Dektak II SOP Revision 1 05/30/12 Page 1 of 5. NRF Dektak II SOP

Size: px
Start display at page:

Download "Dektak II SOP Revision 1 05/30/12 Page 1 of 5. NRF Dektak II SOP"

Transcription

1 Page 1 of 5 NRF Dektak II SOP The Dektak II-A is a sensitive stylus profilometer. A diamond-tipped stylus is moved laterally across the surface while in contact and measures deflections of the tip. It can be used to measure step heights (film thickness) and surface roughness. The vertical measuring range is 50Å to 65µm and can perform a line scan as long as 30mm. Prerequisites for operating the Dektak II: a) Receive one on one training and certification from NRF Staff. Discuss your process with a staff member. b) Obtain a NRF ID (if you do not already have one) by completing the NRF Lab Use Request Form. Safety High Voltage - Voltages supplied to and within certain areas of the system are potentially dangerous and can cause injury to personnel. 1.0 Pre-Operation 1.1 Tool Reservations may be made via the NRF Reservation Page.

2 Page 2 of Operation 2.1 START UP Turn on the unit by pressing the switch located on the back. The documentation included with the DekTak tool states that the tool makes the most accurate measurements 15 minutes after having been turned on. Turn on the unit before logging into TUMI and allow it to warm up! If you do not see anything on the main monitor, press the VIDEO key under DISPLAY/STYLUS to switch on the text screen of the main monitor. 2.2 SAMPLE LOADING Move the stage forward along the Y-axis, by turning the right thumbwheel clockwise, until your sample can be placed near the center of the stage Use the X and Y thumbwheels to move the stage until the sample is positioned below the stylus Press the key under DISPLAY/STYLUS to lower the stylus. The monitor displays the image from the mounted video camera. Turn the focus knob such that the stylus tip and its reflection are sharp and centered on the screen. DO NOT move the stage by a large amount while the stylus is lowered Press the key under DISPLAY/STYLUS to raise the stylus. Locate the general area of measurement. Refocus the sample, if necessary.

3 Page 3 of Lower the stylus again, and use the thumbwheels to CAREFULLY move the feature to be measured to the right of the stylus tip. The scan direction is to the right. CAUTION: Keep the stylus RETRACTED while moving the sample over long distances. Lower the stylus only when making fine adjustments to find the scan's exactstarting point Raise the stylus by pressing the key. The instrument is now ready to make a scan. 2.3 Set/Check Scan Parameters Press the PROG key under SET-UP on the keyboard Program the appropriate scan length, speed, range, and profile. Use the arrow keys under DIRECTION on the keyboard and the numerical keys to change the parameters. Press ENTER whenever making changes. Select NO for Auto Leveling Scan Program. Up to ten programs can be stored in the machine for quick loading of often-used parameters Scan Length. Scan lengths are available from 50 µm to 30 mm Speed. Three scan speeds are available: Low, Medium, and High. Low speed provides high horizontal resolution measurements (i.e., generates more data points). High speed saves time, but offers lower resolution. Medium speed is most commonly used Range. This parameter provides a choice between the Auto Ranging feature and a user determined display range. The maximum display range is 655 kå. Selecting an Auto Ranging feature automatically ranges the profile to fill 80% of the display Profile. Three different profiles are available, depending upon the sample surface characteristics and the measurement range to be selected Valleys - Provides 90% of the Measurement Range below the zero horizontal grid line Hills and Valleys - Provides 50% of the Measurement Range above the zero horizontal grid line and 50% below Hills - Provides 90% of the Measurement Range above the zero horizontal grid line.

4 Page 4 of Auto Leveling. This is used to automatically level a scan where two scan reference points lie on the same plane. Always set to NO Reference Cursor. This sets the horizontal position of the reference cursor and is very useful for measurements on multiple, identical samples or multiple identical measurement locations on the same sample. The cursor position can be moved later as necessary, so it is not critical to position it at this time Measurement Cursor. This sets the horizontal position for the measurement cursor in an identical manner. 2.4 SCAN SAMPLE Make sure your scan parameters are correct, and press the SCAN key under SET-UP to initiate the scan. On the main monitor, you can see a surface profile trace being generated (vertical vs. horizontal distance on a large scale). When the scan is finished, the plot will be redrawn according to the program settings If the trace is at an extreme slope, manual coarse leveling (see section 6.5) is necessary to obtain accurate results Level the trace if coarse leveling is not required Position the R (Reference) and M (Measuring) cursors on two points that should be of equal height along the sloping line. Press the keys labeled R and M under CURSOR to select a cursor. The selected cursor is shown on the lower-right corner of the main monitor. Use the left and right arrow keys under DIRECTION to move the cursors. Holding down the FAST key while moving the cursors makes them travel faster Press the LEVEL key under CONTROL and the profile will be redrawn with the points leveled Once the trace is level, position the R (Reference) cursor at the base of the step and the M (Measuring) cursor at the top of the step to measure the step height. Press the keys labeled R and M under CURSOR to select a cursor. The selected cursor is displayed on the lower-right corner of the main monitor. Use the left and right arrow keys under DIRECTION to move the cursors. Holding down the FAST key while moving the cursors makes them travel faster The absolute height at each cursor is shown on the screen by R CUR and MCUR, and is followed by the horizontal distance from the starting scan point. VERT shows the vertical distance between the

5 Page 5 of 5 height values at the R and M cursors. HORIZ shows the horizontal distance between the two cursors. 2.5 UNLOAD SAMPLE Making sure the stylus is retracted, move the sample stage outward using the Y thumbwheel until the sample is clear from under the stylus Remove the sample with tweezers and load the next sample if applicable Turn off the power. 2.6 COARSE MANUAL LEVELING Using either your own sample or a flat wafer, locate an area of the surface that is known to be level, and position it for measuring. See section 2.2 for help with sample positioning Press the SCAN key under SET-UP As the stage is moving and a trace is being generated on the screen, turn the leveling thumbwheel (shown in Fig. 1.) until the profile trace is tracking in a horizontal line. Clockwise rotation raises the trace and counterclockwise lowers the trace Press SCAN again. The profile must appear completely within the graphic boundaries (with Range set to Auto in the program menu) to achieve the minimum acceptable manual leveling. If not, repeat the manual leveling procedure above Once the stage is level, proceed to step 2.2 to measure or remeasure your sample. NOTE: You may want to adjust the Range parameter in the program menu to a smaller scale for finer leveling.

Approved by Principal Investigator Date: Approved by Super User: Date:

Approved by Principal Investigator Date: Approved by Super User: Date: Approved by Principal Investigator Date: Approved by Super User: Date: Standard Operating Procedure BNC Dektak 3030 Stylus Profilometer Version 2011 May 16 I. Purpose This Standard Operating Procedure

More information

Dektak Step by Step Instructions:

Dektak Step by Step Instructions: Dektak Step by Step Instructions: Before Using the Equipment SIGN IN THE LOG BOOK Part 1: Setup 1. Turn on the switch at the back of the dektak machine. Then start up the computer. 2. Place the sample

More information

Veeco Dektak 6M Profilometer

Veeco Dektak 6M Profilometer Veeco Dektak 6M Profilometer System Ranges/Resolutions Range (Å) Resolution (Å) 50 (5nm) to 65K 1 0.5K to 655K 10 2K to 2620K 40 8K to 10000K (1mm) 160 Maximum sample thickness: 31.75mm Scan range: 50

More information

DektakXT Profilometer. Standard Operating Procedure

DektakXT Profilometer. Standard Operating Procedure DektakXT Profilometer Standard Operating Procedure 1. System startup and sample loading: a. Ensure system is powered on by looking at the controller to the left of the computer.(it is an online software,

More information

Figure 1. MFP-3D software tray

Figure 1. MFP-3D software tray Asylum MFP-3D AFM SOP January 2017 Purpose of this Instrument: To obtain 3D surface topography at sub-nanometer scale resolution, measure contact and friction forces between surfaces in contact, measure

More information

SPM Training Manual Veeco Bioscope II NIFTI-NUANCE Center Northwestern University

SPM Training Manual Veeco Bioscope II NIFTI-NUANCE Center Northwestern University SPM Training Manual Veeco Bioscope II NIFTI-NUANCE Center Northwestern University Introduction: Scanning Probe Microscopy (SPM) is a general term referring to surface characterization techniques that utilize

More information

Quick Start ATOMIC FORCE MICROSCOPE West Campus Imaging Core

Quick Start ATOMIC FORCE MICROSCOPE West Campus Imaging Core Quick Start ATOMIC FORCE MICROSCOPE West Campus Imaging Core 1 Turn On the laser power 2 Open enclosure: - lift the door latch and open the enclosure door. 3 2 1 1.Unlock scanner: Lift the lever to the

More information

STANDARD OPERATING PROCEDURE: ASYLUM MFP-3D AFM

STANDARD OPERATING PROCEDURE: ASYLUM MFP-3D AFM STANDARD OPERATING PROCEDURE: ASYLUM MFP-3D AFM Purpose of this Instrument: To obtain 3D surface topography at sub-nanometer scale resolution, measure contact and friction forces between surfaces in contact,

More information

Quick Start Bruker Dimension Icon AFM

Quick Start Bruker Dimension Icon AFM Do not remove Quick Start Bruker Dimension Icon AFM March 3, 2015 GLA Contacts Harold Fu (hfu@caltech.edu) Weilai Yu (wyyu@caltech.edu) Bruker Tech Support (AFMSupport@bruker-nano.com 800-873-9750) Watch

More information

AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode)

AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode) AFM1 Imaging Operation Procedure (Tapping Mode or Contact Mode) 1. Log into the Log Usage system on the SMIF web site 2. Open Nanoscope 6.14r1 software by double clicking on the Nanoscope 6.14r1 desktop

More information

Scanning Electron Microscopy (FEI Versa 3D Dual Beam)

Scanning Electron Microscopy (FEI Versa 3D Dual Beam) Scanning Electron Microscopy (FEI Versa 3D Dual Beam) This operating procedure intends to provide guidance for basic measurements on a standard sample with FEI Versa 3D SEM. For more advanced techniques

More information

SNG-2150C User s Guide

SNG-2150C User s Guide SNG-2150C User s Guide Avcom of Virginia SNG-2150C User s Guide 7730 Whitepine Road Revision 001 Richmond, VA 23237 USA GENERAL SAFETY If one or more components of your earth station are connected to 120

More information

2.1. Log on to the TUMI system (you cannot proceed further until this is done).

2.1. Log on to the TUMI system (you cannot proceed further until this is done). FEI DB235 ex-situ lift out TEM sample preparation procedure Nicholas G Rudawski ngr@ufledu (805) 252-4916 Last updated: 06/19/15 DISCLAIMER: this procedure describes one specific method for preparing ex-situ

More information

Standard Operating Procedure of nanoir2-s

Standard Operating Procedure of nanoir2-s Standard Operating Procedure of nanoir2-s The Anasys nanoir2 system is the AFM-based nanoscale infrared (IR) spectrometer, which has a patented technique based on photothermal induced resonance (PTIR),

More information

AFM Standard Operating Procedure

AFM Standard Operating Procedure 2013 AFM Standard Operating Procedure Karen Gaskell, David Ramsdell Surface Analysis Centre Department of Chemistry and Biochemistry University of Maryland 1/1/2013 Content Page 1 Hardware 2 1.1 MultiMode

More information

Burlington County College INSTRUCTION GUIDE. for the. Hewlett Packard. FUNCTION GENERATOR Model #33120A. and. Tektronix

Burlington County College INSTRUCTION GUIDE. for the. Hewlett Packard. FUNCTION GENERATOR Model #33120A. and. Tektronix v1.2 Burlington County College INSTRUCTION GUIDE for the Hewlett Packard FUNCTION GENERATOR Model #33120A and Tektronix OSCILLOSCOPE Model #MSO2004B Summer 2014 Pg. 2 Scope-Gen Handout_pgs1-8_v1.2_SU14.doc

More information

Variwrap Controller Manual

Variwrap Controller Manual Variwrap Controller Manual Operation The controller has two operating modes Manual and Auto. The mode is changes by pressing the (F1) Auto/Manual button. The mode setting is displayed in the top right

More information

Cover Page for Lab Report Group Portion. Boundary Layer Measurements

Cover Page for Lab Report Group Portion. Boundary Layer Measurements Cover Page for Lab Report Group Portion Boundary Layer Measurements Prepared by Professor J. M. Cimbala, Penn State University Latest revision: 23 February 2017 Name 1: Name 2: Name 3: [Name 4: ] Date:

More information

Laser Beam Analyser Laser Diagnos c System. If you can measure it, you can control it!

Laser Beam Analyser Laser Diagnos c System. If you can measure it, you can control it! Laser Beam Analyser Laser Diagnos c System If you can measure it, you can control it! Introduc on to Laser Beam Analysis In industrial -, medical - and laboratory applications using CO 2 and YAG lasers,

More information

Computer Graphics Hardware

Computer Graphics Hardware Computer Graphics Hardware Kenneth H. Carpenter Department of Electrical and Computer Engineering Kansas State University January 26, 2001 - February 5, 2004 1 The CRT display The most commonly used type

More information

LeCroy Digital Oscilloscopes

LeCroy Digital Oscilloscopes LeCroy Digital Oscilloscopes Get the Complete Picture Quick Reference Guide QUICKSTART TO SIGNAL VIEWING Quickly display a signal View with Analog Persistence 1. Connect your signal. When you use a probe,

More information

Beginners How to Test DSO138mini

Beginners How to Test DSO138mini Beginners How to Test DSO138mini You have finished assembling your DSO138mini kit. You may be anxious to see it works. But you might not be familiar with oscilloscope and you could encounter unexpected

More information

KRAMER ELECTRONICS LTD. USER MANUAL

KRAMER ELECTRONICS LTD. USER MANUAL KRAMER ELECTRONICS LTD. USER MANUAL MODEL: Projection Curved Screen Blend Guide How to blend projection images on a curved screen using the Warp Generator version K-1.4 Introduction The guide describes

More information

Part 1: Introduction to computer graphics 1. Describe Each of the following: a. Computer Graphics. b. Computer Graphics API. c. CG s can be used in

Part 1: Introduction to computer graphics 1. Describe Each of the following: a. Computer Graphics. b. Computer Graphics API. c. CG s can be used in Part 1: Introduction to computer graphics 1. Describe Each of the following: a. Computer Graphics. b. Computer Graphics API. c. CG s can be used in solving Problems. d. Graphics Pipeline. e. Video Memory.

More information

Lecture 14: Computer Peripherals

Lecture 14: Computer Peripherals Lecture 14: Computer Peripherals The last homework and lab for the course will involve using programmable logic to make interesting things happen on a computer monitor should be even more fun than the

More information

K Service Source. Apple High-Res Monochrome Monitor

K Service Source. Apple High-Res Monochrome Monitor K Service Source Apple High-Res Monochrome Monitor K Service Source Specifications Apple High-Resolution Monochrome Monitor Specifications Characteristics - 1 Characteristics Picture Tube 12-in. diagonal

More information

Part 1: Introduction to Computer Graphics

Part 1: Introduction to Computer Graphics Part 1: Introduction to Computer Graphics 1. Define computer graphics? The branch of science and technology concerned with methods and techniques for converting data to or from visual presentation using

More information

Viewing Serial Data on the Keysight Oscilloscopes

Viewing Serial Data on the Keysight Oscilloscopes Ming Hsieh Department of Electrical Engineering EE 109L - Introduction to Embedded Systems Viewing Serial Data on the Keysight Oscilloscopes by Allan G. Weber 1 Introduction The four-channel Keysight (ex-agilent)

More information

K Service Source. Apple High-Res Monochrome Monitor

K Service Source. Apple High-Res Monochrome Monitor K Service Source Apple High-Res Monochrome Monitor K Service Source Specifications Apple High-Resolution Monochrome Monitor Specifications Characteristics - 1 Characteristics Picture Tube 12-in. diagonal

More information

Wafer defects can t hide from

Wafer defects can t hide from WAFER DEFECTS Article published in Issue 3 2016 Wafer defects can t hide from Park Systems Atomic Force Microscopy (AFM) leader Park Systems has simplified 300mm silicon wafer defect review by automating

More information

Lab Determining the Screen Resolution of a Computer

Lab Determining the Screen Resolution of a Computer Lab 1.3.3 Determining the Screen Resolution of a Computer Objectives Determine the current screen resolution of a PC monitor. Determine the maximum resolution for the highest color quality. Calculate the

More information

University of Minnesota Minnesota Nano Center Standard Operating Procedure

University of Minnesota Minnesota Nano Center Standard Operating Procedure Equipment Name: CHA Evaporator Coral Name: ebevap-cha Revision Number: 12 Model: SEC 600 Revisionist: L. von Dissen Location: PAN Bay 3 Date: 04/13/2018 1 Description The CHA Evaporator is a single source

More information

Viewing Serial Data on the Keysight Oscilloscopes

Viewing Serial Data on the Keysight Oscilloscopes Ming Hsieh Department of Electrical Engineering EE 109L - Introduction to Embedded Systems Viewing Serial Data on the Keysight Oscilloscopes by Allan G. Weber 1 Introduction The four-channel Keysight (ex-agilent)

More information

E X P E R I M E N T 1

E X P E R I M E N T 1 E X P E R I M E N T 1 Getting to Know Data Studio Produced by the Physics Staff at Collin College Copyright Collin College Physics Department. All Rights Reserved. University Physics, Exp 1: Getting to

More information

Activity P32: Variation of Light Intensity (Light Sensor)

Activity P32: Variation of Light Intensity (Light Sensor) Activity P32: Variation of Light Intensity (Light Sensor) Concept DataStudio ScienceWorkshop (Mac) ScienceWorkshop (Win) Illuminance P32 Vary Light.DS P54 Light Bulb Intensity P54_BULB.SWS Equipment Needed

More information

Experiment # 4 Counters and Logic Analyzer

Experiment # 4 Counters and Logic Analyzer EE20L - Introduction to Digital Circuits Experiment # 4. Synopsis: Experiment # 4 Counters and Logic Analyzer In this lab we will build an up-counter and a down-counter using 74LS76A - Flip Flops. The

More information

ME EN 363 ELEMENTARY INSTRUMENTATION Lab: Basic Lab Instruments and Data Acquisition

ME EN 363 ELEMENTARY INSTRUMENTATION Lab: Basic Lab Instruments and Data Acquisition ME EN 363 ELEMENTARY INSTRUMENTATION Lab: Basic Lab Instruments and Data Acquisition INTRODUCTION Many sensors produce continuous voltage signals. In this lab, you will learn about some common methods

More information

Kratos AXIS Ultra DLD X-ray Photoelectron Spectrometer Instructions

Kratos AXIS Ultra DLD X-ray Photoelectron Spectrometer Instructions Kratos AXIS Ultra DLD X-ray Photoelectron Spectrometer Instructions Note: Enter your complete name, CMRF user code, and date into the record book. You must have previously filled out a project information

More information

University of MN, Minnesota Nano Center Standard Operating Procedure

University of MN, Minnesota Nano Center Standard Operating Procedure Equipment Name: Atomic Force Microscope Badger name: afm Revisionist Paul Kimani Model: Dimension 3000 Date: October 1, 2013 Location: Bay 1 A. Description The Dimension 3000 consists of a rigid stage

More information

Spectrum Analyser Basics

Spectrum Analyser Basics Hands-On Learning Spectrum Analyser Basics Peter D. Hiscocks Syscomp Electronic Design Limited Email: phiscock@ee.ryerson.ca June 28, 2014 Introduction Figure 1: GUI Startup Screen In a previous exercise,

More information

Preparing for remote data collection at NE-CAT

Preparing for remote data collection at NE-CAT Preparing for remote data collection at NE-CAT Important Note: The beamtime and remote login privileges are intended just for you and your group. You are not allowed to share these with any other person

More information

Agilent DSO5014A Oscilloscope Tutorial

Agilent DSO5014A Oscilloscope Tutorial Contents UNIVERSITY OF CALIFORNIA AT BERKELEY College of Engineering Department of Electrical Engineering and Computer Sciences EE105 Lab Experiments Agilent DSO5014A Oscilloscope Tutorial 1 Introduction

More information

The BAT WAVE ANALYZER project

The BAT WAVE ANALYZER project The BAT WAVE ANALYZER project Conditions of Use The Bat Wave Analyzer program is free for personal use and can be redistributed provided it is not changed in any way, and no fee is requested. The Bat Wave

More information

PC/HDTV to PC/HDTV converter (CP-251F)

PC/HDTV to PC/HDTV converter (CP-251F) PC/HDTV to PC/HDTV converter (CP-251F) Operation Manual This Converter has been especially modified to also accept RGsB Sync on Green Operation Controls and Functions Front Panel 1. Reset/ and +- The and

More information

University of Utah Electrical & Computer Engineering Department ECE1050/1060 Oscilloscope

University of Utah Electrical & Computer Engineering Department ECE1050/1060 Oscilloscope University of Utah Electrical & Computer Engineering Department ECE1050/1060 Oscilloscope Name:, A. Stolp, 2/2/00 rev, 9/15/03 NOTE: This is a fill-in-the-blanks lab. No notebook is required. You are encouraged

More information

LED Setup Controller XB-R1 USER S MANUAL

LED Setup Controller XB-R1 USER S MANUAL LED Setup Controller XB-R1 USER S MANUAL Rev 1.2 No. 24-004-1656-00 1 Product Description Thanks for your purchase of this XB-R1 product. This is a professional LED Setup Controller with some useful functions.

More information

Quick Start. RSHS1000 Series Handheld Digital Oscilloscope

Quick Start. RSHS1000 Series Handheld Digital Oscilloscope Quick Start RSHS1000 Series Handheld Digital Oscilloscope General Safety Summary Carefully read the following safety precautions to avoid personal injury and prevent damage to the instrument or any products

More information

Connection for filtered air

Connection for filtered air BeamWatch Non-contact, Focus Spot Size and Position monitor for high power YAG, Diode and Fiber lasers Instantly measure focus spot size Dynamically measure focal plane location during start-up From 1kW

More information

INSTRUMENT CATHODE-RAY TUBE

INSTRUMENT CATHODE-RAY TUBE Instrument cathode-ray tube D14-363GY/123 INSTRUMENT CATHODE-RAY TUBE mono accelerator 14 cm diagonal rectangular flat face internal graticule low power quick heating cathode high brightness, long-life

More information

Standard AFM Modes User s Manual

Standard AFM Modes User s Manual Standard AFM Modes User s Manual Part #00-0018-01 Issued March 2014 2014 by Anasys Instruments Inc, 325 Chapala St, Santa Barbara, CA 93101 Page 1 of 29 Table of contents Chapter 1. AFM Theory 3 1.1 Detection

More information

LUMIGEN INSTRUMENT CENTER X-RAY CRYSTALLOGRAPHIC LABORATORY: WAYNE STATE UNIVERSITY

LUMIGEN INSTRUMENT CENTER X-RAY CRYSTALLOGRAPHIC LABORATORY: WAYNE STATE UNIVERSITY Standard Operating Procedure for the Bruker X8 APEX II Single-Crystal X- Ray Diffractometer Contact Manager: Dr. Cassie Ward ward@wayne.edu Office room 061 Chemistry (313) 577-2587 LIC Lab: (313) 577-0518

More information

Experiment 9A: Magnetism/The Oscilloscope

Experiment 9A: Magnetism/The Oscilloscope Experiment 9A: Magnetism/The Oscilloscope (This lab s "write up" is integrated into the answer sheet. You don't need to attach a separate one.) Part I: Magnetism and Coils A. Obtain a neodymium magnet

More information

Standard Operating Procedure II: EDS (Bruker Flat-Quad)

Standard Operating Procedure II: EDS (Bruker Flat-Quad) Standard Operating Procedure II: EDS (Bruker Flat-Quad) ywcmatsci.yale.edu ESC II, Room A119F 810 West Campus Drive West Haven, CT 06516 Version 1.1, October 2018 1 > FOLLOW the SOP strictly to keep the

More information

Technical Reference. TDS 684A, TDS 744A, & TDS 784A Digitizing Oscilloscope Performance Verification and Specifications

Technical Reference. TDS 684A, TDS 744A, & TDS 784A Digitizing Oscilloscope Performance Verification and Specifications Technical Reference TDS 684A, TDS 744A, & TDS 784A Digitizing Oscilloscope Performance Verification and Specifications 070-8990-04 Please check for change information at the rear of this manual. Fifth

More information

S op o e p C on o t n rol o s L arni n n i g n g O bj b e j ctiv i e v s

S op o e p C on o t n rol o s L arni n n i g n g O bj b e j ctiv i e v s ET 150 Scope Controls Learning Objectives In this lesson you will: learn the location and function of oscilloscope controls. see block diagrams of analog and digital oscilloscopes. see how different input

More information

Linkage 3.6. User s Guide

Linkage 3.6. User s Guide Linkage 3.6 User s Guide David Rector Friday, December 01, 2017 Table of Contents Table of Contents... 2 Release Notes (Recently New and Changed Stuff)... 3 Installation... 3 Running the Linkage Program...

More information

USER MANUAL. 28" 4K Ultra HD Monitor L28TN4K

USER MANUAL. 28 4K Ultra HD Monitor L28TN4K USER MANUAL 28" 4K Ultra HD Monitor L28TN4K TABLE OF CONTENTS 1 Getting Started 2 Control Panel/ Back Panel 3 On Screen Display 4 Technical Specs 5 Care & Maintenance 6 Troubleshooting 7 Safety Info &

More information

4. Alignment and Adjustments

4. Alignment and Adjustments 4. Alignment and Adjustments 4-1 Preadjustment 4-1-1 Factory Mode 1. Do not attempt these adjustments in the Video Mode. 2. The Factory Mode adjustments are necessary when either the EEPROM (IC02) or the

More information

University of Minnesota Nano Fabrication Center Standard Operating Procedure

University of Minnesota Nano Fabrication Center Standard Operating Procedure Equipment Name: Focused Ion Beam (FIB) Coral Name: fib Revision Number: 2 Model: FEI Quanta 200 3D Revisionist: Kevin Roberts Location: Area 3 Date: 9/17/2013 1 Description The Quanta 200 3D is a DualBeam

More information

Application Note AN-708 Vibration Measurements with the Vibration Synchronization Module

Application Note AN-708 Vibration Measurements with the Vibration Synchronization Module Application Note AN-708 Vibration Measurements with the Vibration Synchronization Module Introduction The vibration module allows complete analysis of cyclical events using low-speed cameras. This is accomplished

More information

Mechanical aspects, FEA validation and geometry optimization

Mechanical aspects, FEA validation and geometry optimization RF Fingers for the new ESRF-EBS EBS storage ring The ESRF-EBS storage ring features new vacuum chamber profiles with reduced aperture. RF fingers are a key component to ensure good vacuum conditions and

More information

Operating Instructions

Operating Instructions Broadcast A/V Division Model No. M-LYNX-702W Dual 7 High Resolution Rack Mount Display with Waveform Operating Instructions V.1.0 Table of Contents 1. PRODUCT DESCRIPTION... 3 2. MENU SETTING... 6 3. SPECIFICATIONS...

More information

Computer Graphics: Overview of Graphics Systems

Computer Graphics: Overview of Graphics Systems Computer Graphics: Overview of Graphics Systems By: A. H. Abdul Hafez Abdul.hafez@hku.edu.tr, 1 Outlines 1. Video Display Devices 2. Flat-panel displays 3. Video controller and Raster-Scan System 4. Coordinate

More information

University of Utah Electrical Engineering Department EE1050/1060 Oscilloscope. Name:, Lab TA:

University of Utah Electrical Engineering Department EE1050/1060 Oscilloscope. Name:, Lab TA: University of Utah Electrical Engineering Department EE1050/1060 Oscilloscope Name:, Lab TA: A. Stolp, 2/2/00 rev, 9/14/00 NOTE: This is a fill-in-the-blanks lab. No notebook is required. You are encouraged

More information

OPERATING INSTRUCTIONS FOR SYLVANIA. Type I08 Cathode-Ray Oscilloscope. Sylvania Electric Products Inc. Industrial Apparatus. Emporium, Pennsylvania

OPERATING INSTRUCTIONS FOR SYLVANIA. Type I08 Cathode-Ray Oscilloscope. Sylvania Electric Products Inc. Industrial Apparatus. Emporium, Pennsylvania OPERATING INSTRUCTIONS FOR SYLVANIA Type I08 Cathode-Ray Oscilloscope Sylvania Electric Products Inc. Industrial Apparatus Plant Emporium, Pennsylvania OPERATING INSTRUCTIONS FOR Sylvania Type 08 Cathode-Ray

More information

CHA EVAPORATOR. User guidelines. p.1. by Carlos Manzanedo. Last revised: 10/05/2000

CHA EVAPORATOR. User guidelines. p.1. by Carlos Manzanedo. Last revised: 10/05/2000 CHA EVAPORATOR User guidelines by Carlos Manzanedo. Last revised: 10/05/2000 p.1 Introduction. The CHA evaporator has the following characteristics: 1. 2 independent resistive power supplies. 2. 1 Electron

More information

FEI FIB Focused Ion Beam

FEI FIB Focused Ion Beam Operating Manual Part 1 FEI FIB Focused Ion Beam IF IN DOUBT, ASK 1.00 IF ANYTHING UNUSUAL HAPPENS, OR IF THERE IS ANYTHING YOU ARE UNSURE ABOUT, STOP AND CONTACT ME! DO NOT PROCEED OR ATTEMPT TO FIX THE

More information

F7000NV ROBOT VISION OPERATING MANUAL

F7000NV ROBOT VISION OPERATING MANUAL Rev. C Feb 2012 F7000NV ROBOT VISION OPERATING MANUAL Rev. C Feb 2012 This page has intentionally been left blank. Contents Contents Chapter 1. Getting Started... 5 1. Preface... 5 2. Manuals... 5 3. Setting

More information

MestReNova Manual for Chem 201/202. October, 2015.

MestReNova Manual for Chem 201/202. October, 2015. 1. Introduction to 1-D NMR Data Processing with MestReNova The MestReNova program can do all of the routine NMR data processing needed for Chem 201 and 202 and will be available through the Reed downloads

More information

Magnetic Force Microscope (MFM)

Magnetic Force Microscope (MFM) Magnetic Force Microscope (MFM) [Digital Instruments NanoScopeIII] magnetic and topographic resolution ~20 nm [at least several hours/sample (1 cm 2 polished single grains, thin sections, or thin films)]

More information

CARESTREAM VITA/VITA LE/VITA SE CR System Long Length Imaging User Guide

CARESTREAM VITA/VITA LE/VITA SE CR System Long Length Imaging User Guide CARESTREAM VITA/VITA LE/VITA SE CR System Long Length Imaging User Guide Use of the Guide Carestream CR Systems are designed to meet international safety and performance standards. Personnel operating

More information

WAVEJET 300 SERIES OSCILLOSCOPES. Unmatched Performance, Portability, and Value

WAVEJET 300 SERIES OSCILLOSCOPES. Unmatched Performance, Portability, and Value WAVEJET 300 SERIES OSCILLOSCOPES Unmatched Performance, Portability, and Value 1 WAVEJET 300 SERIES Unique Capabilities in a Low Bandwidth Oscilloscope The WaveJet 300 Series features unmatched performance

More information

INSTRUCTION GUIDE. for the. Keysight (formally Agilent; Hewlett-Packard) Model #33509B. and. Tektronix. Fall 2017

INSTRUCTION GUIDE. for the. Keysight (formally Agilent; Hewlett-Packard) Model #33509B. and. Tektronix. Fall 2017 v3 INSTRUCTION GUIDE for the Keysight (formally Agilent; Hewlett-Packard) WAVEFORM GENERATOR Model #33509B 33500 Series and Tektronix OSCILLOSCOPE Model #MSO2004B Fall 2017 Pg. 2 Scope-Gen Handout_part

More information

USER MANUAL. 22" Class Slim HD Widescreen Monitor L215DS

USER MANUAL. 22 Class Slim HD Widescreen Monitor L215DS USER MANUAL 22" Class Slim HD Widescreen Monitor L215DS TABLE OF CONTENTS 1 Getting Started Package Includes Installation 2 Control Panel / Back Panel Control Panel Back Panel 3 On Screen Display 4 Technical

More information

2013, 2014 Hewlett-Packard Development Company, L.P.

2013, 2014 Hewlett-Packard Development Company, L.P. User Guide 2013, 2014 Hewlett-Packard Development Company, L.P. The only warranties for HP products and services are set forth in the express warranty statements accompanying such products and services.

More information

Asylum MFP-3D Standard Operating Procedures

Asylum MFP-3D Standard Operating Procedures Asylum MFP-3D Standard Operating Procedures Document Version 1.5, November 2013 The Asylum MFP-3D is an extremely versatile, research-oriented scanning probe microscope on an inverted optical microscope

More information

Operating Instructions

Operating Instructions Operating Instructions VIA-170 R Video Image Marker- Measurement System Accuracy by Design P/N 700051 ii VIA-170 Video Image Marker-Measurement System User's Manual R iii Copyright 1994-1998 by Boeckeler

More information

HIGH RESOLUTION MONITOR

HIGH RESOLUTION MONITOR HIGH RESOLUTION MONITOR 29" Pure Flat 15K-40Khz Installation Instruction WARNING Primary side and Secondary side The monitor's circuit which is divided into the primary side and secondary side, is insulated.

More information

EASY-MCS. Multichannel Scaler. Profiling Counting Rates up to 150 MHz with 15 ppm Time Resolution.

EASY-MCS. Multichannel Scaler. Profiling Counting Rates up to 150 MHz with 15 ppm Time Resolution. Multichannel Scaler Profiling Counting Rates up to 150 MHz with 15 ppm Time Resolution. The ideal solution for: Time-resolved single-photon counting Phosphorescence lifetime spectrometry Atmospheric and

More information

QUAD Camera Inputs with 8 display options (see pg 14-17) Handbook & Instructions. ...and more!

QUAD Camera Inputs with 8 display options (see pg 14-17) Handbook & Instructions. ...and more! QUAD040 Handbook & Instructions Useful compact quad unit that can be wall mounted or desk mounted. Multiple display options make it indispensable to the professional CCTV engineer. 4 Camera Inputs with

More information

TT AFM LongBeach Procedures and Protocols V2.1

TT AFM LongBeach Procedures and Protocols V2.1 TT AFM LongBeach Procedures and Protocols V2.1 1. Startup Procedure 1. Turn on PC: Allow to boot to Windows. Turn on monitor. Password is afm 2. Turn on second PC that controls the video camera. 3. Turn

More information

Asylum MFP-3D Standard Operating Procedures

Asylum MFP-3D Standard Operating Procedures Asylum MFP-3D Standard Operating Procedures Document Version 2, September 2017 The Asylum MFP-3D is an extremely versatile, research-oriented scanning probe microscope on an inverted optical microscope

More information

FIB Operating Procedure. Effective Date: 08/14/2012 Author(s): Jiong Hua Phone:

FIB Operating Procedure. Effective Date: 08/14/2012 Author(s): Jiong Hua Phone: FIB Operating Procedure Effective Date: 08/14/2012 Author(s): Jiong Hua Phone: 402-472-3773 Email: jhua2@unl.edu 1 1 Introduction 1.1 Key Words Focused Ion Beam (FIB), FEI Strata 201, Ion milling 1.2 Purpose

More information

Analysis of AP/axon classes and PSP on the basis of AP amplitude

Analysis of AP/axon classes and PSP on the basis of AP amplitude Analysis of AP/axon classes and PSP on the basis of AP amplitude In this analysis manual, we aim to measure and analyze AP amplitudes recorded with a suction electrode and synaptic potentials recorded

More information

Uni700 LCD Controller

Uni700 LCD Controller Landmark Technology Inc. Uni700 LCD Controller For TFT LCDs with Resolution up to 1,920 x 1,200 (Version A) January 27, 2009 1 1. Introduction The Uni700 controller board is designed for LCD panels of

More information

LedSet User s Manual V Official website: 1 /

LedSet User s Manual V Official website:   1 / LedSet User s Manual V2.6.1 1 / 42 20171123 Contents 1. Interface... 3 1.1. Option Menu... 4 1.1.1. Screen Configuration... 4 1.1.1.1. Instruction to Sender/ Receiver/ Display Connection... 4 1.1.1.2.

More information

Service manual Cantano W/T

Service manual Cantano W/T Service manual Cantano W/T Here you will see everything that should be included in your Cantano package 2 Prerequisite: Placement and leveling of the drive 5 Setting up the motor and connecting it to the

More information

Expect to Make Waves.

Expect to Make Waves. Expect to Make Waves. The New Oscilloscope Large 10.4" LCD touch screen Long capture time Extensive communication capabilities www.lecroy.com The New Oscillos From its large 10.4" LCD touch screen to its

More information

3D Printing User s Guide

3D Printing User s Guide 3D Printing User s Guide MCLD 257 May 2018 (Revision 2) 1 Table of Contents Content Page Section 1 Summaries Training & Access 3 Quick Guide 4 Section 2 Detailed Guides Training & Access Notes 5 Journal

More information

Manual placement system MPL3100. for BGA, CSP and Fine-Pitch components

Manual placement system MPL3100. for BGA, CSP and Fine-Pitch components Manual placement system MPL3100 for BGA, CSP and Fine-Pitch components Part No: MPL3100BA1.0e Issue Date: 02/2001 You have opted for an ESSEMTEC MPL3100 pick and place system. We thank you for this decision

More information

Screen investigations for low energetic electron beams at PITZ

Screen investigations for low energetic electron beams at PITZ 1 Screen investigations for low energetic electron beams at PITZ S. Rimjaem, J. Bähr, H.J. Grabosch, M. Groß Contents Review of PITZ setup Screens and beam profile monitors at PITZ Test results Summary

More information

Electrical and Electronic Laboratory Faculty of Engineering Chulalongkorn University. Cathode-Ray Oscilloscope (CRO)

Electrical and Electronic Laboratory Faculty of Engineering Chulalongkorn University. Cathode-Ray Oscilloscope (CRO) 2141274 Electrical and Electronic Laboratory Faculty of Engineering Chulalongkorn University Cathode-Ray Oscilloscope (CRO) Objectives You will be able to use an oscilloscope to measure voltage, frequency

More information

User Manual. Applications. See for more detailed information. Manual Version: Software Version: 2.03.

User Manual. Applications. See   for more detailed information. Manual Version: Software Version: 2.03. User Manual Applications Manual Version: 2.00 Software Version: 2.03 Language: English See www.idig-system.com for more detailed information. Safety Information: Magnets Use caution with the powerful magnets

More information

Portable Performance for Debug and Validation

Portable Performance for Debug and Validation WaveJet 300A Oscilloscopes 100 MHz 500 MHz Portable Performance for Debug and Validation A UNIQUE TOOLSET FOR PORTABLE OSCILLOSCOPES Key Features 100 MHz, 200 MHz, 350 MHz and 500 MHz bandwidths Sample

More information

TDS 520B, TDS 540B, TDS 620B, TDS 644B, TDS 680B, TDS 684B, TDS 724A, TDS 744A, & TDS 784A

TDS 520B, TDS 540B, TDS 620B, TDS 644B, TDS 680B, TDS 684B, TDS 724A, TDS 744A, & TDS 784A Technical Reference TDS 520B, TDS 540B, TDS 620B, TDS 644B, TDS 680B, TDS 684B, TDS 724A, TDS 744A, & TDS 784A Digitizing Oscilloscopes Performance Verification and Specifications 070-9384-01 Copyright

More information

Operation Procedure for Phillips XL30 ESEM

Operation Procedure for Phillips XL30 ESEM Operation Procedure for Phillips XL30 ESEM The ESEM will be left in the ON state when not in use. The chamber will be at high vacuum, filament on, stage at home position, VAC and HT buttons lit, and monitor

More information

SC24 Magnetic Field Cancelling System

SC24 Magnetic Field Cancelling System SPICER CONSULTING SYSTEM SC24 SC24 Magnetic Field Cancelling System Makes the ambient magnetic field OK for the electron microscope Adapts to field changes within 100 µs Touch screen intelligent user interface

More information

COMPOSITE VIDEO LUMINANCE METER MODEL VLM-40 LUMINANCE MODEL VLM-40 NTSC TECHNICAL INSTRUCTION MANUAL

COMPOSITE VIDEO LUMINANCE METER MODEL VLM-40 LUMINANCE MODEL VLM-40 NTSC TECHNICAL INSTRUCTION MANUAL COMPOSITE VIDEO METER MODEL VLM- COMPOSITE VIDEO METER MODEL VLM- NTSC TECHNICAL INSTRUCTION MANUAL VLM- NTSC TECHNICAL INSTRUCTION MANUAL INTRODUCTION EASY-TO-USE VIDEO LEVEL METER... SIMULTANEOUS DISPLAY...

More information

16-CH Color Full Duplex Multiplexer Instruction Manual

16-CH Color Full Duplex Multiplexer Instruction Manual 16-CH Color Full Duplex Multiplexer Instruction Manual 707-V1.5(S) Index: 1. Safety Warning 3 2. Introduction 3 3. Features 4 4. Specification 5 5. Front Panel Keypad 6 6. Back Panel Connection 10 7. Menu

More information

CHA EVAPORATOR Short Reference User guidelines

CHA EVAPORATOR Short Reference User guidelines CHA EVAPORATOR Short Reference User guidelines by Carlos Manzanedo. Last revised: 10/05/2000 p.1 Procedure: The CHA evaporator is divided into 16 different panels. These panels are labeled on the machine

More information