1 Operating Manual Part 1 FEI FIB Focused Ion Beam IF IN DOUBT, ASK
2 1.00 IF ANYTHING UNUSUAL HAPPENS, OR IF THERE IS ANYTHING YOU ARE UNSURE ABOUT, STOP AND CONTACT ME! DO NOT PROCEED OR ATTEMPT TO FIX THE PROBLEM YOURSELF! Chemical Safety Users are required to have passed Cornell s on-line laboratory safety training courses before they can use the FIB. Do not bring chemicals into the laboratory. Do not bring hazardous samples or materials into the lab. Contact Mick Thomas or John Grazul with any questions. This is not a storage facility - do not leave or store samples here. You must take specimens, stubs, and raw material with you when you leave. Lab-supplied chemicals are limited to a squirt bottle of isopropanol, methanol, and a can of Aero-duster. Wear safety glasses and read the MSDS s before using them. A First-Aid kit is provided but this is not a substitute for professional medical help. Fire escape routes are documented on the following page. Users failing to follow safety rules will be denied access to the lab. IF IN DOUBT, ASK IF IN DOUBT, ASK
3 1.00 In Case of Fire: Close the door behind you Leave the building First Aid Kit Fire Pull Station Fire Extinguisher Emergency Shower and Eyewash IF IN IF IN DOUBT, ASK
4 1.00 Gas Safety Nitrogen gas cylinders are to be changed by CCMR staff members only Lab Telephone If you don t have a cell phone, there is a land-line phone for your use in the corner of the NION room Campus Police: Emergency: 911 John: Illness If you feel sick, please stay at home. This will require rescheduling your session, but other people will stay healthy. IF IN IF IN DOUBT, ASK
5 2.00 Rules If ANYTHING does not even SEEM right, leave everything as is and get Mick or John DO NOT attempt to repair ANY part of the microscope. DO NOT remove panels, cables, or any part of the FIB. DO NOT add any software or hardware to any computer Flash Drives (Memory Sticks) are forbidden Data MUST be removed by burning a CD DO NOT bring food or drink into 150. Wear gloves when handling samples or shuttles. Individuals failing to follow the rules will be denied access to the instrument. IF IN IF IN DOUBT, ASK
6 2.00 Penalties Failure to follow these guidelines, unprofessional behavior or careless use of the equipment will result in loss of permission to use the FIB, in addition to repair charges, if required. Training requirements No one is permitted to use the FIB until trained by Mick Thomas. Room Access No user is permitted to unlock the door into 150 for the purpose of allowing someone else in to use the FIB. User Status Signup Room Entry Access Novice (<2 sessions) Manager M-F 8am 5pm Trained (>2 sessions) Coral Card Swipe M-F 8am 5pm Experienced (>4 sessions) Coral Card Swipe 24/7 IF IN IF IN DOUBT, ASK
7 2.00 Cancellation If you sign up and then cannot use your time, cancel your time via Coral or contact Mick at least one weekday before your scheduled time. Failing to show up Failure to show up without canceling, unless due to illness or other emergency, can result in use charges If you fail to show up within 30 minutes of your scheduled starting time the FIB becomes available to other users. Publication Acknowledgements When research is published using data from the FIB, the facility and the grant number must be acknowledged, such as: This work made use of the electron microscopy facility of the Cornell Center for Materials Research (CCMR) with support from the National Science Foundation Materials Research Science and Engineering Centers (MRSEC) program (DMR ). Data Storage Data storage and safety is NOT guaranteed! Users are responsible for their data, and must copy their data to a CD in a timely fashion. Data will be periodically erased, leaving only one years data there. IF IN IF IN DOUBT, ASK
8 2.50 About this manual (Version 1.0) This is a limited use manual intended to help new users get started on the FEI Focused Ion Beam (FIB). It provides information on basic imaging, milling, and TEM sample preparation. It is not intended to be exhaustive, and there are many principles about focused ion beam practices as well as features of this microscope that are not covered by this manual. This manual will be updated and expanded as needed.
9 3.00 How to start your FIB session Log into Coral Log on to Windows on the microscope PC Click on the FEI System Control icon. Confirm the four circles in the server window turn solid green If all the circles do not turn solid green then do not proceed contact Mick Click on Start Log on to the xtm Server Click on Determine
10 3.00 Understanding the GUI setup: Main Toolbar Electron Image Ion image Control toolbars TV image Status window Activate an image quadrant by clicking in it. The bottom bar will turn blue, indicating that window is now active.
11 3.50 How to Load a Sample Click on Navigation Button Click on Unload in the software When both lights are green lift the lid and push Unload/Release The arm should move away from the shuttle, allowing you to remove it.
12 3.50 How to Load a Sample Note that the two clamping arms must be retracted BEFORE removing the shuttle Samples must not exceed the allowable height - check height using the height gauge. They must also fit inside the volume of space as shown at left.
13 3.50 How to Load a Sample The shuttle is loaded onto the arm with the three locating balls (rubies) are in the appropriate slots on the arm (inside white circles in picture below). Gently close the lid and push Clamp/Load. When loaded the rotation should be degrees; if not enter that value as shown below. 62.4
14 3.70 How to Find Your Sample in the FIB Standard stage (at right) Enter the coordinates in the boxes for X and Y as shown at right +10, , , +10 0, , +10 0, 0 For TEM stage only: Front position: X = 0 Y = 17 Back position: X = 0 Y = -17
15 4.00 How to Turn On the Beam Select the upper left hand quadrant - this activates the electron beam option Select the voltage and current from the drop -down menus Select the Beam Control icon. Click on Beam On Un-pause the beam
16 4.20 How to Link Your Sample The FIB does not know where the sample is in reference to the polepiece (vertical position). The Z value displayed in the Navigation window will be wrong. Therefore it is critical to tell the FIB where your sample is a process called Linking. Do not adjust specimen height (Z) until you have Linked at low mag. Electron column? Sample Set the magnification to about 50X and focus the beam on the surface of your sample Increase the magnification to about 1000X and refocus Now STOP Go no further until you have clicked on the Link icon
17 4.20 How to Link Your Sample (con t) Coarse alignment Using the Z control, raise the stage to 9mm, then 7mm, then 5mm, checking after each motion that the image refocuses and the Stage motion looks normal. Ifyouseearedcirclearound the Link icon when you reach 5mm then refocus andclickonthelinkicon again.
18 4.20 How to Link Your Sample (con t) Fine alignment Find a distinct feature near the area of interest around 1000X magnification. Tilt the stage by clicking on the drop-down menu and clicking on 15 degrees. Electron column 15 degrees
19 4.20 How to Link Your Sample (con t) If the feature has moved (down, for example) then move the mouse over the TV image, click on the middle button (the wheel) and a yellow line will appear. Move the mouse in the direction necessary to move the feature to the middle of the yellow + sign on the electron image. As you do this a yellow arrow will appear showing you the direction you are going. Repeat at 45 degrees (skip 30 degrees), then finally at 52 degrees. Electron column 45 degrees Reset to 0 degrees tilt
20 5.00 How to Select Imaging Conditions Field-Free: Immersion: EDX: Use for milling and general imaging Use for higher resolution imaging. For use while collecting EDX spectra. How to Select Acquisition Type Live: Average: Integrate: Continuous imaging Uses N number of frames to form an image Takes N frames and then stops
21 6.00 How to Select Detector Type ETD: TTL: STEM: CDEM: CCD: Side mounted detector Through the Lens detector Transmission detector Channeltron Dynamic Electron Multiplier TV image
22 7.00 How to Check Beam Alignment Click on Lens Align to wobble the aperture and reduce wobble, if needed. Click on Crossover to check the beam centering and improve, if needed. Correct stigmatism with hardware controls. You are now ready for electron imaging.
23 8.00 How to Correct Astigmatism Adjust the X-stigmator and Y-stigmator so that when you go in and out of focus there is no stretching of the features Uncorrected Adjust stigmators Corrected
24 9.00 Principles of Ion Milling In the simplest of examples, start with a bulk material with a film on top Add platinum to protect the sample during milling Mill away a volume of space to reveal hidden, sub-surface features
25 9.20 How to set the ion beam conditions Ion milling can occur at any angle up to about 58 degrees. However it is tyically performed at 52 degrees, when the ion beam is at normal incidence to the sample. Click in the upper right hand quadrant the bottom bar should change from gray to blue. Set the beam current to an appropriate value for your sample, keeping it mind that it is milling your sample all the time you are focusing and stigmating. Note that there is a very large range of beam currents, with the highest over 10,000 times that of the lowest, so be careful to choose the correct value for your purpose. The higher the voltage: The sharper the image The faster the milling The more the milling surface will be amorphized The higher the current: The better the S/N The faster the milling The bigger the tails on the beam In the Application drop-down menu select Si
26 9.20 How to turn on the ion beam For general milling, make sure the stage is linked and you can tilt to 52 degrees without the feature moving more than 5 um s Electron column 52 degrees If the bar below the Source is gray, click on Wake Up. It will first turn red, then yellow, then green. When the source is green, click on Beam On. Remember that imaging with the ion beam is inherently destructive. Un-pause the beam
27 9.30 How to mill our sample Select the desired geometric pattern Draw the desired size or enter it in digitally Set the depth (Z value) Click on the Play icon To stop the milling click on the Stop icon
28 9.50 How to Deposit Platinum Choose the Patterning tab Draw a box where you want to deposit platinum. Under Gas Injection > Overview right click on Cold and a Heater option will appear.. Left click on Heater Click on white box to the left of Pt Dep to insert needle
Scanning Electron Microscopy (FEI Versa 3D Dual Beam)
Scanning Electron Microscopy (FEI Versa 3D Dual Beam) This operating procedure intends to provide guidance for basic measurements on a standard sample with FEI Versa 3D SEM. For more advanced techniques
More informationFIB Operating Procedure. Effective Date: 08/14/2012 Author(s): Jiong Hua Phone:
FIB Operating Procedure Effective Date: 08/14/2012 Author(s): Jiong Hua Phone: 402-472-3773 Email: jhua2@unl.edu 1 1 Introduction 1.1 Key Words Focused Ion Beam (FIB), FEI Strata 201, Ion milling 1.2 Purpose
More informationUniversity of Minnesota Nano Fabrication Center Standard Operating Procedure
Equipment Name: Focused Ion Beam (FIB) Coral Name: fib Revision Number: 2 Model: FEI Quanta 200 3D Revisionist: Kevin Roberts Location: Area 3 Date: 9/17/2013 1 Description The Quanta 200 3D is a DualBeam
More informationCHECKLIST FOR VERIOS OPERATION 1. GENERAL The SEM lab is used assuming "operating room" cleanliness, i.e., the SEM lab is a high visibility lab and
CHECKLIST FOR VERIOS OPERATION 1. GENERAL The SEM lab is used assuming "operating room" cleanliness, i.e., the SEM lab is a high visibility lab and must be kept clean and neat so clean up behind yourself
More informationFEI Strata Dual-beam FIB
FEI Strata Dual-beam FIB Quick start Guide Compiled by Mat t hew Hughes and Tony Chen Page 0 Purpose of this guide INTRODUCTION This quick start guide is meant to provide cursory operational knowledge
More information2.1. Log on to the TUMI system (you cannot proceed further until this is done).
FEI DB235 ex-situ lift out TEM sample preparation procedure Nicholas G Rudawski ngr@ufledu (805) 252-4916 Last updated: 06/19/15 DISCLAIMER: this procedure describes one specific method for preparing ex-situ
More informationStandard Operating Procedure for FEI Helios 660 NanoLab Part I: SEM Version
Standard Operating Procedure for FEI Helios 660 NanoLab Part I: SEM Version Helios reservations may be made online using the NERCF website. Note: Always wear gloves when venting the system and exchanging
More informationStandard Operating Procedure II: EDS (Bruker Flat-Quad)
Standard Operating Procedure II: EDS (Bruker Flat-Quad) ywcmatsci.yale.edu ESC II, Room A119F 810 West Campus Drive West Haven, CT 06516 Version 1.1, October 2018 1 > FOLLOW the SOP strictly to keep the
More information1. Check the accelerating voltage, must be at 200 kv (right screen), HT (µa) (left panel) at and Emission (left panel) at
JEOL 2010F MANUAL Quick check list 1. Check the accelerating voltage, must be at 200 kv (right screen), HT (µa) (left panel) at 0.96-0.97 and Emission (left panel) at 155-160. 2. Check the vacuum sequence
More informationPersonal Protective Equipment Wear nitrile gloves, lab coat, and safety glasses as a minimum protection, unless otherwise indicated.
4pt Bending, Mouse This protocol is for standard Jepsen 4pt bending of adult mouse bone. Safety considerations Please reference the Jepsen laboratory when using this protocol. This protocol is subject
More informationTechnical Procedure for Scanning Electron Microscope/ Energy Dispersive X-Ray System (SEM/EDX) for non-gsr Casework
Technical Procedure for Scanning Electron Microscope/ Energy Dispersive X-Ray System (SEM/EDX) for non-gsr Casework 1.0 Purpose This technical procedure shall be followed for the operation of the Scanning
More informationLUMIGEN INSTRUMENT CENTER X-RAY CRYSTALLOGRAPHIC LABORATORY: WAYNE STATE UNIVERSITY
Standard Operating Procedure for the Bruker X8 APEX II Single-Crystal X- Ray Diffractometer Contact Manager: Dr. Cassie Ward ward@wayne.edu Office room 061 Chemistry (313) 577-2587 LIC Lab: (313) 577-0518
More informationJEM 3010 Manual (2011 Edition)
JEM 3010 Manual (2011 Edition) Basic Alignment Instructions Check the vacuum (power supply closet) and write on the log book: Gun pressure SIP2 < 2x10-5 Pa (usually 0.8-1.0x10-5 Pa) Column pressure SIP1
More informationOperation Procedure for Phillips XL30 ESEM
Operation Procedure for Phillips XL30 ESEM The ESEM will be left in the ON state when not in use. The chamber will be at high vacuum, filament on, stage at home position, VAC and HT buttons lit, and monitor
More informationPHI 5000 VersaProbe TM Operator s Guide
PHI 5000 VersaProbe TM Operator s Guide Part No. 705921 Rev. A Copyright 2006 ULVAC-PHI, INC. 370 Enzo, Chigasaki, JAPAN The PHI logo ( ) is a registered trademark of ULVAC-PHI, INC. Physical Electronics,
More informationStandard Operating Procedure of nanoir2-s
Standard Operating Procedure of nanoir2-s The Anasys nanoir2 system is the AFM-based nanoscale infrared (IR) spectrometer, which has a patented technique based on photothermal induced resonance (PTIR),
More informationJEM 2100 Manual. Operation and Basic Alignment Instructions. Check the vacuum levels (power supply closet) Column (blue scale): < 2.
1. Verify the vacuum: JEM 2100 Manual Operation and Basic Alignment Instructions Check the vacuum levels (power supply closet) Column (blue scale): < 2.5x10-5 Pa If vacuum is not good enough, contact somebody
More informationTender Notification for the procurement of a "Dual beam (FIB - FE SEM) system" at IISc (Last Date for submission of tenders: 31st March 2016)
Tender Notification for the procurement of a "Dual beam (FIB - FE SEM) system" at IISc (Last Date for submission of tenders: 31st March 2016) Dear Sir/Madam, Kindly send your best quotation for the following
More informationKratos AXIS Ultra DLD X-ray Photoelectron Spectrometer Instructions
Kratos AXIS Ultra DLD X-ray Photoelectron Spectrometer Instructions Note: Enter your complete name, CMRF user code, and date into the record book. You must have previously filled out a project information
More informationDescription of task... 2 Sample preparation... 2 Sample insertion and SEM settings... 3 Eucentricity... 4
These are simple notes taken by Charlie Sanabria while learning how to use the Focused Ion Beam (FIB) in the Scanning Electron Microscope (SEM) at the Applied Superconductivity Center (ASC). The instructions
More information1.2 Universiti Teknologi Brunei (UTB) reserves the right to award the tender in part or in full.
TENDER SPECIFICATIONS FOR THE SUPPLY, DELIVERY, INSTALLATION AND COMMISSIONING OF ONE UNIT OF VARIABLE PRESSURE ENVIRONMENTAL SCANNING ELECTRON MICROSCOPE (SEM) CUM ENERGY DISPERSIVE SPECTROSCOPY (EDS)
More informationThe Quanta 200 3D Simplified Operation Manual
The Quanta 200 3D Simplified Operation Manual CONTENTS 1. SYSTEM OVERVIEW... 4 2. VACUUM SYSTEM... 5 3. SOFTWARE CONTROL... 6 Pages and Modules... 7 4. BASIC OPERATION OF SEM AND FIB... 12 Specimen Preparation
More informationAmerican DJ. Show Designer. Software Revision 2.08
American DJ Show Designer Software Revision 2.08 American DJ 4295 Charter Street Los Angeles, CA 90058 USA E-mail: support@ameriandj.com Web: www.americandj.com OVERVIEW Show Designer is a new lighting
More informationSTANDARD OPERATING PROCEDURE:
Q-EXACTIVE STANDARD OPERATING PROCEDURE: Purpose of this Instrument: Essential tool for molecular weight identification for small molecules, peptides, proteins, ect. Location: WVU Chemistry Research Laboratory
More informationNOTICE: This document is for use only at UNSW. No copies can be made of this document without the permission of the authors.
Brüel & Kjær Pulse Primer University of New South Wales School of Mechanical and Manufacturing Engineering September 2005 Prepared by Michael Skeen and Geoff Lucas NOTICE: This document is for use only
More informationSPM Training Manual Veeco Bioscope II NIFTI-NUANCE Center Northwestern University
SPM Training Manual Veeco Bioscope II NIFTI-NUANCE Center Northwestern University Introduction: Scanning Probe Microscopy (SPM) is a general term referring to surface characterization techniques that utilize
More informationE X P E R I M E N T 1
E X P E R I M E N T 1 Getting to Know Data Studio Produced by the Physics Staff at Collin College Copyright Collin College Physics Department. All Rights Reserved. University Physics, Exp 1: Getting to
More informationWord Tutorial 2: Editing and Formatting a Document
Word Tutorial 2: Editing and Formatting a Document Microsoft Office 2010 Objectives Create bulleted and numbered lists Move text within a document Find and replace text Check spelling and grammar Format
More informationPHY221 Lab 1 Discovering Motion: Introduction to Logger Pro and the Motion Detector; Motion with Constant Velocity
PHY221 Lab 1 Discovering Motion: Introduction to Logger Pro and the Motion Detector; Motion with Constant Velocity Print Your Name Print Your Partners' Names Instructions August 31, 2016 Before lab, read
More informationDektakXT Profilometer. Standard Operating Procedure
DektakXT Profilometer Standard Operating Procedure 1. System startup and sample loading: a. Ensure system is powered on by looking at the controller to the left of the computer.(it is an online software,
More information9.5 Add or Remove Samples in Single Access Mode
3. Press the Previous or Next button to see the details of the previous or next loaded position. 4. Press the Back button to return to the information screen. The previous menu is displayed. 9.5 Add or
More informationCharacterization of Diamond Samples CHESS Run Summer 2013
Characterization of Diamond Samples CHESS Run Summer 2013 5/22 5/28/2013 Characterization of Diamond Samples CHESS Run Summer 2013 Useful information Goals for this run Conditions of the experiment 1.
More informationINSTRUCTIONAL MANUAL FOR LCD ZOOM MICROSCOPE
INSTRUCTIONAL MANUAL FOR LCD ZOOM MICROSCOPE ? 8 LCD Screen? 10.4 LCD Screen LCD Zoom Microscope Instruction Manual Please read the Instruction Manual carefully before installation and keep it for future
More informationPreparing for remote data collection at NE-CAT
Preparing for remote data collection at NE-CAT Important Note: The beamtime and remote login privileges are intended just for you and your group. You are not allowed to share these with any other person
More informationData Collection Using APEX3. March 30, Chemical Crystallography Laboratory
Data Collection Using APEX3 Page 1 of 10 Data Collection Using APEX3 March 30, 2017 Chemical Crystallography Laboratory Author: Douglas R. Powell Data Collection Using APEX3 Page 2 of 10 Distribution Douglas
More informationSEM- EDS Instruction Manual
SEM- EDS Instruction Manual Double-click on the Spirit icon ( ) on the desktop to start the software program. I. X-ray Functions Access the basic X-ray acquisition, display and analysis functions through
More informationLibrary ebooks and Your Sony ereader
Library ebooks and Your Sony ereader When using your Sony ereader for the first time to obtain library ebooks and audiobooks, download the Overdrive Media Console, Adobe Digital Editions, and the Sony
More informationLab Determining the Screen Resolution of a Computer
Lab 1.3.3 Determining the Screen Resolution of a Computer Objectives Determine the current screen resolution of a PC monitor. Determine the maximum resolution for the highest color quality. Calculate the
More information91.7 The Edge, WSUW-FM Training Manual
91.7 The Edge, WSUW-FM Training Manual Station Policies Overview (Please see the Policy Manual for additional policies and procedures) As a broadcast facility, we come under a number of FCC rules and regulations.
More informationAFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode)
AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode) 1. Log into the Log Usage system on the SMIF web site 2. Open Nanoscope 6.14r1 software by double clicking on the Nanoscope 6.14r1 desktop
More informationQuick Start ATOMIC FORCE MICROSCOPE West Campus Imaging Core
Quick Start ATOMIC FORCE MICROSCOPE West Campus Imaging Core 1 Turn On the laser power 2 Open enclosure: - lift the door latch and open the enclosure door. 3 2 1 1.Unlock scanner: Lift the lever to the
More informationShould you have any questions that aren t answered here, simply call us at Live Connected.
Interactive TV User Guide This is your video operations manual. It provides simple, straightforward instructions for your TV service. From how to use your Remote Control to Video On Demand, this guide
More informationManual Addendum For Rerun V1.1 software 12/12/2006, RERUN-A = Serial #06A068, RERUN-P = Serial #06A031
Manual Addendum For Rerun V1.1 software 12/12/2006, RERUN-A = Serial #06A068, RERUN-P = Serial #06A031 The Rerun product manual was written for V1.0 software. The new release, V1.1, adds a number of new
More informationManual for TV software. TT-Viewer version Figure: TT-budget S2-3200
Manual for TV software TT-Viewer version 2.7.0 Figure: TT-budget S2-3200 Index Manual TT-Viewer 3 1. Starting TT-Viewer software 3 2. General settings 5 3. Assignment of hardware 6 3.1 Unicable 7 4. Renderer
More informationFACILITIES SUPPORT SERVICES
Facilities Directorate FACILITIES SUPPORT SERVICES Mechanical Engineering LT B & Dental LT INTERACTIVE LECTURE THEATRE USER OPERATION GUIDE CONTENTS INTRODUCTION... 2 Main Teaching System:... 2 Collaborative
More informationF7000NV ROBOT VISION OPERATING MANUAL
Rev. C Feb 2012 F7000NV ROBOT VISION OPERATING MANUAL Rev. C Feb 2012 This page has intentionally been left blank. Contents Contents Chapter 1. Getting Started... 5 1. Preface... 5 2. Manuals... 5 3. Setting
More informationFigure 1. MFP-3D software tray
Asylum MFP-3D AFM SOP January 2017 Purpose of this Instrument: To obtain 3D surface topography at sub-nanometer scale resolution, measure contact and friction forces between surfaces in contact, measure
More informationProcedure: Procedures: DIME TEAM: Drop # (circle one) Date: In Lab (either on Level 4, Level 6, or in shop): Installation: Functional Check
Procedure: Red Experiment specific Black - General In Lab (either on Level 4, Level 6, or in shop): Installation: 1. Wear safety goggles when working with the experiment. 2. Perform experiment-specific
More informationLaser Beam Analyser Laser Diagnos c System. If you can measure it, you can control it!
Laser Beam Analyser Laser Diagnos c System If you can measure it, you can control it! Introduc on to Laser Beam Analysis In industrial -, medical - and laboratory applications using CO 2 and YAG lasers,
More informationPerforming a Sound Level Measurement
APPENDIX 9 Performing a Sound Level Measurement Due to the many features of the System 824 and the variety of measurements it is capable of performing, there is a great deal of instructive material in
More informationDektak II SOP Revision 1 05/30/12 Page 1 of 5. NRF Dektak II SOP
Page 1 of 5 NRF Dektak II SOP The Dektak II-A is a sensitive stylus profilometer. A diamond-tipped stylus is moved laterally across the surface while in contact and measures deflections of the tip. It
More informationTL-2900 AMMONIA & NITRATE ANALYZER DUAL CHANNEL
TL-2900 AMMONIA & NITRATE ANALYZER DUAL CHANNEL DATA ACQUISITION SYSTEM V.15.4 INSTRUCTION MANUAL Timberline Instruments, LLC 1880 S. Flatiron Ct., Unit I Boulder, Colorado 80301 Ph: (303) 440-8779 Fx:
More informationThe amazing power of FiOS starts here.
SELF-INSTALLATION GUIDE The amazing power of FiOS starts here. LET S GET STARTED Welcome to a network that s light years ahead. Welcome to life on FiOS. Congratulations on choosing Verizon FiOS! You re
More informationFirst Question: Camera head. Lighting unit. Shooting stage
Elmo P30 Visualiser First Question: Q. Is everyone familiar with exactly what a visualiser is? A. A visualiser is effectively a camera on an arm, usually with a shooting stage and its own lighting source.
More informationThe PHI VersaProbe operates with two essential software programs: PHI Summitt and Vacuum Watcher. A third program, MultiPak, handles data reduction.
PHI VersaProbe Scanning XPS System I. Overview The PHI VersaProbe operates with two essential software programs: PHI Summitt and Vacuum Watcher. A third program, MultiPak, handles data reduction. PHI Summitt
More informationHD Guide. User Manual
HD Guide. User Manual You ve decided you want better TV. Here s how to enjoy it. Welcome to Shaw HD TV. To get the most out of your experience, it s best to know absolutely everything the service offers.
More informationGS122-2L. About the speakers:
Dan Leighton DL Consulting Andrea Bell GS122-2L A growing number of utilities are adapting Autodesk Utility Design (AUD) as their primary design tool for electrical utilities. You will learn the basics
More informationInserting the batteries. Basic settings of the remote control
Inserting the batteries Procedure prior to first use or when changing batteries Remove the back plate to expose the battery tray. Insert 2 x AA 1.5V alkaline batteries. Ensure the polarity of the batteries
More informationREAD ME FIRST. Touchstone TV Lift
Whisper Lift II PRO 2 READ ME FIRST 1. After completing the unpacking and uncrating of the cabinet, you will find the Owner s Manual, TV, installation hardware, and the wireless remote all together and
More informationViewCommander- NVR Version 3. User s Guide
ViewCommander- NVR Version 3 User s Guide The information in this manual is subject to change without notice. Internet Video & Imaging, Inc. assumes no responsibility or liability for any errors, inaccuracies,
More informationOptiflex Interactive Video System
Optiflex Interactive Video System Optiflex Interactive Video System 1 Before You Start...............2 On-site Video Surveillance.......6 Touchscreen Interface Viewing...10 Secure Remote Look-in........16
More informationApproved by: / / R. Battaglia 12/16/2016
Fabrication Laboratory Revision: H Rev Date: 12/16/16 Approved by: Process Engineer / / R. Battaglia 12/16/2016 Equipment Engineer 1 SCOPE The purpose of this document is to detail the use of the Varian
More informationApplication Note AN-708 Vibration Measurements with the Vibration Synchronization Module
Application Note AN-708 Vibration Measurements with the Vibration Synchronization Module Introduction The vibration module allows complete analysis of cyclical events using low-speed cameras. This is accomplished
More informationFilm-Tech. The information contained in this Adobe Acrobat pdf file is provided at your own risk and good judgment.
Film-Tech The information contained in this Adobe Acrobat pdf file is provided at your own risk and good judgment. These manuals are designed to facilitate the exchange of information related to cinema
More informationThe DataView PowerPad III Control Panel
Setting Up a Recording Session in the DataView PowerPad III Control Panel By Mike Van Dunk The DataView PowerPad III Control Panel is designed for working with AEMC PowerPad III Power Quality Analyzers,
More information3.22 Finalize exact specifications of 3D printed parts.
3.22 Finalize exact specifications of 3D printed parts. This is the part that connect between the main tube and the phone holder, it needs to be able to - Fit into the main tube perfectly - This part need
More informationCytoFLEX Flow Cytometer Quick Start Guide
Sheath Waste CLASS 1 LASER PRODUCT COMPLIES WITH 21 CFR 1040.10 AND 1040.11 EXCEPT FOR DEVIATIONS PURSUANT TO LASER NOTICE NO. 50 DATED JUNE 24, 2007 MANUFACTURED Sheath B49008AC February 2015 CytoFLEX
More informationQuick Start Bruker Dimension Icon AFM
Do not remove Quick Start Bruker Dimension Icon AFM March 3, 2015 GLA Contacts Harold Fu (hfu@caltech.edu) Weilai Yu (wyyu@caltech.edu) Bruker Tech Support (AFMSupport@bruker-nano.com 800-873-9750) Watch
More informationDigital Video User s Guide. the Future. now showing
Digital Video User s Guide the Future now showing Welcome the new way to watch Digital TV is TV different than anything you have seen before. It isn t cable it s better. Digital TV offers great channels,
More informationDurham Magneto Optics Ltd. NanoMOKE 3 Wafer Mapper. Specifications
Durham Magneto Optics Ltd NanoMOKE 3 Wafer Mapper Specifications Overview The NanoMOKE 3 Wafer Mapper is an ultrahigh sensitivity Kerr effect magnetometer specially configured for measuring magnetic hysteresis
More informationSZU OPERATING INSTRUCTIONS SAT NAVI
SZU 21-00 O P ER ATI N G I N S T R U C T I O N S SAT NAVI Operation Instructions SZU 21-00 Safety Notes Turn off the receiver or any used power supply before installing, to avoid short-circuit. Installation
More informationU S E R G U I D E HD1000
U S E R G U I D E HD1000 1 W e l c o m e t o R o k u! In This Guide... Bring your HDTV to life with Roku. For the first time, you ll enjoy viewing your favorite digital photos in high-definition on your
More informationVISSIM TUTORIALS This document includes tutorials that provide help in using VISSIM to accomplish the six tasks listed in the table below.
VISSIM TUTORIALS This document includes tutorials that provide help in using VISSIM to accomplish the six tasks listed in the table below. Number Title Page Number 1 Adding actuated signal control to an
More informationMy XDS Receiver- Affiliate Scheduler
My XDS Receiver- Affiliate Scheduler The XDS distribution system represents a marked departure from the architecture and feature set of previous generations of satellite receivers. Unlike its predecessors,
More informationSC24 Magnetic Field Cancelling System
SPICER CONSULTING SYSTEM SC24 SC24 Magnetic Field Cancelling System Makes the ambient magnetic field OK for the electron microscope Adapts to field changes within 100 µs Touch screen intelligent user interface
More informationVideo Conference Classroom Documentation
Updated: 8/18/2017 Video Conference Classroom Documentation Contents About These Classrooms... 2 Where... 2 Podium Overview... 2 On Top of Podium... 2 Inside the Podium... 2 Equipment Information... 2
More informationThe Martin Wood Lecture Theatre. Notes for Speakers on the use of the Audio/Visual equipment
The Martin Wood Lecture Theatre Notes for Speakers on the use of the Audio/Visual equipment The new lecture theatre has a suite of AV equipment intended to aid in the presentation of lectures and conferences.
More informationShow Designer 3. Software Revision 1.15
Show Designer 3 Software Revision 1.15 OVERVIEW... 1 REAR PANEL CONNECTIONS... 1 TOP PANEL... 2 MENU AND SETUP FUNCTIONS... 3 CHOOSE FIXTURES... 3 PATCH FIXTURES... 3 PATCH CONVENTIONAL DIMMERS... 4 COPY
More informationUniversity of St Andrews - Medical and Biological Science Building Handbook LECTURE THEATRE
LECTURE THEATRE 1. Booking. The School of Medicine gets first call to book the lecture theatre during teaching term time 1. Other schools or students can make bookings for the lecture theatre from mid
More informationSC24 Magnetic Field Cancelling System
SPICER CONSULTING SYSTEM SC24 SC24 Magnetic Field Cancelling System Makes the ambient magnetic field OK for the electron microscope Adapts to field changes within 100 µs Touch screen intelligent user interface
More informationModel#: IN-MDRI3MF. Hardware User Manual. 3MP Indoor Mini Dome with Basic WDR, Fixed lens. (PoE) Ver. 2013/02/04
Model#: IN-MDRI3MF 3MP Indoor Mini Dome with Basic WDR, Fixed lens Hardware User Manual (PoE) Ver. 2013/02/04 Table of Contents 0. Precautions 3 1. Introduction 4 Package Contents... 4 Features and Benefits...
More informationD-Lab & D-Lab Control Plan. Measure. Analyse. User Manual
D-Lab & D-Lab Control Plan. Measure. Analyse User Manual Valid for D-Lab Versions 2.0 and 2.1 September 2011 Contents Contents 1 Initial Steps... 6 1.1 Scope of Supply... 6 1.1.1 Optional Upgrades... 6
More informationGustavus Adolphus College. Some Scientific Software of Interest
CHE 372 Gustavus Adolphus College Some Scientific Software of Interest A. Literature Databases There are several literature databases commonly used to conduct scientific literature reviews. The two most
More informationDigital TV. User guide. Call for assistance
Digital TV User guide Call 623-4400 for assistance Table of Contents Watch TV with Tbaytel Digital TV 1 Turn On Your TV and Tbaytel Digital TV 1 Turn Off the Screen Saver 1 Turn Off the TV 1 Use the Set
More informationMestReNova Manual for Chem 201/202. October, 2015.
1. Introduction to 1-D NMR Data Processing with MestReNova The MestReNova program can do all of the routine NMR data processing needed for Chem 201 and 202 and will be available through the Reed downloads
More informationCARESTREAM VITA/VITA LE/VITA SE CR System Long Length Imaging User Guide
CARESTREAM VITA/VITA LE/VITA SE CR System Long Length Imaging User Guide Use of the Guide Carestream CR Systems are designed to meet international safety and performance standards. Personnel operating
More informationModel#: IN-DI2MIRF 2MP Indoor Dome with True Day/Night, IR, Basic WDR, Fixed lens
Model#: IN-DI2MIRF 2MP Indoor Dome with True Day/Night, IR, Basic WDR, Fixed lens Hardware User Manual (PoE) Ver.2013/01/17 Table of Contents 0. Precautions 3 1. Introduction 4 Package Contents...4 Features
More informationSX80 with Touchpad 10 User Guide
SX80 with Touchpad 10 User Guide Rev 11May2017 Page 1 of 19 Table of Contents OVERVIEW.... 3 BASIC NAVIGATION.... 4 GENERAL USE.... 5 Setup... 5 Camera Controls... 6 Microphone.... 8 Volume.... 9 Site
More informationNMS ETA TM Stereo Advanced Digital TV Set-top Box
Digital TV HD Set-top Box STB2-T2 NMS ETA TM Stereo Advanced Digital TV Set-top Box User Guide CONTENTS Safety Information... 3 Introduction... 4 Set-top Box Front Panel... 4 Set-top Box Back Panel...
More informationX-Sign 2.0 User Manual
X-Sign 2.0 User Manual Copyright Copyright 2018 by BenQ Corporation. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, stored in a retrieval system or translated
More informationInitially, you can access the Schedule Xpress Scheduler from any repair order screen.
Chapter 4 Schedule Xpress Scheduler Schedule Xpress Scheduler The Schedule Xpress scheduler is a quick scheduler that allows you to schedule appointments from the Repair Order screens. At the time of scheduling,
More informationRevision Protocol Date Author Company Description 1.1 May 14, Seth LOUTH Revised for formatting
PRODUCT ADC TOPIC ODETICS TCS-2000 CART MACHINE DATE: May 14, 1999 REVISION HISTORY Revision Protocol Date Author Company Description 1.1 May 14, Seth LOUTH Revised for formatting 1999 Olitzky 1.0 Aug.
More informationSR-D8-M, SR-D8-S. (Ver ) SOFTWARE INSTRUCTIONS
SOFTWARE INSTRUCTIONS active l ine array speak er SYStems SR-D8-M, SR-D8-S (Ver. 1.1.1) Thank you for purchasing TOA's Active Line Array Speaker Systems. Please carefully follow the instructions in this
More informationGrid Interactive Advanced Features Self-Supply Mode
Grid Interactive Advanced Features Self-Supply Mode CHILICON POWER MICROINVERTERS AND CP-100 CORTEX GATEWAY SYSTEM OVERVIEW Chilicon Power Advanced Inverter functions Power Factor Control: -0.8 to +0.8
More informationEDL8 Race Dash Manual Engine Management Systems
Engine Management Systems EDL8 Race Dash Manual Engine Management Systems Page 1 EDL8 Race Dash Page 2 EMS Computers Pty Ltd Unit 9 / 171 Power St Glendenning NSW, 2761 Australia Phone.: +612 9675 1414
More informationSetup Guide. Read me BefoRe unpacking!
Setup Guide Read me BefoRe unpacking! Package Contents In The Replicator package The Replicator SD card (in The Replicator SD card slot) In the Accessory Box found within The Replicator frame Single or
More informationConeXus User Guide. HHAeXchange s Communication Functionality
HHAeXchange ConeXus User Guide HHAeXchange s Communication Functionality Copyright 2017 Homecare Software Solutions, LLC One Court Square 44th Floor Long Island City, NY 11101 Phone: (718) 407-4633 Fax:
More informationStandard Operating Procedure Of The CytoFLEX S Flow Cytometer At The Institute Of Virology
Standard Operating Procedure Of The CytoFLEX S Flow Cytometer At The Institute Of Virology Before starting with the CytoFLEX S Flow Cytometer: 1. A new user needs to get an introduction to the flow cytometer
More informationFootnotes and Endnotes
Footnotes and Endnotes Sometimes when writing a paper it is necessary to insert text at the bottom of a page in a document to reference something on that page. You do this by placing a footnote at the
More informationHV/PHA Adjustment (PB) Part
HV/PHA Adjustment (PB) Part Contents Contents 1. How to set Part conditions...1 1.1 Setting conditions... 1 2. HV/PHA adjustment sequence...7 3. How to use this Part...9 HV/PHA Adjustment (PB) Part i
More information