Multi echo Multi slice (MEMS) High Performance fmri at CFMRI... 1

Size: px
Start display at page:

Download "Multi echo Multi slice (MEMS) High Performance fmri at CFMRI... 1"

Transcription

1 Multi echo Multi slice (MEMS) High Performance fmri at CFMRI Table of Contents Multi echo Multi slice (MEMS) High Performance fmri at CFMRI... 1 Introduction... 2 MEMS Protocols... 4 Run MEMS protocol... 5 Set up and prepare... 5 Scan MEMS protocol... 5 End exam and transfer MEMS Data... 8 Reconstruct and pre process MEMS data... 8 Data Requirements... 9 Running the pipeline... 9 Error Logging... 9 Output Appendix I Appendix II: MEMS Pipeline Output files Appendix III: MEICA (More information will be added as they become available)

2 Introduction A major need in the analysis of Blood Oxygen Level Dependent (BOLD) functional MRI (fmri) data is the ability to distinguish BOLD related signals from non BOLD related signals, such as those due to physiological fluctuations or head motion. Previous studies 1 have shown that the amplitude of the BOLD signal variations has a linear dependence on echo time (TE), whereas the amplitude of the non BOLD signal variations does not (Figure 1). A B Non BOLD signal (noise) TE1=15.5 TE2=36.7 TE3=57.9 % Signal Change % Signal Change C Echo Time (TE) BOLD signal Echo Time (TE) Figure 1. A: fmri data acquired at 3 different TEs. B: noise signal amplitude does not depend on TE. C: BOLD signal amplitude has a linear dependence on TE. (Figure credit: Valur Olaffson) 1 Peltier SJ, Noll DC, T2* dependence of low frequency functional connectivity, Neruoimage, 2002; 16(4) 2

3 Kundu et al 2 3 extended this observation to Independent Component Analysis (ICA) of multi echo fmri data, where ICA components that display TE dependencies are considered BOLD signals; ICA components that do not display TE dependencies are considered noise and thus removed from the fmri data. This fmri denoising method, known as multiecho ICA (ME ICA), has been shown to robustly detect motion and other non BOLD related signals, and to significantly improve signal to noise ratio and functional connectivity estimates (Figure 2). Standard denoising MEICA denoising A potential disadvantage of multi echo fmri is the time cost associated with acquiring multiple echoes. However various acceleration techniques are available to speed up fmri data acquisition, such as the simultaneous Figure 2. Comparison of the default mode network maps after applying standard denoising method (i.e. motion correction) and MEICA method (Figure credit: Kundu et al; 2013). multi slice (SMS) technique where multiple ( N>1) slices are excited at a time providing a factor of N reduction in scan time, and parallel imaging methods (such as GRAPPA or SENSE) that shorten the time required for the readout of each echo. At the Center for fmri (CFMRI) we provide an accelerated multi echo multi slice (MEMS) protocol that is capable of acquiring full brain multi echo fmri data with a TR of ~ 1sec. The use of multi echo data with shorter TRs has been shown to improve the ability to detect functional networks. Note that the MEMS protocol requires the use of the Nova Medical 32 channel head coil. 2 Kundu P et al, Differentiating BOLD and non BOLD signals in fmri time series using multi echo EPI. Neuroimage, 2012;60(3) 3 Kundu P et al, Integrated strategy for improving functional connectivity mapping using multi echo fmri, PNAS, 2013; 110(40) 3

4 MEMS Protocols The MEMS protocol can be found on both the 3T West and 3T East scanners under ADULT >HEAD >MEMS. CFMRI offers the protocol at two different resolutions, 3.75mm or 3mm isotropic. Table 1 below summarizes some of the important fmri parameters of the protocols. Table 1. MEMS protocols Protocol 1 (3.75mm 3 ) Protocol 2 (3mm 3 ) Resolution 3.75x3.75x3.75mm 3 3x3x3mm 3 Slice orientation Sagittal (other slice orientations are in development) # slices (must be multiple of 3) FOV 24cm 21.6cm Matrix 64x64 72x72 Multi slice Factor N N =3 N=3 Acceleration Factor R R=2 R=2 # of Echoes 3 3 TR (may increase if # slices increases) 1sec 1.1sec TE [ ]msec [ ]msec Bandwidth ±125KHz ±125Khz Flip angle (adjust with TR) Scan time Adjustable by User (typically 5 10 mins) The MEMS protocol contains the following scans: 1. Localizer (~ 15sec) 2. ASSET calibration (~6 sec) 3. FSPGR (~10mins, adjustable) : T1 weighted high resolution scan 4. HOS(~30 sec): High order shim 5. fm_grass(1 2mins): fieldmap 6. MEMS calibration 1(~18 sec): Calibration 1 for MEMS recon 7. MEMS calibration 2(~18 sec): Calibration 2 for MEMS recon 8. MEMS fmri(~10mins, adjustable): functional or resting fmri scans (can have multiple scans) 9. MEMS_topup_fwd(~6 sec): forward TOPUP scan 4

5 10. MEMS_topup_rvs(~6 sec): reverse TOPUP scan The fm_grass scan and the calibration 1 and 2 scans are required for reconstructing the fmri scan. If there is more than one fmri scan, the graphical prescriptions of all fmri scans have to be matched exactly (use copyrx on the scanner). If the prescription changes, a new set of the fm_grass, calibration 1 and 2, and the topup fwd and rvs scans matching the new prescription will be required. The topup_fwd and topup_rvs scan pair are used to measure a fieldmap which can be applied to the fmri images for correcting geometric distortions. They are typically scanned immediately before or after the fmri scans. In the case of a scan session containing several fmri scans in a row where there is concern about subject motion during the session, users may acquire one topup scan pair before the fmri scans and another pair at the end of the session. The first pair can be used for correcting the fmri data before the motion occurs, and the second pair for correcting the fmri data after the motion. Run MEMS protocol Set up and prepare 1. Place the 32channel coil on the scanner table and plug it in. Make sure the coil is recognized by the scanner by checking the information on the iroc monitor on top of the scanner. 2. Set up peripheral equipment such as the projector, screen, and stimulus laptop etc if needed. 3. Set up the subject on scanner patient table. 4. Setup physiological monitoring if needed. 5. On the console computer, click the downward arrow on the Tools icon. In the drop down menu, select Command Window. In the command window type RTctrl start to start realtime. Drag this window to the lower right corner of the screen so it is easily accessible and not blocking the scan area of the screen. (NOTE: this step must be done before Start Exam in step 6) 6. Register the subject and Start Exam. Scan MEMS protocol 1. Localizer Save Rx and Scan 2. Asset Cal Setup, Prescribe Rx, Save Rx, and Scan 3. FSPGR T1 Setup, Prescribe Rx, Save Rx, and Scan. IMPORTANT: While waiting for the FSPGR to finish, prescribe and save the fm_grass scan below the HOS scan. This step MUST be done before running the next HOS scan. 4. HOS (high order shim) Setup and Save Rx (no need to prescribe slices). Click OK in the popup window saying Running high order shim for clinical Protocol: fm_grass. Then click Scan. 5

6 fm_grass Once the scan finishes, an HOS window will appear. In the HOS window, adjust the size and location of the ROIs to enclose the whole brain (see figure below); Click Calculate Shim, then Done. Make a note of the current RMS and predicted RMS values reported in the lower left of the window. The difference between the two values is indicative of HOS efficacy. Run the HOS a second time by clicking Scan and choose Same Series in the popup window. Once the HOS window appears, click Calculate Shim (do not modify the ROIs). Verify that the new current and predicted RMS values are converging and consistent with the previously predicted value (See figure below). If yes, click Done to finish. The HOS scan can be repeated a third time if needed. We have typically seen RMS value convergence in two iterations. In case the RMS value does not converge in three iterations, please click Quit to skip HOS. Please alert the CFMRI staff of the HOS malfunction, or report the problem using the online webschedule program at your earliest convenience. 5. fm_grass Scan 6

7 6. mems cal 1(see NOTE below) Copy Rx from fm_grass, save Rx, download, and scan. 7. mems cal 2 (see NOTE below) Copy Rx from fm_grass, save Rx, download, and scan. 8. mems_rest_rvs (see NOTE below) Copy Rx from fm_grass, save Rx, download, and scan. 9. mems_topup_fwd (see NOTE below) Copy Rx from fm_grass, save Rx, download and scan. 10. mems_topup_rvs (see NOTE below) Copy Rx from fm_grass, save Rx, download and scan. NOTE: Each of the mems scans (scan 6 10) needs to be downloaded prior to the respective scan. The download step ensures that RDS client is started to receive the acquired MRI data. If the RDS client is not ON, no MRI data will be saved. Below are some useful tools: o o Type ck in the command window to check if the RDS client is ON. This should be checked after each download to make sure RDS client is ON before Scan. Type memslist in the command window to list the raw data files. Anytime during an hcp scan, use memslist to check if data is being saved. If an MEMS scan has to be stopped before it finishes, for example when subject activates the emergency squeeze ball, please do the following: o o Type kk in the command window to kill the RDS client then press Stop Scan button. After the emergency situation or errors are cleared, copy & paste the scan, download and Scan. If the Stop Scan button is pressed before the RDS client is killed, a TPS reset must be performed before the scan can continue. A TPS reset usually takes 2 3 minutes. After the TPS reset, copy & paste the scan, download and Scan. Due to limitations with the RDS server software, after a mems scan finishes, the status of the scan shows Action Failed. You can ignore this status. Additionally, as soon as the next scan is downloaded, the previous mems scan is pushed downward in the scan list as if it has not been scanned. Please pay attention to which scans have already been done and which have not. Appendix I lists all command line tools available for use with the MEMS protocols. 7

8 End exam and transfer MEMS Data On the computer console, click on End > End Exam In the command window, type RTctrl stop to stop realtime. In the command window, type memscopy to transfer mems data (P files). Usage: memscopy s server r raid# d studyfolder login example: memscopy s fmrimems r raid16 d myhcpfolder mylogin Transfer Dicom data using gecopy. Usage: gecopy s server r raid# d studyfolder login example: gecopy s fmrimems r raid16 d myhcpfolder mylogin Transfer physio data using physiocopy as needed. Usage: physiocopy s server d studyfolder starttime endtime login example: physiocopy s fmrimems d myhcpfolder 14:00 15:00 mylogin After data transfer completes, close all command windows and clean up the scanner suite. Reconstruct and pre process MEMS data We provide a Pipeline for reconstructing the MEMS data. The pipeline consists of four functional modules: 1. Data validation: checks if all required data files are present. 2. Data reconstruction: reconstructs the fmri images. 3. Quality control: calculates temporal SNR and estimates T2* maps. 4. Pre processing (optional): performs motion correction, registration and distortion correction (see the diagram below). Due to the use of in plane acceleration in the protocol, the in plane image distortions are typically small, and so this step may be skipped at the discretion of the user. fmri data Echo 1 fmri data Echo 2 fmri data Echo 3 Motion correction (3dvolreg) Apply the transformation matrix from Echo 2 (3dAllineate) Motion correction and Registration to one of the TOPUP pair that has the same phase encoding direction ( align_epi_anat.py) Motion correction (3dvolreg) Apply the transformation matrix from Echo 2 (3dAllineate) Apply distortion correction (TOPUP) Apply distortion correction ( TOPUP) Apply distortion correction (TOPUP) 8

9 5. post processing using MEICA (optional, see Appendix III) : Performs MEICA denoising. System and Data Requirements Add the following path to your ~/matlab/startup.m file. Create the file if the file does not exist. path(path,'/apps/matlabcode/spiralfmap2'); path(path,'/apps/matlabcode/domems'); path(path,'/apps/afni_matlab/matlab'); path(path,'/apps/matlabcode/fmritools'); All data must be located in the same folder. Use unix command ls to check if all the data are present (see example below). The data required include dicom files of the fmap scan and the T1 structural scan (stored in exam/subject directory), and pfiles from the calibration scan 1 and 2, one or more fmri scans, and two TOPUP scans fmrimems2.ucsd.edu:mems_data>> ls e207 (contains the s folders for the fmap_grass scan and the T1 structural scan) P12288_spep_mems_110829_ (Pfile for calibration scan 1) P12800_spep_mems_110829_ (Pfile for calibration scan2 ) P13312_spep_mems_110829_ (Pfile for fmri scan) P13824_spep_mems_110829_ (Pfile for topup foward) P14336_spep_mems_110829_ (Pfile for topup reverse) Running the pipeline 1. Login into your assigned server (either fmrimems.ucsd.edu or fmrimems2.ucsd.edu). 2. Change to the directory where the data are located. 3. Type at the Linux prompt: domems.py [ topup] [ meica] Options: topup: to apply EPI distortion correction meica: to apply MultiEcho ICA (see Appendix III) Your job will be queued. Type qstat to check queue status, or qdel followed by the job number to remove from the queue. Error Logging A log file is automatically saved in the current data directory under the log folder. Automatic notifications will also be sent upon job success or failure to the address registered with the server account (usually the PI s address). 9

10 Output All output is saved under the processed folder. The pre processed fmri data (motion and distortion corrected) are: myhifipa0<study num>_e0<echo num>_afni_al.nii.gz Example: myhifipa01_e01_afni_al.nii.gz myhifipa01_e01_afni_al.nii.gz myhifipa01_e02_afni_al.nii.gz myhifipa01_e03_afni_al.nii.gz For a more complete description of the output files, please see Appendix II. 10

11 Appendix I Table 1: Summary of command line tools for running MEMS protocol Command Usage Description RTctrl start RTctrl Start Start Realtime (Must be done before Start Exam ) RTctrl stop RTctrl Stop End Realtime (Must be done after End Exam ) ck ck Check if RDS client is ON kk kk Kill all active RDS clients memslist memslist List raw data files of the current MEMS scan session. memscopy memscopy s server d studyfolder login Transfer MEMS raw data files to server. gecopy physiocopy cd tail gecopy s server r raid# d studyfolder login physiocopy s server d studyfolder starttime endtime login (*starttime and endtime format: hh:mm) cd /export/home/sdc/rtafni/var/log tail f <last log file name> Transfer DICOM files to server. Transfer physio files to fmrimems server. Check realtime status and log file 11

12 Appendix II: MEMS Pipeline Output files All outputs are saved under processed folder. Below is a list of selected outputs that users may examine for sanity check purpose or for trouble shooting when there are errors in the reconstruction process. Users should contact CFMRI (cfmri@ucsd.edu) for questions regarding these files. bcaipi_<orientation>_<phase encoding dir>_mems0<study num>brik_e0<echo num>+orig.brik Example: bcaipi_sag_rev_mems01brik_e01+orig.brik mems<study num>_e0<echo num>.nii.gz Example: mems01_e01.nii.gz Reconstructed fmri data (all 3 echoes) in BRIK and NIFTI format before motion correction and distortion correction. These images have visible distortions, but should be free of obvious artifacts, and have typical T2* BOLD contrast (the example below shows four brain slices). mems01_e01.nii.gz mems01_e02.nii.gz mems01_e03.nii.gz topup_sag_fwdbrik+orig.brik and topup_sag_revbrik+orig.brik PhaseOne.nii.gz and PhaseTwo.nii.gz Reconstructed images from the TOPUP scan pair in BRIK and NIFTI format. Verify that the distortions in these two datasets are opposite in direction to each other, and there are no obvious imaging artifacts. PhaseOne.nii.gz PhaseTwo.nii.gz coilmap_sag_rev_sess_fmapbrik+orig.brik This is the reconstructed data from calibration 1 which will be used to estimate coil sensitivity profile. This scan is a spin echo, single band and single echo scan. The images should be free of artifacts, and have minimal signal dropout in the OFC area of the brain (see right). anat.nii.gz T1 Structure dataset in NIFTI format (if provided) 12

13 bcaipi_<orientation>_<phase encoding dir>_mems0<study num>brik _T2s.jpg bcaipi_<orientation>_<phase encoding dir>_mems0<study num>brik _T2s_hist.jpg bcaipi_<orientation>_<phase encoding dir>_mems0<study num>brik _e01_tsnr.jpg bcaipi_<orientation>_<phase encoding dir>_mems0<study num>brik _e01_tsnr_hist.jpg bcaipi_<orientation>_<phase encoding dir>_mems0<study num>brik _e02_tsnr.jpg bcaipi_<orientation>_<phase encoding dir>_mems0<study num>brik _e02_tsnr_hist.jpg bcaipi_<orientation>_<phase encoding dir>_mems0<study num>brik _e03_tsnr.jpg bcaipi_<orientation>_<phase encoding dir>_mems0<study num>brik _e03_tsnr_hist.jpg The T2* and tsnr maps generated for Quality Assurance (see examples below). 13

14 b02b0.cnf acqparams5r.txt BothPhases.topup_log Coefficents_fieldcoef.nii.gz Coefficents_movpar.txt TopupField.nii.gz Magnitudes.nii.gz Topup configuration and intermediate files (for more information about the TOPUP distortion correction method please refer to the FSL web site). mems0<study num>_e0<echo num>_vr.nii.gz mems0<study num>_e0<echo num>_afni_al.nii.gz mems0<study num>_e0<echo num>.nii.gz_vr_motion.1d mems0<study num>_e0<echo num>.nii.gz_al_mat.aff12.1d mems0<study num>_e0<echo num>.nii.gz_al_reg_mat.aff12.1d Motion corrected and reregistered fmri data (all three echoes). The 1D files are the motion parameters and transformation matrices. Please see Reconstruct and pre process MEMS data section for information on registration steps. script_<timestamp>.sh Example: script_ :59:34.sh Pre processing script containing motion and distortion correction steps. User can look up the preprocessing steps carried out to the fmri dataset. 14

15 Appendix III: MEICA (More information will be added in a future version of this manual) The Pipeline also supports the MEICA denoising method. To invoke MEICA denoising automatically after the pipeline finishes reconstructing data, use the meica option when calling domems.py. For example domems.py meica The script requires the following python path to be included in your.cshrc file: set path=( $path. ~/bin /usr/local/bin /apps/enthought/canopy_64bit/user/bin ) 15

Troubleshooting Guide. Prep, Scan Errors, and Artifacts

Troubleshooting Guide. Prep, Scan Errors, and Artifacts Troubleshooting Guide Prep, Scan Errors, and Artifacts Preparing for the Study Participant Compliance with MRI Scans Participant Prep Having your participant prepped will allow you to run your study with

More information

Procedures for conducting User QA on the scanner

Procedures for conducting User QA on the scanner Procedures for conducting User QA on the scanner Sample setup: The User QA phantom is clearly labeled and is stored on one of the shelves to the side of the magnet. Note proper orientation of the bottle,

More information

Phantom Test Guidance for Use of the Small MRI Phantom for the MRI Accreditation Program

Phantom Test Guidance for Use of the Small MRI Phantom for the MRI Accreditation Program Phantom Test Guidance for Use of the Small MRI Phantom for the MRI Accreditation Program 1 Contents 0.0 INTRODUCTION 4 0.1 Overview and Purpose 4 0.2 The Phantom 4 0.3 The Required Images 5 0.4 The Image

More information

EPI. Thanks to Samantha Holdsworth!

EPI. Thanks to Samantha Holdsworth! EPI Faster Cartesian approach Single-shot, Interleaved, segmented, half-k-space Delays, etc -> Phase corrections Flyback EPI GRASE Thanks to Samantha Holdsworth! 1 EPI: Speed vs Distortion Fast Spin Echo

More information

PATIENT POSITION IMAGING PARAMETERS

PATIENT POSITION IMAGING PARAMETERS 3 PLANE LOC Patient Entry Feet First Imaging Mode 2D Patient Position Prone Pulse Sequence Gradient Echo Coil Configuration 7breast both MRI Imaging Options Seq, Fast Plane 3-PLANE Acceleration Factor

More information

4/14/2009. The Big Picture of Quality. MRI Quality Assurance and ACR MRI Accreditation Program. Basic Elements for Image Quality.

4/14/2009. The Big Picture of Quality. MRI Quality Assurance and ACR MRI Accreditation Program. Basic Elements for Image Quality. The Big Picture of Quality MRI Quality Assurance and ACR MRI Accreditation Program Chen Lin, PhD Indiana University School of Medicine & Clarian Health Partners Diagnosis accuracy Image quality Knowledge

More information

Topic: Instructional David G. Thomas December 23, 2015

Topic: Instructional David G. Thomas December 23, 2015 Procedure to Setup a 3ɸ Linear Motor This is a guide to configure a 3ɸ linear motor using either analog or digital encoder feedback with an Elmo Gold Line drive. Topic: Instructional David G. Thomas December

More information

Release Notes. Multi-Band EPI C2P. Release February 2013

Release Notes. Multi-Band EPI C2P. Release February 2013 Release Notes Multi-Band EPI C2P Release 008 13 February 2013 Installation 1. Restart the system (reboot host and MRIR) 2. Extract the.zip file to a temporary directory 3. Run the installer.bat file 4.

More information

Multiparametric MRI Prostate Imaging Protocol November 2015 Full Acquisition Protocol with Parameters GE 3T Magnet with Software Version DV25

Multiparametric MRI Prostate Imaging Protocol November 2015 Full Acquisition Protocol with Parameters GE 3T Magnet with Software Version DV25 3Plane Loc SSFSE Multiparametric MRI Prostate Imaging Protocol November 2015 Full Acquisition Protocol with Parameters GE 3T Magnet with Software Version DV25 Save Series Scan After acquisition, scroll

More information

SEM- EDS Instruction Manual

SEM- EDS Instruction Manual SEM- EDS Instruction Manual Double-click on the Spirit icon ( ) on the desktop to start the software program. I. X-ray Functions Access the basic X-ray acquisition, display and analysis functions through

More information

Comparison of Robarts s 3T and 7T MRI Machines for obtaining fmri Sequences Medical Biophysics 3970: General Laboratory

Comparison of Robarts s 3T and 7T MRI Machines for obtaining fmri Sequences Medical Biophysics 3970: General Laboratory Comparison of Robarts s 3T and 7T MRI Machines for obtaining fmri Sequences Medical Biophysics 3970: General Laboratory Jacob Matthews 4/13/2012 Supervisor: Rhodri Cusack, PhD Assistance: Annika Linke,

More information

Operating Procedures for RECO1 & RECO2

Operating Procedures for RECO1 & RECO2 Operating Procedures for RECO1 & RECO2 Reconstruction of Data 1) Open CT Pro 3D a. Remember, when the computer you are working on is receiving files from the scanner, it cannot reconstruct data b. You

More information

Configuring and Troubleshooting Set-Top Boxes

Configuring and Troubleshooting Set-Top Boxes Diagnose RF Tuner Issues, page 1 Protect End-User Privacy, page 4 Apply Services to Device, page 5 View Video Recordings, page 6 View VOD Programs, page 6 View PDL Programs, page 7 View Hard Disk Information,

More information

M R I Physics Course. Jerry Allison Ph.D. Chris Wright B.S. Tom Lavin M.S.M.P. Department of Radiology Medical College of Georgia

M R I Physics Course. Jerry Allison Ph.D. Chris Wright B.S. Tom Lavin M.S.M.P. Department of Radiology Medical College of Georgia M R I Physics Course Jerry Allison Ph.D. Chris Wright B.S. Tom Lavin M.S.M.P. Department of Radiology Medical College of Georgia M R I Physics Course chapter 12 Artifacts and Suppression Techniques Artifacts

More information

RedRat Control User Guide

RedRat Control User Guide RedRat Control User Guide Chris Dodge RedRat Ltd April 2014 For RedRat Control V3.16 1 Contents 1. Introduction 3 2. Prerequisites and Installation 3 3. The First Step Capture of Remote Control Signals

More information

RAD 465 (MRI) Lecture one (Pulse Sequences) Ruba Khushaim MSc

RAD 465 (MRI) Lecture one (Pulse Sequences) Ruba Khushaim MSc RAD 465 (MRI) Lecture one (Pulse Sequences) Ruba Khushaim MSc Outline : Spine echo pulse sequence SE Fast spin echo pulse sequence FSE Inversion recovery pulse sequence IR Gradient pulse sequence GS Pulse

More information

Nuclear Associates and

Nuclear Associates and Nuclear Associates 76-907 and 76-908 AAPM MRI Phantoms Users Manual March 2005 Manual No. 38616 Rev. 3 2003, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks

More information

CMRRpack2. a collection of pulse sequences from the. University of Minnesota, Center for Magnetic Resonance Research (CMRR) Version: 2.

CMRRpack2. a collection of pulse sequences from the. University of Minnesota, Center for Magnetic Resonance Research (CMRR) Version: 2. CMRRpack2 a collection of pulse sequences from the University of Minnesota, Center for Magnetic Resonance Research (CMRR) Version: 2.63_package This document generated: Mon Dec 16 2013 15:55:46 Contents

More information

WU-Minn HCP MEG Initial Data Release: Reference Manual

WU-Minn HCP MEG Initial Data Release: Reference Manual WU-Minn HCP MEG Initial Data Release: Reference Manual Appendix I HCP MEG Session Protocol Details 4 March 2014 MEG1 Reference Manual Appendix I WU-Minn Consortium of the NIH Human Connectome Project Page

More information

SigPlay User s Guide

SigPlay User s Guide SigPlay User s Guide . . SigPlay32 User's Guide? Version 3.4 Copyright? 2001 TDT. All rights reserved. No part of this manual may be reproduced or transmitted in any form or by any means, electronic or

More information

True comfort and flexibility with the power of 3T.

True comfort and flexibility with the power of 3T. True comfort and flexibility with the power of 3T. With a large 71 cm aperture and the quietest exams in the industry, the Vantage Titan 3T is the most comfortable 3T MRI system for all of your patients.

More information

Scope: All CT staff technologist

Scope: All CT staff technologist APPROVED BY: Radiology Technical Director Page 1 of 6 Purpose: The QC program assesses relative changes in system performance as determined by the technologist, service engineer, qualified medical physicist,

More information

Spectrum Analyser Basics

Spectrum Analyser Basics Hands-On Learning Spectrum Analyser Basics Peter D. Hiscocks Syscomp Electronic Design Limited Email: phiscock@ee.ryerson.ca June 28, 2014 Introduction Figure 1: GUI Startup Screen In a previous exercise,

More information

Classical simulations with PJNMR BCMB/CHEM LAB 01/25/12

Classical simulations with PJNMR BCMB/CHEM LAB 01/25/12 Classical simulations with PJNMR BCMB/CHEM LAB 01/25/12 The program PJNMR was written by Paul- Jean Letourneau as a joint project between the PENCE & NANUC projects in Canada. It rotates vectors representing

More information

Image quality in non-gated versus gated reconstruction of tongue motion using Magnetic Resonance Imaging:

Image quality in non-gated versus gated reconstruction of tongue motion using Magnetic Resonance Imaging: This talk was presented 26 June 2008, at the 22nd International Congress and Exhibition of Computer Assisted Radiology and Surgery, in Barcelona at the Hotel Constanza from June 25 to 28, 2008. See http://kochanski.org/gpk/papers/2008/carstalk.html

More information

AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode)

AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode) AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode) 1. Log into the Log Usage system on the SMIF web site 2. Open Nanoscope 6.14r1 software by double clicking on the Nanoscope 6.14r1 desktop

More information

Abstract. Learning Objectives 8/1/2017

Abstract. Learning Objectives 8/1/2017 SAM Practical Medical Physics TU-B-201-0 AAPM Annual Meeting 2017 1 Abstract This course will teach the participant to identify common artifacts found clinically in MR, DR, CT, PET, to determine the causes

More information

Part 1 Basic Operation

Part 1 Basic Operation This product is a designed for video surveillance video encode and record, it include H.264 video Compression, large HDD storage, network, embedded Linux operate system and other advanced electronic technology,

More information

Source/Receiver (SR) Setup

Source/Receiver (SR) Setup PS User Guide Series 2015 Source/Receiver (SR) Setup For 1-D and 2-D Vs Profiling Prepared By Choon B. Park, Ph.D. January 2015 Table of Contents Page 1. Overview 2 2. Source/Receiver (SR) Setup Main Menu

More information

The Kaffeine Handbook. Jürgen Kofler Christophe Thommeret Mauro Carvalho Chehab

The Kaffeine Handbook. Jürgen Kofler Christophe Thommeret Mauro Carvalho Chehab Jürgen Kofler Christophe Thommeret Mauro Carvalho Chehab 2 Contents 1 Kaffeine Player 5 1.1 The Start Window...................................... 5 1.2 Play a File..........................................

More information

Common Spatial Patterns 2 class BCI V Copyright 2012 g.tec medical engineering GmbH

Common Spatial Patterns 2 class BCI V Copyright 2012 g.tec medical engineering GmbH g.tec medical engineering GmbH Sierningstrasse 14, A-4521 Schiedlberg Austria - Europe Tel.: (43)-7251-22240-0 Fax: (43)-7251-22240-39 office@gtec.at, http://www.gtec.at Common Spatial Patterns 2 class

More information

Data Acquisition Instructions

Data Acquisition Instructions Page 1 of 13 Form 0162A 7/21/2006 Superchips Inc. Superchips flashpaq Data Acquisition Instructions Visit Flashpaq.com for downloadable updates & upgrades to your existing tuner (See the next page for

More information

Request for Proposals

Request for Proposals Request for Proposals Reference: ERDFIAI2012-4006A Caring First Ltd, is a limited liability company providing healthcare services. The company is currently commissioning a private new-build hospital complete

More information

Single cable multiswich programmer PC102W

Single cable multiswich programmer PC102W Single cable multiswich programmer PC102W 1. Product description The PC102W - single cable multiswich programmer (in the text - programmer) is useful instrument while configuring and troubleshooting SAT

More information

The Measurement Tools and What They Do

The Measurement Tools and What They Do 2 The Measurement Tools The Measurement Tools and What They Do JITTERWIZARD The JitterWizard is a unique capability of the JitterPro package that performs the requisite scope setup chores while simplifying

More information

Getting Started with the LabVIEW Sound and Vibration Toolkit

Getting Started with the LabVIEW Sound and Vibration Toolkit 1 Getting Started with the LabVIEW Sound and Vibration Toolkit This tutorial is designed to introduce you to some of the sound and vibration analysis capabilities in the industry-leading software tool

More information

HITACHI S FAST SPIN ECHO TECHNOLOGY

HITACHI S FAST SPIN ECHO TECHNOLOGY primefse TECHNOLOGY Yosuke Hitata RT Makoto Sasaki MD Kunio Esashika RT Hiroshi Gakumazawa RT HITACHI S FAST SPIN ECHO TECHNOLOGY Efficacies in Improving Image Quality & Usability Hitachi Medical Systems

More information

Setup Guide. CalMAN Client for SCRATCH. Rev. 1.1

Setup Guide. CalMAN Client for SCRATCH. Rev. 1.1 Setup Guide CalMAN Client for SCRATCH Rev. 1.1 Introduction CalMAN Required Software Version: CalMAN Display Calibration Software interfaces directly with ASSIMILATE SCRATCH software through the CalMAN

More information

Welch Allyn CardioPerfect Workstation Tango+ Interface Notes

Welch Allyn CardioPerfect Workstation Tango+ Interface Notes Welch Allyn CardioPerfect Workstation Tango+ Interface Notes To setup Tango+ with the CardioPerfect stress system, simply follow the directions below. 1. Verify Correct RS-232 and ECG Trigger Cables RS-232

More information

Preparing for remote data collection at NE-CAT

Preparing for remote data collection at NE-CAT Preparing for remote data collection at NE-CAT Important Note: The beamtime and remote login privileges are intended just for you and your group. You are not allowed to share these with any other person

More information

Software manual. ChipScan-Scanner 3.0

Software manual. ChipScan-Scanner 3.0 Software manual ChipScan-Scanner 3.0 Copyright c Langer EMV-Technik GmbH 2016 Version 1.10 Notice No part of this manual may be reproduced in any form or by any means (including electronic storage and

More information

PulseCounter Neutron & Gamma Spectrometry Software Manual

PulseCounter Neutron & Gamma Spectrometry Software Manual PulseCounter Neutron & Gamma Spectrometry Software Manual MAXIMUS ENERGY CORPORATION Written by Dr. Max I. Fomitchev-Zamilov Web: maximus.energy TABLE OF CONTENTS 0. GENERAL INFORMATION 1. DEFAULT SCREEN

More information

C8000. switch over & ducking

C8000. switch over & ducking features Automatic or manual Switch Over or Fail Over in case of input level loss. Ducking of a main stereo or surround sound signal by a line level microphone or by a pre recorded announcement / ad input.

More information

Torsional vibration analysis in ArtemiS SUITE 1

Torsional vibration analysis in ArtemiS SUITE 1 02/18 in ArtemiS SUITE 1 Introduction 1 Revolution speed information as a separate analog channel 1 Revolution speed information as a digital pulse channel 2 Proceeding and general notes 3 Application

More information

Re: ENSC 370 Project Physiological Signal Data Logger Functional Specifications

Re: ENSC 370 Project Physiological Signal Data Logger Functional Specifications School of Engineering Science Simon Fraser University V5A 1S6 versatile-innovations@sfu.ca February 12, 1999 Dr. Andrew Rawicz School of Engineering Science Simon Fraser University Burnaby, BC V5A 1S6

More information

SPM Training Manual Veeco Bioscope II NIFTI-NUANCE Center Northwestern University

SPM Training Manual Veeco Bioscope II NIFTI-NUANCE Center Northwestern University SPM Training Manual Veeco Bioscope II NIFTI-NUANCE Center Northwestern University Introduction: Scanning Probe Microscopy (SPM) is a general term referring to surface characterization techniques that utilize

More information

Procedures to Characterize Maury s Automatic Tuner Using ATS Software Version 5.1 or above

Procedures to Characterize Maury s Automatic Tuner Using ATS Software Version 5.1 or above Procedures to Characterize Maury s Automatic Tuner Using ATS Software Version 5.1 or above Things to check before tuner characterization Make sure tuner is power up and USB cable is connected to the computer

More information

VIDEO GRABBER. DisplayPort. User Manual

VIDEO GRABBER. DisplayPort. User Manual VIDEO GRABBER DisplayPort User Manual Version Date Description Author 1.0 2016.03.02 New document MM 1.1 2016.11.02 Revised to match 1.5 device firmware version MM 1.2 2019.11.28 Drawings changes MM 2

More information

Common Spatial Patterns 3 class BCI V Copyright 2012 g.tec medical engineering GmbH

Common Spatial Patterns 3 class BCI V Copyright 2012 g.tec medical engineering GmbH g.tec medical engineering GmbH Sierningstrasse 14, A-4521 Schiedlberg Austria - Europe Tel.: (43)-7251-22240-0 Fax: (43)-7251-22240-39 office@gtec.at, http://www.gtec.at Common Spatial Patterns 3 class

More information

HV-122-DCA DVB-T 2-Way Diversity Receiver Box Quick Installation Guide

HV-122-DCA DVB-T 2-Way Diversity Receiver Box Quick Installation Guide HV-122-DCA DVB-T 2-Way Diversity Receiver Box Quick Installation Guide PACKAGE CONTENTS 4 FRONT PANEL VIEW 4 BACK PANEL VIEW 4 BOARD VIEW 5 IR REMOTE CONTROLLER-TYPE A 6 FILL BATTERY TO IR CONTROLLERS:

More information

Table of content. Table of content Introduction Concepts Hardware setup...4

Table of content. Table of content Introduction Concepts Hardware setup...4 Table of content Table of content... 1 Introduction... 2 1. Concepts...3 2. Hardware setup...4 2.1. ArtNet, Nodes and Switches...4 2.2. e:cue butlers...5 2.3. Computer...5 3. Installation...6 4. LED Mapper

More information

Dynamic Cable Rating Model Guidelines

Dynamic Cable Rating Model Guidelines lines Introduction This document describes how to set up, configure and run the Dynamic Cable Rating model in Ipsa. The Dynamic Cable Rating model (DCR) has been developed for the WPD FALCON project by

More information

Select source Click the Bio Settings button to modify device settings. Select Fingers Use Ctrl+Left mouse button to select multiple fingers to scan.

Select source Click the Bio Settings button to modify device settings. Select Fingers Use Ctrl+Left mouse button to select multiple fingers to scan. 1. Select Source for SC Biometrics Choose Select Source from the Image Menu. Select the desired image type to link to the SC Biometric image source. Select SC Biometrics from the Custom dropdown list.

More information

HyperMedia User Manual

HyperMedia User Manual HyperMedia User Manual Contents V3.5 Chapter 1 : HyperMedia Software Functions... 3 1.1 HyperMedia Introduction... 3 1.2 Main Panel... 3 1.2.2 Information Window... 4 1.2.3 Keypad... 4 1.2.4 Channel Index...

More information

APP USE USER MANUAL 2017 VERSION BASED ON WAVE TRACKING TECHNIQUE

APP USE USER MANUAL 2017 VERSION BASED ON WAVE TRACKING TECHNIQUE APP USE USER MANUAL 2017 VERSION BASED ON WAVE TRACKING TECHNIQUE All rights reserved All rights reserved. No part of this publication may be reproduced, stored in a retrieval system, or transmitted in

More information

Hardware Setup. If you have any further questions after checking this document, please contact

Hardware Setup. If you have any further questions after checking this document, please contact Hardware Setup If you have any further questions after checking this document, please contact support@cognisens.com Hardware Setup Getting Started: NeuroTracker Pro WHAT TO BUY SETTING UP YOUR 3D TV SETTING

More information

GIfMI MRI user manual

GIfMI MRI user manual AUTHOR: SBOGAERT PAGE: 1/92 DATE: 2017-05-29 VERSION: 1 TITLE: MRI USER MANUAL GIfMI MRI user manual OVERVIEW I. Introduction... 5 II. GIfMI scanning policy MRI safety... 6 Policy regarding personnel -

More information

Experiment: Real Forces acting on a Falling Body

Experiment: Real Forces acting on a Falling Body Phy 201: Fundamentals of Physics I Lab 1 Experiment: Real Forces acting on a Falling Body Objectives: o Observe and record the motion of a falling body o Use video analysis to analyze the motion of a falling

More information

Standard Operating Procedure of nanoir2-s

Standard Operating Procedure of nanoir2-s Standard Operating Procedure of nanoir2-s The Anasys nanoir2 system is the AFM-based nanoscale infrared (IR) spectrometer, which has a patented technique based on photothermal induced resonance (PTIR),

More information

Operating Instructions

Operating Instructions Operating Instructions HAEFELY TEST AG KIT Measurement Software Version 1.0 KIT / En Date Version Responsable Changes / Reasons February 2015 1.0 Initial version WARNING Introduction i Before operating

More information

DETEXI Basic Configuration

DETEXI Basic Configuration DETEXI Network Video Management System 5.5 EXPAND YOUR CONCEPTS OF SECURITY DETEXI Basic Configuration SETUP A FUNCTIONING DETEXI NVR / CLIENT It is important to know how to properly setup the DETEXI software

More information

EDL8 Race Dash Manual Engine Management Systems

EDL8 Race Dash Manual Engine Management Systems Engine Management Systems EDL8 Race Dash Manual Engine Management Systems Page 1 EDL8 Race Dash Page 2 EMS Computers Pty Ltd Unit 9 / 171 Power St Glendenning NSW, 2761 Australia Phone.: +612 9675 1414

More information

Vascular. Development of Trinias FPD-Equipped Angiography System. 1. Introduction. MEDICAL NOW No.73 (2013.2) Yoshiaki Miura

Vascular. Development of Trinias FPD-Equipped Angiography System. 1. Introduction. MEDICAL NOW No.73 (2013.2) Yoshiaki Miura Vascular Development of Trinias FPD-Equipped Angiography System Medical Systems Division, Shimadzu Corporation Yoshiaki Miura 1. Introduction Shimadzu has developed Trinias (one ceiling-mounted type C12

More information

RF Explorer RackPRO. User Manual. Introduction. Greetings fellow traveler on the RF spectrum.

RF Explorer RackPRO. User Manual. Introduction. Greetings fellow traveler on the RF spectrum. RF Explorer RackPRO User Manual Introduction Greetings fellow traveler on the RF spectrum. The RF Explorer RackPRO (referred to in this document in shorthand as RackPRO ) has been designed to be intuitive

More information

Scan Converter Quick Installation Guide

Scan Converter Quick Installation Guide Scan Converter Quick Installation Guide Software Note: No software is required to use your scan converter. Please complete the hardware installation and system setup before you determine the need to the

More information

Pre-processing of revolution speed data in ArtemiS SUITE 1

Pre-processing of revolution speed data in ArtemiS SUITE 1 03/18 in ArtemiS SUITE 1 Introduction 1 TTL logic 2 Sources of error in pulse data acquisition 3 Processing of trigger signals 5 Revolution speed acquisition with complex pulse patterns 7 Introduction

More information

Performing a Sound Level Measurement

Performing a Sound Level Measurement APPENDIX 9 Performing a Sound Level Measurement Due to the many features of the System 824 and the variety of measurements it is capable of performing, there is a great deal of instructive material in

More information

2

2 1 2 3 4 5 6 7 1. Bauhs, J. A., Vrieze, T. J., Primak, A. N., Bruesewitz, M. R., & McCollough, C. H. (2008). CT Dosimetry: Comparison of Measurement Techniques and Devices1. Radiographics, 28(1), 245-253.

More information

Practicum 3, Fall 2010

Practicum 3, Fall 2010 A. F. Miller 2010 T1 Measurement 1 Practicum 3, Fall 2010 Measuring the longitudinal relaxation time: T1. Strychnine, dissolved CDCl3 The T1 is the characteristic time of relaxation of Z magnetization

More information

Brain-Computer Interface (BCI)

Brain-Computer Interface (BCI) Brain-Computer Interface (BCI) Christoph Guger, Günter Edlinger, g.tec Guger Technologies OEG Herbersteinstr. 60, 8020 Graz, Austria, guger@gtec.at This tutorial shows HOW-TO find and extract proper signal

More information

Matrox PowerStream Plus

Matrox PowerStream Plus Matrox PowerStream Plus User Guide 20246-301-0100 2016.12.01 Contents 1 About this user guide...5 1.1 Using this guide... 5 1.2 More information... 5 2 Matrox PowerStream Plus software...6 2.1 Before you

More information

Breast MR Imaging and Quality Control

Breast MR Imaging and Quality Control Breast MR Imaging and Quality Control Donna M. Reeve, MS, DABR, DABMP Department of Imaging Physics Educational Objectives 1. Provide an overview of breast MR imaging and MR-guided biopsy procedures. 2.

More information

administration access control A security feature that determines who can edit the configuration settings for a given Transmitter.

administration access control A security feature that determines who can edit the configuration settings for a given Transmitter. Castanet Glossary access control (on a Transmitter) Various means of controlling who can administer the Transmitter and which users can access channels on it. See administration access control, channel

More information

Positive Attendance. Overview What is Positive Attendance? Who may use Positive Attendance? How does the Positive Attendance option work?

Positive Attendance. Overview What is Positive Attendance? Who may use Positive Attendance? How does the Positive Attendance option work? Positive Attendance Overview What is Positive Attendance? Who may use Positive Attendance? How does the Positive Attendance option work? Setup Security Codes Absence Types Absence Reasons Attendance Periods/Bell

More information

CCE900-IP-TR. User s Guide

CCE900-IP-TR. User s Guide CCE900-IP-TR CCE900-IP-T & CCE900-IP-R User s Guide i-tech Company LLC TOLL FREE: (888) 483-2418 EMAIL: info@itechlcd.com WEB: www.itechlcd.com 1. Introduction The CCE900-IP-T & CCE900-IP-R is a solution

More information

INSTALATION PROCEDURE

INSTALATION PROCEDURE INSTALLATION PROCEDURE Overview The most difficult part of an installation is in knowing where to start and the most important part is starting in the proper start. There are a few very important items

More information

MestReNova Manual for Chem 201/202. October, 2015.

MestReNova Manual for Chem 201/202. October, 2015. 1. Introduction to 1-D NMR Data Processing with MestReNova The MestReNova program can do all of the routine NMR data processing needed for Chem 201 and 202 and will be available through the Reed downloads

More information

DIVERSITY DVB-T RECEIVER (DDR)

DIVERSITY DVB-T RECEIVER (DDR) User s Manual The most important thing we build is trust. DIVERSITY DVB-T RECEIVER (DDR) Cobham Surveillance GMS Products 1916 Palomar Oaks Way Ste 100 Carlsbad, CA 92008 100-M0062X2 T: 760-496-0055 05/15/09

More information

Calibration Using i1 probe with CalMan and Resolve

Calibration Using i1 probe with CalMan and Resolve 20160526 Flanders Scientific, Inc. 6215 Shiloh Crossing Suite G Alpharetta, GA 30005 Phone: +1.678.835.4934 Fax: +1.678.804.1882 E-Mail: Support@FlandersScientific.com www.flandersscientific.com Please

More information

Hidden melody in music playing motion: Music recording using optical motion tracking system

Hidden melody in music playing motion: Music recording using optical motion tracking system PROCEEDINGS of the 22 nd International Congress on Acoustics General Musical Acoustics: Paper ICA2016-692 Hidden melody in music playing motion: Music recording using optical motion tracking system Min-Ho

More information

Getting started with

Getting started with Getting started with Electricity consumption monitoring single phase for homes and some smaller light commercial premises OVERVIEW: The OWL Intuition-e electricity monitoring system comprises of three

More information

Processing data with Mestrelab Mnova

Processing data with Mestrelab Mnova Processing data with Mestrelab Mnova This exercise has three parts: a 1D 1 H spectrum to baseline correct, integrate, peak-pick, and plot; a 2D spectrum to plot with a 1 H spectrum as a projection; and

More information

AmbDec User Manual. Fons Adriaensen

AmbDec User Manual. Fons Adriaensen AmbDec - 0.4.2 User Manual Fons Adriaensen fons@kokkinizita.net Contents 1 Introduction 3 1.1 Computing decoder matrices............................. 3 2 Installing and running AmbDec 4 2.1 Installing

More information

Contents: 1 LANsmart Pro Main Unit 4 Remote Unit: ID1, ID2, ID3, ID4

Contents: 1 LANsmart Pro Main Unit 4 Remote Unit: ID1, ID2, ID3, ID4 LANsmart Pro user manual Introduction LANsmart Pro is a hand-held, multifunction Cable Map Tester and Cable Length Meter. It has an integrated Analog and Digital Tone Generator, Port Finder, and Quick

More information

SC26 Magnetic Field Cancelling System

SC26 Magnetic Field Cancelling System SPICER CONSULTING SYSTEM SC26 SC26 Magnetic Field Cancelling System Makes the ambient magnetic field OK for electron beam tools in 300 mm wafer fabs Real time, wideband cancelling from DC to > 9 khz fields

More information

USER S GUIDE DSR-1 DE-ESSER. Plug-in for Mackie Digital Mixers

USER S GUIDE DSR-1 DE-ESSER. Plug-in for Mackie Digital Mixers USER S GUIDE DSR-1 DE-ESSER Plug-in for Mackie Digital Mixers Iconography This icon identifies a description of how to perform an action with the mouse. This icon identifies a description of how to perform

More information

CI-218 / CI-303 / CI430

CI-218 / CI-303 / CI430 CI-218 / CI-303 / CI430 Network Camera User Manual English AREC Inc. All Rights Reserved 2017. l www.arec.com All information contained in this document is Proprietary Table of Contents 1. Overview 1.1

More information

Housing Inventory Setup Guide

Housing Inventory Setup Guide The following guide should allow users with residential programs to become more familiar with the inventory setup and maintenance functions within the PA HMIS/Client Track system. This guide will walk

More information

MSK Imaging Fundamentals

MSK Imaging Fundamentals MSK Imaging Fundamentals Goals Improve image quality Provide best possible product for our customers Patient Referring clinician Radiologist Reduce number of callback cases Goal reduction of 50% in 3 months

More information

User Guide & Reference Manual

User Guide & Reference Manual TSA3300 TELEPHONE SIGNAL ANALYZER User Guide & Reference Manual Release 2.1 June 2000 Copyright 2000 by Advent Instruments Inc. TSA3300 TELEPHONE SIGNAL ANALYZER ii Overview SECTION 1 INSTALLATION & SETUP

More information

What s New in Raven May 2006 This document briefly summarizes the new features that have been added to Raven since the release of Raven

What s New in Raven May 2006 This document briefly summarizes the new features that have been added to Raven since the release of Raven What s New in Raven 1.3 16 May 2006 This document briefly summarizes the new features that have been added to Raven since the release of Raven 1.2.1. Extensible multi-channel audio input device support

More information

BER MEASUREMENT IN THE NOISY CHANNEL

BER MEASUREMENT IN THE NOISY CHANNEL BER MEASUREMENT IN THE NOISY CHANNEL PREPARATION... 2 overview... 2 the basic system... 3 a more detailed description... 4 theoretical predictions... 5 EXPERIMENT... 6 the ERROR COUNTING UTILITIES module...

More information

Characterization of Diamond Samples CHESS Run Summer 2013

Characterization of Diamond Samples CHESS Run Summer 2013 Characterization of Diamond Samples CHESS Run Summer 2013 5/22 5/28/2013 Characterization of Diamond Samples CHESS Run Summer 2013 Useful information Goals for this run Conditions of the experiment 1.

More information

Reconstruction of Ca 2+ dynamics from low frame rate Ca 2+ imaging data CS229 final project. Submitted by: Limor Bursztyn

Reconstruction of Ca 2+ dynamics from low frame rate Ca 2+ imaging data CS229 final project. Submitted by: Limor Bursztyn Reconstruction of Ca 2+ dynamics from low frame rate Ca 2+ imaging data CS229 final project. Submitted by: Limor Bursztyn Introduction Active neurons communicate by action potential firing (spikes), accompanied

More information

Session 1 Introduction to Data Acquisition and Real-Time Control

Session 1 Introduction to Data Acquisition and Real-Time Control EE-371 CONTROL SYSTEMS LABORATORY Session 1 Introduction to Data Acquisition and Real-Time Control Purpose The objectives of this session are To gain familiarity with the MultiQ3 board and WinCon software.

More information

RSNA 2006 November 26 to December 1 Chicago. Guest author for ImPACT Dr. Koos Geleijns, Medical Physicist, Leiden University Medical Center.

RSNA 2006 November 26 to December 1 Chicago. Guest author for ImPACT Dr. Koos Geleijns, Medical Physicist, Leiden University Medical Center. RSNA 2006 November 26 to December 1 Chicago Guest author for ImPACT Dr. Koos Geleijns, Medical Physicist, Leiden University Medical Center. Once again, more than 60,000 participants (including professional

More information

Quick-Start for READ30

Quick-Start for READ30 Quick-Start for READ30 The program READ30 was written for the purpose of reading and configuring the digital pressure-transmitter of the series 30. The two features are divided into the following parts:

More information

BioGraph Infiniti Physiology Suite

BioGraph Infiniti Physiology Suite Thought Technology Ltd. 2180 Belgrave Avenue, Montreal, QC H4A 2L8 Canada Tel: (800) 361-3651 ٠ (514) 489-8251 Fax: (514) 489-8255 E-mail: mail@thoughttechnology.com Webpage: http://www.thoughttechnology.com

More information

Matrox PowerStream Plus

Matrox PowerStream Plus Matrox PowerStream Plus User Guide 20246-301-0200 2017.07.04 Contents 1 About this user guide... 5 1.1 Using this guide... 5 1.2 More information... 5 2 Matrox PowerStream Plus software... 6 2.1 Before

More information

PRELIMINARY INFORMATION. Professional Signal Generation and Monitoring Options for RIFEforLIFE Research Equipment

PRELIMINARY INFORMATION. Professional Signal Generation and Monitoring Options for RIFEforLIFE Research Equipment Integrated Component Options Professional Signal Generation and Monitoring Options for RIFEforLIFE Research Equipment PRELIMINARY INFORMATION SquareGENpro is the latest and most versatile of the frequency

More information