IAEA RER/9/135 COURSE ON OPTIMIZATION IN COMPUTED TOMOGRAPHY Sofia, Bulgaria, Tube current modulation and dose reduction : How TCM works

Size: px
Start display at page:

Download "IAEA RER/9/135 COURSE ON OPTIMIZATION IN COMPUTED TOMOGRAPHY Sofia, Bulgaria, Tube current modulation and dose reduction : How TCM works"

Transcription

1 IAEA RER/9/135 COURSE ON OPTIMIZATION IN COMPUTED TOMOGRAPHY Sofia, Bulgaria, 2017 Tube current modulation and dose reduction : How TCM works Dean Pekarovič UMC Ljubljana, Institute of Radiology Quality and Safety office Plain Radiography one kv value one mas value What should be on Image? Pulmo, Fat, Air, Bones Do we have same attenuation in all parts of the Image? 1

2 kv increase will Decrease the mas required to achieve a constant detector exposure (AEC on). Decrease ESD and effective dose. Decrease image contrast. Increase scattered radiation. 79 Kv 0,71 mas DAP 0,128 dgy cm 2 EI: 250,9 DI: -0,45 24,98 µgy 67 Kv 0,71mAs DAP 0,085 dgy cm 2 EI 191,5 DI -0,71 34,57 µgy Radiologist to decide contrast ribs/pulmo Ref: ICRP 121, Radiological protection in paediatric diagnostic and interventional radiology mas increase will Increase detector exposure (without AEC) Increase radiation dose to the patient. Improve image quality increased CNR and SNR?. noise / higher mas 75 Kv 1,2 mas DAP 0,224 dgy cm 2 DI 3,23 NO AEC 75 Kv 0,63 mas DAP 0,084 dgy cm 2 DI -0,9 NO AEC 2

3 Lookin for proper ImQ DX 65 kv DAP: 0,021 dgycm 2 0,71 mas DI : 0,35 63 kv DAP: 0,032 dgycm 2 0,9 mas DI : 1,15 58 kv DAP: 0,053 dgycm 2 1 mas DI : 2,26 newborn 3 girl weeks, girl, 2760 g 7 days 2750 g 62 kv DAP: 0,054 dgycm 2 1,21 mas DI : 3,04, + Cu Filter 70 kv DAP: 0,021 dgycm 2 0,56 mas DI : - 0,42, + Cu Filter What we need is how to control the ImQ and Dose kv mas EI DI DAP µgy mgy/s ,53-0,32 0, ,1 256,4 0,12 0,958 71,9 16,6 90 1,2 284,8 0,57 1, ,5 0,32 1, ,8 1246,5-0,03 1, , ,6 217,9-0,6 1, ,6 238,5-0,2 2, ,6 kv mas EI DI DAP ,28 0, ,26 0, ,7-0,07 0,883 How to improve DI 1mAs cca 0,4 DI 1 Kv cca 0,3 DI DI % CsJ Nastavljeni protokoli mas focal Size range kv spot target EI OTROK SMALL 0, ,22 PC 1 KG 68 MEDIUM 70 0, ,22 LARGE 70 0, ,22 range mas spot target EI kv focal SMALL 0, ,22 OTROK PC 2 KG 70 MEDIUM 72 0, ,22 LARGE 72 0, ,22 range mas spot target EI kv focal SMALL 1 226,22 OTROK PC 3 KG 71 0,5 MEDIUM 73 0, ,22 LARGE 73 0, ,22 range mas spot target EI kv focal SMALL 0, ,22 OTROK PC 4 KG 72 MEDIUM 74 0, ,22 LARGE 74 0, ,22 range mas spot target EI kv focal SMALL 0, ,22 OTROK PC 5-10 KG 75 MEDIUM 75 0, ,22 LARGE 75 0, ,22 range mas spot target EI kv focal SMALL 1 226,22 OTROK PC KG 75 0,8 MEDIUM 77 0, ,22 LARGE 77 0, ,22 range spot target EI kv mas focal SMALL 0, ,22 OTROK PC KG 80 MEDIUM 80 0, ,22 LARGE 80 0, ,22 range mas spot target EI kv focal SMALL 0, ,22 OTROK PC KG 80 MEDIUM 85 0, ,22 LARGE 85 0, ,22 range mas spot target EI kv focal SMALL 1 226,22 OTROK PC KG 85 0,9 MEDIUM 90 0, ,22 LARGE 95 1, ,22 3

4 From plain to CT CT -problem 4

5 Tube Current Modulation Purpose is to ensure the same picture quality regardless of the characteristics of the patient. = same CNR in all images Tube Current Determines the number of electrons accelerated across the x-ray tube per unit time Units: milliamperes (ma) CTDI vol is directly proportional to Tube Current CTDI vol Tube Current REF : AAPM Computed Tomography Radiation Dose Education Slides 5

6 TCM along z axis mas changes automatically 140mAs 55mAs 110mAs 130mAs Tube Current Modulation (TCM) / Automatic Exposure Control (AEC) Automatically adapts the Tube Current or Tube Potential according to patient attenuation to achieve a specified image quality Automatic adjustment of Tube Current may not occur when Tube Potential is changed Centering the patient in the gantry is VITAL for most AEC systems AEC aims to deliver a specified image quality across a range of patient sizes. It tends to increase CTDI vol for large patients and decrease it for small patients relative to a reference patient size. The use of Automatic Exposure Control may decrease or increase CTDI vol depending on the patient size and body area imaged and image quality requested REF : AAPM Computed Tomography Radiation Dose Education Slides 6

7 TCM - Principe Reference data is collected from attenuation of anatomical structures during scanned projection radiograph -SPR (one or two) Data collected from previous rotation. TCM -principle TCM could be defined as a CT technique that performs automatic modulation of tube current in the x, y plane (angular modulation), or along the scanning direction, z- axis, (longitudinal modulation), or both (combined modulation). Is really total Automatic? The modification is done according to each patient s size, shape and attenuation of body parts being scanned. The operator must select a required image quality level and then the system can adjust the tube current to obtain the predetermined image quality with improved radiation efficiency. Ref :Kalra MK et all, Computed tomography radiation dose optimization: scanning protocols and clinical applications of automatic exposure control.curr Probl Diagn Radiol 2005; 34:

8 Start info.. TCM needs input WHAT is needed Image quality = radiologist input, but team should be involved; + radiographer + medical physicist Different vendor solutions : Noise = SD CT-numbers on the Image (GE, Toshiba) Reference ma or mas for a standard patient for specified quality (Philips, Siemens) TCM idea AP diameter : 36 cm LAT diameter : 28 cm 8

9 Scan where Tube Current Modulation was used Blue Curve Represents actual instantaneous ma Red Curve Represents avg ma for each image Yellow Curve Represents avg ma over entire scan Overall avg is used for CTDIvol reported in Dose Report Dose Modulation -AEC Many CT scanners automatically adjust the technique parameters (and as a result the CTDI vol ) to achieve a desired level of image quality and/or to reduce dose Dose Modulation and Reduction techniques vary by scanner manufacturer, model and software version. Tip : read the manual and then test the AEC. 9

10 Variations in ma modulation Spatial ma modulation Longitudinal -Z axis modulation Angular / x-y axis / transverse modulation Combined / x-y-z modulation Temporal ma modulation ma adapted at different time points using ECG gating (Cardiac CTA);not included in this PPT TCM Modulations Longitudinal (Z direction ) Is an TCM feature that adjusts the Tube Current as patient size and attenuation changes of the anatomic region in the longitudinal direction. The CT Localizer Radiograph is used to estimate patient attenuation. REF : AAPM Computed Tomography Radiation Dose Education SlideS 10

11 TCM Modulations Angular (X,Y direction) Is an TCM feature that adjusts the Tube Current as the x-ray tube rotates around the patient according to the size, shape, and attenuation of body region being scanned to compensate for attenuation changes with view angle. Angular Tube Current Modulation is used to adjust the Tube Current to attempt to deliver similar dose to the detector at all view angles. REF : AAPM Computed Tomography Radiation Dose Education SlideS 180 -mirror Ref: M.K.Karla, Automatic Exposure Control in Multidetector-row CT TCM Modulations Angular + Longitudinal (x, y, z direction) Is an AEC feature that incorporates the properties of both Angular and Longitudinal Tube Current Modulation to Adjust the Tube Current based on the patient s overall attenuation. Modulate the Tube Current in the angular (X-Y) and longitudinal (Z) dimensions to adapt to the patient s shape. Ref: M.K.Karla, Automatic Exposure Control in Multidetector-row CT REF : AAPM Computed Tomography Radiation Dose Education SlideS 11

12 All in one Ref : Marcus Söderberg; Automatic exposure control in CT Principles are similar, but..? How it works on my CT modality 12

13 Siemens : Care Dose 4D AEC System Operator-Controlled Z-Axis Angular X,Y Vendor Name Parameter Parameter Explanation Principles Technique Modulation Modulation Angular modulation of tube current in the x, Image quality y, and z axes on the Siemens CAREDose4D reference mas mas that would be used for basis of patient size x-y-z/combined Yes Yes an average-sized patient relative to the mas specified by the user for a standard-sized reference patient Implies need for image quality equal to that obtained with the use of specified reference mas in a standard adult (70 kg -80 kg) or child (20 kg). Head Protocols only x,y modulation Quality Reference mas is NOT the max or min. Siemens : Care Dose 4D The degree to which the tube current is adjusted for patient size can be selected, using very weak to very strong in 5 steps (high degree of ma adjustment), compensation settings. Ref : Siemens and Sodberg, Automatic exposure control in CT 13

14 Philips : Dose Right AEC System Operator-Controlled *miss new technique with x, y and z axis modulation Z-Axis Angular X,Y Vendor Name Parameter Parameter Explanation Principles Technique Modulation Modulation Modulation of tube current on the basis of Z-DOM D-DOM Reference image Image quality expressed in patient size to achieve with ACS Phillips Dose Right Ref mas/slice terms of noise level of an the same image noise Yes Yes ACS existing optimal clinical level as in a previously image defined reference image Baseline mas is used as a reference to obtain constant image noise along the z axis. DoseRight ACS (automatic current selector) provides patients based AEC(object size), by use of a reference image DoseRight DOM (Dose Modulation) Z-Dom: Z-axis AEC D-Dom: Tube current is set so that 90% of images will have equal or lower noise than the reference image, with remaining 10% of images in a series having equal or higher noise than the reference image. No ACS for Head CT protocols. Philips : Dose Right* Additional Parameter: Reference Noise Index DoseRight 3D-DOM (three dimensional dose modulation) combines angular and longitudinal patient information to modulate dose in three dimensions (x-y-z-axis). It incorporates modulation of tube current time product (mas) according to changes in individual patient s size and shape in the transverse (x-y-axis; angular) direction during helical scans, in addition to changes in the craniocaudal or caudocranial (z-axis; longitudinal) direction, as the tube rotates. 14

15 GE : AutomA 3D AEC System Operator-Controlled Z-Axis Angular X,Y Vendor Name Parameter Parameter Explanation Principles Technique Modulation Modulation Measure of image quality/ Modulation of tube noise level defined current only in the Auto ma Noise index relative to uniform water longitudinal direction z axis/longitudinal Yes No phantom to maintain a constant NI Measure of image quality/ Modulation of tube GE noise level defined current in the x, y, and Smart ma Noise index relative to uniform water z axes to maintain a x-y axis/angular Yes Yes phantom constant noise index Auto ma 3D* Noise index x-y-z/combined Yes Yes GE : defined with NI (Noise Index) NI : Std.Dev. of HU number in water phantom with standard algorithm used NI CTDI vol Auto ma use last scout if more then one were performed. Smart ma MUST works with Auto ma GE : AutomA 3D Noise Index is Image Quality Parameter which sets the image noise in the image. Scout is used to determine patient attenuation characteristics and size and along with Noise Index the ma per rotation for the acquisition is determined. Noise Index will vary based on the slice thickness selected due to the difference in image noise relative to slice thickness. The same NI should never be used across all slice thicknesses. Minimum and maximum ma - Range of allowed ma to achieve desired noise index 15

16 Toshiba Sure Exposure 3D AEC System Operator-Controlled Z-Axis Angular X,Y Vendor Name Parameter Parameter Explanation Principles Technique Modulation Modulation User prespecifies image quality on the basis of Standard deviation of pixel a patient-equivalent Sure Exposure Target Image quality values in an image water phantom, and x-y-z/combined Yes Yes 3D level Higher SD= higher noise mas is modulated on Always ON the basis of patient Toshiba size to maintain image quality Sure Exposure z axis/longitudinal Sure Exposure 3D is Longitudinal Modulation XY Modulation is Angular Modulation 2 possibilities for radiographer : Standard Deviation SD Image Quality level High, Standard.. SD of pixel values measured in a patientequivalent water phantom. Toshiba Sure Exposure 3D LL (50cm), L (40cm), M (32cm), S (24cm), SS (18cm) AEC setup allows tube current to be limited by max. and min. values. Be careful with reconstructed slice thickness and reconstruction kernel. 16

17 Image Quality Parameters -TCM GE NI (Noise Index) Philips mas/slice Siemens -ref. mas Ref :M.P.Supanich,Tube Current Modulation,3rd CT Dose Summit Scan and Reconstruction parameters Which parameter effect TCM? AEC system Tube Voltage Rotation time Pitch Slice Thickness Reconstruction kernel Localizer** CARE Dose 4D * Yes Yes Yes Dose Right Yes Yes Yes Yes Yes AutomA 3D Yes Yes Yes Yes Yes SureExposure 3d Yes Yes Yes Yes Yes Yes *New Versions of CARE Dose 4D, a change in the tube volatage will result in a change in tube current by the AEC ** add to original Table Ref : 17

18 Localizer -SPR Start and end of scan range. Used for appropriate tube current modulation (AEC in use). Incorrect position in iso center leads to inadequate tube current modulation (higher or lower). Incorrect position lead to incorrect geometry and size of scanned object. 6 cm from Center TH 0 kv 120 mas 190 AEC ON Localizer Both has same exposure parameters kv 120 mas 35 Length 256 mm CTDI vol (32cm) 0,13mGy DLP 3 mgycm Table Height Δ 6cm 1. Spiral mode kv 120 mas/ref mas 204/180 Length 117 mm Slice thickness 5 mm CTDI vol (32cm) 13.6 mgy DLP 145 mgycm 22 % difference in dose from 6 cm table height difference. TH +6 cm kv 120 mas 230 AEC ON Positioned above isocenter (+6 cm) 2. Spiral mode kv 120 mas/ref mas 219/180 Length 117 mm Slice thickness 5 mm CTDI vol (32cm) 16.7 mgy DLP 178 mgycm 18

19 Isocenter noise and CR Cylindrical 16 cm CTDI-head phantom scanned in center position and while lowered by 60 mm. Noise (1SD HU) increases vertically across the phantom in lower position as the beam shape is non-optimally targeted in the scan. isocenter isocenter noise increase slight contrast change Noise (1SD HU) y-coordinate (mm) centered 60 mm below Contrast (HU) y-coordinate (mm) centered 60 mm below Ref : Eurosafe, M. Kortesniemi, D. Pekarovic, D.Sheppard Spr DIRECTION 19

20 Scanning Outside The Localizer Localizer is basic source for TCM If scan region extends outside of localizer we have a problem. Four possible outcomes in scan region not covered by localizer : Tube Current goes to maximum Tube Current goes to minimum Tube Current stays what it was at edge of localizer Tube Current goes to manual setting Ref : TCM, M.Supanich,3rd CT Dose Summit One or two and direction of the localizer 20

21 TCM performance on QA phantom Ref :Iball: A QA test for CT AEC systems, Journal Of Applied Clinical Medical Physics Example how to start 120 kv 146 mas ma ma ma ma ma ma ma ma , ,5 54, ,5 112,5 54, , , , , ,5 54, ,5 58,5 101,5 54, , ,5 54, , ,5 59, ,5 59, ,5 55,5 59, , , , , ,5 83,5 53, ,5 80,5 53, , ,5 53, , , ,5 53, ,5 55,5 60, , , , , , , , ,5 62,5 134,5 54,5 61, ,5 62, ,5 61, ,5 62, , ,5 62,5 120 kv 53 mas 21

22 Example 26 Months ma kv 100 mas 30 Thorax wo CM Topogram Cr/Ca 100kV 35 mas Helical 100 Kv 30 mas AEC off g/rot : 0,5 s CTDIvol: 1.23 mgy DLP : 23 mgy cm Example kv 100 mas ma ma kv 100 mas Months Thorax +Adomen wo CM Topogram Cr/Ca 100kV 100 mas Helical 100 Kv 100 ma AEC off g/rot 0,5 s CTDIvol : 4.50 mgy DLP : 182 mgy cm (+Abd) 22

23 Example ma ma 30 % mas deviation in Z plane kv 100 mas Months Thorax wo CM Topogram Cr/Ca 100kV Helical 100 Kv 30 ref/ 27 mas AEC on g/rot 0,5 s CTDIvol 4.50 mgy DLP 182 mgy cm Example kv 100 mas Months Thorax wo CM Topogram Cr/Ca 120kV Helical 100 Kv 20 ref/ 15 mas AEC on g/rot 0,5 s CTDIvol : 0.80 mgy DLP : 14 mgy cm 23

24 When should it be avoided? A great tool, but sometimes.. Head exams? Brain Orbits Sinuses CT Perfusion Little or no table motion No chance to change cross section shape low manual technique Team Therefore, it is paramount that radiologists know how to use the AEC systems in their own scanners. AEC systems do not reduce radiation dose per se; rather, they control radiation exposure relative to the required image quality. Only Radiologist? 24

25 EMAN - WP 8 :Training & Education 1.The medical practitioners requesting a CT examination. This group requires knowledge about indications for CT, its alternatives and the associated risks and benefits. 2. The core CT team that defines and optimizes the set of standard scan protocols on a specific scanner (radiographer, medical physicist and radiologist).this team will usually start with a standard set of protocols provided by the manufacturer and adapt it to the local needs. This team requires in-depth knowledge of scan parameters and how to optimize them. 3. The professionals (radiologists, radiographers) define the CT protocols. This group has to have knowledge when not to use CT, but another image technique, according to patient clinical indication. They are ultimately responsible for the individual choice of the correct protocol associated with each of the set of available standard protocols at a specific scanner / institution. 4. The radiographers that actually perform the examination. This group requires knowledge about individual routine adaptations required for each patient, such as centring of patients, adapting scan range, adapting protocol to patient size, optimizing modality performance in order to obtain the best diagnostic image at the lowest possible dose. How to evaluate Understanding CTDIvol Dose Reference Levels ( will be disuseed..) 25

26 CT Dose Numbers CTDI vol DLP Effective Dose Why CTDI vol? CTDI vol provides information about the amount of radiation used to perform the study. CTDI vol is a useful index to track across patients and protocols for quality assurance purposes. CTDI vol can be used as a metric to compare protocols across different practices and scanners when related variables, such as resultant image quality, are also taken in account. 26

27 DLP The Dose Length Product (DLP) is also calculated by the scanner. DLP is the product of the length of the irradiated scan volume and the average CTDI vol over that distance. DLP has units of mgy*cm. CTDI Absorbed dose in the slice. Absorbed dose including scatter contributions from outside the slice (CTDI). The CTDI is calculated as the integral of the absorbed dose along the z axis, divided by the nominal slice thickness S. Ref : Easy Guide to low dose,siemens 27

28 CTDIw There are different ways to calculate the CTDI. One of them is to consider the differences between the absorbed dose in the periphery and in the centre of the patient s body by a weighted sum of the central and peripheral CTDI values. Ref : Easy Guide to low dose,siemens CTDIvol -DLP The Dose Length Product (DLP) is the product of CTDIvol and the examination range. The Dose Length Product (DLP) is also calculated by the scanner. DLP is the product of the length of the irradiated scan volume and the average CTDI vol over that distance DLP has units of mgy*cm Ref : Easy Guide to low dose,siemens 28

29 Effective dose Region of body E = E DLP DLP Effective dose E - measured in Sievert (Sv) units, includes the sensitivity to radiation of the different organs. It is the sum of the equivalent doses in all irradiated organs multiplied by the respective tissue weighting factors wi. Normalised effective dose, E DLP (msv mgy -1 cm -1 ) Head Neck Chest Abdomen Pelvis Normalised values of effective dose per dose-length product (DLP) over various body regions. ICRP 60 ICRP 103 The Recommendations of the International Commission on Radiological Protection of 2007 (ICRP 103) has different coefficients than that of 1990 (ICRP 60). In particular, gonads are less radiosensitive and the breast is more radiosensitive than previously assumed. The effective dose E is an approximate measure that was introduced to compare the stochastic risk of a nonuniform exposure of ionizing radiation with the risk caused by a uniform exposure of the whole body. 29

30 Conclusions Users must possess a good understanding of the concepts of noise index, standard deviation, reference images, and reference ma(mas) as they relate to each AEC system. Use User manual and ask vendor for additional papers. Evaluate your own AEC system, for each software and scanner type. Don`t accept one solution. Understanding will improve reports and CT Dose will become controllable. Thank you 30

Understanding CT image quality

Understanding CT image quality IAEA RER/9/135 COURSE ON OPTIMIZATION IN COMPUTED TOMOGRAPHY Sofia, Bulgaria, 2017 Understanding CT image quality Dean Pekarovič UMC Ljubljana, Institute of Radiology Quality and Safety office Role of

More information

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009

Joint ICTP/IAEA Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation May 2009 2033-17 Joint ICTP/ Advanced School on Dosimetry in Diagnostic Radiology and its Clinical Implementation 11-15 May 2009 Dosimetry for CT 3: Practical Experiences Claire-Louise Chapple Freeman Hospital

More information

2

2 1 2 3 4 5 6 7 1. Bauhs, J. A., Vrieze, T. J., Primak, A. N., Bruesewitz, M. R., & McCollough, C. H. (2008). CT Dosimetry: Comparison of Measurement Techniques and Devices1. Radiographics, 28(1), 245-253.

More information

Comparison of Measured Values of CTDI and DPL with Standard Reference values of Different CT Scanners for dose Management

Comparison of Measured Values of CTDI and DPL with Standard Reference values of Different CT Scanners for dose Management 2017 IJSRST Volume 3 Issue 1 Print ISSN: 2395-6011 Online ISSN: 2395-602X Themed Section: Science and Technology Comparison of Measured Values of CTDI and DPL with Standard Reference values of Different

More information

RSNA 2006 November 26 to December 1 Chicago. Guest author for ImPACT Dr. Koos Geleijns, Medical Physicist, Leiden University Medical Center.

RSNA 2006 November 26 to December 1 Chicago. Guest author for ImPACT Dr. Koos Geleijns, Medical Physicist, Leiden University Medical Center. RSNA 2006 November 26 to December 1 Chicago Guest author for ImPACT Dr. Koos Geleijns, Medical Physicist, Leiden University Medical Center. Once again, more than 60,000 participants (including professional

More information

Equipment Quality Control for Digital Radiography February 22, Imaging Physics CancerCare Manitoba

Equipment Quality Control for Digital Radiography February 22, Imaging Physics CancerCare Manitoba Equipment Quality Control for Digital Radiography February 22, 2018 Imaging Physics CancerCare Manitoba Purpose An equipment quality control (QC) program establishes baseline performance levels, tracks

More information

A simple anthropomorphic phantom used to demonstrate the effectiveness of CT dose modulation functions.

A simple anthropomorphic phantom used to demonstrate the effectiveness of CT dose modulation functions. A simple anthropomorphic phantom used to demonstrate the effectiveness of CT dose modulation functions. Lynn Bateman & Peter Hiles North Wales Medical Physics How can we test CT dose modulation? Lynn Bateman

More information

What do we mean by workload? Number of scans Type of scans/mix of scans? Total mas Total dip

What do we mean by workload? Number of scans Type of scans/mix of scans? Total mas Total dip CT & Workload (BIR) David Sutton What do we mean by workload? Number of scans Type of scans/mix of scans? Total mas Total dip What do we do with it? Need to relate workload to total scatter dose in the

More information

Madero Ote. 686, Centro Histórico, C.P Morelia, Michoacán México. #300, Col. Cuauhtémoc, C.P Morelia, Michoacán, México

Madero Ote. 686, Centro Histórico, C.P Morelia, Michoacán México. #300, Col. Cuauhtémoc, C.P Morelia, Michoacán, México Verification of and DLP values for a head tomography reported by the manufacturers of the CT scanners, using a CT dose profiler probe, a head phantom and a piranha electrometer Edith Castillo Corona 1,2,

More information

In recent years, CT technology has undergone profound

In recent years, CT technology has undergone profound PATIENT SAFETY Y.T. Niu M.E. Olszewski Y.X. Zhang Y.F. Liu J.F. Xian Z.C. Wang Experimental Study and Optimization of Scan Parameters That Influence Radiation Dose in Temporal Bone High-Resolution Multidetector

More information

A method for calculating the dose length product from CT DICOM images

A method for calculating the dose length product from CT DICOM images The British Journal of Radiology, 84 (2011), 236 243 A method for calculating the dose length product from CT DICOM images 1 I A TSALAFOUTAS, PhD and 2 S I METALLIDIS, MSc 1 Medical Physics Department,

More information

Standard. Substitute Test or Procedure. Required Test or. 1 Scan Increment Accuracy. Initially and Annually Initially and Annually

Standard. Substitute Test or Procedure. Required Test or. 1 Scan Increment Accuracy. Initially and Annually Initially and Annually Manufacturer s Recommendations for Alternate Dental CBCT QA Program VaTech: Model PAX i3d, i3d Green, i3d Smart Table 6 Medical Physicist s Computed Tomography QC Survey Required Test or Item Procedure

More information

Iterative Reconstruction with Philips idose Characterising Image Quality in Attempting to Realise its Potential

Iterative Reconstruction with Philips idose Characterising Image Quality in Attempting to Realise its Potential Iterative Reconstruction with Philips idose Characterising Image Quality in Attempting to Realise its Potential Julie Smyth & Philip Doyle Regional Medical Physics Service Outline Preamble Image Quality

More information

2012 Computed Tomography

2012 Computed Tomography 2012 Computed Tomography QUALITY CONTROL MANUAL Radiologist s Section Radiologic Technologist s Section Medical Physicist s Section 2012 Computed Tomography QUALITY CONTROL MANUAL Radiologist s Section

More information

NEMA XR 25 COMPUTED TOMOGRAPHY DOSE CHECK

NEMA XR 25 COMPUTED TOMOGRAPHY DOSE CHECK NEMA XR 25 COMPUTED TOMOGRAPHY DOSE CHECK NEMA Standards Publication XR 25-2010 Computed Tomography Dose Check Published by: National Electrical Manufacturers Association 1300 North 17th Street, Suite

More information

2017 Computed Tomography

2017 Computed Tomography 2017 Computed Tomography QUALITY CONTROL MANUAL Radiologist s Section Radiologic Technologist s Section Qualified Medical Physicist s Section 2017 Computed Tomography QUALITY CONTROL MANUAL Radiologist

More information

QUALITY CONTROL AND PATIENT DOSES FROM X-RAY EXAMINATIONS IN SOME HOSPITALS IN THAILAND

QUALITY CONTROL AND PATIENT DOSES FROM X-RAY EXAMINATIONS IN SOME HOSPITALS IN THAILAND QUALITY CONTROL AND PATIENT DOSES FROM X-RAY EXAMINATIONS IN SOME HOSPITALS IN THAILAND P. Plainoi, W. Diswath, N. Manatrakul Ministry of Public Health, Nonthaburi, Thailand XA0101612 Abstract Quality

More information

Brilliance CT 64-channel configuration

Brilliance CT 64-channel configuration Brilliance CT 64-channel configuration with Essence technology The Brilliance CT 64-channel configuration is designed to help you conduct the most advanced multislice CT studies possible. These systems

More information

Request for Proposals

Request for Proposals Request for Proposals Reference: ERDFIAI2012-4006A Caring First Ltd, is a limited liability company providing healthcare services. The company is currently commissioning a private new-build hospital complete

More information

Scope: All CT staff technologist

Scope: All CT staff technologist APPROVED BY: Radiology Technical Director Page 1 of 6 Purpose: The QC program assesses relative changes in system performance as determined by the technologist, service engineer, qualified medical physicist,

More information

Assessment of the Impact of Dark Signal on Image Quality in Computerized Mammography. Introduction

Assessment of the Impact of Dark Signal on Image Quality in Computerized Mammography. Introduction Assessment of the Impact of Dark Signal on Image Quality in Computerized Mammography Alexia Valasi 1, Maria Argyrou 1, Ioannis Floros 1, Elisavet Molyvda 2, Anastasios Siountas 2, Stella Synefia 1, Ioannis

More information

More Info at Open Access Database Process Control for Computed Tomography using Digital Detector Arrays

More Info at Open Access Database  Process Control for Computed Tomography using Digital Detector Arrays Digital Industrial Radiology and Computed Tomography (DIR 2015) 22-25 June 2015, Belgium, Ghent - www.ndt.net/app.dir2015 More Info at Open Access Database www.ndt.net/?id=18082 Process Control for Computed

More information

-Technical Specifications-

-Technical Specifications- Annex I to Contract 108733 NL-Petten: the delivery, installation, warranty and maintenance of one (1) X-ray computed tomography system at the JRC-IET -Technical Specifications- INTRODUCTION In the 7th

More information

3/2/2016. Medical Display Performance and Evaluation. Objectives. Outline

3/2/2016. Medical Display Performance and Evaluation. Objectives. Outline Medical Display Performance and Evaluation Mike Silosky, MS University of Colorado, School of Medicine Dept. of Radiology 1 Objectives Review display function, QA metrics, procedures, and guidance provided

More information

Staff dose and good prac/ces in CT guided IR procedures. Jonas Andersson Medical Physicist, Ph.D.

Staff dose and good prac/ces in CT guided IR procedures. Jonas Andersson Medical Physicist, Ph.D. Staff dose and good prac/ces in CT guided IR procedures Jonas Andersson Medical Physicist, Ph.D. jonas.s.andersson@vll.se Contents Good prac/ces for staff dose management Staff dose reported in the literature

More information

Department of Medical Physics, Graduate School of Nuclear and Allied Science, University of Ghana, Accra - Ghana, West Africa

Department of Medical Physics, Graduate School of Nuclear and Allied Science, University of Ghana, Accra - Ghana, West Africa 2018 IJSRST Volume 4 Issue 5 Print ISSN: 2395-6011 Online ISSN: 2395-602X Themed Section: Science and Technology Assessment of Adult Patient Radiation Dose and Image Quality in CT Examinations Performed

More information

Vascular. Development of Trinias FPD-Equipped Angiography System. 1. Introduction. MEDICAL NOW No.73 (2013.2) Yoshiaki Miura

Vascular. Development of Trinias FPD-Equipped Angiography System. 1. Introduction. MEDICAL NOW No.73 (2013.2) Yoshiaki Miura Vascular Development of Trinias FPD-Equipped Angiography System Medical Systems Division, Shimadzu Corporation Yoshiaki Miura 1. Introduction Shimadzu has developed Trinias (one ceiling-mounted type C12

More information

Ride the Lightning. A Review of Radiographic Physics

Ride the Lightning. A Review of Radiographic Physics Ride the Lightning A Review of Radiographic Physics Overview Basic principles of x-ray production Imaging equipment Quality Control Technical Factors Image QA Preparing for the registry The ARRT just wants

More information

Laser Beam Analyser Laser Diagnos c System. If you can measure it, you can control it!

Laser Beam Analyser Laser Diagnos c System. If you can measure it, you can control it! Laser Beam Analyser Laser Diagnos c System If you can measure it, you can control it! Introduc on to Laser Beam Analysis In industrial -, medical - and laboratory applications using CO 2 and YAG lasers,

More information

Practix. Mobile Radiography. All a Bucky station needs plus mobility. Practix 400

Practix. Mobile Radiography. All a Bucky station needs plus mobility. Practix 400 Practix 400 Specifications Practix Mobile Radiography All a Bucky station needs plus mobility Practix 400 Practix 400 meets the most demanding requirements of medical care, combining the virtues of mobile

More information

Ergonomic design packed with powerful performance MULTIMOBIL 5C. Answers for life.

Ergonomic design packed with powerful performance MULTIMOBIL 5C. Answers for life. Ergonomic design packed with powerful performance MULTIMOBIL 5C Answers for life. MULTIMOBIL 5C Mobile C-arm image intensifier system with high frequency technology Siemens has led the way in development

More information

FDA MAMMOSCAN FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM MAMMOGRAPHY

FDA MAMMOSCAN FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM MAMMOGRAPHY MAMMOGRAPHY MAMMOSCAN FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM Biopsy Attachment џ MAMMOSCAN is a digital scanning technology development taking a special place in ADANI Medical X-ray product line. џ Its

More information

Precision, power, and productivity

Precision, power, and productivity Precision, power, and productivity Philips Brilliance CT Big Bore oncology configuration Philips Healthcare continues its tradition of innovation in oncology with a CT solution that specifically addresses

More information

Intra-oral X-ray unit with high frequency DC generator. Product Data

Intra-oral X-ray unit with high frequency DC generator. Product Data Intra-oral X-ray unit with high frequency DC generator Tubehead Features Product Data High voltage circuit type High frequency (DC) High voltage frequency 30 khz Rated output voltage 60-65 - 70 kvp ± 8%

More information

4/14/2009. The Big Picture of Quality. MRI Quality Assurance and ACR MRI Accreditation Program. Basic Elements for Image Quality.

4/14/2009. The Big Picture of Quality. MRI Quality Assurance and ACR MRI Accreditation Program. Basic Elements for Image Quality. The Big Picture of Quality MRI Quality Assurance and ACR MRI Accreditation Program Chen Lin, PhD Indiana University School of Medicine & Clarian Health Partners Diagnosis accuracy Image quality Knowledge

More information

Promises and Perils of Proton Therapy Beam Delivery (Implications) or Towards Cost Effective Particle Therapy

Promises and Perils of Proton Therapy Beam Delivery (Implications) or Towards Cost Effective Particle Therapy Promises and Perils of Proton Therapy Beam Delivery (Implications) or Towards Cost Effective Particle Therapy Jay Flanz MGH/FBTC Harvard Medical School What is a Beam Delivery? Start with an accelerator

More information

YXLON Cougar EVO PLUS

YXLON Cougar EVO PLUS YXLON Cougar EVO PLUS The best small footprint X-ray inspection system for LABORATORY applications Technology with Passion Choose a custom-built EVO solution for premium inspection Why compromise? As technology

More information

CT Numbers: Think of a number, double it, add 20, divide by 4. Jane Edwards Royal Free Hospital, London

CT Numbers: Think of a number, double it, add 20, divide by 4. Jane Edwards Royal Free Hospital, London CT Numbers: Think of a number, double it, add 20, divide by 4 Jane Edwards Royal Free Hospital, London Background 2 different manufacturers CT scanners available on site: GE Lightspeed Plus; Philips Brilliance

More information

Performance Evaluation of Industrial Computed Radiography Image Display System

Performance Evaluation of Industrial Computed Radiography Image Display System Performance Evaluation of Industrial Computed Radiography Image Display System More info about this article: http://www.ndt.net/?id=21169 Lakshminarayana Yenumula *, Rajesh V Acharya, Umesh Kumar, and

More information

X-ray Tube Housing Assembly H1076X H1076Y Print No.HA Release Date:

X-ray Tube Housing Assembly H1076X H1076Y Print No.HA Release Date: H1076X 0.6 1.5 H1076Y 0.6 1.5 Print No.HA1006 1 Release Date: 2013 07 15 Description X-ray Tube Housing Assembly The tube unit is so constructed that an X-ray tube is sealed in a diagnostic type protective

More information

Creating room. for your visions

Creating room. for your visions BuckyDiagnost Radiography solutions Creating room for your visions 2 What makes a Rad Room simply perfect? Room for all your ideas Efficient, powerful, individual and innovative. This sums up just some

More information

True comfort and flexibility with the power of 3T.

True comfort and flexibility with the power of 3T. True comfort and flexibility with the power of 3T. With a large 71 cm aperture and the quietest exams in the industry, the Vantage Titan 3T is the most comfortable 3T MRI system for all of your patients.

More information

Introduction. Edge Enhancement (SEE( Advantages of Scalable SEE) Lijun Yin. Scalable Enhancement and Optimization. Case Study:

Introduction. Edge Enhancement (SEE( Advantages of Scalable SEE) Lijun Yin. Scalable Enhancement and Optimization. Case Study: Case Study: Scalable Edge Enhancement Introduction Edge enhancement is a post processing for displaying radiologic images on the monitor to achieve as good visual quality as the film printing does. Edges

More information

CARESTREAM VITA/VITA LE/VITA SE CR System Long Length Imaging User Guide

CARESTREAM VITA/VITA LE/VITA SE CR System Long Length Imaging User Guide CARESTREAM VITA/VITA LE/VITA SE CR System Long Length Imaging User Guide Use of the Guide Carestream CR Systems are designed to meet international safety and performance standards. Personnel operating

More information

National Institute of Radiological Sciences. Naoya Saotome

National Institute of Radiological Sciences. Naoya Saotome National Institute of Radiological Sciences Naoya Saotome 1 Contents Introduction History and collaboration KCC i-rock Commissioning of commercial scanning system NIRS Gantry Commissioning of NIRS s Gantry

More information

Optima CT660 Optima CT660

Optima CT660 Optima CT660 Product Data Sheet Rev 2 Page 1/18 Optima CT660 Optima CT660 Milwaukee, USA - Fax: 1 262 544 3384 Tokyo, Japan - Fax: 81 425 85 5490 Paris, France - Fax: 33 1 30 70 94 35 Optima TM CT660 Product Description

More information

FFDM Quality Control in Canada - a Vendor Neutral Approach

FFDM Quality Control in Canada - a Vendor Neutral Approach FFDM Quality Control in Canada - a Vendor Neutral Approach Rasika Rajapakshe, PhD, FCCPM BC Cancer Agency-Center for the Southern Interior Kelowna, BC, Canada 2011 Joint AAPM/COMP Meeting August 4, 2011

More information

Guidelines for Quality Assurance in Radiation Protection for Diagnostic X-Ray Facilities: Large X-Ray Facilities

Guidelines for Quality Assurance in Radiation Protection for Diagnostic X-Ray Facilities: Large X-Ray Facilities Guidelines for Quality Assurance in Radiation Protection for Diagnostic X-Ray Facilities: Large X-Ray Facilities J L Poletti 1995/1 NRL Report GUIDELINES FOR QUALITY ASSURANCE IN RADIATION PROTECTION FOR

More information

Breast Imaging Essentials

Breast Imaging Essentials Breast Imaging Essentials Module 10 Transcript 2016 ASRT. All rights reserved. Breast Imaging Essentials Module 10 Digital Breast Tomosynthesis 1. ASRT Animation 2. Welcome Welcome to Module 10 of Breast

More information

World First Slim Cassette Type Digital Mammo. Upgrade Solution

World First Slim Cassette Type Digital Mammo. Upgrade Solution World First Slim Cassette Type Digital Mammo. Upgrade Solution Still Analog? Slim Cassette Type Digital Mammo. Upgrade Solution Contents 3 P. 5 P. 7 P. 9 P. 11 P. 13 P. 1824C & 2430C, easy digitalization

More information

Y.XST225-VF. INTERCONTROLE Escoffier 1 XYLON MG225VF RX PDS

Y.XST225-VF. INTERCONTROLE Escoffier 1 XYLON MG225VF RX PDS Y.XST225-VF 1 XYLON MG225VF - 0509- RX0905007PDS 1 Table of contents Y.XST225-VF High detail visibility - examples Y.XST225-VF vs. conventional X-ray systems Y.XST225-VF vs. µ-focus X-ray systems Unique

More information

INSTRUMENT CATHODE-RAY TUBE

INSTRUMENT CATHODE-RAY TUBE Instrument cathode-ray tube D14-363GY/123 INSTRUMENT CATHODE-RAY TUBE mono accelerator 14 cm diagonal rectangular flat face internal graticule low power quick heating cathode high brightness, long-life

More information

High Frequency X-Ray Technology. Wide Screen Control with Touch-Technology

High Frequency X-Ray Technology. Wide Screen Control with Touch-Technology High Frequency X-Ray Technology Wide Screen Control with Touch-Technology The Q-VISION HF Series TM The Q-VISION Series TM integrates innovative high frequency design along with superb functionality, resulting

More information

DOSE DELIVERY SYSTEM OF THE VARIAN PROBEAM SYSTEM WITH CONTINUOUS BEAM

DOSE DELIVERY SYSTEM OF THE VARIAN PROBEAM SYSTEM WITH CONTINUOUS BEAM DOSE DELIVERY SYSTEM OF THE VARIAN PROBEAM SYSTEM WITH CONTINUOUS BEAM EUCARD 2 WORKSHOP ON INNOVATIVE DELIVERY SYSTEMS IN PARTICLE THERAPY TORINO, 23 25 FEB 2017 VARIAN PARTICLE THERAPY HOLGER GÖBEL MANGER

More information

Quality Assurance Implementation at the Roberts Proton Therapy Center. James McDonough 3 August 2013

Quality Assurance Implementation at the Roberts Proton Therapy Center. James McDonough 3 August 2013 Quality Assurance Implementation at the Roberts Proton Therapy Center James McDonough 3 August 2013 1 Roberts Proton Therapy Center Machine configuration and layout 4 gantries, 1 fixed beam line, 1 research

More information

The Black Piranha Easy & Fast X-ray Quality Control

The Black Piranha Easy & Fast X-ray Quality Control From Radiation to Information The Black Piranha Easy & Fast X-ray Quality Control The Black Piranha with accessories makes X-ray QA easy and fast. Connection is automatic just plug n play. The Quick Check

More information

Linatron - M9 & M9A. Modular high-energy X-ray source. 2.0 Performance

Linatron - M9 & M9A. Modular high-energy X-ray source. 2.0 Performance The Linatron -M is a modular system. The control console, modulator, and RF unit are common to all model configurations. Only the X-ray head changes to match the application. The Linatron - M is designed

More information

X-Ray Machines, CT Scanners, MRIs: The Pivotal Role of the GE Research and Development Center

X-Ray Machines, CT Scanners, MRIs: The Pivotal Role of the GE Research and Development Center CT Scanner Protoype at UCSF Medical Center 1976 X-Ray Machines, CT Scanners, MRIs: The Pivotal Role of the GE Research and Development Center by Walter L. Robb Early Years In 1895, Professor Wilhelm Conrad

More information

BEAMAGE 3.0 KEY FEATURES BEAM DIAGNOSTICS PRELIMINARY AVAILABLE MODEL MAIN FUNCTIONS. CMOS Beam Profiling Camera

BEAMAGE 3.0 KEY FEATURES BEAM DIAGNOSTICS PRELIMINARY AVAILABLE MODEL MAIN FUNCTIONS. CMOS Beam Profiling Camera PRELIMINARY POWER DETECTORS ENERGY DETECTORS MONITORS SPECIAL PRODUCTS OEM DETECTORS THZ DETECTORS PHOTO DETECTORS HIGH POWER DETECTORS CMOS Beam Profiling Camera AVAILABLE MODEL Beamage 3.0 (⅔ in CMOS

More information

Minimize your cost for Phased Array & TOFD

Minimize your cost for Phased Array & TOFD Minimize your cost for Phased Array & TOFD Latest ultrasonic flaw detector from SIUI, incorporates the latest advancements in Encoder In/Out UT/ TOFD Probe high-performance Phased Array and TOFD detection

More information

BCA 9S. ELECTRICAL CHARACTERISTICS Voltage 230 Vac ±10% monophase standard 105 / 115 / 125 / 220 / 240 Vac ±10% monophase on request

BCA 9S. ELECTRICAL CHARACTERISTICS Voltage 230 Vac ±10% monophase standard 105 / 115 / 125 / 220 / 240 Vac ±10% monophase on request ELECTRICAL CHARACTERISTICS Voltage 230 Vac ±10% monophase standard 105 / 115 / 125 / 220 / 240 Vac ±10% monophase on request Frequency 50 Hz standard 60 Hz on request Absorbed current on Stationary 4.5

More information

Built to do More. Brivo CT slice CT product data sheet

Built to do More. Brivo CT slice CT product data sheet GE Healthcare Built to do More. Brivo CT385 16-slice CT product data sheet Built for your Confidence. BIG Performance, small footprint Smart CT Desk Energy Saving Mode Simply Advance Hilight Scintillator

More information

New appraoch for X-ray weld inspection of pipeline segments

New appraoch for X-ray weld inspection of pipeline segments New appraoch for X-ray weld inspection of pipeline segments Lennart Schulenburg VisiConsult X-ray Systems & Solutions GmbH 1 Overview Weld inspection in heavy industries ( Pipe and Tank ) Analogue Film

More information

Case 1 8/1/2017. Ring Artifact? James M. Kofler, Ph.D. Mayo Clinic Rochester. AAPM Annual Meeting TU-B Artifacts: CT 2

Case 1 8/1/2017. Ring Artifact? James M. Kofler, Ph.D. Mayo Clinic Rochester. AAPM Annual Meeting TU-B Artifacts: CT 2 James M. Kofler, Ph.D. Mayo Clinic Rochester AAPM Annual Meeting 2017 1 Case 1 Ring Artifact? 2 3 1 Star Pattern caused by metal earrings! 4 Same star artifact commonly caused by Dental Amalgam Metal Implants

More information

Australian/New Zealand Standard

Australian/New Zealand Standard AS/NZS 3821:1998 IEC 60406:1997 Australian/New Zealand Standard Cassettes for medical X-ray diagnosis Radiographic cassettes and mammographic cassettes AS/NZS 3821:1998 This Joint Australian/New Zealand

More information

Dental Line. High frequency intraoral x-ray unit. radiology ahead

Dental Line. High frequency intraoral x-ray unit. radiology ahead Dental Line High frequency intraoral x-ray unit radiology ahead enhance your operating process Hi-frequency generator, for shorter exposures and reduced dose Wireless X-ray pushbutton, available for ultimate

More information

BCA 9R. 60 Hz on request Absorbed current on Stationary Vac and Vac in fluoro mode

BCA 9R. 60 Hz on request Absorbed current on Stationary Vac and Vac in fluoro mode ELECTRICAL CHARACTERISTICS RADIOLOGICAL DATA MONOBLOC Voltage 230 Vac ±10% monophase standard 105 / 115 / 125 / 220 / 240 Vac ±10% monophase on request Frequency 50 Hz standard 60 Hz on request Absorbed

More information

Selection Criteria for X-ray Inspection Systems for BGA and CSP Solder Joint Analysis

Selection Criteria for X-ray Inspection Systems for BGA and CSP Solder Joint Analysis Presented at Nepcon Shanghai 2003 Abstract Selection Criteria for X-ray Inspection Systems for BGA and CSP Solder Joint Analysis Dr. David Bernard, Dage Precision Industries, 158-29 Hua Shan Road, Feng

More information

DICOM Correction Proposal

DICOM Correction Proposal DICOM Correction Proposal STATUS Final Text Date of Last Update 2015/11/16 Person Assigned Submitter Name Ulrich Busch (ulrich.busch@varian.com) Michael Moyers Submission Date 2011/05/19 Correction Number

More information

CHAPTER 4: HIGH ENERGY X-RAY GENERATORS: LINEAR ACCELERATORS. Jason Matney, MS, PhD

CHAPTER 4: HIGH ENERGY X-RAY GENERATORS: LINEAR ACCELERATORS. Jason Matney, MS, PhD CHAPTER 4: HIGH ENERGY X-RAY GENERATORS: LINEAR ACCELERATORS Jason Matney, MS, PhD Objectives Medical electron linear accelerators (often shortened to LINAC) The Basics Power Supply Magnetron/Klystron

More information

INSTRUMENT CATHODE-RAY TUBE

INSTRUMENT CATHODE-RAY TUBE INSTRUMENT CATHODE-RAY TUBE 14 cm diagonal rectangular flat face domed mesh post-deflection acceleration improved spot quality for character readout high precision by internal permanent magnetic correction

More information

Collimation and Light Field Brightness

Collimation and Light Field Brightness Collimation and Light Field Brightness Physics and Reasoning Principle 1: Primary X Ray beam is blocked by breast support table. Principle 2: Need to properly imageas as much breast tissue as possible.

More information

Implementing a Proton Beam Scanning System within an Operating Clinical Facility

Implementing a Proton Beam Scanning System within an Operating Clinical Facility Implementing a Proton Beam Scanning System within an Operating Clinical Facility Ben Clasie Many thanks to Hassan Bentefour, Hanne Kooy, and Jay Flanz for their help preparing this presentation 1 Francis

More information

Essentials of Digital Imaging

Essentials of Digital Imaging Essentials of Digital Imaging Module 3 Transcript 2016 ASRT. All rights reserved. Essentials of Digital Imaging Module 3 Display 1. ASRT Animation 2. Welcome Welcome to the Essentials of Digital Imaging

More information

A new shielding calculation method for X-ray computed tomography regarding scattered radiation

A new shielding calculation method for X-ray computed tomography regarding scattered radiation Radiol Phys Technol (2017) 10:213 226 DOI 10.1007/s12194-016-0387-9 A new shielding calculation method for X-ray computed tomography regarding scattered radiation Hiroshi Watanabe 1,2 Kimiya Noto 3 Tomokazu

More information

PRACTICAL APPLICATION OF THE PHASED-ARRAY TECHNOLOGY WITH PAINT-BRUSH EVALUATION FOR SEAMLESS-TUBE TESTING

PRACTICAL APPLICATION OF THE PHASED-ARRAY TECHNOLOGY WITH PAINT-BRUSH EVALUATION FOR SEAMLESS-TUBE TESTING PRACTICAL APPLICATION OF THE PHASED-ARRAY TECHNOLOGY WITH PAINT-BRUSH EVALUATION FOR SEAMLESS-TUBE TESTING R.H. Pawelletz, E. Eufrasio, Vallourec & Mannesmann do Brazil, Belo Horizonte, Brazil; B. M. Bisiaux,

More information

MODE FIELD DIAMETER AND EFFECTIVE AREA MEASUREMENT OF DISPERSION COMPENSATION OPTICAL DEVICES

MODE FIELD DIAMETER AND EFFECTIVE AREA MEASUREMENT OF DISPERSION COMPENSATION OPTICAL DEVICES MODE FIELD DIAMETER AND EFFECTIVE AREA MEASUREMENT OF DISPERSION COMPENSATION OPTICAL DEVICES Hale R. Farley, Jeffrey L. Guttman, Razvan Chirita and Carmen D. Pâlsan Photon inc. 6860 Santa Teresa Blvd

More information

Automatic Defect Recognition in Industrial Applications

Automatic Defect Recognition in Industrial Applications Automatic Defect Recognition in Industrial Applications Klaus Bavendiek, Frank Herold, Uwe Heike YXLON International, Hamburg, Germany INDE 2007 YXLON. The reason why 1 Different Fields for Usage of ADR

More information

G. Pittá(*), S. Braccini TERA Foundation, Novara, Italy (*) Corresponding author.

G. Pittá(*), S. Braccini TERA Foundation, Novara, Italy (*) Corresponding author. Frascati Physics Series Vol. VVVVVV (xxxx), pp. 000-000 XX Conference Location, Date-start - Date-end, Year MATRIX: AN INNOVATIVE PIXEL IONIZATION CHAMBER FOR ON-LINE BEAM MONITORING IN HADRONTHERAPY G.

More information

Data Collection Using APEX3. March 30, Chemical Crystallography Laboratory

Data Collection Using APEX3. March 30, Chemical Crystallography Laboratory Data Collection Using APEX3 Page 1 of 10 Data Collection Using APEX3 March 30, 2017 Chemical Crystallography Laboratory Author: Douglas R. Powell Data Collection Using APEX3 Page 2 of 10 Distribution Douglas

More information

Nuclear Associates and

Nuclear Associates and Nuclear Associates 76-907 and 76-908 AAPM MRI Phantoms Users Manual March 2005 Manual No. 38616 Rev. 3 2003, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks

More information

CARESTREAM DIRECTVIEW Elite CR System

CARESTREAM DIRECTVIEW Elite CR System CARESTREAM DIRECTVIEW Elite CR System Improve workflow, productivity, and patient throughput. The CARESTREAM DIRECTVIEW Elite CR System is small, easy to install and easy to use. This powerful distributed

More information

Drift Tubes as Muon Detectors for ILC

Drift Tubes as Muon Detectors for ILC Drift Tubes as Muon Detectors for ILC Dmitri Denisov Fermilab Major specifications for muon detectors D0 muon system tracking detectors Advantages and disadvantages of drift chambers as muon detectors

More information

CARESTREAM DIRECTVIEW Elite CR System

CARESTREAM DIRECTVIEW Elite CR System CARESTREAM DIRECTVIEW Elite CR System Improve workflow, productivity, and patient throughput. The CARESTREAM DIRECTVIEW Elite CR System is small, easy to install and easy to use. This powerful distributed

More information

Development of New Carbon Therapy Facility and Future Plan of HIMAC

Development of New Carbon Therapy Facility and Future Plan of HIMAC Development of New Carbon Therapy Facility and Future Plan of Koji Noda Research Center for Charged Particle Therapy National Institute of Radiological Sciences 1. Introduction Contents 2. New Carbon-Therapy

More information

Overview. ACR Accreditation Update in Mammography. ACR Topics. Requirements Today. What s Coming For Tomorrow

Overview. ACR Accreditation Update in Mammography. ACR Topics. Requirements Today. What s Coming For Tomorrow ACR Accreditation Update in Mammography Eric Berns, PhD University of Colorado Hospital Denver Health Medical Center Denver, CO *No financial disclosures to report Overview ACR Topics Requirements Today

More information

ISOMET. Compensation look-up-table (LUT) and How to Generate. Isomet: Contents:

ISOMET. Compensation look-up-table (LUT) and How to Generate. Isomet: Contents: Compensation look-up-table (LUT) and How to Generate Contents: Description Background theory Basic LUT pg 2 Creating a LUT pg 3 Using the LUT pg 7 Comment pg 9 The compensation look-up-table (LUT) contains

More information

What to consider when choosing a mammography display

What to consider when choosing a mammography display What to consider when choosing a mammography display Screen size and resolution In digital breast imaging, the quality of the medical display has a direct impact on the decisions you make. Next to display

More information

LightSpeed 16 With Xtream CT Scanner System

LightSpeed 16 With Xtream CT Scanner System Page 1 Introduction The CT Scanner with Xtream TM technology is the next step in GE s CT Continuum and its awardwinning LightSpeed CT platform. The CT Scanner routinely acquires 16 sub-millimeter slices

More information

CHAPTER 4 OSCILLOSCOPES

CHAPTER 4 OSCILLOSCOPES CHAPTER 4 OSCILLOSCOPES 4.1 Introduction The cathode ray oscilloscope generally referred to as the oscilloscope, is probably the most versatile electrical measuring instrument available. Some of electrical

More information

Practical Application of the Phased-Array Technology with Paint-Brush Evaluation for Seamless-Tube Testing

Practical Application of the Phased-Array Technology with Paint-Brush Evaluation for Seamless-Tube Testing ECNDT 2006 - Th.1.1.4 Practical Application of the Phased-Array Technology with Paint-Brush Evaluation for Seamless-Tube Testing R.H. PAWELLETZ, E. EUFRASIO, Vallourec & Mannesmann do Brazil, Belo Horizonte,

More information

Display Quality Assurance: Considerations When Establishing a Display QA Program. Mike Silosky, M.S. 8/3/2017

Display Quality Assurance: Considerations When Establishing a Display QA Program. Mike Silosky, M.S. 8/3/2017 Display Quality Assurance: Considerations When Establishing a Display QA Program Mike Silosky, M.S. 8/3/2017 Objectives and Outline Why, Who, What, When, Where? Discuss the resources that may be needed

More information

Standard Operating Procedure of nanoir2-s

Standard Operating Procedure of nanoir2-s Standard Operating Procedure of nanoir2-s The Anasys nanoir2 system is the AFM-based nanoscale infrared (IR) spectrometer, which has a patented technique based on photothermal induced resonance (PTIR),

More information

Requirements for the Beam Abort Magnet and Dump

Requirements for the Beam Abort Magnet and Dump Requirements for the Beam Abort Magnet and Dump A beam abort kicker (pulsed dipole magnet) and dump are required upbeam of the LCLS undulator in order to protect the undulator from mis-steered and poor

More information

Comed Medical Systems Co., Ltd. Office 707, Woolim Lion s Valley I, 311-3, Sangdaewon-dong, Seongnam-si, Gyeonggi-do, Korea Tel:

Comed Medical Systems Co., Ltd. Office 707, Woolim Lion s Valley I, 311-3, Sangdaewon-dong, Seongnam-si, Gyeonggi-do, Korea Tel: Comed Medical Systems Co., Ltd. Office 707, Woolim Lion s Valley I, 311-3, Sangdaewon-dong, Seongnam-si, Gyeonggi-do, Korea Tel:+82-31-737-2211 Fax:+82-31-737-2210 Factory 236-4 Sangdaewon-dong, Seongnam-si,

More information

Screen investigations for low energetic electron beams at PITZ

Screen investigations for low energetic electron beams at PITZ 1 Screen investigations for low energetic electron beams at PITZ S. Rimjaem, J. Bähr, H.J. Grabosch, M. Groß Contents Review of PITZ setup Screens and beam profile monitors at PITZ Test results Summary

More information

Imaging diagnostico in Sanità Stato attuale e prospettive

Imaging diagnostico in Sanità Stato attuale e prospettive Imaging diagnostico in Sanità Stato attuale e prospettive Sandro Paini AMI & Oncology Pisa 20.12.2016 Fully digital SiPM (dsipm ) invented within Philips Research PMT APD(analog) SiPM(Analog) dsipm( Digital)

More information

Display Quality Assurance: Recommendations from AAPM TG270 for Tests, Tools, Patterns, and Performance Criteria

Display Quality Assurance: Recommendations from AAPM TG270 for Tests, Tools, Patterns, and Performance Criteria Display Quality Assurance: Recommendations from AAPM TG270 for Tests, Tools, Patterns, and Performance Criteria Nicholas B. Bevins, Ph.D. TG270 Co-chair Display Check 2 TG270 Goals Provide an update to

More information

Display Quality Assurance: Recommendations from AAPM TG270 for Tests, Tools, Patterns, and Performance Criteria

Display Quality Assurance: Recommendations from AAPM TG270 for Tests, Tools, Patterns, and Performance Criteria Display Quality Assurance: Recommendations from AAPM TG270 for Tests, Tools, Patterns, and Performance Criteria Nicholas B. Bevins, Ph.D. TG270 Co-chair Display Check 2 1 TG270 Goals Provide an update

More information

SEAMLESS INTEGRATION MULTIPLE SOLUTIONS ONE PRODUCT ONE COMPANY

SEAMLESS INTEGRATION MULTIPLE SOLUTIONS ONE PRODUCT ONE COMPANY SEAMLESS INTEGRATION ONE COMPANY ONE PRODUCT MULTIPLE SOLUTIONS SEAMLESS INTEGRATION ONE COMPANY - ONE PRODUCT - MULTIPLE SOLUTIONS Introduction As leaders in X-Ray Generator Technology we are utilizing

More information