The PHI VersaProbe operates with two essential software programs: PHI Summitt and Vacuum Watcher. A third program, MultiPak, handles data reduction.

Size: px
Start display at page:

Download "The PHI VersaProbe operates with two essential software programs: PHI Summitt and Vacuum Watcher. A third program, MultiPak, handles data reduction."

Transcription

1 PHI VersaProbe Scanning XPS System I. Overview The PHI VersaProbe operates with two essential software programs: PHI Summitt and Vacuum Watcher. A third program, MultiPak, handles data reduction. PHI Summitt controls the movement of the sample stage, the analysis setup and acquisition, and the control of the ion gun and neutralizer gun. Vacuum Watcher is a separate program that interacts with PHI Summitt and is used to load and transfer samples and differentially pump the ion gun for sputtering. Summitt and Watcher should always be open (running) on the instrument. Do not close these programs without specific instructions from a lab supervisor. If Summitt crashes, it can be closed and reopened. If problem persists after restart, contact a lab supervisor. Make sure Summitt is not already open and minimized before opening. Two open versions will cause problems. Never close Watcher. It remains open even if Summitt is restarted. If instructed to restart the computer, open Watcher first, then Summitt. II. Features The VersaProbe XPS is capable of x-ray beam spot sizes from 10um-200um and a High Power setting with a beam spot of roughly 1400x300um. Secondary electron images (SXI) can be acquired to locate specific features to be analyzed. Computer controlled X, Y, Z, Tilt and Rotate stage control. Auto Z alignment for optimization of x-ray impact and electron extraction. Angle-resolved measurements can be done to analyze ultra thin films (with 1 holder only). Sputtering can be done to analyze thicker films or remove surface contamination. System has both Ar+ and C60 sputter guns. Dual beam (electrons and ions) for charge neutralization of insulating samples. 1

2 III. Mounting Samples 1 and 2 mounts are available. CuBe clips or double-sided tape can be used to adhere samples to the mounts. Use the smallest piece of tape possible to avoid excessive outgassing. Silver or Carbon paint (allowed to dry) is also an option for mounting. Molybdenum masks can be secured to avoid using tape or paint. For angle-resolved experiments, must use 1 holder. IV. Sample Introduction (See Figures 1 and 2.) 1. Watcher -> Press Backfill Intro. Will close valve over turbo pump and open N2 gas to fill chamber. Lift cap and place sample onto fork with tongs. For 1 holders, hold tongs with top groove and slide bottom groove onto fork. Check o-ring on underside of cap, remove any dust or fibers with a clean wipe, and replace cap. 2. Watcher -> Press Pump Intro. N2 valve closes, Roughing pump acts on intro chamber, when vacuum is ready, turbo will be opened. A timer starts counting in Watcher. 20 minutes are allowed for roughing, and then a new timer starts for the turbo. Wait for timer to finish and Task Pump Intro complete message in Watcher. The gauge for the Intro chamber will read until ~10-2 Pa and then read 0. While waiting for intro chamber to reach desired vacuum level: Enable the PHI in Badger. (Will allow use of optical microscope and CCD camera on sample.) Set up file for saved data. Summitt -> Push Folder button at bottom left of Acquisition Settings window. Save in C:\Data. Check Summitt to make sure previous user did not leave undesirable settings. (I.e. No sputter is selected, Neutralize is off. In Summitt -> Image window, under SXI/AutoZ tab, make sure Neutralizer and Ion Gun Neut. are deselected if not desired.) Summitt -> Image window: Select desired Xray beam settings to be used later for AutoZ and for acquisition in the SXI/AutoZ and X-Ray tabs. 3. Summitt -> At bottom of Image window, choose Stage tab. Press Intro to move the stage to the home position for sample introduction. An Intro dialog box will open and indicate that the stage is moving. If the X, Y, Z coordinates of the stage are visible in the Summitt Image window, the numbers will be red when the stage is moving. 2

3 When stage has stopped moving, follow instructions in Intro dialog box (see next step below). FIG 1: Vacuum Watcher Intro Procedure FIG 2: Summit Image Window, Stage Tab Intro Procedure (5) (8) (3) 3

4 4. Watcher -> Push Transfer Sample A Task Message window will open. Press Yes to confirm sample transfer. Gate valve (V1) will open. Wait until white line has moved to open position on gate valve to confirm before moving sample. Another Task Message window will open prompting to insert sample. Carefully push transfer rod into chamber by grasping black handle and sliding it until the stop position. Push OK on Task Message Window. Sample should be on fork, positioned just above the stage in the main chamber. 5. Summitt -> Push Up [Z] in the Intro dialog box. The stage will move Z upward to grab the sample holder. The Intro dialog box will indicate that the stage is moving. 6. If the stage has successfully grabbed the sample and is no longer moving, carefully pull the transfer rod out of the chamber to the stop position. 7. The gate valve (V1) will automatically close. If stage is not aligned to the sample correctly, and a problem arises during Up [Z], push Abort in the Intro dialog box and manually drop Z. To drop Z: Summitt Image Window-> type a small number in the Z(mm) box on the Stage tab and push Move. Contact a lab manager for help. 8. Summitt -> In the Intro dialog box, push Set (Sample Inf.) to update sample thickness and holder size, thereby enabling collision protection for the extractor lens and good Z-height alignment. Push OK to close box. Auto Z FIG 3: Summit Image Window, Auto Z FIG 4: Stage Controller 4

5 V. AutoZ Alignment (Figure 3). 1. Manually raise stage until samples are focused in CCD image. Height for a bare holder is 17mm. a. Raise stage to Z= 17- sample thickness (i.e. 16.5mm for typical wafer). b. Summitt -> In Image window, Stage tab, type new value into Z(mm). Push Move button and stage will move. 2. Viewing through glass windows and CCD camera image, translate to desired sample with digital joystick (Stage Controller, Figure 4.). 3. Summitt -> SXI/AutoZ tab in Image window. Pick the desired x-ray settings for AutoZ. 200um file is typical-- larger spot gives better statistics. (HP option is not available for AutoZ.) a. Auto Z viewer should be on and Point should be selected. b. If sample is likely to charge, may need Neutralizer and Ion Gun Neut. See Section VII: Starting the Ion Gun. 4. Summitt -> Image Window, push AutoZ to start routine. If need to stop AutoZ routine, press AutoZ button again. Watch progress counts should increase until maximum. The stage height will rise for routine increments and then decrease and rise again in smaller steps until done. Push Ok and close AutoZ window when finished. a. If final height is not on desired sample, repeat when centered on desired sample. b. If AutoZ routine fails, manually lower Z to a lower starting position (push Move after entering new value) and retry. c. Best result when have more than 1000cps in AutoZ. d. Do not do AutoZ on a tilted sample. AutoZ is already calibrated for tilt. VI. Data Acquisition 1. To perform a survey scan, type SUR into the Region Name box in the Summitt Acquisition Setting window. Change the default parameters as needed. Typical settings include: Pass Energy: 117.4eV (gives 1.5eV energy resolution) Range: eV, or eV if needed Energy Step: 1eV Time/Step: 20ms 1 repeat; 3 total cycles (upper left). 2. To perform a high resolution scan, type desired element and transition into Region Name (i.e. C1s, O1s, Ag3d5 ) a. Press Element Table to open periodic table to select elements/transitions. Push Add for desired transition. b. Change the default parameters as needed. Typical settings include: 5

6 Pass Energy: 23.5eV Range: varies by element, can push Test Acquire to ensure peak is within region. Energy Step: 0.1eV Time/Step: 50ms Repeats: Add repeats to regions that need better statistics. c. For faster option, can choose Unscan at bottom of Summitt-> Acquisition Setting window. 3. Place a check in box at left side of each line to be acquired. 4. Do not collect SUR and high resolution regions at the same time/under same file name. 5. Summitt-> Image Window, XRAY tab, choose the spot size/power to be used for scan. HP or 200um spots give best statistics and most counts, but may need smaller spot for localized area or acquisition within sputtered region. See Figure Summitt-> Acquisition Settings Window (LHS). Push Acquire to start. 7. To use Ion Gun for sputtering or charge compensation, see Section To analyze data, see the MultiPak manual. Choose X-ray spot size FIG 5: Summit Image Window, X-RAY VII. Starting the Ion Gun Neutralization or Sputtering (Figures 6-9) For the purpose of sputtering or neutralizing, must follow steps 1-2e to start the Ion Gun. 1. Watcher -> Push Diff vlv Open, opens V4 to start pumping Ion Gun (will stop pumping on intro chamber). 2. Summitt-> Image window (RHS), Ion Gun tab (FIG 6): a. For charge neutralization, choose Neutralize file under Neutralize heading. 6

7 b. For sputtering, choose desired sputter file. Chart taped to PHI console explains choices. c. Push Standby (will turn on power to sputter gun). d. A dialogue box will open reminding you to push Diff vlv Open. OK. e. On Pfeiffer RVG050 (FIG 7) box near computer monitor, toggle white switch to Set Point. In Ion Gun tab, check Ext.Pressure(mPa) readout and watch until stabilizes ~10mPa. The main chamber pressure will also get worse (1x10-6 ). Choose sputter settings Choose Neutralize file FIG 6: Summit Image Window, Ion Gun FIG 7: Pfeiffer RVG050C Vacuum Control Box 7

8 3. Charge Neutralization a. Summitt-> Image window (RHS), Neutralizer tab (FIG 8). b. Push Standby (will turn on power to electron flood gun). c. Summitt-> Acquisition Settings window (LHS). Under Neutralize, Press ON. Neutralization will begin when spectrum is acquired. d. If need to do AutoZ, check the Neutralize and Ion Gun Neutralize boxes in SXI/AutoZ tab (FIG 3). FIG 8: Summit Image Window, Neutralizer 4. Sputter Modes Manual Sputter o Summitt-> When pressure conditions are ready, pushing Sputter in Ion Gun tab will immediately sputter. o For manual, timed sputter: o choose a low KV sputter file, check timed box, set time, then push Sputter. Depth Profiling (See FIG 9): o Summitt-> Ion gun is in Standby in the Ion Gun tab and desired sputter file is chosen. o Summitt-> Acquisition Setting window -> Check/Enter desired acquisitions. o Summitt-> Acquisition Setting window (LHS)-> Push Alternate in Sputter Mode category. Will open Sputter Setting options below. o Typical Settings: Sputter Time (min)= total time to sputter. 8

9 Interval Time (min)= sputter increments, in units of.5min. Delayed Acq. Time (sec) = Delay before acquisition to allow count rate to stabilize. 3s is typical. Num(Presputter) = How many scans to take before 1 st sputter. 1 is typical. o Now, number of cycles is determined by sputter settings. Add repeats to increase signal for some acquisitions. o Push ACQUIRE to begin depth profile. o The charge neutralizer can be used during depth profiling. FIG 9. Depth Profile Setup 5. To Shut Down Sputtering/Neutralizer a. Summitt-> Acquisition Setting window (LHS)-> Choose No Sputter in Sputter Mode category (FIG 9). b. Summitt-> Acquisition Setting window (LHS)-> Choose Off in Neutralize category (FIG 9). c. Summitt-> Image Window (RHS) -> Ion Gun Tab: Push Off (FIG 6). d. On Pfeiffer RVG050 box (FIG 7), toggle white switch to Limit. Never turn off power to Pfeiffer RVG050! e. Summitt-> Image Window -> Push Off in Neutralizer Tab (FIG 8). 9

10 f. If used neutralizer during AutoZ, go to SXI/AutoZ tab and uncheck Neutralizer and Ion Gun Neut. g. Watcher -> Press Pump Intro so V4 closes and turbo again pumps the intro chamber. VIII. Sample Extraction (See Figures 10-11). 1. Summitt-> Image window-> Press Extract. Sample stage will lower, return to original X, Y and R settings, and then rise to a high Z position. 2. Watcher -> Push Transfer Sample A Task Message window will open. Press Yes to confirm sample transfer. Gate valve (V1) will open. Wait until white line has moved to open position on gate valve to confirm before moving transfer rod. Another Task Message window will open prompting to insert sample. Carefully push transfer rod into chamber and sliding it until the stop position. Fork should slide onto sample holder. Push OK on Task Message Window. 3. Summitt -> Push Down [Z] in the Intro dialog box. The stage will move Z down away from sample holder. The Intro dialog box will indicate that the stage is moving. If the stage is safely below the sample, carefully pull the transfer rod out of the chamber to the stop position. The gate valve (V1) will automatically close. 4. Watcher -> Press Backfill Intro. Will close valve over turbo pump and open N2 gas to fill chamber. With gloves on, remove sample and replace cap. 5. Watcher -> Press Pump Intro. Ensure that Intro chamber is pumped before leaving tool. Disable in Badger. (1) FIG 10: Summit Image Window, Stage Tab Extract Procedure 10

11 FIG 11: Vacuum Watcher Extract procedure (5) (2) (4) 11

PHI 5000 VersaProbe TM Operator s Guide

PHI 5000 VersaProbe TM Operator s Guide PHI 5000 VersaProbe TM Operator s Guide Part No. 705921 Rev. A Copyright 2006 ULVAC-PHI, INC. 370 Enzo, Chigasaki, JAPAN The PHI logo ( ) is a registered trademark of ULVAC-PHI, INC. Physical Electronics,

More information

Kratos AXIS Ultra DLD X-ray Photoelectron Spectrometer Instructions

Kratos AXIS Ultra DLD X-ray Photoelectron Spectrometer Instructions Kratos AXIS Ultra DLD X-ray Photoelectron Spectrometer Instructions Note: Enter your complete name, CMRF user code, and date into the record book. You must have previously filled out a project information

More information

1. Check the accelerating voltage, must be at 200 kv (right screen), HT (µa) (left panel) at and Emission (left panel) at

1. Check the accelerating voltage, must be at 200 kv (right screen), HT (µa) (left panel) at and Emission (left panel) at JEOL 2010F MANUAL Quick check list 1. Check the accelerating voltage, must be at 200 kv (right screen), HT (µa) (left panel) at 0.96-0.97 and Emission (left panel) at 155-160. 2. Check the vacuum sequence

More information

Standard Operating Procedure of nanoir2-s

Standard Operating Procedure of nanoir2-s Standard Operating Procedure of nanoir2-s The Anasys nanoir2 system is the AFM-based nanoscale infrared (IR) spectrometer, which has a patented technique based on photothermal induced resonance (PTIR),

More information

Scanning Electron Microscopy (FEI Versa 3D Dual Beam)

Scanning Electron Microscopy (FEI Versa 3D Dual Beam) Scanning Electron Microscopy (FEI Versa 3D Dual Beam) This operating procedure intends to provide guidance for basic measurements on a standard sample with FEI Versa 3D SEM. For more advanced techniques

More information

Technical Procedure for Scanning Electron Microscope/ Energy Dispersive X-Ray System (SEM/EDX) for non-gsr Casework

Technical Procedure for Scanning Electron Microscope/ Energy Dispersive X-Ray System (SEM/EDX) for non-gsr Casework Technical Procedure for Scanning Electron Microscope/ Energy Dispersive X-Ray System (SEM/EDX) for non-gsr Casework 1.0 Purpose This technical procedure shall be followed for the operation of the Scanning

More information

University of Minnesota Nano Fabrication Center Standard Operating Procedure

University of Minnesota Nano Fabrication Center Standard Operating Procedure Equipment Name: Focused Ion Beam (FIB) Coral Name: fib Revision Number: 2 Model: FEI Quanta 200 3D Revisionist: Kevin Roberts Location: Area 3 Date: 9/17/2013 1 Description The Quanta 200 3D is a DualBeam

More information

CHECKLIST FOR VERIOS OPERATION 1. GENERAL The SEM lab is used assuming "operating room" cleanliness, i.e., the SEM lab is a high visibility lab and

CHECKLIST FOR VERIOS OPERATION 1. GENERAL The SEM lab is used assuming operating room cleanliness, i.e., the SEM lab is a high visibility lab and CHECKLIST FOR VERIOS OPERATION 1. GENERAL The SEM lab is used assuming "operating room" cleanliness, i.e., the SEM lab is a high visibility lab and must be kept clean and neat so clean up behind yourself

More information

2.1. Log on to the TUMI system (you cannot proceed further until this is done).

2.1. Log on to the TUMI system (you cannot proceed further until this is done). FEI DB235 ex-situ lift out TEM sample preparation procedure Nicholas G Rudawski ngr@ufledu (805) 252-4916 Last updated: 06/19/15 DISCLAIMER: this procedure describes one specific method for preparing ex-situ

More information

Standard Operating Procedure II: EDS (Bruker Flat-Quad)

Standard Operating Procedure II: EDS (Bruker Flat-Quad) Standard Operating Procedure II: EDS (Bruker Flat-Quad) ywcmatsci.yale.edu ESC II, Room A119F 810 West Campus Drive West Haven, CT 06516 Version 1.1, October 2018 1 > FOLLOW the SOP strictly to keep the

More information

JEM 2100 Manual. Operation and Basic Alignment Instructions. Check the vacuum levels (power supply closet) Column (blue scale): < 2.

JEM 2100 Manual. Operation and Basic Alignment Instructions. Check the vacuum levels (power supply closet) Column (blue scale): < 2. 1. Verify the vacuum: JEM 2100 Manual Operation and Basic Alignment Instructions Check the vacuum levels (power supply closet) Column (blue scale): < 2.5x10-5 Pa If vacuum is not good enough, contact somebody

More information

HV/PHA Adjustment (PB) Part

HV/PHA Adjustment (PB) Part HV/PHA Adjustment (PB) Part Contents Contents 1. How to set Part conditions...1 1.1 Setting conditions... 1 2. HV/PHA adjustment sequence...7 3. How to use this Part...9 HV/PHA Adjustment (PB) Part i

More information

Operation Procedure for Phillips XL30 ESEM

Operation Procedure for Phillips XL30 ESEM Operation Procedure for Phillips XL30 ESEM The ESEM will be left in the ON state when not in use. The chamber will be at high vacuum, filament on, stage at home position, VAC and HT buttons lit, and monitor

More information

680 System Operator s Guide

680 System Operator s Guide 680 System Operator s Guide For AES PC-ACCESS, Ver. 7 Part No. 647986 Rev. A Copyright 2000 Physical Electronics, Inc. 6509 Flying Cloud Drive Eden Prairie, MN 55344 The PHI logo is a registered trademark

More information

INCA ENERGY EDS TRAINING. System Block Diagram. INCA Energy Software. Xiang Yang EM SMU. Navigators. Point & ID Navigator.

INCA ENERGY EDS TRAINING. System Block Diagram. INCA Energy Software. Xiang Yang EM SMU. Navigators. Point & ID Navigator. INCA ENERGY EDS TRAINING Xiang Yang EM Center @ SMU System Block Diagram INCA Energy Software Navigators Point & ID Navigator Data tree 1 Spectrum Export Bitmap TIFF JPEG Metafile EMSA ISIS Optimum Conditions

More information

HRXRD Analysis of Epitaxial Thin Films

HRXRD Analysis of Epitaxial Thin Films HRXRD Analysis of Epitaxial Thin Films on the Rigaku Smartlab Multipurpose Diffractometer Scott A Speakman, Ph.D Center for Materials Science and Engineering at MIT For assistance in the X-ray lab, contact

More information

DektakXT Profilometer. Standard Operating Procedure

DektakXT Profilometer. Standard Operating Procedure DektakXT Profilometer Standard Operating Procedure 1. System startup and sample loading: a. Ensure system is powered on by looking at the controller to the left of the computer.(it is an online software,

More information

University of Minnesota Minnesota Nano Center Standard Operating Procedure

University of Minnesota Minnesota Nano Center Standard Operating Procedure Equipment Name: CHA Evaporator Coral Name: ebevap-cha Revision Number: 12 Model: SEC 600 Revisionist: L. von Dissen Location: PAN Bay 3 Date: 04/13/2018 1 Description The CHA Evaporator is a single source

More information

Mechanical aspects, FEA validation and geometry optimization

Mechanical aspects, FEA validation and geometry optimization RF Fingers for the new ESRF-EBS EBS storage ring The ESRF-EBS storage ring features new vacuum chamber profiles with reduced aperture. RF fingers are a key component to ensure good vacuum conditions and

More information

Cover Page for Lab Report Group Portion. Boundary Layer Measurements

Cover Page for Lab Report Group Portion. Boundary Layer Measurements Cover Page for Lab Report Group Portion Boundary Layer Measurements Prepared by Professor J. M. Cimbala, Penn State University Latest revision: 23 February 2017 Name 1: Name 2: Name 3: [Name 4: ] Date:

More information

Commissioning the TAMUTRAP RFQ cooler/buncher. E. Bennett, R. Burch, B. Fenker, M. Mehlman, D. Melconian, and P.D. Shidling

Commissioning the TAMUTRAP RFQ cooler/buncher. E. Bennett, R. Burch, B. Fenker, M. Mehlman, D. Melconian, and P.D. Shidling Commissioning the TAMUTRAP RFQ cooler/buncher E. Bennett, R. Burch, B. Fenker, M. Mehlman, D. Melconian, and P.D. Shidling In order to efficiently load ions into a Penning trap, the ion beam should be

More information

FEI Strata Dual-beam FIB

FEI Strata Dual-beam FIB FEI Strata Dual-beam FIB Quick start Guide Compiled by Mat t hew Hughes and Tony Chen Page 0 Purpose of this guide INTRODUCTION This quick start guide is meant to provide cursory operational knowledge

More information

1.2 Universiti Teknologi Brunei (UTB) reserves the right to award the tender in part or in full.

1.2 Universiti Teknologi Brunei (UTB) reserves the right to award the tender in part or in full. TENDER SPECIFICATIONS FOR THE SUPPLY, DELIVERY, INSTALLATION AND COMMISSIONING OF ONE UNIT OF VARIABLE PRESSURE ENVIRONMENTAL SCANNING ELECTRON MICROSCOPE (SEM) CUM ENERGY DISPERSIVE SPECTROSCOPY (EDS)

More information

READ THIS FIRST. Morphologi G3. Quick Start Guide. MAN0412 Issue1.1

READ THIS FIRST. Morphologi G3. Quick Start Guide. MAN0412 Issue1.1 READ THIS FIRST Morphologi G3 Quick Start Guide MAN0412 Issue1.1 Malvern Instruments Ltd. 2008 Malvern Instruments makes every effort to ensure that this document is correct. However, due to Malvern Instruments

More information

CHA EVAPORATOR. User guidelines. p.1. by Carlos Manzanedo. Last revised: 10/05/2000

CHA EVAPORATOR. User guidelines. p.1. by Carlos Manzanedo. Last revised: 10/05/2000 CHA EVAPORATOR User guidelines by Carlos Manzanedo. Last revised: 10/05/2000 p.1 Introduction. The CHA evaporator has the following characteristics: 1. 2 independent resistive power supplies. 2. 1 Electron

More information

SPM Training Manual Veeco Bioscope II NIFTI-NUANCE Center Northwestern University

SPM Training Manual Veeco Bioscope II NIFTI-NUANCE Center Northwestern University SPM Training Manual Veeco Bioscope II NIFTI-NUANCE Center Northwestern University Introduction: Scanning Probe Microscopy (SPM) is a general term referring to surface characterization techniques that utilize

More information

University of Minnesota College of Science and Engineering Characterization Facility Ganesha SAXSLAB User manual

University of Minnesota College of Science and Engineering Characterization Facility Ganesha SAXSLAB User manual University of Minnesota College of Science and Engineering Characterization Facility Ganesha SAXSLAB User manual # Section Pg # 1 Planning a SAXS experiment 1 2 Initial Setup 2 3 Sample mounting and loading

More information

FIB Operating Procedure. Effective Date: 08/14/2012 Author(s): Jiong Hua Phone:

FIB Operating Procedure. Effective Date: 08/14/2012 Author(s): Jiong Hua Phone: FIB Operating Procedure Effective Date: 08/14/2012 Author(s): Jiong Hua Phone: 402-472-3773 Email: jhua2@unl.edu 1 1 Introduction 1.1 Key Words Focused Ion Beam (FIB), FEI Strata 201, Ion milling 1.2 Purpose

More information

TASKI Service Tool Edition: V5.10/2014

TASKI Service Tool Edition: V5.10/2014 Edition: V5.10/2014 Index 1 General 1.1 General information 1 1.1.1 Part reference 1 1.1.2 Consumable supplies 1 1.1.3 Direction description 1 1.1.4 Power source 1 1.2 Required material 2 1.2.1 Tools 2

More information

WELDING CONTROL UNIT: TE 450 USER MANUAL

WELDING CONTROL UNIT: TE 450 USER MANUAL j WELDING CONTROL UNIT: TE 450 USER MANUAL RELEASE SOFTWARE No. 1.50 DOCUMENT NUMBER: MAN 4097 EDITION: MARCH 1998 This page is left blank intentionally. 2 / 34 TABLE OF CONTENTS SUBJECTS PAGE WELDING

More information

XRD-DSC Sample Alignment (BB) Part

XRD-DSC Sample Alignment (BB) Part XRD-DSC Sample Alignment (BB) Part Contents Contents 1. How to set Part conditions...1 1.1 Setting conditions... 1 1.2 Customizing scan conditions and slit conditions... 5 2. Sample alignment sequence...9

More information

PulseCounter Neutron & Gamma Spectrometry Software Manual

PulseCounter Neutron & Gamma Spectrometry Software Manual PulseCounter Neutron & Gamma Spectrometry Software Manual MAXIMUS ENERGY CORPORATION Written by Dr. Max I. Fomitchev-Zamilov Web: maximus.energy TABLE OF CONTENTS 0. GENERAL INFORMATION 1. DEFAULT SCREEN

More information

SEM- EDS Instruction Manual

SEM- EDS Instruction Manual SEM- EDS Instruction Manual Double-click on the Spirit icon ( ) on the desktop to start the software program. I. X-ray Functions Access the basic X-ray acquisition, display and analysis functions through

More information

Non-Invasive Energy Spread Monitoring for the JLAB Experimental Program via Synchrotron Light Interferometers

Non-Invasive Energy Spread Monitoring for the JLAB Experimental Program via Synchrotron Light Interferometers Non-Invasive for the JLAB Experimental Program via Synchrotron Light Interferometers P. Chevtsov, T. Day, A.P. Freyberger, R. Hicks Jefferson Lab J.-C. Denard Synchrotron SOLEIL 20th March 2005 1. Energy

More information

Quick Start Bruker Dimension Icon AFM

Quick Start Bruker Dimension Icon AFM Do not remove Quick Start Bruker Dimension Icon AFM March 3, 2015 GLA Contacts Harold Fu (hfu@caltech.edu) Weilai Yu (wyyu@caltech.edu) Bruker Tech Support (AFMSupport@bruker-nano.com 800-873-9750) Watch

More information

TT AFM LongBeach Procedures and Protocols V2.1

TT AFM LongBeach Procedures and Protocols V2.1 TT AFM LongBeach Procedures and Protocols V2.1 1. Startup Procedure 1. Turn on PC: Allow to boot to Windows. Turn on monitor. Password is afm 2. Turn on second PC that controls the video camera. 3. Turn

More information

Standard Operating Procedure: Dicing Saw

Standard Operating Procedure: Dicing Saw Contents Hardware Description and Principle of Operation... 1 Procedure... 2 Emergency Stop... 12 Allowed Activities... 13 Disallowed Activities... 13 What to watch out for during operation... 13 Common

More information

Standard Operating Procedure for FEI Helios 660 NanoLab Part I: SEM Version

Standard Operating Procedure for FEI Helios 660 NanoLab Part I: SEM Version Standard Operating Procedure for FEI Helios 660 NanoLab Part I: SEM Version Helios reservations may be made online using the NERCF website. Note: Always wear gloves when venting the system and exchanging

More information

A dedicated data acquisition system for ion velocity measurements of laser produced plasmas

A dedicated data acquisition system for ion velocity measurements of laser produced plasmas A dedicated data acquisition system for ion velocity measurements of laser produced plasmas N Sreedhar, S Nigam, Y B S R Prasad, V K Senecha & C P Navathe Laser Plasma Division, Centre for Advanced Technology,

More information

Practical Application of the Phased-Array Technology with Paint-Brush Evaluation for Seamless-Tube Testing

Practical Application of the Phased-Array Technology with Paint-Brush Evaluation for Seamless-Tube Testing ECNDT 2006 - Th.1.1.4 Practical Application of the Phased-Array Technology with Paint-Brush Evaluation for Seamless-Tube Testing R.H. PAWELLETZ, E. EUFRASIO, Vallourec & Mannesmann do Brazil, Belo Horizonte,

More information

BitWise (V2.1 and later) includes features for determining AP240 settings and measuring the Single Ion Area.

BitWise (V2.1 and later) includes features for determining AP240 settings and measuring the Single Ion Area. BitWise. Instructions for New Features in ToF-AMS DAQ V2.1 Prepared by Joel Kimmel University of Colorado at Boulder & Aerodyne Research Inc. Last Revised 15-Jun-07 BitWise (V2.1 and later) includes features

More information

Figure 1. MFP-3D software tray

Figure 1. MFP-3D software tray Asylum MFP-3D AFM SOP January 2017 Purpose of this Instrument: To obtain 3D surface topography at sub-nanometer scale resolution, measure contact and friction forces between surfaces in contact, measure

More information

Personal Protective Equipment Wear nitrile gloves, lab coat, and safety glasses as a minimum protection, unless otherwise indicated.

Personal Protective Equipment Wear nitrile gloves, lab coat, and safety glasses as a minimum protection, unless otherwise indicated. 4pt Bending, Mouse This protocol is for standard Jepsen 4pt bending of adult mouse bone. Safety considerations Please reference the Jepsen laboratory when using this protocol. This protocol is subject

More information

JEM 3010 Manual (2011 Edition)

JEM 3010 Manual (2011 Edition) JEM 3010 Manual (2011 Edition) Basic Alignment Instructions Check the vacuum (power supply closet) and write on the log book: Gun pressure SIP2 < 2x10-5 Pa (usually 0.8-1.0x10-5 Pa) Column pressure SIP1

More information

ivw-ud322 / ivw-ud322f

ivw-ud322 / ivw-ud322f ivw-ud322 / ivw-ud322f Video Wall Controller Supports 2 x 2, 2 x 1, 3 x 1, 1 x 3, 4 x 1 & 1 x 4 Video Wall Array User Manual Rev. 1.01 i Notice Thank you for choosing inds products! This user manual provides

More information

Professional 4-channel Thermocouple Thermometer plus Infrared Thermometer

Professional 4-channel Thermocouple Thermometer plus Infrared Thermometer Professional 4-channel Thermocouple Thermometer plus Infrared Thermometer TIF3340 BUTTONS = Power ON or OFF. = Steps through maximum (MAX), minimum (MIN), and average (AVG) readings. To exit MAX / MIN

More information

Florida State University Thayumanasamy Somasundaram

Florida State University Thayumanasamy Somasundaram 2016 Florida State University 2013-16 Thayumanasamy Somasundaram [PANALYTICAL X-PERT PRO POWDER DATA COLLECTION] A quick start-up procedure for collecting powder x-ray diffraction data from PANalytical

More information

using the Scott A Speakman, Ph.D Center for Materials Science and Engineering at MIT

using the Scott A Speakman, Ph.D Center for Materials Science and Engineering at MIT X-Ray Reflectivity using the PANalytical X Pert Pro MPD Scott A Speakman, Ph.D Center for Materials Science and Engineering at MIT http://prism.mit.edu/xray Modified for configuration used at University

More information

Operation of CEBAF photoguns at average beam current > 1 ma

Operation of CEBAF photoguns at average beam current > 1 ma Operation of CEBAF photoguns at average beam current > 1 ma M. Poelker, J. Grames, P. Adderley, J. Brittian, J. Clark, J. Hansknecht, M. Stutzman Can we improve charge lifetime by merely increasing the

More information

Approved by: / / R. Battaglia 12/16/2016

Approved by: / / R. Battaglia 12/16/2016 Fabrication Laboratory Revision: H Rev Date: 12/16/16 Approved by: Process Engineer / / R. Battaglia 12/16/2016 Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Varian

More information

LUMIGEN INSTRUMENT CENTER X-RAY CRYSTALLOGRAPHIC LABORATORY: WAYNE STATE UNIVERSITY

LUMIGEN INSTRUMENT CENTER X-RAY CRYSTALLOGRAPHIC LABORATORY: WAYNE STATE UNIVERSITY Standard Operating Procedure for the Bruker X8 APEX II Single-Crystal X- Ray Diffractometer Contact Manager: Dr. Cassie Ward ward@wayne.edu Office room 061 Chemistry (313) 577-2587 LIC Lab: (313) 577-0518

More information

CHA EVAPORATOR Short Reference User guidelines

CHA EVAPORATOR Short Reference User guidelines CHA EVAPORATOR Short Reference User guidelines by Carlos Manzanedo. Last revised: 10/05/2000 p.1 Procedure: The CHA evaporator is divided into 16 different panels. These panels are labeled on the machine

More information

LeRIBSS MTC MANUAL. Issue #1. March, MTC Control Unit Definitions, Information and Specifications. MTC Control Unit Electronic Schematics

LeRIBSS MTC MANUAL. Issue #1. March, MTC Control Unit Definitions, Information and Specifications. MTC Control Unit Electronic Schematics LeRIBSS MTC MANUAL Issue #1 March, 2008 Contents: MTC Control Unit MTC Control Unit Definitions, Information and Specifications Programming the MTC Control Unit Program Parameters Initial Setup Measuring

More information

Practicum 3, Fall 2010

Practicum 3, Fall 2010 A. F. Miller 2010 T1 Measurement 1 Practicum 3, Fall 2010 Measuring the longitudinal relaxation time: T1. Strychnine, dissolved CDCl3 The T1 is the characteristic time of relaxation of Z magnetization

More information

E X P E R I M E N T 1

E X P E R I M E N T 1 E X P E R I M E N T 1 Getting to Know Data Studio Produced by the Physics Staff at Collin College Copyright Collin College Physics Department. All Rights Reserved. University Physics, Exp 1: Getting to

More information

High QE Photocathodes lifetime and dark current investigation

High QE Photocathodes lifetime and dark current investigation High QE Photocathodes lifetime and dark current investigation Paolo Michelato INFN Milano - LASA Main Topics High QE photocathode lifetime QE vs. time (measurements on several cathodes, FLASH data) QE

More information

Characterization of Diamond Samples CHESS Run Summer 2013

Characterization of Diamond Samples CHESS Run Summer 2013 Characterization of Diamond Samples CHESS Run Summer 2013 5/22 5/28/2013 Characterization of Diamond Samples CHESS Run Summer 2013 Useful information Goals for this run Conditions of the experiment 1.

More information

Questa Rock Pile Stability Study SOP 26v5 Page 1 STANDARD OPERATING PROCEDURE NO. 26 ELECTRON MICROPROBE ANALYSIS REVISION LOG

Questa Rock Pile Stability Study SOP 26v5 Page 1 STANDARD OPERATING PROCEDURE NO. 26 ELECTRON MICROPROBE ANALYSIS REVISION LOG Questa Rock Pile Stability Study SOP 26v5 Page 1 STANDARD OPERATING PROCEDURE NO. 26 ELECTRON MICROPROBE ANALYSIS REVISION LOG Revision Number Description Date 26.0 Original SOP Oct. 20, 2003 26.1 McLemore

More information

INFN Milano LASA News on Cathode Development

INFN Milano LASA News on Cathode Development INFN Milano LASA News on Cathode Development PITZ Collaboration Meeting December 7 8, 2010 L. Monaco and D. Sertore, INFN Milano LASA Presented dby D. Sertore 1 Outline LASA preparation system status t

More information

POL-200 Semiautomatic Polarimeter. Instruction Manual BANTE INSTRUMENTS CO., LTD

POL-200 Semiautomatic Polarimeter. Instruction Manual BANTE INSTRUMENTS CO., LTD POL-200 Semiautomatic Polarimeter Instruction Manual BANTE INSTRUMENTS CO., LTD POL-200 Semiautomatic Polarimeter 1 Introduction Thank you for selecting the POL-200 semiautomatic polarimeter. This manual

More information

An Overview of Beam Diagnostic and Control Systems for AREAL Linac

An Overview of Beam Diagnostic and Control Systems for AREAL Linac An Overview of Beam Diagnostic and Control Systems for AREAL Linac Presenter G. Amatuni Ultrafast Beams and Applications 04-07 July 2017, CANDLE, Armenia Contents: 1. Current status of existing diagnostic

More information

PRACTICAL APPLICATION OF THE PHASED-ARRAY TECHNOLOGY WITH PAINT-BRUSH EVALUATION FOR SEAMLESS-TUBE TESTING

PRACTICAL APPLICATION OF THE PHASED-ARRAY TECHNOLOGY WITH PAINT-BRUSH EVALUATION FOR SEAMLESS-TUBE TESTING PRACTICAL APPLICATION OF THE PHASED-ARRAY TECHNOLOGY WITH PAINT-BRUSH EVALUATION FOR SEAMLESS-TUBE TESTING R.H. Pawelletz, E. Eufrasio, Vallourec & Mannesmann do Brazil, Belo Horizonte, Brazil; B. M. Bisiaux,

More information

Mini Micro Pulse Lidar System

Mini Micro Pulse Lidar System Mini Micro Pulse Lidar System MiniMPL-532-C Sensor Suite Operations Manual Version: June 2016 THIS PAGE INTENTIONALLY LEFT BLANK 2 Table of Contents MINIMPL SENSOR SUITE SYSTEM: RECORD OF PURCHASE... 4

More information

This guide gives a brief description of the ims4 functions, how to use this GUI and concludes with a number of examples.

This guide gives a brief description of the ims4 functions, how to use this GUI and concludes with a number of examples. Quick Start Guide: Isomet ims Studio Isomet ims Studio v1.40 is the first release of the Windows graphic user interface for the ims4- series of 4 channel synthezisers, build level rev A and rev B. This

More information

FEI FIB Focused Ion Beam

FEI FIB Focused Ion Beam Operating Manual Part 1 FEI FIB Focused Ion Beam IF IN DOUBT, ASK 1.00 IF ANYTHING UNUSUAL HAPPENS, OR IF THERE IS ANYTHING YOU ARE UNSURE ABOUT, STOP AND CONTACT ME! DO NOT PROCEED OR ATTEMPT TO FIX THE

More information

FACSAria I Standard Operation Protocol Basic Operation

FACSAria I Standard Operation Protocol Basic Operation FACSAria I Standard Operation Protocol Basic Operation 1. Checking Lasers Status a. Please check the ON / OFF of the lasers. Sufficient time (~30 minutes) need to be given to allow the laser(s) to warm

More information

Intelligent Pendulum Hardness Tester BEVS 1306 User Manual

Intelligent Pendulum Hardness Tester BEVS 1306 User Manual Intelligent Pendulum Hardness Tester BEVS 1306 User Manual Please read the user manual before operation. PAGE 1 Content 1. Company Profile... 3 2. Product Introduction... 3 3. Operation Instruction...

More information

Quick Start Guide. Multidimensional Imaging

Quick Start Guide. Multidimensional Imaging Quick Start Guide Multidimensional Imaging Printed 11/2012 Multidimensional Imaging Content Quick Start Guide Content 1 Introduction 4 2 Set up multi-channel experiments 5 2.1 Set up a new experiment

More information

Achieving More Efficient Data Review with OpenLAB CDS

Achieving More Efficient Data Review with OpenLAB CDS Achieving More Efficient Data Review with OpenLAB CDS Technical Note Introduction Separations have enjoyed major advances that have significantly reduced chromatographic run times. Ultra high performance

More information

Preparing for remote data collection at NE-CAT

Preparing for remote data collection at NE-CAT Preparing for remote data collection at NE-CAT Important Note: The beamtime and remote login privileges are intended just for you and your group. You are not allowed to share these with any other person

More information

Classroom Setup... 2 PC... 2 Document Camera... 3 DVD... 4 Auxiliary... 5

Classroom Setup... 2 PC... 2 Document Camera... 3 DVD... 4 Auxiliary... 5 Classroom Setup... 2 PC... 2 Document Camera... 3 DVD... 4 Auxiliary... 5 Lecture Capture Setup... 6 Pause and Resume... 6 Considerations... 6 Video Conferencing Setup... 7 Camera Control... 8 Preview

More information

Instructions for Use: Video Inspection Scope with Display

Instructions for Use: Video Inspection Scope with Display Instructions for Use: Video Inspection Scope with Display Brand Name of Product Video Inspection Scope Generic Name of Product Video Inspection Scope Product Code Number(s) FIS-004 Intended Use For visually

More information

Liquid Chromatography- Mass Spectrometer Manual

Liquid Chromatography- Mass Spectrometer Manual Liquid Chromatography- Mass Spectrometer Manual Joshua Willis, Elizabeth Sattely Department of Chemical Engineering Stanford University November 6, 2014 Abstract This manual will explain the LC/MS, its

More information

AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode)

AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode) AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode) 1. Log into the Log Usage system on the SMIF web site 2. Open Nanoscope 6.14r1 software by double clicking on the Nanoscope 6.14r1 desktop

More information

PHGN 480 Laser Physics Lab 4: HeNe resonator mode properties 1. Observation of higher-order modes:

PHGN 480 Laser Physics Lab 4: HeNe resonator mode properties 1. Observation of higher-order modes: PHGN 480 Laser Physics Lab 4: HeNe resonator mode properties Due Thursday, 2 Nov 2017 For this lab, you will explore the properties of the working HeNe laser. 1. Observation of higher-order modes: Realign

More information

with the Field-IQ Crop Input Control System

with the Field-IQ Crop Input Control System with the Field-IQ Crop Input Control System Quick Reference Card CONNECTING THE SYSTEM Ag25 GNSS antenna (P/N 68040-OOS) TNC/TNC right-angle cable (P/N 50449) Cable assembly, display to Field-IQ (P/N 50449)

More information

1.0 ThermoNicolet Nexus 670 FTIR Spectrometer Instructions

1.0 ThermoNicolet Nexus 670 FTIR Spectrometer Instructions 1.0 ThermoNicolet Nexus 670 FTIR Spectrometer Instructions 1.1 Click on the OMNIC icon to open the software. 1.2 Check that a signal is being measured by opening the Experiment Setup menu under the Collect

More information

Durham Magneto Optics Ltd. NanoMOKE 3 Wafer Mapper. Specifications

Durham Magneto Optics Ltd. NanoMOKE 3 Wafer Mapper. Specifications Durham Magneto Optics Ltd NanoMOKE 3 Wafer Mapper Specifications Overview The NanoMOKE 3 Wafer Mapper is an ultrahigh sensitivity Kerr effect magnetometer specially configured for measuring magnetic hysteresis

More information

Veeco Dektak 6M Profilometer

Veeco Dektak 6M Profilometer Veeco Dektak 6M Profilometer System Ranges/Resolutions Range (Å) Resolution (Å) 50 (5nm) to 65K 1 0.5K to 655K 10 2K to 2620K 40 8K to 10000K (1mm) 160 Maximum sample thickness: 31.75mm Scan range: 50

More information

1.5mm amplitude at 10 to 55Hz frequency in each X, Y, Z direction for 2 hours 500m/s² (approx. 50G) in each X, Y, Z direction for 3 times

1.5mm amplitude at 10 to 55Hz frequency in each X, Y, Z direction for 2 hours 500m/s² (approx. 50G) in each X, Y, Z direction for 3 times Color Mark Color Mark Feature Outstanding color matching accuracy - RGB light emitting diodes and 12-bit resolution - 2 detection modes (color only / color + intensity) - -step sensitivity adjustment for

More information

American DJ. Show Designer. Software Revision 2.08

American DJ. Show Designer. Software Revision 2.08 American DJ Show Designer Software Revision 2.08 American DJ 4295 Charter Street Los Angeles, CA 90058 USA E-mail: support@ameriandj.com Web: www.americandj.com OVERVIEW Show Designer is a new lighting

More information

Introducing Purion H, a Scanned Spot Beam High Current Ion Implanter

Introducing Purion H, a Scanned Spot Beam High Current Ion Implanter Introducing Purion H, a Scanned Spot Beam High Current Ion Implanter Bo Vanderberg, Patrick Heres, Edward Eisner, Bruce Libby, Joseph Valinski and Weston Huff Axcelis Technologies, Inc. 108 Cherry Hill

More information

LHC Beam Instrumentation Further Discussion

LHC Beam Instrumentation Further Discussion LHC Beam Instrumentation Further Discussion LHC Machine Advisory Committee 9 th December 2005 Rhodri Jones (CERN AB/BDI) Possible Discussion Topics Open Questions Tune measurement base band tune & 50Hz

More information

SIMET AVIKO D INSTRUCTION MANUAL SORTING Solutions, Ltd.

SIMET AVIKO D INSTRUCTION MANUAL SORTING Solutions, Ltd. SIMET AVIKO D INSTRUCTION MANUAL 1870 SORTING Solutions, Ltd. 1. TABLE OF CONTENTS 1. TABLE OF CONTENTS...1 2. INTRODUCTION...2 2.1. Application...2 2.2. Operating Conditions...2 2.3. Electro - Optical

More information

ISOMET. Compensation look-up-table (LUT) and How to Generate. Isomet: Contents:

ISOMET. Compensation look-up-table (LUT) and How to Generate. Isomet: Contents: Compensation look-up-table (LUT) and How to Generate Contents: Description Background theory Basic LUT pg 2 Creating a LUT pg 3 Using the LUT pg 7 Comment pg 9 The compensation look-up-table (LUT) contains

More information

Waters GCT GCMS Training Manual (Nominal Mass GCMS version) 02/06/2012 S.V.

Waters GCT GCMS Training Manual (Nominal Mass GCMS version) 02/06/2012 S.V. 1234 Hach Hall 515-294-5805 www.cif.iastate.edu Waters GCT GCMS Training Manual (Nominal Mass GCMS version) 02/06/2012 S.V. Location: Contact: 1238 Hach Hall Steve Veysey, 1234 Hach Hall; Kamel Harrata,

More information

HIGH CURRENT IMPLANTER 200KeV - SERIES 1090 TECHNICAL DESCRIPTION

HIGH CURRENT IMPLANTER 200KeV - SERIES 1090 TECHNICAL DESCRIPTION HIGH CURRENT IMPLANTER 200KeV - SERIES 1090 TECHNICAL DESCRIPTION Manufacturer: Danfysik A/S Møllehaven 31 DK-4040 Jyllinge Denmark Tel. +45 4679 0000 Fax +45 4679 0001 danfysik@danfysik.dk www.danfysik.com

More information

The only warranties for HP products and services are set forth in the express warranty statements accompanying such products and services.

The only warranties for HP products and services are set forth in the express warranty statements accompanying such products and services. The only warranties for HP products and services are set forth in the express warranty statements accompanying such products and services. Nothing herein should be construed as constituting an additional

More information

"CHARM Access and operational Procedure

CHARM Access and operational Procedure EDMS NO. REV. VALIDITY 1390490 0.1 DRAFT CERN CH1211 Geneva 23 Switzerland REFERENCE XXXX Date: 2015-09-04 OPERATING PROCEDURE "CHARM Access and operational Procedure ABSTRACT : The aim of this procedure

More information

F7000NV ROBOT VISION OPERATING MANUAL

F7000NV ROBOT VISION OPERATING MANUAL Rev. C Feb 2012 F7000NV ROBOT VISION OPERATING MANUAL Rev. C Feb 2012 This page has intentionally been left blank. Contents Contents Chapter 1. Getting Started... 5 1. Preface... 5 2. Manuals... 5 3. Setting

More information

How to Optimize Ad-Detective

How to Optimize Ad-Detective How to Optimize Ad-Detective Ad-Detective technology is based upon black level detection. There are several important criteria to consider: 1. Does the video have black frames to detect? Are there any

More information

Ford AMS Test Bench Operating Instructions

Ford AMS Test Bench Operating Instructions THE FORD METER BOX COMPANY, INC. ISO 9001:2008 10002505 AMS Test Bench 09/2013 Ford AMS Test Bench Operating Instructions The Ford Meter Box Co., Inc. 775 Manchester Avenue, P.O. Box 443, Wabash, Indiana,

More information

TL-2900 AMMONIA & NITRATE ANALYZER DUAL CHANNEL

TL-2900 AMMONIA & NITRATE ANALYZER DUAL CHANNEL TL-2900 AMMONIA & NITRATE ANALYZER DUAL CHANNEL DATA ACQUISITION SYSTEM V.15.4 INSTRUCTION MANUAL Timberline Instruments, LLC 1880 S. Flatiron Ct., Unit I Boulder, Colorado 80301 Ph: (303) 440-8779 Fx:

More information

Agilent High Capacity RV Pumps & Roots Pumping Systems

Agilent High Capacity RV Pumps & Roots Pumping Systems Agilent High Capacity RV Pumps & Roots Pumping Systems 2-3 MS-Series Rotary Vane Pumps 4-5 RPS-Series Roots Pumping Systems and RPK-Series Roots Pumping Kits 6-7 Typical Applications 8-15 Pump Models MS-Series

More information

Transient Stability Events & Actions

Transient Stability Events & Actions ETAP TIP No. 009 Transient Stability Events & Actions Applicable ETAP Versions: 5.5.0, 5.5.5, 5.5.6 (For lower versions, some of the descriptions and procedures below may differ in some ways) Event is

More information

CyAn User Manual. Flow & Mass Cytometry Center Biomedical Core facility. Technion Faculty of Medicine

CyAn User Manual. Flow & Mass Cytometry Center Biomedical Core facility. Technion Faculty of Medicine CyAn User Manual Flow & Mass Cytometry Center Biomedical Core facility For assistant please call: Technion Faculty of Medicine Amir Grau 054-210-0778 Yaakov Sakoury 050-749-8586 Ofer Shenker 052-456-8846

More information

CESR BPM System Calibration

CESR BPM System Calibration CESR BPM System Calibration Joseph Burrell Mechanical Engineering, WSU, Detroit, MI, 48202 (Dated: August 11, 2006) The Cornell Electron Storage Ring(CESR) uses beam position monitors (BPM) to determine

More information

Instruction manual. KUZMA 4POINT 14 inch TONEARM Serial Number:

Instruction manual. KUZMA 4POINT 14 inch TONEARM Serial Number: Instruction manual KUZMA 4POINT 14 inch TONEARM Serial Number:.. 2016-09 1 KUZMA LTD INSTRUCTION MANUAL FOR 4POINT 14 tonearm The 4POINT 14 tonearm is a very precisely engineered piece of equipment, however,

More information

013-RD

013-RD Engineering Note Topic: Product Affected: JAZ-PX Lamp Module Jaz Date Issued: 08/27/2010 Description The Jaz PX lamp is a pulsed, short arc xenon lamp for UV-VIS applications such as absorbance, bioreflectance,

More information

High Performance DL-60 (Gold Plus) (7 in - 13 in) Dual Lane Spliceable Tape Feeder Part Number: Revision 3 Sep No.

High Performance DL-60 (Gold Plus) (7 in - 13 in) Dual Lane Spliceable Tape Feeder Part Number: Revision 3 Sep No. 8mm High Performance DL-60 (Gold Plus) (7 in - 13 in) Dual Lane Spliceable Tape Feeder Part Number: 50381212 Revision 3 Sep. 2010 No. 0730D-E043 Page i Table of Contents Functional Description...1 Procedures

More information