Operation Procedure for Phillips XL30 ESEM

Size: px
Start display at page:

Download "Operation Procedure for Phillips XL30 ESEM"

Transcription

1 Operation Procedure for Phillips XL30 ESEM The ESEM will be left in the ON state when not in use. The chamber will be at high vacuum, filament on, stage at home position, VAC and HT buttons lit, and monitor turned off. If these conditions are not met, make note of it in the log book. NOTE: When operating the ESEM, if you see any type of error message, make a not of it in the log book. There are many errors that will cause the system to turn off the diffusion pump, so check the VAC button whenever you see an error. If it is not lit then the diffusion pump is off. After making a note of the error message and clicking OK, if the error does not appear again (system returned to its normal state) you may turn on the pump by pressing the VAC button. You will need to wait for the diffusion pump to reheat to operating temperature. Before turning on the beam you will also need to press the HT button to enable high voltage. Important: the SEM computer hard drive should not be considered a back up storage space for your SEM images. Always make your own back ups and then delete all old images on the SEM computer. Go to the end of the procedure to see the Step by Step Operation Guide High Vacuum Operation: 1. Begin your entry in the log book by entering the date, your name, your PI or company name, the Operate current and the SEM session start time. 2. Swipe in to the SEM Start Card reader. 3. Open the Microscope Control software, Mctrl from the desktop. 4. Open the image database software, XL Docu from the desktop. a. Delete all old images in your XL Docu database. b. Click on the light bulb in the Always on top tool bar to toggle between Microscope Control and XL Docu. 5. In Mctrl click on VENT to vent the chamber. The chamber will fill with nitrogen. If the nitrogen cylinder is empty you may change the regulator over to the back up cylinder using the wrench on the cylinder rack. Always mark the empty cylinder with an X on the UCSD Storehouse ID tag and make a note that you changed the N2 in the log book. 6. Open the chamber door and load your samples. Make a mental note of where your samples are, how they are oriented and consider what areas you would like to examine. Remember, the chamber door will be to the left when viewing the SEM image on the screen. 7. If you plan to use the EDS detector, make sure you are aware of its position relative to your samples and make sure that the region of interest is not is not shadowed or blocked from EDS detector by some other part of your sample. 8. If you plan to use the BSE detector, slide it onto the bullet at the end of the pole piece and make sure that it is properly centered. Otherwise it will block the beam.

2 9. Close the chamber door (make sure it s all the way closed!!) and click on Pump. When closing the chamber door, be sure not to slam it shut. 10. When the chamber pressure reaches 1.3e-4mbar the vacuum status will read Vac OK. 11. The default microscope settings are: 10KV beam energy, SE detector, spot 3, TV scan rate averaging 8 frames. If necessary, choose the appropriate detector, beam energy, spot size and scan speed. These parameters can be changed after the beam is already on. 12. Select the desired navigation mode. a. Get mode will use the stage motor to move the scan frame location below 10KX. Above 10KX Get mode will shift the beam to move the scan frame location. When using get mode above 10KX, you may be prompted to allow automatic reset of the beam shift if it has reached its limit. b. Shift mode will shift the beam to move the scan frame and not use the stage motor. Shift mode is best used at magnifications above 30KX since the beam shift range is small. At low magnifications you will reach the beam shift limit almost immediately. You will here a beeping sound when the shift limit has been reached. c. Track mode will use the stage motor to move the scan frame and will not shift the beam. Track mode is best used at low magnifications since at high magnifications the stage motor steps become visible and are comparable or larger than the scan frame size. 13. To change the control panel on the right side of the screen click on Settings. The resulting drop down list shows the control panel display options. Within each control panel display option there are different sections which may also be expanded to access more functions and parameters. Do not press the Filament OFF button!! 14. In the settings control panel, turn on the beam by clicking the button that displays the beam voltage in the Beam portion of the settings panel on the right side of the screen. If the screen is too dark, increase the contrast and/or brightness until you see an image. 15. When you turn on the beam, the Confirm microscope focus pop up window will appear. Do not click OK until you have completed focus confirmation! The system does not know the height of the samples which are loaded in the chamber so, in order to avoid crashing your samples into the pole piece while adjusting the stage height, we must tell the system the maximum height of the samples that have been loaded. We do this by confirming focus at the tallest point of the sample: a. Drive to the tallest point of the samples you have loaded b. Press the right click button of the mouse and drag to the left and right to adjust the focus until the image is clear. c. Increase the magnification to 3000X or higher by pressing the + key on the number pad or selecting a magnification value from the drop down menu. Readjust the focus until the image is clear. d. Now the focal point of the beam is at the sample surface. The system knows the distance from the pole piece to the focal point of the beam and this is indicated on the databar as the working distance (WD), measured in millimeters.

3 e. Click Ok to confirm that the image is in focus. The value of the stage height Z will be replaced by the current value of WD. f. It is imperative that you focus on the highest point of the sample (+-1mm), otherwise you may crash your sample into the pole piece! 16. Set the stage height to 10mm. Now the location at which you confirmed focus will be 10mm below the pole piece. Zoom out and readjust the focus. 17. Drive to the region you wish to examine. 18. Secondary Electron Imaging techniques: a. Use TV scan rate averaging eight frames when imaging with the secondary electron detector. b. Careful attention should be given to the choice of beam acceleration voltage. Different materials will respond in different ways to the beam. Try a variety of beam energies, starting low, to see what gives the best image results. Keep in mind that the beam may damage certain materials. c. Each time you change the beam energy you may also need to adjust the other imaging parameters listed below. d. Adjust the brightness and contrast so that the image contains as many grey levels as possible. The ACB (Auto Contrast and Brightness) button may be used as a starting point. e. Increase or decrease the magnification using the + and keys on the number pad, selecting a preset magnification from the drop down menu or using the slider bar in the imaging panel. f. Adjust the focus by pressing the right click button of the mouse and dragging left or right. A good strategy is to focus the image at twice the magnification at which you intend to capture the image and then zoom back out. Once the image is in clear focus you can zoom out and the image will be in good focus at lower magnifications. g. As you adjust the focus, watch for the features of your sample surface to elongate in perpendicular directions. This indicates astigmatism. If the image simply blurs with no distortion when you adjust the focus in and out, the stigmator may not need adjustment. h. Before adjusting the stigmator you must be certain that the focus is as at the center point. You can identify the ideal position as the exact mid point between the elongation of the image features in either direction as you move the mouse left and right adjusting the focus. Use a round particle for this adjustment. i. To correct astigmatism hold the shift key, press and hold the right click button of the mouse. Green cross hairs will appear. First move the cross hairs in the horizontal (x) direction left or right. Watch for the image quality to steadily improve. If it worsens, move the other way. If you reach the edge of the screen and need further adjustment, release the right click button and while still holding the shift key, press the right click button again. This will reset the position of the cross hairs to the middle of the screen. Between each stigmator adjustment, readjust the focus as well. Repeat this procedure in the y axis. j. For magnifications above 70KX a smaller spot size such as spot 1 or 2 should be selected. At low magnifications a larger spot may be used. Try different spot

4 sizes to see which works best. When using small spot sizes slower scan speeds can increase the signal to noise ratio and make the image clearer. 19. Image capture: a. In TV scan mode the image can be noisy. To increase the signal to noise ratio and get a clear image it is necessary to use a slower scan speed. b. When you are ready to capture the image, press F2 for a single frame scan. c. To adjust the single scan rate, go to the Scan drop down menu and select Change. Select Single scan from the list of presets and set the desired line time and lines per frame. The normal settings for the single scan are 33.2ms line time and 2420 lines per frame. d. After the single scan is complete the image will be frozen. You can save the image by clicking on the file box button on the XL Docu Always on top tool bar. e. Unfreeze the image by clicking on the snowflake button to go back to TV scan and continue examining your sample. 20. End procedure: a. Turn off the beam. b. Go to the Stage drop down menu and select Home to home the stage. c. While the stage is homing, click Vent. d. While the stage is homing and the chamber is venting you can save your images to disc. Click on the light bulb in the XL Docu tool bar to toggle to XL Docu. e. Select all images captured today. f. Click on either the J to export your images as.jpeg files or the T to export your images as.tif files. g. Select the appropriate file path, click Ok. The images will be copied to the specified folder. Be sure to delete old images over time. h. Close XL Docu i. Copy your images to a USB drive. 21. Open the chamber door and unload your samples. 22. If you used the BSE detector, place it back into the holder. 23. Close the chamber door and click on Pump. 24. Close Microscope Control 25. Swipe the Finish SEM card reader 26. Finish your log book entry by entering the session end time. If you encountered any issues during your session, make a note in the log book. 27. Make sure the SEM console is not cluttered. Do not leave a mess! ESEM Step by Step Operation Guide 1. Make log book entry 2. Swipe in 3. Open Microscope Control 4. Vent the chamber 5. Load sample 6. Pump, make sure the chamber door is closed

5 7. Open your XL Docu database (.mdb file) 8. Delete old images 9. Click on the light bulb 10. Maximize Microscope Control 11. Wait for Vac OK 12. Turn on beam 13. Focus at tallest point on sample at 3000X magnification 14. Click OK to confirm focus 15. Set Z=10mm, click Go To. 16. Refocus 17. Find area of interest 18. Zoom to 30KX and find a round particle 19. Focus, stigmate 20. Zoom in to desired magnification, find smaller particle if necessary 21. Focus, stigmate 22. Press F2 for single slow scan 23. Save image with file box button on XL Docu Always on top bar 24. Unfreeze the image When finished imaging your sample: 1. Turn beam off 2. Home the stage 3. Vent 4. Click light bulb to go to XL Docu 5. Select all images 6. Choose J or T (.jpeg or.tif) to export images to your folder on the data storage computer 7. Copy images to your USB drive 8. Unload sample 9. Pump, make sure the chamber door is closed 10. Log off of SEM computer 11. Complete your log book entry with the end time of the session 12. Swipe out In case of PLCB Error: Vacuum button disabled 1. Make note of the text in the error message, ok the error. If the error does not reappear, then the system was able to reach normal configuration 2. Check the VAC and HT buttons on the front panel below the SEM chamber. If they are not lit then the pump has shut off, press the VAC button to turn the pump back on. It may take up to 45 minutes for the diffusion pump to reach operating temperature. 3. When the vacuum status reads Idle, press the HT button and the SEM is ready for use

FIB Operating Procedure. Effective Date: 08/14/2012 Author(s): Jiong Hua Phone:

FIB Operating Procedure. Effective Date: 08/14/2012 Author(s): Jiong Hua Phone: FIB Operating Procedure Effective Date: 08/14/2012 Author(s): Jiong Hua Phone: 402-472-3773 Email: jhua2@unl.edu 1 1 Introduction 1.1 Key Words Focused Ion Beam (FIB), FEI Strata 201, Ion milling 1.2 Purpose

More information

Scanning Electron Microscopy (FEI Versa 3D Dual Beam)

Scanning Electron Microscopy (FEI Versa 3D Dual Beam) Scanning Electron Microscopy (FEI Versa 3D Dual Beam) This operating procedure intends to provide guidance for basic measurements on a standard sample with FEI Versa 3D SEM. For more advanced techniques

More information

University of Minnesota Nano Fabrication Center Standard Operating Procedure

University of Minnesota Nano Fabrication Center Standard Operating Procedure Equipment Name: Focused Ion Beam (FIB) Coral Name: fib Revision Number: 2 Model: FEI Quanta 200 3D Revisionist: Kevin Roberts Location: Area 3 Date: 9/17/2013 1 Description The Quanta 200 3D is a DualBeam

More information

Technical Procedure for Scanning Electron Microscope/ Energy Dispersive X-Ray System (SEM/EDX) for non-gsr Casework

Technical Procedure for Scanning Electron Microscope/ Energy Dispersive X-Ray System (SEM/EDX) for non-gsr Casework Technical Procedure for Scanning Electron Microscope/ Energy Dispersive X-Ray System (SEM/EDX) for non-gsr Casework 1.0 Purpose This technical procedure shall be followed for the operation of the Scanning

More information

Standard Operating Procedure II: EDS (Bruker Flat-Quad)

Standard Operating Procedure II: EDS (Bruker Flat-Quad) Standard Operating Procedure II: EDS (Bruker Flat-Quad) ywcmatsci.yale.edu ESC II, Room A119F 810 West Campus Drive West Haven, CT 06516 Version 1.1, October 2018 1 > FOLLOW the SOP strictly to keep the

More information

AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode)

AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode) AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode) 1. Log into the Log Usage system on the SMIF web site 2. Open Nanoscope 6.14r1 software by double clicking on the Nanoscope 6.14r1 desktop

More information

1.2 Universiti Teknologi Brunei (UTB) reserves the right to award the tender in part or in full.

1.2 Universiti Teknologi Brunei (UTB) reserves the right to award the tender in part or in full. TENDER SPECIFICATIONS FOR THE SUPPLY, DELIVERY, INSTALLATION AND COMMISSIONING OF ONE UNIT OF VARIABLE PRESSURE ENVIRONMENTAL SCANNING ELECTRON MICROSCOPE (SEM) CUM ENERGY DISPERSIVE SPECTROSCOPY (EDS)

More information

PHI 5000 VersaProbe TM Operator s Guide

PHI 5000 VersaProbe TM Operator s Guide PHI 5000 VersaProbe TM Operator s Guide Part No. 705921 Rev. A Copyright 2006 ULVAC-PHI, INC. 370 Enzo, Chigasaki, JAPAN The PHI logo ( ) is a registered trademark of ULVAC-PHI, INC. Physical Electronics,

More information

Dektak Step by Step Instructions:

Dektak Step by Step Instructions: Dektak Step by Step Instructions: Before Using the Equipment SIGN IN THE LOG BOOK Part 1: Setup 1. Turn on the switch at the back of the dektak machine. Then start up the computer. 2. Place the sample

More information

Standard Operating Procedure for FEI Helios 660 NanoLab Part I: SEM Version

Standard Operating Procedure for FEI Helios 660 NanoLab Part I: SEM Version Standard Operating Procedure for FEI Helios 660 NanoLab Part I: SEM Version Helios reservations may be made online using the NERCF website. Note: Always wear gloves when venting the system and exchanging

More information

CHECKLIST FOR VERIOS OPERATION 1. GENERAL The SEM lab is used assuming "operating room" cleanliness, i.e., the SEM lab is a high visibility lab and

CHECKLIST FOR VERIOS OPERATION 1. GENERAL The SEM lab is used assuming operating room cleanliness, i.e., the SEM lab is a high visibility lab and CHECKLIST FOR VERIOS OPERATION 1. GENERAL The SEM lab is used assuming "operating room" cleanliness, i.e., the SEM lab is a high visibility lab and must be kept clean and neat so clean up behind yourself

More information

University of Minnesota Minnesota Nano Center Standard Operating Procedure

University of Minnesota Minnesota Nano Center Standard Operating Procedure Equipment Name: CHA Evaporator Coral Name: ebevap-cha Revision Number: 12 Model: SEC 600 Revisionist: L. von Dissen Location: PAN Bay 3 Date: 04/13/2018 1 Description The CHA Evaporator is a single source

More information

INCA ENERGY EDS TRAINING. System Block Diagram. INCA Energy Software. Xiang Yang EM SMU. Navigators. Point & ID Navigator.

INCA ENERGY EDS TRAINING. System Block Diagram. INCA Energy Software. Xiang Yang EM SMU. Navigators. Point & ID Navigator. INCA ENERGY EDS TRAINING Xiang Yang EM Center @ SMU System Block Diagram INCA Energy Software Navigators Point & ID Navigator Data tree 1 Spectrum Export Bitmap TIFF JPEG Metafile EMSA ISIS Optimum Conditions

More information

1. Check the accelerating voltage, must be at 200 kv (right screen), HT (µa) (left panel) at and Emission (left panel) at

1. Check the accelerating voltage, must be at 200 kv (right screen), HT (µa) (left panel) at and Emission (left panel) at JEOL 2010F MANUAL Quick check list 1. Check the accelerating voltage, must be at 200 kv (right screen), HT (µa) (left panel) at 0.96-0.97 and Emission (left panel) at 155-160. 2. Check the vacuum sequence

More information

Quick Start Bruker Dimension Icon AFM

Quick Start Bruker Dimension Icon AFM Do not remove Quick Start Bruker Dimension Icon AFM March 3, 2015 GLA Contacts Harold Fu (hfu@caltech.edu) Weilai Yu (wyyu@caltech.edu) Bruker Tech Support (AFMSupport@bruker-nano.com 800-873-9750) Watch

More information

2.1. Log on to the TUMI system (you cannot proceed further until this is done).

2.1. Log on to the TUMI system (you cannot proceed further until this is done). FEI DB235 ex-situ lift out TEM sample preparation procedure Nicholas G Rudawski ngr@ufledu (805) 252-4916 Last updated: 06/19/15 DISCLAIMER: this procedure describes one specific method for preparing ex-situ

More information

Approved by: / / R. Battaglia 12/16/2016

Approved by: / / R. Battaglia 12/16/2016 Fabrication Laboratory Revision: H Rev Date: 12/16/16 Approved by: Process Engineer / / R. Battaglia 12/16/2016 Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Varian

More information

LUMIGEN INSTRUMENT CENTER X-RAY CRYSTALLOGRAPHIC LABORATORY: WAYNE STATE UNIVERSITY

LUMIGEN INSTRUMENT CENTER X-RAY CRYSTALLOGRAPHIC LABORATORY: WAYNE STATE UNIVERSITY Standard Operating Procedure for the Bruker X8 APEX II Single-Crystal X- Ray Diffractometer Contact Manager: Dr. Cassie Ward ward@wayne.edu Office room 061 Chemistry (313) 577-2587 LIC Lab: (313) 577-0518

More information

JEM 3010 Manual (2011 Edition)

JEM 3010 Manual (2011 Edition) JEM 3010 Manual (2011 Edition) Basic Alignment Instructions Check the vacuum (power supply closet) and write on the log book: Gun pressure SIP2 < 2x10-5 Pa (usually 0.8-1.0x10-5 Pa) Column pressure SIP1

More information

Kratos AXIS Ultra DLD X-ray Photoelectron Spectrometer Instructions

Kratos AXIS Ultra DLD X-ray Photoelectron Spectrometer Instructions Kratos AXIS Ultra DLD X-ray Photoelectron Spectrometer Instructions Note: Enter your complete name, CMRF user code, and date into the record book. You must have previously filled out a project information

More information

FEI Strata Dual-beam FIB

FEI Strata Dual-beam FIB FEI Strata Dual-beam FIB Quick start Guide Compiled by Mat t hew Hughes and Tony Chen Page 0 Purpose of this guide INTRODUCTION This quick start guide is meant to provide cursory operational knowledge

More information

MestReNova Manual for Chem 201/202. October, 2015.

MestReNova Manual for Chem 201/202. October, 2015. 1. Introduction to 1-D NMR Data Processing with MestReNova The MestReNova program can do all of the routine NMR data processing needed for Chem 201 and 202 and will be available through the Reed downloads

More information

The Quanta 200 3D Simplified Operation Manual

The Quanta 200 3D Simplified Operation Manual The Quanta 200 3D Simplified Operation Manual CONTENTS 1. SYSTEM OVERVIEW... 4 2. VACUUM SYSTEM... 5 3. SOFTWARE CONTROL... 6 Pages and Modules... 7 4. BASIC OPERATION OF SEM AND FIB... 12 Specimen Preparation

More information

SEM- EDS Instruction Manual

SEM- EDS Instruction Manual SEM- EDS Instruction Manual Double-click on the Spirit icon ( ) on the desktop to start the software program. I. X-ray Functions Access the basic X-ray acquisition, display and analysis functions through

More information

Import and quantification of a micro titer plate image

Import and quantification of a micro titer plate image BioNumerics Tutorial: Import and quantification of a micro titer plate image 1 Aims BioNumerics can import character type data from TIFF images. This happens by quantification of the color intensity and/or

More information

The PHI VersaProbe operates with two essential software programs: PHI Summitt and Vacuum Watcher. A third program, MultiPak, handles data reduction.

The PHI VersaProbe operates with two essential software programs: PHI Summitt and Vacuum Watcher. A third program, MultiPak, handles data reduction. PHI VersaProbe Scanning XPS System I. Overview The PHI VersaProbe operates with two essential software programs: PHI Summitt and Vacuum Watcher. A third program, MultiPak, handles data reduction. PHI Summitt

More information

JEM 2100 Manual. Operation and Basic Alignment Instructions. Check the vacuum levels (power supply closet) Column (blue scale): < 2.

JEM 2100 Manual. Operation and Basic Alignment Instructions. Check the vacuum levels (power supply closet) Column (blue scale): < 2. 1. Verify the vacuum: JEM 2100 Manual Operation and Basic Alignment Instructions Check the vacuum levels (power supply closet) Column (blue scale): < 2.5x10-5 Pa If vacuum is not good enough, contact somebody

More information

Durham Magneto Optics Ltd. NanoMOKE 3 Wafer Mapper. Specifications

Durham Magneto Optics Ltd. NanoMOKE 3 Wafer Mapper. Specifications Durham Magneto Optics Ltd NanoMOKE 3 Wafer Mapper Specifications Overview The NanoMOKE 3 Wafer Mapper is an ultrahigh sensitivity Kerr effect magnetometer specially configured for measuring magnetic hysteresis

More information

Standard Operating Procedure of nanoir2-s

Standard Operating Procedure of nanoir2-s Standard Operating Procedure of nanoir2-s The Anasys nanoir2 system is the AFM-based nanoscale infrared (IR) spectrometer, which has a patented technique based on photothermal induced resonance (PTIR),

More information

PulseCounter Neutron & Gamma Spectrometry Software Manual

PulseCounter Neutron & Gamma Spectrometry Software Manual PulseCounter Neutron & Gamma Spectrometry Software Manual MAXIMUS ENERGY CORPORATION Written by Dr. Max I. Fomitchev-Zamilov Web: maximus.energy TABLE OF CONTENTS 0. GENERAL INFORMATION 1. DEFAULT SCREEN

More information

Linkage 3.6. User s Guide

Linkage 3.6. User s Guide Linkage 3.6 User s Guide David Rector Friday, December 01, 2017 Table of Contents Table of Contents... 2 Release Notes (Recently New and Changed Stuff)... 3 Installation... 3 Running the Linkage Program...

More information

Figure 1. MFP-3D software tray

Figure 1. MFP-3D software tray Asylum MFP-3D AFM SOP January 2017 Purpose of this Instrument: To obtain 3D surface topography at sub-nanometer scale resolution, measure contact and friction forces between surfaces in contact, measure

More information

Cover Page for Lab Report Group Portion. Boundary Layer Measurements

Cover Page for Lab Report Group Portion. Boundary Layer Measurements Cover Page for Lab Report Group Portion Boundary Layer Measurements Prepared by Professor J. M. Cimbala, Penn State University Latest revision: 23 February 2017 Name 1: Name 2: Name 3: [Name 4: ] Date:

More information

System Requirements SA0314 Spectrum analyzer:

System Requirements SA0314 Spectrum analyzer: System Requirements SA0314 Spectrum analyzer: System requirements Windows XP, 7, Vista or 8: 1 GHz or faster 32-bit or 64-bit processor 1 GB RAM 10 MB hard disk space \ 1. Getting Started Insert DVD into

More information

SC24 Magnetic Field Cancelling System

SC24 Magnetic Field Cancelling System SPICER CONSULTING SYSTEM SC24 SC24 Magnetic Field Cancelling System Makes the ambient magnetic field OK for the electron microscope Adapts to field changes within 100 µs Touch screen intelligent user interface

More information

ORM0022 EHPC210 Universal Controller Operation Manual Revision 1. EHPC210 Universal Controller. Operation Manual

ORM0022 EHPC210 Universal Controller Operation Manual Revision 1. EHPC210 Universal Controller. Operation Manual ORM0022 EHPC210 Universal Controller Operation Manual Revision 1 EHPC210 Universal Controller Operation Manual Associated Documentation... 4 Electrical Interface... 4 Power Supply... 4 Solenoid Outputs...

More information

Topic: Instructional David G. Thomas December 23, 2015

Topic: Instructional David G. Thomas December 23, 2015 Procedure to Setup a 3ɸ Linear Motor This is a guide to configure a 3ɸ linear motor using either analog or digital encoder feedback with an Elmo Gold Line drive. Topic: Instructional David G. Thomas December

More information

SNG-2150C User s Guide

SNG-2150C User s Guide SNG-2150C User s Guide Avcom of Virginia SNG-2150C User s Guide 7730 Whitepine Road Revision 001 Richmond, VA 23237 USA GENERAL SAFETY If one or more components of your earth station are connected to 120

More information

CHA EVAPORATOR. User guidelines. p.1. by Carlos Manzanedo. Last revised: 10/05/2000

CHA EVAPORATOR. User guidelines. p.1. by Carlos Manzanedo. Last revised: 10/05/2000 CHA EVAPORATOR User guidelines by Carlos Manzanedo. Last revised: 10/05/2000 p.1 Introduction. The CHA evaporator has the following characteristics: 1. 2 independent resistive power supplies. 2. 1 Electron

More information

TL-2900 AMMONIA & NITRATE ANALYZER DUAL CHANNEL

TL-2900 AMMONIA & NITRATE ANALYZER DUAL CHANNEL TL-2900 AMMONIA & NITRATE ANALYZER DUAL CHANNEL DATA ACQUISITION SYSTEM V.15.4 INSTRUCTION MANUAL Timberline Instruments, LLC 1880 S. Flatiron Ct., Unit I Boulder, Colorado 80301 Ph: (303) 440-8779 Fx:

More information

SC24 Magnetic Field Cancelling System

SC24 Magnetic Field Cancelling System SPICER CONSULTING SYSTEM SC24 SC24 Magnetic Field Cancelling System Makes the ambient magnetic field OK for the electron microscope Adapts to field changes within 100 µs Touch screen intelligent user interface

More information

11: PC MACs RealTime Pulldown

11: PC MACs RealTime Pulldown 11: PC MACs RealTime Pulldown This Pulldown Menu is where all the commands having to do with the real time programming of a show can be found. Most of the commands are duplicates of the buttons and checkboxes

More information

Data Collection Using APEX3. March 30, Chemical Crystallography Laboratory

Data Collection Using APEX3. March 30, Chemical Crystallography Laboratory Data Collection Using APEX3 Page 1 of 10 Data Collection Using APEX3 March 30, 2017 Chemical Crystallography Laboratory Author: Douglas R. Powell Data Collection Using APEX3 Page 2 of 10 Distribution Douglas

More information

Log viewer and analyser for switchboard and downhole sensor

Log viewer and analyser for switchboard and downhole sensor Log viewer and analyser for switchboard and downhole sensor Novomet-Perm cc, 2010 2013 User guide User manual 2 http://www.novomet.ru/eng Novomet-Perm cc 395 Cosmonauts st, Perm, 614065, Russia Novomet

More information

Preparing for remote data collection at NE-CAT

Preparing for remote data collection at NE-CAT Preparing for remote data collection at NE-CAT Important Note: The beamtime and remote login privileges are intended just for you and your group. You are not allowed to share these with any other person

More information

Sample BD Tech Concepts LLC

Sample BD Tech Concepts LLC XYZ Corp. Fry Controller FC-1234 Operating Specification Copyright 2014 Brian Dunn BD Tech Concepts LLC Contents Last Modified: 00/00/0000 Introduction 2 Interface 3 Idle 5 Cooking Cycle 5 Displaying and

More information

KRAMER ELECTRONICS LTD. USER MANUAL

KRAMER ELECTRONICS LTD. USER MANUAL KRAMER ELECTRONICS LTD. USER MANUAL MODEL: Projection Curved Screen Blend Guide How to blend projection images on a curved screen using the Warp Generator version K-1.4 Introduction The guide describes

More information

FIRMWARE RELEASE NOTE

FIRMWARE RELEASE NOTE Products VRC300 Date Nov 18, 2015 Version 2.02.08 The version(s) indicated here above has (have) been tested and validated by the Technical Support Department. Version 2.02.08 Technical notes New Features

More information

DektakXT Profilometer. Standard Operating Procedure

DektakXT Profilometer. Standard Operating Procedure DektakXT Profilometer Standard Operating Procedure 1. System startup and sample loading: a. Ensure system is powered on by looking at the controller to the left of the computer.(it is an online software,

More information

DIGITAL MIXING CONSOLE. TF Editor V3.6 User's Guide

DIGITAL MIXING CONSOLE. TF Editor V3.6 User's Guide DIGITAL MIXING CONSOLE TF Editor V3.6 User's Guide EN Special notices Copyrights of the software and this document are the exclusive property of Yamaha Corporation. Copying or modifying the software or

More information

Indiana Center for Biological Microscopy. Zeiss LSM-510. Confocal Microscope

Indiana Center for Biological Microscopy. Zeiss LSM-510. Confocal Microscope Indiana Center for Biological Microscopy Zeiss LSM-510 510-UV Confocal Microscope Microscope and the Attached Accessories Transmission Detector Halogen Lamp House Condenser Eyepiece Stage Scanning and

More information

FEI FIB Focused Ion Beam

FEI FIB Focused Ion Beam Operating Manual Part 1 FEI FIB Focused Ion Beam IF IN DOUBT, ASK 1.00 IF ANYTHING UNUSUAL HAPPENS, OR IF THERE IS ANYTHING YOU ARE UNSURE ABOUT, STOP AND CONTACT ME! DO NOT PROCEED OR ATTEMPT TO FIX THE

More information

READ THIS FIRST. Morphologi G3. Quick Start Guide. MAN0412 Issue1.1

READ THIS FIRST. Morphologi G3. Quick Start Guide. MAN0412 Issue1.1 READ THIS FIRST Morphologi G3 Quick Start Guide MAN0412 Issue1.1 Malvern Instruments Ltd. 2008 Malvern Instruments makes every effort to ensure that this document is correct. However, due to Malvern Instruments

More information

TF5 / TF3 / TF1 DIGITAL MIXING CONSOLE. TF Editor User Guide

TF5 / TF3 / TF1 DIGITAL MIXING CONSOLE. TF Editor User Guide TF5 / TF3 / TF1 DIGITAL MIXING CONSOLE EN Special notices Copyrights of the software and this document are the exclusive property of Yamaha Corporation. Copying or modifying the software or reproduction

More information

JAMAR TRAX RD Detector Package Power Requirements Installation Setting Up The Unit

JAMAR TRAX RD Detector Package Power Requirements Installation Setting Up The Unit JAMAR TRAX RD The TRAX RD is an automatic traffic recorder designed and built by JAMAR Technologies, Inc. Since the unit is a Raw Data unit, it records a time stamp of every sensor hit that occurs during

More information

E X P E R I M E N T 1

E X P E R I M E N T 1 E X P E R I M E N T 1 Getting to Know Data Studio Produced by the Physics Staff at Collin College Copyright Collin College Physics Department. All Rights Reserved. University Physics, Exp 1: Getting to

More information

Fully ly Automaticti. Motorised Satellite t TV System. User s manual REV

Fully ly Automaticti. Motorised Satellite t TV System. User s manual REV REV. 1.0 Fully ly Automaticti Motorised Satellite t TV System User s manual Customer Help Line: 1300 139 255 Support Email: support@satkingpromax.com.au Website: www.satkingpromax.com.au www.satkingpromax.com.au

More information

PYROPTIX TM IMAGE PROCESSING SOFTWARE

PYROPTIX TM IMAGE PROCESSING SOFTWARE Innovative Technologies for Maximum Efficiency PYROPTIX TM IMAGE PROCESSING SOFTWARE V1.0 SOFTWARE GUIDE 2017 Enertechnix Inc. PyrOptix Image Processing Software v1.0 Section Index 1. Software Overview...

More information

using the Scott A Speakman, Ph.D Center for Materials Science and Engineering at MIT

using the Scott A Speakman, Ph.D Center for Materials Science and Engineering at MIT X-Ray Reflectivity using the PANalytical X Pert Pro MPD Scott A Speakman, Ph.D Center for Materials Science and Engineering at MIT http://prism.mit.edu/xray Modified for configuration used at University

More information

A6 OPERATING INSTRUCTIONS

A6 OPERATING INSTRUCTIONS Amerec s A6 control for the AX steamer is a touch screen control intended to be mounted on a wall, generally in or near the steam bath. It may be mounted directly on the wall surface or, using an optional

More information

Contacts: English Department Office 238 Moreland Media Services 109 Kidder

Contacts: English Department Office 238 Moreland Media Services 109 Kidder Contacts: English Department Office 238 Moreland 7-3244 Media Services 109 Kidder September 2006 Welcome...3 Starting Out...5 Unlocking the Station...5 Touch Panel Operation...5 Projector...6 Selecting

More information

USER MANUEL. SNIPE 2 Ref R13

USER MANUEL. SNIPE 2 Ref R13 USER MANUEL SNIPE 2 Ref. 0141317R13 Contents 1. General Information 1-1. Introduction 1-2. Proper use and operation 1-3. Safety notes......... 2 3 3 2. Contents 2-1. Accessory included 2-2. Name of parts......

More information

127566, Россия, Москва, Алтуфьевское шоссе, дом 48, корпус 1 Телефон: +7 (499) (800) (бесплатно на территории России)

127566, Россия, Москва, Алтуфьевское шоссе, дом 48, корпус 1 Телефон: +7 (499) (800) (бесплатно на территории России) 127566, Россия, Москва, Алтуфьевское шоссе, дом 48, корпус 1 Телефон: +7 (499) 322-99-34 +7 (800) 200-74-93 (бесплатно на территории России) E-mail: info@awt.ru, web:www.awt.ru Contents 1 Introduction...2

More information

TASKI Service Tool Edition: V5.10/2014

TASKI Service Tool Edition: V5.10/2014 Edition: V5.10/2014 Index 1 General 1.1 General information 1 1.1.1 Part reference 1 1.1.2 Consumable supplies 1 1.1.3 Direction description 1 1.1.4 Power source 1 1.2 Required material 2 1.2.1 Tools 2

More information

CytoFLEX Flow Cytometer Quick Start Guide

CytoFLEX Flow Cytometer Quick Start Guide Sheath Waste CLASS 1 LASER PRODUCT COMPLIES WITH 21 CFR 1040.10 AND 1040.11 EXCEPT FOR DEVIATIONS PURSUANT TO LASER NOTICE NO. 50 DATED JUNE 24, 2007 MANUFACTURED Sheath B49008AC February 2015 CytoFLEX

More information

Cisco Spectrum Expert Software Overview

Cisco Spectrum Expert Software Overview CHAPTER 5 If your computer has an 802.11 interface, it should be enabled in order to detect Wi-Fi devices. If you are connected to an AP or ad-hoc network through the 802.11 interface, you will occasionally

More information

HyperMedia Software User Manual

HyperMedia Software User Manual HyperMedia Software User Manual Contents V1.2 Chapter 1 : HyperMedia software functions... 2 Chapter 2 : STVR... 3 2.1 System setting and channel setting... 3 2.2 Main panel... 6 2.2.1 Channel list...

More information

HV/PHA Adjustment (PB) Part

HV/PHA Adjustment (PB) Part HV/PHA Adjustment (PB) Part Contents Contents 1. How to set Part conditions...1 1.1 Setting conditions... 1 2. HV/PHA adjustment sequence...7 3. How to use this Part...9 HV/PHA Adjustment (PB) Part i

More information

Manual for TV software. TT-Viewer version Figure: TT-budget S2-3200

Manual for TV software. TT-Viewer version Figure: TT-budget S2-3200 Manual for TV software TT-Viewer version 2.7.0 Figure: TT-budget S2-3200 Index Manual TT-Viewer 3 1. Starting TT-Viewer software 3 2. General settings 5 3. Assignment of hardware 6 3.1 Unicable 7 4. Renderer

More information

U S E R G U I D E HD1000

U S E R G U I D E HD1000 U S E R G U I D E HD1000 1 W e l c o m e t o R o k u! In This Guide... Bring your HDTV to life with Roku. For the first time, you ll enjoy viewing your favorite digital photos in high-definition on your

More information

Processing data with Mestrelab Mnova

Processing data with Mestrelab Mnova Processing data with Mestrelab Mnova This exercise has three parts: a 1D 1 H spectrum to baseline correct, integrate, peak-pick, and plot; a 2D spectrum to plot with a 1 H spectrum as a projection; and

More information

XHD inch Widescreen LCD Monitor USERGUIDE

XHD inch Widescreen LCD Monitor USERGUIDE XHD3000 30-inch Widescreen LCD Monitor USERGUIDE Contents Chapter 1: Checking Out Your Monitor...... 1 Package contents....................................... 2 Monitor features.......................................

More information

Sample BD Tech Concepts LLC

Sample BD Tech Concepts LLC XYZ Corp. Fry Controller FC-1234 Software Test Procedure Copyright 2014 Brian Dunn BD Tech Concepts LLC Last Modified: 00/00/0000 Version Tested: Date Tested: Technician: Results: 1 FC-1234 SW Test Proc.

More information

Liquid Chromatography- Mass Spectrometer Manual

Liquid Chromatography- Mass Spectrometer Manual Liquid Chromatography- Mass Spectrometer Manual Joshua Willis, Elizabeth Sattely Department of Chemical Engineering Stanford University November 6, 2014 Abstract This manual will explain the LC/MS, its

More information

EDL8 Race Dash Manual Engine Management Systems

EDL8 Race Dash Manual Engine Management Systems Engine Management Systems EDL8 Race Dash Manual Engine Management Systems Page 1 EDL8 Race Dash Page 2 EMS Computers Pty Ltd Unit 9 / 171 Power St Glendenning NSW, 2761 Australia Phone.: +612 9675 1414

More information

Waters GCT GCMS Training Manual (Nominal Mass GCMS version) 02/06/2012 S.V.

Waters GCT GCMS Training Manual (Nominal Mass GCMS version) 02/06/2012 S.V. 1234 Hach Hall 515-294-5805 www.cif.iastate.edu Waters GCT GCMS Training Manual (Nominal Mass GCMS version) 02/06/2012 S.V. Location: Contact: 1238 Hach Hall Steve Veysey, 1234 Hach Hall; Kamel Harrata,

More information

BitWise (V2.1 and later) includes features for determining AP240 settings and measuring the Single Ion Area.

BitWise (V2.1 and later) includes features for determining AP240 settings and measuring the Single Ion Area. BitWise. Instructions for New Features in ToF-AMS DAQ V2.1 Prepared by Joel Kimmel University of Colorado at Boulder & Aerodyne Research Inc. Last Revised 15-Jun-07 BitWise (V2.1 and later) includes features

More information

F7000NV ROBOT VISION OPERATING MANUAL

F7000NV ROBOT VISION OPERATING MANUAL Rev. C Feb 2012 F7000NV ROBOT VISION OPERATING MANUAL Rev. C Feb 2012 This page has intentionally been left blank. Contents Contents Chapter 1. Getting Started... 5 1. Preface... 5 2. Manuals... 5 3. Setting

More information

Statement SmartLCT User s Manual Welcome to use the product from Xi an NovaStar Tech Co., Ltd. (hereinafter referred to as NovaStar ). It is our great

Statement SmartLCT User s Manual Welcome to use the product from Xi an NovaStar Tech Co., Ltd. (hereinafter referred to as NovaStar ). It is our great LED Display Configuration Software SmartLCT User s Manual Software Version: V3.0 Rev3.0.0 NS110100239 Statement SmartLCT User s Manual Welcome to use the product from Xi an NovaStar Tech Co., Ltd. (hereinafter

More information

ViewCommander- NVR Version 3. User s Guide

ViewCommander- NVR Version 3. User s Guide ViewCommander- NVR Version 3 User s Guide The information in this manual is subject to change without notice. Internet Video & Imaging, Inc. assumes no responsibility or liability for any errors, inaccuracies,

More information

Outback STX. User Guide Supplement. Parts List. STX Terminal Overview

Outback STX. User Guide Supplement. Parts List. STX Terminal Overview Outback STX User Guide Supplement This supplement details the following changes from STX v1.0 to STX v1.1: Parts List below STX Terminal Overview below Connection Diagram on page 2 AC110 Power Up and Power

More information

USER GUIDE. Table of Contents

USER GUIDE. Table of Contents Table of Contents USER GUIDE USER GUIDE...1 1. Installation of Personal Music Collection Database...2 2. Working with Personal Music Collection Database...4 2.1. General Information...4 2.2. Navigation

More information

CHA EVAPORATOR Short Reference User guidelines

CHA EVAPORATOR Short Reference User guidelines CHA EVAPORATOR Short Reference User guidelines by Carlos Manzanedo. Last revised: 10/05/2000 p.1 Procedure: The CHA evaporator is divided into 16 different panels. These panels are labeled on the machine

More information

Part names (continued) Remote control

Part names (continued) Remote control Introduction Part names (continued) Remote control (1) STANDBY ( 25) (1) (2) ON ( 25) (3) (3) ID - 1 / 2 / 3 / 4 s ( 18) (4) (4) COMPUTER 1 ( 27) (7) (5) COMPUTER 2 * (8) (6) COMPUTER 3 * (10) (13) (7)

More information

Casambi App FAQ. Version Casambi Technologies Oy.

Casambi App FAQ. Version Casambi Technologies Oy. Casambi App FAQ Version 1.3 30.9.2016 Casambi Technologies Oy 1 of 12 GENERAL 3 Q: What is Casambi app used for? 3 Q: Which mobile devices are supported? 3 Q: Where can I get the Casambi app? 3 FIRST TIME

More information

imso-104 Manual Revised August 5, 2011

imso-104 Manual Revised August 5, 2011 imso-104 Manual Revised August 5, 2011 Section 1 Getting Started SAFETY 1.10 Quickstart Guide 1.20 SAFETY 1.30 Compatibility 1.31 Hardware 1.32 Software Section 2 How it works 2.10 Menus 2.20 Analog Channel

More information

Veeco Dektak 6M Profilometer

Veeco Dektak 6M Profilometer Veeco Dektak 6M Profilometer System Ranges/Resolutions Range (Å) Resolution (Å) 50 (5nm) to 65K 1 0.5K to 655K 10 2K to 2620K 40 8K to 10000K (1mm) 160 Maximum sample thickness: 31.75mm Scan range: 50

More information

R.E.A.D.S. INSTRUCTIONS FOR KINDLE ereaders

R.E.A.D.S. INSTRUCTIONS FOR KINDLE ereaders If you have a Kindle Fire, you will need to download the Overdrive Media Console App to your device. Overdrive App instructions are available at the Williamson County Public Library Reference Desk or on

More information

LedSet User s Manual V Official website: 1 /

LedSet User s Manual V Official website:   1 / LedSet User s Manual V2.6.1 1 / 42 20171123 Contents 1. Interface... 3 1.1. Option Menu... 4 1.1.1. Screen Configuration... 4 1.1.1.1. Instruction to Sender/ Receiver/ Display Connection... 4 1.1.1.2.

More information

CCE900-IP-TR. User s Guide

CCE900-IP-TR. User s Guide CCE900-IP-TR CCE900-IP-T & CCE900-IP-R User s Guide i-tech Company LLC TOLL FREE: (888) 483-2418 EMAIL: info@itechlcd.com WEB: www.itechlcd.com 1. Introduction The CCE900-IP-T & CCE900-IP-R is a solution

More information

Classroom Setup... 2 PC... 2 Document Camera... 3 DVD... 4 Auxiliary... 5

Classroom Setup... 2 PC... 2 Document Camera... 3 DVD... 4 Auxiliary... 5 Classroom Setup... 2 PC... 2 Document Camera... 3 DVD... 4 Auxiliary... 5 Lecture Capture Setup... 6 Pause and Resume... 6 Considerations... 6 Video Conferencing Setup... 7 Camera Control... 8 Preview

More information

VeriLUM 5.2. Video Display Calibration And Conformance Tracking. IMAGE Smiths, Inc. P.O. Box 30928, Bethesda, MD USA

VeriLUM 5.2. Video Display Calibration And Conformance Tracking. IMAGE Smiths, Inc. P.O. Box 30928, Bethesda, MD USA VeriLUM 5.2 Video Display Calibration And Conformance Tracking IMAGE Smiths, Inc. P.O. Box 30928, Bethesda, MD 20824 USA Voice: 240-395-1600 Fax: 240-395-1601 Web: www.image-smiths.com Technical Support

More information

Quick reference guide

Quick reference guide Quick reference guide Manufactured by: Esaote Europe B.V. Philipsweg 1 6227 AJ Maastricht The Netherlands Tel. +31 (43) 382 4600 Fax +31 (43) 382 4601 Internet: www.esaote.com Email: international.sales@esaote.com

More information

The only warranties for HP products and services are set forth in the express warranty statements accompanying such products and services.

The only warranties for HP products and services are set forth in the express warranty statements accompanying such products and services. The only warranties for HP products and services are set forth in the express warranty statements accompanying such products and services. Nothing herein should be construed as constituting an additional

More information

MAutoPitch. Presets button. Left arrow button. Right arrow button. Randomize button. Save button. Panic button. Settings button

MAutoPitch. Presets button. Left arrow button. Right arrow button. Randomize button. Save button. Panic button. Settings button MAutoPitch Presets button Presets button shows a window with all available presets. A preset can be loaded from the preset window by double-clicking on it, using the arrow buttons or by using a combination

More information

University of Minnesota College of Science and Engineering Characterization Facility Ganesha SAXSLAB User manual

University of Minnesota College of Science and Engineering Characterization Facility Ganesha SAXSLAB User manual University of Minnesota College of Science and Engineering Characterization Facility Ganesha SAXSLAB User manual # Section Pg # 1 Planning a SAXS experiment 1 2 Initial Setup 2 3 Sample mounting and loading

More information

CNC Router Parts Plasma Software Setup and Usage Guide

CNC Router Parts Plasma Software Setup and Usage Guide Plasma Software Setup and Software Setup for the TMC3in1 Mach3 If your control PC is not currently set up with Mach3 software, follow the CNC Router Parts CNC Software Setup Guide to install the correct

More information

PSC300 Operation Manual

PSC300 Operation Manual PSC300 Operation Manual Version 9.10 General information Prior to any attempt to operate this Columbia PSC 300, operator should read and understand the complete operation of the cubing system. It is very

More information

The BAT WAVE ANALYZER project

The BAT WAVE ANALYZER project The BAT WAVE ANALYZER project Conditions of Use The Bat Wave Analyzer program is free for personal use and can be redistributed provided it is not changed in any way, and no fee is requested. The Bat Wave

More information

VPM2. Operation Manual

VPM2. Operation Manual VPM2 Operation Manual Whip Corporation 361 Farmington Ave. P.O. Box 17183 Louisville, KY 40217-0183 USA 502-637-1451 800-626-5651 Fax 502-634-4512 www.whipmix.com LISTED Features The Whip VPM2 is designed

More information