS HIGH QUALITY GEOMETRY DISTORTION TOOL FOR USE WITH LCD AND DLP PROJECTORS Ronny Van Belle Electronic Design Engineer & Bart Maximus R&D Manager Barco Projection Systems Simulation Department & Philippe

More information

FPA (Focal Plane Array) Characterization set up (CamIRa) Standard Operating Procedure

FPA (Focal Plane Array) Characterization set up (CamIRa) Standard Operating Procedure FPA (Focal Plane Array) Characterization set up (CamIRa) Standard Operating Procedure FACULTY IN-CHARGE Prof. Subhananda Chakrabarti (IITB) SYSTEM OWNER Hemant Ghadi (ghadihemant16@gmail.com) 05 July 2013

More information

The Effect of Time-Domain Interpolation on Response Spectral Calculations. David M. Boore

The Effect of Time-Domain Interpolation on Response Spectral Calculations. David M. Boore The Effect of Time-Domain Interpolation on Response Spectral Calculations David M. Boore This note confirms Norm Abrahamson s finding that the straight line interpolation between sampled points used in

More information

COMP Test on Psychology 320 Check on Mastery of Prerequisites

COMP Test on Psychology 320 Check on Mastery of Prerequisites COMP Test on Psychology 320 Check on Mastery of Prerequisites This test is designed to provide you and your instructor with information on your mastery of the basic content of Psychology 320. The results

More information

Please feel free to download the Demo application software from analogarts.com to help you follow this seminar.

Please feel free to download the Demo application software from analogarts.com to help you follow this seminar. Hello, welcome to Analog Arts spectrum analyzer tutorial. Please feel free to download the Demo application software from analogarts.com to help you follow this seminar. For this presentation, we use a

More information

Heart Rate Variability Preparing Data for Analysis Using AcqKnowledge

Heart Rate Variability Preparing Data for Analysis Using AcqKnowledge APPLICATION NOTE 42 Aero Camino, Goleta, CA 93117 Tel (805) 685-0066 Fax (805) 685-0067 info@biopac.com www.biopac.com 01.06.2016 Application Note 233 Heart Rate Variability Preparing Data for Analysis

More information

Full Disclosure Monitoring

Full Disclosure Monitoring Full Disclosure Monitoring Power Quality Application Note Full Disclosure monitoring is the ability to measure all aspects of power quality, on every voltage cycle, and record them in appropriate detail

More information

DATA COMPRESSION USING THE FFT

DATA COMPRESSION USING THE FFT EEE 407/591 PROJECT DUE: NOVEMBER 21, 2001 DATA COMPRESSION USING THE FFT INSTRUCTOR: DR. ANDREAS SPANIAS TEAM MEMBERS: IMTIAZ NIZAMI - 993 21 6600 HASSAN MANSOOR - 993 69 3137 Contents TECHNICAL BACKGROUND...

More information