1 REVIEW Posterior Astigmatism: Considerations for Cataract Refractive Surgery Planning Milton S. Yogi, MD, MBA1 Bruna V. Ventura, MD, PhD2 Eliane M. Nakano, MD3 1 Head, Cataract Department, Beneficência Portuguesa de São Paulo, Brazil. Director, IPEPO/Instituto da Visão, São Paulo, Brazil. 2 Academic Coordinator of the Altino Ventura Foundation, Recife, Brazil. 3 Assistant Professor, Surgical Optics Sector, Department of Ophthalmology, Federal University of São Paulo, São Paulo, Brazil. Corresponding author: Milton S. Yogi, MD, MBA Avenida Paulista, º andar Bela Vista milton@godoctor.com.br Funding: None Proprietary/financial interes: None Date of submission: : 20/11/2017 Date of Approval: 20/12/2017 Over the last decade, a vast improvement on intraocular lens (IOL) options steered cataract surgery towards a more sophisticated approach concerning the final refractive goal. On a particular note, the availability and adoption of toric IOLs, either in monofocal or multifocal platforms. This allowed for a better final quality of vision, but also addressed several complex steps, especially regarding proper diagnostic assessment during surgical planning. More than 50% of the patients who undergo cataract surgery have corneal astigmatism 0.75 diopter (D), which is within the range of surgically treatable astigmatism and may significantly limit an optimal visual outcome if left uncorrected. 1 When speaking of astigmatism, the main consideration is related to the anterior cornea. Its evaluation in terms of magnitude, vector, and topographic aspect is critical for toric IOL calculation. The influence of the posterior cornea has been noted since the 19th century, as being the difference between the keratometric values from the anterior corneal surface and the total refractive astigmatism. Javal postulated an approximately linear relationship between corneal astigmatism and refractive astigmatism, 2 which became known as Javal s rule: refractive astigmatism equals 1.25 (keratometric astigmatism) In IOL calculations, the cornea is regarded as a single refractive surface with an effective refractive index, which varies from to , assuming that it consists of a fixed ratio between the radius of the anterior and posterior corneal surface. But this approach disregards considerations of corneal thickness and radius of the posterior corneal surface. The role of the posterior cornea was VISION PAN-AMERICA JAN-MAR 2018 VOL. 17 (1) dismissed because of the small difference between the indices of refraction of the cornea and aqueous. 4 A 1992 study revealed that the posterior surface exhibited more toricity than the anterior surface.5 Another study suggested that ignoring posterior corneal astigmatism (PCA) could be a significant factor with regards to postoperative refractive astigmatism after toric IOL implantation. 6 To further evaluate the cornea as a whole, Scheimpflug photography and its analysis have the advantage of being a non-contact technique whereby the anterior surface, the thickness profile, and the posterior surface are determined in one step. 7 Residual errors after toric IOLs implants were scrutinized by Koch and others (Arturo Chayet, personal communication, 2013) to understand better underlying causes. In a study involving 715 corneas of 435 consecutive patients who underwent analysis before cataract or refractive surgery, Koch et al. 8 evaluated total corneal astigmatism using ray tracing, corneal astigmatism from simulated keratometry, anterior corneal astigmatism and PCA. Corneal astigmatism was measured using five devices (IOL Master, Lenstar, Atlas, Manual Keratometer and Galilei) before and three weeks after surgery. They found that the steep axis of PCA was oriented in the horizontal meridian in approximately 87% of eyes and was not factored into what is typically measured as anterior corneal astigmatism. This would overestimate anterior corneal astigmatism by about 0.50 D for with-the-rule (WTR) anterior corneal astigmatism and underestimate 11

2 it by about 0.30 D for against-the-rule (ATR) anterior corneal astigmatism. 9 Selecting toric lenses based on anterior corneal measurements only could lead to overcorrection in eyes that have WTR astigmatism and undercorrection in eyes that have ATR astigmatism. Based on these findings, Koch et al. proposed a nomogram based on a regression analysis derived from direct PCA measurements (Baylor nomogram). Reference In eyes with WTR astigmatism, the nomogram shifts the threshold for selecting a toric IOL up 0.7 D; conversely, in eyes with ATR astigmatism, the shift of the threshold for selecting a toric IOL goes down 0.7 D. This nomogram aims for postoperative WTR astigmatism of about 0.40D to account for the gradual shift of the astigmatism axis with age from WTR to ATR. It requires surgeons to factor in their average surgically induced astigmatism (SIA), which for most surgeons is 0.2 to 0.3 D. It is important to note that the axis of astigmatism can be flipped to achieve the lowest residual magnitude. 10 Flipping the axis is only of concern in glasses, in which meridional aniseikonia and spatial distortion from spectacles occur due to the base curves, power, astigmatism and vertex distance. Besides the nomogram, the study showed that even with advanced tomographic evaluation, validated ways to accurately measure PCA are still missing, with mean prediction errors of 0.57 D in the WTR group and 0.12 D in the ATR group, even when using dual Scheimpflug technology. Also, results of the total corneal astigmatism from various devices cannot be used interchangeably. Deriving from the Baylor nomogram, a hybrid algorithm was coupled with clinical outcomes of trials of AMO toric IOLs (now Johnson & Johnson Vision) and incorporated into its online toric calculator, as an optional field. 11,12 Another study analyzed the total corneal refractive power (TCRP) astigmatism by ray tracing using a rotating Scheimpflug tomographer (Pentacam HR; Oculus) and the factors that may lead to the false calculation of total corneal astigmatism by using anterior curvature measurements only. 13 Although using a different corneal tomograph from the work of Koch et al., they similarly concluded that ignoring posterior astigmatism could lead to miscalculation of total corneal astigmatism, and, in cases of toric IOL implantation, to significant induce visual overcorrection or undercorrection. Studies using different devices also have shown that the steepest corneal meridian on the posterior corneal surface is vertically oriented in most cases and thus generates an ATR astigmatism Posterior corneal astigmatism on average reduced total corneal astigmatism by 13.4% and in 28.8% of eyes total corneal astigmatism differed from anterior corneal astigmatism by >0.5 D or >10 degrees in meridian. 8 The Barrett formula estimates the PCA for each eye using a proprietary mathematical model19, and details over its contents have not been published. The Barrett Toric Calculator (available at uses the Universal II formula to predict the required spherical equivalent IOL power. The calculator derives the posterior corneal curvature based on a theoretical model proposed to explain the behavior of the posterior cornea.20 The toric IOL cylinder power required to correct corneal astigmatism including PCA is calculated from the predicted effective lens position (ELP) using vector calculations for each eye. 21 Although the posterior cornea prediction utilizes a theoretical model as the default mode Predicted PCA to predict the posterior corneal astigmatism for an individual eye, it also allows the user to enter directly measured values from Scheimpflug or Swept Source OCT devices. However, it has to be noted that the Baylor nomogram offers an average adjustment for posterior astigmatism power based on the preoperative anterior astigmatic direction without addressing the possible effect on the net corneal astigmatism meridian. 22 The Barrett toric algorithm is also incorporated in the Alcon online calculator. The Barrett toric formula is also available in the following optical biometers: Zeiss IOLMaster 700, HS Lenstar 900, HS Eyestar 900, Oculus Pentacam AXL, Topcon Alladin e Movu Argos. On an additional note, an accurate preoperative SIA prediction is crucial to hitting refractive outcomes. Whereas traditional mean SIA calculation relies on magnitude alone, a more recent concept considers magnitude and direction. Using vector analysis to find the geometric mean or centroid, yields the highest prediction accuracy of any method. 23,24 By applying vector analysis to a broad sample of postoperative patient astigmatism data, the usual standard value of mean SIA of 0.5 D has given place to the centroid SIA of 0.1 D (recommended for incisions between 2.2 and 2.4 mm). 25 A centroid calculation can be carried out using a tool available at www. SIA-calculator.com. 12

3 Besides the Baylor nomogram and the Barrett Toric Calculator, PCA has also been addressed by other methods such as the subtoric IOL calculator 26 and the Goggin et al. nomogram. 27 Also, a new regression formula (Abulafia-Koch) was developed to calculate the estimated total corneal astigmatism based on standard keratometry measurements and significantly improved the prediction of postoperative astigmatic outcomes in the adjustment of commercial toric IOL calculators. 28,29 Pentacam The Pentacam (Oculus Optikgeräte GmbH) measures both the anterior and posterior corneal surfaces using a rotating Scheimpflug camera. 30 The device calculates the true net power (TNP) of the cornea by adding sagittal curvature values of the anterior and posterior corneal planes; the ray-tracing calculation is also available via the TCRP. However, because of the refractive indices used, the results cannot be applied in the conventional IOL formulas. Posterior corneal data derived from the Pentacam device also can be incorporated into standard toric IOL calculations. As nomogram adjustments based on sample averages of the residual refractive cylinder may not be effective in all cases, the specific analysis of posterior astigmatism yields better results. Because of the variability in the contribution of the PCA in individual patients 31, the specific analysis of the posterior astigmatism yields better results 6,12. Using ray-tracing technology, the corneal power of the anterior surface and posterior surface can be determined, with the result termed the TCRP. This value can be successfully applied to toric IOL calculations. 32 In a paper from Ho 13 with 493 subjects comparing the keratometric corneal astigmatism with the total corneal astigmatism derived by double-angled vector analysis of both corneal surfaces with the Pentacam (Oculus), total corneal astigmatism differed from anterior corneal astigmatism by more than 0.5 D in 28.8% of eyes, and the PCA reduced total corneal astigmatism by an average of 13.4%. In a study comparing the astigmatism prediction errors taken with the Pentacam measurements, Baylor nomogram, and Barrett formula for toric intraocular lens implantation, vector summation using anterior and posterior corneal surface power taken with the Pentacam yielded the least astigmatism prediction error. 33 Besides obtaining data from the posterior cornea, the Pentacam AXL incorporates the Savini Toric formula. It uses TCRP measurements from the 3mm zone, centered on the pupil. 6 Galilei The Galilei (Ziemer Ophthalmic Systems AG, Port, Switzerland) is a Placido-dual Scheimpflug corneal tomographer that provides a range of data for planning cataract and refractive procedures, including anterior and posterior corneal curvature and astigmatism, total corneal power and astigmatism, elevation and pachymetry maps, keratoconus indices, anterior chamber depth (ACD) and angle kappa. 34 Its most recent version, the Galilei G6, incorporated an optical A-scan that provides non-contact biometric measurements, besides the anterior segment parameters. 35 The Galilei uses a proprietary algorithm to derive anterior segment measurements from the images obtained by 20 Placido rings and two rotating Scheimpflug cameras. The two Scheimpflug cameras on opposite sides of the eye capture image simultaneously and compensate for decentration. The device can obtain up to 60 Scheimpflug scans and 2 Placido images in one second, evaluating up to 100,000 data points. 34 Furthermore, the current device uses an 880 nm A-scan to measure the axial length and includes IOL power calculation formulas. 35 Its ACD measurement derives from data obtained by the Scheimpflug cameras and is defined as the distance between the corneal epithelium and the anterior lens surface. 34 The Galilei is a very promising device used worldwide. Previous papers have shown that some anterior segment measurements obtained by the Galilei are highly repeatable, including anterior and posterior corneal power and simulated keratometry. 34,36,37 However, with regards to posterior corneal astigmatism measurements, moderate repeatability was observed 37, suggesting that hardware/ software improvements are still necessary before establishing this device as the gold standard to measure posterior corneal astigmatism with the objective of inputting the exact measurement in the cataract surgical planning. Cassini The Cassini (i-optics, The Hague, Netherlands) was originally launched as a corneal topographer, which acquired data based on the specular reflection of approximately 700 colored light-emitting diodes (LEDs). 38 More recently, seven additional infrared LEDs were added to measure the posterior corneal curvature, and ray tracing began being used to calculate total corneal astigma- 13

4 tism.39 There are a few papers in the literature assessing the outcomes of the Cassini. The repeatability analysis of the anterior corneal measurements has shown that the device needs improvement when compared to others routinely used by ophthalmologists. 34,36,37 However, Klijn et al. 39 have described a better accuracy in astigmatism measurement when using the total corneal astigmatism obtained by the Cassini, than when using its anterior corneal measurement alone. Thus, the results of this latter study are promising, and future data will show if there is a specific corneal tomographer that is accurate enough to be routinely incorporated in the cataract surgical planning. Perspectives Other advances in newer ssoct-based biometry and keratometry, using anterior and posterior corneal surfaces, corneal thickness, axial length, anterior chamber depth and white-to-white distance, led to a retrospective study in which ray tracing-based IOL power calculation was performed using retinal image quality metric (RIQM) criteria in an iterative procedure, and retinal image simulations. In the ray tracing-based method, IOL selection is based on predicted visual acuity and a neuronal weighted RIQM.40 After comparing the spherical equivalent outcomes of ray tracing-based IOL power calculation, using exact IOL design information, with the outcomes of a triple-optimized Haigis formula, it resulted in a 2.5% increase in outcomes within 0.5D and 0.75D of target refraction, and a 10% increase in those coming within 1.00D. New devices, nomograms, and formulas make the posterior astigmatism consideration an exciting opportunity to better refine cataract refractive surgeries outcomes, making the diagnostic preoperative evaluation a key ingredient for success. References 1. Ferrer-Blasco T, Montés-Micó R, Peixoto-de-Matos SC, González-Méijome JM, Cerviño A. Prevalence of corneal astigmatism before cataract surgery. J Cataract Refract Surg.2009;35(1): Javal, E. Memoires D Ophthalmometrie. in: G. Masson (Ed.) Libraire de L Academie de Medecine, Paris; 1890: Bennet AG, Rabbets RB (1998). Clinical visual optics (3rd ed.). Oxford: Butterworth-Heinemann, p Koch D. The posterior cornea: hiding in plain sight. Ophthalmology. 2015;122(6): Dunne MC, Royston JM, Barnes DA. Normal variations of the posterior corneal surface. Acta Ophthalmologica.1992;70: Savini G, N.ser K. An analysis of the factors influencing the residual refractive astigmatism after cataract surgery with toric intraocular lenses. Invest Ophthalmol Vis Sci. 2015;56(2): Brown N. Slit-image photography and measurement of the eye. Medical and Biological Illustration. 1973; 23: Koch DD, Ali SF, Weikert MP, Shirayama M, Jenkins R, Wang L. Contribution of posterior corneal astigmatism to total cornealastigmatism. J Cataract Refract Surg.2012;38(12): Savini G, Naser K, Schiano Lomoriello D, Ducoli P. Total corneal astigmatism measurements: agreement between 2 Scheimpflug cameras. Cornea.2017;36: Holladay JT. Improving toric IOL outcomes. Ocular Surgery News. Published May 25 and June 10, Canovas C et al. Impact of posterior corneal astigmatism in toric IOL power calculations. American Society of Cataract and Refractive Surgery meeting; May 6-10, 2016; New Orleans. 12. Koch D. Posterior corneal astigmatism can be incorporated into toric IOL selection. Ocular Surgery News U.S. Edition, November 10, Ho JD, Tsai CY, Liou SW. Accuracy of corneal astigmatism estimation by neglecting the posterior corneal surface measurement. Am J Ophthalmol.2009; 147: Savini G, Versaci F, Vestri G, Ducoli P, Naser K. Influence of posterior corneal astigmatism on total corneal astigmatism in eyes with moderate to high astigmatism. J Cataract Refract Surg.2014; 40: Tonn B, Klaproth OK, Kohnen T. Anterior surface based keratometry compared with Scheimpflug tomography based total corneal astigmatism. Invest Ophthalmol Vis Sci. 2015;56: Hoffmann PC, Abraham M, Hirnschall N, Findl O. Prediction of residual astigmatism after cataract surgery using swept source fourier domain optical coherence tomography. Curr Eye Res.2014;39: Preussner PR, Hoffmann P, Wahl J. Impact of posterior corneal surface on toric intraocular lens (IOL) calculation. Curr Eye Res.2015;40: Ueno Y, Hiraoka T, Miyazaki M, Ito M, Oshika, T. Corneal thickness profile and posterior corneal astigmatism in normal corneas. Ophthalmology.2015;122: Abulafia A, Barrett GD, Kleinmann G, Ofir S, Levy A, Marcovich AL, Michaeli A, Koch DD, Wang L, Assia EI. Prediction of refractive outcomes with toric intraocular lens implantation. J Cataract Refract Surg.2015;41(5): Barrett GD. Innovative toric IOL calculators and how to use them: Barrett Toric Calculator. Cataract & Refract Surg Today Europe. May 2015 supplement. 21. Caceres V. Barrett Toric Calculator aims for accurate outcomes. February Retrieved in: Reitblat Olga et al. Effect of posterior corneal astigmatism on power calculation and alignment of toric intraocular lenses: Comparison of methodologies. J Cataract Refract Surg ;42(2): Abulafia A, Barrett GD, Kleinmann G et al. Prediction of refractive outcome with toric intraocular lens implants. J Cataract Refract Surg.2015;41(5):

5 24. Abulafia A, Hill WE, Franchina M, Barrett GD. Comparison of methods to predict residual astigmatism after intraocular lens implantation. J Refract Surg. 2015;31(10): Barrett G. A formula for all seasons. The Barrett Universal II and True K and the Barrett Toric Calculator. Supplement to: Cataract & Refractive Surgery Today Europe, October Eom Y, Rhim JW, Kang SY, Kim SW, Song JS, Kim HM. Toric Intraocular Lens Calculations Using Ratio of Anterior to Posterior Corneal Cylinder Power. Am J Ophthalmol.2015;160(4): Goggin M, Zamora-Alejo K, Esterman A, van Zyl L. Adjustment of anterior corneal astigmatism values to incorporate the likely effect of posterior corneal curvature for toric intraocular lens calculation. J Refract Surg.2015;31(2): Abulafia Adi et al. New regression formula for toric intraocular lens calculations. J Cataract Refract Surg. 2016;42(5): Ferreira Tiago B. et al. Comparison of astigmatic prediction errors associated with new calculation methods for toric intraocular lenses. J Cataract Refract Surg. 2017; 43(3): Savini G, Barboni P, Carbonelli M, Hoffer KJ. Comparison of methods to measure corneal power for intraocular lens power calculation using a rotating Scheimpflug camera. J Cataract Refract Surg. 2013; 39: T. Miyake, K. Shimizu, K. Kamiya. Distribution of posterior corneal astigmatism according to axis orientation of anterior corneal astigmatism. PLoS One. 2015;10(1)., article e , Davison JA, Potvin R. Refractive cylinder outcomes after calculating toric intraocular lens cylinder power using total corneal refractive power. Clinical Ophthalmology (Auckland, NZ). 2015;9: Park DY, Lim DH, Hwang S, Hyun J, Chung TY. Comparison of astigmatism prediction error taken with the Pentacam measurements, Baylor nomogram, and Barrett formula for toric intraocular lens implantation. BMC Ophthalmology. 2017;17: Cervino A, Dominguez-Vicent A, Ferrer-Blasco T, Garcia-Lazaro S, Albarran-Diego C. Intrasubject repeatability of corneal power, thickness, and wavefront aberrations with a new version of a dual rotating Scheimpflug-Placido system. J Cataract Refract Surg. 2015;41:

4/9/2016. Sources of. Single-angle vs. double-angle plots for astigmatism data. Commercial Toric IOL calculators. Unexpected residual astigmatism!

4/9/2016. Sources of. Single-angle vs. double-angle plots for astigmatism data. Commercial Toric IOL calculators. Unexpected residual astigmatism! Sources of Corneal astigmatism measurements Methods of calculation Corneal surgically induced astigmatism (SIA) Toric IOL misalignment Unexpected residual astigmatism! Single-angle vs. double-angle plots

More information

* Villegas EL, Alcón E, Artal P. Minimum amount of astigmatism that should be corrected. J Cataract Refract Surg 2014; 40: n My SIA: Ø Centroid

* Villegas EL, Alcón E, Artal P. Minimum amount of astigmatism that should be corrected. J Cataract Refract Surg 2014; 40: n My SIA: Ø Centroid Astigmatism correction in cataract surgery: A work in progress 9 things you should know Douglas D. Koch, M.D. Cullen Eye Institute Baylor College of Medicine Houston, Texas Financial disclosure: AMO Alc

More information

Total corneal astigmatism in older adults taking into account posterior corneal astigmatism by ray tracing

Total corneal astigmatism in older adults taking into account posterior corneal astigmatism by ray tracing ARTICLE Total corneal astigmatism in older adults taking into account posterior corneal astigmatism by ray tracing Alvaro Rodríguez Ratón, MD 1 ; Javier Orbegozo Gárate, MD 1 ; Iñaki Basterra Barrenetxea,OD

More information

Development of a program for toric intraocular lens calculation. considering posterior corneal astigmatism, incisioninduced

Development of a program for toric intraocular lens calculation. considering posterior corneal astigmatism, incisioninduced DOI 10.1007/s00417-016-3446-3 CATARACT Development of a program for toric intraocular lens calculation considering posterior corneal astigmatism, incision-induced posterior corneal astigmatism, and effective

More information

Comparison of the Astigmatic Power of Toric Intraocular Lenses Using Three Toric Calculators

Comparison of the Astigmatic Power of Toric Intraocular Lenses Using Three Toric Calculators Original Article http://dx.doi.org/10.3349/ymj.2015.56.4.1097 pissn: 0513-5796, eissn: 1976-2437 Yonsei Med J 56(4):1097-1105, 2015 Comparison of the Astigmatic Power of Toric Intraocular Lenses Using

More information

Abstract. imedpub Journals Vol.3 No.2:27. Introduction

Abstract. imedpub Journals Vol.3 No.2:27. Introduction Research Article imedpub Journals http://www.imedpub.com/ Journal of Eye & Cataract Surgery DOI: 10.21767/2471-8300.100027 Intraoperative Biometry versus Conventional Methods for Predicting Intraocular

More information

Richard N. McNeely 1,2, Salissou Moutari 3, Eric Pazo 1,2 and Jonathan E. Moore 1,2*

Richard N. McNeely 1,2, Salissou Moutari 3, Eric Pazo 1,2 and Jonathan E. Moore 1,2* McNeely et al. Eye and Vision (2018) 5:7 https://doi.org/10.1186/s40662-018-0103-4 RESEARCH Investigating the impact of preoperative corneal astigmatism orientation on the postoperative spherical equivalent

More information

Premium treatment starts with premium diagnosis

Premium treatment starts with premium diagnosis by i-optics Premium treatment starts with premium diagnosis Complete your cataract-refractive platform The premium IOL opportunity Premium IOL market to nearly double in next 5 years 2013 1.6 million 22

More information

Arthur Cummings FRCSEd

Arthur Cummings FRCSEd How to Improve your Refractive Cataract Surgery Outcomes by Skilful Interpretation of Corneal Mapping Course IC-16 ESCRS Copenhagen 10 th September 2016 Consultant for Alcon / WaveLight/TearLab Arthur

More information

Management of astigmatism at the time of cataract or refractive lens surgery has evolved to include arcuate keratotomy and toric

Management of astigmatism at the time of cataract or refractive lens surgery has evolved to include arcuate keratotomy and toric Management of Astigmatism with the LENSAR Laser System with Streamline Mark Packer Mark Packer MD Consulting, Inc., Boulder, CO, US DOI: https://doi.org/10.17925/usor.2017.12.99 Management of astigmatism

More information

Toric intraocular lenses

Toric intraocular lenses Supplement to EyeWorld August 2015 Driving adoption and outcomes with toric IOLs: Pre-, intra-, and postoperative pearls for success 2014 ASCRS Clinical Survey: Trends in toric IOL implementation Click

More information

Prospective study of toric IOL outcomes based on the Lenstar LS 900 W dual zone automated keratometer

Prospective study of toric IOL outcomes based on the Lenstar LS 900 W dual zone automated keratometer Gundersen and Potvin BMC Ophthalmology 2012, 12:21 RESEARCH ARTICLE Open Access Prospective study of toric IOL outcomes based on the Lenstar LS 900 W dual zone automated keratometer Kjell Gunnar Gundersen

More information

White Paper. Astigmatism Management With Toric IOLs The Importance of Rotational Stability After IOL Implantation. Xiaolin Gu, M.D., PhD.

White Paper. Astigmatism Management With Toric IOLs The Importance of Rotational Stability After IOL Implantation. Xiaolin Gu, M.D., PhD. White Paper Astigmatism Management With Toric IOLs The Importance of Rotational Stability After IOL Implantation Xiaolin Gu, M.D., PhD. Introduction Cataracts, or clouding of the crystalline lens, are

More information

Novel Microscope Mounted Digital Keratoscope for Intra-Operative Toric IOL Alignment

Novel Microscope Mounted Digital Keratoscope for Intra-Operative Toric IOL Alignment Cronicon OPEN ACCESS EC OPHTHALMOLOGY Research Article Novel Microscope Mounted Digital Keratoscope for Intra-Operative Toric IOL Alignment Sviatlana M Ilyina 1 *, Siarhei M Lohash 2 and Alex Artsyukhovich

More information

POST-OPERATIVE ASTIGMATISM AFTER SICS AND PHACOEMULSIFICATION.

POST-OPERATIVE ASTIGMATISM AFTER SICS AND PHACOEMULSIFICATION. 4 POST-OPERATIVE ASTIGMATISM AFTER SICS AND PHACOEMULSIFICATION. Dr.Vijay Damor, Dr.Anupama Mahant, Department of ophthalmology,amc MET Medical college. Ahmedabad pin 380008 Abstract : Background: Astigmatism

More information

Postoperative Astigmatic Outcomes Based on the Haptic Axis of Intraocular Lenses Inserted in Cataract Surgery

Postoperative Astigmatic Outcomes Based on the Haptic Axis of Intraocular Lenses Inserted in Cataract Surgery pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2011;25(1):22-28 DOI: 10.3341/kjo.2011.25.1.22 Original Article Postoperative Astigmatic Outcomes Based on the Haptic Axis of Intraocular Lenses Inserted

More information

AXsys Studay Data and Press Release Reference

AXsys Studay Data and Press Release Reference Clinically Tested to be the World s Most Accurate Toric Marking Device AXsys Studay Data and Press Release Reference Press Release Reference electronic leveling device for implantation of a toric iol The

More information

Irregular Corneal Astigmatism & Cataract

Irregular Corneal Astigmatism & Cataract Costas Karabatsas MD, PhD, MRCOphth, FEBOphth, FRCS Ophth Irregular Corneal Astigmatism & Cataract (assessing ocular surface, IOL selection) In both LRS and Cataract Surgery aim = Emmetropia Refractive

More information

Evaluation of Opposite Clear Corneal Incision in Controlling Astigmatism in Cataract Patients Undergoing Phacoemulsification Surgery

Evaluation of Opposite Clear Corneal Incision in Controlling Astigmatism in Cataract Patients Undergoing Phacoemulsification Surgery Evaluation of Opposite Clear Corneal Incision in Controlling Astigmatism in Cataract Patients Undergoing Phacoemulsification Surgery J.S.Bhalla, Meenakshi Rani, Surbhi Gupta Department of Ophthalmology,

More information

Handout Course Title : Astigmatisme Management with toric IOL

Handout Course Title : Astigmatisme Management with toric IOL Handout Course Title : Astigmatisme Management with toric IOL ESCRS Milano 2012 Level :Basic Course leader : Jerome jean Bovet Course duration : 2 hours Faculty : Jerome Bovet, Warren Hill Keiki Mehta

More information

2nd ESASO Anterior Segment Academy April 2016, Milano/Italy

2nd ESASO Anterior Segment Academy April 2016, Milano/Italy 2nd ESASO Anterior 28 30 April 2016, Milano/Italy Istituto Clinico Humanitas Humanitas Congress Centre Via Manzoni, 56 20089 Rozzano, Milan Congress Chairmen: José L. Güell, Paolo Vinciguerra www.esaso.org/2nd-esaso-anteriorsegment-academy-2016/

More information

Standard for Reporting Refractive Outcomes of Intraocular Lens Based Refractive Surgery

Standard for Reporting Refractive Outcomes of Intraocular Lens Based Refractive Surgery EDITORIAL Standard for Reporting Refractive Outcomes of Intraocular Lens Based Refractive Surgery Dan Z. Reinstein, MD, MA(Cantab), FRCSC; Timothy J. Archer, MA(Oxon), DipCompSci(Cantab); Sathish Srinivasan,

More information

Douglas Katsev MD Sansum Clinic Chairman Ophthalmology Santa Barbara CA

Douglas Katsev MD Sansum Clinic Chairman Ophthalmology Santa Barbara CA Early Outcomes (9 months) with a Toric Accommodating IOL How do They Fit in My refractive Practice Douglas Katsev MD Sansum Clinic i Chairman Ophthalmology Santa Barbara CA 1 Disclosure Consulting Fee:

More information

Index. D DALK, 69, 155 Differential sector index (DSI), 92 Discriminant function analysis, DMEK, 23 Donor factors, 156 DSAEK, 23

Index. D DALK, 69, 155 Differential sector index (DSI), 92 Discriminant function analysis, DMEK, 23 Donor factors, 156 DSAEK, 23 A Abberrometry, intraoperative, 66 Aberrated corneas, topography-guided laser for, 146 Aberration coefficient, 99 corneal, 94 Ablation cylindrical, 131 hyperopic, 130, 131 pattern, design and planning,

More information

AXsys Study Data and Press Release Reference

AXsys Study Data and Press Release Reference Clinically Tested to be the World s Most Accurate Toric Marking Device Takayuki Akahoshi, MD Tokyo Japan Ophthalmologist Anterior Segment I WORKED WITH ASICO LLC TO DESIGN AN AXsys TM TORIC MARKING DEVICE

More information

2Optimizing the Refractive

2Optimizing the Refractive Chapter 02 3/4/11 4:29 PM Page 1 2Optimizing the Refractive Outcome: Correction of Astigmatism in Cataract Surgery By: Robert M. Kershner, M.D., FACS General Considerations Lens extraction with the implantation

More information

ORIGINAL ARTICLE. Primary Topography-Guided LASIK: Treating Manifest Refractive Astigmatism Versus Topography-Measured Anterior Corneal Astigmatism

ORIGINAL ARTICLE. Primary Topography-Guided LASIK: Treating Manifest Refractive Astigmatism Versus Topography-Measured Anterior Corneal Astigmatism ORIGINAL ARTICLE Primary Topography-Guided LASIK: Treating Manifest Refractive Astigmatism Versus Topography-Measured Anterior Corneal Astigmatism Avi Wallerstein, MD, FRCSC; Mathieu Gauvin, BEng, PhD;

More information

Research conducted over the past 15 years has yielded a

Research conducted over the past 15 years has yielded a Visual Psychophysics and Physiological Optics Longitudinal Change and Stability of Refractive, Keratometric, and Internal Astigmatism in Childhood Erin M. Harvey, 1,2 Joseph M. Miller, 1 3 J. Daniel Twelker,

More information

New method of quantifying corneal topographic astigmatism that corresponds with manifest refractive cylinder

New method of quantifying corneal topographic astigmatism that corresponds with manifest refractive cylinder ARTICLE New method of quantifying corneal topographic astigmatism that corresponds with manifest refractive cylinder Noel Alpins, FRANZCO, FRCOphth, FACS, James K.Y. Ong, BOptom, Dr.rer.nat, George Stamatelatos,

More information

CHANGE ON THE HORIZONTAL AND VERTICAL MERIDIANS OF THE CORNEA AFTER CATARACT SURGERY*

CHANGE ON THE HORIZONTAL AND VERTICAL MERIDIANS OF THE CORNEA AFTER CATARACT SURGERY* 15 Merriam Final 11/9/01 11:22 AM Page 187 CHANGE ON THE HORIZONTAL AND VERTICAL MERIDIANS OF THE CORNEA AFTER CATARACT SURGERY* BY John C. Merriam, MD, Lei Zheng, MD (BY INVITATION), Joanna Urbanowicz,

More information

Clinical Study Effect of Pupil Size on Optical Quality Parameters in Astigmatic Eyes Using a Double-Pass Instrument

Clinical Study Effect of Pupil Size on Optical Quality Parameters in Astigmatic Eyes Using a Double-Pass Instrument BioMed Research International Volume 2013, Article ID 124327, 6 pages http://dx.doi.org/1155/2013/124327 Clinical Study Effect of Pupil Size on Optical Quality Parameters in Astigmatic Eyes Using a Double-Pass

More information

The Short Term Effects of a Single Limbal Relaxing Incision Combined with Clear Corneal Incision

The Short Term Effects of a Single Limbal Relaxing Incision Combined with Clear Corneal Incision pissn: 1011-8942 eissn: 2092-9382 Korean J Ophthalmol 2010;24(2):78-82 DOI: 10.3341/kjo.2010.24.2.78 Original Article The Short Term Effects of a Single Limbal Relaxing Incision Combined with Clear Corneal

More information

Non-penetrating Femtosecond Laser. intrastromal astigmatic keratotomy (ISAK) Patients With Mixed Astigmatism After Previous Refractive Surgery

Non-penetrating Femtosecond Laser. intrastromal astigmatic keratotomy (ISAK) Patients With Mixed Astigmatism After Previous Refractive Surgery ORIGINAL ARTICLE Non-penetrating Femtosecond Laser Intrastromal Astigmatic Keratotomy in Patients With Mixed Astigmatism After Previous Refractive Surgery Jan Venter, MD; Rodney Blumenfeld, MD; Steve Schallhorn,

More information

DOWNLOAD ASTIGMATIC TECHNIQUE IN ONE STEP RAINBOW HOLOGRAPHY

DOWNLOAD ASTIGMATIC TECHNIQUE IN ONE STEP RAINBOW HOLOGRAPHY ASTIGMATIC TECHNIQUE IN ONE PDF DOWNLOAD 1 / 5 2 / 5 3 / 5 astigmatic technique in one pdf astigmatic technique in one pdf Signs and symptoms. Although astigmatism may be asymptomatic, higher degrees of

More information

Predicting of Uncorrected Astigmatism from Decimal Visual Acuity in Spherical Equivalent

Predicting of Uncorrected Astigmatism from Decimal Visual Acuity in Spherical Equivalent Journal of the Optical Society of Korea Vol. 17, No. 2, April 2013, pp. 219-223 DOI: http://dx.doi.org/10.3807/josk.2013.17.2.219 Predicting of Uncorrected Astigmatism from Decimal Visual Acuity in Spherical

More information

Assessment & management of irregular astigmatism

Assessment & management of irregular astigmatism Assessment & management of irregular astigmatism Athens, March 2018 D. Epstein, MD, PhD, FARVO No financial interest What is an irregular astigmatism? A question that starts at the wrong end... How about

More information

Refractive, anterior corneal and internal astigmatism in the pseudophakic eye

Refractive, anterior corneal and internal astigmatism in the pseudophakic eye Refractive, anterior corneal and internal astigmatism in the pseudophakic eye Jesper F. Bregnhøj, 1,2 Pourang Mataji 1,2 and Kristian Næser 1,2 1 Department of Ophthalmology, Aarhus University Hospital,

More information

Comparison of Toric Foldable Iris-Fixated Phakic Intraocular Lens Implantation and Limbal Relaxing Incisions for Moderate-to-High Myopic Astigmatism

Comparison of Toric Foldable Iris-Fixated Phakic Intraocular Lens Implantation and Limbal Relaxing Incisions for Moderate-to-High Myopic Astigmatism Original Article Yonsei Med J 216 Nov;57(6):1475-1481 pissn: 513-5796 eissn: 1976-2437 Comparison of Toric Foldable Iris-Fixated Phakic Intraocular Lens Implantation and Limbal Relaxing Incisions for Moderate-to-High

More information

The efficacy of Toric IOL in comparison to LRI in correcting pre-existing astigmatism in phacoemulsification

The efficacy of Toric IOL in comparison to LRI in correcting pre-existing astigmatism in phacoemulsification Original Research Article The efficacy of Toric IOL in comparison to LRI in correcting pre-existing astigmatism in phacoemulsification Parul Singh 1, Ruchika Agarwal 2*, Sanjeev Rohatgi 2, Malini Vohra

More information

Arcuate Keratotomy for High Postoperative Keratoplasty Astigmatism Performed With the IntraLase Femtosecond Laser

Arcuate Keratotomy for High Postoperative Keratoplasty Astigmatism Performed With the IntraLase Femtosecond Laser Arcuate Keratotomy for High Postoperative Keratoplasty Astigmatism Performed With the IntraLase Femtosecond Laser Luca Buzzonetti, MD; Gianni Petrocelli, MD; Antonio Laborante, MD; Emilio Mazzilli, MD;

More information

Clinical outcomes of Transepithelial photorefractive keratectomy to treat low to moderate myopic astigmatism

Clinical outcomes of Transepithelial photorefractive keratectomy to treat low to moderate myopic astigmatism Xi et al. BMC Ophthalmology (2018) 18:115 https://doi.org/10.1186/s12886-018-0775-5 RESEARCH ARTICLE Clinical outcomes of Transepithelial photorefractive keratectomy to treat low to moderate myopic astigmatism

More information

Lin Liu, Jun Zou *, Hui Huang, Jian-guo Yang and Shao-rong Chen

Lin Liu, Jun Zou *, Hui Huang, Jian-guo Yang and Shao-rong Chen Liu et al. Diagnostic Pathology 2012, 7:55 RESEARCH Open Access The influence of corneal astigmatism on retinal nerve fiber layer thickness and optic nerve head parameter measurements by spectral-domain

More information

No financial interest

No financial interest Management of high astigmatism after penetrating keratoplasty Ahmed Sherif MD Assistant Professor of Ophthalmology Cairo University No financial interest 1 Incidence Several reports state that 15-31% of

More information

Femtosecond Cataract Surgery: Correction of Astigmatism and Complex Cases Financial Disclosures Femtosecond Laser Utility in Cataract Surgery

Femtosecond Cataract Surgery: Correction of Astigmatism and Complex Cases Financial Disclosures Femtosecond Laser Utility in Cataract Surgery 1 2 3 4 5 6 7 Femtosecond Cataract Surgery: Correction of Astigmatism and Complex Cases Michael J Taravella, MD Director: Cornea and Refractive Surgery University of Colorado Financial Disclosures Consultant

More information

Circular Keratotomy to Reduce Astigmatism and Improve Vision in Stage I and II Keratoconus

Circular Keratotomy to Reduce Astigmatism and Improve Vision in Stage I and II Keratoconus Circular Keratotomy to Reduce Astigmatism and Improve Vision in Stage I and II Keratoconus Jorg H. Krumeich, MD; Guy M. Kezirian, MD, FACS ABSTRACT PURPOSE: To report the use of circular keratotomy in

More information

Phacoemulsification: The first 50 Cases

Phacoemulsification: The first 50 Cases Phacoemulsification: The first 5 Cases Aneeq Ullah Baig Mirza*, Samina Jehangir**, Wasif Mohy-ud-din Kadri** * Department of Ophthalmology Islamic International Medical College and Railway Hospital Rawalpindi.

More information

STUDY OF ASTIGMATISM IN SMALL INCISSION CATARACT SURGERY BETWEEN TEMPORAL AND SUPERIOR INCISSIONS K. J. N. Sivacharan 1, G.

STUDY OF ASTIGMATISM IN SMALL INCISSION CATARACT SURGERY BETWEEN TEMPORAL AND SUPERIOR INCISSIONS K. J. N. Sivacharan 1, G. STUDY OF ASTIGMATISM IN SMALL INCISSION CATARACT SURGERY BETWEEN TEMPORAL AND SUPERIOR INCISSIONS K. J. N. Sivacharan 1, G. Hanumantharao 2 HOW TO CITE THIS ARTICLE: K. J. N. Sivacharan, G. Hanumantharao.

More information

Clinical results of arcuate incisions to correct astigmatism

Clinical results of arcuate incisions to correct astigmatism Clinical results of arcuate incisions to correct astigmatism Kurt A. Buzard, MD, Eduardo Laranjeira, MD, Bradley R. Fundingsland, BS ABSTRACT Purpose: To evaluate the effectiveness of arcuate incisions

More information

Disclosure. Getting Up to Date with LASIK. Modern advancements LASIK. What we re curing. Changing the corneal surface

Disclosure. Getting Up to Date with LASIK. Modern advancements LASIK. What we re curing. Changing the corneal surface Getting Up to Date with LASIK Disclosure I am not a paid consultant to any drug or device company. Stillwater, MN Adjunct Associate Professor University of Minnesota LASIK Use one instrument to make a

More information

Effect of Pupil Size on Uncorrected Visual Acuity in Pseudophakic Eyes With Astigmatism

Effect of Pupil Size on Uncorrected Visual Acuity in Pseudophakic Eyes With Astigmatism ORIGINAL ARTICLE Effect of Pupil Size on Uncorrected Visual Acuity in Pseudophakic Eyes With Astigmatism Kazuhiro Watanabe, MD; Kazuno Negishi, MD; Murat Dogru, MD; Takefumi Yamaguchi, MD; Hidemasa Torii,

More information

THE CHALLENGES CORNEAL IRREGULARITIES POST-LASIK ECTASIA IS THIS A GOOD LASIK CANDIDATE? 3/5/2015. FITTING THE IRREGULAR CORNEA Challenges & Solutions

THE CHALLENGES CORNEAL IRREGULARITIES POST-LASIK ECTASIA IS THIS A GOOD LASIK CANDIDATE? 3/5/2015. FITTING THE IRREGULAR CORNEA Challenges & Solutions DISCLOSURE STATEMENT No disclosure statement. CORNEAL IRREGULARITIES Course Title: Lecturer: FITTING THE IRREGULAR CORNEA Challenges & Solutions Phyllis Rakow, COMT, NCLM, FCLSA(H) Keratoconus Pseudokeratoconus

More information

Handheld Shack Hartmann Wavefront Sensor. Jim Schwiegerling, Ph.D. Department of Ophthalmology and Optical Sciences The University of Arizona

Handheld Shack Hartmann Wavefront Sensor. Jim Schwiegerling, Ph.D. Department of Ophthalmology and Optical Sciences The University of Arizona Handheld Shack Hartmann Wavefront Sensor Jim Schwiegerling, Ph.D. Department of Ophthalmology and Optical Sciences The University of Arizona COLLABORATORS Erin M. Harvey, PhD Velma Dobson, PhD Joseph M.

More information

ORIGINAL ARTICLE. Corneal and Refractive Error Astigmatism in Singaporean Schoolchildren: a Vector-Based Javal s Rule

ORIGINAL ARTICLE. Corneal and Refractive Error Astigmatism in Singaporean Schoolchildren: a Vector-Based Javal s Rule 1040-5488/01/7812-0881/0 VOL. 78, NO. 12, PP. 881 887 OPTOMETRY AND VISION SCIENCE Copyright 2001 American Academy of Optometry ORIGINAL ARTICLE Corneal and Refractive Error Astigmatism in Singaporean

More information

Results of Intraoperative Manual Cyclotorsion Compensation for Myopic Astigmatism in Patients Undergoing Small Incision Lenticule Extraction (SMILE)

Results of Intraoperative Manual Cyclotorsion Compensation for Myopic Astigmatism in Patients Undergoing Small Incision Lenticule Extraction (SMILE) ORIGINAL ARTICLE Results of Intraoperative Manual Cyclotorsion Compensation for Myopic Astigmatism in Patients Undergoing Small Incision Lenticule Extraction (SMILE) Sri Ganesh, MS, DNB; Sheetal Brar,

More information

LASIK for post penetrating keratoplasty astigmatism and myopia

LASIK for post penetrating keratoplasty astigmatism and myopia Br J Ophthalmol 1999;83:113 118 113 The Eye Institute, Chatswood, NSW, Australia S K Webber M A Lawless G L Sutton C M Rogers Correspondence to: Dr Michael Lawless, Level 3, 7 Victoria Avenue, Chatswood,

More information

Orthokeratology (Ortho-K), or corneal refractive therapy, is. Toric Double Tear Reservoir Contact Lens in Orthokeratology for Astigmatism ARTICLE

Orthokeratology (Ortho-K), or corneal refractive therapy, is. Toric Double Tear Reservoir Contact Lens in Orthokeratology for Astigmatism ARTICLE ARTICLE Toric Double Tear Reservoir Contact Lens in Orthokeratology for Astigmatism Jaume Pauné, M.Sc., Genís Cardona, Ph.D., and Lluïsa Quevedo, Ph.D. Objectives: This study aimed at assessing the performance

More information

Perioperative Modulating Factors on Astigmatism in Sutured Cataract Surgery

Perioperative Modulating Factors on Astigmatism in Sutured Cataract Surgery 접수번호 : 2008-114 Korean Journal of Ophthalmology 2009;23:240-248 ISSN : 1011-8942 DOI : 10.3341/kjo.2009.23.4.240 Perioperative Modulating Factors on Astigmatism in Sutured Cataract Surgery Yang Kyeung

More information

OPTOMETRY. An analysis of the astigmatic changes induced by accelerated o rt ho ke ratolog y I ORIGINALPAPER 1

OPTOMETRY. An analysis of the astigmatic changes induced by accelerated o rt ho ke ratolog y I ORIGINALPAPER 1 OPTOMETRY I ORIGINALPAPER 1 An analysis of the astigmatic changes induced by accelerated o rt ho ke ratolog y Clin Exp Optom ; 85: 5: 84-93 John Mountford* DipAppSc FAAO FVCO FCLS Konrad Pesudovst PhD

More information

Irregular Astigmatism Diagnosis And Treatment

Irregular Astigmatism Diagnosis And Treatment Irregular Astigmatism Diagnosis And Treatment 1 / 5 2 / 5 3 / 5 Irregular Astigmatism Diagnosis And Treatment Irregular Astigmatism: Diagnosis and Treatment. Ming Wang, ed., Thorofare, NJ: Slack Inc.;

More information

A R Sebai Sarhan, Harminder S Dua, Michelle Beach

A R Sebai Sarhan, Harminder S Dua, Michelle Beach Br J Ophthalmol 2000;84:837 841 837 Division of Ophthalmology and Visual Sciences, University of Nottingham, University Hospital, Queen s Medical Centre, Nottingham NG7 2UH A R S Sarhan H S Dua M Beach

More information

A novel method for human Astigmatism formulation and measurement

A novel method for human Astigmatism formulation and measurement Available online at http://www.ijabbr.com International journal of Advanced Biological and Biomedical Research Volume 1, Issue 8, 2013: 874-884 A novel method for human Astigmatism formulation and measurement

More information

AstigmatismamongotherRefractiveErrorsinChildrenofSouthernSriLanka. Astigmatism among other Refractive Errors in Children of Southern Sri Lanka

AstigmatismamongotherRefractiveErrorsinChildrenofSouthernSriLanka. Astigmatism among other Refractive Errors in Children of Southern Sri Lanka : F Diseases Volume 15 Issue 1 Version 1.0 Year 2015 Type: Double Blind Peer Reviewed International Research Journal Publisher: Global Journals Inc. (USA) Online ISSN: 2249-4618 & Print ISSN: 0975-5888

More information

We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors

We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists 3,700 108,500 1.7 M Open access books available International authors and editors Downloads Our

More information

Analysis of WFS Measurements from first half of 2004

Analysis of WFS Measurements from first half of 2004 Analysis of WFS Measurements from first half of 24 (Report4) Graham Cox August 19, 24 1 Abstract Described in this report is the results of wavefront sensor measurements taken during the first seven months

More information

We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors

We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists. International authors and editors We are IntechOpen, the world s leading publisher of Open Access books Built by scientists, for scientists 3,500 108,000 1.7 M Open access books available International authors and editors Downloads Our

More information

Reconstruction of Ca 2+ dynamics from low frame rate Ca 2+ imaging data CS229 final project. Submitted by: Limor Bursztyn

Reconstruction of Ca 2+ dynamics from low frame rate Ca 2+ imaging data CS229 final project. Submitted by: Limor Bursztyn Reconstruction of Ca 2+ dynamics from low frame rate Ca 2+ imaging data CS229 final project. Submitted by: Limor Bursztyn Introduction Active neurons communicate by action potential firing (spikes), accompanied

More information

*Please note that although this product has been approved in Japan, its launch in other countries has not yet been confirmed.

*Please note that although this product has been approved in Japan, its launch in other countries has not yet been confirmed. make News & Information 1-7-1 Konan, Minato-ku, Tokyo 108-0075, Japan Sony Corporation No.13-085E July 23, 2013 Sony Introduces head-mount image processing unit for endoscopic image display - Images from

More information

1. Introduction. Correspondence should be addressed to Edmund Arthur; arthur

1. Introduction. Correspondence should be addressed to Edmund Arthur; arthur Hindawi Publishing Corporation Journal of Ophthalmology Volume, Article ID 989, 7 pages http://dx.doi.org/.//989 Clinical Study Postoperative Corneal and Surgically Induced Astigmatism following Superior

More information

620 Rejwrts Investigative Ophthalmology

620 Rejwrts Investigative Ophthalmology Rejwrts Investigative Ophthalmology August D. E.: Retinal dystrophy in the rat a pigment epithelial disease, INVEST. OPHTHALMOL. :,. Color vision: blue deficiencies in? ANTHONY J. ADAMS,* RICHARD BAL-

More information

Full-Thickness Astigmatic Keratotomy Combined With Small-Incision Lenticule Extraction to Treat High-Level and Mixed Astigmatism

Full-Thickness Astigmatic Keratotomy Combined With Small-Incision Lenticule Extraction to Treat High-Level and Mixed Astigmatism CLINICAL SCIENCE Full-Thickness Astigmatic Keratotomy Combined With Small-Incision Lenticule Extraction to Treat High-Level and Mixed Astigmatism Bu Ki Kim, MD, MS,* Su Joung Mun, MD, PhD,* Dae Gyu Lee,

More information

Cycloplegic Refractions of Infants and Young Children: The Axis of Astigmatism

Cycloplegic Refractions of Infants and Young Children: The Axis of Astigmatism Cycloplegic Refractions of Infants and Young Children: The Axis of Astigmatism Velma Dobson,* Anne B. Fulton, f and S. Lawson Sebris* Review of the cycloplegic refractions of all children who were first

More information

Astigmatism in Children: Changes in Axis and Amount from Birth to Six Years

Astigmatism in Children: Changes in Axis and Amount from Birth to Six Years Astigmatism in Children: Changes in Axis and Amount from Birth to Six Years Jane Gwiazda, Mitchell Scheiman,* Indra Mohindra, and Richard Held Noncycloplegic refractions of, children aged - years revealed

More information

balt5/zov-opx/zov-opx/zov01005/zov a washingd S 12 10/4/05 14:54 Art: OPX Input-nlm ORIGINAL ARTICLE

balt5/zov-opx/zov-opx/zov01005/zov a washingd S 12 10/4/05 14:54 Art: OPX Input-nlm ORIGINAL ARTICLE 1040-5488/05/8210-0001/0 VOL. 82, NO. 10, PP. 1 1 OPTOMETRY AND VISION SCIENCE Copyright 2005 American Academy of Optometry ORIGINAL ARTICLE Progressive Powered Lenses: the Minkwitz Theorem JAMES E. SHEEDY,

More information

Astigmatism: Aberration or ametropia?

Astigmatism: Aberration or ametropia? http://eoftalmo.org.br OPINION OF SPECIALISTS Astigmatism: Aberration or ametropia? Astigmatismo: Aberração ou Ametropia? Astigmatismo: Aberración o ametropía? Sidney Julio Faria e Sousa - Faculdade de

More information

MC-500 Vixi MC-500. Multicolor Scan Laser Photocoagulator. Multicolor Laser Photocoagulator US EDITION

MC-500 Vixi MC-500. Multicolor Scan Laser Photocoagulator. Multicolor Laser Photocoagulator US EDITION Multicolor Scan Laser Photocoagulator MC-500 Vixi MC-500 Multicolor Laser Photocoagulator US EDITION MC-500 / MC-500 The Versatile Laser Photocoagulator The versatility of the MC-500 Vixi / MC-500 speaks

More information

Astigmatic axis and amblyopia in childhood

Astigmatic axis and amblyopia in childhood Astigmatic axis and amblyopia in childhood Maths Abrahamsson and Johan Sjo strand ABSTRACT. Purpose: This study is part of a larger project whose aim is to evaluate the relationship between refractive

More information

Multicolor Scan Laser Photocoagulator MC-500 Vixi. Multicolor Laser PhotocoagulatorMC-500

Multicolor Scan Laser Photocoagulator MC-500 Vixi. Multicolor Laser PhotocoagulatorMC-500 Multicolor Scan Laser Photocoagulator MC-500 Vixi Multicolor Laser PhotocoagulatorMC-500 MC-500 / MC-500 The Versatile Laser Photocoagulator The versatility of the MC-500 Vixi / MC-500 speaks for itself.

More information

Multicolor Scan Laser Photocoagulator MC-500 Vixi

Multicolor Scan Laser Photocoagulator MC-500 Vixi Multicolor Scan Laser Photocoagulator MC-500 Vixi MC-500 The Versatile Laser Photocoagulator Selectable configuration of laser colors and delivery units Multiple scan patterns Enhanced usability LPM (Low

More information

OPTOMETRY INVITED REVIEW. A review of astigmatism and its possible genesis

OPTOMETRY INVITED REVIEW. A review of astigmatism and its possible genesis C L I N I C A L A N D E X P E R I M E N T A L OPTOMETRY INVITED REVIEW A review of astigmatism and its possible genesis Clin Exp Optom 2007; 90: 1: 5 19 Scott A Read PhD Michael J Collins PhD Leo G Carney

More information

The eyes of neonates of all species studied have rather. Severe Astigmatic Blur Does Not Interfere with Spectacle Lens Compensation

The eyes of neonates of all species studied have rather. Severe Astigmatic Blur Does Not Interfere with Spectacle Lens Compensation A R T I C L E S Severe Astigmatic Blur Does Not Interfere with Spectacle Lens Compensation Rhondalyn C. McLean and Josh Wallman PURPOSE. Whether either natural emmetropization or compensation for imposed

More information

Dr Noel Alpins AM Digest of Personal and Professional biography

Dr Noel Alpins AM Digest of Personal and Professional biography Dr Noel Alpins AM Digest of Personal and Professional biography Work Address: 7 Chesterville Road Cheltenham 3192 Dr Noel Alpins AM has been specialising in Cataract and Refractive Surgery since founding

More information

Proposed classification for topographic patterns seen after penetrating keratoplasty

Proposed classification for topographic patterns seen after penetrating keratoplasty Br J Ophthalmol 1999;83:403 409 403 Department of Ophthalmology, Bristol Eye Hospital, Bristol C H Karabatsas S D Cook J M Sparrow Correspondence to: Costas H Karabatsas, PO Box 16757, Athens 115 02, Greece.

More information

Development of OLED Lighting Panel with World-class Practical Performance

Development of OLED Lighting Panel with World-class Practical Performance 72 Development of OLED Lighting Panel with World-class Practical Performance TAKAMURA MAKOTO *1 TANAKA JUNICHI *2 MORIMOTO MITSURU *2 MORI KOICHI *3 HORI KEIICHI *4 MUSHA MASANORI *5 Using its proprietary

More information

Ahigh prevalence of astigmatism has been documented

Ahigh prevalence of astigmatism has been documented Clinical and Epidemiologic Research Prevalence of Corneal Astigmatism in Tohono O odham Native American Children 6 Months to 8 Years of Age Erin M. Harvey, 1,2 Velma Dobson, 1,3 Joseph M. Miller, 1,2,4

More information

Diagnosis and Management of Astigmatism

Diagnosis and Management of Astigmatism Diagnosis and Management of Astigmatism Ray George Diagnosis and Management of Astigmatism "This page is Intentionally Left Blank" Diagnosis and Management of Astigmatism Edited by Ray George Published

More information

White Paper. Uniform Luminance Technology. What s inside? What is non-uniformity and noise in LCDs? Why is it a problem? How is it solved?

White Paper. Uniform Luminance Technology. What s inside? What is non-uniformity and noise in LCDs? Why is it a problem? How is it solved? White Paper Uniform Luminance Technology What s inside? What is non-uniformity and noise in LCDs? Why is it a problem? How is it solved? Tom Kimpe Manager Technology & Innovation Group Barco Medical Imaging

More information

BIBLIOMETRIC REPORT. Bibliometric analysis of Mälardalen University. Final Report - updated. April 28 th, 2014

BIBLIOMETRIC REPORT. Bibliometric analysis of Mälardalen University. Final Report - updated. April 28 th, 2014 BIBLIOMETRIC REPORT Bibliometric analysis of Mälardalen University Final Report - updated April 28 th, 2014 Bibliometric analysis of Mälardalen University Report for Mälardalen University Per Nyström PhD,

More information

Characterisation of the far field pattern for plastic optical fibres

Characterisation of the far field pattern for plastic optical fibres Characterisation of the far field pattern for plastic optical fibres M. A. Losada, J. Mateo, D. Espinosa, I. Garcés, J. Zubia* University of Zaragoza, Zaragoza (Spain) *University of Basque Country, Bilbao

More information

CA-800 Tear Module. Rick Gaudenti. Product Manager, Refraction Mar 2017

CA-800 Tear Module. Rick Gaudenti. Product Manager, Refraction Mar 2017 CA-800 Tear Module Rick Gaudenti Product Manager, Refraction Mar 2017 Meibomian Gland Imaging Meibomian Gland [MEIB] Viewing Meibomian Images are displayed in the center of the screen Two types of an adjustable

More information

Reducing tilt errors in moiré linear encoders using phase-modulated grating

Reducing tilt errors in moiré linear encoders using phase-modulated grating REVIEW OF SCIENTIFIC INSTRUMENTS VOLUME 71, NUMBER 6 JUNE 2000 Reducing tilt errors in moiré linear encoders using phase-modulated grating Ju-Ho Song Multimedia Division, LG Electronics, #379, Kasoo-dong,

More information

This work was supported by FINEP (Research and Projects Financing) under contract

This work was supported by FINEP (Research and Projects Financing) under contract MODELING OF A GRIDDED ELECTRON GUN FOR TRAVELING WAVE TUBES C. C. Xavier and C. C. Motta Nuclear & Energetic Research Institute, São Paulo, SP, Brazil University of São Paulo, São Paulo, SP, Brazil Abstract

More information

Femtosecond laser-assisted astigmatic keratotomy: a review

Femtosecond laser-assisted astigmatic keratotomy: a review Chang Eye and Vision (2018) 5:6 https://doi.org/10.1186/s40662-018-0099-9 REVIEW Open Access Femtosecond laser-assisted astigmatic keratotomy: a review John S. M. Chang Abstract Background: Astigmatic

More information

ORIGINAL ARTICLE. Amblyopia in Astigmatic Infants and Toddlers

ORIGINAL ARTICLE. Amblyopia in Astigmatic Infants and Toddlers 1040-5488/10/8705-0330/0 VOL. 87, NO. 5, PP. 330 336 OPTOMETRY AND VISION SCIENCE Copyright 2010 American Academy of Optometry ORIGINAL ARTICLE Amblyopia in Astigmatic Infants and Toddlers Velma Dobson*,

More information

Eyes with regular astigmatism have two orthogonal focal. Accommodation in Astigmatic Children During Visual Task Performance

Eyes with regular astigmatism have two orthogonal focal. Accommodation in Astigmatic Children During Visual Task Performance Clinical and Epidemiologic Research Accommodation in Astigmatic Children During Visual Task Performance Erin M. Harvey, 1,2 Joseph M. Miller, 1 3 Howard P. Apple, 1 Pavan Parashar, 4 J. Daniel Twelker,

More information

1. Standard Equipment Subjective Eye Tester Name of Parts Details of Auxiliary Lenses Measuring Performance...

1. Standard Equipment Subjective Eye Tester Name of Parts Details of Auxiliary Lenses Measuring Performance... Notification Dear Users, Thank you for your purchase of R 2500 Refractor. Please take time to read our user s manual carefully before use. This guarantees you to make full use of this unit and prolongs

More information

Treatment of astigmatism-related amblyopia in 3- to 5-year-old children

Treatment of astigmatism-related amblyopia in 3- to 5-year-old children Vision Research 44 (2004) 1623 1634 www.elsevier.com/locate/visres Treatment of astigmatism-related amblyopia in 3- to 5-year-old children Erin M. Harvey a, *, Velma Dobson a,b, Joseph M. Miller a,c,d,

More information

Astigmatism is a common refractive error 1 and an important. The Changing Profile of Astigmatism in Childhood: The NICER Study

Astigmatism is a common refractive error 1 and an important. The Changing Profile of Astigmatism in Childhood: The NICER Study Clinical and Epidemiologic Research The Changing Profile of Astigmatism in Childhood: The NICER Study Lisa O Donoghue, Karen M. Breslin, and Kathryn J. Saunders School of Biomedical Sciences, University

More information

Chapter 10 Basic Video Compression Techniques

Chapter 10 Basic Video Compression Techniques Chapter 10 Basic Video Compression Techniques 10.1 Introduction to Video compression 10.2 Video Compression with Motion Compensation 10.3 Video compression standard H.261 10.4 Video compression standard

More information

NIH Public Access Author Manuscript Optom Vis Sci. Author manuscript; available in PMC 2011 May 1.

NIH Public Access Author Manuscript Optom Vis Sci. Author manuscript; available in PMC 2011 May 1. NIH Public Access Author Manuscript Published in final edited form as: Optom Vis Sci. 2010 May ; 87(5): 330 336. doi:10.1097/opx.0b013e3181d951c8. Amblyopia in Astigmatic Infants and Toddlers Velma Dobson,

More information

Fast MBAFF/PAFF Motion Estimation and Mode Decision Scheme for H.264

Fast MBAFF/PAFF Motion Estimation and Mode Decision Scheme for H.264 Fast MBAFF/PAFF Motion Estimation and Mode Decision Scheme for H.264 Ju-Heon Seo, Sang-Mi Kim, Jong-Ki Han, Nonmember Abstract-- In the H.264, MBAFF (Macroblock adaptive frame/field) and PAFF (Picture

More information