Standard Operating Procedure II: EDS (Bruker Flat-Quad)

Size: px
Start display at page:

Download "Standard Operating Procedure II: EDS (Bruker Flat-Quad)"

Transcription

1 Standard Operating Procedure II: EDS (Bruker Flat-Quad) ywcmatsci.yale.edu ESC II, Room A119F 810 West Campus Drive West Haven, CT Version 1.1, October 2018

2 1

3 > FOLLOW the SOP strictly to keep the instrument in good condition. Any violation will lead to user account suspension. > NEVER use your own USB drive on instrument computer. Data can be transferred with the Jump Drive provided by the Core. > NEVER surf the web on the instrument computer to minimize the risk of the computer being hacked. > NEVER allow other users to get access to instrument computer on your reservation. > REPORT any issues to Core director immediately so they can be fixed on time. 1

4 2

5 Table of Contents 1 Introduction Sample Preparation Starting Instrument System Status Check EDS Sample loading EDS setup on computer EDS setup on EDS computer EDS measurement Closing EDS detection:

6 1

7 Energy Dispersive Spectroscopy (EDS) Standard Operating Procedure 1 Introduction 1) Instrument features: > Cold field emission (CFE) e-beam source: high resolution on conductive surfaces (0.8 nm on Au clusters/magnetic tape) > Sliding-in annular Energy Dispersive Spectroscopy (EDS) detector: high elemental mapping resolution > Sliding-in annular Photo Diode PD-BSE detector: high signal intensity from backscattered electron > Scanning Transmission Electron Microscopy (STEM) detector: high resolution compositional contrast imaging, ideal for EDS mapping 2) Location Materials Characterization Core Room E West Campus Drive West Haven, CT ) Primary Staff Contact Dr. Min Li Tel: min.li@yale.edu Office: ESC II, Room E119D Zishan Wu, Lab Assistant zishan.wu@yale.edu (cell) Office: ESC II Yiren Zhong, Lab Assistant yiren.zhong@yale.edu (cell) The Yale West Campus MCC Facilities are operated for the benefit of all researchers. If you encounter any problems with this facility, please contact the staff member listed above immediately. There is never a penalty for asking questions. If the equipment is not behaving exactly the way it should, contact a staff member. Warning: Please follow strictly the SOP to keep the facility at good condition. We DO NOT recommend user explorations on program unless endorsed by core director. 1

8 2 Sample Preparation 1 Note: TEM samples can also be used with a dedicated TEM sample holder for EDS. The EDS mapping resolution can reach below 100 nm with TEM samples using STEM-EDS mode. 1) Always wear gloves for vacuum sample preparation! Change gloves if touched computer keyboard and mouse. 2) Follow instructions from SOP to prepare samples. 3) Align sample surface strictly with the height gauge tip as shown below. Warning: Samples mounted above Height Gauge tip will hit the EDS detector and the repair fee ($40,000) will be charged to PI s account. Note: no need to use height gauge if TEM sample holder is used. 4) Clean sample holder: bring the specimen stub inside the fume hood and blow off loose particles on the sample surface using the N2 gun. 2 3 Starting Instrument Follow Section 3 Starting Instrument in the separate SOP. 4 System Status Check Follow Section 4 System Status Check in the separate SOP. 5 EDS Sample loading 1) Follow Section 5 Loading the Specimen in the separate SOP and leave the sample at exchange position 2) Turn off the small LCD from the back (top left) to avoid EDS detector damage and EDS spectra distortion. Warning: The LCD unit also provides power to the infrared camera inside main chamber. If it is left on during /EDS scan, the camera will continuously emit infrared light to flood the EDS detector even at fully extracted position and shorten its lifetime quickly. 6 EDS setup on computer 1) In PC_ program on the computer (left side), click HOME button to move the sample holder to HOME position 2) Turn off the small LCD by pressing the switch in the back top-left corner of the monitor. 1 Always wear gloves for your sample preparation in your own lab or in MCC! Warnings will be given for violations and the user account will be revoked after three warnings with notice to PI. Further training at PI s expense will be required to resume the account. 2 This step is crucial to keep the chamber vacuum at good pressure, which in turn improves the imaging resolution with less surface contamination and keeps the lens system at good condition. 2

9 Note: This step is crucial, otherwise the EDS detector will be flooded by ambient signals leading to fat peaks in spectra. 3) Click Set button in the Stage tab in the popup window below: 4) Check the FQ-EDX box in the popup window below and hit OK. 5) The stage Z position should move to the default height at 14 mm.as highlighted in the window below: Warning: Never change the Z position < 11 mm. This will cause the sample holder crashing into the EDS detector. This severe SOP violation will lead to user account suspension and charge on PI s account. 6) Check the working distance (W.D.) in PC_ below and make sure it is set at 15 mm. If not, click on the top menu bar to switch imaging mode from LM (low magnification) to high magnification, then change W.D. in the window below. Switch back to LM. 7) Make sure Vacc (e-beam accelerating voltage) is OFF (check in PC_ program) 3

10 Note: it is OK to leave the e-beam ON if decide to do EDS in the middle of measurements. 8) E-beam parameter setup for EDS: a) E-beam voltage (the overvoltage to excite EDS X-ray emission lines) setup: > Check the EDS periodic table and choose a set of characteristic X-ray emission lines for intended sample elements that have least energy overlapping. > The chosen e-beam voltage, the overvoltage should be at least 1.5x the highest X- ray line in matrix elements and below 20 kv. Tip: smaller beam voltage means higher EDS special/mapping resolution. Warning: NEVER choose e-beam voltage above 20 kev!! The EDS detector will be burned!! The user s EDS access will be suspended and the cost ($70k) will be charged on PI s account. b) E-beam current setup: > Switch to High Magnification (HM) mode, set the Probe current to High in the Operation condition window below. > For chosen overvoltage smaller than 12 kv, consider increasing the e-beam current further by lowering the Cond. Lens1 in the window below: select smaller numbers than default EDS setup on EDS computer 1) Log into EDS computer through FOM Screen Locker. a) The EDS software Esprit window should be always kept ON. b) If the Esprit program was closed and the EDS computer was logged off, select the profile PC- with password hitachi 2) Check EDS detector window position on the detector highlighted on the pictures below and make sure it was set by previous user at 20 kv. If not, report to manager immediately. Warning: It is the serious SOP violation if forgot to switch the detector window back to 20 kv after use. This may lead to detector damage and repair charge will be applied to user s PI account. 4

11 20 kv 3) Check to make sure the Specimen Chamber SC pressure is at LE-4 Pascal. 4) Switch EDS detector in operation mode and set up detector configuration parameters: a) In EDS Esprit operating program, click the triangle in the EDS tab at the bottom left corner as indicated below: b) In the EDS DETECTOR CONFIGURATION popup window below: > Check 130 kcps in Pulse throughput and 20 kev in Maximum energy. > Make sure the Thermostat is checked in Cooling setting. > Check Normal operation in Mode setting. Select Yes in the popup INFORMATION window. Warning: the EDS detector cooling should be started only when the Specimen Chamber pressure has reached LE-4 Pascal to minimize detector contamination. 5

12 c) Close both INFORMATION and EDS DETECTOR CONFIGURATION window above. d) The EDS detector will be cooled to the operating temperature of -20 ± 0.5 ºC in about 5 minutes. 5) EDS detector window setting at different overvoltage/e-beam voltage: a) If the beam voltage 6 kv: > Enter Esprit System window by clicking button on the bottom left of the side menu and clicking tab as shown below. Click button to open Detector Data window and select 1µm Mylar from dropdown list. > Go back to EDS detector head on chamber, switch the detector window from 20 kv to 6 kv. 6

13 b) If the beam voltage is between 6 kv and 12 kv: > Enter Esprit System window by clicking button on the left side menu and clicking as shown below. Click to open Detector Data window and select 1+2 µm Mylar from dropdown list. > Go back to EDS detector head on chamber, switch the detector window from 20 kv to 12 kv. c) If the beam voltage is larger than 12 kv (Never change voltage to above 20 kv to damage the detector): > Enter Esprit System window by clicking button on the left side menu and clicking as shown below. Click to open Detector Data window and select 1+6 µm Mylar from dropdown list. > Go back to EDS detector head on chamber, switch the detector window from 20 kv to 20 kv. 8 EDS measurement 1) During EDS detector cooling, turn on e-beam in PC_ program and find the area of interest. 2) Check EDS detector and make sure it has reached the operation temperature -20 ± 0.5 ºC 3) Click button in EDS tab in Esprit program and click OK on DETECTOR POSITION window to move the detector to the acquisition position right above sample. Warning: > If users decide to run EDS in the middle of imaging, make sure to check FQ-EDS mode in PC_ program Stage > Set setting window to bring down the sample stage to required EDS stage height (14 mm). Please check Section 6, Step 3) on page 3. 7

14 > Failure to follow will lead to EDS detector crashing into sample stage. User s access will be suspended and repair cost will be charged on PI s account. 4) EDS dead time adjustment (check steps in Spectrum Acquisition mode below): > As shown in the EDS tab below, the Dead time in percentage denotes the signal processing capability of the EDS controller. The higher number indicates the higher signal counts and the possibility of the signal pileup or false peak appearance. The Dead time should be kept around 30%. Tips: > To improve EDS mapping resolution, the dead time should be kept just above 10%. The input count rate (ICR) should be at least one hundred of kcps to keep a good signal/noise ratio. > For spectral acquisition, the dead time can be kept at around 30%. 5) Spectrum Acquisition Mode: Note: Make sure the imaging area for spectrum collection is homogeneous if using this mode. Consider Objects mode next for inhomogeneous surfaces. a) Go back to PC_ window, decide the image magnification and make an initial adjustment on focus and stigma to get a good image. b) In EDS Esprit window click on the left side menu to enter Spectra workspace c) Click button and check and adjust the Dead time within ~10 30%: > If the Dead time is below 10%, in PC_ window (do the opposite steps if Dead time is above 30%): > Increase the e-beam emission current (maximum 30 µa); > Choose High Probe current; > If necessary, change Cond. Lens 1 setting from 5 to 1 or 2, the smaller number the higher the signal intensity. 8

15 d) Leave spectral acquisition on, go back to PC_ window, adjust focus and stigma to get sharp image. e) If a good signal/noise ratio has achieved on the spectrum, click again to stop acquisition. f) To add collected spectrum into project or report, hit the button on the top right corner of the spectral workspace g) To save the data in Bruker spectra format (*.spx) or export to *.txt or *.xlsx format, click the lower button. 6) Objects Mode: Note: this mode should be chosen for inhomogeneous surface spectral analysis. a) Adjust Dead time within 10 30% in Spectrum Acquisition Mode above b) Hit on the left side menu to enter Object workspace c) In PC_ window adjust focus and stigma to get sharp image d) Go back to EDS Esprit window and click button to capture image. e) Select the desired object type on the bottom menu bar and click on captured image above to specify positions. Tips: > The EDS spatial resolution is e-beam voltage/overvoltage dependent which is around 1 µm. > Choose Point object mode if the interested feature is too small. > Allow longer collection time if the point or small square or circle objects are chosen. f) Click to highlight all objects and click g) If a good signal/noise ratio has achieved on the spectrum, click again to stop acquisition. h) To save object data, click the button on the top right corner of the workspace window. i) To save the spectrum, click the lower spectrum chart. 7) Line Scan Mode: a) Adjust Dead time within ~ 10 30% in Spectrum Acquisition Mode 9

16 b) Hit on the left side menu to enter Line Scan workspace c) In PC_ window adjust focus and stigma to get sharp image d) Go back to EDS Esprit window and click button to capture image. e) Highlight the line and drag and adjust the endpoints to the desired position f) Set Point count of the line scan and click g) Use the icon to identify elements h) If a good signal/noise ratio has achieved on the spectrum, click again to stop acquisition. i) To save line scan data, click the button on the top right corner of the workspace window. j) To save the profile, click the lower profile chart. 8) Mapping Mode: a) Adjust Dead time to be just above 10 % in Spectrum Acquisition Mode b) Click button on Scan tab to activate image drift correction. Make sure the button is highlighted in red to enable drift correction. c) Hit on the left side menu to enter Mapping workspace d) In PC_ window adjust focus and stigma to get sharp image e) Go back to EDS Esprit window and hit button to capture image. f) Click button to start Mapping. g) Use the icon to identify elements in spectrum window during mapping. 10

17 h) Before stop mapping: > Click Spectrum tab above the top right corner of Map window, make sure a good signal/noise ratio on interested elemental peaks in the spectrum. > Check if the mapping resolution has not improved with time. Typical mapping collection time depends on signal intensity (~ mins or even longer). i) Click again to stop Mapping. j) To save map data, click the button on the top right corner of the workspace window. k) To save the map image, click the lower image window. l) To save individual element image in the thumbnail on the bottom, click the thumbnail bar. 9 Closing EDS detection: 1) In PC_ program, click the OFF button to turn off electron beam Note: the beam can be left ON if plan to do further imaging. 2) Click Standby button in EDS DETECTOR CONFIGURATION window to switch the EDS detector to Standby Mode 11

18 3) Check EDS detector window position and make sure it is switched back to 20 kv: Warning: It is crucial to set the detector window back to 20 kv to avoid e-beam damage. Violation will lead to detector damage and repair charge will be applied to user s PI account. 20 kv 4) Fully retract EDS detector by clicking in EDS Esprit program and wait until the detector is fully retracted. 5) In EDS Esprit window, go to tab, click button to open Detector Data window and select 1+µm Mylar from dropdown list. 6) In PC_ window: > Change Vacc back to 10 kv and the Ie back to 10 µa > Change the Probe current back to Norm. > Change the Cond. Lens1 back to 5.0 in the window below: > Click the EXC button to move the specimen stage to the exchange position. Note: click Home button if continue to do imaging. > Click Set button in the Stage tab; uncheck FG-EDX box below. 12

19 7) Wait for the EDS detector temperature rises above 20 C (~ 20 minutes, be patient!) and then take sample out of chamber. Warning: Never start sample transfer while EDS detector is still being cooled. This will induce contamination on the detector if still below 20 C when the Specimen Chamber door is open. 8) During EDS detector warmup: > Make sure sample holder is at exchange (EXC) position > Data transfer is complete > Log off FOM on computer in your FOM account. 9) Check to make sure the EDS detector temperature rises above 20 C and then 10) Turn ON small LCD 11) Finish sample unloading from chamber. 12) Turn OFF small LCD 13) Log off FOM on EDS computer Note: Keep Esprit program ON and maximized on screen 14) Clear the work bench and dispose gloves, Kimwipes 13

20 14

Operation Procedure for Phillips XL30 ESEM

Operation Procedure for Phillips XL30 ESEM Operation Procedure for Phillips XL30 ESEM The ESEM will be left in the ON state when not in use. The chamber will be at high vacuum, filament on, stage at home position, VAC and HT buttons lit, and monitor

More information

Scanning Electron Microscopy (FEI Versa 3D Dual Beam)

Scanning Electron Microscopy (FEI Versa 3D Dual Beam) Scanning Electron Microscopy (FEI Versa 3D Dual Beam) This operating procedure intends to provide guidance for basic measurements on a standard sample with FEI Versa 3D SEM. For more advanced techniques

More information

Technical Procedure for Scanning Electron Microscope/ Energy Dispersive X-Ray System (SEM/EDX) for non-gsr Casework

Technical Procedure for Scanning Electron Microscope/ Energy Dispersive X-Ray System (SEM/EDX) for non-gsr Casework Technical Procedure for Scanning Electron Microscope/ Energy Dispersive X-Ray System (SEM/EDX) for non-gsr Casework 1.0 Purpose This technical procedure shall be followed for the operation of the Scanning

More information

CHECKLIST FOR VERIOS OPERATION 1. GENERAL The SEM lab is used assuming "operating room" cleanliness, i.e., the SEM lab is a high visibility lab and

CHECKLIST FOR VERIOS OPERATION 1. GENERAL The SEM lab is used assuming operating room cleanliness, i.e., the SEM lab is a high visibility lab and CHECKLIST FOR VERIOS OPERATION 1. GENERAL The SEM lab is used assuming "operating room" cleanliness, i.e., the SEM lab is a high visibility lab and must be kept clean and neat so clean up behind yourself

More information

1.2 Universiti Teknologi Brunei (UTB) reserves the right to award the tender in part or in full.

1.2 Universiti Teknologi Brunei (UTB) reserves the right to award the tender in part or in full. TENDER SPECIFICATIONS FOR THE SUPPLY, DELIVERY, INSTALLATION AND COMMISSIONING OF ONE UNIT OF VARIABLE PRESSURE ENVIRONMENTAL SCANNING ELECTRON MICROSCOPE (SEM) CUM ENERGY DISPERSIVE SPECTROSCOPY (EDS)

More information

FEI FIB Focused Ion Beam

FEI FIB Focused Ion Beam Operating Manual Part 1 FEI FIB Focused Ion Beam IF IN DOUBT, ASK 1.00 IF ANYTHING UNUSUAL HAPPENS, OR IF THERE IS ANYTHING YOU ARE UNSURE ABOUT, STOP AND CONTACT ME! DO NOT PROCEED OR ATTEMPT TO FIX THE

More information

Quick Start Bruker Dimension Icon AFM

Quick Start Bruker Dimension Icon AFM Do not remove Quick Start Bruker Dimension Icon AFM March 3, 2015 GLA Contacts Harold Fu (hfu@caltech.edu) Weilai Yu (wyyu@caltech.edu) Bruker Tech Support (AFMSupport@bruker-nano.com 800-873-9750) Watch

More information

PHI 5000 VersaProbe TM Operator s Guide

PHI 5000 VersaProbe TM Operator s Guide PHI 5000 VersaProbe TM Operator s Guide Part No. 705921 Rev. A Copyright 2006 ULVAC-PHI, INC. 370 Enzo, Chigasaki, JAPAN The PHI logo ( ) is a registered trademark of ULVAC-PHI, INC. Physical Electronics,

More information

2.1. Log on to the TUMI system (you cannot proceed further until this is done).

2.1. Log on to the TUMI system (you cannot proceed further until this is done). FEI DB235 ex-situ lift out TEM sample preparation procedure Nicholas G Rudawski ngr@ufledu (805) 252-4916 Last updated: 06/19/15 DISCLAIMER: this procedure describes one specific method for preparing ex-situ

More information

Standard Operating Procedure of nanoir2-s

Standard Operating Procedure of nanoir2-s Standard Operating Procedure of nanoir2-s The Anasys nanoir2 system is the AFM-based nanoscale infrared (IR) spectrometer, which has a patented technique based on photothermal induced resonance (PTIR),

More information

Figure 1. MFP-3D software tray

Figure 1. MFP-3D software tray Asylum MFP-3D AFM SOP January 2017 Purpose of this Instrument: To obtain 3D surface topography at sub-nanometer scale resolution, measure contact and friction forces between surfaces in contact, measure

More information

The PHI VersaProbe operates with two essential software programs: PHI Summitt and Vacuum Watcher. A third program, MultiPak, handles data reduction.

The PHI VersaProbe operates with two essential software programs: PHI Summitt and Vacuum Watcher. A third program, MultiPak, handles data reduction. PHI VersaProbe Scanning XPS System I. Overview The PHI VersaProbe operates with two essential software programs: PHI Summitt and Vacuum Watcher. A third program, MultiPak, handles data reduction. PHI Summitt

More information

Standard Operating Procedure for FEI Helios 660 NanoLab Part I: SEM Version

Standard Operating Procedure for FEI Helios 660 NanoLab Part I: SEM Version Standard Operating Procedure for FEI Helios 660 NanoLab Part I: SEM Version Helios reservations may be made online using the NERCF website. Note: Always wear gloves when venting the system and exchanging

More information

INCA ENERGY EDS TRAINING. System Block Diagram. INCA Energy Software. Xiang Yang EM SMU. Navigators. Point & ID Navigator.

INCA ENERGY EDS TRAINING. System Block Diagram. INCA Energy Software. Xiang Yang EM SMU. Navigators. Point & ID Navigator. INCA ENERGY EDS TRAINING Xiang Yang EM Center @ SMU System Block Diagram INCA Energy Software Navigators Point & ID Navigator Data tree 1 Spectrum Export Bitmap TIFF JPEG Metafile EMSA ISIS Optimum Conditions

More information

University of Minnesota Nano Fabrication Center Standard Operating Procedure

University of Minnesota Nano Fabrication Center Standard Operating Procedure Equipment Name: Focused Ion Beam (FIB) Coral Name: fib Revision Number: 2 Model: FEI Quanta 200 3D Revisionist: Kevin Roberts Location: Area 3 Date: 9/17/2013 1 Description The Quanta 200 3D is a DualBeam

More information

Kratos AXIS Ultra DLD X-ray Photoelectron Spectrometer Instructions

Kratos AXIS Ultra DLD X-ray Photoelectron Spectrometer Instructions Kratos AXIS Ultra DLD X-ray Photoelectron Spectrometer Instructions Note: Enter your complete name, CMRF user code, and date into the record book. You must have previously filled out a project information

More information

HV/PHA Adjustment (PB) Part

HV/PHA Adjustment (PB) Part HV/PHA Adjustment (PB) Part Contents Contents 1. How to set Part conditions...1 1.1 Setting conditions... 1 2. HV/PHA adjustment sequence...7 3. How to use this Part...9 HV/PHA Adjustment (PB) Part i

More information

Part names (continued) Remote control

Part names (continued) Remote control Introduction Part names (continued) Remote control (1) STANDBY ( 25) (1) (2) ON ( 25) (3) (3) ID - 1 / 2 / 3 / 4 s ( 18) (4) (4) COMPUTER 1 ( 27) (7) (5) COMPUTER 2 * (8) (6) COMPUTER 3 * (10) (13) (7)

More information

LUMIGEN INSTRUMENT CENTER X-RAY CRYSTALLOGRAPHIC LABORATORY: WAYNE STATE UNIVERSITY

LUMIGEN INSTRUMENT CENTER X-RAY CRYSTALLOGRAPHIC LABORATORY: WAYNE STATE UNIVERSITY Standard Operating Procedure for the Bruker X8 APEX II Single-Crystal X- Ray Diffractometer Contact Manager: Dr. Cassie Ward ward@wayne.edu Office room 061 Chemistry (313) 577-2587 LIC Lab: (313) 577-0518

More information

1. Check the accelerating voltage, must be at 200 kv (right screen), HT (µa) (left panel) at and Emission (left panel) at

1. Check the accelerating voltage, must be at 200 kv (right screen), HT (µa) (left panel) at and Emission (left panel) at JEOL 2010F MANUAL Quick check list 1. Check the accelerating voltage, must be at 200 kv (right screen), HT (µa) (left panel) at 0.96-0.97 and Emission (left panel) at 155-160. 2. Check the vacuum sequence

More information

Dektak Step by Step Instructions:

Dektak Step by Step Instructions: Dektak Step by Step Instructions: Before Using the Equipment SIGN IN THE LOG BOOK Part 1: Setup 1. Turn on the switch at the back of the dektak machine. Then start up the computer. 2. Place the sample

More information

STANDARD OPERATING PROCEDURE: ASYLUM MFP-3D AFM

STANDARD OPERATING PROCEDURE: ASYLUM MFP-3D AFM STANDARD OPERATING PROCEDURE: ASYLUM MFP-3D AFM Purpose of this Instrument: To obtain 3D surface topography at sub-nanometer scale resolution, measure contact and friction forces between surfaces in contact,

More information

FIB Operating Procedure. Effective Date: 08/14/2012 Author(s): Jiong Hua Phone:

FIB Operating Procedure. Effective Date: 08/14/2012 Author(s): Jiong Hua Phone: FIB Operating Procedure Effective Date: 08/14/2012 Author(s): Jiong Hua Phone: 402-472-3773 Email: jhua2@unl.edu 1 1 Introduction 1.1 Key Words Focused Ion Beam (FIB), FEI Strata 201, Ion milling 1.2 Purpose

More information

Nova NanoSEM Superior Imaging and Analytical Performance

Nova NanoSEM Superior Imaging and Analytical Performance Nova NanoSEM Superior Imaging and Analytical Performance FEI Nova NanoSEM scanning electron microscopes combine best-in-class imaging with superb analytical performance in one easy-to-use instrument.

More information

Topic: Instructional David G. Thomas December 23, 2015

Topic: Instructional David G. Thomas December 23, 2015 Procedure to Setup a 3ɸ Linear Motor This is a guide to configure a 3ɸ linear motor using either analog or digital encoder feedback with an Elmo Gold Line drive. Topic: Instructional David G. Thomas December

More information

Tender Notification for the procurement of a Scanning Electron Microscope" at IISc (Last Date for submission of tenders: 3 rd October 2018)

Tender Notification for the procurement of a Scanning Electron Microscope at IISc (Last Date for submission of tenders: 3 rd October 2018) Tender Notification for the procurement of a Scanning Electron Microscope" at IISc (Last Date for submission of tenders: 3 rd October 2018) Dear Sir/Madam, We are looking for a high-resolution scanning

More information

SEM- EDS Instruction Manual

SEM- EDS Instruction Manual SEM- EDS Instruction Manual Double-click on the Spirit icon ( ) on the desktop to start the software program. I. X-ray Functions Access the basic X-ray acquisition, display and analysis functions through

More information

AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode)

AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode) AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode) 1. Log into the Log Usage system on the SMIF web site 2. Open Nanoscope 6.14r1 software by double clicking on the Nanoscope 6.14r1 desktop

More information

1.0 ThermoNicolet Nexus 670 FTIR Spectrometer Instructions

1.0 ThermoNicolet Nexus 670 FTIR Spectrometer Instructions 1.0 ThermoNicolet Nexus 670 FTIR Spectrometer Instructions 1.1 Click on the OMNIC icon to open the software. 1.2 Check that a signal is being measured by opening the Experiment Setup menu under the Collect

More information

User Manual OVP Raman

User Manual OVP Raman Version 6 User Manual OVP Raman 2006 BRUKER OPTIK GmbH, Rudolf-Plank-Straße 27, D-76275 Ettlingen, www.brukeroptics.com All rights reserved. No part of this manual may be reproduced or transmitted in any

More information

Tender Notification for the procurement of a "Dual beam (FIB - FE SEM) system" at IISc (Last Date for submission of tenders: 31st March 2016)

Tender Notification for the procurement of a Dual beam (FIB - FE SEM) system at IISc (Last Date for submission of tenders: 31st March 2016) Tender Notification for the procurement of a "Dual beam (FIB - FE SEM) system" at IISc (Last Date for submission of tenders: 31st March 2016) Dear Sir/Madam, Kindly send your best quotation for the following

More information

Dektak II SOP Revision 1 05/30/12 Page 1 of 5. NRF Dektak II SOP

Dektak II SOP Revision 1 05/30/12 Page 1 of 5. NRF Dektak II SOP Page 1 of 5 NRF Dektak II SOP The Dektak II-A is a sensitive stylus profilometer. A diamond-tipped stylus is moved laterally across the surface while in contact and measures deflections of the tip. It

More information

Stark Spectroscopy Deanna s Experimental Procedure NWU Hupp Lab Fall 2003

Stark Spectroscopy Deanna s Experimental Procedure NWU Hupp Lab Fall 2003 Stark Spectroscopy Deanna s Experimental Procedure NWU Hupp Lab Fall 2003 1. Generate mixed-valent state of compound check in 1mm cell. Ideally want Abs 1. 2. Setting up the instrument New Dewar i) Approx.

More information

AEROTRAK PORTABLE AIRBORNE PARTICLE COUNTER MODEL 9110 QUICK START GUIDE

AEROTRAK PORTABLE AIRBORNE PARTICLE COUNTER MODEL 9110 QUICK START GUIDE AEROTRAK PORTABLE AIRBORNE PARTICLE COUNTER MODEL 9110 QUICK START GUIDE Thank you for purchasing a TSI AeroTrak Model 9110 Portable Airborne Particle Counter (particle counter). This guide will help you

More information

SC24 Magnetic Field Cancelling System

SC24 Magnetic Field Cancelling System SPICER CONSULTING SYSTEM SC24 SC24 Magnetic Field Cancelling System Makes the ambient magnetic field OK for the electron microscope Adapts to field changes within 100 µs Touch screen intelligent user interface

More information

Digital Projector X66 Operator's Guide

Digital Projector X66 Operator's Guide Digital Projector X66 Operator's Guide Thank you for purchasing this projector. WARNING WARNING CAUTION NOTE Trademark acknowledgment 1 Content About this manual.... 1 Content.... 2 Projector features....

More information

Data Collection Using APEX3. March 30, Chemical Crystallography Laboratory

Data Collection Using APEX3. March 30, Chemical Crystallography Laboratory Data Collection Using APEX3 Page 1 of 10 Data Collection Using APEX3 March 30, 2017 Chemical Crystallography Laboratory Author: Douglas R. Powell Data Collection Using APEX3 Page 2 of 10 Distribution Douglas

More information

SC24 Magnetic Field Cancelling System

SC24 Magnetic Field Cancelling System SPICER CONSULTING SYSTEM SC24 SC24 Magnetic Field Cancelling System Makes the ambient magnetic field OK for the electron microscope Adapts to field changes within 100 µs Touch screen intelligent user interface

More information

1.10mm Height 1210 Package. Bi-Color (Multi-Color) Chip LED. Technical Data Sheet. Part No: S155VBC-V12B-B41B

1.10mm Height 1210 Package. Bi-Color (Multi-Color) Chip LED. Technical Data Sheet. Part No: S155VBC-V12B-B41B .mm Height 2 Package Bi-Color (Multi-Color) Chip LED Technical Data Sheet Part No: S55VBC-V2B-B4B Spec No.: S55 Rev No.: V.3 Date: Jul.//25 Page: OF Features: Package in 8mm tape on 7 diameter reel. Bi-color

More information

FEI Strata Dual-beam FIB

FEI Strata Dual-beam FIB FEI Strata Dual-beam FIB Quick start Guide Compiled by Mat t hew Hughes and Tony Chen Page 0 Purpose of this guide INTRODUCTION This quick start guide is meant to provide cursory operational knowledge

More information

TT AFM LongBeach Procedures and Protocols V2.1

TT AFM LongBeach Procedures and Protocols V2.1 TT AFM LongBeach Procedures and Protocols V2.1 1. Startup Procedure 1. Turn on PC: Allow to boot to Windows. Turn on monitor. Password is afm 2. Turn on second PC that controls the video camera. 3. Turn

More information

Personal Protective Equipment Wear nitrile gloves, lab coat, and safety glasses as a minimum protection, unless otherwise indicated.

Personal Protective Equipment Wear nitrile gloves, lab coat, and safety glasses as a minimum protection, unless otherwise indicated. 4pt Bending, Mouse This protocol is for standard Jepsen 4pt bending of adult mouse bone. Safety considerations Please reference the Jepsen laboratory when using this protocol. This protocol is subject

More information

DektakXT Profilometer. Standard Operating Procedure

DektakXT Profilometer. Standard Operating Procedure DektakXT Profilometer Standard Operating Procedure 1. System startup and sample loading: a. Ensure system is powered on by looking at the controller to the left of the computer.(it is an online software,

More information

Abstract. Keywords INTRODUCTION. Electron beam has been increasingly used for defect inspection in IC chip

Abstract. Keywords INTRODUCTION. Electron beam has been increasingly used for defect inspection in IC chip Abstract Based on failure analysis data the estimated failure mechanism in capacitor like device structures was simulated on wafer in Front End of Line. In the study the optimal process step for electron

More information

IMSERC NMR MANUAL 05: Manual Operation of Agilent NMR Spectrometers (Chem350 Interface)

IMSERC NMR MANUAL 05: Manual Operation of Agilent NMR Spectrometers (Chem350 Interface) IMSERC NMR MANUAL 05: Manual Operation of Agilent NMR Spectrometers (Chem350 Interface) Last updated: October 12, 2011 by Josh Kurutz THIS PAGE = QUICK START GUIDE 0) At the computer, make sure VNMRJ is

More information

FSI Calibration Guide Using I1 / LightSpace / Resolve

FSI Calibration Guide Using I1 / LightSpace / Resolve FSI Calibration Guide Using I1 / LightSpace / Resolve Flanders Scientific, Inc. 6215 Shiloh Crossing Suite G Alpharetta, GA 30005 Phone: +1.678.835.4934 Fax: +1.678.804.1882 E-Mail: Support@FlandersScientific.com

More information

PLASMA MONITOR (PT20 UVVis) USER GUIDE

PLASMA MONITOR (PT20 UVVis) USER GUIDE Thin Film Measurement solution Software, sensors, custom development and integration PLASMA MONITOR (PT20 UVVis) USER GUIDE August 2012 Plasma monitor with VFT probe. INTRODUCTION Plasma Monitor includes

More information

Guide to Analysing Full Spectrum/Frequency Division Bat Calls with Audacity (v.2.0.5) by Thomas Foxley

Guide to Analysing Full Spectrum/Frequency Division Bat Calls with Audacity (v.2.0.5) by Thomas Foxley Guide to Analysing Full Spectrum/Frequency Division Bat Calls with Audacity (v.2.0.5) by Thomas Foxley Contents Getting Started Setting Up the Sound File Noise Removal Finding All the Bat Calls Call Analysis

More information

set. Important Note: the system must be calibrated before meaningful quant results can be obtained!

set. Important Note: the system must be calibrated before meaningful quant results can be obtained! Quant Initialization Before any quantitative results can be obtained, the detector parameters and the analysis parameters must be set. Important Note: the system must be calibrated before meaningful quant

More information

Universal Mode. Survey Settings. The Universal Toolbar. First Run - Initial Configuration via the Setup Button

Universal Mode. Survey Settings. The Universal Toolbar. First Run - Initial Configuration via the Setup Button Universal Mode The Universal Toolbar The Universal Toolbar is the unified (and preferred) method of operating Revolution. From within the Universal Toobar, all functionality of the other 3 modes of Revolution

More information

STANDARD OPERATING PROCEDURE:

STANDARD OPERATING PROCEDURE: Q-EXACTIVE STANDARD OPERATING PROCEDURE: Purpose of this Instrument: Essential tool for molecular weight identification for small molecules, peptides, proteins, ect. Location: WVU Chemistry Research Laboratory

More information

Liquid Chromatography- Mass Spectrometer Manual

Liquid Chromatography- Mass Spectrometer Manual Liquid Chromatography- Mass Spectrometer Manual Joshua Willis, Elizabeth Sattely Department of Chemical Engineering Stanford University November 6, 2014 Abstract This manual will explain the LC/MS, its

More information

Luckylight. 1.9mm (0.8") 8 8 Pure Green Dot Matrix LED Displays Technical Data Sheet. Model No.: KWM-20882XPGB

Luckylight. 1.9mm (0.8) 8 8 Pure Green Dot Matrix LED Displays Technical Data Sheet. Model No.: KWM-20882XPGB .9mm (.8") 8 8 Pure Green Dot Matrix LED Displays Technical Data Sheet Model No.: KWM-88XPGB Spec No: W788C/D Rev No: V. Date: Sep//6 Page: OF 6 Features:.8inch (.mm) Matrix height. Colors: Pure Green.

More information

PulseCounter Neutron & Gamma Spectrometry Software Manual

PulseCounter Neutron & Gamma Spectrometry Software Manual PulseCounter Neutron & Gamma Spectrometry Software Manual MAXIMUS ENERGY CORPORATION Written by Dr. Max I. Fomitchev-Zamilov Web: maximus.energy TABLE OF CONTENTS 0. GENERAL INFORMATION 1. DEFAULT SCREEN

More information

using the Scott A Speakman, Ph.D Center for Materials Science and Engineering at MIT

using the Scott A Speakman, Ph.D Center for Materials Science and Engineering at MIT X-Ray Reflectivity using the PANalytical X Pert Pro MPD Scott A Speakman, Ph.D Center for Materials Science and Engineering at MIT http://prism.mit.edu/xray Modified for configuration used at University

More information

Mechanical aspects, FEA validation and geometry optimization

Mechanical aspects, FEA validation and geometry optimization RF Fingers for the new ESRF-EBS EBS storage ring The ESRF-EBS storage ring features new vacuum chamber profiles with reduced aperture. RF fingers are a key component to ensure good vacuum conditions and

More information

Aurora Grid-Tie Installation Instructions (Model Number: PVI-3.0-OUTD-US-W) Revision 4.1

Aurora Grid-Tie Installation Instructions (Model Number: PVI-3.0-OUTD-US-W) Revision 4.1 Aurora Grid-Tie Installation Instructions (Model Number: PVI-3.0-OUTD-US-W) Revision 4.1 Contents 1) Grid-Tie Installation Block Diagram... 3 2) Installation Steps.... 4 2.1) Initial Setup.... 4 2.1.1)

More information

Features: Descriptions: Applications:

Features: Descriptions: Applications: Features: Package in 8mm tape on 7 diameter reel. Compatible with automatic placement equipment. Compatible with infrared and vapor phase reflow solder process. Mono-color type. The product itself will

More information

Owner's Manual. TOUCH SCREEN CONTROLLER for Air Conditioning Control System. Model BMS-CT5120UL. English

Owner's Manual. TOUCH SCREEN CONTROLLER for Air Conditioning Control System. Model BMS-CT5120UL. English TOUCH SCREEN CONTROLLER for Air Conditioning Control System Model BMS-CT5120UL English Contents 1 Precautions for safety.................................................. 5 2 Main functions........................................................

More information

JEM 3010 Manual (2011 Edition)

JEM 3010 Manual (2011 Edition) JEM 3010 Manual (2011 Edition) Basic Alignment Instructions Check the vacuum (power supply closet) and write on the log book: Gun pressure SIP2 < 2x10-5 Pa (usually 0.8-1.0x10-5 Pa) Column pressure SIP1

More information

University of Minnesota College of Science and Engineering Characterization Facility Ganesha SAXSLAB User manual

University of Minnesota College of Science and Engineering Characterization Facility Ganesha SAXSLAB User manual University of Minnesota College of Science and Engineering Characterization Facility Ganesha SAXSLAB User manual # Section Pg # 1 Planning a SAXS experiment 1 2 Initial Setup 2 3 Sample mounting and loading

More information

JEM 2100 Manual. Operation and Basic Alignment Instructions. Check the vacuum levels (power supply closet) Column (blue scale): < 2.

JEM 2100 Manual. Operation and Basic Alignment Instructions. Check the vacuum levels (power supply closet) Column (blue scale): < 2. 1. Verify the vacuum: JEM 2100 Manual Operation and Basic Alignment Instructions Check the vacuum levels (power supply closet) Column (blue scale): < 2.5x10-5 Pa If vacuum is not good enough, contact somebody

More information

Understanding & Optimising Scanning Electron Microscope Performance

Understanding & Optimising Scanning Electron Microscope Performance W Understanding & Optimising Scanning Electron Microscope Performance hilst the scanning electron microscope with the help of modern computing and sophisticated imaging systems has developed into an instrument

More information

MestReNova Manual for Chem 201/202. October, 2015.

MestReNova Manual for Chem 201/202. October, 2015. 1. Introduction to 1-D NMR Data Processing with MestReNova The MestReNova program can do all of the routine NMR data processing needed for Chem 201 and 202 and will be available through the Reed downloads

More information

Operating Instructions

Operating Instructions Operating Instructions HAEFELY TEST AG KIT Measurement Software Version 1.0 KIT / En Date Version Responsable Changes / Reasons February 2015 1.0 Initial version WARNING Introduction i Before operating

More information

Agilent 83437A Broadband Light Source Agilent 83438A Erbium ASE Source

Agilent 83437A Broadband Light Source Agilent 83438A Erbium ASE Source Agilent 83437A Agilent 83438A Erbium ASE Source Product Overview H Incoherent light sources for single-mode component and sub-system characterization The Technology 2 The Agilent Technologies 83437A (BBLS)

More information

Luckylight Package Pure Green Chip LED. Technical Data Sheet. Part No.: S150PGC-G5-1B

Luckylight Package Pure Green Chip LED. Technical Data Sheet. Part No.: S150PGC-G5-1B 126 Package Pure Green Chip LED Technical Data Sheet Part No.: S15PGC-G5-1B Spec No.: S15 Rev No.: V.3 Date: Jul./1/26 Page: 1 OF 9 Features: Package in 8mm tape on 7 diameter reel. Compatible with automatic

More information

COLOR TFT LCD MONITOR. Manual

COLOR TFT LCD MONITOR. Manual COLOR TFT LCD MONITOR Manual Safety defended: Properly maintains your system to be possible to guarantee its service life and to reduce the damage risk. It should avoid the damp and exceeding temperature

More information

Software Manual Control Panel for Professional Single Booster Units Models: MM3 BW3

Software Manual Control Panel for Professional Single Booster Units Models: MM3 BW3 Software Manual Control Panel for Professional Single Booster Units Models: MM3 BW3 EN Software Manual.. 1-14 1 1. DESCRIPTION 3 2. DISPLAY LAYOUT 4 3. MODES 5 3.1 Power On 5 3.2 Standby 5 3.3 Power off

More information

Part No: 0805-FLWC-DHB

Part No: 0805-FLWC-DHB Features: Package in 8mm tape on 7 diameter reel. Compatible with automatic placement equipment. Compatible with infrared and vapor phase reflow solder process. Mono-color type. The product itself will

More information

MultiQ Digital signage template system for widescreen monitors

MultiQ Digital signage template system for widescreen monitors Technical Note MultiQ Digital signage template system for widescreen monitors This document is intended as a guide for users of the MultiQ Digital Signage Template System for widescreen monitors in landscape

More information

3.0*3.0mm (1.2") 8 8 White Dot Matrix LED Displays Technical Data Sheet. Model No.: KWM-R30881XWB-Y

3.0*3.0mm (1.2) 8 8 White Dot Matrix LED Displays Technical Data Sheet. Model No.: KWM-R30881XWB-Y 3.*3.mm (1.2") 8 8 White Dot Matrix LED Displays Technical Data Sheet Model No.: KWM-R3881XWB-Y Spec No: W1288A/B Rev No: V.2 Date: May/26/29 Page: 1 OF 6 Features: Luckylight 1.2inch (31.7mm) Matrix height.

More information

ABB MEASUREMENT & ANALYTICS. MB3600 The most reliable FT-NIR laboratory analyzer designed for QA/QC

ABB MEASUREMENT & ANALYTICS. MB3600 The most reliable FT-NIR laboratory analyzer designed for QA/QC ABB MEASUREMENT & ANALYTICS MB3600 The most reliable FT-NIR laboratory analyzer designed for QA/QC 2 M B 3 6 0 0 T H E M O S T R E L I A B L E F T- N I R L A B O R ATO R Y A N A LY Z E R D E S I G N E

More information

A-ATF (1) PictureGear Pocket. Operating Instructions Version 2.0

A-ATF (1) PictureGear Pocket. Operating Instructions Version 2.0 A-ATF-200-11(1) PictureGear Pocket Operating Instructions Version 2.0 Introduction PictureGear Pocket What is PictureGear Pocket? What is PictureGear Pocket? PictureGear Pocket is a picture album application

More information

E X P E R I M E N T 1

E X P E R I M E N T 1 E X P E R I M E N T 1 Getting to Know Data Studio Produced by the Physics Staff at Collin College Copyright Collin College Physics Department. All Rights Reserved. University Physics, Exp 1: Getting to

More information

Using the Agilent for Single Crystal Work

Using the Agilent for Single Crystal Work Using the Agilent for Single Crystal Work Generally, the program to access the Agilent, CrysalisPro, will be open. If not, you can start the program using the shortcut on the desktop. There is no password

More information

Luckylight. 1.10mm Height 0805 Package. Warm White Chip LED. Technical Data Sheet. Part No.: S170W-W6-1E

Luckylight. 1.10mm Height 0805 Package. Warm White Chip LED. Technical Data Sheet. Part No.: S170W-W6-1E 1.1mm Height 85 Package Warm White Chip LED Technical Data Sheet Part No.: S17W-W6-1E Spec No.: S17 Rev No.: V.3 Date: Jul./1/26 Page: 1 OF 11 Features: Luckylight Package in 8mm tape on 7 diameter reel.

More information

Noise. CHEM 411L Instrumental Analysis Laboratory Revision 2.0

Noise. CHEM 411L Instrumental Analysis Laboratory Revision 2.0 CHEM 411L Instrumental Analysis Laboratory Revision 2.0 Noise In this laboratory exercise we will determine the Signal-to-Noise (S/N) ratio for an IR spectrum of Air using a Thermo Nicolet Avatar 360 Fourier

More information

Florida State University Thayumanasamy Somasundaram

Florida State University Thayumanasamy Somasundaram 2016 Florida State University 2013-16 Thayumanasamy Somasundaram [PANALYTICAL X-PERT PRO POWDER DATA COLLECTION] A quick start-up procedure for collecting powder x-ray diffraction data from PANalytical

More information

OPERATION MANUAL. TECHNOS Co.,Ltd. Total Reflection X-ray Fluorescence Spectrometer

OPERATION MANUAL. TECHNOS Co.,Ltd. Total Reflection X-ray Fluorescence Spectrometer Total Reflection X-ray Fluorescence Spectrometer OPERATION MANUAL The contents of this system and this manual may change without prior notice. No reproduction, copy or transmission of this publication

More information

SC26 Magnetic Field Cancelling System

SC26 Magnetic Field Cancelling System SPICER CONSULTING SYSTEM SC26 SC26 Magnetic Field Cancelling System Makes the ambient magnetic field OK for electron beam tools in 300 mm wafer fabs Real time, wideband cancelling from DC to > 9 khz fields

More information

Standard Operating Procedure Aria III - Irchel

Standard Operating Procedure Aria III - Irchel 1/6 General note: Only power users having passed the sorter training at the FCF are allowed to operate the Aria III cell sorter. In case of emergency contact the FCF staff under 044-63-50217, -50207 or

More information

PHY221 Lab 1 Discovering Motion: Introduction to Logger Pro and the Motion Detector; Motion with Constant Velocity

PHY221 Lab 1 Discovering Motion: Introduction to Logger Pro and the Motion Detector; Motion with Constant Velocity PHY221 Lab 1 Discovering Motion: Introduction to Logger Pro and the Motion Detector; Motion with Constant Velocity Print Your Name Print Your Partners' Names Instructions August 31, 2016 Before lab, read

More information

FACSAria I Standard Operation Protocol Basic Operation

FACSAria I Standard Operation Protocol Basic Operation FACSAria I Standard Operation Protocol Basic Operation 1. Checking Lasers Status a. Please check the ON / OFF of the lasers. Sufficient time (~30 minutes) need to be given to allow the laser(s) to warm

More information

Vasudevan Agilent 1100 Series HPLC w/ DAD & FLD Detector (nonbuffer

Vasudevan Agilent 1100 Series HPLC w/ DAD & FLD Detector (nonbuffer Vasudevan Agilent 1100 Series HPLC w/ DAD & FLD Detector (nonbuffer solvents) Updated November 14, 2017 Instrument instructions can be found at: http://academic.bowdoin.edu/chemistry/resources/instructions.shtml

More information

Optiflex Interactive Video System

Optiflex Interactive Video System Optiflex Interactive Video System Optiflex Interactive Video System 1 Before You Start...............2 On-site Video Surveillance.......6 Touchscreen Interface Viewing...10 Secure Remote Look-in........16

More information

READ THIS FIRST. Morphologi G3. Quick Start Guide. MAN0412 Issue1.1

READ THIS FIRST. Morphologi G3. Quick Start Guide. MAN0412 Issue1.1 READ THIS FIRST Morphologi G3 Quick Start Guide MAN0412 Issue1.1 Malvern Instruments Ltd. 2008 Malvern Instruments makes every effort to ensure that this document is correct. However, due to Malvern Instruments

More information

McIDAS-V Tutorial Using HYDRA to Interrogate Hyperspectral Data updated September 2015 (software version 1.5)

McIDAS-V Tutorial Using HYDRA to Interrogate Hyperspectral Data updated September 2015 (software version 1.5) McIDAS-V Tutorial Using HYDRA to Interrogate Hyperspectral Data updated September 2015 (software version 1.5) McIDAS-V is a free, open source, visualization and data analysis software package that is the

More information

СВЕТОДИОДЫ BEELED - ТЕХНИЧЕСКОЕ ОПИСАНИЕ

СВЕТОДИОДЫ BEELED - ТЕХНИЧЕСКОЕ ОПИСАНИЕ Features Package in 8mm tape on 7 diameter reel Compatible with automatic placement equipment Compatible with infrared and vapor phase reflow solder process Mono-color type Pb-free Descriptions The 126

More information

FSI Calibration Guide Using CR100 / LightSpace / BoxIO

FSI Calibration Guide Using CR100 / LightSpace / BoxIO FSI Calibration Guide Using CR100 / LightSpace / BoxIO Flanders Scientific, Inc. 6215 Shiloh Crossing Suite G Alpharetta, GA 30005 Phone: +1.678.835.4934 Fax: +1.678.804.1882 E-Mail: Support@FlandersScientific.com

More information

Characterization of Diamond Samples CHESS Run Summer 2013

Characterization of Diamond Samples CHESS Run Summer 2013 Characterization of Diamond Samples CHESS Run Summer 2013 5/22 5/28/2013 Characterization of Diamond Samples CHESS Run Summer 2013 Useful information Goals for this run Conditions of the experiment 1.

More information

Processing data with Mestrelab Mnova

Processing data with Mestrelab Mnova Processing data with Mestrelab Mnova This exercise has three parts: a 1D 1 H spectrum to baseline correct, integrate, peak-pick, and plot; a 2D spectrum to plot with a 1 H spectrum as a projection; and

More information

013-RD

013-RD Engineering Note Topic: Product Affected: JAZ-PX Lamp Module Jaz Date Issued: 08/27/2010 Description The Jaz PX lamp is a pulsed, short arc xenon lamp for UV-VIS applications such as absorbance, bioreflectance,

More information

Configuring the Stack ST8961 VS Module when used in conjunction with a Stack ST81xx series display.

Configuring the Stack ST8961 VS Module when used in conjunction with a Stack ST81xx series display. Configuring the Stack ST8961 VS Module when used in conjunction with a Stack ST81xx series display. Your Stack ST8961 VS module allows you to synchronize, overlay, and record data available on your Stack

More information

CHEMISTRY SEMESTER ONE

CHEMISTRY SEMESTER ONE APPENDIX A USING THE SPECTROMETER FOR AN EMISSION SPECTROSCOPY NANSLO REMOTE WEB-BASED SCIENCE LAB ACTIVITY The following provides information how to use the spectrometer controls for the Emission Spectroscopy

More information

Laser Beam Analyser Laser Diagnos c System. If you can measure it, you can control it!

Laser Beam Analyser Laser Diagnos c System. If you can measure it, you can control it! Laser Beam Analyser Laser Diagnos c System If you can measure it, you can control it! Introduc on to Laser Beam Analysis In industrial -, medical - and laboratory applications using CO 2 and YAG lasers,

More information

Asylum MFP-3D Standard Operating Procedures

Asylum MFP-3D Standard Operating Procedures Asylum MFP-3D Standard Operating Procedures Document Version 1.5, November 2013 The Asylum MFP-3D is an extremely versatile, research-oriented scanning probe microscope on an inverted optical microscope

More information

Asylum MFP-3D Standard Operating Procedures

Asylum MFP-3D Standard Operating Procedures Asylum MFP-3D Standard Operating Procedures Document Version 2, September 2017 The Asylum MFP-3D is an extremely versatile, research-oriented scanning probe microscope on an inverted optical microscope

More information

EMC-Scanner. HR-series

EMC-Scanner. HR-series EMC-Scanner HR-series Seeing high frequencies! Now you can SEE high frequency electromagnetic fields. Visual noise detection The fact that there is no easy way to find the exact location of a radiating

More information

Luckylight Package Warm White Chip LED. Technical Data Sheet. Part No.: S150W-W6-1E

Luckylight Package Warm White Chip LED. Technical Data Sheet. Part No.: S150W-W6-1E 126 Package Warm White Chip LED Technical Data Sheet Part No.: S15W-W6-1E Spec No.: S15 Rev No.: V.3 Date: Jul./1/26 Page: 1 OF 11 Features: Package in 8mm tape on 7 diameter reel. Compatible with automatic

More information