Introduction to System Suitability Tests Michael Wright Bioanalytical Services LGC Fordham, UK
|
|
- Joy Leonard
- 5 years ago
- Views:
Transcription
1 Introduction to System Suitability Tests Michael Wright Bioanalytical Services LGC Fordham, UK Science for a safer world
2 13/10/
3 What is a System Suitability Test (SST)? Also referred to as a System Suitability Sample (SSS) A standard or extracted sample (*in extraction solvent) Usually contains internal standard Stored in a stable manner if possible Sample ID Rack Code Vial Position Reagent Blank DWP96 1 Reagent Blank DWP96 2 SST DWP96 3 Reagent Blank DWP96 4 Injected onto your LC-MS before you submit your sample batch 13/10/2016 3
4 What does it tell us? 1) Retention Time 2) Instrument Response (or Signal:Noise) 3) Peak Shape When combined with a back pressure reading it provides us with a wealth of knowledge.
5 What type? Generic OR Specific 13/10/2016 5
6 What do the SST and Back Pressure tell us? 1) The HPLC is OK Mobile Phases Column Flow rate Column Oven Auto-sampler Injection Seals Valves (Carry-over)
7 What do the SST and Back Pressure tell us? 2) The Mass Spec is OK Ionisation Gas Flow Source Ion transmission Detection
8 What doesn t the SST and Back Pressure tell us? Sample extraction issues Contaminants from the extraction solvents (Reagent Blank/Control Blank)
9 Setting Up the Acceptance Criteria 1) Determine SST concentration (LLOQ or match to LLOQ?) 2) Run before every validation batch to obtain a feel for normal Instrument API3200 Back Pressure at t0 Retention Time Intensity S:N P A/S Maintenance Notes DATE Pump A/B Pump C/D 25OHD3 25OHD2 25OHD3 25OHD3 27/03/ /03/ LLOQ looks poor, QC Low out 28/03/ Fresh Mobile Phases 01/04/ Weekly & Monthly 02/04/ /04/ Weekly 11/04/ Leak in column oven 11/04/ /04/ ) From these experiments set the SST acceptance limits 4) You will need to do this for each type of LC-MS instrument 13/10/2016 9
10 Retention Time: Minimum Intensity 1200cps Minimum Signal:Noise 20:1 Peak Asymmetry (A S ) h A C Time A s = CB/AC B } 10 % of peak height (h) Plate Count (N) t N 5.54 W R 0.5 2
11 Different machines 25OH Vitamin D SST XIC of +MRM (4 pairs): / Da ID: 25OHD3 1 from Sample 137 (2014_08-04 post weekly) of SST.wiff (Heated Nebulizer) Max cps OHD3 RT: 4.64min Older instrument In te n s ity, c p s Lower Limit 2400cps 3-epi- 25OHD Time, min XIC of +MRM (4 pairs): / Da ID: 25OHD3 quant from Sample 28 (2014_08_04) of SST.wiff (Heated Nebulizer) Max. 7.3e4 cps. Newer instrument In te n s ity, c p s 7.3e4 7.0e4 6.5e4 6.0e4 5.5e4 5.0e4 4.5e4 4.0e4 3.5e4 3.0e4 2.5e4 2.0e4 1.5e4 1.0e Lower Limit 2.0e4 cps 25OHD3 RT: 4.75min Time, min epi- 25OHD
12 How much RT shift/ BP change/ Peak symmetry change is still OK? No hard and fast rule batch failures during establishment/validation as a guide If the method is already live or there weren t (m)any failures during method development then. Guess! Try not to make a rod for your own back (not too tight unless you have cause & don t set unnecessarily low LLOQs) Initial criteria are opinions that you then review over time Look at every batch failure as an opportunity to review these criteria to make it a better tool 13/10/
13 Retention Time Ensure that the criteria are set so that the peak doesn t shift into a drastic gradient change Monitor IS in batches with altered RT to look for ionsuppression issues Back Pressure Limits for fittings Check traces to see the % change over the gradient (extrapolate for changes at t0*) Peak Shape (A S and N) Known interferences? Integration Assists in judging column lifetimes 13/10/
14 Example: adding in known interferences XIC of +MRM (4 pairs): / Da ID: 25OHD3 1 from Sample 221 ( TN) of SST.wiff (Heated Nebulizer), Smoothe... Max cps OH OH OHD 3 HO In te n s ity, c p s HO 3epi -25OHD Time, min
15 Example something a little different: Unknown interferences 1 Normetanephrine 2 3 Metanephrine
16 Something has gone wrong.
17 I can t see anything Step 1 start simple Correct method being run? Is there anything in the vial? Is the vial in the correct position? Did the auto-sampler pick up the sample? Needle depth? Wash solvents full? Is the LC connected to the MS? Is the back pressure OK? YES Spray in the source? Correct probe? (ESI vs APCI) corona needle? Divert valve correct? Check ion path Signal Minimum Intensity Retention Time Time NO Correct mobile phases? Lines not purged properly Too High Blockage in lines somewhere? Column too old? Too Low Leak somewhere? Fluctuating Leak on single pumphead Pump seals
18 Signal Signal What is it doing over there? Retention Time Time Retention Time Time No point looking at the Mass Spec Is the back pressure OK? Yes Mobile phase made correctly? Gradual change? (column degradation) Correct wash solvents loaded (peak shape OK?) Injection mode correct? No Mobile phase made correctly? All lines purged correctly? Leak somewhere (early and late peaks) Faulty pump (swap A&B lines and see if RT changes Gradual change? (column degradation) 13/10/
19 Right time - but it is a bit small Signal Minimum Intensity Can t rule out the HPLC or MS Back pressure will often be OK but check anyway Time SST sample mixed properly? Mobile phase made correctly? Auto-sampler working properly?.. blockage/ leak/ wash solvent/ injection mode (*can check with manual injector) Leak somewhere after the column? Correct ion probe? Capillary extended in probe? APCI Corona needle positioned correctly? Mass Spec within calibration? Gradual change across all methods? (MS needs cleaning) Sudden drop MS source/ion guide/q1 may need cleaning Vacuum OK? Gas pressures OK? 13/10/
20 Where did all that come from? Signal Minimum Intensity Time SST sample mixed properly? Mobile phase made correctly? Right amount injected? Mass Spec set to correct resolution? Correct ion probe? Was MS incorrectly calibrated before? Recent maintenance performed? (Source/Gas/Vacuum) Is it really a problem? Can you just inject less? 13/10/
21 It looks ugly SST sample solvent correct? Auto-sampler wash solvents OK? Mobile Phases OK? Pumps OK and delivering gradient? Leak somewhere? Peak tubing cut squarely and fitted correctly? Column degradation? 13/10/
22 It s EVERYWHERE! Signal Minimum Intensity Reagent Blank Sample ID Rack Code Vial Position Reagent Blank DWP96 1 Reagent Blank DWP96 2 Signal Reagent Blank SST DWP96 3 Reagent Blank DWP96 4 Signal SST Ghost Peaks are often a sign that your mobile phase or auto-sampler wash contains a contaminant Signal Reagent Blank Time 13/10/
23 It wont go away Carry-over peaks point the finger at: Auto-sampler wash not working Internal damage to injection port Signal Signal Minimum Intensity Reagent Blank Reagent Blank Issue with peak not being fully eluted from SST Peak tubing not being cut squarely and fitted correctly Signal SST Signal Reagent Blank Time 13/10/
24 Questions? Science for a safer world
Liquid Chromatography- Mass Spectrometer Manual
Liquid Chromatography- Mass Spectrometer Manual Joshua Willis, Elizabeth Sattely Department of Chemical Engineering Stanford University November 6, 2014 Abstract This manual will explain the LC/MS, its
More informationHPLC Tips and Tricks. Tiziana Ladisa Sales Support Specialist for Chromatography Italy Thermo Fisher Scientific, Rodano (MI)
HPLC Tips and Tricks Tiziana Ladisa Sales Support Specialist for Chromatography Italy Thermo Fisher Scientific, Rodano (MI) The world leader in serving science Overview Thermo Scientific UltiMate 3000
More informationSTANDARD OPERATING PROCEDURE:
Q-EXACTIVE STANDARD OPERATING PROCEDURE: Purpose of this Instrument: Essential tool for molecular weight identification for small molecules, peptides, proteins, ect. Location: WVU Chemistry Research Laboratory
More informationStaying Connected in GC
Staying Connected in GC Daron Decker / Eric Pavlich Chromatography Technical Specialists Page 1 Leaks (Oxygen) Damages GC Columns Repair of the GC column is impossible Replacements means re-calibration
More informationAgilent G1978B Multimode Source for 6410 Triple Quad LC/MS
Agilent G1978B Multimode Source for 6410 Triple Quad LC/MS User Guide Agilent Technologies Notices Agilent Technologies, Inc. 2006-2007 No part of this manual may be reproduced in any form or by any means
More informationAIM INTRODUCTION SIMPLIFIED WORKFLOW
CD-R PATENT PROTECTION: BRAND CHARACTERIZATION TO IDENTIFY COUNTERFEIT GOODS USING SIMPLIFIED MASS SPECTROMETRY James Morphet and Eleanor Riches Waters Corporation, Manchester, UK AIM To provide an easy-to-use
More informationAgilent G1978B Multimode Source for 6410 Triple Quad LC/MS
Agilent G1978B Multimode Source for 6410 Triple Quad LC/MS User Guide Agilent Technologies Notices Agilent Technologies, Inc. 2006 No part of this manual may be reproduced in any form or by any means (including
More informationFRACTION COLLECTOR. Microcomputer Controlled CHF122SC
FRACTION COLLECTOR Microcomputer Controlled CHF122SC From open column to HPLC liquid chromatography and general liquid samplings, the Super Fraction Collector, model CHF122SC has a wide range of uses.
More informationSpeed Performance Reliability. Medicinal Chemistry Natural Products Peptides & Polymers Organic Synthesis Purifications
Automated Flash Chromatography Systems Medicinal Chemistry Natural Products Peptides & Polymers Organic Synthesis Purifications Speed Performance Reliability CombiFlash Rf - Making Fl Improve Your Productivity
More informationCustomer Responsibilities. Important Customer Information Infinity LC/1260 Infinity LC Site Preparation Checklist
1290 Site Preparation Infinity LC/1260 Checklist Infinity LC Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this
More informationCreating and fully utilizing the retention time standard processing method for a more automated IC workflow
TECHNICAL NOTE Creating and fully utilizing the retention time standard processing method for a more automated IC workflow No. 178 Jay Lorch, Thermo Fisher Scientific, Sunnyvale, CA Keywords Retention
More informationPerformance. Reliability. Productivity. Automated Flash Chromatography Systems
Performance Reliability Productivity Automated Flash Chromatography Systems CombiFlash Rf+ Family of Purification Systems Flash chromatography is the science of refinement Teledyne Isco continues to refine
More informationEXA PH200/400 and EXA PH202/402 Troubleshooting and Error Code Guide
EXA PH200/400 and EXA PH202/402 Troubleshooting and Error Code Guide Introduction The EXA Series of Instruments (EXA PH200, PH400, PH202, PH402) provide much more than just a measurement. They are also
More informationCustomer Responsibilities. Important Customer Information. Agilent InfinityLab LC Series Site Preparation Checklist
Agilent Site Preparation InfinityLab Checklist LC Series Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification
More informationBitWise (V2.1 and later) includes features for determining AP240 settings and measuring the Single Ion Area.
BitWise. Instructions for New Features in ToF-AMS DAQ V2.1 Prepared by Joel Kimmel University of Colorado at Boulder & Aerodyne Research Inc. Last Revised 15-Jun-07 BitWise (V2.1 and later) includes features
More informationWaters GCT GCMS Training Manual (Nominal Mass GCMS version) 02/06/2012 S.V.
1234 Hach Hall 515-294-5805 www.cif.iastate.edu Waters GCT GCMS Training Manual (Nominal Mass GCMS version) 02/06/2012 S.V. Location: Contact: 1238 Hach Hall Steve Veysey, 1234 Hach Hall; Kamel Harrata,
More informationWaters Synapt High Definition Mass Spectrometry System Quick Start Guide
Waters Synapt High Definition Mass Spectrometry System Quick Start Guide 71500129903/Revision B Copyright Waters Corporation 2007 All rights reserved Copyright notice 2007 WATERS CORPORATION. PRINTED IN
More informationHow Quadrupole Size Affects Research
Quadrupole Size Comparison APPLICATION NOTE NOTE How Quadrupole Size Affects Research Mass Spectrometry is one of the most widely used analytical techniques and the most common mass analyzer is the quadrupole.
More informationOperating pressure should be 6000 psi or better suitable for both analytical and semi-preparative applications
S. No. NIPER-G/PUR/GLOBAL/2018/01 Technical Specification for Analytical HPLC with PDA & Fluorescence detector Fully automated analytical high performance liquid chromatography system with single point
More informationOverview and Interpretation of D7900/D7169 Merge Analysis
Overview and Interpretation of D7900/D7169 Merge Analysis Crude Oil Quality Association New Orleans, LA March 14, 2019 Value of Merged Simdis Analysis Requires very little sample (10-50 mls) Much faster
More informationGC/LC-MS: data acquisition rate and peak reconstruction
GC/LC-MS: data acquisition rate and peak reconstruction Nyquist (Shannon-Kotelnikov-Whittaker) theorem Signal sampling does not involve any loss of information as long as the sampling frequency is at least
More informationAchieving More Efficient Data Review with OpenLAB CDS
Achieving More Efficient Data Review with OpenLAB CDS Technical Note Introduction Separations have enjoyed major advances that have significantly reduced chromatographic run times. Ultra high performance
More informationOnline Process Control using Modular Fluid Delivery and Fast Process Gas Chromatography: From the Sample Point to the DCS Connection
Falcon Analytical makers of the... Online Process Control using Modular Fluid Delivery and Fast Process Gas Chromatography: From the Sample Point to the DCS Connection John Crandall, President Ned Roques,
More informationCross Examination of the Criminalist. Felipe Plascencia
Cross Examination of the Criminalist Prepared by: Felipe Plascencia (Some of this material was copied from the top California DUI lawyer, Don Bartell.) I normally do not write out questions when I do cross-examination
More informationSample Analysis Design. Element2 - Basic Software Concepts (cont d)
Sample Analysis Design Element2 - Basic Software Concepts (cont d) Samples per Peak In order to establish a minimum level of precision, the ion signal (peak) must be measured several times during the scan
More informationTMT data acquisition on the LTQ-Orbitrap XL Mass Spectrometer
TECH TIP # 70 TMT data acquisition on the LTQ-Orbitrap XL Mass Spectrometer Introduction TR0070.0 Successful acquisition of quantitative tandem mass tag (TMT*) data with the Thermo Scientific LTQ-Orbitrap
More informationUNITY 2 TM. Version 1.0. February 2008
UNITY 2 TM ULTRA 50:50 Operators Manual Version 1.0 February 2008 1. Loading tubes onto ULTRA 50:50...3 1.1. DiffLok analytical end caps for sample tubes...3 1.2. Capping tubes with DiffLok caps...3 1.3.
More informationNGC8206 Start-Up Guide
NGC8206 Start-Up Guide MEASUREMENT & CONTROL SYSTEMS Intellectual Property & Copyright Notice 2010 by ABB Inc., Totalflow Products ( Owner ), Bartlesville, Oklahoma 74006, U.S.A. All rights reserved. Any
More informationScanning Electron Microscopy (FEI Versa 3D Dual Beam)
Scanning Electron Microscopy (FEI Versa 3D Dual Beam) This operating procedure intends to provide guidance for basic measurements on a standard sample with FEI Versa 3D SEM. For more advanced techniques
More informationNMR. picospin. Maintenance Guide
NMR picospin Maintenance Guide 269-302600 Revision A January 2013 2013 Thermo Fisher Scientific Inc. All rights reserved. For U.S. Technical Support, please contact: Thermo Fisher Scientific 5225 Verona
More informationModel CB 60 CO 2 incubators with hot air sterilization and heat sterilizable CO 2 sensor
Model CB 60 CO 2 incubators with hot air sterilization and heat sterilizable CO 2 sensor The BINDER CB series CO 2 incubator is the premium class among the CO 2 incubators. It is suitable for all sensitive
More informationThis document is a reference document that shows the menus in the 5500sc, 9610sc and 9611sc analyzers. There are 3 top-level menus:
Controller menus 5500sc, 9610sc and 9611sc analyzers DOC273.53.80566 Introduction This document is a reference document that shows the menus in the 5500sc, 9610sc and 9611sc analyzers. There are 3 top-level
More informationPAL LSI - Robust and Reliable Sample Handling
PAL LSI - Robust and Reliable Sample Handling The PAL System, ingeniously reliable, flexible and productive PAL LSI: Innovative features giving real benefits Easy maintenance and adaptation: The different
More informationDesign of NIR Chromenylium-Cyanine Fluorophore Library for Switch-ON and Ratiometric Detection of Bio-Active Species in Vivo
Supporting information for Design of NIR Chromenylium-Cyanine Fluorophore Library for Switch-ON and Ratiometric Detection of Bio-Active Species in Vivo Yanfen Wei, Dan Cheng, Tianbing Ren, Yinhui Li, Zebing
More informationTL-2900 AMMONIA & NITRATE ANALYZER DUAL CHANNEL
TL-2900 AMMONIA & NITRATE ANALYZER DUAL CHANNEL DATA ACQUISITION SYSTEM V.15.4 INSTRUCTION MANUAL Timberline Instruments, LLC 1880 S. Flatiron Ct., Unit I Boulder, Colorado 80301 Ph: (303) 440-8779 Fx:
More informationReal-time ph and Conductivity Data Acquisition with the ACQUITY UPLC H-Class Bio System, Auto Blend Plus, and the GE Healthcare Monitor ph/c-900
Real-time ph and Conductivity Data Acquisition with the ACQUITY UPLC H-Class Bio System, Auto Blend Plus, and the GE Healthcare Monitor ph/c-900 INTRODUCTION The ACQUITY UPLC H-Class Bio System with Auto
More informationEcoline S series. Precision solutions for quality production. Innovations for a better world.
Ecoline S series. Precision solutions for quality production. Innovations for a better world. Precision solutions for quality production. Die casting with confidence. The Ecoline S is a series of cold-chamber
More informationVasudevan Agilent 1100 Series HPLC w/ DAD & FLD Detector (nonbuffer
Vasudevan Agilent 1100 Series HPLC w/ DAD & FLD Detector (nonbuffer solvents) Updated November 14, 2017 Instrument instructions can be found at: http://academic.bowdoin.edu/chemistry/resources/instructions.shtml
More informationBruce Quimby. Agilent Technologies October 28, For Forensic Use. Page 1
Bruce Quimby Application Scientist Agilent Technologies October 28, 2009 Page 1 Tools For Improving GC/MS Analysis Retention Time Locking (RTL) Method Translation Synchronous SIM/Scan Column effluent split
More information2.1. Log on to the TUMI system (you cannot proceed further until this is done).
FEI DB235 ex-situ lift out TEM sample preparation procedure Nicholas G Rudawski ngr@ufledu (805) 252-4916 Last updated: 06/19/15 DISCLAIMER: this procedure describes one specific method for preparing ex-situ
More informationLinac3 experience for LHC ion runs
Linac3 experience for LHC ion runs G Bellodi for the Linac3 team Keywords: beam performance reliability set up time results of MDs remaining unknowns 1 A year in perspective Source removed: change of main
More informationC200H-AD002/DA002 Analog I/O Units Operation Guide
C200H-AD002/DA002 Analog I/O Units Operation Guide Revised September 1995 Notice: OMRON products are manufactured for use according to proper procedures by a qualified operator and only for the purposes
More informationChromQuest 5.0 Chromatography Data System
ChromQuest 5.0 Chromatography Data System User Guide for the SpectraSYSTEM LC CHROM-97256 Revision A March 2008 2008 Thermo Fisher Scientific Inc. All rights reserved. ChromQuest is a trademark of Thermo
More informationA6 OPERATING INSTRUCTIONS
Amerec s A6 control for the AX steamer is a touch screen control intended to be mounted on a wall, generally in or near the steam bath. It may be mounted directly on the wall surface or, using an optional
More informationInstrument Control and Data Acquisition with Chromeleon
Technical Note 67 Instrument Control and Data Acquisition with Chromeleon Introduction Laboratories generally have a combination of chromatography devices from different manufacturers and, therefore, face
More informationWELDING CONTROL UNIT: TE 450 USER MANUAL
j WELDING CONTROL UNIT: TE 450 USER MANUAL RELEASE SOFTWARE No. 1.50 DOCUMENT NUMBER: MAN 4097 EDITION: MARCH 1998 This page is left blank intentionally. 2 / 34 TABLE OF CONTENTS SUBJECTS PAGE WELDING
More informationDetailed Design Report
Detailed Design Report Chapter 4 MAX IV Injector 4.6. Acceleration MAX IV Facility CHAPTER 4.6. ACCELERATION 1(10) 4.6. Acceleration 4.6. Acceleration...2 4.6.1. RF Units... 2 4.6.2. Accelerator Units...
More informationST-200Pro. Electrolyte Analyzer OPERATOR S MANUAL
ST-200Pro Electrolyte Analyzer OPERATOR S MANUAL ST-200 Pro Electrolyte Analyzer ST-200 Pro Electrolyte Analyzer SensaCore Medical Instrumentation Pvt. Ltd. Plot No. 3, Export Promotion Industrial Park,
More informationMITSUBISHI ELECTRONICS AMERICA, INC.
User Experiences Problem During Operation Success User Resolves Problem Using Troubleshooting Instructions in Operating Manual No Success User Calls Dealer Success Dealer Resolves Problem Over Phone Using
More informationHenkel Installation Handbook LINEGUARD 2001
Henkel Installation Handbook LINEGUARD 2001 HENKEL ITALIA S.r.l. Microprocessor instruments Henkel Installation Handbook Lineguard 2001 Page 2 Summary 1. The Microprocessor Instruments... 4 1.1. Norms
More informationModel CB 160 CO 2 incubators with hot air sterilization and heat sterilizable CO 2 sensor
Model CB 160 CO 2 incubators with hot air sterilization and heat sterilizable CO 2 sensor The BINDER CB series CO 2 incubator is the premium class among the CO 2 incubators. It is suitable for all sensitive
More informationThe PHI VersaProbe operates with two essential software programs: PHI Summitt and Vacuum Watcher. A third program, MultiPak, handles data reduction.
PHI VersaProbe Scanning XPS System I. Overview The PHI VersaProbe operates with two essential software programs: PHI Summitt and Vacuum Watcher. A third program, MultiPak, handles data reduction. PHI Summitt
More informationControl Unit CU 2.1 User Manual V6850A HPLC
Control Unit CU 2.1 User Manual V6850A HPLC 3 Table of Contents Control Unit... 5 Control... 5 Start Screen... 6 Main Menu... 6 Programs and Links... 7 General Settings... 8 Choosing the Network... 8 Setting
More informationAgilent 7000 Series Triple Quadrupole GC/MS System
Agilent 7000 Series Triple Quadrupole GC/MS System Concepts Guide The Big Picture Agilent Technologies Notices Agilent Technologies, Inc. 2011 No part of this manual may be reproduced in any form or by
More informationPAL LSI - Robust and Reliable Sample Handling
PAL LSI - Robust and Reliable Sample Handling PAL LSI: Enhanced productivity due to innovative technology Large Sample Capacity: 162 x 2 ml-vials can be stored on one tray holder. The standard length PAL
More informationFragment Analyzer Genomic DNA Analysis Kit
Fragment Analyzer Genomic DNA Analysis Kit NSTRUMENT Manufacturer: Advanced Analytical Technologies Distributor: Agilent Technologies (customercare_finland@agilent.com, +358 10 855 2465) Technical support:
More informationTechnical Procedure for Scanning Electron Microscope/ Energy Dispersive X-Ray System (SEM/EDX) for non-gsr Casework
Technical Procedure for Scanning Electron Microscope/ Energy Dispersive X-Ray System (SEM/EDX) for non-gsr Casework 1.0 Purpose This technical procedure shall be followed for the operation of the Scanning
More informationLimitations of a Load Pull System
Limitations of a Load Pull System General Rule: The Critical Sections in a Load Pull measurement setup are the sections between the RF Probe of the tuners and the DUT. The Reflection and Insertion Loss
More informationSPECIFICATION FIBER OPTIC SPLICE CLOSURE. Spec No : VSS-1007-BS403A-04A/SD. VSS-0107-BS403A-04A/SD R & D Center Manufacturing Division
SPECIFICATION FIBER OPTIC SPLICE CLOSURE Model Spec. No. Distribution Depts. VSOF-BS403A VSS-0107-BS403A-04A/SD R & D Center Manufacturing Division Sales Division Management Division Revision 10. 07 (Rev.4)
More informationThe Importance of Projector Maintenance.
The Importance of Projector Maintenance www.prometheanworld.com/activcare Contents Introduction 3 Projector Operation 4 Lamp Mode: Normal vs. High 4 Turning the Projector off 4 Projector Environment 5
More informationModel KBF 1020 Constant climate chambers with large temperature / humidity range
Model KBF 1020 Constant climate chambers with large temperature / humidity range The BINDER KBF is the specialist for unconditionally reliable stability testing and precise maintenance of constant climate
More informationICS-3000 System Installation Qualification
IC System IQ ICS-3000 System Installation Qualification 2006 Dionex Corporation Document No. 065080 Revision 02 September 2006 Doc. 065080-02 2006, Dionex Corporation All rights reserved worldwide. Printed
More informationORM0022 EHPC210 Universal Controller Operation Manual Revision 1. EHPC210 Universal Controller. Operation Manual
ORM0022 EHPC210 Universal Controller Operation Manual Revision 1 EHPC210 Universal Controller Operation Manual Associated Documentation... 4 Electrical Interface... 4 Power Supply... 4 Solenoid Outputs...
More informationKratos AXIS Ultra DLD X-ray Photoelectron Spectrometer Instructions
Kratos AXIS Ultra DLD X-ray Photoelectron Spectrometer Instructions Note: Enter your complete name, CMRF user code, and date into the record book. You must have previously filled out a project information
More informationOperation of CEBAF photoguns at average beam current > 1 ma
Operation of CEBAF photoguns at average beam current > 1 ma M. Poelker, J. Grames, P. Adderley, J. Brittian, J. Clark, J. Hansknecht, M. Stutzman Can we improve charge lifetime by merely increasing the
More informationSmartMix Manual V5350A
SmartMix Manual V5350A 3 Table of contents Overview................................................... 4 Features.................................................... 4 Scope of delivery................................................
More informationFinnigan PolarisQ User s Guide. PN , Revision D
Finnigan PolarisQ User s Guide PN 120014, Revision D 2004 Thermo Electron Corporation. All rights reserved. Printing History: Rev D, Oct 2004. Rev C, Sep 2004. Xcalibur, Finnigan TRACE DSQ, Finnigan FOCUS
More informationLadies and Gentlemen,
Ladies and Gentlemen, We have pleasure to present for your kind attention our catalogue of machines and equipment. In case of any special requests or questions concerning our products, please contact our
More informationPHI 5000 VersaProbe TM Operator s Guide
PHI 5000 VersaProbe TM Operator s Guide Part No. 705921 Rev. A Copyright 2006 ULVAC-PHI, INC. 370 Enzo, Chigasaki, JAPAN The PHI logo ( ) is a registered trademark of ULVAC-PHI, INC. Physical Electronics,
More informationEG4015. Digital Generator Governor Controller User Manual
EG4015 Digital Generator Governor Controller User Manual Digital Governor for use in Gas and Diesel Generators with smoke and idle controls. Senses generator frequency, no magnetic pickup unit (MPU) is
More informationAmplification Setup Methods for the Maxprep TM Liquid Handler Technical Manual
Amplification Setup Methods for the Maxprep TM Liquid Handler Technical Manual Instructions for use of Products AS9100, AS9101, AS9200, AS9201 TM526 Revised 2/19 Table of Contents 1 Description...2 2...3
More informationCommissioning the TAMUTRAP RFQ cooler/buncher. E. Bennett, R. Burch, B. Fenker, M. Mehlman, D. Melconian, and P.D. Shidling
Commissioning the TAMUTRAP RFQ cooler/buncher E. Bennett, R. Burch, B. Fenker, M. Mehlman, D. Melconian, and P.D. Shidling In order to efficiently load ions into a Penning trap, the ion beam should be
More informationTVP-20 / RVP-412 CCTV VIDEO-PLEXER
TVP-20 / RVP-412 CCTV VIDEO-PLEXER INSTRUCTION BOOK IB640701 4-1-2003 TABLE OF CONTENTS DESCRIPTION 3 MOUNTING INSTRUCTIONS 3 RVP-412 MOUNTING INSTRUCTIONS 3-4 HOW TO CABLE THE TVP-20 5 HOW TO CABLE THE
More informationRemote Control. degraded, causing unreliable operation. The recommended effective distance for remote operation is about 16 feet (5 meters).
Media Streaming Sound Bar RTS736W User Manual Remote Control using the remote control Point the remote control at the REMOTE SENSOR located on the unit (see Front Panel illustration for precise location).
More informationAnalog Devices Welcomes Hittite Microwave Corporation NO CONTENT ON THE ATTACHED DOCUMENT HAS CHANGED
Analog Devices Welcomes Hittite Microwave Corporation NO CONTENT ON THE ATTACHED DOCUMENT HAS CHANGED www.analog.com www.hittite.com THIS PAGE INTENTIONALLY LEFT BLANK v1.55 Typical Applications The is
More information2. Pin assignment list for EXC-73 Signal conversion
PAGE:1/6 Dear valued customers, Regarding HELIOT series Signal Conversion Unit EXC-73 ULVAC Technologies, Inc. Component Division measurement instrument 1 department 1. Overview Note EXC-73 is a signal
More informationQuick Start Function Summary Instructions for ASHCROFT GC52 Differential Pressure Transmitter Version 6.03 Rev. B
Quick Start Function Summary Instructions for ASHCROFT GC52 Differential Pressure Transmitter Version 6.03 Rev. B (See Complete I&M Manual for Further Detail) LOOK FOR THIS AGENCY MARK ON OUR PRODUCTS
More information9070 Smart Vibration Meter Instruction Manual
9070 Smart Vibration Meter Instruction Manual Overall machine and bearing conditions: vibration values are displayed with color coded alarm levels for ISO values and Bearing Damage (BDU). Easy vibration
More informationInserting System K1600
Inserting Solutions I Data Sheet Kern 1600 7.2018 I V4 I Page 1 / 7 Application Functions Separate cut documents (single sheet) ex stack, portrait or landscape Reading, grouping and collating at request
More informationCover Page for Lab Report Group Portion. Boundary Layer Measurements
Cover Page for Lab Report Group Portion Boundary Layer Measurements Prepared by Professor J. M. Cimbala, Penn State University Latest revision: 23 February 2017 Name 1: Name 2: Name 3: [Name 4: ] Date:
More informationDeveloping an AFM-based Automatic Tool for NanoAsperity Quantification
Developing an AFM-based Automatic Tool for NanoAsperity Quantification September 18, 2008 Sergey Belikov*, Lin Huang, Jian Shi, Ji Ma, Jianli He, Bob Tench, and Chanmin Su Veeco Instruments Inc., Santa
More informationStark Spectroscopy Deanna s Experimental Procedure NWU Hupp Lab Fall 2003
Stark Spectroscopy Deanna s Experimental Procedure NWU Hupp Lab Fall 2003 1. Generate mixed-valent state of compound check in 1mm cell. Ideally want Abs 1. 2. Setting up the instrument New Dewar i) Approx.
More informationPINMRF. Checkout Quiz - Varian Inova-300 Version
PINMRF Checkout Quiz - Varian Inova-300 Version Please carefully read every question and select the best answer(s). Some questions may have more than one correct answer. You must select all the correct
More informationTECHNICAL SPECIFICATION
TECHNICAL SPECIFICATION (FIBER OPTIC SPLICE CLOSURE) Model Spec. No. Distribution Depts. VSOF-BS403A SJP-0609-403A-01A/SD Quality Assurance Team Manufacturing Division Sales Division Management Division
More informationPanelView 1400e CRT Maintenance
Release Note PanelView 1400e CRT Maintenance Maximizing the life of your PanelView 1400e, CRT Terminals To maximize the life of a CRT, the following is strongly recommended: Adjust the external brightness
More informationTender Notification for the procurement of a Scanning Electron Microscope" at IISc (Last Date for submission of tenders: 3 rd October 2018)
Tender Notification for the procurement of a Scanning Electron Microscope" at IISc (Last Date for submission of tenders: 3 rd October 2018) Dear Sir/Madam, We are looking for a high-resolution scanning
More informationProgram Risks Risk Analysis Fallback Plans for the. John T. Seeman DOE PEP-II Operations Review April 26, 2006
Program Risks Risk Analysis Fallback Plans for the PEP-II B-FactoryB John T. Seeman DOE PEP-II Operations Review April 26, 2006 OPS Review Topics Are there any PEP-II program risks? Has the laboratory
More informationAccuLoad II TM - STD Operator Guide
Electronic Preset Controller AccuLoad II TM - STD Operator Guide Issue/Rev. 1. (6/99) Bulletin MN06041 The Most Trusted Name In Measurement Receipt of Equipment When the equipment is received, the outside
More informationRelease Note. Product Name: Current Version: Ver.Type Ver No. Date. RN Date: June 2017 Reference: Kramer Product Affected:
Release Note Product Name: SID-X3N Current Version: Ver.Type Ver No. Date Firmware V8.9.38815 June 2017 Software Web Hardware RN Date: June 2017 Reference: 12658 Kramer Product Affected: Product Name Minimum
More informationModel KBF 115 Constant climate chambers with large temperature / humidity range
Model KBF 115 Constant climate chambers with large temperature / humidity range The BINDER KBF is the specialist for unconditionally reliable stability testing and precise maintenance of constant climate
More informationUp to 85% higher Service Life due to efficient sealing method.
Robot Accessories Feeding through Up to 85% higher Service Life due to efficient sealing method. 346 Robot Accessories Feeding through Feeding through DDF 2 Rotary Feed-through Series Size Page DDF 2 348
More informationOperator s Manual. Ultegra. Health Scale. Fairbanks Scales by Fairbanks Scales Inc. All rights reserved. Revision 5 06/07
Operator s Manual Ultegra Health Scale Fairbanks Scales 2007 by Fairbanks Scales Inc. All rights reserved 50735 Revision 5 06/07 Amendment Record Ultegra Health Scale 50735 Manufactured by Fairbanks Scales
More informationOperation Manual for the Refractive Index detector VE-3580
Operation Manual for the Refractive Index detector VE-3580 October 2012 (MAN0446-1.0) (Version 4.3) Page 1 of 32 pages Table of Contents Table of Contents... 3 Notices, Warnings and Definitions... 5 Figures...
More informationUNIGAS Hand-held combustion gas analyzer with 3 sensors
UNIGAS 2000 Hand-held combustion gas analyzer with 3 sensors Unigas 2000+ is a professional flue gas analyser for 3 gases: O 2, CO & NO with electrochemical sensors, plus CO 2 and NO x calculated. Description
More informationModel KBF P 240 Humidity test chambers with ICH-compliant light source
Model KBF P 240 Humidity test chambers with ICH-compliant light source The KBF P humidity test chamber, equipped with ICH-compliant light source, is an expert when it comes to photostability tests and
More informationFragment Analyzer Genomic DNA Analysis Kit
Fragment Analyzer Genomic DNA Analysis Kit INSTRUMENT Manufacturer: Distributor: Technical support: Advanced Analytical Technologies Kem-En-Tec Nordic (info@kem-en-tec.dk) Klaus T. Nielsen (kn@kem-en-tec.com)
More informationTroubleshooting. 1. Symptom: Status indicator (Red LED) on SSR is constant on. 2. Symptom: Output indicator (Yellow LED) on SSR is flashing.
Product Data Electrical Data SST (Transmitter) SSR (Receiver) Supply voltage 18 30 V dc Max. Voltage ripple 15 % (within supply range) Current consumption 100 ma (RMS) 75 ma Digital - 100 ma Max. outputs
More informationModel ED 400 Drying and heating chambers Classic.Line with natural convection
Model ED 400 Drying and heating chambers Classic.Line with natural convection BENEFITS Uniform drying conditions thanks to APT.line technology Identical test conditions throughout the chamber interior
More informationCR7000. CRT Analyzer & Restorer. Easily Test And Restore CRTs With The Most Complete Tests Available For Added Profit And Security.
CR7000 CRT Analyzer & Restorer Easily Test And Restore CRTs With The Most Complete Tests Available For Added Profit And Security. S1 New Demands From Higher Performance CRTs Require New Analyzing Techniques
More informationFragment Analyzer CRISPR Discovery Gel Kit
Fragment Analyzer CRISPR Discovery Gel Kit INSTRUMENT Manufacturer: Distributor: Technical support: Advanced Analytical Technologies Kem-En-Tec Nordic (info@kem-en-tec.dk) Klaus T. Nielsen (kn@kem-en-tec.com)
More information