IMSERC NMR MANUAL 05: Manual Operation of Agilent NMR Spectrometers (Chem350 Interface)

Size: px
Start display at page:

Download "IMSERC NMR MANUAL 05: Manual Operation of Agilent NMR Spectrometers (Chem350 Interface)"

Transcription

1 IMSERC NMR MANUAL 05: Manual Operation of Agilent NMR Spectrometers (Chem350 Interface) Last updated: October 12, 2011 by Josh Kurutz THIS PAGE = QUICK START GUIDE 0) At the computer, make sure VNMRJ is running under the account named northrup, and that the instrument is locked & not acquiring data 1) Load a new PROTON experiment by pulling down Experiments Proton from the main menu bar 2) Eject the CDCl 3 ( lock ) standard sample from the Start/Standard panel by clicking Eject 3) Remove the sample from the magnet 4) Carefully load your sample into a spinner using the sample positioner 5) Place your sample on top of the magnet so it floats on the (loud) eject air. On Au400, you will need to push it past the silicone grabbers, holding the spinner, not the sample tube. 6) Insert your sample by clicking Insert on the Start/Standard panel 7) Select your solvent in Start/Standard pulldown menu 8) Retrieve the standard shims by clicking the light blue retrieve default shims button. 9) Lock on the sample with the manual locking procedure in the Start/Lock tab. 10) Shim on the sample using the manual shimming procedure. 11) Make sure your sample is locked and that the lock level is between ~15 and ) Move to the Acquire/Default H1 panel by clicking on the Acquire panel s tab and then the Default panel title. 13) Set the number of scans to 1 and acquire a preliminary 1D 1 H spectrum by clicking the green Acquire button. 14) Evaluate the spectrum examine spectral width, phasing, linewidth, referencing, etc., and make necessary changes. 15) If necessary, acquire more scans; in the Acquire/Default H1 panel, pull down the number of scans pulldown to the desired number of scans a. NB: Keep in mind that S/N improves with the square root of the number of scans; increasing them from 4 to 8 will increase S/N by a factor of ~1.4, but from 4 to 16 will improve it by a factor of 2. 16) Save your spectrum by clicking of the floppy disk save button 17) Continue with further experiments and other samples if you have them 18) When you are done with your last sample, eject it using the Eject button 19) Insert the lock standard using the Insert button in the Start/Standard panel 20) Sign out from the pen-and-paper logbook, noting any instrument problems 21) Process your spectrum further and perhaps print it out using an offline processing computer 1

2 2

3 Manual Operation of Agilent NMR Spectrometers (Chem350 Interface) Setup and Acquisition of a 1D 1 H NMR spectrum Step 0) At the computer, make sure VNMRJ is running, and that the instrument is locked & not acquiring data If necessary, wake up the computer by moving the mouse or typing a key on the keyboard (preferably the spacebar). If you are unfamiliar with the layout of the VNMRJ window, you should take a minute to orient yourself: Figure 1. Orientation to the VnmrJ window First, navigate to the Start/Standard panel, indicated in the lower half of the VNMRJ window. 1

4 Toward the upper left corner of the Start/Standard panel, locate the Operator designation. For Chem350, it should read northrup. To determine whether the instrument is locked, look in the lower-left corner of the VNMRJ panel (near the trash can) and read the number under the word Lock. If the number is blue and above ~15, the system is locked. Figure 2. Operator designation & lock status * PROBLEM? If the instrument is not locked on the lock standard when you walk up to the instrument, either the lock circuit is not engaged, there is no sample in the magnet, or the shims are too poor to support locking. Proceed normally, but when you try to eject the sample that should be in the magnet, make sure that a sample emerges. If no sample comes out, contact staff immediately. Usually when you walk up to the instrument, the VNMRJ program is running. If it is not, you ll just see the underlying Linux desktop for the login account northrup (Figure 3): Figure 3. Essential parts of the linux desktop If VNMRJ is not running, start it by clicking on the VNMRJ icon on the Linux desktop (Figure 3). 2

5 Step 1) Load a new PROTON experiment by pulling down Experiments Proton from the VNMRJ main menu. You can see that the spectrometer offers a large number of experiments, but the Proton experiment is the choice on top. Figure 4. Start a new 1H 1D experiment using fresh parameters by pulling down the VnmrJ Manin Menu's pulldown labeled "Experiments" to the "proton" choice. 3

6 After selecting the Proton experiment, your main display window should show you a diagram of the pulse sequence for the PROTON spectrum. Make sure you are viewing the correct panel in the lower part of the display; click on the Start tab and the Standard panel label. Figure 5. The Proton pulse sequence and parameter interface. Step 2) Eject the sample from the Start/Standard panel by clicking Eject Figure 6. The eject button You should see the lock standard floating on top of the magnet. The lock standard is simply a sample of CDCl 3 in a flamed-sealed tube with a red-painted top. IF NOTHING COMES OUT OF THE MAGNET WHEN YOU CLICK EJECT, GET HELP FROM FACILITY STAFF IMMEDIATELY. 4 Figure 7. The Lock Standard floating atop the magnet

7 Step 3) Remove the lock standard from the magnet. Be careful to lift it all the way out before bringing it toward you. The most common way for people to break samples in the magnet it to snap off the tube below the spinner as they remove the sample from the magnet. Figure 8. Carefully removing the sample from atop the magnet. IF YOU EVER BREAK A SAMPLE IN OR NEAR THE MAGNET, YOU MUST ALERT IMSERC STAFF RIGHT AWAY! IF YOU BREAK A SAMPLE IN THE MAGNET BETWEEN 8 AM AND 10 PM, YOU MUST IMMEDIATELY CONTACT JOSH OR YUYANG IN PERSON OR BY PHONE AND AND PLACE A STOP SIGN FROM (obtained from the front of the lab, near the login computers) ON THE KEYBOARD! BEWTWEEN 10 PM AND 8 AM, YOU MUST PLACE A STOP SIGN ON THE INSTRUMENTʼS KEYBOARD AND JOSH AND YUYANG AND SUBMIT A PROBLEM REPORT WHEN LOGGING OUT OF THE FOM. YOU WILL NOT BE CHARGED OR PENALIZED FOR BREAKING A SAMPLE!!! SAMPLE BREAKAGE DOES NOT NECESSARILY CAUSE MUCH DAMAGE TO THE INSTRUMENT, BUT IF THE NEXT SAMPLE IS INSERTED OVER BROKEN GLASS, IT WILL BREAK THE PROBE, CAUSING THOUSANDS OF DOLLARS WORTH OF DAMAGE AND POTENTIALLY SEVERAL WEEKS OF INSTRUMENT DOWNTIME. 5

8 Step 4) Carefully load your sample into a spinner using the sample positioner Figure 9. Position the sample in the spinner. Step 5) Step 6) Place your sample on top of the magnet so it floats on the (loud) eject air. On Au400, you may need to push it past the silicone grabbers; simply hold the spinner not the probe and push it past the grabbers. Insert your sample by clicking Insert on the Start/Standard panel Figure 10. The Insert button. 6

9 Step 7) Select your solvent in Start/Standard pulldown menu Figure 11. Selecting the sample's solvent. Step 8) Retrieve the standard shims by clicking the light blue read default shims button in the Start/Standard panel. Figure 12. Reading the default shims 7

10 Step 9) Lock on the sample with the manual locking procedure in the Start/Lock panel. First, navigate to the Start/Lock panel. Second, turn the Lock off by clicking the OFF button in the lock panel. Third, begin observing your 2 H lock signal by clicking the Lock Scan button: Figure 13. The Lock panel The signal you see is the first part of the 2 H FID. Any waves you observe arise from the beat frequency difference between the fixed spectrometer frequency and the current resonance frequency of your solvent deuterium atoms. Your goal is to adjust the resonance frequency of your solvent so that it matches the transmitter frequency of the spectrometer s lock channel. You make those adjustments by changing the magnetic field strength, which you accomplish by changing the strength of a homogeneous corrective field called Z0 named because its field is the same across the length of the sample, thus has a zeroth-order dependence on Z. Now adjust the magnetic field by changing the value of Z0. Large-scale changes can be effected using the slider bar, but they usually aren t necessary. Left-click to make Z0 lower, right-click to make it higher, and center-click to toggle through the amounts Z0 will change (±1 ±10 ±100 ±1 ). You generally just pick a direction, change Z0, then observe whether the number of beats in your signal increases or decreases. If it decreases, continue changing Z0 in that direction; if it increases, turn around because you re going the wrong way. Figure 14. Making Z0 adjustments 8

11 You should see changes in your 2 H lock signal: Figure 15. Lock signal when Z0 is incorrectly set If you make adjustments to Z0 and do not observe any signals, you may need to increase your signal strength. Do this first by amplifying the lock signal more by increasing the Lock Gain; accomplish this by right-clicking on the Lock gain button. If your signal is too noisy or you ve reached the maximum gain, try applying a stronger lock RF field to increase your 2 H magnetization; increase your lock RF power by right-clicking the Lock power button. Figure 16. Adjustments for lock gain and power You should ultimately get a signal that looks like the one below, or at least one that looks close. When you ve done this, click the On button on the Lock panel to engage the lock feedback loop. The lock status should then read Regulated. Figure 17. G00d-looking lock signal, regulated. 9

12 Step 10) Shim on the sample using the manual shimming procedure. First, navigate to the Start/Shim panel: Figure 18. The Start/Shim panel. The purpose of shimming is to make the magnetic field across the sample more uniform, but we do not generally assess field homogeneity directly. Instead, we use the lock level to gauge shim quality. In the figure above, the lock level reads 13.5 (on a 0-to-100 scale), and on the screen will probably flicker between 13.3 and To shim we adjust the current running through various non-superconducting shim coils in the magnet. We accomplish this by adjusting the setting of various buttons, primarily Z1 and Z2, using the left-, right-, and center-buttons on the mouse just as was done when adjusting Z0 to establish the lock. Left-click to make each shim lower, rightclick to make it higher, and center-click to toggle through the amounts the shim will change (±1 ±10 ±100 ±1 ). One does not know which way to adjust each shim at first it is an empirical process. For each shim, choose one direction of adjustment and assess whether the lock level rises or falls. If it falls, switch to the other direction. If it rises, continue in that direction until it falls, then back up and stay with the shim value that corresponds to the maximum lock level. Then move on to a different shim. Here we start by adjusting Z1 from to to reach a higher lock level (29.8): Figure 19. Adjusting Z1 improves the lock level. 10

13 Then adjusting Z2 from 1458 to 1428 yielded a further gain in lock level (to 35.2). Figure 20. Adjusting Z2 improves the lock level further. Z2 and Z1 shims are dependent on one another, so after adjusting Z2, go back and adjust Z1. If you want further improvement, you may wish to adjust Z3, X, XZ, Y, and YZ. One can spend all day shimming! To avoid that, you probably want to stop when no large improvements can be made. At this point, you should acquire a spectrum and assess whether the observed lineshape is good enough for your purposes. Step 11) Make sure your sample is locked and that the lock level is between ~15 and 90. If you need to adjust it, first change your lock gain (probably setting it lower), then change your lock power. Below a lock level of 15, you risk losing the lock control. Above that, and you may have trouble with your lock level going above 100 if you improve the shims further. There s no magic lock level number you re aiming for when shimming. Be careful to not increase the lock power too much; you may induce saturation of the 2 H signal, in which the lock level fluctuates or is unreliable. Step 12) Move to the Acquire/Default H1 panel by clicking on the Acquire panel s tab and then the Default panel title. Figure 21. The Acquire/Default H1 panel. Step 13) Set the number of scans to 1 and acquire a preliminary 1D 1 H spectrum by clicking the green Acquire button. Under Acquisition Options, pull down the Number of Scans menu to 1. Then click Acquire. 11 Figure 22. Selecting the number of scans.

14 Step 14) Evaluate the spectrum examine spectral width, phasing, linewidth, referencing, etc Figure 23. Spectrum viewing tools. Step 14a) Examine some of the prominent lines in your spectrum more closely by zooming in on them. Place your left cursor to the left of the peak with your left mouse button : Figure 24. The full 1D 1H spectrum, starting to zoom in. Next, place your right cursor to the right of the peaks by right-clicking : Figure 25. Full 1H 1D spectrum, including the right-cursor for zooming. 12

15 Next, click the zoom in button (magnifying glass with + sign, ): Figure 26. The spectrum, zoomed in upon. If you wish, increase the vertical intensity of the spectrum by center-clicking above a peak. This will bring the point in the spectrum directly underneath your cursor right up to the cursor. You can also use the scroller wheel to adjust intensity. Figure 27. Spectrum portion with vertical scale adjusted. Does that look OK to you? It may be hard to tell unless you ve investigated that spectrum before. It s often better to look at the solvent peak, since that shouldn t change significantly from sample to sample. Step 14b) Start gauging your spectrum quality by examining the linewidth of the residual solvent peak (aka, the solvent peak, but which is really the protiated portion of the deuterated solvent. Remember that 99.9%-deuterated CDCl 3 will still have 0.1% CHCl 3, which you observe and can even use as a chemical shift reference.) This is easier for solvents that present single lines, like CDCl 3, than solvents with multiple lines arising from 2 H- 1 H coupling, like dmso-d6. 13

16 Get a better look at your residual solvent line by zooming in on it; you may need to zoom out to the full spectrum first, which you accomplish using the zoom out button:. Figure 28. Zoom in on the solvent line. Next, you need to place your cursor precisely on the solvent peak. You can accomplish this in two steps: first by placing the (left) cursor somewhere on the peak. Figure 29. Place you cursor on the peak. Then, in the Process/Weighting panel, (1) click find nearest line, then (2) display linewidth. Your linewidth will read out in the display windows. Your solvent line should be below 2.0 Hz for routine data, and ~1.0 or below for publication quality. Figure 30. Find nearest line, display linewidth. 14

17 Step 14c) While your cursor is on the tip of the solvent peak, you may also wish to ensure your spectrum is accurately referenced with respect to frequency in ppm. The most definitive way to reference your spectrum is to include a small amount of tetramethyl silane (TMS) in your sample, and ensure that its 1 H peak resides at ppm; this is the IUPAC-approved method for organic samples. If you have no TMS in your sample, which is very common, you can reference the spectrum to the chemical shift of the residual 1 H signal from the solvent. To confer the correct chemical shift on the solvent peak, leave the cursor on top of the peak, navigate to the Process/Default panel, and under the Reference section locate the Reference cursor to box. In that box, type the correct chemical shift of the solvent; for CHCl 3, you would type 7.24 in that box and hit the return key. This changes the spectrum scale so that the cursor s position reads 7.24 ppm. Figure 31. Manually referencing a spectrum by defining the frequency of a solvent peak. Step 14d) Continue evaluating your spectrum by increasing the vertical scale to see if there are any low-lying peaks that you may wish to observe. If so, you may wish to take another spectrum with a larger number of scans to make these peaks more clear. Step 15) If necessary, acquire more scans; in the Acquire/Default H1 panel, pull down the number of scans pulldown to the desired number of scans a. NB: Keep in mind that S/N improves with the square root of the number of scans; increasing them from 4 to 16 will increase S/N by a factor of ~1.4, but from 4 to 16 will improve it by a factor of 2 15

18 Step 16) Save your spectrum by clicking of the floppy disk save button A dialog window will pop up that will show you the files in the location where the last user saved their spectra. Navigate from here to your own folder by A) Clicking on the folder selector pulldown, which opens the file browser dialog B) Clicking the walkon account s folder C) Clicking (or creating) your folder D) Typing in a legal file name in the File Name box; legal = alphanumerics only, upper/lowercase OK, no spaces or special characters like single or double quotes. Step 17) Continue with further experiments other samples if you have them. Step 18) When you are done with your last sample, eject it using the Eject button in the Start/Standard panel 16

19 Step 19) Insert the lock standard using the Insert button in the Start/Standard panel, and establish lock on it. Step 20) Sign out from the pen-and-paper logbook, noting any instrument problems Step 21) Process your spectrum further and perhaps print it out using an offline processing computer 17

Processing data with Mestrelab Mnova

Processing data with Mestrelab Mnova Processing data with Mestrelab Mnova This exercise has three parts: a 1D 1 H spectrum to baseline correct, integrate, peak-pick, and plot; a 2D spectrum to plot with a 1 H spectrum as a projection; and

More information

PINMRF. Checkout Quiz - Varian Inova-300 Version

PINMRF. Checkout Quiz - Varian Inova-300 Version PINMRF Checkout Quiz - Varian Inova-300 Version Please carefully read every question and select the best answer(s). Some questions may have more than one correct answer. You must select all the correct

More information

Vnmrj Training Guide. Vnmrj overview

Vnmrj Training Guide. Vnmrj overview Vnmrj Training Guide Eugenio Alvarado. Department of Chemistry, University of Michigan. 10/29/10, 05/05/11 This manual describes the most basic operation of our Varian NMR spectrometers with Vnmrj 2.2D

More information

MestReNova Manual for Chem 201/202. October, 2015.

MestReNova Manual for Chem 201/202. October, 2015. 1. Introduction to 1-D NMR Data Processing with MestReNova The MestReNova program can do all of the routine NMR data processing needed for Chem 201 and 202 and will be available through the Reed downloads

More information

Introduction to 1D & 2D NMR using VnmrJ

Introduction to 1D & 2D NMR using VnmrJ Introduction to 1D & 2D NMR using VnmrJ Scott R. Burt September 12, 2011 This document is an introductory tutorial for collecting the most important NMR spectra in organic chemistry: (1D) 1 H, 13 C, NOESY,

More information

MestReNova A quick Guide. Adjust signal intensity Use scroll wheel. Zoomen Z

MestReNova A quick Guide. Adjust signal intensity Use scroll wheel. Zoomen Z MestReNova A quick Guide page 1 MNova is a program to analyze 1D- and 2D NMR data. Start of MNova Start All Programs Chemie NMR MNova The MNova Menu 1. 2. Create expanded regions Adjust signal intensity

More information

Procedure for Acquiring 2D Inverse Detection Spectra HMQC, HMBC and HSQC

Procedure for Acquiring 2D Inverse Detection Spectra HMQC, HMBC and HSQC Procedure for Acquiring 2D Inverse Detection Spectra HMQC, HMBC and HSQC This protocol is for qualified users operating Bruker Avance 400 in IBC only. Dr. Jao STOP accepts no responsibility for actions

More information

Basic 13 C Acquisition and Processing 4

Basic 13 C Acquisition and Processing 4 Chapter Basic 13 C Acquisition and Processing 4 Goto Introduction 4.1 This chapter describes the acquisition and processing of a 13 C spectrum acquired with a simple one-pulse sequence with and without

More information

Practicum 3, Fall 2012

Practicum 3, Fall 2012 A.- F. Miller 2012 T1&T2 Measurement 1 Practicum 3, Fall 2012 Measuring the longitudinal relaxation time: T1. Strychnine, dissolved CDCl3 The T1 is the characteristic time of relaxation of Z- magnetization,

More information

Standard Operating Procedure of nanoir2-s

Standard Operating Procedure of nanoir2-s Standard Operating Procedure of nanoir2-s The Anasys nanoir2 system is the AFM-based nanoscale infrared (IR) spectrometer, which has a patented technique based on photothermal induced resonance (PTIR),

More information

Practicum 3, Fall 2010

Practicum 3, Fall 2010 A. F. Miller 2010 T1 Measurement 1 Practicum 3, Fall 2010 Measuring the longitudinal relaxation time: T1. Strychnine, dissolved CDCl3 The T1 is the characteristic time of relaxation of Z magnetization

More information

Quick Start Bruker Dimension Icon AFM

Quick Start Bruker Dimension Icon AFM Do not remove Quick Start Bruker Dimension Icon AFM March 3, 2015 GLA Contacts Harold Fu (hfu@caltech.edu) Weilai Yu (wyyu@caltech.edu) Bruker Tech Support (AFMSupport@bruker-nano.com 800-873-9750) Watch

More information

Class Notes for Cite While You Write Basics. EndNote Training

Class Notes for Cite While You Write Basics. EndNote Training Class Notes for Cite While You Write Basics EndNote Training EndNote X8 Class Notes for Cite While You Write Basics 1 January 3, 2017 Your EndNote data, both on the desktop and online, can be used in Microsoft

More information

GS122-2L. About the speakers:

GS122-2L. About the speakers: Dan Leighton DL Consulting Andrea Bell GS122-2L A growing number of utilities are adapting Autodesk Utility Design (AUD) as their primary design tool for electrical utilities. You will learn the basics

More information

1 H HOMODECOUPLING EXPERIMENT

1 H HOMODECOUPLING EXPERIMENT 1 H HOMODECOUPLING EXPERIMENT This is a simple and quick means of determining if two resonances are coupled. The HOMODEC experiment is most effective for relatively simple spectra where the couplings are,

More information

Analyzing and Saving a Signal

Analyzing and Saving a Signal Analyzing and Saving a Signal Approximate Time You can complete this exercise in approximately 45 minutes. Background LabVIEW includes a set of Express VIs that help you analyze signals. This chapter teaches

More information

Lab experience 1: Introduction to LabView

Lab experience 1: Introduction to LabView Lab experience 1: Introduction to LabView LabView is software for the real-time acquisition, processing and visualization of measured data. A LabView program is called a Virtual Instrument (VI) because

More information

Dektak Step by Step Instructions:

Dektak Step by Step Instructions: Dektak Step by Step Instructions: Before Using the Equipment SIGN IN THE LOG BOOK Part 1: Setup 1. Turn on the switch at the back of the dektak machine. Then start up the computer. 2. Place the sample

More information

Character Users Guide

Character Users Guide Cha r a c t e r Us e r sgui de Character Users Guide Metric Halo $Revision: 1619 $ Publication date $Date: 2012-02-10 20:41:00-0400 (Friday, 10 Feb 2012) $ Copyright 2011 Metric Halo Table of Contents

More information

Defining and Labeling Circuits and Electrical Phasing in PLS-CADD

Defining and Labeling Circuits and Electrical Phasing in PLS-CADD 610 N. Whitney Way, Suite 160 Madison, WI 53705 Phone: 608.238.2171 Fax: 608.238.9241 Email:info@powline.com URL: http://www.powline.com Defining and Labeling Circuits and Electrical Phasing in PLS-CADD

More information

USB Mini Spectrum Analyzer User s Guide TSA5G35

USB Mini Spectrum Analyzer User s Guide TSA5G35 USB Mini Spectrum Analyzer User s Guide TSA5G35 Triarchy Technologies, Corp. Page 1 of 21 USB Mini Spectrum Analyzer User s Guide Copyright Notice Copyright 2011 Triarchy Technologies, Corp. All rights

More information

NMR TRAINING MANUAL LINUX-BASED SPECTROMETERS. And VnmrJ 2.x TEXAS A&M UNIVERSITY DEPARTMENT OF CHEMISTRY

NMR TRAINING MANUAL LINUX-BASED SPECTROMETERS. And VnmrJ 2.x TEXAS A&M UNIVERSITY DEPARTMENT OF CHEMISTRY NMR TRAINING MANUAL LINUX-BASED SPECTROMETERS And VnmrJ 2.x TEXAS A&M UNIVERSITY DEPARTMENT OF CHEMISTRY November 9, 2009 TABLE OF CONTENTS 1. OVERVIEW 1 1.1 WHAT IS FTNMR? 1 1.2 MULTITASKING 1 1.3 DUAL

More information

Figure 1. MFP-3D software tray

Figure 1. MFP-3D software tray Asylum MFP-3D AFM SOP January 2017 Purpose of this Instrument: To obtain 3D surface topography at sub-nanometer scale resolution, measure contact and friction forces between surfaces in contact, measure

More information

Introduction to EndNote Desktop

Introduction to EndNote Desktop Introduction to EndNote Desktop These notes have been prepared to assist participants in EndNote classes run by the Federation University Library. Examples have been developed using Windows 8.1 (Enterprise)

More information

BME 3512 Biomedical Laboratory Equipment List

BME 3512 Biomedical Laboratory Equipment List BME 3512 Biomedical Laboratory Equipment List Agilent E3630A DC Power Supply Agilent 54622A Digital Oscilloscope Agilent 33120A Function / Waveform Generator APPA 95 Digital Multimeter Component Layout

More information

ME EN 363 ELEMENTARY INSTRUMENTATION Lab: Basic Lab Instruments and Data Acquisition

ME EN 363 ELEMENTARY INSTRUMENTATION Lab: Basic Lab Instruments and Data Acquisition ME EN 363 ELEMENTARY INSTRUMENTATION Lab: Basic Lab Instruments and Data Acquisition INTRODUCTION Many sensors produce continuous voltage signals. In this lab, you will learn about some common methods

More information

CMRRpack2. a collection of pulse sequences from the. University of Minnesota, Center for Magnetic Resonance Research (CMRR) Version: 2.

CMRRpack2. a collection of pulse sequences from the. University of Minnesota, Center for Magnetic Resonance Research (CMRR) Version: 2. CMRRpack2 a collection of pulse sequences from the University of Minnesota, Center for Magnetic Resonance Research (CMRR) Version: 2.63_package This document generated: Mon Dec 16 2013 15:55:46 Contents

More information

1.0 ThermoNicolet Nexus 670 FTIR Spectrometer Instructions

1.0 ThermoNicolet Nexus 670 FTIR Spectrometer Instructions 1.0 ThermoNicolet Nexus 670 FTIR Spectrometer Instructions 1.1 Click on the OMNIC icon to open the software. 1.2 Check that a signal is being measured by opening the Experiment Setup menu under the Collect

More information

Using the Agilent for Single Crystal Work

Using the Agilent for Single Crystal Work Using the Agilent for Single Crystal Work Generally, the program to access the Agilent, CrysalisPro, will be open. If not, you can start the program using the shortcut on the desktop. There is no password

More information

AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode)

AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode) AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode) 1. Log into the Log Usage system on the SMIF web site 2. Open Nanoscope 6.14r1 software by double clicking on the Nanoscope 6.14r1 desktop

More information

Contacts: English Department Office 238 Moreland Media Services 109 Kidder

Contacts: English Department Office 238 Moreland Media Services 109 Kidder Contacts: English Department Office 238 Moreland 7-3244 Media Services 109 Kidder September 2006 Welcome...3 Starting Out...5 Unlocking the Station...5 Touch Panel Operation...5 Projector...6 Selecting

More information

Classical simulations with PJNMR BCMB/CHEM LAB 01/25/12

Classical simulations with PJNMR BCMB/CHEM LAB 01/25/12 Classical simulations with PJNMR BCMB/CHEM LAB 01/25/12 The program PJNMR was written by Paul- Jean Letourneau as a joint project between the PENCE & NANUC projects in Canada. It rotates vectors representing

More information

PHI 5000 VersaProbe TM Operator s Guide

PHI 5000 VersaProbe TM Operator s Guide PHI 5000 VersaProbe TM Operator s Guide Part No. 705921 Rev. A Copyright 2006 ULVAC-PHI, INC. 370 Enzo, Chigasaki, JAPAN The PHI logo ( ) is a registered trademark of ULVAC-PHI, INC. Physical Electronics,

More information

NMR picospin User Guide

NMR picospin User Guide NMR picospin User Guide Software Version 0.9.0 269-298002 Revision A June 2014 2013-2014 Thermo Fisher Scientific Inc. All rights reserved. For U.S. Technical Support, please contact: Unity Lab Services

More information

PHY221 Lab 1 Discovering Motion: Introduction to Logger Pro and the Motion Detector; Motion with Constant Velocity

PHY221 Lab 1 Discovering Motion: Introduction to Logger Pro and the Motion Detector; Motion with Constant Velocity PHY221 Lab 1 Discovering Motion: Introduction to Logger Pro and the Motion Detector; Motion with Constant Velocity Print Your Name Print Your Partners' Names Instructions August 31, 2016 Before lab, read

More information

NMR. picospin. Maintenance Guide

NMR. picospin. Maintenance Guide NMR picospin Maintenance Guide 269-302600 Revision A January 2013 2013 Thermo Fisher Scientific Inc. All rights reserved. For U.S. Technical Support, please contact: Thermo Fisher Scientific 5225 Verona

More information

McIDAS-V Tutorial Using HYDRA to Interrogate Hyperspectral Data updated September 2015 (software version 1.5)

McIDAS-V Tutorial Using HYDRA to Interrogate Hyperspectral Data updated September 2015 (software version 1.5) McIDAS-V Tutorial Using HYDRA to Interrogate Hyperspectral Data updated September 2015 (software version 1.5) McIDAS-V is a free, open source, visualization and data analysis software package that is the

More information

UCI NMR MANUAL DRX400 GN500 CRYO500 AVANCE600

UCI NMR MANUAL DRX400 GN500 CRYO500 AVANCE600 UCI NMR MANUAL DRX400 GN500 CRYO500 AVANCE600 Please do not attempt to print this manual directly from the NMR workstations. To obtain or print your own copy, please either ask the NMR facility director

More information

System Requirements SA0314 Spectrum analyzer:

System Requirements SA0314 Spectrum analyzer: System Requirements SA0314 Spectrum analyzer: System requirements Windows XP, 7, Vista or 8: 1 GHz or faster 32-bit or 64-bit processor 1 GB RAM 10 MB hard disk space \ 1. Getting Started Insert DVD into

More information

Operating Instructions for Scanmark insight 4ES Scanner with QuickScore

Operating Instructions for Scanmark insight 4ES Scanner with QuickScore Operating Instructions for Scanmark insight 4ES Scanner with QuickScore This document will guide you through the process of scanning and grading exams and printing their results. It has been written specifically

More information

MIE 402: WORKSHOP ON DATA ACQUISITION AND SIGNAL PROCESSING Spring 2003

MIE 402: WORKSHOP ON DATA ACQUISITION AND SIGNAL PROCESSING Spring 2003 MIE 402: WORKSHOP ON DATA ACQUISITION AND SIGNAL PROCESSING Spring 2003 OBJECTIVE To become familiar with state-of-the-art digital data acquisition hardware and software. To explore common data acquisition

More information

imso-104 Manual Revised August 5, 2011

imso-104 Manual Revised August 5, 2011 imso-104 Manual Revised August 5, 2011 Section 1 Getting Started SAFETY 1.10 Quickstart Guide 1.20 SAFETY 1.30 Compatibility 1.31 Hardware 1.32 Software Section 2 How it works 2.10 Menus 2.20 Analog Channel

More information

iworx Sample Lab Experiment HM-3: The Electrogastrogram (EGG) and the Growling Stomach

iworx Sample Lab Experiment HM-3: The Electrogastrogram (EGG) and the Growling Stomach Experiment HM-3: The Electrogastrogram (EGG) and the Growling Stomach Background Do you ever wonder why your stomach growls, that funny sound it makes when you are really hungry? Stomach growling is the

More information

How to Build A Table of Authorities in Word * By: Morgan Otway

How to Build A Table of Authorities in Word * By: Morgan Otway How to Build A Table of Authorities in Word * By: Morgan Otway Overview: A Table of Authorities (TOA) is a list of all of the sources cited in a legal document that notes the page numbers on which each

More information

Reference. TDS7000 Series Digital Phosphor Oscilloscopes

Reference. TDS7000 Series Digital Phosphor Oscilloscopes Reference TDS7000 Series Digital Phosphor Oscilloscopes 07-070-00 0707000 To Use the Front Panel You can use the dedicated, front-panel knobs and buttons to do the most common operations. Turn INTENSITY

More information

EndNote for Mac. User Guide. UTS Library University of Technology Sydney UTS CRICOS PROVIDER CODE 00099F

EndNote for Mac. User Guide. UTS Library University of Technology Sydney UTS CRICOS PROVIDER CODE 00099F UTS CRICOS PROVIDER CODE 00099F EndNote for Mac User Guide UTS Library University of Technology Sydney EndNote for Mac Table of Contents Part 1 Installing EndNote... 3 Before you begin - Update your mac

More information

3 CHOPS - LIP SYNCHING

3 CHOPS - LIP SYNCHING 3 CHOPS - LIP SYNCHING In this Lesson you will use CHOPs to match lip movements with an existing audio file. You will use blend operations in the SOP editor to create the different facial shapes when saying

More information

Standard Operating Procedure II: EDS (Bruker Flat-Quad)

Standard Operating Procedure II: EDS (Bruker Flat-Quad) Standard Operating Procedure II: EDS (Bruker Flat-Quad) ywcmatsci.yale.edu ESC II, Room A119F 810 West Campus Drive West Haven, CT 06516 Version 1.1, October 2018 1 > FOLLOW the SOP strictly to keep the

More information

Linrad On-Screen Controls K1JT

Linrad On-Screen Controls K1JT Linrad On-Screen Controls K1JT Main (Startup) Menu A = Weak signal CW B = Normal CW C = Meteor scatter CW D = SSB E = FM F = AM G = QRSS CW H = TX test I = Soundcard test mode J = Analog hardware tune

More information

FACSAria I Standard Operation Protocol Basic Operation

FACSAria I Standard Operation Protocol Basic Operation FACSAria I Standard Operation Protocol Basic Operation 1. Checking Lasers Status a. Please check the ON / OFF of the lasers. Sufficient time (~30 minutes) need to be given to allow the laser(s) to warm

More information

Import and quantification of a micro titer plate image

Import and quantification of a micro titer plate image BioNumerics Tutorial: Import and quantification of a micro titer plate image 1 Aims BioNumerics can import character type data from TIFF images. This happens by quantification of the color intensity and/or

More information

Vasudevan Agilent 1100 Series HPLC w/ DAD & FLD Detector (nonbuffer

Vasudevan Agilent 1100 Series HPLC w/ DAD & FLD Detector (nonbuffer Vasudevan Agilent 1100 Series HPLC w/ DAD & FLD Detector (nonbuffer solvents) Updated November 14, 2017 Instrument instructions can be found at: http://academic.bowdoin.edu/chemistry/resources/instructions.shtml

More information

Cisco Spectrum Expert Software Overview

Cisco Spectrum Expert Software Overview CHAPTER 5 If your computer has an 802.11 interface, it should be enabled in order to detect Wi-Fi devices. If you are connected to an AP or ad-hoc network through the 802.11 interface, you will occasionally

More information

Topic: Instructional David G. Thomas December 23, 2015

Topic: Instructional David G. Thomas December 23, 2015 Procedure to Setup a 3ɸ Linear Motor This is a guide to configure a 3ɸ linear motor using either analog or digital encoder feedback with an Elmo Gold Line drive. Topic: Instructional David G. Thomas December

More information

Game Show Physiology

Game Show Physiology iworx Physiology Lab Experiment Experiment HN-8: Game Show Physiology Game Show Physiology Note: The lab presented here is intended for evaluation purposes only. iworx users should refer to the User Area

More information

Experiment HM-8: The Electrogastrogram (EGG) and the Growling Stomach

Experiment HM-8: The Electrogastrogram (EGG) and the Growling Stomach Experiment HM-8: The Electrogastrogram (EGG) and the Growling Stomach Background Do you ever wonder why your stomach growls, that funny sound it makes when you are really hungry? Stomach growling is the

More information

Introduction to EndNote Online

Introduction to EndNote Online Introduction to EndNote Online Creating an EndNote Online account Go to EndNote Online. Click on the Access EndNote Online button and, if prompted, enter your Warwick username and password to confirm you

More information

Please feel free to download the Demo application software from analogarts.com to help you follow this seminar.

Please feel free to download the Demo application software from analogarts.com to help you follow this seminar. Hello, welcome to Analog Arts spectrum analyzer tutorial. Please feel free to download the Demo application software from analogarts.com to help you follow this seminar. For this presentation, we use a

More information

SPM Training Manual Veeco Bioscope II NIFTI-NUANCE Center Northwestern University

SPM Training Manual Veeco Bioscope II NIFTI-NUANCE Center Northwestern University SPM Training Manual Veeco Bioscope II NIFTI-NUANCE Center Northwestern University Introduction: Scanning Probe Microscopy (SPM) is a general term referring to surface characterization techniques that utilize

More information

FIB Operating Procedure. Effective Date: 08/14/2012 Author(s): Jiong Hua Phone:

FIB Operating Procedure. Effective Date: 08/14/2012 Author(s): Jiong Hua Phone: FIB Operating Procedure Effective Date: 08/14/2012 Author(s): Jiong Hua Phone: 402-472-3773 Email: jhua2@unl.edu 1 1 Introduction 1.1 Key Words Focused Ion Beam (FIB), FEI Strata 201, Ion milling 1.2 Purpose

More information

Swinburne University of Technology

Swinburne University of Technology Swinburne University of Technology EndNote X9 for Mac Swinburne Library EndNote resources page: http://www.swinburne.edu.au/library/referencing/references-endnote/endnote/ These notes include excerpts

More information

TL-2900 AMMONIA & NITRATE ANALYZER DUAL CHANNEL

TL-2900 AMMONIA & NITRATE ANALYZER DUAL CHANNEL TL-2900 AMMONIA & NITRATE ANALYZER DUAL CHANNEL DATA ACQUISITION SYSTEM V.15.4 INSTRUCTION MANUAL Timberline Instruments, LLC 1880 S. Flatiron Ct., Unit I Boulder, Colorado 80301 Ph: (303) 440-8779 Fx:

More information

The BAT WAVE ANALYZER project

The BAT WAVE ANALYZER project The BAT WAVE ANALYZER project Conditions of Use The Bat Wave Analyzer program is free for personal use and can be redistributed provided it is not changed in any way, and no fee is requested. The Bat Wave

More information

Amazon Account 3/14 TM 1

Amazon Account 3/14 TM 1 Overdrive ebooks for Kindle Readers: Kindle, Kindle Touch, Kindle Keyboard, Kindle Paperwhite provides a collection of ebooks and digital audiobooks through Overdrive: a distributor of digital books to

More information

Universal Mode. Survey Settings. The Universal Toolbar. First Run - Initial Configuration via the Setup Button

Universal Mode. Survey Settings. The Universal Toolbar. First Run - Initial Configuration via the Setup Button Universal Mode The Universal Toolbar The Universal Toolbar is the unified (and preferred) method of operating Revolution. From within the Universal Toobar, all functionality of the other 3 modes of Revolution

More information

Waters GCT GCMS Training Manual (Nominal Mass GCMS version) 02/06/2012 S.V.

Waters GCT GCMS Training Manual (Nominal Mass GCMS version) 02/06/2012 S.V. 1234 Hach Hall 515-294-5805 www.cif.iastate.edu Waters GCT GCMS Training Manual (Nominal Mass GCMS version) 02/06/2012 S.V. Location: Contact: 1238 Hach Hall Steve Veysey, 1234 Hach Hall; Kamel Harrata,

More information

Kratos AXIS Ultra DLD X-ray Photoelectron Spectrometer Instructions

Kratos AXIS Ultra DLD X-ray Photoelectron Spectrometer Instructions Kratos AXIS Ultra DLD X-ray Photoelectron Spectrometer Instructions Note: Enter your complete name, CMRF user code, and date into the record book. You must have previously filled out a project information

More information

Getting Started with the LabVIEW Sound and Vibration Toolkit

Getting Started with the LabVIEW Sound and Vibration Toolkit 1 Getting Started with the LabVIEW Sound and Vibration Toolkit This tutorial is designed to introduce you to some of the sound and vibration analysis capabilities in the industry-leading software tool

More information

USB Mini Spectrum Analyzer User Manual PC program TSA For TSA5G35 TSA4G1 TSA6G1 TSA12G5

USB Mini Spectrum Analyzer User Manual PC program TSA For TSA5G35 TSA4G1 TSA6G1 TSA12G5 USB Mini Spectrum Analyzer User Manual PC program TSA For TSA5G35 TSA4G1 TSA6G1 TSA12G5 Triarchy Technologies, Corp. Page 1 of 17 USB Mini Spectrum Analyzer User Manual Copyright Notice Copyright 2013

More information

Exercise #1: Create and Revise a Smart Group

Exercise #1: Create and Revise a Smart Group EndNote X7 Advanced: Hands-On for CDPH Sheldon Margen Public Health Library, UC Berkeley Exercise #1: Create and Revise a Smart Group Objective: Learn how to create and revise Smart Groups to automate

More information

TT AFM LongBeach Procedures and Protocols V2.1

TT AFM LongBeach Procedures and Protocols V2.1 TT AFM LongBeach Procedures and Protocols V2.1 1. Startup Procedure 1. Turn on PC: Allow to boot to Windows. Turn on monitor. Password is afm 2. Turn on second PC that controls the video camera. 3. Turn

More information

SR-D8-M, SR-D8-S. (Ver ) SOFTWARE INSTRUCTIONS

SR-D8-M, SR-D8-S. (Ver ) SOFTWARE INSTRUCTIONS SOFTWARE INSTRUCTIONS active l ine array speak er SYStems SR-D8-M, SR-D8-S (Ver. 1.1.1) Thank you for purchasing TOA's Active Line Array Speaker Systems. Please carefully follow the instructions in this

More information

Fall 2014 Chem 636 Lab #4

Fall 2014 Chem 636 Lab #4 latest update: 140922 (cgf) Fall 2014 Chem 636 Lab #4 Assignment due at beginning of lab, week 5 (Sept 30 Oct 2). Use Athena (AC-300) for section I, and Persephone (AVANCE-500) for Sections II and III.

More information

EndNote Tutorial Handout Table of Contents

EndNote Tutorial Handout Table of Contents EndNote Tutorial Handout Table of Contents What is EndNote?... 2 Getting Started... 2 Create a New Library... 2 EndNote Interface... 3 Set the Reference Style... 4 Adding References to Your Library...

More information

Reason Overview3. Reason Overview

Reason Overview3. Reason Overview Reason Overview3 In this chapter we ll take a quick look around the Reason interface and get an overview of what working in Reason will be like. If Reason is your first music studio, chances are the interface

More information

***Please be aware that there are some issues of compatibility between all current versions of EndNote and macos Sierra (version 10.12).

***Please be aware that there are some issues of compatibility between all current versions of EndNote and macos Sierra (version 10.12). EndNote for Mac Note of caution: ***Please be aware that there are some issues of compatibility between all current versions of EndNote and macos Sierra (version 10.12). *** Sierra interferes with EndNote's

More information

Effective Test Procedures for Installing and Maintaining RF Transmitter Sites

Effective Test Procedures for Installing and Maintaining RF Transmitter Sites Product: Hand Held Spectrum Analyzer R&S FSH3 Effective Test Procedures for Installing and Maintaining RF Transmitter Sites This application note describes an effective method for generating test setups,

More information

MultiQ Digital signage template system for widescreen monitors

MultiQ Digital signage template system for widescreen monitors Technical Note MultiQ Digital signage template system for widescreen monitors This document is intended as a guide for users of the MultiQ Digital Signage Template System for widescreen monitors in landscape

More information

Application Note AN-708 Vibration Measurements with the Vibration Synchronization Module

Application Note AN-708 Vibration Measurements with the Vibration Synchronization Module Application Note AN-708 Vibration Measurements with the Vibration Synchronization Module Introduction The vibration module allows complete analysis of cyclical events using low-speed cameras. This is accomplished

More information

cyclenoe 1D NOE Difference Experiment

cyclenoe 1D NOE Difference Experiment cyclenoe 1D NOE Difference Experiment NOTE: If you are interested in qualitative or long-range NOE s, you should use the NOESY1D experiment (see NOESY1D handout for experimental details). However, you

More information

Getting started with EndNote X7

Getting started with EndNote X7 IT Training Getting started with EndNote X7 Sally Swaine, IT Training IT Services Version 3.3 Scope Learning outcomes Develop a better understanding of how EndNote works as a tool. Understand how EndNote

More information

Ultra 4K Tool Box. Version Release Note

Ultra 4K Tool Box. Version Release Note Ultra 4K Tool Box Version 2.1.43.0 Release Note This document summarises the enhancements introduced in Version 2.1 of the software for the Omnitek Ultra 4K Tool Box and related products. It also details

More information

DektakXT Profilometer. Standard Operating Procedure

DektakXT Profilometer. Standard Operating Procedure DektakXT Profilometer Standard Operating Procedure 1. System startup and sample loading: a. Ensure system is powered on by looking at the controller to the left of the computer.(it is an online software,

More information

Gustavus Adolphus College. Some Scientific Software of Interest

Gustavus Adolphus College. Some Scientific Software of Interest CHE 372 Gustavus Adolphus College Some Scientific Software of Interest A. Literature Databases There are several literature databases commonly used to conduct scientific literature reviews. The two most

More information

Optiflex Interactive Video System

Optiflex Interactive Video System Optiflex Interactive Video System Optiflex Interactive Video System 1 Before You Start...............2 On-site Video Surveillance.......6 Touchscreen Interface Viewing...10 Secure Remote Look-in........16

More information

USB Mini Spectrum Analyzer User Manual TSA Program for PC TSA4G1 TSA6G1 TSA8G1

USB Mini Spectrum Analyzer User Manual TSA Program for PC TSA4G1 TSA6G1 TSA8G1 USB Mini Spectrum Analyzer User Manual TSA Program for PC TSA4G1 TSA6G1 TSA8G1 Triarchy Technologies Corp. Page 1 of 17 USB Mini Spectrum Analyzer User Manual Copyright Notice Copyright 2013 Triarchy Technologies,

More information

Word Tutorial 2: Editing and Formatting a Document

Word Tutorial 2: Editing and Formatting a Document Word Tutorial 2: Editing and Formatting a Document Microsoft Office 2010 Objectives Create bulleted and numbered lists Move text within a document Find and replace text Check spelling and grammar Format

More information

SmartScore Quick Tour

SmartScore Quick Tour SmartScore Quick Tour Installation With the packaged CD, you will be able to install SmartScore an unlimited number of times onto your computer. Application files should not be copied to other computers.

More information

STANDARD OPERATING PROCEDURE: ASYLUM MFP-3D AFM

STANDARD OPERATING PROCEDURE: ASYLUM MFP-3D AFM STANDARD OPERATING PROCEDURE: ASYLUM MFP-3D AFM Purpose of this Instrument: To obtain 3D surface topography at sub-nanometer scale resolution, measure contact and friction forces between surfaces in contact,

More information

Library ebooks and Your Sony ereader

Library ebooks and Your Sony ereader Library ebooks and Your Sony ereader When using your Sony ereader for the first time to obtain library ebooks and audiobooks, download the Overdrive Media Console, Adobe Digital Editions, and the Sony

More information

The Kaffeine Handbook. Jürgen Kofler Christophe Thommeret Mauro Carvalho Chehab

The Kaffeine Handbook. Jürgen Kofler Christophe Thommeret Mauro Carvalho Chehab Jürgen Kofler Christophe Thommeret Mauro Carvalho Chehab 2 Contents 1 Kaffeine Player 5 1.1 The Start Window...................................... 5 1.2 Play a File..........................................

More information

EndNote for Mac. EndNote for PC. User Guide. UTS Library University of Technology Sydney UTS CRICOS PROVIDER CODE 00099F

EndNote for Mac. EndNote for PC. User Guide. UTS Library University of Technology Sydney UTS CRICOS PROVIDER CODE 00099F UTS CRICOS PROVIDER CODE 00099F EndNote for Mac EndNote for PC User Guide UTS Library University of Technology Sydney EndNote for PC Table of Contents Part 1 Installing EndNote... 3 What is EndNote?...4

More information

Procedures for conducting User QA on the scanner

Procedures for conducting User QA on the scanner Procedures for conducting User QA on the scanner Sample setup: The User QA phantom is clearly labeled and is stored on one of the shelves to the side of the magnet. Note proper orientation of the bottle,

More information

Precision DeEsser Users Guide

Precision DeEsser Users Guide Precision DeEsser Users Guide Metric Halo $Revision: 1670 $ Publication date $Date: 2012-05-01 13:50:00-0400 (Tue, 01 May 2012) $ Copyright 2012 Metric Halo. MH Production Bundle, ChannelStrip 3, Character,

More information

ExtIO Plugin User Guide

ExtIO Plugin User Guide Overview The SDRplay Radio combines together the Mirics flexible tuner front-end and USB Bridge to produce a SDR platform capable of being used for a wide range of worldwide radio and TV standards. This

More information

SEM- EDS Instruction Manual

SEM- EDS Instruction Manual SEM- EDS Instruction Manual Double-click on the Spirit icon ( ) on the desktop to start the software program. I. X-ray Functions Access the basic X-ray acquisition, display and analysis functions through

More information

SOUND LABORATORY LING123: SOUND AND COMMUNICATION

SOUND LABORATORY LING123: SOUND AND COMMUNICATION SOUND LABORATORY LING123: SOUND AND COMMUNICATION In this assignment you will be using the Praat program to analyze two recordings: (1) the advertisement call of the North American bullfrog; and (2) the

More information

Using EndNote Online: health

Using EndNote Online: health Using EndNote Online: health 1. Registration & access p.1 2. Collecting & organising references p.2 Transferring references from Web of Science (p.2), PubMed (p.5) & others 3. Citing & referencing while

More information

Manual for TV software. TT-Viewer version Figure: TT-budget S2-3200

Manual for TV software. TT-Viewer version Figure: TT-budget S2-3200 Manual for TV software TT-Viewer version 2.7.0 Figure: TT-budget S2-3200 Index Manual TT-Viewer 3 1. Starting TT-Viewer software 3 2. General settings 5 3. Assignment of hardware 6 3.1 Unicable 7 4. Renderer

More information

Operating Guide. ViewClix offers a revolutionary experience for seniors and their families and friends.

Operating Guide. ViewClix offers a revolutionary experience for seniors and their families and friends. ViewClix Mini TM Operating Guide ViewClix offers a revolutionary experience for seniors and their families and friends. To make using ViewClix an easy and fun experience for you and your loved ones, we

More information