Implementing a Proton Beam Scanning System within an Operating Clinical Facility

Size: px
Start display at page:

Download "Implementing a Proton Beam Scanning System within an Operating Clinical Facility"

Transcription

1 Implementing a Proton Beam Scanning System within an Operating Clinical Facility Ben Clasie Many thanks to Hassan Bentefour, Hanne Kooy, and Jay Flanz for their help preparing this presentation 1

2 Francis H. Burr Proton Therapy Center 235 MeV Gantry 1 Gantry 2 DS DS PBS PBS Stereotactic Eye Room 3a Cyclotron manufactured by Ion Beam Applications 2

3 Delivery techniques in Gantry 2 Passive Scattering Broad beam Apertures and range compensators protect healthy tissue and conform dose to the target Pencil Beam Scanning A narrow pencil beam is scanned over the target Apertures and range compensators are optional Universal proton therapy nozzle interchangeable between DS and PBS 3

4 Nozzle in Double Scattering (DS) mode Ion chambers Proton beam Ti window Fixed scatterer Dipole magnets Light-field mirror X-ray box Range-modulator Second Ion chambers wheel scatterer Collimating jaws Snout p D 4

5 Nozzle in Pencil Beam Scanning (PBS) mode 5

6 Treatment room (DS equipment) Gantry * Isocenter First scatterer & Range modulator Ionization chambers Second scatterer 6

7 Treatment room (PBS equipment) Gantry Quadrupole and scanning magnets Ionization chambers * Isocenter 7

8 8

9 In an ideal world. 9

10 Changing modes from DS to PBS Proton beam Range modulator + fixed scatterer Ion. chamber Nozzle quadrupole magnets (optional) 10

11 PBS checklist Date: Initials: Install Clean-up Item Retract RM drawer Change IC cables, HV cables Power to Pyramid devices in the gantry Power to A500 s Power to SMPS Switch B1G2 power supply card to PBS mode Turn DCEU key Install snout Change BeamDeliveryMode and restart TCU2 11

12 General rules for embarking on PBS development within an operating clinical facility 12

13 Rule #1 PBS development can not interrupt normal treatments Example #1 Work is scheduled outside of normal hours Morning Afternoon Evening 6:30-7:30am 6pm-7pm 7pm-8pm 8pm-10pm 10pm-Mid Overnight Mid-2am 2am-4am QA Tx Tx Scanning Ben et. al. Example #2 All work is set up and packed up on each shift without damaging the system Frequent: Time for setting up and packing up Verify the system is operational before morning QA Infrequent: Training Example #3 Upgrades to software and hardware do not affect normal treatments Less Frequent: Design test procedures Perform tests Documentation Compromises: Not all PBS upgrades are compatible with double scattering (Eg extra vacuum chamber in the nozzle) 13

14 Rule #2 Optimize time with the beam Example #1 Priorities, files and devices are prepared before shifts Example #2 When a project is finished at 2am, we start a new project Example #3 Data analysis and reports are prepared in parallel with data collecting, or after shifts If there is time left over, there are also meetings, research projects, , and sleep 14

15 Scanning techniques 15

16 Scanning techniques PBS involves moving a charged particle beam of particular properties and/or changing one or more of the properties of the beam For example: Spot scanning Position is stationary when beam is on. Beam is off between spots. Raster scanning Position is moved from spot to spot but the beam is always on (during and between spots) Continuous scanning Beam is always moving and the beam current is modulated from 0 to 100% Velocity can be modulated Jay Flanz, Beige paper, 2007 Time-driven mode Commands to move the position and change the beam current is synchronized with time Charge-driven mode Commands to move and turn the beam on/off is synchronized with charge collected on the monitor chamber Many names, but still just PBS For example, Spot scanning in the limit of many mini-spots becomes continuous scanning The best choice of operating conditions depends on the condition of the accelerator 16

17 The Pros and Cons of each Regime Spot scanning In addition to the beam-on time, more time is needed for moving the beam and waiting for magnets to settle Simple model of dose deposition Raster scanning Avoids some time delays Needs a model for the dose deposited between spots which will have an uncertainty Continuous scanning May still need time delays to wait for magnets to settle before turning the beam on Safety and clinical requirements may limit beam current and/or scanning speed, especially in fields with large doses Repainting is common Time-driven mode Can be faster than charge-driven mode Requires stable and accurate beam current Charge-driven mode Can be more accurate than time-driven mode There is a small time delay between collecting the appropriate charge and turning the beam off 17

18 Repainting, N times Random Error Systematic Error Constant 1/sqrt(N) N Proportional 1/[N*sqrt(N)] Even Effect of different spot weight errors on the dose distribution uncertainty Repainting may or may not deliver a more accurate dose distribution, it depends on the system and the dominant type of uncertainty Furthermore, repainting usually increases the irradiation time (it may be possible to break even in some system configurations) One can use repainting to reduce interplay effects with organ motion. This benefit is negated, however, if layer-by-layer repainting is done too quickly 18

19 Scanning MGH 19

20 PBS System Schematic Scanning controller, Pyramid Technical Consultants PTC Scanning magnet power supply, JEMA Proton beam Proton source, IBA Beam transport system Ionization chamber Scanning magnets (nozzle) Ionization chambers (nozzle) * Isocenter Hall probes, PTC Electrometers, PTC 20

21 An example of spot scanning and continuous scanning in a single map 21

22 An example of spot scanning and continuous scanning in a single map 22

23 Tolerances in PBS 23

24 Beam Performance requirements for PBS Version 1 (no quads) Requirements for an endometrium tumor Random (peak-topeak) Layer-tolayer (pk-to-pk) Tol: Gamma Index / Dose Prostate Range (mm) ±1.5 ± 1.5 Position ± 20% of σ ± 20% of σ H&N Beam width (%) ± 25 ± 25 Spot weight (%) ± 10 ± 5 Trofimov, Hubeau, and Flanz 24

25 MGH beam performance for the spot scanning technique 25

26 Nozzle and Beam Transport System Gantry magnets p Beam line magnets affect the pencil beam position and width at isocenter Magnets are cycled to reduce hysteresis affects. Proton range is always decreasing or we cycle the magnets Some power supplies overshot requested currents I Some power supplies turn off for safety reasons and have a long settling time when they turn back on t 26

27 Beam Position Performance Our requirement for beam position accuracy is 20% of sigma (2mm) Beamline algorithms are: (1) Set beamline magnets and verify the beam position during irradiations (dead-reckoning), (2) Tune the beamline magnets using position feedback at the first energy layer and use this information to correct subsequent layers, (3) use online position feedback to correct the position when required, and/or (4) Tune beamline magnets at every layer using position feedback. We use a combination of (1), (2) & (3) Horizontal Vertical Position Y accuracy at accuracy isocenter [mm] at X accuracy isocenter at [mm] Range [g/cm 2 ] 270 deg 0 deg 90 deg 180 deg 270 deg 0 deg 90 deg 180 deg 27

28 Pencil Beam Size Performance Beam width, 1 sigma [mm] Horizontal Proton Vertical Range [g/cm2] Proton Range [g/cm2] G270 G000 G090 G180 G315 G045 G135 FIT +/- 15% G270 G000 G090 G180 G315 G045 G135 FIT +/- 10% Observed stability in the beam width is better than 15% Beam Size at Isocenter (1 sigma) without quads is about 13mm. Beam widths (1 sigma) of 6 mm have been achieved by installing a quadrupole doublet and a short vacuum chamber in the nozzle and adjusting the magnet settings in the beam line. This will be implemented for treatments in Phase 2. 28

29 Charge delivered Accuracy Timing and Sensitivity Possible techniques for controlling the spot weight are (1) time driven, (2) charge driven, (3) use feedback from the charge rate to control the beam We choose to use chargedriven mode with charge rate feedback for our first PBS version This choice allows us to meet our requirements for our system with two caveats: accuracy (Gp) Spot weight 4.E-04 2.E-04 0.E+00-2.E-04-4.E Proton current (Gp/s) 1. Lower spot weights require lower proton currents 2. The best spot weight accuracy is 0.8x10-4 Gigaprotons which determines the smallest deliverable spot weight 29

30 Time to Achieve desired Range Ra ange in MLIC - target Range [g/cm2] g/cm2 20 g/cm2 7 g/cm Time from transition [s] The dipole magnets of the energy selection system are set without iterating with a field probe. The field probe verifies the magnets are set correctly and will stop an irradiation if they are outside the allowable tolerance Further improvements to reduce the transition time are planned 30

31 Irradiation time for 2Gy/1L Number of Layers 15 Total number of spots 1,500 Total time changing energy (3 s/layer) Total beam-on time Additional time moving Additional time settling Additional time from total delay at end of spots Sum of all contributions 42 s * 20 s 1.7 s 8.3 s 1.5 s 74 s *Will be upgraded to 1.5 s/layer See more results in a forthcoming paper by Stephen Dowdell and Anatoly Rozenfeld in collaboration with MGH 31

32 What is proton PBS? Summary Arbitrary scanning patterns in (X, Y, Z=E, Q, I, σ, ) PBS offers distal and proximal dose conformation At MGH, PBS is being developed in a Universal nozzle while treating patients with DS Two Rules were described 1. PBS development can not interrupt normal treatments 2. Optimize time with the beam which adds complexity and inconvenience to the development of PBS Further refinements to our PBS system are planned 32

33 Thank you! 33

Promises and Perils of Proton Therapy Beam Delivery (Implications) or Towards Cost Effective Particle Therapy

Promises and Perils of Proton Therapy Beam Delivery (Implications) or Towards Cost Effective Particle Therapy Promises and Perils of Proton Therapy Beam Delivery (Implications) or Towards Cost Effective Particle Therapy Jay Flanz MGH/FBTC Harvard Medical School What is a Beam Delivery? Start with an accelerator

More information

DOSE DELIVERY SYSTEM OF THE VARIAN PROBEAM SYSTEM WITH CONTINUOUS BEAM

DOSE DELIVERY SYSTEM OF THE VARIAN PROBEAM SYSTEM WITH CONTINUOUS BEAM DOSE DELIVERY SYSTEM OF THE VARIAN PROBEAM SYSTEM WITH CONTINUOUS BEAM EUCARD 2 WORKSHOP ON INNOVATIVE DELIVERY SYSTEMS IN PARTICLE THERAPY TORINO, 23 25 FEB 2017 VARIAN PARTICLE THERAPY HOLGER GÖBEL MANGER

More information

Quality Assurance Implementation at the Roberts Proton Therapy Center. James McDonough 3 August 2013

Quality Assurance Implementation at the Roberts Proton Therapy Center. James McDonough 3 August 2013 Quality Assurance Implementation at the Roberts Proton Therapy Center James McDonough 3 August 2013 1 Roberts Proton Therapy Center Machine configuration and layout 4 gantries, 1 fixed beam line, 1 research

More information

National Institute of Radiological Sciences. Naoya Saotome

National Institute of Radiological Sciences. Naoya Saotome National Institute of Radiological Sciences Naoya Saotome 1 Contents Introduction History and collaboration KCC i-rock Commissioning of commercial scanning system NIRS Gantry Commissioning of NIRS s Gantry

More information

PBS Products from Pyramid

PBS Products from Pyramid PBS Products from Pyramid Pyramid was founded in 1986 and is an established supplier of instrument control systems for the medical, semiconductor, physics and biological research markets. The systems typically

More information

A NOVEL GANTRY FOR PROTON THERAPY AT THE PAUL SCHERRER INSTITUTE

A NOVEL GANTRY FOR PROTON THERAPY AT THE PAUL SCHERRER INSTITUTE A NOVEL GANTRY FOR PROTON THERAPY AT THE PAUL SCHERRER INSTITUTE E. Pedroni, T. Böhringer, A. Coray, G. Goitein, M. Grossmann, A. Lomax, S. Lin, M. Jermann Paul Scherrer Institute, CH-5232 Villigen PSI,

More information

G. Pittá(*), S. Braccini TERA Foundation, Novara, Italy (*) Corresponding author.

G. Pittá(*), S. Braccini TERA Foundation, Novara, Italy (*) Corresponding author. Frascati Physics Series Vol. VVVVVV (xxxx), pp. 000-000 XX Conference Location, Date-start - Date-end, Year MATRIX: AN INNOVATIVE PIXEL IONIZATION CHAMBER FOR ON-LINE BEAM MONITORING IN HADRONTHERAPY G.

More information

Particle Beam Production - A Synchrotron-Based System - Prof. Dr. Thomas Haberer Scientific-technical Director Heidelberg Iontherapy Center

Particle Beam Production - A Synchrotron-Based System - Prof. Dr. Thomas Haberer Scientific-technical Director Heidelberg Iontherapy Center Particle Beam Production - A Synchrotron-Based System - Prof. Dr. Thomas Haberer Scientific-technical Director Heidelberg Iontherapy Center Outline Situation/Rationale Requirements Synchrotron choice Functions

More information

Therapy Control and Patient Safety for Proton Therapy

Therapy Control and Patient Safety for Proton Therapy Proceedings of the CAS-CERN Accelerator School: Accelerators for Medical Applications, Vösendorf, Austria, 26 May 5 June 2015, edited by R. Bailey, CERN Yellow Reports: School Proceedings, Vol. 1/2017,

More information

Recent developments in cyclotrons for proton therapy at IBA

Recent developments in cyclotrons for proton therapy at IBA Recent developments in cyclotrons for proton therapy at IBA Yves Jongen. Founder & CRO IBA sa We Protect, Enhance and Save Lives. A typical PT center 30-55 millions for equipment 45-100 millions for the

More information

Development of New Carbon Therapy Facility and Future Plan of HIMAC

Development of New Carbon Therapy Facility and Future Plan of HIMAC Development of New Carbon Therapy Facility and Future Plan of Koji Noda Research Center for Charged Particle Therapy National Institute of Radiological Sciences 1. Introduction Contents 2. New Carbon-Therapy

More information

DICOM Correction Proposal

DICOM Correction Proposal DICOM Correction Proposal STATUS Final Text Date of Last Update 2015/11/16 Person Assigned Submitter Name Ulrich Busch (ulrich.busch@varian.com) Michael Moyers Submission Date 2011/05/19 Correction Number

More information

The PEFP 20-MeV Proton Linear Accelerator

The PEFP 20-MeV Proton Linear Accelerator Journal of the Korean Physical Society, Vol. 52, No. 3, March 2008, pp. 721726 Review Articles The PEFP 20-MeV Proton Linear Accelerator Y. S. Cho, H. J. Kwon, J. H. Jang, H. S. Kim, K. T. Seol, D. I.

More information

In PS 3.3, Section C RT Ion Beams Session Record Module, add the following attributes and make the changes indicated:

In PS 3.3, Section C RT Ion Beams Session Record Module, add the following attributes and make the changes indicated: DICOM Correction Proposal STATUS Letter Ballot Date of Last Update 2016/03/18 Person Assigned Ulrich Busch Submitter Name Olivier Vierlinck Submission Date 2015/02/03 Correction Number CP-1013 Log Summary:

More information

III. Proton-therapytherapy. Rome SB - 3/5 1

III. Proton-therapytherapy. Rome SB - 3/5 1 Outline Introduction: an historical review I Applications in medical diagnostics Particle accelerators for medicine Applications in conventional radiation therapy II III IV Hadrontherapy, the frontier

More information

New Filling Pattern for SLS-FEMTO

New Filling Pattern for SLS-FEMTO SLS-TME-TA-2009-0317 July 14, 2009 New Filling Pattern for SLS-FEMTO Natalia Prado de Abreu, Paul Beaud, Gerhard Ingold and Andreas Streun Paul Scherrer Institut, CH-5232 Villigen PSI, Switzerland A new

More information

SPEAR 3: Operations Update and Impact of Top-Off Injection

SPEAR 3: Operations Update and Impact of Top-Off Injection SPEAR 3: Operations Update and Impact of Top-Off Injection R. Hettel for the SSRL ASD 2005 SSRL Users Meeting October 18, 2005 SPEAR 3 Operations Update and Development Plans Highlights of 2005 SPEAR 3

More information

THE IBA SUPERCONDUCTING SYNCHROCYCLOTRON PROJECT S2C2

THE IBA SUPERCONDUCTING SYNCHROCYCLOTRON PROJECT S2C2 THE IBA SUPERCONDUCTING SYNCHROCYCLOTRON PROJECT S2C2 W. Kleeven, M. Abs, E. Forton, S. Henrotin, Y. Jongen, V. Nuttens, Y. Paradis, E. Pearson, S. Quets, J. Van de Walle, P. Verbruggen, S. Zaremba, IBA,

More information

Commissioning of Accelerators. Dr. Marc Munoz (with the help of R. Miyamoto, C. Plostinar and M. Eshraqi)

Commissioning of Accelerators. Dr. Marc Munoz (with the help of R. Miyamoto, C. Plostinar and M. Eshraqi) Commissioning of Accelerators Dr. Marc Munoz (with the help of R. Miyamoto, C. Plostinar and M. Eshraqi) www.europeanspallationsource.se 6 July, 2017 Contents General points Definition of Commissioning

More information

An Overview of Beam Diagnostic and Control Systems for AREAL Linac

An Overview of Beam Diagnostic and Control Systems for AREAL Linac An Overview of Beam Diagnostic and Control Systems for AREAL Linac Presenter G. Amatuni Ultrafast Beams and Applications 04-07 July 2017, CANDLE, Armenia Contents: 1. Current status of existing diagnostic

More information

Screen investigations for low energetic electron beams at PITZ

Screen investigations for low energetic electron beams at PITZ 1 Screen investigations for low energetic electron beams at PITZ S. Rimjaem, J. Bähr, H.J. Grabosch, M. Groß Contents Review of PITZ setup Screens and beam profile monitors at PITZ Test results Summary

More information

Development of Continuous Line Scanning System Prototype for Proton Beam Therapy

Development of Continuous Line Scanning System Prototype for Proton Beam Therapy Development of Continuous Line Scanning System Prototype for Proton Beam Therapy Ryosuke Kohno, PhD 1 ; Kenji Hotta, PhD 1 ; Takeshi Dohmae, PhD 1 ; Yuka Matsuzaki, PhD 1 ; Teiji Nishio, PhD 1 ; Tetsuo

More information

FIRST SIMULTANEOUS TOP-UP OPERATION OF THREE DIFFERENT RINGS IN KEK INJECTOR LINAC

FIRST SIMULTANEOUS TOP-UP OPERATION OF THREE DIFFERENT RINGS IN KEK INJECTOR LINAC FIRST SIMULTANEOUS TOP-UP OPERATION OF THREE DIFFERENT RINGS IN KEK INJECTOR LINAC M. Satoh #, for the IUC * Accelerator Laboratory, High Energy Accelerator Research Organization (KEK) 1-1 Oho, Tsukuba,

More information

Test beam data analysis for the CMS CASTOR calorimeter at the LHC

Test beam data analysis for the CMS CASTOR calorimeter at the LHC 1/ 24 DESY Summerstudent programme 2008 - Course review Test beam data analysis for the CMS CASTOR calorimeter at the LHC Agni Bethani a, Andrea Knue b a Technical University of Athens b Georg-August University

More information

Proton Engineering Frontier Project

Proton Engineering Frontier Project Proton Engineering Frontier Project OECD Nuclear Energy Agency Fifth International Workshop on the Utilisation and Reliability of High Power Proton Accelerators (HPPA5) (6-9 May 2007, Mol, Belgium) Yong-Sub

More information

LIGHT PROTON THERAPY PROJECT

LIGHT PROTON THERAPY PROJECT 17 th of MAY 2018 LIGHT PROTON THERAPY PROJECT Yevgeniy Ivanisenko on behalf of ADAM team FORM-01040-A AVO-ADAM Advanced Oncotherapy (AVO) is a public company ADAM is R&D center of AVO ~ 100 employees

More information

LHC Beam Instrumentation Further Discussion

LHC Beam Instrumentation Further Discussion LHC Beam Instrumentation Further Discussion LHC Machine Advisory Committee 9 th December 2005 Rhodri Jones (CERN AB/BDI) Possible Discussion Topics Open Questions Tune measurement base band tune & 50Hz

More information

beam dump from P2 losses this morning

beam dump from P2 losses this morning beam dump from P2 losses this morning Some observations on the beam dump from P2 losses this morning 29.10.10 at 01:26:39: - single bunch intensity (average) was ~1.3e11 - significantly higher than previous

More information

Particle Therapy with the Varian / ACCEL 250 MeV S.C. Proton Cyclotron

Particle Therapy with the Varian / ACCEL 250 MeV S.C. Proton Cyclotron Particle Therapy with the Varian / ACCEL 250 MeV S.C. Proton Cyclotron 1st Workshop HADRON BEAM THERAPY OF CANCER ERICE SICILY, 24 APRIL - 1 MAY 2009 D. W. Krischel, A.E. Geisler, J.H. Timmer, Volker Schirrmeister

More information

Brief Description of Circuit Functions. The brief ckt. description of V20 107E5 17 Monitor

Brief Description of Circuit Functions. The brief ckt. description of V20 107E5 17 Monitor Exhibit 4 Brief Description of Circuit Functions The brief ckt. description of V20 107E5 17 Monitor 0. Functional Block Diagram 1. General Description 2. Description of Circuit Diagram A. Power Supply

More information

PRESENT STATUS OF J-PARC

PRESENT STATUS OF J-PARC PRESENT STATUS OF J-PARC # F. Naito, KEK, Tsukuba, Japan Abstract Japan Proton Accelerator Research Complex (J-PARC) is the scientific facility with the high-intensity proton accelerator aiming to realize

More information

P. Emma, et al. LCLS Operations Lectures

P. Emma, et al. LCLS Operations Lectures P. Emma, et al. LCLS Operations Lectures LCLS 1 LCLS Accelerator Schematic 6 MeV 135 MeV 250 MeV σ z 0.83 mm σ z 0.83 mm σ z 0.19 mm σ δ 0.05 % σ δ 0.10 % σ δ 1.6 % Linac-0 L =6 m rf gun L0-a,b Linac-1

More information

INSTALLATION STATUS OF THE ELECTRON BEAM PROFILER FOR THE FERMILAB MAIN INJECTOR*

INSTALLATION STATUS OF THE ELECTRON BEAM PROFILER FOR THE FERMILAB MAIN INJECTOR* TUPB77 INSTALLATION STATUS OF THE ELECTRON BEAM PROFILER FOR THE FERMILAB MAIN INJECTOR* R. Thurman-Keup #, M. Alvarez, J. Fitzgerald, C. Lundberg, P. Prieto, M. Roberts, J. Zagel, FNAL, Batavia, IL 651,

More information

Shifty Manual v1.00. Shifty. Voice Allocator / Hocketing Controller / Analog Shift Register

Shifty Manual v1.00. Shifty. Voice Allocator / Hocketing Controller / Analog Shift Register Shifty Manual v1.00 Shifty Voice Allocator / Hocketing Controller / Analog Shift Register Table of Contents Table of Contents Overview Features Installation Before Your Start Installing Your Module Front

More information

Linatron - M9 & M9A. Modular high-energy X-ray source. 2.0 Performance

Linatron - M9 & M9A. Modular high-energy X-ray source. 2.0 Performance The Linatron -M is a modular system. The control console, modulator, and RF unit are common to all model configurations. Only the X-ray head changes to match the application. The Linatron - M is designed

More information

ABORT DIAGNOSTICS AND ANALYSIS DURING KEKB OPERATION

ABORT DIAGNOSTICS AND ANALYSIS DURING KEKB OPERATION ABORT DIAGNOSTICS AND ANALYSIS DURING KEKB OPERATION H. Ikeda*, J. W. Flanagan, T. Furuya, M. Tobiyama, KEK, Tsukuba, Japan M. Tanaka, MELCO SC,Tsukuba, Japan Abstract KEKB has stopped since June 2010

More information

An extreme high resolution Timing Counter for the MEG Upgrade

An extreme high resolution Timing Counter for the MEG Upgrade An extreme high resolution Timing Counter for the MEG Upgrade M. De Gerone INFN Genova on behalf of the MEG collaboration 13th Topical Seminar on Innovative Particle and Radiation Detectors Siena, Oct.

More information

CHAPTER 4: HIGH ENERGY X-RAY GENERATORS: LINEAR ACCELERATORS. Jason Matney, MS, PhD

CHAPTER 4: HIGH ENERGY X-RAY GENERATORS: LINEAR ACCELERATORS. Jason Matney, MS, PhD CHAPTER 4: HIGH ENERGY X-RAY GENERATORS: LINEAR ACCELERATORS Jason Matney, MS, PhD Objectives Medical electron linear accelerators (often shortened to LINAC) The Basics Power Supply Magnetron/Klystron

More information

Basic rules for the design of RF Controls in High Intensity Proton Linacs. Particularities of proton linacs wrt electron linacs

Basic rules for the design of RF Controls in High Intensity Proton Linacs. Particularities of proton linacs wrt electron linacs Basic rules Basic rules for the design of RF Controls in High Intensity Proton Linacs Particularities of proton linacs wrt electron linacs Non-zero synchronous phase needs reactive beam-loading compensation

More information

Short Introduction to the use of the H6 beam

Short Introduction to the use of the H6 beam SL/EA/KE/ Version 3.0, 2 May 2000 Short Introduction to the use of the H6 beam updated version for 2000: In order to facilitate the search in the H6 zones, the upstream part of the old zone 146 has been

More information

Electrical and Electronic Laboratory Faculty of Engineering Chulalongkorn University. Cathode-Ray Oscilloscope (CRO)

Electrical and Electronic Laboratory Faculty of Engineering Chulalongkorn University. Cathode-Ray Oscilloscope (CRO) 2141274 Electrical and Electronic Laboratory Faculty of Engineering Chulalongkorn University Cathode-Ray Oscilloscope (CRO) Objectives You will be able to use an oscilloscope to measure voltage, frequency

More information

Advanced Photon Source - Upgrades and Improvements

Advanced Photon Source - Upgrades and Improvements Advanced Photon Source - Upgrades and Improvements Horst W. Friedsam, Jaromir M. Penicka Argonne National Laboratory, Argonne, Illinois, USA 1. INTRODUCTION The APS has been operational since 1995. Recently

More information

Precision measurements of beam current, position and phase for an e+e- linear collider

Precision measurements of beam current, position and phase for an e+e- linear collider Precision measurements of beam current, position and phase for an e+e- linear collider R. Corsini on behalf of H. Braun, M. Gasior, S. Livesley, P. Odier, J. Sladen, L. Soby INTRODUCTION Commissioning

More information

30 GHz Power Production / Beam Line

30 GHz Power Production / Beam Line 30 GHz Power Production / Beam Line Motivation & Requirements Layout Power mode operation vs. nominal parameters Beam optics Achieved performance Problems Beam phase switch for 30 GHz pulse compression

More information

OPERATIONAL EXPERIENCE AT J-PARC

OPERATIONAL EXPERIENCE AT J-PARC OPERATIONAL EXPERIENCE AT J-PARC Hideaki Hotchi, ) for J-PARC commissioning team ), 2), ) Japan Atomic Energy Agency (JAEA), Tokai, Naka, Ibaraki, 39-95 Japan, 2) High Energy Accelerator Research Organization

More information

STATUS AND CONCEPTUAL DESIGN OF THE CONTROL SYSTEM FOR THE HEAVY ION THERAPY ACCELERATOR FACILITY HICAT

STATUS AND CONCEPTUAL DESIGN OF THE CONTROL SYSTEM FOR THE HEAVY ION THERAPY ACCELERATOR FACILITY HICAT 10th ICALEPCS Int. Conf. on Accelerator & Large Expt. Physics Control Systems. Geneva, 10-14 Oct 2005, PO1.025-1 (2005) STATUS AND CONCEPTUAL DESIGN OF THE CONTROL SYSTEM FOR THE HEAVY ION THERAPY ACCELERATOR

More information

High Power Cyclotrons

High Power Cyclotrons Accelerator Reliability Workshop ESRF, Grenoble, 04-06.02.02 High Power Cyclotrons P.A.Schmelzbach 1. The PSI Proton Accelerator Facility 2. Failure Analysis 2000/2001 3. The weak Points 4. How to improve

More information

Linac 4 Instrumentation K.Hanke CERN

Linac 4 Instrumentation K.Hanke CERN Linac 4 Instrumentation K.Hanke CERN CERN Linac 4 PS2 (2016?) SPL (2015?) Linac4 (2012) Linac4 will first inject into the PSB and then can be the first element of a new LHC injector chain. It will increase

More information

Transmissive XBPM developments at PSF/BESSY. Martin R. Fuchs

Transmissive XBPM developments at PSF/BESSY. Martin R. Fuchs Transmissive XBPM developments at PSF/BESSY Martin R. Fuchs Acknowledgments PSF Martin Fieber-Erdmann Ronald Förster Uwe Müller BESSY Karsten Blümer Karsten Holldack Gerd Reichardt Franz Schäfers BIOXHIT,

More information

Digital BPMs and Orbit Feedback Systems

Digital BPMs and Orbit Feedback Systems Digital BPMs and Orbit Feedback Systems, M. Böge, M. Dehler, B. Keil, P. Pollet, V. Schlott Outline stability requirements at SLS storage ring digital beam position monitors (DBPM) SLS global fast orbit

More information

arxiv: v1 [physics.ins-det] 1 Nov 2015

arxiv: v1 [physics.ins-det] 1 Nov 2015 DPF2015-288 November 3, 2015 The CMS Beam Halo Monitor Detector System arxiv:1511.00264v1 [physics.ins-det] 1 Nov 2015 Kelly Stifter On behalf of the CMS collaboration University of Minnesota, Minneapolis,

More information

The ESS Accelerator. For Norwegian Industry and Research. Oslo, 24 Sept Håkan Danared Deputy Head Accelerator Division Group Leader Beam Physics

The ESS Accelerator. For Norwegian Industry and Research. Oslo, 24 Sept Håkan Danared Deputy Head Accelerator Division Group Leader Beam Physics The ESS Accelerator For Norwegian Industry and Research Oslo, 24 Sept 2013 Håkan Danared Deputy Head Accelerator Division Group Leader Beam Physics The Hadron Intensity Frontier Courtesy of M. Seidel (PSI)

More information

Simple motion control implementation

Simple motion control implementation Simple motion control implementation with Omron PLC SCOPE In todays challenging economical environment and highly competitive global market, manufacturers need to get the most of their automation equipment

More information

2 Work Package and Work Unit descriptions. 2.8 WP8: RF Systems (R. Ruber, Uppsala)

2 Work Package and Work Unit descriptions. 2.8 WP8: RF Systems (R. Ruber, Uppsala) 2 Work Package and Work Unit descriptions 2.8 WP8: RF Systems (R. Ruber, Uppsala) The RF systems work package (WP) addresses the design and development of the RF power generation, control and distribution

More information

Andrei Seryi, Toshiaki Tauchi. December 15-18, 2008

Andrei Seryi, Toshiaki Tauchi. December 15-18, 2008 ATF2 milestones for discussion Andrei Seryi, Toshiaki Tauchi December 15-18, 2008 7th ATF2 Project Meeting What are natural milestones for ATF2? ATF2 design: Nominal IP β y* =0.1 mm & L * =1 m this give

More information

Hadron Therapy Technologies

Hadron Therapy Technologies Hadron Therapy Technologies S. Peggs, BNL & ESS-S Bevalac 1950-1993 Many figures courtesy of Jay Flanz 1 Consumer demand 1 in 3 Europeans will confront some form of cancer in their lifetime. Cancer is

More information

Characterization and improvement of unpatterned wafer defect review on SEMs

Characterization and improvement of unpatterned wafer defect review on SEMs Characterization and improvement of unpatterned wafer defect review on SEMs Alan S. Parkes *, Zane Marek ** JEOL USA, Inc. 11 Dearborn Road, Peabody, MA 01960 ABSTRACT Defect Scatter Analysis (DSA) provides

More information

STATUS OF THE SWISSFEL C-BAND LINEAR ACCELERATOR

STATUS OF THE SWISSFEL C-BAND LINEAR ACCELERATOR Proceedings of FEL213, New York, NY, USA STATUS OF THE SWISSFEL C-BAND LINEAR ACCELERATOR F. Loehl, J. Alex, H. Blumer, M. Bopp, H. Braun, A. Citterio, U. Ellenberger, H. Fitze, H. Joehri, T. Kleeb, L.

More information

Preparations for Installation, Testing and Commissioning based on Experience at CERN, SNS and Siemens

Preparations for Installation, Testing and Commissioning based on Experience at CERN, SNS and Siemens Preparations for Installation, Testing and Commissioning based on Experience at CERN, SNS and Siemens Eugène Tanke FRIB / MSU ESS Seminar, Lund, 6 March 2013 Outline Project Goal for the Accelerator Path

More information

A Facility for Accelerator Physics and Test Beam Experiments

A Facility for Accelerator Physics and Test Beam Experiments A Facility for Accelerator Physics and Test Beam Experiments U.S. Department of Energy Review Roger Erickson for the FACET Design Team February 20, 2008 SLAC Overview with FACET FACET consists of four

More information

Ion Beam Analyis on the Upgraded CSIRO Nuclear Microprobe

Ion Beam Analyis on the Upgraded CSIRO Nuclear Microprobe Ion Beam Analyis on the Upgraded CSIRO Nuclear Microprobe Jamie S Laird, Chris G Ryan, Robin Kirkham, Peter Siddons, David Parry, Takahiro Satoh, Stephen Gregory and Roland Szymanski MINERAL RESOURCE FLAGSHIP

More information

Figure 2: components reduce board area by 57% over 0201 components, which themselves reduced board area by 66% over 0402 types (source Murata).

Figure 2: components reduce board area by 57% over 0201 components, which themselves reduced board area by 66% over 0402 types (source Murata). 01005 production goes industry wide Satoshi Kataoka, Production Manager, Assembléon Asia Pacific Region and Eric Klaver, Commercial Product Manager, Assembléon, The Netherlands The introduction of the

More information

EUROFEL-Report-2007-DS EUROPEAN FEL Design Study

EUROFEL-Report-2007-DS EUROPEAN FEL Design Study EUROFEL-Report-2007-DS4-095 EUROPEAN FEL Design Study Deliverable N : D 4.3 Deliverable Title: Task: Authors: Generation of 3rd harmonic photons at 90 nm DS-4 see next page Contract N : 011935 Project

More information

Electron Clouds in the SPS: progress in the analysis of cures/mitigations measures and potential schedule of implementation J.M.

Electron Clouds in the SPS: progress in the analysis of cures/mitigations measures and potential schedule of implementation J.M. Electron Clouds in the SPS: progress in the analysis of cures/mitigations measures and potential J.M. Jimenez This talk is a summary of my views meant for a recommendation. For detailed results and pictures,

More information

ESS: The Machine. Bucharest, 24 April Håkan Danared Deputy Head Accelerator Division. H. Danared Industry & Partner Days Bucharest Page 1

ESS: The Machine. Bucharest, 24 April Håkan Danared Deputy Head Accelerator Division. H. Danared Industry & Partner Days Bucharest Page 1 ESS: The Machine Bucharest, 24 April 2014 Håkan Danared Deputy Head Accelerator Division H. Danared Industry & Partner Days Bucharest Page 1 2025 ESS construction complete 2009 Decision: ESS will be built

More information

Beam Loss Detection for MPS at FRIB

Beam Loss Detection for MPS at FRIB Beam Loss Detection for MPS at FRIB Zhengzheng Liu Beam Diagnostics Physicist This material is based upon work supported by the U.S. Department of Energy Office of Science under Cooperative Agreement DE-SC0000661.

More information

Recent APS Storage Ring Instrumentation Developments. Glenn Decker Advanced Photon Source Beam Diagnostics March 1, 2010

Recent APS Storage Ring Instrumentation Developments. Glenn Decker Advanced Photon Source Beam Diagnostics March 1, 2010 Recent APS Storage Ring Instrumentation Developments Glenn Decker Advanced Photon Source Beam Diagnostics March 1, 2010 Ring Diagnostics Overview RF beam position monitor technology Photon beam position

More information

The FAIR plinac RF Systems

The FAIR plinac RF Systems The FAIR plinac RF Systems Libera Workshop Sep. 2011 Gerald Schreiber Gerald Schreiber, GSI RF Department 2 (1) Overview GSI / FAIR (2) FAIR Proton Linear Accelerator "plinac" (3) plinac RF Systems (4)

More information

AREAL- Phase 1. B. Grigoryan on behalf of AREAL team

AREAL- Phase 1. B. Grigoryan on behalf of AREAL team AREAL- Phase 1 Progress & Status B. Grigoryan on behalf of AREAL team Contents Machine Layout Building & Infrastructure Laser System RF System Vacuum System Cooling System Control System Beam Diagnostics

More information

HIGH POWER BEAM DUMP AND TARGET / ACCELERATOR INTERFACE PROCEDURES *

HIGH POWER BEAM DUMP AND TARGET / ACCELERATOR INTERFACE PROCEDURES * HIGH POWER BEAM DUMP AND TARGET / ACCELERATOR INTERFACE PROCEDURES * J. Galambos, W. Blokland, D. Brown, C. Peters, M. Plum, Spallation Neutron Source, ORNL, Oak Ridge, TN 37831, U.S.A. Abstract Satisfying

More information

Audio and Video II. Video signal +Color systems Motion estimation Video compression standards +H.261 +MPEG-1, MPEG-2, MPEG-4, MPEG- 7, and MPEG-21

Audio and Video II. Video signal +Color systems Motion estimation Video compression standards +H.261 +MPEG-1, MPEG-2, MPEG-4, MPEG- 7, and MPEG-21 Audio and Video II Video signal +Color systems Motion estimation Video compression standards +H.261 +MPEG-1, MPEG-2, MPEG-4, MPEG- 7, and MPEG-21 1 Video signal Video camera scans the image by following

More information

Concept of Operations (CONOPS)

Concept of Operations (CONOPS) PRODUCT 0-6873-P1 TxDOT PROJECT NUMBER 0-6873 Concept of Operations (CONOPS) Jorge A. Prozzi Christian Claudel Andre Smit Praveen Pasupathy Hao Liu Ambika Verma June 2016; Published March 2017 http://library.ctr.utexas.edu/ctr-publications/0-6873-p1.pdf

More information

The Elettra Storage Ring and Top-Up Operation

The Elettra Storage Ring and Top-Up Operation The Elettra Storage Ring and Top-Up Operation Emanuel Karantzoulis Past and Present Configurations 1994-2007 From 2008 5000 hours /year to the users 2010: Operations transition year Decay mode, 2 GeV (340mA)

More information

Hall-B Beamline Commissioning Plan for CLAS12

Hall-B Beamline Commissioning Plan for CLAS12 Hall-B Beamline Commissioning Plan for CLAS12 Version 1.5 S. Stepanyan December 19, 2017 1 Introduction The beamline for CLAS12 utilizes the existing Hall-B beamline setup with a few modifications and

More information

Development of beam-collision feedback systems for future lepton colliders. John Adams Institute for Accelerator Science, Oxford University

Development of beam-collision feedback systems for future lepton colliders. John Adams Institute for Accelerator Science, Oxford University Development of beam-collision feedback systems for future lepton colliders P.N. Burrows 1 John Adams Institute for Accelerator Science, Oxford University Denys Wilkinson Building, Keble Rd, Oxford, OX1

More information

Fast Orbit Feedback at the SLS. Outline

Fast Orbit Feedback at the SLS. Outline Fast Orbit Feedback at the SLS 2nd Workshop on Beam Orbit Stabilisation (December4-6, 2002, SPring-8) T. Schilcher Outline Noise Sources at SLS Stability / System Requirements Fast Orbit Feedback Implementation

More information

DC ELV Accelerators: Development and Application

DC ELV Accelerators: Development and Application DC ELV Accelerators: Development and Application Nikolay К. Кuksanov BINP, Novosibirsk, Russia 2017 BINP develops and manufactures ELV accelerators since 1970. These accelerators initially were developed

More information

Madero Ote. 686, Centro Histórico, C.P Morelia, Michoacán México. #300, Col. Cuauhtémoc, C.P Morelia, Michoacán, México

Madero Ote. 686, Centro Histórico, C.P Morelia, Michoacán México. #300, Col. Cuauhtémoc, C.P Morelia, Michoacán, México Verification of and DLP values for a head tomography reported by the manufacturers of the CT scanners, using a CT dose profiler probe, a head phantom and a piranha electrometer Edith Castillo Corona 1,2,

More information

Overview of NIRS Accelerator Activity

Overview of NIRS Accelerator Activity Overview of NIRS Accelerator Activity Koji Noda Research Center for Charged Particle Therapy National Institute of Radiological Sciences Workshop on Hadron Beam Therapy April. 27 09, Erice, Italy Hospital

More information

Scanning Electron Microscopy (FEI Versa 3D Dual Beam)

Scanning Electron Microscopy (FEI Versa 3D Dual Beam) Scanning Electron Microscopy (FEI Versa 3D Dual Beam) This operating procedure intends to provide guidance for basic measurements on a standard sample with FEI Versa 3D SEM. For more advanced techniques

More information

Photo cathode RF gun -

Photo cathode RF gun - Photo cathode RF gun - *),,, ( 05 Nov. 2004 Spring8 UTNL Linac & Mg Photocathode RF Gun Mg photocathode NERL, 18 MeV Linac and the RF gun Electron Beam Mg photocathode Mg photocathode RF gun of SPring8

More information

Non-Destructive Examination Benches and Analysis Laboratories in support to the Experimental Irradiation Process in the Future Jules Horowitz MTR

Non-Destructive Examination Benches and Analysis Laboratories in support to the Experimental Irradiation Process in the Future Jules Horowitz MTR Non-Destructive Examination Benches and Analysis Laboratories in support to the Experimental Irradiation Process in the Future Jules Horowitz MTR D. Parrat 1, P. Kotiluoto 2, T. Jäppinen 2, C. Roure 1,

More information

DIAGNOSTIC INSTRUMENTATION FOR MEDICAL ACCELERATOR FACILITIES

DIAGNOSTIC INSTRUMENTATION FOR MEDICAL ACCELERATOR FACILITIES DIAGNOSTIC INSTRUMENTATION FOR MEDICAL ACCELERATOR FACILITIES M. Schwickert, GSI, Darmstadt, Germany A. Peters, Hit GmbH, Heidelberg, Germany Abstract A number of accelerator facilities are presently emerging

More information

Beam Commissioning of GRIFFIN Beam Line

Beam Commissioning of GRIFFIN Beam Line Document-8677 Beam Commissioning of GRIFFIN Beam Line Document Type: Beam Commissioning Note Release: Release Date: 5--5 Author(s): S. Saminathan, R. Baartman Note: Before using a copy (electronic or printed)

More information

RF2TTC and QPLL behavior during interruption or switch of the RF-BC source

RF2TTC and QPLL behavior during interruption or switch of the RF-BC source RF2TTC and QPLL behavior during interruption or switch of the RF-BC source Study to adapt the BC source choice in RF2TTC during interruption of the RF timing signals Contents I. INTRODUCTION 2 II. QPLL

More information

Oak Ridge Spallation Neutron Source Proton Power Upgrade Project and Second Target Station Project

Oak Ridge Spallation Neutron Source Proton Power Upgrade Project and Second Target Station Project Oak Ridge Spallation Neutron Source Proton Power Upgrade Project and Second Target Station Project Workshop on the future and next generation capabilities of accelerator driven neutron and muon sources

More information

Design of Fault Coverage Test Pattern Generator Using LFSR

Design of Fault Coverage Test Pattern Generator Using LFSR Design of Fault Coverage Test Pattern Generator Using LFSR B.Saritha M.Tech Student, Department of ECE, Dhruva Institue of Engineering & Technology. Abstract: A new fault coverage test pattern generator

More information

NMR. picospin. Maintenance Guide

NMR. picospin. Maintenance Guide NMR picospin Maintenance Guide 269-302600 Revision A January 2013 2013 Thermo Fisher Scientific Inc. All rights reserved. For U.S. Technical Support, please contact: Thermo Fisher Scientific 5225 Verona

More information

Monitor QA Management i model

Monitor QA Management i model Monitor QA Management i model 1/10 Monitor QA Management i model Table of Contents 1. Preface ------------------------------------------------------------------------------------------------------- 3 2.

More information

3 cerl. 3-1 cerl Overview. 3-2 High-brightness DC Photocathode Gun and Gun Test Beamline

3 cerl. 3-1 cerl Overview. 3-2 High-brightness DC Photocathode Gun and Gun Test Beamline 3 cerl 3-1 cerl Overview As described before, the aim of the cerl in the R&D program includes the development of critical components for the ERL, as well as the construction of a test accelerator. The

More information

Laser Beam Analyser Laser Diagnos c System. If you can measure it, you can control it!

Laser Beam Analyser Laser Diagnos c System. If you can measure it, you can control it! Laser Beam Analyser Laser Diagnos c System If you can measure it, you can control it! Introduc on to Laser Beam Analysis In industrial -, medical - and laboratory applications using CO 2 and YAG lasers,

More information

EL302 DIGITAL INTEGRATED CIRCUITS LAB #3 CMOS EDGE TRIGGERED D FLIP-FLOP. Due İLKER KALYONCU, 10043

EL302 DIGITAL INTEGRATED CIRCUITS LAB #3 CMOS EDGE TRIGGERED D FLIP-FLOP. Due İLKER KALYONCU, 10043 EL302 DIGITAL INTEGRATED CIRCUITS LAB #3 CMOS EDGE TRIGGERED D FLIP-FLOP Due 16.05. İLKER KALYONCU, 10043 1. INTRODUCTION: In this project we are going to design a CMOS positive edge triggered master-slave

More information

The FLASH objective: SASE between 60 and 13 nm

The FLASH objective: SASE between 60 and 13 nm Injector beam control studies winter 2006/07 talk from E. Vogel on work performed by W. Cichalewski, C. Gerth, W. Jalmuzna,W. Koprek, F. Löhl, D. Noelle, P. Pucyk, H. Schlarb, T. Traber, E. Vogel, FLASH

More information

LCLS RF Reference and Control R. Akre Last Update Sector 0 RF and Timing Systems

LCLS RF Reference and Control R. Akre Last Update Sector 0 RF and Timing Systems LCLS RF Reference and Control R. Akre Last Update 5-19-04 Sector 0 RF and Timing Systems The reference system for the RF and timing starts at the 476MHz Master Oscillator, figure 1. Figure 1. Front end

More information

Evaluating Oscilloscope Mask Testing for Six Sigma Quality Standards

Evaluating Oscilloscope Mask Testing for Six Sigma Quality Standards Evaluating Oscilloscope Mask Testing for Six Sigma Quality Standards Application Note Introduction Engineers use oscilloscopes to measure and evaluate a variety of signals from a range of sources. Oscilloscopes

More information

Durham Magneto Optics Ltd. NanoMOKE 3 Wafer Mapper. Specifications

Durham Magneto Optics Ltd. NanoMOKE 3 Wafer Mapper. Specifications Durham Magneto Optics Ltd NanoMOKE 3 Wafer Mapper Specifications Overview The NanoMOKE 3 Wafer Mapper is an ultrahigh sensitivity Kerr effect magnetometer specially configured for measuring magnetic hysteresis

More information

STATUS OF THE SwissFEL C-BAND LINAC

STATUS OF THE SwissFEL C-BAND LINAC STATUS OF THE SwissFEL C-BAND LINAC F. Loehl, J. Alex, H. Blumer, M. Bopp, H. Braun, A. Citterio, U. Ellenberger, H. Fitze, H. Joehri, T. Kleeb, L. Paly, J.-Y. Raguin, L. Schulz, R. Zennaro, C. Zumbach,

More information

Jefferson Lab Experience with Beam Halo, Beam Loss, etc.

Jefferson Lab Experience with Beam Halo, Beam Loss, etc. Jefferson Lab Experience with Beam Halo, Beam Loss, etc. Pavel Evtushenko with a lot of input from many experienced colleagues Steve Benson, Dave Douglas, Kevin Jordan, Carlos Hernandez-Garcia, Dan Sexton,

More information

A Fast Magnet Current Change Monitor for Machine Protection in HERA and the LHC

A Fast Magnet Current Change Monitor for Machine Protection in HERA and the LHC 10th ICALEPCS Int. Conf. on Accelerator & Large Expt. Physics Control Systems. Geneva, 10-14 Oct 2005, PO2.042-4 (2005) A Fast Magnet Current Change Monitor for Machine Protection in HERA and the LHC M.Werner

More information

Tolerances on Magnetic Misalignments in SESAME Storage Ring

Tolerances on Magnetic Misalignments in SESAME Storage Ring Tolerances on Magnetic Misalignments in SESAME Storage Ring SES-TE-AP-TN-0003 April 20, 2014 Authored by: Reviewed by: Approved by: Access List : Maher Attal Erhard Huttle Erhard Huttle ---Internal ---------

More information